Upload
others
View
46
Download
0
Embed Size (px)
Citation preview
Operator’s manual
DELTASCOPE® MP10EISOSCOPE® MP10E
DUALSCOPE® MP20E-S RS232 interface
Manual order no. 902-585 Version 2.0 02/2006
DKD-K-33101Certified according to ISO 9001 : 2000
Operating Manual FERITSCOPE® MP30E (2.1 - 01/06) Page 1
Page 2 Operating Manual FERITSCOPE® MP30E (2.1 - 01/06)
� For more detailed information please refer to the operator manual MP10E and MP20E!
Short-Form Operator Manual DELTASCOPE MP10E,
ISOSCOPE MP10E and DUALSCOPE MP20E
®
® ®
ZERO CALNF/Fe NC/NF
milsmm FN
µmMS/m
%
ZERO
CAL
NF/Fe
NC/NF
milsmm FN
µmMS/m
%
Fischer trademark
Normalization is performed
Measurement using the magneticinduction test method is performed
Measurement using the eddy cur-rent test method is performed
Calibration is performed
Restricted operating mode hasbeen selected
"Continuous" display mode hasbeen enabled
Measurement values, errormessages and warnings
Unit of measurement of thedisplayed value
Internal routine is performed
Battery has to be changed orrecharged
Configuration programs have beencalled up
Instrument type
Instrument software version
Symbols and Styles Used
ENTER
ZERO
ON/OFF +
Refers to instrument keys, which have to be pressed
perform Refers to instrument keys or actions, which have to be pressed or performed one
after the other
Refers to instrument keys, which have to be pressed immediately one after the
other (do not keep both keys pressed!)
�
�
Display
��������
������
Performing a Measurement:
ON/OFF
Deleting Measurement Values:
DEL DEL
"Continuous" Display Mode:
Enabling: Disabling
Evaluation (only with MP20E):
FINAL-RES FINAL-RES
ENTER
Normalization:
ZERO
ENTER
Corrective Calibration:
CAL
ENTER
ENTER
ENTER
1. Switch the instrument on:
2. Measuring: Place the probe vertically on the
measuring object
(last value) (all values)
Move the probe over a surface area to deter-
mine the coating thickness distribution.
appears; :
...
(exit the evaluation and delete all
the measurements) or (exit the evaluation
without deleting the measurements)
perform several measurements
on the substrate material
1. perform several measurements on
the substrate material
2. Place standard 1 (foil 1) on the substrate
material perform several measurements
on the standard 1 (foil 1)
3. Set the foil thickness with ,
4. If a 2 point calibration is desired: Place
standard 2 (foil 2) on the substrate mate-
rial perform several measurements on
foil 2 set the foil thickness with , ;
otherwise: after step 3 immediately step 5
5.
�
� � �
� � �
�
�
�
�
�
�
� � �
�
� � �
Measurement and Evaluation:
Normalization / Corrective Calibration:
� For more detailed information please refer to the operator manual MP10E and MP20E!
®
Restricted Operating Mode:
Enabling: ( appears)
Disabling:
ON/OFF + DEL
ON/OFF + ENTER
Configuration
Overview of the Configuration Programs:
10 x ENTER
[ ]157
[ ]159
[ ]FrEE
[ ]FrEE
�
Re-Ini-tiali-
zation
DEL
[ ]30.2 Last measurement value will be displayed;configuration programs have been exited!
Configuration programs are active now!
CAL
MasterCali-
bration
�
Display theRecord of the
InstrumentStatus
� Use and to set the parameters of
the configuration programs; use
to confirm the setting and to
step to the next parameter.
