10
Design of a compact superstrate-loaded slotted implantable antenna for ISM band applications GURPRINCE SINGH * and JASWINDER KAUR Department of Electronics and Communication Engineering, Thapar Institute of Engineering and Technology, Patiala, Punjab, India e-mail: [email protected] MS received 14 January 2021; revised 17 April 2021; accepted 18 July 2021 Abstract. In this paper, the design of a simple rectangular slotted inset-cut fed implantable antenna is pro- posed. By using slotting technique in patch as well as ground plane, the size of the proposed antenna is reduced significantly to 10.2 9 8.61 9 1.92 mm 3 with very small patch size of 6.2 9 4.5 mm 2 suitable for implantable biotelemetry application around 2.45 GHz Industrial, Scientific and Medical (ISM) band. An appreciable fractional bandwidth of 13.7% with a good return loss of - 20.02 dB has been achieved. In-vitro analysis of the antenna prototype is carried out by inserting the prototype into skin mimicking liquid. Keywords. Biotelemetry application; implantable antenna; in-vitro; ISM band; slotted; superstrate-loaded. 1. Introduction Biotelemetry means transferring biological signals from human or animal body and remotely detecting them wired or wirelessly. This is used widely in transferring medical parameters from a patient to external receiving or moni- toring devices. For biotelemetry, two types of devices are used: on-body and body-implantable devices. Many people are dependent on implantable devices like blood glucose monitor [1], cochlear implant [2], retinal implant [3], Functional Electrical Stimulators (FES) [4] and pacemakers [5], etc. Generally, two main frequency bands are used for making these wireless transmission devices which contain antennas working at ISM [2.4–2.48 GHz] band and Medi- cal Implant Communications Service (MICS) [402–405 MHz] band [610]. In our proposed work, we are designing an implantable antenna keeping in mind the following challenges such as miniaturization, biocompati- bility and Specific Absorption Rate (SAR) limitation. As the antenna needs to be inserted inside skin therefore, it has to be as small as possible. For making a compact size antenna, many techniques are used such as Planar Inverted- F Antenna (PIFA) [11], shorting pin [12, 13], patch stack- ing [14, 15], meandered [6, 16], spiral [1720] shaped designs and use of high dielectric materials [2123]. Next challenge faced is that the antenna has to be biocompatible to the human body. As the antenna is inserted permanently inside body therefore it must not react with the nearby tissues causing serious problems to the human body. The reason of antenna reacting with nearby tissues is the metallic layer (patch and ground layer are made of copper) on it, and the human tissues are conductive in nature cre- ating short circuits [6]. For avoiding this problem, the antenna is either made or covered using a high dielectric material called superstrate which does not react with body tissues thus making the antenna biocompatible. Some common superstrate materials are Rogers RT/duroid 3010 (er = 10.2, tand = 0.0022), Rogers RT/duroid 3210 (er= 10.2, tand = 0.003), Rogers RT/duroid 6010 (er = 10.2, tand = 0.0035) and Alumina (er = 9.4, tand = 0.006), etc. Superstrate and substrate layer of antenna can be made from the same material [7, 9]. Another challenge in making an implantable antenna is maximum allowable limit of SAR which ensures patient’s safety. According to IEEE C95. 1-1999 standard the limit of maximum allowable input power given to antenna to protect human body should be kept lower than 1.6 W/kg for any 1-g averaged tissue of cubic shape [SAR 1g,max B 1.6 W/kg] and according to IEEE C95. 1-2005 for any 10-g averaged tissue of cubic shape it should be kept lower than 2 W/kg [SAR 10g,max- B 2 W/kg] [6, 2428]. The antenna is simulated using a three-dimensional simulation software CST MWS before fabrication so that the approximate results can be calcu- lated. For simulation, the antenna is inserted inside single layer or multi-layered [three-layer or five-layer] model using layers of skin, fat, muscle and bone. For testing purposes the antenna can be tested using in-vitro, ex vivo or in-vivo technique whichever possible. If testing of antenna is done inside the whole animal then it is termed as in-vivo. If testing is done by taking a tissue sample of an animal then it is called as ex-vivo testing. Whereas if the testing of *For correspondence Sådhanå (2021)46:164 Ó Indian Academy of Sciences https://doi.org/10.1007/s12046-021-01691-4

Design of a compact superstrate-loaded slotted implantable

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Design of a compact superstrate-loaded slotted implantable

Design of a compact superstrate-loaded slotted implantable antennafor ISM band applications

