Upload
p-oropesa
View
217
Download
2
Embed Size (px)
Citation preview
Applied Radiation and Isotopes 59 (2003) 383–387
ARTICLE IN PRESS
*Correspond
E-mail addr
0969-8043/$ - se
doi:10.1016/j.ap
Comparisons of activity measurements withradionuclide calibrators
P. Oropesaa,*, A.T. Hern!andeza, R. Serraa, E. Mart!ıneza, C. Varelab
aCentro de Is !otopos (CENTIS), P.O. Box 3415, San Jos!e de las Lajas, Havana, CubabCentro de Control Estatal de Equipos M!edicos (CCEEM), Calle 4, No 455 (altos), Vedado CP 10400, Havana, Cuba
Received 4 August 2003; received in revised form 2 September 2003; accepted 3 September 2003
Abstract
The correct administration to a patient of the prescribed activity of a radiopharmaceutical is an important factor to
ensure the confidence in the diagnosis or the therapeutic efficiency, while at the same time keeping the unnecessary
human exposure as low as possible.
Comparisons of activity measurements for 131I, 201Tl and 99mTc with radionuclide calibrators were organized the first
time in Cuba during 2002 with the aim of obtaining information about the quality of administration of
radiopharmaceuticals. Ten Cuban nuclear medicine departments and the laboratories involved in the production of
these kinds of compounds participated in the comparison runs. The results presented in this paper facilitated the
identification of several problems and initiated corrective actions. In addition, they indicate the necessity of establishing
Quality Systems in nuclear medicine in Cuba.
r 2003 Elsevier Ltd. All rights reserved.
Keywords: Radionuclide calibrator; Activimeter; Comparison; Activity measurements; Nuclear medicine
1. Introduction
Before a radiopharmaceutical is administered to a
patient, its activity should be determined on the basis of a
measurement. The correct administration to a patient of
the prescribed activity is an important factor to ensure the
confidence in the diagnosis or the therapeutic efficiency,
while also keeping unnecessary human exposure as low as
possible. In the case of children subjected to a diagnostic
procedure or who undergo therapy with radionuclides, the
administration of well-know activities is of the utmost
importance. That is why regulations and limiting values
for the uncertainties of activity measurements have been
prescribed (European Pharmacopoeia, 1982; Gu!ıa, 2002).
An upper limit of 10% has been established for the
expanded uncertainty of the activity value, with a
coverage factor, k; equal to 3. Larger uncertainties on
ing author.
ess: [email protected] (P. Oropesa).
e front matter r 2003 Elsevier Ltd. All rights reserve
radiso.2003.09.001
these administrations may lead to overdoses or under-
doses, both undesirable due to their high risk-to benefit
ratio, or their risk without any benefits, respectively. In
addition, they could make the comparability of the results
obtained in various hospitals questionable.
During 2002, the Radionuclide Metrology Depart-
ment of the Isotope Center (CENTIS-DMR) has
organized several comparisons with various radio-
nuclides in order to obtain information on the quality
of the activity measurements during production and
administration of radiopharmaceuticals in Cuba. The
participants were expected to measure radioactive
solutions in glass vials and in syringes, to simulate the
administration of a radiopharmaceutical to a patient,
and to report on their instrument measurement and
administration procedures. The comparisons results,
shown in this paper, have uncovered several problems
and initiated corrective actions. Moreover, they reflect
the necessity of establishing Quality Systems in nuclear
medicine in Cuba.
d.
ARTICLE IN PRESSP. Oropesa et al. / Applied Radiation and Isotopes 59 (2003) 383–387384
2. Comparisons
Comparisons focussed on providing information on
the performance of the Nuclear Medicine departments
and the two laboratories involved in the production of
radiopharmaceuticals in Cuba, measuring the activity
of 131I, 201Tl and 99mTc with radionuclide calibrators.
The aim of the comparisons was to contribute to the
improvement of the quality of the measurements made
in the country with that type of instrument. Particularly,
these exercises checked not only the instruments, but
also the procedures used in the routine determinations
and the performance of the personnel involved in the
measurements.
The programme used for the comparisons was the
known-value scheme (ISO/IEC, 1997). The radioactive
samples with a known radionuclide activity were sent to
the participants. The samples sent to a participant were
measured only by that participant. Therefore, it is
possible to evaluate the capability of an individual
laboratory to measure the activity of the sample and to
provide numerical results for comparison with the value
assigned at CENTIS-DMR.
