Upload
lynhan
View
223
Download
0
Embed Size (px)
Citation preview
IAEAInternational Atomic Energy Agency
Slide set of 141 slides based on the chapter authored by
Hee Kwon Song
of the IAEA publication (ISBN 978-92-0-131010-1):
Diagnostic Radiology Physics:
A Handbook for Teachers and Students
Objective:
To familiarize the student with the fundamental concepts of
MRI.
Chapter 14: Physics of Magnetic
Resonance
Slide set prepared
by E. Berry (Leeds, UK and
The Open University in
London)
IAEA Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14 (02/141)
CHAPTER 14 TABLE OF CONTENTS
14.1. Introduction
14.2. Nuclear magnetic resonance
14.3. Relaxation and tissue contrast
14.4. MR spectroscopy
14.5. Spatial encoding and basic pulse sequences
IAEA
14.1 INTRODUCTION14.1
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.1 Slide 1 (03/141)
IAEA
14.1 INTRODUCTION14.1
Nuclear magnetic resonance (NMR)
� Nuclei in a magnetic field absorb applied radiofrequency
(RF) energy and later release it with a specific frequency
� 1920s – Stern and Gerlach
• particles have intrinsic quantum properties
� 1938 – Rabi
• discovered phenomenon of NMR (Nobel prize 1944)
� 1946 – Bloch and Purcell
• measured NMR signal from liquids and solids (Nobel prize 1952)
� But no imaging yet 9
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.1 Slide 2 (04/141)
IAEA Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.1 Slide 3 (05/141)
14.1 INTRODUCTION14.1
Magnetic resonance imaging (MRI)
� 1973 – Lauterbur
• method to spatially encode the NMR signal using linear magnetic
field gradients
� 1973 – Mansfield
• method to determine spatial structure of solids by introducing
linear gradient across the object
� i.e apply magnetic field gradients to induce spatially
varying resonance frequencies to resolve spatial
distribution of magnetization
� Milestone – the beginning of MR Imaging
� Nobel prize in medicine in 2003
IAEA Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.1 Slide 4 (06/141)
14.1 INTRODUCTION14.1
Characteristics of MRI
� No ionizing radiation
• unlike x-rays and CT
� Superior soft tissue contrast compared with other
modalities
� Can control image contrast among different tissues by
adjusting acquisition timing parameters
IAEA
14.2 NUCLEAR MAGNETIC
RESONANCE14.2
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2 Slide 1 (07/141)
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2
Nuclear magnetic resonance
� 14.2.1 The nucleus: spin, angular and magnetic
momentum
� 14.2.2 External magnetic field and magnetization
� 14.2.3 Excitation and detection
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2 Slide 2 (08/141)
IAEA
14.2 NUCLEAR MAGNETIC
RESONANCE14.2.1 THE NUCLEUS: SPIN, ANGULAR AND MAGNETIC
MOMENTUM
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.1 Slide 1 (09/141)
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.1 The nucleus: spin, angular and magnetic momentum
Nuclei used for MRI
� MRI involves imaging the nucleus of hydrogen atom
• = proton
� Hydrogen abundant in human body in water and fat
• Water is 50-70% of total body weight
• Fat is 10-20% of total body weight
� Other nuclei are used in research
• carbon (13C), phosphorus (31P), fluorine (19F), sodium (23Na)
• relatively low abundance in vivo
• limited signal available
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.1 Slide 2 (010/141)
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE
14.2.1 The nucleus: spin, angular and magnetic
momentum
Properties of the nucleus
� Angular momentum
where h is Planck’s constant and I
is the nuclear spin (or quantum
number)
• for the hydrogen nucleus, I =
½
� Because the proton is positively
charged, the angular
momentum also produces a
nuclear magnetic moment
where γ is the gyromagnetic ratio
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.1 Slide 3 (011/141)
hIp =
pγµ =
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE
14.2.1 The nucleus: spin, angular and magnetic
momentum
Gyromagnetic ratio
� Specific to each type of nucleus
� For proton, roughly 42.57 MHz T-1
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.1 Slide 4 (12/141)
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.1 The nucleus: spin, angular and magnetic momentum
Nucleus Relative
Abundance (%)
Spin (I) Gyromagnetic
ratio (Hz/G)
Relative
sensitivity*
Abundance
in human
body (% of
atoms)1H 99.98 1/2 4258 1 6313C 1.11 1/2 1071 0.016 0.1319F 100 1/2 4005 0.83 0.0012
23Na 100 3/2 1126 0.093 0.03731P 100 1/2 1723 0.066 0.1439K 93.1 3/2 199 5.08 x 10-4 0.031
* PER EQUAL NUMBER OF NUCLEI
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.1 Slide 4 (13/141)
Common nuclei for MR
Adapted from Stark & Bradley, Magnetic Resonance Imaging, 2nd edition
IAEA
14.2 NUCLEAR MAGNETIC
RESONANCE14.2.2 EXTERNAL MAGNETIC FIELD AND
MAGNETIZATION
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 14.2.2 Slide 1 (14/141)
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.2 External magnetic field and magnetization
Boltzmann distribution
� Consider a collection of these magnetic moments
• or spins
� No external magnetic field
• random alignment: zero net magnetization
� In external magnetic field B0
• each spin aligns parallel or anti-parallel to direction of applied field
• i.e. polarized
• parallel orientation has lower energy state
• slightly greater number of spins align along that direction
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 14.2.2 Slide 2 (15/141)
IAEA
No external magnetic field
� Random alignment of
spins
� Zero net magnetization
In external magnetic field B0
� Each spin aligns parallel
or anti-parallel to
direction of applied field
� i.e. polarized
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.2 Slide 3 (16/141)
14.2 NUCLEAR MAGNETIC RESONANCE14.2.2 External magnetic field and magnetization
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.2 External magnetic field and magnetization
Boltzmann distribution
� Parallel orientation has lower energy state
� Slightly greater number of spins align along that direction
Where
• N+ is number of spins aligned parallel to direction of applied field
• N- is number of spins aligned anti-parallel to direction of applied field
• ∆E is energy difference between the two states
• k is Boltzmann constant
• T is absolute temperature
• ω0 is Larmor, or resonance, frequency
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.2 Slide 4 (17/141)
kTktE oeeN
N // ωh== ∆
−
+
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.2 External magnetic field and magnetization
Precession and Bloch Equation
� Torque on magnetization causes it to precess about the
direction of the magnetic field
� Will precess at the Larmor frequency,
� Analogous to precession of spinning top about the
direction of gravitational field
• top has angular momentum due to its spin
• precession arises from a torque acting on the top
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.2 Slide 5 (18/141)
B×= µ µ
γdt
d
oo Bγω =
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.2 External magnetic field and magnetization
Larmor frequency
� Under the influence of an
external magnetic field Bo,
the spins precess about
the direction of the field at
the Larmor frequency
which is proportional to Bo
� e.g. at 1.5 T, ω0 = 64 MHz
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.2 Slide 6 (19/141)
oo Bγω =
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.2 External magnetic field and magnetization
Rotating frame of reference
� In stationary (or laboratory) frame of reference,
precession:
• at Larmor frequency ω0
• about the direction of B0 (z-axis)
� In rotating frame of reference, which rotates at Larmor
frequency, precession:
• at Larmor frequency appears to be stationary
• at a different frequency ω in stationary frame, appears to precess
at ωr , where
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 14.2.2 Slide 7 (20/141)
or ωωω −=
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.2 External magnetic field and magnetization
Net magnetization
� Total magnetization within a voxel
• net magnetization
• vector sum of all spins within the voxel
• aligned along +z direction, the direction of B0
� **add diagram**
� Henceforth
• “Magnetization” = net magnetization of a collection of spins
• i.e. not magnetization of a single spin
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 14.2.2 Slide 8 (21/141)
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.2 External magnetic field and magnetization
Bloch equation for net magnetization M
� In the presence of a constant external magnetic field B0
• net magnetization is aligned along the z-axis, remains stationary
and does not precess about any axis
• magnetization is at its equilibrium magnetizationM0.
