Understanding radiation units Answer True or False 1.The same amount of radiation falling on the person at level of breast, head or gonads will have the same biological effects. 2.Effective dose can be easily measured. 3.Diagnostic reference levels are not applicable to paediatric radiology. 3
Citation preview
Biophysics Biophysics
Understanding Radiation UnitsUnderstanding Radiation Units
Dr. Yousif Mohamed Y. AbdallahDr. Yousif Mohamed Y. Abdallah
Understanding radiation units
Educational Objectives
At the end of the lecture, the students should become familiar with the following:
• Why is it important to measure radiation dose ?• How radiation dose can and should be expressed?• Understand the radiation quantities and units used
in diagnostic radiology.
2
Understanding radiation units
Answer True or False
1. The same amount of radiation falling on the person at level of breast, head or gonads will have the same biological effects.
2. Effective dose can be easily measured.3. Diagnostic reference levels are not applicable to
paediatric radiology.
3
Contents• Dose descriptors outside the patient’s body.• Dose descriptors for effects that have threshold
(deterministic effects)• Dose descriptors to estimate stochastic risks• Diagnostic reference levels• Dose descriptors and units for staff dose
assessment
Understanding radiation units
Understanding radiation units
Introduction
• Several quantities and units are used in the field of diagnostic radiology to measure and describe radiation dose
• Some can be measured directly while others can only be mathematically estimated
5
6
Two types of radiation effectsStochastic effects
– Where the severity of the result is the same but the probability of occurrence increases with radiation dose, e.g., development of cancer
– There is no threshold for stochastic effects– Examples: cancer, hereditary effects
Deterministic effects– Where the severity depends upon the radiation dose, e.g.,
skin burns– The higher the dose, the greater the effect– There is a threshold for deterministic effects– Examples: skin burns, cataract
Radiation Protection in Paediatric Radiology L02. Understanding radiation unitsUnderstanding radiation units
Hot Coffee – Energy contained in a sip
Excess Temperature = 60º - 37 = 23º1 sip = 3ml3x 23 = 69 calories
Understanding radiation units
Radiation Dose
Lethal Dose= 4GyLD 50/60 = 4 GyFor man of 70 kg
Energy absorbed = 4 x 70 = 280 J= 280/418= 67 calories= 1 sip
Energy content of a sip of coffee if derived in the form of X-rays can be lethal
Understanding radiation units
X-rays
Dose of Radiation
• Radiation energy absorbed by a body per unit mass.
Understanding radiation units 9
Dose Quantities and Radiation units
- Dose quantities external to the patient’s body.
- Dose quantities to estimate risks of skin injuries and effects that have threshold.
- Dose quantities to estimate stochastic risks.
Understanding radiation units
Why so many quantities?Radiation dose is a complex topic
Understanding radiation units
• 1000 Watt heater giving off heat (IR radiation) - unit is of power which is related with emission intensity
• Heat perceived by the person will vary with so many factors: distance, clothing, room temperature
• As can be seen with the example of heat, the energy transformation is a highly complicated issue
• This is the case with X-rays - radiation can’t be perceived
Understanding radiation units
Basic Radiation Quantities
• Used to quantify a beam of X or γ-rays
• There are:– Quantities to express
total amount of radiation.
– Quantities to express radiation at a specific point
12
Radiation at a specific point•Photon fluence•Absorbed dose•Kerma•Dose equivalent
Total radiation•Total photons•Integral dose
Understanding radiation units
Exposure: X• Exposure is a dosimetric quantity for measuring ionizing
electromagnetic radiation (X-rays & Ɣ-rays), based on the ability of the radiation to produce ionization in air.
Units: coulomb/kg (C/kg)
or roentgen (R)
1 R = 0.000258 C/kg
Understanding radiation units
KERMAKERMA (Kinetic Energy Released in a Material):
– Is the sum of the initial kinetic energies of all charged ionizing particles liberated by uncharged ionizing particles in a material of unit mass
– For medical imaging use, KERMA is usually expressed in air
SI unit = joule per kilogram (J/kg) or gray (Gy)
1 J/kg = 1 Gy
Understanding radiation units
Absorbed dose, D, is the mean energy imparted by ionizing radiation to matter per unit mass.
