By Dinesh Gannerlla

Embed Size (px)

Citation preview

  • 7/28/2019 By Dinesh Gannerlla

    1/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    HOSPITAL TRAINING REPORT

    To be submitted

    ToCARE HOSPITALS

    For partial fulfillment of the requirement for theAward of the Degree

    OfBachelor of Technology

    In

    Bio-medical Engineering

    SUBMITED BY

    B.S SAI SRIDHAR -07211A1118

    Y.MURALI MOHAN -07211A1112

    PULIPATI KALYAN -07211A1108

    G.DINESH -07211A1105A N M SWAMY -07211A1125

    B.HARSHA PRIYA -07211A1106

    K.AARTHI -07211A1101

    CARE HOSPITALS Page 1

  • 7/28/2019 By Dinesh Gannerlla

    2/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    BIOMEDICAL

    INTRODUCTION:

    During the last two decades, there have been tremendous advances in the field of

    biomedical engineering which have been applied to the medical sciences both for diagnosis and

    treatment. As a result of the interaction between these two specialties, a new discipline has

    emerged as bio-medical engineering. (Jacob john-1974)

    It is unique in the sense that this speciality emerged out of a combination of engineering,

    sciences, medicine and biology. (Staewen Williams-1984)

    Definition:

    Bio medical engineering is defined as that branch of applied science which is

    contained with solving and understanding the problems in biology or medicine using principles

    methods and approach drawn from engineering science and technology (Richard johns-1975)

    Historical aspect of Bio-Medical Engineering:

    The field of Medical instrumentation is by no means new. Many instruments were

    developed as early as the nineteenth century, for example the ECG was first used by cinthoren

    at the end of the century. Progress was slow until after World War II when a surplus of

    electronic equipment became available. At that time many technicians and engineers both

    within industry and on their own started experiments with and modify existing equipment for

    medical use. This progress occurred during the 1950's and the results were after disappointing

    for physiological parameters and not measured in the same way as physical parameters. They

    also had severe communication problem with the medical profession. During the next decade

    CARE HOSPITALS Page 2

  • 7/28/2019 By Dinesh Gannerlla

    3/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITmany instrument manufactures philosophy was changed equipment analysis and design were

    applied directly to medical problems. A large measure of help was provided by the US govt. /in

    particular by NASA. The mercury, Gemini and Apollo program needed accurate physiological

    monitoring for astronauts, consequently much research and development and money went into

    this area.

    Some of the concepts of patient monitoring presently used in hospitals through out the

    world evolved from the base of astronauts monitoring.

    The efficiency of modern medicine not only depends upon the clinical acumen of the

    physician but also in optimal utilization of available sophisticated bio medical equipment, and

    such equipments need careful attention as well as maintenance without any breakdown or

    if/there is a breakdown it should be attended within the shortest possible time. Hence the needfor bio medical engineering depart/arises which has to take up the responsibility of maintenance

    equipment.

    Since the new technologies are rapidly creeping into the field of medicine. It is essential to

    have a Bio Medical Engineering unit attached to a modern hospital which can suggest/ can take

    care of planning, organizing, installation, maintenance newly developing biomedical equipment

    for rapid diagnosis and treatment of a number of ailments. Equipments from very large range

    such as the X-RAY, CT-SCAN, MRI, ULTRASOUND, ECHO, IABP, HEART-LUNG

    MACHINE, HOLTER, EEG, EMG, dialysis, PACEMAKER, ANESTHESIA VENTILATORS

    DIATHERMIES. Monitors, ventilators, ABG, etc. have to be regularly checked and maintained

    by Bio Medical Engineering department. Preventive maintenance is necessary for the proper

    operating of equipment but other factors help add to the effectiveness of a good program.

    Apart from the BIOMEDICAL department has the greatest responsibility of maintaining

    major equipment because of service as well as emergency services, which inturn affect the

    patient care. BIOMEDICAL engg dept should also maintain the service record for the major

    equipment to meet the urgent requirement and keep the spare parts in stock for ready usage. The

    purpose of biomedical engg dept in hospital is to guiding/advising the administration and

    CARE HOSPITALS Page 3

  • 7/28/2019 By Dinesh Gannerlla

    4/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITphysicians about the utility, future developments in the equipment to purchased or already

    purchased or conversion of old models to the recent once, extension of the usage of equipment

    for different purposes.

    RESPONSIBILITIES OF A BIO-MEDICAL ENGINEER:

    Attend to emergency breakdown of the essential equipment

    To monitor the available equipment on regular base.

    To keep stock of the spare parts.

    Introducing/briefing the hospital or institute to new technology and new equipment and

    how best to utilize the same.

    In short it is pre purchase evaluation, planning, acceptance testing, inventory control,

    training and maintenance

    Preparing and performing the preventive maintenance schedule.

    PREVENTIVE MAINTAINANCE: PM is defined in the ASHE manual is to clean

    adjust. Check for wear and perhaps replace components that might cause total breakdown

    or serious functional impairment of the equipment before the next scheduled inspection

    (ASHE) PM will help eliminate hazards before they develop.

    The PMprocedure is primarily a performance test to ensure that the equipment

    is operating properly and is calibrated. However problems do occur because of deterioration of

    equipment caused by normal use and again can be detected prior to their causing a malfunction.

    Proper cleaning, Lubrication and repair of replacement of defective parts prior to a serious

    malfunction will prolong the useful life of the equipments.

    CARE HOSPITALS Page 4

  • 7/28/2019 By Dinesh Gannerlla

    5/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    CONTENTS:

    1. NEPHROLOGY: DIALYSIS

    2. NEUROLOGY: EEG

    3. CARDIOLOGY: HEART - LUNG MACHINE

    4. GASTROENTEROLOGY: ENDOSCOPES

    5.RADIOLOGY: MRI,CT,X-RAY,ULTRASOUND

    6.CSSD

    7. OXYGEN PLANT

    CARE HOSPITALS Page 5

  • 7/28/2019 By Dinesh Gannerlla

    6/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    NEPHROLOGY

    Dialysis:

    In medicine, dialysis (from Greek "dialusis", meaning dissolution, "dia", meaning through, and

    "lusis", meaning loosening) is primarily used to provide an artificial replacement for lost kidney

    function (renal replacement therapy) due to renal failure. Dialysis may be used for very sick

    patients who have suddenly but temporarily, lost their kidney function (acute renal failure) or

    for quite stable patients who have permanently lost their kidney function (stage 5 chronic

    kidney disease). When healthy, the kidneys maintain the body's internal equilibrium of water

    and minerals (sodium, potassium, chloride, calcium, phosphorus, magnesium, sulfate) and the

    kidneys remove from the blood the daily metabolic load of fixed hydrogen ions. The kidneys

    also function as a part of the endocrine system producing erythropoietin and 1,25-

    dihydroxycholecalciferol (calcitriol). Dialysis treatments imperfectly replace some of thesefunctions through the diffusion (waste removal) and convection (fluid removal). Dialysis is an

    imperfect treatment to replace kidney function because it does not correct the endocrine

    functions of the kidney

    Principle:

    Dialysis works on the principles of the diffusion and osmosis of solutes and fluid across a semi-

    permeable membrane. Blood flows by one side of a semi-permeable membrane, and a dialysate

    or fluid flows by the opposite side. Smaller solutes and fluid pass through the membrane. The

    blood flows in one direction and the dialysate flows in the opposite. The concentrations of

    CARE HOSPITALS Page 6

  • 7/28/2019 By Dinesh Gannerlla

    7/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITundesired solutes (for example potassium, calcium, and urea) are high in the blood, but low or

    absent in the dialysis solution and constant replacement of the dialysate ensures that the

    concentration of undesired solutes is kept low on this side of the membrane. The dialysis

    solution has levels of minerals like potassium and calcium that are similar to their natural

    concentration in healthy blood. For another solute, bicarbonate, dialysis solution level is set at a

    slightly higher level than in normal blood, to encourage diffusion of bicarbonate into the blood,

    to neutralise the metabolic acidosis that is often present in these patients.

    Types:

    There are two primary types of dialysis, hemodialysis and peritoneal dialysisHemodialysis:

    In hemodialysis,

    the patient's blood

    is pumped through

    the bloodcompartment of a

    dialyzer, exposing

    it to a

    semipermeable

    membrane. The cleansed blood is then returned via the circuit back to the body. Ultrafiltration

    occurs by increasing the hydrostatic pressure across the dialyzer membrane. This usually is

    done by applying a negative pressure to the dialysate compartment of the dialyzer. This

    pressure gradient causes water and dissolved solutes to move from blood to dialysate, and

    allows removal of several litres of excess fluid during a typical 3 to 5 hour treatment. In the US,

    hemodialysis treatments are typically given in a dialysis center three times per week (due in the

    CARE HOSPITALS Page 7

  • 7/28/2019 By Dinesh Gannerlla

    8/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITUS to Medicare reimbursement rules), however, as of 2007 over 2,000 people in the US are

    dialyzing at home more frequently for various treatment lengths. Studies have demonstrated the

    clinical benefits of dialyzing 5 to 7 times a week, for 6 to 8 hours. These frequent long

    treatments are often done at home, while sleeping but home dialysis is a flexible modality and

    schedules can be changed day to day, week to week. In general, studies have shown that both

    increased treatment length and frequency are clinically beneficial

    Peritoneal dialysis:

    In peritoneal dialysis, a sterile solution containing minerals and glucose is run through a tube

    into the peritoneal cavity, the abdominal body cavity around the intestine, where the peritoneal

    membrane acts as a semipermeable membrane. The dialysate is left there for a period of time to

    absorb waste products, and then it is drained out through the tube and discarded. This cycle or

    "exchange" is normally repeated 4-5 times during the day, (sometimes more often overnight

    with an automated system). Ultrafiltration occurs via osmosis; the dialysis solution used

    contains a high concentration of glucose, and the resulting osmotic pressure causes fluid to

    move from the blood into the dialysate. As a result, more fluid is drained than was instilled.