GURPRINCE SINGH* and JASWINDER KAUR

Department of Electronics and Communication Engineering, Thapar Institute of Engineering and Technology,

Patiala, Punjab, India

e-mail: [email protected]

MS received 14 January 2021; revised 17 April 2021; accepted 18 July 2021

Abstract. In this paper, the design of a simple rectangular slotted inset-cut fed implantable antenna is pro-

posed. By using slotting technique in patch as well as ground plane, the size of the proposed antenna is reduced

significantly to 10.2 9 8.61 9 1.92 mm3 with very small patch size of 6.2 9 4.5 mm2 suitable for

implantable biotelemetry application around 2.45 GHz Industrial, Scientific and Medical (ISM) band. An

appreciable fractional bandwidth of 13.7% with a good return loss of - 20.02 dB has been achieved. In-vitro

analysis of the antenna prototype is carried out by inserting the prototype into skin mimicking liquid.

Keywords. Biotelemetry application; implantable antenna; in-vitro; ISM band; slotted; superstrate-loaded.

1. Introduction

Biotelemetry means transferring biological signals from

human or animal body and remotely detecting them wired

or wirelessly. This is used widely in transferring medical

parameters from a patient to external receiving or moni-

toring devices. For biotelemetry, two types of devices are

used: on-body and body-implantable devices. Many people

are dependent on implantable devices like blood glucose

monitor [1], cochlear implant [2], retinal implant [3],

Functional Electrical Stimulators (FES) [4] and pacemakers

[5], etc.

Generally, two main frequency bands are used for

making these wireless transmission devices which contain

antennas working at ISM [2.4–2.48 GHz] band and Medi-

cal Implant Communications Service (MICS)

[402–405 MHz] band [6–10]. In our proposed work, we are

designing an implantable antenna keeping in mind the

following challenges such as miniaturization, biocompati-

bility and Specific Absorption Rate (SAR) limitation. As

the antenna needs to be inserted inside skin therefore, it has

to be as small as possible. For making a compact size

antenna, many techniques are used such as Planar Inverted-

F Antenna (PIFA) [11], shorting pin [12, 13], patch stack-

ing [14, 15], meandered [6, 16], spiral [17–20] shaped

designs and use of high dielectric materials [21–23]. Next

challenge faced is that the antenna has to be biocompatible

to the human body. As the antenna is inserted permanently

inside body therefore it must not react with the nearby

tissues causing serious problems to the human body. The

reason of antenna reacting with nearby tissues is the

metallic layer (patch and ground layer are made of copper)

on it, and the human tissues are conductive in nature cre-

ating short circuits [6]. For avoiding this problem, the

antenna is either made or covered using a high dielectric

material called superstrate which does not react with body

tissues thus making the antenna biocompatible. Some

common superstrate materials are Rogers RT/duroid 3010

(er = 10.2, tand = 0.0022), Rogers RT/duroid 3210 (er =10.2, tand = 0.003), Rogers RT/duroid 6010 (er = 10.2,

tand = 0.0035) and Alumina (er = 9.4, tand = 0.006), etc.

Superstrate and substrate layer of antenna can be made

from the same material [7, 9]. Another challenge in making

an implantable antenna is maximum allowable limit of

SAR which ensures patient’s safety. According to IEEE

C95. 1-1999 standard the limit of maximum allowable

input power given to antenna to protect human body should

be kept lower than 1.6 W/kg for any 1-g averaged tissue of

cubic shape [SAR1g,max B 1.6 W/kg] and according to

IEEE C95. 1-2005 for any 10-g averaged tissue of cubic

shape it should be kept lower than 2 W/kg [SAR10g,max-

B 2 W/kg] [6, 24–28]. The antenna is simulated using a

three-dimensional simulation software CST MWS before

fabrication so that the approximate results can be calcu-

lated. For simulation, the antenna is inserted inside single

layer or multi-layered [three-layer or five-layer] model

using layers of skin, fat, muscle and bone. For testing

purposes the antenna can be tested using in-vitro, ex vivo or

in-vivo technique whichever possible. If testing of antenna

is done inside the whole animal then it is termed as in-vivo.

