BA 053D/06/en/04.0150099247
valid as of software version:V 1.01.XX (Amplifier)V 1.00.XX (Communications)
promag 53PROFIBUS-DP/-PAElectromagneticFlow Measuring System
Operating Instructions
Distributed by:
Ashton House, Stag Lane, Great Kingshill,Astles Control Systems
High Wycombe, Bucks. HP15 6EW +44 (0)1494 714372 http://www.astles.co.uk/
Promag 53 PROFIBUS-DP/-PA
2 Endress+Hauser
Brief operating instructions
These brief operating instructions show you how to configure the measuring device quickly and easily:
Safety instructions Page 5
Installation Page 11
Wiring Page 45
Switching on the measuring device Page 85
Display and operating elements Page 64
“QUICK SETUP” –
Information about the “Quick Setup” menu can be found in the “Description of Device Functions” manual, which is a separate part of this Operating instruction!
Customer specific configuration Page 67 ff.
Complex measurement tasks require the configuration of additional functions, which the user can individually select, set and adapt to his process conditions via the function matrix.All functions are described in detail, as is the function matrix itself, in the “Description of Device Functions” manual, which is a separate part of this Operating Instruction.
Trouble-shooting Page 113 ff.
Always start trouble-shooting with the checklist on Page 113, if faults occur after commissioning or during operation. The routine takes you directly to the cause of the problem and the appropriate remedial measures.
Returning devicesIf you return a measuring device to Endress+Hauser for repair or calibration, you must complete the “Safety regulation” form and enclose it with the device. You will find a preprinted “Safety regulation” form at the back of this manual.
Promag 53 PROFIBUS-DP/-PA Contents
Endress+Hauser 3
Contents
1 Safety instructions . . . . . . . . . . . . . . 5
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . 51.2 Installation, commissioning and operation . . . 51.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . 51.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61.5 Notes on safety conventions and icons . . . . . . 6
2 Identification . . . . . . . . . . . . . . . . . . . 7
2.1 Device designation . . . . . . . . . . . . . . . . . . . . . 72.1.1 Nameplate of the transmitter . . . . . . . . 72.1.2 Nameplate of the sensor . . . . . . . . . . . . .8
2.2 CE mark, declaration of conformity . . . . . . . . . 82.3 Device certification PROFIBUS-DP/-PA . . . . . . 92.4 Registered trademarks . . . . . . . . . . . . . . . . . . 9
3 Installation. . . . . . . . . . . . . . . . . . . . 11
3.1 Incoming acceptance, transport and storage 113.1.1 Incoming acceptance . . . . . . . . . . . . 113.1.2 Transport . . . . . . . . . . . . . . . . . . . . . . 113.1.3 Storage . . . . . . . . . . . . . . . . . . . . . . . . 12
3.2 Installation conditions . . . . . . . . . . . . . . . . . . 133.2.1 Dimensions . . . . . . . . . . . . . . . . . . . . . 133.2.2 Mounting location . . . . . . . . . . . . . . . . 133.2.3 Orientation . . . . . . . . . . . . . . . . . . . . . . .153.2.4 Inlet and outlet runs. . . . . . . . . . . . . . . .163.2.5 Vibrations . . . . . . . . . . . . . . . . . . . . . . 163.2.6 Foundations, supports. . . . . . . . . . . . . .173.2.7 Adapters . . . . . . . . . . . . . . . . . . . . . . . .183.2.8 Nominal diameter and flow rate . . . . . 183.2.9 Length of connecting cable . . . . . . . . .24
3.3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . 253.3.1 Installing the Promag W sensor . . . . . 253.3.2 Installing the Promag P sensor . . . . . . .313.3.3 Installing the Promag H sensor . . . . . . .373.3.4 Turning the transmitter housing. . . . . . .403.3.5 Installing the wall-mount transmitter
housing . . . . . . . . . . . . . . . . . . . . . . . . .413.3.6 Turning the local display . . . . . . . . . . . .43
3.4 Installation check . . . . . . . . . . . . . . . . . . . . . . 44
4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . 45
4.1 Cable specifications for PROFIBUS-DP/-PA . 454.1.1 PROFIBUS-DP: Cable specifications . 454.1.2 PROFIBUS-PA: Cable specifications. . .48
4.2 Connecting the remote version . . . . . . . . . . . 504.2.1 Connecting Promag W / P / H . . . . . . 504.2.2 Cable specifications . . . . . . . . . . . . . . .52
4.3 Connecting the measuring unit . . . . . . . . . . . 534.3.1 Connecting the transmitter . . . . . . . . . 534.3.2 Terminal assignment . . . . . . . . . . . . . 554.3.3 Fieldbus connector . . . . . . . . . . . . . . . .56
4.4 Potential equalisation . . . . . . . . . . . . . . . . . . . 584.4.1 Standard case . . . . . . . . . . . . . . . . . . 58
4.4.2 Special cases . . . . . . . . . . . . . . . . . . . . 594.5 Degree of protection . . . . . . . . . . . . . . . . . . . . 614.6 Electrical connection check . . . . . . . . . . . . . . 62
5 Operation. . . . . . . . . . . . . . . . . . . . . . 63
5.1 Operation at a glance . . . . . . . . . . . . . . . . . . . 635.2 Operation via the local display . . . . . . . . . . . . 64
5.2.1 Display and operating elements . . . . . 645.2.2 Brief description of the function matrix. 675.2.3 Error messages . . . . . . . . . . . . . . . . . . 69
5.3 Communications: PROFIBUS-DP/-PA . . . . . . . 705.3.1 PROFIBUS-DP/-PA technology . . . . . . 705.3.2 PROFIBUS-DP system architecture . . . 715.3.3 PROFIBUS-PA system architecture . . . 725.3.4 Acyclic data exchange. . . . . . . . . . . . . 74
5.4 Operation with the PROFIBUS configuration programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . 755.4.1 FieldTool™ operating program . . . . . . 755.4.2 Commuwin II operating program . . . . . 75
5.5 Hardware configuration . . . . . . . . . . . . . . . . . 835.5.1 Configuration of write protection . . . . . 835.5.2 Configuration of the device address . . 84
6 Commissioning. . . . . . . . . . . . . . . . . 85
6.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . 856.2 Commissioning the PROFIBUS-DP/-PA
interface using the local display . . . . . . . . . . . 866.3 Commissioning using the Class 2 master
(Commuwin II) . . . . . . . . . . . . . . . . . . . . . . . . . 876.3.1 Rescaling the input value . . . . . . . . . . . 89
6.4 System integration . . . . . . . . . . . . . . . . . . . . . 906.4.1 Cyclic data exchange. . . . . . . . . . . . . . 936.4.2 Configuration examples with
Simatic S7 HW-Konfig . . . . . . . . . . . . . 996.4.3 Cycle times . . . . . . . . . . . . . . . . . . . . . 104
6.5 Empty-pipe/full-pipe calibration . . . . . . . . . . 1066.6 Data storage device (DAT, F-Chip™) . . . . . . 107
7 Maintenance . . . . . . . . . . . . . . . . . . 109
8 Accessories . . . . . . . . . . . . . . . . . . . 111
9 Trouble-shooting . . . . . . . . . . . . . . 113
9.1 Trouble-shooting instructions . . . . . . . . . . . . 1139.2 System and process error messages . . . . . . 1159.3 Process errors without messages . . . . . . . . . 1219.4 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . 1229.5 Removing and installing printed circuit
boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1239.6 Replacing the device fuse . . . . . . . . . . . . . . 1279.7 Replacing exchangeable measuring
electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Contents Promag 53 PROFIBUS-DP/-PA
4 Endress+Hauser
9.8 Software history . . . . . . . . . . . . . . . . . . . . . . 130
10 Technical data . . . . . . . . . . . . . . . . 131
10.1 Technical data at a glance . . . . . . . . . . . . . 13110.2 Measuring tube specifications . . . . . . . . . . . 14010.3 Resistance to partial vacuum of
measuring tube lining . . . . . . . . . . . . . . . . . 14210.4 Weight details . . . . . . . . . . . . . . . . . . . . . . . 14310.5 Dimensions of wall-mount housing . . . . . . . 14510.6 Dimensions Promag 53 W . . . . . . . . . . . . . . 14610.7 Dimensions Promag 53 P . . . . . . . . . . . . . . . 15010.8 Dimensions of ground disks (Promag W, P) 15510.9 Dimensions Promag 53 H . . . . . . . . . . . . . . 15610.10 Process connections Promag H (DN 2...25) 16010.11 Process connections of Promag H
(DN 40...100) . . . . . . . . . . . . . . . . . . . . . . . . 168
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Promag 53 PROFIBUS-DP/-PA 1 Safety instructions
Endress+Hauser 5
1 Safety instructions
1.1 Designated use
The measuring device described in this Operating Instruction is to be used only for measuring the flow rate of conductive fluids in closed pipes. A minimum conductivity of 20 µS/cm is required for measuring demineralised water. Most fluids can be metered, provided they have a minimum conductivity of 5 µS/cm, for example:• acids, alkalis, pastes, mashes, pulps,• drinking water, wastewater, sewage sludge,• milk, beer, wine, mineral water, yogurt, molasses, etc.
The manufacturer accepts no liability for damages resulting from incorrect use or use not as designated.
1.2 Installation, commissioning and operation
Note the following points:• Installation, connection to the electricity supply, commissioning and maintenance of
the device must be carried out by trained, qualified specialists authorised to perform such work by the facility's owner-operator. The specialist must have read and under-stood this Operating Instruction and must follow the instructions it contains.
• The device must be operated by persons authorised and trained by the facility's owner-operator. Strict compliance with the instructions in the Operating Instruction is mandatory.
• Endress+Hauser will be happy to assist in clarifying the chemical resistance proper-ties of parts wetted by special fluids, including fluids used for cleaning.
• If welding work is performed on the piping system, do not ground the welding appli-ance through the Promag flowmeter.
• The installer must ensure that the measuring system is correctly wired in accordance with the wiring diagrams. The transmitter must be grounded, unless the power supply is galvanically insulated.
• Invariably, local regulations governing the opening and repair of electrical devices apply.
1.3 Operational safety
Note the following points:• Measuring systems for use in hazardous environments are accompanied by separate
“Ex documentation”, which is an integral part of this Operating instruction. Strict com-pliance with the installation instructions and ratings as stated in this supplementary documentation is mandatory. The symbol on the front of this supplementary Ex docu-mentation indicates the approval and the test center ( f Europe, h USA, g Canada).
• The measuring device complies with the general safety requirements in accordance with EN 61010, the EMC requirements of EN 61326, and NAMUR recommendation NE 21.
• Depending on the application, the seals of the process connections of the Promag H sensor require periodic replacement.
• The manufacturer reserves the right to modify technical data without prior notice. Your E+H distributor will supply you with current information and updates to this Operating Instruction.
1 Safety instructions Promag 53 PROFIBUS-DP/-PA
6 Endress+Hauser
1.4 Return
The following procedures must be carried out before a device requiring repair or cali-bration, for example, is returned to Endress+Hauser:
• Always enclose a duly completed “Safety regulation” form. Only then can Endress+Hauser transport, examine and repair a returned device.
• Enclose special handling instructions if necessary, for example a safety data sheet as per EN 91/155/EEC.
• Remove all residues. Pay special attention to the grooves for seals and crevices which could contain residues. This is particularly important if the substance is hazardous to health, e.g. flammable, toxic, caustic, carcinogenic, etc.
Note:You will find a preprinted “Safety regulation” form at the back of this manual.
Warning:• Do not return a measuring device if you are not absolutely certain that all traces of
hazardous substances have been removed, e.g. substances which have penetrated crevices or diffused through plastic.
• Costs incurred for waste disposal and injury (burns, etc.) due to inadequate cleaning will be charged to the owner-operator.
1.5 Notes on safety conventions and icons
The devices are designed to meet state-of-the-art safety requirements, have been tested, and left the factory in a condition in which they are safe to operate. The devices comply with the applicable standards and regulations in accordance with EN 61010 “Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures”. They can, however, be a source of danger if used incor-rectly or for other than the designated use. Consequently, always pay particular attention to the safety instructions indicated in this Operating Instruction by the following icons:
Warning:“Warning” indicates an action or procedure which, if not performed correctly, can result in injury or a safety hazard. Comply strictly with the instructions and proceed with care.
Caution:“Caution” indicates an action or procedure which, if not performed correctly, can result in incorrect operation or destruction of the device. Comply strictly with the instructions.
Note:“Note” indicates an action or procedure which, if not performed correctly, can have an indirect effect on operation or trigger an unexpected response on the part of the device.
Promag 53 PROFIBUS-DP/-PA 2 Identification
Endress+Hauser 7
2 Identification
2.1 Device designation
The “Promag 53” flow measuring system consists of the following components:• Promag 53 transmitter• Promag W, Promag P or Promag H sensor
In the compact version, transmitter and sensor form a single mechanical unit; in the remote version they are installed separately.
2.1.1 Nameplate of the transmitter
Fig. 1: Nameplate specifications for the “Promag 53” transmitter (example)
1 Ordering code/serial number: See the specifications on the order confirmation for the meanings of the individual letters and digits.
2 Power supply / Frequency: 16...62 V DC / 20...55 V AC / 50...60 HzPower consumption: 15 VA / W
3 Additional functions and software:– EPD: with Empty Pipe Detection– ECC: with Electrode Cleaning Circuitry
4 Outputs / inputs: PROFIBUS-DP/-PA5 Reserved for information on special products6 Ambient temperature range7 Degree of protection
F06-
53P
Bxx
xx-1
8-06
-xx-
xx-0
00
PROMAGENDRESS+HAUSER
Order Code:
53
Ser.No.:
TAG No.:
16-62VDC/20-55VAC50-60Hz
EPD / MSUECCPROFIBUS-PA Profile 3.0
Pat. US 5,323,156 5,479,007
Pat. US 4,382,387 4,704,908 5,351,554
15VA/W
IP67/NEMA/Type4XXXXXX-XXXXXXXXXXXX12345678901ABCDEFGHJKLMNPQRST
–20°C (–4°F) < Tamb < +60°C (+140°F)i
1
76
2
3
4
5
R
2 Identification Promag 53 PROFIBUS-DP/-PA
8 Endress+Hauser
2.1.2 Nameplate of the sensor
Fig. 2: Nameplate specifications for the “Promag” sensor (example)
1 Ordering code/ serial number: See the specifications on the order confirmation for the meanings of the individual letters and digits.
2 Calibration factor: 0.5328; zero point: −53 Nominal diameter: DN 100
Nominal pressure: DIN PN 16 bar4 TMmax +80 °C (max. fluid temperature)5 Materials:
– Lining: hard rubber (HG)– Measuring electrodes: stainless steel 1.4435
6 Additional information (examples):– EPD: with Empty Pipe Detection electrode– R/B: with Reference electrode– EME/AWE: with Exchangeable Measurement Electrodes– 0.2% CAL: with 0.2% calibration
7 Reserved for information on special products8 Ambient temperature range9 Degree of protection10 Flow direction
2.2 CE mark, declaration of conformity
The devices are designed to meet state-of-the-art safety requirements, have been tested, and left the factory in a condition in which they are safe to operate. The devices comply with the applicable standards and regulations in accordance with EN 61010 “Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures”. The measuring system described in this Operating Instruction thus complies with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
i
1
2345
6
7
9
10
8
ABCDEFGHJKLMNPQRSTTAG No.:
Pat. UK EP 219 725 EP 521169
Pat. UK EP 541 878 EP 618 680
IP67/NEMA/Type4X
HG / 1.4435
0.5328 / –5
TMmax.: 80°C
EPD/MSU R/B EME/AWE 0.2% CAL
DN100 DIN PN 16K-factor:
Materials:
ENDRESS+HAUSER
12345678901XXXXX-XXXXXXXXXXXX
PROMAG
Order Code:
Ser.No.:
X
–20°C (–4°F) < Tamb < +60°C+140°F
F06-
xxxx
xxxx
-18-
05-x
x-xx
-000
Promag 53 PROFIBUS-DP/-PA 2 Identification
Endress+Hauser 9
2.3 Device certification PROFIBUS-DP/-PA
The Promag 53 flowmeter has passed all the test procedures implemented and has been certified and registered by the PNO (PROFIBUS User Organisation). The device thus meets all the requirements of the specifications listed below:• Certified for PROFIBUS 3.0
Device certification number: upon request• The instrument meets all of the PROFIBUS 3.0 specifications.• The device may also be operated using certified devices from other manufacturers
(interoperability).
2.4 Registered trademarks
KALREZ ®, VITON ® and TEFLON ® are registered trademarks of E.I. Du Pont de Nemours & Co., Wilmington, USA
TRI-CLAMP ® is a registered trademark of Ladish & Co., Inc., Kenosha, USA
PROFIBUS ® is a registered trademark of PROFIBUS Nutzerorganisation e.V., Karlsruhe, D
S-DAT™, T-DAT™, F-Chip™, FieldTool™, FieldCheck™, Applicator™ are registered trademarks of Endress+Hauser Flowtec AG, Reinach, CH
2 Identification Promag 53 PROFIBUS-DP/-PA
10 Endress+Hauser
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 11
3 Installation
3.1 Incoming acceptance, transport and storage
3.1.1 Incoming acceptance
• Check the packaging and the contents for damage.• Check the shipment, make sure nothing is missing and that the scope of supply
matches your order.
3.1.2 Transport
The following instructions apply to unpacking and to transporting the device to its final location:• Transport the devices in the containers in which they are delivered.• Do not remove the protective plates or caps on the process connections until the
device is ready to install. This is particularly important in the case of sensors with Teflon linings.
Special notes on flanged devices
Caution:• The wooden covers mounted on the flanges before the device leaves the factory pro-
tect the linings on the flanges during storage and transportation. Do not remove these covers until immediately before the device is installed in the pipe.
• Do not lift flanged devices by the transmitter housing, or the connection housing in the case of the remote version.
Transporting flanged devices (DN ≤ 300):Use webbing slings slung round the two process connections (Fig. 3). Do not use chains, as they could damage the housing.
Warning:Risk of injury if the measuring device slips. The center of gravity of the assembled meas-uring device might be higher than the points around which the slings are slung. At all times, therefore, make sure that the device does not unexpectedly turn around its axis or slip.
Fig. 3: Transporting transmitters with DN ≤ 300
F06-
xxxx
xxxx
-22-
00-0
0-xx
-000
3 Installation Promag 53 PROFIBUS-DP/-PA
12 Endress+Hauser
Transporting flanged devices (DN ≥ 350):Use only the metal eyes on the flanges for transporting the device, lifting it and position-ing the sensor in the piping.
Caution:Do not attempt to lift the sensor with the tines of a fork-lift truck beneath the metal casing. This would buckle the casing and damage the internal magnetic coils.
Fig. 4: Transporting sensors with DN ≥ 350
3.1.3 Storage
Note the following points:• Pack the measuring device in such a way as to protect it reliably against impact for
storage (and transportation). The original packaging provides optimum protection.• The permissible storage temperature is −10...+50 °C (preferably +20 °C).• Do not remove the protective plates or caps on the process connections until the
device is ready to install. This is particularly important in the case of sensors with Teflon linings.
F06-
5xFx
xxxx
-22-
xx-x
x-xx
-001
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 13
3.2 Installation conditions
3.2.1 Dimensions
Dimensions and the fitting lengths of the transmitter and sensor are on Page 145 ff.
3.2.2 Mounting location
Correct measuring is possible only if the pipe is full. Avoid the following locations:• Highest point of a pipeline. Risk of air accumulating.• Directly upstream of a free pipe outlet in a vertical pipe.
Fig. 5: Mounting location
Installation of pumpsDo not install the sensor on the intake side of a pump. This precaution is to avoid low pressure and the consequent risk of damage to the lining of the measuring tube. Infor-mation on the lining's resistance to partial vacuum can be found on → Page 142.
It might be necessary to install pulse dampers in systems incorporating reciprocating, diaphragm or peristaltic pumps. Information on the measuring system's resistance to vibration and shock can be found on → Page 133.