    CARE HOSPITALS Page 8

  • 7/28/2019 By Dinesh Gannerlla

    9/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITPeritoneal dialysis is less efficient than hemodialysis, but because it is carried out for a longer

    period of time the net effect in terms of removal of waste products and of salt and water are

    similar to hemodialysis. Peritoneal dialysis is carried out at home by the patient and it requires

    motivation. Although support is helpful, it is not essential. It does free patients from the routine

    of having to go to a dialysis clinic on a fixed schedule multiple times per week, and it can be

    done while travelling with a minimum of specialized equipment.

    Hemofiltration: Hemofiltration is a similar treatment to hemodialysis, but it makes use of a

    different principle. The blood is pumped through a dialyzer or "hemofilter" as in dialysis, but no

    dialysate is used. A pressure gradient is applied; as a result, water moves across the very

    permeable membrane rapidly, facilitating the transport of dissolved substances, importantly

    ones with large molecular weights, which are cleared less well by hemodialysis. Salts and waterlost from the blood during this process are replaced with a "substitution fluid" that is infused

    into the extracorporeal circuit during the treatment. Hemodiafiltration is a term used to describe

    several methods of combining hemodialysis and hemofiltration in one process

    What is dialysis access?

    In the context of this article, dialysis access is an

    entranceway into your bloodstream that lies completely

    beneath your skin and is easy to use. The access is usually in

    your arm, but sometimes in the leg, and allows blood to be

    removed and returned quickly, efficiently, and safely during

    dialysis or, less commonly, for other procedures requiring

    frequent access to your circulation. Dialysis, also called

    hemodialysis, is the most common treatment for kidney

    failure. A dialysis machine is an artificial kidney

    designed to remove impurities from your blood. During dialysis, physicians use the dialysis

    access to remove a portion of your blood to circulate it through the dialysis machine so it can

    CARE HOSPITALS Page 9

  • 7/28/2019 By Dinesh Gannerlla

    10/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITremove impurities and regulate fluid and chemical balances. The purified blood is then returned

    to you, again through the dialysis access.

    Creating the access portal is a minor surgical procedure. There are two types of portals placed

    completely under the skin:

    Fistula, which your vascular surgeon constructs by joining an artery to a vein.

    Graft, which is a man-made tube, consisting of a plastic or other material, that your

    vascular surgeon inserts under the skin to connect an artery to a vein.

    For both fistulas and grafts, the connection between your

    artery and vein increases blood flow through the vein. In

    response, your vein stretches and becomes strengthened. This

    allows an even greater amount of blood to pass through the

    vein and allows your dialysis to proceed efficiently. In the

    weeks after surgery, the fistula begins to mature. The vein

    increases in size and may look like a cord under your skin.

    The whole process of maturation, which is a beneficial feature

    that permits the blood flow to increase in the fistula, typically

    takes 3 to 6 months. Some fistulas may take as long as a year

    or more to develop fully, but this is unusual. Once matured, a fistula should be large and strong

    enough for dialysis technicians and nurses to insert the large dialysis needles easily. If it fails to

    mature in a reasonable period of time, however, you may need another fistula. You can usually

    begin using your graft in 2 to 6 weeks, when it is healed sufficiently. Usually fistulas are

    preferred to grafts, however, because fistulas are constructed using your own tissue, which is

    more durable and resistant to infection than are grafts. However, if your vein is blocked or too

    small to use, the graft provides a good alternative.

    CARE HOSPITALS Page 10

  • 7/28/2019 By Dinesh Gannerlla

    11/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    How do you prepare?

    Before choosing the access site, your surgeon may ask you if you have a history or symptoms of

    arm or leg artery disease. Hardening of the arteries, which reduces blood flow to your arms

    or legs, often can cause these conditions. Your vascular surgeon will not place a dialysis

    access site in an area of the body with reduced circulation because the blood flow will be

    insufficient. For this reason, your surgeon usually places dialysis access sites in the arms

    rather than in the legs because atherosclerosis is more common in the legs. Your vascular

    surgeon may order a blood flow test in your arms and legs, such as an ultrasound exam, or

    an x ray, such as a venogram, to determine whether your veins are large enough to use for

    a fistula. Sometimes a non-invasive pulse volume recording test is used to evaluate the

    flow in your arteries if this issue is a concern to your surgeon. Your vascular surgeon will

    give you the necessary instructions you need to follow before the surgery, such as fasting.

    Usually, your physician will ask you not to eat or drink anything 8 hours before your

    procedure. Your physician will discuss with you whether to reduce or stop any

    medications that might increase your risk of bleeding or other complications.

    Are you eligible for dialysis access?

    If you have chronic kidney failure and need long-term hemodialysis, you may require dialysis

    access. You may not be a good candidate for a fistula if your veins are too small or are

    scarred from frequent placement of intravenous catheters (thin, flexible tubes inserted into

    veins to deliver medicine) or needles to draw blood. In that event, you may be eligible for

    a graft access procedure. You also may not be a good candidate for a fistula if your arteries

    are severely blocked, although they might be repairable if necessary. Your vascular

    CARE HOSPITALS Page 11

  • 7/28/2019 By Dinesh Gannerlla

    12/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITsurgeon will probably be reluctant to use a graft if you have an ongoing infection since the

    graft itself might become infected. If this happens, the infected graft might need to be

    removed in order to clear up the infection.

    What happens during dialysis access?

    Typically you will have the procedure on an outpatient basis. Most often, you will first be

    sedated and then your surgeon will numb the area where the fistula or graft will go. In

    some cases, your anesthesiologist may give you supplemental sedation or put you to sleep.

    Depending upon the quality of your artery and vein, your surgeon will try to construct the

    fistula with one incision using the forearm of the arm that you do not use as frequently. For

    example, if you're left handed, your physician will place the fistula in your right arm, if

    possible. To perform the surgery, your physician joins a large vein under the skin to an

    artery nearby. The physician divides your vein and sews it to an opening made in the side

    of the artery. As a result, the blood flows down the arteries into the hand, as usual, and also

    some of this faster moving blood flows into the veins that lead back to your heart. The

    blood that normally traveled in your divided vein goes back to the heart through other

    veins, and there is usually plenty of blood remaining in your artery to supply your hand. Ifyou cannot receive a fistula because the vein is too small or blocked, your physician may

    construct a graft using a tube of man-made, plastic material. Less commonly, your

    physician may also choose to use a piece of a vein from your leg or a section of artery

    from a cow as alternative graft materials. Your physician sews the graft to one of your

    veins and connects the other end to an artery. Your physician may place the graft material

    straight or form a loop under the skin either in your lower arm, upper arm, or less

    commonly in your leg.

    What can you expect after dialysis access?

    CARE HOSPITALS Page 12

  • 7/28/2019 By Dinesh Gannerlla

    13/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITAfter the operation, you should initially keep the access area raised above your heart to reduce

    swelling and pain. Your surgeon may recommend an over-the-counter painkiller to relieve

    pain, if necessary. Following the suggestions below will help you keep your new access

    site working properly in the weeks after the surgery:

    Keep the incision dry for at least 2 days after the procedure and do not soak or scrub the

    incision until it has healed.

    Avoid lifting more than about 15 pounds or other activities that stress or compress the

    access area, such as digging.

    Report pain, swelling, or bleeding immediately to your physician, especially if these

    symptoms are becoming worse. Some pain or swelling is common and not worrisome if

    decreasing, but you should tell your physician if you have bleeding, drainage or a fever

    higher than 101 degrees Fahrenheit

    You may initially feel some coolness or numbness in the hand with the fistula. These

    sensations usually go away in a few weeks as your circulation compensates for the

    fistula. However, if these sensations are severe, tell your physician as soon as

    possible, because the fistula may be causing too much blood to flow away from your

    hand. You should perform exercises to grow and strengthen your fistula, after the pain

    from the surgery decreases, to make dialysis faster and easier. Your physician may

    recommend squeezing a soft object using the hand on the arm in which the fistula was

    placed. Grafts may mature more quickly than fistulas depending upon the size of the

    vein to which the graft is initially attached. They sometimes can be ready in 2 to 3

    weeks, but many physicians recommend waiting about 4 to 6 weeks before using a

    graft. Grafts have disadvantages over fistulas, however. Grafts are more likely than

    fistulas to become infected. Also, grafts usually last about 1 to 2 years, which is less

    than fistulas, which can often last up to 3 to 7 years. If you care properly for your

    graft, however, you can help it last for many years. Sometimes access portals can take

    CARE HOSPITALS Page 13

  • 7/28/2019 By Dinesh Gannerlla

    14/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITweeks or even months until they are ready for dialysis use. Until the portal is ready,

    you may have to use a catheter for dialysis.

