30
1 Original Papers Comparative Study of the Electrochemical Sensors Based on Enzyme Immobilized into Polyelectrolyte Microcapsules and into Chitozan Gel Yulia V. PLEKHANOVA,* Sergei A. TIKHONENKO,** Alexey V. DUBROVSKY,** Alexander L. KIM,** , *** Egor V. MUSIN,** , *** Gou-Jen WANG,**** Iren E. KUZNETSOVA,***** Vladimir V. KOLESOV,***** and Anatoly N. RESHETILOV* †, ***** * PSCBR RAS G.K. Skryabin Institute of Biochemistry and Physiology of Microorganisms, Russian Academy of Sciences, Prospect Nauki 5, Pushchino, Moscow Region 142290, Russia ** FSBIS Institute of Theoretical and Experimental Biophysics, Russian Academy of Sciences, Institutskaya st. 3, Pushchino, Moscow Region 142290, Russia *** Lomonosov Moscow State University, Leninskie Gory, Moscow 119991, Russia **** Department of Mechanical Engineering, National Chung-Hsing University, Taichung 402, Taiwan ***** A. Kotelnikov Institute of Radio Engineering and Electronics, Russian Academy of Sciences, Mokhovaya st. 11-7, Moscow 125009, Russia To whom correspondence should be addressed. E-mail: [email protected], [email protected] Analytical Sciences Advance Publication by J-STAGE Received April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019 DOI: 10.2116/analsci.19P131

Comparative Study of the Electrochemical Sensors Based on

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

1

Original Papers

Comparative Study of the Electrochemical Sensors Based on Enzyme

Immobilized into Polyelectrolyte Microcapsules and into Chitozan Gel

Yulia V. PLEKHANOVA,*† Sergei A. TIKHONENKO,** Alexey V. DUBROVSKY,**

Alexander L. KIM,**,*** Egor V. MUSIN,**,*** Gou-Jen WANG,**** Iren E.

KUZNETSOVA,***** Vladimir V. KOLESOV,***** and Anatoly N. RESHETILOV*†,*****

* PSCBR RAS G.K. Skryabin Institute of Biochemistry and Physiology of Microorganisms,

Russian Academy of Sciences, Prospect Nauki 5, Pushchino, Moscow Region 142290, Russia

** FSBIS Institute of Theoretical and Experimental Biophysics, Russian Academy of Sciences,

Institutskaya st. 3, Pushchino, Moscow Region 142290, Russia

*** Lomonosov Moscow State University, Leninskie Gory, Moscow 119991, Russia

**** Department of Mechanical Engineering, National Chung-Hsing University, Taichung 402,

Taiwan

***** A. Kotelnikov Institute of Radio Engineering and Electronics, Russian Academy of Sciences,

Mokhovaya st. 11-7, Moscow 125009, Russia

† To whom correspondence should be addressed.

E-mail: [email protected], [email protected]

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

2

Abstract

The characteristics of the electrochemical biosensor based on a Prussian-blue screen-printed

electrode containing glucose oxidase incorporated into polyelectrolyte microcapsules (PMC) are

considered. PMC with the embedded enzyme were formed using sodium polystyrene sulfonate

and poly(allylamine hydrochloride). The characteristics were compared with those of the enzyme

immobilized in chitosan gel. We assessed the dependences of biosensor signals on composition

of the buffer solution, on glucose concentration; the operational and long-term stabilities. The

enzyme immobilized in PMC proved more sensitive to buffer molarity at a maximum within 35–

40 mM. The apparent Michaelis constants were 1.5 and 4.1 mM at the immobilization in,

respectively, chitosan and PMC. The developed biosensors were used to assay commercial juices.

The biosensors’ data on glucose contents were shown to have a high correlation with the standard

spectrophotometric assay (0.92–0.95%), which implies a possible application of the fabricated

biosensors in foodstuff analysis.

Keywords: Electrochemical biosensor, screen-printed electrode, glucose oxidase, polyelectrolyte

microcapsules, chitosan, glucose.

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

3

Introduction

Development of biosensors and biofuel cells is associated with the search for and

development of methods for immobilization of the biocatalyst on the electrode surface, so that

the biocatalyst is active for as long as possible. For this reason, the search for novel biocompatible

materials to create electrodes, the use of new polymers to bind the biocatalyst on the electrode

surface is topical.