SI unit = joule per kg (J/kg) or gray (Gy).
In diagnostic radiology, KERMA and D are equal.
Absorbed dose: D
Harold Gray
Understanding radiation units
Mean absorbed dose in a tissue or organ
The mean absorbed dose in a tissue or organ DT is the energy deposited in the organ divided by the mass of that organ.
Understanding radiation units
Now things get a little more complicated !
17
Radiation Dose Quantities
• Primary physical quantities are not used directly for dose limitation
• The International Council on Radiation Protection (ICRP) has defined values for dose limits in occupational exposure
Radiation Protection in Paediatric Radiology L02. Understanding radiation unitsUnderstanding radiation units
Radiation Dose Quantities
Equivalent Dose:
• Accounts for the type of radiation• Different radiation types have different level of
biologic damage per unit absorbed dose
Radiation Protection in Paediatric Radiology L02. Understanding radiation unitsUnderstanding radiation units
Radiation Weighting Factors, wR
Radiation type Radiation weighting factor, wR
Photons 1Electrons and muons 1Protons and charged pions 2Alpha particle, fission fragments, heavy ions
20
Neutrons A continuous curveas a function ofneutron energy
(Source: ICRP 103) Radiation Protection in Paediatric Radiology L02. Understanding radiation unitsUnderstanding radiation units
Understanding radiation units
Equivalent Dose : HT,R
The absorbed dose in an organ or tissue multiplied by the relevant radiation weighting factor :
where DT,R is the average absorbed dose in the organ or tissue T, and wR is the radiation weighting factor for radiation R.
RTRRT DwH ,,
Understanding radiation units
Radiation Quantities and Units
Equivalent dose (Unit = sievert, Sv )– Compares the biological effects for
different types of radiation, X-rays, Ɣ-rays, electrons, neutrons, protons, α-particles etc.
– For X-rays, Ɣ-rays, electrons : absorbed dose and equivalent dose have the same value Gy = Sv.
22
Rolph Sievert
Understanding radiation units
Detriment
• Radiation exposure to different organs and tissues in the body results in different probabilities of harm and different levels of severity.
• The combination of probability and severity of harm is called “detriment”.
• Effective dose reflects the combined detriment from stochastic effects due to the equivalent doses in all the organs and tissues of the body.
Understanding radiation units
Effective Dose: ET
• Effective dose takes into account the organ specific radio-sensitivity to develop cancer and hereditary effects from radiation
• Unit = sievert, Sv
24
Understanding radiation units
Effective Dose: ET
A summation of the tissue equivalent doses, each multiplied by the appropriate tissue weighting factor:
where HT is the equivalent dose in tissue T and wT is the tissue weighting factor for tissue T.
T
TTHwE
• The organs have different weighting factors, wT.
• These factors are published in ICRP 103 (2007) and have been changed over the years due to increased knowledge.
Understanding radiation units 26
Tissue Weighting Factors, wT
Radiation Protection in Paediatric Radiology L02. Understanding radiation units 2727
• The weighting factors sum up to 1.0.• They are relative and compares one organ with the
other.• They are the same for children and adults!
Tissue Weighting Factors
Understanding radiation units
Understanding radiation units
• Data is primarily taken from knowledge derived from studying the Japanese population exposed to atomic bombs in Hiroshima and Nagasaki
• On going research has changed the weighting factors from 1990 (ICRP 60) to 2007 (ICRP 103).
28
Tissue Weighting Factors
Understanding radiation units
Multipliers of the equivalent dose to an organ or tissue to account for the different sensitivities to the induction of stochastic effects of radiation.
