60
Comparison of Radiofrequency Coil Configurations for Multiple Mouse Magnetic Resonance Imaging by Marc Carias A thesis submitted in conformity with the requirements for the degree of Masters of Science Medical Biophysics University of Toronto © Copyright by Marc Carias 2013

Comparison of Radiofrequency Coil Configurations for ... · Figure 1 - Simple Resonant Circuit Figure 2 - SNR Measurements using Regions of Interest ... discussion on current techniques

Embed Size (px)

Citation preview

Comparison of Radiofrequency Coil Configurations for Multiple

Mouse Magnetic Resonance Imaging

by

Marc Carias

A thesis submitted in conformity with the requirements

for the degree of Masters of Science

Medical Biophysics

University of Toronto

© Copyright by Marc Carias 2013

ii

Comparison of Radiofrequency Coil Configurations for Multiple

Mouse Magnetic Resonance Imaging

Marc Carias

Masters of Science

Marc Carias

University of Toronto

2013

Abstract

Multiple-mouse MRI (MMMRI) accelerates biomedical research by imaging multiple mice

simultaneously. To date, MMMRI has been explored in three ways: shielded transmit-receive

coils, shielded transmits coil with separate unshielded receive coils; and finally shielded transmit-

receive coils with independent gradient coils. However alternative transmit coil configurations and

possible benefits of eliminating shielding have not yet been explored. The goal of this thesis is to

test possible radiofrequency configurations with and without shielding for the purpose of

improving image quality for MMMRI. Results demonstrate that using an unshielded transmit-

receive coil array provided a 20% improvement over an identical shielded coil. A new unshielded

7-coil MMMRI array is presented, minimizing the ghosting between image overlap using mutual

inductance minimization and a sensitivity encoding (SENSE) reconstruction. The final array

provided high resolution images (90µm) of up to seven live mice simultaneously with appropriate

signal-to-noise for automated analysis.

iii

Acknowledgments

First and foremost I would like to acknowledge and thank my supervisor Dr. Brian Nieman for his

constant motivation and support during my thesis. He has introduced and taught me skills that

cannot be memorized or listened to in lecture. This allowed me to grow not only as a scientist but

also as an individual.

I would also like to thank my committee members, Dr. Mark Henkelman and Dr. Alex Vitkin.

Both their expertise and ability to ask the right questions guided my thesis to have a larger impact.

Their guidance and feedback pushed me to become a better student and become more familiar with

various topics.

The Mouse Imaging Centre deserves a special thanks, for providing a great work environment

filled with many areas of expertise. In particular I would like to personally thank Jun Dazai, the

mechanical engineer at MICe. With his help, intuition and motivation, I was able to efficiently

finish my thesis and acquire new skills in device design and machining. I would like to thank

Shoshana Spring, Christine LaLiberté and Dr. Lindsay Cahill for all the help with mouse handling

and support. Finally at MICe I would like to thanks Dr. Jonathan Bishop for his help with MR

sequencing and Michael Wong for some coding support and constant motivation.

I also acknowledge all the funding agencies including the Natural Sciences and Engineering

Research Council of Canada, OICR, and the Faculty of Medicine at University of Toronto.

iv

Table of Contents Chapter One - Introduction

1.1 Mouse Imaging ......................................................................................................................... 1

1.2 MRI ........................................................................................................................................... 2

1.2.1 Fundamentals of MRI Theory ............................................................................................ 2

1.2.2 Radiofrequency Coils ......................................................................................................... 3

1.3 Parallel Imaging ........................................................................................................................ 6

1.3.1 Surface Coils ...................................................................................................................... 6

1.3.2 Phased Array ...................................................................................................................... 7

1.3.3 Sensitivity Encoding .......................................................................................................... 9

1.4 MMMRI and Parallel Imaging ............................................................................................... 12

Chapter Two - Experiments

2.1 Saddle Coil Design ................................................................................................................. 14

2.1.1 Methods ............................................................................................................................ 15

2.1.2 Results ............................................................................................................................. 17

2.2 MMMRI Configurations ......................................................................................................... 20

2.2.1 Methods ............................................................................................................................ 21

2.2.2 Results .............................................................................................................................. 22

2.3 The Effect of Shielding ........................................................................................................... 23

2.3.1 Methods ............................................................................................................................ 24

2.3.1 Results .............................................................................................................................. 24

2.4 Receive Geometries and Coupling ......................................................................................... 25

2.4.1 Rotation ............................................................................................................................ 26

2.4.2 Translation ........................................................................................................................ 28

2.4.3 Depth ................................................................................................................................ 30

2.4.4 Coupling and SNR ........................................................................................................... 32

2.4.4.1 Methods ..................................................................................................................... 32

2.4.4.2 Results ....................................................................................................................... 32

2.5 Theoretical Design of a MMMRI Array ................................................................................. 34

2.6 Imaging with the array ............................................................................................................ 39

2.6.1 Methods ............................................................................................................................ 40

v

2.6.2 Results .............................................................................................................................. 41

Chapter Three - MMMRI Coil Array

3.1 Benefits ................................................................................................................................... 44

3.2 Potential Use ........................................................................................................................... 45

3.3 Custom Coils ........................................................................................................................... 46

3.4 Summary ................................................................................................................................. 48

vi

List of Tables

Table 1 - Orientations (in degrees) of Two Saddle Coils

vii

List of Figures

Figure 1 - Simple Resonant Circuit

Figure 2 - SNR Measurements using Regions of Interest

Figure 3 - Surface Coil Schematic with Field Simulation

Figure 4 - Geometric Decoupling Between Coils

Figure 5 - Sensitivity Encoding Reconstruction

Figure 6 - Saddle Coil Schematic with Parameters

Figure 7 - B1 Profile of a Saddle Coil

Figure 8 - SNR Measurements of Various Saddle Coils

Figure 9 - Saddle Coil Form and Actual Photo

Figure 10 - Radiofrequency Configurations Schematics

Figure 11 - SNR Results of RF Configurations

Figure 12 - Shielding of RF Coils

Figure 13 - SNR and Mutual Inductance of Rotating Saddle Coils

Figure 14 - SNR and Mutual Inductance of Translating Saddle Coils

Figure 15 - SNR and Mutual Inductance of Shifting Saddle Coils

Figure 16 - Comparing SNR and Mutual Inductance

Figure 17 - Mutual Inductance Minimization Results

Figure 18- B1 Field Maps of coils 1 through 4 within the Array

Figure 19 - Saddle Coil Orientation Sensitivity

Figure 20 - CAD drawing and Actual Photo of Final Array

Figure 21 - Pre and Post SENSE Phantom Images

Figure 22 - 7 Live Mice Brain Images Post SENSE Reconstructed

Figure 23 - Three dimensional average image of 7 live mice brains

Figure 24 - Custom Intubated Mouse Imaging Coil and Holder

viii

List of Acronyms

MMMRI - Multiple Mouse MRI

SENSE - Sensitivity Encoding

MICe - Mouse Imaging Centre

OICR - Ontario Institute for Cancer Research

SNR - Signal to Noise Ratio

RF - Radiofrequency

CAD - Computer-aided Design

MRI - Magnetic Resonance Imaging

Tx - Transmit

Rx - Receive

Tx-Rx - Transmit receive

ID - Inner Diameter

OD - Outer Diameter

M - Mutual Inductance

FOV - Field of View

ROI - Region of Interest

TR - Repetition Time

TE - Echo Time

BOLD - Blood Oxygen Level Dependant

3D - Three dimensional

ASL - Arterial Spin labelling

CBV - Cerebral Blood Volume

1

Chapter 1

Introduction

This thesis will focus on comparing different radiofrequency coil configurations for in vivo

multiple mouse MRI. The introduction will begin with a brief section on mouse imaging, followed

by a brief introduction to basic MRI. A more detailed section is provided for radiofrequency coils

and using multiple coils to gain sensitivity in a phased array. This chapter will conclude with a

discussion on current techniques for multiple mouse MRI followed by the motivation for this

thesis.

1.1 Mouse Imaging

Mice in a laboratory can be manipulated and analyzed in a fashion that cannot take place in human

studies. For instance, genetic models, in which genes are added to or removed from the genome,

can be created to study the resulting phenotype (observable characteristics or traits) (1-3). In

addition, diseases such as cancer can be followed longitudinally over their full time course when

using mouse models (4-6). Magnetic resonance imaging (MRI) can be a powerful tool in the

analysis of mouse models by providing an in vivo imaging modality that offers soft tissue contrast

and high resolution capability (7-11). Compared to clinical imaging, high field strengths and longer

scan times are necessary in order to obtain sufficient resolution for mouse imaging. Mouse studies

always require statistical analysis of findings. This is efficiently achieved through increasing the

number of samples within a study. Therefore, there is a need for more effective imaging throughput

(8).

2

A typical small animal MRI system may accomplish two to four live imaging sessions a day. The

duration of each scan is limited by the length of time mice can spend under anesthesia, typically a

max of 3 hours. However by using a multiple-mouse MRI (MMMRI) technique as described in

Bock et al. (8,9), one can accomplish relatively high throughput (up to 16 mice at once) without

sacrificing signal to noise (10,11). This is achieved by using a larger magnet than necessary for a

single mouse and imaging multiple mice simultaneously using independent shielded RF coils. Due

to the high resolution needed in mouse MRI, much importance is still placed in maximizing

efficient use of image signal.

1.2 MRI

1.2.1 Fundamentals of MRI Theory

The basis of MRI is the directional magnetic moment associated with spin 1/2 nuclei. When a

sample is placed in a large magnetic field, a very slight majority of the free hydrogen moments

align themselves with the direction of the magnetic field. This creates a net magnetization aligned

with the main magnetic field (B0) (20). If perturbed from this position, the net magnetization

precesses about the main magnetic field, like a gyroscope. This behavior is termed Larmor

precession (20). The rate of precession is given by the Larmor equation, Eq. 1.

[1]

This introduces a parameter γ, known as the gyromagnetic ratio which is specific to different

nuclei. The precessing magnetization generates electromagnetic radiation that can be detected and

used for MRI. Equation 1 allows calculation of the frequencies of spins in this magnetic field for

MRI (20). For example, at 7 Tesla the hydrogen atom resonates at 300 MHz.

