Upload
others
View
7
Download
0
Embed Size (px)
Citation preview
Confocal characterization and imaging calibration, SIPcharts.
Confocal fluorescence microscopy: characterization of imaging and calibration using thin uniform fluorescence layers, SIPcharts.
G. J. Brakenhoff Swammerdam Institute for Life Sciences, University of
Amsterdam, Amsterdam, Kruislaan 316, 1098 SM Amsterdam, The Netherlands
E-mail: [email protected]
Varenna 12 Juli 2010 International School of Physics Enrico Fermi
Advanced methods in optical fluorescence microscopy towards nanoscopy
objectintensity
lmageintensity
Imaging by a lens
focalpoint
focalpoint
Shading or non-uniform intensity imaging:
focalpoint
focalpoint
objectintensity
lmageintensity
Vignettting:partial block of light rays
Image of uniform layer:
Lower intensity at the edges!
Problems in microscopy affecting quantitative imaging
Shading effects
Imaging conditions difficult to characterize
Imaging conditions difficult to reproduce
���� Non-quantitative intensity imaging
Present both Present both in 2D - regular microscopy
3D - confocal microscopy
Propose correction/calibration procedures based on
Thin uniform fluorescent layers
Sectioned Imaging Property or SIPcharts analysis INTENSITY CORRECTION AND NORMALIZATION INFLUORESCENCE CONFOCAL MICROSCOPY WITH
SECTIONED IMAGING PROPERTY CHARTS OR SIPCHARTS
J. M. Zwier #, G.W.H. Wurpel #, K. Jalink +, L. Oomen+, L. Brocks + & G. J. Brakenhoff #
#Swammerdam Institute for Life Sciences, Section of Molecular Cytology and Center ofAdvanced Microscopy, University of Amsterdam,, The Netherlands,
+ Division of Cell Biology, The Netherlands Cancer Institute, Amsterdam, TheNetherlands
0.01% fluorescing compound in polyvinylalcohol spin-coated on microscope slide.
Protected and sealed with a microscope cover glass
Calibration for 2D and 3D Fluorescence Microscopy
Basis :Spatially uniform fluorescing reference layer
Thickness 100 to 200 nm.Thickness uniform within a few %
Reference layers have to be to a high degree
spatially uniform and reproducible.
• various info
NN
O
O
O
O
concentration of
N,N'-bis(2,5-di-tert-butylphenyl)-3,4,9,10-N,N'-bis(2,5-di-tert-butylphenyl)-3,4,9,10-perylenebis(dicaboximide) in polystyrene/toluene
0.01 w% in polystyrene
concentration measured with absorption spectroscopy (ε(490nm) = 86000 l•mol•cm-1 ) -> A= 0.09 -> 3.0•106 mol/l -> in a layer of 100 nm thickness this leads to
1.8 •1010 molecules/cm2
Average distance between molecules about 100 nm.
No quenching effects!
• Absorption spectrum
• emission spectrum
350
300
250
200
150
100
50
x10
3
550500450400350
wavelength (nm)
abso
rptio
n c
oef
f. (a
.u)
160
120
80
40
0
x10
3
700650600550500
wavelength (nm)
inte
nsi
ty (
cou
nts
)
Fluorescence specimen image:
Ps (x,y) = I(x,y) . D(x,y) . Fs (x,y)
illumination detection specimen fluoresence
Reference layer image: taken under same conditions !
Pr(x,y) = I(x,y) D(x,y) Fr
Reference layer fluorescence : uniform
QuickTime™ and aTIFF (LZW) decompressor
are needed to see this picture.
QuickTime™ and aTIFF (LZW) decompressor
are needed to see this picture.
Arb. Intensity
Arb. Intensity
Ps (x,y)
P (x,y)Calibrated image Pc(x,y)
Ps(x,y F(x,y)Pc(x,y) = ------ = ------- ( F L U ’ S ) ------------ =
Pr(x,y) Fr
expressed in Fluorescence (of the reference) Layer Units
: ” F L U ’s ”
Actual imaging conditions as described by I(x,y) D(x,y) have dropped out !2D
(regular microscopy)
QuickTime™ and aTIFF (LZW) decompressor
are needed to see this picture.
