International Journal of Scientific & Engineering Research, Volume 4, Issue 4, April-2013 1288 ISSN 2229-5518
Development and Validation of Stability Indicating HPLC Assay Method for
Determination of Tapentadol in Tablet Formulation
Gaurang P. Pandya*, Hitendra S. Joshi† Abstract - The purpose of the research described herein was to develop simple, precise and accurate isocratic stability indicating reversed phase HPLC assay method for determination of Tapentadol solid dosage forms. Isocratic RP-HPLC method was developed on Phenomenex Luna C8 150 4.6mm, 5μm column using mobile phase as methanol – 0.002 M potassium dihydrogen phosphate buffer pH 3.0 (60: 40 v/v) at a flow rate of 1.0 ml/min and the detection was carried out at 272 nm using photo-diode array detector. The drug was subjected to oxidation, hydrolysis, photolysis and heat to apply stress condition. The validation element investigated showed that the method has acceptable specificity, accuracy, linearity, solution stability, precision and robustness.
Index Terms - Tapentadol, Stability indicating assay, Method development, Method validation
1. Introduction
Tapentadol is chemically 3-((2R,3R)-1-(dimethylamino)-2-methylpentan-3-yl)phenol (Fig.1). Its molecular formula is C14H23NO having molecular weight 221.34 gm/mol. Tapentadol is centrally acting analgesic with broad analgesic efficacy that was approved by Food and Drug Administration in 2008. The drug has dual mode of action in single molecule as an agonist at µ-opioid receptor and as a norepinephrine reuptake inhibitor [1], [2]. With this mode of action Tapentadol provides analgesia at comparable levels of more potent narcotic analgesics such as morphine,oxycodone and hydrocodone [3],[4] but its µ-sparing effect reduces the frequency and/or severity of side effects. Its other action on noradrenaline re-uptake gives a high analgesic potency [5]. Major sites of action of this drug are the spinal µ-opioid receptors and that the block of the noradrenaline lead to potentially synergistic activation of the spinal ά2 adrenoceptors [6].
According to ICH guidelines stress testing is an integral part of developmental strategy and is carried out under more severe condition than that of accelerated conditions. These studies provide information of drug’s intrinsic stability. Stress testing is useful in developing and validating suitable analytical methods [7],[8],[9]. It is suggested in the ICH guidelines that stress testing should be done including the effect of temperature, light, oxidizing agent and susceptibility to hydrolysis across a wide range of pH values. It is also needed that analysis of stability sample should be carried out with the validated stability testing methods.
Few bioanalytical methods have been reported for the determination of Tapentadol in blood plasma by using LC-MS-MS [10],[11],[12]. Besides, some methods have been
reported for determination of Tapentadol in urine including LC-MS-MS and UPLC-MS.
To the best of our literature survey, so far there is no published report describing validated stability indicating HPLC method for determination of Tapentadol available in literature. This paper deals with forced degradation of Tapentadol under acidic hydrolysis, alkali hydrolysis and oxidation, thermal and photolytic stress condition and the validation of developed method for assay of Tapentadol from its dosage form (tablets).
2. Experimental
2.1 Instrumentation
The chromatographic system used to perform development and validation of this assay method was comprised of a LC-10ATvp binary pump, a SPD-M10Avp photo-diode array detector and a rheodyne manual injector model 7725i with 20μl loop (Shimadzu, Kyoto, Japan) connected to a multi-instrument data acquisition and data processing system (Class-VP 6.13 SP2, Shimadzu).
2.2 Reagents and Reference substance
Tapentadol standard was provided by Ami Life sciences Laboratories Ltd., Baroda (India). Tapentadol tablets containing 50 mg Tapentadol and the inactive ingredient used in drug matrix were obtained from market. HPLC grade methanol was purchased from Spectrochem Pvt. Ltd., Mumbai (India). HPLC grade water was produced in-house by Milli Q (Millipore, Millford, USA) system. Membrane filters of 0.45um (Millipore) were used. Analytical grade ortho-phosphoric acid, hydrochloric acid, sodium hydroxide pellets and 30% v/v hydrogen peroxide
International Journal of Scientific & Engineering Research, Volume 4, Issue 4, April-2013 1289 ISSN 2229-5518 solution were obtained from Ranbaxy Fine Chemicals, New Delhi (India).
2.3 Chromatographic conditions
Chromatographic analysis was performed on Phenomenex Luna C8 (150mm 4.6mm i.d., 5μm particle size) column applying an isocratic elution using methanol – 0.002 M potassium dihydrogen phosphate buffer pH 3.0 (60: 40 v/v) as a mobile phase. The mobile phase was filtered through 0.45μm membrane filter and degassed for 30 minute in an ultrasonic bath prior to its use. Flow rate of mobile phase was adjusted to 1.00 ml/min and injection volume was 20 μL. Detection was performed at 272 nm.
2.4. Standard preparation
Tapentadol standard stock solution containing 500µg/ml was prepared in a 100 ml volumetric flask by dissolving 50.00mg of Tapentadol and then diluted to volume with water as a diluent. Further take 10 ml of this stock solution in 50 ml volumetric flask and make up to mark with diluent (this standard solution of 100µg/ml).
2.5. Test preparation
Twenty tablets were weighed and the average weight of tablet was determined. From these, five tablets were weighed and transfer into a 500 ml volumetric flask. About 50 ml of diluent was added and sonicated for a minimum 30 min. with intermittent shaking. Then content was brought back to room temperature and diluted to volume with diluent. The sample was filtered through 0.45µm nylon syringe filter. Further take 10 ml of this stock solution in 50 ml of volumetric flask and make up to mark with diluent. The concentration obtained was 100 µg/ml of Tapentadol.
Fig. 2 Chromatogram of standard preparation
The degradation samples were prepared by transferring powdered tablets, equivalent to 50 mg of Tapentadol into a 250 ml round bottom flask. Then prepared samples were employed for acidic, alkaline and oxidant media and also for thermal and photolytic conditions. After completing the
degradation treatments, the stress content solutions were allowed to equilibrate to room temperature and diluted with mobile phase to attain 100 µg/ml concentrations of Tapentadol. Specific conditions were described as follows.
2.6.1. Acidic degradation condition
Acidic degradation study was performed by taking the drug content in 0.1 N HCl at room temperature for 2.0 hours and mixture was neutralized.
Fig. 3 Chromatogram of Acid degradation
2.6.2. Alkali degradation condition
Alkaline degradation study was performed by taking the drug content in 0.05 N NaOH at room temperature for 2.0 hours and mixture was neutralized.
Fig. 4 Chromatogram of Alkali degradation
2.6.3. Oxidative degradation condition
Oxidative degradation study was performed by taking the drug content in 30% v/v H2O2 at room temperature for 2 hours.
Fig. 5 Chromatogram of Oxidative degradation
2.6.4. Thermal degradation condition
Thermal degradation was performed by exposing solid drug at 80˚C for 72 hours.
International Journal of Scientific & Engineering Research, Volume 4, Issue 4, April-2013 1290 ISSN 2229-5518
2.6.5. Photolytic degradation condition
Photolytic degradation study was performed by exposing the drug content in UV-light for 72 hours.
2.7. Method validation
2.7.1 Specificity study
The evaluation of the specificity of the method was determined against placebo. The interference of the excipients of the claimed placebo present in pharmaceutical dosage form was derived from placebo solution. Further the specificity of the method toward the drug was established by means of checking the interference of the degradation products in the drug quantification for assay during the forced degradation study.
2.7.2 Linearity
Linearity test solutions for the assay method were prepared at seven concentration levels from 40 to 160 % of assay analyte concentration (40, 60, 80, 100, 120, 140 and 160µg/ml). The peak areas versus concentration data were evaluated by linear regression analysis.
2.7.3 Precision
The precision of the assay method was evaluated in terms of repeatability by carrying out six independent assays of Tapentadol test sample preparation and calculated the % RSD of assay (intraday). Intermediate precision of the method was checked by performing same procedure on the different day (interday) by another person under the same experimental condition.
