Upload
adham08
View
26
Download
1
Embed Size (px)
Citation preview
Chapter 23 Airway Volumes, Flows, and Pressures P.729
Definitions
• Compl iance : Ratio of a change in volume to a change in pressure . It is a
measure of dis tensibi l i ty and is usually expressed in mill i l i ters per centimeter
of water (mL/cm H2O). Compliance commonly refers to the lungs and chest
wal l . Breath ing system components, especia lly breathing tubes and the
reservoir bag, also have compl iance.
• Expiratory Flow Rate : Rate at wh ich gas is exhaled by the patient expressed
as volume per uni t of time.
• Expiratory Flow Time : Time between the beginning and end of expiratory f low
(Fig . 23.1).
• Expiratory Pause Time : Time from the end of exp iratory f low to the s tart of
inspiratory f low (Fig . 23.1).
• Expiratory Phase Time : Time between the start of exp iratory f low and the
start of inspiratory f low. It is the sum of the exp iratory f low and expiratory
pause times (Fig . 23.1).
• Inspiratory Flow Time : Period between the beginning and end of inspiratory
f low (Fig . 23.1).
• Inspiratory Pause Time : That portion of the inspira tory phase time during
which the lungs are held inf lated a t a fixed pressure or volume (i .e., the t ime
of zero f low (Fig. 23 .1). I t is also called the inspira tory hold , inf la tion hold ,
and insp iratory plateau .
• Inspiratory Phase Time : Time between the start of inspiratory f low and the
beginning of expiratory f low (Fig. 23 .1). I t is the sum of the inspiratory f low
and inspiratory pause t imes. The insp iratory pause t ime:insp iratory phase
t ime (T IP :T I) may be expressed as a percentage.
• Inspiratory:Exp iratory Phase Time Ratio (I :E rat io): Ratio of the inspiratory
phase time to the exp iratory phase t ime. For example, an I :E ratio of 1:2
means tha t the inspiratory phase t ime is one thi rd of the venti la to ry cycle
t ime.
• Inspiratory Flow Rate : Rate at which gas f lows in to the pat ient expressed as
volume per uni t of t ime.
• Minute Volume : Sum of al l tidal volumes within 1 minute.
• Peak Pressure : Max imum pressure during the inspiratory phase t ime (Fig.
23.1).
• Plateau Pressure : Rest ing airway pressure during the inspiratory pause.
There is usua lly a lowering of airway pressure f rom peak pressure when
there is an inspiratory pause (F ig. 23.1). This lower pressure is cal led the
plateau pressure .
• Pos it ive End-expiratory Pressure (PEEP): Pos it ive pressure in the ai rway a t
the end of exhalat ion.
• Resis tance : Ratio of the change in driv ing pressure to the change in f low
rate . It is commonly expressed as centimeters of water per l i ter per second
(cm H2O/L/second).
• Tidal Volume : Volume of gas entering or leaving the patien t during the
inspiratory or expira tory phase time, respect ively.
• Venti la to ry (Respira tory) Rate or Frequency: Number of resp iratory cycles
per uni t time, usually per minute.
• Work o f Breathing : Energy expended by the patient and/or venti lator to move
gas in and out of the lungs (1). I t is expressed as the ratio of work to volume
moved, commonly as joules per l i ter. It inc ludes both the work needed to
overcome the elas tic and f low-resistive forces of the both respiratory system
and apparatus.
View Figure
Figure 23.1 Flow, volume, and pressure curves from a ventilator that produces a rectangular inspiratory flow wave. A: This represents controlled ventilation with no inspiratory pause. The end-inspiratory pressure will equal the peak pressure. B: With an inspiratory pause, there is a decrease from peak pressure to a lower plateau pressure. C: This illustrates the effect of continuing fresh gas flow during inspiration. The inspired volume increases, and the peak pressure falls, then rises.
P.730
General Considerations
The Ventilatory (Respiratory) Cycle Airway pressure with controlled venti lat ion is shown in Figure 23.1. There is a r ise
in p ressure with no preced ing negat ive pressure. A fast r ise to peak pressure
sugges ts too high a f low. The peak pressure wi l l increase if tidal volume,
inspiratory f low rate , or res istance increases or compl iance dec reases . A decrease
in peak pressure may resu lt f rom a leak, spontaneous inspiratory effort by the
patient, a dec rease in res istance, or an increase in compliance.
Figure 23.1B shows the respira tory cycle wi th an inspiratory pause. If the pause is
long enough, a p la teau pressure wil l occur a t the end of inspiration . The plateau
pressure is usually preceded by a higher peak pressure .
P.731
Plateau pressure depends on t idal volume and the total s tat ic compl iance but is
independent of resistance (2).
Compliance and Resistance
In the past, compliance and resistance measurements during anesthesia were
diff icul t and involved bulky apparatus. They can now be measured accura tely on a
real-t ime basis with relat ively compac t equ ipment.
Compliance Compl iance measurement may be dynamic or s ta tic . Dynamic compl iance is
ca lcula ted by d iv iding the difference in volume by the dif ference in p ressure at two
points during the vent i latory cycle. Th is is not a true measure of total compliance,
because the ai rway pressure inc ludes the pressure needed to overcome resistance
(3).
Static compliance is ca lculated by using the end-inspiratory occ lusion pressure
(4,5) (Fig. 23.22). Condi t ions of zero gas f low are achieved by employing an
inspiratory ho ld or occluding the expiratory port long enough to allow ai rway
pressure to reach a cons tant value . This pressure, commonly te rmed plateau
pressure , represents the elastic recoil of the tota l respira tory sys tem at end-
inf lat ion volume.
Static Compliance = Tidal volume/Plateau pressure - Posit ive end-exp iratory
pressure (PEEP)
In adul ts , normal total s tatic compliance is 35 to 100 mL/cm H2O. In children,
normal s tat ic compl iance is greater than 15 mL/cm H2O (6).
Total compliance ref lec ts the elast ic propert ies of the lungs, thorax, abdomen, and
the breathing system. Using muscle relaxants wi l l increase chest wall compl iance
but wil l no t affec t lung compliance, so in paralyzed patients, changes in compliance
ref lec t mainly al tera tions in lung compl iance.
Resistance When gas f lows th rough a tube, energy is lost. This is ref lec ted by a decrease in
pressure . The pressure drop can be expressed as the product of resistance and
f low rate . For a given tidal volume, a high resis tance may be overcome by using a
lower f low for a longer time or a higher driv ing pressure . During control led
venti lation , if there is an inc rease in ai rway resistance, the pressure needed to
deliver a given tidal volume wil l increase. This can usually be supplied by the
venti lato r or the person squeezing the reservoir bag so tha t inspira tory f low is not
af fected. Because exhalat ion is passive, expiratory f low depends on the elas tic and
resist ive forces of the lungs and the resistance in the expiratory l imb of the
breathing sys tem and airway dev ice.
Total res istance, wh ich may d if fe r during inspiration and exp irat ion , is determined
predominantly by the res is tance of the pat ient 's ai rway, the tracheal tube, and the
breathing sys tem. Decreased airway ca liber from bronchoconstric tion, secret ions,
tumor, edema, a foreign body, or airway c losure is associated wi th inc reased
resistance. Tracheal tube resistance depends primari ly on its in ternal diameter.
Partial tube obstruction by secretions, kink ing , or other p roblems wil l cause
increased res is tance. Breathing system resistance is affected by the length and
inte rna l diameter of i ts components and is inc reased by sharp bends and
constric t ions.
Total airway resistance can be est imated by using the dif fe rence between peak and
plateau pressures, which is normally 2 to 5 cm H2O. If there is an increase in
resistance, a higher peak pressure wi l l be necessary to produce the same f low.
Plateau pressure, however, depends only on compliance and wi l l not be affected by
resistance. Therefore, if the inspiratory f low and tidal volume remain constant but
resistance increases, there wil l be a greater d if fe rence between the peak and
plateau pressures.
Measured Gas Composition The compos it ion of the gas being measured wil l af fect the accuracy of f low-
measuring devices (7,8). Dif ferences in densi ty and v iscosity of the gases can
induce an error in f low measurement. The composi tion of carrier gas has a greater
impact on v iscosity than volati le anesthetic agents, whereas densi ty is more
inf luenced by volati le agent concentrat ions (9).
For accuracy, the f low-measuring dev ice should be associated wi th a gas monitor
that can make corrections to gas f low caused by changes in gas composi tion . If a
gas such as xenon that is not measured is p resent, f low measurements may be
inaccurate (8).
Respiratory Volume and Flow Measurement A respirometer (sp irometer, vent i lat ion or respira tory meter or mon itor,
venti lometer, volume measuring dev ice, f low monitor, respiratory flowmeter) is a
device tha t measures the volume of gas passing during a period of t ime through a
locat ion in a f low pathway (10).
Moni toring respiratory volumes and f lows can aid in detect ing breath ing system
obstruct ions, disconnect ions, apnea, leaks , venti la tor fai lure , and high or low
volumes in spontaneously breath ing patien ts as wel l as in those whose venti la tion
is contro lled. Some can detec t reversed f low, an indicat ion of an incompetent
unidi rect ional valve or a leak. A discrepancy between expired and inspired tidal
volume should suggest a leak .
P.732
A decrease in t idal volume assoc iated wi th the tracheal tube migrating into a
bronchus may be detec ted (11). Al though there are other ways of detecting these
problems, such as observ ing chest wal l movements, monitoring breath sounds ,
capnometry, and airway pressure moni toring , the use of a volume monitor provides
additional pro tec tion. One study found that for detec ting and c lassify ing breathing
system faul ts , a volume moni tor was better than ai rway pressure or carbon d ioxide
moni toring (12). In 1999, the American Society of Anesthesiologists (ASA) strongly
encouraged qual itat ive moni toring of the volume of expired gas. The anesthesia
workstat ion standard requires a device to moni tor the patient's exhaled tidal or
minute volume or both (13).
Respiratory volume moni toring may fa il to detect some problems. With ai rway
occlusion, there may be enough f low during expiration resul t ing f rom compression
of gas within the breathing system during inspiration to prevent the respirometer
alarm f rom being ac tivated. If the sensor is attached to the patient 's tracheal tube
or supraglottic ai rway device, a disconnect ion between the sensor and the
breathing sys tem wi ll not be detected if the patient is spontaneously breathing . If
the sensor is in the exha la tion s ide of the breathing system, the disconnection wi l l
be detec ted. It is possible to have fai r ly normal f low with esophageal intubat ion.
A high-volume alarm may be useful to de tect unant ic ipa ted increases in t idal
volume (14). This may be due to improper venti la tor settings or inc reased gas f low
into the breathing sys tem during inspirat ion, resu lt ing from a hole in the vent ilator
bellows, an inc reased insp iratory:expiratory (I :E) rat io , or f rom the f lowmeters (if
there is no f resh gas flow compensation or decoupling). During pressure control
venti lation , a decrease in compliance wil l resul t in an increased t idal volume.
Older respirometers were stric tly mechanical dev ices. Newer respirometers convert
f low into an e lectronic signa l that is processed and displayed. Elec tronic processing
enhances alarm capabi l i ty. Alarm limi ts should be set as close as possible to the
displayed tidal or minute volume without producing an unacceptable inc idence o f
false-posi t ive a larms (15).
Equipment Ventilator Bellows Scale
Venti la to rs that are used in anesthesia are discussed in Chapter 12. I f the
venti lato r has a bel lows, there is usual ly a scale on the bel lows housing . This scale
can prov ide a rough estimate of t ida l volume de livered into the breathing system
but is no t an accura te es timate of the volume delivered to the patien t because of
was ted vent ilation secondary to gas compression and dis tension of components o f
the breathing system. I f the fresh gas f low adds to the t idal volume during
inspiration (see Chapter 12), th is added volume wi l l not be represented on this
scale.
View Figure
Figure 23.2 Wright respirometers. A: This small instrument can be handheld or inserted into the breathing system. It has two dials: a large peripheral one and a smaller one on the upper part of the main dial. The small dial indicates volumes up to 1 L and the large dial up to 100 L. Note the reset button on the side. B: This version has three dials. The top small dial reads up to 1 L, the large dial indicates volumes up to 100 L, and the bottom small dial reads up to 10,000 L. Note the on-off control and the directional flow arrow. (Courtesy of Ferraris Medical, Inc.)
Wright Respirometers
Description Typical Wrigh t respirometers are shown in F igure 23.2. They are supp lied wi th
adaptors to fac il i tate connection to a mask, ai rway device, or breathing sys tem.
There is an ON-OFF control in the form of a s liding stud and a spring-loaded reset
button to se t the hands o f the scales to zero .
An infant vers ion that can measure volumes down to 15 mL is avai lable (Fig. 23 .3).
I ts dead space is 15 mL. An e lec tronic vers ion is also available (15).
The internal construction is shown in Figure 23.4 . Gas entering through the outer
casing is di rected through a series of tangential s lots enclosed in a cy lindrical
housing and strikes a vane, causing i t to rotate. The vane is connected by a
mechanical gear system to the hands on the dial so that a reading corresponding to
the volume of gas pass ing through the dev ice is registered.
Evaluation Most studies have found tha t the Wright respirometer over-reads at high flows and
under-reads at low f lows (16,17,18). Pulsati le f lows can cause addit ional over-
reading. It wi l l give s lightly higher readings wi th mixtures of ni trous oxide and
oxygen than for ai r and wi l l s l ightly over-read in the presence of xenon (8).
P.733
View Figure
Figure 23.3 Infant version of Wright respirometer. The outer scale goes up to 500 mL, and the inner scale goes up to 5 L. (Courtesy of Ferraris Medical, Inc.)
Advantages of the Wright respirometer inc lude its small s ize and light weight. Its
low dead space makes it sui table for use between the patien t and the breath ing
system.
