Upload
chao-yang
View
216
Download
1
Embed Size (px)
Citation preview
Biocompatibility of a physiological pressure sensor
Chao Yang a, Chunfeng Zhao a, Lester Wold b, Kenton R. Kaufman a,*a Orthopedic Biomechanics Laboratory, Mayo Clinic, 200 First Street, SW, Rochester, MN 55905, USA
b Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN 55905, USA
Received 18 July 2002; received in revised form 26 February 2003; accepted 11 March 2003
Biosensors and Bioelectronics 19 (2003) 51�/58
www.elsevier.com/locate/bios
Abstract
A newly developed fiber optic micropressure sensor was evaluated for biocompatibility using the International Organization for
Standardization (ISO) test standard 10993-6. The test material and an inert control (fused silica glass) were tested in New Zealand
white rabbits. Four test specimens were implanted in the paravertebral muscles on one side of the spine about 2�/5 cm from the mid-
line and parallel to the spinal column. Similarly, four control specimens were implanted on the opposite side. The implantation
periods were 1, 4, and 12 weeks to ensure a steady state biological tissue response. Four animals were tested at each time period.
Macroscopic and microscopic observations were performed to compare the biological reactions between the test and control
materials. There was an inflammatory reaction at 1 week which subsided at 4 weeks. There was fibrous tissue growth near the
implant that also decreased over time. Most importantly, there was no significant difference in the biological response between the
test and control materials. Therefore, we conclude that the pressure microsensor is biocompatible.
# 2003 Published by Elsevier Science B.V.
Keywords: Biocompatibility; Silica glass; Fiber optic pressure sensor; Intramuscular pressure; Tissue response
1. Introduction
Measurements of biological fluid pressure are of
interest to clinicians in cardiology, pulmonology, gas-
troenterology, urology, neurology and rehabilitation.
Commercially available pressure transducers are too
large for optimum comfort (Crenshaw et al., 1992).
Microsensor technology is now available to construct
transducers that are much smaller. A fiber optic micro-
sensor has been developed for measurement of physio-
logical pressures (Kaufman et al., 2002). The
measurement method is based on an extrinsic Fabry�/
Perot interferometric (EFPI) technique (Murphy et al.,
1991). EFPI technology is a distance measurement
technique based on the formation of a low-finesse
Fabry�/Perot cavity between the polished end of a fiber
and a reflective surface, which is the undersurface of a
diaphragm (Fig. 1). Light is passed through the fiber,
where a portion of the light is reflected off the fiber/air
interface (R1). The remaining light propagates through
the air gap between the fiber and the reflective surface
and is reflected back into the fiber (R2). R1 is the
reference reflection while R2 is the sensing reflection.
Fluid pressure causes displacement of the diaphragm.
These two light waves interfere constructively or
destructively based on the path light difference traversed
by the sensing reflection relative to the reference
reflection, and travel back through the single mode
fiber to the demodulation unit. The return signal is
analyzed with a miniature spectrometer.
The sensor consists of an optical fiber and a dia-
phragm. The optical fiber is made with inert silica glass.
The diaphragm is made from polyimide. These materials
have been carefully selected with biocompatibility issues
in mind. However, biocompatibility testing is a neces-
sary step since this device will be used for measurement
of in-vivo pressures (Black, 1999; ISO 1992�/1994).
Although the main part of this sensor is made of inert
silica glass, the construction of this sensor requires a
special adhesive to attach the diaphragm and optical
fiber together. Therefore, the objective of this study was
to test the in-vivo host response for this sensor.
* Corresponding author. Tel.: �/1-507-284-2262; fax: �/1-507-266-
2227.
E-mail address: [email protected] (K.R. Kaufman).
0956-5663/03/$ - see front matter # 2003 Published by Elsevier Science B.V.
doi:10.1016/S0956-5663(03)00131-3
2. Materials and methods
The International Organization for Standardization
(ISO) test 10993-6, ‘Biological Evaluation of Medical
Devices-Part 6: Tests for local effects after implanta-
tion’, was followed in this test protocol (ISO 1992�/
1994). The test sensor was composed of fused silica
glass, polyimide, and an aliphatic amine-cured mineral
filled epoxy adhesive. The control material was totally
composed of inert silica glass (both of these materials
were provided by Luna Innovations, Blacksburg, VA).