Tapentadol % assay
Interday Intraday (n=6) (n=6)
1 100.25 100.22 2 100.12 99.95 3 101.28 100.10 4 99.58 100.85 5 98.44 101.17 6 100.17 100.87
Mean 99.97 100.50 Standard deviation 0.93 0.50
%RSD 0.93 0.50
2.7.4 Accuracy
An accuracy study was performed by adding known amounts of Tapentadol to the placebo preparation. The actual and measured concentrations were compared. Recovery of the method was evaluated at three different concentration levels (corresponding to 50, 100 and 150 % of test preparation concentration). For each concentration level, three sets were prepared and injected in triplicate.
Level (%)
Theoretical Concentration
(mg/ml)
Observed Concentration
(mg/ml)
% Recovery % RSD
50 0.04993 0.05001 100.15 0.429 100 0.10053 0.10059 100.06 0.614 150 0.15053 0.15004 99.67 0.662
2.7.5. Robustness The robustness of study was carried out to evaluate the influence of small but deliberate variations in the chromatographic conditions. The factors chosen for this study were the flow rate (±0.1 ml/min), mobile phase composition methanol – 0.002 M potassium dihydrogen phosphate buffer pH 3.0 (62: 38 and 58: 42 v/v) and using different lot of LC column.
Table I: Evaluation data of precision study
Table II: Evaluation data of Accuracy Study
International Journal of Scientific & Engineering Research, Volume 4, Issue 4, April-2013 1291 ISSN 2229-5518
Robust conditions % Assay
System suitability parameters
Theoretical plates Asymmetry
Flow 0.9 ml/min 100.56 4235 1.66 Flow 1.1 ml/min 100.84 4430 1.41
Acetonitrile:water:0.1%acetic acid(62:38:0.1) 100.22 4316 1.28
Acetonitrile:water:0.1%acetic acid(58:42:0.1) 100.20 4615 1.36
Column change 99.82 4416 1.12
2.7.6. Solution stability
The stability of solution for test preparation was evaluated. The solution was stored at ambient temperature and 2-5˚C and tested at interval of 12, 24, 36 and 48 hours. The responses for the aged solution were evaluated against a freshly prepared standard solution.
Intervals % Assay for test solution stored at 2 -8˚ C
% Assay for test solution stored at ambient
temperature Initial 100.28 100.03 12 h 100.23 99.74 24 h 100.04 99.45 36 h 99.80 99.35 48 h 99.62 99.22
3. Result and discussion
To develop a rugged and suitable HPLC method for the quantitative determination of Tapentadol, the analytical condition were selected after testing the different parameters such as diluents, buffer, buffer concentration, organic solvent for mobile phase and mobile phase composition and other chromatographic conditions. Our preliminary trials using different composition of mobile phases consisting of water with methanol or acetonitrile, did not give good peak shape. By keeping mobile phase composition as methanol – 0.002 M potassium dihydrogen phosphate buffer pH 3.0 (60: 40 v/v), best peak shape was obtained. For the selection of organic constituent of mobile phase, methanol was chosen and to attain good peak shape. Chromatogram of standard preparation is represented in
(Fig. 2). A system suitability test of the chromatographic system was performed before each validation run. Five replicate injections of standard preparation were injected and asymmetry, theoretical plate and % RSD of peak area were determined for same. For all system suitability injections, asymmetry was less than 2.0, theoretical plate was greater than 4000 and % RSD of peak area was found less than 2.0. The specificity of the method was determined by checking the interference of placebo with analyte and the proposed method was eluted by checking the peak purity of Tapentadol during the force degradation study. There was no interference of any peak of degradation product with drug peak. Major degradation was found in alkaline condition in which product was degraded up to 22%. The major impurity peak was found at 4.263 min. (Fig. 4). In oxidative degradation, it was found that around 12 % of the drug degraded and impurity peak was found at 5.374 min. (fig. 5) and in acidic condition around 7 % of the drug degraded and impurity peak was found at 5.602 min. (fig. 3). Tapentadol was found to be slightly degraded in thermal degradation while it was stable under the photolytic condition. Seven points calibration curve were obtained in a concentration range from 40-160 µg/ml for Tapentadol. The response of the drug was found to be linear in the investigation concentration range and the linear with correlation coefficient 0.999. The result of repeatability and intermediate precision study is shown in Table 1. The developed method was found to be precise as the % RSD values for the repeatability and intermediate precision studies were <0.51 % and <0.94 %, respectively, which confirm that method was precise. The HPLC area responses for accuracy determination are depicted in Table 2. The results show that best recoveries (99.67-100.15%) of the spiked drug were obtained at each added concentration, indicating that the method was accurate. The result of robustness study of the developed assay method was established in Table 3. The result shown that during all variance conditions, assay value of the test preparation solution was not affected and it was in accordance with that of actual. Table 4 shows the results obtained in the solution stability study at different time intervals for test preparation. It was concluded that the test preparation solution was found stable up to 48 hours at 2 - 5˚ C and ambient temperature as during this time the result was not decreased below the minimum percentage. System suitability parameters were also found satisfactory; hence the analytical method would be concluded as robust.
4. Conclusion
A new analytical method has been developed to be routinely applied to determine Tapentadol in pharmaceutical dosage form. In this study, stability of Tapentadol in present dosage form was established
Table IV: Evaluation data of robustness study
Table I: Evaluation data of solution stability study
International Journal of Scientific & Engineering Research, Volume 4, Issue 4, April-2013 1292 ISSN 2229-5518 through employment of ICH recommended stress condition. The developed procedure has been evaluated over the specificity, linearity, accuracy, precision and robustness in order to ascertain the stability of the analytical method. It has been proved that it was specific, linear, precise, accurate and robust and stability indicating. Hence, the method is recommended for routine quality control analysis and also for stability sample analysis.
Acknowledgements
The authors are thankful for facilities & grants given under UGC- Special Assistance Programme Department Research Support (DRS) and Department of Science & Technology (DST) New Delhi, Fund For Improvement of Science & Technology (FIST), National Facility for Drug Discovery (NFDD) and Department of Chemistry, Saurashtra University, Rajkot – 360 005 (INDIA) for providing analytical facilities.
References
[1] C. Kneip, R. Terlinden, H. Beier and G. Chen; Investigations Into the Drug-Drug Interaction Potential of Tapentadol in Human Liver Microsomes and Fresh Human Hepatocytes; Drug Metabolism Letters, (2008), 2: 67-75.
[2] T. Tzschentke, T. Christoph, K. Babette, K. Schiene, H. Hennies, W. Englberger, H. Haurand; (–)-(1R,2R)-3-(3-Dimethylamino-1-ethyl-2-methyl-propyl)-phenol Hydrochloride (Tapentadol HCl): a Novel _-Opioid Receptor Agonist/Norepinephrine Reuptake Inhibitor with Broad-Spectrum Analgesic Properties; The Journal of Pharmacology and Experimental Therapeutics, (2007), 323: 265-276.
[3] R. Rob, F. Carol, H. Julie, W. Gill, M. Kate, S. Michael; Systematic review of tapentadol in chronic severe pain; Current medical research and opinion, (2011), 27(10), 1907-1930.
[4] K. Regina, L. Claudia, S. Achim, B. Peter, G. Jutta, and D. Paul,; Pain Medicine: Single Dose Analgesic Efficacy of Tapentadol in Postsurgical Dental Pain: The Results of a Randomized, Double-Blind, Placebo-Controlled Study; Anesthesia & Analgesia, (2008), 107(6): 2048-2055.
[5] M. Schäfer; Novel concepts for analgesia in severe pain – current strategies and future innovations; European Journal of Pain Supplements, (2009), 3: 6-10.