� �
ENTER
�
�
only for MP10E
only for MP20E
© HELMUT FISCHER GMBH+CO.KG Subject to change. 902-586 Version 1.3 - 11/05
H E L M U T F I S C H E R G M B H + C O . K GInstitut für Elektronik und MesstechnikIndustriestrasse 21, 71069 Sindelfingen, GermanyTel.: (0 70 31) 3 03-0, Fax: (0 70 31) 3 03-79E-Mail: [email protected]: www.helmut-fischer.com
Germanischer LloydCertification
ISO 9001:2000
Certified acc. to
Acoustic Measurement Accept Signal:
Enabling:
Disabling:
ON/OFF +
ON/OFF +
�
�
FINAL-RES
StorageMode*
Switch-OffMode*
2 x �
ENTER
t
t
ZERO
Unit of MeasurementLanguage
Display ModeDisplay ResolutionSwitch-Off Mode
Dual Method*
�
0405
24
Sales and Service Departments- all around the world -GERMANYHELMUT FISCHER GMBH+CO.KGIndustriestraße 21D-71069 Sindelfingen
: (+49) 70 31 303-0Fax: (+49) 70 31 [email protected]
GREAT BRITAINFISCHER INSTRUMENTATION (G.B.) LTD.Gordleton Industrial ParkHannah Way · PenningtonGB-Lymington/Hants SO41 8JD
: (+44) 15 90 68 41 00Fax: (+44) 15 90 68 41 [email protected]
USAFISCHER TECHNOLOGY, INC.750 Marshall Phelps Road Windsor, CT 06095 USA
: (+1) 860 683 07 81Fax: (+1) 860 688 84 96Watts 800 243 84 [email protected]
Sole Agent for Helmut Fischer GmbH+Co.KG, Germany:
SWITZERLANDHELMUT FISCHER Elektronik und Messtechnik AGMoosmattstrasse 1 · Postfach · CH-6331 Hünenberg
: (+41) 41 785 08 00 · Fax: (+41) 41 785 08 [email protected]
Branch Offices of Helmut Fischer AG, Switzerland:
SPAINFISCHER INSTRUMENTS, S.A.C/Almogàvers 157 · 3a PlantaE-08018 Barcelona
: (+34) 93 309 79 16Fax: (+34) 93 485 05 [email protected]
ITALYHELMUT FISCHER S.R.L. Tecnica di Misura Via Columella 40 · I-20128 Milano
: (+39) 0 22 55 26 26Fax: (+39) 0 22 57 00 [email protected]
JAPANFISCHER INSTRUMENTS K.K.Shinmei 1-9-16· Souka-shiSaitama-ken 340-0012 · Japan
: (+81) 489 29 34 55Fax: (+81) 489 29 34 [email protected]
FRANCEFISCHER INSTRUMENTATION ELECTRONIQUEParc d'Activités Nord du Pas du Lac · 5, rue Michaël FaradayF-78180 Montigny le BretonneuxBP 289 · F-78053 St Quentin en Yvelines Cedex
: (+33) (0) 1 30 58 00 58Fax: (+33) (0) 1 30 58 89 [email protected]
HONG KONGFISCHER INSTRUMENTATION (FAR EAST) LTD.Unit 2901 · Level 29Metroplaza Tower 2223 Hing Fong RoadKwai Chung, N.T.Hong Kong
: (+852) 24 20 11 00Fax: (+852) 24 87 02 [email protected]
THE NETHERLANDSHELMUT FISCHER MEETTECHNIEK B.V.Tarasconweg 10NL-5627 GB EindhovenPostbus 1828NL-5602 CA Eindhoven
: (+31) 40 248 22 55Fax: (+31) 40 242 88 [email protected]
CHINANANTONG FISCHER INSTRUMENTATION LTD.7F, No. 2 Building2601 Songhuajiang RoadShanghai 200437 · P.R.C.
: (+86) 21 65 55 74 55 Fax: (+86) 21 65 55 24 [email protected]
SINGAPOREFISCHER INSTRUMENTATION (S) PTE LTD.102 E Pasir Panjang Road #04-04Citilink Warehouse ComplexSingapore 118529
: (+65) 62 76 67 76Fax: (+65) 62 76 76 [email protected]
DKD-D-33101Accredited acc. toDIN EN ISO/IEC 17025
Valid for Helmut Fischer AG and Branch Offices