22
Beijing Nubway S & T Co., Ltd CRYOLIPOLYSIS BODY SLIMMING MACHINE USER MANUAL NBW-C132

Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Embed Size (px)

DESCRIPTION

cryopipolisis

Citation preview

Page 1: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

CRYOLIPOLYSIS BODY SLIMMINGMACHINEUSERMANUAL

NBW-C132

Page 2: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

Content

1. System Overview1-1、Introduction to the system1-2、Introduction to the operation1-3、Treatment Parameters

2. Safety2-1、Electrical and Mechanical safety2-2、Movement and Transportation safety2-3、Warnings3. System Description3-1、product list3-2、System components Description=components3-3、 Warning4. Installation & Adjustment4-1、Installation4-2、Adjustment5. Specifications5-1、System specifications5-2、Relevant requirements6. Operation6-1、Starting the system6-2、Treatment6-3、Turning off the system7. Principle of treatment8. Course setting=parameter setting9. Indications and Contraindication9-1、Therapeutic range9-2、Contraindication10. Post-treatment Care11. Maintenance and Troubleshooting11-1、Cleaning the unit11-2、Cleaning the crystal11—3、Cleaning pump filters11-4、Breakdown maintenance11—5、Error information12. After-sales service13. Contact us

Page 3: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

Chapter 1 Introduction to the systemThanks for your purchasing this NBW-C132 slimming instrument!Believe that the instruments will bring you credible and reliable performance, returnyou with good functionality and satisfactory .This chapter is an overall description of NBW-C132 type slimming instrument.Note: The operation or maintenance personnel of the system read this manualcarefully before operating .

1-1. Introduction to the system

The machine is controlled by computer, mainly composed of the host, the controlpanel and treatment handle.The Operation and control parts of the machine located inside the host, the built-inintegrated microcontroller continuously monitors the operation of equipment.Control system includes a touch screen, handpieces, key switch and emergency stopswitch.The touch screen can show the system operating modes and parameters.

1-2. Introduction to the operationTreatment scopemelting fat dietFirming body sculptingcircumference reduction after liposuctionpostpartum shrink around shaping

1-3, Treatment ParametersThe allowed adjust treatment parameters for users are:Vaccum :0-100kpa continuously adjustableTemperature: 5——-15℃ continuously adjustable

Chapter 2 Safety

2-1 Electrical and Mechanical safety

In order to ensure the safety of usage, The instrument implements theinternational standards in terms of design and manufacturing strictly. It ismanufactured and examined according to the general safety requirements forHairdressing electrical equipments.

The single-phase voltage of the operating system is 220V, the maximum inputcurrent is less than 10A. After turning power off, residual voltage will still exist inthe high-voltage device, it will be danger for you to open shell/protective cover, for itwill be high pressure electric.

When the operating system is running, dangerously high voltages exist in themachine. When the power is turned off, the residual voltage may still exist in the high

Page 4: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

voltage devices, if you open the chassis protective cover, you may run the risk ofhigh-voltage electric shock.

Note: The operating system is powered with a standard three-pin GB connectors,please make sure your network Power socket with protective earthing line, and safelygrounded.

2-2Movement and Transportation safety

The operating system has precision parts, the users are not allowed to move itdistantly. During the transportation, the cushioning packaging provided by originalfactory must be used. Otherwise it will affect the normal use of the machine.

2-3Warning message

Non-professionals are not allowed to enter the treatment roomEnsure the present staffs are familiar with the system operation

Do not use this system in Inflammable and explosive area or other insecure location

Chapter 3 System Description3-1 Main equipment list of product

Page 5: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

The equipment main structure as above pictures, Touch screen, Emergency Switch,Key Switch, mainframe box, cooling system, treatment handle etc.

Page 6: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

3-2. System components DescriptionTouch screen: It is used to display and adjust the specific information of operatingsystem.Emergency Stop Switch: It is used to turn off the power in emergency conditions.Heart rate tester: it used to clip on your fingers to test the heart rate before do thecryolipolysis treatment.Fat measurements pliers: Measuring thickness of the fatMainframe box: Within the cooling system, power system, control system,integrated circuit, etc.Treatment handle: Removable type treatment handle, used to do the treatment forthe customerFunnel: Use for adding water into the water tankSponge Filter: Set on the suction mouth inside the handle, prevent liquid and sundriesbeing suck into the handle.Rubber gasket: Set of interface of the handle, water mouth, to prevent leakageEmergency Stop Switch: Shut down the machine power supply of urgent cases

3-3、Error messageERROR 1: Small water flow WarningERROR 2:Handle A communication outageERROR 3:Handle A excess temperatureERROR 4: Handle B communication outageERROR 5: Handle B communication outage3--4、Advantages