� When additional fields applied, including time varying
fields
• magnetization may deviate from equilibrium position
• magnetization may precess about an effective magnetic field
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.2 Slide 9 (22/141)
BMM
×= γdt
d
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.2 External magnetic field and magnetization
Perturbation from equilibrium
� Arises from applied additional magnetic fields
� Such a perturbation is needed for signal detection
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.2 Slide 10 (23/141)
IAEA
14.2 NUCLEAR MAGNETIC
RESONANCE14.2.3 EXCITATION AND DETECTION
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.3 Slide 1 (24/141)
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.3 Excitation and detection
Effect of external radiofrequency (RF) field B1(t)
� Spins are in a magnetic field B0
� B1(t) is resonating at the Larmor frequency
� From Bloch equation
• magnetization will precess about effective magnetic field
• given by vector sum of static B0 field and time varying B1 field
� In the rotating frame
• B1 field appears stationary
• magnetization is initially aligned along z-axis
• magnetization precesses about the direction of the B1 field
• will continue to do so as long as B1 is applied
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.3 Slide 2 (25/141)
IAEA
� If RF field B1 lies along the x-axis
• magnetization precesses, or
nutates, about the x-axis
� Typically, apply B1 field just long
enough to cause a 90° rotation
� At end of the RF pulse, (2-3 ms),
magnetization is aligned with y-
axis
• note that diagram shows rotating
frame with axes (xr, yr, zr)
Effect of external radiofrequency (RF) field B1(t)
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.3 Slide 3 (26/141)
14.2 NUCLEAR MAGNETIC RESONANCE14.2.3 Excitation and detection
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.3 Excitation and detection
Effect of external radiofrequency (RF) field B1(t)
� Once magnetization is rotated into the transverse (x, y)
plane
� RF is removed
� Precession is again
• about B0
• at the Larmor frequency (in stationary frame of reference)
• according to Bloch equation
� Can detect the rotating magnetization
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 14.2.3 Slide 4 (27/141)
IAEA
Detection of a signal from
rotating magnetization
� Use an RF coil
� By Faraday’s law,
changing magnetic flux
through coil induces
voltage changes
� Changes are detected
by receiver
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.3 Slide 5 (28/141)
14.2 NUCLEAR MAGNETIC RESONANCE14.2.3 Excitation and detection
IAEA
14.2 NUCLEAR MAGNETIC RESONANCE14.2.3 Excitation and detection
Detection of a signal from rotating magnetization
� Overall strength of received signal depends on
• type and size of RF coil used for signal reception
• proximity of coil to the imaged object
• voxel size
� More in Chapter 15
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.2.3 Slide 6 (29/141)
IAEA
14.3 RELAXATION AND
TISSUE CONTRAST14.3
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3 Slide 1 (30/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3
Relaxation and tissue contrast
� 14.3.1 T1 and T2 relaxation
� 14.3.2 Bloch equations with relaxation terms
� 14.3.3 T2* relaxation
� 14.3.4 Contrast agents
� 14.3.5 Free-induction decay (FID)
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3 Slide 2 (31/141)
IAEA
14.3 RELAXATION AND
TISSUE CONTRAST14.3.1 T1 AND T2 RELAXATION
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.1 Slide 1 (32/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.1 T1 and T2 relaxation
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.1 Slide 2 (33/141)
Spin-lattice, or T1, relaxation
� One of two mechanisms that drive magnetization back to
its equilibrium state
� Some of energy absorbed by spins from RF pulse is lost to
their surroundings - the lattice
� Time constant for this phenomenon is T1
• depends on the mobility of the lattice
• efficiency of energy transfer from excited spins to the lattice
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.1 T1 and T2 relaxation
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14. 3.1 Slide 3 (34/141)
Spin-lattice, or T1, relaxation
� The longitudinal component (z-component) of the
magnetization returns to its equilibrium state M0 in an
exponential fashion
where Mz(0) is the longitudinal magnetization immediately
following RF excitation
� For a 90° excitation, Mz(0) is zero
� After a time period of several T1s, the magnetization has
nearly fully returned to its equilibrium state
( ) ( ) ( )11 // 10 Tt
o
Tt
zz eMeMtM −− −+=
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.1 T1 and T2 relaxation
Spin-lattice, or T1, relaxation
� Light gray component
� After a time period of several T1s, the magnetization has
nearly fully returned to its equilibrium state:
• amplitude M0
• aligned with z-axis
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.1 Slide 4 (35/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.1 T1 and T2 relaxation
Spin-spin, or T2, relaxation
� In addition to interactions with the lattice, spins interact
with each other – “spin-spin”
� Each spin is essentially a magnetic dipole
• creates a magnetic field of its own
• slightly alters the field of its surroundings
� Any spin close to another will experience the additional
field
• which slightly alters its precessional frequency
� Spins are in constant motion, so precessional frequency of
each spin changes continuously
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.1 Slide 5 (36/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.1 T1 and T2 relaxation
Spin-spin, or T2, relaxation
� Result is a loss of phase coherence
� Leading to an exponential decay of signal in the
transverse plane
� With time constant T2
where Mxy(0) is the transverse magnetization immediately
following RF excitation
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.1 Slide 6 (37/141)
( ) ( ) 2/0 Tt
xyxy eMtM −=
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.1 T1 and T2 relaxation
Spin-spin, or T2, relaxation
� Dark gray component
� T2 relaxation reduces the transverse component towards
zero
