Upload
tranmien
View
233
Download
1
Embed Size (px)
Citation preview
4
CHAPTER II
INTRODUCTION TO HPLC METHOD DEVELOPMENT
AND VALIDATION
Pharmaceutical analysis simply means analysis of pharmaceuticals.
Webster’ dictionary defines pharmaceutical as a medical drug. It is generally
known that a pharmaceutical is a therapeutic interest. A more appropriate term for
a pharmaceutical is active pharmaceutical ingredient (API) or active ingredient to
distinguish it from a formulated product or drug product is prepared by
formulating a drug substance with inert ingredient (excipient) to prepare a drug
product that is suitable for administration to patients. It is well known in the
pharmaceutical industry that pharmaceutical analysts in research and development
(R&D) play a very comprehensive role in new drug development and follow up
activities to ensure that a new drug product meets the established standards is
stable and continue to approved by regulatory authorities ,assuring that all batches
of drug product are made to the specific standards utilization of approved
ingredients and production method becomes the responsibility of pharmaceutical
analysts in the quality control (QC) or quality assurance department .
The methods are generally developed in an analytical R&D
department and transferred to QC or other departments as needed. At times they
are transferred to other divisions.
5
By now it should be quite apparent that pharmaceutical analysts play a
major role in assuring the identity, safety, efficacy, and quality of drug product
safety and efficacy studies required that drug substance and drug product meet two
critical requirements.
1. Established identity and purity.
2. Established bio availability/dissolution1.
SCOPE AND SIGNIFICANCE OF PHARMACEUTICAL
ANALYSIS:
Pharmaceutical companies rely upon both qualitative and quantitative
chemical analysis to ensure that the raw material used meet all the desired
specifications, and also to check the quality of the final product. The examination
of raw material is carried out to ensure that there is no unusual substance present
which might deteriorate the manufacturing process or appear as a harmful
impurity in the final product. The quantity of required ingredient in raw material is
determined by a procedure known as Assay.
The final manufactured product is subjected to quality control2
to ensure that desired components are present within a range and impurities do not
exceed certain specified limits.
Some specific use of analysis is under mentioned:
(i) Quantitative analysis of air, water, soil samples to determine the level of
pollution.
6
(ii) Chemical analysis to assist diagnosis of illness and monitoring the
condition of patients.
(iii) In farming, nature of soil and level of fertilizer application is analyzed.
(iv) In geology, composition of the rock and soil is carried out.
Types of Analysis:
In general analysis is divided into two major parts:
(a) QUALITATIVE ANALYSIS(what substances are present in the given
sample)
(b) QUANTITATIVE ANALYSIS(to determine the quantity of each
component in the given sample)
Qualitative:
• Qualitative inorganic analysis seeks to establish the presence of a
given element or inorganic compound in a sample.
• Qualitative organic analysis seeks to establish the presence of a
given functional group or organic compound in a sample.
Quantitative:
Quantitative analysis seeks to establish the amount of a given element or
compound in a sample.
The factors which must be taken into account when selecting an appropriate
method of analysis are:
(a) The nature of the information sought
7
(b) The size of sample available and the proportion of the constituent to be
determined
(c) The purpose for which the analytical data is required.
INTRODUCTION TO CHROMATOGRAPHY
The term chromatography 3(Greek kromatos –colour & graphos-written
means colour writing .Mikhail Twestt (1906) - invented the chromatography. The
IUPAC has defined chromatography as “a method used primarily for the
separation of component of a sample, in which the component are distributed
between two phases, one of which is stationary while the other moves .the
stationary may be a solid or liquid supported on a solid or a gel and may be packed
in a column, spread as a layer or distributed as a film. The mobile phase may be
gaseous or liquid”.
Types of Chromatographic Methods.3
Based on modes of chromatography:
Normal phase chromatography
Reverse phase chromatography
Based on principle of separation:
Partition chromatography
Adsorption chromatography
Ion exchange chromatography
Size exclusion chromatography
8
Affinity chromatography
Chiral phase chromatography
Base on elution technique:
Isocratic separation
Gradient separation
Based on the scale of operation:
Analytical HPLC
Preparative HPLC
Partition Chromatography
Partition chromatography was the first kind of chromatography that
chemists developed. The partition coefficient principle has been applied in paper
chromatography, thin layer chromatography, gas phase and liquid-liquid
applications. The 1952 Nobel Prize in chemistry was earned by Archer John Porter
Martin and Richard Laurence Millington Synge for their development of the
technique, which was used for their separation of amino acids. Partition
chromatography uses a retained solvent, on the surface or within the grains or
fibres of an "inert" solid supporting matrix as with paper chromatography; or takes
advantage of some coulombic and/or hydrogen donor interaction with the solid
support. Molecules equilibrate (partition) between a liquid stationary phase and
the eluent. Known as Hydrophilic Interaction Chromatography (HILIC) in HPLC,
9
this method separates analytes based on polar differences. HILIC most often uses
a bonded polar stationary phase and a non-polar, water miscible, mobile phase.
Partition HPLC has been used historically on unbonded silica or alumina supports.
Each works effectively for separating analytes by relative polar differences,
however, HILIC has the advantage of separating acidic, basic and neutral solutes
in a single chromatogram.
The polar analytes diffuse into a stationary water layer associated with the
polar stationary phase and are thus retained. Retention strengths increase with
increased analyte polarity, and the interaction between the polar analyte and the
polar stationary phase (relative to the mobile phase) increases the elution time.