002

Bf

3

There are two steps in acquiring MRI signals. First, the net magnetization must be perturbed from

its equilibrium along the main magnetic field. This is achieved using a radiofrequency (RF) field

that is perpendicular to the main field and operating at the Larmor frequency, known as the B1

field. This rotates the net magnetization about B0. The RF coil in this case is called a transmit coil

and the process is sometimes referred to as transmission. After transmission, the precession of the

net magnetization must be detected. The precessing magnetization induces current in a receiver

coil that is oriented to detect magnetization perpendicular to B0. Afterwards, the signal from the

receive coil is amplified and digitized. The RF receive and transmit functions can be performed

using the same or separate coils.

1.2.2 Radiofrequency Coils

RF coils are responsible for transmission and reception and therefore can be thought of as the

MRI’s antenna. This is achieved via an electrical circuit containing a series of capacitors (C) and

inductors (L, inherent to coil geometry) (25,26) (figure 1A). At resonance the two impedances of

C and L are of equal magnitude. In a very simple model (figure 1) the resonance frequency is given

by Eq. 2 (26). In an MRI system the impedance is typically matched to 50 Ohms, to accommodate

the back end electronics so no losses (reflections) occur. In order to accomplish this, a matching

network (Fig. 1B) is placed on the coil. Here, R, L and Cs represent coil components where R is

the resistance of the coil (copper wire), L, the coil inductance (dependent on size, shape and coil

geometry) and Cs the series capacitance. Many separate series capacitors can be used to reduce

the effective inductance of the coil or coil element (capacitive segmentation) (21). The matching

network contains two components, a variable tuning capacitor (Ct) and a variable matching

capacitor (Cm). Ct creates a parallel resonance and allows tuning of the coil.

4

Figure 1: RF coils for acquiring MRI signals can be modeled by using a series of inductors (L), capacitors (C) and

resistors (R) in a circuit. A. Simple LC resonant circuit for Eq. 2 to achieve the desired resonance at a given

frequency. B. Most types of RF coils for small animal imaging can be modeled with this circuit. It contains three

capacitors, Cs, Ct (variable tuning), and Cm (variable matching), an inductor L (representing total coil inductance), a

resistor (representing all inherent coil resistance) and the measured voltage V.

𝜔0 =1

√𝐿𝐶= 2𝜋𝑓0 [2]

Since the RF coil is responsible for both transmission and reception, much importance is placed

on the design and efficiency of these coils to obtain the highest signal-to-noise ratio (SNR)

possible. SNR can be measured as the average signal within a region of interest in an imaged

sample divided by the standard deviation of the background signal noise, which in MRI is

Gaussian-distributed in the complex image and Rician-distributed in the magnitude image (figure

2). In practice, to maximize sensitivity in RF coil design, the coil is designed to maximize the

magnetic field (B1) produced per unit current over the desired region of space. The principle of

reciprocity (24) ensures that this design will also maximize current induced by a source within the

same region. With an increase in sensitivity of the coil, the SNR of images generally also increases

(21-23). The increase in SNR achieved, if present, will depend upon the dominant noise sources

affecting the image. Noise can be generated within the coil (thermal noise in the copper wire and

components) and from the sample (e.g., tissues within the body). Both sources degrade image

5

quality identically. With respect to coil design, only the sensitivity to sample signal and the coil

noise generated can be altered, which must be maximized and minimized respectively (21, 23).

Figure 2: SNR is used as a measure of image quality in MRI. SNR is measured by dividing the total average signal

intensity within a region of interest (black box, A), by the standard deviation of background noise (white box, B).

This method is how SNR was calculated throughout the thesis.

Whether coil noise or sample noise will be dominant for a given arrangement is determined by the

loading factor (Lf), Eq.3. The loading factor measures the difference in the electrical properties of

the circuit when a sample is placed within the coil versus an empty coil (21).

[3]

[4]

Lf can be determined from the quality factors (Q factor), Eq.4, of the empty coil (Qunloaded) and the

coil loaded with a sample (Qloaded). The Q factor is a metric that describes how the energy in a

resonant circuit dissipates relative to its initial energy. It is calculated as the resonant frequency

(ω0) quantity divided by the width (Δω) of the resonance at half the impendence's (magnitude)

maximum. When the loading factor (Lf) is below 0.5 the coil noise is expected to be greater than

unloaded

loaded

Q

QLf 1

0Q

6

the sample noise and improvements to the coil design will provide an increase in the SNR of the

image.

1.3 Parallel Imaging

1.3.1 Surface Coils

There are many types of coils with different geometries. Surface coils are one of the most common

in the clinic, where more than one are typically used to gain higher sensitivity. A surface coil is a

loop of wire that lays on the surface of a sample. They are typically used for their simplicity and

their ability to acquire high localized SNR (23,27,29). Figure 3 contains a surface coil schematic

with B1 direction drawn and calculated magnetic field map. A surface coil's effective area where

high SNR is acquired is a function of its radius (27-29), corresponding to the green to red colours

(larger than 50%) in figure 3B. For example a 12mm diameter surface coil will have a high SNR

up to ~6mm from the surface. Features below this will have poorer SNR and therefore a different

coil or multiple coils must be used. Additionally, RF coils should ideally be able to provide a

homogeneous B1 field (induced transmission field) throughout the sample (21). Unfortunately

surface coils are not homogenous. Therefore, one typically uses a surface coil for receive only,

and a separate transmit coil (volume coil) to achieve a homogeneous B1 field.

7

Figure 3: Surface coils, the most common RF coils, offer high local SNR however intensity quickly decreases as you

move away from the surface. A. A schematic indicating the direction of the created magnetic field (B1)

perpendicular to the surface coil. B. Simulated B1 map of a centre slice (grey plane shown in A), normalized to the

max intensity (simulation explained in section 2.1.1).

1.3.2 Phased Array

An array of receiver coils that encompasses an area of interest is called a phased array coil (15).

For this to occur, each receive coil must be connected to its own preamplifier and receive channel.

Currently, most new magnetic resonance systems are equipped with 16-32 channels in order to

accommodate phased arrays. When operating a phased array, several challenges need to be

overcome: Transmit (Tx)- receive (Rx) interactions need to be minimized (also a concern for a

single receive coil) (30), Rx-Rx interactions must be minimized (15), and a single image must be

reconstructed using data from all channels.

Allowing for separate Tx and Rx does provide homogeneous B1 with local sensitivity. However,

since the two coils are tuned to the same resonant frequency (f0), coupling can occur and decrease

the sensitivity. During Tx, induced currents in the Rx coil generate B1-fields that destroy the

desired homogeneity in the Tx-field (21). During Rx, since both coils must be resonant at f0, they

can couple into one another and act together like a large single coil, eliminating any local

sensitivity from the Rx coil. Strong coupling splits the coils' resonant peaks such that neither coil

has high sensitivity at f0 (15) (Figure 4).

8

Coupling must be kept at a minimum such that homogeneity is preserved and resonance splitting

does not occur. Geometric decoupling can be used to minimize both these interactions. The main

interaction between Tx-Rx and Rx-Rx coils is mutual inductance. Mutual inductance (M) is a

phenomenon where a change in current in a coil (while acquiring the MR signal) causes a change

in flux (the amount of magnetic field that passes through a surface) in an adjacent coil. It is

expressed by the Neumann formula (Eq.5):

𝑀 =𝜇0

4𝜋∯

𝑑𝑠𝑖∙𝑑𝑠𝑗

|𝑅𝑖𝑗| [5]

where, µ0 is the magnetic constant, followed by the line integrals of the two coils with a dot product

of coil elements (ds) divided by the distance between those coil elements (Rij) (31). Therefore one

can minimize M geometrically through coil orientation or distance. Optimal geometrical

decoupling between coils can be achieved when the magnetic flux of one coil does not pass through

any adjacent coils (15). This can be accomplished by either separating the coils or finding an

orientation where M sums to zero (figure 4). It was shown by Roemer et al. (15) that overlapping

surface coils does allow for decoupling, where the flux of adjacent coils do not interfere with others

(in this case net flux from one coil into another sums up to zero). For example, in a square surface

coil, an overlap of 10% of its length eliminates the double resonance phenomena (15).

9

Figure 4: Geometric decoupling occurs between RF coils when the flux of an adjacent coil does not induce current

in the other, by either summing to zero or being positioned orthogonally. A. No decoupling method between

adjacent coils. The results in a double peaked resonance (displayed as impedance magnitude) where no peak occurs

at f0. B. Two overlapped coils (green and yellow) where flux is eliminated in an adjacent coil through geometric

decoupling. This eliminates the double peak and resonance is restored at f0. This is shown in the figure where fields

from the green coil (red) that enter the yellow coil (fields not shown) sum to zero (both enter and leave, shown by

arrows) thus no coupling occurs.

1.3.3 Sensitivity Encoding

Once sufficient decoupling has been achieved between multiple Rx coils, a method is required to

reconstruct an image from the multiple channels of data. Figure 5 schematically demonstrates the

acquisition and reconstruction process often referred to as Sensitivity Encoding (SENSE) (16). For

a single coil and a fully sampled dataset, the acquired image (Ia) is the product of the coil sensitivity

(Ca1) and the object (ρ1), for every pixel in the image and coil sensitive region, Eq. 6.