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.1 Slide 7 (38/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.1 T1 and T2 relaxation
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.1 Slide 8 (39/141)
Typical T1 and T2 values in human tissues at 1.5 T
� T1 and T2 differ between tissue types
� T1 and T2 are field-strength dependent
Adapted from Bernstein, King and Zhou, Handbook of MRI pulse sequences, 2004
Tissue T1 (ms) T2 (ms)
Muscle 870 50
Fat 260 80
Liver 490 40
Blood (oxygenated) 1200 200
Blood (deoxygenated) 1200 125
White matter 790 90
Gray matter 920 100
Cerebrospinal fluid (CSF) 4000 2000
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.1 T1 and T2 relaxation
TR and image creation
� Repeat process of excitation and signal detection many
times until have sufficient data for image reconstruction
� Time between successive excitations is repetition time
(TR)
� From
value of TR determines the extent to which tissues with
various T1s have returned to their equilibrium state
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.1 Slide 9 (40/141)
( ) ( ) ( )11 // 10 Tt
o
Tt
zz eMeMtM −− −+=
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.1 T1 and T2 relaxation
TR and image creation
� If TR is short
• tissues with short T1s which relax more quickly will appear brighter
than those with longer T1 values
• differences in T1 between tissues will be emphasized
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.1 Slide 10 (41/141)
( ) ( ) ( )11 //10
Tt
o
Tt
zz eMeMtM−− −+=
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.1 T1 and T2 relaxation
TE and image creation
� Time between excitation and data acquisition is echo time
(TE)
� If TE is long
• tissues with short T2s which relax more quickly will appear darker
than those with longer T2 values
• differences in T2 between tissues will be emphasized
� More in Section 14.5.4
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.1 Slide 11 (42/141)
( ) ( ) 2/0 Tt
xyxy eMtM −=
IAEA
14.3 RELAXATION AND
TISSUE CONTRAST14.3.2 BLOCH EQUATIONS WITH RELAXATION TERMS
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.2 Slide 1 (43/141)
IAEA Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.2 Slide 2 (44/141)
14.3 RELAXATION AND TISSUE CONTRAST14.3.2 Bloch equations with relaxation terms
Bloch equations
� If the T1 and T2 relaxation constants are incorporated into
the Bloch equation (slide 17/141)
� Now includes
• effects of both static (B0) and dynamic (B1) magnetic fields
• relaxation of spins due to T1 and T2 relaxation
21
ˆˆˆ
T
yMxMz
T
MM
dt
d yxzo+
−−
+×= BMM
γ
IAEA
14.3 RELAXATION AND
TISSUE CONTRAST14.3.3 T2* RELAXATION
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.3 Slide 1 (45/141)
IAEA Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.3 Slide 2 (46/141)
14.3 RELAXATION AND TISSUE CONTRAST14.3.3 T2* relaxation
Effect of magnetic field inhomogeneities
� In a homogeneous field, the transverse signal decays
• exponentially
• with intrinsic T2 time constant
� Magnetic field inhomogeneities may arise from
• imperfect magnet shimming
• induced field perturbations (e.g. due to susceptibility differences at
tissue boundaries)
� These lead to additional loss of coherence among the
spins
� In an inhomogeneous field, the transverse signal may
decay more rapidly
IAEA Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.3 Slide 3 (47/141)
14.3 RELAXATION AND TISSUE CONTRAST14.3.3 T2* relaxation
Effect of magnetic field inhomogeneities
� Approximation for signal loss within a voxel
� Exponential decay with time constant T2*
• pronounced “T- two star”
• T2* < T2
• represents total transverse relaxation time
• i.e. both intrinsic T2 and the component due to inhomogeneous
field, T2′ “T- two prime”
IAEA Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.3 Slide 4 (48/141)
14.3 RELAXATION AND TISSUE CONTRAST14.3.3 T2* relaxation
T2* relaxation
where
� ∆B is the field inhomogeneity across a voxel
� γ is the gyromagnetic ratio
� T2* is the total transverse relaxation time constant
� T2 is the intrinsic transverse relaxation time constant
� T2′ is the transverse relaxation time constant due to the
inhomogeneous field
′+=∆+=
222
*
2
1111
TTB
TTγ
IAEA Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.3 Slide 5 (49/141)
14.3 RELAXATION AND TISSUE CONTRAST14.3.3 T2* relaxation
Most common incidences of field perturbations in vivo
� Near regions of air-tissue boundaries
� Sinus cavity in the head
� In trabecular (spongy) bone
• susceptibility difference between bone and bone marrow induces
local field inhomogeneities within the marrow
IAEA Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.3 Slide 6 (50/141)
14.3 RELAXATION AND TISSUE CONTRAST14.3.3 T2* relaxation
Reversal of signal loss
� Signal loss due to intrinsic T2 relaxation cannot be
avoided
� Signal loss due to field inhomogeneities can be reversed
� Apply a second RF pulse
• known as a refocusing pulse
� This is basis of spin echo imaging
� Section 14.5.5
IAEA
14.3 RELAXATION AND
TISSUE CONTRAST14.3.4 CONTRAST AGENTS
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.4 Slide 1 (51/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.4 Contrast agents
Contrast Agents
� Further enhance contrast among tissues
� Highly paramagnetic
• enhance the spin-lattice interaction
• shorten T1 time constant
• longitudinal signal restored more rapidly following excitation
� Commonly based on gadolinium 3+ ion
• seven unpaired electrons
• strongly paramagnetic
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.4 Slide 2 (52/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.4 Contrast agents
Relaxation rate R1
where
� T10 is the intrinsic tissue T1 without contrast agent
� r is relaxivity of contrast agent
� [C] is concentration of contrast agent
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.4 Slide 3 (53/141)
[ ]CrTT
R +==101
1
11
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.4 Contrast agents
Relaxivity
� Specific property of each type of contrast agent
� Varies significantly from one agent to another
� Both T1 and T2 are affected by contrast agents
• relaxivities roughly the same for both
� But, T10 >> T20
� i.e. R1 << R2
� So, for a given concentration of contrast agent, relative
effect is greater on longitudinal time constant(T1) than on
transverse (T2) time constant.
