All the back-calculated standard concentrations for both human insulin and insulin detemir in pure solutions were within 95-105% accuracy. Good linearity of the calibration curves was observed. Linearity of the Calibration Curves in Depleted Serum To determine the linearity of the calibration curve in a biological matrix, working solutions spiked in human insulin depleted serum (Figures 7 and 8). The amount of the endogenous human insulin present in such matrix is negligible; therefore, it represents an ideal surrogate for human insulin calibration curve determination. 500 µL of sample was immunoenriched via MSIA technology, eluted and analyzed by LC-MS/MS analysis. Figure 6. Human Insulin Calibration Curve in Depleted Human Serum. Figure 7. Insulin Detemir Calibration Curve in Depleted Human Serum. All the back-calculated standard concentrations for human insulin in depleted plasma were within 95-105% accuracy. Good linearity of the calibration curves was observed. The insulin detemir standards at 6.00 and 10.0 ng/mL deviated from the linear fit. Two hypotheses that would explain the poor accuracy observed at high analyte concentrations were investigated: Insulin detemir adsorption/solubility phenomena at high concentration of analyte in depleted serum Saturation of antibody binding capacity for insulin detemir immunoenrichment Investigation of Non Linearity To control the effect of possible adsorption/solubility phenomena, serial dilutions to generate samples for the calibration curve were made. To reduce possible saturation of the antibody binding capacity the sample volume processed with MSIA technology was reduced from 500 to 250 µL. Figure 8. Human Insulin Calibration Curve. Serial dilution of the stock solution was made to generate standards in human insulin depleted serum. 250 µL of sample was used for immunoenrichment. Figure 9. Insulin Detemir Calibration Curve. Serial dilution of the stock solution was made to generate standards in human insulin depleted serum. 250 µL of sample was used for immunoenrichment. Introduction Insulin is a hormone produced by the pancreas then released in the blood stream where it helps to regulate levels of sugar. The American Diabetes Association recommends that bioanalytical assays, used to measure insulins, should be able to achieve an LLOQ of 0.070 ng/mL. Traditionally ELISA is used for the determination of insulin and insulin isoforms in human serum and plasma based on the inherent sensitivity of this technique. However, in some cases, antibody-based methods may suffer for poor specificity and selectivity or cross reactivity when several isoforms or metabolites are present in the biological matrix. LC-MS/MS approaches are expected to overcome such limitations and enable the simultaneous determination of insulin isoforms in the biological matrices. The aim of this feasibility study was the development of an LC-MS/MS approach for the determination of human insulin and insulin detemir in human plasma. Insulin detemir is a long-acting human insulin analogue used to maintain the basal level of insulin activity in the body. This small protein carries a fatty acid moiety (myristic acid) which is covalently bound to the lysine 29 of the protein beta chain (Figure 1). Figure 1. Human Insulin and Insulin Detemir. Insulin detemir carries fatty acid (myristic acid) that is covalently bound to the lysine 29 of the protein beta chain. The higher complexity of proteins and peptides are often the basis for poor sensitivity and selectivity of the mass spectrometric methods used for their analysis. Recently, the introduction of orthogonal approaches based on immunoenrichment prior to the mass spectrometric determination, alleviates some of these problems. In such methodologies antibodies immobilized on a solid support first capture the molecule of interest from the matrix, then subsequent washing cycles are used to remove contaminants and finally the analyte is eluted prior to its analysis by LC-MS/MS. Samples were immunoenriched prior to LC-MS/MS using MSIA™ technology (Thermo Scientific) (Figure 2). Figure 2. Schematic Representation of MSIA Technology Workflow. Bovine insulin was used as an internal standard. Standards were prepared in insulin immunodepleted serum while QC samples were prepared directly in human plasma. Method Analyte Human insulin and insulin detemir Internal Standard Bovine insulin Matrix Human insulin depleted serum for the calibration curve human plasma for quality control samples Sample Volume 500 μL/250 μL Calibration Range 0.0500 - 10.0 ng/mL Extraction Method Thermo Scientific MSIA LC Method UPLC reverse phase Detection SCIEX Triple Quad 6500 Run Time 7 min LC System Waters ACQUITY UPLC ® I-Class System (SM-FTN) Analytical Column BEH C18, 2.1mm x 50mm, 1.7 μm particles Mobile Phase A Water/Acetic acid (100/0.2) Mobile Phase B Acetonitrile/Acetic acid (100/0.2) Human Insulin (6+) 969.2 ¨ 1130.0 Insulin Detemir (5+) 1184.3 ¨ 1366.7 Bovine Insulin (6+) 956.3 ¨ 1114.8 Solutions of human insulin and insulin detemir were analysed in infusion experiments at 300 µL/min in 30/70/0.2 acetonitrile/water/formic acid. The MRM transitions selected were optimized for CE, DP, CXP and EP. Figure 3 shows a representative MS1 spectrum for insulin detemir. Figure 3. MS1 Spectrum of Insulin Detemir. Impact