If testing is done by taking a tissue sample of an animal

then it is called as ex-vivo testing. Whereas if the testing of*For correspondence

Sådhanå (2021) 46:164 � Indian Academy of Sciences

https://doi.org/10.1007/s12046-021-01691-4Sadhana(0123456789().,-volV)FT3](0123456789().,-volV)

Page 2: Design of a compact superstrate-loaded slotted implantable

antenna is done in body phantom that is made by preparing

homogenous [single layer tissue] or inhomogeneous [multi-

layer tissue] solutions of skin, muscle, bone, fat or their

combinations, then this type of testing is known as in-vitro

testing. In-vitro testing of implantable antennas made for

human body have to go through many tests before inserting

inside the body because insertion of any foreign element in

body can cause serious problems. Therefore, before

inserting anything inside the body, some scientific testing

must be done even after software simulations. The body

phantoms are made by the mixture of water, sugar, salt and

oil, etc. for obtaining the required electrical permittivity

and conductivity which matches the properties of human

body tissues. These solutions are made in cylindrical,

rectangular or square boxes depending upon the type of

implantable area.

Figure 1. Geometrical representation of the proposed antenna. (a) Radiating patch. (b) Ground plane. (c) Fabricated antenna without

superstrate layer. (d) Fabricated antenna with superstrate layer.

Table 1. Geometrical parameters of the proposed antenna in mm

Parameter Value Coordinates

Ls 8.61 (- 4.25,5.1) (4.36,5.1)

Ws 10.2 (- 4.25,- 5.1) (- 4.25,5.1)

PL 4.5 (- 2.25,3.1) (2.25,3.1)

Pw 6.2 (- 2.25,- 3.1) (- 2.25,3.1)

P1 2 (- 1,0) (1,0)

P2 3 (- 1.5,- 3.1) (1.5,- 3.1)

P3 4.86 (1.5,- 0.33) (4.36,- 0.33)

Lg 8.61 (- 4.25,5.1) (4.36,5.1)

Wg 10.2 (- 4.25,- 5.1) (- 4.25,5.1)

r1 1.5 (0,0) (0,1.5)

g2 7.7 (2.25,- 3.1) (2.25,4.6)

g3 4.1 (- 2.25,- 4.1) (- 2.25,0)

Lg1 2.25 (0,- 3.1) (2.25,- 3.1)

Lg2 2.25 (- 2.25,- 4.1) (0,- 4.1)

164 Page 2 of 10 Sådhanå (2021) 46:164

Page 3: Design of a compact superstrate-loaded slotted implantable

(Case 1) (Case 2)

(Case 3) (Case 4)

(a)

(b) (c) (d)

2.0 2.2 2.4 2.6 2.8 3.0-35

-30

-25

-20

-15

-10

-5

0

S 11 (

dB)

Frequency (GHz)

Case 1 Case 2 Case 3 Case 4

2.0 2.2 2.4 2.6 2.8 3.0

-40

-35

-30

-25

-20

-15

-10

-5

0

Ret

urn

loss

(dB)

Frequency (GHz)

r1=0.5 r1=1 r1=1.5 r1=2 r1=2.5 r1=3

2.0 2.2 2.4 2.6 2.8 3.0

-35

-30

-25

-20

-15

-10

-5

0

S 11(

dB)

Frequency (GHz)

g2=7.2 g2=7.45 g2=7.7 g2=7.95 g2=8.2

Figure 2. Evolution of the proposed antenna inside skin layer. (a) Evolution of the ground plane of proposed antenna. (b) Comparison

of return loss for three different configurations of ground plane. (c) Comparison of return loss of proposed antenna with different r1.

(d) Comparison of return loss of proposed antenna with different g2.

Sådhanå (2021) 46:164 Page 3 of 10 164

Page 4: Design of a compact superstrate-loaded slotted implantable

In this paper, a miniaturized implantable patch antenna is

proposed for medical applications like Intracranical Pres-

sure (ICP) monitoring when inserted inside scalp on the top

of brain [7]. Slotting technique is used in the radiating patch

and ground plane for reducing the size of antenna as well as

increasing the bandwidth. The overall size of the proposed

antenna is 10.2 9 8.61 9 1.92 mm3 which is considerably

small. At the same time, the size of patch is reduced to

6.2 9 4.5 mm2 which is among the smallest patch sizes

developed in the present literature. A high dielectric bio-

compatible material Rogers RT/duroid 3010 is used in

making and encasing the proposed antenna. With this small

size and covering the antenna with biocompatible layers on

both sides make the antenna novel as well as best contender

to be used in implantable devices. In-vitro testing is done

for this fabricated antenna using a single layer skin mim-

icking liquid. The proposed antenna works at ISM band and

produces appreciable results in terms of bandwidth, SAR,

volume factor and peak gain. The detailed antenna structure

with simulated and measured results are discussed in the

succeeding sections.

