Upload
briana-malone
View
215
Download
0
Embed Size (px)
Citation preview
Conventional and Conventional and Computed TomographyComputed Tomography
IntroductionIntroduction
Is a radiographic technique that employs Is a radiographic technique that employs motion to show anatomical structures lying motion to show anatomical structures lying in the a plane tissue while blurring or in the a plane tissue while blurring or eliminating the detail in images of eliminating the detail in images of structures above and below the plane of structures above and below the plane of the interestthe interest
PrinciplePrinciple
The principle is based on synchronous The principle is based on synchronous movement of 2 or 3 elements in a movement of 2 or 3 elements in a tomographic system tomographic system
Tomographic units synchronize the Tomographic units synchronize the movements of the x-ray tube and the movements of the x-ray tube and the image receptor in opposite directions image receptor in opposite directions around stationary fulcrum (pivot point) around stationary fulcrum (pivot point) during the exposureduring the exposure
The fulcrum area is sharpThe fulcrum area is sharp
PrinciplePrinciple
The longer the blurring the less The longer the blurring the less opportunity to create a sharp imageopportunity to create a sharp image
The greater distance to the fulcrum the The greater distance to the fulcrum the greater the blurringgreater the blurring
The further the object is from the fulcrum The further the object is from the fulcrum the greater difference between its the greater difference between its projected motion of the image receptor projected motion of the image receptor and the motion of the image receptorand the motion of the image receptor
This cause image more sharpThis cause image more sharp
Tomographic QualityTomographic Quality
The tomographic amplitudeThe tomographic amplitude is the total is the total distance the tube travelsdistance the tube travels
The tomographic amplitude is equal or The tomographic amplitude is equal or greater to the exposure amplitudegreater to the exposure amplitude
Exposure amplitudeExposure amplitude is the total distance is the total distance the tube travels during the exposurethe tube travels during the exposure
BlurBlur is the smearing that results in the is the smearing that results in the loss of nearly all recorded detail of objects loss of nearly all recorded detail of objects outside the focal planeoutside the focal plane
Tomographic QualityTomographic Quality
Increased blurring causes decreased densityIncreased blurring causes decreased density It is effected by tomographic amplitude, It is effected by tomographic amplitude,
distance from the fulcrum, distance from the distance from the fulcrum, distance from the image receptor, and orientation of tube motionimage receptor, and orientation of tube motion
Distance from the fulcrumDistance from the fulcrum has a direct has a direct relationship to blur widthrelationship to blur width
Distance from the image receptorDistance from the image receptor has a has a direct relationship to blur widthdirect relationship to blur width
Tomographic QualityTomographic Quality
Orientation of tube motionOrientation of tube motion has a direct has a direct relationship to blur widthrelationship to blur width
FulcrumFulcrum controls the section level, and it controls the section level, and it may be fixed and patient can move up and may be fixed and patient can move up and down to change section leveldown to change section level
Focal planeFocal plane is the section is the section Section thicknessSection thickness is the width of the focal is the width of the focal
and its control by exposure angleand its control by exposure angle
Tomographic QualityTomographic Quality
Exposure angle inversely proportional to Exposure angle inversely proportional to section thicknesssection thickness
The tomographic images exhibit less The tomographic images exhibit less contrast than static imagecontrast than static image
Section Interval Section Interval is the distance between is the distance between fulcrum levelsfulcrum levels
Types of MotionTypes of Motion
LinearLinear it moving along a straight lineit moving along a straight line The major quality problem is SID and OIDThe major quality problem is SID and OID CurvilinearCurvilinear reduce magnification and reduce magnification and
maintain SID and OIDmaintain SID and OID CircularCircular EllipticalElliptical Figure eightFigure eight TrispiralTrispiral HypocyclodialHypocyclodial
Exposure FactorsExposure Factors
Time it must match the time of the x-ray Time it must match the time of the x-ray tube to complete tube to complete
Complex tomographic motion often require Complex tomographic motion often require 3-6 seconds3-6 seconds
mA usually used below 100 to set long timemA usually used below 100 to set long time 30% more mAs is needed for wider angle 30% more mAs is needed for wider angle
tomographytomography Zonography tomographs usually require Zonography tomographs usually require
the same mAs as static radiographthe same mAs as static radiograph
Exposure FactorsExposure Factors
kV 15% rule is a critical tool to determine kV 15% rule is a critical tool to determine the adjustmentthe adjustment
5% change in kVp is required to produce 5% change in kVp is required to produce visible densityvisible density
Zonography is a narrow angle tomography Zonography is a narrow angle tomography exposure amplitudes less than 10exposure amplitudes less than 10oo
Computed TomographyComputed Tomography
Scanner GenerationsScanner Generations
1- First generation1- First generation2- Second generation2- Second generation3- Third generation3- Third generation4- Fourth generation4- Fourth generation
GantryGantry
Is the movable frame of the CT unitIs the movable frame of the CT unit It contains the x-ray tube and the detectorsIt contains the x-ray tube and the detectors It maintain the alignment of the tube and the It maintain the alignment of the tube and the
