Upload
phungkhuong
View
213
Download
0
Embed Size (px)
Citation preview
Weak Value Amplification of a Post-Selected Single Photon
by
Matin Hallaji
A thesis submitted in conformity with the requirementsfor the degree of Doctor of Philosophy
Graduate Department of PhysicsUniversity of Toronto
© Copyright 2016 by Matin Hallaji
Abstract
Weak Value Amplification of a Post-Selected Single Photon
Matin Hallaji
Doctor of Philosophy
Graduate Department of Physics
University of Toronto
2016
Weak value amplification (WVA) is a measurement technique in which the effect of a pre- and
post-selected system on a weakly interacting probe is magnified. In this thesis, I present the first
experimental observation of WVA of a single photon. We observed that a signal photon - sent
through a polarization interferometer and post-selected by photodetection in the almost-dark
port - can act like eight photons. The effect of this single photon is measured as a nonlinear phase
shift on a separate laser beam. The interaction between the two is mediated by a sample of laser-
cooled 85Rb atoms. Electromagnetically induced transparency (EIT) is used to enhance the
nonlinearity and overcome resonant absorption. I believe this work to be the first demonstration
of WVA where a deterministic interaction is used to entangle two distinct optical systems. In
WVA, the amplification is contingent on discarding a large portion of the original data set.
While amplification increases measurement sensitivity, discarding data worsens it. Questioning
whether these competing effects conspire to improve or diminish measurement accuracy has
resulted recently in controversy. I address this question by calculating the maximum amount
of information achievable with the WVA technique. By comparing this information to that
achievable by the standard technique, where no post-selection is employed, I show that the
WVA technique can be advantageous under a certain class of noise models. Finally, I propose
a way to optimally apply the WVA technique.
ii
Acknowledgements
When I started graduate school, I had to make the tough decision of choosing between ex-
perimental and theoretical quantum optics. I chose experiment. Having no experience in
experimental physics (except a couple of undergraduate lab courses), the first couple of months
working in the lab were extremely difficult. But once I made the transition, I couldn’t be hap-
pier with the choice I made. Working long hours in the lab and the frustration that came with
it, although difficult to bear at times, shaped the person I am now. It taught me one important
lesson: don’t let your opinions bias you towards a desired outcome and always seek the truth.
Doing experiments comes with a reward that no amount of computer simulations and calcula-
tions on the paper can top: watching nature do what you predicted it would. The first time
I saw a cloud of atoms float on a sheet of light, the first time I observed off-resonance Raman
transition, the first time I saw a photon acting as if it was eight photons are the moments I
would never forget in my life.
I should start by thanking my supervisor, Aephraim Steinberg. The amount I learned from
him, his knowledge and superb physical intuition is incalculable. Without his help and guidance
I would have never been able to make the transition of becoming an experimentalist. During
my time in the lab, I had the chance to meet and work with so many amazing and brilliant
people. The first person I closely worked with was Chao Zhuang. He taught me a lot about the
apparatus and optical lattices. I had the privilege to work alongside with Greg Dmochoswki
on the LMI project. I also made some fantastic and dear friends in the physics departments:
Ramon Ramos, Nicolas Quesada and Shreyas Potnis. We shared a lot of moments together
and it saddens me deeply not to be able to see them everyday. My special thanks go to Josiah
Sinclair for his enormous help, support, contributions and many many interesting questions and
discussions. I should also thank Hugo Ferretti, Edwin Tham, David(D1) Schmid, David(D2)
Spierings, Ginelle Johnston, Dylan Mahler, Lee Rozema, Rockson Chang and Alex Hayat for
their helps, supports and priceless conversations.
Being an experimentalist comes with staying long nights in the lab working, debugging and,
if you are lucky, taking data. I consider myself very lucky to have spent most of those long nights
with an extremely smart person and a great friend, Amir Feizpour. None of the milestones we
achieved in the lab in the past two years would have been possible without Amir’s dedication
and hard work. He was an inspiration that kept me motivated to do the same.
iv
Contents
1 Introduction 1
1.1 Background . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Outline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Disambiguation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 Weak Value Amplification of Photon Number - Theory 5
2.1 Introduction to Weak Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1.1 Information versus Disturbance . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1.2 Weak Measurement and Weak Value Amplification . . . . . . . . . . . . . 8
2.2 Weak Value Amplification of a Post-Selected Photon . . . . . . . . . . . . . . . . 13
2.2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.2.2 Schrodinger picture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.2.3 Weak Value Approach . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.2.4 Weak Values and PNR Detectors . . . . . . . . . . . . . . . . . . . . . . . 28
2.2.5 Measuring the Amplified Effect of a Post-selected Single Photon . . . . . 30
2.2.6 Weak Values vs. Strong Values - ABL Rule . . . . . . . . . . . . . . . . . 31
2.3 Weak Measurement as a POVM . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3 Apparatus 36
3.1 Magneto-Optical Trap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3.1.1 Trapping Beam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.1.2 Repumper Beam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
3.1.3 MOT setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
3.1.4 Magnetic fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
3.1.5 Remote Task Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
v
3.1.6 Maintenance and Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . 54
3.2 Probe and Coupling Beams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
3.3 Signal Pulses, Photon Detection and Tagging . . . . . . . . . . . . . . . . . . . . 58
3.3.1 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
3.3.2 Photon Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
3.3.3 Gating the Photon Detection . . . . . . . . . . . . . . . . . . . . . . . . . 59
3.3.4 Tagging the Shots with Successful Photon Detection . . . . . . . . . . . . 60
3.4 Polarization Interferometer and Tomography . . . . . . . . . . . . . . . . . . . . 61
3.5 Atom Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
3.5.1 Level Scheme . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
3.5.2 Optical Pumping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
3.5.3 Atom Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
4 Experimental Results 68
4.1 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
4.1.1 Single Photon Cross-Phase Shift . . . . . . . . . . . . . . . . . . . . . . . 69
4.1.2 XPS Versus Signal Polarization . . . . . . . . . . . . . . . . . . . . . . . . 70
4.1.3 Polarization Tomography on the Signal . . . . . . . . . . . . . . . . . . . 70
4.2 Measurement Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
4.2.1 φclick and φno−click . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
4.2.2 φno−click versus |α|2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
4.2.3 Weak Value Amplification of a Single Post-selected Photon . . . . . . . . 73
4.2.4 Technical Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
4.2.5 Discussion of the Results . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
5 On the Advantage of Weak Value Amplification 78
5.1 Fisher Information as a Figure of Merit . . . . . . . . . . . . . . . . . . . . . . . 79
5.1.1 Covariance Matrix and Fisher Information . . . . . . . . . . . . . . . . . . 79
5.1.2 Fisher Information versus Signal-to-Noise Ratio . . . . . . . . . . . . . . . 80
5.2 Measuring the Effect of a Single Photon . . . . . . . . . . . . . . . . . . . . . . . 81
5.2.1 Gaussian White Noise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
5.2.2 Noise with Time Correlation . . . . . . . . . . . . . . . . . . . . . . . . . 83
5.2.3 Advantage of Post-selection without Weak-Value Amplification . . . . . . 87
vi
5.3 Experimental Proposal and Progress . . . . . . . . . . . . . . . . . . . . . . . . . 89
5.3.1 Fisher Information for Coherent States as the Signal . . . . . . . . . . . . 89
5.3.2 Experimental Progress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
6 Outlook and Summary 92
6.1 Possible Improvements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
6.2 Future of the current Apparatus . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
6.3 Revisiting Optical Lattices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
6.4 Rydberg Atoms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
6.5 Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
A Weak Value Approximations 98
B Alignment Procedures 102
B.1 Polarization Spectroscopy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
B.2 Injection-Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
B.3 Tapered Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
B.4 Polarization Maintaining Fiber . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
B.5 MOT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Bibliography 107
vii
List of Tables
3.1 Typical Trapping parameters for MOT operation. TA running at 1A. . . . . . . 46
3.2 Typical Trapping parameters for molasses operation. TA running at 1A. . . . . 47
3.3 Typical Trapping parameters for imaging. TA running at 1A. . . . . . . . . . . 47
3.4 Typical repumper parameters for different operation. . . . . . . . . . . . . . . . 49
4.1 Polarization tomography on the signal . . . . . . . . . . . . . . . . . . . . . . . . 71
4.2 Experimental parameters used in the WVA experiment. . . . . . . . . . . . . . . 72
viii
List of Figures
2.1 Overlap between two displaced states of the meter: (a) when the displacement
is smaller than the initial uncertainty of the meter, (b) when the displacement
is larger than the initial uncertainty of the meter. . . . . . . . . . . . . . . . . . 8
2.2 Demonstration of WVA as an interference between the meter states: (a) The
result of the interference between ψ+(x) and ψ−(x). The mean value of this
state is much larger than the displacement of either of ψ±(x). (b) The displaced
states ψ+(x) and ψ−(x). It can be seen that the two states are not resolvable.
The vertical lines demonstrated the mean value of each of the distributions. . . 11
2.3 Weak value versus δ. The insets show 12((1 + δ)ψ+(x)− (1− δ)ψ−(x)) for three
different values of δ. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4 Schematic of the setup used to amplify the effect of a single photon. . . . . . . . 13
2.5 A real detector with efficiency η modeled as an ideal detector and a beam splitter
that only transmits a fraction η of the incoming light. In the case where the
incoming beam is a coherent state, the two “detected” and “undetected” modes
are independent of each other and the mean photon number inferred in the
detected mode does not affect the mean photon number in the undetected mode. 15
2.6 The difference between the inferred mean photon number for the “click” events
and |α|2 versus |α|2 for efficiency η = 0.2 (solid line.) The dashed line is when the
effect of dark-counts is added to the inferred photon numbers. The inset shows
the difference between nclick and nno−click versus |α|2, when no dark-counts are
considered (solid) and when the dark-counts are considered (dashed.) . . . . . . 16
2.7 Schematic of the WVA setup. The signal pulse |α〉s interacts with the probe
|β〉probe inside an interferometer. The two arm interact with the probe with
different strengths represented as per photon XPS’s of φ1 and φ2, where φ1 > φ2. 17
ix
2.8 Schematic of the WVA setup. The signal pulse |α〉s interacts with the probe
|β〉probe inside an interferometer. The two arm interact with the probe with
different strengths represented as per photon XPS of φ1 and φ2, where φ1 > φ2.
The effect of a detector with finite efficiency is modeled as a combination of a
beam splitter and a detector with 100% efficiency. . . . . . . . . . . . . . . . . . 24
2.9 Weak value of number of photons in the two arms of the interferometer versus
the post-selection parameter δ. For this plot, |α|2 and η are chosen to be 1 and
0.2 respectively. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
2.10 The weak value of number of photons in arm “1” of the interferometer versus
post-selection parameter, for post-selecting on different n’s at the PNR detector.
For this plot, |α|2 and η are chosen to be 1 and 0.2 respectively. . . . . . . . . . 29
2.11 The weak value of number of photons in arm “1” versus δ, calculated by consid-
ering higher photon terms,∑Pnn, and only post-selecting on a single photon,
n = 1. For this plot, |α|2 and η are chosen to be 1 and 0.5 for a) and 10 and 0.5
for b). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2.12 Comparison between the weak value of mean photon number versus the strong
value calculated using ABL rule. As expected the strong value remains un-
changed as the post-selection parameter changes whereas the weak value increases
as the post-selection parameter decreases. . . . . . . . . . . . . . . . . . . . . . . 32
3.1 Schematic of a one-dimensional MOT setup. Top: Two coils in anti-Helmholtz
configuration create a magnetic field gradient. Two optical fields, with the same
circular polarization (opposite spins due to opposite propagation direction), are
sent through the center of the magnetic field. The two optical fields are red-
detuned from the J = 0→ J′ = 1 transition of the atoms placed near the center
of the magnetic field gradient. Bottom: the Zeeman shifted energy levels of an
atom in the magnetic field gradient. . . . . . . . . . . . . . . . . . . . . . . . . . 38
3.2 Schematic of the trapping beam setup. . . . . . . . . . . . . . . . . . . . . . . . 39
x
3.3 Polarization spectroscopy setup. A circularly polarized pump saturates the
atoms, and a probe, which is vertically polarized, undergoes a polarization rota-
tion through the vapor. A combination of a HWP and a PBS projects the probe
in diagonal and anti-diagonal polarization basis. Two photo-diodes detect the
intensity of the probe in these basis. The subtracted signal is then used as the
error signal to feedback to the master laser for frequency stabilization. . . . . . 40
3.4 Polarization spectroscopy of F = 3→ F ′ transitions. The vertical lines show the
position of different transitions and crossover transitions. The blue circle shows
where the laser is locked to. The horizontal axis of the is detuning of the laser
from F = 3→ F ′ = 4 transition. . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
3.5 Schematic of AOM double-pass setup. . . . . . . . . . . . . . . . . . . . . . . . . 42
3.6 Calibrations of the AOM double-pass setup. (a) Detuning of the laser from
F = 3→ F ′ = 4 transition versus the modulation voltage. (b) Output power of
the double-pass versus the modulation voltage. . . . . . . . . . . . . . . . . . . . 42
3.7 Schematic of injection-lock setup. The output of the double-pass is used to seed
a high power diode laser. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
3.8 Measured power after the PM fiber versus amplitude modulation voltage of the
AOM, when TA running at 1.5A. . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.9 Schematic of inside of the TA box, both top view and side view. . . . . . . . . . 46
3.10 Schematic of the setup to prepare the repumper beam. . . . . . . . . . . . . . . 48
3.11 Saturated absorption spectroscopy error signal for F = 2→ F ′ transitions. The
red circle indicates where the repumper laser is locked to. The transition from
the F = 2 ground state to the F ′ = 1, 2, 3 excites states as well as the crossovers
are marked on the signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
3.12 Calibration of output power of the PM fiber versus the modulation voltage. . . 49
3.13 D2-line of 85Rb atoms. The red arrow indicated the tapping beam and the blue
arrow indicated the repumper beam. . . . . . . . . . . . . . . . . . . . . . . . . 50
3.14 Schematic of optics used for combining the trapping and repumper beam and
capturing a MOT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
3.15 Power versus voltage read by Uberdetect for (a) trapping beam and (b) repumper
beam. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
xi
3.16 Magnetic field measured at the center of one of the OP coils. The field is measured
along the coil axis, z, and the orthogonal directions x and y. . . . . . . . . . . . 53
3.17 Schematic of different stages used to convert the parameters defined in the Lab-
View software to analog signals used to control multiple devices. . . . . . . . . . 54
3.18 Schematic of the probe, the coupling and the signal beam. . . . . . . . . . . . . 56
3.19 Schematic of the signal pulses, SPCM and how the SPCM is gated. . . . . . . . 58
3.20 Probability of photon detection versus number of photons at the SPCM. From a
fit to the data, the efficiency of the SPCM is measured to be 70%. . . . . . . . . 59
3.21 Schematic of the tagging procedure. The result of the AND gate switches on
a 100MHz RF signal to an AOM upon a successful photon detection. The 1st
and 2nd diffraction order of the AOM are sent to the detector which detects the
probe. When a successful detection happens, this AOM sends a flash of light
to the detector which results in a short spike in the measured amplitude (and
phase.) A delay generator controls the relative delay between the expected XPS
and the position of the tag. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
3.22 Schematic of the signal path. Dashed lines marked as 1 and 2 indicated the
beginning and the end of the polarization interferometer. . . . . . . . . . . . . . 62
3.23 Level scheme used for the WVA experiment. . . . . . . . . . . . . . . . . . . . . 63
3.24 Measured OD versus frequency of the probe for three situations: no external field,
no optical pumping with external field and with optical pumping and external
field. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
3.25 The result of fitting the OD versus probe frequency to the weighted sum of five
shifted Lorentzian functions for (a) no optical pumping and (b) with optical
pumping. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
xii
4.1 Schematic of the experimental setup. Counter-propagating probe and signal
beams are focused down to a waist of 13µm inside a cloud of laser cooled 85Rb
atoms confined in a magneto-optical trap. A collimated coupling beam, prop-
agating in a direction orthogonal to that of the signal and probe, sets up EIT
for the probe. The signal beam is prepared in a linear polarization and after its
interaction with the probe beam is post-selected on a final polarization that is
almost orthogonal to the initial polarization. The probe and the signal beams
are separated from one another using a 10:90 beam splitters. . . . . . . . . . . . 69
4.2 XPS versus signal polarization. A quarter-wave plate is used to set the polar-
ization of the signal beam before the interaction. The maximum and minimum
measured phase shifts corresponds to the two circular polarizations σ+ and σ−.
The signal pulses used for this measurement contained around 600 photons. . . 70
4.3 Measured XPS for click events and no-clicks events. Red squares and green
circles are the measured phase shift conditioned on the SPCM not firing (no-
click) and firing (click) respectively. The horizontal axis shows the mean signal
photon number n and the post-selection parameter δ used for each case. . . . . 73
4.4 The measured phase shift for no-click cases versus the average photon number
in the signal beam |α|2 = n. A linear fit reveals a per-photon XPS of φ =
7.6± 0.3µrad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
4.5 XPS difference versus post-selection parameter. Red circles are measured phase
shift difference between click and no-click cases. Gray dashed line is a fit to
φ++φ−2 + φ+−φ−
2δ with φ++φ−2 assumed to be 7.6µrad and φ+ − φ− left as a free
parameter. The fit reveals a value of 11.5± 1.5µrad for φ+−φ−. The cyan solid
line is the full theoretical calculation with the numbers extracted from the two
fits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
5.1 Optimal weights obtained by the covariance matrix Cij = exp(−|i−j|/2.) versus
naive averaging when all data points are weighted equally. . . . . . . . . . . . . 80
5.2 Fisher information in WVT and ST versus post-selection probability. . . . . . . 82
5.3 Fisher information in WVT and ST versus post-selection probability. . . . . . . 83
xiii
5.4 Calculated covariance matrix for (a) 10 points correlated on average and (b) no
time correlation. The noise for (a) is modeled as a slowly varying noise added to
a white noise. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
5.5 Fisher information in the standard technique and the weak value technique.
Figure (a) plots the fisher information when measurements are uncorrelated.
Figures (b),(c) and (d) are results for 0.8, 2 and 20 measurements per correlation
time. A total of 200 data points are used in the data set. The CM’s are calculated
by averaging over 10K trials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
5.6 Two noise traces, with long time correlations(top) and without time correlation
(bottom). The dots represents sampling from the noise with a fixed rate. The
red dots are a 20% random selection from them. . . . . . . . . . . . . . . . . . . 86
5.7 Fisher information in the standard technique, weak value technique, the rejected
subset and correlation between the post-selected subset and rejected subset for
various correlation times. The “optimal” corresponds to summing over the Fisher
information in WVT, rejected data and the correlations. Figure (a) plots the
fisher information when measurements are uncorrelated. Figures (b),(c) and (d)
are results for 0.8, 2 and 20 measurement shots per correlation time. . . . . . . . 88
5.8 The effect of post-selection without weak-value amplification on Fisher informa-
tion versus |α|2. The detector efficiency used for the plots are 100%, 20% and
1% for (a),(b) and (c) respectively. . . . . . . . . . . . . . . . . . . . . . . . . . . 89
5.9 Fisher information for the WVT and ST versus number of correlated measure-
ments. We assumed δ = 0.26, |α|2 = 15 and η = 0.2. . . . . . . . . . . . . . . . . 90
5.10 Schematic of the setup to add slow phase-noise to the probe beam. . . . . . . . 91
6.1 Schematic of the improved level scheme. . . . . . . . . . . . . . . . . . . . . . . 93
B.1 Schematic of setup used to align the input polarization of the PM fiber. . . . . . 105
B.2 (a)Front (left) and side view (right) and (b)3D visualization of relative position
of MOT beams with the cuvette and magnetic field coils. . . . . . . . . . . . . . 106
xiv
Preface
The work presented in this thesis was done as part of the Light Matter Interaction (LMI)
project in the group of Prof. Steinberg. In particular, this weak value amplification project
is in close relation with another project, entitled “Observation of the nonlinear phase shift
due to single post-selected photons” (Nature Physics 11, 905-909) on which I was a principal
contributor and co-author, along with Amir Feizpour and Greg Dmochowski. Therefore, much
of the background to my work has been presented previously in the thesis of Amir Feizpour
(Nonlinear Optics at the Single Photon Level. PhD thesis, University of Toronto, Toronto,
2015.) In order to avoid repeating what has already been published, a few key elements of
this work, such as a full description of the nonlinearity induced by the atomic medium (the
so-called “N-scheme”) and detailed discussion of parts of the experimental setup, are just briefly
introduced; the reader can refer to the references mentioned above and cited later in my thesis
for more detail.
When I started my graduate studies in this group, the cold atom apparatus was shared
between a Lattice project, the goal of which was to study the quantum control techniques using
vibrational states of an optical lattice, and the LMI project. While I had been the lead on
the Lattice project, Amir Feizpour and Greg Dmochowski were setting up the EIT setup and
phase measurement apparatus. The work on the lattice project resulted in “Quantum control of
population transfer between vibrational states in an optical lattice” (arXiv:1510.09186) and my
work on the LMI project contributed to “Observation of EIT-enhanced cross-phase modulation
in the short-pulse regime” (arXiv:1506.07051). Once the lattice project was concluded, I com-
mitted myself fully to the LMI project and after redesigning and improving the apparatus, led
by Amir Feizpour on the “photon” side and me on the “atom” side, we succeeded in measuring
the nonlinear effect due to single post-selected photons. With Amir graduating I took the lead
on the LMI project and the results presented in Chapters 4 grow entirely out of the time during
xv
which I was leading the project. Below is the list of publications on which I’m an author or
co-author:
• Observation of Amplification of the Nonlinear Effect of a Post-Selected Single Photon Us-
ing Weak Measurement Matin Hallaji, Amir Feizpour, Greg Dmochowski, Josiah Sinclair
and Aephraim M. Steinberg (manuscript under preparation)
• Quantum control of population transfer between vibrational states in an optical lattice
Matin Hallaji, Chao Zhuang, Alex Hayat, Felix Motzoi, Botan Khani, Frank K. Wilhem
and Aephraim M. Steinberg, arXiv:1510.09186
• Observation of the nonlinear phase shift due to single post-selected photons Amir Feizpour,
Matin Hallaji, Greg Dmochowski and Aephraim M. Steinberg (Nature Physics 11, 905-909
(2015) , doi:10.1038/nphys3433)
• Optimal estimation, correlated noise, and weak value amplification Josiah Sinclair, Matin
Hallaji, Aephraim M. Steinberg, Jeff Tollaksen and Andrew N. Jordan (manuscript under
preparation)
• Observation of EIT-enhanced cross-phase modulation in the short-pulse regime Greg Dmo-
chowski, Amir Feizpour, Matin Hallaji, Chao Zhuang, Alex Hayat and Aephraim M.
Steinberg, arXiv:1506.07051
• Observing the Onset of Effective Mass Rockson Chang, Shreyas Potnis, Ramon Ramos,
Chao Zhuang, Matin Hallaji, Alex Hayat, Federico Duque-Gomez, John E. Sipe, Aephraim
M. Steinberg, Phys. Rev. Lett 112,170404 (2014)
xvi
Chapter 1
Introduction
1.1 Background
Quantum mechanics is one of the most successful theories to date. One important feature
of quantum mechanics is the way it treats measurements, where instead of deterministically
predicting the outcome of a particular measurement, it assigns probabilities to every possible
outcome. Understanding the nature of measurement in quantum mechanics remains an out-
standing problem[1, 2]. Due to its tremendous success, quantum mechanics has opened up the
field of quantum metrology, which promises to use the theory of quantum mechanics for de-
veloping novel and efficient techniques for parameter estimation and detecting physical effects
with high resolution.
In quantum mechanics, measurement of an observable of a system is accomplished by ob-
serving its effect on a measurement apparatus by entangling them via an interaction, e.g.
measuring the spin of a particle by entangling its spin to its center of mass degree of free-
dom using an inhomogeneous magnetic field[3]. To make the effect resolvable, one needs to
make the system and the meter interact strongly with one another. This resolvability, however,
comes at the cost of disturbing the system. In 1988, Aharonov, Albert and Vaidman (AAV)
introduced a new measurement technique in which the interaction between the system and the
meter is set to be very weak and the effect of the system is measured only if the system is
found to be in a particular final state after the measurement[4, 5, 6, 7]. In this approach, the
system’s disturbance is reduced at the cost of a reduction in the information acquired by the
measurement. The effect of the system on the probe, which is called the weak value, can then
be found by averaging the probe’s readout over many repetitions of the measurement. AAV
1
Chapter 1. Introduction 2
showed that in this type of measurements, the weak value depends equally on both the pre-
selected (initial) and the post-selected (final) states of the system. This feature of the weak
value makes it a powerful tool for exploring fundamental questions and paradoxes in quantum
mechanics[8, 9, 10, 11, 12, 13, 14, 15]. Interest in weak measurement for practical purposes has
grown significantly in past few years[16, 17, 18, 19, 20, 21]. Weak values can be outside of the
eigenvalue spectrum of the observable being measured, and can even be complex. Specifically, as
the overlap between the initial and final states becomes very small, the weak value can become
arbitrarily large 1. This feature, which is called weak value amplification (WVA), has been used
in metrology for parameter estimation and sensitive measurements[22, 23, 24, 25, 26, 27, 28].
Anomalous weak-values observed to date typically utilize two different degrees of freedom of a
single optical system as the “system” and the “probe”[22, 23] , and can therefore be explained
using Maxwell’s equations. Probabilistic weak interactions between two distinct optical systems
has also been demonstrated in [29, 30] where, although the interaction cannot be classically ex-
plained, the probabilistic nature of the interaction greatly reduces their usefulness to quantum
metrology. WVA with a deterministic interaction between two distinct physical systems has
previously only been observed in a transmon qubit system[13] but not, to the best of my knowl-
edge, in optical systems. Here, I present the first observation of weak value amplification of the
photon number in a signal beam. This is accomplished by measuring the nonlinear phase-shift
the signal writes on a probe beam via a deterministic weak interaction that is mediated by a
sample of cold atoms. We believe this work to be the first demonstration of WVA in which a
deterministic weak interaction is used to entangle two distinct optical systems.
Interest in the application of weak measurement to quantum metrology has grown signifi-
cantly in the past few years and debate over its usefulness is still ongoing. While proponents
argue in favor of weak measurement being superior to the standard technique2[31, 32, 33, 34,
35, 36] with experimental demonstrations in[37, 38], opponents claim that weak value tech-
nique is suboptimal[39, 40, 41, 42, 43]. Discussion on imaginary weak values or the effect of
technical noise can be found in[44, 45, 46, 47, 48]. At the core of this debate is the discussion
about quantifying the loss of information when measurement outcomes are rejected in weak
measurement. While it can be shown that in certain regimes the post-selected portion of the
data can contain nearly all the accessible information, some claim that weak measurement is in
1As long as the post-selection success is dominated by the overlap of pre- and post-selected states and notthe measurement back-action.
2I refer to the measurement without pre- and post-selection as the standard technique.
Chapter 1. Introduction 3
general a suboptimal technique. In [39], an optimal measurement technique is suggested, and
it is shown that this optimal technique is always superior to the WVA technique. But one can
argue that the suggested optimal measurement requires a large resource and is not feasible to
be performed experimentally[31]. Furthermore, one ought to assess the cost of implementing
a more complicated procedure against the additional information gained. In this thesis I will
study the performance of weak values under certain noise models and will show that in a certain
regime, the information in the post-selected data can be made to be the same as the information
in the optimal measurement technique.
The curious features of weak measurement have made it an attractive subject to be stud-
ied from many perspectives. Recently, the quantum mechanical nature of WVA has been
challenged[49] and attempts have been made to describe the effect classically based on mea-
surement disturbance. In their work[49], anomalous values are achieved by using a disturbance
model where the probability of disturbing the system depends on the outcome of the measure-
ment; even if the pre- and post-selected states of the systems are orthogonal, which means the
post-selection can only ever occur due to the measurement back-action. In our system, how-
ever, when the post-selection probability is dominated by the measurement disturbance (and
the ovelap between pre- and post-selected states is zero), no anomalous result is expected; hence
our results cannot be explained by the classical model given in the work mentioned[49].