2. Antenna design

A high dielectric material Rogers 3010 (2r = 10.2 and

tand = 0.0022) is used both as a substrate as well as

superstrate for making the antenna compatible to human

body tissues having thickness 0.64 mm each thus making

an overall thickness of antenna as 1.92 mm. figure 1 shows

the geometrical representation of proposed antenna with

dimensions 10.2 9 8.61 mm2 having a small patch size of

6.2 9 4.5 mm2.

Figure 1a shows the radiating patch layer of antenna

having a circular slot at the centre of radius 1 mm and a

semi-circular slot at the bottom of the patch of radius

1.5 mm. An inset feed line of length 2.86 mm and width

0.66 mm is used for feeding the antenna. The advantage of

using an inset feed line over other feeding methods is that it

is simple to model as well as can easily be fabricated. It

also helps to achieve good impedance matching. Two

vertical rectangular slots are introduced in the ground plane

below the patch in which first one is of dimensions

4.1 9 2.25 mm2 and another one is of dimensions

7.7 9 2.25 mm2. Moreover, a circular slot of radius

1.5 mm is introduced at the centre of ground plane which

makes the overall defect to resemble as if it is a quadrant

shaped slot, reason behind is that it merges with the two

rectangular slots shown in figure 1b. The proposed antenna

with and without superstrate is shown in figure 1c and

figure 1d respectively. The detailed parameters of patch and

ground plane are listed in table 1.

2.1 Evolution of the ground plane of proposedantenna

Figure 2a shows evolution of the proposed antenna with

respect to ground. Slotting technique is applied in the

ground plane to obtain the desired result at ISM band of the

proposed antenna. The ground plane basically consists of

two rectangular slots having dimensions 7.7 9 2.25 mm2

and 6.4 9 2.25 mm2 with one circular slot of radius

1.5 mm which has been just introduced below the centre of

radiating patch. The resonant behaviour of these three slots

in the ground plane analysed in terms of return loss is

depicted in figure 2b as four cases: Case 1, Case 2, Case 3

and Case 4 and the results obtained are illustrated in

Table 2. In Case 1, antenna’s ground plane is shown before

applying any slotting technique. From figure 2b it is clear

that without slotting, the antenna does not resonate. In Case

2, the bigger rectangular slot is constructed for which the

resonant frequency obtained is 2.63 GHz covering a

bandwidth of 2.44–2.87 GHz with a return loss of

- 30.79 dB. As can be seen from the bandwidth spectrum

achieved, the desired ISM band (2.4–2.48 GHz) is not fully

covered. Therefore, this configuration of ground plane is

not good enough to satisfy our application. In Case 3, the

smaller rectangular slot is added to the ground plane in

Table 2. Performance characteristics of the ground plane of

antenna for three different cases.

S11 in

dB

Bandwidth

in GHz

Resonant

frequency (GHz)

Case 1 – – –

Case 2 - 30.79 2.44–2.87 2.63

Case 3 - 16.30 2.28–2.57 2.42

Case 4 - 20.02 2.31–2.65 2.475

Table 3. Performance characteristics of the proposed antenna

with different r1.

r1

(mm)

S11 in

dB

Bandwidth

(GHz)

Resonant frequency

(GHz)

0.5 - 16.66 2.28–2.59 2.43

1 - 17.67 2.3–2.6 2.45

1.5 - 20.02 2.31–2.65 2.475

2 - 25.47 2.35–2.69 2.52

2.5 - 37.25 2.4–2.75 2.56

3 - 20.27 2.48–2.84 2.65

Table 4. Parameters of antenna at different values of g2.

g2(mm)

S11 in

dB

Bandwidth

(GHz)

Resonant frequency

(GHz)

7.2 - 34.21 2.43–2.8 2.6

7.45 - 24.26 2.38–2.73 2.54

7.7 - 20.02 2.31–2.65 2.475

7.95 - 17.24 2.25–2.55 2.39

8.2 – – –

164 Page 4 of 10 Sådhanå (2021) 46:164

Page 5: Design of a compact superstrate-loaded slotted implantable

addition to the rectangular slot discussed in Case 2. Case 3

shows a return loss of - 16.30 dB at the resonant fre-

quency 2.42 GHz covering a bandwidth of 2.28–2.57 GHz.