    Are there any complications?

    Complications with dialysis access include clotting, narrowing, aneurysm formation in the

    access itself, infection, and bleeding.

    What can you do to stay healthy?

    Protecting the dialysis access is crucial for you. The following tips will help you care for a

    fistula or a graft:

    Check several times each day to make sure the access is functioning. You should be able

    to feel a vibration in the. Your physician or dialysis center staff will show you how to do

    this.

    Monitor any bleeding after dialysis. If the graft seems to bleed longer than usual from the

    needle sites, you should notify your dialysis center staff. Do not carry heavy items with the arm that has the access.

    Do not sleep on that arm.

    Do not wear any clothing or jewelry that binds that arm.

    Do not let anyone draw blood or measure blood pressure from that arm.

    Do not allow injections to be given into the fistula or graft.

    Keep the site of the fistula or graft clean.

    After dialysis, monitor the access for signs of infection, such as swelling or redness.

    CARE HOSPITALS Page 14

  • 7/28/2019 By Dinesh Gannerlla

    15/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    Electroencephalography

    Electroencephalography (EEG) is the measurement of electrical activity produced by the brain

    as recorded from electrodes placed on the scalp.

    Just as the activity in a computer can be perceived on multiple different levels, from the activity

    of individual transistors to the function of applications, so can the electrical activity of the brain

    be described on relatively small to relatively large scales. At one end are action potentials in a

    single axon or currents within a single dendrite, and at the other end is the activity measured by

    the scalp EEG.

    The data measured by the scalp EEG are used for clinical and research purposes. A technique

    similar to the EEG is intracranial EEG (icEEG), also referred to as subdural EEG (sdEEG) and

    electrocorticography (ECoG). These terms refer to the recording of activity from the surface of

    the brain (rather than the scalp). Because of the filtering characteristics of the skull and scalp,

    icEEG activity has a much higher spatial resolution than surface EEG.

    Source of EEG Activity: Scalp EEG measures summated activity of post-synaptic currents. An

    action potential in a pre-synaptic axon causes the release of neurotransmitter into the

    synapse. The neurotransmitter diffuses across the synaptic cleft and binds to receptors in apost-synaptic dendrite. The activity of many types of receptors results in a flow of ions into

    or out of the dendrite. This results in compensatory currents in the extracellular space. It

    is these extracellular currents which are responsible for the generation of EEG voltages.

    The EEG is not sensitive to axonal action potentials.

    CARE HOSPITALS Page 15

  • 7/28/2019 By Dinesh Gannerlla

    16/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITWhile it is post-synaptic potentials that generate the EEG signal, it is not possible to determine

    the activity within a single dendrite or neuron from the scalp EEG. Rather, surface EEG is

    the summation of the synchronous activity of thousands of neurons that have similar

    spatial orientation, radial to the scalp. Currents that are tangential to the scalp are not

    picked up by the EEG. The EEG therefore benefits from the parallel, radial arrangement

    of apical dendrites in the cortex. Because voltage fields fall off with the fourth power of

    the radius, activity from deep sources is more difficult to detect than currents near the

    skull.Scalp EEG activity is composed of multiple oscillations. These have different

    characteristic frequencies, spatial distributions and associations with different states of

    brain functioning (such as awake vs. asleep). These oscillations represent synchronized

    activity over a network of neurons. The neuronal network underlying some of theseoscillations are understood (such as the thalomocortical resonance underlying sleep

    spindles), while many others are not (e.g., the system that generates the posterior basic

    rhythm is not yet fully understood).

    Method: In conventional scalp EEG, the recording is obtained by placing electrodes on the

    scalp with a conductive gel or paste, usually after preparing the scalp area by light abrasion to

    reduce impedance due to dead skin cells. The technique has been advanced by the use of carbon

    nanotubes to penetrate the outer layers of the skin for improved electrical contact. The sensor is

    known as ENOBIO. however, this technique is not in common research or clinical use. Many

    systems typically use electrodes, each of which is attached to an individual wire. Some systems

    use caps or nets into which electrodes are embedded; this is particularly common when high-

    density arrays of electrodes are needed.

    Computer ElectroencephalographNeurovisor-BMM 40

    CARE HOSPITALS Page 16

    http://en.wikipedia.org/wiki/Image:Electroencephalograph_Neurovisor-BMM_40_%28close_view%29.jpg
  • 7/28/2019 By Dinesh Gannerlla

    17/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITElectrode locations and names are specified by the International 1020 system for most clinical

    and research applications (except when high-density arrays are used). This system ensures that

    the naming of electrodes is consistent across laboratories. In most clinical applications, 19

    recording electrodes (plus ground and system reference) are used. A smaller number of

    electrodes are typically used when recording EEG from neonates. Additional electrodes can be

    added to the standard set-up when a clinical or research application demands increased spatial

    resolution for a particular area of the brain. High-density arrays (typically via cap or net) can

    contain up to 256 electrodes more-or-less evenly spaced around the scalp.Each electrode is

    connected to one input of a differential amplifier (one amplifier per pair of electrodes); a

    common system reference electrode is connected to the other input of each differential

    amplifier. These amplifiers amplify the voltage between the active electrode and the reference(typically 1,000100,000 times, or 60100 dB of voltage gain). In analog EEG, the signal is

    then filtered (next paragraph), and the EEG signal is output as the deflection of pens as paper

    passes underneath. Most EEG systems these days, however, are digital, and the amplified signal

    is digitized via an analog-to-digital converter, after being passed through an anti-aliasing filter.

    Analog-to-digital sampling typically occurs at 256-512 Hz in clinical scalp EEG; sampling rates

    of up to 10 kHz are used in some research applications.The digital EEG signal is stored

    electronically and can be filtered for display. Typical settings for the high-pass filter and a low-

    pass filter are 0.5-1 Hz and 3570 Hz, respectively. The high-pass filter typically filters out

    slow artifact, such as electrogalvanic signals and movement artifact, whereas the low-pass filter

    filters out high-frequency artifacts, such as electromyographic signals. An additional notch filter

    is typically used to remove artifact caused by electrical power lines (60 Hz in the United States

    and 50 Hz in many other countries).As part of an evaluation for epilepsy surgery, it may be

    necessary to insert electrodes near the surface of the brain, under the surface of the dura mater.

    This is accomplished via burr hole or craniotomy. This is referred to variously as

    "electrocorticography (ECoG)", "intracranial EEG (I-EEG)" or "sub-dural EEG (SD-EEG)".

    Depth electrodes may also be placed into brain structures, such as the amygdala or

    CARE HOSPITALS Page 17

  • 7/28/2019 By Dinesh Gannerlla

    18/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIThippocampus, structures which are common epileptic foci and may not be "seen" clearly by

    scalp EEG. The electrocorticographic signal is processed in the same manner as digial scalp

    EEG (above), with a couple of caveats. ECoG is typically recorded at higher sampling rates

    than scalp EEG because of the requirements of Nyquist theoremthe sub-dural signal is

    composed of a higher predominance of higher frequency components. Also, many of the

    artifacts which affect scalp EEG do not impact ECoG, and therefore display filtering is often

    not needed.A typical adult human EEG signal is about 10V to 100 V in amplitude when

    measured from the scalp and is about 1020 mV when measured from subdural

    electrodes.Since an EEG voltage signal represents a difference between the voltages at two

    electrodes, the display of the EEG for the reading encephalographer may be set up in one of

    several ways. The representation of the EEG channels is referred to as a montage.

    Bipolar montage

    Each channel (i.e., waveform) represents the difference between two adjacent electrodes.

    The entire montage consists of a series of these channels. For example, the channel "Fp1-

    F3" represents the difference in voltage between the Fp1 electrode and the F3 electrode.

    The next channel in the montage, "F3-C3," represents the voltage difference between F3

    and C3, and so on through the entire array of electrodes.

    Referential montage

    Each channel represents the difference between a certain electrode and a designated

    reference electrode. There is no standard position at which this reference is always

    placed; it is, however, at a different position than the "recording" electrodes. Midline

    positions are often used because they do not amplify the signal in one hemisphere vs. the

    other. Another popular reference is "linked ears," which is a physical or mathematicalaverage of electrodes attached to both earlobes or mastoids.

    Average reference montage

    The outputs of all of the amplifiers are summed and averaged, and this averaged signal is

    used as the common reference for each channel.

    CARE HOSPITALS Page 18

  • 7/28/2019 By Dinesh Gannerlla

    19/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITLaplacian montage

    Each channel represents the difference between an electrode and a weighted average of

    the surrounding electrodes.

    Limitations: EEG has several limitations. Most important is its poor spatial resolution. EEG is

    most sensitive to a particular set of post-synaptic potentials: those which are generated in

    superficial layers of the cortex, on the crests of gyri directly abutting the skull and radial to the

    skull. Dendrites which are deeper in the cortex, inside sulci, are in midline or deep structures

    (such as the cingulate gyrus or hippocampus) or produce currents which are tangential to the

    skull have far less contribution to the EEG signal.The meninges, cerebrospinal fluid and skull

    "smear" the EEG signal, obscuring its intracranial source.It is mathematically impossible toreconstruct a unique intercranial current source for a given EEG signal, as some currents

    produce potentials that cancel each other out. This is referred to as the inverse problem.

