Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
cobas ® 4800 CT/NG and HPV Workflow Overview
Upon completing this section, you will be able to:
• Perform the required maintenance activities prior to the start of a run
• Understand the assay workflow using the cobas® 4800 System
Objective
Full Workflow
Recovery Workflow
Workflow Overview: Full Run I
System Introduction SW 2.2 System Workflow Design
Daily Maintenance takes app. 10-20 minutes to complete Daily Maintenance
Results are shown and can be printed out Result Reporting
CT/NG ~ 90 minutes
HPV ~ 120 minutes
Real-Time Amplification
&
Detection
CT/NG ~ 120 minutes for 96 samples
HPV ~ 160 minutes for 96 samples
Sample Preparation
& PCR Set Up
Up to 30 minutes to load samples, reagents and consumables for up to 96 samples/controls Instrument Set Up
1. Start up cobas® 4800 System
Start up cobas® 4800 instruments
1. Switch on the cobas® z480 analyzer
2. Switch on heater/shaker unit
3. Switch on the cobas® x480 instrument
Start up and log on to cobas® 4800 software:
1. Switch on the monitor and control unit
• The Windows 7 operating system starts automatically
• Double click the cobas® 4800 software icon on the desktop and log in
• The cobas® 4800 software displays the System tab
2. Click to log on and enter your assigned user ID and password
3. Click OK
2. cobas® x480 Instrument Maintenance
• Select Overview tab and check maintenance status of cobas® x480 instrument
– If weekly maintenance is due, click Run Weekly Maintenance
– If daily maintenance is due, click Run Daily Maintenance
• Follow the software wizard guidance
2. cobas® x480 Instrument Daily Maintenance
The following tasks are part of the daily maintenance:
1. Shutdown and restart the cobas® 4800 system
2. Check if instrument deck is clean
3. Empty the tip waste and the liquid waste container, clean the tip eject plate
4. The cobas® x480 instrument automatically checks the tightness of the pipetting channels and verifies the functioning of the liquid level detection
3. Remove Samples and Reagents from Storage
• Remove HPV or CT/NG samples & HPV or CT/NG reagents from storage
• The run size is either 24 or 96 batch for Sample Preparation kits
Note: MMx and Mn2+ can be loaded directly from the fridge onto cobas® x480
instrument
4. Start New Run
Start a new run
1. Click
2. Select Workflow Type:
– Full
– Recovery (Only for CT/NG and HPV)
3. Select test
4. Type run name (optional)
5. Click OK The default as shown
5. Load Specimens (1)
Load specimens
1. Uncap specimens
2. Place specimens in corresponding carrier
3. Insert sample carriers on autoload tray
4. Click Load Specimens
A) Specimen barcodes facing to the right
B) Sample carrier
C) Tracks 17 - 34 are reserved for sample carriers
B
A
C
5. Load Specimens (2)
Unloading Samples/Sample Carriers Option
Default is checked—unloading of sample carriers:
6. Define Work Order (without LIS)
Define work order
1. Specimen IDs are scanned by
the autoloader
2. Select Specimen
3. Select Specimen type:
CT/NG – PC, Swab, Urine
HPV – PC, SP
4. Select Requested Result
• CT/NG:
a. CT
b. NG
c. CT and NG
• HPV:
a. HPV High Risk Panel
b. HPV High Risk Panel
+ Genotyping
5. Select Next
7. Load Consumables
1. Place listed consumables on appropriate
carriers
2. Insert carriers on autoload tray
3. Click Load Consumables:
Use the following tracks:
• Plate carrier: tracks 1 through 6
• Left tip carrier: tracks 11 through 16
• Right tip carrier: tracks 35 through 40
8. Load Reagents (1)
To minimize handling errors the reagent reservoirs are filled and placed using
the scan-scan-pour principle:
1. Scan the barcode of the required reagent
2. Scan the barcode of an unused reagent reservoir
3. Pour the reagent in the scanned reagent reservoir
4. Place the filled reagent reservoir onto the required position of the reagent
reservoir carrier as indicated in the software
• The reagent reservoirs are available in the two sizes: 200 mL and 50 mL
• The reagent reservoir barcodes must align with the barcode of the carrier (face to
the right of the carrier)
8. Load Reagents (2)
Continued
• All reagents except MGPs are homogeneous and do not require
vortexing/shaking
• Each reagent has a specific location assigned to it on the carriers
• All reagents except MMX and Mg/Mn must be at ambient
temperature prior to loading on the instrument
• The MMXs and Co-factors may be taken directly from 2-8 ºC storage
8. Load Reagents (3)
Continued
• The reagent inventory marks a reagent as used as soon as it is assigned to
a reservoir. From this time point on, the reagent is dedicated to this run and
cannot be used later on another run even if the reagent is not used due to aborting
the run
• The cobas® 4800 System has an internal timer to monitor reagent on-board
stability
• Once the wash buffer is scanned, 1 hour (60 minutes) is allowed to
complete the loading process and to start the sample preparation run
• A countdown timer is displayed on the Workplace tab
8. Load Reagents (4)
Wash Buffer
200 mL reagent reservoir carrier
1. Load a 200 mL reagent reservoir
on the carrier:
Scan – Scan – Pour - Place
2. Insert carrier on the autoload tray:
200 mL reagent reservoir
carrier: tracks 48 through 49
3. Click Load Reagents
8. Load Reagents (5)
SDS, Lysis, MGP & Elution Buffer
50 mL reagent reservoir carrier
1. Load reagents on 50 mL reagent
reservoir carrier as indicated in
the wizard
2. Insert the carrier onto the
autoload tray:
50 mL reagent reservoir
carrier is placed on track 50
3. Select Load Reagents
8. Load Reagents (6)
Controls, MMX & Metal(s)
Reagent carrier
1. Uncap reagent vials and load them
on reagent carrier as indicated in
the wizard
2. Insert carrier on autoload tray:
CT/NG:
Reagent carrier: track 51
IC, Control Dil, RMECs,
MMx and Co-factor