[6] L. Bee, K. Bannister, W. Rahman and A. Dickenson; Mu-opioid and noradrenergic a2-adrenoceptor contributions to the effects of tapentadol on spinal electrophysiological measures of nociception in nerve-injured rats; Pain, (2011),152: 131-139
[7] S. Singh and M. Bakshi,; Guidance on conduct of stress tests to determine inherent stability of drugs; Pharmaceutical Technology On-line (2000); 24: 1-14
[8] S. Klick, V. Muijselaar, J. Waterval, T. Eichinger, C. Korn, T. Gerding; Toward a Generic Approach for Stress Testing of Drug Substances and Drug Products; Pharmaceutical Technology, (2005);February:48-66
[9] S. Baerstichi, Pharmaceutical Stress Testing Predicting Drug Degradation, 1st edition, Chapter 6 Informa Healthcare,London,U.K.,(2005), pp. 459–469.
[10] Y. Jin, X. Jiang, N., Zhai, P. Xu, B. Yuan, and H. Xu,; LC-MS/MS method for the determination of tapentadol in rats; Shenyang Yaoke Daxue Xueba,; (2012), 29(2): 143-146.
[11] Y. Liu, C. Uboh, L. Soma, X. Li, F. Guan, and J. Chen,; Efficient Use of Retention Time for the Analysis of 302 Drugs in Equine Plasma by Liquid Chromatography-MS/MS with Scheduled Multiple Reaction Monitoring and Instant Library Searching for Doping Control; Analytical Chemistry, (2011), 83(17): 6834-6841.
[12] R. Terlinden, B. Koegel, W. Englberger and T. Tzschentke; In vitro and in vivo characterization of tapentadol metabolites; Methods and Findings in Experimental and Clinical Pharmacology, (2010), 32(1): 31-38.
[13] X. Jiang, Y. Jin, H. Xu, P. Xu, N. Zhai and B. Yuan,; Analysis of metabolites of tapentadol in rat urine by liquid chromatography tandem mass spectrometry; Shenyang Yaoke Daxue Xuebao, (2012), 29(1): 64-69.
[14] C. Coulter, M. Taruc, J. Tuyay and C. Moore; Determination of tapentadol and its metabolite N-desmethyltapentadol in urine and oral fluid using liquid chromatography with tandem mass spectral detection; Journal of Analytical Toxicology, (2010), 34(8); 458-463.
[15] R. Terlinden, J. Ossig, F. Fliegert, C. Lange, and K. Goehler; Absorption, metabolism, and excretion of 14C-labeled Tapentadol HCl in healthy male subjects; European Journal of Drug Metabolism and Pharmacokinetics; (2007), 32(3), 163-169.
[16] J. Bourland, A. Collins, S. Chester, S. Ramachandran and R. Backer; Determination of tapentadol (Nucynta®) and N-desmethyltapentadol in authentic urine specimens by ultraperformance liquid chromatography-tandem mass spectrometry; Journal of analytical toxicology (2010), 34(8): 450-457.
[17] J. Tuyay, C. Coulter, W. Rodrigues and C. Moore, Disposition of opioids in oral fluid: Importance of chromatography and mass spectral transitions in LC-MS/MS. Drug Test Analysis; Drug Testing and Analysis, doi: 10.1002/dta.1324
* Gaurang P. Pandya is currently persuing Ph.D. Degree program in Department of Chemistry, Saurashtra University, Rajkot-360 005, Gujarat, India. E-mail: [email protected]
† Hitendra S. Joshi is professor in Department of Chemistry, Saurashtra University, Rajkot-360 005, Gujarat, India. E-mail: [email protected]
Available online at www.pelagiaresearchlibrary.com
Pelagia Research Library
Der Pharmacia Sinica, 2013, 4(2):145-152
ISSN: 0976-8688 CODEN (USA): PSHIBD
145 Pelagia Research Library
Development and validation of stability indicating HPLC assay method for simultaneous determination of amlodipine besylate, olmesartan medoxomil
and hydrochlorothiazide in tablet formulation
Gaurang P. Pandya and Hitendra S. Joshi*
Department of Chemistry, Saurashtra University, Rajkot, Gujarat, India ____________________________________________________________________________________________
ABSTRACT The purpose of the research described herein was to develop simple, precise and accurate isocratic stability indicating reversed phase HPLC assay method for determination of simultaneous determination of Amlodipine besylate, Olmesartan medoxomil and Hydrochlorothiazide solid dosage forms. Isocratic RP-HPLC method was developed on Phenomenex Gemini C18 250× 4.6mm, 5µm column using mobile phase as 0.02M ammonium acetate buffer pH 4.5 - Acetonitrile (60:40, v/v) at a flow rate of 1.0 ml/min and the detection was carried out at 241 nm using photo-diode array detector. The drug was subjected to oxidation, hydrolysis, photolysis and heat to apply stress condition. The validation element investigated showed that the method has acceptable specificity, accuracy, linearity, solution stability, precision and robustness. Keywords: Amlodipine besylate, Olmesartan medoxomil, Hydrochlorothiazide Stability indicating assay, Method development, Method validation _____________________________________________________________________________________________
INTRODUCTION
Stress testing is a part of developmental strategy under International Conference on Harmonization (ICH) requirements and is carried out under more severe conditions than accelerated conditions. These studies serve to give information on a drug’s inherent stability and assist in the validation of analytical methods to be used in stability studies (1–3). It is suggested that stress testing should include the effects of temperature, light, oxidizing agents and susceptibility across a wide range of pH values. It is also recommended that analysis of stability samples should be accomplished through the use of a validated stability-testing method. Amlodipine besylate is chemically described as 3-Ethyl-5-methyl (±)-2-[(2-aminoethoxy)methyl]-4-(2-chlorophenyl)-1,4-dihydro-6-methyl-3,5-pyridinedicarboxylate, monobenzenesulphonate. Its empirical formula is C20H25CIN2O5•C6H6O3S, and its structural formula is shown in figure 1. Amlodipine belongs to the dihydropyridine (DHP) class of calcium channel blockers (CCBs), the most widely used class of CCBs. Similar to other DHP CCBs, amlodipine binds directly to inactive L-type calcium channels stabilizing their inactive conformation. Since arterial smooth muscle depolarizations are longer in duration than cardiac muscle depolarizations, inactive channels are more prevalent in smooth muscle cells. Amlodipine decreases arterial smooth muscle contractility and subsequent vasoconstriction by inhibiting the influx of calcium ions through L-type calcium channels. The vasodilatory effects of amlodipine result in an overall decrease in blood pressure. Amlodipine is a long-acting CCB that may be used to treat mild to moderate essential hypertension and exertion-related angina (chronic stable angina).
Hitendra S. Joshi et al Der Pharmacia Sinica, 2013, 4(2):145-152 _____________________________________________________________________________
146 Pelagia Research Library
Figure-1 Structure of Amlodipine Besylate
Olmesartan medoxomil is chemically (5-methyl-2-oxo-2H-1,3-dioxol-4-yl)methyl 4-(2-hydroxypropan-2-yl)-2-propyl-1-({4-[2-(2H-1,2,3,4-tetrazol-5-yl)phenyl]phenyl}methyl)-1H-imidazole-5-carboxylate [figure 2]. Its molecular formula is C29H30N6O6 having molecular mass 558.58 gm/mole. Olmesartan blocks the vasoconstrictor effects of angiotensin II by selectively blocking the binding of angiotensin II to the AT1 receptor in vascular smooth muscle. Its action is, therefore, independent of the pathways for angiotensin II synthesis.