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.4 Slide 4 (54/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.4 Contrast agents
Imaging sequences in contrast-enhanced MRI
� Highlight enhancing structures by using short TR
� Most tissues do not recover sufficiently after excitation
� Only those tissues affected by contrast agent will recover
sufficiently to produce a high signal
• as they have drastically shortened T1 values
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 14.3.4 Slide 5 (55/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.4 Contrast agents
Clinical applications of contrast-enhanced MRI
� Tumour detection
• inject contrast agent into blood stream through vein in arm
• most tumours have a rich blood supply
• contrast agent diffuses from vessels into extra vascular space
• local reduction in T1
• enhance lesion signal intensity compared with surrounding tissue
• degree of signal enhancement may be used to assess tumour
perfusion
� MR angiography
� Imaging of myocardial infarction
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.4 Slide 6 (56/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.4 Contrast agents
Example of tumour detection
� Pre-contrast, no lesion is visible
� Post-contrast (intra-vascular administration), the lesion is
enhanced due to high vascularity within the lesion
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.4 Slide 7 (57/141)
IAEA
14.3 RELAXATION AND
TISSUE CONTRAST14.3.5 FREE-INDUCTION DECAY (FID)
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.5 Slide 1 (58/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.5 Free-induction decay (FID)
Free induction decay (FID)
� Occurs following excitation with an RF pulse
� Evolution of the signal in the transverse plane
• exponential decay with T2 time constant (ignoring the T2* effect)
• spins precess at Larmor frequency in laboratory frame of reference
� In homogeneous field, detected signal is
• perfectly sinusoidal
• modulated by exponential decay with T2 time constant
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.5 Slide 2 (59/141)
IAEA
14.3 RELAXATION AND TISSUE CONTRAST14.3.5 Free-induction decay (FID)
Frequency spectrum of free induction decay
� Take Fourier transform of the FID
� Result is a function, whose real component is a Lorentzian
• peak of Lorentzian function centred at the Larmor frequency
• full-width-at-half-maximum =
� Rapidly decaying signal
• broader spectrum
� Longer T2
• sharper peak in spectrum
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.3.5 Slide 3 (60/141)
21 Tπ
IAEA
14.4 MR SPECTROSCOPY14.4 MR SPECTROSCOPY
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.4 Slide 1 (61/141)
IAEA
14.4 MR SPECTROSCOPY14.4 MR spectroscopy
Proton spectroscopy
� All hydrogen nuclei (protons) have same properties
• spin number, angular moment
� In a homogeneous magnetic field expect to precess at
same frequency
� But local magnetic field differs for hydrogen nuclei of
different chemical species
• different magnetic shielding from electron clouds
� So even if a homogeneous external field
• slight difference in precessional frequency for hydrogen nuclei in
different molecules
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.4 Slide 2 (62/141)
IAEA
14.4 MR SPECTROSCOPY14.4 MR spectroscopy
Proton spectroscopy – water and fat
� e.g. water and fat, frequency difference of
• 3.35 ppm (parts-per-million)
• 215 Hz at 1.5 T
� FID for an object containing both water and fat
• sum of two decaying sinusoids
• slightly different resonant frequencies
• different T2 time constants
� Fourier transform of FID – frequency spectrum
• superposition of two Lorentzians
• separated by 3.35 ppm
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.4 Slide 3 (63/141)
IAEA
14.4 MR SPECTROSCOPY14.4 MR spectroscopy
MR spectroscopy
� Within single molecule such as fat
• hydrogen nuclei resonate at different frequencies
• protons bound to carbon atom with single bonds vs. carbon atom
with double bonds
• T2 values could also vary
� Complex frequency spectrum
• multiple peaks of different amplitudes and widths
� MR spectroscopy
• use the information in spectrum to determine the chemical and
structural properties of molecules
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.4 Slide 4 (64/141)
IAEA
14.4 MR SPECTROSCOPY14.4 MR spectroscopy
Chemical shift
� Separations between resonant peaks of different protons
proportional to external magnetic field strength
• high field systems advantageous for MR spectroscopy
• higher spectral resolution
� Frequency shift due to different electronic environments
• known as chemical shift
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 14.4 Slide 5 (65/141)
IAEA
14.4 MR SPECTROSCOPY14.4 MR spectroscopy
Clinical applications of MR spectroscopy
� Monitor biochemical changes in tumours, stroke, metabolic
disorders
� Commonly used nuclei
• proton, phosphorous, carbon
� Higher peak in spectrum associated with higher
concentration of a compound or molecule
� E.g. in proton spectroscopy
• increased ratio of choline:creatine may indicate malignant disease
• high lactate levels may indicate cell death and tissue necrosis
� More in Chapter 15
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.4 Slide 6 (66/141)
IAEA
14.5 SPATIAL ENCODING AND
BASIC PULSE SEQUENCES14.5
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5 Slide 1 (67/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5
Spatial encoding and basic pulse sequences
� 14.5.1 Slice-selection
� 14.5.2 Frequency and phase encoding
� 14.5.3 Field-of-view and spatial resolution
� 14.5.4 Gradient echo imaging
� 14.5.5 Spin echo imaging
� 14.5.6 Multi-slice imaging
� 14.5.7 Three-dimensional imaging
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5 Slide 2 (68/141)
IAEA
14.5 SPATIAL ENCODING AND
BASIC PULSE SEQUENCES14.5.1 SLICE-SELECTION
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.1 Slide 1 (69/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.1 Slice-selection
Slice selection
� In previous discussion RF pulse caused entire imaging
volume to be excited
� In most applications though, an image of a thin slice of
object is desired
� Achieve thin slice by
• applying a linear magnetic field gradient across the object during
the RF pulse
• shaping the RF pulse to give desired slice profile
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.1 Slide 2 (70/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.1 Slice-selection
Slice profile
� RF pulse applied simultaneously with linear magnetic field
gradient
� Excited slice profile in direction of gradient resembles
Fourier transform of the shape of the RF pulse