FLU’s Intensity
QuickTime™ and aTIFF (LZW) decompressor
are needed to see this picture.
QuickTime™ and aTIFF (LZW) decompressor
are needed to see this picture.
Specimen: Lipid DPPCtwo phase monolayer segregation
Pr(x,y)
QuickTime™ and aTIFF (PackBits) decompressorare needed to see this picture.
Microscope20x image
Microscope 40 x image
Images of object
Images of
Shading correction
are needed to see this picture.
Images with different magnifications can be well correllated!
Images of reference layer
Calibrated images
2D(regular microscopy)
500
400
300
4000
3000
A 500
400
3001.21.0
B
before shading correction after shading correction
Application:Shading correction Specimen:
Lipid DPPC monolayer,two phases with low (LC) and high (LE) NPB fluorescencer concentration
300
200
100
0
6004002000
3000
2000
1000
x-pixel
y-pi
xel 300
200
100
0
6004002000
1.00.80.60.40.2
x-pixely-
pixe
l
1200
1000
800
600
400
200
0
40003000200010000 fluorescence intensity
c) 1200
1000
800
600
400
200
0
1.20.80.40.0relative fluorescence intensity
d)
Are the reference layers really uniform?
Image Reference image
Two images of the reference layer taken at two different locations are used one as image to be calibrated and the other as reference image.
Calibrated image
drawing
Referenceimage
Image
Proves that our fluorescent reference layers for the calibration that are to a high degree
spatially uniform
reference layer
2D(regular microscopy)
2DIm age calibra tion in fluore scen ce microscopyJ. M. Zwier, G . J. VAN R oo ij, J . W , Ho fst raa t & G . J. Brak enhof fJ. of Micr., 216 , 15 -24 , 2004
3D
Confocal fluorescence microscopy sectioned images
(Anaphase, Allium Cepa (onion), DNA stain: mitrmycine)
Integrated or Total Confocal Image
QuickTime™ and aTIFF (LZW) decompressor
are needed to see this picture.
Method: Making a z-stack (through focus) of the uniform polymer fluorescence reference layer (ca. 100 nm)
Sectioned image characterization with thin uniform fluorescent layers
Through focus stack
coverslip
reference layer
reference image
slide
Sectioned image characterization with thin uniform fluorescent layers
1
1 2
4
3 5
z
Through focus stack
x
z
z1
1 24
3 5
Axial Point Spread Function
Axial PSF
Through-focus stack of “ thin”uniform fluorescence reference layer
LaterallyIntegrated Intensity
z
yx
Sectioned image characterization with thin uniform fluorescent layers
0.00E+00
1.00E+02
2.00E+02
3.00E+02
4.00E+02
5.00E+02
6.00E+02
1 6
11 16 21 26 31 36 41 46 51 56 61 66 71 76 81 86 91 96
Series1
Series2
0.00E+00
1.00E+02
2.00E+02
3.00E+02
4.00E+02
5.00E+02
6.00E+02
7.00E+02
1 6
11 16 21 26 31 36 41 46 51 56 61 66 71 76 81 86 91 96
Series1
Series2
0.00E+00
1.00E+02
2.00E+02
3.00E+02
4.00E+02
5.00E+02
6.00E+02
1 6
11 16 21 26 31 36 41 46 51 56 61 66 71 76 81 86 91 96
Series1
Series2
0.00E+00
1.00E+02
2.00E+02
3.00E+02
4.00E+02
5.00E+02
6.00E+02
7.00E+02
1 6
11
16
21
26
31
36
41
46
51
56
61
66
71
76
81
86
91
96
Series1
Series2
0.00E+00
1.00E+02
2.00E+02
3.00E+02
4.00E+02
5.00E+02
6.00E+02
7.00E+02
Series1
Series2
1
43
5
2
I intensity
0.00E+00
1.00E+02
2.00E+02
3.00E+02
4.00E+02
5.00E+02
6.00E+02
7.00E+02
Series1
Series2
Series3
Series4
Series5
Confocal trough-focus response: local off axis responses
Super position plot axial PSF responses over image field
a)
b) c)
Axial PSF
zconfocal PSF 700nm and up
z
Sectioned image characterization with thin uniform fluorescent layers
LaterallyIntegrated Intensity
Layer thickness thin with respect to axial confocal dimensions.Essential !
fluorescer layer thickness 100nm
m max intensitys =a/b skew
a b
1
1 2
4
3 5 z
LaterallyIntegrated Intensity
How to characterize the spatial imaging conditions in confocal and (later) multi-photon microscopy?
