Upload
john
View
464
Download
2
Embed Size (px)
Citation preview
CH2205 - Electroanalytical Techniques
Electrochemical Cells
Typically a cell is galvanic if it produces electrical energy or electrolytic if it consumes electrical
energy.
A cell is made of two conductors, or electrodes, each immersed in a suitable electrolyte. For
electricity to flow it is necessary that the electrodes are connected externally by a metal conductor
and that the two electrolyte solutions are in contact to permit movement of ions from one solution
to the other.
Two common cell set-ups are;
Electricity is conducted in various ways in the cell. There is a migration of cations and anions as
electrons move from one electrode to the other, at the electrode surfaces an oxidation or reduction
process provides a mechanism by which the ionic conduction of the solution is coupled with the
electron conduction of the electrodes, providing a complete circuit for the flow of electricity.
By definition;
An anode is where oxidation occurs.
A cathode is where reduction occurs.
The liquid junction (frit or salt bridge) is used to avoid the direct reaction of the components of the
two half-cells, there would be a direct deposition of one metal onto the other. There is a small
potential, the junction potential, arises at the interface between two electrolytic solutions of
differing compositions.
Schematic representations of cells are often used to simplify the diagrams using the following
notation;
M(s) І M2+
(aq)||Me2+(aq) І Me(s)
Where each line represents a phase boundary at which a potential may develop, two lines (||)
represents a salt bridge and a long single line (|) represents a frit.
Cell Potentials
Using the example cell;
2AgCl(s)+H2(g)⇋2Ag(s)+2Cl-+2H+
The equilibrium constant for this reaction is;
Salt Bridge
V
Frit
V
K =[H+]2[Cl−]2
p(H2)
To consider the value at any point during the reaction (using instantaneous concentration values)
using Q;
𝑄 =[H+]a
2[Cl−]a2
p(H2)a
The change of the free energy in the cell is given by;
∆G = ∆Geq − ∆Ga
∆G = −RTlnK and ∆Ga = −RTlnQ
∴ ∆G = RTlnQ − RTlnK
The magnitude of free energy for the system depends on how far away the system is from
equilibrium.
Since;
∆G = −nFEcell
−nFEcell = RTlnQ − RTlnK
Ecell =−RTlnQ
nF+
RT
nFlnK
RT
nFlnK = Ecell
o
∴ Ecell = Eo −RT
nFlnQ
Electrode Potentials
A cell is made of two half-cell reactions, conventionally written as reductions (a species gaining
electrons).
To obtain the cell, the second is subtracted from the first to cancel out electrons. To obtain the cell
potential, strictly speaking, it must be done via free energy of the system, however if n is the same
for each half-cell;
Ecell=Ecathode-Eanode
Calculating Half-Cell Potentials
It is rare that a cell will be ‘standard’ and its potential will often have to be calculated via the Nernst
equation.
For the reaction;
pP+qQ+ne-⇋rR+sS
Then;
E = E0 −RT
nFln
[R]r[S]s
[P]p[Q]q
The standard potential is often defined as the electrode potential of a half-cell when all reactants
and products exist at unit activity. It is a physical constant that gives a quantitative description of the
relative driving force of a half-cell reaction.
Potentiometry
Potentiometry has been used for a long time in detecting the end-point during titrimetric analysis.
More recently it has been used for the quantitative analysis of specific ions in solution.
The method simply involves the determination of the potential between two half-cells, if one half-
cell is constant then the potential of the other is effectively being measured and this potential is
related to the ion concentration through the Nernst equation.
The equipment required is similar to that of the above cells, one electrode is replaced with a
standard reference electrode and the other an indicator electrode.
Reference Electrodes
The reference electrode should have a known potential that is constant and insensitive to the
composition of the solution being studied. It is, however, important to remember that the potential
will vary with temperature.
