12
International Journal of Bio-Science and Bio-Technology Vol.6, No.5 (2014), pp.55-66 http://dx.doi.org/10.14257/ijbsbt.2014.6.5.07 ISSN: 2233-7849 IJBSBT Copyright ⓒ 2014 SERSC Design of CMOS Two-stage Operational Amplifier for ECG Monitoring System Using 90nm Technology Fateh Moulahcene, Nour-Eddine Bouguechal, Imad Benacer and Saleh Hanfoug Advanced Electronics Laboratory (LEA), University of Batna, Avenue Chahid Boukhlouf, 05000, Batna, Algeria [email protected] Abstract This paper presents a high performance Two-stage operational amplifier for biomedical applications. This Two-stage is designed for low noise, low power, high PSRR and high CMRR. The Miller compensation technique (Cc) is used with a nulling active resistance (Rz) implemented using Transmission gate transistors for stable operation in feedback mode. The operational amplifier was manufactured in a SPECTRE using GPDK 0.90-um CMOS technology with threshold voltages of a 0.17 V and - 0.14 V achieve a low power 2.6uW, high CMRR up to 130dB and PSRR up to 70dB at 1V power supply. Keywords: Two-stage amplifier, Miller compensation, Transmission gate, CMRR (Common mode Rejection Ratio) 1. Introduction The trend in the use of low-voltage power supplies for CMOS integrated circuits is by the reliability issue of small size MOSFET transistors and the increasing use of low-weight long life battery-operated portable electronic systems [1, 2]. With the rapid development of microelectronics in the recent past years, more and more applications require an ultra-low amplitude signal measurement module, such as implantable devices in biomedical application to monitor several Neuro-muscular activities [3-6]. Monitoring biomedical signals of the human body is a very interesting topic since it can be used to know vital health information of the body from the acquired data. These data capacitates medical practitioners to diagnose diseases [7, 8]. Biomedical signals, such as electrocardiogram (ECG), Electromyogram (EMG), Electroencephalogm (EEG), are characterized by their low voltage-levels and low frequency, Table 1 shows the characteristics of these signals [8 -12] Table 1. Most Commonly used Biomedical Signals The Famous architecture used in a biomedical monitoring system, are an operational amplifier, must exhibit very low-referred noise, low power and high Common Mode Rejection Ratio (CMRR). Signal Frequency Amplitude ECG 0.05-250 Hz 5 uV-8 mV EEG 0.5-200 Hz 2 uV-200 uV EMG 0.01-10 KHz 50 uV-10 mV

Design of CMOS Two-stage Operational Amplifier for ECG

  • Upload
    buitruc

  • View
    235

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014), pp.55-66

http://dx.doi.org/10.14257/ijbsbt.2014.6.5.07

ISSN: 2233-7849 IJBSBT

Copyright ⓒ 2014 SERSC

Design of CMOS Two-stage Operational Amplifier for ECG

Monitoring System Using 90nm Technology

Fateh Moulahcene, Nour-Eddine Bouguechal, Imad Benacer and Saleh Hanfoug

Advanced Electronics Laboratory (LEA), University of Batna, Avenue Chahid

Boukhlouf, 05000, Batna, Algeria

[email protected]

Abstract

This paper presents a high performance Two-stage operational amplifier for biomedical

applications. This Two-stage is designed for low noise, low power, high PSRR and high

CMRR. The Miller compensation technique (Cc) is used with a nulling active resistance (Rz)

implemented using Transmission gate transistors for stable operation in feedback mode. The

operational amplifier was manufactured in a SPECTRE using GPDK 0.90-um CMOS

technology with threshold voltages of a 0.17 V and - 0.14 V achieve a low power 2.6uW, high

CMRR up to 130dB and PSRR up to 70dB at 1V power supply.

Keywords: Two-stage amplifier, Miller compensation, Transmission gate, CMRR

(Common mode Rejection Ratio)

1. Introduction

The trend in the use of low-voltage power supplies for CMOS integrated circuits is by the

reliability issue of small size MOSFET transistors and the increasing use of low-weight long

life battery-operated portable electronic systems [1, 2].

With the rapid development of microelectronics in the recent past years, more and more

applications require an ultra-low amplitude signal measurement module, such as implantable

devices in biomedical application to monitor several Neuro-muscular activities [3-6].

Monitoring biomedical signals of the human body is a very interesting topic since it can be

used to know vital health information of the body from the acquired data. These data

capacitates medical practitioners to diagnose diseases [7, 8].