In 2011, we proposed that WVA could be used to amplify the number of photons in a signal
pulse[44]. By sending a single photon through an interferometer and post-selecting on the rare
event of detecting the photon in the nearly-dark port of the interferometer, one can “amplify”
the mean photon number in each of the arms of the interferometer. By making a probe coherent
state interact with the split single photon in one of the arms, one can measure a nonlinear phase
shift equal to that which would be expected from several photons on the probe despite the fact
that the signal beam only ever contained a single photon. This thesis reports on experimental
observation of weak value amplification of photon numbers, as well as a detailed theoretical
study of the phenomenon and its applications.
1.2 Outline
In chapter 2, I present theoretical calculations of the weak value amplification of a post-selected
single photon. First, I will briefly introduce weak measurement and weak value amplification of
Chapter 1. Introduction 4
a single photon. I will then extend the discussion to an experiment where coherent states are
used instead of a single photon and comment on the effect of post-selection. A full quantum
mechanical calculation of the problem will be presented as well as an equivalent weak value
treatment. I will compare the weak value results to a similar experimental situation but with
strong interactions, and finally I will introduce a POVM picture of weak measurement.
In chapter 3 I will introduce the apparatus used for this experiment in detail. The key
components and elements used in the setup will be discussed alongside with maintenance and
diagnosis tips for the most important parts of the setup. This chapter is written with the goal
of it being a reference for students working on the apparatus in the future.
The main experimental results of the thesis are in presented in chapter 4. Calibration data
as well as a discussion of the main result will be provided. I will show how a single post-selected
photon can be made to act like 8 photons. I will comment on how this experiment outperformed
a similar former experiment we did on observing the effect of a single post-selected photon[50]
without weak value amplification.
In chapter 5 I will discuss the advantage of WVA in the presence of time-correlated noise. I
will discuss how the information is distributed between the post-selected data and the rejected
data as a function of various measurement parameters. I will show how weak measurement
and weak value amplification can outperform standard measurement techniques as the time
correlation in the noise increases.
Finally, in chapter 6, I discuss various changes that could be made to improve the current
setup as well as different future directions to be pursued.
1.3 Disambiguation
In this thesis, when discussing measurement and weak measurement theory, I refer to the mea-
surement apparatus as “probe” and the quantum system that the measurement is performed on
as “system”. When I discuss the experiment, the “system” is an optical pulse the mean photon
number of which is being measured. This pulse is referred to as “signal” in the experiment. The
“probe” is a coherent state which measures the number of photons in the “signal” (“system”)
which in the experiment is still called the “probe”.
Chapter 2
Weak Value Amplification of Photon
Number - Theory
In this chapter, I present the theory of amplifying the nonlinear effect of post-selected single
photons, using weak measurement. I first introduce weak measurement and weak value ampli-
fication and discuss how it can be used to amplify the number of photons in a signal beam.
I present exact calculation of this effect using the Schrodinger’s equation as well as the weak
value approach to the problem. A comparison between the weak values and the strong values
(expectation values from a strong measurement) and a POVM approach to weak measurement
will be discussed.
2.1 Introduction to Weak Measurement
In this section I will briefly introduce weak measurement and weak value. I first discuss the
relation between the information gain from a measurement and its disturbance. I then introduce
the weak measurement and weak value amplification.
2.1.1 Information versus Disturbance
In quantum mechanics, measuring an observable of a system is done by coupling the system
to a meter via a Hamiltonian so that the change in the state of the meter depends on the
value of the observable being measured. Then, by looking at the final state of the meter, one
gains information about the observable of interest. This way of treating measurements was
5
Chapter 2. Weak Value Amplification of Photon Number - Theory 6
introduced by von Neumann. The Hamiltonian describing the interaction between the meter
and system in such measurement is given by
H = ~χP .A (2.1)
where P is the momentum operator of the meter, A is the observable being measured of the
system and χ is the coupling strength between the meter and the system. This Hamiltonian
causes a displacement in the meter’s position proportional to the eigenvalues of observable A.
A strong measurement is when the coupling strength between the meter and the system is large
enough so that the displacement due to each eigenvalue of A is larger than the uncertainty in
the meter’s position. In this limit, displacements due to different eigenvalues are well resolved
and a single measurement reveals the information about the value of A. This information,
however, comes at the price of disturbing the coherences of the system. To illustrate this, let’s
consider a simple example where the observable A has only two eigenvalues α+ and α− with
eigenvectors |+〉 and |−〉, respectively. The meter’s position wavefunction is initialized to be
|ψ0〉. Let’s assume that the system is initially in the state |i〉 = a|+〉+ b|−〉. This corresponds
to an initial density matrix given by
ρi =
|a|
2 ab
ab |b|2
(2.2)
Here I use overbars to indicate complex conjugates. After the interaction the combined state
of the meter and system will be
|Ψf 〉 = exp(− i~~∫χP .Adt)|ψ0〉 ⊗ (a|+〉+ b|−〉)
=a|ψ+〉|+〉+ b|ψ−〉|−〉(2.3)
where |ψ+〉 and |ψ−〉 are the displaced meter states due to eigenvalues α+ and α−, respectively.
To see the effect of the measurement on the system, I trace over the meter’s Hilbert space.
Chapter 2. Weak Value Amplification of Photon Number - Theory 7
Using position eigenstates |x〉 as a basis for the meter, we will have
Trmeter[|Ψf 〉〈Ψf |
]=
∫dx〈x|Ψf 〉〈Ψf |x〉
=
∫dx〈x|
[a|ψ+〉|+〉+ b|ψ−〉|−〉
][a〈ψ+|〈+|+ b〈ψ−|〈−|
]|x〉
=
∫dx[aψ+(x)|+〉+ bψ−(x)|−〉
][aψ+(x)〈+|+ bψ−(x)〈−|
].
(2.4)
Hence the density matrix of the system after the interaction becomes:
ρf =|a|2|+〉〈+|∫dx|ψ+(x)|2 + ab|+〉〈−|
∫dxψ+(x)ψ−(x)
ba|−〉〈+|∫dxψ−(x)ψ+(x) + |b|2|−〉〈−|
∫dx|ψ−(x)|2
(2.5)
Since the meter’s wavefunctions are assumed to be normalized we have∫dx|ψ±(x)|2 = 1. The
other two integrals are equal to the spatial overlap of the two displaced wavefunctions and can
be written as∫dxψ+(x)ψ−(x) = 〈ψ−|ψ+〉 and
∫dxψ−(x)ψ+(x) = 〈ψ+|ψ−〉. Therefore, the
final density matrix of the system can be written as
ρf =
|a|2 ab〈ψ−|ψ+〉ab〈ψ+|ψ−〉 |b|2
(2.6)
It’s evident that the coherences (off-diagonal terms in the density matrix) in the system are
reduced by a factor equal to the overlap between the two displaced meter states. If the two
displaced states are well resolved (have a very small overlap, figure 2.1(b)), then the system will
get completely decohered, but on the plus side, one measurement on the meter will reveal all
the information about the observable being measured. This is known as strong measurement.
On the other hand, if the two states have a large overlap, figure 2.1(a), the system will remain
coherent, but a measurement on the meter won’t give much information because the two meter
states are indistinguishable. In this regime, in order to be able to resolve the two displaced
states, one needs to repeat the measurement many times. At a first glance, the advantage of a
measurement without well resolved meter states may not be obvious, but, since the state of the
system is left almost undisturbed in these measurements, one could make use of the system after
it interacts with the meter. Specifically, one can initialize (pre-select) the system in a particular
state and after its interaction with the meter, can perform a projective measurement on the
system and condition reading the value of the meter only if a certain outcome is achieved (post-
Chapter 2. Weak Value Amplification of Photon Number - Theory 8
1.0 0.5 0.0 0.5 1.0(a)
0.0
0.2
0.4
0.6
0.8
1.0
1.0 0.5 0.0 0.5 1.0(b)
0.0
0.2
0.4
0.6
0.8
1.0
Figure 2.1: Overlap between two displaced states of the meter: (a) when the displacement issmaller than the initial uncertainty of the meter, (b) when the displacement is larger than theinitial uncertainty of the meter.
selection.) This is known as weak measurement, which was introduced by Aharonov, Albert
and Vaidman[4]. We will discuss weak measurement in more detail in the next section.
2.1.2 Weak Measurement and Weak Value Amplification
Assume a system with initial state |i〉 that is interacting with a meter with initial state |ψ0〉 via
a Hamiltonian given by equation 2.1. After the interaction, the combined state of the system
and meter can be written as
|Ψf 〉 = exp(− i~~∫χP .Adt)|i〉|ψ0〉 = exp(−igP .A)|i〉|ψ0〉 (2.7)
where g =∫χdt determines the strength of the interaction. Assuming weak measurement,
g 1, one can expand the exponential as exp(−igP .A) = I − igP .A. By post-selecting the
system on final state |f〉, the state of the probe can be calculated to be
|ψf 〉 = 〈f |(I− igP .A)|i〉|ψ0〉
= (〈f |i〉 − igP 〈f |A|i〉)|ψ0〉
= 〈f |i〉(I− igP 〈f |A|i〉〈f |i〉 )|ψ0〉
= 〈f |i〉 exp(−igAwP )|ψ0〉
(2.8)
Chapter 2. Weak Value Amplification of Photon Number - Theory 9
Where
Aw =〈f |A|i〉〈f |i〉 (2.9)
is defined as the weak value (for a discussion on the validity of the approximations made in
derivation of equation 2.9, see A.) Throughout the thesis, I will refer to the overlap of the initial
and final state as the post-selection parameter
δ ≡ 〈f |i〉, (2.10)
where δ 1 indicates post-selecting on an almost orthogonal state and δ = 1 indicates post-
selecting on the initial state1. The propagator exp(−igAwP ) displaces the state of the meter
|ψ0〉 by an amount equal to the weak value times g. So, if ψi(x) ∝ exp(−x2/2σ2), then after
the post-selection, the state of the meter becomes ψf (x) ∝ exp(−(x− gAw)2/2σ2) [4].
As mentioned in the previous section, for this technique to be valid, the disturbance of
measurement on the system should be negligible. To quantify this disturbance, we look at
the density matrix of the system after the interaction. This time I will use the momentum
eigenstates |p〉 as a basis for the meter
ρsystem =
∫dp〈p|e−igP A|ψ0〉|i〉〈i|〈ψ0|eigP A|p〉
=
∫dpe−igpA〈p|ψ0〉〈ψ0|p〉|i〉〈i|eigpA
=
∫dp|ψ(p)|2e−igpA|i〉〈i|eigpA,
(2.11)
where ψ(p) represents the momentum wavefunction of the meter. Equation 2.11 shows that for
each momentum component of the meter’s wavefunction p, the state of the system is displaced by
exp(−igpA). This disturbance can be made very small by choosing the momentum wavefunction
of the meter to be very narrow and centered around p = 0. Specifically, if ψ(p) = δ(p),
where δ represents the delta function, then the system will be completely undisturbed and
ρsystem = |i〉〈i|. Narrowing the momentum distribution, increases the position uncertainty of
the meter, which is the same criteria derived in the previous section. Therefore, for equation
2.9 to be a valid description of the measurement outcome, the measurement strength, g, and
the meter’s position uncertainty, σ, should be chosen so that the meter’s displacement due to
1In this thesis, I will assume the post-selection parameter is a real number.
Chapter 2. Weak Value Amplification of Photon Number - Theory 10
different eigenvalues of A have large overlaps.
The weak value given in equation 2.9, has some curious features. Firstly, it is not necessarily
a real value. Also, as the overlap between the initial and final states of the system (〈f |i〉)decreases, the weak value can get (almost) arbitrarily large, which will be discussed later in
more detail. Weak values taking an imaginary value may be at odds with the nature of them
representing a measurement outcome, but as it will soon become clear, the imaginary part of
the weak value comes about due to the measurement backaction and the disturbance of the
system. By writing the weak value in terms of its real and imaginary part as Aw = AR + iAI ,
the final state of the probe, up to a normalization factor, becomes
ψf (x) = exp(−(x− gAw)2
2σ2) = exp(−(x− gAR − igAI)2
2σ2)
∝ exp(igAIσ2
x) exp(−(x− gAR)2
2σ2).
(2.12)
As it can be seen in equation 2.12, the position of the meter is displaced by an amount pro-
portional to the real part of the weak value. The imaginary part of the weak value acts as
exp(igAIσ2 x). Since this a position dependent phase on the meter’s wavefunction, it acts as a
displacement in momentum. Therefore, the momentum of the probe is shifted by an amount
equal to gAI/σ2. The Hamiltonian in equation 2.1, however, cannot cause a momentum shift
on the meter. This change of momentum is purely a consequence of post-selection. As shown
in equation 2.11, the backaction of the measurement on the system depends on the momentum
of the meter. The backaction due to a certain momentum component of the meter can make a
specific post-selection more probable. So when the system is post-selected in that specific final
states, the momentum distribution of the meter is biased towards the momentum components
which made that post-selection more likely. Therefore, by looking at the displaced momentum
state of the meter, one can learn about the backaction of the measurement on the system.
As seen in equation 2.9, as the overlap between the initial and final state approaches zero,
the weak value can become very large. It can even become larger than the largest eigenvalue
of the observable being measured. This property of the weak value is referred to as Weak
Value Amplification (WVA.) While it may seem to be at odds with the standard measurement
in quantum mechanics, where the outcome is bounded within the eigenvalues of the observ-
able, this amplification can be easily seen as an interference phenomenon between the meter’s
displaces states. To illustrate this, let’s assume that the system was initially prepared in an
Chapter 2. Weak Value Amplification of Photon Number - Theory 11
equal superposition of eigenstates corresponding to two of the eigenvalues of the observable,
|i〉 = 1/√
2(|+〉 + |−〉). After the interaction, the meter’s state will be in a superposition of
the displaced states due to each of the eigenvalues. Post-selecting the system in an almost
orthogonal final state, |f〉 = 1/√
2((1 + δ)|+〉− (1− δ)|−〉), for post-selection parameter δ 1,
can cause the still coherent meter states to destructively interfere with each other. We will have
ψmeter = 〈f | 1√2
(ψ+(x)|+〉+ ψ−(x)|−〉)
=1
2((1 + δ)ψ+(x)− (1− δ)ψ−(x)).
(2.13)
The mean value of the resulting state can be much larger than the mean value of either of
the displaced meter states. In figure 2.2, an example of this phenomenon is demonstrated. In
figure 2.2(b), the two displaced meter states ψ±(x) are shown. In figure 2.2(a), the result of
12((1 + δ)ψ+(x)− (1− δ)ψ−(x)) is plotted. The vertical lines demonstrated the mean value of
each of the distributions. Note that δ = 〈f |i〉 is the overlap between the initial and final states
100 50 0 50 100(a)
0.02
0.00
0.02
0.04
0.06
0.08
20 10 0 10 20(b)
0.80
0.85
0.90
0.95
1.00
1.05
Figure 2.2: Demonstration of WVA as an interference between the meter states: (a) The resultof the interference between ψ+(x) and ψ−(x). The mean value of this state is much larger thanthe displacement of either of ψ±(x). (b) The displaced states ψ+(x) and ψ−(x). It can be seenthat the two states are not resolvable. The vertical lines demonstrated the mean value of eachof the distributions.
of the system, and the mean value of 12((1 + δ)ψ+(x)− (1− δ)ψ−(x)) depends on the value of
the parameter δ.
From equation 2.9, it may look like that the weak value can grow indefinitely large as
δ → 0. But the weak value starts to deviate from equation 2.9 when δ becomes so small that
the probability of post-selection is not given by δ2 anymore and instead is dominated by the
Chapter 2. Weak Value Amplification of Photon Number - Theory 12
remaining, although tiny, backaction of the measurement. Figure 2.3 shows the weak value
versus δ. It can be seen that the weak value is inversely growing with δ up to a maximum
value. After that it start to become smaller as δ decreases. The insets of the plot show the
result of the interference between ψ±(x) for three different value of δ. The behavior of weak
values in figure 2.3 is the result of interference between two displaced Gaussian functions with
different weights, where the weight is set by the parameter δ. For δ = 0, the two Gaussian
functions have equal (in magnitude) but opposite weights and the destructive interference will
result in a distribution with zero mean (see the left inset of figure 2.3.) As δ increases, the
imbalance between the two interfering Gaussian functions will result in a distribution with a
mean that is larger than the mean of the original Gaussian functions (see the middle inset of
figure 2.3.) Further increase in δ, decreases the mean of the distribution resulting from the
interference between the two displaces Gaussians (the right inset.) For δ = 1, the two Gaussian
functions have equal wights and the constructive interference will also result in a distribution
with zero mean.
Figure 2.3: Weak value versus δ. The insets show 12((1 + δ)ψ+(x) − (1 − δ)ψ−(x)) for three
different values of δ.
Chapter 2. Weak Value Amplification of Photon Number - Theory 13
2.2 Weak Value Amplification of a Post-Selected Photon
2.2.1 Introduction
WVA of a single photon
Here, I will show how weak measurement can be used to amplify the effect of a single photon.
Due to this amplification, the effect of a single photon on a separate probe beam can be similar
to the effect due to multiple photons[44]. A single-photon signal pulse is sent through an
interferometer as shown in figure 2.4. Once inside the interferometer, the state of the signal
can be written as |i〉 = 1/√
2(|0〉a|1〉b + |1〉a|0〉b). A coherent state probe weakly measures the
number of photons in path a 2 The interferometer is made a little imbalanced so that an ideal
single-photon detector, in (what used to be) the dark port of the interferometer, can post-select
the signal in |f〉 = 1/√
2((1 + δ)|0〉a|1〉b − (1 − δ)|1〉a|0〉b), with δ 1. With this pre- and
post-selection, the weak value of number of photons in arm a can be calculated to be
〈na〉w =〈f |a†aaa|i〉〈f |i〉 =
1
2+
1
2δ(2.14)
where δ = 〈f |i〉 1 is the overlap between the initial and final state of the signal. The
parameter δ is called post-selection parameter hereafter.
Imbalanced Beam-Splitter
a
b
|ipr
|1is
Nonlinear medium
Figure 2.4: Schematic of the setup used to amplify the effect of a single photon.
Without the post-selection, one would expect that the mean value of the photon number in
arm a to be na = 1/2. But with the post-selection, the effect of the photon is amplified by the
term 1/2δ. It is worth noting that the weak value of photon number in arm b can be shown to
2Later I will discuss how one can realize such photon measurements by using a photon-photon interaction ina nonlinear medium[51].
Chapter 2. Weak Value Amplification of Photon Number - Theory 14
be 〈nb〉w = 1/2 − 1/2δ. Therefore, the total number of photons inside the interferometer will
be 〈na〉w + 〈nb〉w = 1; no extra photons are generated in the process of WVA. Note that the
presented formulas for the weak value are only valid in the limit δ 1.
Experimental realization of this amplification effect with single photons is extremely chal-
lenging. Single photon sources that are bright enough for this experiment to be done in a
feasible time are hard to set up and maintain. On top of that, losses in the path of the single
photon and inefficiencies in photon detection add to the difficulties of this experiment. These
experimental challenges can be overcome by using coherent states instead of single-photon fock
states for the signal. In what follows, I will show how doing post-selection (photon detection)
on a coherent state can result in an inferred additional photon in the mean value of photon
number, and then I will show how the effect of those additional photons can be amplified by
weak measurement.
The effect of post-selection on a coherent state
Here, I will show how by using a single-photon detector to detect the presence of a photon
in a coherent state signal pulse, in a certain limit, the inferred mean photon number of the
pulse, conditioned on successful detection, can increase by one photon. A careful analysis of
this problem using Bayesian statistics can be found in [52]. Here I will take a simple approach
to this problem by modeling the detector as the following. A single-photon detector with finite
efficiency η, can be modeled as an ideal detector with 100% efficiency preceded by a beam
splitter that transmits a fraction η of the incoming beam (“detected mode”) and reflects a
fraction 1− η of it (“undetected mode”), see figure 2.5
When the beam that is sent to the detector is a coherent state with mean photon number
n = |α|2, the two fields in the detected and undetected mode will also be coherent states |√ηα〉and |√1− ηα〉, respectively. Since the two coherent states in the two modes are independent of
each other, a photon detection on the detected mode wouldn’t change the inferred mean photon
number in the undetected mode. Therefore, the inferred mean photon number in the undetected
mode will always be nundetected = (1−η)|α|2. Now, since the detector after the the beam splitter
is assumed to be ideal, in the cases it doesn’t fire, with probability of Pno−click = e−η|α|2, the
inferred mean photon number in the detected mode will be zero. Therefore, when the detector
does not fire, referred to as “no-click” hereafter, the inferred mean photon number in the initial
coherent state will be nno−click = (1− η)|α|2 With the information above, we now calculate the
Chapter 2. Weak Value Amplification of Photon Number - Theory 15
Real Detector
Ideal Detector
|p↵idetected
|p
1 ↵iundetected
|↵i
Figure 2.5: A real detector with efficiency η modeled as an ideal detector and a beam splitterthat only transmits a fraction η of the incoming light. In the case where the incoming beamis a coherent state, the two “detected” and “undetected” modes are independent of each otherand the mean photon number inferred in the detected mode does not affect the mean photonnumber in the undetected mode.
mean photon number in the cases when the detector fires, referred to as “click”. We know that
the mean photon number calculated from both “click” and “no-click” cases should be equal to
n = |α|2. Therefore, we have
n = Pclick × nclick + Pno−click × nno−click
|α|2 = (1− e−η|α|2)nclick + e−η|α|2[(1− η)|α|2]
⇒ nclick =1
1− e−η|α|2 ×[|α|2 − e−η|α|2 [(1− η)|α|2]
].
(2.15)
It is easy to show that in the limit where η|α|2 1, the inferred mean photon number for “click”
cases become nclick = |α|2 +1. In this limit, the inferred mean photon number in the undetected
mode is approximately nundetected = |α|2 and the mean photon number3 in the detected mode is
very close to 0. In the unlikely events when the detector fires, the best estimate of mean photon
number in the detected mode will be ndetected = 1. Therefore, the overall mean photon number
when detector fires is nclick = ndetected + nundetected = |α|2 + 1. Figure 2.6 plots the difference
between the inferred mean photon number nclick and |α|2 for η = 0.2. It is evident that for
low |α|2 this difference approaches 1. As |α|2 becomes larger, this difference becomes smaller
3not the inferred photon number
Chapter 2. Weak Value Amplification of Photon Number - Theory 16
and approaches zero. This is due to the fact that for a large enough η|α|2, the higher photon
number terms in the coherent state in the detected mode become important and upon a photon
detection, the inferred mean photon number becomes ndetected = η|α|2 instead of simplify being
1. Therefore, the overall inferred mean photon number will be |α|2. In practice, there will
always be dark-counts, which cause the detector to fire when ideally it should not. The dark
counts degrade the effect of the post-selection, as shown in figure 2.6 with dashed line. A more
detailed study of the effect of the dark-counts can be found in [52].
0 5 10 15 20| |2
0.0
0.2
0.4
0.6
0.8
1.0
n clic
k|
|2
0 5 10 15 20| |2
01234
n clic
kn n
ocl
ick
Figure 2.6: The difference between the inferred mean photon number for the “click” events and|α|2 versus |α|2 for efficiency η = 0.2 (solid line.) The dashed line is when the effect of dark-counts is added to the inferred photon numbers. The inset shows the difference between nclickand nno−click versus |α|2, when no dark-counts are considered (solid) and when the dark-countsare considered (dashed.)
An experimental observation of the effect of this post-selected single photon was performed
in our group[50]. In the next section, I will show how this added photon by post-selection can
undergo weak value amplification.
2.2.2 Schrodinger picture
I will now extend the ideas presented in section 2.2 to when the signal pulse is a coherent state
instead of a single photon fock state. As shown in figure 2.7, the interaction between the signal
beam and the probe beam happens inside an interferometer. I will assume that the two arms
of the interferometer interact with the probe with different strengths. The propagator for the
interaction can therefore be written as
U = exp(− i~×∫H(t)dt) = exp(iφ1nprn1 + iφ2nprn2) (2.16)
Chapter 2. Weak Value Amplification of Photon Number - Theory 17
where npr, n1 and n2 are the number operators for the probe field and the field in the two
arms of the interferometer, respectively. φ1 and φ2 represent the interaction strength between
the probe and the signal in the two arms of the interferometer. The interaction causes a phase
shift on the probe which is linear in the number of photons in the signal, and φ1 and φ2 are the
magnitude of this phase shift due to a single photon in each arm. Therefore, by measuring the
total cross-phase shift (XPS) of the probe, one can measure the mean photon number in the
signal4. We assume the initial combined state of the probe and signal, before the interferometer,
Imbalanced Beam-Splitter
|ipr
Nonlinear medium
2
1
3
4
|↵is
Figure 2.7: Schematic of the WVA setup. The signal pulse |α〉s interacts with the probe|β〉probe inside an interferometer. The two arm interact with the probe with different strengthsrepresented as per photon XPS’s of φ1 and φ2, where φ1 > φ2.
is
|Ψi〉 = |α〉s|β〉pr. (2.17)
The creation operators for the signal pulse before the interferometer and the two modes of the
interferometer are related to one another by a†s = 1√2(a†2 − ia†1). Therefore, the state of the
signal inside the interferometer can be written as
|α〉s = exp(αa†s − h.c.)|vac〉
= exp(α√2
(−ia†1 + a†2)− h.c)|vac〉
= |−iα√2〉1|
α√2〉2.
(2.18)
4Note that probe beam also writes a nonlinear phase shift on the signal.
Chapter 2. Weak Value Amplification of Photon Number - Theory 18
After the interaction the state of the combined system becomes
|Ψf 〉 = exp(iφ1nprn1 + iφ2nprn2)|−iα√2〉1|
α√2〉2|β〉pr
=∑
n
[exp(iφ1nprn1 + iφ2nprn2)
]|−iα√
2〉1|
α√2〉2[e−|β|2
2βn√n!|n〉pr
]
=∑
n
e−|β|2
2βn√n!
[exp(−inφ1n1)|−iα√
2〉1][
exp(−inφ2n2)| α√2〉2]|n〉pr
=∑
n
e−|β|2
2βn√n!|−iαe
inφ1
√2〉1|αeinφ2
√2〉2|n〉pr
(2.19)
Before I continue with calculating the nonlinear phase shift written on the probe (the measured
number of photons of the signal), I will introduce the criteria for the weakness of the interaction.
Weakness of the interaction
The interaction mediated by the propagator in equation 2.16 is considered weak if the state
after the interaction is almost indistinguishable from the initial state, 〈Ψi|Ψf 〉 ≈ 1. Therefore,
we have
〈Ψi|Ψf 〉 =[
1〈−iα√
2|2〈
α√2|pr〈β|
][∑
n
e−|β|2
2βn√n!|−iαe
inφ1
√2〉1|αeinφ2
√2〉2|n〉pr
]
=∑
n
e−|β|2
2βn√n!〈−iα√
2|−iαe
inφ1
√2〉1〈
α√2|αe
inφ2
√2〉2〈β|n〉pr.
(2.20)
Knowing 〈β|n〉pr = exp(−|β|2/2) βn√n!
and 〈α1|α2〉 = exp(− |α1|22 − |α2|2
2 + α1α2), equation 2.20
becomes
〈Ψi|Ψf 〉 =∑
n
e−|β|2 |β|2nn!
exp(−|α|2
2(1− einφ1)) exp(
−|α|22
(1− einφ2)). (2.21)
To simplify equation 2.21, I assume |β|2, which is the mean photon number in the probe, and
both φ1 and φ2 are sufficiently small, so that we have nφ1,2 1. It is worth noting that although
n is unbounded and can be arbitrarily large, |β|2 can be chosen so that the probability of having
an n-photon term that violates nφ1,2 1 is negligible. Under these assumption, exp(inφ1,2)
Chapter 2. Weak Value Amplification of Photon Number - Theory 19
can be written as 1 + inφ1,2 for all probable n’s in the probe. Hence, equation 2.21 becomes
〈Ψi|Ψf 〉 =∑
n
e−|β|2 |β|2nn!
exp(in|α|2φ1
2) exp(
in|α|2φ2
2)
=∑
n
e−|β|2 |β|2nn!
exp(in|α|2φ)
=∑
n
e−|β|2 (|β|2 exp(i|α|2φ))n
n!
