Anesthetic Gas ModulesBy Pioneer Biomedical
HP &Philips
GE
Other monitor equipment
Introduction
Thank you KAMI for letting us host your meeting!
Mike CorumDan ChapmanBrendan Healy
www.AGM-parts.comDedicate website for Anesthetic Gas Monitoring, AGM-Parts.com
M1026A support
M1026A & M1026B
M1026A & M1026B
Inside look
M1026A Insides
Flow Diagram
Major Problems
Problem Hierarchy
Unit without errors
Common Problem #1
Agent ID replacement
Common Problem # 2
Common Problem # 3
Other IR Bench errors
Measurement Optic Path shows failure and Main board failure
Noisy Baseline on CO2, N2O, or AgentGas Analyzer Equipment Malfunction or
communication problem
Replacing the IR Bench
Transferring Data
Check out procedure
1 Perform Mechanical/Visual Checks 2 Check Error Contents 3 Check System Ventilator Fan Operation 4 Perform Pump Test 5 Replace Internal Nafion Tubing 6 Replace Manifold Seals 7 Replace Room Air Filter 8 Replace Pump Filter 9 Pump Hours: 10 Check Pneumatic System Leakage 11 Check/Calibrate Pneumatic System Flow rates and set (150 and 310 ml/min) 12 Perform Zero Calibration 13 Perform Barometric Pressure Check/Calibration 14 Perform another Zero Calibration 15 Perform Span Calibration Check on all gas channels and agents 16 Repeat Zero Calibration, check displayed values 17 Perform Electrical Safety Check
Quality assurance- what customers get from us
Checking Nafion Tubing
Leak Check
Checking flows
Need:Flow meterTubing
Adjusting flows
Normal Flow Adjustment
Purge Flow Adjustment
Barometric Pressure Check
SPAN calibration of all gases
GE
Other monitor equipment
HP &Philips
Conclusion
For parts, tech help, PM kits, Pioneer Biomedical 1-888-644-3274
Air Liquide, 1-215-766-7407, x1235, for combined M1026A & M1026B cal gas
Part # 47068SR 3% isoflurane, 5% C02, 40% Nitrous Oxide, Bal O2 works for both M1026A and M1026B