The test material and control were a flat-shaped
cylinder, 10 mm in length and 360 mm in diameter
(Fig. 2).
Twelve healthy adult New Zealand white rabbits were
tested. Their body weight ranged from 3.0�/4.0 kg
(average 3.2 kg). The animals were divided into three
implantation periods (1, 4, and 12 weeks), with four
animals tested at each time period. The Mayo Institu-
tional Animal Use and Care Committee (IACUC)
approved all animal procedures. NIH guidelines for
the care and use of laboratory animals (NIH Publication
#85-23 Rev. 1985) have been observed. The persons
who performed the surgical procedure were trained to
obey animal welfare and handling protocols. The rabbits
were anesthetized using Ketamine (30�/50 mg/kg) and
then placed in sternal recumbency. The hair over the
dorsum, from mid-thorax to caudal lumbar region was
shaved. The shaved region then received a surgical
preparation with Betadine scrub and then Betadine
solution. The tip of the spinous process was palpated
to identify the lumbar spine (the space between the
lumbar spinous process is larger than that between the
thoracic spinous process). The sterile test sensors and
controls were percutaneously inserted at eight sites
along the lumbar paraspinal muscles (four on each
side). The implant was injected through an 18 gauge, 51
Fig. 1. Block diagram of the low-frequency, high resolution, fiber optic pressure measurement system. Light is conveyed via a fiber optic cable to the
sensor. Fluid pressure causes a diaphragm to deflect. The light is reflected back at the fiber optic interface and the sensor diaphragm. The returned
light is analyzed with a miniature spectrometer to determine the relative displacement of the diaphragm. The graphs represent the light spectrum that
is returned from the sensor before it is processed to determine the gap measurement.
Fig. 2. Configuration of test and control samples used in biocompat-
ibility tests.
C. Yang et al. / Biosensors and Bioelectronics 19 (2003) 51�/5852
mm I.V. needle (Terumo Corporation Tokyo, Japan) at
a 458 into the muscle, with the needle tip about 1.5 cm
beneath the skin. The implant was advanced using the
needle until the implant exited the tip of the I.V. needle.The implantation sites were recorded in reference to the
nearest vertebra. The injection sites were then covered
with sterile tissue. After the implantations were com-
plete, the animal was given a dose of Buprenorphine
(0.02�/0.05 mg/kg by I.V. means) for pain relief and
taken off the inhalation anesthesia. The animals recov-
ered from surgery on a warming pad and were placed in
their cages when fully recovered from the anesthesia.The rabbits were observed daily during the first post-
operative week and weekly thereafter. The observation
included body temperature, posture, activity, appetite,
and sensitivity to touch. If needed, analgesics were
administrated.
At 1, 4, and 12 weeks after implantation the animals
was anesthetized with Ketamine or Xylazine and
euthanized with Pentobarbital (�/100 mg/kg by I.V.).The rabbits were placed in a prone position. The hairs
over the dorsum, from mid-thorax to caudal lumbar
region, were shaved again. Any skin changes near the
implantation site were recorded. A vertical incision was
made along the dorsum midline. After exposure of the
paraspinal muscle, the nature and extent of any tissue
reaction was observed with the help of a low magnifica-
tion lens. The whole longissimus lumborum was dis-sected and wrapped with physiological saline-soaked
gauze, and fixed with 10% formalin solution for at least
48 h. The implants were found by sectioning the tissue
until the implant was identified. Approximately 1 cm3 of
paraspinal muscle at each implantation site was then
isolated and sent for histological preparation.
All specimens were formalin fixed and processed
routinely resulting in the tissue being embedded inparaffin. The slides were cut from the paraffin blocks
perpendicular to the long axis of the implant. The
section orientation in relation to the implant dimensions
and implant orientation was recorded.