1. Non-surgical, non-narcotic2. Non-invasive,no infection, safety3. No need garment, no recovery4. No effect the normal work

Chapter 4 Installation and Commissioning4-1, installation

The system can be installed in hospitals, medical clinics and beauty salons, littlepreparation is needed before installation. When you purchase our company'sinstruments, you can enjoy our free onsite installation and commissioningservices.The system should be installed and commissioned by professional and technicalpersonnel authorized by our company. Installation work includes:

4-1-1, check equipment list:Mainframe (1)Treatment heads (2)Heart rate test finger-cot (1)

Page 7: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

Fat measuring caliper (1)Measurement scale (1)Switch key (2)User Manual (1)Water funnel (1)Drain socket (1)Power cord (1)Rubber ring (8)Handle bracket (1)Package box (1)Sponge Filter(10)

4-1-2, Be familiar with the installation requirements:Before unpacking the Instrument, you should check whether the accessories arecomplete, the installation location meets the requirements of this section.

The system should be installed away from heat vents and keep an appropriatedistance from other objects.

The instrument is preset the rated voltage of the installation place before deliveryaccording to the user's requirements. The corresponding electrical equipmentrequirements are as follows:

Single phase 220V ± 10%, 10A 50Hz/110v ± 10% The equipment lands the ground through the ground wire of the power

sheath, passing the wall line. Good ground connection is essential forsafe operation.

The input voltage can not have transients, voltage and current spikes,collapse or surge.

To ensure that the system is in the best condition, the room temperatureshould be maintained at about 20 ℃, relative humidity should notexceed 80% .The thermal power of the system is about 200W duringwork, so it’s the best if the treatment rooms are air-conditioned.

Do not allow the instrument being irradiated by prolonged directsunlight, nor being placed near a heater.

4-1-3 installation steps 1) Carefully open the package of the machine, place the machine in a preselected

position 2) Take out the handle bracket, and install it in the corresponding position

Page 8: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

Notice:When installing the handle, please do the

following thing as the pictures show.there are 2

springs beside each connector, the springs

should be press down before install the handle,

or the Rubber ring will be damaged. Don’t pull

Page 9: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

the handle during operation.

Take the Water funnel, insert the filler, inject 5 liter of water

Notice: When adding water, please open the VENTS in advance,add enough water until water spill out from the VENTS.

Handle cannot beinstalled in this case

Handle can beinstalled

Page 10: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

Put the filter cotton on the vacuum inlet mouth to protect the liquid be suckedinto the handle.

Insert the handhold emergency switch

Notice: please confirm the handhold emergency switch in“ON”station before operate the machine

Page 11: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

ON OFFPlug in the power cord

Insert the switch key and release the red emergency button

4-2 Debugging Adjustments4-2-1 Fulfill waterFirst fulfill distilled or purified water by the water inlet on the back of the

machine.Running water or mineral water is not allowed. Pls do not stopfulfilling until the water flow out from the over flow port (shown aspicture1). Then turn the key switch to the right, and observe whether theinterface of the treatment handle is leaking water or not. The water inside

Page 12: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

the machine should be replaced every month.After replacing the water, themachine should keep working for 3 minutes at first to observe if thewaterway is smooth.

Picture 1

4-2-2 DebuggingOpen the pulse oximeter, clamp the forefinger and wait 3 minutes to see if the displayed number

is normal. Less than 50 or more than 120 is not suitable for cryolipolysis treatments (if receiving

this test after a strong exercise, the result is not accurate and generally the number will be more

than 120). The heart rate test should be done in a calm state.

Switch the key to the right, the whole machine starts working. The touch screenwill be same as picture 2.

Page 13: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

Picture 2If picture2 appears, it indicates the machine’s starting is normal.

Cryolipolysis+vacuum refer to cryolipolysis slimming system.

Click , the system will enter into picture 3.

图 3

Use the fat measuring caliper to measure the fat thickness of the treated area.Based on the actual thickness, you can select cryolipolysis slimming mode, or

Page 14: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

enter into expert mode, adjusting the treatment parameters of the 2 handlesfreely.

Then click entering into picture 4. Under this mode, the set treatment

time and temperature are locked. Only pressure intensity can be adjusted.

Notes:The system has 3 minutes heating by default. The temperature will beno more than 40℃. 3 minutes later, the system will enter into cooling state

automatically. If you want to cancel the heating program, slightly click

turning into ,then the heating process will be cancelled.

Picture 4

After entering into picture 4, you can click A or B on the screen to control 2 handlesrespectively.