Fig. 6: Installation of pumps
F06-
5xxx
xxxx
-11-
00-0
0-xx
-000
F06-
5xxx
xxxx
-11-
00-0
0-xx
-001
3 Installation Promag 53 PROFIBUS-DP/-PA
14 Endress+Hauser
Partially filled pipesPartially filled pipes with gradients necessitate a drain-type configuration. The Empty Pipe Detection function (see Page 106) offers additional protection by detecting empty or partially filled pipes.
Caution:Risk of solids accumulating. Do not install the sensor at the lowest point in the drain. It is advisable to install a cleaning valve.
Fig. 7: Installation in partially filled pipe
Down pipesInstall a siphon or a vent valve downstream of the sensor in down pipes longer than 5 meters. This precaution is to avoid low pressure and the consequent risk of damage to the lining of the measuring tube. These measures also prevent the system losing prime, which could cause air inclusions.Information on the lining's resistance to partial vacuum can be found on Page 142.
Fig. 8: Measures for installation in a down pipe (a = vent valve; b = siphon)
F06-
5xxx
xxxx
-11-
00-0
0-xx
-002
F06-
5xxx
xxxx
-11-
00-0
0-xx
-003
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 15
3.2.3 Orientation
An optimum orientation position helps avoid gas and air accumulations and deposits in the measuring tube. Promag, nevertheless, supplies a range of functions and accesso-ries for correct measuring of problematic fluids:• Electrode Cleaning Circuit (ECC) for applications with accretive fluids, e.g. electrically
conductive deposits → “Description of Device Functions” manual.• Empty Pipe Detection (EPD) ensures the detection of partially filled measuring tubes,
e.g. in the case of degassing fluids or varying process pressures (see Page 106)• Exchangeable measuring electrodes for abrasive fluids (see Page 128)
Vertical orientationThis is the ideal orientation for self-emptying piping systems and for use in conjunction with Empty Pipe Detection.
Fig. 9: Vertical orientation
Horizontal orientationThe measuring-electrode plane should be horizontal. This prevents brief insulation of the two electrodes by entrained air bubbles.
Caution:Empty Pipe Detection functions correctly with the measuring device installed horizon-tally only when the transmitter housing is facing upward (Fig. 10). Otherwise there is no guarantee that Empty Pipe Detection will respond if the measuring tube is only partially filled or empty.
Fig. 10: Horizontal orientation
1 EPD electrode for the detection of empty pipes (not with Promag H, DN 2...8)2 Measurement electrodes for the signal acquisition3 Reference electrode for the potential equalisation (not with Promag H)
F06-
5xxx
xxxx
-11-
00-0
0-xx
-004
F06-
5xxx
xxxx
-11-
00-x
x-xx
-000
3 Installation Promag 53 PROFIBUS-DP/-PA
16 Endress+Hauser
3.2.4 Inlet and outlet runs
If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc. Compliance with the following requirements for the inlet and outlet runs is necessary in order to ensure measuring accuracy:• Inlet run ≥ 5 x DN• Outlet run ≥ 2 x DN
Fig. 11: Inlet and outlet runs
3.2.5 Vibrations
Secure the piping and the sensor if vibration is severe.
Caution:It is advisable to install sensor and transmitter separately if vibration is excessively severe. Information on resistance to vibration and shock can be found on → Page 133.
Fig. 12: Measures to prevent vibration of the measuring device
F06-
5xxx
xxxx
-11-
00-0
0-xx
-005
F06-
5xxx
xxxx
-11-
00-0
0-xx
-006
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 17
3.2.6 Foundations, supports
If the nominal diameter is DN ≥ 350, mount the transmitter on a foundation of adequate load-bearing strength.
Caution:Risk of damage. Do not support the weight of the sensor on the metal casing:the casing would buckle and damage the internal magnetic coils.
Fig. 13: Correct support for large nominal diameters (DN ≥ 350)
F06-
5xFx
xxxx
-11-
05-x
x-xx
-000
3 Installation Promag 53 PROFIBUS-DP/-PA
18 Endress+Hauser
3.2.7 Adapters
Suitable adapters to (E) DIN EN 545 (double-flange reducers) can be used to install the sensor in larger-diameter pipes. The resultant increase in the rate of flow improves measuring accuracy with very slow-moving fluids.
The nomogram shown here can be used to calculate the pressure loss caused by cross-section reduction:
Note:The nomogram applies to fluids of viscosity similar to water.
1. Calculate the ratio of the diameters d/D.2. From the nomogram read off the pressure loss as a function of flow velocity
(downstream from the reduction) and the d/D ratio.
Fig. 14: Pressure loss due to adapters
3.2.8 Nominal diameter and flow rate
The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum velocity of flow is 2...3 m/s. The velocity of flow (v), moreover, has to be matched to the physical properties of the fluid:• v < 2 m/s: for abrasive fluids such as potter's clay, lime milk, ore slurry, etc.• v > 2 m/s: for fluids producing build-up such as wastewater sludge, etc.
Note:Flow velocity can be increased, if necessary, by reducing the nominal diameter of the sensor (see Page 18).
F06-
5xxx
xxxx
-05-
05-x
x-xx
-000
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 19
Promag W
Flow rate characteristic values - Promag W (SI units)
Nominal diameter
Recommendedflow rate
Factory setting
[mm] [inch]min./max. full scale value
(v ~ 0.3 or 10 m/s)Full scale value
(v ~ 2.5 m/s)Low flow cutoff(v ~ 0.04 m/s)
25 1" 9...300 dm3/min 75 dm3/min 1 dm3/min
32 1 1/4" 15...500 dm3/min 125 dm3/min 2 dm3/min
40 1 1/2" 25...700 dm3/min 200 dm3/min 3 dm3/min
50 2" 35...1100 dm3/min 300 dm3/min 5 dm3/min
65 2 1/2" 60...2000 dm3/min 500 dm3/min 8 dm3/min
80 3" 90...3000 dm3/min 750 dm3/min 12 dm3/min
100 4" 145...4700 dm3/min 1200 dm3/min 20 dm3/min
125 5" 220...7500 dm3/min 1850 dm3/min 30 dm3/min
150 6" 20...600 m3/h 150 m3/h 2.5 m3/h
200 8" 35...1100 m3/h 300 m3/h 5.0 m3/h
250 10" 55...1700 m3/h 500 m3/h 7.5 m3/h
300 12" 80...2400 m3/h 750 m3/h 10 m3/h
350 14" 110...3300 m3/h 1000 m3/h 15 m3/h
400 16" 140...4200 m3/h 1200 m3/h 20 m3/h
450 18" 180...5400 m3/h 1500 m3/h 25 m3/h
500 20" 220...6600 m3/h 2000 m3/h 30 m3/h
600 24" 310...9600 m3/h 2500 m3/h 40 m3/h
700 28" 420...13500 m3/h 3500 m3/h 50 m3/h
– 30" 480...15000 m3/h 4000 m3/h 60 m3/h
800 32" 550...18000 m3/h 4500 m3/h 75 m3/h
900 36" 690...22500 m3/h 6000 m3/h 100 m3/h
1000 40" 850...28000 m3/h 7000 m3/h 125 m3/h
− 42" 950...30000 m3/h 8000 m3/h 125 m3/h
1200 48" 1250...40000 m3/h 10000 m3/h 150 m3/h
– 54" 1550...50000 m3/h 13000 m3/h 200 m3/h
1400 – 1700...55000 m3/h 14000 m3/h 225 m3/h
− 60" 1950...60000 m3/h 16000 m3/h 250 m3/h
1600 – 2200...70000 m3/h 18000 m3/h 300 m3/h
− 66" 2500...80000 m3/h 20500 m3/h 325 m3/h
1800 72" 2800...90000 m3/h 23000 m3/h 350 m3/h
− 78" 3300...100000 m3/h 28500 m3/h 450 m3/h
2000 – 3400...110000 m3/h 28500 m3/h 450 m3/h
3 Installation Promag 53 PROFIBUS-DP/-PA
20 Endress+Hauser
Flow rate characteristic values - Promag W (US units)
Nominal diameter Recommendedflow rate
Factory setting
[inch] [mm]min./max. full scale value
(v ~ 0.3 or 10 m/s)Full scale value
(v ~ 2.5 m/s)Low flow cutoff(v ~ 0.04 m/s)
1" 25 2.5...80 gal/min 18 gal/min 0.25 gal/min
1 1/4" 32 4...130 gal/min 30 gal/min 0.50 gal/min
1 1/2" 40 7...190 gal/min 50 gal/min 0.75 gal/min
2" 50 10...300 gal/min 75 gal/min 1.25 gal/min
2 1/2" 65 16...500 gal/min 130 gal/min 2.0 gal/min
3" 80 24...800 gal/min 200 gal/min 2.5 gal/min
4" 100 40...1250 gal/min 300 gal/min 4.0 gal/min
5" 125 60...1950 gal/min 450 gal/min 7.0 gal/min
6" 150 90...2650 gal/min 600 gal/min 12 gal/min
8" 200 155...4850 gal/min 1200 gal/min 15 gal/min
10" 250 250...7500 gal/min 1500 gal/min 30 gal/min
12" 300 350...10600 gal/min 2400 gal/min 45 gal/min
14" 350 500...15000 gal/min 3600 gal/min 60 gal/min
16" 400 600...19000 gal/min 4800 gal/min 60 gal/min
18" 450 800...24000 gal/min 6000 gal/min 90 gal/min
20" 500 1000...30000 gal/min 7500 gal/min 120 gal/min
24" 600 1400...44000 gal/min 10500 gal/min 180 gal/min
28" 700 1900...60000 gal/min 13500 gal/min 210 gal/min
30" – 2150...67000 gal/min 16500 gal/min 270 gal/min
32" 800 2450...80000 gal/min 19500 gal/min 300 gal/min
36" 900 3100...100000 gal/min 24000 gal/min 360 gal/min
40" 1000 3800...125000 gal/min 30000 gal/min 480 gal/min
42" − 4200...135000 gal/min 33000 gal/min 600 gal/min
48" 1200 5500...175000 gal/min 42000 gal/min 600 gal/min
54" – 9...300 Mgal/d 75 Mgal/d 1.3 Mgal/d
– 1400 10...340 Mgal/d 85 Mgal/d 1.3 Mgal/d
60" − 12...380 Mgal/d 95 Mgal/d 1.3 Mgal/d
– 1600 13...450 Mgal/d 110 Mgal/d 1.7 Mgal/d
66" − 14...500 Mgal/d 120 Mgal/d 2.2 Mgal/d
72" 1800 16...570 Mgal/d 140 Mgal/d 2.6 Mgal/d
78" − 18...650 Mgal/d 175 Mgal/d 3.0 Mgal/d
– 2000 20...700 Mgal/d 175 Mgal/d 3.0 Mgal/d
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 21
Promag P
Flow rate characteristic values - Promag P (SI units)
Nominal diameter
Recommendedflow rate
Factory setting
[mm] [inch]min./max. full scale value
(v ~ 0.3 or 10 m/s)Full scale value
(v ~ 2.5 m/s)Low flow cutoff(v ~ 0.04 m/s)
15 1/2" 4...100 dm3/min 25 dm3/min 0.5 dm3/min
25 1" 9...300 dm3/min 75 dm3/min 1 dm3/min
32 1 1/4" 15...500 dm3/min 125 dm3/min 2 dm3/min
40 1 1/2" 25...700 dm3/min 200 dm3/min 3 dm3/min
50 2" 35...1100 dm3/min 300 dm3/min 5 dm3/min
65 2 1/2" 60...2000 dm3/min 500 dm3/min 8 dm3/min
80 3" 90...3000 dm3/min 750 dm3/min 12 dm3/min
100 4" 145...4700 dm3/min 1200 dm3/min 20 dm3/min
125 5" 220...7500 dm3/min 1850 dm3/min 30 dm3/min
150 6" 20...600 m3/h 150 m3/h 2.5 m3/h
200 8" 35...1100 m3/h 300 m3/h 5.0 m3/h
250 10" 55...1700 m3/h 500 m3/h 7.5 m3/h
300 12" 80...2400 m3/h 750 m3/h 10 m3/h
350 14" 110...3300 m3/h 1000 m3/h 15 m3/h
400 16" 140...4200 m3/h 1200 m3/h 20 m3/h
450 18" 180...5400 m3/h 1500 m3/h 25 m3/h
500 20" 220...6600 m3/h 2000 m3/h 30 m3/h
600 24" 310...9600 m3/h 2500 m3/h 40 m3/h
3 Installation Promag 53 PROFIBUS-DP/-PA
22 Endress+Hauser
Promag H
Flow rate characteristic values - Promag P (US units)
Nominal diameter Recommendedflow rate
Factory setting
[inch] [mm]min./max. full scale value
(v ~ 0.3 or ~ 10 m/s)Full scale value
(v ~ 2.5 m/s)Low flow cutoff(v ~ 0.04 m/s)
1/2" 15 1.0...27 gal/min 6 gal/min 0.10 gal/min
1" 25 2.5...80 gal/min 18 gal/min 0.25 gal/min
1 1/4" 32 4...130 gal/min 30 gal/min 0.50 gal/min
1 1/2" 40 7...190 gal/min 50 gal/min 0.75 gal/min
2" 50 10...300 gal/min 75 gal/min 1.25 gal/min
2 1/2" 65 16...500 gal/min 130 gal/min 2.0 gal/min
3" 80 24...800 gal/min 200 gal/min 2.5 gal/min
4" 100 40...1250 gal/min 300 gal/min 4.0 gal/min
5" 125 60...1950 gal/min 450 gal/min 7.0 gal/min
6" 150 90...2650 gal/min 600 gal/min 12 gal/min
8" 200 155...4850 gal/min 1200 gal/min 15 gal/min
10" 250 250...7500 gal/min 1500 gal/min 30 gal/min
12" 300 350...10600 gal/min 2400 gal/min 45 gal/min
14" 350 500...15000 gal/min 3600 gal/min 60 gal/min
16" 400 600...19000 gal/min 4800 gal/min 60 gal/min
18" 450 800...24000 gal/min 6000 gal/min 90 gal/min
20" 500 1000...30000 gal/min 7500 gal/min 120 gal/min
24" 600 1400...44000 gal/min 10500 gal/min 180 gal/min
Flow rate characteristic values - Promag H (SI units)
Nominal diameter
Recommendedflow rate
Factory settings
[mm] inch]min./max. full scale value
(v ~ 0.3 or 10 m/s)Full scale value
(v ~ 2.5 m/s)Low flow cutoff(v ~ 0.04 m/s)
2 1/12" 0.06...1.8 dm3/min 0.5 dm3/min 0.01 dm3/min
4 5/32" 0.25...7 dm3/min 2 dm3/min 0.05 dm3/min
8 5/16" 1...30 dm3/min 8 dm3/min 0.1 dm3/min
15 1/2" 4...100 dm3/min 25 dm3/min 0.5 dm3/min
25 1" 9...300 dm3/min 75 dm3/min 1 dm3/min
40 1 1/2" 25...700 dm3/min 200 dm3/min 3 dm3/min
50 2" 35...1100 dm3/min 300 dm3/min 5 dm3/min
65 2 1/2" 60...2000 dm3/min 500 dm3/min 8 dm3/min
80 3" 90...3000 dm3/min 750 dm3/min 12 dm3/min
100 4" 145...4700 dm3/min 1200 dm3/min 20 dm3/min
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 23
Flow rate characteristic values - Promag H (US units)
Nominal diameter Recommendedflow rate
Factory settings
[inch] [mm]min./max. full scale value
(v ~ 0.3 or 10 m/s)Full scale value
(v ~ 2.5 m/s)Low flow cutoff(v ~ 0.04 m/s)
1/12" 2 0.015...0.5 gal/min 0.1 gal/min 0.002 gal/min
5/32" 4 0.07...2 gal/min 0.5 gal/min 0.008 gal/min
5/16" 8 0.25...8 gal/min 2 gal/min 0.025 gal/min
1/2" 15 1.0...27 gal/min 6 gal/min 0.10 gal/min
1" 25 2.5...80 gal/min 18 gal/min 0.25 gal/min
1 1/2" 40 7...190 gal/min 50 gal/min 0.75 gal/min
2" 50 10...300 gal/min 75 gal/min 1.25 gal/min
2 1/2" 65 16...500 gal/min 130 gal/min 2.0 gal/min
3" 80 24...800 gal/min 200 gal/min 2.5 gal/min
4" 100 40...1250 gal/min 300 gal/min 4.0 gal/min
3 Installation Promag 53 PROFIBUS-DP/-PA
24 Endress+Hauser
3.2.9 Length of connecting cable
In order to ensure measuring accuracy, comply with the following instructions when installing the remote version:• Secure the cable run or route the cable in a conduit. Movement of the cable can falsify
the measuring signal, particularly if the fluid conductivity is low.• Route the cable well clear of electrical machines and switching elements.• Ensure potential equalisation between sensor and transmitter, if necessary. • Permissible cable length Lmax depends on the fluid conductivity (Fig. 15). A minimum
conductivity of 20 µS/cm is required for measuring demineralised water.
Fig. 15: Permissible cable length for the remote version
Gray shaded area = permissible area Lmax = connecting cable length in [m]Fluid conductivity in [µS/cm]
200
100
5
10 100
L max
200[m]
[ S/cm]µ
L max
F06-
xxxx
xxxx
-05-
xx-x
x-xx
-006
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 25
3.3 Installation
3.3.1 Installing the Promag W sensor
Note:Bolts, nuts, seals, etc. are not included in the scope of supply and must be supplied by the customer.
The sensor is designed for installation between the two piping flanges. Always tighten all threaded fasteners to the specified torques → Page 27 ff.
Fig. 16: Installing the Promag W sensor
SealsComply with the following instructions when installing seals:• Hard rubber lining → additional seals are always necessary!• Polyurethane lining → additional seals are recommended• For DIN flanges, use only seals acc. to DIN 2690.• Make sure that the seals do not protrude into the piping cross-section.
Caution:Risk of short circuit. Do not use electrically conductive sealing compound such as graphite. An electrically conductive layer could form on the inside of the measuring tube and short-circuit the measuring signal.
Grounding cable (DN 15...2000)If necessary, the special grounding cable for potential equalisation can be ordered as an accessory (see Page 111). Detailled assembly instructions → Page 59 ff.
F06-
5xFx
xxxx
-17-
05-x
x-xx
-000
3 Installation Promag 53 PROFIBUS-DP/-PA
26 Endress+Hauser
Assembly with ground disks (DN 15...300)Depending on the application, e.g. with lined or ungrounded pipes (see Page 58 ff.), it may be necessary to mount ground disks between the sensor and the pipe flange for potential equalisation:
Caution:• In this case, when using ground disks (including seals) the total fitting length
increases! The dimensions plus information about the material can be found on Page 155.
• Hard rubber lining → install additional seals between the sensor and grounding disk and between the grounding disk and pipe flange.
• Polyurethane lining → only install additional seals between the grounding disk and pipe flange.
1. Place ground disks and seals between the instrument and the pipe flange (Fig. 17).2. Insert the bolts through the flange holes. Tighten the nuts so that they are still loose.3. Now rotate the grounding disk as shown in Fig. 17 until the handle strikes the bolts.
This will center the grounding disk correctly.4. Now tighten the bolts to the required torque (see Page 27 ff.)5. Connect the grounding disk to ground.
Fig. 17: Assembly with ground disks (Promag W, DN 15 …300)F0
6-5x
Fxxx
xx-1
7-05
-xx-
xx-0
01
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 27
Torques of threaded fasteners (Promag W)Note the following points:• The tightening torques listed below are for lubricated threads only.• Always tighten threaded fasteners uniformly and in diagonally opposite sequence.• Overtightening the fasteners will deform the sealing faces or damage the seals.• The tightening torques listed below apply only to pipes not subjected to tensile stress.