N
N
CH3
CH3
HO
O
O
O
O
O
CH3
N
NH
N
N
Figure-2 Structure of Olmesartan medoxomil
Hydrochlorothiazide is chemically6-chloro-1,1-dioxo-3,4-dihydro-2H-1,2,4,benzothiadiazine-7-sulfonamide. Its molecular formula is C7H8N3O4S2Cl having molecular mass 297.74 gm/mole. Hydrochlorothiazide belongs to the thiazide class of diuretics. It reduces blood volume by acting on the kidneys to reduce sodium (Na) reabsorption in the distal convoluted tubule. The major site of action in the nephron appears on an electroneutral Na+-Cl- co-transporter by competing for the chloride site on the transporter. By impairing Na transport in the distal convoluted tubule, hydrochlorothiazide induces a natriuresis and concomitant water loss. Thiazides increase the reabsorption of calcium in this segment in a manner unrelated to sodium transport.
S
NH
HNCl
S
H2N
O OO O Figure-3 Structure of Hydrochlorothiazide
Many methods, either individual or in a combination of Amlodipine, Olmesartan and Hydrochlorothiazide have been reported, including simultaneous determination of Amlodipine with Olmesartan, Olmesartan with Hydrochlorothiazide and Amlodipine with Olmesartan. Few methods are also reported for simultaneous determination of Amlodipine with Olmesartan and Hydrochlorothiazide in combination formulations by ultraviolet (UV) absorption and UV derivative spectrophotometry, spectrophotometric determination with artificial neural network, high performance liquid chromatography (10–12), ultra high performance liquid chromatography and thin-
Hitendra S. Joshi et al Der Pharmacia Sinica, 2013, 4(2):145-152 _____________________________________________________________________________
147 Pelagia Research Library
layer chromatographic determination (13). Furthermore these methods are not impressionable to achieve the high throughput study which can be possible by optimizing the method in such a way which includes shortest runtime with maximum selectivity. Hence, it can be maximum utilize for the analysis of formulation development and stability testing as well as at quality control laboratory for routine use. This paper deals with forced degradation of Amlodipine, Olmesartan and Hydrochlorothiazide under acidic hydrolysis, alkali hydrolysis and oxidation, thermal and photolytic stress condition and the validation of developed method for assay of Amlodipine with Olmesartan and Hydrochlorothiazide from its dosage form (tablets).
MATERIALS AND METHODS
2.1 Instrumentation The chromatographic system used to perform development and validation of this assay method was comprised of a LC-10ATvp binary pump, a SPD-M10Avp photo-diode array detector and a rheodyne manual injector model 7725i with 20µl loop (Shimadzu, Kyoto, Japan) connected to a multi-instrument data acquisition and data processing system (Class-VP 6.13 SP2, Shimadzu). 2.2 Reagents and Reference substance Olmesartan medoxomil and Amlodipine besylate standards were provided by Cadila pharmaceuticals Ltd., Ahmadabad (India) and Hydrochlorothiazide standard was provided by Alembic pharmaceuticals Ltd., Baroda (India). Olmesartan medoxomil, amlodipine besylate and hydrochlorothiazide tablets containing 40mg Olmesartan medoxomil, 10mg amlodipine besylate and 25mg hydrochlorothiazide and the inactive ingredient used in drug matrix were obtained from market. HPLC grade acetonitrile was obtained from Spectrochem Pvt. Ltd., Mumbai (India). Analytical grade ammonium acetate, hydrochloric acid, glacial acetic acid, sodium hydroxide pellets and 30% v/v hydrogen peroxide solution were obtained from Ranbaxy Fine Chemicals, New Delhi (India). 2.3 Chromatographic conditions Chromatographic analysis was performed on Phenomenex Gemini C18 (250mm × 4.6mm i.d., 5µm particle size) column applying an isocratic elution using a mixture of 0.02M ammonium acetate buffer pH 4.5 - Acetonitrile (60:40, v/v) as a mobile phase. The mobile phase was filtered through 0.45µm membrane filter and degassed for 30 minute in an ultrasonic bath prior to its use. Flow rate of mobile phase was adjusted to 1.00 ml/min and injection volume was 20 µL. Detection was performed at 241 nm. Normal run time was chosen as 20 minutes but for degradation study chromatographic run was carried out up to 60 minutes to confirm that any degradation peak is eluted after 20 minutes or not. 2.4. Standard preparation Olmesartan medoxomil standard stock solution containing 400µg/ml was prepared in a 100 ml volumetric flask by dissolving 40.00 mg of Olmesartan medoxomil and then diluted to volume with diluent. Further take 25 ml of this stock solution in 50 ml volumetric flask and make up to mark with diluent (this standard solution of 200µg/ml). Amlodipine besylate standard stock solution containing 100µg/ml was prepared in a 100 ml volumetric flask by dissolving 10.00 mg of Amlodipine besylate and then diluted to volume with diluent. Further take 25 ml of this stock solution in 50 ml volumetric flask and make up to mark with diluent (this standard solution of 50µg/ml). Hydrochlorothiazide standard stock solution containing 250µg/ml was prepared in a 100 ml volumetric flask by dissolving 25.00 mg of Hydrochlorothiazide and then diluted to volume with diluent.Further take 25 ml of this stock solution in 50 ml volumetric flask and make up to mark with diluent (this standard solution of 125µg/ml). . 2.5. Test preparation Twenty tablets were weighed and the average weight of tablet was determined. From these, five tablets were weighed and transfer into a 500 ml volumetric flask. About 50 ml of diluent was added and sonicated for a minimum 30 minute with intermittent shaking. Then content was brought back to room temperature and diluted to volume with diluent. The sample was filtered through 0.45µm nylon syringe filter. .Further take 25 ml of this stock solution in 50 ml volumetric flask and make up to mark with diluent.The concentration obtained was 200 µg/ml of Olmesartanmedoxomil, 50µg/ml ofAmlodipine besylateand 125 µg/ml of Hydrochlorothiazide.
Hitendra S. Joshi et al Der Pharmacia Sinica, 2013, 4(2):145-152 _____________________________________________________________________________
148 Pelagia Research Library
Figure -3 Chromatogram of standard preparation
2.6 Degradation Study 2.6.1. Acidic degradation condition Acidic degradation study was performed by taking the drug content in 0.1 N HCl at room temperature for 2.0 hours and mixture was neutralized. 2.6.2. Alkali degradation condition Alkaline degradation study was performed by taking the drug content in 0.05 N NaOH at room temperature for 2.0 hours and mixture was neutralized. 2.6.3. Oxidative degradation condition Oxidative degradation study was performed by taking the drug content in 30% v/v H2O2 at room temperature for 2 hours. 2.6.4. Thermal degradation condition Thermal degradation was performed by exposing solid drug at 80˚C for 72 hours. 2.6.5. Photolytic degradation condition Photolytic degradation study was performed by exposing the drug content in UV-light for 72 hours. 2.7. Method validation 2.7.1 Specificity study The evaluation of the specificity of the method was determined against placebo. The interference of the excipients of the claimed placebo present in pharmaceutical dosage form was derived from placebo solution. Further the specificity of the method toward the drug was established by means of checking the interference of the degradation products in the drug quantification for assay during the forced degradation study. 2.7.2 Linearity Linearity test solutions for the assay method were prepared at seven concentration levels from 40 to 160 % of assay analyte concentration. The peak areas versus concentration data were evaluated by linear regression analysis. 2.7.3 Precision The precision of the assay method was evaluated in terms of repeatability by carrying out six independent assays of test sample preparation and calculated the % RSD of assay (intraday). Intermediate precision of the method was checked by performing same procedure on the different day (interday) by another person under the same experimental condition. 2.7.4 Accuracy An accuracy study was performed by adding known amounts of Amlodipine besylate, Olmesartan medoxomil and Hydrochlorothiazide to the placebo preparation. The actual and measured concentrations were compared. Recovery
Hitendra S. Joshi et al Der Pharmacia Sinica, 2013, 4(2):145-152 _____________________________________________________________________________
149 Pelagia Research Library
of the method was evaluated at three different concentration levels (corresponding to 50, 100 and 150 % of test preparation concentration). For each concentration level, three sets were prepared and injected in triplicate. 2.7.5. Robustness The robustness of study was carried out to evaluate the influence of small but deliberate variations in the chromatographic conditions. The factors chosen for this study were the flow rate (±0.1 ml/min), mobile phase composition 0.02M ammonium acetate buffer pH 4.5 - Acetonitrile (62: 38 and 58: 42 v/v) and using different lot of LC column. 2.7.6. Solution stability The stability of solution for test preparation was evaluated. The solution was stored at ambient temperature and 2-5˚C and tested at interval of 12, 24, 36 and 48 hours. The responses for the aged solution were evaluated against a freshly prepared standard solution.