Adjust and display the intensity of vacuum, that’s pressure. It can

be adjusted by pressing and button; its adjusting scope is

0-100KPA.

Page 15: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

Adjusting treatment time, its adjusting scope is 1-60min (under

this mode, it cannot be adjusted).

Adjusting the set temperature of treatment handles, the

adjusting scope is -15-5℃.(under this mode, it cannot be adjusted)

The big font shows the real temperature of treatment handles;

the below small font shows the set temperature at the beginning.

After the suitable treatment parameters are adjusted, click into the

machine starts working.The system is in working state at this moment; the selected handle starts heatingwith vacuum. Put the handle on belly, you will have a feeling of being sucked.3minutes later, the handle starts cooling to the set temperature. Meanwhile, the 4LED lights change their colors from red to blue according to heating and cooling.

It means Cryolipolysis slimming system is working normally. Click , turning

the system to standby state like picture 5. Click the button on the

lower left of the screen, back to picture 3.

Choose , entering into manual mode. Under this mode,pressure, set temperature and treatment time all can be adjusted, like picture 5.

The other operations are the same as above.

Page 16: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

Picture 5

When end, click back to the home page, picture 2.Notes: When doing treatments under this mode, the clients should hold theemergencystop switch all the time. If feeling uncomfortable, they can press the switch,

andthe machine will stop working. Press it again, the machine will start working.

In the mode setting interface, slightly click button, the machine will enter

into vacuum calibration interface, like below picture shows.Notes:Do not enter into vacuum calibration interface randomly to do adjustments.If needed, pls contact our engineers.

Page 17: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

Water flow setting adjustments:Alarming water flow can be adjusted in the vacuum calibration interface. Slightly

touch the adjusting buttons of the two sides, which can adjust the

water flow alarming value, its setting scope is 5-10.

When the actual water flow is lower than the set alarming value, the machine will

warn you automatically by “Error 1”. After finish adjustments, press button

to store the adjusted parameters.

Vacuum calibration adjustments:

In this calibration interface, slightly touch into to start air pump.

Choose A1, open the back cover and adjust the corresponding knob to a certain value.

Similarly, touch A2 slightly and adjust A2 knob. Then you can adjust the others one

by one. After finish adjustments, slightly touch into , and then press back

button.Reference parameters are as below:

A1 A2 A3 A4 B1 B2 B3 B4ReferenceValue

56 37 31 25 56 37 31 25

Chapter 5 Technical parameters5-1、Technical parameters:

Input Power 1800WFrozen head pressure 0-100kpaCryolipolysis handle screen size 3.5 InchTreatment handle’s quantity 2 units( Optional)Touch screen size 10.4 InchFrozen head temperature -15-5℃Cooling System Semiconductor Cooling + water-cooling + air

coolingDimension 60*41*50 cmNet weight 34kgFuse specification Ø5×25 10APower AC220V±10%,10A,50HZ / AC110V±10%10A,

60HZ

Page 18: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

5-2、Related Request:The system running environment requirements:

Environment temperature: 10℃~30℃Relative humidity: less than 80%

The system storage environmental requirement:Environment temperature: 0℃~40℃Relative humidity: less than 80%

Operating Procedure6-1 Opening1. make sure that if the heart rate of patient is normal, and has no other diseasesthat not adapted to do this treatment.2. measure the fat thickness by using the fat measuring clamp.

3. clean the treatment parts (if where there is hair, remove it.)4. stick the antifreeze membrane on the treatment part.5. Put the Cling film on the membranes6. Open the machine to start the treatment

6-2 choosing Cryolipolysis mode1).Choose precise parameter to enter according to the thickness of the fat

Adjust the pressure, then click to start to work.

The handle must be near to the skin, and the button in the handle can controlworking and stopping.1) Note that in the same time course, If continue, previous process starts again afterthe first stop.

Page 19: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

If you want to start again begin heating the entire course of treatment, the need toreturn to the main screen and re-enter.2) During the treatment, the client should hold the emergency stop tightly. If feeluncomfortable, press the stop. The machine will stop working. Then press stop oncemore, the machine works.

6-3 turn off1. place back the treatment handle to its place.2. change working state into "standby".3. turn the key switch to turn off the system.4. clean the handle with clean cotton or soft cloth.

Note:Frozen fat cells disappear fat technology no invasive wound, after the majority ofpeople do not have any abnormal reaction, but the individual constitution is different,a few people applied parts will be a little red and swollen, but will disappear within afew hours, a few people may have a slight sense of paralysis, with slight frozen or thecondition of stasis table, this is a temporary phenomenon, a few days to weeks ordisappear, do not use special too anxious.

