Page 1

Clouds and the Earthʼs Radiant Energy System"

CERES FM6 Calibration Subsystem Performance Issues

Kory Priestley - Project Scientist Audra Bullock - Chief Engineer

CERES Science Team Meeting

NASA LaRC May 7, 2013

Page 2: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 2

Clouds and the Earthʼs Radiant Energy System"

Discussion Topics

•  CERES Instrument • Description

•  Issues uncovered during FM6 ground calibration (May 2012)

• Solar Diffuser non-uniformity • SWICS Lamp instability • SWICS Reference Detector

instability •  Recovery Status •  Programmatics

• Schedule/Workplan

Page 3: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 3

Clouds and the Earthʼs Radiant Energy System"

CERES Instrument

•  Designed, manufactured and tested by TRW, Redondo Beach, CA (currently Northrop Grumman Aerospace Systems)

•  Contains three sensor assemblies with cassegrain optics and thermistor bolometer detectors •  Sensors measure thermal radiation in the near-visible through far-infrared spectral region •  Sensor channels are coaligned and mounted on a spindle that rotates about the elevation axis •  Hemispherical sampling obtained with an azimuthal axis drive system

Orbits 705 km altitude, 10:30 a.m. descending node (Terra) or 1:30 p.m. ascending node (PM-1), sun-synchronous, near-polar; 350 km altitude, 35o inclination (TRMM)

Spectral Channels

Solar Reflected Radiation (Shortwave): 0.3 - 5.0 µm Window: 8 - 12 µm , 5 – 40 µm (FM6) Total: 0.3 to > 100 µm

Swath Dimensions Limb to limb

Angular Sampling Cross-track scan and 360o azimuth biaxial scan

Spatial Resolution 20 km at nadir (10 km for TRMM, 28 km for NPP)

Mass 45 kg

Duty Cycle 100%

Power 45 W

Data Rate 10 kbps

Size 60 x 60 x 70 cm (deployed)

Design Life 6 years

Base

Pedestal

MAM Baffles

Sensors Total, Longwave, Shortwave

Azimuth Axis

Elevation Axis

Main Electronics

Alignment Cube

Page 4: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 4

Clouds and the Earthʼs Radiant Energy System"

97 98 99 00 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20 21

Missions with ERB Observations PFM FM-1,2 FM-5 FM-6 RBI Sensors:

CY:

CERES

FM-6

RBI

FM-5

Initial Studies/Reqmts Development

Sensor Fab, Assembly, Test

Sensor in Storage

Spacecraft I&T

Nominal Mission Lifetime

Operational Lifetime

TRMM (11/97)

Terra (12/99)

Aqua (5/02)

NPP (10/11)

JPSS-1 (11/16)

JPSS-2 (11/21)

FM-3,4

Enabling Climate Data Record Continuity

CERES Flight Schedule

Page 5: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 5

Clouds and the Earthʼs Radiant Energy System"

CERES FM6 Instrument Performance Issues

•  The CERES FM6 Team continues to pursue resolution of three issues:

1.  Solar Diffuser Mirror Attenuator Mosaic (MAM) scatter non-uniformity 2.  Internal Calibration Module (ICM) lamp brightening 3.  Internal Calibration Module (ICM) reference detector photodiode (PD)

response decrease

•  In May of 2012, FM6 had completed its baseline calibration with the following open issues:

•  SWICS lamp appears to have increased in brightness throughout the calibration campaign whereas the source monitor photodiode indicated a decrease in source output throughout the calibration campaign

•  Data from MAM scatter testing indicate spatial non-uniformity of >4.5% vs. 1.5% measured at component level.

