4
BIO 2+ Fumigation Options D DB2 DRS R RRS RD • Integral Formalin Vapouriser with auto timer O O S • Integral Formalin Vapouriser with auto timer and exhaust sealing valve O S O • Auto-purge to allow full automatic fumigation cycle O O S O S O • 5m fumigant extract flexible duct including connection O S • Mobile carbon filter unit O O • H 2 0 2 vapour phase biodecontamination system including connections and sealing valves O O O O O O Note: When a formalin vaporiser is fitted to a recirculating (R or RRS) cabinet, it must be used either in conjunction with the 5m fumigant extract flexible duct including connection (if there is a window or fume cupboard nearby through which it can exhaust) or with the Mobile carbon filter unit. Useful Information - Definitions BS EN 12469 2000 ACDP (Advisory Committee for Dangerous Pathogens) Microbiological safety cabinet (MSC): Ventilated enclosure intended to offer protection to the user and the environment from the aerosols arising from the handling of potentially hazardous and hazardous micro-organisms, with air discharged to the atmosphere being filtered. Hazard Group 1: A biological agent unlikely to cause disease. MSC Class I: Safety cabinet with a front aperture through which the operator can carry out manipulations inside the cabinet and which is constructed so the worker is protected and the escape of airborne particulate contamination generated within the cabinet is controlled by means of an inward airflow through the working front aperture and filtration of the exhaust air. Hazard Group 2: A biological agent that can cause human disease and may be a hazard to employees; it is unlikely to spread to the community and there is usually effective prophylaxis or effective treatment available. MSC Class II: Safety cabinet with a front aperture through which the operator can carry out manipulations inside the cabinet and which is constructed so the worker is protected, the risk of product and cross contamination is low and the escape of airborne particulate contamination generated within the cabinet is controlled by means of an appropriate filtered internal airflow and filtration of the exhaust air. NOTE A typical way of achieving this is by means of a unidirectional downward laminar airflow inside the cabinet and an air curtain at the front aperture. Hazard Group 3: A biological agent that can cause severe human disease and presents a serious risk to employees; it may present a risk of spreading to the community, but there is usually effective prophylaxis or treatment available. MSC Class III: Safety cabinet in which the working area is totally enclosed and the operator is separated from the work by a physical barrier (i.e. gloves mechanically attached to the cabinet). Filtered air is continuously applied to the cabinet and the exhaust air is treated to prevent release of micro-organisms. Hazard Group 4: A biological agent that causes severe human disease and is a serious hazard to employees; it is likely to spread to the community and there is usually no effective prophylaxis or treatment available. Essential reading: Advisory Committee on Dangerous Pathogens: The management, design and operation of microbiological containment laboratories, HSE Books, 2001 Envair York Avenue Haslingden Rossendale Lancashire BB4 4HX United Kingdom T: +44 (0) 1706 228 416 F: +44 (0) 1706 242 205 email: [email protected] website: www.envair.co.uk Product Sheet 102 April 2006

BIO 2+ - Envair · NSF49 designation for ‘TotalExhaust’ class II microbiological safety cabinets). DRS The Bio 2+ DRS is similar to the Bio 2+ D but is generously sized and has

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: BIO 2+ - Envair · NSF49 designation for ‘TotalExhaust’ class II microbiological safety cabinets). DRS The Bio 2+ DRS is similar to the Bio 2+ D but is generously sized and has

BIO 2+

Fumigation Options D DB2

DRS

R RRS

RD

• Integral Formalin Vapouriser with auto timer O O S

• Integral Formalin Vapouriser with auto timer and exhaust sealing valve O S O

• Auto-purge to allow full automatic fumigation cycle O O S O S O

• 5m fumigant extract flexible duct including connection O S

• Mobile carbon filter unit O O

• H202 vapour phase biodecontamination system including connections and sealing valves O O O O O O

Note:When a formalin vaporiser is fitted to a recirculating (R or RRS) cabinet, it must be used either in conjunction with the 5mfumigant extract flexible duct including connection (if there is a window or fume cupboard nearby through which it canexhaust) or with the Mobile carbon filter unit.

Useful Information - Definitions

BS EN 12469 2000 ACDP (Advisory Committee for Dangerous Pathogens)

Microbiological safety cabinet (MSC): Ventilated enclosureintended to offer protection to the user and the environment fromthe aerosols arising from the handling of potentially hazardous andhazardous micro-organisms, with air discharged to the atmospherebeing filtered.

Hazard Group 1: A biological agent unlikely to cause disease.