Promag WNominal diameter
DINPressure rating
Threaded fasteners
Max. tightening torque [Nm]
[mm] [bar] Hard rubber Polyurethane
25 PN 40 4 x M 12 − 8
32 PN 40 4 x M 16 − 14
40 PN 40 4 x M 16 − 18
50 PN 40 4 x M 16 − 25
65 PN 16 4 x M 16 65 22
65 PN 40 8 x M 16 32 18
80 PN 16 8 x M 16 40 13
80 PN 40 8 x M 16 40 23
100 PN 16 8 x M 16 43 14
100 PN 40 8 x M 20 59 39
125 PN 16 8 x M 16 56 19
125 PN 40 8 x M 24 83 63
150 PN 16 8 x M 20 74 27
150 PN 40 8 x M 24 104 84
200 PN 10 8 x M 20 106 35
200 PN 16 12 x M 20 70 28
200 PN 25 12 x M 24 104 57
250 PN 10 12 x M 20 82 27
250 PN 16 12 x M 24 98 48
250 PN 25 12 x M 27 150 93
300 PN 10 12 x M 20 94 34
300 PN 16 12 x M 24 134 67
300 PN 25 16 x M 27 153 97
350 PN 10 16 x M 20 112 47
350 PN 16 16 x M 24 152 68
350 PN 25 16 x M 30 227 141
400 PN 10 16 x M 24 151 65
400 PN 16 16 x M 27 193 95
400 PN 25 16 x M 33 289 192
450 PN 10 20 x M 24 153 59
450 PN 16 20 x M 27 198 96
450 PN 25 20 x M 33 256 185
3 Installation Promag 53 PROFIBUS-DP/-PA
28 Endress+Hauser
500 PN 10 20 x M 24 155 66
500 PN 16 20 x M 30 275 132
500 PN 25 20 x M 33 317 228
600 PN 10 20 x M 27 206 93
600 PN 16 20 x M 33 415 202
600 PN 25 20 x M 36 431 342
700 PN 10 24 x M 27 246 105
700 PN 16 24 x M 33 278 202
700 PN 25 24 x M 39 449 397
800 PN 10 24 x M 30 331 150
800 PN 16 24 x M 36 369 283
800 PN 25 24 x M 45 664 589
900 PN 10 28 x M 30 316 159
900 PN 16 28 x M 36 353 299
900 PN 25 28 x M 45 690 619
1000 PN 10 28 x M 33 402 210
1000 PN 16 28 x M 39 502 401
1000 PN 25 28 x M 52 970 871
1200 PN 6 32 x M 30 319 138
1200 PN 10 32 x M 36 564 289
1200 PN 16 32 x M 45 701 575
1400 PN 6 36 x M 33 430 181
1400 PN 10 36 x M 39 654 368
1400 PN 16 36 x M 45 729 675
1600 PN 6 40 x M 33 440 208
1600 PN 10 40 x M 45 946 503
1600 PN 16 40 x M 52 1007 915
1800 PN 6 44 x M 36 547 262
1800 PN 10 44 x M 45 961 566
1800 PN 16 44 x M 52 1108 1023
2000 PN 6 48 x M 39 629 316
2000 PN 10 48 x M 45 1047 636
2000 PN 16 48 x M 56 1324 1241
Promag WNominal diameter
DINPressure rating
Threaded fasteners
Max. tightening torque [Nm]
[mm] [bar] Hard rubber Polyurethane
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 29
Promag WNominal diameter
AWWAPressure rating
Threaded fasteners
Max. tightening torque [Nm]
[mm] [inch] Hard rubber Polyurethane
700 28" Class D 28 x 1 1/4" 247 109
750 30" Class D 28 x 1 1/4 287 126
800 32" Class D 28 x 1 1/2" 394 170
900 36" Class D 32 x 1 1/2" 419 186
1000 40" Class D 36 x 1 1/2" 420 200
1050 42" Class D 36 x 1 1/2" 528 226
1200 48" Class D 44 x 1 1/2" 552 240
1350 54" Class D 44 x 1 3/4" 730 345
1500 60" Class D 52 x 1 3/4" 758 359
1650 66" Class D 52 x 1 3/4" 946 436
1800 72" Class D 60 x 1 3/4" 975 449
2000 78" Class D 64 x 2" 853 552
Promag WNominal diameter
ANSIPressure rating
Threaded fasteners
Max. tightening torque [Nm]
[mm] [inch] [lbs] Hard rubber Polyurethane
25 1" Class 150 4 x 1/2" − 3
25 1" Class 300 4 x 5/8" − 8
40 1 1/2" Class 150 4 x 1/2" − 6
40 1 1/2" Class 300 4 x 3/4" − 20
50 2" Class 150 4 x 5/8" − 12
50 2" Class 300 8 x 5/8" − 13
80 3" Class 150 4 x 5/8" 60 20
80 3" Class 300 8 x 3/4" 38 30
100 4" Class 150 8 x 5/8" 42 15
100 4" Class 300 8 x 3/4" 58 45
150 6" Class 150 8 x 3/4" 79 33
150 6" Class 300 12 x 3/4" 70 57
200 8" Class 150 8 x 3/4" 107 51
250 10" Class 150 12 x 7/8" 101 57
300 12" Class 150 12 x 7/8" 133 78
350 14" Class 150 12 x 1" 135 105
400 16" Class 150 16 x 1" 128 102
450 18" Class 150 16 x 1 1/8" 204 147
500 20" Class 150 20 x 1 1/8" 183 142
600 24" Class 150 20 x 1 1/4" 268 218
3 Installation Promag 53 PROFIBUS-DP/-PA
30 Endress+Hauser
Promag WNominal diameter
JISPressure rating
Threaded fasteners
Max. tightening torque [Nm]
[mm] Hard rubber Polyurethane
25 20K 4 x M 16 − 9
32 20K 4 x M 16 − 12
40 20K 4 x M 16 − 13
50 10K 4 x M 16 − 13
50 20K 8 x M 16 − 9
65 10K 4 x M 16 55 18
65 20K 8 x M 16 28 14
80 10K 8 x M 16 29 10
80 20K 8 x M 20 42 22
100 10K 8 x M 16 35 12
100 20K 8 x M 20 56 32
125 10K 8 x M 20 60 20
125 20K 8 x M 22 91 52
150 10K 8 x M 20 75 25
150 20K 12 x M 22 81 48
200 10K 12 x M 20 61 23
200 20K 12 x M 22 91 69
250 10K 12 x M 22 100 39
250 20K 12 x M 24 159 118
300 10K 16 x M 22 74 38
300 20K 16 x M 24 138 116
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 31
3.3.2 Installing the Promag P sensor
Caution:• The protective covers mounted on the two sensor flanges guard the Teflon (PTFE) lin-
ing, which is turned over the flanges. Consequently, do not remove these covers until immediately before the sensor is installed in the pipe.
• The covers must remain in place while the device is in storage.• Make sure that the lining is not damaged or removed from the flanges.
Note:Bolts, nuts, seals, etc. are not included in the scope of supply and must be supplied by the customer.
The sensor is designed for installation between the two piping flanges. Always tighten all threaded fasteners to the specified torques → Page 34.
Fig. 18: Installing the Promag P sensor
SealsComply with the following instructions when installing seals:• Measuring tube linings with PFA or PTFE → No seals are required.• For DIN flanges, use only seals acc. to DIN 2690.• Make sure that the seals do not protrude into the piping cross-section.
Caution:Risk of short circuit. Do not use electrically conductive sealing compound such as graphite. An electrically conductive layer could form on the inside of the measuring tube and short-circuit the measuring signal.
Grounding cable (DN 15...600)If necessary, the special grounding cable for potential equalisation can be ordered as an accessory (see Page 111). Detailled assembly instructions → Page 59 ff.
F06-
5xFx
xxxx
-17-
05-x
x-xx
-000
3 Installation Promag 53 PROFIBUS-DP/-PA
32 Endress+Hauser
Assembly with ground disks (DN 15...300)Depending on the application, e.g. with lined or ungrounded pipes (see Page 58 ff.), it may be necessary to mount ground disks between the sensor and the pipe flange for the potential equalisation:
Caution:• In this case, when using ground disks (including seals) the total fitting length
increases! The dimensions plus information about the material can be found on Page 155.
• PTFE and PFA lining → only install additional seals between the grounding disk and pipe flange.
1. Place ground disks and seals between the instrument and the pipe flange (Fig. 19).2. Insert the bolts through the flange holes. Tighten the nuts so that they are still loose.3. Now rotate the grounding disk as shown in Fig. 19 until the handle strikes the bolts.
This will center the grounding disk correctly.4. Now tighten the bolts to the required torque (see Page 34 ff.)5. Connect the grounding disk to ground.
Fig. 19: Assembly with ground disks (Promag P, DN 15…300)
F06-
5xFx
xxxx
-17-
05-x
x-xx
-001
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 33
Installing the high-temperature version (with PFA/PTFE lining)The high-temperature version has a housing support for the thermal separation of sen-sor and transmitter. The high-temperature version is always used for applications in which high ambient temperatures are encountered in conjunction with high fluid tempe-ratures. The high-temperature version is obligatory if the fluid temperature exceeds +150 °C.
Note:Information on permissible temperature ranges → Page 134
InsulationPipes generally have to be insulated if they carry either very hot or cryogenic fluids, in order to avoid energy losses and to prevent accidental contact with pipes at tempera-tures that could cause injury. Guidelines regulating the insulation of pipes have to be taken into account.
Caution:Risk of measuring electronics overheating! The housing support dissipates heat and its entire surface area must remain uncovered. Make sure that the sensor insulation does not extend past the top of the two sensor shells (Fig. 20).
Fig. 20: Promag P (high-temperature version): Insulating the pipe
F06-
5xP
xxxx
x-17
-05-
00-x
x-00
0
3 Installation Promag 53 PROFIBUS-DP/-PA
34 Endress+Hauser
Tightening torques for threaded fasteners (Promag P)Note the following points:• The tightening torques listed below are for lubricated threads only. • Always tighten threaded fasteners uniformly and in diagonally opposite sequence.• Overtightening the fasteners will deform the sealing faces or damage the seals.• The tightening torques listed below apply only to pipes not subjected to tensile stress.
Promag PNominal diameter
DINPressure rating
Threaded fasteners
Max. tightening torque [Nm]
[mm] [bar] PTFE PFA
15 PN 40 4 x M 12 11 −
25 PN 40 4 x M 12 26 19
32 PN 40 4 x M 16 41 33
40 PN 40 4 x M 16 52 42
50 PN 40 4 x M 16 65 55
65 PN 16 4 x M 16 87 73
65 PN 40 8 x M 16 43 36
80 PN 16 8 x M 16 53 44
80 PN 40 8 x M 16 53 44
100 PN 16 8 x M 16 57 45
100 PN 40 8 x M 20 78 63
125 PN 16 8 x M 16 75 59
125 PN 40 8 x M 24 111 87
150 PN 16 8 x M 20 99 73
150 PN 40 8 x M 24 136 104
200 PN 10 8 x M 20 141 100
200 PN 16 12 x M 20 94 67
200 PN 25 12 x M 24 138 104
250 PN 10 12 x M 20 110 −
250 PN 16 12 x M 24 131 −
250 PN 25 12 x M 27 200 −
300 PN 10 12 x M 20 125 −
300 PN 16 12 x M 24 179 −
300 PN 25 16 x M 27 204 −
350 PN 10 16 x M 20 188 −
350 PN 16 16 x M 24 254 −
350 PN 25 16 x M 30 380 −
400 PN 10 16 x M 24 260 −
400 PN 16 16 x M 27 330 −
400 PN 25 16 x M 33 488 −
450 PN 10 20 x M 24 235 −
450 PN 16 20 x M 27 300 −
450 PN 25 20 x M 33 385 −
500 PN 10 20 x M 24 265 −
500 PN 16 20 x M 30 448 −
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 35
500 PN 25 20 x M 33 533 −
600 PN 10 20 x M 27 345 −
600 PN 16 20 x M 33 658 −
600 PN 25 20 x M 36 731 −
Promag PNominal diameter
ANSIPressure rating
Threaded fasteners
Max. tightening torque [Nm]
[mm] [inch] [lbs] PTFE PFA
15 1/2" Class 150 4 x 1/2" 6 −
15 1/2" Class 300 4 x 1/2" 6 −
25 1" Class 150 4 x 1/2" 11 9
25 1" Class 300 4 x 5/8" 14 11
40 1 1/2" Class 150 4 x 1/2" 24 19
40 1 1/2" Class 300 4 x 3/4" 34 28
50 2" Class 150 4 x 5/8" 47 38
50 2" Class 300 8 x 5/8" 23 19
80 3" Class 150 4 x 5/8" 79 67
80 3" Class 300 8 x 3/4" 47 40
100 4" Class 150 8 x 5/8" 56 49
100 4" Class 300 8 x 3/4" 67 58
150 6" Class 150 8 x 3/4" 106 91
150 6" Class 300 12 x 3/4" 73 67
200 8" Class 150 8 x 3/4" 143 121
250 10" Class 150 12 x 7/8" 135 −
300 12" Class 150 12 x 7/8" 178 −
350 14" Class 150 12 x 1" 260 −
400 16" Class 150 16 x 1" 246 −
450 18" Class 150 16 x 1 1/8" 371 −
500 20" Class 150 20 x 1 1/8" 341 −
600 24" Class 150 20 x 1 1/4" 477 −
Promag PNominal diameter
JISPressure rating
Threaded fasteners
Max. tightening torque [Nm]
[mm] PTFE PFA
15 20K 4 x M 12 16 −
25 20K 4 x M 16 32 −
32 20K 4 x M 16 38 −
40 20K 4 x M 16 41 −
50 10K 4 x M 16 54 −
50 20K 8 x M 16 27 −
65 10K 4 x M 16 74 −
Promag PNominal diameter
DINPressure rating
Threaded fasteners
Max. tightening torque [Nm]
[mm] [bar] PTFE PFA
3 Installation Promag 53 PROFIBUS-DP/-PA
36 Endress+Hauser
65 20K 8 x M 16 37 −
80 10K 8 x M 16 38 −
80 20K 8 x M 20 57 −
100 10K 8 x M 16 47 −
100 20K 8 x M 20 75 −
125 10K 8 x M 20 80 −
125 20K 8 x M 22 121 −
150 10K 8 x M 20 99 −
150 20K 12 x M 22 108 −
200 10K 12 x M 20 82 −
200 20K 12 x M 22 121 −
250 10K 12 x M 22 133 −
250 20K 12 x M 24 212 −
300 10K 16 x M 22 99 −
300 20K 16 x M 24 183 −
Promag PNominal diameter
JISPressure rating
Threaded fasteners
Max. tightening torque [Nm]
[mm] PTFE PFA
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 37
3.3.3 Installing the Promag H sensor
The Promag H is supplied to order, with or without pre-installed process connections. Pre-installed process connections are secured to the sensor with hex-head threaded fasteners.
Caution:• If you intend to use your own process connections, make up the process adapters as
specified on Page 160 ff..• The sensor might require support or additional attachments, depending on the appli-
cation and the length of the piping run. A wall-mounting kit can be ordered separately from E+H as an accessory (see Page 111).
Fig. 21: Promag H process connections (DN 2...25, DN 40...100)
A: DN 2...25 / process connections with O-rings:Welding flanges (ISO 2463, IPS), flanges (DIN 2635, ANSI B16.5, JIS B2238), PVDF flanges (DIN 2501, ANSI B16.5, JIS B2238), external and internal pipe threads (ISO / DIN), hose con-nections, PVC adhesive fitting
B: DN 2...25 / process connections with aseptic gasket seals:Weld nipples (DIN 11850, ODT), Tri-Clamp, Clamp (ISO 2852, DIN 32676), coupling (DIN 11851, DIN 11864-1, SMS 1145), flange DIN 11864-2
C: DN 40...100 / process connections with aseptic gasket seals:Weld nipples (DIN 11850, ODT), Tri-Clamp, Clamp (ISO 2852, DIN 32676), coupling (DIN 11851, DIN 11864-1, ISO 2853, SMS 1145), flange DIN 11864-2
SealsWhen installing the process connections, make sure that the seals are clean and cor-rectly centered. Firmly tighten the threaded fasteners. The process connection forms a metallic connection with the sensor, which ensures a defined compression of the seal.
Caution:The seals must be replaced periodically, depending on the application, particularly in the case of gasket seals (aseptic version). The period between changes depends on the frequency of cleaning cycles, the cleaning temperature and the fluid temperature.Replacement seals can be ordered as accessories → Page 111.
F06-
xxH
xxxx
x-17
-05-
xx-x
x-00
0
3 Installation Promag 53 PROFIBUS-DP/-PA
38 Endress+Hauser
Usage and assembly of ground rings (DN 2...25)In case the process connections are made of plastic (e.g. flanges or adhesive fittings), the potential between the sensor and the fluid must be equalised using additional ground rings. If the ground rings are not installed this can affect the accuracy of the measurements or cause the destruction of the sensor through the electrochemical erosion of the elec-trodes.
Caution:At the factory, plastic rings will be inserted in the process connections instead of the ground rings. These are only intended as “placeholders” and do not provide potential equalisation.
Before installing the ground rings, please observe the following points:• ground rings can be ordered separately from E+H as accessories (see Page 111).
– When placing the order, make certain the included o-ring seals are of the same material as the seals that are used for the process connections.
– When placing the order, make certain that the grounding ring is compatible with the material used for the electrodes. Otherwise, there is a risk that the electrodes may be destroyed by electrochemical corrosion! Information about the materials can be found on page 137.
• The dimensions of the ground rings can be found on page 167.• ground rings, including the seals, are mounted within the process connections. There-
fore, the fitting length is not affected.• For plastic flange connections, ground disks, which are mounted between the
flanges, can be used instead of ground rings.
1. Loosen the four hexagonal headed bolts (1) and remove the process connection from the sensor (5).
2. Remove the plastic disk (3), including the two o-ring seals.3. Place the new seal (2) in the groove of the process connection.4. Place the metal ground ring (3) on the process connection.5. Now place the second seal (4) in the groove of the ground ring.6. Finally, mount the process connection on the sensor again.
Fig. 22: Installing ground rings with a Promag H (DN 2...25)
1 Hexagonal headed bolts (process connection)2 O-ring seal for the process connection3 Plastic disk (placeholder) or ground ring4 O-ring seal for the ground ring5 Sensor Promag H
1
3 4
5
2
F06-
xxH
xxxx
x-17
-xx-
xx-x
x-00
1
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 39
Welding the sensor into the piping (weld nipple)
Caution:Risk of destroying the measuring electronics. Make sure that the welding machine is not grounded via the sensor or the transmitter.
1. Tack-weld the Promag H sensor into the pipe. A suitable welding jig can be ordered separately from E+H as an accessory (see Page 111).
2. Remove the threaded fasteners from the process-connection flange. Remove the sensor complete with seal from the pipe.
3. Weld the process connection into the pipe.4. Reinstall the sensor in the pipe. Make sure that everything is clean and that the seal
is correctly seated.
Note:• If thin-walled foodstuffs pipes are not welded correctly, the heat could damage the
installed seal. It is therefore advisable to remove the sensor and the seal prior to welding.
• The pipe has to be spread approximately 8 mm to permit disassembly.
Cleaning with pigsIf pigs are used for cleaning, it is essential to take the inside diameters of measuring tube and process connection into account (see Page 161 ff.).
3 Installation Promag 53 PROFIBUS-DP/-PA
40 Endress+Hauser
3.3.4 Turning the transmitter housing
Turning the aluminum field housing
Warning:The turning mechanism in devices with EEx d/de or FM/CSA Cl. I Div. 1 classification is not the same as that described here. The procedure for turning these housings is described in the Ex-specific documentation.
1. Loosen the two securing screws.2. Turn the bayonet catch as far as it will go.3. Carefully lift the transmitter housing as far as it will go.4. Turn the transmitter housing to the desired position (max. 2 x 90° in either direction).5. Lower the housing into position and re-engage the bayonet catch.6. Retighten the two securing screws.
Fig. 23: Turning the transmitter housing (aluminum field housing)
Turning the stainless-steel field housing
1. Loosen the two securing screws.2. Carefully lift the transmitter housing as far as it will go.3. Turn the transmitter housing to the desired position (max. 2 x 90° in either direction).4. Lower the housing into position.5. Retighten the two securing screws.
Fig. 24: Turning the transmitter housing (stainless steel field housing)
F06-
xxxx
xxxx
-17-
06-x
x-xx
-000
F06-
xxxx
xxxx
-17-
06-x
x-xx
-001
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 41
3.3.5 Installing the wall-mount transmitter housing
There are various ways of installing the wall-mount transmitter housing:• Mounted directly on the wall (without mounting kit)• Installation in control panel (with separate mounting kit, accessories → Page 111)• Pipe mounting (with separate mounting kit, accessories → Page 111)
Caution:• Make sure that ambient temperature does not exceed the permissible range
(–20 °...+60 °C). Install the device at a shady location. Avoid direct sunlight.• Always install the wall-mount housing in such a way that the cable entries are pointing
down.
