DATALOGGERS Kipp & Zonen pyranometers such as the CM3, CM6, CM11. CM21 etc are all compatible with the Skye DataHog2 logger. There voltage output can be connected to either a single ended voltage DataHog socket (sometimes labelled 0-2V) or a differential voltage socket. Skye recommend the differential voltage option, although either is possible. Differential V oltage Input Pin 1 +5V power supply not connected Pin 2 not used not connected Pin 3 Negative signal input Negative output Pin 4 Positive signal input Positive output Pin 5 Ground Cable screen Before configuration of a DataHog2 differential voltage channel for a pyranometer, you will first need to calculate the logger's Full Scale Value and Zero Offset values as shown in the DataHog manual, Chapter 3.2.10 and Technical Notes 2.0. Pyr anomet er CM21 with a sensiti vit y of 15 V/Wm (or 66. 67 Wm per mV) These sensors do not have a zero offset. In this example the maxi mum pyran ometer out put at 1200 Wm is 18 mV So the DataHog range required is 0-20 mV, i.e. a gain of 100, Gain Code 2 Dat aHog Full Sca le = [S ens itiv ity (Wm per mV) / Ga in] *2000 = [66.67 / 100] *2000 = DataHo g Zero Off set = Sensor of fset (mV) * Gain * 9.5 = 0 * 100 * 9.5 = Enter these calculated values as follows: 1. W ake up the DataHog2 as usual, to reveal the Main Menu of the logger 2. Press '9 ' to ch oose 'O ption 9 - Set Ax+B c alibrat ion fa ctors' A) Ent er the sof twar e c han nel you wis h t o configure (e. g. 00 or 01 etc. ) B) Enter the Full Scale V alue for the pyran ometer associ ated with that softwar e chann el (E.g. 1333.3 as in example above). You must use the format of 5 digits plus a decimal  point, with no leading zeros. C) Ent er the Off set Cou nt for the pyr anomete r ( e.g. 000 0 as abo ve) WIRING INSTRUCTI ONS DATAHOG2 CONFIGURATION Example: 1333.4 +0000  DataHog socket Function Pyranometer Connector -2 -2 -2 -2 Updated 30/11 /04 Configuring for Kipp & Zonen Pyranometers Link pins 3 & 5

DataHog2 Configure for Kipp Zonen Pyranometers

Embed Size (px)

Citation preview

7/31/2019 DataHog2 Configure for Kipp Zonen Pyranometers

http://slidepdf.com/reader/full/datahog2-configure-for-kipp-zonen-pyranometers 1/2

DATALOGGERS

Kipp & Zonen pyranometers such as the CM3, CM6, CM11. CM21 etc are all compatible with the SkyeDataHog2 logger. There voltage output can be connected to either a single ended voltage DataHog socket(sometimes labelled 0-2V) or a differential voltage socket. Skye recommend the differential voltageoption, although either is possible.

Differential Voltage Input

Pin 1 +5V power supply not connectedPin 2 not used not connectedPin 3 Negative signal input Negative outputPin 4 Positive signal input Positive outputPin 5 Ground Cable screen

Before configuration of a DataHog2 differential voltage channel for a pyranometer, you will first need to

calculate the logger's Full Scale Value and Zero Offset values as shown in the DataHog manual, Chapter 3.2.10 and Technical Notes 2.0.

Pyranometer CM21 with a sensitivity of 15 V/Wm (or 66.67 Wm per mV)These sensors do not have a zero offset.

In this example the maximum pyranometer output at 1200 Wm is 18 mVSo the DataHog range required is 0-20 mV, i.e. a gain of 100, Gain Code 2

DataHog Full Scale = [Sensitivity (Wm per mV) / Gain] *2000= [66.67 / 100] *2000=

DataHog Zero Offset = Sensor offset (mV) * Gain * 9.5= 0 * 100 * 9.5=

Enter these calculated values as follows:

1. Wake up the DataHog2 as usual, to reveal the Main Menu of the logger 

2. Press '9' to choose 'Option 9 - Set Ax+B calibration factors'A) Enter the software channel you wish to configure (e.g. 00 or 01 etc.)B) Enter the Full Scale Value for the pyranometer associated with that software channel

(E.g. 1333.3 as in example above). You must use the format of 5 digits plus a decimal point, with no leading zeros.

C) Enter the Offset Count for the pyranometer (e.g. 0000 as above)

WIRING INSTRUCTIONS

DATAHOG2 CONFIGURATION

Example:

1333.4

+0000

 DataHog socket Function Pyranometer Connector 

m -2 -2

-2

-2

Updated 30/11/04

Configuring for Kipp & Zonen Pyranometers

Link pins3 & 5

7/31/2019 DataHog2 Configure for Kipp Zonen Pyranometers

http://slidepdf.com/reader/full/datahog2-configure-for-kipp-zonen-pyranometers 2/2

D) Enter the Offset Sign + or - of the Offset Count (e.g. + as above)E) The figures you have just entered will be displayed for you to confirm. If OK, press 'Y' a

nd you will return to the Main Menu.

F) Repeat steps a) to e) for each differential voltage input to be configured.

3. Make sure that this software channel is set up with the correct Gain Code as used in the calculationfor the Full Scale Value.A) Choose Option A - Set channel configuration.B) Enter the software channel and hardware channel (from the logger's Hardware Configuration

Certificate)C) Enter the Gain Code (e.g. 2 in the above example)D) Enter the Scale Code - choose 1 for Ax+B scaling If correct, type Y to accept.E) It is advisable to check that the changes you have just made are in place before returning the

DataHog 2 to logging mode.

4. From the Main Menu, choose Option 1 to display the current setup.A) Choose sub-menu '3' to check the Ax+B Scaling Factors. The software channels will be

displayed with Full Scale Value and Zero Offsets you have entered.B) Choose sub-menu '0' to check the Gain Code and Scale Code for the software channel you

have reconfigured.C) Choose '6' to return to Main Menu

5. If no more changes are to be made, then press ESCAPE to return the DataHog to logging mode.

6. Write the new Full Scale Values and Zero Offset counts on the Hardware Configuration

Certificate for your future record.

EXAMPLE

Software Hardware Configured Gain Termination Scale Full ZeroChannel Channel for Input Type Code Code Code Scale Offset

00 02 Differential 02 00 01 1333.3 +0000Voltage

Configuring for Kipp & Zonen Pyranometers

(continued)

SKYE INSTRUMENTS LTD

21, Ddole Enterprise Park, Llandrindod Wells, Powys, LD1 6DF, UK

Tel: +44(0)1597 824811 Fax: +44(0)1597 824812

Email: [email protected] Web: www.skyeinstruments.com

Updated 30/11/04