Figure-5 Chromatogram of Acidic forced degradation study
Figure-6 Chromatogram of Alkali forced degradation study
Hitendra S. Joshi et al Der Pharmacia Sinica, 2013, 4(2):145-152 _____________________________________________________________________________
150 Pelagia Research Library
Figure-7 Chromatogram of Oxidative forced degradation study
Table-1 Evaluation Data of Precision study
Amlodipine besylate (% Assay) Olmesartanmedoxomil (% Assay) Hydrochlorothiazide (% Assay)
Intraday (n = 6) Interday (n = 6) Intraday (n = 6) Interday (n = 6) Intraday (n = 6) Interday (n = 6)
Set Intraday (n = 6) Interday (n = 6) Intraday (n = 6) Interday (n = 6) Intraday (n = 6) Interday (n = 6) 1 100.9 100.2 100.8 100.9 100.2 100.4 2 100 99.51 99.47 99.9 99.86 99.73 3 100 100.1 101.8 99.74 100.2 100 4 99.21 101.1 100.2 98.95 98.85 100.5 5 99.43 98.94 101.1 100.5 101 101.3 6 99.67 100.1 100 99.21 100.9 101.1
Mean 99.9 100 100.6 99.9 100.2 100.5 Standard deviation 0.596 0.734 0.832 0.736 0.773 0.625
% RSD 0.597 0.734 0.827 0.737 0.771 0.622
Table-2 Evaluation Data of Accuracy study
Drug Level (%)
Amount added concentration a (mg/ml)
Amount found concentration a (mg/ml)
% Recovery
% RSD
Amlodipine 50 0.0248 0.0248 99.72 0.619 100 0.0500 0.0498 99.67 0.414 150 0.0750 0.0740 98.70 0.686
Olmesartan 50 0.1002 0.0995 99.35 0.426 100 0.1997 0.1979 99.09 0.594 150 0.3002 0.2986 99.47 0.832
Hydrochlorothiazide 50 0.0625 0.0620 99.23 0.990 100 0.1255 0.1251 99.69 0.546 150 0.1877 0.1870 99.67 0.564
a Each value corresponds to the mean of three determinations
Table-3 Evaluation Data of Solution Stability study
Intervals %Assay for test solution stored at 2-8˚ C
% Assay for test solution stored at ambient temperature
Amlodipine Olmesartan HCTZ Amlodipine Olmesartan HCTZ
Initial 100.96 99.98 100.09 100.99 99.94 100.13 12 h 100.92 99.76 99.99 100.90 99.67 99.67 24 h 100.82 99.67 99.66 100.58 99.32 99.62 36 h 100.53 99.56 99.62 100.07 99.06 99.31 48 h 100.16 99.37 99.42 99.64 98.78 99.08
RESULTS AND DISCUSSION
To develop a rugged and suitable HPLC method for the quantitative determination of amlodipine, olmesartan and hydrochlorothiazide, the analytical condition were selected after testing the different parameters such as diluents, buffer, buffer concentration, organic solvent for mobile phase and mobile phase composition and other
Hitendra S. Joshi et al Der Pharmacia Sinica, 2013, 4(2):145-152 _____________________________________________________________________________
151 Pelagia Research Library
chromatographic conditions. Our preliminary trials using different composition of mobile phases consisting of water with methanol or acetonitrile, did not give good peak shape. By keeping mobile phase composition as 0.02M ammonium acetate buffer pH 4.5 - Acetonitrile (60:40, v/v), best peak shape was obtained. For the selection of organic constituent of mobile phase, Acetonitrile was chosen to reduce run time and to attain good peak shape. Chromatogram of standard preparation is represented in (Fig. 2). A system suitability test of the chromatographic system was performed before each validation run. Five replicate injections of standard preparation were injected and asymmetry, theoretical plate and % RSD of peak area were determined for same. For all system suitability injections, asymmetry was less than 2.0 and % RSD of peak area was found less than 2.0. The specificity of the method was determined by checking the interference of placebo with analyte and the proposed method was eluted by checking the peak purity of all drug peaks during the force degradation study. There was no interference of any peak of degradation product with drug peak. Major degradation was found in oxidative condition in which drug products were degraded up to 24%. (Fig. 3). In oxidative degradation, it was found that around 15 % of the drugs contents were degraded (fig. 5) and in alkaline condition around 8 % of the drugs contents were degraded and impurity peak was found at 5.602 min (Fig. 5). Drug contents were found to be degraded around 2% in thermal degradation while slightly degradation was observed under the photolytic condition. For linearity seven points calibration curve were obtained in a concentration range from 50.0-200.0 µg/ml for Amlodipine, from 80-320 µg/ml for Olmesartan and from 50-200 µg/ml for Hydrochlorothiazide. The response of the drug was found to be linear in the investigation concentration range and the linear regression equation was y = 2E+07x + 3802 with correlation coefficient 0.999 for Amlodipine, y = 7E+07x + 15600 with correlation coefficient 0.999 for Olmesartan and was y = 1E+08x + 53918 with correlation coefficient 0.999 for Hydrochlorothiazide. The result of repeatability and intermediate precision study is shown in Table 1. The % RSD values for intraday precision study and interday precision study was < 2.0 % for all drugs which confirm that the method is precise. The HPLC area responses for accuracy determination are depicted in Table 2. The results show that best recoveries (99.72-98.70% for Amlodipine, 99.09-99.47% for olmesartan and 99.23-99.67% for Hydrochlorothiazide) of the spiked drug were obtained at each added concentration, indicating that the method was accurate. Table 3 shows the results obtained in the solution stability study at different time intervals for test preparation. It was concluded that the test preparation solution was found stable up to 48 hours at 2 - 5˚ C and ambient temperature as during this time the result was not decreased below the minimum percentage. The result of robustness study showed that during all variance conditions, assay value of the test preparation solution was not affected and it was in accordance with that of actual. System suitability parameters were also found satisfactory; hence the analytical method would be concluded as robust.
CONCLUSION
A new analytical method has been developed to be routinely applied to simultaneous determination of Amlodipine besylate, Olmesartan medoxomil and Hydrochlorothiazide in pharmaceutical dosage form. In this study, stability of Amlodipine besylate, Olmesartan medoxomil and Hydrochlorothiazide in present dosage form was established through employment of ICH recommended stress condition. The developed procedure has been evaluated over the specificity, linearity, accuracy, precision and robustness in order to ascertain the stability of the analytical method. It has been proved that it was specific, linear, precise, accurate and robust and stability indicating. Hence, the method is recommended for routine quality control analysis and also for stability sample analysis. Acknowledgements The authors are thankful for facilities & grants given under UGC- Special Assistance Programme Department Research Support (DRS) and Department of Science & Technology (DST) New Delhi, Fund For Improvement of Science & Technology (FIST), National Facility for Drug Discovery (NFDD) and Department of Chemistry, Saurashtra University, Rajkot – 360 005 (INDIA) for providing analytical facilities.
REFERENCES
[1] Bakshi, M.; Singh, S. Journal of Pharmaceutical and Biomedical Analysis 2004, 36, 769–775. [2] Ojha, T.; Bakshi, M.; Chakraborti, A. K.; Singh, S. Journal of Pharmaceutical and Biomedical Analysis 2003, 31, 775–783. [3] Bakshi, M.; Singh, B.; Singh, A.; Singh, S. Journal of Pharmaceutical and Biomedical Analysis 2001, 26, 891–897. [4] Handrock, R.; Herzig, S. Deutsche Medizinische Wochenschrift 1998, 123, 19–20.