-Calomel electrodes
Represented as follows;
||Hg2Cl2(sat), KCl(xM) ІHg
Where x is the molar concentration of KCl, the electrode reaction is;
Hg2Cl2(s)+2e-⇋2Hg(l)+2Cl-
The potential of this electrode will vary with [Cl-], hence the concentration of KCl must be quoted.
The saturated calomel electrode (SCE) is the most commonly used calomel electrode due to the ease
with which it can be prepared.
At 25oC the SCE has a potential of 0.244V
-Silver/silver chloride electrodes
Analogous to calomel the cell is written as;
||AgCl(s)(sat),KCl(xM)ІAg
The electrode reaction is;
AgCl(s)+e-⇋Ag(s)+Cl-
The potential of the saturated Ag/AgCl at 25OC is 0.199V.
Precaution in the uses of Reference Electrodes
The level of the internal solution must always be above that of the solution in which it is immersed.
This prevents contamination of the electrode or plugging of the junction by reaction of analytes with
Ag(I) or Hg(I) salts. Junction plugging is the most common cause of erratic cell behaviour.
With the level of the electrode solution above the analyte solution some sample contamination may
occur. This is normally so slight it is of little concern, however for the determination of Cl-, K+ or Ag+
the precaution of having a second salt bridge may be necessary. This bridge should contain a high
concentration of non-interfering electrolyte.
Indicator Electrodes
-Metallic electrodes of the first kind
These are in direct equilibrium with the cation derived from the metal electrode;
e.g. Cu2++2e-⇋Cu(s)
Therefore E=E0Cu – 0.0591/2 log10 (1/[Cu2+])
Thus the copper electrode provides a direct measurement of copper concentration in solution.
Similarly Hg, Ag, Cd, Zn and Pb may be used, however not all metals behave reversibly and hence not
all metals are suitable.
-Metallic electrodes of the second kind
A metal electrode can be made responsive to the concentration of anion with which it forms a
precipitate or a stable complex ion.
e.g. AgCl(s)+e-⇋Ag(s)+Cl-
Therefore E=E0AgCl – 0.0591/2 log10 [Cl-]
Thus a convenient way of preparing a Cl- sensitive electrode is to make a pure silver wire the anode
in an electrolytic cell containing KCl. The wire is coated with an adherent AgCl deposit which will
rapidly equilibrate with the surface of the solution in which it is immersed. As the AgCl solubility is
low an electrode may be used for numerous measurements.
An important electrode is that used for the measurement of [edta4-] based on the response of a Hg
electrode when a small amount of Hg(edta)2- is present.
[edta= ]
The half-reaction is;
Hg(edta)2-+2e-⇋Hg(l)+edta4-
Thus E=E0 – 0.0591/2 log ([edta4-]/[Hg(edta)2-])
Hg(edta)2- is very stable and its concentration will not change, giving rise to a new constant.
K=E0+0.0591/2 log10[Hg(edta)2-]
-Metallic electrodes of the third kind
A metal electrode can, under certain circumstances, be made to respond to a different cation. An
example is the determination of the calcium concentration using a mercury electrode.
A small amount of Hg(edta)2- complex is introduced into the solution, hence the potential is given by;
E=K – 0.0591/2 log10[edta4-]
In addition if a small volume of Ca(edta)2- is introduced a new equilibrium is established;
Ca(edta)2-⇋Ca2++edta4-
Kf = Ca2+ [edta4−]
Ca edta 2−
∴ edta4− = Kf
Ca edta 2−
Ca2+
Substituting this into the previous equation for E;
E = K − 0.0591
2log10Kf
Ca edta 2−
Ca2+
𝐸 = 𝐾 −0.0591
2log10Kf Ca edta 2− −
0.0591
2log10(
1
Ca2+ )
Hence if a constant concentration of Ca(edta)2- is used;
E = K′ −0.0591
2log10 1/[Ca2+]
Membrane Electrodes
These may be broadly classified into ion selective and molecular selective electrodes.