of Modifier on Analyte Signal Intensities To determine the effects of the modifier on the signal intensities, human insulin and insulin detemir were infused in reconstitution solvent (30/70 acetonitrile/water) modified either with 0.2% formic acid or 0.2% acetic acid at chromatographic compatible flow rate. Whilst no clear difference in charge state distribution was evident, an overall sensitivity difference was observed. For both human insulin and insulin detemir, acetic acid favoured ionization of the analytes and increased signal intensities of 140-170%. Linearity of the Calibration Curves in Pure Solution Working solutions were prepared and diluted to a concentration similar to that reached in the samples after elution of the analytes from the MSIA tips. Solutions were analyzed by LC-MS/MS using a linear gradient of water, acetonitrile modified with 0.2% acetic acid. Bioanalytical Determination of Human Insulin and Insulin Detemir in Human Plasma by Immunoenrichment and LC-MS/MS A. Maiolica, S. Wood and P. Struwe Celerion Switzerland AG, 8320 Fehraltorf, Switzerland Poster presentation at EBF 2015 8 th Open Meeting Barcelona, Nov 18-20, 2015 Human Insulin All the calibration standards were within 85-115% accuracy and were measured with a precision better that 7% Good linearity of the calibration curves was observed Insulin Detemir Deviation from a linear fit of the calibration curve was observed. Although linearity was overall slightly improved The poor linearity for the insulin detemir calibration curve cannot be attributed only to possible adsorption or antibody saturation phenomena. Sufficient sensitivity could be achieved using 250 µL samples; therefore, this value was applied to the method. Method Evaluation 250 µL of human depleted serum at concentration of 50 pg/mL was immunoenriched using MSIA technology and analyzed by LC-MS/MS. Figure 10. Standard Sample of Human Insulin at 0.0500 ng/mL was Immunoenriched via MSIA Technology and Analyzed by LC-MS/MS. Figure 11. Standard Sample of Insulin Detemir Spiked at 0.0500 ng/mL was Immunoenriched via MSIA Technology and Analyzed by LC-MS/MS. Good sensitivity was achieved for both human insulin and insulin detemir using 250 µL samples. Quality control samples were prepared in human plasma. The endogenous concentration of the human insulin in a control sample was determined by a validated ELISA method developed in house (DAKO). The concentration was determined to be 0.527 ng/mL. QC control samples spiked at 0.050 and 0.150 ng/mL. Table 2. Precision and Accuracy of Human Insulin and Insulin Detemir Quality Control Samples. 1140 1160 1180 1200 1220 1240 1260 1280 1300 1320 1340 1360 1380 1400 1420 1440 1460 1480 1500 1520 1540 1560 m/z, Da 0.0 1.0e6 2.0e6 3.0e6 4.0e6 5.0e6 6.0e6 7.0e6 8.0e6 9.0e6 1.0e7 1.1e7 1.2e7 1.3e7 1.4e7 1.5e7 1.6e7 Intensity, cps 1480.2 1184.3 1124.0 1191.8 1200.9 1495.5 1485.1 1137.4 1170.1 1196.9 1322.7 1504.1 1231.1 1297.0 1145.9 1498.8 1236.7 1494.5 1265.8 1224.7 1144.5 1344.7 1182.3 1427.6 1283.1 1292.0 1211.0 1220.1 1310.4 1348.4 1404.0 1390.3 1448.9 1510.1 1364.9 1252.1 1162.4 1529.1 1255.4 1550.3 1506.5 1503.5 1293.6 1453.7 1332.0 1438.8 1468.7 4+ 5+ Human Insulin 0.0500 Insulin Detemir Analyte Concentration (ng/mL) % CV Accuracy Analyte Concentration (ng/mL) % CV Accuracy 1.23 98.2 0.0500 4.89 96.3 0.250 1.80 112 0.250 2.22 115 0.625 2.46 93.8 0.625 9.02 98.7 1.25 4.18 104 1.25 13.6 128 2.50 0.518 88.2 2.50 2.14 90.9 5.00 1.82 107 5.00 4.90 98.8 8.00 6.04 99.7 8.00 6.90 87.8 10.0 1.18 97.3 10.0 1.08 84.4 Human Insulin 0.500 0.587 106 Insulin Detemir 0.0500 8.38 71.9 0.600 2.86 96.2 0.150 0.332 104 Analyte QC Concentration (ng/mL) % CV Accuracy Analyte QC Concentration (ng/mL) % CV Accuracy Human Insulin The endogenous concentration of human insulin measured in control plasma was 0.450 ng/mL. Good correlation between the ELISA and LC-MS/MS estimation was measured (15.8% from the mean concentration value) Good accuracy and precision was measured for the QC samples Insulin Detemir Insulin detemir spiked in human plasma at 0.0500 and 0.150 ng/mL. Poor accuracy was measured for Insulin Detemir at the LLOQ Conclusions In this feasibility study a protocol for the immunoenrichment approach prior to mass spectrometric was applied for the determination of the human insulin and insulin detemir. A good linearity over the quantitation range was demonstrated for the human insulin. The data showed that accuracy at high concentration of insulin detemir decreased considerably. The experiments performed excluded analyte adsorption\solubility or saturation of antibody binding capacity as the main sources of the problem. One possible explanation of the phenomenon could be related to the physiochemical characteristics of insulin detemir different from human insulin. This insulin analogue carries a fatty acid chain attached on the lysine 29 of the protein beta chain. This could influence the binding to the antibody used for its immunopurification when insulin detemir is present at high concentration in the sample. Experiments are continuing to further optimize this method. Figure 4. Human Insulin Calibration Curve in Pure Solutions. Figure 5. Insulin Detemir Calibration Curve in Pure Solutions. Table 1. Precision and Accuracy of Human Insulin and Insulin Detemir Calibration Curve.