Hitendra S. Joshi et al Der Pharmacia Sinica, 2013, 4(2):145-152 _____________________________________________________________________________
152 Pelagia Research Library
[5] Puscas, I.; Gilau, L.; Coltau, M.; Pasca, R.; Domuta, G.; Baican, M.; Hecht, A. Cardiovascular Drugs and Therapy 2000, 14, 523–528. [6] Fujiwara, T. Japanese Pharmacology & Therapeutics 2006, 34, 19–56. [7] Xu, Y.; Lin, L.; Peng, F. Guangdong Yaoxueyuan Xuebao 2004, 20, 289–292. [8] Fritel, D.; Quichaud, J.; Hodara, M.; Truffert, J. Revue francaise d’etudes cliniques et biologiques 1961, 6, 560–573. [9] Puscas, I.; Coltau, M.; Baican, M.; Pasca, R.; Domuta, G.; Hecht, A. Drugs under Experimental and Clinical Research 2001, 27, 53–60. [10] Sharma, H. K.; Jain, N.; Jain, S. K. Pharmaceutica Analytica Acta 2011, 2, 126. [11] Patel, S. R.; Patel, C. N. Novel Science International Journal of Pharmaceutical Science 2012, 1, 317–321. [12] Darwish, H. W.; Backeit, A. H. Journal of fluorescence 2013, 23, 79–91. [13] Rai, M.; Kawde, P. B. Research Journal of Pharmaceutical, Biological and Chemical Sciences 2013, 4, 560–567. [14] Doshi, N.; Sheth, A.; Patel, C. N. International Research Journal of Pharmacy 2012, 3, 178–182. [15] Dubey, N.; Jain, A.; Sardiya, J.; Jain, D. K. Asian Journal of Chemistry 2012, 24, 4644–4646. [16] Rao, J. R.; Rajput, M. P.; Yadav, S. S.; Mulla, T. S.; Bharekar, V. V International Journal of PharmTech Research 2011, 3, 1435–1440. [17] Dubey, N.; Jain, A.; Raghuwanshi, A. K.; Jain, D. K. Journal of Planar Chromatography--Modern TLC 2012, 25, 475–480.
IOSR Journal of Pharmacy and Biological Sciences (IOSR-JPBS)
e-ISSN: 2278-3008. Volume 5, Issue 5 (Mar. – Apr. 2013), PP 22-28 www.iosrjournals.org
www.iosrjournals.org 22 | Page
Stability Indicating HPTLC Method for Estimation of Modafinil
in the Bulk and Tablet Formulation
Gaurang P. Pandya, Dr. Hitendra S. Joshi* Department of Chemistry, Saurashtra University, Rajkot-360 005, Gujarat, India.
Abstract: A simple, selective, precise and stability-indicating high-performance thin layer chromatographic
method for analysis of Modafinil, both as the bulk drug and in a tablet formulation has been developed and
validated. Aluminum foil TLC plates precoated with silica gel 60F 254 were used as stationary phase ethyl
acetate, acetone and methanol in the volume ratio of (7:2:1 v/v) respectively as mobile phase. A compact band
(Rf 0.42±0.02) was obtained for modafinil. Densitometric analysis was performed in absorbance mode at 232
nm. Linear regression analysis revealed a good linear relationship (r2=0.9995) between peak area and
concentration in the range of 80-320 ng /spot. The method was validated for precision, recovery, and
robustness. The limits of detection and quantitation were 15 and 50 ng/spot, respectively. Modafinil was
subjected to acid and alkaline hydrolysis, oxidation, photochemical and thermal degradation and underwent
degradation under all these conditions. Statistical analysis proved the method enables repeatable, selective, and
accurate analysis of the drug. It can be used for identification and quantitative analysis of Modafinil in the bulk
drug and in tablet formulations.
Key Words: Modafinil HPTLC Validation Stability-indicating Degradation
I. Introduction Modafinil is chemically 2-[(diphenylmethyl)sulfinyl]acetamide having molecular formula C15H15NO2S
and molecular mass 273.35 gm/mol (Figure 1). Modafinil is an analeptic drug manufactured by Cephalon, and
is approved by the U.S. Food and Drug Administration (FDA) for the treatment of narcolepsy, shift work sleep
disorder and excessive daytime sleepiness associated with obstructive sleep apnea. Narcolepsy is caused by
dysfunction of a family of wakefulness-promoting and sleep-suppressing peptides, the orexins. Orexin neurons
are activated by modafinil. Modafinil also affects tuberomammillary nucleus [1-2].
Figure-1 Structure of Modafinil
However extensive research into the interaction of modafinil with a large number of neurotransmitter systems
has been done, an exact mechanism or set of mechanisms of action remains uncertain. It seems that modafinil,
like other stimulants, increases the release of monoamines [3-4], specifically the catecholamine4,
norepinephrine [5-6] and dopamine from the synaptic terminals. Though, modafinil also elevates hypothalamic
histamine levels, leading some researchers to consider Modafinil a "wakefulness promoting agent" rather than a
classic amphetamine-like stimulant [7-8]. Even though modafinil's histaminergic action, it still partially shares
the actions of amphetamine-class stimulants due to its effects on norepinephrine and dopamine [9-11]. Modafinil has been prohibited by World Anti Doping Agency since 2004 so that there are so many bio
analytical methods are reported to detect Modafinil in urine and blood plasma by GC-MS[12], LC-MS/MS[12-
14],UPLC-MS[15-17] and HPLC[22]. Some methods are also reported for chiral resolution of modafinil
enantiomers by HPLC [18-21]. Besides bioanalytical methods few methods are reported to determine Modafinil
in tablet dosage form by HPLC [23-25]. Furthermore these methods are not impressionable to achieve the high
throughput study which can be possible by optimizing the method in such a way which includes cost
effectiveness with optimum selectivity. Hence, it can be maximum utilize for the analysis of formulation
development and stability testing as well as at quality control laboratory for routine use.
In recent times, there is an increase tendency towards the development of stability-indicating assay,
using the approach of stress testing as mentioned in the ICH guidelines (Q1A). It is also recommended to carry
Stability Indicating HPTLC Method for Estimation of Modafinil in the Bulk and Tablet Formulation
www.iosrjournals.org 23 | Page
out stress testing on the drug substance to establish its inherent stability characteristics and to hold up the
suitability of the proposed analytical method. In this study our approach is to study degradation behavior of
Modafinil and development of a validated stability indicating HPTLC assay method.
II. Experimental 2.1 Chemicals and reagents
The modafinil working standard was provided as a gift sample by Alembic pharmaceutical Ltd,
Baroda. All chemicals and reagents used were of AR grade. The formulation, 100 mg and 200 mg is available in
market by brand name PROVIGIL.
2.2 HPTLC Instrumentation and conditions Chromatography was performed on Merck TLC plates pre-coated with silica gel 60 F254 (10 cm ×10
cm with 250 μm layer thicknesses) from E. Merck, Germany. Before use the plates were prewashed with
methanol then dried in the current of dry air and activated at 110 °C for 5 min. The samples were applied onto
the plates as a band with 4 mm width using Camag 100 μl sample syringe (Hamilton, Switzerland) with a
Linomat 5 applicator (Camag, Switzerland). The mobile phase consisted of ethyl acetate-acetone-methanol
(7:2:1 v/v/v). Linear ascending development was carried out in a twin trough glass chamber (for 10 x 10 cm)
previously saturated with mobile phase vapour for 30 min at room temperature and relative humidity 60 ± 5%.