The main disadvantage is that i t has no alarms. I t is somewhat di ff icul t to read and
does not give respiratory rate. A clock is necessary to de termine minute volume. It
does not read bidi rectional f low. Maintenance can be expensive. Many instruments
that are in use are inaccurate because of poor mechanical cond ition. I ts portabil i ty
can resu lt in inaccuracy due to pocket di rt and a high incidence of damage f rom
being dropped. Guards that are designed to reduce damage from phys ical abuse
are avai lable. I t needs to be c leaned and dis infected between patien ts .
Spiromed
Description
The Spiromed is an electron ic respirometer that is designed for use wi th North
American Drager breathing systems (19,20). As gas f lows through the monitor, i t
forces a pair of rotors to counter-rotate (Fig. 23 .5). Attached to the axle of one of
the rotors is a four-pronged armature with a smal l magnet at the tip of each prong.
As exhaled gas f lows through the sensor, the rotor and armature spin in un ison.
Located at approximately the 12 and 7 o'c lock posi tions are two transistors that
turn ON in the presence of a magnetic f ield. As the armature that carries the
magnets rotates, the trans is tors are turned ON and OFF. These paired pulses are
transmitted through the sensor cable to the inte rface panel and then to the
processor.
View Figure
Figure 23.4 Internal construction of Wright respirometer. Gas entering the casing is directed through a series of tangential slots and strikes the vane in the center, causing it to rotate.
View Figure
Figure 23.5 Spiromed. (See text for details.)
The number of paired pulses is related to the volume of gas tha t passes through
the sensor over t ime. The to tal number of pulse pairs counted during each
exhalation determines the tidal volume. The speed a t which the exhaled gas f lows
through the sensor determines the dura tion of each pulse pa ir. Rapid gas f low
causes the ro tors and armature to spin quick ly, and shorter pulses are produced as
the trans istors rap id ly cycle. With slower gas f low, longer pulses are produced. The
processor analyzes the pulse lengths and displays the information as the exhaled
waveform.
Sixty seconds of cont inuous data is required for the in it ia l display of the respira tory
rate , and the displayed read ing is recalculated af ter each exhalation. For an
exhalation to be counted as a “valid b reath,” the processor mus t count at least 80
mL. All exhaled gas volumes, regardless of s ize, are counted and inc luded in
ca lcula ting the minute volume. The sensor in the breath ing system is shown in
Figure 23.6.
The Spiromed sensor recognizes the di rection of gas flow by monitoring the phase
rela tionship between the pulses in each pu lse pair. When the gas f lows “forward,”
the pulse from one transisto r leads the pulse from the o ther transducer because of
the armature's rotat ional direct ion. I f gas f lows in the wrong direc tion, the order
P.734
of the pu lses is reversed and the processor recogn izes this as reverse f low. If two
consecutive pulse pairs are in reverse order, a reverse f low alarm is generated.
View Figure
Figure 23.6 Spiromed in place in breathing system.
Evaluation This instrument is programmed to measure t idal volumes equal to or g reater than
150 mL. If the t ida l volume is less than 150 mL, the instrument wi l l automatica lly
add two or more consecut ive tidal volumes and reduce the recorded frequency
accordingly. The minute-volume display remains correct.
The accuracy o f tidal volume measurement is reported as ±0 .04 L, minute volume
as ±10% of reading or 0.1 L, and resp iratory rate as ±8% of reading or 1
breath/minute.
D-Lite Gas Sampler and Flow Sensor
Description The f low sensor (Fig. 23.7) fo r this device is a modif ied F leisch pneumotach with a
two-s ided Pitot tube (21,22,23,24). The Fleisch pneumotach measures f low by
measuring the pressure difference across a f low resisto r (capil lary tube) in a tube.
The Pi tot tube uses two sensing tubes to make a d if fe rential p ressure
measurement. One tube faces the di rec tion of f low (to tal pressure), and the other
faces the opposi te di rect ion to measure the static pressure . The dif ference in
pressure between the total pressure and sta tic pressure is the dynamic pressure,
which is proportiona l to the square of gas f low.
The sensor body (Fig. 23 .8) consists of a straight tube wi th a combined 15-mm
female/22-mm male connector on the pat ient end and a 15-mm male connector on
the machine end. Two smal l hollow pressure tubes perforate the side of the tube
and extend in to the lumen. Each makes a 90-degree turn ins ide the lumen so that
the end of one tube faces the breathing system and the o ther end faces the pat ient.
A gas sampling port is also present. A double-lumen tube conduc ts the f low s igna l
as a pressure dif ference to the pressure sensor ins ide the moni to r.
The sensor is placed between the breathing system and the pat ient. A f il ter or heat
and moisture exchanger may be placed on e ither s ide of the sensor. If placed
between the patient and the sensor, mucus and humid ity wil l be prevented from
entering the gas sampling tube. If the sensor is placed between the patient and the
heat and moisture exchanger, a higher compl iance wi l l be observed than if i t is
placed between the heat and moisture exchanger and the Y-piece (25).
View Figure
Figure 23.7 D-Lite flow sensor and gas sampler. The patient end has a 15-mm internal and 22-mm outside diameter connector to fit a mask or tracheal tube connector. The other end has a 15-mm outside diameter connector. Because the pressure tubes point in opposite directions, gas flows can be measured during both inspiration and exhalation. Note that one pressure tube is larger than the other to avoid misconnection of the tubings. There is a gas sampling port on the opposite site of the sensor.
P.735
View Figure
Figure 23.8 Sensor for D-Lite flow sensor and gas sampler A: The gas sample port is at the top. Note that the pressure line attachments utilize male and female connectors. B: Attachment for pressure tubings.
During inspirat ion, gas moves f rom the breath ing system toward the patient. The
pressure in the hollow tube fac ing the breathing system and the pressure in the
tube that faces away from the di rect ion of gas f low are measured. Since the
pressure tubings face in opposi te di rections, s imilar measurements can be made
during exhala tion, when the gas flow is reversed.
In the moni tor (Fig . 23.9), concentra tions of carbon dioxide, oxygen, and anes thetic
agents are determined. The moni tor u ti l izes the gas composi tion data to
P.736
compensate for changes due to densi ty and v iscosity (7 ). A correct ion factor must
be applied if helium is in the respira tory mixture (26). From the derived f lows (f low
rate , peak f low) and measured pressures (end-expiratory, plateau, minimum, and
maximum), the inspiratory and expiratory t idal and minute volumes , compliance,
and resistance are ca lcula ted and displayed, and f low-volume and pressure-volume
loops are displayed. Inspired and expired gas concentrat ions are also d isplayed.
View Figure
Figure 23.9 The monitor utilizes the gas composition data to compensate for changes due to density and viscosity. From the derived flows and measured pressures, the inspiratory and expiratory tidal and minute volumes, compliance, and resistance are calculated and displayed, and flow-volume and pressure-volume loops are displayed. Inspired and expired gas concentrations are also displayed.
Alarms inc lude a h igh PEEP alarm wi th a default value of 10 cm H2O. High pressure
and low inspira tory pressure alarms have defau lt settings of 40 and 0 cm H2O.
There are high and low expira tory minute vo lume a larms as wel l as messages for
leak, disconnect ion, and obstruction .
Calibration The sensor needs to be ca librated a t least every 6 months (27). One indica tion that
ca libration needs to be performed is an open or overshoot pressure- and flow-
volume loop (Figs. 23 .61 and 23.62). The opera tion manua l should be consulted for
the complete procedure. It should be perfo rmed wi th the equ ipment in the
configuration tha t wil l be used wi th the next patient. Accessories, such as heat and
moisture exchangers, placed proximal to the sensor wil l no t af fec t calibra tion.
However, if the c linic ian wishes to p lace an accessory on the distal end of the
sensor, then the uni t should be cal ibrated wi th this in place. A dif ferent size of
tracheal tube or the omission of the connector could affect the ca libration value and
resul t in incorrec t volume measurements.
Evaluation The D-Li te is used to measure f lows in ranges common for adul ts and chi ld ren down
to 3 kg. Res istance to f low is 0.5 cm H2O at 30 L/minute. I ts volume is 9.5 mL.
Tidal volumes of 150 to 2000 mL and minute volumes of 2.5 to 30 L/minute can be
measured. The pedi-l i te sensor can measure t idal volumes in the range from 15 to
300 mL. The range of measurement for a irway pressure is -20 to +80 cm H2O. The
range of f low rates is -100 to +100 L/minute. I t over-reads when xenon is used (8).
Regular v isual inspec tion and water removal is required for troublefree
performance. I f used fo r extended periods with heavy humidif icat ion, condensed
water may occlude the pressure sensing or gas sampl ing tubes. The pressure and
sample l ine tubings shou ld be on the upper s ide of the sensor.
The D-Li te sensor has a simple and robust construct ion, l igh t weight, low dead
space, and no moving parts. I t is not posit ion dependent and al lows bidi rec tiona l
gas f low measurement. Small amounts of mucus and water d roplets do not affect
the measurements . Only one adaptor is needed for respirometry and gas sampling.
I t can be used with both c i rc le and Mapleson breathing sys tems. Another important
advantage is the abil i ty to moni tor f low-volume and pressure-volume loops.
View Figure
Figure 23.10 Novometrics sidestream sensor.
Novometrics Sidestream Sensor The Novometrics s idestream sensor (Fig. 23 .10) combines measurement of f low,
pressure , and carbon d ioxide. Flow is determined f rom a different ial pressure
measurement across a f ixed orif ice wi th a spl i t obs truct ion to one side of the optical
window used for mainstream carbon dioxide measurement. The sensor has a dead
space of less than 0 .8 mL, mak ing i t useful for the neonate (28).
Heated Wire Anemometer In the heated-wire anemometer (thermal diss ipat ion device) (Figs. 23.11, 23.12),
gas f lows around a th in wire (usually plat inum or a p lat inum al loy) that is heated to
a constant temperature (10). Heat is diss ipated when gas f lows past this wire. The
greater the volume of gas f lowing pas t per uni t time, the more heat wi l l be
diss ipa ted. The current level is usual ly low so that the outs ide of the sensor does
not become heated.
View Figure
Figure 23.11 Heated wire anemometer.
P.737
View Figure
Figure 23.12 Heated wire anemometer. One wire is for measuring flow, and one is for reference.
The hot-wire sensor tends to be more accurate at low f low rates. Since it is
insensi tive to f low d irect ion, two heated wires are needed to determine the f low
di rection . The heat dissipa ted by the second wire is determined when there is no
gas f low (e.g., during inhala tion in the exhalat ion s ide of the breathing system). I t
under-reads in the presence of xenon and sl ightly over-reads wi th ni trous oxide (8).
Ultrasonic Flow Sensor An ul trasound sensor (Figs. 23.13, 23 .14, 23.15) measures the inf luence of gas
f low on the transmission times of pulses between two crystals. Sending and
receiv ing transducers are used to transmit s ignals through the f low. The s ignal
travels fas ter when moving wi th the f low stream rather than aga inst the f low stream.
The dif fe rence between the two transmission times is used to calculate the f low
rate .
The f low sensor has no mov ing parts, is easy to c lean, is autoclavable, and i ts
accuracy is independent of gas composit ion. The f low measurement range is 0 to
120 L /minute. The accuracy is specif ied as ±10% or 15 mL, whichever is greater.
Resis tance is less than 2 cm H2O at a f low of 60 L/minute.
Variable Orifice Flow Sensor The variable orif ice f low sensor is used wi th the Ohmeda 7900 series venti la tors
(Chapter 12) and the ci rcle system. Sensors a t both connect ions to the carbon
dioxide absorber a re used to measure insp iratory and expira tory f lows . These
sensors can be used to generate pressure- and f low-volume loops. The 7900
venti lato r ut i l izes the information f rom these sensors to a llow it to del iver accurate
t ida l volumes.
Construction Each sensor (F igs. 23.16 , 23.17) uses the princ iple of p ressure drop across an
orif ice. A plast ic flap that opens wi th increasing f lows is placed across the di rec tion
of gas f low. Two sensors and a transducer
P.738
inside the anes thesia machine measure pressure proximal and distal to the f lap.
Volume is calculated from these f lows . The sensor on the inspiratory s ide is
connec ted to a pressure sensor so that b reathing system pressure is measured.
The information is used by the venti la tor to compensate for changes in fresh gas
f low.
View Figure
Figure 23.13 Ultrasonic flow sensor in breathing system.
View Figure
Figure 23.14 Inside of ultrasonic flow sensor.
Use Before use, the tubes should be checked to make certain that they are clear. The
pressure l ines should poin t up, and there should be no k inks , cracks , or other
problems. The use of f i l ters is recommended to protec t the sensors f rom
contamination. Calibrat ion by using a menu on the vent ilator is recommended on a
weekly basis.
View Figure
Figure 23.15 Diagram of ultrasonic flow sensor.
P.739
View Figure
Figure 23.16 Variable orifice sensor. Gas flow causes the Milar flap to bend. There is a pressure drop across the flap. A transducer inside the ventilator converts the pressure drop into a flow.
Accuracy The sensor can measure f lows from 1 to 120 L/minute. There are no respiratory
rate l imi ts . The instrument wi l l read high with xenon and ni trous oxide (8).
Since both inspired and exhaled volumes are measured, the vent i lator can make
adjustments so that changes in f resh gas f low do not affect the del ivered volumes.
Since the sensors are loca ted at the absorber, they cannot compensate for gas
compression or expansion in the breath ing system. This is a small error un less very
compliant breathing tubes are used or the breathing system contains a large
volume of gas.