Biomedical signals, such as electrocardiogram (ECG), Electromyogram (EMG),

Electroencephalogm (EEG), are characterized by their low voltage-levels and low frequency,

Table 1 shows the characteristics of these signals [8 -12]

Table 1. Most Commonly used Biomedical Signals

The Famous architecture used in a biomedical monitoring system, are an operational

amplifier, must exhibit very low-referred noise, low power and high Common Mode

Rejection Ratio (CMRR).

Signal Frequency Amplitude

ECG 0.05-250 Hz 5 uV-8 mV

EEG 0.5-200 Hz 2 uV-200 uV

EMG 0.01-10 KHz 50 uV-10 mV

Page 2: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014)

56 Copyright ⓒ 2014 SERSC

In this paper, a low voltage, low noise and high CMRR operational amplifier for portable

monitoring system is proposed. The operational amplifier is able to work under 1-V supply

and has high CMRR.

2. ECG Data Acquisition System

ECG data acquisition begins with the use of electrodes that attached to body skin. Three

electrodes are used as sensors to detect the heart signals from a human body. Two electrodes

are placed each on the left and right wrist while the third electrode is placed on the ankle of

the leg as ground [13].

The ECG signal that acquired is in the range of 5uV to 8 mV. Due to the weak voltage

level, the signal is fed into an amplifier circuit to be amplified to a desirable voltage level.

Output from the amplifier is then fed into a bandpass filter circuit and a High Q notch filter.

The purpose of this filter is to filter out the very low and high frequency noise components of

the signals and the 60-Hz power line interference. The desirable analog output from the filter

is then sent to S/H and ADC to become a digital signal. After that, these digital data will

processed in PCs or microprocessors. Figure.1 shows the proposed ECG data acquisition

system [7, 13-16].

Figure 1. The Proposed ECG Data Acquisition

3. ECG Signal

Electrocardiogram (ECG) signal plays a vital role in the monitoring and diagnosis of the

health conditions of the heart. An electrocardiogram is a recording of the small electric waves

being generated during heart activity. The main tasks in the ECG signal analysis are the

detection of QRS complex and the estimation of the instantaneous heart rate. Figure 2 shows

an example of a normal ECG trace, which consists of a P wave, a QRS complex and a T

wave.

Biomedical

signal ECG LowPass

filter Electrode Amp

Digital

Output

Sample –and-

Hold S/H

Analog-to-Digital

Converter(ADC)

Figure 2. Shows an Example of a Normal ECG Trace

Page 3: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014)

Copyright ⓒ 2014 SERSC 57

The P wave represents the atrial contractions.

QRS complex represents the ventricular contractions. The R peak indicates a

heartbeat.

The T wave is the last common wave in an ECG. This electrical signal is produced

when the ventricles are re-polarizing.

The heart is a muscular pump made up of four chambers. The two upper chambers are

called atria, and the two lower chambers are called ventricles. A natural electrical system

causes the heart muscle to contract and pump blood through the heart to the lungs and the rest

of the body [13].

We have collected data from MIT-BIH Arrhythmia Database [17-19], for training and

testing of proposed ECG amplifier Design. See Figure 3 of the input (Vin) ECG signal.

4. ECG Amplifier Design

A normal ECG signal falls in the range of 5uV – 8mV. The amplifier is required to

increase this weak signal into an acceptable level. The main operational amplifier for

biomedical applications is illustrated in Figure 4. This Two-stage architecture with miller

compensation capacitor (Cc) and nulling active resistor (Rz), is used for the design of the

operational amplifier with the first stage being a differential input pair and the second, a gain

stage. Because the Biomedical signal is so weak, the noise will affect the real biomedical

signal. Due to the low frequency of biomedical signal the flicker noise dominates, it has

strong dependence on the width and length product of a CMOS transistor. The flicker noise is

modeled in Eq (1), the spectral density of flicker noise is inversely proportional to the

transistor area, WL. In other words, 1/f noise can have a lesser effect on larger devices.

However the input pair (M1 and M2) of the input stage has been designed using large PMOS

transistors. A P-channel current source (M5) and an n-channel current mirror load (M3 and

M4) are used in the input stage.

Figure 3. ECG Signal from MIT-BIH Arrhythmia Database.(Vin(max)=1.5mV)

Page 4: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014)

58 Copyright ⓒ 2014 SERSC

fWLC

KV

OX

n

1.2 (1)

The second stage of the operational amplifier includes an n-channel common-source

amplifier (M6) with a p-channel current source load (M7).