-Ion selective electrodes (ISE’s)
Properties;
Minimal solubility – the solubility in analyte solutions should approach zero. Typically membranes
are formed from large molecules or molecular aggregates (e.g. glasses or polymeric resins). Inorganic
compounds of low solubility (e.g. silver halides) can be converted into membranes.
Electrical conductivity – the membrane must have some electrical conductivity (generally via the
migration of ions)
Selective reactivity with analyte – membrane must be able to selectively bind the analyte of interest.
Typical bindings include ion exchange and complexation.
Principles and design;
Cell design – the cell consists of a reference electrode and a membrane electrode (both of which are
dipped into the analysed solution). These two electrodes are connected to a voltmeter. The
membrane electrode consists of an active membrane sealed at one end of a tube, the tube holds a
standard solution of the species to be analysed, X+, and immersed in this standard solution is
reference electrode 2.
The standard solution within the membrane electrode serves a dual purpose. It bathes the internal
surface of the membrane with a fixed concentration of X+ and it also serves as part of the reference
electrode.
For example the internal solution of a calcium selective electrode is typically CaCl2 that is saturated
with AgCl. When a silver wire is dipped into this solution a Ag/AgCl reference electrode is formed. In
addition the Ca2+ ions expose the inner membrane to a constant concentration of analyte.
Electrical conduction – unlike metallic electrodes where conduction occurs via redox processes,
conduction in membrane electrodes occurs via ion transfer. This is made possible by the ionic nature
of the membrane or a species within the membrane. A common ion exchanger for IS membranes is a
silicate glass which consists of a 3D infinite network of oxygen atoms held by silicon atoms. If the
glass has a negative charge this is balanced out by cations residing in open regions of the structure.
Singly charged cations within the glass are mobile enough to provide a mechanism by which
electricity may be carried through the glass.
Conduction across the two glass-solution interfaces of M+ (e.g. H+) will constitute the current. In the
absence of a current the equilibria, on the two glass surfaces, is determined by the pH of the internal
and external solutions. When these equilibria are different the surface where greater dissociation
has occurred will be negative with respect to the other surface. Thus a potential develops (the
boundary potential) whose magnitude depends on the difference in pH on the two sides of the
membrane, this provides the basis for potentiometric pH measurement.
Non-Crystalline Ion Selective Membrane Electrodes
-Glass electrodes
A glass electrode is made by sealing a thin, pH sensitive glass tip to the end of a piece of heavy
walled glass tubing. The resulting bulb is filled with HCl (0.1M) which is saturated in AgCl. A silver
wire is immersed in the solution to form reference electrode 2.
Composition of glass membrane – originally Corning 015 glass (22%Na2O, 6%CaO, 72%SiO2) was
used, this glass shows strong specificity towards H+ upto pH 9, above this value it will start to
respond to Na+.
Hygroscopity of glass membrane – the surface of the glass membrane must be hydrated in order to
have pH activity. Non-hygroscopic glasses, such as a Pyrex and quartz show no pH function.
Dehydrated Corning 015 glass shows little activity but its activity is restored after a few hours
standing in water.
The difference in pH on either side of the glass gives rise to the reading (since the pH inside is fixed
and only the outside varies, it forms a useable electrode).
-Liquid membrane electrodes
These are formed from immiscible liquids that selectively bond certain ions. Membranes of this type
are important as they permit the direct potentiometric determination of several polyvalent cations
and certain singly charged anions and cations.
Early liquid membranes were prepared from immiscible ion exchangers, which were retained in a
porous inert solid support. A porous hydrophobic plastic disk serves to hold the organic layer
Ag
AgCl
HCl
Inner surface of glass;
Binding Site+H+⇋BSH+
Outer surface of glass;
Binding Site+H+⇋BSH+
between the two aqueous layers and a wick action caused the pores of the disk to be filled with
organic liquid contained in an outer reservoir;
Again a potential is set-up, the porous material has a known MXn concentration above binding to
Ca2+ and a varying MXn concentration in the solution outside.