Activity of enzymes can be preserved at a high level by encapsulation. The enzymes are

immobilized in capsules from various materials, such as liposomes1, polymers2, sol-gels3,

hydrogels4, etc. The layer-by-layer coating of oppositely charged polyelectrolytes for

immobilization of enzymes5–7 enables their use as microcontainers and microreactors in various

industries, for example, food8, cosmetic9, pharmaceutical10. Earlier, for urease and lactate

dehydrogenase, it has been shown that immobilization of enzymes in polyelectrolyte

microcapsules (PMC) makes it possible to maintain enzyme activity at a high level for several

months and also protects encapsulated enzymes from proteinase K11; encapsulated catalase is

protected by PMC from the action of microorganisms and protease12. Enzymes immobilized in

polyelectrolyte capsules can be used to create biosensors, e.g., for potentiometric assays of urea13–

15 or acetylcholine16, optical (fluorescent) determination of glucose.17

The catalytic properties of glucose oxidase (GOx) have been well investigated. At the same

time, immobilization of the enzyme in polyelectrolyte capsules may change its structure and, as

a consequence, its catalytic activity. In work18 the binding of a number of negatively charged

polyelectrolytes to lactate dehydrogenase and glutamate dehydrogenase has been shown to break

down the tertiary and partially secondary structure of the enzymes and, in this way, to totally

inactivate their activity. In study20, the use of polyelectrolytes of opposite chirality has also led to

change the activity of immobilized horseradish peroxidase and GOx.

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

4

One of the methods to control the catalytic activity of the enzyme in biosensors is the

determination of hydrogen peroxide released during the oxidation of the substrate. An

electrocatalyst of hydrogen peroxide reduction is iron hexacyanoferrate (Prussian blue), which is

used to modify the measurement electrodes.20,21 The possibility of selective determination of

hydrogen peroxide by its reduction reaction in the presence of oxygen on electrodes with iron

hexacyanoferrate at a zero potential and pH close to neutral has been demonstrated in work.22

The aim of this work was to compare the characteristics of the glucose biosensor prepared

using two techniques of immobilization of the enzyme glucose oxidase on the surface of a

Prussian-blue screen-printed electrode – encapsulation of the enzyme in polyelectrolyte layers or

in chitosan gel. Immobilization into chitosan for comparison was chosen because, due to the

attractive properties of chitosan (biocompatibility, biodegradability and ease of use), this

technique is often used to develop biosensors.23,24

Experimental

Reagents

The following reagents were used: dibasic trihydrate potassium phosphate, sodium

hydroxide, sodium chloride, glucose, acetic acid (Diakon, Russia); iron(III) chloride, potassium

chloride, calcium chloride, sodium carbonate (Khimmed, Russia); hydrochloric acid, hydrogen

peroxide (30% solution), sodium carbonate (Reakhim, Russia); chitosan (low molecular weight),

potassium hexacyanoferrate(III), sodium polystyrene sulfonate (PSS, 70 kDa), poly(allylamine

hydrochloride) (PAH, 70 kDa), ethylenediaminetetraacetic acid (EDTA), 4-aminoantipyrine,

phenol, glucose oxidase (EC 1.1.3.4) from Aspergillus niger (activity, 185000 U/g), horseradish

peroxidase (1.11.1.7) (activity, 1 kU/g) (Sigma-Aldrich, USA).

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

5

Formation of a Prussian blue-based chemical sensor

Screen-printed three-contact electrodes (Color Electronics, Moscow, Russia) were used as

working electrodes. Prussian blue was precipitated on the electrode surface from the reaction

mixture containing 0.1 M FeCl3 and 0.1 M K3[Fe(CN)6] in the background electrolyte (0.1 M KCl,

0.1 M HCl). The electrode was activated (a potentiodynamic treatment of the Prussian blue

electrode in a cyclic mode in the background electrolyte within the range of potentials from – 0.05

up to 0.35 V) in accordance with the method described in work.22

Fabrication of polyelectrolyte microcapsules with incorporated glucose oxidase

Formation of composite CaCO3–protein microspherolites. An equal volume of a sodium

carbonate solution (0.33 M) was added to a 0.33-M calcium chloride solution containing 3 mg/ml

GOx and intensively stirred on a magnetic mixer. The mixing was continued for 30 s, after which

the suspension was kept up to the complete sedimentation of formed particles.25 Maturation of

microspherolites was monitored by a light microscope. Then the supernatant was decanted, the

sediment was washed with water and used to produce PMC. The particles formed were 5±1 μm

in size.