Proposed parameters for charcterization:
0
5000000
10000000
15000000
20000000
25000000
30000000
35000000
40000000
45000000
0 20 40 60 80 100 120
Series1
zm axial position of maximum
fwhm resolution
a b
integrated intensity
40000000
45000000
m max intensitys =a/b skew
a b
Analysis of confocal imaging conditionsResolution variation over image
50 1.15
bin_fwhmAvg = 1.05 µm
SD = 0.048 µm
1
1 2
4
3 5
0
5000000
10000000
15000000
20000000
25000000
30000000
35000000
0 20 40 60 80 100 120
Series1
zm axial position of maximum
fwhm resolution
exp 230603 100nm layer pinhole 1 Airy
40
30
20
10
0
y
40200x
1.10
1.05
1.00
0.95µm
Beam scanning:
Confocal imaging determined by overlap of
illumination and detection distributions
50
40
30
20
10
0
y
40200x
8.8
8.4
8.0
7.6
x106
bin_i Avg = 8.20187e+06SD = 306548
x,y confocal intensity distribution
Sectioned image characterization with thin uniform fluorescent layers
1
1 2
4
3 5
m max intensitys =a/b skewBeam scanning:no aberration off-axis chromatic aberration
xxInt.
Int.
0
5000000
10000000
15000000
20000000
25000000
30000000
35000000
40000000
45000000
0 20 40 60 80 100 120
Series1
zm axial position of maximum
m max intensitys =a/b skew
fwhm resolution
a b
How to standardize the characterization of the spatial imaging conditions in confocal and (later) multi-photon microscopy?
Propose:Sectioned Imaging Property charts
Sectioned image characterization with thin uniform fluorescent layers
Sectioned Imaging Property charts or
SIPcharts
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ItotalAvg = 90.6395SD = 4.68024
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ImaxAvg = 79.9347SD = 10.0326
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
6150
6100
6050
6000
5950
5900
5850
zmaxAvg = 6012.89SD = 75.4975
60
50
40
1400
1300
fwhm Avg = 1026.68SD = 92.209
30000000
35000000
40000000
45000000
Imax max intensity
a b Axial PSF
skew : s =a + b
a − bskew = a − b
a + b Sectioned Imaging
Property chart
SIPchart
Sectioned image characterization with thin uniform fluorescent layers
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
-30
-20-10
0
1020
30
x10-3
skew Avg = -0.0488328SD = 0.0299356
40
30
20
10
0
y (
pix)
6040200
x (pix)
1200
1100
1000
900
12
10
8
6
4
2
0
Flu
ores
cenc
e In
t. (1
03 )
121086420
z (knm)
I z(16,16)
I z(16,48)
I z(32,32)
I z(48,16)
I z(48,48)
0
5000000
10000000
15000000
20000000
25000000
30000000
35000000
40000000
45000000
0 20 40 60 80 100 120
Series1
Zmax axial position of maximum
Imax max intensity
fwhm resolution
a b
Itot integrated intensityunder axial PSF
AxialPSF
skew : s =a + b
a − b
�SIP Chart SectionedImagingPropertyChart
Confocal System 1 Obj. 63x, oil, 1,4 NA pinhole 1 Airy NKI 12-08-04 Bin 8 from 512*512
0
5000000
10000000
15000000
20000000
25000000
0 20 40 60 80 100 120
Series1
Zmax axial position of maximum
fwhm resolution
Itot integrated intensityunder axial PSF
PSF SIPchart
Analysis 3D image formation with SIPcharts
Sectioned image characterization with thin uniform fluorescent layers
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ItotalAvg = 90.6395SD = 4.68024
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ImaxAvg = 79.9347SD = 10.0326
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
6150
6100
6050
6000
5950
5900
5850
zmaxAvg = 6012.89SD = 75.4975
60
501400
fwhm Avg = 1026.68SD = 92.209
4000 SIP Chart SectionedImagingPropertyChart