10

𝐼𝑎 = 𝐶𝑎1𝜌1 [6]

It is common practice when using SENSE to acquire data from multiple coils and to sample less

data to reduce acquisition time. This has the effect of aliasing the image on top of itself. Each pixel

now contains information from several points in the object, ρ1-4, and signal from other pixels. It

can be written as (Eq. 7):

𝐼𝑎 = 𝐶𝑎1𝜌1 + 𝐶𝑎2𝜌2 + 𝐶𝑎3𝜌3 + 𝐶𝑎4𝜌4 [7]

In this equation, Ia is known (acquired), all four coil sensitivities are known (Ca), however all four

object pixels are unknown. If multiple receiver coils are introduced, with their own known

sensitive regions, corresponding equations here will be introduced per coil (figure 5). Therefore,

if we introduced three more coils (coils b, c and d) we have the dataset represented in Eq. 8 or 9:

𝐼𝑎 = 𝐶𝑎1𝜌1 + 𝐶𝑎2𝜌2 + 𝐶𝑎3𝜌3 + 𝐶𝑎4𝜌4

𝐼𝑏 = 𝐶𝑏1𝜌1 + 𝐶𝑏2𝜌2 + 𝐶𝑏3𝜌3 + 𝐶𝑏4𝜌4

𝐼𝑐 = 𝐶𝑐1𝜌1 + 𝐶𝑐2𝜌2 + 𝐶𝑐3𝜌3 + 𝐶𝑐4𝜌4 [8]

𝐼𝑑 = 𝐶𝑑1𝜌1 + 𝐶𝑑2𝜌2 + 𝐶𝑑3𝜌3 + 𝐶𝑑4𝜌4

Or

𝐼 = �̂� ∙ �⃗� [9]

The full image can be reconstructed using Eq. 10 (16) (figure 5):

�⃗� = �̂�−1 ∙ 𝐼 [10]

In order to obtain an accurate coil sensitivity map, a large field of view image is acquired to

determine a sensitivity map for each coil over the whole object. This sensitivity map includes the

spatial sensitivity inherent to the coil geometry and any residual coupling between coils. It is

important to obtain an accurate sensitivity map, so several post processing steps are usually

11

performed to reduce noise in the map. For example, since a smooth coil sensitivity is expected, we

can smooth the results using a polynomial fit to achieve an accurate map (16). In work described

in section 2.6 of this thesis, we will take the coil sensitivity C to be piecewise constant over

multiple samples in an array. This allows a very simple form of SENSE as described in Bock et

al. (8), referred to as a SENSE-like reconstruction. Note that this makes the assumption that the

dominant source of signal in the sensitivity map is residual coupling between coils. In this mode,

the key objective of SENSE is to separate coils signals from one another. This has the benefit of

also subtracting any noise likewise coupled, so can also improve SNR.

A measure known as the coil geometry factor tells us how well the SENSE method may separate

pixel intensities to corresponding images or how much our sensitivity maps overlap/relate with

one another (16) (Eq.11):

𝑔 =𝑆𝑁𝑅𝑐𝑜𝑖𝑙

𝑆𝑁𝑅𝑟𝑒𝑐𝑜𝑛 [11]

The g factor is computed by dividing the SNR of the initial coil image by the final reconstructed

image with all other coils present. The larger the g factor, the more coil sensitivities overlap.

SENSE is commonly used in clinical systems where there is often an excess of SNR that could be

traded off for lower scan time. This is not the case for mouse and small animal imaging where

extra sensitivity is usually desired to increase resolution.

12

Figure 5: SENSE reconstruction without artifact relies on accurate individual coil sensitivity maps to reconstruct the

desired image. Here, four surface coils are used to acquire an image. Each pixel in these images (Ia-d) contains

spatial data from four points in the object (ρ1-4). Using coil sensitivity maps (Ca-d1-4) at corresponding pixel

locations, equation 9 can be solved to acquire the desired reconstructed image.

1.4 MMMRI and Parallel Imaging

The aim in MMMRI is to generate separate images for many samples simultaneously. Initially

four configurations for performing MMMRI were considered (8). These included, i) imaging mice

with separate receivers and separate gradient coils, ii) imaging with separate unshielded RF coils

with a common gradient with a field of view (FOV) encompassing all mice, iii) imaging with

separate shielded RF coils with a common gradient with a small FOV and iv) imaging with separate

shielded RF coils with a common gradient with a large FOV. It was shown that using an individual

shielded RF coil per mouse with a common gradient coil (small FOV) allowed the highest

throughput without sacrificing image quality (8). A shield is a conductive material that surrounds

13

the RF coil to limit RF signal within a desired volume. This made it possible to run several

independent experiments in a single imaging session. It was later shown that this configuration

provides 3.3 times greater SNR when compared to imaging all samples in a single large image

with a large RF coil (12). A different approach has been described as a 'super-parallel microscope'

which introduces a separate gradient field per sample in a multiple imaging sample session (19).

A historic limitation of MMMRI is the availability of having multiple channels for transmit and

receive. Traditionally, MRI scanners were equipped with only one or two transmit channels,

favoring the large single transmit receive RF coil configuration, a major limitation for imaging

mice efficiently. However, this is no longer a challenge as most MRI systems are installed with

multiple channels (13,14). This dramatically increases the feasibly of implementing MMMRI,

which may allow it to become more common practice, even on human clinical systems.

Increasing the sensitivity of MMMRI has recently been investigated by adapting a multi-coil

interface to allow for parallel imaging or increased sensitivity. A multi-coil MMMRI (three receive

and one transmit coil per mouse) has recently been demonstrated and showed an acceleration factor

of 1.8 compared to the same image quality obtained using a single channel configuration (18). This

now allows receive coil arrays to be tailored to MMMRI experiments as seen in clinical systems

(14,18). However, possible transmit configurations with respect to image quality have not been

explored. Furthermore, SENSE and similar algorithms to separate signal from multiple coils are

now commonly used. This suggests that additional MMMRI configurations without shielding

between mice are possible, and they may have some image quality benefits. It is therefore the goal

of this thesis to explore possible radiofrequency coil configurations with and without shielding for

the purpose of improving image quality for MMMRI.

14

Chapter 2

Experiments

This chapter will focus on experiments conducted to determine a RF coil configuration for

MMMRI. I begin with an experiment involving the design of a RF coil specifically for in vivo

mouse brain imaging. Afterwards several RF configurations are compared using this newly

designed coil, accompanied by a discussion and further experiments involving shielding. Once a

configuration was chosen, I explored methods of geometric decoupling for design of a final array.

This chapter concludes with the construction of the final multiple mouse brain imaging array and

high resolution brain images of seven live mice.

2.1 Saddle Coil Design

At first I designed a volume coil capable of imaging the mouse brain with high sensitivity. A

volume coil is classified as a RF coil that encompasses a specific region of a sample with a fairly

uniform field over the region of interest, for example the head. Volume coils offer excellent

homogeneity compared to surface coils (15). The most effective volume coils are solenoids, which

offer the highest SNR and have excellent homogeneity (32,33). Solenoids are helical wound wires

with a number of turns around a cylinder that produces a magnetic field along the helical axis

(32,33). However, a major disadvantage of solenoids is that they must be positioned 90 degrees

relative to the main magnetic field B0. This constraint makes solenoids impractical for loading

live mice. A solution to the loading issue is to use a saddle coil. A saddle coil has a cylindrical

body coaxial with the magnet and turns of wire on the cylinders curved frame. The field produced

is perpendicular to the axis of the cylinder (figure 6).

15

2.1.1 Methods

Due to its simplicity and expected high performance for mouse imaging at 7 Tesla, a linear saddle

coil was chosen. Saddle coils of a given diameter only have two parameters that can be altered:

the length of the saddle (the length down the cylinder) and the aperture angle (how far the saddle

extends along the circumference of the cylinder) (figure 6). The diameter (D) of a saddle is usually

set to be the minimum radius that will allow the coil to encompass the entire sample. Previous

studies have shown that in order to provide an optimally homogenous B1 the saddle must have an

aperture angle (α) of 120 degrees and the length (Lo) must be at least four times its radius (34-36).

In addition, it has also been demonstrated that when coil noise is dominant, adding more loops to

a saddle coil improves the Q factor and SNR (37). These results however were for lower fields,

while at higher fields with small rodents both sample and coil effects must be considered equally.

Therefore, I investigated a saddle coil design at 7 Tesla for the mouse brain.

For this investigation, I introduced three variables to be tested: i) saddle aperture, ii) the effects of

adding loops and iii) connecting the loops in a parallel circuit arrangement compared to the

traditional series. A total of five coils were built and tested. All coils were built on a cylinder with

radius 11mm and a length of 32mm. This length is the maximum coil length possible for an adult

mouse head before the shoulders must also fit within the coil diameter. In addition, an inductance

calculation was performed using a MATLAB (The Mathworks Inc., Natick, MA) script that

allowed a wire pattern for each coil to be made and then the entire coil discretized into 5,000 to

10,000 elements. This allowed us to calculate total inductance and mutual inductance using

Neumann's formula (Eq. 5). This method was compared to measured values of the inductance of

our saddle coils. The calculation of multi turn saddle coils with parallel loops used the same

method. However, for the parallel configuration, current was divided between the loops based on

16

their inductive impedance. In addition, the inductance of coils after construction was measured

using a network analyzer (Aligent, Palo Alto, CA). In every case, the measured value was within

5% of the calculated inductance. The geometry of the five coils is as follows (calculated coil

inductance, in parentheses):

1. Single loop with an aperture of 120 degrees (120nH)

2. Single loop with an aperture of 180 degrees (180nH)

3. Double loop with apertures of 180 and 120 degrees (310nH)

4. Triple loop with apertures of 180, 120 and 90 degrees (630nH)

5. Double loop with apertures of 180 and120 degrees connected in parallel (220nH)

Figure 6: A schematic of a single loop saddle coil with parameters: D, diameter (mm), Lo, coil length (mm) and α

aperture in degrees. Coils with multiple loops will have additional α and Lo values. The red arrow indicates the

direction of the produced field (B1).

Simulated B1 images were computed using the Biot-Savart Law (Eq. 12) where, µ0 is the magnetic

constant, I is current and r is the distance from the wire element to where the field is being

17

calculated. This calculation was performed using a MATLAB script that allowed specification of

a wire pattern for each coil, which was discretized into 2,000 to 5,000 elements.

