Upload
rosamund-nichols
View
223
Download
3
Embed Size (px)
Citation preview
Design of Low Power Instrumentation Amplifier
Instrumentation Amplifier
Design Goal
• I want to design an amplifier with more than 40 dB in Gain
• Less than 10 uA of current
Top Level Schematic
1st stage amplifier 2nd stageBias circuit
Design Strategy
• Get a voltage gain of 10 from the first stage (By choosing R1 and RG carefully)
• Get the rest of the gain from the second stage
• Reuse amplifier cell
Active Loaded Differential Amplifier
2 uA of currentVoltage gain of 24.56
Biasing of M3
2 uA of current
Testing of the ECG Circuit
(Biasing Circuit)
Low Voltage Cascode Bias Cicuit
To keep M2 in saturation: Vx>Vb-Vth→Vx+Vth2>Vb
To keep M1 in saturation: VA>Vx-Vth1
Since VA=Vb-VGS2, Vb>Vx-Vth1+VGS2
Design criteria for M2
Vb Requirement
Vb=VOD3+VGS4 to produce a minimum outputVoltage of VOD3 and VOD4.
By design, VGS4=VGS2, VA=VB
Vb=VOD2+VTH2+VOD1.
Vb Generation
Requirement:Vb=VOD2+VTH2+VOD1
VGS5=VGS2
VOD1=VGS6-VTH7
Problem: M5 suffers from no body effectM2 suffers from body effect
Design M7 (Large W7/L7) so that VGS7 is approx. VTH7
Need to have sufficiently largeVGS6, otherwise M6=triode since M7=Sat.
Testing of the ECG Circuit
(Biasing Circuit)
Biasing Circuitry
Top Level Schematic
1st stage amplifier 2nd stageBias circuit
2 uA
2 uA
2 uA
2 uA
Gain of the Amplifier
Trasient Simulation
Vin,diff=2 mVVout,single=336 mVVout,single/Vin,diff=168