The interaction strength depends on the functional groups in the analyte molecule
which promote partitioning but can also include coulombic (electrostatic)
interaction and hydrogen donar capability. Use of more polar solvents in the
mobile phase will decrease the retention time of the analytes, whereas more
hydrophobic solvents tend to increase retention times.
Normal Phase Chromatography
Also known as normal-phase HPLC (NP-HPLC), or adsorption
chromatography, this method separates analytes based on adsorption to a
stationary surface chemistry and by polarity. It was one of the first kinds of HPLC
that chemists developed. NP-HPLC uses a polar stationary phase and a non-polar,
non-aqueous mobile phase, and works effectively for separating analytes readily
10
soluble in non-polar solvents. The analyte associates with and is retained by the
polar stationary phase. Adsorption strengths increase with increased analyte
polarity, and the interaction between the polar analyte and the polar stationary
phase (relative to the mobile phase) increases the elution time. The interaction
strength depends not only on the functional groups in the analyte molecule, but
also on steric factors. The effect of sterics on interaction strength allows this
method to resolve (separate) structural isomers.
The use of more polar solvents in the mobile phase
will decrease the retention time of the analytes, whereas more hydrophobic
solvents tend to increase retention times. Very polar solvents in a mixture tend to
deactivate the stationary phase by creating a stationary bound water layer on the
stationary phase surface. This behavior is somewhat peculiar to normal phase
because it is most purely an adsorptive mechanism (the interactions are with a
hard surface rather than a soft layer on a surface).
Displacement Chromatography
The chromatography matrix (the displacer) will compete effectively for
binding sites, and thus displace all molecules with lesser affinities.There are
distinct differences between displacement and elution chromatography. In elution
mode, substances typically emerge from a column in narrow, Gaussian peaks.
Wide separation of peaks, preferably to baseline, is desired in order to achieve
maximum purification. The speed at which any component of a mixture travels
11
down the column in elution mode depends on many factors. But for two
substances to travel at different speeds, and thereby be resolved, there must be
substantial differences in some interaction between the biomolecules and the
chromatography matrix. basic principle of displacement chromatography is: A
molecule with a high affinity for Operating parameters are adjusted to maximize
the effect of this difference. mode chromatography, especially at the preparative
scale, are operational complexity, due to gradient solvent pumping, and low
through put, due to low column loadings. Displacement chromatography has
advantages over elution chromatography in that components are resolved into
consecutive zones of pure substances rather than “peaks”. Because the process
takes advantage of the nonlinearity of the isotherms, a larger column feed can be
separated on a given column with purified components recovered at significantly
higher concentrations.
Size Exclusion Chromatography
Size exclusion chromatography (SEC), also known as gel permeation
chromatography or gel filtration chromatography, separates particles on the
basis of size. It is generally a low resolution chromatography and thus it is often
reserved for the final, "polishing" step of a purification. It is also useful for
determining tertiary structure and quaternary structure of purified proteins. SEC is
used primarily for the analysis of large molecules such as proteins or polymers.
SEC works by trapping these smaller molecules in the pores of a particle. The
12
larger molecules simply pass by the pores as they are too large to enter the pores.
Larger molecules therefore flow through the column quicker than smaller
molecules, that is, the smaller the molecule, the longer the retention time.
This technique is widely used for the molecular weight
determination of polysaccharides. SEC is the official technique (suggested by
European pharmacopeia) for the molecular weight comparison of different
commercially available low-molecular weight heparins.
Ion- Exchange Chromatography
In ion-exchange chromatography (IC), retention is based on the attraction
between solute ions and charged sites bound to the stationary phase. Ions of the
same charge are excluded. Types of ion exchangers include:
Polystyrene resins – These allow cross linkage which increases the
stability of the chain. Higher cross linkage reduces swerving, which increases the
equilibration time and ultimately improves selectivity.
Cellulose and dextran ion exchangers (gels) – These possess larger
pore sizes and low charge densities making them suitable for protein separation.
Controlled-pore glass or porous silica
In general, ion exchangers favour the binding of ions of higher charge and
smaller radius.An increase in counter ion (with respect to the functional groups in
resins) concentration reduces the retention time. A decrease in pH reduces the
13
retention time in cation exchange while an increase in pH reduces the retention
time in anion exchange. By lowering the pH of the solvent in a cation exchange
column, for instance, more hydrogen ions are available to compete for positions on
the anionic stationary phase, thereby eluting weakly bound cations.
This form of chromatography is widely used in the
following applications: water purification, preconcentration of trace components,
ligand-exchange chromatography, ion-exchange chromatography of proteins,
high-pH anion-exchange chromatography of carbohydrates and oligosaccharides,
and others.
Bioaffinity Chromatography
This chromatographic process relies on the property of biologically active
substances to form stable, specific, and reversible complexes. The formation of
these complexes involves the participation of common molecular forces such as
the Van der Waals interaction, electrostatic interaction, dipole-dipole interaction,
hydrophobic interaction, and the hydrogen bond. An efficient, biospecific bond is
formed by a simultaneous and concerted action of several of these forces in the
complementary binding site.
Aqueous Normal-Phase Chromatography
Aqueous normal-phase chromatography (ANP) is a chromatographic
technique which encompasses the mobile phase region between reversed-phase
chromatography (RP) and organic normal phase chromatography (ONP). This
14
technique is used to achieve unique selectivity for hydrophilic compounds,
showing normal phase elution using reverse-phase solvents. Normal – phase
chromatography uses a polar (hydrophilic) stationary phase and a non-polar
(usually with no water) mobile phase.