(2.22)
where φ = (φ1 + φ2)/2. Equation 2.22 can be simply written as
〈Ψi|Ψf 〉 = exp(−|β|2) exp(|β|2ei|α|2φ). (2.23)
Expanding ei|α|2φ to second order in φ yields
〈Ψi|Ψf 〉 = exp(i|α|2|β|2φ) exp(−|β|2|α|4φ2
2). (2.24)
From equation 2.24, it is evident that for the interaction between the probe and the signal to
be weak, 〈Ψi|Ψf 〉 ≈ 1, we should have
|β|2|α|2φ 1,
|α|2φ 2
β.
(2.25)
The first term in equation 2.25 can be seen as an overall phase and can be compensated for
(see [44].) The second term is the more interesting condition: |α|2φ is the average XPS written
on the probe, and 2/β is the minimum phase uncertainty of a coherent state with mean photon
number of |β|2, imposed by the number-phase uncertainty ∆φ∆n ≤ 1/2. Therefore, the second
condition in equation 2.25 indicates that for the interaction to be weak, the average XPS on the
probe should be much smaller than the initial phase uncertainty of the probe, which is same
condition mentioned in sections 2.1.1 and 2.1.2.
Expected XPS on the probe
I now continue from equation 2.47 to calculate the expected XPS written on the probe, when
a single photon is detected in the nearly-dark port of the interferometer. I model the imbal-
anced interferometer by writing the transformation matrix of the final beam splitter in the
Chapter 2. Weak Value Amplification of Photon Number - Theory 20
interferometer as a1
a2
=
1+δ√2
i1−δ√2
i1−δ√2
1+δ√2
a3
a4
(2.26)
for δ 1, where the parameter δ quantifies the imbalance in of the interferometer. With this
transformation, we have
|−iαeinφ1
√2〉1|αeinφ2
√2〉2
=|−iαeinφ1
√2
(1 + δ√
2) +−iαeinφ2
√2
(1− δ√
2)〉3|−iαeinφ1
√2
(−i1− δ√2
) +αeinφ2
√2
(1 + δ√
2)〉4
=| − iαeinφ
2
[e−in
φ1−φ22 (1 + δ) + ein
φ1−φ22 (1− δ)
]〉3|αeinφ
2
[− e−in
φ1−φ22 (1− δ) + ein
φ1−φ22 (1 + δ)
]〉4
(2.27)
where φ = (φ1 + φ2)/2. I define (φ1 − φ2) ≡ ∆φ. If we again assume nφ1,2 1, equation 2.27
becomes
|−iαeinφ1
√2〉1|αeinφ2
√2〉2
=|−iαeinφ
√2
[(1 + i
n∆φ
2)(1 + δ) + (1− in∆φ
2)(1− δ)
]〉3
|αeinφ
√2
[− (1− in∆φ
2)(1− δ) + (1 + i
n∆φ
2)(1 + δ)
]〉4
=| − iαeinφ〉3|αeinφ(in∆φ
2+ δ)〉4.
(2.28)
Next, we assume n∆φ δ. The importance of this assumption will be further discussed later.
With this assumption, one can write
in∆φ
2+ δ ≈ δe in∆φ
2δ . (2.29)
Equation 2.28 then becomes
|−iαeinφ1
√2〉1|αeinφ2
√2〉2 = | − iαeinφ〉3|αδein(φ+ ∆φ
2δ)〉4 (2.30)
and the final state of the probe and signal, equation 2.47, will be
|Ψf 〉 =∑
n
e−|β|2
2β2
√n!| − iαeinφ〉3|αδein(φ+ ∆φ
2δ)〉4|n〉pr (2.31)
Chapter 2. Weak Value Amplification of Photon Number - Theory 21
Now I calculate the expected XPS written on the probe, when the measurement is condi-
tioned on the single-photon detection in the dark port (mode “4” in figure 2.7.) For now I
assume a detector with 100% efficiency and later will discuss the effect of finite efficiency. For
αδ 1, the field in mode 4 of equation 2.30 can be written as
|αδein(φ+ ∆φ2δ
)〉4 = |0〉4 + αδein(φ+ ∆φ2δ
)|1〉4. (2.32)
Detecting a photon in the dark port is equivalent to post-selecting the field in mode 4 to be in
|1〉4. Therefore, upon post-selection, the combined state of the probe and signal, equation 2.31,
becomes
4〈1|Ψf 〉 = αδ∑
n
e−|β|2
2β2
√n!ein(φ+ ∆φ
2δ)|ineinφ〉3|n〉pr. (2.33)
This post-selection happens with probability Pps = |α|2δ2. After renormalizing 4〈1|Ψf 〉 and
writing the coherent state in mode 3 in terms of number states we will have
|Ψclick〉 =4〈1|Ψf 〉|αδ| =
∑
n,m
e−|β|2−|α|2
2
[βei(φ+∆φ/2)
]n√n!
[− iαeiφ
]m√m!
|m〉3|n〉pr
=∑
m
e−|α|2
2(iα)m√m!|m〉3|βei
[(m+1)φ+ ∆φ
2
]〉pr.
(2.34)
Therefore, for each m-photon term in mode 3, the XPS written on the probe is φm = (m+1)φ+
∆φ/2δ (for more information on operational definition of phase measurement see [53, 54].) I
define Pm = exp(−|α|2)×|α|2m/m! to be the probability of having m photons in mode 3. Then
the average XPS on the probe will be
Φclick =∑
m
Pm × φm = φ∑
m
mPm + (φ+∆φ
2δ)∑
m
Pm
Φclick = |α|2φ+ φ+∆φ
2δ.
(2.35)
It is evident, from equation 2.35, that when the detector fires, the XPS inversely grows
with the parameter δ. To compare the equation 2.35 with equation 2.14, we set φ2 = 0 in
equation 2.35, which is equivalent to only measuring the mean photon number in arm 1 in
the interferometer. With this assumption, we will have φ = φ1/2 and ∆φ = φ1. Therefore,
Chapter 2. Weak Value Amplification of Photon Number - Theory 22
equation 2.35 becomes
Φclick =[ |α|2
2+
1
2+
1
2δ
]× φ1
⇒〈n1〉w =Φclick
φ1=|α|2
2+
1
2+
1
2δ,
(2.36)
which gives a similar result as equation 2.14, apart from the term |α|2/2. The added term is
due to the fact that we used a coherent state for the signal pulse. The other term, 12 + 1
2δ , is
the added photon due to post-selection and it’s amplified effect. To make this more clear, I will
investigate what the XPS on the probe would be, if the detector failed to fire, which is to say
we post-select on |0〉4 in equation 2.32. With this post-selection, equation 2.31 becomes
|Ψno−click〉 =4 〈0|Ψf 〉 =∑
n
e−|β|2
2βn√n!| − iαeinφ〉3|n〉pr
=∑
n,m
e−|β|2−|α|2
2βn√n!
(−iα)m√m!
eimnφ|m〉3|n〉pr
=∑
m
e−|α|2
2(−iα)m√
m!|m〉3|βeimφ〉pr.
(2.37)
Following the steps taken in derivation of equation 2.35, it is easy to show that the average
XPS when the detector fails to click is given by
Φno−click = |α|2φ. (2.38)
When the backaction dominates
In deriving the expected XPS’s for “click” and “no-click” cases, I made the assumption of
n∆φ δ, which led to equation 2.29. Here I will elaborate further on the meaning of this
assumption. In order to post-select on an unlikely final state, we made the interferometer
shown in figure 2.7 imbalanced. The amount by which the interferometer is imbalanced is
quantified by parameter δ, and the probability of post-selection happening, for an ideal detector,
is Pps = δ2|α|2 (under the δ 1 assumption.) What I have neglected here is the fact that
the post-selection could have happened due to the backaction of the measurement. Since the
two arms of the interferometer interact with the probe with different strength, the probe writes
different XPS’s on the two signal fields in the two arms of the interferometer. This difference
in phase shifts makes the interferometer inherently imbalance. For each n-photon term in the
Chapter 2. Weak Value Amplification of Photon Number - Theory 23
probe, the difference between the two phase shifts is given by n∆φ. Therefore, if the post-
selection is dominated by the back-action of the measurement and not by the parameter δ,
when n∆φ > δ, one should not expect any weak-value amplification. To further illustrate this,
I will calculate the expected XPS on the probe in the extreme case where δ = 0, where the
post-selection only happens due to the back-action.
With δ = 0, equation 2.28 becomes |− iαeinφ〉3|αeinφ(in∆φ)/2〉4. Hence the combined state
of the probe and the signal after post-selecting on detecting a single photon in mode 4 becomes
|Ψps〉 =∑
n
e−|β|2
2βn√n!
(inα∆φ
2einφ)| − iαeinφ〉3|n〉pr (2.39)
It is easy to show that the relative phase between different n-photon terms in probe for each
m-photon term in the mode 3 is given by (m + 1)φ. Therefore, the expected XPS when the
post-selection is dominated by the measurement backaction is given by
Φδ=0 = (|α|2 + 1)φ. (2.40)
It can be seen that there is still one photon worth of extra XPS due to the post-selection, but
the amplification effect is absent.
It is worth noting that the wavefunction given in equation 2.39 is not normalized. The
normalization factor is given by
〈Ψps|Ψps〉 =∑
m,n
e−|β|2/2 β
m
√m!
(−imα∆φ
2e−imφ)3〈−iαeimφ|pr〈m|
e−|β|2/2 β
n
√n!
(inα∆φ
2einφ)| − iαeinφ〉3|n〉pr
=∑
m,n
e−|β|2 βmβn√
m!n!(nm|α|2 ∆φ2
4)ei(n−m)φ〈−iαeimφ| − iαeinφ〉3δn,m
=∑
n
e−|β|2 |β|2nn!
n2 |α|2∆φ2
4=|α|2∆φ2
4〈n2pr〉β
(2.41)
where 〈n2pr〉β is the expectation value of the square of the number operator of the probe field
given coherent state |β〉pr. This quantity is related to the variance of the photon number
distribution in the probe field. This normalization factor can also be seen as the probability
of the post-selection happening due to the backaction, and as discussed in section 2.1.2, this
probability depends on the variance of the photon number distribution of the probe field.
Chapter 2. Weak Value Amplification of Photon Number - Theory 24
In this section, I used the Schrodinger equation to calculate the amplified effect of a post-
selected single photon on a probe in the limit where δ 1. In the next section, I will take an
equivalent approach, the weak value approach, to calculate the weak value of photon number
with arbitrary post-selection parameter. This approach will enable us to easily calculate the
expected XPS on a probe for every δ and to consider the effect of detector with finite efficiency.
2.2.3 Weak Value Approach
In this section, I will use the weak value formula given in equation 2.9 to calculate the ex-
pectation value of the mean photon number in the two arms of the interferometer, for all
post-selection parameters δ. I will also consider the effect of a detector with finite efficiency.
The operator being measured is n1(2) = a†1(2)a1(2), the photon number operator in each arm
of the interferometer. The initial state of the signal is given by |i〉 = |α〉s. The final state is
detecting a single photon at the single-photon detector, which is detecting a photon in mode
“5” in figure 2.8. The weak values are then defined as
〈n1(2)〉w =〈f |a†1(2)a1(2)|i〉
〈f |i〉 (2.42)
2
1
3
4
|ipr
|↵is
Imbalanced Beam-Splitter
5
6
Ideal Detector
Detector with finite efficiency
Nonlinear medium
Figure 2.8: Schematic of the WVA setup. The signal pulse |α〉s interacts with the probe|β〉probe inside an interferometer. The two arm interact with the probe with different strengthsrepresented as per photon XPS of φ1 and φ2, where φ1 > φ2. The effect of a detector with finiteefficiency is modeled as a combination of a beam splitter and a detector with 100% efficiency.
Chapter 2. Weak Value Amplification of Photon Number - Theory 25
I first model the imbalanced interferometer by writing the transformation matrix of the
imbalanced beam splitter as
a1
a2
=
cos θ sin θ
− sin θ cos θ
a3
a4
(2.43)
where the parameter θ ∈ [−3π/4,−π/4] determines the amount by which the beam splitter in
imbalanced. The parameter θ is related to the previously introduced post-selection parameter
δ by δ = (cos θ+sin θ)/√
2; θ = −3π/4 corresponds to δ = 1 (post-selecting on the initial state)
and θ = −π/4 corresponds to δ = 0 (post-selecting on a completely orthogonal state.) With
this model, the state of the signal after the interferometer can be written as
|α〉s = | α√2
(cos θ − sin θ)〉3|α√2
(cos θ + sin θ)〉4 (2.44)
Similar to figure 2.5, the detector in figure 2.8 can be modeled as an ideal detector proceeded
with a beam splitter that partially transmits the beam. This beam splitter can be modeled as
a4 = ra5 + ta6 (2.45)
with t2 = η and r2 = 1− η, η being the efficiency of the detector. With this model it’s easy to
show that the initial state of the signal can be written as
|i〉 = | α√2
(cos θ − sin θ)〉3|tα√2
(cos θ + sin θ)〉5|rα√2
(cos θ + sin θ)〉6 (2.46)
The final state is detecting a single photon at the detector
|f〉 = | α√2
(cos θ − sin θ)〉3|1〉5|rα√2
(cos θ + sin θ)〉6 (2.47)
Now, in order to calculate the weak values given in equation 2.42, we need to write n1
and n2 in terms of creation and annihilation operators in modes “3”,“5” and “6”. With some
Chapter 2. Weak Value Amplification of Photon Number - Theory 26
algebra, it can be shown that
a†1a1 = cos2 θa†3a3 + r sin θ cos θa†3a6 + t sin θ cos θa†3a5
r sin θ cos θa†6a3 + r2 sin2 θa†6a6 + rt sin2 θa†6a5
t sin θ cos θa†5a3 + rt sin2 θa†5a6 + t2 sin2 θa†5a5
(2.48)
If we now define α3 = α√2(cos θ− sin θ), α5 = t α√
2(cos θ+ sin θ) and α6 = r α√
2(cos θ+ sin θ),
the weak value of number of photons in mode 1 becomes
〈f |a†1a1|i〉〈f |i〉 = cos2 θ|α3|2 + r sin θ cos θα3α6 + t sin θ cos θα3α5
+ r cos θ sin θα3α6 + r2 sin2 θ|α6|2 + rt sin2 θα6α5
+ t sin θ cos θα3
α5+ rt sin2 θ
α6
α5+ t2 sin2 θ
(2.49)
which can be simplified to
〈n1〉w =|α|2
2+
sin θ
cos θ + sin θ− η|α|2
2[sin θ(sin θ + cos θ)] (2.50)
A similar calculation reveals
〈n2〉w =|α|2
2+
cos θ
cos θ + sin θ− η|α|2
2[cos θ(sin θ + cos θ)] (2.51)
To see the relation between equations 2.50 and 2.51 and the weak value calculated in equa-
tion 2.36, I will first assume that η|α|2 1. Then, by expanding θ near θ0 = −π/4, and after
replacing θ with δ (δ = (cos θ + sin θ)/√
2), we arrive at
〈n1〉w =|α|2
2+
1
2+
1
2δ
〈n2〉w =|α|2
2+
1
2− 1
2δ
(2.52)
Assuming an XPS of φ1(2) for mode 1(2) of the interferometer, the expected XPS written on
the probe will be
Φclick = φ1〈n1〉w + φ2〈n2〉w = (|α|2 + 1)φ1 + φ2
2+φ1 − φ2
2δ(2.53)
which is exactly the same as what was calculated in equation 2.35, with φ = (φ1 + φ2)/2 and
Chapter 2. Weak Value Amplification of Photon Number - Theory 27
∆φ = φ1 − φ2
It is worth noting that the total number of photons inside the interferometer is given by
〈n1〉w + 〈n2〉w = |α|2 − η|α|2δ2 + 1 (2.54)
where |α|2− η|α|2δ2 is sum of mean photon numbers in the undetected modes 3 and 6, and the
added photon is due to post-selection in mode 5, see figure 2.8. Figure 2.9 plots 〈n1〉w,〈n2〉wand 〈n1〉w + 〈n2〉w versus post-selection parameter δ for |α|2 = 1 and η = 0.2.
0.0 0.2 0.4 0.6 0.8 1.015
10
5
0
5
10
15
n
n1 w
n2 w
n1 w + n2 w
Figure 2.9: Weak value of number of photons in the two arms of the interferometer versus thepost-selection parameter δ. For this plot, |α|2 and η are chosen to be 1 and 0.2 respectively.
It is easy to show that equations 2.50 and 2.51 can be written in terms of the post-selection
parameter δ as
〈n1(2)〉w =|α|2
2+
1
2(1±
√1
δ2− 1)− η|α|2
2δ(δ ±
√1− δ2) (2.55)
So far I assumed that the post-selection is done by using a single-photon detector in port 4
of the interferometer in figure 2.8, which is to say we post-select on a single-photon fock state
in that mode. But in reality, the detectors used in a laboratory usually only detect the presence
of a photon and cannot distinguish between detecting a single photon and detecting multiple
photons. Therefore, for my treatment to be accurate, I need to assume that mean photon
number at the ideal detector in mode 5 is very low, so that detecting more than one photon is
Chapter 2. Weak Value Amplification of Photon Number - Theory 28
extremely unlikely, |α|2ηδ2 1. In the next section, I will replace the single-photon detector
with a Photon-Number Resolving (PNR) detector[55] and will calculate the weak values when
the measurement is conditioned on detecting an n-photon state at the PNR.
2.2.4 Weak Values and PNR Detectors
In this section, I will calculate the weak value of photon number in the two arms of an interfer-
ometer as shown in figure 2.8, by post-selecting on detecting an n-photon state at the detector
in mode 5. As before, the initial state of the signal is given by equation 2.46. The final state
in which the signal is post-selected is given by
|f〉 = | α√2
(cos θ − sin θ)〉3|n〉5|rα√2
(cos θ + sin θ)〉6. (2.56)
Following a calculation similar to the one carried out in the previous section, the weak values
can be shown to be
〈n1〉w∣∣∣n
=|α|2
2+ n
sin θ
cos θ + sin θ− η|α|2
2[sin θ(sin θ + cos θ)] (2.57)
and
〈n2〉w∣∣∣n
=|α|2
2+ n
cos θ
cos θ + sin θ− η|α|2
2[cos θ(sin θ + cos θ)] (2.58)
These equations are very similar to equations 2.50 and 2.51, except that the inferred mean
photon number has increased by a factor of n. Most importantly, these equations show that the
effect of all n detected photons will be amplified due to the weak value amplification. Figure
2.10 shows the weak value of photon number in arm “1” of the interferometer (figure 2.8) versus
post-selection parameter δ. It can be seen that post-selecting on higher photon numbers at the
PNR detector results in larger amplification.
As mentioned in section 2.2.3, when using a single-photon detector instead of a PNR detec-
tor, we should either consider the effect of detecting higher photon numbers or limit ourselves to
the regime where the probability of detecting higher photons are low. Making the |α|2ηδ2 1
assumption, ensured the latter in previous section. Here I will investigate the effect of detecting
higher photon term on the weak value. For this, I only consider two detection outcomes at the
single-photon detector: the detector either detects the presence of a photon (“click”) or it does
not (“no-click”). The “no-click” case corresponds to 〈n1〉w∣∣∣n
for n = 0. Given a signal coherent
Chapter 2. Weak Value Amplification of Photon Number - Theory 29
0.0 0.2 0.4 0.6 0.8 1.00
2
4
6
8
10
12
14
n 1w
| n
n = 0n = 1n = 2n = 3n = 4
Figure 2.10: The weak value of number of photons in arm “1” of the interferometer versuspost-selection parameter, for post-selecting on different n’s at the PNR detector. For this plot,|α|2 and η are chosen to be 1 and 0.2 respectively.
state |α|2, efficiency η and post-selection parameter δ, the probability of a “no-click” outcome
is
Pno−click =∣∣〈0|√ηαδ〉
∣∣2 = e−η|α|2δ2
(2.59)
Therefore, the probability of a “click” outcome is given by
Pclick = 1− Pno−click = 1− e−η|α|2δ2(2.60)
The probability of each n-photon detection is given by
Pn =∣∣〈n|√ηαδ〉
∣∣2 = e−η|α|2δ2 (η|α|2δ2)n
n!(2.61)
Now we can calculate the effect higher photon detection events on the weak value of photon
number by
〈n1〉w∣∣∣click
=∑
n=1
Pn × 〈n1〉w∣∣click
Pclick(2.62)
Figure 2.11 compares the weak value of photon number in mode “1” in two cases: when the
higher photons terms are considered and when only the detecting a single photon is considered.
The parameters in figure 2.11a are chosen so that the assumption |α|2ηδ2 1 remains valid
Chapter 2. Weak Value Amplification of Photon Number - Theory 30
0.0 0.2 0.4 0.6 0.8 1.00
2
4
6
8
10
n 1w
Pn ×nn =1
(a)
0.0 0.2 0.4 0.6 0.8 1.00
2
4
6
8
10
n 1w
Pn ×nn =1
(b)
Figure 2.11: The weak value of number of photons in arm “1” versus δ, calculated by consideringhigher photon terms,
∑Pnn, and only post-selecting on a single photon, n = 1. For this plot,
|α|2 and η are chosen to be 1 and 0.5 for a) and 10 and 0.5 for b).
for most δ’s and it can be seen that the two methods don’t deviate from each other by much.
The difference, however, is more pronounced in 2.11b, where the parameters are chosen so that
the assumptions is only valid for δ 1. As it can be seen, in 2.11b, one would underestimate
the weak value when not considering the effect of detection of higher photon terms.
2.2.5 Measuring the Amplified Effect of a Post-selected Single Photon
The weak value of photon numbers in arm 1 and 2 of the interferometer for “click” and “no-
click” cases from equations 2.57 and 2.58 can be written as
〈n1(2)〉click =|α|2 + 1
2± 1
2δ
〈n1(2)〉no−click =|α|2
2
(2.63)
The term |α|2/2 is common in both “click” and “no-click” cases. Therefore, by subtracting
the measured effect of mean photon number when detector fires from when it doesn’t fire, one
can directly measure the effect of a single photon. If we assume a per photon XPS of φ1(2) for
arm 1(2) of the interferometer, looking at the difference between the measured XPS when the
detector fires, φclick, and the measured XPS when it doesn’t, φno−click, will result in
φclick − φno−click =φ1 + φ2
2+φ1 − φ2
2δ. (2.64)
Chapter 2. Weak Value Amplification of Photon Number - Theory 31
Evidently, this quantity is independent of |α|2 and only shows the amplified the post-selected
photon.
2.2.6 Weak Values vs. Strong Values - ABL Rule
In this section, I will calculate the mean photon number in arm 1 of the interferometer, as shown
in figure 2.8, conditioned on post-selecting a single photon in arm 4, but this time assuming a
strong interaction between the signal and probe in arm 1. When the interaction is strong, the
two fields in arm 1 and 2 will no longer be coherent, hence no interference will happen at the
beam splitter that closes the interferometer. Because of this, the weak value formalism will no
longer be applicable and one should use the ABL rule to calculate the mean photon number in
arm 1.
The ABL rule, introduced by Aharonov, Bergmann and Lebowitz[56], evaluates the proba-
bility of a certain measurement outcome, n, when the measurement is conditioned on having a
certain initial state |i〉 and post-selecting on a certain final state |f〉
P (n|i, f) =|〈f |n〉〈n|i〉|2
∑
n
|〈f |n〉〈n|i〉|2(2.65)
Knowing the conditional probability of each outcome, the mean value can be calculated as
n =∑P (n|i, f)× n.
I will now calculate the mean value of number of photons in mode 1, using the ABL rule.
To make the calculations easier, I will limit the treatment to the case where the interferometer
is only slightly imbalanced, where the WVA occurs. Since the detected photon in port 4 could
have come from either of the arms of the interferometer, I should first calculate P (n1, n2|i, f),
which the conditional probability of having n1 photons in arm 1 and n2 photons in arm 2 given
the pre- and post-selection. For |i〉 = |α/√
2〉1|α/√
2〉2 and post-selecting on |f〉 = |α〉3|1〉4, one
can show
〈f |n1, n2〉1,2〈n1, n2|i〉 =
(α/√
2)2(n1+n2)
n1!n2!α
(− n1e
−δ + n2eδ). (2.66)
Knowing 〈f |n1, n2〉1,2〈n1, n2|i〉, one can numerically evaluate P (n1, n2|i, f) using equation 2.65.
Once P (n1, n2|i, f) is known, the photon number distribution in arm 1, conditioned on i and
Chapter 2. Weak Value Amplification of Photon Number - Theory 32
f , can be calculated by
P (n1|i, f) =∑
n2
P0(n2)P (n1, n2|i, f), (2.67)
where P0(n2) is the initial photon number distribution in arm 2. Figure 2.12 plots the numerical
result of the mean photon number number using ABL rule for |α|2 = 2, and compares the result
with the weak values calculated for the same |α|2. As it can be seen, the weak value of the
mean photon number increases as δ gets smaller, whereas the mean value calculated using the
ABL rule remains constant.
0.00 0.05 0.10 0.15 0.20
0
2
4
6
8
10
12
14
n
ABL ruleWeak Value
Figure 2.12: Comparison between the weak value of mean photon number versus the strongvalue calculated using ABL rule. As expected the strong value remains unchanged as the post-selection parameter changes whereas the weak value increases as the post-selection parameterdecreases.
Figure 2.12 shows that if one were to measure the mean photon number in arm “1” of the
interferometer conditioned upon detecting a single photon in the nearly-dark port, the result
will be amplified if the measurement is done weakly. But if the measurement is performed
strongly, no anomalous result should be expected.
2.3 Weak Measurement as a POVM
Positive-Operator Valued Measure (POVM) is the most generalized formulation of measurement
in quantum mechanics. Often the probe and the system in a measurement belong to different
Hilbert spaces, and because of that a projective measurement on the probe cannot be written
Chapter 2. Weak Value Amplification of Photon Number - Theory 33
as a projective measurement on the system. Every measurement in quantum mechanics can be
described by a POVM and weak measurement is no exception. Here I will present a POVM
treatment of the photon number measurement described in this chapter. For simplicity I make
the assumption that the measurement can only have two results and will limit the calculation
to real post-selection parameter and real weak values. A more detailed calculation of what
follows, for a different system, can be found in[57, 49].
As described in section 2.2.2, the propagator of the interaction, in the weak limit and when
only one arm interacts with the probe, can be written as
U = eiφ0nsnpr = I + iφ0nsnpr (2.68)
where φ0 1 is the per photon phase shift (describing the interaction strength) and ns and
npr are photon number operator for the signal and probe respectively. For a probe initialized
in a state |ψ〉pr, we define the Kraus operator for a phase measurement on the probe as
Kφ = 〈φ|eiφ0nsnpr |ψ〉pr
= 〈φ|ψ〉pr + iφ0ns〈φ|npr|ψ〉pr.(2.69)
With analogy to the relation between position and momentum operators x and p, one can
write the photon number operator in phase basis as 〈φ|npr|φ′〉 = −iδ(φ − φ′) ∂∂φ . Assuming a
Gaussian function for the initial wavefunction of the probe, ψpr(φ) = Nexp(φ2/2σ2), N being
the normalization factor, we will have
〈φ|npr|ψ〉pr = −i φσ2ψpr(φ) (2.70)
and equation 2.69 becomes
Kφ = (I + φ0nsφ
σ2)ψpr(φ). (2.71)
I will consider a coarse grained measurement so that for all φ ≤ 0 the readout is −1 and
for φ > 0 the readout is 1. The corresponding quantum operations on a signal described by
Chapter 2. Weak Value Amplification of Photon Number - Theory 34
density matrix ρs can be written as
E−ρs =
∫ 0
−∞dφ KφρsK
†φ
E+ρs =
∫ ∞
0dφ KφρsK
†φ
(2.72)
Performing the integrals, to the first order in φ0, the operations become
E±ρs =1
2
[ρs ±
φ0
σ√
2(nsρs + ρsns)
]. (2.73)
The term (nsρs+ρsns) describes the measurement disturbance on the system. It is evident that
in the limit where no information is gained from the measurement, φ0 σ, the measurement
will not disturb the system. Defining measurement outcome O ∈ −1, 1 and C = φ0
σ√π
,
equation 2.73 can be written as
EOρs =1
2
[ρs +OC(nsρs + ρsns)
]. (2.74)
We are interested in finding the expectation value of O conditioned on pre-selecting the
system in state |i〉 and post-selecting the system in state |f〉. This expectation value is described
by
〈O〉 =∑
OOP (O
∣∣f, i) (2.75)
where P (O∣∣f, i) is the conditional probability of outcome O given the initial and final state of
the system. Using the identity P (O∣∣f, i) = P (O, f
∣∣i)/P (f∣∣i), equation 2.75 becomes
〈O〉 =∑
OO〈f |EO|i〉〈i|f〉|〈f |i〉|2
=∑
O
O2
|〈f |i〉|2 +OC[〈f |ns|i〉〈i|f〉+ 〈f |i〉〈i|ns|f〉]|〈f |i〉|2
=∑
O
O2
[1 +OC 〈f |ns|i〉〈f |i〉
].