Direct wall mounting
1. Drill the holes as illustrated in Fig. 25.2. Remove the cover of the connection compartment (a).3. Push the two securing screws (b) through the appropriate bores (c) in the housing.
– Securing screws (M6): max. Ø 6.5 mm– Screw head: max. Ø 10.5 mm
4. Secure the transmitter housing to the wall as indicated.5. Screw the cover of the connection compartment (a) firmly onto the housing.
Fig. 25: Mounted directly on the wallF0
6-xx
xxxx
xx-1
7-03
-xx-
xx-0
00
3 Installation Promag 53 PROFIBUS-DP/-PA
42 Endress+Hauser
Panel installation
1. Prepare the opening in the panel.2. Slide the housing into the opening in the panel from the front.3. Screw the fasteners onto the wall-mount housing.4. Place the threaded rods in the fasteners and screw them down until the housing is
seated tightly against the panel. Afterwards, tighten the locking nuts. Additional support is not necessary.
Abb. 26: Panel Installation (wall-mount housing)
Pipe mountingThe assembly should be performed by following the instructions in Fig. 27.
Caution:If the device is mounted to a warm pipe, make certain that the ambient temperature does not exceed +60 °C, which is the maximum permissible temperature.
Fig. 27: Pipe mounting (wall-mount housing)
245
~110
+0.
5–
0.5
210+0.5– 0.5
F0
6-xx
xxxx
xx-0
6-03
-06-
xx-0
02
Ø 2
0...7
0
~155
F0
6-xx
xxxx
xx-0
6-03
-06-
xx-0
01
Promag 53 PROFIBUS-DP/-PA 3 Installation
Endress+Hauser 43
3.3.6 Turning the local display
1. Remove the cover of the electronics compartment.2. Press the side latches on the display module and remove it from the electronics
compartment cover plate.3. Rotate the display to the desired position (max. 4 x 45° in each direction), and place
it back into the electronics compartment cover plate.4. Screw the cover of the electronics compartment firmly onto the transmitter housing.
Fig. 28: Turning the local display (field housing)
F06-
xxxx
xxxx
-07-
xx-0
6-xx
-000
3 Installation Promag 53 PROFIBUS-DP/-PA
44 Endress+Hauser
3.4 Installation check
Perform the following checks after installing the measuring device in the pipe:
Device condition and specifications Notes
Is the device damaged (visual inspection)? −
Does the device correspond to specifications at the measuring point, including process temperature and pressure, ambient temperature, minimum fluid conductivity, measuring range, etc.?
see Page 131 ff.
Installation Notes
Does the arrow on the sensor nameplate match the direction of flow through the pipe?
−
Is the plane of the measuring-electrode axis correct? horizontal
Is the position of the Empty Pipe Detection (EPD) electrode correct? see Page 15
Were all threaded fasteners tightened to the specified torques when the sensor was installed?
see Section 3.3
Hard rubber lining and ground disks:Were the correct seals installed (type, material, installation)?
Promag W → Page 25Promag P → Page 31Promag H → Page 37
Are the measuring point number and labeling correct (visual inspection)?
−
Process environment / process conditions Notes
Are the inlet and outlet runs respected? Inlet run ≥ 5 x DNOutlet run ≥ 2 x DN
Is the measuring device protected against moisture and direct sunlight?
−
Is the sensor adequately protected against vibration (attachment, sup-port)?
Acceleration up to 2 gby analogy with IEC 68-2-6
Promag 53 PROFIBUS-DP/-PA 4 Wiring
Endress+Hauser 45
4 Wiring
Warning:• When connecting Ex-certified devices, see the notes and diagrams in the Ex-specific
supplement to this Operating Instruction. Please do not hesitate to contact your E+H representative if you have any questions.
• If you use remote versions, connect each sensor only to the transmitter having the same serial number. Measuring errors can occur if the devices are not connected in this way.
4.1 Cable specifications for PROFIBUS-DP/-PA
4.1.1 PROFIBUS-DP: Cable specifications
Two types of cable are specified for the bus in the EN 50 170 standard. Cable type A can be used for all transmission rates up to 12 Mbit/s. The cable parameters can be taken from the following table:
When setting up the bus, observe the following points:• The maximum cable length (segment length) of a PROFIBUS-DP system depends on
the transmission rate. With PROFIBUS-RS 485 Cable Type A, this value is:
• A maximum of 32 stations are permitted per segment.• Each segment is terminated at both ends with a terminating resistor.• The bus length or number of users can be increased by installing a repeater.• The first and last segments can support a max. of 31 devices. The segments between
repeaters can support a max. of 30 stations.• The maximum distance achievable between two bus users is calculated as:
(NUM_REP + 1) x segment lengthNUM_REP = maximum number of repeaters, which can be placed in series, depend-ent on the respective repeater.
Example:According to the manufacturer's information, a maximum of 9 repeaters may be placed in series on a standard line.The maximum distance between two bus users at a transmission rate of 1.5 MBit/s is thus: (9 + 1) x 200 m = 2000 m
Cable Type A
Characteristic impedance 135... 165 Ω at a measurement frequency of 3...20 MHz
Cable capacitance <30 pF/m
Wire size >0.34 mm2, equals AWG 22
Cable type twisted pairs, 1 x 2, 2 x 2 or 1 x 4 conductors
Loop resistance 110 Ω/km
Signal attenuation max. 9 dB over the entire length of the line segment
Shielding Copper braided shield or braided shield and foil screen
Transmission Rate [kBit/s] 9.6...93.75 187.5 500 1500 300...12000
Cable length [m] 1200 1000 400 200 100
4 Wiring Promag 53 PROFIBUS-DP/-PA
46 Endress+Hauser
Stubs (PROFIBUS -DP)Note the following points:• Total combined length of all stubs < 6.6 m (at a max. of 1.5 MBit/s)• At transmission rates >1.5 MBit/s, stubs should not be used.
The line between the cable connector and the bus driver in the field device is called a stub. Our experience with the systems, indicates that you should be quite careful with the length of the stubs when planning your project. Therefore, we recommend that you do not attempt to utilise the full theoretical maximum total combined length of 6.6 m for all stubs at 1.5 MBit/s. The order of the respective field devices makes more of a difference in this case. We recommend that at transmission rates >1.5 MBit/s you avoid using stubs.
• If you must use stubs, do not install terminating resistors on them.
Shielding and grounding (PROFIBUS-DP/-PA)When planning the shielding and grounding for a field bus system, there are three important points to consider:• Electromagnetic compatibility (EMC)• Explosion protection• Safety of the personnel
To ensure the optimum electromagnetic compatibility of systems, it is important that the system components and above all the cables, which connect the components, are shielded and that no portion of the system is unshielded.Ideally, the cable shields will be connected to the field devices' housings, which are usually metal. Since these housings are generally connected to the protective ground conductor, the shield of the bus cable will thus be grounded many times.This approach, which provides the best electromagnetic compatibility and personnel safety, can be used without restriction in plants with good potential equalisation.In the case of plants without potential equalisation, a mains frequency (50 Hz) equalis-ing current can flow between two grounding points, which in unfavorable cases, e.g. when it exceeds the permissible shield current, may destroy the cable. To suppress the low frequency equalising currents on systems lacking potential equal-isation, it is therefore advisable to connect the cable shield directly to the building (or protective ground conductor) at only one end and to use capacitive coupling to connect it to all other grounding points.
Promag 53 PROFIBUS-DP/-PA 4 Wiring
Endress+Hauser 47
Setting the terminatorsSince mismatches in the impedance result in signal reflections on the line and can thus lead to communication errors, it is important to terminate the lines properly.
Warning:Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power supply is switched off before you remove the cover of the electronics com-partment.
The terminator switches are located on the I/O board (see Fig. 29).
• For baud rates of up to 1.5 MBaud, terminate the last transmitter on the bus by setting the terminator switch SW 1 to: ON – ON – ON – ON.
• If the device is to be operated at over 1.5 Mbaud, you can tap the supply voltage for an external terminator from terminals 24 (GND) and 25 (+5 V) (see Fig. 33).
• If the device is to be operated at a baud rate >1.5 Mbaud, an external terminator is necessary, e.g. with a 9-pin Sub D cable connector combination, with an integrated series inductance to compensate for the station's capacitive load and minimise the resulting line reflections.
Fig. 29: Setting the terminators (PROFIBUS-DP)
A = Factory settingB = Setting on the last transmitter
Note:As a rule, we recommend that an external terminator be used, since a defect in an inter-nally terminated device can disrupt the entire segment.
F06-
53xP
Bxx
x-04
-00-
xx-x
x-00
2
1 2
O N
3 41 2
O N
3 4
1 2
O N
3 41 2
O N
3 4
390 Ω
SW1
220 Ω390 Ω
+5V
ONOFF
4321
A
SW1
390 Ω220 Ω390 Ω
+5V
ONOFF
4321
B
4 Wiring Promag 53 PROFIBUS-DP/-PA
48 Endress+Hauser
4.1.2 PROFIBUS-PA: Cable specifications
Cable typeTwin-core cable is required for connecting the device to the fieldbus. By analogy with IEC 61158-2 protocol four different cable types (A, B, C, D) can be used with PROFIBUS protocol, only two of which (cable types A and B) are shielded.• Cable types A or B are particularly preferable for new installations. Only these types
have cable shielding that guarantees adequate protection from electromagnetic inter-ference and thus the most reliable data transfer. On multi-pair cables (Type B), it is permissible to operate multiple fieldbuses (with the same degree of protection) on one cable. No other circuits are permissible in the same cable.
• Practical experience has shown that cable types C and D should not be used due to the lack of shielding, since the freedom from interference generally does not meet the requirements described in the standard.
The electrical data of the fieldbus cable has not been specified but determines impor-tant characteristics of the design of the fieldbus, such as distances bridged, number of participants, electromagnetic compatibility, etc.
Suitable fieldbus cables from various manufacturers for the non-hazardous area are listed below:• Siemens: 6XV1 830-5BH10• Belden: 3076F• Kerpen: CeL-PE/OSCR/PVC/FRLA FB-02YS(ST)YFL
Cable type A Cable type B
Cable structure twisted pair, shielded one or more twisted pairs, fully shielded
Wire size 0.8 mm2 (AWG 18) 0.32 mm2 (AWG 22)
Loop resistance (DC) 44 Ω/km 112 Ω/km
Impedance at 31.25 kHz 100 Ω ± 20% 100 Ω ± 30%
Attenuation at 39 kHz 3 dB/km 5 dB/km
Capacitive asymmetry 2 nF/km 2 nF/km
Envelope delay distortion (7.9 through 39 kHz)
1.7 µs/km *
Shield coverage 90% *
Max. cable length (inc. spurs >1 m) 1900 m 1200 m
* not specified
Promag 53 PROFIBUS-DP/-PA 4 Wiring
Endress+Hauser 49
Maximum overall cable lengthThe maximum network expansion depends on the type of ignition protection and the cable specifications. The overall cable length is made up of the length of the main cable and the length of all spurs (>1 m). Note the following points:• The maximum permissible overall cable length depends on the cable type used.
• If repeaters are used the maximum permissible cable length is doubled.A maximum of four repeaters are permitted between station and master.
Maximum spur lengthThe line between distribution box and field device is described as a spur.In the case of non ex-rated applications the max. length of a spur depends on the number of spurs (>1 m):
Number of field devicesIn systems that meet FISCO in the EEx ia type of protection, the line length is limited a max. of 1000 m.A maximum of 32 stations per segment in non-explosive areas or a maximum of 10 sta-tions in an Ex-area (EEx ia IIC) are possible. The actual number of stations must be determined during the project planning.
Bus terminationThe start and end of each fieldbus segment are always to be terminated with a bus ter-minator. With various junction boxes (not ex-rated) the bus termination can be activated via a switch. If this is not the case a separate bus terminator must be installed. Note the following points in addition:• In the case of a branched bus segment the device furthest from the segment connec-
tor represents the end of the bus.• If the fieldbus is extended with a repeater then the extension must also be terminated
at both ends.
Shielding and grounding(see Cable specifications PROFIBUS-DP Page 46)
Further informationGeneral information and further notes regarding the wiring can be found in the Operat-ing Instructions BA 198F/00/en “Field communications - PROFIBUS-DP/-PA:Guides for project planning and commissioning”.
Type A 1900 m
Type B 1200 m
Number of spurs 1 ... 12 13... 14 15... 18 19... 24 25... 32
Max. length per spur 120 m 90 m 60 m 30 m 1 m
4 Wiring Promag 53 PROFIBUS-DP/-PA
50 Endress+Hauser
4.2 Connecting the remote version
4.2.1 Connecting Promag W / P / H
Warning:• Risk of electric shock. Switch off the power supply before opening the device. Do not
install or wire the device while it is connected to the power supply. Failure to comply with this precaution can result in irreparable damage to the electronics.
• Risk of electric shock. Connect the protective conductor to the ground terminal on the housing before the power supply is applied.
Procedure (Fig. 30, Fig. 31):
1. Transmitter: Loosen the screws and remove cover (a) from the connection compartment.
2. Sensor: Remove cover (b) from the connection housing.3. Feed signal cable (c) and coil cable (d) through the appropriate cable entries.
Caution:– Make sure the connecting cables are secured.– Risk of damaging the coil driver. Always switch off the power supply before
connecting or disconnecting the coil cable.
4. Establish the connections between sensor and transmitter in accordance with the wiring diagram:→ Fig. 30, Fig. 31→ wiring diagram inside the cover
Caution:Insulate the shields of cables that are not connected to eliminate the risk of short-circuits with neighboring cable shields inside the sensor connection housing.
5. Transmitter: Secure cover (a) on the connection compartment.6. Sensor: Secure cover (b) on the connection housing.
Promag 53 PROFIBUS-DP/-PA 4 Wiring
Endress+Hauser 51
Fig. 30: Connecting the Promag W/P remote version
a = cover of the connection compartment, b = cover of the sensor connection housing, c = signal cable, d = coil current cable, n.c. = not connected, insulated cable shields
Fig. 31: Connecting the Promag H remote version
a = cover of the connection compartment, b = cover of the sensor connection housing, c = signal cable, d = coil current cable, n.c. = not connected, insulated cable shields
6 5
5
7
7
8 4
4
37
37
36 42
42
41
41
E1 E2 GND E
2 1
a
c d
b
S1 E1 E2 S2 GND E S
n.c. n.c.n.c.
brn
wht
grn
yel
EPD Coil circuitPipe
Electrode circuit
Meas. signal
F06-
5xFx
xxxx
-04-
xx-x
x-en
-000
6 5
5
7
7
8 4
4
37
37
36 42
42
41
41
E1 E2
DN 40...100 DN 2...25
GND E
2 1
a
c d
b
S1 E1 E2 S2 GND E S
n.c. n.c.n.c.
brn
wht
grn
yel
EPD Coil circuitPipe
Electrode circuit
Meas. signal
F06-
5xH
xxxx
x-04
-xx-
xx-e
n-00
0
4 Wiring Promag 53 PROFIBUS-DP/-PA
52 Endress+Hauser
4.2.2 Cable specifications
Coil cable• 2 x 0.75 mm2 PVC cable with common, braided copper shield (Ø approx. 7 mm)• Conductor resistance: ≤ 37 Ω/km• Capacitance: core/core, shield grounded: ≤ 120 pF/m• Permanent operating temperature: –20...+80 °C
Signal cable:• 3 x 0.38 mm2 PVC cable with common, braided copper shield (Ø approx. 7 mm) and
individually shielded cores• With Empty Pipe Detection (EPD): 4 x 0.38 mm2 PVC cable with common, braided
copper shield (Ø approx. 7 mm) and individually shielded cores• Conductor resistance: ≤ 50 Ω/km• Capacitance: core/shield: ≤ 420 pF/m• Permanent operating temperature: –20...+80 °C
Fig. 32: a = Signal cable, b = Coil current cable
1 = Core2 = Core insulation3 = Core shield4 = Core jacket5 = Core reinforcement6 = Cable shield7 = Outer jacket
As an option, E+H can also deliver reinforced connecting cables with an additional, reinforcing metal braid. We recommend such cables for the following cases:• Directly buried cable• Cables endangered by rodents• Device operation which should comply with the IP 68 standard of protection
Operation in zones of severe electrical interference:The measuring device complies with the general safety requirements in accordance with EN 61010, the EMC requirements of EN 61326, and NAMUR recommendation NE 21.
Caution:Grounding is by means of the ground terminals provided for the purpose inside the con-nection housing. Keep the stripped and twisted lengths of cable shield to the terminals as short as possible.
1234567
a b
F06
-5xW
xxxx
x-04
-11-
08-x
x-00
3
Promag 53 PROFIBUS-DP/-PA 4 Wiring
Endress+Hauser 53
4.3 Connecting the measuring unit
4.3.1 Connecting the transmitter
Warning:• Risk of electric shock. Switch off the power supply before opening the device.
Do not install or wire the device while it is connected to the power supply. Failure to comply with this precaution can result in irreparable damage to the electronics.
• Risk of electric shock. Connect the protective conductor to the ground terminal on the housing before the power supply is applied (not necessary if the power supply is gal-vanically isolated).
• Compare the specifications on the nameplate with the local voltage supply and fre-quency. The national regulations governing the installation of electrical equipment also apply.
Procedure (Fig. 33, Fig. 34):
1. Remove the cover of the connection compartment (f) from the transmitter housing.2. Feed the power-supply cable (a) and PROFIBUS cable (b) through the appropriate
cable entries.
Note:The Promag 53 can also be supplied with the option of a ready-mounted fieldbus connector. More information on this can be found on Page 56.
3. Connecting the cables:– Wiring diagram (aluminum and stainless steel housings) → Fig. 33– Wiring diagram (wall-mount housing) → Fig. 34
Caution:– The PROFIBUS cable can be damaged!
If the shielding of the cable is grounded at more than one point in plants without additional potential equalisation, mains frequency equalisation currents can occur that damage the cable or the shielding. In such cases the shielding of the cable is to be grounded on only one side, i.e. it must not be connected to the ground terminal of the housing. The shield that is not connected should be insulated!
– We recommend that the PROFIBUS not be looped using conventional cable glands. If you later replace even just one measuring device, the bus communica-tion for the entire bus will have to be interrupted.
Note:– The terminals for the PROFIBUS-PA connection (26/27) have an integral polarity
protection. This ensures correct signal transmission via the fieldbus even if lines are confused.
– Conductor cross-section: max. 2.5 mm2
– Observe the plant’s grounding concept.