Evaluation The main advantage of this dev ice is that i t al lows the venti lator to automatica lly
compensate for changes in f resh gas f low. Disadvantages include the need for two
sensors and f il te rs . A break in one of the pressure l ines can cause a leak in the
breathing sys tem
P.740
(29). The sensor i tself may be the source of a leak (30,31,32,33). The sensors are
sensi tive to humidity (34,35).
View Figure
Figure 23.17 A: Variable orifice sensor. Tubings are attached on either side of the flap. B: Tubings from the sensors attached to the anesthesia machine. It is important that they are attached to the proper connection.
Fixed Orifice Flow Sensor This dev ice cons is ts of a restric tor and two pressure sensors, one on ei ther side of
the res tric to r. A zeroing valve compensates for pressure sensor drif t .
Respirometer Position in the Breathing System Figure 23.18 shows possib le locat ions for a respirometer in the c irc le sys tem. From
the standpoin t of accuracy, the most desirable locat ion is be tween the breathing
system and the patient (posit ion C). In this loca tion, read ings are not affected by
breathing sys tem leaks , expansion of breath ing system components, o r gas
compression. Both inspired and expired volumes can be measured. Placing the
sensor a t th is si te wil l increase the dead space, and water condensation may be a
problem. Th is posi tion may resul t in inc reased l ikelihood of damage, disconnect ion,
or tracheal tube k ink ing.
A disadvantage of posi tion C is that if a disconnection occurs between the sensor
and the breathing system during spontaneous venti lat ion, the spirometry
measurements wil l no t be af fec ted. I f the sensor is loca ted a t posit ion B , the
disconnection wi l l be detected by the change in volume.
A common pract ice is to locate the respirometer in the exhalation l imb upstream or
downstream of the unidirect ional valve (posi tions A and B). An advantage of these
posit ions is tha t if the respirometer can sense reverse f low, a malfunc tioning
unidi rect ional valve can be detec ted. If a d isconnection that prevents exhaled
gases from passing down the exhalation tubing occurs, the resp irometer wi l l not
sense a gas f low, and an apnea message and a larm wil l be ac tivated. A
respirometer in this locat ion wi l l usually read accura tely during spontaneous
respira tion, but during control led respirat ion, i t wil l usual ly g ive erroneously high
readings
P.741
(36,37). This is due to expansion of components of the breathing system and gas
compression. If a venti lato r with a hang ing bellows is used, a respirometer in this
posit ion may st il l indicate f low when a disconnection occurs (38).
View Figure
Figure 23.18 Possible sites for a respirometer in the circle system. (See text for details.) PEEP, positive end-expiratory pressure; APL, adjustable pressure limiting.
I f the resp irometer is located downstream of the absorber (posi tion E), the volume
of gas measured wi l l be dec reased by the amount of carbon dioxide absorbed in the
absorber.
Another possible loca tion fo r the respirometer is on the inspiratory s ide of the
system (pos it ion D). In this locat ion, the respirometer wil l disp lay erroneously high
readings , as gas that does not inf late the patient 's lungs wil l also pass th rough it .
During contro lled venti lation , a disconnect ion may not be detected.
I t is common to loca te pressure-f low sensors at bo th posi tions B and D. This allows
both the inspiratory and exhalation volumes and pressures to be measured. This
provides the info rmation to produce a f low-volume or pressure-volume loop.
Sensors in both these posi tions are used wi th the Datex-Ohmeda 7900 venti lato r to
compensate for changes in t idal volume due to fresh gas f low or leaks .
When one o f these devices is used in a Map leson system wi th an adul t pat ient, it
should be placed between the patient connect ion port and the patien t. In smal le r
patients, i t should be placed in the expiratory l imb to avoid an increase in dead
space (11,15).
Airway Pressure Monitoring Airway pressure moni tors (vent i lator or respiratory monitors or alarms; pressure
alarms; pressure alarm systems; anesthesia, pat ient, or breathing c i rcui t moni to rs;
venti lato r moni to ring alarms ; breathing gas inte rrupt ion moni tors; d isconnect
moni tors ; breathing pressure moni to rs) are used to warn of high- or low-pressure
condi tions in the breathing system (39,40).
High- o r low-pressure condi tions in the breath ing system have been a major cause
of anesthesia mortali ty and morb id ity. A device tha t responds to pressure changes
wi th in the breath ing system and provides warning of a problem is s trongly
recommended. Other parameters such as exhaled carbon dioxide and exha led
volumes may remain rela tively normal in the presence of dangerously abnormal
ai rway pressures.
Continuous airway pressure monitoring is now the norm in both the opera ting room
and in crit ical care areas . It is a s imple and noninvasive technique tha t helps in
assess ing the patient 's mechanical and spontaneous venti lation and determining
the presence of PEEP.
P.742
View Figure
Figure 23.19 Virtual pressure gauge on anesthesia machine display. This can be displayed on demand.
Equipment An airway pressure moni to r may be frees tanding or incorporated in to a venti la to r or
an anesthesia machine. Most of these devices are inexpensive, robust, easy to use,
and rel iable . They may be powered ei ther from the main electrica l system wi th
battery backup or by a batte ry wi th ba ttery test capabi li ty .
Most new anes thesia machines have a buil t- in da ta sc reen where information
including ai rway pressure is avai lable . Usually, there wi l l be an airway pressure
versus time waveform displayed on th is screen. A v i rtual electron ic pressure gauge
may be disp layed on the moni tor screen (F ig. 23.19). Many machines in use are
st i l l equipped wi th a mechanica l pressure gauge on the absorber. Th is does not
al low e lectronic recording or the data to be integra ted wi th o ther parameters to
al low compl iance calcula tions . Alarms were not assoc ia ted wi th these manometers,
so they had to be repeatedly scanned by the anesthesia provider.
Alarms Most pressure alarms have an aud io pause (delay, mute, s ilencing) control that wil l
delay the aud ib le s ignal fo r some or a ll functions. This pause should not prevent
the v isual s ignal f rom func tioning. Some units ' aud ib le alarms can be comple tely
turned OFF (41). Other features on some devices include automatic act ivation when
a pressure is detected, the abil ity to de tect impending battery fai lure , and
protect ion f rom accidental inactivation or power fai lure (42,43). Some airway
pressure monitors have the abi l i ty to display pressure waveforms .
Low Peak Inspiratory Pressure A low peak inspira tory pressure (minimum airway pressure, low airway pressure ,
venti lation failure, apnea, cycling, pressure fa ilure, disconnect, vent i lator
disconnect, minimum venti la tory, venti lat ion pressure, threshold pressure , low-
pressure , peak a irway, fai l -to-cycle, low pressure, low c i rcui t pressure) alarm is
ac tivated i f the pressure detected does not exceed a preset min imum within a f ixed
t ime.
Bas ic mon itoring s tandards adopted by the ASA and the American Association of
Nurse Anesthetis ts (AANA) state that when venti lation is controlled by a mechanical
venti lato r, there sha ll be a means of de tecting disconnection of breath ing system
components in cont inuous use. Such an alarm has been recommended by other
responsible bodies around the world (44). The low peak insp iratory pressure alarm
is one of the means to fu lf i l l th is requirement. However, the pressure moni tor is not
foolproof . Under certain c i rcumstances, i t may fail to de tect anesthetic ci rcui t
disconnections. Pressure moni toring wi l l fail to detect a disconnect ion that occurs
during spontaneous vent ilat ion (45).
The airway pressure mus t exceed a threshold value (l imi t) to prevent an alarm. If
this l imit is not reached over a period of time, usual ly around 15 seconds, the alarm
is act ivated. The thresho ld needs to be set at a value s l igh tly below the peak
ai rway pressure. This value varies, depending on the c linical s i tuat ion. The l imi t
may be se t manua lly or au tomatically around a value s light ly lower than the peak
pressure determined by several successive breaths. When the peak ai rway
pressure becomes higher or lower, the th reshold may be automatical ly al tered or
there may be a means to move the thresho ld c loser to the new peak pressure (Fig.
23.20). On some machines , there wi l l be a v isual informationa l signa l that the l imit
is set too low.
The low peak pressure ala rm is enabled when the venti lato r is turned ON. I t is
inact ive when the venti la tor is no t being used.
Condit ions that can cause a low peak pressure alarm inc lude a disconnec tion or
major leak in the breathing system; an obs truct ion upstream of the pressure sensor;
inadequate fresh gas f low (disconnection of the f resh gas l ine, an inte rna l mach ine
obstruct ion, o r loss or reduc tion of pipe line pressure); the bag/venti lator selec tor
P.743
valve in the bag pos it ion; a leaking tracheal tube cuf f ; ex tubat ion; a faulty, poorly
se t, o r unconnec ted vent ilator; fai lure of the gas or power supply to the venti lator; a
malfunct ioning scavenging system; increased compl iance; reduced resistance; and
a suction dev ice mistakenly placed wi th in the gas f low pathway (46 ,47). Low
pressure alarms are of l i tt le o r no use during spontaneous breath ing when the
pressure in the system does not rise and fall appreciably (45).
View Figure
Figure 23.20 The pressure (top waveform) exceeds the threshold (dotted line) by a small amount. At the bottom is a touch control marked “AUTO PRESSURE THRESHOLD.” If the threshold is too high or too low, it can be altered by using the control or by touching the threshold on the screen and moving it to the desired location.
The alarm threshold should be set jus t below the minimum peak pressure expec ted
during inspira tion (41,42,48,49). This peak pressure wi l l vary not only from patient
to patien t but also during a given case. Of ten, the th reshold is se t lower in an
attempt to prevent false-pos itive ala rms. If the alarm limi t is set too low, a fa lse
negat ive may occur (50 ,51,52). I t has been sugges ted that a pressure threshold of
less than 8 to 10 cm H2O is unacceptab le (53). On most modern de livery sys tems,
the ci rcu it p ressure waveform and the low-pressure alarm threshold can be
displayed, mak ing i t easy fo r the operator to ad just the threshold properly (Fig .
23.2). On some moni tors , an adv isory s ignal wi l l be act ivated if the thresho ld is set
a certain amount be low the peak pressure. Some units automatically set the
threshold based on the pressure sensed during prev ious breaths. There may be a
manual threshold reset contro l.
Problems wi th these moni tors have been reported. A disconnection or leak may not
be detec ted if the alarm is not swi tched ON (54) or the th reshold is se t too low. A
false-negat ive condi tion may occur if the end-exp iratory pressure is above the
threshold pressure. Other condit ions that may produce a pressure high enough to
exceed the th reshold when a disconnec tion occurs inc lude the breathing system
connec tor's obstruction by a pi l low, sheet, or surgical drape; a high-resistance
component such as a heat and moisture exchanger, capnometer cuvette, or
humidif ier; ai r entrainment into the breathing system (especial ly wi th a venti lato r
bellows descending during expirat ion); partial extubat ion; compression of an empty
venti lato r bel lows; and a Mapleson system wi th a high resistance (55,56,57,58,59).
I f a vent i lator that uses a ram of oxygen to produce inspira tion is used wi th a T-
piece system, a disconnec tion at the common gas out let may not be detected due
to the high resistance of the f resh gas tubing (60). Those devices opera ting on
batteries wi l l not a larm if the batte ries fail . I t is essential tha t the a larm be checked
before use by mak ing a disconnection at the patient connector whi le the venti lator
is cycling (Chapter 33) (50). Unfortunately, s tudies show tha t this test is not
performed rout inely or correct ly (48). It is importan t tha t another means of
detec ting a disconnec tion (such as a capnograph or volume or f low monitor) be
used.
Sustained Elevated Pressure A sus tained (continuous, continuing) pressure monitor ac tivates an alarm if the
pressure does not fall be low a certain level during part of the respiratory cyc le.
Most are always enabled. Some incorporate a valve tha t opens to relieve the
pressure af ter a certain t ime.
Severa l mechanisms can produce a sus tained e levated pressure: accidental
ac tivat ion of the oxygen flush valve; occlusion or obstruc tion of the expiratory l imb;
an improperly ad justed adjustable pressure l imi ting (APL) valve ; occlusion of the
scavenging sys tem; a malfunct ioning venti la to r; or a malfunctioning or incorrect ly
se t PEEP valve (42,43,55).
High Pressure A high-pressure alarm is act ivated if the pressure exceeds a certain limit . On some
devices, the threshold
P.744
is fixed (usua lly 50 to 80 cm H2O); on o thers, it is adjus table (41). Some
instruments automatica lly set the alarm thresho ld at a set amount above the
average peak pressure for several p rev ious breaths. Mos t of these a la rms are
always enabled. There shou ld be no delay on the high-pressure a larm. Some
anesthesia delivery sys tems are f i tted wi th pressure-l imi t ing valves that vent gas
f rom the breath ing system when a high pressure is detected (41).
Possible causes of high pressure include airway obs truction, reduced compl iance,
increased res is tance, oxygen flush ac tivat ion during the inspiratory phase, a
punctured venti lator be llows , occlusion or obs truct ion of the expira tory l imb of the
breathing sys tem, scavenger malfunct ion , or the pat ient coughing or s training
(42,61). Even in the presence of complete obstruct ion , th is alarm wi l l not be
ac tivated i f the peak inspiratory pressure does not reach the set l imi t (49). High
compliance, low res is tance, leaks, low inspira tory flow rates, high respira tory ra tes ,
low I:E rat ios, low tidal volumes, and low f resh gas f lows can al l dec rease the peak
inspiratory pressure so that there is no ala rm cond it ion (2,62). During pressure
control venti lat ion, the inspira tory ai rway pressure is preset and thus cannot act as
a warning of tracheal tube occlusion (63).