𝐵 = 𝜇0

4𝜋∮

𝐼𝑑𝑙×𝑟

|𝑟|3 [12]

For all of the coils with more than one loop, the loops were spaced one wire diameter apart to

reduce capacitive coupling and proximity effect between loops (43). All coils loops were cut once

per loop (at the top of the saddle) for capacitive segmentation. All coils were tuned and matched

using the variable capacitors in the circuit shown in fig 1. Phantoms (10mL 1% agar, 0.9% saline)

were scanned using a gradient echo sequence with the following imaging parameters: repetition

time (TR)=50ms, echo time (TE)=4.55ms, Matrix:512x512, FOV:5.12x5.12cm, flip angle= 90

degrees, scan time = 26 seconds, slice thickness= 3mm and one average, using our 7 Tesla Varian

(Varian NMR Instruments, Palo Alto, CA) equipped with a 290 mm inner bore gradient (Tesla

Engineering Ltd., Sussex, UK) multiple mouse imaging system capable of imaging up to 16

channels. SNR measurements were taken using a region of interest (ROI) within the centre of the

phantom for the signal and the top right corner for background noise as seen in figure 2.

2.1.2 Results

The ideal B1 profile should have large and uniform amplitudes in each direction (figure 7).

Although not feasible to build (due to space restrictions) a four loop design was also simulated,

with the inner most loop having an aperture of 60 degrees. Here all the plots are normalized to the

highest B1 value in the 4 loop calculation. The single loop of 120 degrees aperture coil provided

comparably low magnitude magnetic field. Increasing the aperture to 180 degree increased B1

magnitude in the central plane, although inhomogeneity is clearly present, i.e. the profiles are not

flat. Two loops increased B1 magnitude, with a parallel configuration showing better homogeneity

18

but lower B1 magnitude. Three loops provided the highest B1 magnitude of the coils that could be

built, with excellent homogeneity.

B1 Line Profile Across Z Axis

-15 -10 -5 0 5 10 150.0

0.2

0.4

0.6

0.8

1.0120 Deg

180 Deg

2 Loops

2 Loops Parallel

3 Loops

4 Loops

Z Position (mm)

No

rmalized

B1

Figure 7: Calculated B1 values show that a three loop saddle coil provides good B1 homogeneity (flat profile) with

high B1 values. A. Normalized B1 line profile across the X axis of saddle coils (blue line in diagram). B: Normalized

B1 line profile across the Z axis of saddle coils (red line in diagram). The four loop saddle coil is included for

comparison but is not feasible for this study because of space restrictions.

SNR results show that our three loop design resulted in the highest SNR (figure 8). Figure 8 shows

normalized SNR (normalized to the three loop coil) as a function of calculated coil inductance. As

expected, the increase in B1 magnitude resulted in an increase in SNR. Overall a 77% increase in

SNR was seen between a single loop 120 degrees aperture and a three loop coil. In addition, a large

B1 Line Profile Across X Axis

-30 -25 -20 -15-10 -5 0 5 10 15 20 25 30-1.0

-0.5

0.0

0.5

1.0 120 Deg

180 deg

2 Loop

2 Loop Parallel

3 Loop

4 Loop

X Position (mm)

No

rmalized

B1

X

Z

A

B

19

increase of 50% was seen between the 120 and 180 degrees aperture with only a smaller difference

between the 2 and 3 loops coils (33%). This suggests that adding more loops (or increasing coil

inductance) at this point offers diminishing SNR returns. This indicates a relationship where the

sensitivity (ability to acquire signal) of adding more loops is offset by the addition of more coil

resistance, a result not unexpected for high fields. In addition, adding more inner loops, (with

progressively smaller diameter) will cause these loops to have smaller sensitive regions, reducing

the benefit through much of the coil.

Figure 8: The three loop saddle coil provided the highest SNR when compared with four other feasible

configurations. Normalized SNR as a function of coil inductance (nH). The joining curve connects all coils

connected in series (circles) while the square point represents the parallel connection saddle.

In all cases Lf values remained the same at around 0.5±0.05. This suggests that coil and sample

noise contribute about equally in all cases. Therefore, when adding more loops, the relative sum

of coil and sample noise may be reduced if coil noise increases less, providing a higher SNR in

the image.

With the support of this data, we concluded that a three loop saddle coil would be appropriate for

in vivo mouse brain imaging. Construction of these coils is difficult, and in order to maintain

Rx Coil Inductance

0 200 400 600 8000.0

0.2

0.4

0.6

0.8

1.0

1 Loop 120 deg

1 Loop 180deg

2 Loops

3 Loops

2 Loops Parallel

Coil Inductance (nH)

No

rmalized

SN

R

20

consistency and to be able to create many of them for MMMRI a form was designed (courtesy of

Jun Dazai) for the three loop saddle coil (Fig. 9). Coils for the remainder of this thesis were

constructed using this form.

Figure 9: Final three loop saddle coil design using a custom form. This coil is used throughout thesis. A: A

Computed-aided design (CAD) drawing of a three loop saddle coil (courtesy of Jun Dazai). Loop wires are

positioned one diameter apart to minimize capacitive and proximity effects. B: Photo of actual three loop saddle coil

in a clear form showing segmenting capacitors.

2.2 MMMRI Configurations

To compare possible Tx and Rx coil configurations four cases were tested using the simple case

of two samples placed 60mm apart. I compared four configurations, shown in figure 10. Briefly,

Case A is composed of a shielded Tx coil surrounding individual Rx saddle coils. In case B, saddle

coils operate in a transmit-receive mode surrounded by a copper shield. In case C, one single large

Tx coil encompasses all individual saddle Rx coils. In case D, saddle coils operate in a transmit-

receive mode without any shields. I used SNR measurements to compare these configurations.

21

2.2.1 Methods

Phantom based SNR measurements were taken with 10mL 1% agar phantoms electrically loaded

with 0.9% saline to simulate a mouse's head. In Case A, (Fig. 10A) two shielded birdcage Tx coils

(outer diameter [OD] 60mm, inner diameter [ID] 40mm, RAPID MR International) were mounted

adjacent to one another, with two three loop saddle coils (ID 19mm) to serve as Rx coils. In this

configuration, only Tx-Rx interactions must be considered and not Rx-Rx interactions due to

shielding of the Tx coils. Therefore Tx and Rx coils were placed so that their fields are orthogonal

to each other (geometrically decoupled) to avoid any coupling effects. This approach is similar to

the method used in Bock et al (8), where each mouse is isolated from one another and allows us to

conduct isolated experiments. Case B, (Fig. 10B) consisted of the three loop saddle coils operating

in a transmit-receive (Tx-Rx) mode with a 60mm diameter copper shield surrounding each coil.

This case provides similar benefits to Case A however no coil-to-coil interactions need to be

considered. Case C, (Fig. 10C) involved a single large-volume shielded birdcage Tx coil (OD

154mm, ID 135mm, RAPID MR International) of sufficient size to image at least 7 mice

simultaneously, surrounded by unshielded three loop saddle Rx coils (positioned in the same

orientation). Once again, the large Tx coil and Rx coils with positioned orthogonally to one another

to prevent any coupling between the two. In addition to the Tx-Rx interactions described above,

this configuration must also consider Rx-Rx interactions. Here, the ability to run separate

independent experiments is lost. For example, using separate transmit frequencies per sample is

not possible. Case D, (Fig. 10D) consisted of two adjacent three loop saddle coils in the same

orientation operating in a Tx-Rx mode without any shielding. Here, Rx-Rx interactions must be

considered, as described above, and Tx-Tx interactions are also possible. Large field of view

(LFOV) images were acquired in all cases using a gradient echo sequence with a field of view

(FOV) encompassing all samples, with the following parameters: (TR=50ms, TE=4.55ms,

22

Matrix:128[readout]x704[phase encoding] FOV:5.12x14.1cm, flip angle=90degrees, number of

averages= 2, slice thickness= 3mm and scan time = 1 minute 10 seconds). LFOV images

eliminated the need to further post process images using a SENSE algorithm, enabling assessment

of each configuration based on SNR.

Figure 10: Four RF coil configurations tested for MMMRI. A: Two shielded birdcage Tx coils mounted next to one

another with saddle coils as Rx. B: Two saddles coils operating in a Tx-Rx mode with 60mm diameter shield

surrounding them. C: One large Tx birdcage coil surrounding two Rx saddle coils. D: Two exposed saddle coils

operating in a Tx-Rx mode. B1 generating by the saddle coils and birdcage coils are shown in red and blue

respectively, indicating that coils are geometrically decoupled.

2.2.2 Results

The RF coil configuration experiment showed two distinct pairings based on the effect of shielding

(figure 11). Here the average of three measurements is plotted against normalized SNR (with

standard deviations). First, cases A and B both represent individually shielded experiments and

have a normalized average SNR of 83±6% and 80±8% respectively. Second, case C and D both

23

eliminate shielding effects however introduce Rx-Rx interactions and have a normalized SNR of

100±4% and 93±2%, respectively.

Figure 11: RF configuration tests in an agar phantom indicate that either a large single Tx coil or RF coils in a Tx-

Rx mode without shielding provided the highest SNR. Values are normalized to the average of the large single Tx

coil. Errors bars indicate the standard deviation of three measurements.

2.3 The Effect of Shielding Results from above indicate that shield effects may degrade image quality by up to 20%. To further

understand coil-to-shield interactions, a further experiment was conducted using our three loop

saddle coil in a Tx-Rx mode with the same phantoms described above. This shielding effect is very

likely to have a geometry dependence, therefore shields of varying diameter were tested.

SNR of Different RF Coil Configurations

A, I

ndivid

ual T

x

B, T

R w

ith S

hield

ing

C, L

arge

Sin

gle T

x

D, T

R N

o Shie

ldin

g

0.0

0.5

0.6

0.8

1.0

RF Coil Configuration

No

rmalized

SN

R

24

2.3.1 Methods

In this experiment images were acquired using a two-dimensional gradient echo sequence with

parameters: TR=50ms, TE=4.55ms, Matrix:512x512 FOV:5.12x5.12cm, flip angle= 90 degrees,

slice thickness= 3mm, number of averages= 2 and scan time = 50 seconds. The shield diameters

that were tested were: 45mm, 60mm, 75mm, 120mm and 290mm (diameter of our gradient bore).

Three different coils were tested with all shields and SNR measured.

