51
ID/AAPM/10 Indra J. Das, PhD, FACR Department of Radiation Oncology Indiana University of School of Medicine & Midwest Proton Radiation Therapy Institute (MPRI), Indiana Commissioning and Calibrating a Linear Accelerator – State-of-the-Art in 2010

Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

  • Upload
    trandat

  • View
    222

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Indra J. Das, PhD, FACRDepartment of Radiation OncologyIndiana University of School of Medicine & Midwest Proton Radiation Therapy Institute (MPRI), Indiana

Commissioning and Calibrating a Linear Accelerator – State-of-the-Art in 2010

Page 2: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Preface

Med. Phys. 35(9), 4186-4214, 2008

Page 3: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Page 4: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

2008

Page 5: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Planning for Commissioning Data

Page 6: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Time= [(PDD + 5 profiles)/beam energy]x ( open + 4 wedges) x 60 points/scanx [(1 s/pts + (1s/movement and delay)]x (15 fields x 2 energies)

∼105 s∼ 30 h

Planning for Commissioning Time

Page 7: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Rational For Commissioning Beam DataFirst, it is not evident that manufacturing procedures for all linear accelerators have produced a level of reproducibility acceptable for clinical use. For example, variations in beam parameters have been noted between beams with the same nominal energies.

Second, on-site changes made during installation and acceptance of the user’s accelerator e.g., changes in beam energy and/or profiles from beam steering will not be modeled in the golden data.

Third, the beam characteristics of the soft wedges are made by moving jaws that depend on the speed parameters of the jaws and a deviation at site could affect the beam profile of the soft wedge.

Fourth, although acceptable agreement with the golden data set may be found in individual checks, it may be that some clinical setups will have multiple errors, which combine to produce unacceptable results.

TG-106

Page 8: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Hrbacek, et al Med. Phys. 34, 2917–2927, 2007.

Rational For Not Using Golden Data

Page 9: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Hrbacek, et al “Quantitative evaluation of a beam-matching procedure using one dimensional gamma analysis,” Med. Phys. 34, 2917–2927, 2007.

Rational For Not Using Golden Data

|xi − x¯ | <∆, ∀ xi

∆ =? (0.5, 1.0 0r 2%)

Page 10: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Time required for commissioning adual energy linear accelerator with photon and electron beam is

Answer: 4Reference: Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008

Question

1. 1 day2. 3 days3. 1 week4. 4-6 weeks5. 2 months

Page 11: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Standard chamber 10−1 cm3—The active volume for a standard Farmer-type ionization chamber is on average 0.6 cm3.

Minichamber 10−2 cm3—The active volume for a mini-ionization chamber is on average 0.05 cm3.

Microchamber 10−3 cm3—The active volume for a microionization chamber is on average 0.007 cm3

and ideally suited for small field dosimetry such as radiosurgery,gamma knife, CyberKnife, and IMRT

Definition of Detectors

Page 12: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Water

Air

1 2 3 4 5

Setting Water Tank & Detector

Page 13: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Know Your Connectors

Page 14: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Understand Detector, Connector & Cable

See Poster:Srivastava et al, SU-GG-T-270, 2010

Page 15: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Quality of Cables

Page 16: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

When setting ion chamber in water tank, the correct position of the chamber as viewed in water tank (as seen in figure) is:

Water

Air

1 2 3 4 5

Answer: 3Reference : Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008

Question

1. Position # 12. Position # 23. Position # 34. Position # 45. Position # 5

Page 17: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Setup and Possible Errors

Page 18: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Electrometer Null Setting Cable subtraction Proper bias >300 V for ion chamber 100 V for diamond 0 v for all diodes

Proper gain Proper mode

Page 19: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Choose Consistent & Correct Polarity

Page 20: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Chambers & Gain

Page 21: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Selection of detector for beam data

Page 22: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

XYZ

XY

Z

Scan Direction

Chamber Orientations

Radiation beam

Choice of Detector Orientation

Page 23: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Detector Orientation

Page 24: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

0102030405060708090

100110

0 0.5 1 1.5 2

Rel

ativ

e Dos

e

Distance Off Axis (cm)

6MV, 2x2 cm2 field, Illustration of chamber volume effects

Diode, dmax

PTW Pinpoint, dmax

RK chamber, dmax

Page 25: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Page 26: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

The possible setup error that causes the photon beam dose profile in figure is due to:

0

10

20

30

40

50

60

70

80

90

100

-4 -3 -2 -1 0 1 2 3 4Distance (cm)

Rel

ativ

e D

ose 1.5 cm

10 cm20 cm

Answer: 3Reference: Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008

Question

1. Gantry tilt

2. Collimator rotation

3. Tank arm tilt

4. Gantry and arm tilt5. Gantry tilt, collimator rotation and

tank arm tilt

Page 27: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0

Depth (mm)

10

20

30

40

50

60

70

80

90

100

0

Dos

e (%

)

Surface & Buildup Dose, 6 MV

0.6 cc0.3 cc0.125 ccMarkusA16IC-4PinPointDiamondPFDSFD

Reference: Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008

Answer: 2

Question

The accurate measurement of the buildup and surface dose could be achieved by a:

1. Diode detector2. Parallel plate chamber3. Cylindrical chamber4. Diamond detector5. PinPoint chamber

Page 28: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

0

50

100

150

200

250

-20 -15 -10 -5 0 5 10 15 20

Rel

ativ

e D

ose

Distance Off Axis(cm)

