S O U T H W E S T R E S E A R C H I N S T I T U T E
Department of Quality Assurance Calibration Laboratory
CERTIFICATE OF CALIBRATION
Device No: 2028 Issued to: DIV20 B57 NARASI SRIDHAR
Manufacturer: METTLER Model: UMT2
Nomenclature: ELECTRONIC BALANCE
Serial Number: 211001-142 SwRI No: NONE Cal interval 6 Mo.
Remarks
Accuracy: MFGR SPECS Procedure: SWRI
ENVIRONMENT Temperature: 0 Humidity: 0 Location: SWRI DIV20 B57
CONCLUSION Tolerance/Remarks: Received into the system, introduced or reactivated
LIMITED CALIBRATION ; to +/- -030 milli grams. CALIBRATED TO +/- 30 micro grams.
Calibration was in accord with requirements of MIL-STD-45662A. Measurements are traceable to the National Institute of Standards and Technology. Inspection and test data are on
Signed
Calibr
Cal interval: 6 Months
Record Number: 00010556 Next Calibration Due: 10/29/93
S O U T H W E S T R E S E A R C H I N S T I T U T E
Department of Quality Assurance Calibration Laboratory
Device Serial No: 211001-142
STANDARDS --------- Standard No: 1704 Manufacturer: RICE LAKE Nomenclature: WEIGHT STANDARD Serial No: C859 Cal.Due: 07/08/93
Standard No: 1706 Manufacturer: RICE LAKE Nomenclature: WEIGHT STANDARD Serial No: C861 Cal.Due: 07/08/93
Standard No: 1720 Manufacturer: RICE LAKE Nomenclature: WEIGHT STANDARDS Serial No: C858 Cal.Due: 07/08/93
Calibration Date: 04/29/93
Model: 1 GRAM
Cal.Rec.No: 00000000
Model: 2 GRAM
Cal.Rec.No: 00000000
Model: 1 - 500 MG
Cal.Rec.No: 00000000
S O U T H W E S T F l E S E A R C H I N S T I T U T E
Department of Quality Assurance Calibration Laboratory
CERTIFICATE OF CALIBRATION
Issued to: DIV20 B57 NARASI SRIDHAR
Manufacturer: METTLER Model: UMT2
Nomenclature: ELECTRONIC BALANCE
Serial Number: 211001-142 SwRI No: NONE
Remarks
Accuracy: MFGR SPECS
Device No: 2028
Cal interval 6 Mo.
Procedure: SWRI
ENVIRONMENT Temperature: 70 Humidity: 40 Location: SWRI DIV20 BLDG. 57
CONCLUSION Tolerance/Remarks: Received in tolerance, no adjustments made
LIMITED CALIBRATION- +/- .030mg
Calibration was in accord with requirements of MIL-STD-45662A. Measurements are traceable to the Nation Inspection and test data are on fi available f o r inspection,
Cal interval: 6 Months
Record Number: 00012628 Next Calibration Due: 04/22/94
S O U T H W E S T It E S E A R C H I N S T I T U T E
Department of Quality Assurance Calibration Laboratory
Device Serial No: 211001-142 Calibration Date: 10/22/93
STANDARDS ---------
Standard No: 1704 Manufacturer: RICE LAKE Model: 1 GRAM Nomenclature: WEIGHT STANDARD Serial No: C859 Cal.Due: 06/30/94 Cal.Rec.No: 00011756
Standard No: 1705 Manufacturer: RICE LAKE Model: 2 GRAM Nomenclature: WEIGHT STANDARD Serial No: C860 Cal.Due: 06/30/94 Cal.Rec.No: 00011757
Standard No: 1720 Manufacturer: RICE LAKE Model: 1 - 500 MG Nomenclature: WEIGHT STANDARDS Serial No: C858 Cal.Due: 06/30/94 Cal.Rec.No: 00011772
I BALANCE CAUBRATION VERIFICATION FORM
MFGR: &?k772dk/W MODEL:
SER. NO.Z//@ RANGE: <, CALIBRATION DATES /
I TEMPERATURE: 96 HUMIDITY: yb
I BALANCE TOLERANCES
// BALANCE CAL #: LINE. TOL.: REP. TOL.: I
I DATA PTS.
RUN #1
RUN #2
RUN #3
I MEAN
I STD. DEV.
n A h i n p 8 trnirin A T I ~ L I I ~ N Y U C v c n i r i b n I IUIY
POSITION GUIDE RANGE #l RANGE #2 RANGE #3
CALIBRATION POINTS I CALIBRATION POINTS I CALIBRATION POINTS I rn 1 1 2
COMMENTS: l / M f l k p SHIFT VERIFICATION
,03&5pZacrr SELF CALIBRATION Y/N INTERNAL: 1 EXTERNAL -
WEIGHT PAN POSITION VALUE
/
1 2 3 i SIGNATURE: - -- SWRI FORM QA-153-0
S O U T H W E S T F l E S E A R C H I N S T I T U T E
Department of Quality Assurance Calibration Laboratory
CERTIFICATE OF CALIBRATION
Device No: 2028 Issued to: DIV20 B57 NARASI SRIDHAR
Manufacturer: METTLER Model: UMT2
Nomenclature: ELECTRONIC BALANCES
Serial Number: 211001-142 SwRI No: NONE Cal interval 6 Mo.
Remarks
Accuracy: MFGR SPECS Procedure: SWRI
Temperature: ENVIRONMENT
Humidity : Location: SWRI DIV20 BLDG. 57
CONCLUSION Tolerance/Remarks: Received in tolerance, no adjustments made
Limited Calibration- +/-0.030mg
Calibration was in accord with requirements of MIL-STD-45662A. Measurements are traceable to the National Institute of Standards and Technology. Inspection and test data are on file available for inspection.
Cal interval: 6 Months
Record Number: 00013952 Next Calibration Due: 10/26/94
S O U T H W E S T R E S E A R C H I N S T I T U T E
Department of Quality Assurance Calibration Laboratory
Device Serial No: 211001-142 Calibration Date: 04/26/94
STANDARDS --------- Standard No: 1704 Manufacturer: RICE LAKE Model: 1 GRAM Nomenclature: WEIGHT STANDARD Serial No: C859 Cal.Due: 06/30/94 Cal.Rec.No: 00011756
Standard No: 1706 Manufacturer: RICE LAKE Model: 2 GRAM Nomenclature: WEIGHT STANDARD Serial No: C861 Cal.Due: 06/30/94 Cal.Rec.No: 00011758
Standard No: 1720 Manufacturer: RICE LAKE Model: 1 - 500 MG Nomenclature: WEIGHT STANDARDS Serial No: C858 Cal.Due: 06/30/94 Cal.Rec.No: 00011772
I BALANCE CALIBRATION VERIFICATION FORM
&L/TLt/e MODEL: &m/cz MFGR:
SER. NO. 2 // 00 @/qZ RANGE: 9 CALIBRATION DATES - 6
HUM I DlTY: I TEMPERATURE:
I CALIBRATED BY: /o b !& BALANCE TOLERANCES d/e ,~d>
BALANCE CAL #: /3J5-z I RANGE VERIFICATION
I RANGE #1 I RANGE #2 I RANGE #3
I CALIBRATION POINTS 1 CALIBRATION POINTS I CALIBRATION POINTS
MEAN
STD. DEV.