Tissue Weighting Factors
Tissue weighting
factorwT
*
∑ wT
Bone-marrow (red), Colon, Lung, Stomach, Breast, Remainder Tissues**(nominal weighting factor applied to the average dose to 14 tissues)
0.12 0.72
Gonads 0.08 0.08Bladder, Esophagus, Liver, Thyroid 0.04 0.16Bone surface, Brain, Salivary glands, Skin 0.01 0.04*ICRP 103
**Remainder Tissues (14 in total): Adrenals, Extrathoracic (ET) region, Gall bladder,Heart, Kidneys, Lymphatic nodes, Muscle, Oral mucosa, Pancreas, Prostate, Small intestine, Spleen, Thymus, Uterus/cervix..
Understanding radiation units 3030
Dose to lungs times their weighting factor; DL x wL
+Dose (mean absorbed dose)
to gastrointestinal tract times their weighting factor;
DGI x wGI+
....(summation over organ after organ)
=Effective dose
Effective Dose (E)
where T stands for tissuewhere T stands for tissue
T
TTHwE
Understanding radiation units
Effective Dose (E)
We can compare different paediatric imaging procedures through their different effective doses, E.
31
Understanding radiation units
Radiation Quantities and Units used in Diagnostic Radiology
– Incident air kerma– Entrance surface air kerma– Air kerma-area product– Air kerma-length product– Dosimetric quantities for CT – Dosimetric quantities for interventional radiology
32
Understanding radiation units
Incident Air Kerma
Measured Free in Air on the central beam axis at the focal spot to surface distance.
Only primary beam is considered, that is, no scatter contribution.
Unit: joule/kg or gray (Gy)
Understanding radiation units
Entrance Surface Air Kerma (ESAK)
• ESAK measured on the surface of the patient or phantom where X-ray beam enters the patient or phantom.
• Includes a contribution from photons scattered back from deeper tissues, which is not included in free in air measurements.
Understanding radiation units
Entrance Surface Air Kerma (ESAK)
• If measurements are made at other distances than the true focus - to - skin distance, doses must be corrected by the inverse square law and backscatter factor incorporated into the calculation.
References:– Dosimetry in Diagnostic Radiology: An International code of
practice, TRS 457, IAEA, 2007– Phys. Med. Biol. 43 (1998) 2237-2250.
Understanding radiation units
Kerma in X-ray field can be measured using calibrated: • Ionization chamber
• Semiconductor dosimeter
• Thermoluminescent dosimeter (TLD)
Dose Measurement
36
Understanding radiation units
Kerma-Area Product: KAP
• The kerma - area product (KAP) is defined as the kerma in air in a plane perpendicular to the incident beam axis, integrated over the area of interest.
• This is the dose related quantity measured and displayed on all modern X-ray equipment excluding CT.
KAP meter
Understanding radiation units
Kerma-Area Product: KAP• The KAP (Gy·cm2) is constant
with distance since the cross section of the beam is a quadratic function which cancels the inverse quadratic dependence on dose .
• KAP remains constant along the beam axis as long as it is not measured close to the patient/phantom surface which introduces backscatter.
Understanding radiation units
KAP = K x Area
the SI unit of KAP is the Gy·cm2
Kerma-Area Product: KAP
39
Area = 1Dose = 1
Area = 4Dose = 1/4
d1=1
d2=2
Understanding radiation units
Kerma-Area Product: KAP
40
KAP is independent of distance from the X-ray source, as:
Air Kerma decreases with the inverse square law.
Area increase with the square distance
KAP is usually measured at the level of the tube diaphragms
Area = 1Dose = 1
Area = 4Dose = 1/4
d1=1
d2=2
Understanding radiation units
KAPKAP (kerma(kerma--area product)area product)
UnitUnit: : GyGy··cmcm22This is a picture of a KAP meter which measures the kerma area product
Understanding radiation units 42
Example of a dose display during fluoroscopy or cine
runs with dose rate as shown
Understanding radiation units
In paediatric radiology KAP may be used for:
– Diagnostic reference levels (DRLs)– By use of conversion factors, it can be converted to
skin dose and/or effective dose
43
Kerma-Area Product
Understanding radiation units
Dosimetric Quantities for CT• Computed Tomography Dose
Index (CTDI) - determined using scan protocol
parameters.-useful for comparison of different scanners.