• most accurate for small flip angles (< 30°)
• acceptable for 90° excitation
• deviates at larger flip angles, but often assumed to hold
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.1 Slide 3 (71/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.1 Slice-selection
Slice profile
� To get rectangular slice profile
� Use sinc-shaped RF pulse
� The dotted box (left) indicates how the sinc pulse is often
truncated to limit its width, and the ends may be tapered to
create a smooth transition region in the slice profile
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.1 Slide 4 (72/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.1 Slice-selection
Shaping the excitation pulse
� Sinc pulse is infinitely long, so is truncated in practice
• usually include only one or two lobes on either side of main lobe
� To taper to zero at the ends, apply low pass smoothing
• Hanning or Hamming filter
� Without a low pass filter get abrupt truncations
• can lead to Gibbs ringing
• and excitation outside the desired slice
� Gaussian RF pulse shape could be used too
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.1 Slide 5 (73/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.1 Slice-selection
Slice thickness
� The thickness of the excited slice ∆z is a function of
• bandwidth of the sinc pulse, BW
• amplitude of slice selection gradient, Gz
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.1 Slide 6 (74/141)
zG
BWz
γ=∆
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.1 Slice-selection
Slice selection gradient
� In MRI have three sets of linear gradient coils
• used for spatial encoding
• Gx, Gy and Gz
• provide linear gradients long the x, y and z axes respectively
� Slice selection gradient may be any one of these
� Or can obtain a slice plane in any spatial orientation by
using more than one of the gradients
� Excited plane is perpendicular to the direction of the net
gradient
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.1 Slide 7 (75/141)
IAEA
14.5 SPATIAL ENCODING AND
BASIC PULSE SEQUENCES14.5.2 FREQUENCY AND PHASE ENCODING
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 1 (76/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.2 Frequency and phase encoding
Spatial encoding within the selected slice
� After exciting desired slice
• use linear gradients along the two perpendicular in-plane directions
• Gx and Gy for an axial slice plane from Gz gradient
� Linear gradient field will cause spins at different positions
along direction of gradient
• to precess at different frequencies
� Frequency of precession varies linearly with position
where Gx is the applied magnetic field gradient along x axis and ωx
is the resonance frequency of the spins at position x
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 2 (77/141)
xGxx γω =
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.2 Frequency and phase encoding
Rotating frame of reference
� The precession frequency in the equation is relative to the
rotating frame
• the additive frequency term due to the static B0 field is not included
� Note that this also applies to subsequent equations
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 3 (78/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.2 Frequency and phase encoding
Frequency-encoding
� During MRI data acquisition
• spatial gradient applied
• along x-axis by convention
• spins resonate at linearly varying frequencies
� This is frequency-encoding
� Applied gradient is called readout gradient
• since data points are acquired while gradient is applied
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 4 (79/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.2 Frequency and phase encoding
Phase-encoding
� Arises from spatial gradient applied along y-axis by
convention
� Applied gradient is called phase-encoding gradient
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 5 (80/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.2 Frequency and phase encoding
The detected signal, frequency encoding only
� The detected signal S(t) is the sum of all the excited spins
in the imaging volume
� Each spin is resonating at a frequency corresponding with
its position along gradient direction
where ρ(x) is the spin density
• T1 and T2 dependence have been ignored for simplicity
� Term in the exponent is a phase term
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 6 (81/141)
( ) ( ) ( ) dxexdxextSxtGjtj xx ∫∫ == γω ρρ
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.2 Frequency and phase encoding
The phase term in the detected signal
� The phase term represents the relative phase
accumulated
• in the rotating frame
• due to the frequency encoding gradient
� Typically, readout gradient is constant
� More generally, readout gradient may be time varying
• so the accumulated phase term must be replaced by a time
integral of the readout gradient
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 7 (82/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.2 Frequency and phase encoding
The phase term in the detected signal for a time varying
readout gradient
� Define
� Then
� i.e. position variable x and the variable kx are a Fourier pair
� kx represents spatial frequency, in “k-space”
� Thus image space and k-space are a Fourier pair
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 8 (83/141)
( ) ( ) ( ) dxexkStSxjk
xx∫== ρ
( ) dttGk xx ⋅= ∫γ
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.2 Frequency and phase encoding
Extend to 2D, i.e. both frequency and phase encoding
� Define
� Then
� Image space and k-space are a Fourier pair
� Once a sufficient amount of data are acquired in k-
space
� Can obtain imaging object by Fourier
transformation
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 9 (84/141)
( ) dttGk yy ⋅= ∫γ
( ) ( ) ( )dxdyeyxkkS
ykxkj
yx
yx∫+= ,, ρ
( ) dttGk xx ⋅= ∫γ
( )yx kkS ,
( )yx,ρ
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.2 Frequency and phase encoding
Data acquisition in 2D k-space
� Desired k-space data, , acquired by navigating
through k-space
� Use Gx and Gy gradients, where
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 10 (85/141)
( )yx kkS ,
( ) dttGk xx ⋅= ∫γ
( ) dttGk yy ⋅= ∫γ
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.2 Frequency and phase encoding
Data acquisition in 2D k-space
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 11 (86/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.2 Frequency and phase encoding
Data acquisition strategies in 2D k-space
� Can acquire data to fill k-space after a single excitation
• but signal may be small during data acquisition due to T2 or T2*
decay