2.3.1 Results

Normalized SNR was found to be directly related to the shield diameter (Figure 12, slope = -0.94

and R2=0.92). These results are consistent with the empirical relationship in equation 13, where

this line should have slope of -1. Equation 13 indicates that the effect of shield on a coil is geometry

dependant (38). The bigger the shield, the better the coil performance. A total measured loss of

15% was seen in decreasing the shield diameter from 290mm (gradient bore diameter) to 60mm

(diameter of shield and Tx coil used in figure 11). This suggests that the 20% difference in SNR

seen between our cases (A,B and C,D) is due to the effect of shielding.

25

𝑁𝑜𝑟𝑚𝑎𝑙𝑖𝑧𝑒𝑑𝑆𝑁𝑅 = 1 −𝐶𝑜𝑖𝑙 𝐷𝑖𝑎𝑚𝑒𝑡𝑒𝑟

𝑆ℎ𝑖𝑒𝑙𝑑 𝐷𝑖𝑎𝑚𝑒𝑡𝑒𝑟 [13]

Figure 12: Shielding of RF coils lowers SNR, this difference of SNR accounts for the variation in SNR in figure 13

(configurations A,B versus C,D ). Inverse shield diameter (1/mm) is plotted against normalized SNR. The red

continuous line represents a regression line with 95% confidence interval (dash curve) to compare results to the

expected relationship in equation 13.

The results from this section and the previous section favor a mouse array of a single large Tx coil

encompassing all mice with individual Rx coils (Case C), or of saddle coils in a Tx-Rx mode

without any shields (case D). We will favour case C, due to its simplicity, since case D would

required additional electronics to ensure that Tx-Rx interactions are eliminated in a MMMRI array,

which can reduce SNR by up to 10% (21). In case C, both Tx-Rx and Rx-Rx interactions must be

minimized, through mutual inductance.

2.4 Receive Geometries and Coupling

Induced B1-field is dependent on the rotation of our saddle coil, therefore coupling can be reduced

by coil separation and coil orientation. This can be accomplished either by rotating or separating

adjacent coils. Three more experiments were conducted: i) rotating two coils at a fixed separation,

Shielding of Radiofrequency Coils

0.000 0.005 0.010 0.015 0.020 0.0250.0

0.5

0.7

0.8

0.9

1.0

1.1

1/Shield Diameter (mm)

No

rma

lize

d S

NR

26

ii) changing the distance between coils with two fixed rotations and iii) shifting the coils along Lo

(figure 6, the B0 or z direction) at a fixed rotation. It is well known that placing linear coils

orthogonal to one another will eliminate inductive coupling between the two. However, in a

multiple mouse array, not all coils can be placed orthogonal to one another and many interactions

can take place. Therefore, other possible configurations must be investigated.

2.4.1 Rotation

In this experiment two three loop saddle coils were placed adjacent to one another at a fixed centre-

to-centre distance apart of 60mm. By using symmetry only a select number of orientations were

tested (Table 1), these included: 0-0, 0-30, 0-60, 0-90, 30-30, 30-60, 30-90, 60-60,60-90 and 90-

90 degrees. Each orientation was measured as the counter clockwise rotation from the 0 degrees

orientation, where the induced B1-field in the saddle coil is vertical. Images were then acquired

using a two dimensional gradient echo (TR=50ms, TE=4.55ms, Matrix:256x704

FOV:5.12x14.1cm, flip angle= 90 degrees, number of acquisitions= 2, slice thickness= 3mm and

scan time = 1 minute 10 seconds). This experiment was repeated using three combinations of three

different saddle coils. Once again large FOV images were acquired so that no post processing was

necessary. We used the same calculation in section 2.1.1 to predict the mutual inductance between

two coils. Mutual inductances were calculated with both coils rotated from zero to 90 degrees and

plotted on a contour plot (figure 13B).

27

Table 1: Orientation (in degrees) of adjacent saddle coils for Rx coil rotation test

Coil Degree 0 30 60 90

0

30 -

60 - -

90 - - -

Figure 13: The relative orientation of two adjacent saddle coils causes changes in mutual inductance that were found

to significantly affect image quality. A. SNR using 10 different orientations of two Rx coils. The error bars

represent standard deviation values of three measurements. B. A contour plot of calculated mutual inductance

between two coils with varying degrees of rotation.

The ideal rotation was found to be 0-90 degrees with a 25% increase in SNR relative to the poorest

rotation of 90-90 (figure 13). For the most part, expected trends (calculated in Figure 13B) are

seen. For example, an increase in SNR is seen from coils approaching the ideal 0-90 degrees

orientation and away from the non-ideal 90-90 degrees orientation. The 30-30 and 30-60

orientations did not perform as well as expected, which may imply coupling through means other

than mutual inductance.

Rx Coil Rotation

0-0

0-30

0-60

0-90

30-3

0

30-6

0

30-9

0

60-6

0

60-9

0

90-9

0

0

700800

900

1000

1100

1200

1300

Rotation (Degrees)

SN

R

A B

28

2.4.2 Translation

Our next experiment tested how significantly the perpendicular distance between coils affected

image quality. Two three loops saddle coils were oriented in a 0-0 degrees position and the

separation between the two coils was tested at 40mm, 50mm, 60mm and 70mm. Images were

acquired using a two dimensional gradient echo (TR=50ms, TE=4.55ms, Matrix:128x704

FOV:5.12x14.1cm, flip angle= 90 degrees, number of acquisitions= 2, slice thickness= 3mm and

scan time = 1 minute 10 seconds). Again three combinations using three different saddle coils were

used and LFOV images acquired. This experiment was then repeated in a 0-90 degrees orientation

to further characterize the interaction. A similar calculation of mutual inductance to the one

described above was also conducted at fixed rotations and by varying the centre-to-centre coil

translations. These results are plotted in figure 14B.

29

Figure 14: Changing the centre-to-centre distance between coils increases SNR while the mutual inductance is

decreased. A. SNR with centre-to-centre distance of two Rx saddle coils at 0-0 degrees (black) and 0-90 degrees

(red). The error bars represent standard deviation values of three measurements. B. Calculated mutual inductance

values between two coils at 0-0 degrees (black line) and 0-90 degrees (red line).

As the distance increases between the two adjacent coils, the mutual inductance and coupling

between them decreases (figure 14). In the first of these experiments the coils were placed in a 0-

0 degrees orientation where mutual inductance is more significant. We observed the expected trend

of decreased inductance and higher SNR. A 25% difference in SNR was seen between 40 and

Translation of Rx Coils

30 40 50 60 70 800

500600

800

1000

1200

1400

0 0 Degrees

0 90 Degrees

Distance Apart (mm)

SN

R

Mutual Inductance Between 2 Saddle Coils

30 40 50 60 70 800

5

10

15

20

250-0 Deg

0-90 Deg

Centre-to-Centre Distance (mm)

Mu

tual In

du

cta

nce (

nH

)

A

B

30

70mm separation. When the pair of coils was placed orthogonal to one another, the mutual

inductance was minimized and a 16% difference in SNR was seen between coils 40 and 70 mm

apart.

2.4.3 Depth

Our final experiment tested how shifting the coils along Lo (figure 6, the B0 or z direction) affected

mutual inductance and SNR. Two three loops saddle coils were oriented at 0 degree (0-0 pair) and

with a centre-to-centre separation of 60mm. One coil was adjusted and moved with respect to Z at

centre-to-centre distances of 0mm, ±20mm, ±40mm, and ±60mm. Images were acquired with the

following parameters: TR=50ms, TE=4.55ms, Matrix:128x704 FOV:5.12x14.1cm, flip angle= 90

degrees, number of acquisitions= 2, slice thickness= 3mm and scan time = 1 minute 10 seconds.

Afterwards a mutual inductance calculation was conducted using the method described above.

These results are plotted in figure 15.

31

Figure 15: Another method to reduce coupling and increase SNR is by shifting the coils along Lo (the z axis). A.

SNR depth results using two Rx coil rotation 0-0 degrees. The error bars represent standard deviation values of

three measurements. B. Calculated mutual inductance values between two coils of different depths.

As the Lo separation increases between the two adjacent coils, the mutual inductance and coupling

between them decreases. Here a 29% decrease in SNR was seen at 0mm compared to ±60 mm.

The results also demonstrate the expected results of symmetry about 0mm.

Rx Coil Shift

-100 -50 0 50 1000

800900

1000

1100

1200

1300

1400

Centre-to-Centre Distance (mm)

SN

R

Mutual Inductance Between 2 Saddle Coils

-100 -50 0 50 1000

1

2

3

4

Z Distance (mm)

Mu

tual In

du

cta

nce (

nH

)

A

B

32

2.4.4 Coupling and SNR

The above experiments confirm that mutual inductance between two coils affects image quality.

This section aims to quantify this relationship more generally. This is achieved through a simple

circuit model (figure 1) which states that mutual inductance will increase the coil impedance, thus

decreasing image SNR.

2.4.4.1 Methods

A simple model expresses SNR as a function of: (i) both coil and sample resistance (R1) and (ii)

the coupling of resistance from adjacent coils (R2) by mutual inductance, equation 14. Thus it can

be shown without any coil coupling (zero mutual inductance) the expected value of only coil

resistance is achieved. Using a multi meter, coil resistance was measured to be one ohm, and

assuming a loading factor of 0.5, sample resistance should also be one Ohm. This yields a total

resistance in each coil of two Ohms. In our model, mutual inductance allows for additional

resistance to be introduced within the coil and thus lower image quality. This can be calculated

using a LRC model of two adjacent three loop saddle coils, as expressed in equation 14:

𝑆𝑁𝑅 𝛼 1

√𝑅1+𝜔2𝑀2

𝑅2

[14]

It is important to indicate that the calculated value from equation 14 only accounts for mutual

inductance between the coils and that capacitive coupling between coils may also occur.

2.4.4.2 Results

Figure 16, 1/ normalized SNR2 is plotted as a function of mutual inductance squared (M2) shown

in equation 14, which should provide a linear relationship. The solid line in the figure represents a

linear regression with 95% confidence intervals shown.