4. Screw the cover of the connection compartment (f) firmly back onto the transmitter housing.
4 Wiring Promag 53 PROFIBUS-DP/-PA
54 Endress+Hauser
Fig. 33: Connecting the transmitter (field housing), conductor cross-section: max. 2.5 mm2
A = Aluminium field housingB = Stainless steel housing
a Cable for power supply: 85...260 V AC, 20...55 V AC, 16...62 V DC Terminal No. 1: L1 for AC, L+ for DC; Terminal No. 2: N for AC, L− for DC
b PROFIBUS-DP/-PA cable (see Page 55): Terminal No. 26: DP(B) / PA+ Terminal No. 27: DP(A) / PA –DP(A) = RxD/TxD-N, DP(B) = RxD/TxD-P
c Ground terminal for protective conductord Ground terminal for signal-cable shielde Service adapter for connecting service interface FXA 193 (FieldCheck™, FieldTool™)f Cover of the connection compartmentg Cable for external termination:
Terminal No. 24: DGNDTerminal No. 25: +5V
h ClampF0
6-53
xPB
xxx-
04-0
6-xx
-xx-
000
b
b
c
d
a
a27
25
23
21
21
26
24
22
20
L1 (L+)
DP(B) / PA(+)
N (L-)
DP(A) / PA(-)
h
f
e
f
b
a
A
B
(DGND)(+5 V)
g
Promag 53 PROFIBUS-DP/-PA 4 Wiring
Endress+Hauser 55
Fig. 34: Connecting the transmitter (wall-mount housing), conductor cross-section: max. 2.5 mm2
a Cable for power supply: 85...260 V AC, 20...55 V AC, 16...62 V DC Terminal No. 1: L1 for AC, L+ for DC; Terminal No. 2: N for AC, L− for DC
b PROFIBUS-DP/-PA cable (see Page 55):Terminal No. 26: DP(B) / PA+ Terminal No. 27: DP(A) / PA –DP(A) = RxD/TxD-N, DP(B) = RxD/TxD-P
c Ground terminal for protective conductord Ground terminal for signal-cable shielde Service adapter for connecting service interface FXA 193 (FieldCheck™, FieldTool™)f Cover of the connection compartmentg Cable for external termination:
Terminal No. 24: DGNDTerminal No. 25: +5V
4.3.2 Terminal assignment
F06-
53xP
Bxx
x-04
-03-
xx-x
x-00
0
1 2
c d
e
aa bbg
222320 21 2425 26 27
f
DP (B) / PA (+)DP (A) / PA (–)
(DGND)(+5 V)
g
L1 (L+)N (L–)
Order variant Terminals No. (Outputs/Inputs)
26: DP(B) / PA+27: DP(A) / PA–
53***-***********W PROFIBUS-DP/-PA
53***-***********P PROFIBUS-DP/-PA
53***-***********H PROFIBUS-DP/-PA
Connected load PROFIBUS-DP/-PA
PROFIBUS-PA:Vi = 30 V AC, Ii = 500 mA, Pi = 5.5 W, Li = 10.0 µH, Ci = 5.0 nF
PROFIBUS-DP:To connect on device with Vmax = 260 V and Imax = 500 mA
4 Wiring Promag 53 PROFIBUS-DP/-PA
56 Endress+Hauser
4.3.3 Fieldbus connector
Note:This connector can only be used for a PROFIBUS-PA device.
The connection technology of PROFIBUS-PA allows measuring devices to be con-nected to the fieldbus via uniform mechanical connections such as T-boxes, junction boxes, etc. This connection technology using prefabricated distribution modules and plug-in connectors offers substantial advantages over conventional wiring:• Field devices can be removed, replaced or added at any time during normal opera-
tion. The communications will not be interrupted.• This simplifies installation and maintenance significantly.• Existing cable infrastructures can be used and expanded instantly, e.g. when con-
structing new star distributors using 4-channel or 8-channel junction boxes.
The Promag 53 can therefore be supplied with a ready-mounted fieldbus connector. Fieldbus connectors for retrofitting can be ordered from E+H as a spare part (see Page 111).
Fig. 35: Connectors for connecting to the PROFIBUS-PA
A = Aluminum field housingB = Stainless steel field housingC = Protective cap for connectorD = Fieldbus connectorE = Adapter PG 13.5 / M 20.5F = Connector on housing (male)G = Connector (female)
Pin assignment / color codes:1 = Brown wire: PA+ (Terminal 26)2 = Not connected3 = Blue wire: PA– (Terminal 27)4 = Black wire: Grounding (Notes about connection → Page 54, 55)5 = Female contact in the center not connected6 = Positioning groove7 = Positioning mark
F06-
xxxP
Bxx
x-04
-xx-
xx-x
x-00
0
3 4
150/300 mm45.0 mm(1.766")
PG 13.5M 12 x 12 1
B
C D E
4
1 2
F
G
3
5
6
7
A
Promag 53 PROFIBUS-DP/-PA 4 Wiring
Endress+Hauser 57
Technical data (fieldbus connector):
Conductor size 0.75 mm2
Connector thread PG 13.5
Degree of protection IP 67 in accordance with DIN 40 050 IEC 529
Contact surface CuZnAu
Housing material Cu Zn, surface Ni
Flammability V - 2 in accordance with UL - 94
Operating temperature –40...+85 °C
Ambient temperature –40...+150 °C
Nominal current per contact 3 A
Nominal voltage 125...150 V DC in accordance with the VDE standard 01 10/ISO Group 10
Comparative tracking KC 600
Volume resistance ≤8 mΩ in accordance with IEC 512 Part 2
Dialectric resistance ≤1012 Ω in accordance with IEC 512 Part 2
4 Wiring Promag 53 PROFIBUS-DP/-PA
58 Endress+Hauser
4.4 Potential equalisation
4.4.1 Standard case
Proper measurement is only ensured when the fluid and sensor are at the same electri-cal potential. The majority of Promag sensors have, as part of their standard configura-tion, a built-in reference electrode, which ensures the required potential equalisation. As a rule, this is sufficient to eliminate the need for ground disks or other measures.
Promag W:Reference electrode is standard
Promag P:Reference electrode is optional, depending on material
Promag H:• No reference electrode. The metallic process connection provides a permanent elec-
trical connection to the fluid.• If the process connections are made of a synthetic material, ground disks have to be
used to ensure that potential is equalised (see Page 38). Ground rings are optional accessories, which must be ordered separately → Page 111.
Note:When installed in metal pipes, we recommend that the transmitter housing's ground ter-minal be connected to the pipe.
Fig. 36: Potential equalisation by means of the transmitter's ground terminal
Caution:If the sensor does not have a reference electrode or metallic process connections, the potential equalisation must be established in the manner described in the following spe-cial cases. These special measures are in particular necessary when the usual ground-ing is not ensured or when it is likely that there will be an excessive equalisation current.
F06-
5xxx
xxxx
-04-
xx-x
x-xx
-002
Promag 53 PROFIBUS-DP/-PA 4 Wiring
Endress+Hauser 59
4.4.2 Special cases
Equalising currents in metallic, ungrounded pipesTo prevent disturbances to the measurement, we recommend that both sensor flanges be connected with a grounding cable to the adjacent pipe flange and to ground. Con-nect the transmitter or sensor connection housing, as applicable, to ground potential by means of the ground terminal provided for the purpose (Fig. 37).
Note:The grounding cable for flange-to-flange connections can be ordered separately as an accessory from E+H → Page 111.• DN ≤ 300: the grounding cable is in direct connection with the conductive flange coa-
ting and is secured by the flange screws.• DN ≥ 350: the grounding cable connects directly to the metal transport bracket.
Fig. 37: Potential equalisation with equalising currents in metallic, non-grounded piping systems
6 mm² Cu
DN 300≤ DN 350≥
F06-
5xxx
xxxx
-04-
xx-x
x-xx
-003
4 Wiring Promag 53 PROFIBUS-DP/-PA
60 Endress+Hauser
Plastic pipes or pipes lined with insulating materialNormally the potential equalisation is accomplished via the reference electrodes in the measuring tube. However, it is still possible that in some cases due to a plant's ground-ing design that a large equalisation current will flow over the reference electrodes. This can cause the destruction of the sensor, e.g. through the electrochemical erosion of the electrodes. In such cases, e.g. with fiberglass or PVC pipes, we recommend that you use ground disks to improve the potential equalisation (Fig. 38).Mounting of ground disks → Page 26, 32.
Caution:Risk of damage by electrochemical corrosion. Note the electrochemical series of met-als, if the ground disks and measuring electrodes are made of different materials.
Fig. 38: Potential equalisation / ground disks with plastic pipes or lined pipes
Lined pipes (cathodic protection)In such cases, the sensor should be installed in the pipes in a potential-free manner:• When installing the measuring device, make sure that there is an electrical connection
between the two piping runs (copper wire, 6 mm2).• Make sure that the installation materials do not establish a conductive connection to
the measuring device and that the installation materials withstand the tightening tor-ques applied when the threaded fasteners are tightened.
• Also comply with the regulations applicable to potential-free installation.
Fig. 39: Potential equalisation and cathode protection
a = power supply isolating transformator, b = electrically isolated
F06-
5xxx
xxxx
-04-
xx-x
x-xx
-004
F06-
5xxx
xxxx
-04-
xx-x
x-xx
-005
Promag 53 PROFIBUS-DP/-PA 4 Wiring
Endress+Hauser 61
4.5 Degree of protection
The devices fulfill all the requirements for IP 67. Compliance with the following points is mandatory following installation in the field or servicing, in order to ensure that IP 67 pro-tection is maintained:
• The housing seals must be clean and undamaged when inserted into their grooves. The seals must be dried, cleaned or replaced if necessary.
• All threaded fasteners and screw covers must be firmly tightened.• The cables used for connection must be of the specified outside diameter
(see Page 132).• Firmly tighten the cable entries (Fig. 40). • The cables must loop down before they enter the cable entries (“water trap”, Fig. 40).
This arrangement prevents moisture penetrating the entry. Always install the measur-ing device in such a way that the cable entries do not point up.
• Remove all unused cable entries and insert plugs instead.• Do not remove the grommet from the cable entry.
Fig. 40: Installation instructions, cable entries
Caution:Do not loosen the threaded fasteners of the Promag sensor housing, as otherwise the degree of protection guaranteed by Endress+Hauser no longer applies.
Note:The Promag W and Promag P sensors can be supplied with IP 68 rating (permanent immersion in water to a depth of 3 meters). In this case the transmitter must be installed remote from the sensor.
F06-
xxxx
xxxx
-04-
xx-x
x-xx
-005
4 Wiring Promag 53 PROFIBUS-DP/-PA
62 Endress+Hauser
4.6 Electrical connection check
Perform the following checks after completing electrical installation of the measuring device:
Device condition and specifications Notes
Are cables or the device damaged (visual inspection)? −
Electrical connection of meter Notes
Does the supply voltage match the specifications on the nameplate? 85...260 V AC (45...65 Hz)20...55 V AC (45...65 Hz)16...62 V DC
Do the cables comply with the specifications? see Page 45
Do the cables have adequate strain relief? −
Are the cables correctly segregated by type?Without loops and crossovers?
−
Are the power-supply and fieldbus cables correctly connected? see the wiring diagram inside the cover of the terminal compartment
Are all screw terminals firmly tightened? −
Have the measures for grounding/potential equalisation been correctly implemented?
see Page 58
Are all cable entries installed, firmly tightened and correctly sealed? Cables looped as “water traps”?
−
Are all housing covers installed and firmly tightened? −
Electrical connection - PROFIBUS-DP/-PA Notes
Are all the connecting components (T-boxes, junction boxes, connec-tors, etc.) connected with each other correctly?
–
Has each fieldbus segment been terminated at both ends with a bus terminator?
–
Has the max. length of the fieldbus cable been observed in accordance with the PROFIBUS specifications?
see Page 48
Has the max. length of the stubs been observed in accordance with the PROFIBUS specifications?
see Page 46
Is the fieldbus cable fully shielded and correctly grounded? see Page 46
Promag 53 PROFIBUS-DP/-PA 5 Operation
Endress+Hauser 63
5 Operation
5.1 Operation at a glance
You have a number of options for configuring and commissioning the device:
1. Local display (Option) → Page 64The local display enables you to read all of the important parameters directly at the measuring point, configure device-specific parameters in the field and commission the instrument.
2. Configuration program → Page 75The configuration of profile and device-specific parameters is primarily done via the PROFIBUS-DP/-PA interface. You can obtain special configuration and operating pro-grams from the various manufacturers for these purposes.
3. Jumpers and miniature switches for hardware settings → Page 83You can make the following hardware settings for the PROFIBUS-DP/-PA using jumpers or miniature switches on the I/O board:• Set the device bus address• Switch the hardware write protection on or off
Fig. 41: Options for operating the Promag 53 PROFIBUS-DP/-PA
1 Configuration / operation programs for operating the device via the PROFIBUS-DP/-PA2 Jumpers or miniature switches for hardware settings (write protection, device address)3 Local display for the operation of the device in the field (option)
F06-
xxxP
Bxx
x-19
-xx-
xx-x
x-00
0
1
2
3E- +
5 Operation Promag 53 PROFIBUS-DP/-PA
64 Endress+Hauser
5.2 Operation via the local display
5.2.1 Display and operating elements
The local display enables you to read all important parameters directly at the measuring point and configure the device using the function matrix.
The display consists of four lines; this is where measured values and/or status variables (direction of flow, empty pipe, bar graph, etc.) are displayed. You can change the assignment of display lines to different variables to suit your needs and preferences (→ see the “Description of Device Functions” manual).
Fig. 42: Display and operating elements
Liquid-crystal display (1)
The backlit, four-line liquid-crystal display shows measured values, dialog texts and both system and process error messages. The display as it appears when normal measuring is in progress is known as the HOME posi-tion (operating mode).
Optical sensors for Touch Control (2)
Plus / Minus keys (3)
– HOME position → Direct access to totalizer values and actual values of inputs/outputs– Enter numerical values, select parameters– Select different blocks, groups or function groups within the function matrix
Press the +/− keys simultaneously to trigger the following functions:– Exit the function matrix step by step → HOME position– Press and hold down +/– keys for longer than 3 seconds → Return directly to the HOME position– Cancel data entry
Enter key (4)
– HOME position → Entry into the function matrix– Save the numerical values you input or settings you change
+24.502+1863.97
l
m3
s
–50 +50 %
v
3Σ
Esc
E+-
+24.502+1863.97
l
m3
s
–50 +50 %
v
3Σ1
2
3 4
F06-
53xx
xxxx
-07-
xx-x
x-xx
-000
Promag 53 PROFIBUS-DP/-PA 5 Operation
Endress+Hauser 65
DisplayThe display area consists of three lines in all; this is where measured values are dis-played, and/or status variables (direction of flow, partially filled pipe, bargraph, etc.). You can change the assignment of display lines to variables at will in order to customize the display to suit your needs and preferences (→ see the “Description of Device Func-tions” manual).
Multiplex mode:A maximum of two different display variables can be assigned to each line.Variables multiplexed in this way alternate every 10 seconds on the display.
Error messages:The modes of presentation for system and process error messages are described in detail on Page 69 ff.
Fig. 43: Typical display for normal operating mode (HOME position)
1 Main line shows primary measured values, e.g. volume flow in [l/s].2 Supplementary line shows supplementary measured variables, e.g. totalizer No. 3 in [m3]3 Information line shows additional information on the measured variables, e.g. bar graph of the limit value
reached by the volume flow.4 “Info icons” field: Icons representing additional information on the measured values are shown in this
field. See Page 65 for a full list of the icons and their meanings.5 “Measured values” field: The current measured values appear in this field.6 “Unit of measure” field: The units of measure and time defined for the current measured values appear
in this field.
IconsThe icons which appear in the field on the left make it easier to read and recognise measured variables, device status, and error messages.
Icon Meaning
S System error
P Process error
Fault message (with effect on outputs)
! Notice message (without effect on outputs)
Σ 1...n Totalizers 1...n (system totalizer)
Note:The totalizer amounts with the symbols Σ 1...Σ 3 do not correspond to those that will be sent to the automation control system (→ see the “Description of Device Functions” manual)
1
4 5 6
2
3
+24.502+1863.97
l
m3
s
–50 +50 %
v
3Σ
F06-
53xx
xxxx
-07-
xx-x
x-xx
-001
5 Operation Promag 53 PROFIBUS-DP/-PA
66 Endress+Hauser
Other icons
a Measuring mode = PULSATING FLOWb Counting mode, totalizer = BALANCE (forwards and reverse flow)c Measuring mode = SYMMETRY (bidirectional)d Counting mode, totalizer = forwardse Measuring mode = STANDARDf Counting mode, totalizer = reverseg Icon for volume flowh Icon for mass flowi Status of Analog Output 1 (display value) with status OKj Status of Analog Output 1 (display value) with status UNC = uncertaink Status of Analog Output 1 (display value) with status BADl Status of Analog Input 1...2 with status OKm Status of Totalizers 1...3 with status OKn Status of Analog Input 1...2 with status UNC = uncertaino Status of Totalizers 1...3 with status UNC = uncertainp Status of Analog Input 1...2 with status BADq Status of Totalizers 1...3 with status BAD
a b
e f
ji
c
g
k
l
n
p
m
o
q
d
h
F06-
53xP
Bxx
x-07
-xx-
xx-x
x-00
0
Promag 53 PROFIBUS-DP/-PA 5 Operation
Endress+Hauser 67
5.2.2 Brief description of the function matrix
Note:• See the general notes on Page 68.• Function descriptions → see the “Description of Device Functions” manual
1. HOME position → → Enter the function matrix2. Select a block (e.g. USER INTERFACE)3. Select a group (e.g. CONTROL)4. Select a function group (e.g. BASIC CONFIGURATION)5. Select function (e.g. LANGUAGE)
Change parameter / enter numerical values:key → Select/enter release code, parameters, counter values key → Save the entries
6. Exit the function matrix:– Press and hold down Esc key () for longer than 3 seconds → HOME position– Repeatedly press Esc key () → return step by step to HOME position
Fig. 44: Selecting functions and configuring parameters (function matrix)
F06-
x3xx
xxxx
-19-
xx-x
x-xx
-000
5 Operation Promag 53 PROFIBUS-DP/-PA
68 Endress+Hauser
General notesA few parameters must be entered for the commissioning (see Commissioning Page 86). Complex measuring operations on the other hand necessitate additional functions that you can configure as necessary and customize to suit your process parameters. The function matrix, therefore, comprises a multiplicity of additional func-tions which, for the sake of clarity, are arranged on a number of menu levels (blocks, groups, and function groups).
Comply with the following instructions when configuring functions:• You select functions as described on Page 67. Each cell in the function matrix is iden-
tified by a numerical or letter code on the display.• You can switch off certain functions (OFF). If you do so, related functions in other
function groups will no longer be displayed.• Certain functions prompt you to confirm your data entries. Press to select
“SURE [ YES ]” and then press again to confirm. This saves your setting or starts a function, as applicable.
• Return to the HOME position is automatic if no key is pressed for 5 minutes.• Programming mode is disabled automatically if you do not press a key within
60 seconds following automatic return to the HOME position.
Note:• The transmitter continues to measure while data entry is in progress, i.e. the current
measured values are output via the signal outputs in the normal way.• If the power supply fails, all preset and parameterised values remain safely stored in
the EEPROM.
Caution:All of the functions are described in detail, including the function matrix itself, in the “Description of Device Functions” manual, which is a separate part of this Operating Instruction!
Enabling the programming modeThe function matrix can be disabled. Disabling the function matrix rules out the possi-bility of inadvertent changes to device functions, numerical values or factory settings. You must first enter a numerical code (factory setting = 53) before you can change the settings. If you use a code number of your choice, you exclude the possibility of unau-thorised persons accessing data (→ see the “Description of Device Functions” manual).
Comply with the following instructions when entering codes:• If programming is disabled and the keys are pressed in any function, a prompt for
the code will appear automatically on the display.• If “0” is entered as the customer's code, programming is always enabled.• The E+H service organisation can be of assistance if you mislay your personal code.
Caution:Changing certain parameters such as all sensor characteristics, for example, influences numerous functions of the entire measuring system, particularly measuring accuracy. There is no need to change these parameters under normal circumstances and conse-quently, they are protected by a special code known only to the E+H service organisa-tion. Please contact Endress+Hauser if you have any questions.
Promag 53 PROFIBUS-DP/-PA 5 Operation
Endress+Hauser 69
Disabling the programming modeProgramming mode is disabled if you do not press a key within 60 seconds following automatic return to the HOME position.You can also disable programming in the “ACCESS CODE” function by entering any number (other than the customer's code).
5.2.3 Error messages
Type of errorErrors which occur during commissioning or measuring operation are displayed imme-diately. If two or more system or process errors occur, the error with the highest priority is the one shown on the display.
The measuring system distinguishes between two types of error:• System errors: This group comprises all device errors, e.g. communication errors,
hardware errors, etc. → see Page 115• Process errors: This group comprises all application errors, e.g. empty pipe, etc.
→ see Page 115
Fig. 45: Error messages on the display (example)
1 Error type: P = process error, S = system error2 Error message type: $ = fault signal, ! = notice message (definition: → Page 115 ff.)3 Error designation: e.g. EMPTY PIPE = measuring tube is only partly filled or completely empty4 Error number: e.g. #4015 Duration of most recent error occurrence (in hours, minutes and seconds)
Error message typeThe device assigns system error and process error to two error message types (fault message or notice message) in accordance with a predefined algorithm and classifies them accordingly → Page 115 ff.Serious system errors, e.g. module defects, are always identified and classed as “fault messages” by the measuring device.