    However, much work has been done to produce remarkably good estimates of, at least, a

    localized electric dipole that represents the recorded currents.

    Advantages: EEG has several strong sides as a tool of exploring brain activity; for

    example, its time resolution is very high (on the level of a single millisecond). Other

    methods of looking at brain activity, such as PET and MRI have time resolution between

    seconds and minutes. EEG measures the brain's electrical activity directly, while other

    methods record changes in blood flow (e.g., SPECT, MRI) or metabolic activity (e.g.,

    PET), which are indirect markers of brain electrical activity. EEG can be used

    simultaneously with MRI so that high-temporal-resolution data can be recorded at the

    same time as high-spatial-resolution data, however, since the data derived from each

    occurs over a different time course, the data sets do not necessarily represent the exact

    same brain activity. There are technical difficulties associated with combining these two

    modalities, including the need to remove RF pulse artifact and ballistocardiographic

    CARE HOSPITALS Page 19

  • 7/28/2019 By Dinesh Gannerlla

    20/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITartifact (a results from the movement of pulsed blood) from the EEG. Furthermore,

    currents can be induced in moving EEG electrode wires due to the magnetic field of the

    MRI.EEG can be recorded at the same time as MEG so that data from these

    complimentary high-time-resolution techniques can be combined.

    Normal activity: The EEG is typically described in terms of (1) rhythmic activity and (2)

    transients. The rhythmic activity is divided into bands by frequency. To some degree, these

    frequency bands are a matter of nomenclature (i.e., any rhythmic activity between 8-12 Hz

    can be described as "alpha"), but these designations arose because rhythmic activity

    within a certain frequency range was noted to have a certain distribution over the scalp or

    a certain biological significance.Most of the cerebral signal observed in the scalp EEG

    falls in the range of 1-20 Hz (activity below or above this range is likely to be artifactual,

    under standard clinical recording techniques).

    Comparison table

    Comparison of EEG bands

    TypeFrequency

    (Hz)Location Normally Pathologically

    Delta up to 3 frontally in

    adults,

    posterior in

    children;

    high

    adults slow

    wave sleep

    in babies

    subcortical lesions

    diffuse lesions

    metabolic

    encephalopathy

    hydrocephalus

    CARE HOSPITALS Page 20

  • 7/28/2019 By Dinesh Gannerlla

    21/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    amplitude

    waves

    Deep midline

    lesions.

    Theta 4 - 7 Hz

    young

    children

    drowsiness or

    arousal in

    older children

    and adults

    focal subcortical

    lesions

    metabolic

    encephalopathy

    deep midline

    disorders

    some instances of

    hydrocephalus

    Alpha 8 - 12 Hz

    Posterior

    regions of

    head, both

    sides, higher

    in amplitude

    on dominant

    side. Central

    sites (c3-c4)

    at rest.

    Closing the

    eyes and by

    relaxation.

    coma

    Beta 12 - 30 Hz both sides,

    symmetrical

    distribution,

    most evident

    frontally;

    low

    amplitude

    active, busy or

    anxious

    thinking,

    active

    concentration

    benzodiazepines

    CARE HOSPITALS Page 21

  • 7/28/2019 By Dinesh Gannerlla

    22/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    waves

    Gamma 26100

    certain

    cognitive or

    motor

    functions

    Wave patterns:

    Delta is the frequency range up to 3 Hz. It tends to be the highest in amplitude and the

    slowest waves. It is seen normally in adults in slow wave sleep. It is also seen normally in

    babies. It may occur focally with subcortical lesions and in general distribution with

    diffuse lesions, metabolic encephalopathy hydrocephalus or deep midline lesions. It is

    usually most prominent frontally in adults (e.g. FIRDA - Frontal Intermittent Rhythmic

    Delta) and posteriorly in children e.g. OIRDA - Occipital Intermittent Rhythmic Delta).

    Theta is the frequency range from 4 Hz to 7 Hz. Theta is seen normally in young

    children. It may be seen in drowsiness or arousal in older children and adults; it can also

    be seen in meditation. Excess theta for age represents abnormal activity. It can be seen as

    a focal disturbance in focal subcortical lesions; it can be seen in generalized distribution

    in diffuse disorder or metabolic encephalopathy or deep midline disorders or some

    instances of hydrocephalus.

    CARE HOSPITALS Page 22

  • 7/28/2019 By Dinesh Gannerlla

    23/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    Alpha is the frequency range from 8 Hz to 12 Hz. Hans Berger named the first rhythmic

    EEG activity he saw, the "alpha wave." This is activity in the 8-12 Hz range seen in the

    posterior regions of the head on both sides, being higher in amplitude on the dominant

    side. It is brought out by closing the eyes and by relaxation. It was noted to attenuate with

    eye opening or mental exertion. This activity is now referred to as "posterior basic

    rhythm," the "posterior dominant rhythm" or the "posterior alpha rhythm." The posterior

    basic rhythm is actually slower than 8 Hz in young children (therefore technically in the

    theta range). In addition to the posterior basic rhythm, there are two other normal alpha

    rhythms that are typically discussed: the mu rhythm and a temporal "third rhythm". Alpha

    can be abnormal; for example, an EEG that has diffuse alpha occurring in coma and is

    not responsive to external stimuli is referred to as "alpha coma".

    Mu rhythm is alpha-range activity that is seen over the sensorimotor cortex. It

    characteristically attenuates with movement of the contralateral arm (or mental imagery

    of movement of the contralateral arm).

    CARE HOSPITALS Page 23

  • 7/28/2019 By Dinesh Gannerlla

    24/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    Beta is the frequency range from 12 Hz to about 30 Hz. It is seen usually on both sides in

    symmetrical distribution and is most evident frontally. Low amplitude beta with multiple

    and varying frequencies is often associated with active, busy or anxious thinking andactive concentration. Rhythmic beta with a dominant set of frequencies is associated with

    various pathologies and drug effects, especially benzodiazepines. Activity over about 25

    Hz seen in the scalp EEG is rarely cerebral (i.e., it is most often artifactual). It may be

    absent or reduced in areas of cortical damage. It is the dominant rhythm in patients who

    are alert or anxious or who have their eyes open.

    Gamma is the frequency range approximately 26100 Hz. Because of the filtering

    properties of the skull and scalp, gamma rhythms can only be recorded from

    electrocorticography or possibly with magnetoencephalography. Gamma rhythms are

    thought to represent binding of different populations of neurons together into a network

    for the purpose of carrying out a certain cognitive or motor function.

    HEART LUNG MACHINE

    The coronary arteries supply blood to the heart. The most common cause of death for is heart

    attack which is usually caused by blockages in arteries. When the condition cannot be

    effectively treated with medicines or catheter-based angioplasty and stents(blood flows through

    CARE HOSPITALS Page 24

  • 7/28/2019 By Dinesh Gannerlla

    25/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITstents after opening clogged arteries), a form of open-heart surgery called coronary artery

    bypass grafting is recommended which usually takes 3 to 5 hours.. Coronary artery bypass

    grafting (CABG) uses arteries and/or veins from other parts of the body to bypass the blocked

    coronary arteries on the surface of the heart. This technique and technology make operations

    more effective and safer for patients.

    Open-Heart Surgery: For Open-Heart surgery, patient is made unconscious and various

    parameters like heartbeat, blood pressure, oxygen levels, breathing etc are monitored. A

    breathing tube is placed in lungs through throat and connected to a ventilator and surgeon

    makes a 6- to 8-inch incision in the center of chest wall. Then the chest bone is cut and rib cage

    is opened to access heart. Then a medicine is given to thin blood and keeps it from clotting. A

    heart-lung bypass machine is connected to heart and takes over for heart by replacing the heart's

    pumping action. The bypass machine allows the surgeon to operate on a heart that isn't moving

    and full of blood. Then medicines are given to stop the heartbeat once patient is connected to

    the heart-lung machine. A pipe is placed in heart to drain blood to the machine which removes

    carbon dioxide from blood, & adds oxygen, and then pumps the blood back into body. Tubes

    are then inserted into chest to drain fluid. Once the machine begins to work, the surgeon

    performs the surgery to repair heart problem.

    CARE HOSPITALS Page 25

    http://en.wikipedia.org/wiki/Image:PTCA_stent_NIH.gif
  • 7/28/2019 By Dinesh Gannerlla

    26/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    At the end of the surgery, heart is restarted using mild electric shocks. The pipes and tubes are

    removed from heart, and the heart-lung machine is stopped. Then again a medicine is given to

    allow the blood to clot and chest bone is closed with wires. Stitches or staples are used to close

    the incision. The breathing tube is removed. An advantage of open-heart surgery is that it's

    easier for the surgeon to operate for long and complex surgeries.

    Applications: Heart surgery is done to repair or replace valves that control blood flow through

    the heart, repair structures in the heart, implant devices to regulate heart rhythms, or replace a

    damaged heart with a healthy heart from a donor, bypass blocked arteries, treat arrhythmiasrepairaneurysms, treat angina (chest pain or discomfort) etc.

    Sarns 8000 Heart Lung Machine: Cardiothoracic surgical program is the most important goal

    of open-heart program. The Sarns 8000 is a modular heart lung machine that is available in a

    variety of configurations. It has 4 pumps units that are available with built-in centrifugal pump

    and power input is 220-240VAC, 50Hz,/440 V 3 Phase as appropriate fitted.