33

Mutual Inductance and SNR Between 2 Saddle Coils

0 20 40 60 80 1000.0

0.5

1.0

1.5

2.0

M2(nH

2)

1/S

NR

2(x

10

-6)

Figure 16: Modeling SNR from M indicates that image quality is decreased due to the addition of impedance from

adjacent coils. 1/SNR2 as a function of predicted mutual inductance squared. This figure contains the results of all

data in 2.4.1, 2.4.2 and 2.4.3. The black line is a linear regression with 95% confidence interval. Colour points

represent rotations: 0-0 red, 60-60 green, 90-90 yellow and 30-60 purple.

Qualitatively, the linear relationship from equation 14 holds. As mutual inductance increases, extra

impedance from the adjacent coil is added and lowers SNR. However, the variance around the fit

line is somewhat high, likely due to other types of coupling, i.e. capacitive coupling. For example,

within the rotation result values (0-0 red, 60-60 green, 90-90 yellow and 30-60 purple), coils are

positioned in a parallel orientation to one another, meaning the rungs along the saddle are all in

the same direction and at the closest possible distance to one another. This may provide increased

capacitive interactions and thus result in a lower SNR than expected. Also, points within the circled

region all have very low mutual inductances and fall below (larger SNR than expected) the 95%

confidence interval. These positions have very low mutual inductances, so that coupling by other

means may dominate SNR outcome.

Despite the possibility of other effects, figure 16 shows that mutual inductance is an important

factor in coupling coils, resulting in a decreased image SNR. This is an important consideration

34

when designing an array to image multiple mice simultaneously, especially if coils are unshielded

and operating in a Tx-Rx mode. Figure 16 suggests that minimization of mutual inductance during

array design provides a means of optimizing the expected SNR. Note that it is likely that a SENSE

reconstruction, which was not used in Figures 13 through 16, can recover some of this SNR and

decrease the SNR dependence on mutual inductance. However, this benefit of SENSE

reconstruction—and indeed SENSE reconstruction in general—is dependent on orthogonal coil

sensitivities, a property also degraded by coil coupling by mutual inductance or other means.

Hence, the degree to which SNR can be recovered is fundamentally limited, and minimization of

mutual inductance (and any other means of coupling) is an important consideration for multiple-

mouse coil arrays.

Array design with respect to coils must consider two dimensions; orientation and distance apart.

With respect to coil rotation, our model of computed SNR as a function of mutual inductance

(figure 16) can be used to determine the ideal coil rotation. In addition, the distance apart must be

kept as large as possible to minimize inductance (and capacitive) coupling to keep image quality

high. Additional factors to take into consideration include the number of mice, bore diameter and

animal handling.

2.5 Theoretical Design of a MMMRI Array

A total of seven mice was chosen to be imaged at once. This number is convenient geometrically

in the dimensions of our gradient bore. The most efficient method to position seven mice within a

bore of 210mm in diameter is to have three rows of mice, with two on the first row (top), three on

the second (middle) row and two on the third (bottom) role. In this configuration the maximum

distance apart is 60mm. However, to accommodate an existing rail system that allows efficient

35

placement of coils and mice in the bore, a centre-to-centre separation distance of 58mm was

chosen. In addition, different depth positions can be considered for possible array configurations.

To determine which orientations are ideal in a seven coil array that minimize the mutual

inductance, we computed all pair wise mutual inductances in the array as seen in section 2.4.

Within a seven coil array system there will be six pairs of nearest coil interactions (adjacent coils

separated by 58mm) and 15 pairs of second order interactions (coils separated by more than

58mm). Here, all 21 mutual inductance pairs functions were used as inputs in a nonlinear least

squares minimization (the Levenberg-Marquardt algorithm) (39) with all 7 coil rotations and z

positions as independent variables. Shown in equation 15:

min𝑥,𝑧

‖𝑓(𝑥)‖22 = 𝑚𝑖𝑛𝑥1−7,𝑧1−7

∑ ∑ 𝑓2(𝑥𝑖 , 𝑥𝑗 , 𝑧𝑖 , 𝑧𝑗)𝑁𝑗=𝑖+1

𝑁−1𝑖=1 [15]

Here each function f represents the mutual inductance of a pair of coils where x is the rotation of

each coil, z the depth position and i,j index the coils for a total of 21 pairs. The sum of all 21 pairs

of mutual inductance squared was then minimized. To ensure we remain within the gradient

volume, a maximum z distance between any two coils was set to be 50mm. Several starting

positions were chosen at random, all of which resulted in similar orientations with symmetric

properties. Shown in figure 17 is the orientation that minimized all 21 local pairs. There is a

consistent symmetry in all the solutions from random start points (figure 17).

36

Figure 17: Random start points for the minimization in equation 15 all produced the same result, shown

above. All results would either have to be rotated or mirrored to produce the above result. Here coil

degrees are shown (A) within each coil and coils 1,3,6 and coils 2,7,4,5 are separated by 50mm in the z

direction (B).

All results of three random start points resulted in a single solution shown in figure 17. The coils

would either have to be rotated or mirrored (in some cases both) to have the same orientation in

figure 17. All orientations have coils at two different z positions, ±25mm or 50mm apart, also coils

that are across from one another (coils, 2-7, 3-6 and 4-7) have the same degree where coil one

would have one of these rotation values (and be placed in the same z plane). Given this solution,

there are three coils pairs, 2-5, 3-6 and 4-7, that have identical properties due to the symmetry of

the array.

Minimizing mutual inductance reduces coupling between coils by minimizing the magnetic flux

from adjacent coils so that no coupling can occur. This is shown in figure 18, where the solid lines

37

represent the general direction of the magnetic field produced in coils one through four. Here the

magnetic field produced by a certain coil offers an almost perpendicular orientation to adjacent

coils, decreasing the flux and therefore mutual inductance.

Figure 18: Coils are orientated such that the produced B1 field (red) crosses adjacent coils at nearly an orthogonal

orientation, thus reducing flux and coupling. These B1 field maps represent the final orientation of coils 1-4 (A-D).

Coils that are positioned within the same plane are coloured the same (1-3-6 and 2-4-5-7). B1from coil 5,6 and 7 (not

shown) have the same results as coils 2,3 and 4 respectively due to symmetry of the array.

The question of how sensitive the minimization is to coil orientation could be quite important in

the design of a MMMRI array, for two reasons: i) it determines tolerance on coil orientation in

array construction and assembly and ii) array usage could permit unintentional coil orientation

shifts. To predict tolerance, total mutual inductance (the sum of all 21 pairs squared) was calculated

for ±30 degrees for all coils from the orientation in figure 17. Results are shown in figure 19. These

results show that all coils are quite stable and have a ±10 degrees tolerance for 0.5 nH of extra

38

total mutual inductance. Therefore we can conclude that our reference position is at a favoured

position and is quite stable. We now have an orientation where all seven coils in the array receive

equally low mutual inductance values and are stable with respect to their orientation of ±10

degrees.

Coil Orientation Sensitivity

-40 -20 0 20 400

1

2

3

4Coil 1

Coil 2

Coil 3

Coil 4

Degree Change

To

tal M

utu

al In

du

cta

nce (

nH

)

Figure 19: Total mutual inductance as a function of degree change of all four coils in our final orientation. Coil

orientations are shown to be robust as ±10 degree change from figure 17 only adding 0.5nH of total M. Coils 2,3,4

are identical with coils 5,6 and 7 (not shown) respectively due to the symmetry of the array.

Next we designed an array for MMMRI. In this component I worked closely with Jun Dazai, a

mechanical engineer at the Mouse Imaging Centre. The array was designed using the CAD

software Autodesk Inventor (San Rafael, California, U.S) and incorporated all necessary

components for live mouse imaging. These include: anesthesia gas delivery, gas scavenging,

temperature monitoring, respiratory monitoring, heat delivery and proper holding for the mouse.

A fully labelled CAD image is shown in figure 20A (courtesy of Jun Dazai) accompanied by a

photo (B). Each coil contains its own tooth bar, where the mouse is held securely in place and

anesthesia delivered. The coil form itself acts to contain the anesthesia so that efficient scavenging

39

can occur, this feeds onto a main manifold where scavenging is achieved through one of the four

outer rods. Respiratory and temperature monitoring lines are fed through each individual mouse's

line. The mouse's body sits in a semi-circular exposed tube that permits imaging of both small and

large mice. Heat is provided through hot air delivered through the bottom two outer rods. All coils

matching circuits are shielded from one another using copper foil to avoid any unnecessary cross

talk. Finally the array is held secured in the magnet bore via a pneumatic compressed air clamp.

The array was constructed using a CNC milling machine with non-magnetic components. Coils

were rotated and positioned to the orientation shown in figure 17.

Figure 20: A. A labelled CAD drawing of a seven multiple mouse live head imaging array, courtesy of Jun Dazai.

The array includes seven three loop saddle RF coil for imaging, semi circular mouse bed to accommodate large

mice, individual anesthesia, scavenging and monitoring per mouse and global heat delivery. B. Photo of the array

with the two lower beds connected and all lines and coils shown.

2.6 Imaging with the array

Once the array was constructed initial tests were completed in phantoms. This was to ensure our

SENSE-like reconstruction was sufficient to eliminate ghosts and to observe any trade off in image

40

quality when having seven coils in an array. Afterwards high resolution live images were acquired

using seven mice and images reconstructed using our SENSE-like method.

2.6.1 Methods

To demonstrate that the SENSE algorithm could eliminate ghosting, phantoms images where

acquired using a cylindrical gradient echo sequence. A cylindrical k-space acquisition, where the

corners of k-space are not acquired in 2 dimensions, was used to save time (40). The cylindrical

acquisition allows us to image in ~28% less time than a standard acquisition. A low resolution

large FOV image that encompassed the entire range to capture all seven phantoms was also

acquired (T1 weighted, TR=60ms, TE=6.05ms, Matrix:388x75x75 FOV:3.5x13.5x13.5cm, flip

angle= 45 degrees, number of acquisitions=1, scan time= 4 minutes 33 seconds). This image was

used to calculate our sensitivity matrix, which was used to eliminate ghosts in our desired high

resolution small field of view (SFOV) images (T1 weighted, TR=60ms, TE=6.05ms,

Matrix:388x222x278 FOV:3.5x2.0x2.5cm, flip angle= 45 degrees, number of acquisitions= 3,

scan time = 2 hour 28 minutes). Once the LFOV images were acquired a 7x7 sensitivity matrix

was created. This matrix contains cross talk coefficients, from each coil to every other coil (with

diagonal entries equal to unity). This matrix was computed using the large FOV data, where for

each coil each ghost average intensity value was divided by its own average intensity value, much

like a normalization. This was completed for both the magnitude and phase (complex conjugate)

of the data. This provided our sensitivity matrix or C values in equation 9.