Subambient Pressure A subambient (subatmospheric) pressure alarm is act ivated by a pressure that fal ls
below atmospheric pressure by a predetermined amount. Subatmospheric pressure
can be generated by a pat ient attempting to inhale against a collapsed reservoir
bag or increased resistance; a blocked inspiratory l imb (during the venti lator's
expiratory phase); a malfunct ioning active closed scavenging sys tem; suct ion
applied to a nasogastric tube placed in the tracheobronchial tree or to the work ing
channe l of an endoscope passed into the airway; a sidestream gas analyzer; or the
ref i l l ing of a hanging bel lows venti la tor bel lows (14,42,49,64,65,66).
Monitoring Site The location where pressure is sensed wil l af fect i ts usefulness. Figure 23.21
shows possible s i tes. Idea lly, the s i te should be c lose to the patient 's airway
(posi tion C). Many disposab le breathing systems have a smal l port a t the Y-piece
that can serve as the connect ion s ite fo r tub ing that transmits the pressure to a
moni toring device (57). Pressures during both inspirat ion and exha la tion can be
measured at this s ite. The D-Li te and Novometrics sensors discussed earlier in this
chapter
P.745
are placed at this s i te . Pressure- and f low-volume loops can a lso be generated from
pressures sensed at this s ite. While p lacement between the pat ient and the
breathing sys tem is best f rom this standpoin t, in practice i t may present problems
wi th dead space, disconnect ions, tracheal tube kinking, and water bui ldup in the
pi lot l ine . The l ines must be connected for every case.
View Figure
Figure 23.21 Possible sites for monitoring airway pressure in the circle system. (See text for details.) PEEP, positive end-expiratory pressure; APL, adjustable pressure limiting.
The more distant the measurement s i te is from the patien t, the less usefu l it is as
an es timate of ai rway pressure (2,67). Breathing sys tem resistance and
compliance, leaks, obstructions , and other mechan ical factors may cause the
measured pressure to be qu ite different f rom the pressure in the patient 's ai rway
(56).
Frequently, the monitoring s ite is in the breathing system (posit ions A, B, and D).
An occlusion in the breathing system wil l cause a low-pressure state distal to the
obstruct ion and a high-pressure state proximal to i t, so certain types of problems
may be missed (57). I f PEEP is used, i t wi l l not be indicated on a pressure moni tor
located at posi tion B. Posi tions A and D are frequently used to moni tor pressure
during inspira tion and exhalat ion. These locat ions may be used to provide
pressures for p ressure-volume loops.
In the past, the sensor was sometimes located in the venti lato r (posi tion E). This is
unsatis fac tory because under certain c i rcumstances, suff ic ient back pressure to
inhibi t the minimum pressure alarm may be genera ted a t the bellows even when
there is a disconnec tion (55,56). Placing the sensing point in the vent i lator may
also resul t in fai lure to detect an incorrect ly set bag/vent i lator selector valve .
Spirometry Loops
A loop is a graphic representat ion of the dynamic relationship between two
variables (p ressure and volume or f low and volume) during both insp iration and
exhalation (6,23,68). The f low, volume, and pressure curves i l lustrated in Figure
23.1 are the bases of spi rometry loops.
Pressure- and f low-volume loops are available on certain physiologic moni tors as
an option. Later-generat ion anes thesia machines and most physiologic moni tors
now offer these loops , usually as an option. The authors bel ieve that the
info rmation provided just if ies the extra cos t.
Illustrative Loops The Pressure-volume Loop The pressure-volume (compliance) loop shows volume on the vertica l ax is and
ai rway pressure on the horizonta l ax is (F ig . 23.22). With controlled vent i lat ion, the
pressure in the breathing system increases during inspirat ion. At the same time, the
inspired volume of gas inc reases . The t idal volume is the poin t on the vertical ax is
that corresponds to the highest poin t on the loop. The peak pressure is the highest
value on the horizontal axis. The shape of the inspiratory phase is determined by
the type of respira tion being monitored.
View Figure
Figure 23.22 Pressure-volume loop. The pressure-volume relationship reflects pulmonary and tracheal tube mechanics. During controlled ventilation, a line drawn from the zero point through the point of end inspiration represents the compliance, which is determined by dividing the tidal volume by the pressure at end inspiration. With good compliance, that line forms an angle of 45 degrees or less with the volume scale. A loop that becomes more horizontal indicates a decrease in compliance.
A line drawn from the zero po in t th rough the po in t of end inspirat ion during
control led vent ilation (Fig. 23.22) represents the compl iance. With good
compliance, that l ine fo rms an angle of 45 degrees or less with the volume scale. A
loop that becomes more horizontal indica tes a decrease in compliance.
The port ion of the loop represent ing exhalation starts at the point of highest volume
and moves downward toward zero. The area ins ide the loop is related to the work of
breathing (1 ,69).
The Flow-volume Loop The f low-v olume (resistance) loop (Fig. 23.23) has volume on the horizontal axis
and f low on the vert ical ax is . The zero point for volume is to the righ t on the
horizontal axis, corresponding to func tional residual capacity. During inspiration ,
f low rate increases (plotted downward). The inspiratory f low drops to zero as
inspiration ends. The tidal volume is reached at the point where f low returns to zero
and the loop crosses the horizonta l ax is. The shape of this part of the loop depends
on the mechanism of respirat ion (e.g., vo lume controlled, pressure controlled,
manual , or spontaneous).
P.746
View Figure
Figure 23.23 Flow-volume loops with controlled ventilation.
Exhalation is represented by the part of the loop above the horizontal ax is. The
shape of this portion of the loop is determined by the rate of passive lung def lat ion,
which is in turn de termined by elastic recoil of the lung and chest wal l and by the
total flow resistance offered by the bronchial tree , ai rway device tube, expiratory
l imb of the breathing sys tem, and any addi tiona l equ ipment. With a normal loop, the
f low rate during exhalation increases rapidly at the beginning, quickly reaches a
peak, then slows and gradual ly returns to zero.
Figure 23.24 shows another way of i l lustrating a f low-volume loop that is used by
some manufacturers. The zero point is at the junction of the horizontal wi th the
vertical ax is . Inhalation is above the horizontal axis, and exha lat ion is below. Flow-
volume loops i l lustrated in this chapter employ the representation used by
pulmonologists . This loop may be presented in other ways, bu t the basic loop is the
same.
View Figure
Figure 23.24 Alternative method of displaying flow-volume loops. (See text for details.)
Representative Normal Loops Loops that are i l lustrated in this chapter do not come from actual patien ts bu t are
styl ized to i l lus trate certain aspects of respira tory mechanics. They are based on
ac tual loops as much as possible. The reader should not expect to see an exact
reproduc tion of these loops when moni toring a pat ient. Clinica l condit ions and
venti lato r function are rarely stra ightforward and usually inc lude a number of
factors . The loop represents a composi te of mechanical and phys io log ic factors. As
famil iari ty with the loops inc reases , the user can progress to the finer details of
inte rpretat ion .
Volume-controlled Ventilation with No Inspiratory Pause
Pressure-volume Loop The pressure-volume loop for volume-contro lled venti lation wi th no insp iratory
pause is shown in Figure 23.22. It begins at zero volume and near or at zero
pressure . During insp irat ion, both pressure and volume inc rease, so the loop moves
up and to the right. At the end of inspirat ion, both peak pressure and t ida l volume
are attained. The end of inspirat ion represents a fall in pressure that occurs before
exhalation can begin. The loop returns downward and toward the lef t to i ts original
s tarting posit ion.
Flow-volume Loop The f low-v olume loop seen wi th volume-control led vent i la tion wi th a constant f low
generator is shown in Figure 23.23. The flow quick ly r ises to a level tha t is
constant, producing a f la t inspira tory portion . At the end of insp iration, the f low
drops rapidly to zero. The tidal volume is reached as the loop crosses the volume
l ine.
As exhalation begins, there is a rapid ascent to a peak. Flow then dec reases , and
the loop falls smoothly toward zero f low and volume. The angle a t the top of the
loop is narrow.
Controlled Ventilation with Positive End-expiratory
Pressure
Pressure-volume Loop The addi tion of PEEP causes the starting po int of the pressure-volume loop to shif t
to the righ t by the amount of PEEP that is appl ied (F ig. 23.25). PEEP may cause an
increase in compliance and is represented by the loop becoming more vert ical wi th
a dec rease in peak pressure. If the loop tends more to the right, PEEP may not be
benef ic ial and may need to be wi thdrawn.
The loop disp lay makes i t easy to detect inadvertent PEEP. This may be due to the
PEEP valve having been inadvertently turned ON, a partial obstruc tion in the
breathing sys tem, a malfunc tioning exhalation un id irectional valve, fai lu re of the
posit ive-pressure re lief in the scavenging inte rface, or an incorrect ly set APL valve.
P.747
View Figure
Figure 23.25 With PEEP, the loop is shifted to the right. PEEP, positive end-expiratory pressure.
Flow-volume Loop The f low-v olume loop wi th PEEP during control led vent i la tion wi th a constant f low
venti lato r is shown in F igure 23.26. PEEP wil l decrease the exp iratory driv ing
pressure , producing lower f lows during exha la tion so that the loop appears f latter
(4).
Controlled Ventilation with an Inspiratory Pause
Pressure-volume Loop Figure 23.27 shows the pressure-volume loop during control led vent i lation with an
inspiratory pause. After the peak pressure is reached, the venti la to r pauses fo r a
short t ime wi th the lungs inf la ted. During this pause, the pressure in the breathing
system drops to a pla teau level , usually 2 to 5 cm H2O lower than peak pressure.
Fresh gas entering the breathing system from the anesthesia mach ine wi thout fresh
gas decoupl ing (Chapter 12) wi l l inc rease the inspired volume. During exhalat ion,
the pressure and volume wi l l drop in the expected manner.
Flow-volume Loop The f low-v olume loop seen wi th an insp iratory pause is shown in Figure 23.28.
There is a drop in f low near the end of inspira tion wi th a small increase in t ida l
volume during the pause resu lt ing from f resh gas f lowing into the breathing system
i f there is no fresh gas decoupl ing. This pattern should not be confused wi th a
spontaneous breath during control led vent i lat ion (Fig. 23.60). The increase in
volume due to fresh gas f low is s traight, f rom right to lef t , and s tays near the zero
f low l ine. Exhalation is s imi lar to the loop wi thout an inspiratory pause.
View Figure
Figure 23.26 PEEP produces a decrease in expiratory flow. PEEP, positive end-expiratory pressure.
P.748
View Figure
Figure 23.27 During an inspiratory pause, it is common for the airway pressure to decline 2 to 5 cm H2O. The lower pressure is called the plateau pressure. If the ventilator does not block fresh gas flow during inspiration (fresh gas decoupling) there will be an increase in tidal volume during the inspiratory pause.
Pressure-controlled Ventilation
Pressure-control led vent ilat ion differs f rom volume-control led respirat ion in that the
inspiratory f low is not constant. Pressure-control led vent i lation is d iscussed in
Chapter 12.
Pressure-volume Loop The pressure-volume loop is shown in Figure 23.29. The loop is wider than that
seen with volume-controlled venti lat ion and starts off wi th a greater pressure rise
than volume increase. As inspiration proceeds, volume rises faster than wi th
volume-control led vent i lat ion.
Flow-volume Loop The pressure-contro lled mode of vent i lat ion has an accelerating-decelera ting
inspiratory f low prof i le in contras t to the constant insp iratory f low seen wi th volume-
control led vent ilation (Fig. 23.30). The exhala tion part of the loop is s imi lar to that
wi th vo lume-control led vent i la tion.
View Figure
Figure 23.28 The blip near the end of inspiration represents the increase in tidal volume during the inspiratory pause due to fresh gas continuing to flow into the breathing system. This will not be seen if the ventilator has fresh gas decoupling.
P.749
View Figure
Figure 23.29 Pressure-volume loop with pressure-controlled ventilation. Pressure rises rapidly to the set pressure during inspiration.
Spontaneous Respiration without Positive End-expiratory
Pressure
Pressure-volume Loop With spontaneous respirat ion and no PEEP, the pressure-volume loop (Fig. 23 .31)
starts ou t at zero pressure and volume. During inspira tion ai rway, pressure is
negat ive, so the loop moves in a c lockwise di rection. At the end of inspirat ion, the
pressure returns to zero. At this point, the loop crosses the t idal volume point on
the ord inate . During exhalat ion, ai rway pressure becomes posit ive, and the loop
moves to the right. At the same t ime, the volume drops. At the end of exhala tion,
the pressure and volume return to zero. The shape of the loop is st il l double
convex, but i ts s lope is d if fe rent from tha t seen wi th contro lled venti lation.
Compl iance cannot be calcu lated f rom this loop because the insp iratory pressure is
negat ive. The area of the loop represents the work of breath ing.
Flow-volume Loop The f low-v olume loop wi th spontaneous resp iration is shown in Figure 23.32. The
f low rate during inspiration varies more than wi th mechan ica l venti lation. Peak f low
occurs near the middle of inspira tion. At the end of inspiration , f low becomes zero,
and the loop crosses the horizonta l line at a volume correspond ing to the t idal
volume. The f low during exhala tion is s imi lar to that found in other forms of
respira tion.
View Figure
Figure 23.30 Flow-volume loop with pressure-controlled ventilation. Flow is rapid at the beginning of inspiration, then decreases.
P.750
View Figure
Figure 23.31 With spontaneous ventilation, the shape of the loop is double convex, but the slope is different from that seen with controlled ventilation. PEEP, positive end-expiratory pressure.
Spontaneous Respiration with Positive End-expiratory
Pressure
Pressure-volume Loop
I f PEEP is appl ied during spontaneous venti la tion, the pressure-volume loop wi l l
s tart out a t the PEEP value and move to the lef t (F ig. 23.33). Inspira tion cannot
begin un ti l the pressure has become negative. At this point, the t idal volume
increases rap id ly. During exhalat ion, pressure increases rap id ly, and the loop
moves toward the right and down ward to the po in t of origin . The loop has a
rectangu lar shape. The larger internal area of the loop indicates the inc reased work
of b reathing .