    CARE HOSPITALS Page 26

    http://www.nhlbi.nih.gov/health/dci/Diseases/arr/arr_whatis.htmlhttp://www.nhlbi.nih.gov/health/dci/Diseases/arm/arm_what.htmlhttp://www.nhlbi.nih.gov/health/dci/Diseases/Angina/Angina_WhatIs.htmlhttp://www.nhlbi.nih.gov/health/dci/Diseases/arr/arr_whatis.htmlhttp://www.nhlbi.nih.gov/health/dci/Diseases/arm/arm_what.htmlhttp://www.nhlbi.nih.gov/health/dci/Diseases/Angina/Angina_WhatIs.html
  • 7/28/2019 By Dinesh Gannerlla

    27/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    Equipment consist of following :

    5- Pump Console

    Temperature Control Module (Hypo-Hyper thermia unit)

    Monitors: a) Pressure monitor arterial and cardioplegia with transducers

    b) Time

    c) Temperature Monitor with probes

    d) Display of total volume of each infusion with delivery time

    Air- Oxygen Blender with hoses and Flow meter

    Safety Devices

    Ultrasonic air sensor

    Level Sensor

    Technical Specifications:

    A. 5- Pump Console:

    1. It has 5-pump console compactly arranged with separate power supply and control modules

    with easy access connectors for interchanging the pump.

    2. Each individual roller pump is capable of running independently on 220 V/50Hz supply

    with a spill proof base. The unit is supplied with a Battery backup for at least two pumps

    & all safety systems for a minimum of 15 minutes. Switch over from main power to battery

    backup is automatic and immediate.

    CARE HOSPITALS Page 27

  • 7/28/2019 By Dinesh Gannerlla

    28/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT3. Individual pump heads have Harvey Roller pumps with facility for tubing to be used

    adjustable from to 5/8 through 3/8 & by easily changeable mechanism. The total

    infusion volume in litres and delivery time, the flow rates in LPM and in RPM.

    4. Each Pump has easy mechanism for occlusion setting for different thickness of tubes with

    unidirectional hand crank facility as a critical safety feature.

    5. The Console has a compact base mount for the entire pump heads together, with pole and

    handles and variable, changeable tubing holders in each pump head.

    6. It has a movable oxygenator holder and roller pump has a self diagnostic circuit with

    provision to detect and display critical alarm conditions and optional pulsatile module

    which can be mounted on any of the blood pump.

    7. It has a venous control module with single pole mast with electronic venous line occluder,a monitor mount with adjustable monitoring arm and an instrument tray positionable with

    long monitoring arm.

    B. Temperature Control Module: Temperature control and monitor system with Cardioplegia

    supply and remote temperature display has following features:

    1. Simultaneous delivery of water for arterial and cardioplegia heat exchangers and to

    thermal blankets and works with a power supply of 220 20 V 50 Hz.

    2. Pressure regulated blanket ports maintain the temperature of arterial port with

    temperature display range of 0- 50 Celsius and remote accuracy of 0.3 Celsius.

    3. Microprocessor based unit control, cool, rewarm and maintains temperature. Water outlet

    temperature of heat exchanger and blanket range 0-42 C.

    4. Maximum flow performance of heat exchanger port 15 22 LPM; 480mmHg maximum

    pressure; Blanket 1.5 to 2.5 LPM at zero head.5. Ice generation and rewarming facility with venous difference mode settable at 6 to 10 C

    gradients to hold the water bath at higher than venous blood temperature.

    CARE HOSPITALS Page 28

  • 7/28/2019 By Dinesh Gannerlla

    29/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    6. Temperature probe module for the operating ranges of 0-50 C. and temperature probes

    to fit in standard oxygenators (bubble / membrane).

    C. Monitors: They have optional capability for computer interface to retrieve perfusion data.

    1. Pressure Monitor: It can monitor one arterial line pressure and two cardioplegia line

    pressures (total 3); along with necessary pressure transducers, cables and domes reusable,

    with accurate digital display and alarms.

    2. Time Monitor: 4 time displays -- 2 for arterial and 2 for cardioplegia delivery with stop,

    reset and start function.

    3. Temperature: 6 displays- 3 for patient monitoring & 3 for cardioplegia monitoring with

    digital display in Celsius with 6 necessary compatible temperature probes with 3 of them

    for nasal, rectal and esophageal use.

    D. Air-Oxygen Blender: Works at 50-60 PSI for membrane oxygenator with water trap

    attached with hoses & connections of min. of 5 meters length and triple flow glass flow meters.

    E. Safety and monitoring devices:

    1. Ultrasonic Air Sensor: It detect bubbles to work equally well with crystalloid and blood

    and can fit anywhere in the circuit easily.

    2. Level Sensor System: Ultrasonic transducers to work well with crystalloid and blood

    with adhesive pads & with alarm settings.

    F. Environmental factors:

    The unit is capable of operating continuously in ambient temperature of 10 -400 C and relative

    humidity of 15-90% and is capable of being stored continuously in ambient temperature of 0

    -500 C and relative humidity of 15-90% with general requirements of safety for Electromagnetic

    Compatibility.

    CARE HOSPITALS Page 29

  • 7/28/2019 By Dinesh Gannerlla

    30/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    VENTRICULAR ASSIST DEVICES

    Ventricular assist devices are innovative external pumps that temporarily support the circulation

    inpatients who are believed to have reversible heart failure or who are candidates for heart

    transplantation. The Cardiothoracic Surgeons use the Abiomed BVS 5000 Blood Pump, which

    can take over pumping action for the right, left, or both ventricles. Ventricular assist devices are

    used relatively rarely and are very expensive for the hospital, but are mandatory for any

    comprehensive cardiac program. They provide support for prolonged heart recovery and, in

    addition, can be an effective bridge to further therapy including transplantation.

    CARE HOSPITALS Page 30

  • 7/28/2019 By Dinesh Gannerlla

    31/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    CARE HOSPITALS Page 31

  • 7/28/2019 By Dinesh Gannerlla

    32/74

  • 7/28/2019 By Dinesh Gannerlla

    33/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    ENDOSCOPE

    INTRODUCTION:

    Endoscopy is a minimally invasive diagnostic medical procedure that uses tube like instruments

    (called endoscopes) to assess interior surfaces of an organ. This procedure is different from

    imaging tests, such as X-Rays, which also look inside of the body but usually do not place

    instruments inside the body.

    An endoscope is a device that uses fibre optics and powerful lens system s to provide lighting

    and visualization of the interior of a joint. The portion of the endoscopes inserted into the body

    may be rigid or flexible, depending upon the medical procedure. The rigid or flexible tube not

    only provide an image for visual inspection and photography, but also enable taking biopsies

    and retrieval o foreign objects.

    Endoscopes uses two fibre optic lines. A light fibre carries light into the body cavity and an

    image fibre carries the image of the body cavity back to the physicians viewing lens. There is

    also a separate port to allow for administration of drugs, suction, and irrigation. This port may

    also be used to introduce small folding instruments such as forceps, scissors, brushes, snares

    and baskets for tissue incision (removal), sampling or other diagnostic and therapeutic work.

    Endoscopes may be used in conjunction with a camera or video recorder to document images of

    the inside of the joint or chronicle an endoscopic procedure. New endoscopes have digital

    capabilities for manipulating and enhancing the video images.

    CARE HOSPITALS Page 33

  • 7/28/2019 By Dinesh Gannerlla

    34/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITMany endoscopic procedures are considered to be relatively painless and, at worst, associated

    with mild discomfort; for eg. As in esophagogastroduodenoscopy, most patients tolerate the

    procedure with only topical anaesthesia of the oropharynx using lignocaine spray

    Complications are not common (only 5% of all operations ) but can include perforation of the

    organ under inspection with the endoscope or biopsy instrument. If that occurs open surgery

    may be required to repair the injury

    ENDOSCOPE

    There are many different kinds of endoscopes, or scopes. Some are hollow, allowing the

    doctor to see directly into the body, while others use fibre optic (flexible glass or plastic fibers

    that transmit light). Still others have small video camera on some endoscopes are stiff while

    others are flexible. Endoscopes also vary in length. Each type is specially designed for looking

    at a different part of the body. Depending on the area of the body being looked at, the

    endoscope may be inserted through an opening like the mouth, anus, or urethra. In some cases,

    endoscope is inserted through small incision.

    TYPES OF ENDOSCOPES:

    Type of Endoscope Inserted into or

    through

    Body area

    Examined

    Name of procedures

    Bronchoscope mouth or nose Trachea

    (windpipe) and

    bronchi (tubes

    inside the lungs)

    Bronchoscopy,

    flexible

    bronchoscopy

    Colonoscope Anus Colon (large

    intestine)

    Colonoscopy, lower

    endoscopy

    CARE HOSPITALS Page 34

  • 7/28/2019 By Dinesh Gannerlla

    35/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    Cystoscope Urethra Urinary bladder Cystoscopy,

    cystourethroscopy

    Esophagogastroduode-

    noscope

    Mouth Esophagus,

    stomach and

    duodenum

    Esophagogastro-

    uodenoscopy (EGD),

    panendoscopy,

    upper endoscopy

    gastroscopy

    Hysteroscope Vagina Inside of uterus Hysteroscopy

    Laproscope Incision in

    abdomen

    Space inside

    abdomen and

    pelvis

    Laproscopy,

    peritoneal endoscopy

    Laryngoscope Mouth or nose Larynx (voice

    box)

    Layngoscopy

    Mediastinoscope Incision above

    sternum (breast

    bone)

    Mediastinum

    (space between

    lungs)

    Mediastinoscopy

    Sigmoidoscope Anus Rectum and

    sigmoid colon

    Sigmoidoscopy,

    flexible

    CARE HOSPITALS Page 35

  • 7/28/2019 By Dinesh Gannerlla

    36/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    (lower part of

    large intestine)

    sigmoidoscopy,

    proctosigmoidoscopy

    Thoracoscope Incision Space between

    lungs and chest

    wall

    Thoracoscopy,

    pleuroscopy

    Gastroscopy viewing of the stomach. Used to check for ulcers, bleeding or tumours.