• leading to blurring or other image artefacts
� Instead use multiple TRs
� Acquire data for a single line of k-space following each
excitation
� Details of typical data acquisition strategies in sections
14.5.4, 14.5.5, 14.5.6
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.2 Slide 12 (69/141)
IAEA
14.5 SPATIAL ENCODING AND
BASIC PULSE SEQUENCES14.5.3 FIELD-OF-VIEW AND SPATIAL RESOLUTION
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.3 Slide 1 (88/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.3 Field-of-view and spatial resolution
Field of view (FOV)
� Determined by manner in which k-space is sampled
� Image space and k-space are
• Fourier pair
• have an inverse relationship
• image space, units of distance – spatial position
• k-space, units of 1/distance – spatial frequency
� Larger FOV: finer sampling in k-space
� Smaller FOV: coarser sampling in k-space
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.3 Slide 2 (89/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.3 Field-of-view and spatial resolution
Field of view (FOV)
� FOV given by inverse of the distance between adjacent
sampled k-space points
� Can be different for the two in-plane directions
where ∆kx and ∆ky are the spacing between adjacent k-
space points along kx and ky
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.3 Slide 3 (90/141)
x
xk
FOV∆
=1
y
yk
FOV∆
=1
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.3 Field-of-view and spatial resolution
Spatial resolution / pixel width
� The range of the sampled k-space region is important
� High spatial resolution requires larger region of k-space to
be sampled
� Expressed in terms of pixel widths ∆x and ∆y
where kx, max and ky, max are the maximum positions, along kx
and ky respectively, that are sampled
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 14.5.3 Slide 4 (91/141)
max,2
1
xkx =∆
max,2
1
yky =∆
IAEA
14.5 SPATIAL ENCODING AND
BASIC PULSE SEQUENCES14.5.4 GRADIENT ECHO IMAGING
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 1 (92/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Pulse sequences
� Set of instructions that control how MRI data are acquired
• application of Gx, Gy and Gz magnetic field gradients
• the RF pulse
• data acquisition
� Many available
� Some optimized e.g. for
• rapid imaging
• estimating diffusion or flow
• robustness against motion
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 2 (93/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Gradient echo pulse sequence
� One of the simplest imaging pulse sequences
• Slice-selective RF pulse which rotates the magnetization into the
transverse plane (conventionally along the z-axis)
• Frequency-encoding, or readout, gradient (x-axis)
• Phase-encoding gradient (y-axis)
� Plus
• Slice “rewinder”
• Readout pre-phase gradient
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 3 (94/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Gradient echo pulse sequence
� Following slice excitation,
phase-encoding and readout
pre-phase gradients are
applied.
� Data acquisition subsequently
follows in which one line of k-
space data is acquired.
� The phase encoding gradient
amplitude is linearly increased
from one TR to the next in
order to fill different ky lines of
k-space.
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 4 (95/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Slice “rewinder”
� Opposite polarity to slice-selection gradient amplitude
� Needed to undo the phase accumulation that occurs
during RF pulse when a slice-selection gradient present
• spins at different positions along the z-axis will accumulate
different amounts of phase during the RF pulse
� Apply rewinder gradient with area approximately half that
of slice selection gradient
• reverses undesired phase dispersion
• aligns spins along same direction in the transverse plane
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 5 (96/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Readout pre-phase gradient
� For reconstruction of an MR image, require k-space to be
filled symmetrically about its centre
• i.e. about kx = ky = 0
� Can be achieved efficiently by acquiring a full line of data
in a single readout
� Pre-phase gradient moves the k-space location to –kx, max
prior to each data acquisition
• position in k-space is the integral of the spatial gradients
• a negative Gx pre-phase gradient will move k-space position along
negative kx direction
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 6 (97/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Readout pre-phase gradient
a) The pre-phase gradient of the readout axis moves the current k-space
position to –kx, max
b) Subsequently, one complete line of data is acquired during readout
c) Also apply the various phase-encoding gradients prior to readout to
move along the ky axis
• two different phase-encoded acquisitions are shown, one positive and one negative
Continue until the desired k-space region is filled (gray box).
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 7 (98/141)
a) b) c)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Readout pre-phase gradient
� Without a pre-phase gradient
• data readout would begin at kx = 0
• to fill the other half of k-space would need second readout in
opposite direction
� Pre-phase gradient causes de-phasing of spins
• that element of de-phasing is then reversed by the readout
gradient
• spins come back into phase and result in the “echo” signal
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 8 (99/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
When to apply gradients
� One line of data acquired during each TR
� During each TR apply different phase-encoding
amplitudes to sample different lines along ky axis
� Phase-encoding, pre-phase and slice rewinders must be
applied
• after the end of the excitation pulse
• prior to beginning of readout window
� Could be applied simultaneously
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 9 (100/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Gradient amplitude and duration
� Total area under pulses is important
• since k is time integral of the gradient
� Area under the readout pre-phase gradient should be one-
half of the area of the readout gradient
• echo will occur midway through the readout window
� Achieve using
• same amplitude, but opposite polarity and half pulse width
or
• maximum possible amplitude and a shorter pulse width
• good if short TE is desired
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 10 (101/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Image reconstruction
� Acquire desired data
• e.g.