Flow-volume Loop The corresponding f low-vo lume loop during spontaneous respiration with PEEP is
shown in Figure 23.34. Both the inspira tory and exhalation port ions of the loop are
f lat tened. The exhalation port ion is more rounded than when PEEP is not p resent.
This conf igurat ion is s imilar to the loop demons trat ing a f ixed inspiratory and
expiratory obstruction (Fig . 23.52).
Face Mask Positive-pressure Ventilation
Pressure-volume Loop The basic shape of the pressure-volume loop is s ti l l doub le convex (Fig. 23.35).
The insp iratory port ion of the loop is more rounded. If there is a signif icant leak
around the mask, an open loop may be seen.
Flow-volume Loop The f low-v olume loop wi th mask venti lation (Fig . 23.36) is more rounded during
both inspira tion and exhalat ion than that seen wi th intubat ion. Th is can vary with
the way in wh ich the anesthesia provider squeezes the bag.
Intermittent Mandatory Ventilation Inte rmittent mandatory vent i lat ion produces a combination of loops representing
both spontaneous and control led breaths .
View Figure
Figure 23.32 With spontaneous ventilation, the flow rate during inspiration varies more than with mechanical ventilation. Inspiration and exhalation tend to mirror each other. Tidal volume during spontaneous ventilation is usually lower than with controlled ventilation.
P.751
View Figure
Figure 23.33 With spontaneous ventilation and positive end-expiratory pressure, the loop is shifted leftward and becomes rectangular. PEEP, positive end-expiratory pressure.
Pressure-volume Loop The pressure-volume loop (Fig. 23 .37) shows both a spontaneous breath (solid l ine)
and a control led breath (dashed line). Mos t moni to rs wi l l display these loops
consecutively and not on the same sc reen as il lustrated un less one of the loops has
been previously saved.
Flow-volume Loop The f low-v olume loop (Fig. 23 .38) shows a spontaneous breath (solid l ine) and a
control led breath (dashed line). Each has the characteris tics of the normal loop for
this type of respiration.
Patient-triggered Ventilation I f the venti lator is in a tr iggering mode (pressure support vent ilat ion), a
spontaneous breath wi l l be necessary to in itia te a pos it ive-pressure respirat ion.
The pressure-volume loop wil l s tart out negat ively, but as the vent i la tor is engaged,
the loop becomes rap idly posi tive fo r
P.752
the duration of inspiration (Fig. 23.39). The exhala tion port ion of the loop wi l l be
s imilar to that found in controlled venti lation . The f low-vo lume loop associated wi th
pressure support vent i lat ion is shown in Figure 23.40.
View Figure
Figure 23.34 PEEP during spontaneous respiration results in lower flows during both inspiration and exhalation. PEEP, positive end-expiratory pressure.
View Figure
Figure 23.35 With mask ventilation, the pressure rises more slowly during inspiration. During expiration, the absence of the tracheal tube decreases resistance to flow and volume and pressure drops rapidly. The shape of the upstroke will vary.
Spontaneous-assisted Ventilation I f the spontaneous ly breathing pat ient is not p roducing a satis fac tory t idal volume,
respira tion is often manually assisted. Figure 23.41 shows one of the conf igurat ions
of the pressure-volume loop that may be seen. As the patien t init ia tes the
venti lation , there wi l l f i rs t be a negative pressure . As the bag is squeezed, p ressure
and t idal volume inc rease.
Loops Representative of Patient Factors Variations f rom a normal loop can be caused by anyth ing that affects the way gas
moves past the sensor during inspirat ion or exp iration . This may be a pat ient
factor, a c ircui t variable, o r a venti lator variable.
Changes in Compliance
Pressure-volume Loop A major advantage of p ressure-volume loops is the ir ab il i ty to detect changes in
compliance. If the lungs or chest wa ll become s ti ffer, increased pressure wil l be
necessary to del iver the same t ida l volume. This causes the pressure-volume loop
to be
P.753
displaced c lockwise (Fig. 23 .42). I f PEEP is introduced, there may or may not be an
increase in compliance (Fig. 23 .43). I f PEEP does not increase compl iance or
makes the s ituat ion worse, this can be determined by subsequent loops .
View Figure
Figure 23.36 Mask ventilation. The inspiratory flow is more variable when ventilation is manually controlled than when a ventilator is used. During exhalation, the lower resistance due to the absence of a tracheal tube results in higher flow.
View Figure
Figure 23.37 Pressure-volume loop with intermittent mandatory ventilation.
Flow-volume Loop Decreases in compl iance wi l l affec t the f low-v olume loop (Fig. 23 .44). F low wi l l be
increased during exhalat ion, wi th a higher peak and a s teeper s lope.
Decreases in compl iance can result f rom inadequate muscle relaxation; a ir
embolism; d iseases and tumors that invade la rge areas of the lung or al ter i ts
dis tens ibi l i ty; narcotics ; bronchial in tubat ion ; bronchoconstric tion ; pneumothorax;
reduc tion pneumoplasty; lateral decubi tus, l i tho tomy, or Trendelenburg posit ions;
ex ternal pressure on the ches t or abdomen; abdomina l re tractors or packing;
abdominal enlargement; curvature of the spine; obesi ty; prone posi tion;
pressuriza tion in the peri toneal cavi ty during laparoscopic surgery; or adul t
respira tory distress syndrome (ARDS) (6,68,70,71,72,73,74,75,76,77,78,79,80).
Dec reases in compl iance can be found during part ial coronary bypass. The loop
returns to the prebypass s ta te af te r the bypass is discont inued. Compliance is
lower in chi ldren than in adul ts (Fig. 23 .45) (80).
Fac tors that increase compl iance include PEEP, emphysema, and resolution of the
factors that decrease compl iance. Since changes in compliance often occur
gradually, they may not be recognized unless the change is la rge . It is useful,
therefore, to store a loop f rom the beginning of a case for comparison.
View Figure
Figure 23.38 Flow-volume loop with intermittent mandatory ventilation.
P.754
View Figure
Figure 23.39 Patient-triggered ventilation. As the patient takes a spontaneous breath, the loop becomes positive for the duration of inspiration.
Changes in Resistance An increase in res is tance may be caused by tracheal tube obs truct ion (kinking ,
dislodgment, or secre tions ), b ronchoconstric t ion, airway col lapse f rom loss of
elast ic recoil or by obstruc tion in a large airway caused by secretions, blood,
fore ign body, neoplasm, inf lammation, or using a trachea l tube that is too small .
Whi le mi ld bronchospasm causes on ly s l ight changes in the f low-volume loop, as it
increases there wi l l be changes in both the inspiratory and exhalat ion portions .
With severe expiratory res is tance, expira tory f low may stop abruptly before the next
mechanical inf lat ion . The effects of treatment for b ronchospasm can be assessed
by observ ing the loop after trea tment.
Pressure-volume Loop During contro lled venti lation , increased resistance means that higher inspira tory
pressures wil l be required to del iver a given f low. Tidal volume may be reduced. As
shown in Figure 23.46 (sol id l ine), the pressure-volume loop is shif ted to the right
and downward with a large internal area. The pressure falls rapid ly af ter inspira tion
is complete. The loop may be open if there is air trapping . With spontaneous
venti lation , the inspiratory l imb is displaced leftward (Fig. 23.47).
View Figure
Figure 23.40 Pressure support ventilation.
P.755
View Figure
Figure 23.41 Spontaneous-assisted ventilation. As the patient begins to inspire, a negative pressure is seen. Then the bag is squeezed and the pressure becomes positive. The shape of the inspiratory portion will depend on how the bag is squeezed.
Flow-volume Loop I f res is tance is inc reased, the f low-v olume loop wi l l show decreased f low
throughout exhalation (4,21) (Fig . 23.48). As resistance increases fu rther (Fig.
23.49), there wil l be changes in bo th the inspiratory and exhalation port ions, and
the tidal volume may be dec reased. With severe exp iratory resistance, expiratory
f low may stop abruptly before the nex t mechanical inf lation.
Chronic Obstructive Lung Disease Emphysema is charac terized by a progressive loss of elastic t issue in the lung.
These pat ients have no problem with in flating the lungs but must work to exha le .
During mechan ical venti lat ion , pat ients with airf low obstruction may develop
inadvertent PEEP (auto-PEEP, occult or in trinsic PEEP, dynamic hyperinf lat ion, a ir
t rapping) if there is not enough t ime for complete
P.756
exhalation (81,82,83,84,85). The exhaled vo lume wi l l be less than the inspired
volume.
View Figure
Figure 23.42 Low compliance causes the loop to be moved closer to the horizontal axis. High compliance causes the loop to move closer to the vertical axis. The dotted line shows normal compliance. The solid line shows decreased compliance.
View Figure
Figure 23.43 The dotted line represents decreased compliance. With the addition of PEEP, the loop is moved to the left, and the increased compliance results in a more normal-looking loop. If the loop does not improve with PEEP, the PEEP may not be beneficial and may need to be removed. PEEP, positive end-expiratory pressure.
Pressure-volume Loop The pressure-volume loop seen wi th this condit ion is shown in Figure 23.50. A t the
beginning of inspiration, the pressure rises s lowly.
During exha lat ion, the pressure drops wi th l i tt le change in volume unt i l the end of
exhalation . An open loop may be seen.
Flow-volume Loop The corresponding f low-vo lume loop is shown in Figure 23.51. During expiration ,
there is a severe reduc tion in f low. The loop may be open if the patien t does not
have suff icient time to exhale completely. In terrupted expiratory f low may sugges t
the presence of intr insic PEEP (auto-PEEP).
Pat ients with obstructive a irway disease may not complete a ful l exhalation prio r to
the start of the next inhala tion, result ing in pers istent posi t ive pressure. Th is wi l l be
indicated by the absence of a period of zero f low before the next inhalation (Fig .
23.63).
Airway Obstruction Flow-volume loops may be he lpful in identi fying ai rway obs tructions (86,87,88). The
inspiratory l imb of the loop
P.757
is useful in diagnosing extra thorac ic ai rway obs truct ion, and the expiratory l imb is
sensi tive to intra thoracic obstruc tion (87,89). When the c ross-sectional area of the
airway is decreased to a c ri tical level , characteris tic patterns of f low occur with
spontaneous venti la tion. Typical ly, the flow ra te wi l l pla teau. The value of the flow
rate at this plateau wi l l depend on the cross-sectional a rea of the f low-l imi t ing
segment in the airway.
View Figure
Figure 23.44 The dotted line represents decreased compliance. Flow is greater at the beginning of exhalation due to the increased pressure.
View Figure
Figure 23.45 Pressure-volume loop in pediatric patient.
With a f ixed intrathorac ic or thorac ic obstruct ion and spontaneous venti la tion (F ig .
23.52), bo th the inspira tory and expiratory l imbs of the f low-volume curve are
f lat tened (86,87,90).
A variable extrathoracic obs truct ion (Fig. 23.53) wi l l af fect inspirat ion as the
negat ive pressure causes the obstruction to inc rease. During exhalation, posi t ive
pressure in the a irway wi l l keep the trachea open at the si te of the les ion, leaving
the expiratory curve unaffec ted (87).
A variable intra thoracic obstruc tion (F ig. 23.54) wi l l show a normal inspiratory
curve as the negat ive intrathorac ic pressure wi l l keep the ai rway open. During
expirat ion, the in trathoracic pressure becomes pos itive
P.758
and thus decreases the ai rway diameter so that the exp iratory f low is reduced.
View Figure
Figure 23.46 With an increase in resistance, a higher pressure is needed to deliver the same volume (solid curve). Tidal volume may be reduced.
View Figure
Figure 23.47 Spontaneous respiration with increased resistance. The normal loop is shown with dotted lines. With increased resistance, greater pressure (more negative during inspiration, more positive during exhalation) will be needed to move the same volume of gas.
Restrictive Disease The increase in elastic recoil with restric tive defects increases the force driv ing
expiratory f low. Thus, the flow-volume loop usual ly shows a high expiratory f low
associated wi th a steep descending l imb (Fig. 23.55).
Secretions Secret ions in the tracheal tube wil l cause a sawtooth pattern in the pressure- and
f low-vo lume loops (83,91) (Figs. 23.56, 23.57).
Pediatric Patients Pediatric pa tients require a different scale for p ressure- and f low-v olume loops.
Smal l pa tients require re la tively high ai rway pressures because of the small
diameter of the trachea l tube. An example is found in Figure 23.45.
Spontaneous Breathing during Controlled Ventilation I t is possible for a nonparalyzed pat ient to b reathe spontaneously during controlled
venti lation . This can occur at any time during the resp iratory cycle. The
spontaneous breath usual ly has a lower inspiratory f low than the mechanical
breath.
View Figure
Figure 23.48 Flow-volume loop with increased resistance. The dotted line represents the curve with normal resistance. With increased resistance, there is diminished expiratory flow. The convex configuration of the expiratory limb reflects uneven lung emptying.
P.759
View Figure
Figure 23.49 With a severe increase in resistance, the ventilator cannot fully compensate, and inspiratory flow will be diminished. Tidal volume may be decreased. Expiratory flow is also severely decreased. The dotted line represents the normal curve.
Pressure-volume Loop Figure 23.58 shows a pressure-volume loop with a spontaneous breath during
exhalation . As the spontaneous breath occurs , the pressure drops below the
expected level whi le the volume rises above the usual curve. As the spontaneous
breath is exhaled, the pressure increases brie fly and the volume drops rapidly. The
remainder of the loop follows the expec ted shape.