    Colonoscopy and Sigmoidoscopy viewing of the colon. Used to check for tumours or

    polyps (small growths).

    Bronchoscopy viewing of the airways and lung tissue. Used to remove foreign bodies,

    take samples to diagnose lung cancer or other lung diseases and to assess the condition of

    the airways following smoke inhalation.

    Cystoscopy viewing of the bladder. Used to check for tumors, stones and other

    abnormalities.

    Hysteroscopy viewing of the uterus. Used to check for adhesions or polyps and to

    investigate excessive menstrual bleeding.

    In other types of endoscopy, the skin is opened to allow the scope to enter an area:

    Arthroscopy to look into a joint such as the knee. Used to check for torn ligaments or

    damaged cartilage.

    CARE HOSPITALS Page 36

  • 7/28/2019 By Dinesh Gannerlla

    37/74

  • 7/28/2019 By Dinesh Gannerlla

    38/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITmicroscope or tested in other ways to know for sure whether or not cancer is present. A biopsy

    is the best way to find out if a growth is cancer or something else.

    In some cases endoscopes are used to help find out how far a cancer may have spread.

    Thoracoscopy and laparoscopy can be especially useful in finding out whether certain cancershave spread into one of the body cavities (thorax or abdomen). They let the surgeon look at

    these spaces without making a large incision in the skin.

    Some types of endoscopy can help make imaging tests more accurate. This can e especially

    helpful when trying to find the stage (extent) of cancer within the body. Most people are

    familiar with ultrasound, an imaging test in which a transducer is moved over the skin. It sends

    sound waves into the body, which bounce back in a pattern that creates an image of the inside

    of the body.

    ENDOSCOPIC ULTRASOUND (ESU) is a procedure in which a small transducer on the tip of

    an endoscope is inserted into the mouth or rectum. By putting the transducer on the tip of the

    endoscope, it can get closer to the area where the tumor is to take pictures. It is used to get

    information about problems in the digestive tract and surrounding organs. Because the

    transducer is close to the organ being studied, it can make very detailed pictures. EUS can be

    used to see how deeply a tumor may have penetrated into the rectum or esophagus, or into an

    organ like the pancreas. It can also help show whether certain lymph nodes are enlarged. EUS

    can also help a doctor guide a needle into a lymph node or other suspicious area to do biopsy.

    CARE HOSPITALS Page 38

  • 7/28/2019 By Dinesh Gannerlla

    39/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    TREATING CANCER:

    Endoscopes can be used to remove or destroy small cancers. Small instruments passed through

    an endoscope can be used to cut out small growths. Some forms of endoscopy allow doctors to

    use a cautery or laser through the tip of the endoscope to burn or vaporize growths.

    Over the last decade or so, a wide variety of endoscopic tools have been developed that allow

    doctors to perform minimally invasive surgery. This type of surgery is sometimes called

    keyhole surgery or when used for the abdomen, is called laparoscopic surgery. Instead of

    using one long surgical incision, it involves making several small incisions in the skin usually inthe chest or abdomen. Long, thin instruments are then inserted through the holes to reach the

    inside of the body. A video endoscope is placed through one of the holes to allow the surgeon to

    see inside during the operation.

    This type of surgery was first used for fairly minor procedures such as gall bladder removal, but

    in recent years have begun using it to treat some types of cancer. It is also used to treat early

    cancers of the lung, colon, prostate, and some other organs.

    There are pros and cons to keyhole surgeries. There is generally less blood loss during the

    operation, and patients often recover faster and with less pain because the incisions are smaller

    than in regular surgery. Some forms of keyhole surgery use robotic arms, which a surgeon

    controls from a console. This technique allows for better magnification of the area and more

    precision in working with the delicate surgical instruments.

    CARE HOSPITALS Page 39

  • 7/28/2019 By Dinesh Gannerlla

    40/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITAFTER THE ENDOSCOPY PROCEDURE:

    After the procedure the patient will be observed and monitored by a qualified individual in the

    endoscopy or a recovery area until a significant portion of the medication has worn off

    Occasionally a patient is left with a mild sore throat, which promptly responds to saline gargles

    or a feeling of distention from the insufflated air that was used during the procedure. Both

    problems are mild and fleeting. When fully recovered, the patient will be instructed when to

    resume his/her usual diet (probably within a few hours) and will be allowed to be taken home.

    Because of the use of sedation, most facilities mandate that the patient is taken home by another

    person and not to drive on his/her own or handle machinery for the remainder of the day.

    RISKS

    Infection

    Punctured organs

    Allergic reactions due to Contrast agents or dyes (such as those used in a CT scan)

    Over-sedation

    COMPONENTS OF ENDOSCOPE:

    CARE HOSPITALS Page 40

    http://en.wikipedia.org/wiki/Image:Am_ulcer.gif
  • 7/28/2019 By Dinesh Gannerlla

    41/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITAn endoscope can consist of

    a rigid or flexible tube

    A light delivery system to illuminate the organ or object under inspection. The light

    source is normally outside the body and the light is typically directed via an optical fibresystem

    a lens system transmitting the image to the viewer from the fiberscope

    an additional channel to allow entry of medical instruments or manipulators

    DISTAL END PART OF ENDOSCOPE

    A distal end part of an endoscope including an end cap provided with a nozzle for spouting a

    fluid in a predetermined direction. The end cap is detachable with respect to a distal end block

    which is provided at the distal end of an insert part of the endoscope. Engagement portions are

    provided on the end cap and the distal end block to regulate the condition in which the end cap

    is fitted to the distal end block. The engagement portion of the end cap and the engagement

    portion of the distal end block are formed such that, among different models of endoscopes in

    which the distal end blocks have approximately the same outer diameter, the engagementportion of the end cap off one model cannot be engaged with

    CARE HOSPITALS Page 41

  • 7/28/2019 By Dinesh Gannerlla

    42/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITA system of different models of endoscopes, each endoscope having a distal end part, each

    distal end part comprising:

    an end cap provided with a nozzle for spouting a fluid in a predetermined direction, said end

    cap being detachable with respect to a distal end block which is provided at a distal end of an

    insert part of an endoscope; and

    engagement means provided on said end cap and said distal end block to regulate a condition in

    which said end cap is fitted to said distal end block, said engagement means including mating

    portions which are engageable with each other, said mating portions also positioning the end

    cap with respect to said distal end block in a direction of rotation,

    said engagement means provided on said end cap and said engagement means provided on said

    distal end block being formed such that, among said different models of endoscopes in said

    system in which distal end blocks have approximately the same outer diameter, said

    engagement means provided on said end cap of one model cannot be engaged with said

    engagement means provided on said distal end block of another model.

    Distal endoscope part having light emitting source such as light emitting diodes as

    illuminating means

    A substrate having a plurality of light emitting diodes united therewith lies on a plane

    containing the longitudinal axis of an insertion unit of an endoscope and its neighborhood.

    Likewise, part of a first objective surface lies on the plane containing the longitudinal axis of

    the insertion unit of the endoscope and its neighborhood. As long as the diameter of the

    insertion unit remains unchanged, the plane containing the longitudinal axis of the insertion unit

    of the endoscope and its neighborhood provides the largest area for the light emitting diodes.

    CARE HOSPITALS Page 42

  • 7/28/2019 By Dinesh Gannerlla

    43/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITThe light emitting diode sub-assembly is therefore placed on the plane, whereby the outer

    diameter of a distal endoscope part can be made as small as possible.

    BRONCHOSCOPE

    BRONCHOSCOPY is a medical procedure where a tube is inserted into the airways, usually

    through the nose or mouth. This allows the practitioner to examine inside a patients airway for

    abnormalities such as foreign bodies, bleeding, tumors, or inflammation. The practitioner often

    takes samples from inside the lungs: biopsies, fluid (bronchoalveolar lavage), or endobronchial

    brushing. The practitioner may use either a rigid bronchoscope or flexible bronchoscope.

    CARE HOSPITALS Page 43

  • 7/28/2019 By Dinesh Gannerlla

    44/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    Rigid Bronchoscopy:

    A rigid bronchoscope is a straight, hollow, metal tube. Doctors perform rigid bronchoscopy less

    often today, but it remains the procedure of choice for removing foreign materials, as its

    increased thickness allows instruments to be more easily inserted through it. Rigid

    bronchoscopy also becomes useful when bleeding interferes with viewing the examining area

    and allows for more interventions, such as cautery to stop the bleeding.