(2.76)
Doing the summation over O, we find
〈O〉 = C〈f |ns|i〉〈f |i〉 . (2.77)
Chapter 2. Weak Value Amplification of Photon Number - Theory 35
Hence the weak value of ns is defined as
〈ns〉w =〈f |ns|i〉〈f |i〉 = 〈O
C〉. (2.78)
The parameter C = φ0/σ√π can be thought of as a calibration factor that readjusts the outcome
of the coarse grained measurement of the probe for different measurement strength. A simple
example can illustrate the importance of the calibration factor C. Assume that we have built
a photon number measurement apparatus, with only two outcomes as mentioned above. For
calibration, we use this device to measure the number of photons in a signal pulse that contains
a single photon. If we get the outcome “1” 55% of the time and measure “-1” 45% of the time,
the expectation value of O will be 0.1. Therefore, in order for this measurement apparatus to
correctly evaluate the mean photon number in that pulse, we need to divide the mean value of
O by 0.1. With this calibration factor, if for a different pulse with unknown photon number we
measure 〈O〉 = 0.01, we know that the mean photon number is 0.1 and if we measure 〈O〉 = 0.5,
we know that the mean photon number is 5.
Chapter 3
Apparatus
As introduced in the previous chapter, in order to observe the amplified effect of a post-selected
single photon, one needs to send the signal through an interferometer. Once inside the in-
terferometer, the signal beam interacts with a coherent state probe via a nonlinear medium,
where the interaction strength is different in each arm. The effect of the signal on the probe
manifests itself as a cross-phase shift (XPS), which is linear in the number of photons in the
signal. After the interaction, a single-photon detector detects the presence of a single photon in
the nearly-dark port of the interferometer. The “darkness” of this port is controlled by using
an imbalanced beam splitter. The information about the number of signal photons inside the
interferometer is obtained by measuring the nonlinear phase shift written on the probe by the
signal, See figure 2.7. Therefore, the key elements for realizing this amplification effect can be
listed as
• A nonlinear medium
• An interferometer in which the signal beam in each arm interacts with the probe with
different strengths
• A beam splitter whose reflectance and transmittance can be precisely controlled
• A single photon detector
• A precise phase measurement apparatus
In practice, the nonlinear interaction between the signal and probe beam is mediated by a
sample of cold 85Rb atoms. Using electromagnetically induced transparency (EIT), we enhance
36
Chapter 3. Apparatus 37
the nonlinearity by using atomic resonance while eliminating absorption of the probe. The
two paths of the interferometer are made up of two orthogonal circular polarizations, which
overlap spatially. A certain choice of beam geometry and atomic level scheme ensures that
the interaction between the probe and the signal pulse is different for the two polarizations.
Combination of a quarter-wave plate (QWP), a half-wave plate (HWP) and a polarizer enables
us to “post-select” the polarization of the signal on an arbitrary final polarization. A Single-
Photon Counting Module (SPCM) is used to detect the presence of a single photon in the dark
port of the interferometer. The phase of the probe is measured using a beat-note interferometry
technique. In what follows in the rest of this chapter, all of these elements will be introduced
and discussed in detail.
3.1 Magneto-Optical Trap
The element used to mediate the interaction between the signal and the probe in this experiment
is a cloud of cold rubidium 85 atoms. Rubidium is an alkali metal that has one free electron
in the 5s orbital. Due to fine-structure splitting, the excited state 5p is split into a doublet
and the two transitions 52S1/2→ 52P1/2
and 52S1/2→ 52P3/2
, are referred to as the D1 line
(794.98nm) and the D2 line (780.24nm), respectively. Each of these transitions, in turn, have
additional hyperfine structure due to the coupling between the total angular momentum of
the electron and the angular momentum of the nucleus. The details of the hyperfine structure
depends on the isotope of interest, since different isotopes have nuclei with different spin. The
two naturally occurring isotopes of rubidium are 85Rb (stable) with nucleus angular momentum
of I = 5/7 and natural abundance of 72.2% and 87Rb (half-life of 49 billion years) with nucleus
angular momentum of I = 3/2 and natural abundance of 27.8%. For 85Rb, which is used for
this experiment, the hyperfine splitting splits the ground state into two ground states with total
angular momenta of F = 2 and F = 3. The excited state of the D2 line splits into 4 excited
states with total angular momenta of F ′ = 1 , 2 , 3 and 4, with lifetime of 26.23ns (decay rate
of Γ = 2π× 6.06 MHz.) In what follows I will show how these hyperfine structure of D2 line of
85Rb can be used for laser cooling and trapping as well as mediating the interaction between
the probe beam and the signal pulse (for more information about Rubidium atoms, see [58].)
Magneto-Optical Trapping (MOT)1 is a way of cooling and trapping atoms that makes
1also not very commonly called Zeeman Assisted Radiation Pressure Trap (ZARPT) or Zeeman assistedOptical Trap (ZOT)
Chapter 3. Apparatus 38
use of a combination of an optical force and a magnetic field gradient. The scattering force
,+ ,I I-1 0 +1
z
E
J=0
J=1
J=0
J=1
mj = 0
mj = 0
mj = -1
mj = +1
+
B(z)
Figure 3.1: Schematic of a one-dimensional MOT setup. Top: Two coils in anti-Helmholtzconfiguration create a magnetic field gradient. Two optical fields, with the same circular polar-ization (opposite spins due to opposite propagation direction), are sent through the center ofthe magnetic field. The two optical fields are red-detuned from the J = 0 → J′ = 1 transitionof the atoms placed near the center of the magnetic field gradient. Bottom: the Zeeman shiftedenergy levels of an atom in the magnetic field gradient.
experienced by a cloud of atoms in a MOT is both velocity dependent (due to the Doppler
effect) and position dependent (due to the position dependent Zeeman shifts in the presence of
the magnetic field gradient.) Therefore, the atoms, which feel this force, can be seen as damped
harmonic oscillators. A certain choice of optical field intensity, detuning and magnetic field
gradient can result in cooling the atoms down to (usually) below their Doppler temperature
(defined as TDoppler = ~Γ/2kB, where Γ is the natural linewidth of the transition and kB is the
Boltzmann’s constant) and trapping them at the center of the magnetic field gradient. Figure
3.1 shows the schematic of a one-dimensional MOT setup.
The MOT setup used in our experiment has five main components: trapping beam, re-
pumper beam, magnetic field coils, optical pumping beam and the computer which controls the
tasks and atomic cycles. Each of these components will be discussed in detail in what follows
in this section.
Chapter 3. Apparatus 39
3.1.1 Trapping Beam
The key component of the MOT setup is the trapping beam, which provides the radiation
pressure needed for laser cooling and optical molasses. Having a stable trapping beam in power
and frequency, enables maintaining a healthy MOT for a long period of time, which is a necessity
for long data runs. In designing a stable apparatus, one should avoid long free space optical
paths, specially in a lab that is not climate controlled. To achieve stability, we have designed
our trapping beam setup in a modular way with very short paths between each module. Figure
3.2 shows different modules used in the setup and their relative position.
Master LaserOI
AOM
H-O
I
TAAO
M
-
Master Laser
AOM Double Pass
Injection Lock
Spectroscopy
AOMSingle
Pass
Tapered Amp
Figure 3.2: Schematic of the trapping beam setup.
Master laser and Spectroscopy
The master laser used for the trapping beam is a New Focus 6013 Vortex which is an External
Cavity Diode Laser (ECDL.) The wavelength of the laser is centered at 780.24nm. This laser
is controlled by a laser driver and controller New Focus TLB 6000, which provides current
modulation, piezo modulation and temperature controller. The output power of the laser is
typically around 7.5mW (6mW after the Optical Isolator (OI) that is used immediately after
Chapter 3. Apparatus 40
the laser to protect the laser from back-reflections.) A small portion of the output (100µW) is
branched off using a Polarizing Beam Splitter (PBS) and is sent to the polarization spectroscopy
setup. The error signal from the spectroscopy is sent to a homemade PID circuit which provides
a feedback to the current and piezo modulations to stabilize the frequency of the laser. Details
of the PID circuit and how to set the PID parameters can be found in [59]. Figure 3.3 shows
the schematic of the polarization spectroscopy setup.
Rb VaporCell
QWP
HWP
PBS
BSTo PID feedback
Photo Diode
100uWV-polarized
Probe Pump
20uWV-Polarized
80uWR or L-Polarized
Figure 3.3: Polarization spectroscopy setup. A circularly polarized pump saturates the atoms,and a probe, which is vertically polarized, undergoes a polarization rotation through the vapor.A combination of a HWP and a PBS projects the probe in diagonal and anti-diagonal polariza-tion basis. Two photo-diodes detect the intensity of the probe in these basis. The subtractedsignal is then used as the error signal to feedback to the master laser for frequency stabilization.
The master laser is locked to F = 3→ F ′ = 4 transition. Figure 3.4 shows the spectroscopy
signal for the transitions between the ground state F = 3 to the excited states F ′ = 2, 3 and 4
as well crossovers between the excited states. The blue circle shows where the master laser is
typically locked to (for more information about polarization spectroscopy, see[60, 61].)
The polarization spectroscopy setup is usually very stable and doesn’t drift significantly
within a few week time. Reduction in the size of the error signal or appearance of a Doppler
background are indicators that suggest the alignment of the spectroscopy setup is degrading,
which can result in an unstable lock. See B.1 for alignment procedure of the spectroscopy setup.
If after locking the laser, you observe that the error signal is oscillating on the scope, check the
Chapter 3. Apparatus 41
X343 ! 3 3 ! 43 ! 2 X24X23
Figure 3.4: Polarization spectroscopy of F = 3 → F ′ transitions. The vertical lines show theposition of different transitions and crossover transitions. The blue circle shows where the laseris locked to. The horizontal axis of the is detuning of the laser from F = 3→ F ′ = 4 transition.
PID parameters of the locking circuit. See [59] on how to set the PID parameters.
Important note: Avoid placing magnetic elements, such as an optical isolator, close to the
Rubidium vapor cell when doing polarization spectroscopy setup. Polarization spectroscopy
works based on spin polarizing the atoms and external magnetic fields can degrade the signal.
AOM Double-Pass
As mentioned before, the master laser is locked to the F = 3→ F ′ = 4 transition. However, the
trapping beam is required to be about 20 MHz detuned from the F = 3→ F ′ = 4 transition for
best cooling and trapping performance. To finely adjust the frequency of the trapping beam,
we use an Acousto-Optic Modulator (AOM) in double-pass configuration and a second AOM
in single-pass configuration. Double-passing a beam through an AOM enables changing the
frequency of the beam without changing its alignment. Figure 3.5 shows the schematic of the
AOM double-pass.
The radio frequency (RF) by which the AOM is driven is set to be centered at -71MHz
(the minus sign indicates that the frequency of the diffracted beam is shifted down.) The RF
can be modulated by an external control voltage between -1V and 1V which corresponds to a
frequency range of 42MHz to 104MHz. Hence, the frequency of the output of the double-pass
can be changed in a rage of -208MHz to -84MHz. To have the ability to change the frequency
of the trapping beam around the F = 3 → F ′ = 4 transition, we use another AOM in single-
Chapter 3. Apparatus 42
AOM
71MHz
Beam BlockBeam Block
Quarter-Wave Place (QWP)
PBS
4mW
0.2 to 1.9 mW
Figure 3.5: Schematic of AOM double-pass setup.
pass configuration, driven at a fixed frequency of +90 MHz.2 Calibration of detuning of the
trapping beam from F = 3→ F ′ = 4 transition and the output power of the double-pass versus
modulation voltage V are plotted in figure 3.6.
1.0 0.5 0.0 0.5Modulation Voltage (V)
140
120
100
80
60
40
20
0
20
Detu
ning
34
(MHz
)
(a)
1.0 0.5 0.0 0.5Modulation Voltage (V)
0.0
0.5
1.0
1.5
2.0Ou
tput
pow
er (m
W)
(b)
Figure 3.6: Calibrations of the AOM double-pass setup. (a) Detuning of the laser from F =3 → F ′ = 4 transition versus the modulation voltage. (b) Output power of the double-passversus the modulation voltage.
Alignment advice: If the double-pass is well aligned, the alignment of the output of the
double-pass should not change when varying the RF. Therefore, by monitoring the alignment
of beam at the output after some distance while frequency-modulating the RF, the goodness of
the alignment can be tested and improved.
Important notes:
• In setting up an AOM double-pass, never put the QWP before the AOM. The crystal
inside the AOM is highly birefringent. Its birefringence changes as the temperature of
2The AOM is located before the tapered amplifier.
Chapter 3. Apparatus 43
the crystal changes. If the input polarization of the AOM is anything but horizontal or
vertical (with respect to the crystal inside the AOM), one would notice large polarization
drifts as the temperature of the crystal increases after few hours of operation.
• Since the output of the double-pass is used to seed an injection-locked laser (to be ex-
plained later), make sure that the other diffraction orders of the AOM don’t leak into
the output. They can cause a mode competition and instability of the injection-lock.
Specially for RF ranges where the output power is low.
Injection-Lock
The output power of the double-pass is too low to be used as the trapping beam or to seed
a tapered amplifier. Therefore, it is first used to seed an injection-locked diode laser. In this
technique, a seed light with low power is used to force a high power slave laser to emit light
with high power at exactly the frequency of the seed light. In our setup, as low as 100µW of
power (the output of the double-pass, considering the losses through its path to the slave diode
laser) is used to seed a diode laser with output power of 50 mW. Figure 3.7 shows the schematic
of the setup used for injection-locking.
Output PolarizerFaraday
Rotator
B
HWP1
PBS
Optical Isolatorw/o input polarizer
Diode Laser
Temperature- Controlled Housing
H D
DV
HWP2
0.2 to 1.8 mW
35 mW
2
1
Figure 3.7: Schematic of injection-lock setup. The output of the double-pass is used to seed ahigh power diode laser.
In order to separate the input beam from the output beam, a combination of PBS, HWP
and an OI with its input polarizer removed (referred to as H-OI) is used. The current running
through the slave diode laser and the temperature it’s kept at can be varied to ensure that the
Chapter 3. Apparatus 44
frequency of the injection-locked laser is following the seed. To confirm that the slave laser is
following the seed laser, the frequency spectrum of the injection-locked laser is compared to that
of the seed light by sending them to a Fabry-Perot cavity with free spectral range of 150MHz.
The frequency spectra are measured by detecting the intensity of the output of the cavity while
scanning one of its mirrors. If the injection-locked laser is following the seed, one should see a
single peak in the spectra. Blocking the seed should make that single peak disappear and turn
the spectra into multiple random peaks. The quality of the injection-lock is very sensitive to
the input alignment. See B.2 for the alignment procedure.
Important note: Other diffraction orders of the AOM double-pass could leak into the seed light
for the injection-lock, and could win over the wanted seed to force the slave laser to follow them.
While scanning the current on the diode laser, if you notice that the diode can be injection-
locked to more than one frequency, check for leakage at the double-pass. If you cannot eliminate
the leakage, make sure you are locking the diode to the correct seed.
Tapered Amplifier
The last stage of trapping beam preparation is a Tapered Amplifier (TA) preceded by the
single-passed AOM mentioned before (See figure 3.2.) The TA amplifies the beam to the power
needed for a MOT and the AOM does the final adjustment of the frequency as well as giving
us a way to quickly switch the trapping beam on and off and to adjust its power. The RF
driving the AOM is amplitude modulated with a modulation voltage ranging between 0V to
1V, which corresponds to an RF amplitude of 0V to 20V. The 1st diffraction order of the AOM,
with maximum power of 20 mW is sent through a TA, while the 0th order is used for the optical
pumping and tagging light, which will be discussed later. The tapered amplifier is an SDL-
8630, with maximum power of 500mW. The typical output power of the TA, when run with
1A current, is around 200mW. By turning the current up to a maximum of 1.5A, the output
power can be >350mW. The output of the TA is then coupled to a Polarization Maintaining
(PM) single-mode Fiber, through which the beam is guided to the vacuum chamber. Figure
3.8 shows the calibration of the measured power after the PM fiber (the power used in trapping
beam) versus the amplitude modulation voltage.
The output power of the TA is extremely sensitive to the input alignment and polarization.
The input polarization is controlled using a HWP. The input alignment can be optimized by
overlapping the input beam with the Amplified Spontaneous Emission (ASE) light coming out
Chapter 3. Apparatus 45
0.0 0.1 0.2 0.3 0.4 0.5 0.6Modulation Voltage (V)
0
20
40
60
80
100
120
140
Powe
r (m
W)
Figure 3.8: Measured power after the PM fiber versus amplitude modulation voltage of theAOM, when TA running at 1.5A.
of the input port of the TA. If however, you find that the improving the input alignment does
not improve the output power, or the output power is degrading with time, it indicates that
the alignment of the optical elements inside the TA box need to be improved. Figure 3.9 shows
the optical elements used inside the TA box. If you need to improve the alignment of these
elements, follow the steps in B.3 with extreme care. The optical elements inside the TA are
usually stable for a long time and the alignment procedure needs to be done once or twice a
year.
Polarization-Maintaining Fiber
The laser beam after the TA has the correct frequency and enough optical power to be used as
the trapping beam. To sent this beam to the vacuum chamber, where the MOT is captured,
we use a single-mode PM fiber. Two lenses are employed to match the spatial mode between
the output of the TA and the mode accepted in the fiber. An OI is used to prevent the beam
to reflect back into the TA. Overall, we manage to couple 40% of the output of the TA into the
single-mode PM fiber.
For the PM fiber to maintain the polarization, the input polarization must be set to be
linear and along either the fast or the slow axis of the PM fiber. If the polarization of the input
beam is not along either of the two, the polarization of the beam at the output of the fiber will
fluctuate dramatically. These polarization fluctuations depend on the tension on various parts
of the fiber, and any change in the position of the fiber will change the output polarization.
Therefore, it is extremely important to make sure the input polarization is set correctly. This
Chapter 3. Apparatus 46
Top View:
Side View:
z
x
y
z
Input Lens Output lens Slit Cylindrical Lens
Semiconductor Chip
Figure 3.9: Schematic of inside of the TA box, both top view and side view.
change in polarization also depends on the frequency of the beam that is sent through the fiber.
One can take advantage of this dependence to set the input polarization. See B.4 for an easy
way to find the correct input polarization. The input polarization almost never changes and
once it is set correctly, it can be trusted for a long time. The efficiency of the coupling to the
fiber, however, needs to be optimized on a daily basis.
Typical Trapping Beam Parameters
The driver of the two AOM’s mentioned in previous sections are controlled by a computer,
which controls the tasks in the experiment using a Virtual Instrument (VI) program see 3.17.
An analogue voltage between -1V and 1V from the computer, with a controlled timing, sets the
RF frequency of the AOM double-pass (see figure 3.6a). Another analogue voltage, also with
controlled timing, between 0V and 1V sets the amplitude of the RF frequency which drives the
AOM single-pass.
AOM VI Voltage Corresponding Value
Double-Pass -0.45V 20MHz red-detunedfrom F = 3→ F ′ = 4
single-Pass 0.6V 70mW
Table 3.1: Typical Trapping parameters for MOT operation. TA running at 1A.
Chapter 3. Apparatus 47
AOM VI Voltage Corresponding Value
Double-Pass -0.2V → 0.1V 35MHz → 60MHz red-detunedfrom F = 3→ F ′ = 4
single-Pass 0.3V → 0.1V 10mW → 50mW
Table 3.2: Typical Trapping parameters for molasses operation. TA running at 1A.
AOM VI Voltage Corresponding Value
Double-Pass -0.6V On Resonancewith F = 3→ F ′ = 4
single-Pass ≤ 0.6V 70mWshould be lowered if the cam-era is saturated
Table 3.3: Typical Trapping parameters for imaging. TA running at 1A.
Typically the trapping beam is used for three purposes: MOT, Molasses Cooling and Fluores-
cence Imaging. Tables 3.1, 3.2 and 3.3 show the typical parameters used for the MOT, molasses
and imaging stages respectively. The current to the TA is set to be at 1A for all these opera-
tions. note: Make sure the center frequency of the AOM drivers (71MHz for double-pass and
90 MHz for single-pass) are checked periodically. The center frequencies can drift with time.
3.1.2 Repumper Beam
The essence of laser cooling is to use a cycling transition so that the atoms can scatter multiple
photons to reduce their momentum. For the cooling to work, the electrons in 85Rb atoms,
should remain in the F = 3 ground state and scatter photons off of F ′ = 4. The F ′ = 4 excited
state cannot decay to the F = 2 ground state. However, the atoms can scatter a photon off
of the F ′ = 3 excited state, and the electron can decay to the F = 2 ground state. Once the
electron is in the F = 2 ground state, the cooling stops. To prevent the electrons to spend time
in the F = 2 ground state, a repumper beam is employed to pump the electrons out of F = 2
and keep this ground state empty. The frequency of the repumper beam is set to be around
20MHz red-detuned from F = 2→ F ′ = 3 transition.
Figure 3.10 shows the setup used for preparing the repumper beam. The laser used for
the repumper beam is a New focus Vortex II TLB 6913 with output power of 48mW and
a wavelength which is tunable around 780.24nm. The driver used for this laser is similar to
Chapter 3. Apparatus 48
OI
-
HWP
Polarization-Maintaining Fiber
Rb CellErrorSignal
Laser
AO
M AO
M
46mW
22mW
48mW
0.3mW
Figure 3.10: Schematic of the setup to prepare the repumper beam.
the driver used for the trapping beam. After the OI, 300µW of the power is used to perform
saturated absorption spectroscopy.
2 ! 1
X12
2 ! 2
X13
X23
2 ! 3
Figure 3.11: Saturated absorption spectroscopy error signal for F = 2 → F ′ transitions. Thered circle indicates where the repumper laser is locked to. The transition from the F = 2 groundstate to the F ′ = 1, 2, 3 excites states as well as the crossovers are marked on the signal.
The result of the saturated spectroscopy for F = 2 → F ′ transitions are shown in figure
3.11. The laser is locked around 35MHz below F = 2→ F ′ = 3 transition.
Important Note: The red circle in figure 3.11 indicates where the laser is typically locked to.
To find the best place to lock the repumper laser, it’s best to change the offset of the lock while
measuring the density of the MOT, either by OD measurement or by fluorescence imaging after
molasses cooling. Lock the repumper where the highest density is achieved.
Chapter 3. Apparatus 49
Two single-pass AOM’s with frequencies of +100 MHz and -84MHz are used to bring the
frequency of the repumper to be around 20MHz below the F = 2 → F ′ = 3 transition. After
the two AOM, 22mW of power is coupled into a PM fiber, similar to that of the trapping beam.
The coupling efficiency is typically 60%. A modulating voltage between 0V and 1V is used to
amplitude modulate the RF signal used to drive the -84MHz AOM single-pass. Figure 3.12
shows the calibration for the output power of the PM fiber versus the voltage used to set the
amplitude of the RF signal.
0.0 0.2 0.4 0.6 0.8 1.0Modulation Voltage (V)
0
2
4
6
8
10
12
14
16
Outp
ut p
ower
(mW
)
Figure 3.12: Calibration of output power of the PM fiber versus the modulation voltage.
Operation VI Voltage Corresponding Value
MOT 0.6V-0.8V 12mW to 14mW
Molasses 0.1V to 0.3V 1mW to 5mW
Imaging 0.8V 14mW
Table 3.4: Typical repumper parameters for different operation.
Typical repumper powers used for different operations are shown in 3.4.
3.1.3 MOT setup
In sections 3.1.1 and 3.1.2 the preparation of trapping beam and repumper beam were discussed.
Figure 3.13 shows the D2-line of 85Rb as well as were the trapping beam and repumper beams
are locked to. Figure 3.14 shows the schematic of how the two beams are combined as well as
the optics used for guiding the beams towards the center of the trap to capture a MOT.
Once the trapping beam leaves the fiber, out of a fiber to free-space adapter, it expands
for a distance of about 15cm. It is then collimated to have a beam size of about 1 inch
Chapter 3. Apparatus 50
120M
Hz
63M
Hz
29M
HZ
3.03
5 G
Hz
790.
24nm
(384
TH
z)
1.77
0GH
z10
0MH
z
F = 2
52S1/2
F 0 = 3
F 0 = 2
F = 3
F 0 = 1
52P3/2
F 0 = 4
20M
Hz
20M
Hz
Figure 3.13: D2-line of 85Rb atoms. The red arrow indicated the tapping beam and the bluearrow indicated the repumper beam.
in diameter (1/e2) using an achromatic lens with focal length of 150mm. Using the same
technique, with a 45mm achromatic lens, the repumper beam is expanded to be about 1cm
in diameter. The polarization of the two beams are controlled using two HWP’s inside the
cage system (H1 and H2 in figure 3.14), so that the trapping beam and the repumper beam
are split with 2 : 1 power ration in the two outputs of the PBS1. The output with the higher
power is in turn split into two paths with 1 : 1 power ratio. The three resulting beams, with
the same power, are guided towards the center of the trap, which is defined by the magnetic
field gradient. Three QWPs (Q1,Q3 and Q5) are used change the polarizations of the beams to
circular polarizations. Three mirrors (M1,M3 and M5) are used to retro-reflect the MOT beams.
Another three QWPs (Q2,Q4 and Q6) are used to ensure that the retro-reflected beams have
the same circular polarization (opposite spin) as the incoming beam. Two magnetic coils, with
1.6A of current running through them in opposite directions (anti-Helmholtz configuration),
create a magnetic field gradient of around 20G/cm in the z-direction. Three set of Helmholtz
coils are used to compensate the stray magnetic fields down to a mG level. The density and the
temperature of the cloud heavily depends on the quality of the alignment of the MOT beams.
See B.5 for alignment procedure.
Chapter 3. Apparatus 51
Half-Wave plate
Quarter-Wave plate
Lens
coil
Polarizing Beam Splitter
MOT
x
y
z
PBS2
H3 H1
H2
Q1Q1
Q2
Q3
Q4
Q5
Q6
Mirror
PBS1
M1
M2
M3
M4
M5
Repumper
Trapping
3
1 2
Figure 3.14: Schematic of optics used for combining the trapping and repumper beam andcapturing a MOT.
For convenient power measurements of the MOT beams, a cage-mountable detector is de-
signed and calibrated, which we named Uberdetect. It consists of a photodiode detector and
an ND (OD=3) filter. The calibration is done with a 5kΩ resistor in parallel with the photodi-
ode. Figure 3.15 shows the calibration results for the trapping beam and the repumper beam.
The results indicate a 3.8mW/mV power to voltage conversion for the trapping beam and a
0.7mW/mV conversion factor for the repumper beam. Different beam sizes account for the
difference between the two conversion factors.
Chapter 3. Apparatus 52
0 5 10 15 20Voltage (mV)
0
10
20
30
40
50
60
70
80
Powe
r (m
W)
(a)
0 2 4 6 8 10 12 14 16 18Voltage (mV)
0
2
4
6
8
10
12
Powe
r (m
W)
(b)
Figure 3.15: Power versus voltage read by Uberdetect for (a) trapping beam and (b) repumperbeam.
