2
“It’s the small things that count.” Airmodus Oy Erik Palménin aukio 1 00560 Helsinki, Finland +358 50 5666043 www.airmodus.com [email protected] A23 bCPC www.airmodus.com Designed for vehicle emission measurements Airmodus A23 Condensation Particle Counter is a user-friendly tool for all applications where counting aerosol particles larger than 23 nm is a necessity. The A23 CPC is compliant with the Particle Measurement Protocol (PMP) for EURO 5/6. It fulfills the requirements of UN/ECE R49 and UN/ECE R83. A versatile particle counter The A23 can be used both as a stand-alone instrument for measuring the total particle number concentration, and as a counter in different kinds of aerosol measurement systems. It is easy to use and handle. All settings can be quickly adjusted from a handy touch screen, which also displays the current concentration reading and instrument diagnostics. The A23 is also compatible with the Airmodus Particle Size Magnifier A10. Benefits of the A23 - The saturator is made of robust and inert stainless steel in order to ensure stable operation - Narrow pulse width for higher counting accuracy - Easy to use touch screen - Improved usability and design: all connections optimized for easy access Regulation requirements A23 Dp 50% (±12%) = 23 nm ± 1 nm ü 23 nm Dp >90% = 41 nm ± 1 nm ü very sharp cut-off curve Linear concentration response between 0 - 10 000 #/cm3 ü 0 - 30 000 #/cm3

Designed for vehicle emission measurements€œIt’s the small things that count.” Airmodus Oy Erik Palménin aukio 1 00560 Helsinki, Finland +358 50 5666043 [email protected] A23

Embed Size (px)

Citation preview

Page 1: Designed for vehicle emission measurements€œIt’s the small things that count.” Airmodus Oy Erik Palménin aukio 1 00560 Helsinki, Finland +358 50 5666043 info@airmodus.com A23

“It’s the small things that count.”

Airmodus Oy Erik Palménin aukio 1 00560 Helsinki, Finland

+358 50 5666043 www.airmodus.com [email protected]

A23 bCPC www.airmodus.com

Designed for vehicle emission measurements Airmodus A23 Condensation Particle Counter is a user-friendly tool for all applications where counting aerosol particles larger than 23 nm is a necessity. The A23 CPC is compliant with the Particle Measurement Protocol (PMP) for EURO 5/6. It fulfills the requirements of UN/ECE R49 and UN/ECE R83.

A versatile particle counter The A23 can be used both as a stand-alone instrument for measuring the total particle number concentration, and as a counter in different kinds of aerosol measurement systems. It is easy to use and handle. All settings can be quickly adjusted from a handy touch screen, which also displays the current concentration reading and instrument diagnostics.

The A23 is also compatible with the Airmodus Particle Size Magnifier A10. Benefits of the A23

- The saturator is made of robust and inert stainless steel in order to ensure stable operation

- Narrow pulse width for higher counting accuracy - Easy to use touch screen - Improved usability and design: all connections optimized for easy access

Regulation requirements A23

Dp50% (±12%) = 23 nm ± 1 nm ü 23 nm

Dp>90% = 41 nm ± 1 nm ü very sharp cut-off curve

Linear concentration response between 0 - 10 000 #/cm3

ü 0 - 30 000 #/cm3

Page 2: Designed for vehicle emission measurements€œIt’s the small things that count.” Airmodus Oy Erik Palménin aukio 1 00560 Helsinki, Finland +358 50 5666043 info@airmodus.com A23

Airmodus Oy Erik Palménin aukio 1 00560 Helsinki, Finland

+358 50 5666043 www.airmodus.com [email protected]

A23 bCPC Specifications www.airmodus.com

*) Cut-off size in mobility equivalent diameter. See calibration certificate. **) With high relative humidity, an aerosol drier should be used to prevent excess water condensation inside the instrument. Microsoft and Windows are registered trademarks of Microsoft Corporation.

Particle size range 23 nm – 2.5 µm Dp50% = 23 nm*

Concentration 0 – 100 000 #/cm3 Up to 30 000 #/cm3 in single particle counting mode with coincidence <10%; higher concentrations with Total Scattering Mode Correction

Aerosol sample flow 1.00 ± 0.06 lpm, controlled with a critical orifice

Response time t90 < 3 s

False counts <0.0001 #/cc

Working fluid n-Butanol (>99.5%)

Sample conditions

Pressure: 75 to 105 kPa Relative humidity: 0 to 95% non-condensing (preferably<40%)**

Environmental conditions

Temperature: 15oC to 35oC Pressure: 75 to 105 kPa Relative humidity: 0 to 95% non-condensing

Communication

Analog in: BNC connector, 0 to 10 V (reading data of external sensor) Analog out: BNC connector, 0 to 10 V, user-selectable function output (linear concentration, also DMA voltage control) Pulse out: BNC connector Serial: RS-232 Ethernet: RJ45 USB: type B connector All communication based on ASCII character-encoding scheme.

Fittings External Vacuum: 1/4 in. stainless steel tube Inlet: 1/4 in. stainless steel tube

Software Airmodus A2X software for online data acquisition (for Microsoft Windows)

External vacuum requirement

100 - 400 mbar pressure at NTP (or <40% of inlet pressure)

Power requirements 100 - 240 VAC max. 320 W universal AC input/full range

Dimensions and weight

260x230x400 (height x width x depth in mm) 10.5 kg

Shipping conditions Temperature: 0 - 40oC Relative humidity: <95% non condensing The instrument should be shipped in upright position and should be protected against tremor and blows.