For divalent cation determinations, the inner tube contains a standard solution of MCl2 where M is
the ion to be determined. The solution is also saturated with AgCl to allow the formation of an
Ag/AgCl reference electrode.
As an alternative to the porous disk it is possible to immobilise the ion exchangers in rigid PVC
membranes. Currently most liquid membrane electrodes are of this type.
The active compounds in membranes are of three types;
-cation exchanger; ammonium salts
-anion exchanger; sulphate salts
-macrocyclic compound which complexes a specific ion
One of the most important liquid membrane electrodes is selective towards Ca2+ (in neutral media).
The active species is a cation exchanger consisting of an aliphatic diester of phosphoric acid
dissolved in a polar solvent. The diester contains a single acidic proton and two molecules of it react
with Ca2+ to form a complex in which the two molecule form donor bonds to the Ca2+.
Crystalline Ion Selective Membrane Electrodes
These may be single crystal membranes of polycrystalline heterogeneous membranes. Typically
single crystal homogeneous membranes have better reproducibility, selectivity, linearity of response
and have longer lifetimes.
-Single crystal membranes
Most ionic crystals are insulators or do not have sufficient electrical conductivity to serve as
membrane electrodes. Those that conduct are characterised by having a small ion that is mobile in
the solid phase, for example fluorides of rare earths, Ag+ in silver halides or sulphides and copper I in
Cu2S. Conduction typically occurs by ions jumping to holes (defects) within the crystal lattice. As
these holes are very size specific, these single crystal membranes exhibit excellent specificity.
-The fluoride electrode
LaF3 is an ideal compound for the preparation of membrane electrodes for F- determination.
Membranes are prepared by cutting disks from a single crystal of the compound doped with
Ag
AgCl
Liquid ion exchanger
Porous, plastic hydrophobic
disk
MXn
i.e. CaCl2
europium difluoride to create defects or ‘holes’. Again the mechanism by which a potential is
developed is analogous to the glass electrode.
The magnitude of the charge is dependent on the F- concentration in solution. Thus the side of the
membrane with the lowest F- concentration is positive with respect to the other surface. It is this
difference which allows the measurement in the difference of concentration between the two
solutions.
Most commercial electrodes are rugged, may be used between 0-80oC and give a linear response
down to 20ppb, the only ion that interferes with the electrode is the hydroxide ion and
measurements above a pH of 8 are problematic.
Molecular Selective Electrodes
-Gas sensing probes
A gas sensing probe is not an electrode but a cell containing both ion specific and reference
electrodes. A thin gas permeable membrane allows the gas to pass into the electrode and so be
analysed.
Gas effuses into the pores of the membrane and rapidly establishes an equilibrium. Within the pores
it is also in equilibrium with the internal solution of the electrode.
CO2 ⇋ CO2 ⇋ CO2
External⇋Membrane⇋Internal
Here another equibrium is established that causes the internal pH to change.
CO2+H2O⇋H++HCO3-
The glass electrode immersed in the internal solution detects this change.
The concentration of H+ will be equal to the rate constant times concentration of CO2 divided by the
concentration of HCO3-. If the concentration of HCO3
- is made relatively high (so that it doesn’t
change significantly);
K’=[H+]/[CO2(aq)]
Hence [H+]=K’[CO2(aq)]
Ag/AgCl2
NaCl,NaHCO3
Ag/AgCl2
HCl
pH electrode
Thin glass
Glass permeable membrane
Thus the pH of the internal solution (and the potential across the glass membrane) to the amount of
dissolved CO2 in the sample can be related.
Enzyme Electrodes
By combining the specificity of an ISE and the selectivity of enzyme catalysed reactions we may
obtain apparatus for the determination of compounds of biological and biochemical interest.