•  If not resolved, CERES FM-6 will not meet its on-orbit SW channel performance accuracy requirement

Page 6: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 6

Clouds and the Earthʼs Radiant Energy System"

CERES Instrument Description

Power Converter Assembly

Elevation Control

Assembly

Sensor Module

Assembly

X

ZMain Cover

Actuator

Instrument Control

Assembly

DAA Power Conditioning

Assembly

Data Acquisition Assembly

Alignment Cube

Azimuth Control

Assembly

Pedestal Assembly

Base Plate

Main Cover

SWICS / Blackbody Assembly

Solar Presence Sensor

MAM Baf f le

Assembly

Mirror Attenuator

Mosaic (MAM)

Internal Calibration Module (ICM)

Page 7: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 7

Clouds and the Earthʼs Radiant Energy System"

Performance Resolution Approach

ICM Resolution •  Isolate Performance Problems

•  ICM Vacuum Test determined the Lamp and PD performance issues are confined to the ICM

•  ICM Diagnostic Test to further isolate performance issues

•  Select replacement flight Lamp and PD from CERES heritage or CERES FM6 parts

MAM Resolution (Complete) •  Isolate Performance Problem

•  Diamond-Turned Tooling marks have been identified as the source of MAM performance issue

•  Select replacement flight MAM from CERES heritage MAMs •  Pe-condition MAM using AO asher from GRC

•  Verify ICM performance in vacuum

•  Verify Instrument Performance

•  Conduct SAR/PSRR

Page 8: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 8

Clouds and the Earthʼs Radiant Energy System"

CERES FM6 Instrument Status

♦  CERES FM6 Calibration subsystem issues: 1.  Solar Diffuser Mirror Attenuator Mosaic (MAM)

scatter non-uniformity

2.  Internal Calibration Module (ICM) lamp brightening

3.  Internal Calibration Module (ICM) photodiode reference detector response decrease

MAM Assembly

2 MAMs installed in CERES Instrument: SW & Total

ICM

Page 9: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 9

Clouds and the Earthʼs Radiant Energy System"

CERES FM6 MAM Overview

•  Mirror Attenuator Mosaics (MAMs) diffuse and attenuate incident solar energy so that the CERES Shortwave and Total Channels are presented with repeatable, on-scale radiance scenes over the range of incident angles experienced during periodic, on-orbit, solar calibrations.

•  MAMs fabricated for FM6 are different than prior flight units –  Substrate is Diamond Turned Aluminum vs. Electroformed Nickel –  Coatings updated for improved resistance to Atomic Oxygen

•  Why did we change Fab process? –  Diamond turned aluminum substantially easier to fabricate than electroformed nickel and electroform/grinding

process has low yield ~25%.

02S00220-0039b-154A

Figure 3.1-3. CERES Mirror Attenuator Mosaic

02S00220-0039b-154A

Figure 3.1-3. CERES Mirror Attenuator Mosaic

02S00220-0039b-154A

Figure 3.1-3. CERES Mirror Attenuator MosaicCERES MAM Assembly

t = 0

t = ~6 mins

MAM Goal minimize variability in reflected solar energy

over all incident angles

Page 10: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 10

Clouds and the Earthʼs Radiant Energy System"

Uniformity Requirement ±1% over 4.5° Solar Angle

MAM Scatter Uniformity Issue (Instrument Level Testing)

0.9

0.91

0.92

0.93

0.94

0.95

0.96

0.97

0.98

0.99

1

-4 -3 -2 -1 0 1 2 3 4

Nor

mal

ized

Sen

sor

Cha

nnel

Res

pons

e

Solar Angle (Degrees)

1999 FM5 Calibration Data: MAM Scatter Uniformity

Total ChannelSW Channel

0.90

0.92

0.94

0.96

0.98

1.00

-­‐4 -­‐2 0 2 4

Norm

alized  Detector  Signal

Solar  Angle  (degrees)

SW  MAM,  Shuttered  DataSW MAM in FM6

0.80

0.84

0.88

0.92

0.96

1.00

-­‐4 -­‐2 0 2 4

Norm

alized  Detector  Signal

Solar  Angle  (degrees)

TC  MAM,  Shuttered  DataTot MAM in FM6

Two contributors identified: •  Solar Simulator non-uniformity •  Diamond-turning resulted in fringed

scatter pattern

Requirement Requirement

Requirement

Electroform MAM Diamond-turned MAM

-4 -2 0 2 4 Solar Incident Angle (deg)

-4 -2 0 2 4

-4 -2 0 2 Solar Incident Angle (deg) Solar Incident Angle (deg)