3.1.4 Magnetic fields
Magnetic field gradient
The magnetic field gradient is generated by two coils, each with 200 turns and resistance of
1Ω, in an anti-Helmholtz configuration. The current in the coils is generated and controlled
by a Bipolar Operational Power (BOP) supplier made by KEPCO: model 20-10M. In order to
quickly switch the field gradient on and off, the BOP is operated in current-controlled mode.
In normal MOT operation, we run 1.6A current through the coils which generate a magnetic
field gradient of 20G/cm along the axis of the coils (10G/cm in the orthogonal directions.) The
BOP is remotely controlled by a computer and the current can be ramped up to 5A to obtain
higher field gradients. But it is advised not to maintain such high currents for a long time to
avoid over-heating the coils. The current generated by the BOP can be turned off in <500µs.
To eliminate fast oscillation of the current after switching on or off, a 3µF capacitor and a 67Ω
resistor are put in parallel with the output of the BOP.
Optical Pumping Coil
As mentioned before, three pairs of coils in Helmholtz configuration are used to compensate
stray magnetic fields. One of the three pairs is also used to generate a constant magnetic field
necessary for optical pumping, referred to as Optical Pumping (OP) coils. The currents for OP
coils are generated using another BOP made by KEPCO: model 36-12M. This BOP, which is
Chapter 3. Apparatus 53
3 2 1 0 1 2 3Control Voltage (V)
6
4
2
0
2
4
6
B (G
)
Bz
Bx
By
Figure 3.16: Magnetic field measured at the center of one of the OP coils. The field is measuredalong the coil axis, z, and the orthogonal directions x and y.
remotely controlled by the computer, is also set to run in current-controlled mode with a 3µF
capacitor and a 67Ω resistor put in parallel with its output. The field generated by the coils
can be switched off in 240µs. Figure 3.16 plots the measured magnetic field at the center of one
of the OP coils in the direction of the axis of the coil, Bz as well as the orthogonal directions.
3.1.5 Remote Task Control
A typical experiment contains different stages and each stage requires certain experimental
parameters. It is important to be able to vary these parameters between different stages with
precise timing. Therefore, being able to control all of the experimental parameters remotely
and in an automated way is a crucial element for performing complicated experiments. Most
elements used in this experiment can be controlled using an analog voltage as a control signal.
These analog signals are generated by two National Instrument: PCI 6713 cards. Each card
has 8 analog output channels (16 in total) with voltage range of ±10V and update rate of
1MS/s. The precise timing of the experiment is controlled using the 20KHz clock of a National
Instrument: PCI 1200 digital card, which corresponds to a minimum time step of 50µs. The
digital card also outputs Transistor-to-Transistor Logic (TTL) signals with controllable delay,
which are used for triggering purposes.
The control values needed for an experiment along with their timing information (altogether
called a task) are set using National Instrument LabVIEW software, through Virtual Instru-
ments (VI’s.) Once a task is executed, the software writes the user-defined control values on the
PCI 6713 cards and the timing information on the PCI 1200 card. The analog card transfers
Chapter 3. Apparatus 54
LabViewV
t1 t2 t3
PCI 1200
PCI6713 BNC 2110
Output to devices
Figure 3.17: Schematic of different stages used to convert the parameters defined in the LabViewsoftware to analog signals used to control multiple devices.
the information to an NI BNC 2110 device. The PCI 1200 converts the timing information
into a train of TTL’s that are spaced in accordance to the specified timings. These triggers
are then sent to the NI BNC 2110. Every time NI BNC 2110 receives a trigger, it updates
the output analog voltages. The analog outputs of the NI BNC 2110 are sent to the multiple
devices using coaxial cable with BNC connection. Figure 3.17 shows the schematic of how the
information from a VI get converted to analog outputs.
Important Note: The two analog cards can be timed using two different clocks, e.g. internal
clock of the PC and the clock on the digital card. The timing source of the cards can be
specified within the VI.
3.1.6 Maintenance and Diagnosis
To keep the setup in good condition, periodic checks and maintenance are necessary. Slow drifts
and fluctuations can lead to an unstable setup, if not corrected for as they happen. In this
section some maintenance details will be discussed and a diagnosis checklist will be provided.
Rubidium dispenser. To maintain the rubidium vapor pressure necessary for capturing a
MOT in the vacuum chamber, the temperature of the dispenser should be kept at around 37C
to 40C. This temperature should be readjusted at the beginning and the end of winter every
year. At the end of winter, the ventilation system in the lab is switched to cooling and the lab
gets colder. As a result, the temperature of the dispenser drops. This results in a lower vapor
pressure and one may notice that the MOT gets smaller and takes longer to load. Increasing the
voltage applied to the heating tapes should resolve this problem. At the beginning of winter,
Chapter 3. Apparatus 55
the ventilation system is switched to heating and the temperature of the dispenser will rise. In
this case, the voltage applied to the heating tapes should be lowered to adjust the temperature
of the dispenser.
Running low on rubidium. The rubidium in the dispenser should be replaced once every
two to three years. Insufficient rubidium in the dispenser results in a smaller MOT and longer
loading times. Heating the dispenser harder will not resolve these symptoms.
Mode-hopping of the master laser. The trapping and repumper lasers are designed to have
a mode-hop free range of several GHz. While doing spectroscopy, you should be able to observe
all 4 hyperfine transitions of 85Rb and 87Rb. If you cannot see all 4 transitions, it indicates that
the laser is mode-hoping. Also, if you observe that the size of the error signal changes while you
scan through one transition, it usually indicates that the laser is multi-mode and it is about to
mode-hop. To find a mode-hop free region, change the current running through the diode and
if the problem persists, use temperature as a second nob to find a mode-hop free region.
Diagnosis Checklist: If you find that the MOT has disappeared and you have difficulty
getting a MOT, run through the following checklist
• 1) Make sure the magnetic field gradient is turned on and the VI is running.
• 2) Make sure the trapping laser is locked to the correct transition. The spectroscopy
signals for 85Rb F = 3 → F ′ = 2, 3 and 4 and 87Rb F = 3 → F ′ = 1, 2 and 3 are nearly
identical in shape can be confused with each other. Note that the frequency spacing
between the transitions in 87Rb are larger than the spacings in 85Rb.
• 3) Check the injection lock and make sure it is following the correct output order of the
double-pass.
• 4) Measure the repumper and trapping beam powers and make sure the power is balanced
among the three pairs of MOT beams.
• 5) Check the frequency of the RF signals driving the AOMs.
• 6) Use an IR-viewer to look for fluorescence in the cuvette.
• 7) Check the alignment of the MOT beam. Start with the back-reflected beams and make
sure they are overlapping with the incoming beam. Then make sure the three beams are
overlapping at the center of the cloud.
Chapter 3. Apparatus 56
• 8) Check the polarization of the MOT beams.
3.2 Probe and Coupling Beams
In section 3.1, the MOT setup was discussed in detail. In this section, a brief description of
probe and coupling beam preparation will be provided. A more detailed description of this part
of the experiment can be found in [52]. Figure 3.18 shows the schematic of the probe beam,
the coupling beam and the signal pulses at the interaction region.
To single photon detector
To phase measurement detector
Probe
Coupling
Signal
MOT
13um focus10:90
Figure 3.18: Schematic of the probe, the coupling and the signal beam.
Probe
Preparation. The probe beams is composed of two frequency component: a component that is
on-resonance with the F = 2 → F ′ = 3 transition and an off-resonance component which is
tuned 100MHz above that transition. The probe beam is branched off of a master laser (New
Focus Vortex ) that is locked ≈30MHz red of the F = 2 → F ′ = 3 transition. By using an
AOM that is driven at +130MHz (double-passing at +65MHz) the off-resonance component
of the probe beam is generated. The RF signal driving this double-passed AOM is amplitude
modulated to switch the probe beam on and off. This off-resonance component is then sent
through another AOM at -100MHz. The 1st diffraction order of the AOM is used as the on-
resonance probe component. The RF signal driving this AOM is locked to a stable 10MHz clock.
Chapter 3. Apparatus 57
The two beams are then combined on a beam splitter and are sent towards the interaction region
using a single-mode optical fiber. Typical probe power used in this experiment is 3 nW for the
on-resonance probe and about 20-30 nW of off-resonance probe.
Phase Measurement. After passing through the interaction region, the probe beam reflects
off of a 10:90 beam splitter and is then detected using a fast Avalanche Photodiode Detector
(APD.) The APD detects the 100MHz beating signal of the two components of the probe beam.
Any change in the phase and/or the amplitude of the on-resonance probe beam results in a
change in the phase and/or the amplitude of this 100MHz beating signal (This technique is
referred to as beat-note interferometry[62].) We use an IQ-demodulator, at 100MHz, to then
read off the phase (arctan( IQ)) and amplitude (
√I2 + Q2) of the beating signal. The same
stable 10MHz clock that was used for preparing the probe beam, is used to lock-in this 100MHz
demodulation. The resulting I and Q are then digitized with a sampling frequency of 1.5MHz.
To eliminate slow drifts in the phase measurement, the phase of the probe is averaged for 3
samples (200ns) before and 3 samples after when the XPS is expected to happen. This value
is then subtracted from the average of the probe phase in the interval when XPS is expected.
The resulting number is reported as the measured XPS. The probe beam contains about 2000
photons, which corresponds to 22mrad phase uncertainty (shot-noise.) We measure a single-
shot phase uncertainty of around 100mrad. We believe the source of the added uncertainty is
due to a combination of detector noise used to detect the beating signal and the electronics
afterwards. In order to measure the phase shift down to few µrad uncertainty level, we repeat
the measurement about half a billion times.
Coupling
To increase the interaction strength, the probe is tuned to be on resonance with F = 2→ F ′ = 3
transition. In order to eliminate absorption of the probe, a coupling beam is used to create
an electromagnetically-induced transparency (EIT) situation. EIT is a coherent effect in which
a destructive interference prevents the two involved laser beams from being absorbed by the
atoms[63, 64]. In order to create an EIT situation, the probe and the coupling lasers should
be phase-locked to one another. To achieve the phase-lock condition, the remainder portion of
the master laser used for preparing the probe beam is frequency-shifted down using an electro-
optic modulator (EOM) driven at 3GHz. It is then used to seed an injection-locked diode laser.
As a result, the second diode laser, called the coupling laser, is phase-locked to the master
Chapter 3. Apparatus 58
laser. A combination of an AOM (single-passed at +103MHz) and fine tuning the frequency
by which the EOM is driven, sets the frequency of the coupling beam to be on resonance with
F = 3→ F ′ = 3 transition. The intensity of the probe and coupling beams are chosen so that
the width of the resulting EIT feature is measured to be 2MHz. The polarization of the probe
and coupling beams are set to be σ+ and π, respectively.
3.3 Signal Pulses, Photon Detection and Tagging
3.3.1 Preparation
To generate the signal pulses, 40mW of the total 55mW output power of the injection-locked
diode laser (coupling laser) is branched off and sent to an AOM double-pass at +80MHz. The
output of the double-pass is then sent through an AOM single-pass at +75MHz, which sets
the frequency of the signal beam to be around 18MHz blue-detuned from F = 3 → F ′ = 4
transition. By amplitude modulating the RF that drives the AOM single-pass we generate signal
pulses with 40-100ns duration. These pulses are then sent to the “Atoms Lab” (interaction
region) using a single-mode fibers.
AND
SPCM
Delay GeneratorDetector
ND Filter
Figure 3.19: Schematic of the signal pulses, SPCM and how the SPCM is gated.
3.3.2 Photon Detection
Before sending the signal pulses to the interaction region, they are sent through a PBS. Around
95% of the total power is transmitted and detected on a Thorlabs, PDA10A detector. The signal
from this detector is then used for gating the photon detections at a single-photon detector,
figure 3.19. The 5% remaining power is sent to the interaction region using a single-mode fiber.
Chapter 3. Apparatus 59
An ND filter with eOD = 613± 12 is used to attenuate the signal pulses to very low intensities.
This way, signal pulses with very low mean photon numbers, < 0.2, can be prepared. After
passing through the interaction region, the signal pulses are collected in a multi-mode fiber with
30% overall collection efficiency. A Single-Photon Counting Module (SPCM) is used to detect
the presence of photons in the signal pulses. Figure 3.20 plots the probability of a click, defined
as number of photon detection events over the total number of pulses, versus mean photon
number per pulse. The data is fit to Pclick = 1− e−ηn where n is the mean photon number per
pulse and η is the efficiency. The efficiency of the SPCM is inferred to be 70%.
0 2 4 6 8 10 12 14 16Number of photons at the SPCM
0.0
0.2
0.4
0.6
0.8
1.0
P clic
k
Figure 3.20: Probability of photon detection versus number of photons at the SPCM. From afit to the data, the efficiency of the SPCM is measured to be 70%.
3.3.3 Gating the Photon Detection
Upon detecting a photon, the SPCM outputs a 30ns trigger. These triggers can be used to tag
the measured phases that correspond to successful photon detections. However, due to dark
counts, the SPCM sends triggers even when no signal pulse is present, which makes the tagging
procedure challenging. To eliminate these dark counts, we use a logic AND gate between the
output triggers of the SPCM and the output of the PDA detector that was mentioned before.
This is to say that we gate the SPCM with separately detected signal pulses. A delay generator
is used to correct for the time delay between the detected signals from the PDA and the triggers
from SPCM, see figure 3.19. The result of the AND gate is then used for tagging.
This technique removes all the dark counts that happen outside the time window in which
we expect to have a signal pulse. But a dark count can also happen during this time window.
The probability of getting a dark count in this time window depends on the temporal length
Chapter 3. Apparatus 60
of the signal pulses. Any photon scattering event from the probe and the coupling beam has
a chance of ending up being detected at the SPCM through the signal path. Therefore, the
probability of a background photon detection also depends on the presence of the probe, the
coupling beam and the atoms in the interaction region. To measure the probability of a dark
count, we block the signal beams entering the interaction region and count the detection events
with the probe and coupling beam present. If we gate the SPCM with 100ns windows, we
observe that without atoms, we get <2% dark counts (without the atoms, there could not be
any scattering.) This number, however, increases to 10-12% when the atoms are switched on.
To decrease the dark counts, we gate the SPCM with 40ns time windows, which is the shortest
signal pulse the AOM can generate, and we observe that the dark counts reduce to 6%. To
observe the dark counts due to scattered photons, both the probe beam and the coupling beam
need to be present to maintain a certain population ratio in F = 2 and F = 3 ground states.
Therefore, in the absence of either of the two beams, the dark counts decrease to 2%.
It is worth mentioning that at first we tried to gate the SPCM with the electrical signals
that generate the optical signal pulses. To use those electrical signals, we had to send them
through 20m of coaxial cable. These long cables, however, caused broadening in the electrical
signals (due to the capacitance of the cable which is significant when the cable is long enough)
and back-reflection of the signal due to impedance mismatch between the AND gate and the
signal generator that generates the electrical signals. These imperfections could leak into the
measured beating signal of the probe beam through the AND gate and the tagging procedure.
Given that we were to measure a signal that was 10,000 smaller that the noise, any added noise
and systematic could not be tolerated. To eliminate these problems, we decided to use a PDA
to detect the optical signal pulses after they have traveled the distance and use the detected
signal to gate the SPCM.
3.3.4 Tagging the Shots with Successful Photon Detection
In order to measure the phase shift due to “click” and “no-click” events separately, we need a
way to distinguish the two events; we need to tag the shots when “click” happens. The easiest
way to tag the shots with successful post-selection is to use a stable and fast clock to time-tag
the triggers from the SPCM and signal pulses. Unfortunately we were not in possession of
such a device. Therefore, we decided to we write the tags on the same demodulated signal
that contains the measured phase of the probe. This way, each measured shot contains the tag
Chapter 3. Apparatus 61
RFSwitch
AOMVCO
Delay Generator Tagging Light
Probe
To Phase and AmplitudeMeasurement
XPS
Tags
Output ofAND gate
Figure 3.21: Schematic of the tagging procedure. The result of the AND gate switches on a100MHz RF signal to an AOM upon a successful photon detection. The 1st and 2nd diffractionorder of the AOM are sent to the detector which detects the probe. When a successful detectionhappens, this AOM sends a flash of light to the detector which results in a short spike in themeasured amplitude (and phase.) A delay generator controls the relative delay between theexpected XPS and the position of the tag.
information with it and no clocked time-tag is needed.
To tag the shots with successful photon detection, we use a short flash of light, that is
shone on the detector which detects the beating signal of the probe. Since the measurement
bandwidth used for the experiment is 2MHz centered at 100MHz, that short burst of light
should also have a 100MHz component. Therefore, the we use the 1st and the 2nd diffraction
orders of an AOM driven at 100MHz, to generate the flash of light. The RF to the AOM is
switched on and off using the output of the AND gate between the SPCM and the detected
signal pulses, so that the RF is only on when the output of the AND gate is high. The triggers
coming from the AND gate are delayed using a delay generator to precisely control the position
of tags with respect to the XPS. The amplitude of the tagging light is set to be high enough
for the tags to be single shot resolvable. Figure 3.21 shows the schematic of the tagging setup.
3.4 Polarization Interferometer and Tomography
To implement the weak value amplification of photon numbers, the signal pulse should be sent
through an interferometer. The interferometer used in this experiment is a polarization interfer-
ometer where the two paths of the interferometer are the two orthogonal circular polarizations
Chapter 3. Apparatus 62
(right-handed and left-handed.) Since the two paths are spatially overlapping, this interferom-
eter is immune to many instabilities that occur in interferometers with non-overlapping paths.
Figure 3.22 shows the schematic of the signal path from before it is sent through the interac-
tion region to when it is collected in the multi-mode fiber for detection at the SPCM. The first
element inside the signal telescope is a polarizer to ensure the initial polarization of the signal
is linear. Well-calibrated QWP and HWP are used to control the initial polarization of the
signal. We set the QWP at 205 and the HWP at 79.
1
2
Signal
ProbeCoupling
To the SPCM
10:90 BS
Lens
Glan-ThompsonPolarizer
Half-Wave plateQuarter-Wave plate
Signal Telescope
Probe TelescopePolarizer
Figure 3.22: Schematic of the signal path. Dashed lines marked as 1 and 2 indicated thebeginning and the end of the polarization interferometer.
A linear polarization is an equal superposition of right-handed circular polarization, σ+,
and left-handed circular polarization, σ−. After the interaction region and before the signal is
detected at the SPCM, a combination of a QWP (fast axis at 62), a HWP (fast axis at 355 )
and a Glan-Thompson Polarizer (GTP) is used to project the output of the GTP to an arbitrary
polarization. This also gives us a tool to perform polarization tomography on the signal and
see how the polarization of the signal is affected during its propagation in the presence and
absence of the atoms, the probe beam and the coupling beam. Once the initial polarization of
the signal is well understood, the QWP and the HWP are set to project the output of the GTP
to the final polarization, corresponding to a certain post-selection parameter.
The polarization tomography is performed by measuring the signal power at the output
of the GTP, when projected in H-basis (QWP at 62 and HWP at 355), V-basis (QWP at
Chapter 3. Apparatus 63
62 and HWP at 40), D-basis (QWP at 62 and HWP at 17.4) and R-basis (QWP at 62
and HWP at 332.5.) From the tomography result, we calculate the stokes parameters for the
initial polarization of the signal beam. Knowing the initial polarization, we calculate the final
polarization that corresponds to the desired post-selection parameter. Correct setting of the
QWP and the HWP projects the output of the GTP to the calculated final polarization.
3.5 Atom Preparation
3.5.1 Level Scheme
18 M
Hz
F=2
F=3
F'=1
F'=2
F'=3
F'=4
Probe
Coupling
Signal
Figure 3.23: Level scheme used for the WVA experiment.
A requirement for the WVA experiment is for the signal in the two paths to interact with
the probe with different strengths. Therefore, we should chose the level structure so that
the interaction of the probe with the σ+-polarized signal is different from its interaction with
σ−-polarized signal. Figure 3.23 shows the level scheme used for the WVA experiment. The
reason for this polarization dependence is the following: The size of the XPS is proportional
Chapter 3. Apparatus 64
to Ω2sΩ
2pr/Ω
2C (in the low OD limit) where Ωi’s are the Rabi frequency of the signal, probe
and coupling beams. The Clebsch-Gordon (CG) coefficient for σ+ polarization of the probe
is largest for the F = 2,mF = 2 → F ′ = 3,mF ′ = 3 transition. Likewise, the CG coefficient
for π-polarized coupling beam is largest for the F = 3,mF = 3 → F ′ = 3,mF ′ = 3 transition.
Therefore, the Λ-system created between F = 2,mF = 2 and F = 3,mF = 3 ground states is the
dominant Λ-system3. One can easily show that for equal population distribution among F = 2
ground state, one should expect the XPS for the σ+ polarization of the signal to be around
3.6 times larger than the XPS due to σ− polarization. If one initializes all the population in
F = 2,mF = 2 ground state (via optical pumping), the ratio between the two XPS’s is expected
to be 28 (for more details see [52].)
3.5.2 Optical Pumping
0.0 0.2 0.4 0.6 0.8 1.0frequency (a.u.)
0.0
0.1
0.2
0.3
0.4
0.5
0.6
OD
B × ,OP ×fitB ,OP ×fitB ,OP fit
Figure 3.24: Measured OD versus frequency of the probe for three situations: no external field,no optical pumping with external field and with optical pumping and external field.
As mentioned in section 3.5.1, pumping all the population to be in F = 2,mF = 2 can
make the XPS for the σ+ polarization to be 28 time larger than that of the σ− polarization.
Therefore, we did optical pumping in F = 2 ground state to prepare the population in mF = 2.
Two optical beams were used for optical pumping: a pump beam and a repumper beam, which
we are going to call F3-repumper to avoid confusion with the repumper beam used for the MOT.
The pump beam was branched off of the repumper beam for the MOT. Using an AOM, the
frequency of the pump beam was set to be on resonance with F = 2 → F ′ = 2. F3-repumper
3The CG coefficient for F = 3,mF = 1 → F ′ = 3,mF ′ = 1 and F = 3,mF = −1 → F ′ = 3,mF ′ = −1 arethe same but the CG coefficient for F = 2,mF = −2 → F ′ = 3,mF ′ = −1 is 15 times smaller than the CGcoefficient for F = 2,mF = 2→ F ′ = 3,mF ′ = 3.
Chapter 3. Apparatus 65
was branched off of the trapping beam and its frequency was set to be on resonance with
F = 3 → F ′ = 3. An external DC magnetic field of around 5G was created using the optical
pumping coils mentioned in 3.1.4. Around 200µW of pump and 1mW of F3-repumper were
combined at a beam splitter and were sent through the atoms along the DC magnetic field.
The polarization of both beams were set to be σ+. We did optical pumping for 200µs.
The see the result of the optical pumping, we measured the OD as seen by the probe beam,
with σ+ polarization, as we scanned its frequency. In the absence of the external magnetic field,
the OD versus frequency of the probe is expected to be a Lorentzian function. If, however, the
external magnetic field is turned on, the OD should have a much broader feature due to the
Zeeman shifts of the sub-levels of the F = 2 ground state. If all the atoms are pumped into
the desired ground state, in the presence of the external field, the OD versus frequency of the
probe should be a Lorentzian function with the original width (without the magnetic field)
but displaced in frequency. Figure 3.24 shows the result of the OD measurement versus probe
frequency for the three situations mentioned above.
0.0 0.2 0.4 0.6 0.8 1.0frequency (a.u.)
0.0
0.1
0.2
0.3
0.4
0.5
OD
datafitmF =2mF =1mF =0mF = 1mF = 2
(a)
0.0 0.2 0.4 0.6 0.8 1.0frequency (a.u.)
0.0
0.1
0.2
0.3
0.4
0.5
0.6
OD
datafitmF =2mF =1mF =0mF = 1mF = 2
(b)
Figure 3.25: The result of fitting the OD versus probe frequency to the weighted sum of fiveshifted Lorentzian functions for (a) no optical pumping and (b) with optical pumping.
Knowing the strength of the external magnetic field and the magnetic moment g-factor for
F = 2 ground state, we can fit the measured OD versus probe frequency in the presence of
the magnetic field to a weighted sum of five Lorentzian functions, each centered at a different
frequency. The weight for each Lorentzian reflects the population of the atoms in that particular
Zeeman sub-level. Figure 3.25 shows the result of this fit for the case of no optical pumping
and the case where optical pumping was performed. The fits suggest that initially we’d had
21% of the population in mF = 2, 43% in mF = 1, 19% in mF = 0, 16% in mF = −1 and no
detectable population in mF = −2. With optical pumping we measure 55% in mF = 2 and
45% in mF = 1 with no detectable population in the rest of the Zeeman sub-levels.
Chapter 3. Apparatus 66
Note: We ended up not using optical pumping for the WVA experiment. Due to what we
believe is inhomogeneity of the external magnetic field, we couldn’t obtain an EIT situation
that was nearly good enough for the experiment to move forwards. Since time was of essence
and WVA could be done without optical pumping, we decided to move on with the experiment
and postpone a more detailed study of the optical pumping for later. The setup for optical
pumping has since been removed from the table due to equipment shortage for the tagging
setup.
3.5.3 Atom Cycle
Our measurement cycle has three parts: cooling for 20ms, population preparation for 0.5ms,
and free expansion for 1.5 ms. The cooling time can be as small as 6ms without any observable
effect in the density of the cloud. After the cooling stage, the population is mainly in F = 3
ground state. To pump the population to F = 2 ground state, We turn the repumper beam
off, lower the trapping beam power and increase its detuning for 0.5ms. This stage gives us
enough time to make sure the magnetic field gradient is also turned off. Then, we turn all the
MOT beams off and sent a trigger to the probe beam and coupling beam to turn them on. The
same trigger also triggers a function generator to start sending the signal pulses. A separate
trigger, which is simultaneous to the first trigger, sets the starting point for the IQ-analyzer to
take measurements. The measurement is done for 1.5ms during the free expansion of the cloud.
The first 300-400 µs of this time is used to ensure that the EIT is established and the initial
phase dynamics of the probe are suppressed. We then start sending the signal pulses every 2.4
µs. The collected data by the IQ-analyzer is sent to a computer via an Ethernet cable for the
extraction of the phase, the amplitude and the tags.
Terminology. We call each of the 1.5ms measurements a trace. A trace is divided in 2.4µs
intervals, each containing a signal (an XPS.) Each interval is called a shot. We take the data
in many parts each with several traces. Each part is referred to as a bin.
note: A task on the computer that controls the experiment (see section 3.17) consists of
the three mentioned parts, cooling, population preparation and free expansion, that is repeated
many times (usually 100 times but the number of repetitions can be modified.) Once the
repetitions are over, the computer generates another similar task and runs it again. Be advised
that the computer takes some time, typically around 10ms, to prepare in between each two
tasks. Therefore, there will be a delay between the triggers sent from the computer every time
Chapter 4
Experimental Results
In this chapter calibration results, details of the experiment and a discussion of the results
of the thesis will be provided. Here, I present clear data which demonstrates that a single
post-selected photon, using weak value amplification, can act like eight photons. I believe this
experiment to be the first ever weak value experiment that utilizes a deterministic interaction
between two distinct optical systems.
Figure 4.1 shows the schematic of the setup used for this experiment. (Details of the setup
can be found in chapter 3.) The nonlinear interaction between the signal and probe beams is
mediated by a sample of laser-cooled 85Rb atoms in a magneto-optical trap (MOT). Presence of
a coupling beam sets up electromagnetically induced transparency (EIT) for the probe beam.
By using EIT, the nonlinear interaction can be enhanced by tuning the probe beam to be on
resonance with the atoms while substantially reducing the accompanying absorption. As the
signal pulse passes through the medium, it causes an AC-Stark shift on the ground state which
the coupling beam addresses. Due to this energy shift, the probe and coupling beam will no
longer be on two-photon resonance and therefore, the probe beam sees a modified index of
refraction. As a result, the probe picks up a phase shift relative to what the phase would have
been in the absence of the signal pulse (Kerr effect[65, 66, 67, 68, 69, 70].) The change in the
index of refraction, and therefore, the phase shift of the probe, depends linearly on the number
of photons (intensity) of the signal pulse and the slope of the EIT feature at the frequency of
the probe. Therefore, measurement of the phase of the probe is equivalent to a non-demolition
photon number measurement[51, 71].