Preparation of PMC. Polyelectrolyte microcapsules were produced by alternate adsorption

of oppositely charged polyelectrolytes on disperse microparticles (nuclei) followed by the

dissolution of the nuclei. Alternate adsorption of PAH and PSS on the surfaces of CaCO3

microspherolites was carried out in polyelectrolyte solutions (2 mg/ml) containing 0.5-M NaCl.

Each adsorption step was followed by a triple washing of fabricated capsules with a 0.5-M NaCl

solution, which was required to discard non-adsorbed polymer molecules. Particles were

separated from the supernatant by centrifugation. After a required number of layers was deposited,

carbonate nuclei were dissolved in a 0.2-M solution of EDTA for 12 h. Produced capsules were

washed three times with water to remove nuclei’ decomposition products. Capsules contained 6

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

6

polyelectrolyte layers. The concentration of protein in a capsule was 0.66 ng/capsule. The

suspension of capsules at a concentration of 3×109 particles/ml was stored in bidistilled water at

a temperature of 4С. Formation of polyelectrolyte capsules around the enzyme is shown

schematically in Fig. 1.

Formation of the glucose biosensors

1) Polyelectrolyte capsules in the amount of 5 μl with incorporated GOx were applied onto

the surface of a Prussian blue working electrode and dried at room temperature for 30 min. The

enzyme concentration on the electrode surface was 139 μg/cm2.

2) A mixture of GOx (6 mg/ml) and chitosan (2% solution in 1% acetic acid26) at a volume

ratio of 1:1 was prepared; 5 μl of the mixture was applied onto the surface of a Prussian blue

working electrode and air-dried for 30 min at room temperature. The concentration of the enzyme

on the electrode surface was 211 μg/cm2.

Between measurements, the biosensors were stored at a temperature of 4ºС in the dark. The

measurements were carried out at a temperature of 20ºС in a 1-ml cuvette at constant stirring in a

sodium–potassium–phosphate buffer solution. The molarity, pH, content of sodium chloride in

the buffer were varied. The measurements were performed on an EmStat 3 potentiostat (PalmSens,

Netherlands) at a constant zero potential fed to the working electrode relative to the reference

electrode (Ag/AgCl). The measured parameter was the amplitude of the signal; it was assessed

by the difference between the initial and final levels before and after the introduction of glucose.

As real samples to determine the concentration of glucose, we used beverages diluted

tenfold with a buffer solution to a concentration required for the assay: 1) Global Village

reconstituted clarified apple juice, Yuzhnaya Sokovaya Kompaniya, Russia; 2) Fruto-Nyanya

reconstituted clarified apple juice, Progress Ltd, Russia; 3) Fruktovyi Sad orange pulp nectar,

Lebedyansky Ltd, Russia; 4) Fruktovyi Sad multifruit pulp nectar, Lebedyansky Ltd, Russia; 5)

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

7

Home-made wine, 2-year-aged, Izabella grape; 6) Aloe-Fresh non-alcoholic non-carbonated

pasteurized beverage, Velta-Penza Ltd, Russia; 7) El Kurador Sangria sugary wine drink, Hauser

Weinimport GmbH, Germany.

Spectrophotometric assay of glucose

The assay is based on the photometric determination of a compound coloured during the

oxidation of chromogenic substrates.27 Oxidation of glucose by air oxygen in the presence of

glucose oxidase produces hydrogen peroxide. Hydrogen peroxide oxidizes the chromogenic

substrate (4-aminoantipyrine) in the presence of phenol and peroxidase (catalyst) to produce a

coloured compound, quinoneimine. The coloration intensity of this compound is directly

proportional to the concentration of glucose in the sample. Measurements were carried out using

a Specord m40 spectrophotometer at a wavelength of 500 nm.

Results and Discussion

Dependence of biosensor signals on buffer solution

The optimal pH for glucose oxidase to function is pH 5.6.28 The maximal amplitudes of

biosensor responses during the immobilization of the enzyme in chitosan gel were observed

within the range of pH 6.0–7.0; during the immobilization in PMC, pH 5.5–6.0 (Fig. 2 (A)). An

insignificant shift of the optimum to the alkaline region and a broadening of the optimal pH range

is, most probably, due to conditions created by the immobilization of the enzyme.