The host response to the implant was determined by
measurement of the distance from the implant/tissue
interface to unaffected areas, i.e. normal tissue and
normal vascularity. The biological response parametersassessed and recorded included the: (a) extent of
fibrosis/fibrous capsule and inflammation; (b) degenera-
tion determined by changes in tissue morphology; (c)
number and distribution of inflammatory cells types,
namely macrophages, polymorphonuclear leukocytes,
lymphocytes, plasma cells, giant cells, as a function of
distance from the material/tissue interface; (d) presence
of necrosis determined by nuclear debris and/or capil-lary wall breakdown; (e) and other parameters such as
material debris, fatty infiltration and presence of
granulomata. All analyses were done with a digital
image analysis system (Microscope: Axioplan2; camera:
Axiocam; software: KS400. Carlzeiss Inc. Oberkochen
Germany).
A two-sample t-test was used to test for significant
differences between the two groups at each time period.The dependent variables were thickness of reaction
capsule and density of inflammatory cells. Statistical
significance was set at P�/0.05.
3. Result
3.1. General and macroscopic observations
All 12 rabbits recovered well after the surgery. All
grossly showed swelling around the insertion area on the
1st post-operative day. The swelling gradually subsided
by the 2nd day post-operatively. From the 2nd post-
operative day to the day of sacrificing, no abnormal
conditions, i.e. temperature, eating, and activity, were
observed. During post-mortem dissection, no gross
abnormalities were identified in the skin or the hostmuscle. No local swelling or fluid accumulation was
found around the implant sites and no suppurative
changes were identified. No infection occurred. For all
rabbits, the insertion hole could still be found on the
muscle fascia for all implants after 1, 4 and 12 weeks
(Fig. 3).
3.2. Cell responses and histology
Although there were no significant regimes of necro-
sis, granuloma formation, or soft tissue calcification
around any of the implants, apoptosis could occasion-
ally be seen in stromal cells in the 1 week test group.
However, the number of such observations was small
and did not consistently occur. For the 1 week group,
the inflammatory reaction was mild (Fig. 4). Monocytes,macrophages and eosinophils were the main inflamma-
tory cell types present for both test and control groups.
The capsule wall was composed of three layers. The
inner layer was mostly composed of macrophages and
monocytes. Eosinophils were located in the middle
layer. The outer layer was mainly composed of fibro-
blasts. There was no significant difference in inflamma-
tory cell composition or capsule thickness between thetest and control groups (Figs. 5�/7). At 1 week the mean
capsule thickness of test and control groups were 90.0
and 94.7 mm, respectively (t�/0.755, P �/0.05). The
mean density of macrophages and monocytes were 5479
and 5610/mm2, for the test and control groups, respec-
tively (t�/0.218, P �/0.05). The mean density of eosi-
nophils was 1655 and 1619/mm2, respectively (t�/0.201,
P �/0.05). The density of fibroblasts was 1621 and 1938/mm2, respectively (t�/1.409, P �/0.05).
For the week-4 group, the capsule thickness decreased
sharply. Unlike the week-1 sample, the capsule consisted
C. Yang et al. / Biosensors and Bioelectronics 19 (2003) 51�/58 53
Fig. 3. Comparison of insertion sites for test and control materials at 1, 4 and 12 weeks after implantation. The small box in each figure indicates the
region shown in Fig. 4.
Fig. 4. Close-up view of test and control sites at 1, 4 and 12 weeks after implantation. There was no significant difference in thickness of the fibrous
capsule or type and density of cellular response.
C. Yang et al. / Biosensors and Bioelectronics 19 (2003) 51�/5854
of two clearly distinctive layers (Fig. 4). The inner layer
was an inflammatory cell layer. The cell composition
was mainly macrophages and round cells. Lymphocytes
could be occasionally noticed at this time. Eosinophils
decreased. The outer layer was a thin fibrous layer
formed by fibroblasts and collagen fibers. The mean
capsule thickness for the test group (47.2 mm) and
control group (47.3 mm) was not significantly different
(t�/0.51, P �/0.05) (Fig. 5). The macrophages and otherinflammatory cell decreased at 4 weeks, so we counted
all the inflammatory cells together. The mean density of
inflammatory cells were 3243 and 3330/mm2, for the test
and control groups, respectively (t�/0.089, P �/0.05)
(Fig. 6). The mean density of fibroblasts was 4211 and
4655/mm2, respectively (t�/0.161, P �/0.05) (Fig. 7).