detectors and contains the equipment detectors and contains the equipment necessary to perform the scanning movementnecessary to perform the scanning movement
The aperture for the patient is 50-58 cmThe aperture for the patient is 50-58 cm Do not force the obese patient into the Do not force the obese patient into the
aperture aperture
GantryGantry
Most gantrys can be angled up to 30Most gantrys can be angled up to 30oo
Table angulation can be sometimes be Table angulation can be sometimes be used in place of gantry angulationused in place of gantry angulation
Positioning lights are usually mounted on Positioning lights are usually mounted on the gantry intense white halogen lights and the gantry intense white halogen lights and low power laser lights low power laser lights
There are often 3 positioning lights for There are often 3 positioning lights for accurate sagittal, coronal, and transverse accurate sagittal, coronal, and transverse centering centering
X-ray TubesX-ray Tubes
CT images produce massive amounts of heat CT images produce massive amounts of heat in the x-ray tubein the x-ray tube
Focal spots size use 0.6 mm and 1.2 mmFocal spots size use 0.6 mm and 1.2 mm 0.6 mm use a pulsed beam to reduce the heat 0.6 mm use a pulsed beam to reduce the heat
loadload Modern CT pulsed scanner tubes operate at Modern CT pulsed scanner tubes operate at
120 kVp, 1-5 msec pulses120 kVp, 1-5 msec pulses X-ray tube produce 0.5-5 million heat unitX-ray tube produce 0.5-5 million heat unit Liquid and air-cooled tube housing design Liquid and air-cooled tube housing design
have been developed have been developed
X-ray TubesX-ray Tubes The radiation beam is double collimated to The radiation beam is double collimated to
assist in eliminating scatter informationassist in eliminating scatter information Slight misalignment can create ring artifact Slight misalignment can create ring artifact
imageimage The anode is aligned with their long axis The anode is aligned with their long axis
perpendicular to the scanner plane, this to perpendicular to the scanner plane, this to prevent heel effectprevent heel effect
Collimation controls voxel lengthCollimation controls voxel length Can be vary between 1mm to 13 mm usually Can be vary between 1mm to 13 mm usually
control by the software programcontrol by the software program The dimension width determine by section The dimension width determine by section
thickness or voxel lengththickness or voxel length
DetectorsDetectors
CT detectors should have:CT detectors should have: (a) high capture efficiency i.e. how well the (a) high capture efficiency i.e. how well the
detectors receive the photon from the detectors receive the photon from the patient and that can be control by patient and that can be control by detectors size and the distance between detectors size and the distance between detectorsdetectors
(b) high absorption efficiency i.e. how well (b) high absorption efficiency i.e. how well the detectors convert incoming photons, it the detectors convert incoming photons, it determine by the material used as well the determine by the material used as well the size and thickness of the detectorsize and thickness of the detector
DetectorsDetectors
(c) high conversion efficiency i.e. how well (c) high conversion efficiency i.e. how well the detector convert the absorb photon to the detector convert the absorb photon to analog or digital signalanalog or digital signal
CT detectors should have also high stability, CT detectors should have also high stability, fast response time, and wide dynamic range fast response time, and wide dynamic range which is the ratio between the largest signal which is the ratio between the largest signal to the smallest that can be measuredto the smallest that can be measured
Typical modern scanners are capable of Typical modern scanners are capable of dynamic range of 1,000,000:1 dynamic range of 1,000,000:1
ComputerComputer
Is design to control the data acquisition, Is design to control the data acquisition, process and display, and storageprocess and display, and storage
The computer should be in enclosed room The computer should be in enclosed room with controlled temperature and humiditywith controlled temperature and humidity
CT console provide the radiographer CT console provide the radiographer access to the software program that data access to the software program that data acquisition, controls data processing and acquisition, controls data processing and display and storage functions display and storage functions
ComputerComputer
A system program is used to start up the A system program is used to start up the CT unit, this program turns on and perform CT unit, this program turns on and perform quality assurance, and record various quality assurance, and record various problemsproblems
The CT console operate from the menu The CT console operate from the menu simply uses a keyboard, light pensimply uses a keyboard, light pen
Data AcquisitionData Acquisition
Controls the tube and detector collimation Controls the tube and detector collimation (pixel size), matrix size, gantry angle, table (pixel size), matrix size, gantry angle, table top entrance, section increment top entrance, section increment movement, kVp, mA, scan speedmovement, kVp, mA, scan speed
Display ConsoleDisplay Console
Controls the digital image production Controls the digital image production process, that compile the image and process, that compile the image and display parameters, such as window width display parameters, such as window width and leveland level
Exposure FactorExposure Factor
Most CT performed at 120 kVpMost CT performed at 120 kVp Time is not a factor as it must be Time is not a factor as it must be
controlled by scanning programcontrolled by scanning program mA should be setupmA should be setup Dual energy scanning units required Dual energy scanning units required
usually 80 kVp and 140 kVpusually 80 kVp and 140 kVp
ArtifactsArtifacts
MotionMotion Metal or StarMetal or Star Beam HardeningBeam Hardening Partial Volume EffectPartial Volume Effect Ring artifactsRing artifacts