Notice message (!)• The error in question has no effect on measurement currently in progress.• Displayed as → Exclamation mark (!), error type (S: system error, P: process error).• PROFIBUS-DP/-PA → Errors of this type are registered in the Transducer Block with
an “UNC(ERTAIN)” status for the process variable in question.
Fault message ( )• The error in question interrupts or stops measurement currently in progress.• Displayed as → Lightning flash ( ), error designation (S: system error, P: process
error).• PROFIBUS-DP/-PA → Errors of this type are registered in the Transducer Block with a
“BAD” status for the process variable in question.
1
2 4 5 3
+24.502y
XXXXXXXXXX#401 00:00:05
P
+24.502y
F06-
53xx
xxxx
-07-
xx-x
x-xx
-002
5 Operation Promag 53 PROFIBUS-DP/-PA
70 Endress+Hauser
Confirming error messagesFor the sake of plant and process safety, the measuring device can be configured in such a way that fault messages () always have to be rectified and acknowledged locally by pressing .Only then do the error messages disappear from the display.This option can be switched on or off by means of the “ACKNOWLEDGE FAULT MES-SAGES” function (see the “Description of Device Functions” manual).
Note:Notice messages (!) do not require acknowledgment. Note, however, that they remain visible until the cause of the error has been rectified.
5.3 Communications: PROFIBUS-DP/-PA
5.3.1 PROFIBUS-DP/-PA technology
PROFIBUS (Process Field Bus) is a standardised bus system based on the European standard EN 50170, Volume 2, which has been successfully used for many years in pro-duction and process automation (chemical industry and process engineering).The PROFIBUS is a multi-master bus system with high performance, which is particu-larly suitable for medium to large plants.
PROFIBUS-DPPROFIBUS-DP is a MASTER/SLAVE bus system. The master function is handled by an automation control system (Class 1 master) or by one or more personal computers (Class 2 master). Using cyclic data telegrams, the automation control system has full access to all of the bus stations assigned to it. With a personal computer (Class 2 mas-ter), you can, if needed, use acyclic data telegrams to exchange data with all of the con-nected stations. In accordance with the standard, up to 126 stations can be connected to a PROFIBUS-DP network. The PROFIBUS-DP can operate at a transmission rate of 9.6 kBit/s...12 MBit/s and, at a transmission rate of 1.5 MBit/s, can support network expan-sions of up to 2000 m with copper cables or 21,730 m using optical link modules.
PROFIBUS-PAPROFIBUS-PA expands the PROFIBUS-DP by using an optimised transmission technol-ogy for field devices while retaining the communications functions of the PROFIBUS-DP. With the selected transmission technology, field devices can be connected via the PROFIBUS-PA to automation control systems over great distances, even in hazardous areas. PROFIBUS-PA is the communications compatible extension of PROFIBUS-DP.
PROFIBUS-PA = PROFIBUS-DP + optimised transmission technology for field devices
Promag 53 PROFIBUS-DP/-PA 5 Operation
Endress+Hauser 71
5.3.2 PROFIBUS-DP system architecture
Fig. 46: PROFIBUS-DP diagram
1 = Automation control system, 2 = Commuwin II operating program, 3 = PROFIBUS-DP RS 485 line (max. 12 MBaud)
General informationThe Promag 53 can be equipped with a PROFIBUS-DP interface (decentral peripheral) in accordance with the fieldbus standard PROFIBUS-DP (EN 50170 Volume 2), which utilises the RS485 technology.As a consequence, the Promag 53 can exchange data with automation control systems, which satisfy this standard. The integration in an automation control system must be in accordance with the PROFIBUS-PA Profile 3.0 specifications.
Data transmission rateThe Promag 53's maximum PROFIBUS-DP data transmission rate is 12 Mbaud.
Note:• The device is able to automatically determine the data transmission rate. We recom-
mend that, before you load a new data transmission rate, you reset the device:– via the function: SUPERVISION → SYSTEM → OPERATION → SYSTEM RESET– by switching the supply voltage off and back on again
• Information about bus terminators can be found on Page 47.
Communications partnerIn a automation control system, the Promag always serves as a slave and can thus, depending on the type of application, exchange data with one or more masters.The master can be a automation control system, a PLC or a PC with a PROFIBUS-DP communications adapter card.
Filling functionIn contrast to the functionality of a Promag 53 that does not support the PROFIBUS, the internal filling function is not integrated in the PROFIBUS-DP devices, since the device does not have any relay functions.However, it is still possible to use a filling function, for some applications, using the total-izer functionality.
F06-
xxxP
Bxx
x-02
-xx-
xx-x
x-00
3
1
2
3
5 Operation Promag 53 PROFIBUS-DP/-PA
72 Endress+Hauser
Note:For additional project planning information about the PROFIBUS-DP fieldbus, seethe Operating Instructions BA 198F/00/en “Field communications - PROFIBUS-DP/-PA:Guides for project planning and commissioning”.
Function blocksThe PROFIBUS uses predefined function blocks to describe the function blocks of a device and to specify uniform data access.The function blocks implemented in the fieldbus devices provide information on the tasks which a device can perform as part of the whole automation strategy. The follow-ing blocks can be implemented in field devices in accordance with Profile 3.0:• Physical Block:
The Physical Block contains all the device-specific characteristics.• Transducer Block (transmission block):
One or more transducer blocks contain all the device measurement and device-spe-cific parameters. In the transducer blocks, the measurement principles (e.g. flow rate and temperature) are mapped in accordance with the PROFIBUS Profile 3.0.
• Function Block:One or more function blocks contain the device's automation functions. We distin-guish between different function blocks, e.g. Analog Input Block, Analog Output Block, Totalizer Block, etc. Each of these function blocks is used to process different applications.
5.3.3 PROFIBUS-PA system architecture
Fig. 47: PROFIBUS-PA system architecture
1 = Automation control system, 2 = Commuwin II operating program, 3 = Segment coupler, 4 = PROFIBUS-DP RS 485 line (max. 12 MBit/s), 5 = PROFIBUS-PA IEC 61158-2 (max. 31.25 kbit/s)
General informationThe Promag 53 can be equipped with a PROFIBUS-PA interface in accordance with the fieldbus standard PROFIBUS-DP (EN 50170 Volume 2).As a consequence, the Promag 53 can exchange data with automation control systems, which satisfy this standard. The integration in a automation control system must be in accordance with the PROFIBUS-PA Profile 3.0 specifications.The selection of the international IEC 61158-2 (International Electrotechnical Commis-sion) transmission standard ensures a future-proof field installation with PROFIBUS-PA.
PROFIBUS PA
3
1
2
4
5
5
F06-
xxxP
Bxx
x-02
-xx-
xx-x
x-00
0
Promag 53 PROFIBUS-DP/-PA 5 Operation
Endress+Hauser 73
Communications partnerIn an automation control system, the Promag always serves as a slave and can thus, depending on the type of application, exchange data with one or more masters.The master can be an automation control system, a PLC or a PC with a PROFIBUS-DP communications adapter card.
Note:During the project planning, please remember that the Promag 53 consumes 11 mA.
Caution:To prevent severe device failures (e.g. short-circuits) from having effect on the PROFIBUS-PA segment, the IEC 61158-2 interface is equipped with a fuse. If the fuse blows, the device is permanently disconnected from the bus. In this case, the I/O mod-ule must be replaced (see Page 123).
Filling functionIn contrast to the functionality of a Promag flow meter that does not support the PROFIBUS, the internal filling function is not integrated in the PROFIBUS-PA devices, since the device does not have any relay functions.However, it is still possible to use a filling function, for some applications, using the total-izer functionality.
Note:For additional project planning information about the PROFIBUS-PA fieldbus, seethe Operating Instructions BA 198F/00/en “Field communications - PROFIBUS-DP/-PA:Guides for project planning and commissioning”.
Function blocksThe PROFIBUS uses predefined function blocks to describe the function blocks of a device and to specify uniform data access. The function blocks implemented in the fieldbus devices provide information on the tasks which a device can perform as part of the whole automation strategy.The following blocks can be implemented in field devices in accordance with Profile 3.0:• Physical Block:
The Physical Block contains all the device-specific characteristics.• Transducer Block (transmission block):
One or more transducer blocks contain all the device measurement and device-spe-cific parameters. The measurement principles (e.g. flow rate and temperature) are mapped in the Transducer blocks in accordance with the PROFIBUS specification.
• Function Block:One or more function blocks contain the device's automation functions. We distin-guish between different function blocks, e.g. Analog Input Block, Analog Output Block, Totalizer Block, etc. Each of these function blocks is used to process different applications.
More information can be found in the “Description of Device Functions” manual.
5 Operation Promag 53 PROFIBUS-DP/-PA
74 Endress+Hauser
5.3.4 Acyclic data exchange
The acyclic data transmission is used to transfer parameters during the commissioning, during maintenance or to display other measured variables that are not contained in the cyclic data traffic.
Generally, a distinction is made between Class 1 and Class 2 master connections.Depending on the implementation of the field device, it is possible to simultaneously establish several Class 2 connections.• Theoretically, a maximum of 49 Class 2 connections can be established to the same
field device.• Two Classe 2 masters are permitted with a Promag. This means that two Class 2 mas-
ters can access the Promag 53 at the same time. However, you must make certain that they do not both attempt to write the same data, since otherwise the data consistency cannot be guaranteed.
• When a Class 2 master reads parameters, it will send an interrogation telegram to the field device specifying the field device address, the slot/index and the expected record length. The field device will answer with the requested record, if the record exists and is the correct length (byte).
• When a Class 2 master writes parameters, it will send the address of the field device, the slot/index, record length (byte) and the record. The field device will acknowledge this write job after completion.
A Class 2 master can access the blocks that are shown in the illustration below.The parameters, which can be accessed by the Endress+Hauser operating program (Commuwin II), are shown Page 75 ff. in the form of a matrix.
Fig. 48: Function block model for the Promag PROFIBUS-DP/-PA
F06-
x3P
Bxx
xx-0
2-xx
-xx-
en-0
01PBPhysicalBlock
TBTransducerBlock
TOTTotalizerFunctionBlock 1...3
AIFunctionBlock
PROFIBUS-DP/-PA
Automation controlsystem
Promag 53 PROFIBUS-DP/-PA 5 Operation
Endress+Hauser 75
5.4 Operation with the PROFIBUS configuration programs
The user can obtain special configuration and operating programs offered by the differ-ent manufacturers for use in configuration. These can be used for configuring both the PROFIBUS-DP/-PA parameters and all of the device-specific parameters. The prede-fined function blocks allow uniform access to all the network and device data.
Note:A step-by-step description of the procedure for commissioning the PROFIBUS interface is given on Page 86 together with information about the configuration of device-specific parameters.
5.4.1 FieldTool™ operating program
FieldTool™ is a universal service and configuration software package designed for the PROline measuring devices. Connection is by means of the PROline service interface (service adapter).
The functionality of FieldTool™ includes the following:• Configuration of device functions• Visualisation of measuring values (including data logging)• Data backup of device parameters• Advanced device diagnosis• Measuring-point documentation
Note:You can find more information on FieldTool™ in the following E+H document:• System information: SI 031D/06/en “FieldTool™”
5.4.2 Commuwin II operating program
Commuwin II is a program for remote operation of field and control-room equipment. Commuwin II can be used irrespective of the device type and the mode of communica-tion (HART or PROFIBUS).
Note:You can find more information on Commuwin II in the following E+H documents:• System information: SI 018F/00/en “Commuwin II”• Operating manual: BA 124F/00/en “Commuwin II” operating program• An exact description of the data types can be found in the slot/index lists in the
“Description of Device Functions” manual.
As an aid in programming with Commuwin II, all of the Promag 53's device functions are clearly arranged in a matrix.Using the “MATRIX SELECTION” (VAH5) function, you can call up various parts of the matrix.
5 Operation Promag 53 PROFIBUS-DP/-PA
76 Endress+Hauser
Device matrix
H9
H8
H7
DIS
P.VA
L.S
TATU
S(D
isp
lay)
H6
EP
D E
LEC
TRO
DE
(Op
tions
)
DIS
PLA
Y V
ALU
E(D
isp
lay)
OV
ER
VO
LT. T
IME
(Ent
ry, S
ervi
ce)
DE
VIC
E N
AM
E(D
isp
lay)
H5
OU
T S
TATU
S(D
isp
lay)
SY
STE
M R
ES
ET
(Sel
ectio
n)
ME
AS
UR
ING
P
ER
IOD
(Ent
ry, S
ervi
ce)
MAT
RIX
SE
LEC
TIO
N(O
ptio
ns)
H4
UN
IT L
EN
GTH
(Op
tions
)
ER
D R
ES
PO
NS
E T
IME
(Ent
ry)
OU
T VA
LUE
(Dis
pla
y)
CH
EC
K C
ON
FIG
..(D
isp
lay)
H3
OFF
-VA
L.LF
-CU
TOFF
(Ent
ry)
ER
D A
DJU
STM
EN
T(A
usw
ahl)
BLO
CK
SE
LEC
TIO
N(O
ptio
ns)
DE
VIC
E ID
(Dis
pla
y)
PO
S. Z
ER
O R
ETU
RN
(Op
tions
)
NO
MIN
AL
DIA
ME
TER
(Op
tions
, Ser
vice
)
H2
DE
NS
ITY
(Ent
ry)
UN
IT D
EN
SIT
Y(O
ptio
ns)
STA
TUS
AC
CE
SS
(Dis
pla
y)
ON
-VA
L.LF
-CU
TOFF
(Ent
ry)
SE
T U
NIT
TO
BU
S(O
ptio
ns)
AC
TUA
L B
AU
DR
ATE
(Dis
pla
y)
INTE
GR
ATIO
N T
IME
(Ent
ry)
ZE
RO
PO
INT
(Ent
ry, S
ervi
ce)
H1
CA
LC. M
AS
SFL
OW
(Dis
pla
y)
UN
IT M
AS
SFL
OW
(Op
tions
)
DE
F. P
RIV
ATE
CO
DE
(Ent
ry)
SE
LEC
TIO
N G
SD
(Op
tions
)
PR
OFI
LE V
ER
SIO
N(D
isp
lay)
SY
STE
M D
AM
PIN
G(O
ptio
ns)
K-F
AC
TOR
NE
GAT
IV(E
ntry
, Ser
vice
)
H0
VO
LUM
E F
LOW
(Dis
pla
y)
UN
IT V
OLU
ME
FLO
W(O
ptio
ns)
AC
CE
SS
CO
DE
(Ent
ry)
AS
SIG
N L
F-C
UTO
FF(O
ptio
ns)
EM
PTY
PIP
E D
ET.
(Op
tions
)
WR
ITE
PR
OTE
CT
(Dis
pla
y)
BU
S-A
DR
ES
SE
(Dis
pla
y)
INS
TL.D
IR S
EN
SO
R(O
ptio
ns)
K-F
AC
TOTR
PO
SIT
IV(E
ntry
, Ser
vice
)
TAG
NA
ME
(Ent
ry)
V0
ME
AS
UR
ING
VA
LUE
S
V1
SY
STE
M U
NIT
S
V2
US
ER
INTE
RFA
CE
V3
LF-C
UTO
FF P
AR
AM
.
V4
EP
D P
AR
AM
ETE
R
V5
V6
PR
OFI
BU
S-D
P/P
A
V7
PR
OFI
BU
S_I
NFO
V8
SY
STE
M
PAR
AM
ETE
R
V9
SE
NS
OR
DAT
A
VA ME
AS
UR
ING
PO
INT
Promag 53 PROFIBUS-DP/-PA 5 Operation
Endress+Hauser 77
Display functions (partial matrix)
H9
H8
H7
H6
DE
VIC
E ID
(Dis
pla
y)
H5
MAT
RIX
SE
LEC
TIO
N(O
ptio
ns)
H4
H3
DIS
PLA
Y M
OD
E(O
ptio
ns)
DIS
PLA
Y M
OD
E(O
ptio
ns)
DIS
PLA
Y M
OD
E(O
ptio
ns)
DIS
PLA
Y M
OD
E(O
ptio
ns)
H2
STA
TUS
AC
CE
SS
(Anz
eig
e)
CO
NTR
AS
T LC
D(E
ntry
)
FOR
MAT
(Op
tions
)
FOR
MAT
(Op
tions
)
FOR
MAT
(Op
tions
)
FOR
MAT
(Op
tions
)
FOR
MAT
(Op
tions
)
FOR
MAT
(Op
tions
)
H1
DE
F. P
RIV
ATE
CO
DE
(Ent
ry)
DIS
PLA
Y D
AM
PIN
G(E
ntry
)
100%
-VA
LUE
(Ent
ry)
100%
-VA
LUE
(Ent
ry)
V10
0%-V
ALU
E(E
ntry
)
100%
-VA
LUE
(Ent
ry)
100%
-VA
LUE
(Ent
ry)
100%
-VA
LUE
(Ent
ry)
H0
AC
CE
SS
CO
DE
(Ent
ry)
LAN
GU
AG
E(O
ptio
ns)
AS
SIG
N L
INE
1(O
ptio
ns)
AS
SIG
N L
INE
1(O
ptio
ns)
AS
SIG
N L
INE
2(O
ptio
ns)
AS
SIG
N L
INE
2(O
ptio
ns)
AS
SIG
N L
INE
3(O
ptio
ns)
AS
SIG
N L
INE
3(O
ptio
ns)
TAG
NA
ME
(Ent
ry)
V0
V1
V2
DIS
PLA
Y
V3
DIS
PLA
Y F
UN
CTI
ON
V4
MA
IN L
INE
V5
MU
LTIP
LEX
V6
AD
DIT
ION
LIN
E
V7
MU
LTIP
LEX
V8
INFO
RM
ATIO
N L
INE
V9
MU
LTIP
LEX
VA ME
AS
UR
ING
PO
INT
5 Operation Promag 53 PROFIBUS-DP/-PA
78 Endress+Hauser
Diagnosis/Alarm/Simulation/Version Info/Service&Analysis (partial matrix)
H9
H8
H7
H6
DE
VIC
E ID
(Dis
pla
y)
H5
SW
-RE
V_S
-DAT
(Ent
ry, S
ervi
ce)
SW
-RE
V_T
-DAT
(Ent
ry, S
ervi
ce)
MAT
RIX
_SE
LEC
TIO
N(O
ptio
ns)
H4
PR
OD
-NR
.AM
P.(D
isp
lay,
Ser
vice
)
H3
ALA
RM
DE
LAY
(Ent
ry)
SW
-ID
EN
T. A
MP.
(Dis
pla
y, S
ervi
ce)
SW
-RE
V. I/
O(D
isp
lay,
Ser
vice
)
H2
STA
TUS
AC
CE
SS
(Dis
pla
y)
SIM
. FA
ILS
AFE
MO
DE
(Op
tions
)
SW
-RE
V.A
MP.
(Ent
ry, S
ervi
ce)
H1
DE
F.P
RIV
ATE
CO
DE
(Ent
ry)
VALU
E S
IM.M
EA
S..
(Ent
ry)
SE
NS
OR
TY
PE
(Dis
pla
y)
HW
-ID
EN
T. A
MP.
(Dis
pla
y, S
ervi
ce)
HW
-RE
V. I/
O(D
isp
lay)
H0
AC
TUA
L S
YS
. CO
ND
.(D
isp
lay)
AC
CE
SS
CO
DE
(Ent
ry)
SIM
.ME
AS
UR
AN
D(O
ptio
ns)
SE
RIA
L N
UM
BE
R(E
ntry
, Ser
vice
)
HW
-RE
V.A
MP.
(Dis
pla
y, S
ervi
ce)
I/O T
YP
E(D
isp
lay)
TAG
NA
ME
(Ent
ry)
V0
DIA
GN
OS
E/A
LAR
M
V!