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)
High performance liquid chromatography(HPLC)3-7 is a process, which
separates mixture containing two or more components under high pressure. in this
the stationary phase is packed in a column one end of which is attached to a source
of pressurized liquid mobile phase.High performance liquid chromatography is the
fastest growing analytical technique for the analysis of drugs. Its simplicity, high
specificity and wide range of sensitivity makes it ideal for the analysis of many
drugs in both dosage forms and biological fluids.
HPLC is also known as high pressure liquid
chromatography .It is essential form of column chromatography in which the
stationary phase is consist of small particles (3-50μm) pickings contained in a
column with a small pore(2-5mm) one end of which is attached to a source of
pressurized liquid eluent (mobile phase). The three form of high performance
liquid chromatography most often used are ion-exchange, partition and adsorption.
Advantages of HPLC
i. It provides a specific, sensitive and precise method for analysis of
different
15
complicated samples.
ii. There is speed of analysis.
iii. The analysis by HPLC is specific, accurate and precise.
iv. It offers advantage over gas chromatography in analysis of many polar
substances, metabolic products and thermo labile as well non-volatile substances.
It is presently used in pharmaceutical research and developments in the
following ways:
To purify synthetic or natural products,
To characterize metabolites,
To assay active ingredients, impurities, degradation products and
in dissolution assays
In pharmacodynamics and pharmacokinetic studies.
Fig.1: Instrumentation of HPLC
16
HPLC includes
1) Mobile phase reservoir and solvent treatment systems
2) Pumps:
i. Displacement pumps
ii. Reciprocating pumps
iii. Pneumatic pumps
3) Precolumn
4) Sample injectors
a. Syringe injection
b. Stop flow injection
c. Solvent flowing
5) Liquid chromatographic columns
a) Analytical columns
b) Preparative columns
6) Column packing materials
a) Pellicular
b) Porous particle
7) Detectors
a) Photometric detectors
b) Fluorescence detectors
c) Refractive index detectors
d) Electrochemical detectors
17
8) Recorders
Mobile phase reservoir and solvent treatment systems
A modern HPLC apparatus is equipped with one or more glass or stainless steel
reservoirs each of which contain 500 ml or more of solvent. The reservoirs are
often equipped with a means of removing dissolved gases usually O2andN2 that
interfere by forming bubbles in the columns and detector systems. These bubbles
cause band spreading; in addition they interfere with the performance of the
detector.
Isocratic And Gradient Elution
A separation in which the mobile phase composition remains constant throughout
the procedure is termed isocratic (meaning constant composition). The word was
coined by Csaba Horvath who was one of the pioneers of HPLC.
The mobile phase composition does not have to remain constant. A separation in
which the mobile phase composition is changed during the separation process is
described as a gradient elution.One example is a gradient starting at
10% methanol and ending at 90% methanol after 20 minutes. The two components
of the mobile phase are typically termed "A" and "B"; A is the "weak" solvent
which allows the solute to elute only slowly, while B is the "strong" solvent which
rapidly elutes the solutes from the column. In reverse-phase chromatography,
solvent A is often water or an aqueous buffer, while B is an organic solvent
miscible with water, such as acetonitrile, methanol, THF, or isopropanol.
18
Pumps:
The pumps are used to pass mobile phase through the column at high pressure and
at controlled flow rate .in addition to this its performance directly effects the
retention time, reproducibility and detector sensitivity.
The pumps used in HPLC should have the following features:
I. The generation of pressures up to 6000psi
II. Flow rates ranging from 0.1 to 10 ml/min
III. Flow control and flow reproducibility of ±0.5%
IV. It should be composition resistant and give a pulse free out put.
Fig. 2: Structure of pump
a. Displacement pumps
It consists of a large, syringe like chamber equipped with a plunger that is
activated by a screw driven mechanism powered by a stepping motor.
19
b. Reciprocating pumps:
It consists of a small chamber in which the solvent is pushed back force with the
help of a motor driven piston or pressure may be transmitted by a diaphragm
which is hydraulically pumped by a reciprocating piston.
c. Pneumatic pumps:
In this the mobile phase is contained in a collapsible container housed in a vessel
that can be pressurized by a compressed gas.
Pre column
Some HPLC instruments are equipped with a precolumn, which contains a
packing chemically identical to that in a analytical column. Particle size is a large
hence the pressure drop across the pre column is negligible with respect to the
analytical column. The precolumn is mainly used to remove the impurities from
the solvent and thus prevent contamination of the analytical column.
Sample injectors
Often the limiting factor in the precision of liquid chromatographic measurements
lies in the reproducibility with which samples can be introduced in to column
packing. It must be noted that overloading of the sample causes band broadening.
The sample injectors are of the following types
a) Syringe injection
b) Stop flow injection
c) Solvent flowing
20
Fig.3: Loading Mode
Inject Mode
Columns
The column is one of the most important components of the HPLC chromatograph
because the separation of the sample components is achieved when those
components pass through the column. HPLC columns are made of high quality
stainless steel, polished internally to a mirror finish. Standard analytical columns
are 4-5mm internal diameter and 10-30cm in length.Normally, columns are filled
with silica gel because its particle shape, surface properties, and pore structure
help to get a good separation. Silica is wetted by nearly every potential mobile
21
phase, is inert to most compounds and has a high surface activity which can be
modified easily with water and other agents. Silica can be used to separate a wide
variety of chemical compounds, and its chromatographic behaviours generally
predictable and reproducible.
In HPLC, generally two types of columns are used; normal phase columns
and reversed phase columns. Using normal phase chromatography, particularly
polar and moderately non-polar substances can make excellent separation using
normal phase columns and polar eluents. While reversed phase chromatography,
particularly polar substances can make excellent separation using reversed phase
columns and non-polar eluents.