6 MV 60 Deg Wedge, 10 cm depth: water vs Profiler

Diode Array Profile

Water Profile

Page 29: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

0

20

40

60

80

100

120

140

160

180

200

-20 -15 -10 -5 0 5 10 15 20

Rel

ativ

e D

ose

Distance Off Axis(cm)

18 MV 60 Degree Wedge, 10 cm depth, 100 cm SAD: Water vs Profiler

Diode Array profile

Water Profile

Page 30: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Scanning Speed

Page 31: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Scanning Speed

Page 32: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

0

20

40

60

80

100

0 5 10 15 20 25 30 35

Perc

ent D

epth

Dos

e

Depth(cm)

Arithmetic Mean (AM) smoothing: 60 Degree Wedge PDD

UnsmoothedAM x 1AM x 2AM x 3AM x 5

Page 33: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Page 34: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Electron beam depth doses shown in figure represents the problem of

0

0.2

0.4

0.6

0.8

1.0

1.2

0 2 4 6 8 10 12 14 16 18 20

Depth, z (cm)

Dep

th D

ose

(%)

6 MeV9 MeV

12 MeV

15 MeV

18 MeV

21 MeV

Answer: 4Reference: Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008

Question

1. Noise in the cable2. Electrometer gain3. Bias on the electrometer4. Speed of scanning5. Tuning of accelerator

Page 35: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Future of Beam Data Commissioning

Standardization of linear acceleratorsMonte Carlo based commissioning Newer Radiation Detectors & Cables Newer Scanning Systems Smart algorithms

Page 36: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Aubin et al., Med Phys, 37(5), 2279-2288, 2010

Page 37: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Monte Carlo Codes

Aubin et al., Med Phys, 37(5), 2279-2288, 2010

Page 38: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Simulation of intensity at target

Aubin et al., Med Phys, 37(5), 2279-2288, 2010

Page 39: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Simulated Profiles

Aubin et al., Med Phys, 37(5), 2279-2288, 2010

Page 40: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Profiles for different fields

Page 41: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Depth Dose Simulation

Aubin et al., Med Phys, 37(5), 2279-2288, 2010

Page 42: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Detector Manufacturer Type volumeSFD Scanditronix Photon diode 1.7x10-5cm3

PFD Scanditronix Photon diode 1.9x10-4cm3

Exradin A-16 Standard Imaging Ion chamber 0.007cm3

Wellhofer-IC4 Scanditronix Ion chamber 0.40 cm3

Pinpoint PTW Ion chamber 0.015cm3

0.125cc PTW Ion chamber 0.125cm3

0.3cc PTW Ion chamber 0.3 cm3

0.6cc PTW Ion chamber 0.6 cm3

Diamond PTW Diamond 0.003cm3

Markus PTW Parallel plate 0.055cm3

Edge Detector Sun Nuclear Diode 10-5cm3

Other

Detectors

Page 43: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Sensitivity vs Volume of Detectors

0

10

20

30

40

50

60

70

80

90

100

110

120

130

140

1.0E-04 1.0E-03 1.0E-02 1.0E-01 1.0E+00

Volume (cm3)

Rela

tive s

ensit

ivity

PFD

SFD

A-16 PinPointMarkus

IC4

0.125cc

0.3cc

0.6 cc

Page 44: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Beam Profile

0

10

20

30

40

50

60

70

80

90

100

110

0 10 20 30 40 50 60 70 80Distance (mm)

Dos

e (%

)

0.6 cc0.3 cc0.125 ccMarkusA16IC-4PinPointDiamondPFDSFD

Answer: 5Reference: Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008

Question

Photon beam dose profiles taken with various detectors as shown in figure is possibly due to:

1. Speed of scanning2. Beam asymmetry3. Pb piece in the beam4. Hysteresis of scanning system5. Orientation of scanning detector

Page 45: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Comparison of Large Tank and Small SRS Cylinder Tank for SRS, TMR & Profiles

ARM Inc., Port Saint Lucie, FL 34983

Moving Tank System for TMR

Page 46: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

No SSD to SAD calculations required, No cubic spline fit of a limited number of fixed data points needed

0

0.2

0.4

0.6

0.8

1

1.2

0

0.2

0.4

0.6

0.8

1

1.2

0 50 100 150 200 250 300 350

Calculated TPR

Measured TPR

Calculated TPR ~2% less at depth Cubic spline fit of 12 data points

Nikesch et al, CyberKnife Center , Palm Beach, FL

Direct TMR Data Acquisition

Page 47: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Scanner Orientation Advantage

Page 48: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Sun Nuclear 3D Scanner

2

1

1

32

1. Ring drive maintains consistent scanning direction

2. Diameter drive has maximum scanning range of 640mm

3. Vertical drive has maximum travel of 400mm

Page 49: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Conclusions Golden Data should be taken as a

reference only Understand time and amount of data

to be taken View each parameters properly,

double check by another individual Use proper detector for each type of

data collection Set optimum speed for scanning, do

not rush

Page 50: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

-Conclusions

Understand the limits and measuring condition

Question every unusual data set Do not smooth data too much Write report for future reference Future technology & resources

could help commissioning simpler

Page 51: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical

ID/AAPM/10

Thanks