Confocal System 1
Comparing confocal systems, system1
Sectioned image characterization with thin uniform fluorescent layers
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
-30
-20-10
0
1020
30
x10-3
skew Avg = -0.0488328SD = 0.0299356
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
1300
1200
1100
1000
900
12
10
8
6
4
2
0
Flu
ores
cenc
e In
t. (1
03 )
121086420
z (knm)
I z(16,16)
I z(16,48)
I z(32,32)
I z(48,16)
I z(48,48)
Point params bin_avg bin_sd
0 I_max 79.9347 10.0326
1 z_max 6012.89 75.4975
2 fwhm 1026.68 92.209
3 skew -0.0488328 0.0299356
4 I_total 90.6395 4.68024
3000
2000
1000
0
fwhm
(nm
)
Confocal System 1Obj . 63x, oil, 1,4 NA pinhole 1 AiryNKI 12-08-04 Bin 8 from 512*512
0
5000000
10000000
15000000
20000000
25000000
30000000
35000000
40000000
45000000
0 20 40 60 80 100 120
Series1
Zmax axial position of maximum
Imax max intensity
fwhm resolution
a b
Itot integrated intensityunder axial PSF
Axial PSF
skew : s = a + b
a − b
Comparing confocal systems, system 2
60
40
20
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ItotalAvg = 81.7096SD = 9.1452
60 1350
fwhm Avg = 1194.62SD = 66.1558
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ImaxAvg = 83.4493SD = 8.27537
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
6000
5500
5000
4500
4000
3500
zmaxAvg = 4656.19SD = 666.415
4000 SIP Chart SectionedImagingPropertyChart
Confocal System 2
Sectioned image characterization with thin uniform fluorescent layers
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
1350
1300
1250
1200
1150
1100
105030
0020
0010
000
fwhm
(nm
)
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
-0.15
-0.10
-0.05
0.00
0.05
0.10
0.15
skew Avg = 0.0526943SD = 0.0480111 200
150
100
50
0
Flu
ores
cenc
e In
t. (1
03 )
80006000400020000
z (nm)
I z(16,16)
I z(16,48)
I z(32,32)
I z(48,16)
I z(48,48)
Point params bin_avg bin_sd
0 I_max 83.4492 8.27537
1 z_max 4656.19 666.415
2 fwhm 1194.62 66.1558
3 skew 0.0526943 0.0480111
4 I_total 81.7096 9.1452
0
5000000
10000000
15000000
20000000
25000000
30000000
35000000
40000000
45000000
0 20 40 60 80 100 120
Series1
Zmax axial position of maximum
Imax max intensity
fwhm resolution
a b
Itot integrated intensityunder axial PSF
Axial PSF
skew: s =a + b
a − b
Confocal System 2Obj . 63x, oil, 1,4 NA pinhole 1 AiryCLSM 04-05-04 no4 Bin 8 from 512*512
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ItotalAvg = 90.6395SD = 4.68024
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ImaxAvg = 79.9347SD = 10.0326
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
6150
6100
6050
6000
5950
5900
5850
z maxAvg = 6012.89SD = 75.4975
60
50
40
y (
pix) 10
20
30
x10
skew Avg = -0.0488328SD = 0.0299356
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
1400
1300
1200
1100
1000
900
fwhm Avg = 1026.68SD = 92.209
12
10
8
Flu
ores
cenc
e In
t. (1
03 )
I z (16,16)
I z (16,48)
I z (32,32)
I (48,16)
Point params bin_avg bin_sd
0 I_max 79.9347 10.0326
1 z_max 6012.89 75.4975
2 fwhm 1026.68 92.209
3 skew -0.0488328 0.0299356
4 I_total 90.6395 4.68024
4000
3000
2000
1000
0
fwhm
(nm
)
SIP Chart SectionedImagingPropertyChart
Confocal System 1Obj . 63x, oil, 1,4 NA pinhole 1 AiryNKI 12-08-04 Bin 8 from 512*512
30000000
35000000
40000000
45000000
Imax max intensity
a b
Axial
skew : s = a + b
a − b
60
40
20
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ItotalAvg = 81.7096SD = 9.1452