I I I I I I I I.
POSITION GUIDE
1 COMMENTS: SHIFT VERIFICATION
SELF CALIBRATION Y/N INTERNAL: EXTERNAL
WEIGHT PAN POSITION
1 2 3 4 VALUE
SWRI FORM QA-153-0
SOUTHWEST RESEARCH INSTITUTE
Department of Quality Assurance Calibration Laboratory
CERTIFICATE OF CALIBRATION ll/Ol/94
Issued to: NARASI SRIDHAR DIW2O ,B57 Manufacturer: METTL Nomenclature: ELECTRONIC BALANCE Serial Number: 211001-142
Asset Number: 002028 Model Number: UMT2
SwRI Capital Number: NONE
ENVIRONMENTAL CONDITIONS
Temperature: 69.OF Relative Humidity: 51 %
CALIBRATION INFORMATION
Location: B57 Procedure Number: SWRI Remarks : LIMITED TO +/-0.030 MIL GRAMS.
Technician: 7213 Accuracy: MFGR SPECS Received IN Tolerance
Calibration was in accordance with requirements of MIL-STD-45662A. Measurements are traceable to the National Institute of Standards and Tech- nology. Inspection and test data are on file and available for inspection.
STANDARDS USED FOR CERTIFICATION
001720 C858 R I C E 1 - 500 BIG WEIGHT STANDARDS
001704 C859 R I C E 1 GRAM WEIGHT STANDARD
001705 C 8 6 0 R I C E 2 GRAM WEIGHT STANDARD
07/20/94 12 07/20/95 07/20/94 12 07/20/95 07/20/94 12 07/20/95
Work Order: 15529 Chibration Date: 10/27/94 Interval: 6 months Next Calibration Due: 04/27/95
37 r TEMPERATURE: 69 HUM I DITY: I
t RANGE .vERiF[cATioN I
0 -c. .430-9n& I COMMENTS: // mw2-P 7 SHIFT VERIFICATION
SELF CALIBRATION Y/N INTERNAL: EXTERNAL
WEIGHT PAN POSITION VALUE
~
SWRI FORM QA-153-0
4
SOUTHWEST RESEARCH INSTITUTE
Department of Quality Assurance Calibration Laboratory
CERTIFICATE OF CALIBRATION os/os/9s
Issued to : DARRELL DUNN D IV2 0 I B57 Manufacturer: METTL Nomenclature: ELECTRONIC BALANCE Serial Number: 211001-142
Asset Number: 002028 Model Number: UMT2
SwRI/Div. I.D. #: NONE
ENVIRONMENTAL CONDITIONS
Temperature: 70.OF Relative Humidity: 56%
CALI BRAT I ON INFORMATION
Procedure Number: SWRI Remarks : LIMITED TO +/-O0O3C) MIL GRAMS
Accuracy: MFGR SPECS Received IN Tolerance
Calibration was in accordance with requirements of MIL-STD-45662A. Measurements are traceable to the National Institute of Standards and Tech- nology. Inspection and test data are on file and available for inspection.
STANDAFLDS USED FOR CERTIFICATION
Asset # Serial # Mfg Model # Nomenc 1 a ture Cal Date Int. Cal Due
001720 C858 RICE 1 - 500 MG WEIGHT STANDARDS 07/20/94 12 07/20/95 001704 C859 R I C E 1 GRAM WEIGHT STANDARD 07/20/94 12 07/20/95 001705 C860 R I C E 2 GRAM WEIGHT STANDARD 07/20/94 12 07/20/95
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
A Certified by :
/
Certificate#: 17215 I / Calibration Date: 05/05/95
Interval: 6 months Next Calibration Due: 11/05/95
-- - -~
M C E CAU8RATION VERIFICATION FORM
SER. NO. z / /m - /Yz RANGE: z c CALIBRATION DATES /
TEMPERATURE: 7 d b HUMIDITY:
1
BALANCE CAL #:
BALANCE TOLERANCES C/.S ,TG LINE. TOL.: REP. TOL.:
RANGE VERIFICATION
1 I I RANGE #1 RANGE #2 RANGE #3 POSITION GUIDE
- - ~ -~ ~
I
COMMENTS: d24,7&B /'a ,030 a y SHIFT VERIFICATION
SELF CALIBRATION Y/N INTERNAL EXTERNAL - WEIGHT
4
A
SWRI FORM QA-153-0
SOUTHWEST RESEARCH INSTITUTE .
Department of Quality ASSUranc8 Calibration Laboratory
CER!PIFICATE OF CALIBRATION 11/03/95
Issued to: DARRELL DUNN D IV2 0 I B57 Manufacturer/Model: METTL/UMT2 Nomenclature: ELECTRONIC BALANCE Serial Number: 211001-142 Asset Number: 002028 Notes:
ENVIRONMENTAL CONDITIONS
Temperature: 75.OF Relative Humidity: 66%
CALIBRATION INFORMATION
Procedure Number: WI-9-30-WT01 Remarks : LIMITED TO +/-0.03 MGRAMS
Accuracy: MFGR SPECS Received IN Tolerance
Calibration was in accordance with requirements of MIL-STD-45662A. Measurements are traceable to the National Institute of Standards and Tech- nology. Inspection and test: data are on file and available for inspection.
STANDARDS USED FOR CERTIFICATION
Certif icateX: 19168 Calibration Date : 11/02 /95 Interval: 6 months Next Calibration Due: 05/02/96
Southwest Research Institute 6220 Culebra Road
San Antonio, TX 78238 IDepartment of Quality Assurance Calibration Laboratory
Total Pages Printed: 1
Certificate of Calibration 3 May 1996
Issued to: DARRELL DUNN DIV20 B57 Man ufact u rer/M ode I : MElrTL UMT2 Description: ELECTRONIC BALANCE Serial Number: Asset Number: 002028
21 1001 -1 42
[E n v i ro n menta I C o n d it i o n s I Tem pe rat u re : 78.0 Humidity: 59%
[Calibration Information I Calibration was in accordance with requirements of MIL-STD-45662A and ANSVNCSL 2540-1 -1 994. Measurements are traceable to tlhe National Institute of Standards and Technology (NIST). This report may not be reproduced except in full without written approval of the originator. Inspection and test data are on file and available for inspection.