Live mouse imaging were acquired using the same two sequences describes above. 24 hours before

imaging mice were injected with a manganese chloride solution (0.4 mmol/kg) to provide the

desired contrast. The mice were anesthetized using 2% isoflurane, maintained at 1-1.5%

isoflurane, and kept at a constant temperature of 29 degrees Celsius. All animal experiments were

41

approved by the Animal Care Committee at the Toronto Centre for Phenogenomics (animal use

protocol 0227).

2.6.2 Results The pre (figure 21 A) and post SENSE (figure 21 B) phantoms images are shown for all coils.

Ghosts were measured by dividing the signal intensity from a ROI within the phantom by another

signal intensity ROI of the ghost. Before processing the ghosts values were on average 5% per

sample, however after the SENSE-like reconstruction ghosts were eliminated to less than

detectable levels (within the noise of the image ~1%). These images provided an average of 6±2%

decrease in SNR when compared to a single isolated coil.

Figure 21: Ghosting artifacts are fully removed by using our SENSE like reconstruction method. Pre (A) and post

(B) SENSE reconstruction for all coils. Ghosting (pointed by white arrows) signal intensity drops from ~5% per

sample to undetectable levels (<1%) using a modified SENSE reconstruction.

Figure 22 shows post SENSE reconstruction images of seven live mice. This figure demonstrates

the capability of the new array, here high resolution (90µm) and high image quality is made

possible in a respectable scan time (2 hours 28 minutes). Panel A is displays a horizontal slice of

seven mice and panel B is a three section view (coronal, horizontal and sagittal) of the mouse in

42

coil three. All images provide sufficient SNR for image analysis (32±5, panel A) (41). In these

images layers in the cortex (right side) and the hippocampus (center white bands) are clearly

visible. In addition an average image was created using all seven mice, figure 23.

Figure 22: High resolution (90µm) and high SNR (32±5) brain images of seven live mice post SENSE

reconstruction using the array from figure 20. These images offer excellent image contrast and a high level of detail

that can be used for image analysis. A. Horizontal slice of all 7 mice. B. Enlarged 3 axis view (sagittal, coronal and

horizontal) for the mouse in coil 3.

43

Figure 23: Three dimensional average image of all seven mice from figure 22. Coronal (top), horizontal (middle)

and sagittal (bottom). This average can be used as a template for image analysis for future experiments.

44

Chapter 3

MMMRI Coil Array

3.1 Benefits My investigation of RF coil configurations for live mouse brain imaging resulted in the design of

a MMMRI array with three loop saddle coils operating in a Tx-Rx mode. Even without shields and

when operating in a seven coil array, coil coupling could be minimized using geometry of coil

placement and then corrected using a SENSE-like reconstruction. One can trade off this added

benefit, according to the relationship:

𝑆𝑁𝑅 ∝ ∆𝑥∆𝑦∆𝑧√𝑁𝑎𝑣𝑔𝑁𝑝𝑒 [16]

One can i) increase resolution (by decreasing Δx, Δy and Δz) and/or ii) decrease scan time (reduce

number of averages, Navg or number of phase encoding steps, Npe). For example, given equation

16, a 50% increase in SNR will provide a 17% increase in resolution for an isotropic 3D image.

Live mouse imaging is typically SNR limited and there is a limited time a mouse could be

anesthetized, scan time for mice must be limited to ~3 hours. Therefore, resolution is often

sacrificed in order to acquire sufficient signal. When resolution is sufficient, the increased SNR

provided by this array can benefit many imaging protocols. For example, functional MRI uses

blood oxygen level dependant (BOLD) signal and depends on the detection of small signal changes

(a few percent). Having higher SNR results in higher sensitivity to the BOLD effect.

Some longitudinal studies may now be more feasible. This involves increasing resolution while

decreasing scan time. The high resolution capability of sensitive coils will make it possible to

image young mice and still be able to detect brain differences. Here the increased coil sensitivity

may be traded for shorter scans (25% shorter), allowing more frequent scanning sessions to occur.

45

For example, a young mouse could be imaged at post natal day 14 with subsequent imaging every

two to three days afterwards to track more rapid changes in early development. I anticipate that

this is one of the greatest benefit of the new saddle-coil array for mouse imaging.

3.2 Potential Use

Interest in 3D imaging in biology has increased throughout the past 10 years. However, standard

imaging modalities based on histological techniques and microscopy are cumbersome to adapt to

3D. Three-dimensional MRI minimizes distortion artifacts produced during fixation, embedding,

sectioning and mounting for histological methods. MRI also provides excellent soft tissue contrast

and the ability to perform in vivo experiments.

My seven coil live mouse imaging array provides high throughput and high quality images and

could be implemented in any laboratories that already have access to a modern MRI scanner. The

RF design and matching circuit can easily be constructed and do not require expensive

components. Labs inexperienced with multiple-mouse MRI may encounter three difficulties

associated with implementation of the seven coil array: i) Instrumentation, ii) Construction and iii)

Image reconstruction. With respect to instrumentation, current clinical MRI scanners are being

equipped with multiple Tx and Rx channels to allow for more efficient human imaging. In many

cases, a scanner will have up to 16 Tx and 32 Rx channels, therefore up to 16 samples could be

imaged in a Tx-Rx mode for multiple specimens. In addition, clinical scanners already have large

bore sizes that would provide enough room for a multiple sample array. However two limiting

factors of clinical systems are their magnetic field strength and gradient strength. The lower field

strength of clinical systems results in a lower SNR whereas a smaller gradient strength would limit

image resolution. The saddle design would promote high sensitivity to overcome low field

46

strength, however gradient strength may still limit resolution. Overall, however, modern MRI

systems are equipped with the instrumentation required to operate the multiple-mouse coil array

designed in this thesis. Construction of the array is an issue for labs that do not have their own mill

nor CAD software. However, this could potentially be outsourced for a relatively low cost. As

shown in my second chapter, coils should be placed as far apart as possible given the number of

coils and gradient boundaries. Coil rotation can be determined using the mutual inductance least

square minimization method. This will ensure that coupling is at a minimum and image quality

high. Finally image reconstruction will be achieved using a SENSE-like method. Clinical systems

do come equipped with standard SENSE methods, where my method could easily be adapted;

since coil sensitivities are kept constant over an image, ghosts can be easily removed via a simple

subtraction.

3.3 Custom Coils

This thesis introduces an idea that has been practised in clinical systems over the past few years.

That is, custom coils or arrays tailored to a specific applications or imaging protocols. In this thesis,

a custom coil was designed for live mouse head imaging. After a coil was designed and built, a

multiple mouse imaging array was constructed. Generic systems with large volume coils have the

potential to suffer in two ways, i) image quality and ii) imaging efficiency (handling and

throughput). Creating custom coils and/or arrays solves these issues. During my time at the Mouse

Imaging Centre, an investigation was completed where a custom set up was needed and

constructed for intubated mice for arterial spin labeling (ASL) and cerebral blood volume (CBV)

measurements. Projects were completed with Dr. Cahill from the Mouse Imaging Centre and

apparatus design and construction was completed by our mechanical engineer Jun Dazai.

47

The purpose of the ASL and CBV study is to observe structural changes in the brain vascular

network of mice that have undergone a regime of voluntary exercise. The mice are exposed to

hypercapnic conditions (5% CO2) to measure changes in blood flow and volume. For strict

control of the gas conditions, the mice are intubated during scanning. However intubated mice

required additional lines and components that were not incorporated in the existing head imaging

array. Two additional separate gas lines (inhale and exhale) were required to control breathing

and inhalation of different gas mixtures. None of our current coil arrays allowed us to image

intubated mice and the smaller three loop saddle coil did not provide sufficient space for the

mouse and additional lines, therefore a large three loop saddle coil was constructed with an ID of

25mm. Also an open concept removable mouse bed was built, that allowed for the imaging for

large mice and easy mobility from the bench (where the intubation was done) to the

scanner. Figure 24 displays a photo of the coil apparatus. We now have an efficient method to

image a live intubated mouse.

Figure 24: Custom coils allow for the design of very specific and unique experiments. Here, a larger three loop

saddle coil is used to accommodate an intubated mouse for dedicated brain imaging. Equipped with respiratory

monitoring (blue tube), temperature monitoring (brown wire) and intubation lines (white tubes).

48

3.4 Summary

Mice are the mammals of choice in the laboratory for cancer and genetic studies because they can

be manipulated genetically. MRI offers a method to quantify and track possible changes in models

using imaging. To ensure findings are statically meaningful, many mice must be imaged for a

particular study. MMMRI offers a solution, where multiple mice are imaged simultaneously. This

makes it possible to undertake large mouse studies in a reasonable time frame. There have been

many RF configurations of MMMRI, however possible transmit and shielding configurations have

not yet been explored. The purpose of this thesis was to investigate possible RF configurations to

image the brains of multiple live mice.

The study began with designing a multiple loop saddle coil that proved to provide increased SNR

and sufficient B1 homogeneity over the mouse brain. Following, four possible RF configurations

were explored, resulting in a Tx-Rx saddle coil without shielding. This prompted an investigation

of coupling between exposed coils through a mechanism of mutual inductance. It was shown that

coupling could be minimized through two means, i) rotating adjacent coils and ii) separating the

coils, both of which resulted in increased image quality. The final multiple live imaging array was

then designed using a minimization algorithm which minimized mutual inductance between all

seven coils in the array. The array resulted in only a 6% loss in image quality when compared to a

single coil. High resolution images (90µm) were acquired and a modified SENSE reconstruction

method allowed single images to be properly reconstructed and analyzed. Alternatively, the added

sensitivity could be used to reduce scan time or allow for other types of scanning protocols such

as BOLD imaging which requires high SNR.