Figure 23.59 shows a series of pressure-volume loops wi th the sol id l ine (loop 1)
representing a normal loop produced wi th mechanical vent i lation. Loops 2 and 3
show the patient b reathing against the venti lato r. The loop moves toward the
negat ive s ide of the pressure axis and then changes to posi t ive. The pressures
generated are qui te h igh and the t idal volume is dec reased s ince the patient is
exhal ing during insp iration .
Flow-volume Loop Figure 23.60 dep icts a f low-volume loop wi th a spontaneous breath near the end of
inspiration . There is a sudden increase in inspira tory flow and volume. As the
breath is exhaled, the f low and volume move toward zero .
View Figure
Figure 23.50 With severe COPD, resistance during expiration is greatly increased. The patient may have difficulty exhaling completely before the next inspiration, producing an open loop. COPD, chronic obstructive pulmonary disease.
P.760
View Figure
Figure 23.51 With severe COPD, expiratory flow is severely reduced. COPD, chronic obstructive pulmonary disease.
View Figure
Figure 23.52 Flow-volume loop with fixed intra- or extrathoracic obstruction.
View Figure
Figure 23.53 A variable obstruction located outside the thorax will cause a plateau during inspiration. The expiratory portion of the curve is close to normal. The dotted line shows a normal loop.
P.761
View Figure
Figure 23.54 With a variable intrathoracic obstruction (such as a tumor in the trachea or a mediastinal mass), inspiratory flow may be relatively normal, but during expiration, flow rises to a plateau instead of the usual rise to and descent from peak flow. The dotted line shows a normal loop.
View Figure
Figure 23.55 With a restrictive defect, the increase in elastic recoil is associated with higher expiratory flows. As the process becomes more severe and lung volumes are decreased, the flow-volume curve becomes tall and narrow. The dotted line shows a normal loop.
View Figure
Figure 23.56 Pressure-volume loop with secretions in the tracheal tube.
P.762
View Figure
Figure 23.57 Flow-volume loop with secretions in the tracheal tube.
Open Loop A loop should return to i ts s ta rt ing point at the end of the resp iratory cycle. An open
loop (F ig . 23.61) has a gap between the end and s tart ing points , indicat ing that the
exhaled volume is less than the inhaled volume. Whi le the pressure-volume loop
appears to c lose, i t actual ly returns to zero pressure along the vert ical ax is rather
than at the starting point. The amount of gas loss can be read between zero and
the return point of the loop.
An open loop means that more gas has passed the sensor during inspirat ion than
returns during exhalation . Most often, this is because of a leak distal to the sensor.
Incorrect ca libration should also be considered. An open loop is of ten seen wi th
mask anesthesia , an uncuffed tracheal tube, or a supraglottic ai rway device. A
double-lumen tube may impose suff ic ient res istance to resul t in the lungs not
comple tely emptying, even in patients wi th normal ai rways (22 ,92).
Leaks of ten occur after lung reduc tion or o ther thoracic surgery (93). The extent of
the leak and changes in the volume of gas los t can be tracked wi th the f low-volume
loop.
An open loop may be due to incomple te exhalat ion caused by chronic obstruct ive
pulmonary disease (COPD), inc reased resistance caused by apparatus, a tension
pneumothorax, lung retrac tion, or a flap-valve obstruction in a large ai rway or a
double-lumen tube (94).
Overshoot Loop An overshoot loop ind ica tes that the exhaled volume is greater than the inspired
volume. This usually indicates
P.763
that the moni to r needs to be recal ibra ted . It can also resul t if pressure on the
thorax causes some of the func tiona l residual volume to be added to the tida l
volume.
View Figure
Figure 23.58 This spontaneous breath occurs during expiration.
View Figure
Figure 23.59 Loop 1 represents the normal loop. Loop 2 shows a spontaneous breath during inspiration. The pressure drops, and the volume increases briefly. There is a decrease in compliance caused by an increase in tension in the chest wall muscles. In loop 3, the patient inhales at the beginning of the respiratory cycle, so the loop moves to the left of the vertical axis. There is a further decrease in compliance.
Pressure-volume Loop The pressure-volume overshoot loop (Fig . 23.62) can appear as a normal loop, but
the beginning poin t is to the right of the endpoint. The dif ference between the
beginning and endpoint is the amount o f excess gas that is exhaled.
Flow-volume Loop On the overshoot f low-vo lume loop (Fig . 23.63), the endpoin t is to the right of the
zero volume po int. The d is tance to the right represents the excess gas that has
been exha led.
Intrinsic Positive End-expiratory Pressure and Air Trapping Intr insic (au to, occu lt) PEEP (PEEPi) resul ts f rom a difference between the ac tual
expiratory t ime and the
P.764
expiratory t ime required fo r complete exhala tion of the tida l volume so that some
air is trapped in the lungs (Fig. 23.63). It may be generated by a very short
expiratory t ime and/or s low exp iration due to high resistance or abnormal ly high
compliance. Air trapping is l ikely to occur in patien ts wi th ai rf low limitat ion, inverse
ratio venti la tion, or when us ing a h igh respiratory rate .
View Figure
Figure 23.60 The spontaneous breath occurs near the end of inspiration. The thin, dotted line represents the normal loop. Instead of returning to zero at the end of inspiration, the flow increases. There is a small increase in volume as well.
View Figure
Figure 23.61 There is a leak, so exhaled volume is approximately 150 mL less than the inhaled volume. This produces open pressure-volume and flow-volume loops.
Loops Representing Equipment Problems Tubing Misconnection Some older sensors had a long and a short nipple des igned to connec t to the short
and long lumens of the pressure tubing, respectively. It was possib le to connect the
tubings long to long and short to short. If this were done, the moni tor would sense
exhalation as inha la tion and v ice versa. The loops would be drawn backward and
upside down (F ig. 23.64). It would take two respira tory cycles to produce a full
loop, and the moni tor would be unable to compute the compliance. Extremely high
PEEP values would be recorded. Newer sensors wi th male and female connections
make tubing misconnec tions less l ikely (Fig. 23 .8).
Disconnection between the Sensor and the Breathing
System I f there is a d isconnection between the spirometry sensor and the breathing
system, there wi l l be no f low
P.765
through the sensor during mechanical vent ilat ion and a loop wil l no t be generated
(Fig . 23.65).
View Figure
Figure 23.62 The exhaled volume exceeds the inhaled volume, producing an overshoot loop.
View Figure
Figure 23.63 Intrinsic positive end-expiratory pressure and air trapping. The gap in the flow-volume loop indicates that there was still expiratory flow when the next inspiration commenced. PEEP, positive end-expiratory pressure.
I f the patien t breathes spontaneously with a d isconnection between the sensor and
the breathing system, loops seen wi th spontaneous respira tion wi l l be generated. I f
a c i rc le breathing sys tem also conta ins a f low sensor mounted on the exhala tion
s ide near the absorber, this sensor wi l l not indicate f low during spontaneous
venti lation and wil l help to localize the disconnec tion.
Leak between the Sensor and the Breathing System
Pressure-volume Loop
I f there is a leak or partial d isconnection between the sensor and the breathing
system, some gas wil l be lost. The pressure-v olume loop (Fig. 23 .66) wi l l be normal
in shape but wi l l show a dec rease in t idal vo lume and a decrease in peak a irway
pressure . This differs f rom the leak associated with the open loop in that the
inhaled volume is dec reased as a resul t of the leak, but the amount of gas exha led
is equal to the amount of gas inhaled.
Flow-volume Loop The f low-v olume loop during a part ial disconnection (Fig. 23.66) wi l l show a
decreased t idal volume and a decrease in peak expira tory f low. The loop wil l no t be
open because both inspired and exhaled tidal volumes wi l l be the same.
View Figure
Figure 23.64 Misconnection of the tubings will cause the loop to be drawn backward and upside down.
P.766
View Figure
Figure 23.65 A disconnection between the sensor and the breathing system will result in no flow through the sensor.
Disconnection between the Sensor and the Patient With a disconnection between the sensor and the patient, during controlled
venti lation there wil l be f low during inspiration but not exhalat ion (Fig. 23 .67). The
pressure-volume loop wi l l show a high tida l volume but almost no pressure. Since
there wi l l be no return gas f low, there wi l l be only half of a p ressure- or f low-v olume
loop.
Bronchial Intubation Bronch ia l in tubat ion can occur anytime a trachea l tube is in place. It is the most
f requently occurring problem leading to hypox ia (95). If this happens , the
compliance decreases and the pressure needed to deliver a set t idal volume rises
(21,96,97). Bronchial intubation is discussed in Chapter 19 . It is most l ikely to
occur wi th a change in the patient 's posi tion. If a sudden change occurs in the
pressure-volume loop at this t ime, bronchial intubation should be suspected.
Withdrawing the tracheal tube s l ight ly wi l l remedy the prob lem and wi l l be
demonstra ted wi th the nex t loop. Pressure-volume loops may be among the best
ways of detect ing this problem.
Pressure-volume Loop The pressure-volume loop wil l show a decrease in compl iance wi th a rightward and
downward shif t and a h igh peak pressure (Fig. 23.68, sol id l ine).
View Figure
Figure 23.66 With a leak between the sensor and the breathing system, the loop will have a normal shape but the tidal volume, peak pressure, and expiratory flows will be decreased. The dotted line represents the normal loop, and the solid line represents the loop with a leak.
P.767
View Figure
Figure 23.67 A disconnection distal to the sensor will result in flow during inspiration but none during exhalation.
Flow-volume Loop The f low-v olume loop associa ted wi th bronchial intubat ion is shown in F igure 23.69
(sol id l ine). The increase in peak pressure wil l increase the peak f low during
exhalation .
Double-lumen Bronchial Tube Problem
Double-lumen bronchial tubes are of ten placed incorrec tly or may be displaced
during pat ient positioning or surgery (6,21,22,98,99,100). Continuous spirometric
moni toring can help to detect the problem. Basel ine flow-volume and pressure-
volume loops should be es tablished and recorded for each pat ien t whi le in the
supine posi t ion during two-lung vent i lat ion to al low comparison with la ter loops.
Pressure-volume Loop When one-lung venti lat ion is begun, the pressure-volume loop should show a s light
sh if t of the s lope to the right, ref lect ing decreased compl iance (98,99,100). With
surgical handling of the nondependent lung, compliance may decrease further
(100).
I f a doub le-lumen tube is placed too deeply, there wil l appear to be a dec rease in
compliance (Fig. 23.70). Tidal volume may be decreased. Incomplete lung emptying
wi l l result in an open loop (22).
The loops shown in Figure 23.71 resu lted f rom the t ip of the bronchial lumen
impinging on the wa ll of the bronchus. Loop 1 (solid l ine) is the normal loop. In loop
3 (dotted l ine), the tube t ip has impinged on the wal l of the bronchus . Pressure
rises rap idly wi th l i t tle increase in volume unt i l suff icien t pressure has been
P.768
P.769
exerted to move the t ip away from the wal l . At this point, the volume rises rapidly.
There wi l l be an increase in peak pressure and/or a decrease in compliance. If
there is a bal l-valve action , during exhalat ion the pressure wil l dec rease rapidly
wi th l i t t le change in volume unt i l the pressure in the bronchial tube has fallen
suff ic iently to allow exhala tion.
View Figure
Figure 23.68 Bronchial intubation (solid loop) will result in a decrease in compliance.
View Figure
Figure 23.69 Bronchial intubation. The solid loop shows an increase in expiratory flow, especially during the early part of expiration.
View Figure
Figure 23.70 The dotted line represents the loop when the double-lumen tube is correctly positioned. If the bronchial lumen is inserted too deeply, a severe reduction in compliance and tidal volume will be seen.
View Figure
Figure 23.71 Impingement of the end of the tube on the bronchial wall has created a ball-valve obstruction. The pressure rises rapidly with little increase in volume until the pressure is sufficient to overcome the obstruction. The volume then increases, and the pressure drops. If the pressure drops low enough, there will again be obstruction to flow, creating another notch on the upswing of the loop.
Flow-volume Loop The f low-v olume loop during proper doub le-lumen tube placement wi l l show a
s light ly decreased expira tory f low ra te. If the tube is p laced too deeply into the
bronchus, the loop wi l l show dimin ished inspiratory and expiratory f lows . If there is
a ball -va lve obstruction to the t ip of the double-lumen tube, the f low-volume loop
wi l l be irregular (F ig. 23.72).
The f low-v olume loop that would be generated if there were a disconnect ion of one
l imb of a double-lumen tube or a leaking bronchial cuff is shown in Figure 23.73
(sol id l ine). The inspira tory portion is normal. Because much of the insp ired tida l
volume is lost through the leak, the loop is open. In addi t ion, there wi l l be
decreased f low during exhalation .
View Figure
Figure 23.72 The loop represents repeated ball-valve obstruction to flow during both inspiration and expiration.
P.770
View Figure
Figure 23.73 With a leak in the bronchial cuff or a disconnection of one limb (solid loop), there will be an open loop, with the inhaled volume exceeding the exhaled volume. Exhaled flows will be decreased.
Esophageal Intubation
With esophageal intubat ion, the pressure-vo lume loop wi l l usually show a decrease
in compliance (Fig . 23.74, solid l ine), although compliance may be inc reased or
normal (21). Gas that enters the stomach and is not returned wi l l c reate an open
loop. The f low-volume loop wi l l be dis to rted and show small inspiratory and
expiratory volumes, often wi th an open loop.
Obstructed Tube A nearly completely obstructed tracheal tube or an incorrect ly p laced suprag lo tt ic
ai rway device wi l l resul t in a pressure-vo lume loop that shows a high pressure wi th
l i t tle or no t idal volume (Fig. 23 .75). This is usually a s ignal tha t the device should
be removed and reinserted.