Calibration Date: 2 May 96 Calibration Procedure: WI-9-30-WTO 1
Interval: 6 months Accuracy: MFGR SPECS
Next Calibration Due: 2 Nov 96 Received: In Tolerance
Remarks: LIMITED TO +\- 0.03 MILLIGRAMS
Certificate ## 21114
Southwest Research Institute 6220 Culebra Road
San Antonio, TX 78238 Department of Quality Assurance Calibration Laboratory
Certificate of Calibration 11 November 1996
Issued to: DARRELL DUNN DIV20 B57 Manufacturer/Model: METTL UMT2 Description: ELECTRONIC BALANCE Serial Number: 211001-142 Asset Number: 002028
c ~~~ rr- nvironmental ~~ - Conditions
Temperature: 71 .O Deg. F Humidity: 55%
ICalibration Information LI
Calibration was in accordance with requirements of MIL-STD-45662A and ANSVNCSL 2540-1 -1 994. Measurements are traceable to the National Institute of Standards and Technology (NIST). This report may not be reproduced except in full without written approval of the originator. Inspection and test data are on file and available for inspection.
Calibration Date: 1 Nov 96 Cali brat ion Proced u re : CLC P-WB-00 1
Interval: 6 months Accuracy: MFGR SPECS
Next Calibration Due: 1 May 97 Received: In Tolerance
Remarks: LIMITED TO +/-0.03 MILLIGRAMS
Certificate # 23192
/ ’ LAST PAGE OF REPORT Total Pages Printed: 1
Southwest Research Institute 6220 Culebra Road
San Antonio, TX 78238 Department of Quality Assurance Calibration Laboratory
1
Total Pages Printed: 1
Certificate of Calibration 13 May 1997
Issued to: DARRELL DUNN DIV20 857 Manufacturer/Model: ME'TTL UMT2 Description: ELECTRONIC BALANCE Serial Number: Asset Number: 002028
21 1001 -1 4P!
I Environmental Conditions C
Temperature: 80.0 Deg. F Humidity: 57%
ICalibration Information C
Calibration was in accordance wiith requirements of MIL-STD-45662A and ANSVNCSL 2540-1 -1 994. Measurements are traceable to the National Institute of Standards and Technology (NIST). This report may not be reproduced except in full without written approval of the originator. Inspection and test data are on file and available for inspection.
Calibration Date: 1 May 97 Cali brat ion Procedure: C LC P-WB-00 1
Interval: 6 months Accuracy: MFGR SPECS
Next Calibration Due: 1 Nov 97 Received: In Tolerance
Remarks: MEASURED UNCElRTAINTY OF BALANCE IS +/- 0.034MG AT 80 DEG. F.
Standards Used I Asset MFR Model Description Serial No. Due Cal 001704 RICE 1 GRAM WEIGHT STANDARD C859 19 Jul97 001706 RICE 2 GRAM WEIGHT STANDARD C86 1 19 Jul97 001720 RICE 1 - 500 MG WEIGHT STANDARDS C858 19 Jul97
Certificate # 25285
Southwest Research Institute 6220 Culebra Road
San Antonio, TX 78238 Department of Quality Assurance Calibration Laboratory
Certnficate of Calibration 4 November 1997
Issued to: DARRELL DUNN DIV20 857 Manufacturer/Model: METTLER UMT2 Description: ELECTRONIC BALANCE Serial Number: Asset Number: 002028
21 1001 -1 42
!Environmental Conditions b ~ ~
Temperature: 72.00 Deg. F Humidity: 33 % RH
1 Cali brat ion Information c Calibration was in accordance with requirements of MIL-STD-45662A and ANSVNCSL 2540-1 -1 994. Measurements are traceable to the National Institute of Standards and Technology (NIST). This report may not be reproduced except in full without written approval of the originator. Inspection and test data are on file and available for inspection.
The uncertainty of the calibration was sufficient to determine that the instrument met the manufacturer% specifications.
Calibration Date: 3 Nov 97 Cali bration Procedure: CLCP-WT-001
Interval: 6 months
Next Calibration Due: 3 Nlay 98 Received: In Tolerance
Remarks:
!Standards Used -LI
Asset MFR Model Description
001704 RICELAKE 1G WEIGHT STANDARD 001705 RICELAKE 2G WEIGHT STANDARD 001 720 RICE LAKE 1 MG-500 MG WEIGHT SET
Serial No. Due Cal
C859 C860 C858
n
21 Jul98 21 Jut98 21 Jul98
LAST PAGE OF REPORT Total Pages Printed: 1
Certificate # 27411
Soil t li wes t Resea 1-c h I rist i t ii te 6220 Cri1ebi.a Road
Saii Antonio, TX 78238
C a 1 i b ration L 21 bo 1-23 to r y Del) it rt me 11 t of Qu ;I I i t y Assu I - i l IICC
Certificate of Calibration 1 May 1998
Issued to: DARRELL DUNN DIV20 Manufacture r/M ode I : ME'TTLER UMT2 Description: ELECTRONIC BALANCE Serial Number: 21 1001 -142 Asset Number: 002028
B57
v ACCREDITED
c ~- ~
IEnvironmental Conditions
Temperature: 73.00 Deg. F Humidity: 62 O/O RH
Calibration was in accordance with requirements of MIL-STD-45662A and ANSVNCSL 2540-1 -1 994. Measurements are traceable to the National Institute of Standards and Technology (NET). This report may not be reproduced except in full without written approval of the originator. Inspection and test data are on file and available for inspection.
The uncertainty of the calibration was sufficient to determine that the instrument met the manufacturer's speci fi cations.
Calibration Date: 4 May 98 Ca I i brat i o n P roced u re : CLCP-WT-001
Interval: 6 months
Next Calibration Due: 4 Nov 98 Received: In Tolerance
Remarks: UNCERTAINTY OF +/-0.034 MIL GRAMS
IStandards Used c Asset MFR Model Description Serial No. Due Cal
001706 RICE LAKE 2G WEIGHT STANDARD 001 720 RICE LAKE 1 MG-500 MG WEIGHT SET
LAST PAGE OF REPORT Total Pages Printed 1
C861 C858
21 Jul98 21 Jul 98
Certificate # 29572
\
Iss ed to
Southwest Research Institute 6220 Culebra Road
San Antonio, TX 78238 Department of Quality Assurance Calibration Laboratory
ACCREDITED - Certificate #
09 72-0 1
Certificate of Calibration 27 October 1998
DARRELL DUNN Manufacturer/Model: METTLER UMT2 Description: ELECTRONIC BALANCE Serial Number: 21 1001-142 Asset Number: 002028
DIV20 B57
kalibration Information I
Calibration was in accordance with requirements of MIL-STD-45662A and ANSVNCSL 2540-1 -1 994. Measurements are traceable to the National Institute of Standards and Technology (NIST). This report may not be reproduced except in full without written approval of the originator. Inspection and test data are on file and available for inspection.