The catalytic activity of the immobilized enzyme was also affected by the molarity of the

buffer solution (Fig. 2 (B)). Within the investigated range of its changes for the enzyme

immobilized in chitosan, biosensor signals are seen to grow monotonically; the signal amplitudes

reach the greatest values at 40–50 mM of the buffer solution. At the same time, for the

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

8

encapsulated enzyme the amplitude depends on the value of molarity in a bell-shaped way; the

maximum in this case is observed at a buffer solution concentration of 30–40 mM.

A high ionic strength of a solution (greater than 100 mM) affects negatively the stability of

polyelectrolyte complexes29 but increases its conductance by facilitating the transfer of protons

onto the electrode. This is reflected on the generation of biosensor and biofuel cell signals30;

addition of sodium chloride leads to decrease the internal resistance of the system and to increase

power.31 Figure 2 (C) plots dependences of biosensor signals on NaCl concentrations in the buffer.

An increase of NaCl concentration increases the signal, whereas for the encapsulated enzyme at

a NaCl concentration of 25 mM and higher the signal decreases. The observed dependence can

be partially explained by the data of Fig. 2 (B): at a buffer solution molarity rise the dependence

of the signal amplitude passes through a maximum; as the concentration of sodium chloride is

increased, we observe in fact a non-monotonic dependence, which also passes through a

maximum. Thus, it can be summed up that high concentrations of sodium chloride (greater than

80 mM) have a negative effect on the activity of the enzyme, which leads to decrease biosensor

signals; an increase of the concentration beyond 30 mM has a negative effect on the activity of

the enzyme in the capsule. As seen from the comparison of Figs 2 (B) and 2 (C), potassium ions

render a more negative effect (the buffer molarity increases due to both potassium and sodium

ions), whereas an increase in the concentration of sodium ions (Fig. 2 (C)) beyond the level of 30

mM leads to a drop in the level of biosensor signals by ~10% (and in the case of a joint increase

of the concentrations of potassium ions and sodium ions beyond the level of 30 mM the drop is

~44%).

Calibration dependences of biosensor signals on glucose concentration

Dependences of biosensor signals on glucose concentration are plotted in Fig. 3. Major

characteristics of the investigated biosensors are given in Table 1. The ranges of glucose

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

9

determination using the encapsulated enzyme differ from those for the enzyme immobilized in

chitosan gel. Thus, the lower and upper detection limits are higher, and the linear range is broader,

in the case of the encapsulated enzyme. The sensitivity coefficient, however, is higher when using

chitosan for immobilization (1.17 vs 0.12 μA/mM). Works6,32 have shown that the layer-by-layer

application of oppositely charged polyelectrolytes does not prevent the penetration of low-

molecular compounds into the interior of the capsule. Glucose is a low-molecular compound, so

the change of enzyme activity can be due only to immobilization. As is known, the Michaelis

constant shows the extent of affinity of substrate and enzyme33: the lower the value of the constant,

the higher the affinity of the enzyme to substrate is. Thus, immobilization of the enzyme in PMC

leads to decrease the affinity of the enzyme to substrate (Table 1).

The reproducibility of the results of measurements by biosensors was assessed based on

the statistical treatment of a sampling of 15 measurements for a glucose concentration of 1

mM. The variation coefficient for the biosensor with the encapsulated enzyme was 4.5%;

the mean value of the signal and the standard deviation were 151.31 ± 6.88 nA. For a

biosensor with GOx immobilized in a chitosan gel, the variation coefficient was 4.6%; the

mean value of the signal and the standard deviation 587.37 ± 27.08 nA.

Assessment of long-term stability

For GOx incorporated into polyelectrolyte capsules, the signal was ~75% of the initial level

after 5 months of storage of the biosensor between measurements at 4С; after 8 months, 20%

(Fig. 4 (A)). For GOx immobilized in chitosan gel, the biosensor signal was 43% of the initial

level after 2 months; 13%, after 1 year (Fig. 4 (B)). Table 2 shows changes in the major parameters

of the biosensors in storage.

For capsules with the enzyme stored in bidistilled water at 4С for 11 months followed by

their immobilization on the electrode the biosensor signal was about 50% of the initial level (data

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

10

not shown). Therefore, the activity of the enzyme decreases slower for capsules with the enzyme

stored in solution than those immobilized on the electrode, the biosensor being stored dry at 4С.