Glass vials and 5ml capacity plastic syringes of the
type routinely used in nuclear medicine were filled with
single nuclide solutions (the glass vials containing 5ml
and the syringes containing approximately 2.5ml of the
radioactive solution). The reference activity of each
individual sample was established as the resulting value
of its measurements in an activimeter, that had been
previously calibrated at CENTIS-DMR for the specific
nuclides and sample geometries. The reference sources
for the calibration of this activimeter were prepared in
appropriate geometries by weighing aliquots from
solutions calibrated using a germanium spectrometer.
For each radionuclide of interest, the gamma-spectro-
metry determination of the activity concentration of the
solution, is carried out with n (usually eight–ten) point
sources prepared by weighing 20/50-mg drops, with the
mass of each drop being determined by the pychnometer
method (Oropesa et al., 2002). Traceability to SI units is
guaranteed by using point gamma sources certified by
the Hungarian National Office of Measures (OMH) for
the efficiency calibration of the germanium spectro-
meter. Finally, the activity measurements of the
activimeter are traced with a relative combined standard
uncertainty of 2% to the calibrated gamma spectro-
meter. All 131I and 201Tl comparison samples were also
measured in this activimeter at CENTIS-DMR after
they were received back from the participants. This type
of check was not possible for the 99mTc samples because
of the short half-life of that nuclide. As a conclusion of
these controls and the results of previous stability
studies (Oropesa et al., 2002), all samples used for the
comparisons were considered stable enough for the
evaluation of the participant’s performance.
Ten Cuban nuclear medicine departments and the two
laboratories involved in the production of radiophar-
maceuticals participated in the comparisons. They were
asked to apply their established routine procedures for
measurements in the radionuclide calibrator and for the
activity administration to the patient. The selection of
the geometries and nuclides to be sent to a particular
participant was done in correspondence with the ones
used in routine activity measurements at the respective
institution. Specifically, samples in a glass vial geometry
were distributed only to the two laboratories involved in
the production of radiopharmaceuticals. The nuclear
medicine departments were provided with 3 radioactive
samples of each nuclide used by them, two of these
samples in the glass vial geometry and another one in the
syringe geometry. All participants were asked to
determine the activity in the received vials and syringes.
Supplementary information was requested about the
instrument and measuring procedures used.
Table 1 summarizes the results of the comparisons for
the determination of the activity of 131I, 201Tl and 99mTc
in the glass vials and syringes. Xmed and Xref are the
activity values (related to the same reference date)
reported by the participants and measured at CENTIS-
DMR, respectively. As had been expected, the results for
the vials and syringes are in good agreement, as the
attenuation of the radiation at a photon energies of 140
and 364 keV depends only slightly on the materials and
thickness of the sources and their containers (Debertin
and Schrader, 1992). In the 131I and 201Tl comparisons,
80% or more of the participants obtained results that
agreed within about 10% with the value measured at
CENTIS-DMR and thus their performance can be
evaluated as satisfactory. Particularly, only two nuclear
medicine laboratories, one during the 131I comparison,
and another during the 201Tl comparison reported
unsatisfactory results. The unsatisfactory performance
of these two laboratories was caused by the same error:
the wrong choice of the switch-selectable radionuclide
option during the sample measurements. There were no
available written procedures for any of the radionuclide
calibrator measurements at either of these two labora-
tories.
During the 99mTc comparison, 100% of the partici-
pating laboratories reported results that agreed to within
10% of the value measured at CENTIS-DMR for both
glass vial and syringe. This comparison was organized at
the end of 2002, after the results of the previous two
comparisons (131I and 201Tl) had been discussed with the
participants.
Thirteen radionuclide calibrators from 5 different
manufactures were used during comparisons, one of
which predominated (see Table 2). The comparison
results demonstrated that the calibration condition of
all of the calibrators used was satisfactory. Here is
necessary to point out that, at the beginning of the 2002
ARTICLE IN PRESS
Table 2
Manufacturers of the radionuclide calibrators used during
comparisons
Manufacturer No. of instruments
PTW–Freiburg 8
Biodex Medical Systems 2
Victoreen 1
Alfanuclear 1
Capintec 1
Table 1
Results of the comparisons for the determination of the activity of 131I, 201Tl and 99mTc in the glass vials and syringes
Nuclide 131I 201Tl 99mTc
No. of participants 9 5 9
Percentage of participants reporting satisfactory results 88.9% 80% 100%
1. Glass vials
No. of reported values 17 8 18
Percentage of results with Xmed � Xrefj j=0:033Xref p3 88.2% 75% 100%
2. Syringes
No. of reported values 5 3 8
Percentage of results with Xmed � Xrefj j=0:033Xref p3 100% 66.7% 100%
Xmed-activity reported by the participant; Xref-activity measured at CENTIS-DMR.