The high output resistances of these two transistors (M7, M6) equate to a relatively large

gain for this stage and an overall moderate gain for the complete amplifier. Because the

operational amplifier inputs are connected to the gates of MOS transistors, the input

resistance is essentially infinite. The sizes of the transistors were designed for a bias current

of 0.5μA to provide for sufficient output voltage swing, output-offset voltage, slew rate, and

gain-bandwidth product.

4.1. Current Mirrors

Current mirrors are used extensively in MOS analog circuits both as biasing elements and

as active loads to obtain high AC voltage gain [20, 21]. Enhancement mode transistors remain

in saturation when the gate is tied to the drain, as the drain-to source voltage (VDS) is greater

than the gate-to-source voltage (VGS) due to the threshold voltage (Vth) drop, i.e.,

ThGSDS VVV (2)

Based on Eq (2), constant current sources are obtained through current mirrors designed

by passing a reference current through a diode-connected (gate tied to drain) transistor.

We choose a simple current mirror because low voltage (1-V), Figures 5(a) and (b) show

the PMOS and NMOS current mirrors design. A PMOS mirror serve as a current source while

the NMOS acts as a current sink. The voltage developed across the diode-connected transistor

is applied to the gate and the source of the second transistor, which provides a constant output

current. Since both the transistors have the same gate to source voltage, the currents when

both transistors are in the saturation region of operation, are governed by the following Eq (3)

and Eq(4) assuming matched transistors. The current ratio Iout/Iref is determined by the

𝑽𝒊𝒏+

𝐌𝟏𝟎

M8

𝐌𝟒

𝑪𝑪

Iref

𝐌𝟕

𝐌𝟗

Figure 4. Schematic of Two-stage Operational Amplifier

𝑽𝒊𝒏−

𝑽𝑫𝑫

𝑽𝑺𝑺

𝐌𝟏 𝐌𝟐

𝐌𝟑

𝐌𝟓

𝐌𝟔

𝑽𝒐𝒖𝒕

𝑪𝑷

Page 5: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014)

Copyright ⓒ 2014 SERSC 59

aspect ratios of the transistors. The reference current that was used in the design is 0.5μA. The

desired output current is 1μA.

For the PMOS current mirror, we can write,

8

8

7

7

L

WL

W

I

I

REF

OUT (3)

For the NMOS current mirror, we can write,

3

3

4

4

L

WL

W

I

I

REF

OUT (4)

Figure 5. (a) PMOS Current Mirror, (b) NMOS Current Mirror

4.2. Active Resistors

There are two active resistors used in the design. Firstly, the reference current that is

applied to the current mirror is obtained by means of an active resistor. The resistor here is

obtained by simply connecting the gate of an NMOS (M11) to its drain as shown in Figure 6.

This connection forces the MOSFET to operate in saturation.

The second active resistor a transmission gate (TG) shown in Figure 7 has been used to

realize the nulling active resistance (Rz) to reduce the effects of the right hand plane zero in

the transfer function. The gate of these transistors M9, M10 is biased at VDD, VSS

respectively. Its small signal output resistance is obtained from Eq (5) and Eq(6) (Vds very

small).

The dynamic range limitations associated with single channel MOS switches can be

avoided with CMOS switch shown in figure 7. Using the CMOS technology, a switch usually

M8 M7

𝐌𝟑 𝐌𝟒

𝑰𝑹𝑬𝑭=0.25uA

𝑰𝑶𝑼𝑻=0.25uA

𝐕𝐬𝐬 𝑰𝑹𝑬𝑭=0.5uA

𝑰𝑶𝑼𝑻=1uA

𝑽𝑫𝑫

(a) (b)

Figure 6. Active Resistor

M11

Page 6: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014)

60 Copyright ⓒ 2014 SERSC

constructed by connecting P-channel and N-channel enhancement mode in parallel as

illustrated.

For this configuration, When Q is low, both transistors are off, creating an effective open

circuit. When Q is high both transistors are on, giving a low impedance state. The bulk

potentials of the p-channel and n-channel devices are taken at the highest and lowest

potentials respectively. The primary advantage of the CMOS switch over the single channel

MOS switch is that the dynamic analog-single range in the ON state is greatly increased [21].