There are various columns that are secondary to the separating column or
stationary phase. They are guard, Derivatizing, Capillary, and preparatory
columns.
22
Fig.4: Structure of Column
Guard columns are placed anterior to the separating column. This serves as a
protective factor that prolongs the life and usefulness of the separation column.
They are dependable columns designed to filter or remove:
1. Particles that clog the separation column
2. Compounds and ions could ultimately cause “baseline drift”, decreased
resolution, decreased sensitivity, and create false peaks .Compounds that may
cause precipitation upon contact with the stationary or mobile phase and
3. Compound that might co-elute and cause extraneous peaks and interfere
with detection and / or quantification.
These columns must be changed on a regular basis in order to optimize their
protection function. Size of the packing varies with the type of protection needed.
23
Derivatizing columns – Pre or post – primary column derivatization can be an
important aspect to the sample analysis. Reducing or altering the parent compound
to a chemically related daughter molecule or fragment elicits potentially tangible
data, which may complement other results or prior analysis. In few cases, the
derivatization step to cause data to become questionable, which is one reason why
HPLC was advantageous over gas chromatography. Because GC requires volatile,
thermally stabile, or nonpolar analytes , derivatization was usually required for
those samples, which did not contain these properties, Acetylation, Silylation, or
concentrated acid hydrolysis are a few derivatization techniques.
Capillary columns – Advances in HPLC led to smaller analytical columns. Also
known as microcolumns, capillary columns have a diameter much less than a
millimeter and there are three types: open – tubular, partially packed, and tightly
packed. They allow the user to work with nanonliter sample volumes, decreased
flow rate, and decreased solvent volume usage which may lead to cost
effectiveness.
Microbore and small – bore Columns are also analytical and small volumes
assays. A typical diameter for a small – bore column is 1-2mm. Like capillary
columns, instruments must usually be modified to accommodate these smaller
capacity columns (i.e decreased flow rate).
Fast columns – One of the primary reasons for using these columns is to obtain
improve sample through put (amount of compound per unit time). Fast columns
24
are designed to decreases time of the chromatographic analysis without forsaking
significant deviations in results. These columns have the same internal diameter
but much shorter length than most other columns, and they are packed with
smaller particles that are typically 3 mcg/ml in diameter. Advantages include
increased sensitivity, deceased analysis time, decreased mobile phase usage, and
increased reproducibility.
Preparatory Columns - These columns are utilized when the objective its prepare
bulk (milligrams) of sample for laboratory preparatory applications. A facilitate
large volume injection into the HPLC system. Accessories important to mention
are the back-pressure regulator and the fraction collector. The back-pressure
regulator is placed immediately posterior to HPLC detector. The fraction collector
is an automated device that collects uniform increment of the HPLC output.
Microbe columns of 1-2mm internal diameter and 10-25 cm in length have
certain advantages of lower detection limits and lower consumption of solvent, the
latter being important if expensive HPLC grade solvents are used.
25
Table 1.COLUMN DIMENSIONS
Detectors:
The most widely used detectors for liquid chromatography are based upon
absorption of ultraviolet or visible radiation. Photometers and columns are
available from commercial sources. The former often makes use of the 254 nm to
280 nm lines from a mercury source because many organic functional groups
absorb in this region. Deuterium or tungsten filament sources with interference
filters also provide a simple means of detecting absorbing species. Some modern
filters, which can be rapidly switched in to place. Spectrophotometer detectors are
considerably more versatile than photometry and are also widely used in high
performance instruments. Often these are diode-array instruments that can display
an entire spectrum as an analyte exits the column. Another detector, which has
Type
Internal
Diameter
(cm)
Length
(cm)
Particle
Size
(µm)
Analytical 0.3 - 0.46 3-28 3-10
Semimicro 0.1 – 0.21 10 – 25 3 – 18
Semipreparative 0.8 – 1.0 10 – 25 5 – 10
Preparative 2.0 – 5.0 10 – 25 10 – 20
26
found considerable application, is based up on the changes in the refractive index
of the solvent that is caused by analyte molecules. In contrast to most of the other
detector is its some what limited sensitivity. Several electrochemical detectors
have also been introduced that are based on potentiometric, conductometric and
voltametric measurements.
Recorders
The signals from a detector are recorded as deviations from a base line. Two pen
recorder are used with instruments having two detectors. The peak position along
the curve relative to the starting point denotes the particular component .with
proper calibration, the height or area of the peak is a measure of amount of
component in a sample.
Chromatographic Parameters3,4:
Retention time (tr):
This is the time of emergence of the peak maximum of the component after
injection. This is the sum of the times the component spends in the mobile phase
(tM) and in the stationary phase.
Adjusted retention time:
It is the time the component spends in the stationary phase and is given by t1
r=tR-tM
The value of tM is obtained by measuring the time to elute an un retained
substance, e.g. air or Methane.
27
It is the ratio of the time the component spends in the stationary phase to the time
in the mobile phase.
K= tR-tM/tM
Retention volume (VR):
This is the volume of carrier gas required to elute one
half of the compound from the column by the peak maximum and is given by:
VR=tR x f
Adjusted retention volume (VR):
This allows for the gas hold up volume of the column which is due to the
interstitial volume of the column and the volume of the injector and detector
systems .It is given by:
V'R=t'R x f
Relative retention volume:
Retention volumes for compounds are expressed relative to the retention volume
of a standard compound on the same column under the same conditions of a
standard compound examined. Therefore, this ratio is given by:
VN(Sample)/VN(Standard)=t’R(Sample)/t’R(Standard)
Relative retention volumes can there fore be represented by ratios of the distances
on the recorder chart and are the same as relative retention times.