Chapter 7 Treatment Principle

Cryolipolysis Technology: As triglyceride in fats will be converted into solid inparticular low temperatures (below 5℃), It uses advanced cooling technology toselectively target fat bulges and eliminate fat cells through a gradual process thatdoes not harm the surrounding tissues, reduce unwanted fat, When fat cells areexposed to precise cooling, they trigger a process of natural removal that graduallyreduces the thickness of the fat layer. And The fat cells in the treated area are gentlyeliminated through the body's normal metabolism process, to eliminate unwantedfat.

Chapter 8 Treatment Setting

1-3 treatments per treatment course; do one treatment for 30-60 minute. After onetreatment, you can do the second treatment on the same area after 45-60 days. Thetreatment course, treatment time and temperature setting are depend on thethickness of the fat. You can set 1-3 treatments per treatment course. If your clientneed more treatment, please set the second treatment course after at least 6 months.

Notice:1, When using the cryolipolysis handles, you must put the anti free membrane on thetreatment area and also cover a piece a plastic membrane on it to protect the anti

Page 20: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

freeze liquid lossing. Or you may freeze and hurt the skin.2, when using the Cryolipolysis handles, you can do treatment on different areas in atreatment course, but don’t do repeat treatments on a same area within 45days. Forexample, you can do treatment on the abdomen, waist, buttock, thigh etc. in a sametreatment course, but don’t do treatment on the abdomen again in 45days.

Chapter 9 Indications and Contraindication

10-1 Indications

The system can be treated:

Body shaping: Shaping of the four limbs, buttocks, abdomen, recovery of Postpartumand liposuction surgery.Stretch marks: eliminate stretch marks efficientlyBody sliming: reduce the cellulite, reduce the fat, improve skin laxity

10-2 Contraindication

Pregnant womanThe patient who has the heart bypass surgery.Heart diseaseHypertension

Chapter 10 Post-treatment care10-1: The temperature of Cryolipolysis instrument must be controlled well, not toolow, otherwise it will still hurt the skin, after doing this treatment, please payattention to diet and do some exercise, so that it is not easy to rebound.10-2: Please pay attention to use -15——40℃ anti-membranes membranes.

Chapter 11 Maintenance and Repair11-1 Device CleaningClean the exterior surface of this machine with soft wet cotton at least once a week.Neutral cleaner could be used, but pls make sure no liquid flows into the machine.

The purified or distilled water inside the water pathway should be changed onceevery month.

11-2 Light-guide Crystal CleaningThe treatment handle should be always kept clean and should be cleaned after eachtreatment.

Page 21: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

When cleaning, first wipe it clean with a soft cloth, and then clean it with another softcloth moistened with water-free alcohol and let it dry thoroughly.

11-3 Pump filter CleaningAfter the machine works a period of time, if liquid is found inside the pump filter, youshould clean it timely.Unscrew the transparent container at the bottom of the filter, re-install the filter afterpouring the liquid inside.

12-4 TroubleshootingThe table below is a list of some possible faults and solutions accordingly.

12-5 Fault information

Faults Check parts

No powerCheck the power 220V inlet wire to seewhether the fuse wire is broken or not

No touch Re-calibrate the screen

Cooling temperature is lo Check power supply voltageCheck water circulation

Leaking waterbetween machine’s joints

Check rubber ring

If any other problems, please contact with our after-sale team instantly.Don’t take the machine apart please.

Right RotationTo open them

Page 22: Cryolipolysis Body Slimming Machine Nbw-c132-User Manual

Beijing Nubway S & T Co., Ltd

When starting the machine, or in the process of treatment, if some faults orabnormalities occur, the machine should stop working immediately. Then pleasecontact with our maintenance department, the technicians authorized by ourcompany could do troubleshooting only.

12-6 Replace the fuse wire:Turn off all power switch, unplug the power cord; open the fuse holder with a smallflathead screwdriver, and then take the fuse wire in the fuse holder. Take out thespare (15A / 250V) fuse wire in the accessories sent to you along with the machine.

12-7 Touch screen calibrationWhen turn on the machine, touch the touch screen 20 times in a position of the initialinterface, then the system will automatically enter the calibration interface. The LCDscreen turns blue after entering the calibration state. The upper left corner of the screenwill appear white "cross", touch the "cross", there will be "cross" in the upper right corner,doing the same operation, the lower right corner will appear "cross", Please touch it again.After finish the above three "cross", restart the treatment, touch-screen calibration isfinished.

Chapter 12Contact Us

Beijing Nubway S&T Co.LtdWeb: www.nubway.netEmail: [email protected]: Room 805 No.12,Yabao Rd Chaoyang District Beijing 100020 ChinaTel: 010-10-5120-6522Fax: 010-10-5120-6522