From 4 to 12 week, the thickness of the capsule
decreased slightly. The mean capsule thickness for thetest group was 30.0 mm, and the mean thickness of the
control group was 29.0 mm (Fig. 5). There was no
significantly difference between the groups (t�/0.249,
P �/0.05). The capsule was composed of 2�/3 layers of
cells, mainly fibroblasts. Macrophages were present but
decreased. The mean density of inflammatory cells was
2625 and 2500/mm2, for the test and control groups,
respectively (t�/0.238, P �/0.05) (Fig. 6). The meandensity of fibroblasts was 3970 and 3879/mm2, respec-
tively (t�/0.078, P �/0.05) (Fig. 7).
4. Discussion
This biocompatibility test focused on the direct
interactions between the substance of the implant andthe chemical and biological aspects of the host environ-
ment. Histologic and morphologic in-vivo examinations
are well-accepted methods to assess biocompatibility
giving reliable unequivocal results. Standards that
govern these types of tests have been detailed by various
federal and international agencies (ASTM, 1997; ISO
1992�/1994). The accepted international standard for
biocompatibility testing is ISO 10993, which is thestandard used in this study (ISO 1992�/1994).
This study followed the ISO 10993-6 standard with
respect to the type of animal used, study period, implant
site and surgical procedure (ISO 1992�/1994). However,
the dimension of the implant was 10 mm by 360 mm,
which was slightly thinner than the ISO 10993 recom-
mendation. We chose this sized implant as it reflects the
actual full-sized implant that would be used in futureclinic practice. In addition, the implant size allowed us
to insert it through a needle probably resulting in a
decrease in the acute inflammatory response as com-
pared to an open surgery procedure. The sensor will
probably always be inserted with a needle in clinical
practice. We chose silica glass as the negative control
since this material is similar to our test material. In
addition, silica is a well-accepted inert material used as acontrol in many previous studies (Hakkinen et al., 1988;
Ito et al., 1987b; Kobayashi et al., 1997a, 1999, 1997b;
Matsuda et al., 1987; Seitz et al., 1982).
Fig. 5. Capsule thickness surrounding test and control implants at 1, 4
and 12 weeks after implantation. There was no significant difference in
capsule thickness between test and control at each time point studied.
Fig. 6. Cellular density of inflammatory cells in the reaction zone.
There was no significant difference in cellular response surrounding the
test and control implants at each time period.
Fig. 7. Cellular density of fibroblasts in the reaction zone throughout
the test period. There was no significant difference in cellular density
between the test and control implants at each time period.
C. Yang et al. / Biosensors and Bioelectronics 19 (2003) 51�/58 55
Biocompatibility testing can be done using sensitive
in-vitro cell culture techniques. These tests can serve as
precursors for more involved, more costly, and more
time-consuming animal biocompatibility tests. We chosean in-vivo animal test because the morphology and
function of cells in-vivo depend to a degree on the
chemistry of the pericellular environment and the nature
of the substrate supporting and surrounding the cells.
Lack of these environmental factors may cause cells to
become inactive or to revert, at least functionally (Black,
1999). Tissue culture experiments are considered by
many to be too sensitive if used alone since failure inthese tests would eliminate many useful materials that
are presently in use. Effects on division, growth, and
function of cells in culture can be easily and quantita-
tively estimated. But such tests should not be considered
‘fail safe’. A material that displays some toxic signs in
culture may yet prove acceptable because of the capacity
of the whole animal to buffer local effects. In the present
study, an in-vivo test was adopted to circumvent someof these problems (Wilson et al., 1981).