To get better results in terms of return loss within ISM

band, Case 3 has been considered in which one more cir-

cular slot of radius 1.5 mm is added in addition to the two

rectangular slots of Case 3. Further, Case 4 depicts a return

(d)

2.0 2.2 2.4 2.6 2.8 3.0-35

-30

-25

-20

-15

-10

-5

0

Ret

urn

Loss

(dB)

Frequency (GHz)

skin brain three layer

(a) (b) (c)

Figure 3. Simulation of antenna. (a) Antenna inside skin layer using CST MWS software. (b) Antenna inside three layer using CST

MWS software. (c) Antenna inside brain layer using CST MWS software. (d) Comparison of simulated return loss characteristics of the

proposed antenna.

Table 5. Dielectric properties of human tissues at ISM band.

Human tissue

layer Permittivity

Conductivity (S/

m)

Density (kg/

m3)

skin 38.007 1.464 1020

fat 5.280 0.105 918

muscle 53.574 1.810 1040

Table 6. Performance of the proposed antenna in brain layer,

single skin layer and three layer phantom models.

S11 in

dB

Bandwidth in

GHz

Resonant frequency

(GHz)

Skin - 20.02 2.31–2.65 2.475

Brain - 30.61 2.35–2.68 2.51

Three

layer

- 17.04 2.24–2.56 2.38

Sådhanå (2021) 46:164 Page 5 of 10 164

Page 6: Design of a compact superstrate-loaded slotted implantable

loss of - 20.02 dB at the resonant frequency 2.475 GHz

covering a bandwidth of 2.31–2.65 GHz.

2.2 Parametric sweep with different r1 and g2

Figure 2c shows the change in return loss of antenna with

change in radius of circular slot in ground plane from

r1 = 0.5 to r1 = 3. The return loss of r1 = 1.5 is best suited

for the proposed antenna as its resonance frequency falls in

the desired bandwidth allocated to ISM band. Table 3

depicts the performance characteristics of the proposed

antenna with different values of radius of circular slot

embedded at the centre of ground plane. Moreover, fig-

ure 2d shows the change in return loss of antenna with

change in length g2 of bigger rectangular slot in ground

plane from g2 = 7.2 to g2 = 8.2. From table 4 it is clear that

return loss of g2 = 7.7 and its corresponding resonance

frequency is best suited for our desired ISM band.

3. Simulation set up of the proposed antenna

Figure 3 shows the simulation setup of the proposed

antenna. The proposed antenna has been simulated using

Computer Simulation Technology Microwave Studio

(CST MWS). Figure 3a shows the antenna inside single-

layer model of skin phantom with 4 mm skin layers at both

sides whereas, in figure 3b the antenna is simulated inside

three-layer model of skin, fat and muscle with height

4 mm, 4 mm and 8 mm respectively and in figure 3c the

antenna is simulated inside single-layer model of 4 mm

brain layers at both sides.

Figure 3d illustrates the simulated return loss charac-

teristic of the proposed antenna inside the brain layer,

single-skin layer and three-layer from which it is clearly

observed that in all the models, the antenna is resonating

at desired ISM band. There are minimal changes in the

return loss of these three models which shows that the

antenna is perfectly and evenly working in brain layer,

skin layer as well as in three layered phantoms. Table 5

shows the dielectric properties of these human tissues at

ISM band. Further, Table 6 shows the values of return

loss and bandwidth achieved by the proposed antenna

within brain layer, single skin layer and three-layer

phantom models.

3.1 Surface current distribution

Figure 4a shows the surface current distribution of the radi-

ating element and ground plane of the proposed compact

Figure 4. Surface current and SAR of the proposed antenna. (a) Surface current distribution on radiating element and ground plane.

(b) SAR value for 1-g averaged and 10-g averaged.

164 Page 6 of 10 Sådhanå (2021) 46:164

Page 7: Design of a compact superstrate-loaded slotted implantable

superstrate loaded slotted implantable antenna. This gives an

idea about those parts of antenna that are mostly responsible

for radiation. It can be observed that the current intensity is

maximumacross the feedlinewhich is attached to the radiating

patch and across a semi-circular slot embedded at the bottom

corner of the patch. Two rectangular slots and one circular slot

are etched from the ground plane to attain the property of the

defected ground structure. The current density is increased

near the edges of slots,whichmeans that the inserted slots play

a vital role to obtain the desired ISMbandwithout affecting the

bandwidth.

3.2 Specific absorption rate (SAR)

The impact of the proposed compact superstrate loaded

slotted implantable antenna on human tissue, described by

the SAR was also examined. SAR value depends on the

geometry of the human tissue, antenna positioning,

dielectric properties of tissue, transmitting input power and

spacing between the antenna and the human tissue. The

basic formula of SAR is given in equation (1).