To achieve this the enzymes must be immobilised on the surface of the electrode and ideally should
be long-lived and useable for several measurements.
An example is the measurement of Blood Urea Nitrogen (BUN), a routine clinical test. In neutral
media, urea is hydrolysed in the presence of the enzyme urease;
(NH2)2CO + 2H2O + H+ ⇋ 2NH4+ + HCO3
-
The products may then be determined via an ammonium selective electrode. Other electrodes exist
for the determination of glucose, lactose, sucrose, galactose, cholesterol and insecticides.
Applications of Ionic/Molecular Selective Electrodes
-Calibration curves for concentration measurement
An obvious method of correcting potentiometric measurements to give measurements in
concentration is to make an empirical calibration curve (Voltage against log[X]). For this approach to
be a success, it is desirable that the matrix of the standards is as similar as possible to those of the
samples. Sometimes it is helpful to swamp out both standards and samples with an excess of inert
electrolyte. Under these circumstances the additional electrolyte in the sample is negligible. A
commercially available solution, Total Ionic Strength Adjusting Buffer (TISAB), is used for this
purpose.
-Activity vs. concentration
Electrode response is in fact related to activity not concentration. At higher concentrations activity
‘falls off’
-Standard addition
While typically used for colorimetric, atomic absorption and atomic emission spectroscopy analysis
via standard addition in potentiometry is also possible. Standard addition works by measuring the E
value of the unknown solution (with a known volume), adding a small volume of standard solution
(with a concentration much higher than that of the analyte), then measuring the change in E.
Assuming a linear change in log[M+] vs. E the original concentration can then be calculated by;
For the original solution;
E=Eo + 59.1/n log[M+]
For the solution with added standard solution;
E=Eo + 59.1/n log (v1/v2 [M+] + v3/v2 x [Standard])
where v1 is the original volume of solution, v2 is the volume plus the volume of standard and v3 is the
volume of added standard solution.
By subtracting the first equation from the second the concentration of M+ can be calculated.
Voltammetry
Current-Voltage Measurements at Micro-Electrodes
-Micro-electrodes; typically a micro-electrode has a surface area of 1-10mm2, combined with the
fact the typical concentration of an active species is low (<10-4M), the current normally observed at
the electrode is quite small (μA scale). As the microelectrode is small with respect to the volume of
the solution analysed, the quantity of species oxidised or reduced is negligible to the total quantity
and so during the course of analysis the concentration of the species of interest does not change.
Voltammetry is defined as the measurement of a current at a microelectrode as a function of the
potential (vs. a reference electrode) of the micro-electrode.
The measurement is normally carried out in an electrolysis cell containing the analyte solution. A
potential, E, is applied to a microelectrode called the working electrode. This potential is regulated
very precisely and accurately versus a reference electrode. A current flows between the working
electrode and an auxiliary electrode.
The Current-Voltage Curve
The output from a voltammetric instrument is a form of current-voltage curve;
Conventionally, E is plotted with more negative values to the right and a cathodic current
corresponding to a reduction is plotted with increasing values upward. The general name for such a
plot is a voltammogram.
Voltammetry has many forms, the most well known being polarography, another useful analytical
form is anodic stripping voltammetry. Cyclic voltammetry, differential pulsed voltammetry and
square wave voltammetry are routinely used to probe the electrochemistry of newly discovered
Measurement of
Current i Reference electrode
Auxiliary (counter) electrode
Working electrode
Electrolysis cell
Measurement of
Potential
I
V
materials. In addition AC polarography, fast linear sweep voltammetry and differential pulsed anodic
stripping voltammetry are new techniques being increasingly used.
-Polarography
A branch of voltammetry in which the working electrode consists of drops of mercury issuing from
the bottom of a piece of capillary tube (a dropping mercury electrode, DME).
The Hg drop of the DME falls off after every few seconds and is replaced by a new drop. This occurs
naturally or by the help of a drop knocker, the current-voltage curve (or polarogram) is
discontinuous and has a saw-tooth shape.