Thus, encapsulation contributes to preserving the activity of the enzyme, including as a biosensor

component, at a sufficiently high level for a long time, which is consistent with data for other

enzymes, e.g., for urease.11,15

Both methods of immobilizing GOx on the surface of a screen-printed electrode may find

use in the development of biosensors. The simplicity of immobilizing the enzyme in chitosan gel

and small time required to form a biosensor enable speeding up and simplifying the assay based

on this immobilization technique; at the same time, immobilization in polyelectrolyte capsules

makes it possible to increase the lifetime of the biosensor and to maintain the activity of the

enzyme on the initial level.

Effects of real-sample components on biosensor signals

When assaying real samples, it is required to take into account the effects of their major

components. Thus, biosensor signals in glucose assays of wine beverages can be affected by

ethanol contents. We investigated biosensor signals for the introduction of glucose in buffers with

ethanol contents within the range of 0.1–2.0 M. At the introduction of glucose at a concentration

of 1.0 and 5.0 mM, the signal of the biosensor did not change; the variation coefficient did not

exceed 5%. Thus, the content of ethanol in a sample within the concentration range of 0.1–2.0 M

(0.5–9.2%) shall not affect the biosensor signal or introduce additional measuring error. Therefore,

the developed biosensor can be used to measure glucose contents in products that contain alcohol

up to 9.2%.

The content of acid in juices should not exceed 3 g/l; in nectars, 5 g/l. We investigated the

effect of citric acid within the concentration range of 0.06–0.9 mM on biosensor signals during

the introduction of glucose at a concentration of 5 mM. The occurrence of citric acid did not

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

11

change the biosensor signal; therefore, its presence in samples up to a concentration of 31 mM (6

g/l) shall not affect the glucose assay.

Use of the biosensors for assays of glucose in beverages

Monitoring of the level of sugar in foodstuffs is a necessary procedure for assessing the

quality of product. The sugar contained in juices consists mainly of glucose, fructose and sucrose.

Glucose increases the sugar level in the blood and stimulates the secretion of insulin faster than

other sugars.34 Therefore, its content in foodstuffs is an important parameter, especially for

diabetic patients. The developed biosensors were tested on real samples of juice and alcoholic

beverages (Table 3). The results of measurements were compared with the data of the standard

glucose assay of samples by spectrophotometry. The correlation coefficient of the data between

measurements made by the biosensor with GOx immobilized in chitosan gel and by the standard

technique is 0.95. The correlation coefficient of the data between measurements made by the

biosensor with GOx immobilized in polyelectrolyte capsules and by the standard method is 0.92.

The correlation coefficient of the data between measurements made by two types of biosensors is

0.88. The correlation data are shown in Fig. 5. Thus, the developed biosensor is capable of

determining the glucose content in complex matrices comprising both other sugars and

compounds of other classes, which is indicative of the specificity of the biosensor.

Conclusions

Thus, we showed that the activity of the enzyme in capsules depended on the molarity and

sodium chloride concentration in the buffer solution; the presence of potassium ions had a greater

effect on the decrease of the signal than sodium ions. The sensitivity coefficient of the biosensor

based on GOx immobilized in chitosan gel was by an order of magnitude higher than that of the

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

12

biosensor based on GOx immobilized in polyelectrolyte capsules (1.17 and 0.12 μA/mM,

respectively). However, the linear range of glucose determination for the biosensor based on PMC

is broader (0.10–5.50 mM and 0.05–0.60 mM, respectively). In work35, its authors used quartz

sensors also to study GOx-containing polyelectrolyte capsules. Application of natural

polyelectrolytes, alginate and chitosan, led to swelling of films and, correspondingly, to a change

of the reaction to glucose, as it can be incorporated into produced polyelectrolyte complexes.

They showed that, by varying the film crosslinking time, pH, extent of alginate acetylation, the

sensitivity of capsules to glucose can be controlled. We showed that the use of PMC based on

sodium polystyrene sulfonate and poly(allylamine hydrochloride) to immobilize the enzyme

enables increasing the lifetime of the biosensor for at least up to half a year (the apparent

Michaelis constant of 4.1 mM did not practically change over 5 months of storage). We have

shown earlier36 that a modification of polyelectrolyte capsules with the enzyme by carbon

nanotubes enables increasing the biosensor signal, sensitivity, glucose-detection lower limit. At

the same time, many studies point to a toxicity of nanomaterials.37,38 In the present work, we

increase the amount of capsules on the biosensor surface from 17.5✕104 (used in work36) up to