Height equivalent to a theoretical plate (HETP):
The column is considered as being made up of a large number of parallel layers or
‘theoretical plates’, and when the mobile phase passes down the column the
28
components of a mixture on the column distribute themselves between the
stationary and mobile phases in accordance with their partition that equilibrium is
established with in each plate. The equilibrium however is dynamic and the
components move down the column at a definite rate depending on the rate of
movement of the mobile phase.
A column may be considered as being made up of a large number of theoretical
plates where distribution of sample between liquid and gas phase occurs.
The number of theoretical plates (n) in a column is given by the relationship.
n=16(tR/W)2=5.54(tR/W1/2)2
W= peak width, i.e the segment of the peak base formed by projecting the straight
sides of the peak to the base line.
W1/2=peak width at half height
Resolution:
Chromatographers measure the quality of separation by resolution of adjacent
bands. T1 and t2retention times of the first and second adjacent bands.
RS=2(t2-t1)/W1-W2
W1 and W2 are the base line band width.
Column Efficiency (N):
Two related terms are widely used as quantitative measures of the efficiency of the
chromatographic columns.
I. Plate height
II. Number of theoretical plates
29
The two are related by the equation
N =L/H
Selectivity:
It measures relative retention of two components selectivity is the function of
chromatographic surface (column), melting point and temperature.
α =K’2/K’1=V2-V0/V1-V0
Method:
For achieving of stable base line Equilibration of the column with the prescribed
mobile phase and flow rate and room temperature or at the temperature specified
in the monograph and preparation of sample solution and to be examined and the
reference solution is require d the solutions must be free from solid particles.
Optimization of the method:
During optimization of the method5 the initial set of conditions have evolved from
the first stages of development are improved or maximized in terms of resolution
and peak shape plate counts asymmetry capacity elution time detection limits limit
of quantification and overall ability to quantify the specific analyte of interest.
The various parameters is that include to be optimized during method
development
Modes of separation
Selection of stationary phase
Selection of mobile phase
Selection of the detector
30
Selection modes of separation:
Reverse phase mode the mobile phase is comparatively more polar than the
stationary phase for the separation of polar or moderately polar compounds the
most preferred mode is reverse phase The nature of the analyte is the primary
factor in the selection of modes of separation. The second factor is based on nature
of the matrix.
Selection of stationary phase /column:
Selection of the column is the first and the most important step in method
development. The appropriate choice of separation column includes three different
approaches
1. Selection of separation system
2. The particle size and the nature of the column packing
3. The physical parameters of the column i.e the length and the diameter
The important parameters which should be selected the chromatographic column
Length and diameter of the column
Packing material
Shape of the particles
Size of the particles
% of the carbon loading
Pore volume
Surface area
End capping
31
The column is selected depending on the nature of the solute and the information
about the Analyte.Reverse phase mode of chromatography facilitates a wide range
of columns like dimethyl silane(C2) butylsilane(C4) octylsilane
(C8),octadecacylsilane(C18),base deactivated silane , BDS
phenyl,cyanopropyl(CN) nitro amino etc c18 was for this study since it is most
retentive one. The sample manipulation becomes easier with this type of column.
Due to higher theoretical plates the higher columns provide
better separation surface area available for coating increases as the particle size
decreases for the better efficacy reproducibility and reliability size of 5 μm the
column which we have selected should be of 5m and internal diameter of 4.6 mm.
Peak shape is equally important in method development columns
that provide symmetrical peaks are always preferred while peaks with poor
asymmetry result in
• Accurate plate number and resolution measurement
• Imprecise quantitation
• Degraded and undetected minor bands in the peak tail
• Poor retention reproducibility
A useful and practical measurement of peak shape is peak asymmetry factor and
peak tailing factor peak asymmetry is measured at 10% of full peak height and
peak tailing factor 5% reproducibility of retention times and capacity factor is
important for developing a rugged and repeatable method.
32
Selection of mobile phase:
The primary objective in selection and optimization of mobile phase is to achieve
optimum separation of all the individual impurities and degradants from each
other and from analyte peak.
In liquid chromatography the solute retention is governed by the
solute distribution factor, which reflects the different interactions of the solute-
stationary phase, solute-Mobile phase and the mobile phase –stationary phase for
the given stationary phase the retention of the given solute depends directly up on
the mobile phase, the nature and the composition of which has to be judiciously
selected in order to get appropriate and required solute retention.
The mobile phase has to be adopted in terms of elution strength
(solute retention) and the solvent selectivity (solute separation) solvent polarity is
the key word in chromatographic separations since a polar mobile phase will give
rise to low solute retention in normal phase and high solute retention in reverse
phase LC.
The selectivity will be particularly altered if the buffer pH is close to
the pKa of the Analytes: the solvent strength is a measure of its to pull an analyte
from the column it is generally controlled by the concentration of the solvent with
the highest strength.
The following parameters which shall be taken in to consideration while selecting
and optimizing the mobile phase.
33
• Buffer
• pH of the buffer
• Mobile phase composition
Selection of the detector:
The detector was chosen depend up on the some characteristic property of the
Analyte like UV absorbance, Fluorescence Conductance, Oxidation, Reduction
etc.