MSC Class I: Safety cabinet with a front aperture through whichthe operator can carry out manipulations inside the cabinet andwhich is constructed so the worker is protected and the escape ofairborne particulate contamination generated within the cabinet iscontrolled by means of an inward airflow through the working front

aperture and filtration of the exhaust air.

Hazard Group 2: A biological agent that can cause human diseaseand may be a hazard to employees; it is unlikely to spread to thecommunity and there is usually effective prophylaxis or effectivetreatment available.

MSC Class II: Safety cabinet with a front aperture through which the operator can carry out manipulations inside the cabinetand which is constructed so the worker is protected, the risk ofproduct and cross contamination is low and the escape of airborneparticulate contamination generated within the cabinet is controlledby means of an appropriate filtered internal airflow and filtration ofthe exhaust air. NOTE A typical way of achieving this is by meansof a unidirectional downward laminar airflow inside the cabinet andan air curtain at the front aperture.

Hazard Group 3: A biological agent that can cause severe humandisease and presents a serious risk to employees; it may present arisk of spreading to the community, but there is usually effectiveprophylaxis or treatment available.

MSC Class III: Safety cabinet in which the working area is totally enclosed and the operator is separated from the work by a physical barrier (i.e. gloves mechanically attached to the cabinet). Filtered air is continuously applied to the cabinet and theexhaust air is treated to prevent release of micro-organisms.

Hazard Group 4: A biological agent that causes severe humandisease and is a serious hazard to employees; it is likely to spreadto the community and there is usually no effective prophylaxis ortreatment available.

Essential reading:Advisory Committee on Dangerous Pathogens: The management, design and operation of microbiological containment laboratories, HSEBooks, 2001

Envair York AvenueHaslingdenRossendaleLancashireBB4 4HXUnited Kingdom

T: +44 (0) 1706 228 416 F: +44 (0) 1706 242 205 email: [email protected] website: www.envair.co.uk

Product Sheet 102 April 2006

Page 2: BIO 2+ - Envair · NSF49 designation for ‘TotalExhaust’ class II microbiological safety cabinets). DRS The Bio 2+ DRS is similar to the Bio 2+ D but is generously sized and has

CLASS II MICROBIOLOGICAL SAFETY CABINET - BIO 2+

Principal Design Features

• Unique double HEPA filtration system comprising a primary (exhaust and main) HEPA filter and providing valuablebenefits:

* All fans in HEPA filtered air for long life and safe maintenance* Integral double HEPA filtration of downflow on all models to ensure air quality in the event of downflow filter failure* Integral secondary exhaust HEPA filter on recirculating (R)models * Replacement of primary (exhaust and main) HEPA filter without dismantling fan, electrics or ductwork

• Clean and attractive design in white polyester coated steel construction• Very high levels of retention efficiency (operator protection) and product protection• Very high level of protection against cross-contamination • Simple hinged visor, therefore no difficult or inaccessible areas to clean• Cabinet body under negative pressure and filters mounted correctly for safety in accordance with

BS EN 12469: 2000• Self-sealing light-weight night door/closure panel for safe fumigation • Comprehensively alarmed and interlocked for safety• Access for set up and all maintenance from front of cabinet • Built in BMS (Building Management System) and auto-damper signal outputs

Compact, highly specified, high performance Class II microbiological safety cabinetswith unique double HEPA filtration.

www.envair.co.uk

ApplicationsCell culture operations and handling of biological agents up toHazard Group 3* where both retention efficiency (operatorprotection) and the provision of clean air conditions over the work area are required as well as total compliance with all therequirements of BS EN 12469: 2000.

* ACDP Categorisation

Benefits • Carefully designed and extensively tested to provide assurance

of microbiological safety when used with biological agents upto ACDP Hazard Group 3 (subject to proper siting, operationand maintenance)

• Unique double HEPA filtration system and many other design features to facilitate safe and easy operation, cleaning, testingand maintenance

• Aerodynamically independent downflow and exhaust air fan systems to facilitate quick and easy set up

• Wide range of sizes, configurations and fumigation options to provide a solution for every laboratory and application

Standards Compliance• BS EN 12469: 2000 • Certified by CAMR (now HPA) at Porton Down• All HEPA filters H14 to BS EN 1822-1: 1998 • IEC 61010-1: 2001 • Installation to BS 5726: 2005 is recommended• Air quality to ISO Class 5 (BS EN ISO 14644-1: 1999) and EC

GMP Grade A for a high level of product protection

Page 3: BIO 2+ - Envair · NSF49 designation for ‘TotalExhaust’ class II microbiological safety cabinets). DRS The Bio 2+ DRS is similar to the Bio 2+ D but is generously sized and has