Table 3
Observed differences between the installed by the producer, Ai ;and the determined by the CENTIS-DMR, Ai0; calibrationcoefficients for the PTW-Freiburg’s instruments, under the
specific conditions used in the laboratories
Nuclide Source
geometry
No. of
results
Mean of
Ai=Ai0
Standard
deviation
131I Glass vial 6 0.968 0.003
Syringe 7 0.984 0.009
201Tl Glass vial 6 0.879 0.007
Syringe 6 0.879 0.007
99mTc Glass vial 7 0.912 0.002
Syringe 6 0.892 0.007
P. Oropesa et al. / Applied Radiation and Isotopes 59 (2003) 383–387 385
all instruments were calibrated by the CENTIS-DMR
for the nuclides and geometries used in their routine
activity measurements. During these calibrations per-
formed under the specific conditions used in the nuclear
medicine departments, as can be seen from Table 3,
cases of significant differences were found between the
calibration coefficients installed by the producer, Ai; andthose determined by the CENTIS-DMR, Ai0; for PTW-
Freiburg instruments. A lead shield was used for the
reference calibration in the manufacture and the
chamber is employed without any optional shield in all
eight laboratories with this type of chamber. Therefore,
the significantly low amount of backscatter could
explain the obtained less than 1 values for the quotient
Ai=Ai0: Although this fact was noted and an approxi-mated correction factor equal to 1.1 applied to the
‘‘without shield measurements’’ specified in the instru-
ment manual, very often the activity value shown by the
instrument was used without any correction during the
administration of the pharmaceutical to the patient in
the hospital. The new Ai0 calibration coefficients,
determined for the specific conditions of use, were
installed in the eight PTW-Freiburg instruments. For the
instruments of the other four manufacturers, no
significant differences, in compliance with the IEC
1145 standard (IEC, 1992), were found between the
installed by the producer and the determined by
the CENTIS-DMR calibration coefficients, under
the specific conditions used in the nuclear medicine
departments.
The nuclear medicine laboratories were additionally
requested to simulate the administration to a patient by
injection, of a prescribed activity. Sealed glass vials,
containing a non-radioactive solution with a chemical
composition similar to the composition of the compar-
ison samples of each employed nuclide, were also sent to
the participants for this administration test. These vials
simulated the patient to whom the activity should be
administered and the non-radioactive solution contained
inner the vials played the role of the blood for carrying
out the typical ‘washing’ or ‘rinsing’ of the syringe
during the injection of the radiopharmaceutical. For this
test, the activity to be ‘‘administered’’ must be taken
with a syringe of a type commonly used by the
participants, from one of the received comparison
samples in the glass vial geometry and injected through
ARTICLE IN PRESSP. Oropesa et al. / Applied Radiation and Isotopes 59 (2003) 383–387386
the rubber stopper to the vial with the non-radioactive
solution. The participants were asked to report:
* the prescribed activity value to be administered,
Xplan,* the activity in the original vial, Xini,* measurement result of the activity taken in the
syringe, Xsyr,* measurement result of the ‘‘true’’ administered
activity, transferred to the sealed glass vial with the
non-radioactive solution from the syringe, Xadm,* residual activity in the original vial,* activity of the radioactive wastes generated during
this process, and* supplementary information about the administration
procedure employed.
The sum of the Xsyr plus the residual activity in the
original vial was always statistically the same as the Xinivalue; that is, no activity losses were detected during the
test. The z-core statistic
z ¼ Xadm � Xplan
��
��=0:033Xplan
was calculated to evaluate the performance. The results
of this simulated administration study are presented
in Table 4 and Fig. 1. Only 47% of the reported
‘administered’ activity results, Xadm, agreed to within
10% of the intended value, Xplan. Two of the results with
Table 4
Results of the simulated administration test
No. of participants 8
Percentage of satisfactory results,
Xadm � Xplan
��
��=0:033Xplan p3
47%
No. of reported values 15
No. of results with Xadm � Xplan
��
��=0:033Xplan p3 7
No. of results with Xadm � Xplan
��
��=0:033Xplan p6 12
No. of results with Xadm � Xplan
��
��=0:033Xplan > 6 3
-12.00
-9.00
-6.00
-3.00
0.00
3.00
6.00
9.00
12.00
Z-s
core
z-scor
Fig. 1. Individual z-score values for the administration test.
the z-score value greater than 6 are in correspondence
with the above-mentioned measurements using the
wrong choice of the switch-selectable radionuclide
option.