V2
US
ER
INTE
RFA
CE
V3
V4
SIM
ULA
TIO
N
V6
SE
NS
OR
INFO
V7
AM
PLI
FIE
R IN
FO
V8
I/O M
OD
ULE
INFO
V9
VA ME
AS
UR
ING
PO
INT
Promag 53 PROFIBUS-DP/-PA 5 Operation
Endress+Hauser 79
Physical Block (operation via profile)
H9
H8
H7
H6
H5
ST
RE
VIS
ION
(Dis
pla
y)
H4
MA
NU
FAC
TUR
ER
ID(D
isp
lay)
H3
HA
RD
W V
ER
SIO
N(D
isp
lay)
DE
VIC
E C
ER
TIFI
CAT
(Dis
pla
y)
DIA
G M
AS
K E
XTE
NS
(Dis
pla
y)
DIA
GN
OS
IS E
XTE
NS
(Dis
pla
y)
PE
RM
ITTE
D(D
isp
lay)
PR
OFI
LE V
ER
SIO
N(D
isp
lay)
H2
SO
FTW
VE
RS
ION
(Dis
pla
y)
ME
SS
AG
E(E
ntry
)
LOC
AL
OP
ER
ATIO
N(E
ntry
)
MA
SK
2(D
isp
lay)
DIA
GN
OS
IS 2
(Dis
pla
y)
NO
RM
AL
(Dis
pla
y)
ALE
RT
KE
Y(E
ntry
)
H1
SE
RIA
L N
UM
BE
R(D
isp
lay)
INS
TALL
ATIO
N D
ATE
(Dis
pla
y)
HW
WR
ITE
PR
OTE
C(O
ptio
ns)
MA
SK
1(D
isp
lay)
DIA
GN
OS
IS 1
(Dis
pla
y)
AC
TUA
L(D
isp
lay)
DIS
AB
LE(D
isp
lay)
STR
ATE
GY
(Ent
ry)
H0
DE
VIC
E ID
(Dis
pla
y)
DE
SC
RIP
TOR
(Ent
ry)
SO
FTW
AR
E R
ES
ET
(Ent
ry)
WR
ITE
LO
CK
ING
(Ent
ry)
IDE
NT
NU
MB
ER
(Op
tions
)
MA
SK
(Dis
pla
y)
DIA
GN
OS
IS(D
isp
lay)
TAR
GE
T M
OD
E(E
ntry
)
CU
RR
EN
T(D
isp
lay)
TAG
(Ent
ry)
V0
DE
VIC
E D
ATA
V1
DE
SC
RIP
TIO
N
V2
SO
FTW
RE
RE
SE
T
V3
SE
CU
RIT
Y L
OC
KIN
G
V4
DE
VIC
E D
ATA
V5
DIA
GN
OS
IS M
AS
K
V6
DIA
GN
OS
IS
V7
V8
BLO
CK
MO
DE
V9
ALA
RM
CO
NFI
G
VA BLO
CK
PA
RA
ME
TER
5 Operation Promag 53 PROFIBUS-DP/-PA
80 Endress+Hauser
Transducer Block flow (operation via profile)
H9
H8
UN
IT(E
ntry
)
H7
NO
MIN
AL
SIZ
E(E
ntry
)
UN
IT M
OD
E(O
ptio
ns)
H6
CA
LIB
. FA
CTO
R(E
ntry
)
H5
UN
IT(O
ptio
ns)
ST
RE
VIS
ION
(Dis
pla
y)
H4
UP
PE
R_R
AN
GE
_VA
LUE
(Ent
ry)
ZE
RO
PO
INT
AD
JUS
T(E
ntry
)
H3
LOW
ER
_RA
NG
E_V
ALU
E(E
ntry
)
ZE
RO
PO
INT
(Ent
ry)
PE
RM
ITTE
D(D
isp
lay)
UN
RE
PO
RTE
D(D
isp
lay)
PR
FILE
VE
RS
ION
(Dis
pla
y)
H2
UN
IT(E
ntry
)
UN
IT(E
ntry
)
LOW
FLO
W C
UTO
FF(E
ntry
)
NO
RM
AL
(Dis
pla
y)
UN
AC
KN
OW
LED
GE
D(D
isp
lay)
ALE
RT
KE
Y(E
ntry
)
H1
STA
TUS
(Dis
pla
y)
STA
TUS
(Dis
pla
y)
FLO
W D
IRE
CTI
ON
(Op
tions
)
AC
TUA
L(D
isp
lay)
DIS
AB
LE(D
isp
lay)
STR
ATE
GY
(Ent
ry)
H0
VO
LUM
E_F
LOW
(Dis
pla
y)
VO
LUM
E F
LOW
(Dis
pla
y)
ME
AS
UR
ING
MO
DE
(Op
tions
)
TAR
GE
T_M
OD
E(E
ntry
)
CU
RR
EN
T(D
isp
lay)
TAG
(Ent
ry)
V0
VO
LUM
E_F
LOW
V1
MA
SS
_FLO
W
V2
DE
NS
ITY
V3
TEM
PE
RAT
UR
E
V4
ULT
RA
SO
NIC
V5
VO
RTE
X
V6
SA
MP
LIN
G_F
RE
Q V7
SY
STE
M_P
AR
AM
. V8
BLO
CK
MO
DE
V9
ALA
RM
_CO
NFI
G
VA BLO
CK
_PA
RA
ME
TE
Promag 53 PROFIBUS-DP/-PA 5 Operation
Endress+Hauser 81
Analog Input Block (operation via profile)
H9
H8
RIS
ING
_TIM
E(E
ntry
)
H7
FAIL
SA
FE V
ALU
E(E
ntry
)
DE
C_P
OIN
T_O
UT
Ent
ry)
UN
IT_M
OD
E(O
ptio
ns)
BAT
CH
_OP
ER
ATIO
N(O
ptio
ns)
H6
FAIL
SA
FEA
CTI
ON
(Op
tions
)
US
ER
_UN
IT(E
ntry
)
BAT
CH
_PH
AS
E(E
ntry
)
H5
OU
T_U
NIT
(Ent
ry)
CH
AN
NE
L(O
ptio
ns)
ST_
RE
VIS
ION
(Dis
pla
y)
BAT
CH
_UP
(Ent
ry)
H4
OU
T LI
MIT
(Dis
pla
y)
OU
T_S
CA
LE_M
AX
(Ent
ry)
SW
ITC
H_O
FF_P
OIN
T(E
ntry
)
SW
ITC
H_O
FF_P
OIN
T(E
ntry
)
SW
ITC
H_O
FF_P
OIN
T(E
ntry
)
SW
ITC
H_O
FF_P
OIN
T(E
ntry
)
BAT
CH
_ID
(Ent
ry)
H3
OU
T_S
UB
_STA
TUS
(Dis
pla
y)
OU
T_S
CA
LE_M
IN(E
ntry
)
SW
ITC
H_O
N_P
OIN
T(E
ntry
)
SW
ITC
H_O
N_P
OIN
T(E
ntry
)
SW
ITC
H_O
N_P
OIN
T(E
ntry
)
SW
ITC
H_O
N_P
OIN
T(E
ntry
)
PE
RM
ITTE
D(D
isp
lay)
PR
OFI
LE V
ER
SIO
N(D
isp
lay)
H2
OU
T_S
TATU
S(D
isp
lay)
TYP
E_O
F_LI
N(O
ptio
ns)
ALA
RM
_STA
TE(D
isp
lay)
ALA
RM
_STA
TE(D
isp
lay)
ALA
RM
_STA
TE(D
isp
lay)
ALA
RM
_STA
TE(D
isp
lay)
SIM
ULA
TIO
N_M
OD
E(O
ptio
ns)
NO
RM
AL
(Dis
pla
y)
ALE
RT_
KE
Y(E
ntry
)
H1
OU
T_S
TATU
S(D
isp
lay)
PV
_SC
ALE
_MA
X(E
ntry
)
VALU
E(D
isp
lay)
VALU
E(D
isp
lay)
VALU
E(D
isp
lay)
VALU
E(D
isp
lay)
SIM
ULA
TIO
N_S
TAT.
(Op
tions
)
AC
TUA
L(D
isp
lay)
DIS
AB
LE(D
isp
lay)
TAR
GE
T(E
ntry
)
H0
OU
T_VA
LUE
(Dis
pla
y)
PV
_SC
ALE
_M
IN(E
ntry
)
ALA
RM
_HY
STE
RE
SIS
(Ent
ry)
HI_
HI_
LIM
(Ent
ry)
HI_
LIM
(Ent
ry)
LO_L
IM(E
ntry
)
LO_L
O_L
IM(E
ntry
)
SIM
ULA
TIO
N_V
ALU
E(E
ntry
)
TAR
GE
T_M
OD
E(E
ntry
)
CU
RR
EN
T(D
isp
lay)
TAG
(Ent
ry)
V0
OU
T
V1
SC
ALI
NG
V2
ALA
RM
_LIM
ITS
V3
HI_
HI_
ALA
RM
V4
HI_
ALA
RM
V5
LO_A
LAR
M
V6
LO_L
O-A
LAR
M
V7
SIM
ULA
TIO
N
V8
BLO
CK
_MO
DE
V9
ALA
RM
_CO
NFI
G
VA BLO
CK
_PA
RA
ME
TE
5 Operation Promag 53 PROFIBUS-DP/-PA
82 Endress+Hauser
Totalizer Block (operation via profile)
H9
H8
H7
UN
IT_M
OD
E(O
ptio
ns)
BAT
CH
_OP
ER
ATIO
N(O
ptio
ns)
H6
FAIL
_TO
TFA
ILS
AFE
_MO
DE
BAT
CH
_PH
AS
E(E
ntry
)
H5
CH
AN
NE
L(E
ntry
)
ST_
RE
VIS
ION
(Dis
pla
y)
BAT
CH
_RU
P(E
ntry
)
H4
TOTA
L_LI
MIT
SW
ITC
H-O
FF_P
OIN
T(E
ntry
)
SW
ITC
H-O
FF_P
OIN
T(E
ntry
)
SW
ITC
H-O
FF_P
OIN
T(E
ntry
)
SW
ITC
H-O
FF_P
OIN
T(E
ntry
)
BAT
CH
_ID
(Ent
ry)
H3
TOTA
L_S
UB
_STA
TUS
MO
DE
_TO
TALI
ZE
R
SW
ITC
H-O
N P
OIN
T(E
ntry
)
SW
ITC
H-O
N P
OIN
T(E
ntry
)
SW
ITC
H-O
N P
OIN
T(E
ntry
)
SW
ITC
H-O
N P
OIN
T(E
ntry
)
PE
RM
ITTE
D(D
isp
lay)
PR
FILE
VE
RS
ION
(Dis
pla
y)
H2
TOTA
L_S
TATU
S
PR
ES
ET_
TOTA
LIZ
ER
ALA
RM
STA
TE(D
isp
lay)
ALA
RM
STA
TE(D
isp
lay)
ALA
RM
STA
TE(D
isp
lay)
ALA
RM
STA
TE(D
isp
lay)
NO
RM
AL
(Dis
pla
y)
ALE
RT_
KE
Y(E
ntry
)
H1
TOTA
LS
tatu
sTO
TAL_
STA
TUS
SE
T_TO
TALI
ZE
R
VALU
E(D
isp
lay)
VALU
E(D
isp
lay)
VALU
E(D
isp
lay)
VALU
E(D
isp
lay)
AC
TUA
L(D
isp
lay)
DIS
AB
LE(D
isp
lay)
STR
ATE
GY
(Ent
ry)
H0
TOTA
L_VA
LUE
TOTA
L U
NIT
ALA
RM
_HY
STE
RE
SIS
HI_
HI_
LIM
(Ent
ry)
HI_
LIM
(Ent
ry)
LO_L
IM(E
ntry
)
LLO
_LO
_LIM
(Ent
ry)
TAR
GE
T_M
OD
E(E
ntry
)
CU
RR
EN
T(D
isp
lay)
TAG
(Ent
ry)
V0
TOTA
LIZ
ER
V1
CO
NFI
GU
RAT
ION
V2
ALA
RM
LIM
ITS
V3
HI H
I ALA
RM
V4
HI A
LAR
M
V5
LO A
LAR
M
V6
LO L
O A
LAR
M
V7
V8
BLO
CK
MO
DE
V9
ALA
RM
CO
NFI
G
VA BLO
CK
PA
RA
ME
TE
Promag 53 PROFIBUS-DP/-PA 5 Operation
Endress+Hauser 83
5.5 Hardware configuration
5.5.1 Configuration of write protection
A jumper on the I/O board provides the means of activating or deactivating hardware write protection.
Warning:Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power supply is switched off before you remove the cover of the electronics com-partment.
1. Switch off power supply.2. Remove I/O board → Page 123, 125.3. Configure hardware write protection appropriately using the jumpers (Fig. 49).4. Installation of the I/O board is the reverse of the removal procedure.
Fig. 49: Hardware configuration (I/O board)
1 Jumper for hardware write protection:1.1 Enabled (factory setting) = access to flowmeter parameters via PROFIBUS possible1.2 Disabled = access to flowmeter parameters via PROFIBUS not possible
2 Jumper with no function
LED (LED at the back of the board):– lit continuously → ready for operation– not lit → not ready– flashes → critical fault (no connection to measuring amplifier)
1
1.21.1
2
LED
F06-
xxxP
Bxx
xx-1
6-xx
-xx-
xx-0
00
5 Operation Promag 53 PROFIBUS-DP/-PA
84 Endress+Hauser
5.5.2 Configuration of the device address
Note the following points:• In the case of a PROFIBUS-DP/-PA flowmeter, it will always be necessary to configure
the address. Valid flowmeter addresses lie within the range 0…125. Each address can only be used once in a PROFIBUS-DP/-PA network. If an address is not configured correctly, the flowmeter will not be recognised by the master.The address 126 is used for initial commissioning and for service purposes.
• All flowmeters have the address 126 and software addressing on leaving the factory.
Addressing PROFIBUS-DP/-PA using the local display → Page 86
Addressing PROFIBUS-DP/-PA using miniature switches:
Warning:Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power supply is switched off before you remove the cover of the electronics com-partment.
1. Loosen the Allen-head screw securing the clamp (3 mm key).2. Remove the cover of the electronics compartment.3. Remove the local display (if existing) by loosening the fixing screw of the display
module.4. Use a pointed object to alter the positions of the miniature switches on the I/O
board.5. Installation is the reverse of the removal procedure.
Fig. 50: Addressing using the miniature switches on the I/O board
a Miniature switches Nos. 1–7 for defining the bus address (for example: 1 + 16 + 32 = 49)b Switch for the address mode (type of addressing):
OFF = software addressing via the local displayON = software addressing via miniature switches No. 1–7
OFF ON
11
7
22
8
43
84
165
326
64
b
a
OFF ON
F06-
xxxP
Bxx
x-16
-xx-
xx-x
x-00
1
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 85
6 Commissioning
6.1 Function check
Make sure that all final checks have been completed before you start up your measuring point:• Checklist for “Installation check” → Page 44• Checklist for “Electric connection check” → Page 62
Note:• The technical function data of the PROFIBUS interface in acc. with IEC 61158-2 must
be maintained (FISCO model).• The bus voltage of 9 ... 32 V and the current consumption of 11 mA at the device can
be checked using a normal multimeter.• Using the LED on the I/O board (see Page 113) it is possible to carry out a simple
function check on the fieldbus communication in the non-hazardous area.
Switching on the deviceOnce the connection checks (see Page 62) have been successfully completed, it is time to switch on the power supply. The device is now operational. The measuring device performs a number of power on self-tests. As this procedure progresses the following sequence of messages appears on the local display:
Normal measuring mode commences as soon as start-up completes. Various measured values appear on the display (HOME position).
Note:If start-up fails, an error message indicating the cause is displayed.
PROMAG 53
START UPIS RUNNING
Start-up message
PROMAG 53
SW-AMPLIFIERXX.XX.XX
Current software version
SYSTEM OK
→ OPERATIONBeginning of normal measuring mode
6 Commissioning Promag 53 PROFIBUS-DP/-PA
86 Endress+Hauser
6.2 Commissioning the PROFIBUS-DP/-PA interface using the local display
Note:It will be necessary to enter the numerical code (factory setting: 53) before altering device functions, numerical values or factory settings → Page 68.
The following steps have to be carried out one after the other:
1. Check the hardware write protection:BASIC FUNCTION (G) → PROFIBUS-DP/-PA (GBA) → CONFIGURATION (610) → WRITE PROTECT (6102)
2. Enter the tag name:BASIC FUNCTION (G) → PROFIBUS-DP/-PA (GBA) → CONFIGURATION (610) → TAG NAME (6100)
3. Assign a bus address if this has not already been effected via the corresponding miniature switches (see Page 84):BASIC FUNCTION (G) → PROFIBUS-DP/-PA (GBA) → CONFIGURATION (610) → BUS ADDRESS (6101)
4. Configuring the system unit of the volume and mass flow:– Via the function group “System unit”: MEASURED VARIABLES (A) → SYSTEM
UNITS (ACA) → CONFIGURATION (040) → UNIT MASS FLOW (0400) → UNIT MASS (0401) → UNIT VOL....
– The system unit which has been configured will become effective in the control system once the function SET UNIT TO BUS has been activated: BASIC FUNC-TION (G) → PROFIBUS-DP/-PA (GBA) → OPERATION (614) → SET UNIT TO BUS (6141)
Note:The measured values are transmitted in system units - as described in the table on Page 93 ff. - to the automation control system during the cyclic data exchange.If the system unit of a measured value is changed using the local display, this will not have an immediate effect on the output of the AI block and therefore will not influence the measured value which is transmitted to the automation control system.The altered system unit of the measured value will only be transmitted to the auto-mation control system once the “SET UNIT TO BUS“ function in block G (BASIC FUNCTIONS) → PROFIBUS-DP/-PA (GBA) → OPERATION (614) → SET UNIT TO BUS (6141) has been activated.
5. Configuration of the totalizers 1-3:Promag 53 disposes of 3 totalizers. The following description provides an example for totalizer 1.– Selection of the process variables with the parameter CHANNEL e.g. volume
flow: BASIC FUNCTION (G) → PROFIBUS-DP/-PA (GBA)→ TOTALIZER (613) → CHANNEL (6132)
– Entry of the required units for the totalizer:BASIC FUNCTION (G) → PROFIBUS-DP/-PA (GBA) → TOTALIZER (613) → UNIT TOTALIZER (6133)
– Configuration of the totalizer (define totalizer status) e.g. for Totalizing: BASIC FUNCTION (G) → PROFIBUS-DP/-PA (GBA) → TOTALIZER (613) → SET TOT (6134)
– Configuration of the totalizer mode e.g. for Balancing: BASIC FUNCTION (G) → PROFIBUS-DP/-PA (GBA) → TOTALIZER (613) → TOTALIZER MODE (6136)
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 87
6. Selection of GSD file:BASIC FUNCTION (G) → PROFIBUS-DP/-PA (GBA) → OPERATION (614)→ SELECT GSD (6140)
Note:The selection possibilities and the predefined values/parameters are described in more detail in the “Description of Device Functions” manual.
6.3 Commissioning using the Class 2 master (Commuwin II)
Operation with Commuwin II is described in the E+H documentation BA 124F/00/a2/... Steps 1-5 can be dealt with in the same order as described in Chap. 6.2 “Commission-ing using the local display”.
The configuration parameters can be found in the Commuwin II operation matrix in the manufacturer-specific matrix or in the individual blocks:• in the Physical Block → Page 79• in the manufacturer-specific device matrix lines V6 and V7 → Page 76• in the Analog Input Block → Page 81• in the Totalizer Block line V1 → Page 82
1. Parametering the “Physical Block”:– Open the Physical Block.– Software and hardware write protection is disabled in Promag 53 so that you can
access all the write parameters. Check this status with the parameters WRITE LOCKING (V3H0, software write protection) and HW WRITE PROTECT (V3H1, hardware write protection).
– Enter the tag name.
2. Parametering the manufacturer-specific device parameters of the Transducer Block “PROMAG53”:– Open the manufacturer-specific Transducer Block “PROMAG53”:– Enter the desired name for the block (tag name).
Factory setting: No block name (tag name)– Now configure the device-specific parameters for flow measurement.
Note: Other matrixes can be selected in the cell VAH5 if you wish to configure further manufacturer-specific parameters.Please note that alterations made to device parameters will only be activated once a valid release code has been entered. The release code can be entered in the matrix cell V3H0 (factory setting: 53).
3. Parametering the “Analog Input function block”:Promag 53 disposes of two Analog Input function blocks. The Analog Input 1 block contains the process variable “Volume flow” and the Analog Input 2 block the “Mass flow”. These are selected using the connection clearance list. The following description provides an example for the Analog Input function block No. 1.– Enter the required name for the Analog Input function block 1.