Bioanalytical Determination of Human Insulin and Insulin ... · MSIA technology, eluted and analyzed by LC-MS/MS analysis. Figure 6. Human Insulin Calibration Curve in Depleted Human

  • Upload
    lamnga

  • View
    215

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Bioanalytical Determination of Human Insulin and Insulin ... · MSIA technology, eluted and analyzed by LC-MS/MS analysis. Figure 6. Human Insulin Calibration Curve in Depleted Human

All the back-calculated standard concentrations for both human insulin and insulin detemir in pure solutions were within 95-105% accuracy. Good linearity of the calibration curves was observed.

Linearity of the Calibration Curves in Depleted Serum

To determine the linearity of the calibration curve in a biological matrix, working solutions spiked in human insulin depleted serum (Figures 7 and 8). The amount of the endogenous human insulin present in such matrix is negligible; therefore, it represents an ideal surrogate for human insulin calibration curve determination. 500 µL of sample was immunoenriched via MSIA technology, eluted and analyzed by LC-MS/MS analysis.

Figure 6. Human Insulin Calibration Curve in Depleted Human Serum.

Figure 7. Insulin Detemir Calibration Curve in Depleted Human Serum.

All the back-calculated standard concentrations for human insulin in depleted plasma were within 95-105% accuracy. Good linearity of the calibration curves was observed.

The insulin detemir standards at 6.00 and 10.0 ng/mL deviated from the linear fit. Two hypotheses that would explain the poor accuracy observed at high analyte concentrations were investigated: Insulin detemir adsorption/solubility phenomena at high concentration of analyte in depleted serum

Saturation of antibody binding capacity for insulin detemir immunoenrichment

Investigation of Non Linearity

To control the effect of possible adsorption/solubility phenomena, serial dilutions to generate samples for the calibration curve were made. To reduce possible saturation of the antibody binding capacity the sample volume processed with MSIA technology was reduced from 500 to 250 µL.

Figure 8. Human Insulin Calibration Curve. Serial dilution of the stock solution was made to generate standards in human insulin depleted serum. 250 µL of sample was used for immunoenrichment.

Figure 9. Insulin Detemir Calibration Curve. Serial dilution of the stock solution was made to generate standards in human insulin depleted serum. 250 µL of sample was used for immunoenrichment.