These results demonstrate that the administration to
the patient of the exact planned activity should be
considered to be a problem to resolve. In particular, a
written procedure for administration of the radio-
pharmaceutical does not exist in most of the nuclear
medicine departments and in some cases when it does
exist, it is not fully available for routine work. The
criteria for acceptance of an activity measurement
during the administration process were not clearly
stated in any of the existing procedures.
3. Summary and conclusions
In the present study, an attempt has been made to
ascertain the quality of activity measurements in the
field of nuclear medicine in Cuba. When the value of
10% required in the Pharmacopoeia is used as a limit for
the uncertainty (corresponding to about three standard
deviations) of the measurement, the following conclu-
sions could be made based on these findings:
* In most nuclear medicine laboratories, the accuracy
of the activity measurements with radionuclide
calibrators suffices for 131I, 99mTc and 201Tl radio-
pharmaceuticals. Nevertheless, the deviations de-
tected in individual cases show that the possibility
of the occurrence of spurious errors is present.
Consequently, the necessary preventive actions
should be taken to reduce these types of errors to
the minimum.* The calibration of the instruments under the specific
measurement conditions used in the nuclear medicine
laboratories is an important prerequisite to guarantee
the adequate quality of the measurements results.
Otherwise, the influences of the measuring geometry
e |z|=3
The 15th result (z ¼ �28) is not shown in the figure.
ARTICLE IN PRESSP. Oropesa et al. / Applied Radiation and Isotopes 59 (2003) 383–387 387
and the type and form of the source on the
measurand are often not taken into account by these
laboratories.* It is estimated from the responses received from the
questionnaires and the discussion of the results with
the participants, that written procedures for radio-
nuclide calibrator measurements and the administra-
tion of the radiopharmaceuticals do not exist in most
of the nuclear medicine laboratories and in some
cases when they exist, they are not available for the
routine work. The criteria for the acceptance of an
activity measurement during the administration
process were not clearly stated in any of the existing
procedures.* It is estimated from the responses received from the
questionnaires that in most of the nuclear medicine
laboratories, the test measurements recommended
for the radionuclide calibrators are often not carried
out at all or only at irregular intervals.* The administration of the accurate prescribed activity
is still a problem to be resolved for improving the
quality of service to the patient in the nuclear
medicine laboratories. The adequate qualification of
the technicians in this area is of the highest
importance for this purpose.
To improve the quality of the measurements, compar-
isons with the most important nuclides should be
organized at regular intervals. These comparisons have
to be used as an important tool for the identification of
the problems and to initiate corrective actions. The
interest in this kind of comparisons should be awakened
in all laboratories working in the field of nuclear medicine.
The obtained results reflect the necessity of establish-
ing Quality Systems in the Cuban Nuclear Medicine.
Acknowledgements
We thank all analysts and participating institutions
for their collaboration during the execution of this
study.
References
Debertin, K., Schrader, H., 1992. Intercomparisons for quality
assurance of activity measurements with radionuclide
calibrators. Nucl. Instrum. Methods A 312, 241.
European Pharmacopoeia, 1982. 2nd Edition, Maisonneuve,
Sainte Ruffine, France.
Gu!ıa para la implementaci !on de los reglamentos de seguridad
en la pr!actica de la medicina nuclear, 2002. 1st Edition.
Ministerio de Ciencia, Tecnolog!ıa y Medioambiente.
CNSN. Havana, Cuba.
IEC (International Electrotechnical Committee), 1992. Calibra-
tion and usage of ionization chamber systems for assay of
radionuclides. IEC 1145.
ISO/IEC (International Standards Organization and Interna-
tional Electrotechnical Committee), 1997. Proficiency test-
ing by interlaboratory comparisons—Part 1: Development
and operation of proficiency testing schemes. ISO/IEC
Guide 43-1.
Oropesa, P., Serra, R., Guti!errez, S., Hern!andez, A.T., 2002. A
procedure for the standardization of gamma reference
sources for quality assurance in activity measurements of
radiopharmaceuticals. Appl. Radiat. Isot. 56, 787.