This laboratory is accredited by the American Association for Laboratory Accreditation (A2LA) and the results shown in this calibration certificate have been determined in accordance with the laboratory's terms of accreditation unless stated otherwise in the report. The uncertainty of the calibration was sufficient to determine that the instrument met the manufacturer's specifications.
Temperature: 70.0 Deg. F Humidity: 65 % RH
Calibration Date: 27 Oct 98 Calibration Procedure: CLCP-WT-001
Interval: 6 months Received: IN TOLERANCE
Next Calibration Due: 27 Apr 99
Remarks:
lSGnd&ds Used b
Asset MFR Model Description Serial No. Due Cal
00170p RICE LAKE 1G WEIGHT STANDARD RICE LAKE 2G WEIGHT STANDARD RICE LAKE 1 MG-500 IblG WEIGHT SET
C859 25 Jun 99 C861 25 Jun 99 C858 25 Jun 99
n
LAST PAGE OF REPORT Total Pages Printed: 1
Certificate # 31805
Southwest Research Institue 6220 Culebra Road
San Antonio, TX 78238
Department of Quality Assurance (210) 522-5215
Calibration Laboratory
Certificate of Calibration 7 April 1999
Accredited
Certificate ## 0972-01
Issued to: DARRELLDU" DIV20 B57 ManufacturerModel: ME"LER UMT2 Description: ELECTRONIC BALANCE
Asset Number: 002028 Serial Number: 211001-142
This certifies the above item was calibrated in compliance with MIL-STD-45662A and ANSVNCSL 2540-1-1994. Standards used in this calibration, described in the referenced calibration procedure with associated uncertainties or tolerances, ame traceable to the National Institute of Standards and Technology (NIST). Supporting documentation relative to traceatility is on file and is available for examination upon request. This certificate is not to be reproduced, except in firll, without the written approval of the Southwest Research Institute Department of Quality Assurance Calibration Laboratory.
This laboratory is accredited by the American Association for Laboratory Accreditation (A2LA) and the results of this calibration certificate were: determined in accordance with the terms of accreditation unless stated o thedse below.
The uncertainty of the calibration was sufficient to determine that the item met the manufacturer's published specifications unless stated otherwise below.
Ambient Conditions: Temperature: 67.0 Degrees Fahrenheit Humidity: 68 'YO RH
Calibration Date: 3 1 Mar 99 Calibration Procedure: CLCP-WT-001 AUG 97
Condition as Received: IN TOLERANCE
Condition as Released: IN TOLERANCE, ADJUSTED
Remarks: MEASURING UNCERTAINTY IS +/-0.034 MILLIGRAMS.
Approved by: _ _
: Jim Patterson, Supervisor or Walt Hill, :Metrologist
Certificate ## 33828 m:\a2la.rpt Rev date 10 Mar 99
I I / Page 1 of 1
Southwest Research Institue 6220 Culebra Road
San Antonio, TX 78238
Department of Quality Assurance (210) 522-5215
Calibration Laboratory
Accredited m
Certificate # 0972-0 1
Certificate of Calibration 30 September 1999
Issued to: DARRELLDUNN DIV20 B57 ManufacturerRModel: METTLER lJMT2 Description: ELECTRONIC BALANCE
Asset Number: 002028 Serial Number: 211001-142
This certifies the above item was calibrated in compliance with MIL-STD-45662A and ANSI/NCSL 2540-1-1994. Standards used in this calibration, described in the referenced calibration procedure with associated uncertainties or tolerances, are traceable to the National Institute of Standards and Technology 0. Supporting documentation relative to traceability is on file and is available for examination upon request. This certificate is not to be reproduced, except in full, without the written approval of the Southwest Research Institute Department of Quality Assurance Calibration Laboratory.
This laboratory is accredited by the American Association for Laboratory Accreditation (A2LA) and the results of t h i s calibration certificate were determined in accordance with the terms of accreditation unless stated otherwise below.
The uncertainty of the calibration was siflicient to determine that the item met the manufacturer's published specifications unless stated otherwise below.
Ambient Conditions: Temperature: 78.0 Degrees Fahrenheit Humidity: 37 % RH
Calibration Date: 30 Sep 99 Calibration Procedure CLCP-WT-001 AUG 97
Condition as Received: IN TOLERANCE
Condition as Released: IN TOLERANCE
Remarks:
n Approved by: d
Jim Patterson, S u w Certificate # 36118 m:Wlrrpt Rev date 10 Mar 99
Page 1 of 1
SOUTHWEST RESEARCH INSTITUTE CALIBRATION LABORATORY
6220 CULEBRA ROAD POST OFflCE DRAWER 28510 SAN ANTONIO, TEXAS 78228-0510 TEL (210) 522-5215 FAX (210) 5223692
Receipt for Work Order 40566 Date Received 9/14/00 ................................... ............................................
Asset No. 002028 Manufacturer METTLER Model UMT2 ........................................................... ............................. ............................................ ....................................
Description ELECTRONIC BALANCE Serial Number 21 1 001 -1 42 ......................................................................................................................................................................................................................... ........................................................
Accessory Received/Required NONE ........................................................................................................................................................................................
Div/CC ID NONE Accessory to Asset No NIA Accuracy MFG SPECS ............................................................................................. ....................................................................................... .........................................................................................................
Custodian DARRELL DUNN Tel. 6090 ................................................... ........................................................... ...................... B57 ............................................. Div/CC DIV20 Location
Char g e/P ro j ec t N 0. 2 0.0 0 7 5 1 . 0 0 6 P ro p r i e t a ry /Co n f i de n ta I ..... N.. ... Date R eq u i red R 0 U TI N E .................................................................................................. ..................................
Work Requested CALIBRATION ONSlTEf ................................................................................................................................................................................................................................................................��
Rece i vi n g I ns pect i o n
Delivered By N/A ....................................................................................................................................................................................................................................................... Tel. 6090
CUSTOMER COMMENTS REQUESTED The Calibration Laboratory staff is committed to good laboratory practice and providing our customers quality calibration services. Please rate how well we are meeting our goal and provide us with your recommendations for im prov em en t .