15✕106 and, thus, succeed in changing biosensor’s main parameters, sensitivity (which changes

from 0.05 μА/mM up to 0.12 μА/mM) and detection range (which changes from 1.0–25.0 mM to

0.1–10.0 мМ), as well as in decreasing the glucose detection limit (from 1.0 mM down to 0.1

mM) without using nanotubes. Although the sensitivity of the biosensor with nanotubes is still

higher (0.30 μA/mM) and the glucose detection lower limit is lower (0.05 mM), the biosensors

presented in this work can find practical application. The biosensors we developed were used to

assay glucose in real samples of juice; the obtained data were shown to have a high correlation

with the data by the standard spectrophotometric glucose assay. Strong retention of

polyelectrolyte capsules with the enzyme on the surface of a graphite electrode enables producing

a multiple use biosensor, which makes it economically viable and competitive with other glucose

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

13

assay techniques.

Acknowledgements

The authors acknowledge Victor Selivanov for providing language help.

The work was supported by a Russian Science Foundation grant No 18-49-08005.

References

1. J.M. Rodriguez-Nogales, and A. Delgadillo, J. Mol. Catalysis B: Enzym., 2005,

33(1–2), 15.

2. A. Ramanavicius, A. Kausaite, and A. Ramanavicien, Sens. Actuators B, 2005,

111–112, 532.

3. A.C. Pierre, J. Biocatal. Biotransform., 2004, 22(3), 145.

4. X. Zhu, Y. Ma, C. Zhao, Z. Lin, L. Zhang, R. Chen, and W. Yang, Langmuir,

2014, 30(50), 15229.

5. M.J. McShane, Methods Mol. Biol., 2011, 679, 147.

6. K. Etsuo, Bioseparation, 1998, 7, 241.

7. S. Yabuki, Anal. Sci., 2011, 27, 695.

8. M.Â. Cerqueira, A.C. Pinheiro, O.L. Ramos, H. Silva, A.I. Bourbon, and A.A.

Vicente, in “Micro and Nano Technologies, Emerging Nanotechnologies in Food

Science”, ed. R. Busquets, 2017, Elsevier, 11.

9. A. Munin and F. Edwards-Lévy, Pharmaceutics, 2011, 3(4), 793.

10. A. Larrañaga, M. Lomora, J.R. Sarasua, C.G. Palivan, and A. Pandit, Progr. Mat.

Sci., 2017, 90, 325.

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

14

11. B.I. Sukhorukov, S.A. Tikhonenko, E.A. Saburova, A.V. Dubrovsky, Yu.N.

Dybovskaya, and L.I. Shabarchina, Biophysics, 2007, 52(6), 575.

12. F. Caruso, D. Trau, H. Mohwald, and R. Renneberg, Langmuir, 2000, 16(4), 1485.

13. V.I. Ternovsky, Yu.V. Chernokhvostov, M.G. Fomkina, and M.M. Montrel,

Biofizika, 2007, 52(5), 825.

14. M.M. Montrel, A.I. Petrov, V.I. Ternovsky, and M.G. Fomkina, RF Patent, 2008,

2333231.

15. A.N. Reshetilov, Yu.V. Plekhanova, S.A. Tikhonenko, and A.V. Dubrovskii, J.

Anal. Chem., 2015, 70(11), 1368.

16. L.O. Yagodina, and Yu.V. Chernokhvostov, Vestn. Kazan Technological

University, 2013, 16(1), 117.

17. L.I. Kazakova, N.P. Sirota, T.V. Sirota, and L.I. Shabarchina, Rus. J. Phys. Chem.

A, 2017, 91(9), 1828.

18. E.A. Saburova, B.I. Sukhorukov, M.E. Bobreshova, and L.I. Elphimova,

Biochemistry (Moscow), 2000, 65(8), 976.

19. C. Ding, H. Sun, J. Ren, and X. Qu, Analyt. Chim. Acta, 2017, 952, 88.

20. P.A. Fiorito, C.M.A. Brett, and S.I.C. de Torresi, Talanta, 2006, 69, 403.

21. J. Li, J.D. Qiu, J.J. Xu, H.Y. Chen, and X.H. Xia, Adv. Functional Mat., 2007, 17,

1574.

22. A.A. Karyakin, O.V. Gitelmacher, and E.E. Karyakina, Anal. Lett., 1994, 27(15),

2861.