• Dose-Length Product (DLP)- measure of dose to patient- used to estimate effective dose
CT and Risk
Understanding radiation units
KERMA (in phantom)
CTDI (dose in phantom per slice)
Length of scan and pitch
DLP
Effective dose
Risk
Understanding radiation units
Measurement of Dosimetric Quantities in CT
• Pencil ionisation chamber with active length of 100 mm.
• • Measurements free-in-air or in
standard dosimetry phantom.
• Alternatives: TLD, solid state detectors.
• CTDIVOLshould be displayed on the console, reflecting the conditions of operation selected (IEC, 2003)
Dose Indicators in Interventional Radiology
• For quality assurance purposes• To estimate the probability of occurrence of
stochastic effects use:
Kerma-air product rate (KAP, PKA)
Understanding radiation units
Dose Indicators in Interventional Radiology
• For quantifying the threshold and severity of deterministic effects use:
• Maximum skin dose (MSD)• Cumulative dose (CD) to Interventional
Reference Point (IRP)
• In a complex procedure skin dose is highly variable
Understanding radiation units
Understanding radiation units
• In some procedures, patient skin doses approach those used in radiotherapy fractions
• Maximum skin dose (MSD) or peak skin dose is the maximum dose received by a portion of the exposed skin.
Interventional Procedures: Skin Dose
49
Radiodermatitis in the right arm. 7 year-old patient. Photograph taken 4 months after radiofrequency ablation. Surce: ICRP 85
Cumulative Dose to Interventional Reference Point*
• IRP is located 15 cm from the isocentre towards the focal spot
• The air kerma accumulated at a specific point in space relative to the fluoroscopic gantry (IRP) during a procedure
• Cumulative dose does not include tissue backscatter and is measured in Gy.
• Cumulative dose is sometimes referred to as cumulative air kerma
Understanding radiation units
*IRP
Cumulative dose to Interventional Reference Point
Cumulative dose to IRP is measured with a flat ion chamber or calculated by the system and displayed in the angiography room
Understanding radiation units
15 cm
Isocenter
IRP
15 cm
Isocenter
IRP
(IEC-60601-2-43)
MSD vs. Cumulative dose
• In some procedures, cumulative dose to IRP is well correlated with MSD
• Cumulative dose to IRP can be a good indicator of doses higher than the thresholds for skin injures
• A “trigger value” for cumulative dose can be adopted to alert interventionalists the threshold for skin erythema could be reached.
• A follow-up protocol can be adopted.
Understanding radiation units
Understanding radiation units
Other related dose parametersFluoroscopy time:
• Has a weak correlation with KAP• But, in a quality assurance programme it can be
adopted as a starting unit for– comparison between operators, centres,
procedures– for the evaluation of protocol
optimization, and– to evaluate operator skill
53
Understanding radiation units
Other related dose parametersNumber of acquired images and number of series:
– Patient dose is a function of total acquired images
– But dose/image can have big variations– There is an evidence of large variation in
protocols adopted in different centres
54
Diagnostic reference levels (DRLs)
• ICRP, IAEA, EC: introduced the concept of diagnostic reference levels (DRLs) for patients
• DRLs are a form of investigation level, apply to an easily measured quantity at the surface of a simple standard phantom or a representative patient.
• An optimisation tool, not dose limits
Understanding radiation units
Diagnostic Reference Levels (DRLs)
• DRLs calls for local investigation (often very simple) if constantly exceeded
• DRLs: Management of patient doses must be consistent with the required clinical imaging information
Understanding radiation units
Quantities for Establishment of DRLs
• Incident air kerma and entrance-surface air kerma
• Incident air kerma rate and entrance-surface air kerma rate
• Air kerma–area product• CT Dose index, CT Dose–length product
Understanding radiation units
Quantities and Units for Staff Dose Assessment
• Personal dosimetry services typically provide monthly estimates of Hp(10) (mSv), the dose equivalent in soft tissue at 10 mm depth. This is in most of the cases used to estimate the effective dose.
