In the Laboratory 1182 Journal of Chemical Education Vol. 80 No. 10 October 2003 JChemEd.chem.wisc.edu The use of Pasteur pipets for microscale column chro- matography is commonly reported in this Journal (1). How- ever, obtaining excellent separation depends on forming uniform columns, a daunting challenge for many students. This Journal has reported the use of a mercury-filled pres- surization system for column chromatography (2), a battery- operated, portable, aeration pump for packing ion-exchange resin into a large column (3), a balloon filled with air (4), an aquarium pump (5), and a 50-mL syringe (6). A more re- cent article describes the use of a plastic syringe barrel as a pressurized filtration system (7). Pressurization of a Pasteur pipet is easily accomplished with a plastic syringe and a length of Tygon tubing. Even nonmajor chemistry students can pres- sure-pack uniform columns on their first try (Figure 1). The apparatus costs about two dollars per student. By using pressure, the column can be dry-packed rather than slurry-packed. This simplifies the process of adding cotton Manual Microscale Column Chromatography Pressurization Apparatus W Bruce W. Baldwin Chemistry Department, Spring Arbor University, Spring Arbor, MI 49283; [email protected] The Microscale Laboratory edited by R. David Crouch Dickinson College Carlisle, PA 17013-2896 wadding, dry-packing material, and sand into the Pasteur pi- pet followed by the eluent. Transient pressures of 20 to 50 psi at the air–liquid interface were estimated by applying Boyle’s law to the observed air compression ratios of 1.5 to 3. Because the system was open (the pipet tip is plugged with a piece of cotton, not stoppered), these pressures were easily relieved and no breakage occurred. After assembling the col- umn, solvent was easily added to the headspace above the sand and quickly forced through the column using the sy- ringe to pressurize the column until the packing material be- came uniformly wetted. The ability to easily pressurize the column also dramatically speeds the passage of a sample through the column, decreasing the time necessary for ob- taining the fractions. Chemistry students at Spring Arbor University have used this technique successfully on a wide variety of separations, such as the separation of carotene from chlorophyll in spin- ach extract and the separation of ferrocene from acetyl- ferrocene. The fine control of column pressurization afforded by this apparatus allows students from first year through se- nior levels to obtain uniformly dry-packed columns useful for excellent separation of a wide range of chemical prod- ucts. Hazards The materials used in the manual pressurization appa- ratus are inherently safe. The pressures obtained inside the Pasteur pipets are never sufficient to burst the glass. Students have reported no problems with the apparatus. Appropriate safety goggles should be worn when working with chemicals. W Supplemental Material A student handout and notes for the instructor are avail- able in this issue of JCE Online. Literature Cited 1. Reynolds, R. C.; O’Dell, C. A. J. Chem. Educ. 1992, 69, 989– 991. 2. Chen, S. C. J. Chem. Educ. 1972, 49, 318. 3. Williams, H. P.; Moore, D. H.; Byrd, F. A. J. Chem. Educ. 1974, 51, 370. 4. Bell, W. L.; Edmondson, R. D. J. Chem. Educ. 1986, 63, 361. 5. Jacobson, B. M. J. Chem. Educ. 1988, 65, 459. 6. Krause, J. G. J. Chem. Educ. 1991, 68, 790. 7. Begtrup, M. J. Chem. Educ. 2001, 78, 543. Figure 1. The manual microscale column chromatography pressur- ization apparatus. sand packing material cotton syringe and plunger headspace Tygon tubing

Bruce W. Baldwin- Manual Microscale Column Chromatography Pressurization Apparatus

Embed Size (px)

Citation preview

Page 1: Bruce W. Baldwin- Manual Microscale Column Chromatography Pressurization Apparatus

In the Laboratory

1182 Journal of Chemical Education • Vol. 80 No. 10 October 2003 • JChemEd.chem.wisc.edu

The use of Pasteur pipets for microscale column chro-matography is commonly reported in this Journal (1). How-ever, obtaining excellent separation depends on forminguniform columns, a daunting challenge for many students.This Journal has reported the use of a mercury-filled pres-surization system for column chromatography (2), a battery-operated, portable, aeration pump for packing ion-exchangeresin into a large column (3), a balloon filled with air (4), anaquarium pump (5), and a 50-mL syringe (6). A more re-cent article describes the use of a plastic syringe barrel as apressurized filtration system (7). Pressurization of a Pasteurpipet is easily accomplished with a plastic syringe and a lengthof Tygon tubing. Even nonmajor chemistry students can pres-sure-pack uniform columns on their first try (Figure 1).

The apparatus costs about two dollars per student. Byusing pressure, the column can be dry-packed rather thanslurry-packed. This simplifies the process of adding cotton

Manual Microscale Column ChromatographyPressurization Apparatus WBruce W. BaldwinChemistry Department, Spring Arbor University, Spring Arbor, MI 49283; [email protected]

The Microscale Laboratoryedited by

R. David CrouchDickinson College

Carlisle, PA 17013-2896

wadding, dry-packing material, and sand into the Pasteur pi-pet followed by the eluent. Transient pressures of 20 to 50psi at the air–liquid interface were estimated by applyingBoyle’s law to the observed air compression ratios of 1.5 to3. Because the system was open (the pipet tip is plugged witha piece of cotton, not stoppered), these pressures were easilyrelieved and no breakage occurred. After assembling the col-umn, solvent was easily added to the headspace above thesand and quickly forced through the column using the sy-ringe to pressurize the column until the packing material be-came uniformly wetted. The ability to easily pressurize thecolumn also dramatically speeds the passage of a samplethrough the column, decreasing the time necessary for ob-taining the fractions.

Chemistry students at Spring Arbor University have usedthis technique successfully on a wide variety of separations,such as the separation of carotene from chlorophyll in spin-ach extract and the separation of ferrocene from acetyl-ferrocene. The fine control of column pressurization affordedby this apparatus allows students from first year through se-nior levels to obtain uniformly dry-packed columns usefulfor excellent separation of a wide range of chemical prod-ucts.

Hazards

The materials used in the manual pressurization appa-ratus are inherently safe. The pressures obtained inside thePasteur pipets are never sufficient to burst the glass. Studentshave reported no problems with the apparatus. Appropriatesafety goggles should be worn when working with chemicals.

WSupplemental Material

A student handout and notes for the instructor are avail-able in this issue of JCE Online.

Literature Cited

1. Reynolds, R. C.; O’Dell, C. A. J. Chem. Educ. 1992, 69, 989–991.

2. Chen, S. C. J. Chem. Educ. 1972, 49, 318.3. Williams, H. P.; Moore, D. H.; Byrd, F. A. J. Chem. Educ.

1974, 51, 370.4. Bell, W. L.; Edmondson, R. D. J. Chem. Educ. 1986, 63, 361.5. Jacobson, B. M. J. Chem. Educ. 1988, 65, 459.6. Krause, J. G. J. Chem. Educ. 1991, 68, 790.7. Begtrup, M. J. Chem. Educ. 2001, 78, 543.

Figure 1. The manual microscale column chromatography pressur-ization apparatus.