)(

1

ThnGSOXn

ON

VVL

WCU

R

(5)

)(

1

ThpSGOXP

OP

VVL

WCU

R

(6)

5. Results

5.1. Device Parameter

Table 2 gives the device parameter for a Two-stage operational amplifier.

Table 2. Device Parameter for Two-stage Amplifier

Device Type W(um) L(um)

M1,M2 P 10 0.378

M3,M4 N 18 0.556

M5,M8 P 35 0.460

M7 P 80 0.530

M6 N 15 0.100

M9 N 1 0.100

M10 P 10 0.100

M11 N 3 0.800

Figure 7. Transmission Gate (TG)

Q

M9

M10

Q

Page 7: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014)

Copyright ⓒ 2014 SERSC 61

5.2. Frequency Response, Compensation

Operational amplifier architectures are generally two types: single-stage externally

compensated or two stage internally compensated amplifiers, both of which can have a

dominant two pole type frequency response [21]. Ignoring higher order poles, the small signal

behavior of the two stage amplifier can be modeled by the general two poles, one zero small

signal equivalent circuit shown in Figure 8. Referring back to the model of the input stage

Figure 4 (Without (Cc) and (Rz)), We have two poles these are Eq (7) and Eq (8)

3

1

1 1

1

mgC

P (7)

)(

1

04022

2rrC

P (8)

The compensation of the two stage CMOS amplifier can be carried out using a pole

splitting capacitor. The goal of the compensation task is to achieve a phase margin greater

than 45 degree. The circuit can be approximately represented by the small signal equivalent

circuit of Figure 8 (without Rz), if the non-dominant poles which may exist on the circuit are

neglected.

Figure 8. Simplified Small Signal Model of the basic Two Stage Op-amp Added with the Nulling Resistor

The circuit displays two poles and a half right plane zero, which under an assumption that

the poles are widely separated can be shown to be approximately located at (Eq (9), Eq (10)

and Eq (11)).

122

11

1

RCRgP

Cm

(9)

CC

Cm

CCCCCC

CgP

2112

22

(10)

C

m

C

gZ 2 (11)

Note that the pole due to the capacitive loading of the first stage by the second, p1, has

been pushed down to a very low frequency by the miller effect in the second stage, while the

Page 8: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014)

62 Copyright ⓒ 2014 SERSC

pole due to the capacitance at the output node of the second stage p2, has been pushed to a

very high frequency due to the shunt feedback. For this reason, the compensation technique is

called pole splitting.

Physically, the zero arises because the compensation capacitor provides a path for the

signal to propagate directly through the circuit to the output at high frequencies. Since there is

no inversion in that signal path as there is in the inverting path dominant at low frequencies,

stability degraded.

An even simpler approach is to insert an active nulling resistor (Rz) in series with the

compensation capacitor as shown in Figure 4 and Figure 8. If Rz is assumed to be less than R1

or R2 and the poles are widely spaced, then the poles are: (Eq(12),Eq(13),Eq(14) and Eq(15)).

212122

1

1

)1(

1

RRCgRCRgP

CmCm

(12)

2

2

1221

22

C

g

CCCCCC

CgP m

CC

Cm

(13)

1

13

CRP

Z

(14)

)1

(

1

2 Zm

CRg

C

Z

(15)

The resistor RZ= (Ron||Rop) allows independent control over the placement of the zero.

The zero vanishes when RZ is made equal to 1/gm2. In fact, the resistor can be further

increased to move the zero into the left half-plane and place it on top of p2 to improve the

amplifier phase margin. RZ can be realized by a CMOS switch transistor in the triode region.

Using CMOS switch is to increase dynamic range. Where the resistance of CMOS switch

is plotted as a function of the input voltage figure 9, In this figure, the p-channel M10 and n-

channel M9 devices are sized in such a way that they have an equivalent resistance to

identical terminal conditions.

The peak behavior at midrange (near VDD/2), is due the parallel combination of the two

devices M9 and M10. In the right n-channel M9 dominating when E is low, and in the left p-

channel M10 dominating when E is high (near VDD).

Figure 10 shows a typical variation of the gain and phase versus frequency, with Cc =0.6

pF, Cp=1 pF.