The in-vivo biocompatibility of a material can be
evaluated by analyzing the cell population present,
measuring the mediator and metabolite cells excreted,
and analyzing the morphologic characteristics of the
tissue and the capsule thickness around the implant.
Different methods or scoring systems have been used to
evaluate the host response (Wood et al., 1970; Boswaldet al., 1999; Ito et al., 1987a; Wilson et al., 1981;
Schmalz and Schmalz, 1981; Haggerty and Lusted,
1989; Mathur et al., 1997; Allen et al., 2001; Woodward
and Salthouse, 1986). No one system seems to be well
accepted. In our study, the average capsule thickness
decreased sharply from the 1st week to the 4th week for
both the sensor and control, indicating that the tissue
went from an acute, insertion procedure related inflam-matory phase to a relatively stable phase. At week 1,
macrophages and a small amount of monocytes were at
the inner layer. Their number was much higher at 1
week compared to 4 and 12 weeks. Woodward and
Salthouse observed similar changes (Woodward and
Salthouse, 1986). They pointed out that activated
macrophages will be observed at the surface of all
implant biomaterials, but their numbers will be propor-tional to the toxicity of the implant material. With
smooth-walled, nontoxic materials, the macrophage
population will be largely replaced by fibroblasts and
later by a fibrous capsule (Woodward and Salthouse,
1986). This is the type of cellular cascade we observed. It
was very important to cut the samples perpendicular to
the long axis of the implant. This way we could measure
the real thickness of the capsule. Yet, some samples werenot sectioned ideally. In these cases, we measured the
thinnest part of the capsule, which represented the real
thickness of the capsule. The area chosen for cell
counting was also important. We chose a rectangular
area with the long axis perpendicular to the capsule.
This area included all the layers of the reaction zones
from the inner part to outer part. This avoided
inappropriate sampling of a particular area and omis-sion of another area within the reaction zone. This
method ensured that every cell type was counted in the
evaluation.
Measuring the capsule membrane thickness around
the implant is a basic and important tool for estimating
biocompatibility. Investigators have used different mea-
suring techniques (Ellies et al., 1988; Christel and
Meunier, 1989; Benghuzzi, 1996; Ryhanen et al.,1998), but the results show similar outcomes. In our
study, the capsule thickness dropped sharply from 1 to 4
weeks, and slowly to 12 weeks. Even at 12 weeks, the
capsule was still about 30 mm for the test group and 29
mm for control group. This supports the contention that
after 4 weeks, no new chemicals were released from the
implant, and the tissue underwent a stable inflammation
process. As Woodward pointed out, the interfacebetween the host and the foreign body is permanent,
and the cells and collagen participating in it turn over
extremely slowly after the interface response is estab-
lished (Woodward and Salthouse, 1986). This was
confirmed by Ryhanen et al. (1998), where after 26
weeks there was still a thin layer of fibrous capsule
around the implant.
In this study, although the difference in number ofinflammatory cells between the two groups was not
significant, the cell numbers of the control group
appeared to be higher than the test group at 1 and 4
weeks. For the 1 week group, some apoptosis was
noticed in the test group, which is probably due to
chemical release. Since the amount of apoptosis was
very small, it could not be used for statistical analysis.
More importantly, there was no significant difference inthe inflammatory process of the two groups at either 4
or 12 weeks.
The cells at the interface were mainly macrophages,
eosinophils, and fibroblasts. Other monocytes, giant
cells and lymphophils were only occasionally present.
This cellular distribution has been reported previously
(Wood et al., 1970; Wilson et al., 1981; Christel and
Meunier, 1989; Mathur et al., 1997; Ryhanen et al.,1998). Polymorphonuclear granulocytes migrate into the
damaged tissues, where their phagocytic activity re-
moves tissue debris, deleterious foreign material, and
pathogens. Like some other studies, most of the poly-
morphonuclear cells present in our case were eosinophils
but not neutrophils. This indicates that no infection was
present. Although eosinophils were present at the
implant-tissue interface early, the eosinophils tended todisappear by the 12 weeks observations.