Quality of service provided Turn-around time Certificates Calibration labels Recall notification Condition of returned equipment Overall service
Rating Scale 4 = Excellent
2 = Acceptable 1 = Needs improvement
3 = Good
Please check all that apply: Equipment user Delivery person Cus tod ian
IF WE PLEASE YOU, TELL OTHERS; IF NOT, TELL US.
COMMENTS:
Name Tel.
workreq.rpt Rev 30 Jan 00
Y
SOUTHW EST RESEARCH INS' 6 2 2 0 C u l e b r a R o a d , P.O. D r a w e r 2 8 5 1 0
Ins t i tu te Q u a l i t y S y s t e m s Ins t i tu te Cal ibra t ion L a b o r a t o r y
P h o n e : 2 1 0 - 5 2 2 - 5 2 1 5 F a x 2 1 0 - 5 2 2 - 3 6 9 2
Certificate of Calibration
ITU T E T M
Submitted By: D I V 2 0 Address: B57 Contact: D A R R E L L D U N N
Manufacturer Model: M E T T L E R U M T 2 Description: BALANCE
Serial No: 2 1 1 0 0 1 - 142 Asset No: 002028
Procedure: CLCP-WT-00 1, 12/99
Work Order: 444049935 Date Issued: 8/20/2002
Calibration Date: Aug 1 9 , 2 0 0 2 ""Calibration Due: F e b 1 9 , 2 0 0 3
Calibration Location: B57 L113 Environment: Temp. 70.O"F Hum. 62 %RH
*As Found: LIMITED CALIBRATION *As Left: LIMITED CALIBRATION
This certificate documents traceability to the National Institute of Standards and Technology (NIST) and the International System of Units (SI). The Laboratory quality system conforms to ISO/IEC 17025, 1999 and ANSI/NCSL 2540-1-1994 which are equivalent to relevant requirements of the IS0 9000-1994 series of standards. This certificate may not be reproduced, except in full, without the written approval of the Southwest Research Institute Calibration Laboratory. The results of this calibration relate only to the individual instrument described above. This certificate shall not be used to claim product endorsement by the American Association for Laboratory Accreditation (A2LA) or any agency of the U. S. Government.
Uncertainty evaluation includes the item under test and is calculated in accordance with the IS0 "Guide to the Expression of Uncertainty in Measurement" (GUM). The uncertainty represents an expanded uncertainty using a coverage factor of k=2 to approximate a 95% confidence level. The calibration process provides a Test Uncertainty Ratio (TUR) of less than or equal to 25% (4: 1) of the test limit unless otherwise stated in remarks or an attachment.
*The client has sole responsibility for determination of inlout olf tolerance or compliancehoncompliance. An in/out of tolerance opinion is provided for your convenience based only on the Test Instrument (TI) reading(s) and limits as reported. The reported uncertainty relates only to the results at the time of calibration and does not imply any short or long term stability of the TI.
**Calibration interval is determined by the client and does not assure the instrument will remain within tolerance until this date. Any number of factors may cause the instrument to be out of tolerance before the next calibration date.
Remarks: U S E R STATED A C C U R A C Y IS +/-0.034. MILLIGRAMS.
Standards Used Asset Manufacturer Model Description Cal Due
00 1705 RICE LAKE 2G WEIGHT, CLASS 2 Jun 14 ,03 00 1704 RICE LAKE 1G WEIGHT, CLASS 1 Jun 14,03 001720 RICE LAKE 1MG-500 MG WEIGHT SET, CLASS 1 Jun 14,03
Approved by: Walt Hill Metrology Group Leader m:Wona2lal.rpt Rev date 15, August 02
Metrology Technician Page 1 of 1
Southwest Research lnst it u te Calibration Laboratory Calibration Data Sheet
Location: Bldg. 57/ Lab 113 I IAmbient Conditions: 70 F 62 %RH 14.18 PSlA
Operational Check: Limits +/- : 0.0000340 g Uncertainty: 0.0000746 g
STD Mass As Found Instrument Load I n d ica t ion Error
Post Calibration Check:
STD Mass Post calibration Instrument
Repeatability Check: Mass Load: 1.0000000 g
1 2 3 4 5 , Std Dwiation Tolerance
0.0000074 g 0.0000340 g I : k: Mass Load: 1 .OOOOOOO g Uncertainty: 0.0000746 g
Instrument I n d i cat io n Error +/- Limits Resu Its
1 2 3 4
I I
Non-Linearity Check: Fta ng e : 2.0000000 g Uncertainty: 0.0000746 g
STD Mass Instrument Load In d icat i o n Error +/- Limits Results
Datasheets, Mettler.xls 20 August 2002, Rev. 01 Page 1 of 1
L
SOUTHW EST RESEARCH INS' 622:O C u l e b r a Road, P.O. Drawer 285 10
Ins t i t u t e Qual i ty S y s t e m s Ins t i t u t e C a l i b r a t i o n L a b o r a t o r y
Phone: 210-522-5215 F a x 210-522-3692 Certificate of Calibration
TITU TETM
Submitted By: DIV20 Address: B57 Contact: DARRELL DUNN
Manufacturer Model: METTLER UMT2 Description: BALANCE
Serial No: 21 1001-142 Asset No: 002028
Procedure: CLCP-WT-001, 12/99
Work Order: 444052194 Date Issued: Jan 30,2003
Calibration Date: Jan 29, 2003 **Calibration Due: Jul29, 2003
Calibration Location: B57 LAB 113 Environment: Temp. 72.O"F Hum. 53 %RH
*As Found: SEE ATTACHED DATA *As Left: SEE ATTACHED DATA
This certificate documents traceability to the National Institute of Standards and Technology (NIST) and the International System of Units (SI). The Laboratory quality system conforms to ISO/IEC 17025, 1999 and ANSI/NCSL 2540-1-1994 which are equivalent to relevant requirements of the IS0 900G1994 series ofstandards. This certificate may not be reproduced, except in full, without the written approwl ofthe Southwest Research Institute Calibration Laboratory. The results of this calibration relate only to the individual instrument described above. This certificate shall not be used to claim product endorsement by the American Association for Laboratory Accredtation (A2LA) or any agency ofthe U. S. Government
Uncertainty evaluation includes the item under test and is calculated in accordance with the IS0 "Guide to the Expression of Uncertainty in Measurement" (GUM). The uncertainty represents an expanded uncertainty using a coverage factor ofk=2 to approximate a 95% confidence level. The calihation process provides aTest Uncertainty Ratio (TUR) of less than or equal to 25% (4: 1) ofthe test limit unless otherwise stated in remarks or an attachment
*The client has sole responsibility for determination of in/out oftolerance or compliance/noncompliance. An in/out of tolerance opinion is povided for your convenience based on@ on the Test Instrument (TI) reading(s) and limits as reported. The reported uncertainty relates only to the results a t the time ofcalibration and does not imply any short or long term stability ofthe TI.