N-Channel dominates P-Channel dominates

Figure 9. Rz of figure 7, as a Function of E

Page 9: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014)

Copyright ⓒ 2014 SERSC 63

Figure. 10 Shows a Typical Variation of the Gain and Phase versus Frequency

5.3. Common Mode Rejection Ratio CMRR

Due to the large amount of 60-Hz human in bio-potential measurements, common-mode

rejection is as important as low-noise operation. CMRR is defined as the ratio of the voltage

gain for a differential-mode input signal to the voltage gain for a common-mode input signal,

and can be stated as [22], Therefore, measuring the dc CMRR involves determining the

change in the offset voltage due to any change in the applied common-mode voltage. But,

power supply voltage variation may affect the output as well as the offset voltage. In order to

ensure that the measured change in the offset voltage is only due to a change in common-

mode input voltage, the output voltage of the Two-stage operational amplifier is maintained at

a fixed voltage for the complete test sequence.

We simulate the complete circuit in single power supply for CMRR. The results of CMRR

are up to 130dB as shown in Figure 11.

Figure 11. Show a Variation of the CMRR versus Frequency

5.4. Noise

The input stage dominates the noise of the op amp. PMOS devices offer lower flicker noise

which is advantageous in our op amp design. Also, the noise component of the load devices is

scaled by the ratio of their transconductance to that of the input pair devices. So, the

transconductance of the input differential pair needs to be larger than that of the load pair in

order to ensure low input-referred noise. There exists a trade-off between noise and the output

swing of the stage. Generally, the overdrive voltage of the current loads is minimized to

realize a wide output swing, but for a fixed current bias this increases the transconductance

(due to increased W) and thereby results in a larger input-referred noise.

Page 10: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014)

64 Copyright ⓒ 2014 SERSC

Figure 12. Simulated Input-refred Noise Spectrum

The simulation results of the input referred noise spectrum is depicted in figure 12, flicker

noise below 100Hz is clearly visible, it achives input referred noise density of

at 10Hz.

This a standard two-stage 11-transistor amplifier Figure 4 with miller compensation and

zero-nulling active resistor Rz=(Rn//Rp). All the devices operate in saturation region except

the transistors M9 and M10 operate in the linear region. The input differential stage draws

1uA , while the output stage current is 1.1uA. This results in a total current consumption of

2.1uA plus the current reference Iref=0.5uA , i.e., a 2.6 uW power consumption from single

1V supply.

Figure 13 shows the amplified ECG (Vout) signals, note the original ECG is only 1.5mV

in its peak-to-peak magnitude.

Figure 13. Simulated Results of the Amplified ECG (Vout) Signals

The overall performance summary of the complete circuit two-stage operational amplifier

for biomedical applications is given in Table 3.

138𝑛𝑉 𝐻𝑧

Page 11: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014)

Copyright ⓒ 2014 SERSC 65

Table 3. Simulation Results of Schematic Two-stage Amplifier

Process 90nm CMOS

Power supply 1V

Power dissipation 2.6uW

CMRR 131 dB

Bandwidth 15.3kHz

PSRR+ 70.7 dB

Ad 55.1 dB

Input offset voltage (Vos) 6.42um

PM 69.46

Input-referred-noise at 10 Hz

6. Conclusion

In this paper, we present a Two-stage operational amplifier topology for low voltage and

low power, ECG Monitoring System applications, This two-stage amplifier with Miller

compensation can be used in low power, low voltage High CMRR and PSRR applications

such Biomedical instrument and a small battery operated devices. The circuit has been

designed in a SPECTRE using 0.90um CMOS technology.

To reduce the noise of the amplifier, we used the P-channel input devices with N-channel

load, because it’s flicker noise is less than that the N-channel input devices with P-channel

load.

We have described an ECG amplifier with low input-referred noise, 131-dB CMRR, less

than 3uW of power consumption, and good cardiac signal fidelity.

Proposed two-stage operational amplifier with Miller compensation (Cc) and nulling active

resistor (Rz), is well suited to biomedical systems such as cardiac pacemaker,

electrocardiogram (ECG) where low-power consumption is of primary concern.

It seems that the input-referred-noise not minimized well, we can reducing flicker noise by

using techniques, such as auto-zero-technique, or chopper-stabilization technique.

References

[1] T. H. Ning, P. W. Cook, R. H. Dennard, C. M. Osburn, S. E. Schuster and H.N.Yu, “1um MOSFET VLSI

technology, Part IV-Hot-electron design constraints”, IEEE Trans.Electron.Devices, vol. 26,(1979), pp. 346-

353.

[2] M.F.Li, Uday Dasgupta, X. W. Zhang and Yong Ching Lim, “A low-Voltage CMOS OTA with Rail-to-Rail

Differential Input Range”, IEEE Transactions on circuits and systems, (2000) January, vol 47, pp.1-8.