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
1350
1300
1250
1200
11501100
1050
fwhm Avg = 1194.62SD = 66.1558
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ImaxAvg = 83.4493SD = 8.27537
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
6000
5500
5000
4500
4000
3500
zmaxAvg = 4656.19SD = 666.415
4000
3000
2000
1000
0
fwhm
(nm
)
60
50
40
y (
pix)
0.00
0.05
0.10
0.15
skew Avg = 0.0526943SD = 0.0480111 200
150
100
Flu
ores
cenc
e In
t. (1
03 )
I z(16,16)
I z(16,48)
I z(32,32)
I z(48,16)
I (48,48)
Point params bin_avg bin_sd
0 I_max 83.4492 8.27537
1 z_max 4656.19 666.415
2 fwhm 1194.62 66.1558
3 skew 0.0526943 0.0480111
4 I_total 81.7096 9.1452
20000000
25000000
30000000
35000000
40000000
45000000
Series1
Imax max intensity
a b
Axial PSF
skew: s = a + b
a − b
SIP Chart SectionedImagingPropertyChart
Confocal System 2Obj . 63x, oil, 1,4 NA pinhole 1 AiryCLSM 04-05-04 no4 Bin 8 from 512*512
Comparing confocal systems
Sectioned image characterization with thin uniform fluorescent layers
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
1400
1300
1200
1100
1000
900
fwhm Avg = 1026.68SD = 92.209
60
50
40
30
20
10
0y
(pi
x)
6040200
x (pix)
1350
1300
1250
1200
1150
1100
1050
fwhm Avg = 1194.62SD = 66.1558
30
20
10
0
y (
pix)
6040200
x (pix)
-30
-20-10
0
x10-3
6
4
2
0
Flu
ores
cenc
e In
t. (1
0
121086420
z (knm)
I z (48,16)
I z (48,48)0
5000000
10000000
15000000
20000000
25000000
0 20 40 60 80 100 120
Series1
Zmax axial position of maximum
fwhm resolution
Itot integrated intensityunder axial PSF
PSF 30
20
10
0
y (
pix)
6040200
x (pix)
-0.15
-0.10
-0.05
0.00
50
0
Flu
ores
cenc
e In
t. (1
0
80006000400020000
z (nm)
I z(48,48)
0
5000000
10000000
15000000
0 20 40 60 80 100 120
Zmax axial position of maximum
fwhm resolution
Itot integrated intensityunder axial PSF
Comparison resolution
4000
3000
2000
1000
0
fwhm
(nm
)
4000
3000
2000
1000
0
fwhm
(nm
)
System 1 System 2
Comparison of lenses:Overview from 4 SIPcharts each taken with a different 63x objective
Sectioned image characterization with thin uniform fluorescent layers
Conclusion: easy to spot differences between lenses: quality control!
160
140
120
100
80
60
40
20
0
x10
3
806040200
z1616 z1648 z3232 z4816 z4848
150
100
50
0
x10
3
806040200
z1616 z1648 z3232 z4816 z4848
60
50
40
30
20
10
0
y
6040200
x
12.0
11.0
10.0
bin_fwhm Avg = 10.5374SD = 0.551831
60
50
40
30
20
10
0y
6040200
x
22
20
18
16
14
12
bin_fwhm Avg = 16.844SD = 2.11699
objective 2 phase contrast63x oil immersion170204 fred012
objective 163x oil immersion170204 fred010
4 Objectives: resolution at pinhole 1 Airy equiv.
160
140
120
100
80
60
40
20
0
x10
3
806040200
z1616 z1648 z3232 z4816 z4848
140
120
100
80
60
40
20
0
x10
3
806040200
z1616 z1648 z3232 z4816 z4848
60
50
40
30
20
10
0
y
6040200
x
11.0
10.0
9.0
8.0
bin_fwhm Avg = 9.27375SD = 0.640423
60
50
40
30
20
10
0
y
6040200
x
11
10
9
8
7
bin_fwhm Avg = 7.90227SD = 0.738263
x
objective 363x oil immersion170204 fred013
objective 4 100 x oil immersion170204 fred014
Confocal course with partial beam block
Leica TCSNA= 1.463Xwith partial beam block
lmageintensity
with partial block
Confocal course without partial beam block
Leica TCSNA= 1.463XWithout partial beam block
wiih partialblock removed
lmageintensity
Confocal course function NA
QuickTime™ and aTIFF (LZW) decompressor
are needed to see this picture.