This shape is explained by the behaviour of the background solution, in which only the solvent and
electrolyte are present (the electrolyte is added to prevent electrostatic forces acting on the
electroactive species in solution ensuring only diffusional processes are involved in the migration of
analyte to working electrode). Only a small residual current will be observed as a negative potential
is applied potentially due to traces of impurity in the cell. A sharp rise will occur at a voltage in which
the first species is reduced, when a second species is added an s-shaped curve is obtained. This
reaction occurs at a potential sufficiently negative to reach a decomposition potential, at which i is
clearly higher than the residual current. This current will then reach a plateau and the different
between this current and the residual current is referred to as the limiting current. The current is
limited by the rate at which the ions of interest are transported to the electrode surface.
When diffusion is the only mechanism of motion (as is preferred), the limiting current is called the
diffusion current, id.
The voltage on the polarogram where i= ½ id is characteristic of the analyte and the media it is
dissolved in. This voltage is the half-wave potential, E1/2. When the analyte forms an amalgam E1/2 is
independent from the concentration of analyte.
-Analysis using polarography
-Qualitative; the value of E1/2 is indicative of the identity of a species and id provides quantitative
information;
Tl (I) Cd(II) Zn(II)
V
I
-Quantitative; The values of id can be used to generate calibration curves for ions and so give
concentrations of species. By measuring the potential at different concentrations a calibration curve
can be formed of id vs. [A], any unknown solution can then have its id measured and read against the
curve to determine concentration.
Standard addition can again be used in polarography, this time however since the relationship is
linear and idα[M+] and so c (in y=mx+c) will equal 0, y=mx. m is equal to the change in y over the
change in x and can be calculated leading to a calculation of the concentration of the ion in question.
-Stripping voltammetry
The analyte is deposited on the surface of a mercury drop by an electrochemical reaction occurring
at a specifically selected potential. This is done for several minutes, the analyte thus concentrated on
the working electrode surface is then stripped off in a few seconds by a rapid shift in potential.
Using a mercury drop, maintained at a potential more negative than E1/2 for ~5 minutes the metal
may be deposited in a concentrated form on the surface layer of the drop. During deposition the
solution is stirred to maximise the metal that is reduced (note only a small percentage of analyte is
reduced so it is important to have control and reproducibility to get accurate results).
After preconcentration, the potential is scanned to a more positive potential at a relatively rapid rate
(25mV/s). Near E1/2 the reverse reaction occurs. As a relatively large amount of analyte has been
preconcentrated a large anodic (negative i) current is observed. It reaches a maxima and decays back
to a low background level.
Stripping voltammetry is much more sensitive than ordinary polarography, and with very long
deposition times (1 hour) analytes wit a concentration of 10-9M can be determined.
Coulometry and Electrogravimetric Determination
-Electrogravimetric
If a well defined product is deposited at the electrode, the weight of the substance may be
measured, which if carried out exhaustively constitutes an electrogravimetric analysis.
In this method the quantity of electricity needed is not measured, only the mass of the product is
important.
This form of analysis is conceptually the most fundamental and simple method of analysis known as
it simply involves the isolation and weighing of the analyte.
-Coulometry
An electrolytic method of analysis where the quantity of electricity required to carry out an
electrolysis is measured. As one mole of electrons is 96485C (a Faraday), for a given half-equation,
we can relate the number of moles of analyte to the number of coulombs required to carry out a
process.
There are two forms of Coulometry, constant-potential coulometry and constant-current
coulometry.
The general equipment required for this process is the same as with voltammetry.
In all cases, the analytical reaction will occur at the working electrode. A reference electrode is
required for the regulation of the potential. Since the solution cannot be used again this form of
analysis must be achieved in one attempt.