• Sometimes, Hp(0.07) (mSv) is also reported: the dose equivalent in soft tissue at 0.07 mm depth)
• Personal dosememters (film, thermoluminescent...)
Understanding radiation units
Personal Dosimetry Methods
• Single dosimeter worn– above the apron at neck level
(recommended) or under the apron at waist level
• Two dosimeters worn (recommended in intrevational procedures)– one above the apron at neck
level – another under the lead apron
at waist level
Understanding radiation units
Lens dose, optional
Finger dose, optional
Second dosemeter
outside and above the apronat the neck, optional
Personal dose
dosemeter behind the lead apron
X-ray
tube
Image intensifier
Patient
Radiation
protectionmeasures
Dose limits of occupational exposure
(ICRP 60)
Effective dose 20 mSv in a yearaveraged over a period of 5 years
Anual equivalent dose in the lens of the eye 150 mSv
skin 500 mSvhands and feet 500 mSv
Dose Measurement
Dose due to scatter radiation at a point occupied by the operator can be measured with a portable ionization chamber
Understanding radiation units
Understanding radiation units
Summary
• Dosimetric quantities are useful to know the potential hazard from radiation and to determine radiation protection measures to be taken
• Physical quantities - Directly measurable• Protection quantities - Defined for dose limitation
purposes, but not directly measurable.• Application specific quantities - Measurable in medical
imaging. • Diagnostic Refernce Levels
Understanding radiation units
Answer True or False
1. The same amount of radiation falling on the person at level of breast, head or gonads will have same biological effects.
2. Effective dose can be easily measured. 3. Diagnostic reference levels are not applicable to
paediatric radiology.
62
Radiation Protection in Paediatric Radiology L02. Understanding radiation units 6363
Answer True or False
1. False -Different organs have different radio-sensitivity and tissue weighting factors as given by ICRP.
2. False -It can be only calculated using different methods.
3. False - DRLs apply for paediatric radiology, but these are age-specific.
Understanding radiation units
Understanding radiation units
References
• INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS, Patient Dosimetry for X Rays Used in Medical Imaging, ICRU, Rep. 74, ICRU, Bethesda, MD (2006).
• INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Radiological Protection in Medicine, Publication 105, Elsevier, Oxford (2008)
• INTERNATIONAL COMMISSION ON RADIOLOGICAL PROTECTION, Recommendations of the ICRP, Publication 103, Elsevier, Oxford (2008)
• EUROPEAN COMMISSION, Guidance on Diagnostic Reference Levels (DRLs) for Medical Exposure, Radiation Protection 109, Office for Official Publications of the European Communities, Luxembourg (1999)
• INTERNATIONAL ATOMIC ENERGY AGENCY, Dosimetry in Diagnostic Radiology: an International Code of Practice, Technical Report Series No 457, IAEA, Vienna (2007)
Additional information
Understanding radiation units
Quantities for radiation measurement
• Physical quantities - Directly measurable
• Protection quantities - Defined for dose limitation purposes, but not directly measurable
• Application specific quantities - Measurable in medical imaging
Understanding radiation units
Radiation quantities and units• Fundamental dosimetric quantities• Protection quantities
– Equivalent dose– Effective dose
• Application specific dosimetric quantities used in DR– Incident air kerma– Entrance surface air kerma– Air kerma area product– Air kerma length product– Dosimetric quantities in CT and mammography
Understanding radiation units
Physical Quantities
Understanding radiation units
Physical quantities
• Fluence
• Exposure
• Kerma
• Absorbed dose
Understanding radiation units
Fluence : f
The fluence, f , is the quotient of dN by da, where dN is the number of particles incident on a sphere of cross section da, thus
f = dN/da
The unit of fluence is m-2
Understanding radiation units
Exposure: X
where dQ is the absolute value of the total charge of ions produced in air when all the electrons liberated in air of mass dm are completely stopped in air.