Leica SP2-2 NA= 1.2
Confocal course function NA
Leica SP2-4 NA= 1.4100X
Sipchart_before_cleaning objective
Sipchart_after_cleaning objective
Spinning Disk -1.1.pdf
Dependence of imaging on proper immersion oil.
At temperatures for live cell imaging –37 C-regular oils provide not optimal imaging.
Sipchart representations show:Total photon yield is about 23% higher in the DF2 oil for 37 C then for the regular DF23 and Leica oils.then for the regular DF23 and Leica oils.
With proper immersion oil: resolution is improved together with imaging field uniformity
Dependence of imaging on proper immersion oil.Imaging at 37C with standard immersion oil.
Dependence of imaging on proper immersion oil.Imaging at 37C with immersion oil for 37 C.
60
40
20
0
y (
pix)
6040200
x (pix)
6100
6000
5900
zmaxAvg = 5985.81SD = 87.651
60
40
20
0
y (
pix)
6040200
x (pix)
5900
5800
5700
5600
zmax Avg = 5767.09SD = 79.6776
Imaging wavelengths: Imaging wavelengths :
Axial image plane dependence on wavelength settings
Sectioned image characterization with thin uniform fluorescent layers
Imaging wavelengths: ext: λ 488, detection: λ bp 505-530
Imaging wavelengths : exitation: λ 543, detection: λ 560-615
Shift axial image plane with wavelength settings
60
40
20
0
pix
6040200
pix
z-max(b-a)
240
200
160
Avg= 218.723SD= 18.7466
nm
Important for co-localizationand FRET etc. studies!
Multi-excitation, multi-detector confocal microscope
variablepinhole
detectionchannels
(PMT) dichroics
computer
display
lasers
scanmirror
dichroic
pinhole
objective
pinhole
BRAKENHOFF, G. J., WURPEL, G. W. H., JALINK, K., OOMEN, L., BROCKS, L. &ZWIER, J. M. (2005)Characterization of sectioning fluorescence microscopy with thin uniformfluorescent layers: Sectioned Imaging Property or SIPcharts.Journal of Microscopy 219 (3), 122-132.
3D characterization reference:
Sectioned image characterization with thin uniform fluorescent layers
Next:
Sectioning microscopy image Sectioning microscopy image calibrationcalibration
• SIPchart based relative and absolute image calibration
– Relative image calibration• uniform sensitivity of the lateral image field (= shading compensation)• images can be related by expressing their intensities in
fluorescence layer units (FLU’s)
Calibrated image Pc(x,y)
– Examples:• fluorescent beads• shading correction BPAE cells
Bovine pulmonary artery endothelial cells.
• comparison of images taken with different magnifications
Ps(x,y) F(x,y) . PSF(x,y) F(x,y)Pc(x,y) = ------ = --------------------- = --------- ( F L U ’s )
Pr(x,y) Fr . PSF(x,y) Fr
Shading correction, beads
I total SIPchartbased
• SIPchart based relative and absolute image calibration– Using the SIPchart data to do correction of the total intensities of image
stacks• proof of principle on fluorospheres
15
12
3.0
2.5
coun
ts)
1 3
befo
re s
hadi
ng c
orre
ctio
n
afte
r s
hadi
ng c
orre
ctio
n
12
9
6
3
0
15129630I (total) field 2 ( Irel)
I (to
tal)
field
1 (
Ire
l)2.5
2.0
1.5
1.0
0.5
0.0
3.02.52.01.51.00.50.0
I(to
tal)
field
1 (
x10
6 co
unts
)
I(total) field 2 (x106
counts)
0 1
2
3
4
56
78
befo
re s
hadi
ng c
orre
ctio
n
afte
r s
hadi
ng c
orre
ctio
n
Calibration imaging with different magnifications / NA
Reference layer image Pr(x,y)
from I total SIPchart!
I total SIPchart’s
Before calibration
from I total SIPchart!