[3] M. Swan, K. Arabi and B. Provost, “Implantable volume monitor and miniaturized simulator dedicated to

bladder control”, Artificial Organs, vol. 21, no. 3, (1997), pp. 219-222.

[4] R. Stein, B. Charles, D. Davis, L. Jhamandas, J. A. Mannard and T. R. Nichols, “Principles underlying new

methods for chronic neural recording”, Can. J. Of Neuro.SCI., (1975), pp. 235-244.

[5] D. Popovic, B. Stein, R. B. Jovanovic, K. L. Dai, A. R. Kostovand and W.W. Armstrong, “Sensory nerve

recording for closed-loop control to restore motor function”, IEEE Trans. Biomed.Eng, vol. 40, no. 10,

(1993), pp. 1024-1031.

[6] Y. Hu. and M. Sawan, “CMOS Front-end Amplifier Dedicated to Monitor Very low Amplitude Signal from

Imlantable Sensors”,Kluwer Academic Publishers, vol. 33, (2002), pp. 29-41.

[7] W.-S. Wang, Z.-C. Wu, H.-T. Huang and C.-H. Luo, “Low Power Instrumental Amplifier for Portable ECG”,

IEEE Circuits and Systems International Conference, (2009), pp. 1-4.

[8] C.-T. Ma, P.-I. Mak, M.-I. Vai, P. Mak, S.-H. Pun, W. Feng and R.P. Martins, “A 90nm CMOS Bio-Potential

Signal Readout Front-End with Improved Powerline Interference Rejection”, IEEECircuits and Systems,

(2009), pp. 665-668.

[9] C.-C. Huang, S.-H. Hung and J.-F. Chung, “Front-End Amplifier of low-Noise and Tunable B/W Gain for

Portable Biomedical Signal Acquidition”, IEEE Circuits and Systems, (2008), pp. 2717 – 2720.

𝟏𝟑𝟖𝐧𝐕 𝐇𝐳

Page 12: Design of CMOS Two-stage Operational Amplifier for ECG

International Journal of Bio-Science and Bio-Technology

Vol.6, No.5 (2014)

66 Copyright ⓒ 2014 SERSC

[10] D.Yates, E.Lopez-Morillo and R.G.Carvajal, “A Low-Voltage Low-Power Front-End for wearable EEG

Systems”, Conference of the IEEE EMBS, (2007).

[11] L.Lentola, A. Mozzi, A. Neviani, A. Baschirotto, “A 1uA Front-End for Pacemaker Atrial Sensing

Channels”,IEEE , (2009).

[12] C. Bronskowski and D. Schroeder, “An ultra low–noise CMOS Operational Amplifier with programmable

Noise-Power Trade-off ”, IEEESolid-State Circuits Conference, (2006), pp. 368 - 371

[13] J.G. Webster, “Medical instrumentation : Application and design”, 3rd ed, John Wiley & sons Inc,(1998).

[14] H.-C. Chow and B.-S. Tang, “A high Performance Current-mode Instrumentation Amplifier For Biomedical

applications”, 2nd International Conference on Signal Processing Systems, (2010).

[15] C. Ken and L. Xiaoying, “A Zigbee Based Mesh Network for ECG Monitoring System”, Bioinformatics and

Biomedical Engineering (iCBBE), (2010), pp.1-4.

[16] Y.F. Low, I. B. Mustaffa, N. B. Mohd Saad and A. H. Bin Hamidon, “Development of PC-Based ECG

Monitoring System”, IEEE Research and Development, 4th Student Conference , (2006), pp.66-69.

[17] PhysioBank, physiologic signal archives for biomedical research, http://www.physionet.org/physiobank/

[18] The MIT-BIH Arrhythmia Database, http:// physionet.org/physiobank/database/mitdb/

[19] The MIT-BIH Compression Test Database, http:// www.physionet.org/physiobank/database/cdb/

[20] R.L. Geiger, P.E. Allen and N.R. Strader, “VLSI Design Techniques for Analog and Digital Circuits”,

McGraw-Hill, (1990).

[21] P.E. Allen and D.R Holberg, “CMOS Analog Circuit Design”, Second Edition, Oxford University Press,

(2002).

[22] E. J. Kennedy, “Operational Amplifier Circuits, Theory and Applications”, Holt, Rinehart and Winston, Inc,

New York, (1988).