• 256 phase-encoding lines
• 256 readout points per line
� Apply 2D discrete fast Fourier transform to k-space data
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 11 (102/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
TE, echo time
� Time between peak of excitation pulse and centre of
readout window (kx=0)
• above a certain minimum, can select value of TE to achieve
desired image contrast
� During TE
• transverse magnetization decays with instrinsic T2 relaxation time
constant
• also loss of phase coherence among the spins from field
inhomogeneity
• result is T2* decay of signal
• the gradient echo image is weighted by the factor exp(-TE/T2*)
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 12 (103/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
T1-weighted imaging
� T1 relaxation takes place following each excitation and
longitudinal magnetization recovers
� If TR short
• longitudinal magnetization for tissue with relatively long T1s
remains low (insufficient time for recovery)
• lower signal in tissues with long T1s
• longitudinal magnetization greater for tissue with shorter T1s
• higher signal in tissues with short T1s
� T1-weighted images
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 13 (104/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
T1-weighted imaging
� Image contrast results from difference in T1 among the
tissues
� TE kept to a minimum in T1-weighted imaging to minimise
contrast from T2 or T2*
� Both TR and TE are short
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 14 (105/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Fast imaging
� Applications such as dynamic imaging
� Multiple images are acquired rapidly to detect changes in
the images over time
� TR needs to be short
• but short TR reduces time available for magnetization to recover
� So, use small excitation angles (e.g., 5o – 30o) so that
acquired signal is maximised
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 15 (106/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Ernst angle
� θE
� Flip angle that yields highest signal intensity for
• given TR
• known or approximate tissue T1
� Obtained by solving Bloch equation
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 16 (107/141)
( )1/1cosTTR
E e−−=θ
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.4 Gradient echo imaging
Drawback of gradient echo imaging
� Sensitive to magnetic field inhomogeneities
• imperfect shimming of main B0 field
• along air tissue boundaries where susceptibility difference between
regions distorts the local magnetic field
� Leads to additional signal loss from dephasing of spins
(T2* decay) within affected voxels
� Results in reduced signal
� Or even complete loss of signal at long TE
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.4 Slide 17 (108/141)
IAEA
14.5 SPATIAL ENCODING AND
BASIC PULSE SEQUENCES14.5.5 SPIN ECHO IMAGING
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.5 Slide 1 (109/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.5 Spin echo imaging
Spin echo imaging
� Avoids reduction of signal due to magnetic field
inhomogeneities
� A second RF pulse is applied following the initial 90°
excitation pulse
• at time TE/2
• 180° flip angle
� Called a “refocusing pulse”
� Effect is to reverse the phase that a spin may have
accumulated due to the inhomogeneous field
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.5 Slide 2 (110/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.5 Spin echo imaging
After the refocusing pulse
� After 180° refocusing pulse, magnetization again
accumulates phase
� But, because of the phase reversal from the refocusing
pulse, the total phase is reduced
� At the echo time TE the phase accumulated after the
refocusing pulse cancels the phase accumulated before
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.5 Slide 3 (111/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.5 Spin echo imaging
Spin echo pulse sequence
� The spin echo sequence is
similar to the gradient echo
sequence
• phase-encoding, rewinder and pre-
phase gradients
� Plus a 180° refocusing pulse
• to refocus any spin that may have
dephased due to magnetic field
inhomogeneities
� The refocusing pulse reverses
any accumulated phase• so polarities of both the phase
encoding gradient and the pre-
phase gradient of the readout are
reversed
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.5 Slide 4 (112/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.5 Spin echo imaging
Spin echo pulse sequence in
k-space
� Effect of refocusing pulse
is
• to move current k-space
location to its conjugate
position
• i.e. to reflect the point about
the k-space origin
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.5 Slide 5 (113/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.5 Spin echo imaging
Acquired signal at TE
� Spin echo imaging
• evolves with the intrinsic T2 time constant
• not affected by inhomogeneous fields
� Gradient echo imaging
• evolves with T2*
• is affected by inhomogeneous fields
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.5 Slide 6 (114/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.5 Spin echo imaging
Fast imaging
� In gradient echo, use short TR and small angle excitation
� In spin echo, short TR not recommended due to rapid
reduction of available magnetization
• 180° refocusing pulse inverts any positive longitudinal
magnetization to –z direction
• a short TR does not allow sufficient time for recovery
� Instead, reduce scan time by acquiring multiple lines of
data following a single excitation pulse
� “Fast spin echo” or “turbo spin echo”
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.5 Slide 7 (115/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.5 Spin echo imaging
“Fast spin echo” or “turbo spin echo”
� For each excitation pulse
• apply multiple refocusing pulses
• acquire multiple lines of data
� Typically 16, 32 or more lines of data per TR
� Single shot acquisitions possible
• e.g. 128 refocusing pulses and 128 readout lines following a single
excitation pulse
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.5 Slide 8 (116/141)
IAEA
14.5 SPATIAL ENCODING AND
BASIC PULSE SEQUENCES14.5.6 MULTI-SLICE IMAGING
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.6 Slide 1 (117/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.6 Multi-slice imaging
Two approaches to get more than one slice
� Acquire multiple 2D slices by
• repeating the single slice strategy
• at shifted slice positions
� Use 3D imaging
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.6 Slide 2 (118/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.6 Multi-slice imaging
Multiple slice technique
� All gradient waveforms, including slice selection gradients,
remain unchanged
� Vary location of each slice by modulating frequency of the
excitation RF pulse
• modify the carrier frequency of the RF pulse
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.6 Slide 3 (119/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.6 Multi-slice imaging
Multiple slice technique
� Centre location of excited slice with respect to scanner
isocentre, zslice, given by
where ∆f is the offset frequency of the RF pulse relative to
resonance frequency at isocentre and Gsl is the slice
selection gradient amplitude
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.6 Slide 4 (120/141)
sl
sliceG
fz
γ∆
=
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.6 Multi-slice imaging
Multiple slice technique
� Typically acquire multiple 2D slices in an interleaved
fashion
� Do not need to wait until the end of a TR following data
acquisition from one slice before acquiring data from other
slices
• there is “dead time” between end of data acquisition and next
excitation pulse for a given slice
• excite a different slice, and acquire data, during “dead time” of the
previous slice
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.6 Slide 5 (121/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.6 Multi-slice imaging
Interleaved slice acquisition
� Three slices S1, S2 and S3 are acquired
� Each box can represent any imaging sequence
� Effective TR for each slice is period between acquisitions
of the same slice
• corresponds to the time between excitations for any given spin
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.6 Slide 6 (122/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.6 Multi-slice imaging
Interleaved slice acquisition
� Total scan time can be equal to scan time for a single slice
if the dead time is sufficiently long
� Otherwise, repeat scan until all remaining slices are
acquired
� Following data acquisition, images are reconstructed by
applying separate 2D Fourier transforms for each slice
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.6 Slide 7 (123/141)
IAEA
14.5 SPATIAL ENCODING AND
BASIC PULSE SEQUENCES14.5.7 THREE-DIMENSIONAL IMAGING
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.7 Slide 1 (124/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.7 Three-dimensional imaging
3D imaging
� Alternative approach to acquisition of multiple slices
� Excite a single large volume (instead of a single slice)
� Add phase encoding along slice direction
� Now have phase encoding in
• in-plane, and
• through-plane (slice) directions
� Just as in-slice spatial information is encoded using
phase-encoding along y-axis
� Slice information is encoded using phase-encoding along
z-direction
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.7 Slide 2 (125/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.7 Three-dimensional imaging
3D gradient echo sequence
� Similar to 2D sequence
� But has additional phase
encoding along the slice-
encoding direction, Gz
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.7 Slide 3 (126/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.7 Three-dimensional imaging
3D imaging
� After data acquisition, images are obtained by 3D Fourier
transform of the acquired k-space data
� Data for all slices acquired simultaneously
• in 2D scheme each data acquisition window receives data from
only one slice
� Scan time in 3D is greater than that for 2D (for a single
slice with the same TR) by a factor equal to the number of
slices
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.7 Slide 4 (127/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.7 Three-dimensional imaging
Advantages/Disadvantages of 3D imaging
� 3D imaging preferred when multiple contiguous slices
needed
• because slices are contiguous
� 3D imaging preferred if thin slices required
• because excitation of a large 3D volume requires much lower
gradient amplitude than needed to excite individual thin slices
� 2D imaging preferred if arbitrary slice positions desired
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.7 Slide 5 (128/141)
IAEA
14.5 SPATIAL ENCODING AND
BASIC PULSE SEQUENCES14.5.8 MEASUREMENT OF RELAXATION TIME
CONSTANTS
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.8 Slide 1 (129/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.8 Measurement of relaxation time constants
Why measure T1 and T2?