HPV:
Reagent carrier: track 51
PK, RMECs, MMx and
Co-factor
3. Click Load Reagents
9. Start Sample Preparation Run
Select “Start run”
10. Unload & Seal Microwell Plate
Sample Preparation Run Complete (1)
• Amplification and detection should be started as soon as possible but no
later than 90 minutes (1 h 30 min) after the end of the cobas® x480
instrument run
• A countdown timer is displayed on the Workplace tab
To see flags
• Click Sample Prep
Results to review the
results of the sample
preparation
• The Sample Preparation
Results window pops up
• Samples that had clots or
pipetting errors will have
“failed” and will show
flags
10. Unload & Seal Microwell Plate
Sample Preparation Run Complete (2)
Sample Workflow Overview: Full Run Part II
10. Unload & Seal Microwell Plate
1. Click Unload
2. Seal the microwell plate:
• Remove the protection layer from
the sealing film
• Cover the microwell plate with the
adhesive side of the sealing film
• Firmly press the sealing film to the
plate surface using the sealing
film applicator
• Remove both ends of the sealing
film alongside the perforation
3. Click Next
11. Load Microwell Plate
1. Press Load button on cobas® z480 analyzer
2. Place the sealed microwell plate into the extended loader
3. Press load button again. The amplification and detection run starts automatically
12. Review and Accept Results
Review and accept results:
1. Click Show Results
2. Review and accept results in
Results tab
3. Select results and click to
print the results report
13. With LIS, send the results to LIS
Consider that depending on the
configuration, all results are
transferred to LIS or only accepted
results are transferred to LIS.
• Control results are always
uploaded to LIS
• Select a result or group of
results and choose the Send
results to LIS button
14. Unload cobas® z480 analyzer
Unload cobas® z480 analyzer:
• After the run has finished, unload the microwell plate from the cobas®
z480 analyzer as soon as practical
• The sealing film adhesiveness may degrade with time and without
pressure. Removing the microwell plate promptly minimizes any possible
contamination of the cobas® z480 analyzer
Full Workflow
Recovery Workflow
Recovery Run
Overview
• The recovery workflow allows the user to recover failed runs where the sample
has been successfully prepared
• A run can only be recovered once
Note: Recovery Workflow is available only for CT/NG and HPV (all
specimen types)
Not available for C. diff, HSV 1/2 and MRSA/SA or Virology assays
Recovery Run
Overview
• User can perform the recovery workflow anytime a full run fails, under the
following conditions:
– DWP preparation stage is completed during sample preparation
– Recovery workflow is started within 24 hours of the start of the full run
• Possible recovery scenarios
– If run was aborted during sample preparation (after DWP prep is completed)
– If run was aborted during plate transfer
– If run was aborted during amplification and detection
A full run has been aborted by a user (M2 flag) or the analyzer (Z1 flag)
Recovery Run
Overview
• The recovery workflow is intended for repeat amplification from the remaining
eluate in the deepwell plate
• For the recovery workflow, the microwell plate is manually prepared with working
master mix reagent, Mn reagent and residual eluate from the deepwell plate
Sample Workflow Overview: Recovery Run
1. Start a recovery run
1. Choose (New run)
2. Select the Recovery option
3. Select the appropriate test check
box
4. Optionally, type a run name
5. Choose the OK button
2. Select the run to recover and add new IDs
1. Choose the run to recover
2. In the Microwell plate ID field,
scan the microwell plate barcode
3. In the Master Mix ID field, scan
the master mix reagent barcode
4. In the Mn Reagent ID field, scan
the Mn reagent barcode
3. Print the microwell plate layout
1. To print the work order file for
microwell plate setup, choose
(Print) from the global navigation bar
2. In the software, choose the Next
button
4. Retrieve the deepwell plate
Do one of the following:
– If the deepwell plate has been unloaded by the
instrument, remove it from the plate carrier, or
– If the deepwell plate has been stored, retrieve it from
storage, or
– If the deepwell plate has not been unloaded by the
instrument, unload the deepwell plate manually
5. Set up the new microwell plate
Set up the microwell plate for the recovery run in the following way:
1. Perform manual PCR setup according to the test-specific package insert
2. Seal microwell plate
3. Centrifuge microwell plate
4. In the software, choose the Next button
6. Load microwell plate into the analyzer
1. Press the load button on the
analyzer
2. Place the sealed microwell plate
into the microwell plate loader
3. Press the load button again to
close the drawer
The amplification and detection run
starts automatically
4. Check the timer in the wizard
7. Review and accept results
1. Choose the Show result button
2. Review and accept results in
Results work area
3. Select results and choose
(Print) to print the results report, if
required
8. With LIS, send the results to LIS
Consider that depending on the
configuration, all results are
transferred to LIS or only accepted
results are transferred to LIS
• Control results are always
uploaded to LIS
• Select a result or group of results
and choose the Send results to
LIS button
9. Unload the microwell plate from the cobas®
z480 analyzer
Unload cobas® z480 analyzer:
• After the run has finished, unload the microwell plate from the cobas® z
480 analyzer as soon as practical
• The sealing film adhesiveness may degrade with time and without
pressure. Removing the microwell plate promptly minimizes any possible
contamination of the cobas® z480 analyzer
Doing now what patients need next