Advantages of Loop Technology Pressure- and f low-volume loops prov ide the c l inic ian with real-time information .
There are many problems that can occur unexpectedly during a case (e.g., k inked
tracheal tube, disconnection , migrat ion of a tracheal tube into a bronchus). Without
loops, these problems might no t be recogn ized promptly and correct ive act ion may
be de layed. Whi le many of the problems that occur during respira tion could be
determined by watching
P.771
pressure , tidal volume, and f low moni to rs, loop technology offers a graphic
representation tha t integra tes that information , making changes more obv ious.
View Figure
Figure 23.74 The loops associated with esophageal intubation may vary greatly. Compliance may be increased, decreased, or normal.
View Figure
Figure 23.75 With a nearly obstructed tube, there will be a high pressure with little volume.
References 1. Banner MJ. Respiratory muscle loading and the work of breathing. J
Cardiothorac Vasc Anesth 1995;9:192–204.
[CrossRef]
[Med line Link]
2. Ell io tt WR, Topulas GP. The inf luence of the mechan ics of anesthesia breathing
c i rcui ts on respiratory moni toring . Biomed Inst Tech 1990;24:260–265.
3. Khorasan i A, Candido KD, Saatee S, e t al . The re la tionship between dynamic
compliance and inspiratory f low. Anes th Ana lg 1999;88:465.
[Fu ll text Link]
[CrossRef]
[Med line Link]
4. Tobin MJ . Moni toring of p ressure, f low and volume during mechan ical venti lation .
Resp Care 1992;37:1081–1096.
[Med line Link]
5. Bardoczky GI, Engelman EE. Comparison o f ai rway pressures from different
measuring s ites. Anesth Analg 1996;83:887.
[Fu ll text Link]
[CrossRef]
[Med line Link]
6. Bardoczky G, de Vries J, Meri lainen P, et a l. S ide s tream spirometry. Mon itoring
patient venti lat ion during anaesthesia . Datex Div ision Ins trumentat ion Corp .,
Helsink i , Finland.
7. F ilmyer WG, Sardar A, Weissman C. Respira tory f low measurements and
anesthet ic drugs : some in v i tro observations . J Clin Anes th 1999;11:355–359.
[CrossRef]
[Med line Link]
8. Goto T, Saito H, Nakata Y, e t al . Effec ts of xenon on the performance of various
respira tory f lowmeters. Anesthesiology 1999;90:555–563.
[Fu ll text Link]
[CrossRef]
[Med line Link]
9. Habre W, Asztalos T, Sly PD, et al . Viscosi ty and dens ity of common anaesthetic
gases: implications fo r f low measurements . Br J Anaesth 2001;87:602–607.
[Full text Link]
[CrossRef]
[Med line Link]
10. Jaffe MB. F low measurement. In : Gravens tein JS, Jaffe MB, Paulus DA, eds .
Capnography: c l inical aspec ts. Cambridge, MA: Cambridge Universi ty Press, 2004:
413–421.
11. Wilson S, Frerk C. A paed iatric ai rway moni tor for use with the T-piece.
Anaesthesia 1997;52:185.
[Fu ll text Link]
[Med line Link]
12. Orr JA, Westenskow DR, Farre ll RM. Information content of three breathing
c i rcui t moni to rs: a neural ne twork ana lysis. Anesthesiology 1992;77:A517.
[Fu ll text Link]
[CrossRef]
13. American Society for Testing and Materials . Standard specif ica tion for
part icula r requirements for anesthesia workstat ions and their components (ASTM F-
1850-00). West Conshohocken, PA: Author,2000.
14. Eisenkraf t JB. Prob lems wi th anesthesia gas del ivery systems (ASA Refresher
Course #212). Park Ridge, IL: ASTM, 2004.
15. Hatch DJ, Jackson EA. A new crit ical inc ident moni tor for use wi th the
paediatric T-piece. Anaesthesia 1996;51:839–842.
[Fu ll text Link]
[CrossRef]
[Med line Link]
16. Meeke R, Wren W, Davenport J , e t al . The measurement of tidal volumes in
spontaneously breathing children during genera l anaesthesia using a Haloscale
infant Wrigh t respirometer. Acta Anaesth Scand 1984;28:696–699.
[Med line Link]
17. Bushman JA. Effect of d if fe rent flow patterns on the Wrigh t respirometer. Br J
Anaesth 1979;51:895–898.
[CrossRef]
[Med line Link]
18. Ilsley AH, Hart JD, Withers RT, et al. Evaluation of five small turbine-type
respirometers used in adul t anes thesia. J Cl in Monit 1993;9:196–201.
[CrossRef]
[Med line Link]
19. Cicman J, Himmelwright C, Skibo V, et a l. Opera ting princ iples of Narkomed
anesthesia systems. Telford , PA: North American Drager 1993.
20. Chung DC, Ho AMH, Tay BA. “Apnea-volume” warning during normal venti lation
of the lungs : an unusual leak in the Narkomed 4 anesthesia sys tem. J Clin Anes th
2001;13:40–43.
[CrossRef]
[Med line Link]
21. Bardoczky GI, Engelman E, D'Hollander A. Continuous spirometry: an aid to
moni toring venti lat ion during operation. Br J Anaesth 1993;71:747–751.
[CrossRef]
[Med line Link]
22. Bardoczky GI, deFrancquen P, Engelman E, et al. Continuous moni toring of
pulmonary mechanics wi th the s ides tream spirometer during lung transp lan tation. J
Cardiothorac Vasc Anesth 1992;6:731–734.
[CrossRef]
[Med line Link]
23. Huffman LM. AANA journal course : new techno logies in anesthesia: update fo r
nurse anesthetis ts—monitoring vent i lat ion and compliance with Side S tream
SpirometryTM . J Am Assoc Nurse Anesth 1991;59:249–258.
24. Meri lainen P, Hanninen H, Tuomaala L. A novel sensor for rout ine cont inuous
sp irometry of intubated patients. J Cl in Moni t 1993;9 :374–380.
[CrossRef]
[Med line Link]
25. Puolakka JJ, Jouse la IT. The effects of a heat and moisture exchange f i l te r on
s ides tream spirometry in c ri tica lly i l l pa tients. Int J Cl in Monit Comput
1993;11:217–222.
[CrossRef]
[Med line Link]
26. Sondergaard S, Karason S, Lundin S, et a l. Evaluat ion of a Pitot type
sp irometer in helium/oxygen mixtures. J Cl in Moni t 1998;14:425–431.
27. Anonymous An at las of sp irometry. Helsinki , Finland Datex Corporation .
28. Jaffe MB. Performance of a combined neonatal f low/CO2 a irway adapter under
loaded condi tions. J Cl in Mon it 1998;14:519.
29. Dhar P, George I, Mankad A, e t al . Flow transducer gas leak detected af te r
induction. Anesth Analg 1999;89:1587.
[Fu ll text Link]
[CrossRef]
[Med line Link]
30. Aldridge J. Leak on Datex Aestiva/5 anaesthet ic machine. Anaes thesia
2005;60:420–421.
[Fu ll text Link]
[CrossRef]
[Med line Link]
31. Hari MS, Jennings M. Flow sensor faul t causing venti la tor malfunc tion.
Anaesthesia 2005;60:1049–1050.
[Fu ll text Link]
[CrossRef]
[Med line Link]
32. Krock J, Padda G, Moore J. Ohmeda 7900 vent i lator f low sensor. Anesth Analg
1998;86:231–232.
[Fu ll text Link]
[CrossRef]
[Med line Link]
33. Shogase A, Mizutani K, Toyoda Y. Protect ing Ohmeda 7900 venti lator f low
sensor. Anesth Analg 1999;88:234.
[Fu ll text Link]
[CrossRef]
[Med line Link]
34. Cantil lo J, Domsky R, Gra tz I , et al . Venti lato ry failures wi th the Datex-Ohmeda
7900 SmartVent. Anesthesiology 2002;96:766–768.
[Fu ll text Link]
[CrossRef]
[Med line Link]
35. Snyders S. Fai lure of a f low sensor of a Datex Ohmeda S/5 Aespire.
Anaesthesia 2005;60:941.
[Fu ll text Link]
[CrossRef]
[Med line Link]
36. Purnel l RJ. The posi t ion of the Wrigh t anemometer in the ci rc le absorber
system. Br J Anaesth 1968;40:917–918.
[Med line Link]
37. Feldman JM, Muller J . Tidal volume measurement e rrors—the impact of lung
compliance and a c i rcuit humidif ier. Anesthesio logy 1990;73:A468.
38. Thorpe CM. Venti la to rs, c ircle sys tems and respirometers. Anaesthesia
1992;47:913.
[CrossRef]
[Med line Link]
39. Anonymous . Overv iew of ai rway pressure alarms . Technol Anesth 2002;22:8–
10.
40. Alo i A, Burns SM. Continuous ai rway pressure moni to ring in the crit ical care
se tt ing. Crit Care Nurse 1995;15:66–74.
[Med line Link]
41. Anonymous . An overv iew of ai rway pressure alarms . Technol Anes th
1995;16:1–4.
42. Myerson KR, I ls ley AH, Runciman WB. An evaluat ion of venti lato r moni to ring
alarms. Anaesth Intens Care 1986;14:174–185.
[Med line Link]
43. Mimpriss T, Spivey A . A s imple disconnect alarm. J Med Eng Technol
1989;13:222–224.
[Med line Link]
44. Winter A, Spence AA. An in ternat ional consensus on moni to ring. Br J Anaes th
1990;64:263–266.
[CrossRef]
[Med line Link]
45. Epstein RA. The elus ive “disconnec t alarm” examined. APSF Newslett
1988;3:39.
46. Raphael DT, Weller RS, Coran DJ . A response a lgori thm for the low-pressure
alarm condi tion . Anes th Ana lg 1988;67:876–883.
[Fu ll text Link]
[CrossRef]
[Med line Link]
47. Raphael DT. The low-pressure alarm condit ion: safety considerations and the
anesthesiologist 's response. APSF News le tt 1998–1999;13:33–40.
48. Campbell RM, Sheikh A, Crosse MM. A study of the incorrect use of venti lator
disconnection alarms. Anaesthesia 1996;51:369–370.
[Fu ll text Link]
[CrossRef]
[Med line Link]
49. Eisenkraf t JB. Compl ications of anesthesia del ivery systems (ASA Annual
Ref resher Course #255). Park Ridge, IL: ASA, 1996.
50. Pryn SJ, Crosse MM. Ventilato r disconnex ion alarm failu res . Anaesthesia
1989;44:978–981.
[CrossRef]
[Med line Link]
51. Bourke AE, Snowdon SL, Ryan TDR. Fai lure of a venti lator alarm to detect
patient d isconnect ion. J Med Eng Technol 1987;11:65–67.
[Med line Link]
P.772
52. Murphy PJ, Rabey PG. The Humphrey ADE breathing system and venti lato r
alarms. Anaesthesia 1991;46:1000.
[CrossRef]
[Med line Link]
53. Lawrence JC. Breathing system gas pressure monitoring and vent ing, vent i lator
moni tors and ala rms. Anaes th Intens Care 1988;16:38–40.
[Med line Link]
54. Russell WJ, Webb RK, Van der Wal t JH, e t al . Problems with vent i la tion: an
analys is of 2000 inc ident reports. Anaesth Intens Care 1993;21:617–620.
[Med line Link]
55. McEwen JA, Smal l CF, Jenkins LC. Detect ion of interruptions in the breath ing
gas of venti la ted anaes thetized patients. Can J Anaes th 1988;35:549–561.
[Med line Link]
56. Sinclair A, VanBergen J. Flow res is tance of coaxial breathing systems:
investigat ion of a ci rcui t disconnec t. Can J Anaesth 1992;39:90–94.
[Med line Link]
57. Raphael DT. An algori thmic response for the breath ing system low-pressure
alarm condi tion . Prog Anesthesiol 1997;11:219–244.
58. Slee TA, Pav lin EG. Fai lure of low pressure alarm associa ted wi th the use o f a
humidif ier. Anesthesiology 1988;69:791–793.
[Fu ll text Link]
[CrossRef]
[Med line Link]
59. Mil l igan KA. Disablement of a venti lato r disconnec t alarm by a heat and
moisture exchanger. Anaesthes ia 1992;47:279.
[CrossRef]
[Med line Link]
60. French R, Kennedy R. Disconnect a larm fai lure in detect ion of common gas
outlet disconnection . Anaesth Intens Care 1998;26:665–670.
[Med line Link]
61. Anonymous . Heat/moisture exchange humid if ie rs. Technol Anesth 1991;11:5.
62. Bashein G, MacEvoy B . Anes thesia venti lators shou ld have adjustable high-
pressure alarms. Anesthesiology 1985;63:231–232.
[Fu ll text Link]
[CrossRef]
[Med line Link]
63. Tung A, Morgan SE. Model ing the effect o f progressive endotrachea l tube
occlusion on tida l volume in pressure-contro l mode. Anes th Ana lg 2002;95:192–
197.
[Fu ll text Link]
[CrossRef]
[Med line Link]
64. Hodgson CA, Mos tafa SM. Riddle of the pers istent leak. Anaesthesia
1991;46:799.
[CrossRef]
[Med line Link]
65. Walker T. Another problem with a c ircle sys tem. Anaes thesia 1996;51:89.
[Fu ll text Link]
[CrossRef]
[Med line Link]
66. Spielman FJ, Sprague DH. Another benef i t of the subatmospheric alarm.
Anesthesiology 1981;54:526–527.
[Fu ll text Link]
[CrossRef]
67. Sola A , Farina D, Rodriguez S, et al. Lack of relat ionship between the true
ai rway pressure and the pressure displayed with an in fan t venti la to r. Crit Care Med
1992;20:778–781.