The SI unit of exposure is the coulomb per kilogram (C/kg)
The special unit of exposure is the röntgen (R).1R = 2.58 x 10-4 C kg-1
dmdQX
Understanding radiation units
KERMAThe KERMA (Kinetic Energy Released in a MAterial)
where dEtrans is the sum of the initial kinetic
energies of all charged ionizing particles liberatedby uncharged ionizing particles in a material ofmass dm The SI unit of kerma is the joule per kilogram (J/kg),termed gray (Gy).
.
dmdEK trans
Understanding radiation units
Exposure and KERMA
Exposure, X, in units of C kg-1, is related to air kerma as follows:
where W is the average energy spent by an electron to produce an ion pair, g is the fraction of secondary charged particles that is lost to bremsstrahlung radiation production and e is the electronic charge
W
egKX a
1
Understanding radiation units
The fundamental dosimetric quantity absorbed dose, D, is defined as:
where is the mean energy imparted by ionizing radiation to matter in a volume element and dm is the mass of matter in the volume element.
The SI unit of absorbed dose is the joule per kilogram (J/kg), termed the gray (Gy)
In diagnostic radiology, KERMA and D are equal
Absorbed Dose: D
dmdD
d
Understanding radiation units
Exposure and Absorbed Dose or KERMA
• Exposure can be linked to air dose or kerma by suitable conversion coefficients.
• For example, 100 kV X-rays that produce an exposure of 1 R at a point will also give an air kerma of about 8.7 mGy and a tissue kerma of about 9.5 mGy at that point.
Understanding radiation units
Application Specific Quantities
Understanding radiation units
Imaging modality
Measurement subject
Measured radiation quantity
Remark
Radiography PhantomPatient
Incident air kermaESAK, KAP Calculated from X-ray
tube outputFluoroscopy/ Interventional procedures
PhantomPatient
ESAKKAP/Peak skin dose
CT PhantomPatient
CT air kerma indexCT air kerma- length product
Measured in PMMA head and body phantom
Mammogra-phy
PhantomPatient
Incident air kerma, ESAKIncident air kerma
Calculation of mean glandular dose
Dental radiography
Patient Incident air kermaAir kerma-length product
Understanding radiation units
Application Specific Quantities
Patient thickness
Focal-spot to image receptor distance (FFD)
Focal-spot to patient skin distance (FSD)
Incident air kerma (no backscatter)
Entrance surface air kerma (including backscatter)
X-ray tubefocal spot position
Image receptor
Schematic diagram showing some dosimetric and geometric quantities
Entrance Surface Air Kerma (ESAK)
where Y(kVp, FFD) is tube output for actual kVp used during examination, mAs is actual tube current-time product used during examination and FFD is focus-to-film distance. BSF is the backscatter factor that depends on kVp and total filtration of X-rays
Understanding radiation units
BSFtFFD
FDDmAsFDDkVpYESAKp
2
),(
Understanding radiation units
Backscatter Factors (Water)
HVL Field size (cm x cm)
mmAl 10 x 10 15 x 15 20 x 20 25 x 25 30 x 30
2.0 1.26 1.28 1.29 1.30 1.30
2.5 1.28 1.31 1.32 1.33 1.34
3.0 1.30 1.33 1.35 1.36 1.37
4.0 1.32 1.37 1.39 1.40 1.41
Understanding radiation units
• If the KAP is calculated by the system, you must know if the user added filtration you use is included or not !