The Feasibility of Electrolytic Analysis
An optimum potential for an analysis may have to be determined. This is especially true when other
electroactive species are present in solution with the analyte.
For accurate results the analyte must be reduced (or oxidised) ~100% while the interfering ion must
not be reduced to any significant degree.
If this is the case the Nernst equation can be used. Consider a solution of 0.0005M Cd2+ and an
unknown molarity of Pb2+.
Given the half-equations;
Pb2++2e-⇋Pb(s) Eo=-0.368V (vs. SCE)
Cd2++2e-⇋Cd(s) Eo=-0.645V (vs. SCE)
The lead ion is more easily reduced, thus it is necessary to calculate the percentage of Pb2+
remaining in solution as Cd2+ begins to deposit.
First the potential at which cadmium ions will deposit is calculated;
E = Eo +0.0591
2log Cd2+
E = −0.645 +0.0591
2log 0.005 = −0.713V
At this potential the value of [Pb2+] is;
E = −0.368 +0.0591
2log Pb2+ = −0.713
Pb2+ = 2.11x10−12M
And so the percentage of Pb2+ remaining is;
2.11x10−12
0.005x100 = 4.22x10−8%
Thus there will be a complete separation of lead ions from cadmium ions in any electrolytic assay
performed.
Alternatively voltammetry can give an idea of how viable electrolytic analysis will be, if a plot of
current against voltage shows rises too close together the method will not be feasible, if the steps
are well separated then one can be cleanly reduced before the other, making the method feasible.
-Electrogravimetry
Several metals can be determined by this method; Cu, Cd, Ni, Co, Ag, Bi, Zn and Pb.
Normally a platinum gauze electrode in a cylindrical configuration is employed as a working
electrode. The nature of the deposit is important to the success of the method, the deposit must be
strongly adherent. It should be a fine-grained smooth deposit with a metallic lustre, indicative of a
high quality metal deposit that can withstand subsequent washing, drying and weighing operations
without loss or decomposition.
Several factors influence the quality of the deposit. Stirring rate is one, another being that
complexes of CN- or NH3 give better quality deposits. A low current density (<0.1A/cm2) is beneficial
and the optimum temperature must be experimentally determined.
Hydrogen production at the cathode must be avoided, this is achieved by the addition of anitrate ion
(a cathode depolariser) which is reduced at the cathode to a non-gaseous product;
NO3- + :OH+ + 8e- → NH4
+ + 3H2O
Nitrite ions lead to poor quality deposits and must also be avoided and removed from solution, this
can be done chemically by adding sulphamic acid;
H+ + NH2S(O)2O- + NO2- → N2(g) + HSO4
- + H2O
-Constant-Potental Coulometry
This measures the analyte by determining the quantity of electricity that is involved in reducing or
oxidising the analyte at a working electrode set a constant potential (vs. a reference electrode).
The current is not constant and decreases during the electrolysis (with time less material exists in
the solution to reach the electrode quickly enough to maintain the current). The total change is
obtained by integrating the current-time curve over the time, t, of the reaction.
C = idt
t
0
-Constant-Current Coulometry
This involves an electrolytic reaction that occurs at a constant rate. It is impossible for the analyte to
react with 100% current efficiency at a relatively high constant current. A mediator species is added
at high concentrations that can be oxidised or reduced at the working electrode. This species can
then oxidise or reduce the analyte in solution, away from the electrode.
Consider the analysis of Fe2+ by oxidation to Fe3+, to maintain the current with decreasing Fe2+
concentration a mediator is added, in this case Ce4+ in high concentration.
This reacts with the Fe2+ to form Fe3+ and Ce3+, at the end-point the Ce4+ will no longer be used up,
using an ion-selective electrode to detect the presence of Ce4+ this end-point can be recognised. The
reaction is then halted, the solution stirred and the Ce4+ given a chance to ‘find’ more Fe2+ and
continue reacting, this is repeated until the Ce4+ is no longer used up after stirring.