Upload
sreenivasa-murthy
View
212
Download
7
Embed Size (px)
Citation preview
WelcomeBasics of Quality Controls
Rajiv Gandhi University of Health and Sciences
10-04-09
• A control is a “Patient-Like” sample composed of one or many constituents whose concentrations are known. The constituent or analyte concentrations are represented by a range of acceptable values with upper and lower limits. Controls are used to monitor the precision and accuracy of the diagnostic testing system.
• When the observed control values fall within the acceptable range, the laboratories can confirm the test system is working properly and the patient results can be reported with confidence.
What is a Control?What is a Control?
Why do we do QC testing?Precision Testing
Determination of Relative Bias
Objectives
Quality Controls and Regulatory Standards
• Quality Council Of India ( QCI ) : Essential Standards for Registration of Medical Testing Laboratories In India
- 3.5.2 : Medical Laboratories shall perform internal Quality Control . Use of third party human matrix quality control is recommended for all analytes .Laboratories are encouraged to participate in interlab comparisons / External Quality Assessment Schemes EQAS/ Proficiency Testing (PT).
• ISO 15189 Medical Laboratories-Particular requirements for quality and competence– 5.6.1 “The laboratory shall design internal quality control
systems that verify the attainment of the intended quality of results.”
– 5.6.3 “…other means of providing confidence in the results shall be applied, including but not limited to the following: a) participation in a suitable programme of interlaboratory comparisons…”
Quality Controls and Regulatory Standards
Quality Controls and Regulatory Standards
Clinical Laboratory Improvement Act (CLIA)
• Laboratories must test controls at prescribed intervals
• At least two different concentrations of controls should be tested
• Laboratories must have established criteria for acceptable QC
• Laboratories are responsible for the quality of testing
Quality Controls:Important Characteristics
• Unbiased Assessment or Third Party• Long Shelf-Life and Open Vial Stability• Clinically Relevant Levels• Human Based Matrix• Extensive Assayed Values Provided• Comprehensive Range of Products
“In-kit control” = “Manufacturer’s control”
Supplied with test kit or by the manufacturer of the laboratory’s instrument
In-Kit Controls: Can problems arise?
Why Choose Third Party Controls?Independent of Kit and Reagent Lot Changes
Bio-Rad Controls have up to 3 years of shelf life
-Fewer lot crossovers to perform
Goal of Quality Control:Evaluate and minimize analytical error
Use a system approach
• Minimize error through following operating protocols
– Calibration, maintenance, etc.• Assess daily performance
– Internal statistical quality control• Validate accuracy and precision periodically
– EQA (proficiency testing)
QC Vs Calibrators & Standards
• Calibration is “setting” the analyzer to give correct results– Using calibrating materials, not QC materials
• QC is checking to see if the analyzer is producing correct results– Checking the instrument’s calibration and other analytical
processes
• Standards also cannot be used as QC material- Multi-analyte standards are not available
How does basic QC work?
• Run a control sample • Compare result with expected range of values derived
from previous testing• Check to see if the result is right
– yes -- system is working• report good patient results
– no -- system is not working• do not report any bad data
Quality control- Frequency
• Should be run periodically
– NABL specific criteria 112╕Kdrr sg`m 14 r` l okdr
├nmd kdudk `s kd`rs nmbd ` c`x╕L nqd sg`m 14 r` l okdr
├sv n kdudkr `s kd`rs nmbd ` c`x╕L nqd sg`m 64 r` l okdr
├Sv n kdudkr `s kd`rs sv hbd ` c`x
• The closeness of the agreement between the measured values and the “true value” of the constituent. The best estimate of accuracy can be measured in two ways:
1. Reference Method2. Peer Group Comparison
Accuracy
The degree of replication or reproducibility.
The best estimate of replication uses both S.D. and CV to describe Precision.
Precision
• Mean
• Standard Deviation (SD)
• Coefficient of Variation (CV)
• Coefficient Of Variation Ratio (CVR)
• The Standard Deviation Index (SDI)
QC Statistics
• Mean-- Arithmetic average of set of data points e.g.. Set LDH QC values of 121,115,118,119,120
• Sum of the set-- 603 and themean-603/5=120.6 IU/L
• At least 20 data points to be collected from 20 or more separate runs to establish target values for control materials
Mean
• This quantifies the degree of dispersion of STANDARD data points about the mean and is used to set limits upon which control result acceptability is determined
Standard deviation is valuable when comparing methods or evaluating new instruments.
• A method/instrument --low standard deviation--consistent results• A method/instrument -high standard deviation --less certainty of accuracy of
diagnosis and therefore less effectiveness of treatment
Standard DeviationStandard Deviation
F nnc OqdbhrhnmMn Ah̀ r
F nnc OqdbhrhnmAh̀ r Oqdrdms
>
@bbt q̀ bx >>>Ghf g Hl oqdbhrhnm
>
) ) NABL 112 Req
How many levels of controls should a Laboratory be running?