Total Shortwave

FM5 System Level Test 1.00

0.98

0.96

0.94

0.92

0.90

1.00

0.98

0.96

0.94

0.92

0.90

1.00

0.96

0.92

0.88

0.84

0.80

~5”

Page 11: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 11

Clouds and the Earthʼs Radiant Energy System"

Lesson Learned: The diamond-turning process is plagued with issues of tooling marks

Diamond-turned vs Nickel MAM Laser Illuminating Single Dimple

Diamond-turned MAM Tooling marks cause fringed scatter

Electroformed MAM Scatter is much more uniform

Page 12: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 12

Clouds and the Earthʼs Radiant Energy System"

MAM Resolution

♦ NG located 2 legacy MAMs of the FM4 vintage   Parts removed from flight status in late 90’s for surface blemishes

  Parts bagged and stored in controlled environment during interim

♦ Scatter uniformity meets requirements with margin ♦ AO coating susceptibility corrected

  LaRC & NG collaborated with GRC to fully oxidize protective coating (Complete)

♦ MAMs upscreened and ready for installation 3 weeks ahead of schedule   Coating adhesion testing (Complete)

  Component scatter test post AO exposure (Complete)

S/N E1/B3

S/N C4

Page 13: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 13

Clouds and the Earthʼs Radiant Energy System"

CERES FM6 Instrument Status

♦  CERES FM6 Calibration subsystem issues: 1.  Solar Diffuser Mirror Attenuator Mosaic (MAM) scatter

non-uniformity

2.  Internal Calibration Module (ICM) lamp brightening

3.  Internal Calibration Module (ICM) photodiode reference detector response decrease

ICM

Page 14: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 14

Clouds and the Earthʼs Radiant Energy System"

CERES Internal Calibration Module

Tungsten Filament Lamp

Reference Photodiode

Light out to Shortwave Sensor

♦  Shortwave sensor stability verified independently ♦  Shortwave sensor indicated SWICS brightening ♦  Reference photodiode indicated SWICS dimming

Instrument vacuum calibration results:

ICM contains 2 subassemblies Shortwave Internal Calibration Source (SWICS):

Lamp and focusing optics Reference Detector Folding Mirror

Internal BlackBody (IBB): Concentric grooved blackbodies Heaters and PRTs

During Calibration we view the SWICS lamp with two independent detectors

CERES SW sensor and Reference Photodiode

Page 15: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 15

Clouds and the Earthʼs Radiant Energy System"

Instability of SWICS Lamp and Photodiode (Instrument Level Testing)

2505

2515

2525

2535

2545

2555

12 20 28 36 44 52

SW

Res

pons

e to

SW

ICS

(cou

nts)

Elapsed Time Under Vacuum (Days)

FM6 SW Sensor Response to SWICS Level 2

Level 2 @ Hot Accept

Level 2 @ Cold Accept

Level 2 @ Nominal

1340

1350

1360

1370

1380

12 20 28 36 44 52

SW

ICS

Mon

itor C

ount

s

Elapsed Time Under Vacuum (Days)

2012 SWICS Monitor Response During Internal Calibrations

SWICS Monitor Response to Lamp Level 2 at Hot AcceptSWICS Monitor Response to Lamp Level 2 at Cold AcceptSWICS Monitor Response to Lamp Level 2 at Nominal

♦  ICM removed from instrument and tested under vacuum (Oct 2012, Jan 2013) ♦  Module-level test results consistent with instrument-level calibration (FM6 Instrument exonerated) ♦  Shifts with temperature are artifact of sensor and BB heater drive ground bias effect ♦  PD Data uncorrected for apparent source drift ♦  PD Response Drift Rate is a function of Instrument Temperature

The trends in the data below should be consistent or preferably flat.