Characteristics that are to be fulfilled by a detector to be used in HPLC
determination are
Higher sensitivity, Facilitating trace analysis
Negligible base line noise to facilitate lower detection
Large linear dynamic range
Low dead volume
Non destructive to sample
Inexpensive to purchase and operate
Pharmaceutical ingredients do not all absorb UV light equally. So the selection of
detection wave length is important .An understanding of the UV light absorptive
properties of the organic impurities and the active pharmaceutical ingredient is
very helpful.
Fur the greatest Sensitivity λmax should be used .UV wave lengths
below 200nm should be avoided because detector noise increases in this region.
Higher wavelengths give greater selectivity.
34
HPLC METHOD DEVELOPMENT:
Systematic approach to HPLC method development5 should be based on the
knowledge of the chromatographic process. In most cases, a considerable amount
of experimentation may be needed. A good method development strategy should
require only as many experimental runs as are necessary to achieve desired final
result.
Fig. 5: Flow chart for Method development
1. Information on sample defines separation
Goals
2. Need for special HPLC procedure sample
Pre-treatment etc.
3. Choose of detector.
4. Choose LC method; preliminary runs; estimate
Best separation conditions.
5. Optimize separation conditions.
35
6. Requirements for separation procedures
7a. Recovery of purified material 7b.Quantitative method &
7c. Qualitative method
8. Validated method for laboratories, released to routine
36
Table 2. Choice of operating conditions to obtain the adequate
resolution of the mixture
Separation Variable Preferred Initial Choice
Column
Dimensions (length, ID) 15 X 0.46 cm
Particle size 5 μma
Stationary phase C8 or C18
Mobile phase
Solvents A and B Buffer- acetonitrile
% B 80-100%b
Buffer (compound, PH, concentration)
25mM potassium phosphate, 2.0<pH<3.0c
Additives (e.g., amine modifiers, ion- pair reagents)
Do not use initially
Flow rate 1.5-2.0 mL/min
Temperature 35-45oC
Sample Size
Volumed < 25μL
Weight <100 μg
a: 3.5 μm particles are an alternative, using a 7.5 cm column. b : For an initial isocratic run; an initial gradient run is preferred c No buffer required for neutral samples; for pH <2.5, pH-stable columns are recommended. d Smaller values required for smaller-volume columns (e.g., 7.5 x 0.46-cm, 3.5-μm column).
37
HPLC METHOD VALIDATION
Definition:
Method validation8-10 is defined as a process of providing that an analytical
method is acceptable for its intended use. Method validation provides the method
development extremely specific, linear, precise, accurate and sensitive.
OBJECTIVE OF THE VALIDATION
The primary objective of validation is to from a basis for written procedure
for production and process control which are designed to assure that the
drug products have the identity, quality, and purity they purport are
represented to possess.
Assurance of Quantity
Government Regulation
CONCEPT OF VALIDATION
The basic principle of quality assurance has as goal the production of
articles that are fit for their intended use.The principle may be stated as quality,
safety and effectiveness must be designed and built into the product and quality
cannot be inspected or tested to the finished product.Each step of the
manufacturing process must be controlled to maximize the probability that the
finished product meets all quality and design specification.
38
IMPORTANCE OF VALIDATION
As the quality of product cannot always be assured by routine quality
control because of testing of statically insignificant number of sample, the
validation thus should provide adequacy and reliability of a system or
product to meet the pre – determined criteria or attributes to provide high
degree of confidence that the same level of quality is consistently built into
each of finished product from batch to batch and to take action in case of
non compliance.
Retrospective Validation is useful for trend comparison of results
complains to cGMP to cGLP.
Fig.6: Types of Validation
VALIDATION
Analytical Method validation
Instrumental Validation
Process Validation
Prospective Validation
Retrospective Validation
Revalidation
39
ANALYTICAL METHOD VALIDATION
Method validation is the process for establishing that performance
characteristics of the analytical method are suitable for the intended application.
Chromatographic methods need to be validation before first routine use. To obtain
the most accurate results, all of the variables of the method should be considered,
including sampling procedure, sample preparation, chromatographic separation,
detection and data evaluation, using the same matrix as that of the intended
sample. The validity of an analytical method can only be verified by laboratory
studies. All validation experiments used to make claims or conclusions about
validity of the method should be documented in report.
Types of analytical procedures to be validated
Identification test for impurities
Quantitative test for impurities
Limit test control of impurities
Quantitative test for the active moiety in samples of drug substance or drug
product, or other selected components (s) in the drug product.
Dissolution testing.
BENEFITS OF VALIDATION
Regulatory compliance
Minimize rejection and reworking
Minimize utility cost
40
Minimize complaints
Reduce testing requirements
More rapid and reliable start – up new equipment
Easier scale – up from development
Easier maintenance of equipment
More rapid automation
The different parameters of analytical method development are discussed below:
System Suitability:
System suitability tests are an integral part of chromatographic methods.
These tests are used to verify that the resolution and reproducibility of the system
are adequate for the analysis to be performed. System suitability tests are based on
the concept that the equipment, electronics, analytical operations, and samples
constitute an integral system that can be evaluated as a whole. The purpose of the
system suitability test is to ensure that the complete testing system (including
instrument, reagents, columns, analysts) is suitable for the intended application.
Similar to the analytical method development, the system suitability test
strategy should be revised as the analysts develop more experience with the
assay. In general, consistency of system performance (e.g., replicate injections of
the standard) and chromatographic suitability (e.g. tailing factor, column
efficiency and resolution of the critical pair) are the main components of system
suitability.