SAR W/kgð Þ ¼ r� E2

qð1Þ

where r is the conductivity of tissue (S/m), E is root mean

square value of the electric field (V/m) induced in the tissue

and q is the tissue density (kg/m2). The American standard

for SAR analysis is used in this i.e. the SAR value averaged

over 1 g of tissue should be less than 1.6 W/kg. For vali-

dating the results of the proposed antenna, SAR (1-g

averaged and 10-g averaged) are calculated at the res-

onating frequency. For checking that if the antenna is

biocompatible inside the human body or not, the SAR

values for 1 W are shown in figure 4b. The SAR value for

1 g-averaged cubic tissue obtained for supply power of

1 W is 900 W/kg means a supply of less than 1.78 mW

should be applied to obtain a value of SAR below 1.6 W/kg

limit. Similarly, the SAR value for 10 g-averaged cubic

tissue obtained for supply power of 1 W is 82.1 W/kg

means a supply of less than 24.4 mW should be applied to

obtain a value of SAR below 2 W/kg limit.

Figure 5 shows the impedance versus frequency plot of

the proposed antenna within the range 2–3 GHz. A marker

Figure 5. Impedance vs frequency plot of proposed antenna. (a) Real part. (b) Imaginary part.

Sådhanå (2021) 46:164 Page 7 of 10 164

Page 8: Design of a compact superstrate-loaded slotted implantable

is shown to depict the resonating frequency of the proposed

antenna satisfying ISM band. The real part of impedance is

49.33 ohm as shown in figure 5a whereas the imaginary

part of impedance is - 6.02 ohms at the resonating fre-

quency as shown in figure 5b. These results clearly justify

the radiated behaviour of antenna with perfect impedance

matching of approximately 50 ohms and negligible reactive

power, thus making the proposed antenna suitable for ISM

band applications.

Figure 6a depicts the simulated gain vs frequency plot of

antenna in single skin layer. A peak gain of - 17.08 dBi at

2.475 GHz is reported. Figure 6b and figure 6c illustrate the

simulated E-plane and H-plane polar plots in relation to co-

polarization and cross-polarization at phi = 90� and phi =

0� of the proposed antenna, respectively. The E-plane

pattern is bidirectional whereas H-plane pattern is omni-

directional in nature.

Figure 7 shows the testing of antenna using Vector

Network Analyzer (VNA) and comparison of simulated

and measured return loss of the proposed antenna. The

return loss of the fabricated antenna is measured using

Agilent E5071C vector network analyser available in

Antenna Research Laboratory, Department of Electronics

and Communication Engineering, Thapar Institute of

Engineering and Technology, Patiala as shown in fig-

ure 7a. To test the antenna, a homogenous phantom of skin

mimicking liquid (made using 50% water and 50% sugar)

is made with approximately the same electrical properties

as that of skin whose recipe is illustrated in [8]. This

phantom is inserted inside a container of size

150 9 100 9 50 mm3 and filled such that the antenna can

be easily dipped inside the liquid phantom. Figure 7b

shows the comparison of simulated and measured return

loss of the proposed antenna inside a single layer skin

phantom. Figure 7b clearly depicts a close agreement

between the simulated and measured results of the pro-

posed antenna at ISM band inside skin except for some

small differences in return loss and bandwidth as illus-

trated in Table 7. Table 8 illustrates the comparison of

performance characteristics of the proposed antenna with

recently reported literature. Bandwidth, peak gain, SAR

(1 g-averaged and 10 g-averaged) and volume factor are

some of the performance measures used for comparing the

proposed antenna with previous literature. Volume factor

of antenna is the ratio of bandwidth (kHz) to the area of

antenna (mm3) as depicted in equation (2) [21]. Compar-

ing this parameter with the previously reported literature it

is clear that the proposed antenna is much better.

Volume factor VFð Þ ¼Bandwidth in kHzð Þ=Antenna area mm3

� � ð2Þ

4. Conclusion

In this proposed work, a compact implantable antenna

of patch size 6.2 9 4.5 mm2 and peak gain

- 17.08 dBi is made for biotelemetry applications at

(b)

(c)

2.0 2.2 2.4 2.6 2.8 3.0-20

-15

-10

-5

0

Gai

n (d

Bi)

Frequency (GHz)

Gain

030

60

90

120

150180

210

240

270

300

330

cross polar co polar

030

60

90

120

150180

210

240

270

300

330

cross polar copolar

(a)

Figure 6. Radiation pattern of proposed antenna. (a) Simulated

gain vs frequency pattern of the proposed antenna. (b) Simulated

co-polarization and cross-polarization at E-plane for phi = 90�.(c) Simulated co-polarization and cross-polarization at H-plane for

phi = 0�.