Light Source : SWICS Lamp Detector : CERES SW Channel Conclusion : SWICS Lamp appears to be getting brighter

Light Source : SWICS Lamp Detector : SWICS Photodiode Reference Detector Conclusion : Photodiode Response Decreasing

Page 16: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 16

Clouds and the Earthʼs Radiant Energy System"

Instability of SWICS Lamp and Photodiode (Instrument & Module Level Testing)

Instrument Level Calibration Test Module Level PD Test

♦  ICM removed from instrument and tested under vacuum (Oct 2012, Jan 2013) ♦  Module-level test results consistent with instrument-level calibration (FM6 Instrument exonerated) ♦  PD Data uncorrected for apparent source drift ♦  PD Response Drift Rate is a function of Instrument Temperature

The trends in the data below should be consistent or preferably flat.

-1.0%

-0.8%

-0.6%

-0.4%

-0.2%

0.0%

0.2%

0.4%

0.6%

0.8%

1.0%

15 20 25 30 35 40 45

Pe

rce

nt

Ch

an

ge

Cumulative Days under Vacuum

SW - Hot Acceptance SW - Cold Acceptance SW - Nominal Acceptance PD - Hot Acceptance PD - Cold Acceptance PD - Nominal Acceptance

SW Sensor viewing SWICS Lamp

Reference PD viewing SWICS Lamp

Stability Requirement over 5 year mission

0.0

- 0.1

- 0.2

- 0.3

- 0.4

- 0.5

- 0.6

Hours in Vacuum

FM6 1DB55 Spare 1DB55

Legacy 1DB26

Different lamp, with known instability

Page 17: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 17

Clouds and the Earthʼs Radiant Energy System"

69

Fishbone Diagram: PD Response Degradation

Plausible Contributing Root Cause Probable Root Cause No Supporting Data As Root Cause

Defective PD Vacuum Sensitivity

SWICS PD Response

Degradation in Vacuum

Electrical Opto-Mechanical

Contamination

Temperature Sensitivity

PD Temperature Drift

Contamination of PD filter

Filter Bandpass Shifting

Localized outgassing

Charge buildup on PD Case

Crosstalk with new PRT circuitry

Lamp migration within holder

Responsivity instability

Compromised PD package hermeticity

Localized outgassing

Instrument Temp Sensitivity

Electrical

Contamination of PD Window

Insufficient venting within PD Assy

Damaged during Elect. Qualification

Amplifier Gain Temperature

sensitivity

Lamp Filament sagging/shifting

Structural Failure Optical Alignment

Drifting

Contamination within PD package

Optical Alignment

PD Electrical Isolation compromised

Mirror thru-hole obstruction

Optical Alignment Increasing NIR

Absorption in optical path

PD Responsivity

Optical

Stray Light

Vignetting

51

52

53

54 55

56

57

58

59

60

62

63

61

64

65

66

67

68

70

71

72

73

74

75 76

77

78

79

69 Lamp migration within holder

observed

observed

None observed Amplifier Gain

Identified path Tested negative

Test in progress

Lamp Color Temperature 81

80

Plausible by analysis

None observed in LaRC Test

Mirror Thru-hole misalignment possible contributor

Plausible by analysis

Page 18: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 18

Clouds and the Earthʼs Radiant Energy System"

12

Fishbone Diagram: Increasing SWICS Radiance

Improbable Root Cause Plausible Contributing Root Cause Probable Root Cause No Supporting Data As Root Cause

Defective Lamp Vacuum Sensitivity

SWICS Radiance Increasing with time in

Vacuum

Electrical Opto-Mechanical

Contamination

Temperature Sensitivity

Lamp thermal gradients

Lamp operating Temperature

Contaminant sublimation from optics

Bubbles in lamp potting RTV

Localized outgassing

Diffusion through glass envelope

Ground instability

Drift in lamp current supply

Filament sagging/shifting

Compromised Lamp Envelope Hermeticity

Localized outgassing

Instrument Temp Sensitivity

RCF Pressure Sensitivity

RCF Source Temp Sensitivity

Vacuum contaminant sublimation Venting

obstruction

Lamp Aging effects

Crosstalk with ICSBB heaters

Temperature sensitivity Gravity

sensitivity

Structural Failure

Optics Alignment Drifting

Lamp Storage Conditions

Optical Alignment

Tungsten plating

May also explain apparent PD response degradation

Optical

Stray Light

1

2

3

4

5 7

8

9

10

11 13

14

15

16

17

18

20 21

22

23 24

25 26

27

28

Increasing refl /fwd scatter from Lamp Assy

29

Increasing refl /fwd scatter from lamp cavity

6 Lamp Migration within holder 19

29 Potting RTV Debonding

Page 19: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 19

Clouds and the Earthʼs Radiant Energy System"