**Calibration interval is determined by the client and does not assure the instrument will remain within tolerance until this date. Any number of factors may cause the instrument to be out of tolerance before the next calibration date. Remarks: None
Standards Used Asset Manufacturer Model Description Cal Uue
007 102 TROEMNER 1MGTO lOOG WEIGHT SET, CLASS El Jun 13,03
Approved by: Walt Hill Metrology Group Leader n:\Nona2lal.rpt Revdate 15, August 02
Metrology Technician 1 of 1
So ut hwest Research Institute Calibration La bora tory Calibration Data Sheet
. Work Order 444052194 Mfr. Mettler Technician Mark A. Romero Asset No. 002028 Model UMT2 Procedure CLCP-WT-001. 12/99
Location: Bldg. 57/ Lab 113 I IAm bient Conditions: 72 F 53 %RH 14.21 PSlA
Limits +/- : 14.3 ug Uncertainty: 20.4 ug
Nominal Conventional Weight Value Value (Di'sola v) Deviation
Post Calibration Check:
Nominal Conventional Weight Value Adjustment Value Mass De via tion
Repeatability Check: Mass Load: 2000000.0 ug
1 2 3 4 5
Std Deviation Tolerance 0.8 ug 0.5 ug I
I Off -Centerline Check: Mass Load: 1000000.0 ug Uncertainty:
In strum en t Indication Error +/- Limits
1 2 3 4
20.4 ug
2000000.0 ug Uncertainty: 20.4 ug
Nominal Conventional Weight Value
Datasheets, Mettler 20 January 2003, Rev. 01 Page 1 of 1
, J
S O U T H W E S T R E S E A R C H INSTITUTETM 6220 C u l e b r a R o a d , P . O . D r a w e r 28510
Ins t i t u t e Q u a l i t y S y s t e m s Inst i tute Cal ibra t ion L a b o r a t o r y
P h o n e : 210-522-5215 F a x 210-522-3692
0972-01 Certificate of Calibration
Submitted By: DIV20 Address: B57 Contact: DARRELL DUNN
Manufacturer Model: METTLER UMT2 Description: BALANCE
Serial No: 2 1 100 1 - 142 Asset No: 002028
Procedure: CLCP-WT-00 1, DEC/99
Work Order: 444055472 Date Issued: Oct 2,2003
Calibration Date: Oct 2, 2003 **Calibration Due: Apr 2, 2004
Calibration Location: B57 Environment: Temp. 73.O"F Hum. 48 %RH
*As Found: SEE ATTACHED DATA *As Left: SEE ATTACHED DATA
This certificate documents traceability to the National Institute of Stimdards and Technology (NIST) and the International System of Units (SI). The Laboratory quality system conforms to ISO/IEC 17025, 1999 and ANSWCSL 2540-1-1994 which areequivalent to relevant requirements of the IS0 900@ 1994 series ofstandards. This certificate may not be reprodwed, except in full, without the written approml ofthe Southwest Research Institute Calibration Laboratory. The results of this calibration relate only to the individual instrument described above. This certificate shall not be used to claim poduct endorsement by the American Association for Laboratory Accreditation (A2LA) or any agency ofthe U. S. Government
Uncertainty evaluation includes the item under test and is calculated in accordance with the IS0 "Guide to the Expession of Uncertainty in Measurement" (GUM). The uncertainty represents an expanded uncertainty using a coverage factor ofk=2 to approximate a 95% confidence level. The calikation process provides aTest Uncertainty Ratio (TUR) of less than or equal to 25% (4: 1) of the test limit unless otherwise stated in remarks or an attachment
*The client has sole responsibility for determination of idout oftolerance or compliance/noncompliance. An in/out of tolerance opinion is povided for your convenience tased only on the Test Instrument (TI) reading(s) ard limits as reported. The reported uncertainty relates only to the results at the time of calibration and does not imply any short or long term stability ofthe TI.
**Calibration interval is determined bj the client and does not a s w e the instrument will remain within tolerance until this date. Any number of factors may cause the instrument to be cut of tolerance before the next calibration date. Remarks: None
Standards Used Asset Manufacturer Model Description Cal Due
007 102 TROEMNER 1MG TO lOOG WEIGHT SET, CLASS E l Aug 29,04
Approved by: Walt Hill Metrology Group Leader m:\dlal.rpt Rev date 15, August 02 1 of 1
c . L.
Southwest Research Institute Calibration La bora tory Calibration Data Sheet
IAmbient Conditions: 73 F 48 %RH 14.35 PSlA
Operational Check: Limits +/- : 14.3 ug Uncertainty: 20.4 ug
Nominal Conventional Weight Value Value (Display) De via tion
2000000.0 1999995.6 u 2000399.2 ug 403.6 ug I Post Calibration Check:
Nominal Conventional Weight Value Adjustment Value (Disda Y ) De via tion
Re peata bi I ity Check: Mass Load: 2000000.0 ug
1 2 3 4 5
Std Deviation To le ra n ce I 19.1 uq 0.5 uq I
Off-Centerline Check: Mass Load: 1000000.0 ug Uncertainty: 20.4 ug
lnstrumen t Indication Error +/- Limits
I 2 3
Rang e : 2000000.0 ug Uncertainty: 20.4 ug
m 4 Non-Linearity Check:
Conventional Weight Value Mass (Dimla v) De via tion +/- Limits
1500000.0 500002.2 ug 499991.1 ug -1.5 ug 2.0 2000000.0 499994.6 ug 499992.3 ug -0.3 ug 2.0
__-____ Remarks: * Data is provided without regard to ‘pass” or “fail“. The end user is responsible ____ in determining __ whether the results meet
their - . requirements. _. - - - . . - _ _ . _.