164 Page 8 of 10 Sådhanå (2021) 46:164

Page 9: Design of a compact superstrate-loaded slotted implantable

ISM band. Slotting strategy is utilized both in patch as

well as ground plane because of which an appreciable

bandwidth of 13.7% is accomplished. The measured

results obtained from the in-vitro testing of fabricated

antenna done inside skin phantom are in close agree-

ment to the simulated results with a little yet satis-

factory variation. SAR value for 1-g and 10-g averaged

cubic tissue is considered for the security of patient

from unfavourable impact of radio frequency radiations

on human body. Table 8 gives a point to point corre-

lation of this work with the ongoing literature in terms

of performance measures such as SAR, peak gain,

bandwidth and volume factor which shows that the

presented antenna is quite adequate for

implantable devices and circuits.

(a) (b)

2 2.2 2.4 2.6 2.8 3

-30

-25

-20

-15

-10

-5

0

5

Ret

urn

loss

(dB)

Frequency (GHz)

Simulated Measured

Figure 7. Testing of antenna and comparison of simulated and measured return loss. (a) Fabricated antenna testing inside skin

mimicking liquid. (b) Comparison of simulated and measured return loss of the proposed antenna in skin.

Table 7. Simulated and measured return loss and bandwith of the

proposed antenna.

S11 in dB Bandwidth in GHz

Skin (simulated) - 20.02 2.31–2.65

Skin (measured) - 29.28 2.22–2.61

Table 8. Comparison of performance of proposed antenna with recent literature.

Refs.

Dimensions

(mm3)

Bandwidth

(|S11| B - 10 dB) at

ISM band

Peak

gain

(dBi)

SAR (1 g-average)

at 1 W power

SAR (10 g-average)

at 1 W power

Volume

factor

Superstrate

layers

[8] 27.5 9 21 9 1.6 12.57% - 27.46 – – 335 Nil

[9] 10 9 10 9 1.27 10.84% - 21 710.65 84.6 826 Single

[10] 16.5 9 16.5 9 2.54 4.4% - 9 292 – 159 Single

[23] 10 9 15 9 2 – - 16.3 – – – Both sides

[27] 22 9 16 9 1.27 1.6% - 19.5 2.15 9 10-3 – 89.5 Nil

This

work

10.2 9 8.61 9 1.92 13.7% - 17.08 900 82.1 2016 Both sides

Sådhanå (2021) 46:164 Page 9 of 10 164

Page 10: Design of a compact superstrate-loaded slotted implantable

References

[1] Shults M C, Rhodes R K, Updike S J, Gilligan BJ and

Reining WN 1994 A telemetry-instrumentation system for

monitoring multiple subcutaneously implanted glucose sen-

sors. IEEE Transactions on Biomedical Engineering 41:

937–942

[2] Buchegger T, Obberger G, Reisenzahn A, Hochmair E,

Stelzer A and Springer A 2005 Ultra-wideband transceivers

for cochlear implants. EURASIP Journal on Applied SignalProcessing 2005: 3069–3075

[3] Gosalia K, Lazzi G and Humayun M 2005 Investigation of a

microwave data telemetry link for a retinal prosthesis. IEEETransactions on Microwave Theory and Techniques 52:

1925–1933

[4] Guillory K S and Normann R A 1999 A 100-channel system

for real time detection and storage of extracellular spike

waveforms. Journal of Neuroscience Methods 91: 21–29[5] Merli F, Bolomey L, Zurcher J F, Corradini G, Meurville E

and Skrivervi A K 2011 Design, realization and measure-

ments of a miniature antenna for implantable wireless

communication systems. IEEE Transactions on Antennasand Propagation 59: 3544–3555

[6] Kiourti A and Nikita K S 2012 2012 A review of

implantable patch antennas for biomedical telemetry: chal-

lenges and solutions [wireless corner]. IEEE Antennas andPropagation Magazine 54: 210–228

[7] Cho Y and Yoo H 2016 Miniaturised dual-band

implantable antenna for wireless biotelemetry. ElectronicsLetters 52: 1005–1007

[8] Sukhija S and Sarin R K 2017 Low-profile patch antennas for

biomedical and wireless applications. Journal of Computa-tional Electronics 16: 354–368