ICM & Breadboard Test Summary

1DB55 Decays Lamp (Brightens?) Noise on Reference Detector 7 day test, Jan 24-31, 2013

Repeated Instrument Behavior: 1DB55 Decays Lamp Brightens 16 day test, Oct. 8-23, 2012

ICM Baseline Test

1DB55 FM6 (43)

Lamp FM6 (50)

1DB26 Decays Lamp (Brightens?) Noise on Reference Detector 13 day test, Feb 7-19, 2013

Filterless 1DB26 Decays Lamp Brightens Reference Detectors Behaving 12 day test, Mar 1-12, 2013

Filterless 1DB55 -/Flat Lamp -/Flat Reference Detectors Behaving 12 day test, Mar 1-12, 2013

ICM Test #2

1DB55 New (94)

Lamp FM6 (50)

ICM Test #3

1DB26 Legacy

Lamp FM6 (50)

BB Test w. ICM #4

1DB55 New (94)

Lamp Non-flight

BB Test w. ICM #6

1DB55 FM6 (43)

Lamp New (16)

ICM Test #4

1DB26 Legacy

Lamp New (23)

ICM Test #5

1DB55 FM6 (43)

Lamp New (16)

ICM Test #6

1DB55 New (94)

Lamp Non-flight

BB Test w. ICM #3

1DB55 New (94)

Lamp Non-flight

BB Validation Test

1DB26 Legacy

Lamp Non-flight

BB Test w. ICM #5

1DB55 New (94)

Lamp Non-flight

Filterless 1DB55 Decays Lamp Brightens Reference Detectors Behaving 7 day test, Mar 14-20, 2013

Current Test: Filterless 1DB55(94) BB Non-flight Lamp Mar 23- April 3

1DB26 Decays Lamp (-/Flat) Noise on Reference Detector 11 day test, Jan 11-21, 2013

1DB55 Decays Lamp (Brightens?) Noise on Reference Detector 13 day test, Feb 7-19, 2013

Placed Filter 1DB55 Decays Lamp Flat Reference Detectors Behaving 7 day test, Mar 14-20, 2013

Current Test: Filterless FM6 1DB55 Spare Flight Lamp Mar 23-April 3

Page 20: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 20

Clouds and the Earthʼs Radiant Energy System"

ICM & BB Vacuum Test #7 Results

BB Test w. ICM #7

1DB55 New (77)

Lamp New (24)

ICM Test #7

Lamp New (16)

1DB55 New (94)

y  =  -­‐0.035%x  +  1430.214%  

y  =  0.010%x  -­‐  418.856%  

-­‐1.000%  

-­‐0.800%  

-­‐0.600%  

-­‐0.400%  

-­‐0.200%  

0.000%  

0.200%  

0.400%  

4/2   4/4   4/6   4/8   4/10   4/12   4/14   4/16   4/18  

ICM:  Flight  candidate  Lamp  SN16,  1DB55  SN94  PD  w  placed  Barr  filter,  SW  Ref  

SWCSMON1   SWREF   Linear  (SWCSMON1)   Linear  (SWREF)  

y  =  -­‐0.042%x  +  1741.646%  

y  =  0.004%x  -­‐  175.540%  

-­‐1.000%  

-­‐0.800%  

-­‐0.600%  

-­‐0.400%  

-­‐0.200%  

0.000%  

0.200%  

0.400%  

4/2   4/4   4/6   4/8   4/10   4/12   4/14   4/16   4/18  

BB:  Flight  candidate  Lamp  SN25,    1DB55  SN77  PD  w/o  filter,  TOT  Ref  

SWCSMON2   TOTREF   Linear  (SWCSMON2)   Linear  (TOTREF)  