IntroductionInsulin is a hormone produced by the pancreas then released in the blood stream where it helps to regulate levels of sugar. The American Diabetes Association recommends that bioanalytical assays, used to measure insulins, should be able to achieve an LLOQ of 0.070 ng/mL. Traditionally ELISA is used for the determination of insulin and insulin isoforms in human serum and plasma based on the inherent sensitivity of this technique. However, in some cases, antibody-based methods may suffer for poor specificity and selectivity or cross reactivity when several isoforms or metabolites are present in the biological matrix. LC-MS/MS approaches are expected to overcome such limitations and enable the simultaneous determination of insulin isoforms in the biological matrices.

The aim of this feasibility study was the development of an LC-MS/MS approach for the determination of human insulin and insulin detemir in human plasma. Insulin detemir is a long-acting human insulin analogue used to maintain the basal level of insulin activity in the body. This small protein carries a fatty acid moiety (myristic acid) which is covalently bound to the lysine 29 of the protein beta chain (Figure 1).

Figure 1. Human Insulin and Insulin Detemir. Insulin detemir carries fatty acid (myristic acid) that is covalently bound to the lysine 29 of the protein beta chain.

The higher complexity of proteins and peptides are often the basis for poor sensitivity and selectivity of the mass spectrometric methods used for their analysis. Recently, the introduction of orthogonal approaches based on immunoenrichment prior to the mass spectrometric determination, alleviates some of these problems. In such methodologies antibodies immobilized on a solid support first capture the molecule of interest from the matrix, then subsequent washing cycles are used to remove contaminants and finally the analyte is eluted prior to its analysis by LC-MS/MS. Samples were immunoenriched prior to LC-MS/MS using MSIA™ technology (Thermo Scientific) (Figure 2).

Figure 2. Schematic Representation of MSIA Technology Workflow.

Bovine insulin was used as an internal standard. Standards were prepared in insulin immunodepleted serum while QC samples were prepared directly in human plasma.

MethodAnalyte Human insulin and insulin detemirInternal Standard Bovine insulinMatrix Human insulin depleted serum for the calibration curve human plasma for quality control samplesSample Volume 500 μL/250 μLCalibration Range 0.0500 - 10.0 ng/mLExtraction Method Thermo Scientific MSIALC Method UPLC reverse phaseDetection SCIEX Triple Quad 6500Run Time 7 minLC System Waters ACQUITY UPLC® I-Class System (SM-FTN)Analytical Column BEH C18, 2.1mm x 50mm, 1.7 μm particlesMobile Phase A Water/Acetic acid (100/0.2)Mobile Phase B Acetonitrile/Acetic acid (100/0.2)Human Insulin (6+) 969.2 ¨ 1130.0Insulin Detemir (5+) 1184.3 ¨ 1366.7Bovine Insulin (6+) 956.3 ¨ 1114.8

Solutions of human insulin and insulin detemir were analysed in infusion experiments at 300 µL/min in 30/70/0.2 acetonitrile/water/formic acid. The MRM transitions selected were optimized for CE, DP, CXP and EP. Figure 3 shows a representative MS1 spectrum for insulin detemir.

Figure 3. MS1 Spectrum of Insulin Detemir.

Impact of Modifier on Analyte Signal Intensities

To determine the effects of the modifier on the signal intensities, human insulin and insulin detemir were infused in reconstitution solvent (30/70 acetonitrile/water) modified either with 0.2% formic acid or 0.2% acetic acid at chromatographic compatible flow rate.

Whilst no clear difference in charge state distribution was evident, an overall sensitivity difference was observed. For both human insulin and insulin detemir, acetic acid favoured ionization of the analytes and increased signal intensities of 140-170%.

Linearity of the Calibration Curves in Pure Solution

Working solutions were prepared and diluted to a concentration similar to that reached in the samples after elution of the analytes from the MSIA tips. Solutions were analyzed by LC-MS/MS using a linear gradient of water, acetonitrile modified with 0.2% acetic acid.

Bioanalytical Determination of Human Insulin and Insulin Detemir in Human Plasma by Immunoenrichment and LC-MS/MSA. Maiolica, S. Wood and P. StruweCelerion Switzerland AG, 8320 Fehraltorf, Switzerland

Poster presentation at EBF 2015 8th Open Meeting Barcelona, Nov 18-20, 2015

Human Insulin All the calibration standards were within 85-115% accuracy and were measured with a precision better that 7%

Good linearity of the calibration curves was observed

Insulin Detemir

Deviation from a linear fit of the calibration curve was observed. Although linearity was overall slightly improved

The poor linearity for the insulin detemir calibration curve cannot be attributed only to possible adsorption or antibody saturation phenomena. Sufficient sensitivity could be achieved using 250 µL samples; therefore, this value was applied to the method.