Terminology: I will refer to the the intrinsic per photon nonlinear phase shift of σ+-polarized
68
Chapter 4. Experimental Results 69
Figure 4.1: Schematic of the experimental setup. Counter-propagating probe and signal beamsare focused down to a waist of 13µm inside a cloud of laser cooled 85Rb atoms confined ina magneto-optical trap. A collimated coupling beam, propagating in a direction orthogonalto that of the signal and probe, sets up EIT for the probe. The signal beam is prepared ina linear polarization and after its interaction with the probe beam is post-selected on a finalpolarization that is almost orthogonal to the initial polarization. The probe and the signalbeams are separated from one another using a 10:90 beam splitters.
signal as φ+ and σ−-polarized signal as φ−. φclick refers to the average XPS on the probe
when the measurement is conditioned on the single-photon detector firing and φno−click is the
measured phase shift when the detector fails to fire. I will use “click” and “successful post-
selection” interchangeably.
4.1 Calibration
4.1.1 Single Photon Cross-Phase Shift
As a first calibration, we measured the nonlinear phase shift written on the probe as we varied
the mean photon number on the signal beam. The calibration was done using circularly po-
larized (σ+) signal pulses. We measured a per photon nonlinear phase shift of 13 ± 1.5µrad.
We also measured the effect of individual photons by post-selecting on a subsequent detection
of a photon after the interaction between the signal and the probe. We measured the effect of
individual post-selected photons to be 18± 4µrad. Details of these measurement can be found
Chapter 4. Experimental Results 70
in [50, 52].
4.1.2 XPS Versus Signal Polarization
150 200 250 300 350 400QWP Angle (Degrees)
0.51.01.52.02.53.03.54.04.55.0
XPS
(mra
d)
Figure 4.2: XPS versus signal polarization. A quarter-wave plate is used to set the polarizationof the signal beam before the interaction. The maximum and minimum measured phase shiftscorresponds to the two circular polarizations σ+ and σ−. The signal pulses used for thismeasurement contained around 600 photons.
As mentioned in section 3.4, for the WVA experiment to work, we need the two paths of
the polarization interferometer, which are σ+ and σ−, to interact with the probe with different
strengths. The polarization of the probe can be controlled before the interaction by using a
QWP. To calibrate the polarization dependence of the XPS, we measured the XPS written on
the probe by approximately 600 signal photons as we varied the polarization of the signal from
σ+ to linear to σ− by rotating the QWP. Figure 4.2 show the measured XPS versus the angle
of the QWP. The calibration shows that the measured XPS is largest for σ+ and smallest for
σ− polarization of the signal. For the WVA experiment, the angle of the QWP is set to be at
205, which corresponds to a linear polarization.
4.1.3 Polarization Tomography on the Signal
In section 3.4 I discussed how to perform polarization tomography on the signal beam. Under-
standing the initial polarization of the probe and minimizing the effect of the atoms is crucial
Chapter 4. Experimental Results 71
for post-selecting on the correct final polarization. Table 4.1 shows the result of the tomography
on the signal beam in various situations. The parameters S0, S1, S2 and S3 are the measured
stokes parameters of the signal polarization.
Atoms Probe Coupling S0 S1 S2 S3
off blocked blocked 10.1 -1.4 9.0 4.4
on blocked blocked 10.1 -1.4 9.0 4.5
on blocked on 10.2 -1.4 9.1 4.5
on 3nW blocked 5.7 -3.4 0.2 4.2
on 3nW on 9.9 -2.5 7.7 5.7
on 6nW on 8.7 -3 6 5.6
Table 4.1: Polarization tomography on the signal
The results indicate that in the absence of the coupling and the probe beam, the polarization
of the signal in the presence and the absence of the atoms remains the same. The reason is that
the population is initially mostly in the F = 2 ground state and the cloud is almost transparent
for the signal. Adding the coupling beam only helps with pumping the population out of the
F = 3 ground state and the polarization of the signal remains unchanged. On the other hand,
if the probe beam is present and the coupling beam is blocked, the polarization of the probe
dramatically changes and the signal gets absorbed by 44%. This is due to the fact that the
probe beam will pump the population out of the F = 2 ground state and the signal beam will
see a large OD. When the coupling beam and the probe beam (with 3nW of power) are both
present, the population will mostly remain in the F = 2 ground state and the polarization of
the signal won’t change by much and we observe less that 2% absorption. However if the probe
power is increased to 6nW, a larger fraction of the population will be in F = 3 ground state
and the signal will see a larger OD and will get absorbed by 14%.
A large population in the F = 3 ground state means a larger polarization rotation of the
signal. Therefore, any fluctuation in the OD will result in an unstable post-selection parameter.
The tomography results suggest that we need to turn the probe power down as much as possible.
Turning the probe power down, however, results in a larger phase-noise uncertainty which makes
the measurement of XPS harder. We chose the probe power to be 3nW for this experiment
because the polarization rotation of the signal with this probe power becomes almost negligible,
while the phase uncertainty remains low enough for the experiment to remain feasible to be
done. We measured shot-noise uncertainty for this probe power to be around 100mrad.
Chapter 4. Experimental Results 72
4.2 Measurement Results
4.2.1 φclick and φno−click
The experiment is performed using the measurement cycle introduced in 3.5.3. We analyze
the data by separating the measured shots into two groups of “click” and “no-click”, based on
the tags from the SPCM, see 3.3.4. The parameters were chosen in such a way to keep the
probability of a successful post-selection around 22-25%, 6% of which being background photon
detection. Table 4.2 shows the combination of parameters used for each data point along with
the total number of measurement used, Ntot, and the measured post-selection probability, Pclick.
In this table, n is the mean photon number for signal, η is the overall efficiency and δ is the
Data Point n δ η Ntot Pclick
1 95 0.10 0.2 42,111,000 31%
2 45 0.14 0.2 104,111,000 23%
3 20 0.22 0.2 102,380,000 25%
4 10 0.32 0.2 204,112,000 23%
5 40 1 0.03 374,443,000 20%
Table 4.2: Experimental parameters used in the WVA experiment.
post-selection parameter as defined in 2.10. As stated before, δ = 1 refers to post-selecting on
the initial state.
Figure 4.3 shows the results of the measured φclick and φno−click for various experimental
parameters. It can be seen that the measured phase shift for the “click” cases are always
larger than the measured phase shift for the “no-click” cases. The difference between the two
gets smaller as the post-selection parameter gets larger. In what follow, I will investigate the
dependence of this difference versus post-selection parameter in more detail.
4.2.2 φno−click versus |α|2
As a first step in analyzing the data, we extract the value of φ = φ++φ−2 . From equation 2.38,
we know φno−click should have a linear dependence on |α|2. The value of the slope of this linear
function is equal to φ.
Figure 4.4 plots the φno−click data versus |α|2 and the linear fit to the data. From the fit
we measure φ = 7.6± 0.3µrad. If we assume that the atomic population is equally distributed
between Zeeman sub-levels of F = 2 ground state, we expect φ+ = 3.6 × φ−, see 3.5.1. With
Chapter 4. Experimental Results 73
n
95 45 20 10 40
0.10 0.14 0.22 0.32 1.00
Figure 4.3: Measured XPS for click events and no-clicks events. Red squares and green circlesare the measured phase shift conditioned on the SPCM not firing (no-click) and firing (click)respectively. The horizontal axis shows the mean signal photon number n and the post-selectionparameter δ used for each case.
this assumption and the value measured for φ, we can infer φ+ and φ− to be 11.9±0.6µrad and
3.3±0.2µrad, respectively. However, in the next section I will show how this assumption is not
valid in our experiment.
4.2.3 Weak Value Amplification of a Single Post-selected Photon
In Section 2.2.5 (equation 2.64) we discussed how the quantity φclick − φno−click is independent
of |α|2 and directly shows the effect of a single post-selected photon. Figure 4.5 plots the
measured φclick − φno−click versus post-selection parameters δ, given in table 4.2. The plot
clearly shows that as δ becomes smaller, the effect of the post-selected single photon becomes
larger. For δ = 0.1, the smallest post-selection parameter used in our experiment, we measure
φclick − φno−click = 63± 18µrad. From a fit to φno−clicks versus |α|2 we measured a per photon
phase shift of φ = 7.6 ± 0.3µrad, which gives an amplification factor of of 8 ± 2. Hence, a
post-selected single photon can act like 8 photons.
To extract the value of φ+ and φ−, we fit the first 4 data points to the function φ++φ−2 +
Chapter 4. Experimental Results 74
0 20 40 60 80 100n
0
100
200
300
400
500
600
700
800
nocli
ck (
rad)
Figure 4.4: The measured phase shift for no-click cases versus the average photon number inthe signal beam |α|2 = n. A linear fit reveals a per-photon XPS of φ = 7.6± 0.3µrad.
φ+−φ−2δ . We use the measured value of 7.6µrad for φ++φ−
2 and leave φ+−φ− as a free parameter.
From the fit we infer φ+ − φ− to be 11.5± 1.5µrad. Knowing the sum and the difference of φ+
and φ−, we can infer their values to be 13±1µrad and 2±1µrad, respectively, which is different
from what one would have expected if the population was equally distributed in F = 2. The
values we measure for φ+ and φ− suggest that the the population is more likely to be in Zeeman
sub-levels with larger mF ’s. We believe that the probe beam is optically pumping the atoms
towards mF = 2 which results in φ+/φ− > 3.6.
To directly compare the effect of varying δ without changing the mean photon number, we
measure φclick − φno−click for δ = 0.14 and δ = 1, with almost the same mean photon number
(|α|2 = 45 and |α|2 = 40 respectively.) We measured 45.5 ± 13.0µrad for the former and
9 ± 10µrad for the latter. It is worth noting that for δ = 1 we do not expect to observe any
WVA effect and φclick − φno−click should be equal to φ.
4.2.4 Technical Details
Weakness Criteria. In derivation of weak value formulas in chapter 2, we had to make a few
assumptions to guarantee that the measurement is weak. First assumption was that |β|2×∆φ1, where |β|2 is the mean photon number in the probe beam and ∆φ = φ+ − φ−. Using 2000
Chapter 4. Experimental Results 75
0.0 0.2 0.4 0.6 0.8 1.00
20
40
60
80cli
ckno
click
(ra
d)
Datatheorytheory with 1intrinsic per photon phase shift
Figure 4.5: XPS difference versus post-selection parameter. Red circles are measured phaseshift difference between click and no-click cases. Gray dashed line is a fit to φ++φ−
2 + φ+−φ−2δ
with φ++φ−2 assumed to be 7.6µrad and φ+ − φ− left as a free parameter. The fit reveals a
value of 11.5± 1.5µrad for φ+ − φ−. The cyan solid line is the full theoretical calculation withthe numbers extracted from the two fits.
photons in the probe beam and ∆φ = 11µrad ensures that this condition is met. The most
important criteria for the weakness is given by equation 2.25 which says that the phase-shift
written on the probe should be smaller than the uncertainty in the phase of the probe. The
largest mean photon number used for the signal was |α|2 = 100, and φ = 7.6µrad. Therefore,
|α|2φ = 7.6× 10−4rad which is much smaller than 22mrad calculated phase uncertainty of the
probe. Experimentally, we measured 100mrad of phase uncertainty on the probe.
Background Counts. In section 3.3.3 we discussed how gating the SPCM with 40ns windows
reduced the background counts to about 6%. For every data point, the post-selection parameter,
detector efficiency and mean photon number in signal are set so that without background
photons, we expect the post-selection probability to be 19%. Table 4.2 shows the measured
probability of “click” for each data point. As you can see, expect for data point number 1 and
number 5, the rest agree very well with the expected background counts of 6% and true counts
of 19%. The reason background counts are much smaller for data point number 5 is that we
used an ND filter to reduced the overall collection efficiency of the detector, and as a result,
the background counts dropped to around 1-2 %. For data point 1, in the absence of the signal
pulses we measured the expected 6% dark counts and in the absence of probe (and presence of
the signal) we measured 20% true counts. But the presence of the two beams resulted in 30%
total counts. The reason for the discrepancy between the 30% measured background counts
for data point 1 and the expected 25% is a bit subtle. We believe that the main source of the
Chapter 4. Experimental Results 76
background photons is the scattered probe and coupling photons; if the probe sees a higher OD,
it will scatter more photons and as a result background counts will increase. For data point
1, we sent signal pulses with on average 95 photons. The ac-Stark shift due to 95 photons,
shifts the probe about 100KHz outside the transparency window and as a result the probe
sees a higher OD, and therefore, scatters more photons. The fact that these added background
photons only exist in the presence of the signal pulses makes them difficult to be further studied.
Stability Monitoring and Data Rejection Criteria. During the data collections, we continu-
ously monitored the OD and the “click” rates. The OD was monitored at two points in a trace:
once in between the shots when the probe was in the EIT (transparency OD), and once at the
end of the trace. The OD at the end of the trace was measured after turning the coupling beam
off.1 Typically we had transparency OD of 0.7-1 and final OD of 2.5-3. These three measured
quantities always fluctuate. We decided that when any one of the three deviated more than
10% from their initial value, we would stop the run and discard any measured XPS after that
point.
4.2.5 Discussion of the Results
I will now discuss the advantage of using the WVA over a similar experiment in which we
only used post-selection with no amplification[50]. In that experiment a combination of around
1 billion trials (300 million successful post-selection) were used to measure the XPS due to
single post-selected single photons with σ+ polarization. By looking at the difference between
the XPS measured for “click” and “no-click” events, we measured φ+ of 18 ± 4µrad. In this
experiment, where we use WVA technique, we used a total of around 830 million trials (200
million successful post-selection) to extract a φ+ of 13±1µrad. It is evident that the WVA
technique yielded a better signal-to-noise ratio. This advantage comes from two sources: first, by
doing a linear fit to the subset where the post-selection failed (“no-click” events), we measured
φ++φ−2 = 7.6±0.3µrad. The small uncertainty is due to using around 600 million measurement
and multiple photons for each data point, which is the first advantage. Second, we fit the
function 7.6 + φ+−φ−2δ to the measured φclick − φno−click for every δ. From the fit we inferred
φ+ − φ− to be 11.5± 1.5µrad. By combining this two results, we infer 13±1µrad. The second
advantage is due to the fact that the quantity φclick−φno−click gets amplified as δ gets smaller,
1Turning the coupling beam off removes the EIT condition on the probe and the measured OD would correpondto the “actual” OD of the cloud.
Chapter 4. Experimental Results 77
but since the probability of post-selection is proportional to δ2|α|2, one can maintain a certain
post-selection probability by increasing |α|2. Hence, the amplification is not accompanied by a
reduction in the size of “click” events, which would otherwise increase the uncertainty of the
post-selected data set. Due to this compensation, the SNR is improved.
Chapter 5
On the Advantage of Weak Value
Amplification
In this chapter, I will discuss the usefulness of weak measurement to the precision measure-
ment. I will introduce Fisher Information as a measure of the maximal information that could
be gained through measurement, and by using it as a figure of merit, I will show how weak
measurement performs compared to a standard measurement, where no post-selection is em-
ployed. I will focus on a specific noise model with long time correlation and demonstrate a
regime where weak value amplification can outperform a standard measurement.
The question of whether weak value amplification is useful or not, has gained a lot of
attention in the past few years. Finding a general and final answer to the question has proven
to be difficult and one can find arguments both in favor and against the usefulness of the
weak value technique (see the references in 1.1.) Some of the elements that contribute to
the difficulty of this question are difficulty in introducing a general enough noise model, using
optimal statistical techniques for data analysis, fair resource counting and consideration of
experimental complexities. Therefore, before I continue, I should clarify the context in which
I am comparing the weak value technique to the standard technique. In what follows in this
chapter, I will mostly consider weak value amplification of the nonlinear effect of a single
photon by sending it through an interferometer. For simplicity I will only consider the case
where only one of the paths of the interferometer interacts with the probe. The standard version
of this measurement would be to eliminate the interferometer and directly let the single photons
interact with the probe. No post-selection is employed in the standard technique. To make the
78
Chapter 5. On the Advantage of Weak Value Amplification 79
comparison between the two techniques fair, I assume that we have only N single photons to
perform the measurement with. The shot-noise of both techniques are assumed to be the same.
In the next section I will introduce the technique I will be using for data analysis.
5.1 Fisher Information as a Figure of Merit
A common problem in statistics is to know the minimum uncertainty that can be achieved for
measuring a quantity given a data set. In order to extract the maximal amount of information
(minimum uncertainty) from a data set, one needs to carefully study how correlated the data
is. When there are correlations (covariances) in a set of measurement results, not all the data
points have the same importance. Therefore, not all data points should be weighted equally
when trying to find the average of a quantity. Below, I will show how knowing the covariances
can help in finding the optimal weighting of different data points.
5.1.1 Covariance Matrix and Fisher Information
To find the optimal weights, which corresponds to an estimator with minimum uncertainty, one
should construct a matrix called the Covariance Matrix (CM). CM is a square matrix whose
size is equal to the size the data set. For a data set with ith element given by di ± σi, the
elements of the CM are calculated as Cij = 〈σiσj〉. The diagonal elements of the CM are equal
to the variance of the corresponding data point and the off diagonal elements are equal to the
covariance between two data points. Knowing the CM, the average of the data set can be found
by[31]
d =
∑i,j C−1
ij dj∑i,j C−1
ij
. (5.1)
Equation 5.1 reveals the optimal way of weighting the the data points in a data set. As an
example, figure 5.1 shows the calculated weights for 10 measurements by using a CM given
by Cij = exp(−|i − j|/2.) (exponentially decaying covariances.) As it can be seen, due to
correlation between the measurements, the initial and final measurement are given a larger
weight than the measurements in the middle, while the naive averaging would give each data
point a weight of 1/10.
From a maximum likelihood approach it can be shown the variance of a data set has a
lower bound called the Cramer-Rao Bound (CRB). It can be shown that the formula given in
Chapter 5. On the Advantage of Weak Value Amplification 80
0 2 4 6 8 10measurement index
0.00
0.05
0.10
0.15
0.20
0.25
0.30
weig
ht
optimalnaive
Figure 5.1: Optimal weights obtained by the covariance matrix Cij = exp(−|i − j|/2.) versusnaive averaging when all data points are weighted equally.
equation 5.1 is the unbiased estimator for the mean value that minimizes the variance. This
minimum variance can be calculated to be
Var(d)min =(∑
i,j
C−1ij
)−1(5.2)
One can formulate the CRB as Var(d) ≥ 1/I, where I is called the Fisher Information. There-
fore, given a CM, the Fisher information can be calculated as
I =∑
i,j
C−1ij . (5.3)
5.1.2 Fisher Information versus Signal-to-Noise Ratio
Often to quantify the information in a measurement, one would use the Signal-to-Noise Ratio
(SNR), which, as the name suggests, is the ratio between the size of the signal to the uncertainty
in the signal. Therefore, for a measurement result given by s ± σ1, the SNR is calculated as
SNR1 = s/σ1. If the uncertainty σ1 for this measurement is obtained by the technique given
by equation 5.1, then the Fisher information for this measurement is I1 = 1/σ21. If a different
approach was used to measure the same quantity, for example by amplifying the signal, and the
CM analysis of this second technique revealed A× s± σ2, where A is the amplification factor,
Chapter 5. On the Advantage of Weak Value Amplification 81
the SNR would have been SNR2 = A × s/σ2. However, for the second measurement, defining
the Fisher information as I2 = 1/σ22 is erroneous, because A×s is not an unbiased estimator for
the signal s. In order to correctly calculate the Fisher information, one should first make the
second measurement unbiased by writing it as s±σ2/A and the Fisher information can correctly
be evaluated to be I2 = A2/σ22. With the correct definition of the Fisher information, one can
see that I2/I1 =(SNR2/SNR1
)2. Therefore, to compare different measurement techniques, as
long as optimal estimators are used to calculate the SNR and unbiased estimators are used to
find the Fisher information, one can use either the SNR or the Fisher information to make the
comparison and get the same result.
5.2 Measuring the Effect of a Single Photon
In this section I will compare the Weak Value Technique (WVT) to the Standard Technique
(ST) for measuring the nonlinear effect of a single photon.1 I will use the Fisher information
as a figure of merit when doing the comparison. I will first compare the two in the case of
Gaussian white noise and then I will study the effect of adding correlations. At the end, I will
introduce an optimal weak value technique which also makes use of the rejected data and the
correlations between the post-selected data and the rejected data.
5.2.1 Gaussian White Noise
The nonlinear effect of a photon can be measured as a cross-phase shift (XPS) on a probe
coherent state, as studied in detail in chapter 2. In the ST, no post-selection is used and the
measured XPS on the probe will be given by φ0, the intrinsic per photon nonlinear phase shift.
If we assume the shot-noise of the phase measurement is given by σ0 and the measurement is
performed for N times, then the measured phase in the ST would be Sstd = φ0 ± σ0/√N . The
Fisher information can be found to be Istd = N/σ20.
In the WVT, the measured signal is given by 〈n〉w × φ0, where 〈n〉w is the weak value of
photon number. By a similar calculation to what resulted in equation 2.50, one can easily show
that
〈n〉w =1
cot θ + 1(5.4)
with probability of post-selection given by Pclick = (cos θ + sin θ)2/2 for θ ∈ [−3π/4,−π/4].
1It can be thought of as estimating φ0, the per photon nonlinear phase shift, or estimating χ(3).
Chapter 5. On the Advantage of Weak Value Amplification 82
Therefore, the weak value signal is given by Swv = φ0/(1 + cot θ) ± σ0/√PclickN . The Fisher
information can be shown to be Iwv = (N/2σ20)× sin2 θ.
0.0 0.2 0.4 0.6 0.8 1.0P
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4 /
(N/
2 0)
weak valuerejectedstandardweak value + rejected
Figure 5.2: Fisher information in WVT and ST versus post-selection probability.
Figure 5.2 plots the Fisher information in the WVT and the ST versus post-selection param-
eter. As it can be seen, the Fisher information in the WVT, only gets half as large as the Fisher
information in the ST. The reason is that in ST, one photon is sent through the interaction
region whereas in the WVT, because of the interferometer, only half a photon is sent through
the interaction region. Therefore, the number of photons used in the ST is a factor of two larger
than the number of photons used in the WVT, which results in the Fisher information to be
twice as large. Second striking feature in figure 5.2 is that the maximum Fisher information
is achieved for probability of post-selection of 50% and not for when the the probability of
post-selection is much less than one, i.e. where WVA happens. To see the reason behind this,
we should look at the Fisher information contained in the rejected portion of the data. The
size of the rejected data set is given by (1− Pclick)N . We can find the size of the signal of the
rejected data the fact that Pclick × Swv + (1 − Pclick) × Srejected = φ0/2. Therefore, the Fisher
information of the the rejected data can be calculated to be Irejected = (N/2σ20)× cos2 θ(figure
5.2). It is easy to show that for P=50%, the rejected data set averages to zero, and as a result,
carries no information. Because of that, the information in the post-selected portion of the data
is maximized at P=50%.
The reason behind both of the features mentioned above, is that we set the system up in
Chapter 5. On the Advantage of Weak Value Amplification 83
such a way so that only one of the arms of the interferometer is interacting with the probe.
To illustrate this point better, let’s compare this system with a similar system in which the z-
component of a spin-12 particle is being measured. Figure 5.3 shows the result of the calculated
0.0 0.2 0.4 0.6 0.8 1.0P
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4 /
(N/
2 0)
weak valuerejectedstandardweak value + rejected
Figure 5.3: Fisher information in WVT and ST versus post-selection probability.
Fisher information in the ST, the WVT and the rejected data for spin-12 particle. Evidently,
the Fisher information in WVT is maximized when the probability of post-selection vanishes
and is equal to the Fisher information in the ST. It is worth noting that for the weak value
of a single photon, if instead of measuring a signal that is proportional to the weak value of
the photon number in only of the arms of the interferometer, we were measuring the photon
number difference between the two arms - that is instead of measuring n1 we were measuring
n1 − n2 - the Fisher information versus post-selection probability would look exactly like what
is plotted in figure 5.3.
5.2.2 Noise with Time Correlation
In this section I will study the effect of noise with time correlation on the Fisher information
in the WVT and the ST. I will first introduce the method used for simulating the CM and
calculating the corresponding Fisher information. Next I will compare the Fisher information
in the WVT and the ST.
Chapter 5. On the Advantage of Weak Value Amplification 84
(a) (b)
Figure 5.4: Calculated covariance matrix for (a) 10 points correlated on average and (b) notime correlation. The noise for (a) is modeled as a slowly varying noise added to a white noise.
Method
To simulate a CM for a noise with certain correlation time and to find the Fisher information,
we take the following steps: First, we generate a noise array with a certain time correlation by
filtering a Gaussian white noise with a filter width that corresponds to the desired correlation
time. We create a matrix by multiplying the noise array by its transpose. This procedure is
then repeated over many trials and the final CM is calculated by averaging over all matrices
generated in each trial. Figure 5.4 shows two example of the simulated CM’s with and without
time correlations. Once the CM is calculated, we use equation 5.3 to calculate the Fisher
information.
To simulate the CM for WVT, in each trial we randomly select a subset of the noise. The
size of this subset is chosen to reflect the post-selection probability. The remainder of the noise
array constitutes the rejected subset. We then separately calculate the CM for the post-selected
group and rejected group.
At each trial, we also concatenate the post-selected noise array and the rejected noise array.
We also calculate a CM for the mixture of the two. This concatenated CM is used for the
optimal weak value technique. Note that the concatenated CM has two main diagonal blocks
that are equal to the post-selected CM and the rejected CM. The off-diagonal blocks of the
concatenated CM corresponds to the covariance between the post-selected data set and the
rejected data set.
Chapter 5. On the Advantage of Weak Value Amplification 85
WVT vs. ST
We now compare the Fisher information in the WVT to the Fisher information in the ST. The
Fisher information in the ST is calculated by summing over the elements of the inverse of CM
calculated above. The Fisher information in the WVT is calculates similarly by summing over
elements of the inverse of the post-selected CM. To make sure an unbiased estimator is used
for the WVT, the sum is multiplied by 〈nw〉2.
0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9P
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
Fish
er In
form
ation
1e14weak valuestandard
(a)
0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9P
2.0
2.5
3.0
3.5
4.0
4.5
5.0
5.5
6.0
Fish
er In
form
ation
1e13weak valuestandard
(b)
0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9P
1.0
1.5
2.0
2.5
3.0
3.5
Fish
er In
form
ation
1e13weak valuestandard
(c)
0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9P
0.2
0.4
0.6
0.8
1.0
Fish
er In
form
ation
1e13weak valuestandard
(d)
Figure 5.5: Fisher information in the standard technique and the weak value technique. Figure(a) plots the fisher information when measurements are uncorrelated. Figures (b),(c) and (d)are results for 0.8, 2 and 20 measurements per correlation time. A total of 200 data points areused in the data set. The CM’s are calculated by averaging over 10K trials.
Figure 5.5 plots the simulation results for having 0, 0.8, 2 and 20 correlated measurements.
Evidently, as the number of correlated measurements increases, the Fisher information in both
the WVT and the ST decreases. Fisher information in the ST is more sensitive to correlations
Chapter 5. On the Advantage of Weak Value Amplification 86
and decreases faster than it does in the WVT. Therefore, as the correlation time increases, it
is advantageous to use the WVT over the ST. Also, it is worth noting that as the correlation
time increases, the maximum of the Fisher information in the WVT occurs at lower values
of post-selection parameter (P→0). Therefore, for long time correlations, not only the WVT
becomes superior over the ST, it is also advantageous to chose the post-selection parameter to
be in the weak value amplification regime.