23. A. Baranwal, A. Kumar, A. Priyadharshini, G.S. Oggu, I. Bhatnagar, A.

Srivastava, and P. Chandra, Int. J. Biol. Macromol., 2018, 110, 110.

24. M. Hasanzadeh, N. Shadjou, and M. de la Guardia, Trends Anal. Chem., 2018,

102, 210.

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

15

25. A.I. Petrov, D.V. Volodkin, and G.B. Sukhorukov, Biotechnol. Prog., 2005,

21(3), 918.

26. X. Wang, H. Gu, F. Yin, and Y. Tu, Biosens. Bioelectronics, 2009, 24(5), 1527.

27. D. Barham, and P. Trinder, Analyst, 1972, 97, 142.

28. A.P. Nechayev, S.E. Traubenberg, and A.A. Kochetkova, “Food Chemistry”, 4th

edn, 2007, GIORD Publishers, St.-Petersburg, RF, 640.

29. S.A. Tikhonenko, E.A. Saburova, E.N. Durdenko, and B.I. Sukhorukov, Russian

J. Phys. Chem. A, 2009, 83(10), 1781.

30. J.-Y. Nam, H.-W. Kim, K.-H. Lim, H.-S. Shin, and B.E. Logan, Biosens.

Bioelectronics, 2010, 25, 1155.

31. O. Lefebvre, Z. Tan, S. Kharkwal, and H.Y. Ng, Bioresource Technol., 2012, 112,

336.

32. M.J. McShane, Methods Mol. Biol., 2011, 679, 147.

33. A. Fersht, “Enzyme Structure and Mechanism”, Reading, San Francisco, 1977,

371.

34. B.M. Lee and T.M.S. Wolever, Eur. J. Clin. Nutrition, 1998, 52, 924.

35. S. Belbekhouche, S. Charaabi, L. Picton, D. Le Cerf, and B. Carbonnier,

Carbohydrate Polym., 2018, 184, 144.

36. A. Reshetilov, Y. Plekhanova, S. Tarasov, S. Tikhonenko, A. Dubrovsky, A. Kim, V.

Kashin, A. Machulin, G.-J. Wang, V. Kolesov and I. Kuznetsova, Membranes, 2019, 9,

53, doi:10.3390/membranes9040053

37. R. Madannejad, N. Shoaie, F. Jahanpeyma, M.H. Darvishi, M. Azimzadeh and H. Javadi,

Chemico-Biological Interactions, 2019, doi: https://doi.org/10.1016/j.cbi.2019.04.036

38. L. Yang, X.-B. Luo and S.-L. Luo, “Assessment on Toxicity of Nanomaterials”, ed. X.

Luo and F. Deng, 2019, Chap. 9, Elsevier Inc., 273.

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

16

Table 1 Characteristics of the fabricated glucose biosensors

Parameter In polyelectrolyte capsules In chitosan gel Values of parameters of the calibration

dependence Vmax = 1.069;

h = 1.223; KM = 4.071

Vmax = 1.398; h = 1.176; KM = 1.517

Linear detection range, mM 0.10–5.50 0.05–0.60 Regression equation for the linear

segment y = 0.1184x + 0.0456 y = 1.1749x – 0.0032

Sensitivity coefficient, μA/mM 0.12 1.17 Minimum detection limit, mM 0.10 0.05 Detection range, mM 0.10–10.00 0.05–5.00

An equation describing the calibration dependences: . Correlation coefficient for

calibration dependences and for regression equation for the linear segment of R2 is 0.99.

max

h

h h

M

V SV

K S=

+

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

17

Table 2 Changes in major parameters of the calibration curve in storage of the enzyme biosensor

Parameter Immobilization in polyelectrolyte capsules Immobilization in chitosan gel