[CrossRef]
[Med line Link]
68. Huffman LM. Monitoring compl iance: a sensit ive ind icator of change. J Am
Assoc Nurse Anes th 1992;60:217–220.
69. Heyer L, Louis B, Isabey D, et al. Noninvasive est imate of work of b reathing
due to the endotrachea l tube. Anesthes io logy 1996;85:1324–1333.
[Fu ll text Link]
[CrossRef]
[Med line Link]
70. Oikkonen M, Tallgren M. Changes in respiratory compl iance at laparoscopy:
measurement using s ide stream spirometry. Can J Anaes th 1995;42:495–497.
[Med line Link]
71. Bardoczky GI, Engelman E, Levarle t M, et al. Venti latory ef fects of
pneumoperi toneum moni to red wi th cont inuous spirometry. Anaesthesia
1993;48:309–311.
[CrossRef]
[Med line Link]
72. Bennett JA, Abrams JT, VanRiper DF, et al . Diff icul t venti lation af ter opioid
induction is caused by la ryngospasm. Anesthesio logy 1995;83:A74.
73. Chassard D, Berrada K, Tournadre J-P, et al . The effects of neuromuscular
block on peak ai rway pressure and abdominal e las tance during pneumoperitoneum.
Anesth Analg 1996;82:525–527.
[Fu ll text Link]
[CrossRef]
[Med line Link]
74. Fahy BG, Barnas GM, Nagle SE, et a l. Changes in lung and chest wal l
propert ies wi th abdominal insufflat ion of carbon diox ide are immedia te ly revers ible.
Anesth Analg 1996;82:501–505.
[Fu ll text Link]
[CrossRef]
[Med line Link]
75. Fahy BG, Barnas GM, Nagle SE, et a l. Effects of Trendelenburg and reverse
Trende lenburg postures on lung and chest wall mechanics. J Cl in Anesth
1996;8:236–244.
[CrossRef]
[Med line Link]
76. Makinen M-T, Yl i-Hankala A. The effec t of laparoscopic cholecys tec tomy on
respira tory compliance as determined by cont inuous spirometry. J Cl in Anesth
1996;8:119–122.
[CrossRef]
[Med line Link]
77. Kytta J, Rande ll T , Tanskanen P, et al . Mon itoring lung compl iance and end-
t ida l oxygen content for the detect ion of venous ai r embolism. Br J Anaes th
1995;75:447–451.
[Med line Link]
78. Makinen M-T, Yl i-Hankala A, Kansanaho M. Early detect ion of CO2
pneumothorax wi th continuous spirometry during laparoscopic fundoplicat ion . Acta
Anaesthes Scand 1995;39:411–413.
[Med line Link]
79. Lu Q, Rouby JJ. Measurement of pressure-vo lume curves in pa tients on
mechanical vent i la tion: methods and s ignif icance. Cri t Care 2000;4 :91–100.
[CrossRef]
[Med line Link]
80. Bardoczky GI, de Vries J, Meri la inen P. Pat ient sp irometryTM . Monitoring of
patient venti lat ion during anesthesia. City, STATE: Datex Ohmeda, Year.
81. Bardoczky G, d Hol lander A , Yernaul t J-C, et al. On-l ine expira tory f low-volume
curves during thoracic surgery: occurrence of auto-PEEP. Br J Anaesth
1994;72:25–28.
[CrossRef]
[Med line Link]
82. Patel H, Yang KL. Variabil i ty of intr ins ic posi t ive end-expiratory pressure in
patients receiv ing mechanical veni tlat ion. Crit Care Med 1995;23:1074–1079.
[Fu ll text Link]
[CrossRef]
[Med line Link]
83. Anonymous . Pat ient spi rometry. Clinical interpre ta tion in c ri tical care. Hels ink i,
Finland: Datex Ohmeda.
84. Bardoczky GI, d ' Hol lander AA, Cappello M, et al. Interrupted expira tory flow on
automatical ly cons tructed flow-volume curves may determine the presence of
intr insic posi t ive end-exp iratory pressure during one-lung vent i lat ion. Anesth Analg
1998;86:880–884.
[Fu ll text Link]
[CrossRef]
[Med line Link]
85. Dueck R, Cooper S, Kapelansk i D, et a l. A pi lot s tudy of expiratory f low
l imi ta tion and lung volume reduction surgery. Ches t 1999;16:1762–1771.
[CrossRef]
[Med line Link]
86. Pul le ri ts J , Holzman R. Anesthesia fo r pat ients wi th mediast inal masses. Can J
Anaesth 1989;36:681–688.
[Med line Link]
87. Pinsonneaul t C, Fort ier J , Donati F. Tracheal resect ion and reconstruction. Can
J Anaesth 1999;46:439–455.
[Med line Link]
88. Weissman C. Flow-volume re la tionship between spontaneous breathing through
endotracheal tubes. Cri t Care Med 1992;20:615–620.
[CrossRef]
[Med line Link]
89. Dahlberg CGW, Hales CA. Pulmonary funct ion tes ts to evalua te the airway. In :
Roberts JT, ed. Clinical management of the ai rway. Philadelphia: WB Saunders,
1994: 56–62.
90. Brodsky JB. Bronchoscopic procedures for central a irway obs truction. J
Cardiothorac Vasc Anesth 2003;17:638–646.
[CrossRef]
[Med line Link]
91. Jubran A, Tobin MJ . Use of f low-volume curves in detecting secret ions in
venti lato r-dependent patients. Am J Respir Crit Care Med 1994;150:766–769.
[Med line Link]
92. Yokota K, Toriumi T, Sari A, et al . Auto-posit ive end-exp iratory pressure during
one-lung venti lat ion using a double-lumen endobronchial tube. Anesth Analg
1996;82:1007–1010.
[Fu ll text Link]
[CrossRef]
[Med line Link]
93. Robinson RJS. The use of s ide-stream spirometry to assess ai r leak during and
af te r lung volume reduct ion surgery. Anesthesiology 1999;91:571–573.
[Fu ll text Link]
[CrossRef]
[Med line Link]
94. Breen PH, Serina ER, Barker SJ . Exhaled f low moni toring can detec t bronch ial
f lap-valve obstruct ion in a mechanical lung model. Anesth Analg 1995;81:292–
296.
[Fu ll text Link]
[CrossRef]
[Med line Link]
95. Cooper J. ICPAMM meeting review of safety includes new worldwide expans ion .
J Cl in Moni t 1997;13:205–210.
96. Campos C, Naguib SS, Chuang AZ, e t al . Endobronchial intubation causes an
immediate increase in peak in flation pressure in ped iatric pa tients. Anesth Analg
1999;88:268–270.
[Fu ll text Link]
[CrossRef]
[Med line Link]
97. Mahajan A, Wald SH, Schroeder R, et al. Detect ion of endobronchia l intubation
in infants and ch ildren using sp irometry. Anes thesiology 2002;96:A1268.
98. Simon BA, Hurford WE, Alf i lle PH, et al. An a id in the d iagnosis of
malposit ioned double-lumen tubes. Anesthesiology 1992;76:862–863.
[Fu ll text Link]
[CrossRef]
[Med line Link]
99. Bardoczky GI, Levarlet M, Engelman E, et al. Continuous spirometry for
detec tion of double-lumen endobronchial tube d isplacement. Br J Anaesth 1993;70:
499–502.
[CrossRef]
[Med line Link]
100. Iwasaka H, Itah K, Miyakawa H, et al . Continuous moni toring of vent ilatory
mechanics during one-lung venti lation . J Clin Moni t 1996;12:161–164.
[CrossRef]
[Med line Link]
P.773
Questions For the fol lowing quest ions, select the correct answer
1. Normal static compliance in an adult is
A. 15 mL/cm H2O
B. 35 to 100 mL/cm H2O
C. 20 to 35 mL/cm H2O
D. 35 to 50 mL/cm H2O
E. 75 to 125 mL/cm H2O
View Answer2. For greatest accuracy, the gas flow sensor should be placed
A. In the inspiratory l imb a t the absorber
B. Between the insp iratory l imb and the Y-piece
C. Between the Y-piece and the pat ient
D. Between the Y-piece and the expiratory l imb
E. On the absorber on the exha lat ion s ide
View Answer3. The preferred site for monitoring a irway pressure is
A. At the vent ilator
B. In the inspiratory l imb a t the canister
C. At the connec tion between the patient and the breathing system
D. On the expiratory l imb at the canister
E. At the bag mount connection
View AnswerFor the following quest ions, answer
• i f A, B, and C are correct
• i f A and C are correct
• i f B and D are correct
• is D is correct
• i f A, B, C, and D are correct.
4. What does the difference between the peak and plateau pressure measure? A. Plateau pressure
B. Total ai rway resistance
C. Compliance
D. Insp iratory f low
E. Resistance in the breathing system
View Answer5. Use of muscle relaxants
A. May cause the compliance of the lungs to increase
B. May cause the compliance of the ches t wal l to increase
C. May cause a decrease in res istance
D. May resu lt in a decrease in plateau pressure
View Answer6. A rise in peak airway pressure may occur as a result of A. An increase in t ida l volume
B. An increase in inspiratory f low rate
C. An increase in res istance
D. An increase in compliance
View Answer7. With inadvertent bronchial intubation, A. There wi l l be a decrease in compliance
B. The pressure-volume loop wil l be shif ted to the right and downward
C. The peak pressure wi l l be increased
D. The f low-volume loop wil l show an increase in peak expiratory f low
View Answer8. Increased resistance to breathing during controlled respiration can be overcome by A. Adding or increasing PEEP
B. Increasing the driv ing pressure
C. Increas ing the inspired oxygen
D. Decreasing inspira tory f low
View Answer9. Tota l res istance during inspiration and expiration is influenced by A. The pat ien t's ai rway
B. The exha lat ion portion of the breathing sys tem
C. The s ize and charac teris t ics of the tracheal tube
D. The amount of PEEP added to the breathing system
View Answer10. Means to detect a leak around a tracheal tube include A. Airway pressure moni tors
B. Respiratory breath sound monitoring
C. Respiratory volume measurement
D. Capnography
View Answer11. Respiratory volume monitoring may fa il to detect
A. Occlusion of the ai rway
B. A disconnec tion during spontaneous breathing
C. Esophageal intubation
D. Apnea
View Answer12. Which factors are components of total compliance? A. Elas tic properties of the lungs
B. Elas tic properties of the thorax
C. Elastic properties of the abdomen
D. Elastic properties of the breathing sys tem
View Answer13. Conditions that can be detected by using a minimum pressure alarm include
A. An unconnected venti lato r
B. A major leak in the breathing system
C. A malfunct ioning scavenging system
D. An increase in res istance
View Answer14. The minimum airway pressure alarm should be set A. At the lowes t setting
B. At dif ferent sett ings during a case
C. Greater than 6 cm H2O
D. Slight ly less than the peak pressure during inspirat ion
View Answer15. Conditions that may prevent activation of the minimum pressure alarm with a disconnection include
A. Obstruct ion of a breathing sys tem connec tor
B. Part ia l extubation
C. High resistance of components of the breathing system
D. Air en trained in to the breathing system
View Answer16. A sustained pressure may be caused by A. Occlusion of the scaveng ing system
B. Ac tivat ion of the oxygen f lush valve
C. Improper adjustment of the APL valve
D. Occlusion of the inspira tory l imb of the breathing system
View Answer17. An excessively high pressure in the breathing system may be caused by A. The pat ien t coughing or straining
B. Use o f the oxygen f lush during the inspiratory phase of the vent ilator cycle
C. A punctured venti lator be llows
D. Increased compl iance
View AnswerP.774
18. Subambient pressure in the breathing system may be caused by A. A nasogastric tube p laced in the trachea and attached to suc tion
B. A blocked expiratory l imb
C. Insp irat ion wi th an empty reservoir bag
D. Refi l l ing of a venti lato r with a standing bel lows wi th low f resh gas f lows
View Answer19. On a pressure-volume loop, A. The farthest poin t to the righ t on the horizontal axis represents the tidal volume
B. The slope of the inspira tory portion is determined by the resistance
C. The h ighest point on the curve represents the peak pressure
D. The curve can s lope to the right or lef t during inspirat ion
View Answer20. Concerning a flow-volume loop,
A. The portion below the horizonta l l ine represents insp irat ion using the
representation used by pulmonologists
B. The portion above the horizontal l ine represents the passive def lat ion as
determined by the e lastic recoi l of the lungs and chest wal l in the representa tion
used by pulmonologis ts
C. The t idal volume is that poin t where the loop crosses the horizontal l ine
D. This loop is known as a compl iance loop
View Answer21. If a double-lumen tube is placed too deeply, A. The pressure-volume loop wil l be shif ted to the right and downward
B. There wi l l be a decrease in compliance
C. An overshoot loop may resul t
D. Resis tance wi l l inc rease
View Answer22. Possible causes of inadvertent PEEP include
A. Part ia l obs truct ion of the breath ing system
B. Malfunct ioning scavenging device
C. Obs tructive ai rway d isease
D. Airf low obstruct ion
View Answer23. Possible causes of an open loop include A. Improper cal ibrat ion o f the monitor
B. An uncuffed tracheal tube
C. A leak distal to the sensor
D. Tension pneumothorax
View Answer24. Decreases in compliance may be caused by
A. Inadequate muscle relaxat ion
B. Bronchial intubat ion
C. Obesi ty
D. Reverse Trendelenburg posi tion
View Answer25. With restrictive lung disease,
A. Peak exp iratory f low wi l l be increased
B. The f low-volume loop may become tall and narrow
C. The f low-volume loop may have a rela tively normal shape but may appear
smal ler in all dimensions
D. Resis tance is inc reased
View Answer