The development distance was approximately 80 mm. After development the plates were dried in current of air
by use of an air dryer. Densitometric scanning, at 232 nm, was performed with a Camag TLC scanner III,
operated by Wincats Software (V 1.4.2, Camag) in absorbance mode. The selection of wavelength was based on
maximum absorbance for optimum sensitivity. The source of radiation was a deuterium lamp emitting a
continuous UV spectrum in the range 190–400 nm. The slit dimensions were 5 mm × 0.45 mm.
2.3 Preparation of standard solution The standard modafinil 10 mg was weighed accurately and transferred to volumetric flask 100 ml. It
was dissolved and sonicated for few minutes and diluted up to the mark with methanol to obtain final
concentration of 100 μg/ml and the resulting solution was used as working standard stock solution.
2.4 Preparation of sample solution To determine the content of modafinil in tablets formulations, 20 branded tablets were weighed; their
mean weight was calculated and crushed to fine powder. Tablet powder equivalent to 1 tablet of modafinil was
weighed and transferred to 100 ml volumetric flask then dissolved with methanol and further diluted with
methanol. It was kept for ultrasonication for few min; the solution was then filtered through Whatmann filter
paper No. 41 and further dilution was made with methanol to get the final sample solution of 100 μg/ml. The resultant solution was used as such for analysis.
2.5 Method validation
The method was validated in compliance with ICH guidelines. The following parameters were
performed and evaluated for the validation of developed method.
2.5.1 Linearity Appropriate volume of aliquots from standard modafinil stock solutions were prepared and applied on
the TLC plate covering the range from 200-800 ng/spot with the help of micro liter syringe using an automatic
sample applicator. The plates were developed, dried and scanned densitometrically at 232 nm. Each
concentration was applied seven times to the plate and the plate was developed as described above. Peak areas were plotted against corresponding concentrations to furnish the calibration plot
.
Table 1
Regression analysis of calibration curve
Parameters Result
Linearity Range(ng/Band) 200-800
correlation coefficient 0.998
Slope 82.46
Intercept 120.2
Stability Indicating HPTLC Method for Estimation of Modafinil in the Bulk and Tablet Formulation
www.iosrjournals.org 24 | Page
2.5.6 Accuracy To check the recovery of the drug at different levels in formulations, samples were analysed at 3
different level 50, 100, and 150% of modafinil standard by proposed method and the recovered amount of drugs were calculated. The experiment was conducted in triplicate.
Table 2
Result of recovery studies
Sr. no. Level of
Recovery
Amount of
drug
taken(mg)
Amount of
drug found(mg)
%
Recoverya) % RSD
1 50 50 50.11 100.22 1.859
2 100 100 100.31 100.31 1.797
3 150 150 149.97 99.38 1.491
a) Mean of three determinations
2.5.3 Precision
Precision of the developed method was measured by repeatability, intra-day and inter-day precision. In
repeatability study analyses of six replicates of same concentration (500 ng) were done by developed method.
For intra-day and inter-day precision, three sample sets of tablet sample and standard were analyzed in triplicate by developed method on the same day and on three different days, respectively.
2.5.3 Robustness
Small changes in the chromatographic conditions were introduced and the effects on the results were
examined. Slight changes in the composition of mobile phase (±0.1ml) were made in the used optimized mobile
phase. The time from spotting to chromatography and time from chromatography to scan was varied at ±5
minutes. The chromatographic development distance was varied in the range of ±5 mm. The effects of these
changes on both RF value and peak area were assessed by calculating relative standard deviation (%RSD).
Table 3
Result of robustness studies
Condition RSD (%)
mobile phase composition 1.34
amount of mobile phase 0.95
chamber saturation time 1.21
time from application to chromatography 1.58
time from chromatography to scanning 0.66
Distance from application to solvent front 0.83
2.5.4 Limits of Detection and Quantification To determine the limits of detection (LOD) and limit of quantification (LOQ), different dilution of
standard solution of drug were applied along with methanol as the blank. The LOD and LOQ were determined
on the basis of signal to noise ratio until the average responses of triplicate analysis were obtained
approximately 3 and10 times the responses of the blank respectively.
2.5.5 Specificity
The specificity of the method was determined by analysis of drug standards and samples. The band for
modafinil in the sample was identified by comparing the Rf value and spectrum of the band with those of the
band from a standard. The peak purity of modafinil was assessed by comparing spectra acquired at three
different positions on the peak, i.e. the peak start (S), peak apex (M), and peak end (E) positions of the peak.
2.6 Analysis of Marketed Formulation
Assay of marketed tablet was done through developed HPTLC method. The sample solution was made
as depicted in section 2.4. Same volume (5 µL) of Standard solution and sample solution were applied on TLC.
The experiment was carried out for six sample solutions. The chromatogram obtained is shown in Figure 2.
Stability Indicating HPTLC Method for Estimation of Modafinil in the Bulk and Tablet Formulation
www.iosrjournals.org 25 | Page
Table 4
Summary of validation parameters
Parameter Result %RSD
λ (nm) 232
RF 0.42
Linear Range (ng/Band) 200-800
Accuracy(at 100% level) 100.31 1.797
Precision
Intraday precision 99.98 1.579
interday precision 99.57 0.983
Limit Of Detection (ng/Band) 15
Limit Of Quantitation (ng/Band) 50
Figure-2 Chromatogram of Modafinil drug sample
Table 5
Analysis of tablet formulation
Method Drug Lable claim
[mg/tablet]
Amount found
[mg/tablet]a)
Drug Assay
[%]a) %RSD
TLC-Densitometry
Modafinil 100 99.84 99.84 0.786
a) Mean of six determinations
2.7 Forced Degradation studies A stock solution containing 10 mg modafinil in 10 ml methanol was prepared. This solution was used
for forced degradation to provide an indication of the stability indicating property and specificity of the method.
In all degradation studies the average peak area of modafinil after application (500 ng/spot) of seven replicates
was obtained after development and scanning of the plate as described above.
2.6.1 Acid and Base Degradation Modafinil (10 mg) was separately dissolved in 10 ml of methanolic solution of 0.2 N HCl and 0.5 N
NaOH. These solutions were refluxed at 80ºC for 2 h in the dark in order to exclude the possible degradative
effect of light. 1 ml of treated solution were taken and neutralized and then diluted up to 10 ml with methanol. Then this diluted sample was used to apply on TLC plate in triplicate (500 ng/spot). The plates were
chromatographed and analysed by proposed method.
Stability Indicating HPTLC Method for Estimation of Modafinil in the Bulk and Tablet Formulation
www.iosrjournals.org 26 | Page
2.6.2 Oxidative Degradation Modafinil (10 mg) was dissolved in 10 ml of methanolic solution of hydrogen peroxide (30%, v/v) and
the mixture was kept for 48 h at room temperature in the dark, to exclude the possible degradative effect of light. The solutions (1 ml) were diluted to 10 ml with methanol and further proceed as described for acid and
base degradation.
2.6.3 Dry Heat Degradation
Dry heat studies were performed by keeping powdered drug in oven (600 C) for 6 hrs. Samples were
withdrawn, dissolved in methanol, diluted suitably and appropriate volumes of resultant solution were applied
on TLC plate and densitograms were developed.
2.6.4 Photo Degradation
Photo degradation studies were carried out by keeping modafinil drug sample in sunlight for 4 days.
Sample solution was taken and diluted in methanol, and appropriate volume of the resultant solution were
applied and chromatographed as described above.
III. Results And Discussion 3.1 Development of the optimum mobile phase
TLC procedure was optimized with a view to develop a stability-indicating assay method. The drug
reference standards were spotted on the TLC plates and developed in different solvent systems. Different mobile
phases were tried to resolve modafinil and degradation products. Best suited mobile phase was found to ethyl
acetate, acetone and methanol in the volume ratio of (7:2:1 v/v) respectively. Developed mobile phase enabled
good resolution, and a sharp and symmetrical peak of RF 0.42 for modafinil (Fig. 2) form a compact and non-
diffused band. It was observed that prewashing of TLC plates with methanol (followed by drying and activation) and pre-saturation of TLC chamber with mobile phase for 30 min (the optimum saturation time) ensured good
reproducibility and peak shape.