NA 1.2540x
Planapo
NA 1.3263x
Planapo
SIPchart’s
after calibrationin FLU’s
Calibration between two confocal microscopes: • SP2 Leica system• NT Leica system
– 63x objectives (1.32 NA) (exc 488 nm)– Only detection filters were somewhat different: NT - lp530 nm
SP2 - +530 nm
500
4001.0
nt-microscope (63x, 1.32 NA)500
4001.0
sp2-microscope (63x, 1.32 NA)
Calibration imaging with between different microscopes
NT SP2300
200
100
0
5004003002001000
0.8
0.6
0.4
0.2
0.0
FLU's
300
200
100
0
5004003002001000
0.8
0.6
0.4
0.2
0.0
FLU's
• After calibration (FLU intensity scales are comparable)
• Intensities within 30%! > Small differences in spectral filtering conditions
– object image:
– reference layer image:
Fobj (x,y ) = QE obj * σref .Nobj (x,y ).Peff (x,y )
fluorescence quantum yield
absorption cross section
number of molecules
eff,PSF
500
400
300
200
100
0
5004003002001000
imagestack-bg(counts)
160x103
120
80
40
0
500
400
I total (counts)
140x103
SIPchart based absolute image calibrationSIPchart based absolute image calibration
– reference layer image:
– ratio image:
Fref (x ,y ) = QE ref * σref .N ref .Peff (x ,y )
Fobj(x,y)
Fref (x,y)= QEobj
QEref.σobj
σref.Nobj(x ,y)
Nref.
300
200
100
0
5004003002001000
140x101301201101009080
Nobj(x,y) = 1.7∗104.Fobj(x,y)
Fref (x,y)mol /µm2
SIPchart based absoluteimage calibration
–ratio image:
then:
Fobj(x,y)
Fref (x,y)= QEobj
QEref.σobj
σref.Nobj(x ,y)
Nref.
SIPchart based absolute image calibrationSIPchart based absolute image calibration
Used parameter values:QE (BodipyFL) = 0.8 QE (dtbpd ref layer) = 0.95 dtbpd (488 nm) : ε = 84000 l.mol-1cm-1=> σ = 3.2 * 10-16 cm 2
bodipyFL (488 nm) : ε = 60000 l.mol-1cm-1 => σ = 2.3 * 10-16 cm (σ = ε * ln 10/ NA = 3.825 * 10-21 * ε)
Nref = 1.0 *104 molecules / µm2
• SIPchart based relative and absolute image calibration
500
400
molecules bodipyFL
(micrometer-2
)
Nobj(x,y) = 1.7∗104.Fobj(x,y)
Fref (x,y)mol /µm2
SIPchart based absolute image calibrationSIPchart based absolute image calibration
400
300
200
100
0
5004003002001000
(micrometer-2
)
2.0x104
1.5
1.0
0.5
0.0
Total number of BodipyFl
fluorophores in the cell:
4.7 * 107
100µm
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ItotalAvg = 90.6395SD = 4.68024
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ImaxAvg = 79.9347SD = 10.0326
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
6150
6100
6050
6000
5950
5900
5850
z maxAvg = 6012.89SD = 75.4975
60
50
40
y (
pix) 10
20
30
x10
skew Avg = -0.0488328SD = 0.0299356
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
1400
1300
1200
1100
1000
900
fwhm Avg = 1026.68SD = 92.209
12
10
8
Flu
ores
cenc
e In
t. (1
03 )
I z (16,16)
I z (16,48)
I z (32,32)
I (48,16)
Point params bin_avg bin_sd
0 I_max 79.9347 10.0326
1 z_max 6012.89 75.4975
2 fwhm 1026.68 92.209
3 skew -0.0488328 0.0299356
4 I_total 90.6395 4.68024
4000
3000
2000
1000
0
fwhm
(nm
)
SIP Chart SectionedImagingPropertyChart
Confocal System 1Obj . 63x, oil, 1,4 NA pinhole 1 AiryNKI 12-08-04 Bin 8 from 512*512
30000000
35000000
40000000
45000000
Imax max intensity
a b
Axial
skew : s = a + b
a − b
60
40
20
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ItotalAvg = 81.7096SD = 9.1452
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
1350
1300
1250
1200
11501100
1050
fwhm Avg = 1194.62SD = 66.1558
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
100
90
80
70
60
50
40
ImaxAvg = 83.4493SD = 8.27537