� Determine disease status
� Track disease progress following treatment
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.8 Slide 2 (130/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.8 Measurement of relaxation time constants
Measurement of T1
� Several approaches
� One example is the inversion recovery (IR) technique
• accurate and widely used
� Apply 180° pulse
• inverts spins from equilibrium position into the –z axis
� Magnetization recovers from its inverted state towards
equilibrium magnetization
• in exponential fashion, with longitudinal time constant T1
• according to
where
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.8 Slide 3 (131/141)
( ) ( ) ( )11 //10
Tt
o
Tt
zz eMeMtM−− −+=
( ) 00 MM z −=
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.8 Measurement of relaxation time constants
Inversion recovery (IR)
sequence
� Invert magnetization to the –z
axis by a 180° pulse
� Acquire data following an
inversion time (TI)
� Repeat for different TI values
(open circles)
� Fit data to equation to find T1
� Typically a fast imaging
technique used
• such as turbo spin echo
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 14.5.8 Slide 4 (132/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.8 Measurement of relaxation time constants
Measurement of T1
� Series of images acquired, each with different inversion
time TI
� Measure signal intensity and fit to
� Derive T1
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.8 Slide 5 (133/141)
( ) ( ) ( )11 //10
Tt
o
Tt
zz eMeMtM−− −+=
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.8 Measurement of relaxation time constants
Measurement of T1
� Any imaging sequence could be used
� Rapid sequence preferred to reduce total scan time
• e.g. turbo spin echo
� Long TR used
• ensures spins are nearly fully relaxed prior to subsequent TRs
• ideally TR
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.8 Slide 6 (134/141)
15 T×≈
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.8 Measurement of relaxation time constants
Measurement of T2
� Two approaches
� Use a spin echo sequence
1. repeatedly with different TEs
2. with one excitation and multiple refocusing pulses
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.8 Slide 7 (135/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.8 Measurement of relaxation time constants
Measurement of T2, first approach
� Spin echo sequence
• expect signal to decay exponentially with T2 time constant
� Repeat using increasing echo times
� Fit measured signal intensities to decaying exponential
� Could use just two different echo times, but four or five
more typical
� TR >> T1
• ensures magnetization returns to equilibrium state
• magnetization at start of each excitation is identical and not a
function of TR or TE
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.8 Slide 8 (136/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.8 Measurement of relaxation time constants
Measurement of T2, second approach
� Spin echo sequence
• expect signal to decay exponentially with T2 time constant
� Use one excitation and multiple refocusing pulses
� e.g. 90° excitation followed by
• 6 refocusing pulses (180°)
• 6 data acquisition periods
• so get set of images with different TEs within the same TR
� Sequence similar to fast spin echo
• except acquired echoes used to reconstruct different images
� Method is efficient
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.8 Slide 9 (137/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.8 Measurement of relaxation time constants
Measurement of T2 - Inaccuracies in the 180° refocusing
pulses
� Do not affect T2 measurement accuracy in single echo spin
echo sequences
• signal at different echo times is equally affected by the
inaccuracies, so signal amplitudes are scaled in the same way for
all echoes
� Do affect T2 measurement accuracy in multiple echo spin
echo sequences
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.8 Slide 10 (138/141)
IAEA
14.5 SPATIAL ENCODING AND BASIC PULSE
SEQUENCES14.5.8 Measurement of relaxation time constants
Measurement of T2*
� Similar to measurement of T2, but use gradient echo
sequence in lieu of spin echo
• multiple gradient echoes follow a single excitation pulse
• each echo at a different TE
� No concerns about accuracy of RF pulse affecting
accuracy of measurement
• as no refocusing pulses used in gradient echo
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14.5.8 Slide 11 (139/141)
IAEA
14. BIBLIOGRAPHY14.
Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14. Bibliography Slide 1 (140/141)
IAEA Diagnostic Radiology Physics: A Handbook for Teachers and Students - 14. Bibliography Slide 2 (141/141)
14. BIBLIOGRAPHY14.
� BERNSTEIN, M.A., KING, K.F., ZHOU, X.J., Handbook of MRI pulse
sequences, Elsevier Inc, Amsterdam (2004).
� BUSHBERG, J.T., SEIBERT, J.A., LEIDHOLDT, E.M.J., BOONE, J.M., The
Essential Physics of Medical Imaging, 2nd Ed edn, Williams and Wilkins.
(2002).
� HAACKE, E.M., BROWN, R.W., THOMPSON, M.R., VENKATESAN, R.,
Magnetic resonance imaging: Physical principles and sequence design, John
Wiley & Sons, Inc, New York (1999).
� PAULY, J., NISHIMURA, D., MACOVSKI, A., A k-space analysis of small-tip-
angle excitation, J. Magn. Reson. 81 43-56 (1989).
� SPRAWLS, P., Magnetic Resonance Imaging: Principles, Methods, and
Techniques, 2nd edn, Medical Physics Publishing, Madison, Wisconsin
(2000).
� STARK, D.D., BRADLEY, W.G., Magnetic resonance imaging, 2nd edn,
Mosby-Year Book, St. Louis (1992).