1

1. Clear, easy to clean work-space with no inaccessible voids 2. One piece, easily removable, 316L stainless steel work-tray 3.Formalin vapouriser* with automatic purge valve* 4. Thimble Exhaust System, as used with the RD model 5. Membrane control panel 6. Easy maintenance access for controls and light tubes 7. Splashproof sockets* 8. Microswitch safety interlock for nightdoor * optional

Dimensions

BIO 2+

* Except DRS and RRS models. Drawing and dimensions of these models available on request

5

2

6

3

7

4

8

Bio 2+* A (mm) B (mm) C (mm) Weight (Kg)Power(Watts) Exhaust rate m3/s Exhaust rate DB2 m3/s

1.2m 1250 1060 1000 195 900 0.12 0.30

1.5m 1675 1485 1425 235 1200 0.15 0.50

1.8m 1870 1680 1620 285 1200 0.18 0.56

Page 4: BIO 2+ - Envair · NSF49 designation for ‘TotalExhaust’ class II microbiological safety cabinets). DRS The Bio 2+ DRS is similar to the Bio 2+ D but is generously sized and has

BIO 2+

Specification D DB2

DRS

R RRS

RD

• Primary (exhaust and main) HEPA filter to H14 (BS EN 1822) S S S S S

• Downflow HEPA filter to H14 (BS EN 1822) S S S S S S

• Secondary exhaust HEPA filter to H14 (BS EN 1822) S S S

• Single 100% exhaust HEPA filter to H14 (BS EN 1822) S

• Integral anti-blowback damper S S S

• Integral exhaust fan box S S

• Integral exhaust fan box and thimble exhaust spigot S

• 316L stainless steel worktray with 4 litre spillage capacity S S S S S S

• Hinged toughened glass visor, with gas strut supports for opening S S S S S S

• Light weight 100% sealable night door (closure panel) S S S S

• Night door interlocked with formalin / UV light for safety S S S S

• Light weight night door (closure panel) suitable for sealing with tape S S

• Easy to clean touch control membrane on hinged lockable control access panel S S S S S S

• Solid state fan speed controllers, selectable automatic fan speed control S S S S S S

• Latched audible and visible alarms for low and high airflow and for visor open S S S S S S

• Alarm for downflow air supply failure, alarm mute and alarm test S S S S S S

• Digital inflow air monitoring, filter soil pressure indicator, hours run counter S S S S S S

• Switchable auxillary circuits for gas, UV and formalin vapouriser options, plus 3 spare switches S S S S S S

• Internal flourescent lighting behind sealed acylic panel S S S S S S

• Robotic sampler transfer chassis on roller and rail S S

• Guide rails for loading and removal of robotic sampler S S

• 1 No RS 232 computer data point (LHS) S S

• Electrical supply: 230v/50Hz/1Ph S S S S S S

• Support frame with adjustable levelling feet O O S O

• Support frame with lockable castors O S

• 316L stainless steel polished internally and externally O O O O O O

• UV light O O O O O O

• Solenoid controlled fail-safe gas valve O O O O O O

• 13 amp single splashproof socket (2 fitted LHS & RHS as standard on DRS and RS) S S S S S S

• Service taps (i.e. Vacuum, Nitrogen, Compressed Air) O O O O O O

S Standard Feature O Optional Feature

Airflow Diagram

MODEL DESIGNATIONS

D The Bio 2+ D is designed to exhaust HEPA filtered air to atmosphere via a dedicated ducting system and remote mounted extract fan.

DB2

The Bio 2+ DB2 is a ‘Total Exhaust’ cabinet and 100% of the combined volume of the inflow and downflow air is exhausted via a HEPA filter to atmosphere via a dedicated ducting system and remote mounted extract fan. (B2 is the AmericanNSF49 designation for ‘Total Exhaust’ class II microbiological safety cabinets).

DRS The Bio 2+ DRS is similar to the Bio 2+ D but is generously sized and has additional features in order to accommodate Robotic Samplers and other equipment that might require Class II protection.

RThe Bio 2+ R is designed to exhaust HEPA filtered air back to the laboratory via a dual in line HEPA filter and an integral exhaust fan.

RRSThe Bio 2+ RRS is similar to the Bio 2+ R but is generously sized and has additional features in order to accommodate Robotic Samplers and other equipment that might require Class II protection.

RD

The Bio 2+ RD is designed to exhaust HEPA filtered air into a thimble extractsystem (provided by the client) via a dual in line HEPA filter, an integral exhaustfan and a special exhaust spigot. A thimble extract system takes the exhaust of a microbiological safety cabinet into a specially designed room extract in such away that the volume of air taken out of the room remains the same whether thecabinet is working or not.