INFUSION PUMPS
• Prepared for:
Clinical Engineering Professional
Plymouth HospitalSeptember 2, 2011
BiomedicalBenchmark™
Technology Maintenance Report
8/10/2019 BiomedicalBenchmark Technology Maintenance Report Sample Infusion Pumps
http://slidepdf.com/reader/full/biomedicalbenchmark-technology-maintenance-report-sample-infusion-pumps 2/11
© 2011 ECRI Institute
Executive Summary—Infusion Pumps
You’re doing well here:
Plymouth Hospital’s IPM Intervals are equal to or exceed the BB
average.
Repairs are equal to or less frequent than the BB average for allpump categories.
Issues to consider:
Opportunities to reduce IPM and Repair time.
Model 6201 exceeds typical expected life and Model 6301 isentering the range for the “end of life”.
Six models of your pumps have been discontinued.
8/10/2019 BiomedicalBenchmark Technology Maintenance Report Sample Infusion Pumps
http://slidepdf.com/reader/full/biomedicalbenchmark-technology-maintenance-report-sample-infusion-pumps 3/11
8/10/2019 BiomedicalBenchmark Technology Maintenance Report Sample Infusion Pumps
http://slidepdf.com/reader/full/biomedicalbenchmark-technology-maintenance-report-sample-infusion-pumps 4/11
8/10/2019 BiomedicalBenchmark Technology Maintenance Report Sample Infusion Pumps
http://slidepdf.com/reader/full/biomedicalbenchmark-technology-maintenance-report-sample-infusion-pumps 5/11
© 2011 ECRI Institute
You’re doing well here
Your IPM interval of 24 months is longer than the 12-monthinterval most frequently reported for the 6201 and 6301 infusion
pumps and their repair rate is the same as your colleagues.
The IPM intervals and times for 6101, Colleague 3CX, 2001 and
2010 are consistent with the BB data.
Repairs are less frequent than or equal to the BB average for allpump categories.
8/10/2019 BiomedicalBenchmark Technology Maintenance Report Sample Infusion Pumps
http://slidepdf.com/reader/full/biomedicalbenchmark-technology-maintenance-report-sample-infusion-pumps 6/11
© 2011 ECRI Institute
Issues to consider
IPM times for 6201 and 6301 are higher than the BB average.You may want to double check your procedures as you may be
able to save over ½ hour for each model.
Your average repair times are significantly higher for all of yourmodels when compared to BB data.
BB average times should reflect working time only (not travel
time or wait time). If your time recording procedures reflectthose other time categories, your averages will be higher than
the BB average.
8/10/2019 BiomedicalBenchmark Technology Maintenance Report Sample Infusion Pumps
http://slidepdf.com/reader/full/biomedicalbenchmark-technology-maintenance-report-sample-infusion-pumps 7/11
© 2011 ECRI Institute
Issues to consider
Model 6201 has exceeded and Model 6301 is entering the
range for the typical expected life. If they meet your clinical
needs, they do not have to be replaced only for this reason. Ifyou replace them, ECRI Institute recommends using pumps with
Dose Error Reduction Systems.
8/10/2019 BiomedicalBenchmark Technology Maintenance Report Sample Infusion Pumps
http://slidepdf.com/reader/full/biomedicalbenchmark-technology-maintenance-report-sample-infusion-pumps 8/11
© 2011 ECRI Institute
Issues to consider
The following models appear on your inventory and on our
discontinued devices list:
--Flo-Gard 6201, 6301
--Colleague CX3 (your listing is “Colleague 3CX”)
--Med Fusion 2001, 2010, 3500
Check with each manufacturer to ensure that you will have
access to parts and service for discontinued devices or identify
alternative sources for parts and service, if required.
8/10/2019 BiomedicalBenchmark Technology Maintenance Report Sample Infusion Pumps
http://slidepdf.com/reader/full/biomedicalbenchmark-technology-maintenance-report-sample-infusion-pumps 9/11
© 2011 ECRI Institute
BiomedicalBenchmark™ Recommendations
Inspection and Preventive Maintenance:Review your current procedures and compare them to ECRI
Generic Procedures and User Contributed Procedures in BB.
Consider the use of automatic pump testers if you are not
already using them.
Repair:
Determine the amount of repair time that is associated with
battery failures and replacement.
8/10/2019 BiomedicalBenchmark Technology Maintenance Report Sample Infusion Pumps
http://slidepdf.com/reader/full/biomedicalbenchmark-technology-maintenance-report-sample-infusion-pumps 10/11
© 2011 ECRI Institute
Potential annual savings-Infusion Pumps
Infusion pumps represent less than 5% of your inventory;analysis of other device categories may lead to further savings.
Category Savings
IPM Time 35 hours
Repair Time 150 hours
Total 185 hours