    Flexible Bronchoscopy:

    A flexible bronchoscope is a long tube that contains small clear optical fibers that transmit light

    images as the tube bends. Its flexibility allows this instrument to reach further into the airway.

    The procedure can be performed easily and safely under local anesthesia. As flexible

    bronchoscopes become more advanced, it is likely that they will replace rigid bronchoscopes for

    most procedures.

    CARE HOSPITALS Page 44

  • 7/28/2019 By Dinesh Gannerlla

    45/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    GASTROSCOPE

    GASTROSCOPY is an examination of the inside of the gullet, stomach and duodenum. It is

    performed by using a thin, flexible fibre optic instrument that is passed through the mouth and

    allows the doctor to see whether there is any damage to the lining of the oesophagus (gullet) or

    stomach, and whether there are any ulcers in the stomach or duodenum.

    Gastroscope

    The gastroscope is a flexible plastic tube approximately four feet long and one half inch wide

    The gastroscope contains optic fibers with a light source that allow the gastroscope to function

    like a video camera.

    CARE HOSPITALS Page 45

  • 7/28/2019 By Dinesh Gannerlla

    46/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITThe use of the gastroscope is esophagogastroduodenoscopy or EGD. An EGD is done when the

    doctor suspects that there is a problem with the swallowing tube, stomach, or small intestine. In

    the ICU, the two most common reasons for the using the gastroscope are to evaluate a patient

    suspected of bleeding from the stomach or intestines and to help place a gastrostomy tube.

    Occasionally, a special gastroscope is used for patients who may have an inflamed pancreas or

    blocked gallbladder drainage system.

    CARE HOSPITALS Page 46

  • 7/28/2019 By Dinesh Gannerlla

    47/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    RADIOLOGY

    MAGNETIC RESONANCE IMAGING

    Medical Imaging has created a revolutionary change in the field of medicine, the imaging of the

    human body is very much useful in determining the diseases and used for all the treatments.

    The magnetic resonance phenomenon can be described by both classical and quantum

    mechanical approaches. Magnetic resonance imaging is based on the techniques of nuclear

    magnetic resonance. MRI scanner uses powerful magnets to polarize and excite hydrogen

    nuclei (singleproton) in water molecules in human tissue, producing a detectable signal which

    is spatially encoded resulting in images of the body. MRI involves the use of three kinds of

    electromagnetic field: Avery strong (of the order of units of teslas) static magnetic field to

    polarize the hydrogen nuclei, called the static field, A weaker time-varying (of the order of 1

    kHz) for spatial encoding, called the gradient fields, A weak radio-frequency (RF) field for

    manipulation of the hydrogen.

    The scanner first aligns the nuclear spins of hydrogen atoms in the patient and starts rotating

    them in a perfect concert. The nuclei emit maximum-strength electromagnetic waves at the

    start, but over time the rotating spins get out of synch, simply due to small differences in localmagnetic fields. The unsynchronized spins cause the combined electromagnetic signal to decay

    with time, a phenomenon called relaxation. A slice is selected applying a gradient in a particular

    direction (X, YorZ). Magnetic resonance signals are then formed by means of the application of

    magnetic field gradients along three different directions. Finally, the signals are acquired and

    Fourier transformed to form a two-dimensional or three-dimensional image.

    MRI Construction and operation

    CARE HOSPITALS Page 47

    http://en.wikipedia.org/w/index.php?title=MRI_scanner&action=edithttp://en.wikipedia.org/wiki/Hydrogenhttp://en.wikipedia.org/wiki/Protonhttp://en.wikipedia.org/wiki/Electromagnetic_fieldhttp://en.wikipedia.org/wiki/Tesla_(unit)http://en.wikipedia.org/wiki/Radio-frequencyhttp://en.wikipedia.org/wiki/RFhttp://en.wikipedia.org/w/index.php?title=MRI_scanner&action=edithttp://en.wikipedia.org/wiki/Hydrogenhttp://en.wikipedia.org/wiki/Protonhttp://en.wikipedia.org/wiki/Electromagnetic_fieldhttp://en.wikipedia.org/wiki/Tesla_(unit)http://en.wikipedia.org/wiki/Radio-frequencyhttp://en.wikipedia.org/wiki/RF
  • 7/28/2019 By Dinesh Gannerlla

    48/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    Magnet: The imaging magnet is the most expensive component of the magnetic resonance

    imaging system. Most magnets are of the superconducting type. A superconducting magnet is

    an electromagnet made of superconducting wire. Superconducting wire has a resistance

    approximately equal to zero when it is cooled to a temperature close to absolute zero (-273.15 o

    C or 0 K) by immersing it in liquid helium. Once current is caused to flow in the coil it will

    continue to flow as long as the coil is kept at liquid helium temperatures. The length of

    superconducting wire in the magnet is typically several miles. The coil of wire is kept at a

    temperature of 4.2K by immersing it in liquid helium. The coil and liquid helium is kept in a

    large dewar. The typical volume of liquid Helium in an MRI magnet is 1700 liters. In early

    magnet designs, this dewar was typically surrounded by a liquid nitrogen (77.4K) dewar which

    acts as a thermal buffer between the room temperature (293K) and the liquid helium. In later

    magnet designs, the liquid nitrogen region was replaced by a dewar cooled by a refrigerator.

    This design eliminates the need to add liquid nitrogen to the magnet. It is expected that

    cryocoolers for the helium reservoir will soon be available

    CARE HOSPITALS Page 48

  • 7/28/2019 By Dinesh Gannerlla

    49/74

  • 7/28/2019 By Dinesh Gannerlla

    50/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITAn imaging coil must resonate, or efficiently store energy, at the Larmor frequency. All

    imaging coils are composed of an inductor, or inductive elements, and a set of capacitive

    elements. The resonant frequency, of an RF coil is determined by the inductance (L) and

    capacitance (C) of the inductor capacitor circuit.

    .

    RF Coil for 4.7T Magnet

    There are many types of imaging coils. Volume coils surround the imaged object while surface

    coils are placed adjacent to the imaged object. An internal coil is one designed to record

    information from regions outside of the coil, such as a catheter coil designed to be inserted into

    a blood vessel. Some coils can operate as both the transmitter of the B 1 field and the receiver of

    the RF signal. Other coils are designed as only the receiver of the RF signal. When a receive

    only coil is used, a larger coil on the imager is used as the transmitter of RF energy to producing

    the 90o and 180o pulses.

    CARE HOSPITALS Page 50

  • 7/28/2019 By Dinesh Gannerlla

    51/74

  • 7/28/2019 By Dinesh Gannerlla

    52/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    Slice selection

    A single frequency RF pulse is applied to the whole sample; only a narrow plane perpendicular

    to the longitudinal axis at the centre of the sample will absorb the RF energy. Everywhere else

    in the sample is receiving the wrong frequency of excitation for resonance to occur. This

    technique allows a slice, with thickness determined by the magnetic field gradient strength, to

    be selected from a sample.

    Technical Specifications

    Tim

    Up to 76 seamlessly integrated coil elements with up to 18 RF channels

    205 cm FoV. Whole Body

    PAT. Unlimited

    Workflow automation

    Examples are: Phoenix, Auto Align, Inline Technology

    Compact magnet

    Ultra compact 1.5T magnet (length: 120 cm)

    Wide, patient-friendly inner bore diameter (70 cm)

    Magnet weight including Helium only approx. 3,800 kg

    Unique CT-like 70 cm Open Bore diameter, with high homogeneity over 30 cm

    CARE HOSPITALS Page 52

  • 7/28/2019 By Dinesh Gannerlla

    53/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    Computer

    syngo-speaking user interface syngo is the common software platform for all imaging

    modalities. Enhanced productivity with minimized user interactions per operation step. Based

    on a powerful Pentium 4/3 GHz Panoramic Recon Image Processor reconstructing up to 3,226

    images per second (256 x 256, 25% recFoV) in combination with a Pentium 4-based Host

    Computer with two CPUs/3 GHz and 2 GB RAM capacity.

    Cost-effective siting

    30 qm/325 sq. ft. floor space only

    No computer room required for a total of just two electronic cabinets (water-cooled).

    Computer tomography(ct scan)

    CT scanningsometimes called CAT scanningis a noninvasive medical test that helps

    physicians diagnose and treat medical conditions.

    CT scanning combines special x-ray equipment with sophisticated computers to produce

    multiple images or pictures of the inside of the body. These cross-sectional images of the areabeing studied can then be examined on a computer monitor, printed or transferred to a CD.

    CT scans of internal organs, bones, soft tissue and blood vessels provide greater clarity and

    reveal more details than regular x-ray exams.

    Using specialized equipment and expertise to create and interpret CT scans of the

    body, radiologists can more easily diagnose problems such as cancers, cardiovascular disease,

    infectious disease, appendicitis, trauma and musculoskeletal disorders.

    CARE HOSPITALS Page 53

    http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=17http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=169http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=17http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=169
  • 7/28/2019 By Dinesh Gannerlla

    54/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    CT-SCAN MACHINE

    The CT scanner is typically a large, box like machine with a hole, or short tunnel, in the center.

    You will lie on a narrow examination table that slides into and out of this tunnel. Rotating

    around you, the x-ray tube and electronic x-ray detectors are located opposite each other in a

    ring, called a gantry. The computer workstation that processes the imaging information is

    located in a separate room, where the technologist operates the scanner and monitors your

    examination.