Monocytes migrated to the interface where they
assumed the role of macrophages. Macrophages play a
very important role in acute inflammation and probably
C. Yang et al. / Biosensors and Bioelectronics 19 (2003) 51�/5856
in final biocompatibility. They can release mediators,
which can in turn activate other cells. Macrophages are
found to affect fibroblast and lymphocyte activity. The
amounts of macrophages present at the different timeperiods did not differ between the test and control
material. Very few giant cells were identified, suggesting
that the amount of ‘foreign body debris’ was small.
Fibroblasts migrated to the implant-tissue interface in
the early phase of healing and gathered around the
implants. Other granulation tissue was slowly replaced
by fibroblast proliferation and collagen deposition. In
optimal situations, an inert biomaterial forms a thin,relatively avascular and acellular fibrous scar capsule at
the tissue�/plant interface. This kind of capsule was
similar to what we saw in our 12 weeks group.
The basic biomaterial�/cell interaction is still poorly
understood. If a material is clearly toxic, it induces a
strong local response. On the other hand, if there are
toxic elements in the material which dissolve slowly,
they might give symptoms many years after the implantis inserted. Thus, the results of a single study are only
suggestive in nature. However, our test material was
predominantly composed of materials that are known to
be inert. Therefore, we do not anticipate that a longer
duration study will change the cell response documented
in this study.
5. Summary
Through our study, we conclude that this microsensoris biocompatible. The microsensor has a bio-reaction
which was histopathologically similar to the inert silica
glass in term of inflammatory migration and fibro-
capsule formation. Based on these results, this device
can be used for in-vivo human applications.
Acknowledgements
This study was supported by NIH grant R01H03476.We thank James E. Tarara for his help with the semi-
quantitative image analysis. Appreciation is also ex-
pressed to Tom Wavering and Luna Innovations for
their collaboration on the design, development and
manufacturing of this pressure microsensor.
References
ISO 1992�/1994. ISO 10993.
Allen, M., Myer, B., Rushton, N., 2001. In vitro and in vivo
investigations into the biocompatibility of diamond-like carbon
(DLC) coatings for orthopedic applications. Journal of Biomedical
Materials Research 58 (3), 319�/328.
ASTM, 1997. vol. 13.01. ASTM, West Conshohocken, PA.
Benghuzzi, H., 1996. Cytological evaluation of capsular tissue
surrounding TCPL implant in adult rats. Biomedical Sciences
Instrumentation 32, 81�/86.
Black, J., 1999. In: Black, J. (Ed.), Biological Performance of
Material�/Fundamentals of biocompatibility. Marcel Dekker Inc,
New York, Basel, pp. 321�/453.
Boswald, M., Mende, K., Bernschneider, W., Bonakdar, S., Ruder, H.,
Kissler, H., Sieber, E., Guggenbichler, J.P., 1999. Biocompatibility
testing of a new silver-impregnated catheter in vivo. Infection 27
(Suppl. 1), S38�/S42.
Christel, P., Meunier, A., 1989. A histomorphometric comparison of
the muscular tissue reactions to high-density polyethylene in rats
and rabbits. Journal of Biomedical Materials Research 23 (10),
1169�/1182.
Crenshaw, A.G., Styf, J.R., Hargens, A.R., 1992. Intramuscular
pressures during exercise: an evaluation of a fiber optic transdu-
cer-tipped catheter system. European Journal of Applied Physiol-
ogy and Occupational Physiology 65 (2), 178�/182.
Ellies, L.G., Carter, J.M., Natiella, J.R., Featherstone, J.D., Nelson,
D.G., 1988. Quantitative analysis of early in vivo tissue response to
synthetic apatite implants. Journal of Biomedical Materials Re-
search 22 (2), 137�/148.
Haggerty, H.S., Lusted, H.S., 1989. Histological reaction to polyimide
films in the cochlea. Acta Oto-Laryngologica 107 (1�/2), 13�/22.