To understand why post-selecting on a small portion of the data in the WVT is advantageous
in this regime, we should first note that the reason behind repeating a measurement to carefully
study the noise by sampling it as much as possible. But if the noise has time correlation,
sampling the noise multiple times within one time correlation will not better our understanding
of the noise, and one point within a correlation time is as good as many. To illustrate this
point, figure 5.6 shows two noise traces, one with long time correlations (top) and the other is
white noise (bottom.) The dots represent sampling both noises 100 times at a fixed rate. The
red dots are a random selection from these 100 measurements. It can be seen that in the case
of white noise, the randomly selected portion doesn’t sample the noise as well as the original
100 points. This loss in sampling is much less pronounced in the noise trace with long time
correlations. Hence, by sampling the noise once per correlation time and amplify the signal at
the same time, which is what weak value amplification naturally does, one would get a better
SNR (higher Fisher information.)
Figure 5.6: Two noise traces, with long time correlations(top) and without time correlation(bottom). The dots represents sampling from the noise with a fixed rate. The red dots are a20% random selection from them.
Chapter 5. On the Advantage of Weak Value Amplification 87
Optimal Weak Value Technique
In previous subsection, I ignored the role of the rejected data. I now calculate the Fisher
information that is left in the rejected portion of the data and in the correlation between
the post-selected data and the rejected data. To find the Fisher information, I first calculate
the inverse of the concatenated CM. Summing over the elements that correspond to the post-
selected data (and multiplying by 〈nw〉2 to make it unbiased) reveals the Fisher information
in the post-selected data. Note that this Fisher information is different from the one calcu-
lated in the previous subsection, because it is calculated from a different CM. Summing over
the elements that correspond to the rejected data and correlations (and multiplying them by
appropriate factors to make them unbiased) reveals the Fisher information in the rejected data
and the correlations. The total Fisher information can be found by summing over the individual
Fisher informations in each subset. I will show that the optimal weak value technique always
outperforms both the WVT and the ST.
Figure 5.7 plots the calculated Fisher information in the ST, the WVT, rejected dataset
and the correlations versus post-selection parameter, as the correlation time of the noise in-
creases. Evidently, as the time correlation becomes longer, the information in ST dramatically
decreases, whereas the maximum information in the optimal technique remains almost constant.
Therefore, the optimal weak value technique can eliminate the effect of the added correlations.
5.2.3 Advantage of Post-selection without Weak-Value Amplification
In this section, I will show that in a certain regime just doing post-selection, without weak value
amplification, can be advantageous over the standard technique. As an example, consider that
you want to measure the per photon phase shift of a nonlinear medium (measuring χ(3)) but
as a resource you have only coherent states with on average a tenth of a photon (|α|2 = 0.1.)
In the ST, the measured signal would be 0.1×φ0±σ0/√N . However, by using a single-photon
detector with 100% efficiency, one could condition the measurement on a successive photon
detection of the signal. The post-selection probability would be 10% and the measured signal
would be φ0±σ0/√
0.1×N . It can be seen that in this case the SNR has improved by a factor
of√
10 by just doing the post-selection.
Figure 5.8 plots the Fisher information in the ST and in post-selected data versus |α|2
for three different detector efficiency. The probability of post-selection is given by Pps = 1 −
Chapter 5. On the Advantage of Weak Value Amplification 88
0.0 0.2 0.4 0.6 0.8 1.0P
0
1
2
3
4
5
6Fi
sher
Info
rmati
on
1e13weak valuerejectedcorrelationsstandardOptimal
(a)
0.0 0.2 0.4 0.6 0.8 1.0P
0.0
0.5
1.0
1.5
2.0
2.5
3.0
Fish
er In
form
ation
1e13weak valuerejectedcorrelationsstandardOptimal
(b)
0.0 0.2 0.4 0.6 0.8 1.0P
0.0
0.5
1.0
1.5
2.0
2.5
Fish
er In
form
ation
1e13weak valuerejectedcorrelationsstandardOptimal
(c)
0.0 0.2 0.4 0.6 0.8 1.0P
0.0
0.5
1.0
1.5
2.0
2.5
3.0
Fish
er In
form
ation
1e13weak valuerejectedcorrelationsstandardOptimal
(d)
Figure 5.7: Fisher information in the standard technique, weak value technique, the rejectedsubset and correlation between the post-selected subset and rejected subset for various corre-lation times. The “optimal” corresponds to summing over the Fisher information in WVT,rejected data and the correlations. Figure (a) plots the fisher information when measurementsare uncorrelated. Figures (b),(c) and (d) are results for 0.8, 2 and 20 measurement shots percorrelation time.
exp(−η|α|2) and the mean photon number conditioned on post-selection is given in equation
2.15. As it can be seen, in the case of a detector with 100% efficiency it is always advantageous
to use post-selection.2 As the efficiency of the detector decreases, the range of |α|2’s in which
post-selection is advantageous decreases. The noise is considered to be white in all these cases.
2for large |α|2, when Pps → 1, the two techniques will have the same Fisher Information.
Chapter 5. On the Advantage of Weak Value Amplification 89
0.0 0.2 0.4 0.6 0.8 1.0| |2
3.0
2.5
2.0
1.5
1.0
0.5
0.0
log(
/(N
/0
))
StandardPost-Selection
(a)
0.0 0.2 0.4 0.6 0.8 1.0| |2
3.0
2.5
2.0
1.5
1.0
0.5
0.0
log(
/(N
/0
))
StandardPost-Selection
(b)
0.0 0.2 0.4 0.6 0.8 1.0| |2
3.0
2.5
2.0
1.5
1.0
0.5
0.0
log(
/(N
/0
))
StandardPost-Selection
(c)
Figure 5.8: The effect of post-selection without weak-value amplification on Fisher informationversus |α|2. The detector efficiency used for the plots are 100%, 20% and 1% for (a),(b) and(c) respectively.
5.3 Experimental Proposal and Progress
In the previous section I showed how in the presence of noise with time correlation, weak value
amplification can be advantageous. In this section I will propose an experiment to demonstrate
this advantage. First I will extend this idea to the regime where a coherent state is used
instead of single photon. Then I will discuss how to generate a controlled time-correlated noise
experimentally. Finally I will report on the experimental progress made so far.
5.3.1 Fisher Information for Coherent States as the Signal
Here, I will calculate the Fisher information in the post-selected data in the experiment de-
scribed in chapter 4 and will compare it to the Fisher information in a similar experiment
without post-selection. For the weak value technique, we use the weak values given in equation
2.50. The signal for the ST is equal to |α|2 × φ0. Figure 5.9 plots the calculated Fisher infor-
mation for the two techniques as the number of measurements per correlation time increases. I
have used parameters δ = 0.26, |α|2 = 15 and η = 0.2 for this calculation. Evidently, when the
number of measurements per correlation time is very small (white noise regime) the ST outper-
forms the WVT. This is due to the fact that the amplification factor for the chosen parameter
is not large enough to compensate for the reduction in the size of the dataset. However, as the
number of correlated measurements increases, the Fisher information in the WVT is affected
less than the Fisher information in the ST. Therefore, in the regime where the measurements
are highly correlated, WVT becomes superior.
Chapter 5. On the Advantage of Weak Value Amplification 90
0.002 0.02 0.2 1.0 2.0 4.0 10.0 20.0 40.0 100.0Number of measurements per correlation time
12.0
12.5
13.0
13.5
14.0
14.5
log(
)
WVTST
Figure 5.9: Fisher information for the WVT and ST versus number of correlated measurements.We assumed δ = 0.26, |α|2 = 15 and η = 0.2.
5.3.2 Experimental Progress
In order to use our setup to demonstrate the advantage of the WVT over the ST, we need to add
noise with time correlations to our phase measurement. It is very important to add the noise
in such a way that it doesn’t get amplified when doing weak measurement. The added noise
should not cause further fluctuations in the interaction region and should not interfere with the
post-selection probability. Since the off-resonance portion of the probe does not interact with
any part of the setup and is only used as a reference for the probe beam, it is the perfect place
to add the slow noise. Adding slow phase noise to the off-resonance portion of the probe causes
slow drifts in the measured phase without interrupting the interaction between the signal and
the probe.
When we were first designing the phase measurement apparatus, we made it to be robust
to slow phase drifts. By doing reference measurements before and after when we the XPS is
expected, we filter out any slow phase drift. This, however, is an obstacle in front of adding slow
phase noise to the probe; the slow phase noise added to the probe would get completely filtered
out by our phase measurement. To get around this, we decided to only add a sudden phase jump
during the time we expect the XPS to occur. By slowly changing the magnitude of the jumps,
in accordance to a slow noise trace, we managed to add the slow noise drift to the measured
phase while keeping the measurement robust to all other slow phase drifts. Figure 5.10 shows
the schematic of the setup used to add slow phase noise. We have programmed the task control
computer so that the two analog cards controlling the setup are timed independently from each
Chapter 5. On the Advantage of Weak Value Amplification 91
RFSwitch
Phase-Shifter AOM
DCOffset
Slow Noise
Signal Gates
Gates Noise
10MHzReference
70MHz RF with Phase-Noise
Probe withPhase-Noise
Figure 5.10: Schematic of the setup to add slow phase-noise to the probe beam.
other, see section 3.17. While one of the cards control the experiment, the other card generates
a stream of slow noise. We can control the time correlation of this noise as well as its offset and
amplitude. We can choose between filtered Gaussian white noise or 1/f noise as well as different
modulation functions (square, saw-tooth, sinusoidal, etc.). We can also load noise from a saved
file. The generated noise is then gated using an RF switch to only be on for a 100ns every 2.4µs
(length of a shot.) An RF switch switches between the noise trance and a DC offset and uses a
100ns square modulation signal as the control. The DC offset is used to ensure that the gated
phase averages to zero. We employ a Phase-shifter circuit[72], to generate a 70MHz RF signal,
where the phase of the signal is controlled by the gated phase. A stable 10MHz references is
used to lock-in the RF to the phase measurement apparatus for demodulation purposes. The
RF drives an AOM, where the 1st order of the AOM, which contains the phase jumps, is used
as the off-resonance probe and the 0th order is used as the on-resonance probe. Therefore, the
phase of the beating signal of the on- and off-resonance probe will change in accordance to the
added slow noise. A delay in gating is used to ensure that the phase jumps happen exactly
when the XPS is expected. The apparatus mentioned above has been built and tested. The
Phase-shifter circuit is well calibrated and all the elements are in good shape for the experiment
to be carried out.3
3A sudden and unfortunate death of the turbo pump that maintains the vacuum in the chamber preventedus from moving forward and the experiment needed to be postponed until the vacuum problem is resolved.
Chapter 6
Outlook and Summary
In this chapter I will discuss the limitations of the current apparatus and possible routes to
overcoming them. These improvements primarily seek to increase the size of the nonlinearity.
A larger nonlinearity will open up the path to many experiments in quantum nonlinear optics,
which due to the small size of the current nonlinearity are experimentally inaccessible. I will
also discuss alternative experiments outside of quantum nonlinear optics that can be studied in
the future. In the last section, I will summarize the achievements of the thesis.
6.1 Possible Improvements
Alternative Level Scheme
As mentioned in section 5.2.2, the Fisher information in the post-selected data will be maximum
at very small post-selection parameter, if we were measuring n1 − n2 instead of just n1. To
perform this measurement, the two paths of the interferometer should interact with the probe
with the same strength but opposite signs, so that the XPS written on the probe would be
φ0 × (n1 − n2). The other advantage of having such interaction is that since φ1 = −φ2 = φ0,
the XPS conditioned on a successful post-selection (from equation 2.35) becomes φclick = φ0/δ
and the effect of the coherent state will be automatically canceled out. One way to realize such
interaction is by using the level scheme shown in figure 6.1. First, one needs to optically pump
all the population to F = 2,mF = 2 ground state. A combination of a σ+-polarized probe
and a π-polarized coupling beam creates the EIT. The signal is an equal superposition of σ+
and σ− polarization. Using a DC magnetic field, we can control the Zeeman shift between the
92
Chapter 6. Outlook and Summary 93
sub-levels of the F ′ = 4 excited state, and therefore, we can make σ+ polarization component of
the signal to be red-detuned and σ− component to be blue-detuned, which makes the ac-Stark
shift of he two polarizations to have opposite sign. By fine tuning the magnitude of δ1 and δ2
in accordance to the CG coefficients of the two transitions, we can ensure that the ac-Stark
shifts due to the two polarizations have the same magnitude. This will result in φ1 = −φ2, and
therefore, by using such level scheme, we can measure n1 − n2.
9/64" 3/8"
mF = 2 mF = 3 mF = 4
F 0 = 4
F = 3
F 0 = 3
F = 2
12
Figure 6.1: Schematic of the improved level scheme.
Filtering Background Photons
Eliminating the background photon detections will enable us to perform the experiment with
lower signal photon numbers and/or to use much smaller post-selection parameter to gain larger
amplification factors. Although gating the SPCM with smaller window can reduce the chance
of a background photon detection, but we are at the limit of the smallest time window we can
use. One way to reduce the background photons is to couple the signal to a single-mode fiber
instead of a multi-mode fiber. The downside, however, is that the collection efficiency of the
signal will also be reduced. Recently, D.T. Stack et al.[73] demonstrated that by using a vapor
cell containing 85Rb and a buffer gas, they could attenuate noise photons by up to two orders
of magnitude in their quantum memory scheme. A similar technique can be pursued to filter
the background photons in our setup.
Chapter 6. Outlook and Summary 94
Brighter Single Photon Sources
We are currently equipped with a single photon source that is compatible with Rb atom[74].
The best single photon rate we obtain from this source, however, is 100 per second. Observing
the effect of these single photons with our current setup needs continuously measuring for 10
days. Therefore, exploring alternative narrow-band single photon sources with high brightness
is necessary. One promising alternative sources uses four-wave-mixing in a atomic sample to
generate a narrow-band pair of photons[75, 76]. We are currently equipped with necessary
elements for generating photon pairs in atomic ensembles. Exploring this system might prove
it to be an easy and efficient alternative for the photon source we currently possess.
Different MOT Geometries and BEC
Higher optical densities could enhance the size of the nonlinearity. Although, due to group
velocity mismatch and signal absorption, one may not benefit from a high OD. Nonetheless, a
proper choice of level scheme, or using EIT schemes other than N-scheme[77, 78, 79], could put
us in a regime where the size of the nonlinearity grows with OD. However, due to the radiation
pressure from the scattered photons within the cloud, the density of a typical MOT cannot
surpass a certain amount. To obtain higher densities or OD’s, one need to explore different
MOT geometries[80, 81, 82, 83].
An alternative option is using a Bose Einstein Condensate (BEC) to mediate the interaction
between the probe and the signal. However, since the smallest photon scattering could destroy
the BEC, one needs to be careful when trying to obtain an EIT in a BEC without scattering
photons. Also, the BEC apparatus typically has a low duty cycle, therefore it is not suitable
for measurements where hundreds of millions of trials are needed. A BEC could be desirable if
one could obtain large enough nonlinearity to be close to single shot resolvability.
6.2 Future of the current Apparatus
In this section I discuss possible experiments that can be carried out in the near future with
small modifications to the current setup.
Chapter 6. Outlook and Summary 95
Imaginary Weak Values
In this thesis, I limited my discussions to real weak values. But as the formula for weak values
suggests, the weak value can be imaginary as well. In an experiment similar to what was carried
out in this thesis, imaginary weak values can be measured if one post-selects on a polarization
that has an imaginary overlap with the initial polarization, which will result in an imaginary
δ. The effect of the imaginary weak value can be observed as a shift in the amplitude of the
probe instead of its phase. Observing an imaginary weak value is fundamentally interesting as
well useful in quantum metrology[45, 46].
PNR detectors and Observing the Effect Higher Photon Detections
In section 2.2.4 we discussed the effect of using a PNR detector on the inferred photon number.
We also showed that post-selecting on an n-photon state in weak measurement results in an
amplification of all n photons. Therefore, using a PNR detector[55] or branching SPCMs with
beam splitters can be a possible experiment for the future to observe the effect of the n-photon
detections.
6.3 Revisiting Optical Lattices
The apparatus used in this thesis has the capability of carrying out optical lattice experiments.
Several coherent control, state manipulation and quantum chaos experiments were performed
using vibrational states of optical lattices on this setup[84, 85, 86]. Here we discuss two possible
experiments that could be done with optical lattices.
Rabi Oscillation and Breakdown of Rotating Wave Approximation
The coupling between different vibrational states of the optical lattice can be done by displacing
the lattice. A periodic displacement can act like a dipole coupling between adjacent vibrational
states, and one can observe Rabi oscillation. The Rabi frequency in our system can be made
large enough to be comparable with the energy difference between the two levels. Our sim-
ulations indicate that in this regime, the fast oscillating terms in the Hamiltonian cannot be
ignored and one should expect to see fast oscillations on top the expected Rabi oscillations. This
could be the first demonstration of breakdown of rotating wave approximation by observing
the fast population oscillations in real time.
Chapter 6. Outlook and Summary 96
Quantum Heat Engine
In 2010, Linden, Popescu and Skrzypczyk proposed a scheme to make the smallest possible self-
contained heat engine[87, 88]. Due to its need of three-body interactions and thermalization
of qubits, however, the scheme is difficult to be experimentally realized. Using a combination
of a one dimensional optical lattice and the internal electronic states of atom, one can tailor
an effective interaction between three “degrees of freedom” of a single atom. Thermalization
could be achieved by making an EIT dark state between two electronic states that constitute
the qubit, where the population ratio (temperature) can be controlled by adjusting the relative
power of the beams creating the EIT. Further research is needed to confirm the feasibility of
the proposed scheme. If implemented, it would be the first ever observation of the smallest
possible heat engine.
6.4 Rydberg Atoms
One way to enhance the nonlinearity at the single photon level, is to use exotic systems such as
cavity quantum electrodynamics[89], atoms inside hollow core fiber[90, 91, 92], atoms trapped
around a tapered nanofiber[93], a single atom strongly coupled to a micro-resonator[94, 95] and
very recently, using Rydberg atoms. Long-range strong interactions between two Rydberg atoms
and Rydberg blockade effect have been used to demonstrate strong nonlinearities[96, 97, 98, 99].
Transitioning to use Rydberg atoms is a natural next step for the future of this apparatus for
further studying of interaction between two single photons, and even generation of non-classical
states of light. Experiments involving Rydberg atoms distance themselves from the scheme pro-
posed by Monro and Nemoto[100, 101], in which the goal is to use relatively weak nonlinearities
and non-demolition measurement to perform a phase-gate. However, the possibility of obtain-
ing large enough nonlinearities using Rydberg atoms is still an open questions and experimental
investigation in needed to shed light on the subject[102, 103].
6.5 Summary
In this thesis, I reported on the experimental observation of amplifying the effect of a single
post-selected photon using weak measurement. I showed that a single photon, when prop-
erly pre- and post-selected, can write eight photons’ worth of phase shift on a coherent state
Chapter 6. Outlook and Summary 97
probe. I demonstrated that this technique, with comparable resources, resulted in a better
signal-to-noise ratio than a similar experiment which only utilized post-selection but no weak
value amplification. The presented experiment is the first ever demonstration of weak value
amplification using a deterministic weak interaction between two optical systems that cannot
be explained classically.
A detailed theoretical background of the experiment was also provided in this thesis. Using
the weak value approach, I calculated the weak value of photon number inside the two arms of
an interferometer versus the post-selection parameter. I also showed that the same result can
be derived from the Schrodinger equation. I studied the effect of detectors with and without
photon number resolvability and showed amplification of higher photon number states can be
achieved by using a photon number resolving detector. I also studied the effect of measurement
back-action on the expected weak values, and compared the weak value to what one would
expect to measure in a similar experiment but with strong interactions.
Finally, I used the Fisher information as a metric to compare weak measurement to a
standard measurement. I showed that in the white noise regime, the maximum information
weak values contain can be as large as the information in the standard measurement. By
using a noise model with time correlations, I demonstrated that as the number of correlated
measurements increases, the Fisher information in the standard technique decreases much faster
than it does in the weak value technique. An optimal weak value technique was introduced
in which one makes use of all the information in the weak value data set, the rejected data
set and the correlation between the two data sets. It was shown that the optimal weak value
technique is always superior. I proposed an experiment to demonstrate the advantage of weak
measurement and reported on the progress made towards its implementation.
Appendix A
Weak Value Approximations
The weak value formula was derived in equation 2.8. Two approximations were made in this
derivation: first, I approximated exp(−igP .A) as I− igP .A. Second, I wrote (I− ig 〈f |A|i〉〈f |i〉 )|ψ0〉as exp(−igAwP )|ψ0〉, where Aw is the weak value defined as 〈f |A|i〉/〈f |i〉. I now discuss the
conditions necessary for these approximations to be valid.
The combined state of the system and probe after the interaction can be written as
e−igP .A|i〉|ψ0〉 =∞∑
n=0
(−ig)n
n!PnAn|i〉|ψ0〉. (A.1)
After post-selecting the system in final state |f〉, the state of the probe becomes
|ψf 〉 =
∞∑
n=0
(−ig)n
n!Pn〈f |An|i〉|ψ0〉. (A.2)
Due to the post-selection, this state is not normalized. In section 2.1.2, I normalized the state
by the normalization factor 1/〈f |i〉, assuming that the probability of the post-selection is given
by |〈f |i〉|2. One can also find the normalization factor from equation A.2
N2 =
∞∑
n,m=0
〈ψ0|[(ig)m
m!Pm〈i|Am|f〉
][(−ig)n
n!Pn〈f |An|i〉
]|ψ0〉
=
∞∑
n,m=0
(−1)n(ig)m+n
m!n!〈i|Am|f〉〈f |An|i〉〈Pm+n〉ψ0 ,
(A.3)
where 〈P k〉ψ0 is the expectation value of P k with respect to the probe’s initial state |ψ0〉, and
98
Appendix A. Weak Value Approximations 99
1/N is the normalization factor. Equation A.3 can be written as
N2 = |〈f |i〉|2 − ig〈P 〉ψ0
(〈i|A|f〉〈f |i〉 − 〈f |A|i〉〈i|f〉
)
+ (ig)2〈P 2〉ψ0
(1
2〈i|A2|f〉〈f |i〉+
1
2〈f |A2|i〉〈i|f〉 − 〈i|A|f〉〈f |A|i〉
)+O(g3〈P 3〉ψ0),
(A.4)
which can be simplified to
N2 = |〈f |i〉|2[1 + 2g〈P 〉ψ0 Im(Aw)− g2〈P 2〉ψ0
(Re(A(2)
w )− |Aw|2)]
+O(g3〈P 3〉ψ0), (A.5)
where A(k)w is defined as 〈f |Ak|i〉/〈f |i〉. The same result was derived in [6]. Evidently the
normalization factor is given by 1/〈f |i〉 to the 0th order in g〈P 〉ψ0 . The first order will vanish
if one choses an initial probe state with zero mean momentum, 〈P 〉ψ0 = 0, or a post-selection
parameter which results in a real weak value. In this case, the deviation of |〈f |i〉|2 from equation
A.5 will be in the order of g2〈P 2〉ψ0 . For a probe state with zero mean momentum, 〈P 2〉ψ0 is
equal to the variance in the momentum distribution. Hence, the following condition ensures
that the deviation of |〈f |i〉|2 from equation A.5 would be vanishingly small
VarP 1
g2(Re(A(2)w )− |Aw|2)
. (A.6)
If the system is post-selected in the same state as the initial state, Re(A(2)w ) − |Aw|2) will be
the variance of A with respect to the initial state of the system, |i〉. Therefore, condition A.6
will become VarP 1/(g2VarA). This condition can more intuitively be understood as the
variance in the position of the probe should be much larger than g2VarA.
I now derive the condition necessary for validity of equation 2.8, assuming the condition
given in A.6 is satisfied. Using 1/〈f |i〉 as the normalization factor, the final state of the probe
after the post-selection (equation A.2) becomes
|ψf 〉 =
∞∑
n=0
(−ig)n
n!Pn〈f |An|i〉〈f |i〉 |ψ0〉. (A.7)
The final state of the probe in equation 2.8 was approximately written as
|ψ(a)f 〉 = e−igAwP |ψ0〉 =
∞∑
n=0
(−ig)n
n!PnAnw|ψ0〉. (A.8)
Appendix A. Weak Value Approximations 100
For |ψ(a)f 〉 to be a valid approximation we must have 〈ψf |ψ(a)
f 〉 ≈ 1. Therefore
〈ψf |ψ(a)f 〉 = 〈ψ0|
[ ∞∑
n=0
(ig)n
n!Pn〈i|An|f〉〈i|f〉
][ ∞∑
m=0
(−ig)m
m!PmAmw
]|ψ0〉
=∞∑
n,m=0
(−1)m(ig)m+n
m!n!〈Pm+n〉ψ0A
(n)w Amw .
(A.9)
It is easy to show that for m = n = 0, (−1)m (ig)m+n
m!n! 〈Pm+n〉ψ0A(n)w Amw = 1. Therefore, for
〈ψf |ψ(a)f 〉 to be approximately 1, all other terms in the summation should be vanishingly small.
I assume that the initial momentum wavefunction of the probe is a Gaussian function centered
at zero with a width given by σp. For this wavefunction, it is easy to show
〈PN 〉ψ0 =
(N − 1)!!σNp if N is even
0 if N is odd,
(A.10)
where (N − 1)!! = 1 × 3 × 5 × ... × (N − 1). Hence, for m + n = (2k + 1) , k = 1, 2, 3, ..., the
terms in the summation in equation A.9 are zero. For m+ n = 2k , k = 1, 2, 3, ..., we have
gm+n
m!n!〈Pm+n〉ψ0A
(n)w Amw =
gm+n(m+ n− 1)!!
m!n!σm+np A(n)
w Amw 1. (A.11)
Since (n − 1)!! < n!, for inequality A.11 to be true, we need to have gm+nσm+np A
(n)w Amw 1.
Therefore, the condition necessary for the validity of equation 2.8 is
σp 1
g ×(A
(n)w Amw
)( 1m+n
). (A.12)
Unless one can prove that the terms in the summation in equation A.9 approach zero faster
than a geometric series, the condition A.12 should be true for all m’s and n’s. No general
general statement can be given on that regard unless A(n)w is calculated as function of n for a
specific system.
In the original AAV paper [4], the condition for validity of weak values was presented as
σp maxn|〈f |i〉|
|〈f |An|i〉| 1n. (A.13)
However, no derivation for this derivation was presented. Later, in [5], the authors claimed that
Appendix A. Weak Value Approximations 101
the condition introduced by AAV was incorrect and instead introduced two other conditions.
One condition simply says σp 1/Aw and the second condition is
σp minn
∣∣∣ |〈f |A|i〉||〈f |An|i〉|∣∣∣1/(n−1)
. (A.14)
This condition was also not rigorously derived in [5].
Appendix B
Alignment Procedures
B.1 Polarization Spectroscopy
The following steps make the alignment of the polarization spectroscopy easier and ensure a
healthy signal. See figure 3.3 for reference.
• 1) Make sure that the probe and pump beams are overlapping inside the rubidium vapor
cell. You may find it easier to first overlap the beams and then to put the cell in place.
Also make sure the probe is well aligned with the two photo diodes
• 2) Block the beams hitting the photo diodes and look at the error signal on an oscilloscope
• 3) Turn the offset nob on the front of the PID circuit box to make sure the error signal is
zero
• 4) Block the pump beam.
• 5) Make sure the frequency of the laser is not resonant with any of the Rubidium transi-
tions and unblock the two beams hitting the two photo diodes
• 6) By adjusting the HWP before the PBS adjust the power balance of the two beams
hitting the photo diodes; if they are balanced, the error signal should be zero
• 7) Unblock the pump beam, vary the piezo voltage on the laser driver until you see a
transition. If you have followed the previous steps correctly, you should not be seeing any
Doppler background
102
Appendix B. Alignment Procedures 103
• 8) Once you find the wanted transition, rotate the QWP to adjust the slope and shape of
the wanted feature
B.2 Injection-Lock
Take the following steps to improve the alignment of the injection lock. The optical elements
mentioned below refer to figure 3.7.
• 1)Place the H-OI in front of the slave diode laser. Rotate the polarizer until the power
passing through the H-OI is maximized. This ensures that the polarizer is aligned to the
output polarization of the diode laser.
• 2)Rotate the HWP2 before the PBS until the output power, which reflects off of the PBS,
is maximized.
• 3)Rotate the HWP1 at the seed input to maximize the input transmission through the
PBS. If the first two steps are done correctly, the input should also be passing thought he
H-OI with maximum power.