Single Level Only
Two Levels of Control
Three Levels of Control
l Better Informed Decisions
l Reduces Number of Repeats
WESTGARD RULES- WHY ?
Westgard Rules
Random Error Systematic Error
Troubleshooting and Westgard Rules
Problem identified by Westgard rule violated
Systematic Error (bias) 1 3s, 2 2s, 4 1s, 10xShifts, drifts, or trendsCalibration failure
ORRandom Error (imprecision) 1 3s, R4s
Poor reproducibility / Outliers
EQAS Programs available with Bio-Rad
1. Clinical Chemistry Monthly EQAS2. Clinical Chemistry Biweekly EQAS3. Immuno Assay Monthly EQAS4. Immuno Assay Biweekly EQAS5. Hemoglobin EQAS6. Hematology EQAS7. Urine Chemistry EQAS8. Therapeutic Drug Monitoring EQAS
Coming Soon…..1. Coagulation EQAS2. Infectious Serology EQAS
Why to use an EQA Program?
• Increase confidence in your testing system• Tool for testing accuracy
• May be required for accreditation
• External, unbiased assessment
• Additional tool for trouble shooting
Bio-Rad EQAS Programs
• Set of 12 unknown samples• Assayed at regular intervals ( 1 vial every month OR
once in 2 weeks)• Results are sent to Bio-Rad every month • Results are compared to your peer group• Results are compared to your previous performance• Monthly reports are given to the labs
Your EQAS ReportsEssential Information for Busy Laboratory Managers
Your EQAS Reports
Bi-Weekly/Monthly Report
Provides informationfor each sample
in a cycle
End of CycleReport
Provides a summaryof laboratory performance
for the entire cycle
Cover Page (Lab No. and contact information)Summary of Your ResultsData Analysis by Individual Analyte (also referred to as the “Histogram Page”)Summary of Data on File
Bi-Weekly/Monthly Report
How your results compare to your Comparator ValueYour Results, and Units listed for each Analyte
“Summary of Your Results”Lists a summary of information for all analytes for this sample
Your Method Codes, Description and any Conversion Factors to be applied
Levels of Comparison for Statistics
METHOD (Instrument)e.g.Glucose Oxidase on Hitachi 911
Method GROUPe.g. All Glucose Oxidase assays
ALL RESULTS or MODEe.g. All Glucose methods
Example for Glucose:
If insufficient data for comparison
If insufficient data for comparison
All Results / Mode
Mode 1
Mode 2
Mode 3
Example of Multi-Modal Data Distribution:
Your Result
“Histogram Page”Provides statistics and easy-to-read graphs for each analyte
Your MethodYour GroupAll Results
Analyte Statistics provide theMean and SD for Your Method,
Your Group and All Results
Deviation from All ResultsDeviation from Your GroupDeviation from Your Method
… and your deviation from Your Method mean, Your Group mean and the All Results mean
Clinical assessment
… and as assigned by other labs (Your Method, Your Group and All Results)
Your Lab:
Histogram
Your Result
Different shading is used to identify Your Method, Your Group and All Results
Levey-Jennings Chartdisplays last 12 data points
The shaded boxes indicate the comparator used for plotting the Levey-Jennings and calculating
your Accuracy Score
Different symbols are used to indicate a change in comparator or late data
Your method information listed for each analyteYour Results submitted so far this cycle, listed for each analyte
“Summary of Data on File”Lists a summary of all information so far this cycle
Developing a QC ProgramFAQ’s
My control values are not on/Close to the mean as provided in the package insert range. What should I do?
- Establish your mean and ranges- Obtain peer information
Developing a QC ProgramFAQ’s
My control values do not fall within your package insert range. What should I do?
- CALL The Manufacturer: Instrument or QC- Obtain peer information- Any changes to reagents / software?- Confirm change affects patients / QC?
How do I establish a range to use?
Brand New ControlNCCLS - At least 20 points from separate
runs - Standard recommendation NCCLS - May use fewer runs (provisional)
but replace with data from all 20 runs
Developing a QC ProgramFAQ’s
How do I establish a range to use?
New lot of same control typeNCCLS - At least 20 points from separate
runs - Standard recommendation NCCLS - Perform parallel testing with old
lot over 20 runs prior to beginning new lot
Developing a QC ProgramFAQ’s
My IQC data is good however EQA/ PT shows poor accuracy scores?
Daily QC- Indicator of precisionEQA / PT – True Indicator of Accuracy
Developing a QC ProgramFAQ’s