Method Evaluation250 µL of human depleted serum at concentration of 50 pg/mL was immunoenriched using MSIA technology and analyzed by LC-MS/MS.

Figure 10. Standard Sample of Human Insulin at 0.0500 ng/mL was Immunoenriched via MSIA Technology and Analyzed by LC-MS/MS.

Figure 11. Standard Sample of Insulin Detemir Spiked at 0.0500 ng/mL was Immunoenriched via MSIA Technology and Analyzed by LC-MS/MS.

Good sensitivity was achieved for both human insulin and insulin detemir using 250 µL samples.

Quality control samples were prepared in human plasma. The endogenous concentration of the human insulin in a control sample was determined by a validated ELISA method developed in house (DAKO). The concentration was determined to be 0.527 ng/mL. QC control samples spiked at 0.050 and 0.150 ng/mL.

Table 2. Precision and Accuracy of Human Insulin and Insulin Detemir Quality Control Samples.

1140 1160 1180 1200 1220 1240 1260 1280 1300 1320 1340 1360 1380 1400 1420 1440 1460 1480 1500 1520 1540 1560m/z, Da

0.0

1.0e6

2.0e6

3.0e6

4.0e6

5.0e6

6.0e6

7.0e6

8.0e6

9.0e6

1.0e7

1.1e7

1.2e7

1.3e7

1.4e7

1.5e7

1.6e7

Inte

nsity

, cps

1480.2

1184.3

1124.01191.8

1200.9 1495.51485.11137.4 1170.1 1196.9 1322.7 1504.11231.1 1297.01145.9 1498.81236.7 1494.51265.81224.71144.5 1344.71182.3 1427.61283.1 1292.01211.0 1220.1 1310.4 1348.4 1404.01390.3 1448.9 1510.11364.91252.11162.4 1529.11255.4 1550.31506.51503.51293.6 1453.71332.0 1438.8 1468.7

4+

5+

Human Insulin

0.0500

InsulinDetemir

AnalyteConcentration

(ng/mL) % CV Accuracy AnalyteConcentration

(ng/mL) % CV Accuracy1.23 98.2 0.0500 4.89 96.3

0.250 1.80 112 0.250 2.22 1150.625 2.46 93.8 0.625 9.02 98.71.25 4.18 104 1.25 13.6 1282.50 0.518 88.2 2.50 2.14 90.95.00 1.82 107 5.00 4.90 98.88.00 6.04 99.7 8.00 6.90 87.8

10.0 1.18 97.3 10.0 1.08 84.4

Human Insulin

0.500 0.587 106 InsulinDetemir

0.0500 8.38 71.90.600 2.86 96.2 0.150 0.332 104

AnalyteQC Concentration

(ng/mL) % CV Accuracy AnalyteQC Concentration

(ng/mL) % CV Accuracy

Human Insulin

The endogenous concentration of human insulin measured in control plasma was 0.450 ng/mL. Good correlation between the ELISA and LC-MS/MS estimation was measured (15.8% from the mean concentration value)

Good accuracy and precision was measured for the QC samples

Insulin Detemir Insulin detemir spiked in human plasma at 0.0500 and 0.150 ng/mL. Poor accuracy was measured for Insulin Detemir at the LLOQ

ConclusionsIn this feasibility study a protocol for the immunoenrichment approach prior to mass spectrometric was applied for the determination of the human insulin and insulin detemir. A good linearity over the quantitation range was demonstrated for the human insulin. The data showed that accuracy at high concentration of insulin detemir decreased considerably. The experiments performed excluded analyte adsorption\solubility or saturation of antibody binding capacity as the main sources of the problem. One possible explanation of the phenomenon could be related to the physiochemical characteristics of insulin detemir different from human insulin. This insulin analogue carries a fatty acid chain attached on the lysine 29 of the protein beta chain. This could influence the binding to the antibody used for its immunopurification when insulin detemir is present at high concentration in the sample. Experiments are continuing to further optimize this method.

Figure 4. Human Insulin Calibration Curve in Pure Solutions.

Figure 5. Insulin Detemir Calibration Curve in Pure Solutions.

Table 1. Precision and Accuracy of Human Insulin and Insulin Detemir Calibration Curve.