41
During the early stage of the method development process some of the
more sophisticated system suitability tests may not be practical due to the lack of
experience with the method. In this stage, usually a more "generic" approach is
used. For example, evaluation of the tailing factor to check chromatographic
suitability, and replicate injections of the system suitability solution to check
injection precision may be sufficient for an HPLC impurities assay. As the method
matures more experience is acquired for this method, a more sophisticated system
suitability test may be necessary.
System suitability is the checking of a system to ensure system
performance before or during the analysis of unknowns. Parameters such as plate
count, tailing factors, resolution and reproducibility (%RSD retention time and
area for six repetitions) are determined and compared against the specifications set
for the method. These parameters are measured during the analysis of a system
suitability "sample" that is a mixture of main components and expected by-
products.
The following table lists the terms to be measured and their recommended
limits obtained from the analysis of the system suitability sample as per current
FDA guidelines on "Validation of Chromatographic Methods".
42
Table 3.System Suitability Parameters and Recommendations
Parameter Recommendation
Capacity Factor (k’) The peak should be well-resolved from other peaks and the void volume, generally k’>2.0
Repeatability RSD < 1% for N > 5 is desirable.
Relative retention Not essential as long as the resolution is stated.
Resolution (Rs) Rs of > 2 between the peak of interest and the closest eluting potential interferent (impurity, excipient, degradation product, internal standard, etc.
TailingFactor(T) T of < 2
Theoretical Plates (N)
N > 2000
Accuracy:
Accuracy is the measure of exactness of an analytical method, or the
closeness of agreement between the value which is accepted either as a
conventional true value or an accepted reference value and the value found. It is
measured as the percent of analyte recovered by assay, by spiking samples in a
blind study. For the assay of the drug substance, accuracy measurements are
obtained by comparison of the results with the analysis of standard reference
material or by comparison to a second, well – characterized method. For the assay
of the drug product, accuracy is evaluated by analyzing synthetic mixtures spiked
with known quantities of components. For the quantitation of impurities, accuracy
43
is determined by analyzing samples (drug substance or drug product) spiked with
known amounts of impurities are not available, see specificity.)
To document accuracy the ICH guideline on methodology recommends
collecting data from a minimum of nine determinations over a minimum of three
concentration levels covering the specified range (for example, three
concentrations, three replicates each).The data should be reported as the percent
recovery of the known, added amount, or as the difference between the mean and
true value with confidence intervals.
Precision:
Precision is the measure of the degree of repeatability of an analytical
method under normal operation and is normally expressed as the percent relative
standard deviation for a statistically significant number of samples. According to
the ICH precision should be performed at three different levels: repeatability,
intermediate precision and reproducibility. Repeatability is the results of the
method operating over a short time interval under the same conditions (inter-assay
precision). It should be determined from a minimum of nine determinations
covering the specified range of the procedure (for example, three levels, three
repetitions each) or from a minimum of six determinations at 100 % of the test or
target concentration. Intermediate precision is the results from within lab variation
due to random events such as different days, analysts, equipment, etc. In
determining intermediate precision, experimental design should be employed so
that the effects (if any) of the individual variables can be monitored.
44
Documenting precision:
Reproducibility refers to the results of collaborative studies between
laboratories. Documentation in support of precision studies should include the
standard deviation relative standard deviation, coefficient of variation, and the
confidence interval.
Specificity:
Specificity is the ability to measure accurately and specifically the analyte
of interest in the presence of the other components that may be expected to b
present in the sample matrix. It is a measure of the degree of interference from
such things as other active ingredients, excipients, impurities and degradation
products, ensuring that a peak response is due to a single component only, i.e. that
no co- elution exist.
Specificity is measured and documented in a separation by the resolution,
plate count (efficiency), and tailing factor. Specificity can also be evaluated with
modern photodiode array detectors that compare spectra collected across a peak
mathematically as an indication of peak homogeneity ICH also use the term
specificity, and divide it in to two separate categories: identification and
assay/impurity tests.
For identification purposes, specificity is demonstrated by the ability to
discriminate between compounds of closely related structures, or by comparison to
known reference materials. For assay and impurity tests, specificity is
demonstrated by the resolution of the two closest eluting compounds. These
45
compounds are usually the major component or active ingredient and an impurity.
If impurities are available, it must be demonstrated that the assay is unaffected by
the presence of spiked materials (impurities and /or excipients). If the impurities
are not available, the test results are compared to a second well- characterized
procedure. For impurity tests, the impurity profiles are compared head- to-head.
In case of the assay, demonstration of specificity requires that the
procedure is unaffected by the presence of impurities or excipients. In practice,
this can be done by spiking the drug substances or product with appropriate levels
of impurities or excipients and demonstrating that the assay is unaffected by the
presence of these extraneous materials. If the degradation product impurity
standards are unavailable, specificity may be demonstrated by comparing the test
results of samples containing impurities or degradation products to a second well-
characterized procedure. These comparisons should include samples stored under
relevant stress conditions (e.g. light, heat humidity, acid/base hydrolysis,
oxidation).
Limit of Detection:
The limit of detection (LOD) is defined as the lowest concentration of an
analyte in a sample that can be detected, not quantitated. It is a limit test that
specifies whether are not an analyte is above or below a certain value. It is
expressed as concentration at a specified signal-to-noise ratio, usually two-or
46
three-to-one. The ICH has recognized the signal- to- noise ratio convention, but
also lists two other options to determine LOD: Visual non- instrumental methods
and a means of calculating the LOD. Visual non- instrumental methods may
include LOD’S determined by techniques such as thin layer
chromatography(TLC) or titration .LOD’s may also be calculated based on the
standard deviation of the response (SD) and the slope of the calibration curve(S) at
levels approximating the LOD according to the formula :
LOD=3.3(SD/S)
The standard deviation of the response can be determined based on standard
deviation of the blank, on the residual standard deviation of the regression line, or
the standard deviation of y- intercepts of regression lines.