Summary: Lamp SN24 and 1DB55 SN94 w placed filter performed best

Page 21: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 21

Clouds and the Earthʼs Radiant Energy System"

FM6 ICM vs. BB#1 ICM: PD Response

♦ Photodiodes have trended negative in every test configuration ♦ 1DB55 PDs tend to exhibit decreasing slope with time in vacuum ♦ 1DB55 SN94 remains leading PD candidate

Note- Results uncorrected for measured lamp drift

!"#$"%&

!"#'"%&

!"#("%&

!"#)"%&

!"#*"%&

!"#+"%&

!"#,"%&

!"#-"%&

"#""%&

"#-"%&

"#"& ,#"& *#"& (#"& $#"& -"#"& -,#"& -*#"& -(#"&

!"#$%&

"'()

$*+"'

,$-.'/*0"12$3445'

6(7'2$3445'8".9:';)<9<0=<0"';"1><15$*3"'(<5?$1=.<*'

./01&2-&345(&67849&

./01&2,&3:;<*&67849&

./01&2+&3-7=,(&67849&

./01&2*&3-7=,(&67>49&

./01&2)&345(&67>49&

./01&2(&3:;<*&67>49&

./01&2'&3:;<*&678?49&

!"#$"%&

!"#'"%&

!"#("%&

!"#)"%&

!"#*"%&

!"#+"%&

!"#,"%&

!"#-"%&

"#""%&

"#-"%&

"#"& ,#"& *#"& (#"& $#"& -"#"& -,#"&

!"#$%&

"'()

$*+"'

,$-.'/*0"12$3445'

66'2$3445'7".89':);8;0<;0"':"1=;15$*3"'(;5>$1<.;*'

./0&1234&5-67,(&869:;&

1234&<+&5=>?*&869:;&

1234&<*&5=>?*&86@:;&

1234&<)&5=>?*&869A:;&

1234&<(&5:B(&86@:;&

1234&<'&5=>''&86@:;&

Page 22: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 22

Clouds and the Earthʼs Radiant Energy System"

FM6 ICM vs. BB#1 ICM: Lamps

♦ Data from Test #3 is suspect due to mechanical crosstalk causing dual-off-state level

♦ Non-flight BB lamp is the only one to show a negative trend ♦ Lamps appear to have a vacuum stabilization period in the ICM

Results measured with Thermopile reference detectors

!"#$"%&

!"#'"%&

!"#("%&

!"#)"%&

"#""%&

"#)"%&

"#("%&

"#'"%&

"#$"%&

"#"& (#"& $#"& *#"& +#"& )"#"& )(#"&

!"#$%&

"'()

$*+"'

,$-.'/*0"12$3445'

66'2$3445'7".89':$5;'<"1=>15$*3"'(>5;$1?.>*'

,-.&/012&345&66&789:&

/012&;'&345&66&789:<&

/012&;$&345&66&789:&

/012&;=&345&66&789:&

/012&;*&3789&>4)*:&

/012&;?&3789&>4($:&

!"#$"%&

!"#'"%&

!"#("%&

!"#)"%&

"#""%&

"#)"%&

"#("%&

"#'"%&

"#"& )#"& (#"& '#"& $#"& *"#"& *)#"& *(#"& *'#"&

!"#$%&

"'()

$*+"'

,$-.'/*0"12$3445'

6(7'2$3445'8".9:';$5<'="1>?15$*3"'(?5<$1@.?*'

+,-.&/*&012'&3456&

+,-.&/)&012'&3456&

+,-.&/7&012'&34568&

+,-.&/(&0345&9:)76&

+,-.&/;&0345&9:*'6&

+,-.&/'&0:1&<<&3456&

+,-.&/=&0345&9:*'6&

Page 23: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 23

Clouds and the Earthʼs Radiant Energy System"

ICM Major Accomplishments since Summer 2012

♦ Performed Ambient Instrument-level testing  Testing showed possible air recovery of PD response  Testing eliminated gravity effects as root cause for anomaly