Day 1 After 5

months After 8

months Day 1 After 2 months

Vmax, μА 1.069 0.775 0.445 1.398 0.491 KM, мМ 4.071 4.142 12.641 1.517 1.166 Enzyme activity, % 100 75 20 100 43 Minimum detection

limit, mM 0.10 0.16 0.80 0.05 0.10

Detection range,

mM 0.10–10.00 0.16–15.00 0.80–18.00 0.05–5.00 0.10–4.00

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

18

Table 3 Results of measurements of real samples

Technique Immobilization in

polyelectrolyte capsules Immobilization in chitosan

gel Spectrophotometry, λ = 500 nm

Sample Biosensor

readings,

a SD, mM

Measure-

ment

error, %

Biosensor

readings,

a SD, mM

Measure-

ment

error, %

Standard method

readings,

a SD, mM

Measure-

ment error,

%

1 129.00±1.87 1.5 142.18±0.35 0.2 136.95±2.56 1.9 2 136.24±5.10 3.7 167.43±6.67 4.0 147.51±5.83 4.0 3 156.04±6.10 3.9 192.41±3.00 1.6 149.46±6.43 4.3 4 180.01±1.44 0.8 171.00±4.52 2.6 159.88±11.00 6.9 5 204.20±11.33 5.5 198.67±3.44 1.7 178.38±1.03 0.6 6 37.20±1.96 5.3 24.22±0.94 3.9 28.82±2.05 7.1 7 306.94±15.62 5.1 216.16±8.45 4.0 192.25±12.4 6.5

a, The mean value of measurements performed in 3–5 repeats; SD, standard deviation. Samples

are described in Experimental.

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

19

Figure Captions

Fig. 1 Scheme of fabricating polyelectrolyte microcapsules with the enzyme.

Fig. 2 Dependence of biosensor signals on composition of the buffer solution for a biosensor

based on glucose oxidase immobilized in polyelectrolyte capsules (1) and in chitosan gel (2): A,

on pH of the buffer solution; B, on molarity of the buffer solution; C, on the concentration of

NaCl in the buffer solution. Concentration of glucose, 5 mM for a biosensor with polyelectrolyte

capsules, and 1.24 mM for a biosensor with chitosan gel.

Fig. 3 Dependence of biosensor signals on the concentration of glucose for a biosensor based

on glucose oxidase immobilized in polyelectrolyte capsules (1) and in chitosan gel (2). Inset, the

initial linear-range segments of the concentration dependences.

Fig. 4 Long-term stability of the glucose biosensors: A, glucose oxidase immobilized in

polyelectrolyte capsules (1, The first 24 h of biosensor formation; 2, in 5 months; 3, in 8 months);

B, glucose oxidase immobilized in chitosan gel (1, The first 24 h of biosensor formation; 2, in 2

months; 3, in 12 months).

Fig. 5 Graphical correlation of data produced by various glucose assays of real samples: A,

spectrophotometric assay, glucose oxidase immobilized in chitosan gel; B, spectrophotometric

and biosensor assays, glucose oxidase immobilized in polyelectrolyte microcapsules; C, two types

of biosensors.

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

20

Fig. 1 Scheme of fabricating polyelectrolyte microcapsules with the enzyme.

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

21

Fig. 2 A

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

22

Fig. 2 B

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

23

Fig. 2 C

Fig. 2 Dependence of biosensor signals on composition of the buffer solution for a biosensor

based on glucose oxidase immobilized in polyelectrolyte capsules (1) and in chitosan gel (2): A,

on pH of the buffer solution; B, on molarity of the buffer solution; C, on the concentration of

NaCl in the buffer solution. Concentration of glucose, 5 mM for a biosensor with polyelectrolyte

capsules, and 1.24 mM for a biosensor with chitosan gel.

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

24

Fig. 3 Dependence of biosensor signals on the concentration of glucose for a biosensor based on

glucose oxidase immobilized in polyelectrolyte capsules (1) and in chitosan gel (2). Inset, the

initial linear-range segments of the concentration dependences.

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

25

Fig. 4 A

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

26

Fig. 4 B

Fig. 4 Long-term stability of the glucose biosensors: A, glucose oxidase immobilized in

polyelectrolyte capsules (1, The first 24 h of biosensor formation; 2, in 5 months; 3, in 8 months);

B, glucose oxidase immobilized in chitosan gel (1, The first 24 h of biosensor formation; 2, in 2

months; 3, in 12 months).

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

27

Fig. 5 A

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

28

Fig. 5 B

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

29

Fig. 5 C

Fig. 5 Graphical correlation of data produced by various glucose assays of real samples: A,

spectrophotometric assay, glucose oxidase immobilized in chitosan gel; B, spectrophotometric

and biosensor assays, glucose oxidase immobilized in polyelectrolyte capsules; C, two types of

biosensors.

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131

30

Graphical Index

Analytical SciencesAdvance Publication by J-STAGEReceived April 14, 2019; Accepted May 23, 2019; Published online on May 31, 2019DOI: 10.2116/analsci.19P131