[9] Ahlawat S, Srivastava G and Kumar G 2019 Design of skin

implantable radiator with capacitive and CSRR loadings for

ISM band applications. International Journal of InformationTechnology 12: 1–10

[10] Liu C, Guo Y X and Xiao S 2012 Compact dual-band

antenna for implantable devices. IEEE Antennas and Wire-less Propagation Letters 11: 1508–1511

[11] Liu W C, Chen S H and Wu C M 2008 Implantable broad-

band circular stacked PIFA antenna for biotelemetry com-

munication. Journal of Electromagnetic Waves andApplications 22: 1791–1800

[12] Soontornpipit P, Furse C M and Chung Y C 2004 Design of

implantable microstrip antenna for communication with

medical implants. IEEE Trans Microwave Theory Techn.52: 1944–1951

[13] Lee S, Seo W, Ito K and Choi J 2011 Design of an implanted

compact antenna for an artificial cardiac pacemaker system.

IEICE Electron Express 8: 2112–2117[14] Liu W, Yeh F and Ghavami M 2008 Miniaturized

implantable broadband antenna for biotelemetry communi-

cation. Microw. Opt. Technol Lett. 50: 2407–2409

[15] Lee C M, Yo T C and Luo C H 2007 Compact broadband

stacked implantable antenna for biotelemetry with medical

devices. Electronics Lett. 43: 660–662[16] Karacolak T, Cooper R and Topsakal E 2009 Electrical

properties of rat skin and design of implantable antennas for

medical wireless telemetry. IEEE Trans Antennas Propag.57: 2806–2812

[17] Basir A, Bouazizi A and Zada M et al 2018 A dual-band

implantable antenna with wide-band characteristics at MICS

and ISM bands. Microw. Opt. Technol. Lett. 60: 2944–2949[18] Kim J and Rahmat-Samii Y 2014 Implantable antennas

inside a human body: simulations, designs, and characteri-

zations. IEEE Trans Microwave Theory Techn. 52:

1934–1943

[19] Bakogianni S and Koulouridis S 2015 An implantable pla-

nar dipole antenna for wireless MedRadio band bioteleme-

try devices. IEEE Antennas Wireless Propag. Lett. 15:

234–237

[20] Lee J H and Seo D W 2019 Compact and tissue-insensitive

implantable antenna on magneto-dielectric substrate for

wireless biotelemetry. Journal of Electromagnetic Wavesand Applications 33: 2339–2461

[21] Ikonen P M T, Rozanov K N, Osipov A V, Alitalo P and

Tretyakov S A 2006 Magnetodielectric substrates in antenna

miniaturization: potential and limitations. IEEE TransAntennas Propag. 54: 3391–3399

[22] Abdelrahman E M, Ali H M and Mohammad S S 2017

Superstrate loaded miniaturized patch for biomedical teleme-

try. OMicrow Opt. Technol. Lett. 59: 1212–1218[23] Aslam B, Khan U H, Azam M A, Amin Y, Loo J and

Tenhunen H A 2017 A compact implantable RFID tag

antenna dedicated to wireless health care. InternationalJournal of RF and Microwave Computer-Aided Engineering27: e21094

[24] Kovar S, Spano I, Gatto G, Valouch J and Adamek M 2017

SAR evaluation of wireless antenna on implanted cardiac

pacemaker. Journal of Electromagnetic Waves andApplica-tions 31: 627–635

[25] Zhang Y, Liu C, Zhang K, Cao H and Liu X 2019 Design

and in-vivo testing of a low-cost miniaturized capsule

system for body temperature monitoring. InternationalJournal of RF and Microwave Computer-Aided Engi-neering 29: e21793

[26] Zhang H, Li L, Liu C, Guo Y X and Wu S 2017 Miniaturized

implantable antenna integrated with split resonate rings for

wireless power transfer and data telemetry. Microwave andOptical Technology Letters 59: 710–714

[27] Yeap K, Voon C, Hiraguri T and Nisar H 2019 A compact

dual-band implantable antenna for medical telemetry. Mi-crowave and Optical Technology Letters 61: 2105–2109

[28] Das S and Mitra D 2020 Design of a compact circular

polarized implantable ring slot antenna for biomedical

applications. Electromagnetics 40: 1–10

164 Page 10 of 10 Sådhanå (2021) 46:164