Readability is 0. lug. Standard used 7702. __________
Datasheets, Mettler.xls 20 January 2003, Rev. 01 Page 1 of 1
&
S O U T H W E S T R E S E A R C H INSTITUTETM 6 2 2 0 C u l e b r a R o a d , P . O . D r a w e r 2 8 5 1 0
Ins t i tu te Qua l i ty S y s t e m s Ins t i tu te C: a l ibra t ion Labora to ry
P h o n e : 2 1 0 - 5 2 2 - 5 2 1 5 F a x 210-522-3692
Certificate of Calibration 0972-01
Submitted By: DIV20 Address: B57 Contact: DARRELL DUNN
Manufacturer Model: METTLER UMT2 Description: BALANCE
Serial No: 2 1 100 1 - 142 Asset No: 002028
Procedure: CLCP-WT-00 1 , DEC/99
Work Order: 444058 18 1 Date Issued: Apr 6,2004
Calibration Date: Apr 2, 2004 **Calibration Due: Oct 2, 2004
Calibration Location: B57 Environment: Temp. 74.O"F Hum. 65 %RH
*As Found: SEE ATTACHED DATA *As Left: SEE ATTACHED DATA
This certificate documents traceability to the National Institute of Standards and Technology (NIST) and the International System of Units (SI). The Laboratory quality system conforms to ISO/IEC 17025, 1999 and ANSL'NCSL 2540-1-1994 which areequivalent to relevant requirements of the IS0 900@ 1994 series ofstandards. This certificate may not be reproduced, except in full, without the written approml of the Southwest Research Institute Calibration Laboratory. The results of this calibration relate only to the individual instrument described above. This certificate shall not be used to claim pcduct endorsement by the American Association for Laboratory Accreditation (A2LA) or any agency ofthe U. S. Government
Uncertainty evaluation includes the item under test and is calculated in accordance with the IS0 "Guide to the Expession of Uncertainty in Measurement" (GUM). The uncertainty represents an expanded uncertainty using a coverage factor ofk=2 to approximate a 95% confidence level. The calibration process provides aTest Uncertainty Ratio (TUR) of less than or equal to 25?? (4: 1) of the test limit unless otherwise stated in, remarks or an attachment
*Theclient has sole responsibility for determination of idout of tolerance or compliance/noncompliance. An in/out of tolerance opinion is provided for ycur convenience based only on the Test Instrument (TI) reading@) a d limits as reported. The reported uncertainty relates only to the results at the time ofcalibration and does not imply any short or long term stability oftheTI.
**Calibration interval is determined by the client and does not assure the instrument will remain within tolerance until this date. Any number of factors may cause the instrument to be cut of tolerance before the next calibration date. Remarks: None
Standards Used Asset Manufacturer Model Description Cal Due
00 1705 RICE LAKE 001704 RICE LAKE 001720 RICE LAKE
2G 1G 1MG-500 MG
WEIGHT, CLASS 2 WEIGHT, CLASS 1 WEIGHT SET, CLASS 1
Jun 23 ,04 Jun 23 ,04 Jun 20,04
f
Approved by: Walt Hill Metrology Group Leader m:\a2lal .rpt Rev date 15, August 02 W 1 of 1
Southwest Research Institute Calibration Laboratory Calibration Data Sheet
Found / Left
Location: Bldg. 571 Lab L105 I IAmbient Conditions: F %RH PSlA
Operational Check: Limits +/- : 14.3 ug Uncertainty: 20.4 ug
I Nominal Conventional Weight Value
Repeatability Check: Mass Load: 2000000.0 ug
I 2 3 4 5
Std Deviation Tolerance 0.5 ug I I 3.8 ug
Mass Load: 1000000.0 ug Uncertainty: 20.4 ug
lnstrumen t Indication Error +/- Limits
Non-Linearity Check: I Fta n g e: 2000000.0 ug Uncertainty: 20.4 ug
Nominal Conventional Weighlt Value
Page 1 of 1 Datasheets, Mettler.xls 20 January 2003, Rev. 01
~ ~~ ~ ~~~~~~~ ~
SOUTHWEST RESEARCH INSTITUTETM 6220 C u l e b r a R o a d , P . O . Drawer 285 1 0
Ins t i tu te Q u a l i t y Systems Ins t i tu t e C a l i bra ti on Lab o r a t o r y
P h o n e : 2 10-522-52 1 5 Fax 2 1 0 - 5 2 2 - 3 6 9 2 Certificate of Calibration
Submitted By: DIV20 Address: B57 Contact: DARRELL DUNN
Manufacturer Model: METTLER UMT2 Description: BALANCE
Serial No: 2 1 1001-142 Asset No: 002028
Procedure: CLCP-WT-00 1, DEC/99
Work Order: 444061018 Date Issued: Oct 1, 2004
Calibration Date: Oct 1, 2004 **Calibration Due: Apr 1,2005
Calibration Location: B57 Environment: Temp. 74.O"F Hum. 73 %RH
*As Found: SEE ATTACHED DATA *As Left: SEE ATTACHED DATA
This certificate documents traceability to the National Institute of Standards and Technology (NIST) and the International System of Units (SI). The Laboratory quality system conforms to ISO/IEC 17025, 1999 and ANSI/NCSL 2540-1-1994 which are equivalent to relevant requirements of the I S 0 900@ 1994 series ofstandards. This certificate may not be reproduced, except in full, without the written approval ofthe Southwest Research Institute Calibration Laboratory. The results of this calibration relate only to the individual instrument described above. This certificate shall not be used to claim poduct endorsement by the American Association for Laboratory Accreditation (A2LA) or any agency ofthe U. S. Government
Uncertainty evaluation includes the item under test and is calculated in accordance with the IS0 "Guide to the Expession of Uncertainty in Measurement" (GUM). The uncertainty represents an expanded uncertainty using a coverage factor ofk=2 to approximate a 95% confidence level. See Remarks a attached Calibration Report with thesame Work Order number for calilxation data
*The client has sole responsibility for determination of idout ofto1ei:ance or compliancehoncompliance. An in/out of tolerance opinion is povided for your convenience bsed only on the Test Instrument (TI) reading(s) and limits as reported. The reported uncertainty relates only to the results at the time ofcalibration and does not imply any short or long term stability of the TI.
**Calibration interval is determined by the client and does not assure the instrument will remain within tolerance until this date. Any number of factors may cause the instrument to be out of tolerance before the next calibration date. Remark: Readings are without pass or fail indication. Customer must determine if measurements meet their needs.