49

Possible improvements with the current array may include incorporating capacitive coupling (in

addition to mutual inductance) minimization which may further decrease coupling and improve

image quality. Also, the multiple loop saddle coil may be further optimized by configuring a

possible quadature design and investigating the configuration and number of segmentation

capacitors. In addition, we limited our design to seven coils (samples). Further investigations could

be done to accommodate more coils, which would inherently have more coupling and a different

solution. In this thesis only one type of pulse sequence was used, a gradient echo. Experiments

need to be completed with different pulses sequences (for example a spin echo) where coil

coupling on transmit could result in image artifacts and poor image quality. Finally, the benefits

of having a higher resolution and SNR still need to be investigated. For example, in the detection

of anatomical brain differences between groups of mice, perhaps higher resolution may show

differences not previously seen with lower resolution images.

50

References

1. R.M. Henkelman. Systems Biology through Mouse Imaging Centers: Experience and New

Directions. Annual Review of Biomedical Engineering 12:143:166, 2010.

2. B.J. Nieman, M.D. Wong, R.M. Henkelman. Genes into Geometry for Mouse Development in

3D. Current Opinion in Genetics and Development 21:638-646, 2011.

3. Nadeau JH, Balling R, Barsh G, Beier D, Brown SDM, Bucan M, Camper S, Carlson G,

Copeland N, Eppig J, Fletcher C, Frankel N, Ganten D, Goldowitz D, Goodnow C, Guenet J,

Hicks G, Hrabe de Angelis M, Jackson I, Jacob HJ, Jenkins N, Johnson D, Justice M, Kay S,

Kingsley D, Leharch H, Magnuson T, Meisler M, Poustka A, Rinchik EM, Rossant J, Russell LB,

Schimenti J, Shiroishi T, Skarnes WC, Soriano P, Stanford W, Takahashi JS, Wurst W, Zimmer

A. Functional annotation of mouse genome sequences. Science;291:1251–1255, 2001.

4. Moats RA, Velan-Mullan S, Jacobs R, Gonzalez-Gomez I, Dubowitz DJ, Taga T, Khankaldyyan

V, Schultz L, Fraser S, Nelson MD, Laug WE. Micro-MRI at 11.7 T of a murine brain tumor

model using delayed contrast enhancement. Mol Imaging;2:150–158,2003

5. Bock NA, Zadeh G, Davidson LM, Qian B, Sled JG, Guha A, Henkelman RM. High-resolution

longitudinal screening with magnetic resonance imaging in a murine brain cancer model.

Neoplasia, 2003; 5(6): 546–554.

6. L.M. Gazdzinski, B.J. Nieman. Cellular Imaging and Texture Analysis Distinguish Differences in

Cellular Dynamics in Mouse Brain Tumors. Magn. Reson. Med. (2013)

7. B.J. Nieman, J.B. Bishop, J. Dazai, J.P. Lerch, A. Feintuch, J.G. Sled, R.M. Henkelman. MR

Technology for Biological Studies in Mice. NMR in Biomedicine 20(3):291-303, 2007.

8. N.A Bock, N.B Konyer, R.M. Henkelman. Multiple Mouse MRI. Magn. Reson. Med. 49(1):158-

167, 2003.

9. N.A Bock, B.J. Nieman, J.B. Bishop, R.M. Henkelman. In Vivo Multiple Mouse MRI at 7 Tesla.

Magn. Reson. Med. 54(5):1311-1316, 2005.

10. J.P. Lerch, J.G. Sled, R.M. Henkelman. MRI Phenotyping of Genetically Altered Mice. Method

in Molecular Biology 711:349-361, 2011.

11. J.E. Bishop, S. Spring, J. Dazai, B.P. Chugh, S. Portnoy, S. Suddarth, G.R. Morris GR, R.M.

Henkelman. Multiple mouse imaging of 16 live mice. Proceedings of the 16th Annual Meeting

ISMRM, Toronto, ON, Canada,1176, 2008.

12. X. Zhang, J.E. Schneider, S. Portnoy, S. Bhattacharya, R.M. Henkelman. Comparative SNR for

High-Throughput Mouse Embryo MR Microscopy. Magn. Reson. Med. 63:1703–1707, 2010.

13. M. Mitsuda, M. Yamaguchi, T. Furuta, A. Nabetani, A. Hirayama, A. Nozaki, M. Niitsu, H. Fujii.

Multiple-animal MR Imaging using a 3T Clinical Scanner and Multi-channel Coil for Volumetric

Analysis in a Mouse Tumor Model. Magnetic Resonance in Medical Sciences: 10(4): 229-237,

2011.

14. A. Nabetani, M. Yamaguchi, A. Hirayama. Multiple-animal MR imaging performed using a 3-

Tesla whole-body scanner: A feasibility study using a 16-chanel array coil. Proc Intl Soc Mag

Reson Med; 18:1488, 2010.

15. P.B Roemer, W.A. Edelstein, C.E Hayes, S.P. Souza, O.M. Mueller. The NMR phased array.

Magn. Reson. Med:16(2): 192–225, 1990.

16. K.P. Pruessmann, M. Weiger, M.B. Scheidegger, P. Boesiger. SENSE: sensitivity encoding for

fast MRI. Magn. Reson. Med; 42(5): 952–962, 1999.

51

17. M.S. Ramirez, E. Esparza-Coss, J.A. Bankson. Multiple-mouse MRI with multiple arrays of

receive coils. Magn. Reson. Med; 63(3): 803–810, 2010.

18. M.S. Ramirez, S.Y. Lai, J.A. Bankson. A throughput-optimized array system for multiple-mouse

MRI. NMR Biomed; 26: 237–247, 2013.

19. Y. Matsuda, S. Utsuzawa, T. Kurimoto, T. Haishi, Y. Yamazaki, K. Kose, I. Anno, M. Marutani.

Super-Parallel MR Microscope. Magn. Reson. Med; 50:183–189, 2003.

20. D.G. Nishimura. Principles of Magnetic Resonance Imaging. Lulu 1.1, 2010.

21. F. D. Doty, G. Entzminger, J. Kulkarni, K. Pamarthy and J.P. Staab. Radio frequency coil

technology for small-animal MRI. NMR Biomed; 20: 304–325, 2007.

22. D.I. Hoult, R.E. Richards. The Signal-to-Noise Ratio of the NMR Experiment. J. Magn. Reson;

24: 71–85, 1976.

23. D.I. Hoult. Sensitivity of the NMR Experiment. In Encyclopaedia of NMR Vol. 7. Wiley:

Chichester; 4256–4266, 1996.

24. J.S. Tropp. Radiation damping and reciprocity in magnetic resonance reception. Presented at 12th

ISMRM. Poster 1646. Kyoto, 2004.

25. E. Fukushima, B. Stephen, W. Roeder. Experimental pulse NMR, A nuts and bolts approach.

Addison Wesley p.297 386,1981.

26. D.I. Hoult. The NMR receiver: a description and analysis of design. Progress in NMR

Spectroscopy; 12: 41–77, 1978.

27. J.J.H. Ackerman. Surface (local) coils as NMR receivers. Concepts in Magnetic Resonance; 2: 23–

42, 1990.

28. J. Hyde. Surface and other local coils for in vivo studies. In Encyclopaedia of NMR Vol. 7. Wiley:

Chichester, ;4656–4664, 1996.

29. B.H Suits, A.N. Garroway, J.B. Miller. Surface and gradiometer coils near a conducting body: the

lift-off effect. J. Magn. Reson; 135: 373–379, 1998.

30. W.A. Edelstein, C.J. Hardy, O.M. Mueller. Electronic decoupling of surface-coil receivers for

NMR imaging and spectroscopy. J Magn Reson; 67: 156–161, 1886.

31. Tropp J. Mutual inductance in the birdcage resonator. J Magn Reson; 126: 9–17, 1997.

32. W. Edelstein. RF Systems and Coils for MRI and MRS. In Encyclopaedia of NMR Vol 6. Wiley:

Chichester, 3950-3954, 1996.

33. Y. Li, A.G. Webb, S. Saha, W.W. BreyW, C. Zachariah, A.S. Edison. Comparison of the

performance of round and rectangular wire in small solenoids for high-field NMR. Magnetic

Resonance in Chemistry; 44: 255–262, 2006.

34. D.M. Ginsberg, M.J. Melchner. Optimum geometry of saddle shaped coils for generating a

uniform magnetic field. Rev. Sci. Instrum; 41: 122–123,1970.

35. F.D. Doty, G. Entzminger Jr. Center-fed paralleled saddle coils for multinuclear double-

resonance NMR or MRI. U.S. Pat. 6,175,237, 2001.

36. C.E. Garrido-Salomon, E.L. Vidoto, M.J. Martins, A. Tannuns. Optimization of saddle coils for

magnetic resonance imaging. Braz J Phys 36(1A):4–8, 2006.

37. W. Domingguez-Viqueira, W. Berger, J. Parra-Robles, G.E. Santyr. Litz Wire Radiofrequency

Receive Coils for Hyperpolarized Noble Gas MR Imaging of Rodent Lungs at 73.5 mT Concepts

in Magnetic Resonance Part B: Vol. 37B(2) 75–85, 2010.

38. C.M. Collins, L. Shizhe, X. Qing, M.B. Smith. A Method for Accurate Calculation of B1 Fields

in Three Dimensions, Effects of Shield Geometry on Field Strength and Homogeneity in the

Birdcage Coil. Magn. Reson. Med. 125(1):233-241, 1997.

39. J.E. Dennis. Nonlinear Least-Squares. State of the Art in Numerical Analysis ed. D. Jacobs,

Academic Press, pp. 269–312. 1977.

52

40. B.J. Nieman, N.A. Bock, J. Bishop, J.G. Sled, X. J. Chen, R.M. Henkelman. Fast Spin-Echo for

Multiple Mouse Magnetic Resonance Phenotyping. J. Magn. Reson; 54: 532–37, 2005.

41. S.C. Kale, J. P. Lerch, R.M. Henkelman, X.J. Chen. Optimization of the SNR-Resolution Trade

off for Registration of Magnetic Resonance Images. Human Brain Mapping; 29: 1147-58, 2008.