    CARE HOSPITALS Page 54

  • 7/28/2019 By Dinesh Gannerlla

    55/74

  • 7/28/2019 By Dinesh Gannerlla

    56/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITIn many ways CT scanning works very much like other x-ray examinations. X-rays are a form

    of radiationlike light or radio wavesthat can be directed at the body. Different body parts

    absorb the x-rays in varying degrees.

    In a conventional x-ray exam, a small burst of radiation is aimed at and passes through the

    body, recording an image on photographic film or a special image recording plate. Bones

    appear white on the x-ray; soft tissue shows up in shades of gray and air appears black.

    With CT scanning, numerous x-ray beams and a set of electronic x-ray detectors rotate aroundyou, measuring the amount of radiation being absorbed throughout your body. At the same

    time, the examination table is moving through the scanner, so that the x-ray beam follows a

    spiral path. A special computer program processes this large volume of data to create two-

    dimensional cross-sectional images of your body, which are then displayed on a monitor. This

    technique is called helical or spiral CT.

    CT imaging is sometimes compared to looking into a loaf of bread by cutting the loaf into thin

    slices. When the image slices are reassembled by computer software, the result is a very

    detailed multidimensional view of the body's interior.

    Refinements in detector technology allow new CT scanners to obtain multiple slices in a singlerotation. These scanners, called "multislice CT" or "multidetector CT," allow thinner slices to

    be obtained in a shorter period of time, resulting in more detail and additional view capabilities.

    BENEFITS AND RISKS OF CT-SCAN

    Benefits

    CT scanning is painless, noninvasive and accurate.

    A major advantage of CT is its ability to image bone, soft tissue and blood vessels all at the

    same time.

    Unlike conventional x-rays, CT scanning provides very detailed images of many types of

    tissue as well as the lungs, bones, and blood vessels.

    CT examinations are fast and simple; in emergency cases, they can reveal internal injuries

    and bleeding quickly enough to help save lives.

    CT has been shown to be a cost-effective imaging tool for a wide range of clinical

    problems.

    CT is less sensitive to patient movement than MRI.

    CT can be performed if you have an implanted medical device of any kind, unlike MRI. A diagnosis determined by CT scanning may eliminate the need for exploratory surgery and

    surgical biopsy.

    No radiation remains in a patient's body after a CT examination.

    X-rays used in CT scans usually have no immediate side effects.

    CARE HOSPITALS Page 56

  • 7/28/2019 By Dinesh Gannerlla

    57/74

  • 7/28/2019 By Dinesh Gannerlla

    58/74

  • 7/28/2019 By Dinesh Gannerlla

    59/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    The equipment typically used for bone x-rays consists of an x-ray tube suspended over a table

    on which the patient lies. A drawer under the table holds the x-ray film orimage recording

    plate. Sometimes the x-ray is taken with the patient standing upright, as in cases of knee x-rays.

    A portable x-ray machine is a compact apparatus that can be taken to the patient in a hospital

    bed or the emergency room. The x-ray tube is connected to a flexible arm that is extended over

    the patient while an x-ray film holder or image recording plate is placed beneath the patient.

    PROCEDURE

    X-rays are a form of radiation like light or radio waves. X-rays pass through most objects,

    including the body. Once it is carefully aimed at the part of the body being examined, an x-ray

    machine produces a small burst of radiation that passes through the body, recording an image

    on photographic film or a special digital image recording plate.

    Different parts of the body absorb the x-rays in varying degrees. Dense bone absorbs much of

    the radiation while soft tissue, such as muscle, fat and organs, allow more of the x-rays to pass

    through them. As a result, bones appear white on the x-ray, soft tissue shows up in shades of

    gray and air appears black.

    Until recently, x-ray images were maintained as hard film copy (much like a photographic

    negative). Today, most images are digital files that are stored electronically. These storedimages are easily accessible and are frequently compared to current x-ray images for diagnosis

    and disease management.

    The technologist, an individual specially trained to perform radiology examinations, positions

    the patient on the x-ray table and places the x-ray film holder or digital recording plate under

    the table in the area of the body being imaged. When necessary, sandbags, pillows or other

    CARE HOSPITALS Page 59

    http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=695http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=695http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=695http://www.radiologyinfo.org/en/glossary/glossary1.cfm?gid=695
  • 7/28/2019 By Dinesh Gannerlla

    60/74

  • 7/28/2019 By Dinesh Gannerlla

    61/74

  • 7/28/2019 By Dinesh Gannerlla

    62/74

  • 7/28/2019 By Dinesh Gannerlla

    63/74

  • 7/28/2019 By Dinesh Gannerlla

    64/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    CSSD EQUIPMENT

    Sterilizers:

    We manufacture a wide array of medical sterilizer and hospital sterilizer that are

    available with different features to facilitate the process of sterilizing which is one of

    the most important aspects of medical practice. These are made of superior quality

    material like stainless steel that makes them hygienic by preventing them from being

    prey to rust. The different variety includes, Triple Chamber Autoclave, Double

    Drum Autoclave SS, Single Drum Autoclave Aluminum etc.

    Horizontal Rectangular High Pressure Steam Sterilizer

    In our medical sterilizer horizontal autoclave is available in

    triple-walled (jacketed). Triple-walled execution is preferably

    used in hospitals for sterilization of all goods, suitable for the

    dressing drums size up to 12x15 Inches especially of infectious

    garbage & laboratory waste. The garbage sacks are sterilized inbuckets, which are placed one above the other. Our medical

    sterilizer hasinner and outer wall made of heavy gauge of stainless steel and

    Construction of the unit totally made of 304 stainless steel heavy gauge sheet middle

    jacket made of brass or stainless steel and gun metal

    Technical Data:

    Steam Working Pressure 1.26 kgf/cm2 (2.2 kgf/cm2 in case High speed

    sterilizer

    Steam Working Temperature 121oC

    Operating Voltage 240, single phase, AC supplies, 50 Hz.

    CARE HOSPITALS Page 64

  • 7/28/2019 By Dinesh Gannerlla

    65/74

  • 7/28/2019 By Dinesh Gannerlla

    66/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITabove the other. In hospital sterilizer inner and outer wall made of heavy gauge of

    stainless steel and middle jacket made of brass and gun metal

    Technical Data:

    Steam Working Pressure 1.26 kgf/cm2 (2.2 kgf/cm2 in case High speed

    sterilizer

    Steam Working Temperature 121oC

    Operating Voltage 240, single phase, AC supply, 50 Hz.

    Sterilization Period 45 to 50 minutes.

    Radiant locking system, cold water drainage system Steam

    Exhaust:

    5 to 7 provided with timer from 1 to 60 Min with double safety valve, steam

    releasing valve, pressure gauge for the pressure inside the chamber, and one pressure

    gauge to show the pressure of outer chamber both the chamber made of stainless

    steel lid of the unit made of heavy duty gun metal piece for extra safety, gasket,

    water level indicator to see the water inside, heavy duty power plug with socket

    provided with the unit

    Accessories:

    One spare element, and two gaskets, our common size 400x1200 mm

    Vertical Sterilizer

    CARE HOSPITALS Page 66

  • 7/28/2019 By Dinesh Gannerlla

    67/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRITSterilization accessories:

    Auto claves

    Ultra sonic cleaners

    Autoclave racks

    Instrument Trays

    Autoclave Chemicals

    Autoclave supplies and testing

    The main objectives of the Central Sterile Supply Department are: To provide sterilized material from a central department where sterilizing

    practice is conducted under conditions, which are controlled, thereby

    contributing to a reduction in the incidence of hospital infection.

    To take some of the work of the Nursing staff so that they can devote more

    time to their patients.

    To maintain record of effectiveness of cleaning disinfections and sterilization

    process.

    To monitor and enforce controls necessary to prevent cross infection

    according to infection control policy.

    To maintain an inventory of supplies and equipment.

    To stay updated regarding developments in the field in the interest of

    efficiency, economy, accuracy and provision of better patient care.

    To provide a safe environment for the patients and staff.

    Technical Data:

    CARE HOSPITALS Page 67

  • 7/28/2019 By Dinesh Gannerlla

    68/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    CARE HOSPITALS Page 68

    Steam

    Working

    Pressure

    1.26 kgf/cm2 (2.2 kgf/cm2 in case High

    speed sterilizer

    Steam

    Working

    Temp

    121oC (134o in case of High speed

    sterilizer)

    Hyd. Tested at Jacket-Twice the working pressure.

    Chamber-One & half time the working

    pressure.

    Operating

    Voltage

    400/440, 3 phase, AC supply, 50c/s.

    Sterilization

    Period

    20 to 25 minutes(in case of High speed

    sterilizer 5 to 7 minutes).

    Steam Exhaust 5 to 7 minutes (in case of High speed

    sterilizer 1 minute)

  • 7/28/2019 By Dinesh Gannerlla

    69/74

    MINI PROJECT BIOMEDICAL ENGINEERING BVRIT

    INDICATORS

    In the process of sterilization of medical instruments a indicator is used by the

    operator to find out weather the material in the sterilizer is completely sterile or not. A paper like

    substance which is applied with some chemical is used as an indicator. Initially the tape will be

    yellow in color, after sterilization it turns black.

    It is called AUTO CLAVE. This indicator is attached to the

    material to be sterilized and verified after sterilization process.

    CARE HOSPITALS Page 69