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
6000
5500
5000
4500
4000
3500
zmaxAvg = 4656.19SD = 666.415
4000
3000
2000
1000
0
fwhm
(nm
)
60
50
40
y (
pix)
0.00
0.05
0.10
0.15
skew Avg = 0.0526943SD = 0.0480111 200
150
100
Flu
ores
cenc
e In
t. (1
03 )
I z(16,16)
I z(16,48)
I z(32,32)
I z(48,16)
I (48,48)
Point params bin_avg bin_sd
0 I_max 83.4492 8.27537
1 z_max 4656.19 666.415
2 fwhm 1194.62 66.1558
3 skew 0.0526943 0.0480111
4 I_total 81.7096 9.1452
20000000
25000000
30000000
35000000
40000000
45000000
Series1
Imax max intensity
a b
Axial PSF
skew: s = a + b
a − b
SIP Chart SectionedImagingPropertyChart
Confocal System 2Obj . 63x, oil, 1,4 NA pinhole 1 AiryCLSM 04-05-04 no4 Bin 8 from 512*512
Comparing confocal systems
Sectioned image characterization with thin uniform fluorescent layers
60
50
40
30
20
10
0
y (
pix)
6040200
x (pix)
1400
1300
1200
1100
1000
900
fwhm Avg = 1026.68SD = 92.209
60
50
40
30
20
10
0y
(pi
x)
6040200
x (pix)
1350
1300
1250
1200
1150
1100
1050
fwhm Avg = 1194.62SD = 66.1558
30
20
10
0
y (
pix)
6040200
x (pix)
-30
-20-10
0
x10-3
6
4
2
0
Flu
ores
cenc
e In
t. (1
0
121086420
z (knm)
I z (48,16)
I z (48,48)0
5000000
10000000
15000000
20000000
25000000
0 20 40 60 80 100 120
Series1
Zmax axial position of maximum
fwhm resolution
Itot integrated intensityunder axial PSF
PSF 30
20
10
0
y (
pix)
6040200
x (pix)
-0.15
-0.10
-0.05
0.00
50
0
Flu
ores
cenc
e In
t. (1
0
80006000400020000
z (nm)
I z(48,48)
0
5000000
10000000
15000000
0 20 40 60 80 100 120
Zmax axial position of maximum
fwhm resolution
Itot integrated intensityunder axial PSF
Comparison resolution
4000
3000
2000
1000
0
fwhm
(nm
)
4000
3000
2000
1000
0
fwhm
(nm
)
System 1 System 2
How to make a SIPchart?
• Web-application– Accepts large 3D data sets– Automatic generation of SIPcharts– SIPchart delivered to user via
website:
www.SIPchart.org
www.SIPchart.com
Sipchart creation
Get a calibration layer.Collect 3D data set from thiscalibration layerGo to:
Click restricted area:
• We-application:– uplboad page:
• Webapplication– Return email:
• Webapplication:– SIPchart
Sipchart creationYou need a uniform fluorescence calibration layer
Send me an email at: [email protected] can then send you a layer. For free or else?
The “layer project” has ended succesfully.No real grounds for a continuation grant.
If the layers are usefull for maintaining and calibrating microscope performancethey should be of some value to people.
I want to keep the project going by inviting people to buy a few layers.comes with unlimited website access to produce SIPcharts.Euro 175 each?
Money goes to the University, to pay people etc.
20071113-14094813553
Sipchart Double Layer.pdf
20071113-141826-13644Sipchartbollen.pdf
Application:Generation of loacl 3D PSF’sCorrection of datasets acquired at different wavelengths for chromatic axial and lateral image shifts
Applications SIPcharts :
Testing and Evaluation , Optimization and alignmentMaintenance of confocal systems
Co-localization studiesidentify axial image plane shifts with spectral settings
Suitable basis for more detailed analysis
of sectioned imaging systems
Suitable basis for more detailed analysis
Correction of sectioned imaging first order correction fluorescence intensity variations
optimization of de-convolution algorithms
Basis for: absolute calibration of sectioned imaging
Email : [email protected]
Close