3.2 Validation of the developed stability-indicating method
The response for the drug was found to be linear in the concentration range 200-800 ng/band for
modafinil. The results of regression analysis are depicted in Table 1. The accuracy of the method was
determined at 50, 100 and 150% level. Results of accuracy study are expressed in terms of % recovery (Table
2). According to results of robustness studies shown in Table 3. %RSD value does not exceed more than 2. So
the developed method is robust. The %RSD value for precision study was found to be not more than 2 % for
modafinil, thus confirming precision of the method. The Limit of Detection and Limit of Quantitation were
found to be 15 ng/band and 50 ng/band respectively for analyte. The results are shown in Table 4. The
specificity of the method was ascertained by peak purity profiling studies. In analysis of marketed formulation the mean assay result was found to be 99.84% (Table 5). From the result obtained we can conclude that there is
no interference of the excipients present in the formulation. The low %RSD suggests the suitability of developed
method for routine analysis.
3.3 Degradation observed
After applying different stress conditions upon modafinil, following results were observed by analyzing
samples using the developed method. About 21.50% drug was degraded (Figure 3) under oxidative condition
and major degradation peaks were observed at 0.18, 0.20 and 0.23 RF values. In acid hydrolysis modafinil was
degraded (Figure 4) about 17.35% and degraded product was observed at 0.14 and 0.19 RF value, however in
alkali hydrolysis some harsh condition was required compared to that in acid degradation. Initially, 0.1 N NaOH
was used and refluxed at 800C for 2 hours, yet no degradation was observed; then strength of alkali was increased to 0.5 N and refluxed at 800C for 2 hours resulting in 8.20% degradation (Figure 5) and degradation
peaks were observed at 0.23, 0.27 and 0.32 RF values. Under dry heat (Oven, 80 0C, 6 hr), drug was found to be
degraded up to 9.25% with decrease in area only, whereas under Photolytic studies, no additional peaks were
observed and drug peak area remained constant. This indicates stability of modafinil in UV light. In all forced
degradation condition the drug peak was well resolved from degradation products.
Stability Indicating HPTLC Method for Estimation of Modafinil in the Bulk and Tablet Formulation
www.iosrjournals.org 27 | Page
Figure-3 Chromatogram of Modafinil after Oxidative degradation
Figure-4 Chromatogram of Modafinil after Acid degradation
Figure-5 Chromatogram of Modafinil after Alkali degradation
Stability Indicating HPTLC Method for Estimation of Modafinil in the Bulk and Tablet Formulation
www.iosrjournals.org 28 | Page
IV. Conclusion Based on the above study and result, it can be concluded that the developed HPTLC method is simple,
precise, accurate, rapid, selective and sensitive for the determination of modafinil in tablet formulation as well
as in bulk drugs. The products formed after forced decomposition studies were resolved from the bulk drug
response. From the peak purity profile studies, it was confirmed that the peak of the degradation product was not
interfering with the peak of drugs. It confirms that peak for degradation product of drug can be resolved from
the drug peak by this method; hence it can be used as stability indicating method. The method can minimize the
time and cost of analysis. It has also utilized the merit of applying several samples on TLC plates which can be
advantageous for regulatory quality control laboratories. The method does not require expensive chemical and
solvents, thus it can stand for a good alternative for already existing hplc methods. It is proposed for routine
analysis of these drugs in presence of degradation products in stability study.
Acknowledgements The authors are thankful for facilities & grants given under UGC- Special Assistance Programme
(SAP),Department Research Support (DRS), Department of Science & Technology (DST) New Delhi, Fund For
Improvement of Science & Technology (FIST), National Facility for Drug Discovery (NFDD) and Department
of Chemistry, Saurashtra University, Rajkot – 360 005 (INDIA) for providing analytical facilities.
References [1] R. Chemelli, J. Willie, C. Sinton, J. Elmquist, T. Scammell, C. Lee, J. Richardson, S. Williams, Y. Xiong, Y. Kisanuki et al, Cell
(Cambridge, Massachusetts) 98 (1999) 437-451.
[2] T. Scammell, I. Estabrooke, M. McCarthy, R. Chemelli, M. Yanagisawa, M. Miller, C. Saper, Journal of Neurosci. 20 (2000) 8620-
8628.
[3] D. Zara, M. Orosco, C. Rouch, B. Gerard, S. Nicolaidis, NeuroReport, 12 (2001) 3533-3537.
[4] J. Duteil, F. Rambert, J. Pessonnier, J. Hermant, R. Gombert, E. Assous, Euro. J. Pharmaco. 180 (1990) 49-58.
[5] B. Madras, Z. Xie, Z. Lin, A. Jassen, H. Panas, L. Lynch, R. Johnson, E. Livni, T. Spencer, A. Bonab, et al Journal of Pharmaco.
Experi. Therap. 319 (2006) 561-569.
[6] M. Minzenberg, A. Watrous, J. Yoon, S. Ursu, C. Carter, Science (Washington, DC, U.S.), 322 (2008) 1700-1702.
[7] J. Lin, B. Roussel, H. Akaoka, P. Fort, G. Debilly, M. Jouvet, Brain Res. 591 (1992) 319-26.
[8] J. Lin, Y. Dauvilliers, I. Arnulf, H. Bastuji, C. Anaclet, R. Parmentier, L. Kocher, M. Yanagisawa, P. Lehert, X. Ligneau et al,
Neurobio. Disease 30 (2008) 74-83.
[9] T. Ishizuka, Y. Sakamoto, T. Sakurai, A. Yamatodani, Neurosci. Letters. 339 (2003) 143-146.
[10] T. Ishizuka, Y. Sakamoto, T. Sakurai, A. Yamatodani, Euro. J. Pharmaco. 578 (2008) 209-215.
[11] R. Parmentier, C. Anaclet, C. Guhennec, E. Brousseau, D. Bricout, T. Giboulot, D. Bozyczko-Coyne, K. Spiegel, H. Ohtsu, M.
Williams et al, Biochem. Pharmaco. 73 (2007) 1157-1171.
[12] S. Dubey, S. Ahi, I. Reddy, T., Kaur, A. Beotra, S. Jain, Ind. J. Pharmaco. 41 (2009) 278-283.
[13] M. Mazzarino, X. Torre, F. Botre, Anal. Bioanal. Chem. 392 (2008) 681-698.
[14] M. Mazzarino,F. Botre, Rapid Com. Mass Spectr. 20 (2006) 3465-3476.
[15] G. Jimenez, K. Deventer, K. Roels, E. Van, Anal. Chim. Acta 721 (2012) 137-146.
[16] E. Deconinck, K. Verlinde, P. Courselle, J. De Beer, J. Pharm. Biomed. Anal. 59 (2012) 38-43.
[17] H. Thoerngren, F. Oestervall, M. Garle J. Mass Spectr. 43 (2008) 980-992.
[18] T. Lourenco, D. Armstrong, Q. Cass, Chromatographia 71 (2010) 361-372.
[19] Y. Ji, X., Fan, X. Zhu, Zhongguo Yiyao Gongye Zazhi 34 (2010) 251-256.
[20] R. Rao, D. Shinde, Biomed. Chrom. 23 (2009) 811-816.
[21] J. Donovan, R. Malcolm, J. Markowitz, C. DeVane, Therap. Drug Monit. 25 (2003) 197-202.
[22] H. Schwertnera, S. Kong, J. Pharm. and Biomed. Anal. 37 (2005) 475–479.
[23] X. Li, M. Jin, J. Yi, Zhongguo Yaofang 18 (2007) 1729-1730.
[24] C. Li, H. Lu, Guangdong Yaoxueyuan Xuebao, 18 (2002) 90-92.
[25] H. Ma, T. Guo, M. Yan, Zhongguo Yiyao Gongye Zazhi, 40 (2009) 377-378.