♦ Module-level ICM Diagnostic Vacuum Performance  Completed 7 vacuum test

•  Isolated problem to the ICM •  Tested FM6 Lamp and flight candidate lamp assemblies SN 23, 16, & 24 •  Testing of FM6 PD and flight candidate PD assemblies SN 94, 77

 Discovered and corrected filter assembly issues •  Reworked filter mounting to shim space between filter and PD window •  Devised/implemented temporary filter mounting to eliminate epoxy

♦ Performed initial stray light test   Remove and replace ICM Cover does not affect PD Signal

♦ Exonerated ICM Electronics under vacuum ♦ Performed filter stability assessment

  Measured shift does not affect PD Signal

Page 24: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 24

Clouds and the Earthʼs Radiant Energy System"

Near Term Path Forward with Photodiode

♦ Complete PD sensitivity to lamp de-centering using Breadboard ICM ♦ Prepare test set for ambient thermal sensitivity testing ♦ Configure ICM for Vacuum Test #8

 Lamp SN24  PD 1DB55 SN94 (assuming uniformity is achieved, test with placed

filter) ♦ Complete leak test of FM6 PD

 NG is working the planning to remove PD from mount for this testing ♦ Build-up and vacuum test of non-flight ODC 1DB55 in BB ♦ Build-up TVAC capability for final module test ♦ Resolve PD radiation sensitivity issue

Page 25: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 25

Clouds and the Earthʼs Radiant Energy System"

CERES FM6 Critical Path Schedule Wk 1 Wk 2 Wk 3 Wk 4 Wk 1 Wk 2 Wk 3 Wk 4 Wk 1 Wk 2 Wk 3 Wk 4 Wk 1 Wk 2 Wk 3

Jan Feb March Wk 4

CFT Aligmt + Docs TVAC

April

Sign-off

Wk 1 Wk 2 Wk 3 Wk 4 Wk 1 Wk 2 Wk 3 Wk 4 Wk 1 Wk 2 Wk 3 Wk 4 Wk 1 Wk 2 Wk 3 Sept Oct Nov

Wk 4 Dec

Calibration

ETC #3 Proposal

Wk 1 Wk 2 Wk 3 Wk 4 Wk 1 Wk 2 Wk 3 Wk 4 Wk 1 Wk 2 Wk 3 Wk 4 Wk 1 Wk 2 Wk 3 May June July

Wk 4 Aug

RCF Validation

Legacy MAM TV test & AO exposure

MAM Decision Gate

Flight ICM prep ICM & PD Diagnostics

Lamp Screening

Final Flight ICM Test

Candidate PD and lamp prep

Install ICM onto

instrmt PER

Instrument Storage at NG

Cal Report Final Verification Report

Final EIDP

Requirement Verification

TACR Telescope Replacement TACR Radiometric Analysis

FM6 MAM Removal

Legacy MAM Assembly

Legacy MAM Installation

Vibe Aligmt + Docs TVAC Calibration

Calibration Report

Additional PD Screening

EMI/EMC

PSRR EIDP

CFT MLI CLEAN

ICM/BB

Additional Testing See next chart

Page 26: CERES FM6 Calibration Subsystem Performance Issues · -4 -2 0 2 4 -4 -2 0 2 ... Lesson Learned: The diamond-turning process is plagued with issues of tooling marks ... uniform . Page

Page 26

Clouds and the Earthʼs Radiant Energy System"

24 25 26 27 28 29 30 1 2 3 4 5 6 7 8 Feb

9 April

26 27 28 29 30 31 1 2 3 4 5 6 7 8 9 Oct

10 Dec

10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 June July

25 Aug

Electrical Exoneration Part B (ICM under Vacuum)

April May

May

May Jun

Optical Path Sensitivity Evaluation (BB ambient pressure/temp)

Temperature Dependency Testing (BB ambient pressure) Set up for Temperature Dependency Testing

Design and Fab Filter Installation Kit

Original PD Leak Test

Set up for ICM Thermal Vacuum Testing

Flight ICM Integration (PD94 w/filter & Lamp 24)

BB Integration (OptiDiode no filter & Lamp 26) ICM Thermal Vacuum Testing

Schedule to Final ICM Test