Standards Used Asset Manufacturer Model Descrip tion Cal Due
005566 007290 001 704 001705 001720
MERIAN VAISALA RICE LAKE RICE LAKE RICE LAKE
A0030P HM34F 1G 2G 1MG-500 MG
Approved by: Walt Hill Metrology Group Leader n:\Nona2lal. rpt Rev date 1 1, May 04
PRESSURE GAUGE, ABSOLUTE Oct 15,04 HUMIDTY/ TEMPERATURE METER Apr 07 ,05 WEIGHT, CLASS 1 Jun 23,05 WEIGHT, CLASS 2 Jun 23 ,05 WEIGHT SET, CLASS 1 Jun 23,05
Measurements by: Scott Kester Metrology Technician
1 of 1
Southwest Research Institute Calibration Laboratory Measurement Report
Work Order: 444061018 Mfr. Mettler Asset No. 002028 Model UMT2
Technician SRK
Serial No. 21 1001-142 ITyple. Balance (Cat Date. 0 1 -0ct-04 Remarks: Readings are without pass or fail indications. Customer must determine if measurements meet their needs. Ambient Conditions 74 deg F 73 % RH 14.22 PSlA
FunctionIRange Test Point TI Reading Difference +/-Limit +/-Uncertainty
Repeatability Check ug ug ug ug ug 1 2000000.0 200001 3.2 2 2000000.0 2000014.7 3 2000000.0 2000002.6 4 2000000.0 200001 1.6 5 2000000.0 1999998.3 6 2000000.0 2000007.9 7 2000000.0 2000006.2 8 2000000.0 2000031 .O 9 2000000.0 2000004.7 10 2000000.0 1999996.9
Std Deviation 9.8
Front 1000000.0 1000014.6 14.6
Rear 1000000.0 1000014.4 14.4 Left 1000000.0 1000008.6 8.6
Offcenter Check
Right 1000000.0 100001 1.1 11.1
Non I i neari ty Check 0 0.0 0.0 0.0
500000 499996.3 :500005.1 8.8
1500000 499983.2 50001 5.9 32.7 2000000 500012.0 500036.5 24.5
End of Report
1000000 499986.9 1499968.2 -1 8.7
0.5
2.0 2.0 2.0 2.0
2.0 2.0 2.0 2.0 2.0
3.0 3.0 3.0 3.0
2.4 2.5 3.0 3.1 2.8
Rev, OcVO4 Page 1 of 1 Mettler U MT2.xls
S O U T H W E S T R E S E A R C H INS' 6220 Culebra Road , P . O . Drawer 2 8 5 1 0
Institute Quality Sys t ems Institute Calibration Laboratory
Phone : 210-522-5215 Fax 210-522-3692
Certificate of Calibration
TIT U T E T M
Certificate #
0972-01
Submitted By: DIV20 Address: B57 Contact: DARRELL D U "
Manufacturer Model: METTLER UMT2 Description: BALANCE
Serial No: 2 1 100 1 - 142 Asset No: 002028
Procedure: BALANCES & SCALES, DEC/04
Work Order: 303063529 Date Issued: Apr 4,2005
Calibration Date: Apr 1 , 2005
Calibration Location: B57 **Calibration Due: Sep 30, 2005
Environment: Temp. 73.O"F Hum. 22 %RH *As Found: SEE ATTACHED DATA
*As Left: SEE ATTACHED DATA
This certificate documents traceability to the National Institute of Standards and Technology (NIST) and the International System of Units (SI). The Laboratory quality system conforms to ISO/IEC 17025, 1999 and ANSIOJCSL 2540-1-1994 which are equivalent to relevant requirements of the IS0 9000-1994 series of standards. This certificate may not be reproduced, except in full, without the written approval of the Southwest Research Institute Calibration Laboratory. The results of this calibration relate only to the individual instrument described above. This certificate shall not be used to claim product endorsement by the American Association for Laboratory Accreditation (A2LA) or any agency of the U. S. Government.
Uncertainty evaluation includes the item under test and is calculated in accordance with the IS0 "Guide to the Expression of Uncertainty in Measurement" (GUM). The uncertainty represents an expanded uncertainty using a coverage factor of k=2 to approximate a 95% confidence level. See Remarks or attached Calibration Report with the same Work Order number for calibration data.
T h e client has sole responsibility for determination of inlout of tolerance or compliance/noncompliance. An idout of tolerance opinion is provided for your :onvenience based only on the Test Instrument (TI) reading(s) and limits as reported. The reported uncertainty relates only to the results at the time of calibration md does not imply any short or long term stability of the TI.
"Calibration interval is determined by the client and does not assure the instrument will remain within tolerance until this date. Any number of factors may cause h e instrument to be out of tolerance before the next calibration date.
Remarks: None
standards Used Manufacturer Model Description Cal Due Asset
007290 00 1704 00 1705 001720 005566
VAISALA RICE LAKE RICE LAKE RICE LAKE MERIAN
HM34F 1G 2G 1MG-500 MG A0030P
4pproved by: Walt Hill vletrology Group Leader n:\a2lal .rpt Rev date 1 1, May 04
HUMIDTY/ TEMPERATURE METER Apr 07,05 WEIGHT, CLASS 1 Jun 23,05 WEIGHT, CLASS 2 Jun 23,05 WEIGHT SET, CLASS 1 Jun 23,05 PRESSURE GAUGE, ABSOLUTE Apr 14,05
I
Measurements by: Scott Kester Metrology Technician
Page 1 of 1
k
Work Order: 303063529 Asset No. 002028
Southwest Research Institute Cali bration Laboratory Measurement Report
Mfr. Mettler Technician SRK Model UMT2
Serial No. 21 1001-142 IType. Balance lCal Date. 01 -Apr-05 Remarks: Readings are witholut pass or fail indications. Customer must determine if measurements meet their needs. Ambient Conditions 73 O F 22 % RH 14.44 PSlA
Function/Range Test Point TI Reading Difference +/-Limit +/-U ncertai n ty
Repeatability ug ug ug ug ug 1 1000000.0 !399977.7 2 1000000.0 !3 9 9 9 82.4 3 1000000.0 !39 998 3.0 4 1000000.0 99 99 80.4 5 1000000.0 999983.1 6 1000000.0 99 99 80.6 7 1000000.0 13 99 96 8.8 8 1000000.0 $399981.6 9 1000000.0 $399984.3 10 1000000.0 $399970.5
Std Deviation 5.39 0.50 Nonlinearity Check
0 ug 0.0 0.0 0.0 2.0 8.9 200000 ug 199999.8 II 99999.8 0.0 2.0 8.9 400000 ug 399994.0 3 99999.4 5.4 2.0 8.9 600000 ug 599999.0 600007.8 8.8 2.0 8.9 800000 ug 799998.8 €!00004.1 5.3 2.0 8.9 1000000 ug 999986.9 99 9986.8 -0.1 2.0 8.9 1200000 ug 1 199986.7 1 199986.7 0.0 2.0 8.9 1400000 ug 1399980.9 1$399986.3 5.4 2.0 8.9 1600000 ug 1599985.9 11599981.5 -4.4 2.0 8.9 1800000 ug 1799985.7 1'799997.4 11.7 2.0 8.9 2000000 ug 2000012.0 200001 8.0 6.0 2.0 8.9
End of Report
Rev, OW04 Page 1 of 1 Mettler UMT2.xls