36
Anesthetic Gas Modules By Pioneer Biomedical

By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Embed Size (px)

Citation preview

Page 1: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Anesthetic Gas ModulesBy Pioneer Biomedical

Page 2: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

HP &Philips

Page 3: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

GE

Page 4: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Other monitor equipment

Page 5: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Introduction

Thank you KAMI for letting us host your meeting!

Mike CorumDan ChapmanBrendan Healy

Page 6: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

www.AGM-parts.comDedicate website for Anesthetic Gas Monitoring, AGM-Parts.com

Page 7: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

M1026A support

Page 8: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

M1026A & M1026B

Page 9: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

M1026A & M1026B

Page 10: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Inside look

Page 11: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

M1026A Insides

Page 12: By Pioneer Biomedical. HP &Philips GE Other monitor equipment
Page 13: By Pioneer Biomedical. HP &Philips GE Other monitor equipment
Page 14: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Flow Diagram

Page 15: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Major Problems

Page 16: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Problem Hierarchy

Page 17: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Unit without errors

Page 18: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Common Problem #1

Page 19: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Agent ID replacement

Page 20: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Common Problem # 2

Page 21: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Common Problem # 3

Page 22: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Other IR Bench errors

Measurement Optic Path shows failure and Main board failure

Noisy Baseline on CO2, N2O, or AgentGas Analyzer Equipment Malfunction or

communication problem

Page 23: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Replacing the IR Bench

Page 24: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Transferring Data

Page 25: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Check out procedure

1 Perform Mechanical/Visual Checks  2 Check Error Contents 3 Check System Ventilator Fan Operation 4 Perform Pump Test 5 Replace Internal Nafion Tubing 6 Replace Manifold Seals 7 Replace Room Air Filter 8 Replace Pump Filter 9 Pump Hours: 10 Check Pneumatic System Leakage 11 Check/Calibrate Pneumatic System Flow rates and set (150 and 310 ml/min) 12 Perform Zero Calibration 13 Perform Barometric Pressure Check/Calibration 14 Perform another Zero Calibration 15 Perform Span Calibration Check on all gas channels and agents 16 Repeat Zero Calibration, check displayed values 17 Perform Electrical Safety Check

Page 26: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Quality assurance- what customers get from us

Page 27: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Checking Nafion Tubing

Page 28: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Leak Check

Page 29: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Checking flows

Need:Flow meterTubing

Page 30: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Adjusting flows

Normal Flow Adjustment

Purge Flow Adjustment

Page 31: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Barometric Pressure Check

Page 32: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

SPAN calibration of all gases

Page 33: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

GE

Page 34: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Other monitor equipment

Page 35: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

HP &Philips

Page 36: By Pioneer Biomedical. HP &Philips GE Other monitor equipment

Conclusion

For parts, tech help, PM kits, Pioneer Biomedical 1-888-644-3274

Air Liquide, 1-215-766-7407, x1235, for combined M1026A & M1026B cal gas

Part # 47068SR 3% isoflurane, 5% C02, 40% Nitrous Oxide, Bal O2 works for both M1026A and M1026B