If LOD is determined based on visual evaluation or based on signal to noise ratio,
the presentation of the relevant chromatograms is considered acceptable for
justification.
In cases where an estimated value for the detection limit is obtained by
calculation or extrapolation, this estimate may subsequently be validated by the
independent analysis of a suitable number of samples known to be near or prepared
at the detection limit.
Limit of quantitation:
The limit of quantitation (LOQ) is defined as the lowest concentration of an
analyte in a sample that can be determined with acceptable precision and accuracy
47
under the stated operational conditions of the method. Like LOD, LOQ is
expressed as a concentration with the precision and accuracy of the measurement
also reported. Sometimes a signal- to-noise ratio of ten-to- one is used to
determine LOQ. That is as the LOQ concentration level decreases the precision
increases .If better precision is required, a higher concentration must report for
LOQ. This compromise is dictated by the analytical method and its intended use.
The ICH has recognized the ten-to-one-signal-to-noise ratio as typical, and also,
like LOD, lists the same two additional options that can be used to determine
LOQ, visual non-instrumental methods and a means of calculating the LOQ. The
method is again based on the standard deviation of the response (SD) and the slope
of the calibration curve (S) according to the formula:
LOQ=10(SD/S)
Again, the standard deviation of the response can be determined based on the
standard deviation of the blank, on the residual standard deviation of the
regression line, or the standard deviation of y intercepts of regression lines.
Linearity and Range:
Linearity is the ability of the method to elicit test results that are directly
proportional to analyte concentration within a given range .Linearity is generally
reported as the variance of the slope of the regression line. Range is the interval
between the upper and lower levels of analyte (inclusive) that have been
demonstrated to be determined with precision, accuracy and linearity using the
method as written. The range is normally expressed in the same units as the test
48
results obtained by the method. The ICH guidelines specify a minimum of five
concentration levels, along with certain minimum specified ranges.
For assay, the minimum specified range is from 80-120 % of the target
concentration. For an impurity test, the minimum range is from the reporting level
of each impurity, to 120 % of the specification (for toxic or more pote nt
impurities, the range should be commensurate with the controlled level).
For content uniformity testing, the minimum range is from 70-130 % of the
test or target concentration, and for dissolution testing ±20 % over the specified
range of the test. That is, in the case of an extended release product dissolution
test, with a Q- factor of 20 % dissolved after six hours, and 0 % dissolved after 24
hrs, the range would be 0-100 %.
Ruggedness:
Ruggedness, according to the USP, is the degree of reproducibility of the
results obtained under a variety of conditions, expressed as % RSD. These
conditions include different laboratories, analysts, instruments, reagents, days, etc.
In the guide line on definitions and terminology, the ICH did not address
ruggedness specifically. This apparent omission is really a matter of semantics,
however, as ICH chose instead to cover the topic of ruggedness as precision, as
discussed previously.
Robustness:
Robustness is the capacity of a method to remain unaffected by small
deliberate variations in method parameters. The robustness of a method is
49
evaluated by varying method parameters such as mobile phase ratio, pH, ionic
strength, temperature, flow rate etc and determining the effect (if any) on the
results of the method. As documented in the ICH guidelines, robustness should be
considered early in the development of the method. In addition, if the results of a
method or other measurements are susceptible to variations in method parameters,
these parameters should be adequately controlled and a precautionary statement
included in the method documentation.
50
REFERENCES
1. Satinder, A., Stephen, S., Hand Book of Modern Pharmaceutical Analysis.,
Published by
Academic Press, London., (2001),(3), 1-2.
2. Yang H, Feng Y and Luan Y., Simultaneous Determination of Simvastatin and
Ezetimibe in
Tablets by HPLC.,J Chromatogr B., 2003, 785, 369.
3. Srivastava, VK., and Srivastava, KK., Introduction to Chromatography Theory
and Practice,
14th Edition., S.Chand and Company limited, New Delhi., (1991), 66-67.
4. Sethi PD., Quantitative Analysis of Pharmaceutical Formulations., 1st ed., CBS
Publishers
and Distributors, New Delhi. (2001), 3-5.
5. Snyder, LR., Joseph Kirkland, J., Joseph Glajch, L., Practical HPLC Method
Development.,
2nd ed., John Wiley and Sons, INC, Canada., (1997), 2-11.
6. Melani L, Mills R and Hassman D., Efficacy and safety of ezetimibe co-
administered with
pravastatin in patients with primary hypercholesterolemia: a prospective,
randomized, Double-
blind trial., Eur Heart J., 2003, 24, 717-728.
51
7. Curlucci G, Mazzeo P, Biordi L and Bologna M., Simultaneous determination
of simvastatin
and its hydroxy acid form in human plasma by high performance liquid
chromatography with
UVdetection., J. Pharm. Biomed. Anal., 1992, 10 (9),693-697.
8. International Conference on Harmonization, Draft Guideline on Validation of
Analytical
Procedures: Definitions and Terminology, Federal Register, 60 (1995)
11260.,1996(1-8).
9. Center for Drug Evaluation and Research, Food and Drug
Administration,Reviewer Guidance,
Validation of Chromatographic Methods. 1994.
10. Guideline for Submitting Samples and Analytical Data for Methods
Validation. Food and
Drug Administration, 1987.