• 4)overlap the input beam with the output of the diode laser at a point close to the diode
by adjusting the mirror 1.
• 5)Rotate the HWP2 for a small amount just to let a small portion of the output of the
diode laser to leak into the input port.
• 6)Using mirror 2, overlap this leaked portion with the input light at a point as close to
the input port as possible.
• 7)Repeat steps 4 and 6 until the two beams completely overlap at the two points.
• 8)Rotate HWP2 back to maximize the power in the output
• 9)Look at the output spectra at the Fabry Perot cavity and ensure that the slave laser is
following the seed by adjusting the current to the diode and its temperature
B.3 Tapered Amplifier
• 1)Make sure that the TA chip is running at low current (<0.2A) to avoid
exposing yourself to high optical power
Appendix B. Alignment Procedures 104
• 2)Remove the cylindrical lens and the slit. You should unscrew the cage mounts holding
them in place and slide them out.
• 3)Align the input beam to the ASE coming out of the input port. Make the spatial mode
of the two beams are matching.
• 4)The input lens is mounted on a z-translation mount. While monitoring the output
power of the TA, gently move the lens in the z direction.
• 5)Realign the input beam and try to maximize the output power. If you see an improve-
ment, keep moving the input lens in the same direction. If the power is getting lower,
move the lens in the opposite direction. Repeat these two steps until the maximum output
power is achieved.
• 6)The output lens in mounted on a x,y,z-translation mount. Move it in the z direction to
collimate the beam in vertical direction (see figure 3.9 side view.)
• 7)Slide the slit back in place. The correct position for the slit is the point where the
horizontal component of the beam is focused (see figure 3.9 top view.) Move the output
lens in x and y direction if necessary. If the slid is in its correct position, you should not
be able to see any interference fringes along the path of the beam.
• 8)Slide the cylindrical lens back in place and move in z direction to collimate the beam
in horizontal direction.
B.4 Polarization Maintaining Fiber
• 1) Couple the light into the fiber. Make sure the input beam has linear polarization.
• 2) Place a HWP at the input and a second HWP, a polarizer and a photo-detector at the
output. See figure B.1.
• 3) Frequency modulate the input beam. Select the largest possible frequency range for
modulation. (Modulating the piezo of the Master Laser can modulate the frequency of
the beam in a several GHz range.)
• 4) You should see the power that is measured at the photo-detector to be modulated at the
same modulation frequency. Rotate the HWP at the output to maximize the amplitude
Appendix B. Alignment Procedures 105
Frequency
time
Power
time
HWP HWPPM Fiber
Polarizer
Detector
Figure B.1: Schematic of setup used to align the input polarization of the PM fiber.
of this modulation. This indicated that the polarization inside the fiber is changing with
frequency and the input polarization is not right.
• 5) Rotate the HWP at the input to minimize the power modulation amplitude seen on
the photo-diode.
• 6) Rotate the HWP at the output to maximize the amplitude of the modulation. If the
changes are being made in the correct direction you should observe an overall decrease in
the amplitude of power modulation.
• 7) Repeat steps 5 and 6 until you don’t observe any power modulation at the output.
B.5 MOT
(The x,y and z direction and optical elements mentioned in what follows, refer to the directions
and corresponding elements indicated in figure 3.14)
• 1) Make sure the center of two coils are perfectly aligned in the z direction.
• 2) Mount the cage system and overlap the trapping beam and the repumper beam. The
fiber to free space adapters are mounted on xy-translation cage mounts. Use them to
Appendix B. Alignment Procedures 106
x y
6cm
4.5cm
2.5cm
3.7cm
8.3cm
z
5.5cm
4cm
6cm
Coil
Cuvette
(a)
(b)
Figure B.2: (a)Front (left) and side view (right) and (b)3D visualization of relative position ofMOT beams with the cuvette and magnetic field coils.
ensure the repumper beam is perfectly at the center of the trapping beam.
• 3) Place two cage-mountable iris diaphragms in the outputs of the cage system.
• 4) Put PBS2 in place and make sure the reflected beam (path 1) is sent through the
center of the magnetic field gradient. You can measure where the beam is entering and
exiting the cuvette to ensure the beam is roughly passing through the center of the trap.
See figure B.2
• 5) Using the same measurements, align the paths 2 and 3. Now all beams in x, y and z
directions should be roughly overlapping at the center of the cuvette. (if the Rb pressure
inside the cuvette is high enough, you should be able to see them overlapping using an
IR-viewer.)
• 6) Place Q1, Q3 and Q5 in paths 1 , 2 and 3. Set the angles so that the polarization of
the beams in the three paths are circular. Important Note: Paths 1and 2 should have
the same circular polarization and path 3 should have the opposite circular polarization.
• 7) Place Q2, Q4 and Q6 in place. If the polarization in the paths are perfectly circular,
the angle of Q2, Q4 and Q6 should not be important. However, if the polarization
in those paths are elliptical, it is better to set the angle of Q2, Q4 and Q6 so that the
linear polarization component of the reflected light is orthogonal to the linear polarization
Appendix B. Alignment Procedures 107
component of the incoming beam. This results in lower temperature when doing molasses
cooling.
• 8) Overlap the retro-reflected beams with the incoming beams.
• 9) Turn on the magnetic field gradient and open the iris diaphragms all the way. You
should be able to see the MOT now. (If you don’t, the problem might be that the circular
polarization setting doesn’t match the direction of the magnetic field gradient. Flip the
direction of the magnetic field by flipping the direction of the current. If the problem
persists, check the polarization setting.)
• 10) Once you see a cloud, start closing the iris diaphragms slowly until the MOT start to
disappear.
• 11) Improve the alignments of the three paths along with the retro-reflected beams to
improve the MOT.
• 12) Repeat steps 10 and 11 until the relative alignment of the three beams are improved.
Bibliography
[1] Yakir Aharonov and Lev Vaidman. Properties of a quantum system during the time
interval between two measurements. Phys. Rev. A, 41:11–20, Jan 1990.
[2] Yakir Aharonov and Lev Vaidman. The two-state vector formalism: An updated review.
In J.G. Muga, R.Sala Mayato, and .L. Egusquiza, editors, Time in Quantum Mechanics,
volume 734 of Lecture Notes in Physics, pages 399–447. Springer Berlin Heidelberg, 2008.
[3] Walther Gerlach and Otto Stern. Der experimentelle nachweis der richtungsquantelung
im magnetfeld. Zeitschrift fr Physik, 9(1):349–352, 1922.
[4] Yakir Aharonov, David Z. Albert, and Lev Vaidman. How the result of a measurement of
a component of the spin of a spin- 1/2 particle can turn out to be 100. Phys. Rev. Lett.,
60:1351–1354, Apr 1988.
[5] I. M. Duck, P. M. Stevenson, and E. C. G. Sudarshan. The sense in which a ”weak
measurement” of a spin-½ particle’s spin component yields a value 100. Phys. Rev. D,
40:2112–2117, Sep 1989.
[6] Justin Dressel, Mehul Malik, Filippo M. Miatto, Andrew N. Jordan, and Robert W. Boyd.
Colloquium : Understanding quantum weak values: Basics and applications. Rev. Mod.
Phys., 86:307–316, Mar 2014.
[7] Xuanmin Zhu, Yuxiang Zhang, Shengshi Pang, Chang Qiao, Quanhui Liu, and Shengjun
Wu. Quantum measurements with preselection and postselection. Phys. Rev. A,
84:052111, Nov 2011.
[8] K.J. Resch, J.S. Lundeen, and A.M. Steinberg. Experimental realization of the quantum
box problem. Physics Letters A, 324(23):125 – 131, 2004.
108
Bibliography 109
[9] Sacha Kocsis, Boris Braverman, Sylvain Ravets, Martin J. Stevens, Richard P. Mirin,
L. Krister Shalm, and Aephraim M. Steinberg. Observing the average trajectories of
single photons in a two-slit interferometer. Science, 332(6034):1170–1173, 2011.
[10] Lee A. Rozema, Ardavan Darabi, Dylan H. Mahler, Alex Hayat, Yasaman Soudagar, and
Aephraim M. Steinberg. Violation of heisenberg’s measurement-disturbance relationship
by weak measurements. Phys. Rev. Lett., 109:100404, Sep 2012.
[11] J. S. Lundeen, B. Sutherland, A. Patel, C. Stewart, and C. Bamber. Direct measurement
of the quantum wavefunction. Nature, 474:188–191, 2011.
[12] J. S. Lundeen and A. M. Steinberg. Experimental joint weak measurement on a photon
pair as a probe of hardy’s paradox. Phys. Rev. Lett., 102:020404, Jan 2009.
[13] J. P. Groen, D. Riste, L. Tornberg, J. Cramer, P. C. de Groot, T. Picot, G. Johans-
son, and L. DiCarlo. Partial-measurement backaction and nonclassical weak values in a
superconducting circuit. Phys. Rev. Lett., 111:090506, Aug 2013.
[14] K. Yokota, T. Yamamoto, M. Koashi, and N. Imoto. Direct observation of Hardy’s
paradox by joint weak measurement with an entangled photon pair. New Journal of
Physics, 11(3):033011, March 2009.
[15] Justin Dressel, Kevin Lyons, Andrew N. Jordan, Trent M. Graham, and Paul G. Kwiat.
Strengthening weak-value amplification with recycled photons. Phys. Rev. A, 88:023821,
Aug 2013.
[16] Xiao-Ye Xu, Yaron Kedem, Kai Sun, Lev Vaidman, Chuan-Feng Li, and Guang-Can Guo.
Phase estimation with weak measurement using a white light source. Phys. Rev. Lett.,
111:033604, Jul 2013.
[17] G. G. Gillett, R. B. Dalton, B. P. Lanyon, M. P. Almeida, M. Barbieri, G. J. Pryde, J. L.
O’Brien, K. J. Resch, S. D. Bartlett, and A. G. White. Experimental feedback control of
quantum systems using weak measurements. Phys. Rev. Lett., 104:080503, Feb 2010.
[18] David J. Starling, P. Ben Dixon, Andrew N. Jordan, and John C. Howell. Precision
frequency measurements with interferometric weak values. Phys. Rev. A, 82:063822, Dec
2010.
Bibliography 110
[19] Luis Jose Salazar-Serrano, David A. Guzman, Alejandra Valencia, and Juan P. Torres.
Demonstration of a highly-sensitive tunable beam displacer with no use of beam deflection
based on the concept of weak value amplification. Opt. Express, 23(8):10097–10102, Apr
2015.
[20] Sumei Huang and Girish S. Agarwal. Weak value amplification of atomic cat states.
arXiv:1501.02359, 2015.
[21] Kevin Lyons, Justin Dressel, Andrew N. Jordan, John C. Howell, and Paul G. Kwiat.
Power-recycled weak-value-based metrology. Phys. Rev. Lett., 114:170801, Apr 2015.
[22] Onur Hosten and Paul Kwiat. Observation of the spin hall effect of light via weak mea-
surements. Science, 319(5864):787–790, 2008.
[23] P. Ben Dixon, David J. Starling, Andrew N. Jordan, and John C. Howell. Ultrasensitive
beam deflection measurement via interferometric weak value amplification. Phys. Rev.
Lett., 102:173601, Apr 2009.
[24] Marcel Pfeifer and Peer Fischer. Weak value amplified optical activity measurements.
Opt. Express, 19(17):16508–16517, Aug 2011.
[25] Itay Shomroni, Orel Bechler, Serge Rosenblum, and Barak Dayan. Demonstration of weak
measurement based on atomic spontaneous emission. Phys. Rev. Lett., 111:023604, Jul
2013.
[26] G. Jayaswal, G. Mistura, and M. Merano. Observation of the imbert – fedorov effect via
weak value amplification. Opt. Lett., 39(8):2266–2269, Apr 2014.
[27] Omar S. Magana Loaiza, Mohammad Mirhosseini, Brandon Rodenburg, and Robert W.
Boyd. Amplification of angular rotations using weak measurements. Phys. Rev. Lett.,
112:200401, May 2014.
[28] Luis Jose Salazar-Serrano, Davide Janner, Nicolas Brunner, Valerio Pruneri, and Juan P.
Torres. Measurement of sub-pulse-width temporal delays via spectral interference induced
by weak value amplification. Phys. Rev. A, 89:012126, Jan 2014.
[29] G. J. Pryde, J. L. O’Brien, A. G. White, T. C. Ralph, and H. M. Wiseman. Measurement
of quantum weak values of photon polarization. Phys. Rev. Lett., 94:220405, Jun 2005.
Bibliography 111
[30] M. E. Goggin, M. P. Almeida, M. Barbieri, B. P. Lanyon, J. L. OBrien, A. G. White, and
G. J. Pryde. Violation of the leggettgarg inequality with weak measurements of photons.
Proceedings of the National Academy of Sciences, 108(4):1256–1261, 2011.
[31] Andrew N. Jordan, Julian Martınez-Rincon, and John C. Howell. Technical advantages
for weak-value amplification: When less is more. Phys. Rev. X, 4:011031, Mar 2014.
[32] G. Bie Alves, B. M. Escher, R. L. de Matos Filho, N. Zagury, and L. Davidovich. Weak-
value amplification as an optimal metrological protocol. Phys. Rev. A, 91:062107, Jun
2015.
[33] David J. Starling, P. Ben Dixon, Andrew N. Jordan, and John C. Howell. Optimizing the
signal-to-noise ratio of a beam-deflection measurement with interferometric weak values.
Phys. Rev. A, 80:041803, Oct 2009.
[34] Yuki Susa, Yutaka Shikano, and Akio Hosoya. Optimal probe wave function of weak-value
amplification. Phys. Rev. A, 85:052110, May 2012.
[35] Shengshi Pang and Todd A. Brun. Improving the precision of weak measurements by
postselection. arXiv:1409.2567, 2014.
[36] J. Calsamiglia, B. Gendra, R. Munoz-Tapia, and E. Bagan. Probabilistic metrology
defeats ultimate deterministic bound. arXiv:1407.6910, 2014.
[37] Gerardo I. Viza, Julin Martnez-Rincn, Gabriel B. Alves, Andrew N. Jordan, and John C.
Howell. Experimentally quantifying the advantages of weak-values-based metrology.
arXiv:1410.8461, 2014.
[38] Gerardo I. Viza, Julian Martınez-Rincon, Gregory A. Howland, Hadas Frostig, Itay Shom-
roni, Barak Dayan, and John C. Howell. Weak-values technique for velocity measure-
ments. Opt. Lett., 38(16):2949–2952, Aug 2013.
[39] Christopher Ferrie and Joshua Combes. Weak value amplification is suboptimal for esti-
mation and detection. Phys. Rev. Lett., 112:040406, Jan 2014.
[40] Joshua Combes, Christopher Ferrie, Zhang Jiang, and Carlton M. Caves. Quantum limits
on postselected, probabilistic quantum metrology. Phys. Rev. A, 89:052117, May 2014.
Bibliography 112
[41] George C. Knee, G. Andrew D. Briggs, Simon C. Benjamin, and Erik M. Gauger. Quan-
tum sensors based on weak-value amplification cannot overcome decoherence. Phys. Rev.
A, 87:012115, Jan 2013.
[42] Saki Tanaka and Naoki Yamamoto. Information amplification via postselection: A
parameter-estimation perspective. Phys. Rev. A, 88:042116, Oct 2013.
[43] Lijian Zhang, Animesh Datta, and Ian A. Walmsley. Precision metrology using weak
measurements. Phys. Rev. Lett., 114:210801, May 2015.
[44] Amir Feizpour, Xingxing Xing, and Aephraim M. Steinberg. Amplifying single-photon
nonlinearity using weak measurements. Phys. Rev. Lett., 107:133603, Sep 2011.
[45] Nicolas Brunner and Christoph Simon. Measuring small longitudinal phase shifts: Weak
measurements or standard interferometry? Phys. Rev. Lett., 105:010405, Jul 2010.
[46] Y. Kedem. Using technical noise to increase the signal-to-noise ratio of measurements via
imaginary weak values. Phys. Rev. A, 85:060102, Jun 2012.
[47] Shengshi Pang and Todd A. Brun. Suppressing technical noise in weak measurements by
entanglement. Phys. Rev. A, 92:012120, Jul 2015.
[48] George C. Knee and Erik M. Gauger. When amplification with weak values fails to
suppress technical noise. Phys. Rev. X, 4:011032, Mar 2014.
[49] Christopher Ferrie and Joshua Combes. How the result of a single coin toss can turn out
to be 100 heads. Phys. Rev. Lett., 113:120404, Sep 2014.
[50] A. Feizpour, M. Hallaji, G. Dmochowski, and Steinberg A.M. Observation of the nonlinear
phase shift due to single post-selected photons. Natre Physics, 11:905–909, 2015.
[51] N. Imoto, H. A. Haus, and Y. Yamamoto. Quantum nondemolition measurement of the
photon number via the optical kerr effect. Phys. Rev. A, 32:2287–2292, Oct 1985.
[52] Amir Feizpour. Nonlinear Optics at the Single Photon Level. PhD thesis, University of
Toronto, Toronto, 2015.
[53] J. W. Noh, A. Fougeres, and L. Mandel. Measurement of the quantum phase by photon
counting. Phys. Rev. Lett., 67:1426–1429, Sep 1991.
Bibliography 113
[54] J. W. Noh, A. Fougeres, and L. Mandel. Operational approach to the phase of a quantum
field. Phys. Rev. A, 45:424–442, Jan 1992.
[55] Danna Rosenberg, Adriana E. Lita, Aaron J. Miller, and Sae Woo Nam. Noise-free high-
efficiency photon-number-resolving detectors. Phys. Rev. A, 71:061803, Jun 2005.
[56] Yakir Aharonov, Peter G. Bergmann, and Joel L. Lebowitz. Time symmetry in the
quantum process of measurement. Phys. Rev., 134:B1410–B1416, Jun 1964.
[57] Jonathan A Gross, Ninnat Dangniam, Christopher Ferrie, and Carlton M Caves. On the
novelty, efficacy, and significance of weak measurements for quantum tomography. arXiv
preprint arXiv:1506.08892, 2015.
[58] Daniel A. Steck. http://steck.us/alkalidata/.
[59] Alan Stummer. Schlosser laser lock. http://www.physics.utoronto.ca/ astum-
mer/pub/mirror/projects/archives/.
[60] Huy Diep Do, Geol Moon, and Heung-Ryoul Noh. Polarization spectroscopy of rubidium
atoms: Theory and experiment. Phys. Rev. A, 77:032513, Mar 2008.
[61] M. L. Harris, C. S. Adams, S. L. Cornish, I. C. McLeod, E. Tarleton, and I. G. Hughes.
Polarization spectroscopy in rubidium and cesium. Phys. Rev. A, 73:062509, Jun 2006.
[62] Yong-Fan Chen, Yu-Chen Liu, Zen-Hsiang Tsai, Shih-Hao Wang, and Ite A. Yu. Beat-
note interferometer for direct phase measurement of photonic information. Phys. Rev. A,
72:033812, Sep 2005.
[63] Michael Fleischhauer, Atac Imamoglu, and Jonathan P. Marangos. Electromagnetically
induced transparency: Optics in coherent media. Rev. Mod. Phys., 77:633–673, Jul 2005.
[64] K.-J. Boller, A. Imamoglu, and S. E. Harris. Observation of electromagnetically induced
transparency. Phys. Rev. Lett., 66:2593–2596, May 1991.
[65] Danielle A. Braje, Vlatko Balic, G. Y. Yin, and S. E. Harris. Low-light-level nonlinear
optics with slow light. Phys. Rev. A, 68:041801, Oct 2003.
[66] Hoonsoo Kang and Yifu Zhu. Observation of large kerr nonlinearity at low light intensities.
Phys. Rev. Lett., 91:093601, Aug 2003.
Bibliography 114
[67] Hsiang-Yu Lo, Yen-Chun Chen, Po-Ching Su, Hao-Chung Chen, Jun-Xian Chen, Ying-
Cheng Chen, Ite A. Yu, and Yong-Fan Chen. Electromagnetically-induced-transparency-
based cross-phase-modulation at attojoule levels. Phys. Rev. A, 83:041804, Apr 2011.
[68] M. D. Lukin and A. Imamoglu. Nonlinear optics and quantum entanglement of ultraslow
single photons. Phys. Rev. Lett., 84:1419–1422, Feb 2000.
[69] N. Matsuda, R. Shimizu, Y. Mitsumori, H. Kosaka, and K. Edamatsu. Observation
of optical-fibre Kerr nonlinearity at the single-photon level. Nature Photonics, 3:95–98,
February 2009.
[70] H. Schmidt and A. Imamoglu. Giant Kerr nonlinearities obtained by electromagnetically
induced transparency. Optics Letters, 21:1936–1938, December 1996.
[71] Andreas Reiserer, Stephan Ritter, and Gerhard Rempe. Nondestructive detection of an
optical photon. Science, 342(6164):1349–1351, 2013.
[72] Alan Stummer. Phase shifter. http://www.physics.utoronto.ca/ astum-
mer/pub/mirror/projects/archives/.
[73] D. T. Stack, P. J. Lee, and Q. Quraishi. Simple and efficient absorption filter for single
photons from a cold atom quantum memory. Optics Express, 23:6822, March 2015.
[74] Xingxing Xing. A Quantum Light Source for Light-Matter Interaction. PhD thesis,
University of Toronto, Toronto, 2013.
[75] Pavel Kolchin, Shengwang Du, Chinmay Belthangady, G. Y. Yin, and S. E. Harris. Gen-
eration of narrow-bandwidth paired photons: Use of a single driving laser. Phys. Rev.
Lett., 97:113602, Sep 2006.
[76] Andrew John MacRae. An Atomic Source of Quantum Light. PhD thesis, University of
Calgary, Calgary, 2012.
[77] Yanxu Han, Jintao Xiao, Yonghong Liu, Chunhong Zhang, Hai Wang, Min Xiao, and
Kunchi Peng. Interacting dark states with enhanced nonlinearity in an ideal four-level
tripod atomic system. Phys. Rev. A, 77:023824, Feb 2008.
Bibliography 115
[78] Shujing Li, Xudong Yang, Xuemin Cao, Chunhong Zhang, Changde Xie, and Hai Wang.
Enhanced cross-phase modulation based on a double electromagnetically induced trans-
parency in a four-level tripod atomic system. Phys. Rev. Lett., 101:073602, Aug 2008.
[79] Bor-Wen Shiau, Meng-Chang Wu, Chi-Ching Lin, and Ying-Cheng Chen. Low-light-level
cross-phase modulation with double slow light pulses. Phys. Rev. Lett., 106:193006, May
2011.
[80] Wolfgang Ketterle, Kendall B. Davis, Michael A. Joffe, Alex Martin, and David E.
Pritchard. High densities of cold atoms in a dark spontaneous-force optical trap. Phys.
Rev. Lett., 70:2253–2256, Apr 1993.
[81] Y.-W. Lin, H.-C. Chou, P. P. Dwivedi, Y.-C. Chen, and I. A. Yu. Using a pair of
rectangular coils in the MOT for the production of cold atom clouds with large optical
density. Optics Express, 16:3753, March 2008.
[82] W. Petrich, M. H. Anderson, J. R. Ensher, and E. A. Cornell. Behavior of atoms in a
compressed magneto-optical trap. Journal of the Optical Society of America B Optical
Physics, 11:1332–1335, August 1994.
[83] M. T. DePue, S. Lukman Winoto, D. J. Han, and D. S. Weiss. Transient compression of
a MOT and high intensity fluorescent imaging of optically thick clouds of atoms. Optics
Communications, 180:73–79, June 2000.
[84] Samansa Maneshi, Chao Zhuang, Christopher R. Paul, Luciano S. Cruz, and Aephraim M.
Steinberg. Coherence freeze in an optical lattice investigated via pump-probe spec-
troscopy. Phys. Rev. Lett., 105:193001, Nov 2010.
[85] Chao Zhuang, Christopher R. Paul, Xiaoxian Liu, Samansa Maneshi, Luciano S. Cruz,
and Aephraim M. Steinberg. Coherent control of population transfer between vibra-
tional states in an optical lattice via two-path quantum interference. Phys. Rev. Lett.,
111:233002, Dec 2013.
[86] J. F. Kanem, S. Maneshi, M. Partlow, M. Spanner, and A. M. Steinberg. Observation
of high-order quantum resonances in the kicked rotor. Phys. Rev. Lett., 98:083004, Feb
2007.
Bibliography 116
[87] Noah Linden, Sandu Popescu, and Paul Skrzypczyk. How small can thermal machines
be? the smallest possible refrigerator. Phys. Rev. Lett., 105:130401, Sep 2010.
[88] Noah Linden, Sandu Popescu, and Paul Skrzypczyk. The smallest possible heat engines.
arXiv:1010.6029, 2010.
[89] Wenlan Chen, Kristin M. Beck, Robert Bcker, Michael Gullans, Mikhail D. Lukin, Haruka
Tanji-Suzuki, and Vladan Vuletic. All-optical switch and transistor gated by one stored
photon. Science, 341(6147):768–770, 2013.
[90] M. Bajcsy, S. Hofferberth, V. Balic, T. Peyronel, M. Hafezi, A. S. Zibrov, V. Vuletic, and
M. D. Lukin. Efficient all-optical switching using slow light within a hollow fiber. Phys.
Rev. Lett., 102:203902, May 2009.
[91] Ephraim Shahmoon, Gershon Kurizki, Michael Fleischhauer, and David Petrosyan.
Strongly interacting photons in hollow-core waveguides. Phys. Rev. A, 83:033806, Mar
2011.
[92] V. Venkataraman, K. Saha, and A. L. Gaeta. Phase modulation at the few-photon level
for weak-nonlinearity-based quantum computing. Nature Photonics, 7:138–141, February
2013.
[93] S. T. Dawkins, R. Mitsch, D. Reitz, E. Vetsch, and A. Rauschenbeutel. Dispersive optical
interface based on nanofiber-trapped atoms. Phys. Rev. Lett., 107:243601, Dec 2011.
[94] Barak Dayan, A. S. Parkins, Takao Aoki, E. P. Ostby, K. J. Vahala, and H. J. Kimble.
A photon turnstile dynamically regulated by one atom. Science, 319(5866):1062–1065,
2008.
[95] Danny O’Shea, Christian Junge, Jurgen Volz, and Arno Rauschenbeutel. Fiber-optical
switch controlled by a single atom. Phys. Rev. Lett., 111:193601, Nov 2013.
[96] Simon Baur, Daniel Tiarks, Gerhard Rempe, and Stephan Durr. Single-photon switch
based on rydberg blockade. Phys. Rev. Lett., 112:073901, Feb 2014.
[97] D. E. Chang, V. Vuletic, and M. D. Lukin. Quantum nonlinear optics – photon by photon.
Nature Photonics, 8:685–694, September 2014.
Bibliography 117
[98] Ofer Firstenberg, Thibault Peyronel, Qi-Yu Liang, Alexey V. Gorshkov, Mikhail D. Lukin,
and Vladan Vuletic. Attractive photons in a quantum nonlinear medium. Nature, 502:71–
75, 2013.
[99] Thibault Peyronel, Ofer Firstenberg, Qi-Yu Liang, Sebastian Hofferberth, Alexey V. Gor-
shkov, Thomas Pohl, Mikhail D. Lukin, and Vladan Vuletic. Quantum nonlinear optics
with single photons enabled by strongly interacting atoms. Nature, 488:56–60, 2012.
[100] W. J. Munro, K. Nemoto, and T. P. Spiller. Weak nonlinearities: a new route to optical
quantum computation. New Journal of Physics, 7:137, May 2005.
[101] Kae Nemoto and W. J. Munro. Nearly deterministic linear optical controlled-not gate.
Phys. Rev. Lett., 93:250502, Dec 2004.
[102] Bing He, Andrew MacRae, Yang Han, A. I. Lvovsky, and Christoph Simon. Transverse
multimode effects on the performance of photon-photon gates. Phys. Rev. A, 83:022312,
Feb 2011.
[103] Jeffrey H. Shapiro. Single-photon kerr nonlinearities do not help quantum computation.
Phys. Rev. A, 73:062305, Jun 2006.