81
Kerma-Area Product: KAP
Understanding radiation units
Kerma-Area Product: KAP
• It is always necessary to calibrate and to check the transmission chamber for the X-ray installation in use
• In some European countries, it is compulsory that new equipment is equipped with an integrated ionization transmission chamber or with automatic calculation methods
Understanding radiation units
Dosimetric Quantities for CT
• Computed Tomography Dose Index (CTDI)
• CT air kerma index
• Dose-Length Product (DLP)• Air kerma-length product
Understanding radiation units
ICRU 74 / IAEA TRS 457
• CT air kerma index– Free-in-air (Ck)– In phantom (Ck,PMMA)
• Air kerma length product (PKA)
Understanding radiation units
Dosimetric Quantities for CTPrincipal dosimetric quantity in CT is CT air kerma index:
where K(z) is air kerma along a line parallel to the axis of rotation of the scanner over a length of 100 mm.
N = Number of detectors in multi-slice CT T = Individual detector dimension along z-dimension
The product NT defines the nominal scan beam width/collimation for a given protocol.
50
50100, )(1 dzzK
NTCa
Understanding radiation units
Dosimetric Quantities for CT
Weighted CT air kerma index, CW, combines values of CPMMA,100 measured at the centre and periphery of a standard CT dosimetry phantoms
pPMMAcPMMAw CCC ,100,,100, 231
Understanding radiation units
Dosimetric Quantities for CT
Pitch (IEC, 2003):
T= Single detector dimension along z-axis in mm.
N=Number of detectors used in a given scan protocol (N>1 for MDCT), N x T is total detector acquisition width or collimation
I=table travel per rotation
Radiographic, 2002, 22:949-62
NTIp
Understanding radiation units
• Volume CTDI describes the average dose over the total volume scanned in sequential or helical sequence, taking into account gaps and overlaps of dose profiles (IEC, 2003):
• Average dose over x, y and z direction• Protocol-specific information
Dosimetric Quantities for CT
lNTCC WVOL
Understanding radiation units
Dosimetric Quantities for CT
• Kerma-length product (PKL):
where L is scan length is limited by outer margins of the exposed scan range (irrespective to pitch)
• PKL for different sequences are additive if refer to the same type of phantom (head/body)
LCP VOLKL
Maximum Skin Dose (MSD)
• Measurement/evaluation of MSD• Point or area detectors • Cumulative dose at IRP (interventional radiology point)• Calculation from technical data
• Off line methods• Area detectors: TLD array, slow films, radiochromic
films• From KAP and Cumulative dose measurement
Understanding radiation units
Understanding radiation units
Method for MSD Evaluation: Radiochromic Large Area Detector
Example: Radiochromic films type Gafchromic XR R 14”x17”• useful dose range: 0.1-15 Gy • minimal photon energy dependence (60 - 120 keV)• acquisition with a flatbed scanner:b/w image, 12-16
bit/pixel or, measure of OD measurement with a reflection densitometer
Benefits of Radiochromic Films• The radiochromic film:
– displays the maximum dose and its location– shows how the total dose is distributed– provides a quantitative record for patient files– provides physician with guidance to enable safe planning of
future fluoroscopically guided procedures– improves fluoroscopic technique and patient safety– possible rapid semi-quantitative evaluation
Understanding radiation units
Example of an exposed radiochromic film in a cardiac interventional procedure
Rapid Semi-Quantitative Evaluation: Example
• For each batch number (lot #) of gafchromic film a Comparison Tablet is provided
• In the reported example we easily can recognise that the darkness area of the film, corresponding to the skin area that has received the maximum local dose, has an Optical Density that correspond at about 4 Gy
Understanding radiation units
Understanding radiation units
DRLs for Complex Procedures
3rd level “Patient risk”
2nd level “Clinical protocol”
1st level“Equipment
performance”
94
Dose rate and dose/image(BSS, CDRH, AAPM)
Level 1 + No. images + fluoroscopy
time
Level 2 + DAP + Peak Skin Dose (MSD)
Reference levels (indicative of the state of the practice): to help operators to conduct optimized procedures with reference to patient exposure
For complex procedures reference levels should include: • more parameters • and, must take into account the complexity of the procedures.(European Dimond Consortium recommendations)