Hakkinen, L., Yli-Urpo, A., Heino, J., Larjava, H., 1988. Attachment
and spreading of human gingivival fibroblasts on potentially
bioactive glasses in vitro. Journal of Biomedical Materials Re-
search 22 (11), 1042�/1059.
Ito, G., Matsuda, T., Inoue, N., Kamegai, T., 1987a. A histological
comparison of the tissue interface of bioglass and silica glass.
Journal of Biomedical Materials Research 21 (4), 485�/497.
Ito, G., Matsuda, T., Inoue, N., Kamegai, T., 1987b. A histological
comparison of the tissue interface of bioglass and silica glass.
Journal of Biomedical Materials Research 21 (4), 485�/497.
Kaufman, K.R., Wavering, T., Morrow, D., Davis, J., Lieber, R.L.,
2002. Performance characteristics of a pressure microsensor.
Journal of Biomechanics, in press.
Kobayashi, M., Kikutani, T., Kokubo, T., Nakamura, T., 1997a.
Direct bone formation on alumina bead composite. Journal of
Biomedical Materials Research 37 (4), 554�/565.
Kobayashi, M., Nakamura, T., Shinzato, S., Mousa, W.F., Nishio, K.,
Ohsawa, K., Kokubo, T., Kikutani, T., 1999. Effect of bioactive
filler content on mechanical properties and osteoconductivity of
bioactive bone cement. Journal of Biomedical Materials Research
46 (4), 447�/457.
Kobayashi, M., Nakamura, T., Tamura, J., Iida, H., Fujita, H.,
Kokubo, T., Kikutani, T., 1997b. Mechanical and biological
properties of bioactive bone cement containing silical glass powder.
Journal of Biomedical Materials Research 37 (1), 68�/80.
Mathur, A.B., Collier, T.O., Kao, W.J., Wiggins, M., Schubert, M.A.,
Hiltner, A., Anderson, J.M., 1997. In vivo biocompatibility and
biostability of modified polyurethanes. Journal of Biomedical
Materials Research 36 (2), 246�/257.
Matsuda, T., Yamauchi, K., Ito, G., 1987. The influence of bioglass on
the growth of fibroblasts. Journal of Biomedical Materials
Research 21 (4), 499�/507.
Murphy, K.A., Gunther, M.F., Vengsarkar, A.M., Claus, R.O., 1991.
Quadrature phase-shifted, extrinsic Fabry�/Perot optical fiber
sensors. Optics Letters 16 (4), 273.
Ryhanen, J., Kallioinen, M., Tuukkanen, J., Junila, J., Niemela, E.,
Sandvik, P., Serlo, W., 1998. In vivo biocompatibility evaluation of
nickel�/titanium shape memory metal alloy: muscle and perineural
tissue responses and encapsule membrane thickness. Journal of
Biomedical Materials Research 41 (3), 481�/488.
Schmalz, G., Schmalz, C., 1981. Toxicity tests on dental filling
materials. International Dental Journal 31 (3), 185�/192.
C. Yang et al. / Biosensors and Bioelectronics 19 (2003) 51�/58 57
Seitz, T.L., Noonan, K.D., Hench, L.L., Noonan, N.E., 1982. Effect of
fibronectin on the adhesion of an established cell line to a surface
reactive biomaterial. Journal of Biomedical Materials Research 16
(3), 195�/207.
Wilson, J., Pigott, G.H., Schoen, F.J., Hench, L.L., 1981. Toxicology
and biocompatibility of bioglasses. Journal of Biomedical Materi-
als Research 15 (6), 805�/817.
Wood, N.K., Kaminski, E.J., Oglesby, R.J., 1970. The significance of
implant shape in experimental testing of biological materials: disc
vs. rod. Journal of Biomedical Materials Research 4 (1), 1�/12.
Woodward, S.C., Salthouse, T.N., 1986. In: Recum, A.F.V. (Ed.),
Handbook of Biomaterials Evaluation. Macmillan Publishing
Company, New York, pp. 364�/377.
C. Yang et al. / Biosensors and Bioelectronics 19 (2003) 51�/5858