Factory setting: VOLUMEFLOW BLOCK– Open the Analog Input function block.– The input value or the input range can be scaled in accordance with the require-
ments of the automation control system → Page 89 “Rescaling the input value”.– If necessary, set a limit value.
6 Commissioning Promag 53 PROFIBUS-DP/-PA
88 Endress+Hauser
4. Parametering the “Totalizer Block”:Promag 53 comprises three Totalizer function blocks. These are selected using the profile blocks “Totalizer Block” in the connection clearance list.– Enter the required name for the Totalizer function block.
Factory setting: TOTALIZER BLOCK– Select the process variable (e.g. Volume flow) with the parameter CHANNEL
(V8H5).– Enter the required units for the totalizer (UNIT TOTALIZER, V1H0).– Configure the totalizer, e.g. for totalizing (SET TOT, V1H1).– Configure the totalizer mode, e.g. balancing (TOTALIZER MODE, V1H3).
5. Configuration of the cyclic data traffic:– All the relevant data is described in Chapter “System Integration” (see Page 90).– We recommend that the “Coupling Documentation” be used for step-by-step
configuration. This can be obtained from Endress+Hauser Process Solutions for various automation control systems and programmable logic controls.
– The files required for commissioning and network configuration can be obtained as described on Page 90.
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 89
6.3.1 Rescaling the input value
In the Analog Input function block the input value or input range can be scaled in accordance with the automation requirements.
Example:• The system unit in the Transducer Block is m3/h.• The measurement range of the sensor is 0 ... 30 m3/h.• The output range to the automation control system should be 0 ... 100%.• The measured value from the Transducer Block (input value) is rescaled linearly via
the input scaling PV_SCALE to the desired output range OUT_SCALE.
Parameter group PV_SCALE (see “Description of Device Functions” manual)
Parameter group OUT_SCALE (see “Description of Device Functions“ manual)
The result is that with an input value of, for example 15m3/h, a value of 50% is output via the parameter OUT.
Fig. 51: Rescaling the input value in the AI function block
Note:The OUT_UNIT does not have any effect on the scaling. It should, however, be config-ured that it can, for example, be shown on the display.
PV_SCALE_MIN (V1H0) → 0PV_SCALE_MAX (V1H1) → 30
OUT_SCALE_MIN (V1H3) → 0OUT_SCALE_MAX (V1H4) → 100OUT_UNIT (V1H5) → [%]
F06-
53P
Bxx
xxx-
05-x
x-xx
-en-
002
Esc
E- +
0...30 m /h3
Esc
E- + 15 m /h3
m /h3 PV_SCALE OUT_SCALE
100
00
30
100%
00
100%
OUT50%
Volumeflow
Analog Input Function blockTransducerblock
6 Commissioning Promag 53 PROFIBUS-DP/-PA
90 Endress+Hauser
6.4 System integration
The device will be ready for system integration once commissioning has been effected via the local display or the Class 2 master (Commuwin II). The PROFIBUS-DP system will require a description of the device parameters, e.g. output data, input data, data for-mat and supported transmission rate so that it can integrate the field devices into the bus system.This data is contained in a PROFIBUS description file (GSD file) which will be placed at the disposal of the PROFIBUS-DP master while the communication system is being commissioned.Device bitmaps, which appear as symbols in the network tree, can also be integrated.The Profil 3.0 PROFIBUS description file (GSD) allows field devices from various manu-facturers to be exchanged without having to repeat the configuration process.
Generally, a distinction can be made between three different types of GSD (Factory set-ting: manufacturer-specific GSD):
Manufacturer-specific GSD: This GSD guarantees the unlimited functionality of the field device. Device-specific process parameters and functions are therefore available.
Profile GSD: This GSD is different in terms of the number of Analog Input blocks (AI) and the measuring principles. If a system is configured with profile GSDs, it will be pos-sible to exchange devices that are supplied by various manufacturers. It is, however, essential that the cyclic proces values follow the same sequence.
Example:Promag 53 supports Profile GSD PA039741.gsd (RS 485) or PA139741.gsd (IEC 61158-2). This GSD file contains two AI blocks and a totalizer block. The first AI block is always assigned to the volume flow. This guarantees that the first measured variable agrees with the field equipment of other manufacturers. The second AI block can be freely selected as Promag 53 is capable of producing a calculated mass flow (see configura-tion example on Page 98 ff.).
Profile GSD (multivariable) with the ID number 9760Hex: This GSD contains all func-tion blocks such as AI, DO, DI.... This GSD is not supported by Promag.
Note:• A decision should be made with respect to which GSD is to be used before configu-
ration takes place.• The configuration can be altered using the local display or a Class 2 master!
Configuration using the local display → Page 86 ff.
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 91
Promag 53 supports the following GSD files:
Each device is assigned an identification number by the Profibus User Organisation (PNO). The name of the PROFIBUS description file (GSD) is derived from this.At Endress+Hauser, this ID No. starts with the manufacturer ID 15xx. In order to ensure clarity, the GSD names (with the exception of type files) at Endress+Hauser are as follows:
The GSD files for all Endress+Hauser devices can be acquired in the following manner:• Internet (Endress+Hauser) → http://www.endress.com (Products → Process
Solutions → PROFIBUS → GSD files)• Internet (PNO) → http://www.profibus.com (GSD library)• On CD-ROM from Endress+Hauser: Order Number 50097200
Structure of GSD files from Endress+HauserIn the case of the Endress+Hauser field transmitter with PROFIBUS interface, all the files which are needed for configuration are contained in one file. Once unpacked, the file will create the following structure:• Version #xx stands for the corresponding device version. Device-specific bitmaps
can be found in the directories “BMP” and “DIB”. The utilisation of these will depend on the configuration software that is being used.
• The GSD files are saved in the subdirectories “Extended” and “Standard” which can be found in the “GSD” folder. Information relating to the implementation of the field transmitter and any dependencies in the device software can be found in the “Info” folder. Please read this carefully before configuration takes place. The files with the extension *.200 have been saved in the “TypDat” folder.
Name of device Manufacturer-specific ID No.
Profile 3.0 ID No. Manufacturer-specific GSD
Promag 53 PAPROFIBUS-PA(IEC 61158-2)
1527 (Hex) 9741 (Hex) EH3_1527.gsdEH3X1527.gsd
Profile 3.0 GSD Type file Bitmaps
PA139741.gsd EH_1527.200 EH_1527_d.bmp/.dibEH_1527_n.bmp/.dibEH_1527_s.bmp/.dib
Manufacturer-specific ID No.
Profile 3.0 ID No. Manufacturer-specific GSD
Promag 53 DPPROFIBUS-DP(RS 485)
1526 (Hex) 9741 (Hex) EH3_1526.gsdEH3X1526.gsd
Profile 3.0 GSD Type file Bitmaps
PA039741.gsd EH_1526.200 EH_1526_d.bmp/.dibEH_1526_n.bmp/.dibEH_1526_s.bmp/.dib
EH3_15xx EH = Endress + Hauser3 = Profile 3.0_ = standard identification15xx = ID No.
EH3x15xx EH = Endress + Hauser3 = Profile 3.0x = extended identification15xx = ID No.
6 Commissioning Promag 53 PROFIBUS-DP/-PA
92 Endress+Hauser
Standard and Extended formatsThe modules of some GSD files are transmitted with an extended identification (e.g. 0x42, 0x84, 0x08, 0x05). These GSD files can be found in the “Extended” folder.All GSD files that have a standard identification (e.g. 0x94) can be found in the “Stand-ard” folder.When integrating field transmitters, the GSD files with the extended identification should be used first. If, however, the integration is not successful, the standard GSD should be used. This differenciation is the result of a specific implementation in the master systems.
Contents of the download file from the internet and the CD-ROM:• All Endress+Hauser GSD files• Endress+Hauser type files• Endress+Hauser bitmap files• Useful information relating to the devices
Working with GSD / type filesThe GSD files must be integrated into the automation control system.Depending on the software that is being used, the GSD files can be copied to the pro-gram-specific directory or can be read into the database using the import function within the configuration software.
Example 1:In the case of the configuration software Siemens STEP 7 (Siemens PLC S7-300 / 400) the files are copied to the subdirectory ...\ siemens \ step7 \ s7data \ gsd.The GSD files also contain bitmap files. These bitmap files are used to display the meas-uring points in image form. The bitmap files will have to be saved to the directory ...\ sie-mens \ step7 \ s7data \ nsbmp.
Example 2:If you have a PLC Siemens S5 where the PROFIBUS-DP network is configured with the configuration software COM ET 200, you will have to use the type files (x.200 files).
If you are using configuration software other than that referred to above, ask your PLC manufacturer which directory you should use.
Compatibility of Profile 2.0 and 3.0 devicesIt is possible to operate Profile 2.0 and 3.0 devices with different GSD files in one system using one DP master as the cyclic data for the automation control system in both profile versions is compatible.
Note:In most cases, it is also possible to replace Profile 2.0 devices with Profile 3.0 devices of the same type without having to repeat the configuration process.This does not apply when replacing an E+H Profile 2.0 flow device with a Profile 3.0 flow device. The functionality of the new generation of devices PROline (Profile 3.0) differs from that of Profile 2.0 devices as do the names. For this reason, PROline devices (Profile 3.0) have been issued with a new identification number and as a result, an exchange of devices will involve the reconfiguration of the automation control system.
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 93
6.4.1 Cyclic data exchange
In the case of PROFIBUS-DP/-PA, the cyclic transmission of analog values to the auto-mation control system is effected in data blocks of 5 bytes. The measured value is por-trayed in the first 4 bytes in the form of flowing point numbers in accordance with IEEE 754 standard (see IEEE floating point number). The 5th byte contains status infor-mation pertaining to the measured value which is implemented in accordance with the Profile 3.0 specifications (see Page 90). The status will be indicated on the display of the device (see Page 65).
Note:A detailed description of the data types can be found in the Slot/Index lists in the sepa-rate “Description of Device Functions” manual.
IEEE floating point numberConversion of a hexadecimal value into an IEEE floating point number for measured value detection.The measured values are shown in numerical format IEEE-754 in the following manner and are transferred to Class 1 master:
Formula value = (-1)Sign * 2(exponent -127) * (1 + Mantisse)
Example: 40 F0 00 00 hex = 0100 0000 1111 0000 0000 0000 0000 0000 binaryValue = (-1)0 * 2(129-127) * (1 + 2-1 + 2-2 + 2-3)
= 1 * 22 * (1 + 0.5 + 0.25 + 0.125)= 1 * 4 * 1.875 = 7.5
Byte n Byte n+1 Byte n+2 Byte n+3
Bit 7 Bit 6 Bit 0 Bit 7 Bit 6 Bit 0 Bit 7 Bit 0 Bit 7 Bit 0
Sign 27 26 25 24 23 22 21 20 2-1 2-2 2-3 2-4 2-5 2-6 2-7 2-8 2-9 2-10 2-11 2-12 2-13 2-14 2-15 2-16...2-23
Exponents Mantisse Mantisse Mantisse
6 Commissioning Promag 53 PROFIBUS-DP/-PA
94 Endress+Hauser
Block modelThe analog values transferred by Promag 53 during the cyclic data exchange are:• Volume flow• Totalizers 1-3 and the corresponding control parameters• Calculated mass flow (see “Description of Device Functions” manual)• Display value• Control blocks for manufacturer-specific functions
The block model shows the input and output data which Promag 53 provides for data exchange.
Fig. 52: Block model Promag 53 PROFIBUS-DP/-PA Profile 3.0
...
...
...
...
Local operation
COMMUNICATION
TAGNAME
BUSADDRESS
WRITEPROTECT
SELCTIONGSD
SET UNITTO BUS
PROFILEVERSION
TOTALIZER
SELECTTOTALIZER
TOTALIZEROUT VALUE
CHANNEL UNITTOTALIZER
SETTOTALIZER
TOTALIZE
RESET
PRESET
PRESETTOTALIZER
PR
OF
IBU
S-D
P/P
A
Signal processing
Transducer Block
DISPLAY VALUE
CONTROL
Value/Status
PROFILEParameter
Manufacturerspecific
parameterPhysical Block
Analog InputFunction block
OUT VALUE
Sensor
Value/Status
TOT. OUT VALUEValue/Status
TotalizerFunction block
F06-
50P
Bxx
xxx-
05-x
x-xx
-en-
006
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 95
Input data The following are examples of input data: volume flow, totalizers 1-3 and calculated mass flow.The current volume flow, totalizers 1-3 and calculated mass flow can be displayed on the basis of these measured variables. The calculated mass flow is derived from the vol-ume flow and a fixed density.
Data transfer from Promag to the automation control systemThe input and output bytes are structured in a fixed sequence. If addressing is effected automatically using the configuration program, the numerical values of the input and output bytes may deviate from the values in the following table.
Note:• The system units in the table correspond to the predefined scales which are trans-
ferred during the cyclic data exchange.• The assignment of the measured variables to the respective variables can be effected
using the parameter Channel, the local controls or a Class 2 master.• Totalizers 1-3 can be configured individually. The following settings are possible
(factory setting: volume flow in m3):– Off – Mass flow– Volume flow
The parameter “Channel” is described in more detail in the separate “Description of Device Functions” manual.
Inputbyte
Processparameter
Accesstype
Comment/Data format Factory setting unit
0, 1, 2, 3 Volume flow read32-bit floating point number (IEEE-754)Illustration → Page 93
m3/h
4Status
volume flowread Status codes → Page 103 –
5, 6, 7, 8 Totalizer 1 read32-bit floating point number (IEEE-754)Illustration → Page 93
* m3 or kg
9Status
totalizer 1read Status codes → Page 103 –
10, 11, 12, 13 Totalizer 2 read32-bit floating point number (IEEE-754)Illustration → Page 93
* m3 or kg
14Status
totalizer 2read Status codes → Page 103 –
15, 16, 17, 18 Totalizer 3 read32-bit floating point number (IEEE-754)Illustration → Page 93
* m3 or kg
19Status
totalizer 3read Status codes → Page 103 –
20, 21, 22, 23 Mass flow read32-bit floating point number (IEEE-754)Illustration → Page 93
kg/h
24 Status mass flow read Status codes → Page 103
6 Commissioning Promag 53 PROFIBUS-DP/-PA
96 Endress+Hauser
Output data Display valueThe display value allows you to transfer a measured value which has been calculated in the automation control system directly to the Promag. This measured value is a dis-play value which can be assigned to the main line, the secondary line and the info line of the display. The display value contains 4 bytes measured value and 1 byte status.The status is displayed as being OK, UNCERTAIN or BAD (see Page 65).
Data transfer from the automation control system to Promag 53 (display value)
Note:The status can be entered freely and will be interpreted in accordance with the status coding in Profile Specification 3.0.
Example: The concentration is calculated in % f(temperature/density) in the automation control system. The temperature and density status is transferred with the two cyclic measured values and can therefore be shown directly in the automation control system together with the calculated concentration.
Control variables (output data) manufacturer-specificPromag 53 is capable of processing control variables during the cyclic data exchange e.g the switching on positive zero return.The following table shows the control variables (output data) that can be transferred to Promag 53.
Data transfer from the automation control system to Promag 53 (Control block)
Note:A control variable can be executed through the cyclic data exchange each time the out-put byte changes from “0” to another bit pattern. It will then be necessary to reset to “0” before a further control variable can be executed. The transition from any bit pattern to “0” will not have any effect.
Output byte
Process parameter Access type
Comment/Data format Factory set-ting unit
6, 7, 8, 9 Display value write32-bit floating point number (IEEE-754)Illustration → Page 93
ao
10 Status Display value read - –
Output byte
Processparameter
Access type
Comment/Control variable Factory set-ting unit
11 Control variable
write This parameter is manufacturer-specific and can control the following control vari-ables:
0 → 1: Reserved0 → 2: Positive zero return ON0 → 3: Positive zero return OFF0 → 4: Reserved0 → 5: ECC ON 0 → 5: ECC OFF
–
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 97
Control variables for totalizers 1-3 (output data)These functions allow totalizers 1-3 to be controlled from the automation control system.The following control variables are possible: totalizing, resetting, activation of a prede-fined value, balancing, positive flow detection, negative flow detection and stopping of totalizing.
Data transfer from the automation control system to Promag 53 (Control variables of totalizers)
Note:• A control variable can be executed through the cyclic data exchange each time the
output byte changes from one bit pattern to any other bit pattern. It will not be neces-sary to reset to “0” to execute a control variable.
• It is only possible to preset a predefined totalizer value via the local display or the Class 2 master!
Example of SET_TOT and MODE_TOT: If the control variable SET_TOT is set to “1” (1 = Reset the totalizer), the value of the total-izer will be set to “0”. The value of the totalizer will now be added up starting from “0”.If the totalizer is to retain the value “0”, it will be necessary to set the control variable MODE_TOT to “3” (3 = STOP totalizing). The totalizer will now stop adding up. The con-trol variable SET_TOT can be set to “1” at a later point in time (1 = Reset the totalizer).
Factory settings of cyclic measured variablesThe following measured variables are configured in Promag 53 at the factory:• Volume flow• Totalizer 1 (with control variable SET_TOT and MODE_TOT)• Totalizer 2 (with control variable SET_TOT and MODE_TOT)• Totalizer 3 (with control variable SET_TOT and MODE_TOT)• Mass flow• Display value (input value)• Manufacturer-specific control block
If all measured variables are not required, you can use the placeholder “EMTY_MODULE” (0x00) - which can be found in the GSD file - to deactivate individual measured variables using the configuration software of the Class 1 master. Configuration example → Page 98 ff.
Note:Only activate the data blocks which are to be processed in the automation control sys-tem. This will improve the data throughput rate of a PROFIBUS-DP/-PA network.A blinking double-arrow symbol will appear on the display to show that Promag 53 is communicating with the automation control system.
Output byte
Processparameter
Access type
Comment/Control variable Factory set-ting unit
024
SET_TOT 1SET_TOT 2SET_TOT 3
writewritewrite
The following control variables can be entered for totalizers 1-3 using these parameters.
Control variable for SET_TOT:0: Totalizing1: Reset totalizer2: Preset totalizer
–
135
MODE_TOT 1MODE_TOT 2MODE_TOT 3
writewritewrite
Control variable for MODE_TOT:0: Balancing1: Only positive flow detection2: Only negative flow detection3: Stop totalizing
–
6 Commissioning Promag 53 PROFIBUS-DP/-PA
98 Endress+Hauser
Caution:• It is essential that the following sequence be adhered to when configuring the meas-
ured variables – volume flow, totalizers 1...3, mass flow, display value and control!• The device will have to be reset after the new measured variables have been config-
ured. This can be effected in either of two ways:– Via the local display: HOME → Block J (SUPERVISION) → Group JAA (SYSTEM)
→ Function group 804 (OPERATION) → Function 8046 (SYSTEM RESET)– Switch supply voltage off and then on again.
System unitsThe measured values are transmitted in system units - as described in the table on Page 95 - to the automation control system during the cyclic data exchange.If the system unit of a measured value is changed using the local display, this will not have an immediate effect on the output of the AI block (Analog Input Block) and there-fore will not influence the measured value which is transmitted to the automation control system.The altered system unit of the measured value will only be transmitted to the automation control system once the “SET UNIT TO BUS” function in block G (BASIC FUNCTIONS) → Group GBA (PROFIBUS-DP/-PA) → function group 614 (OPERATION)→ Function 6141 (SET UNIT TO BUS) has been activated. This can also be activated with a Class 2 master (e.g. Commuwin II).
Configuration examplesThe configuration of a PROFIBUS-DP system is normally effected in the following manner:
1. The field device (Promag 53) which is to be configured is integrated into the con-figuration program of the automation control system via the PROFIBUS-DP network. The GSD file is used here. The configuration software can be used to configure measured variables “offline”.
2. The automation control system's user program will have to be programmed now. The input and output data is controlled in the user program and the location of the measured variables is defined so that they can be processed further.An additional measured value configuration module may have to be used in the case of automation control systems which do not support the IEEE-754 floating point format.It may also be necessary to change the byte sequence (byte swapping) depending on the type of data management employed in the automation control system (little-endian format or big-endian format).
3. When configuration has been completed, this will be transferred to the automation control system in the form of a binary file.
4. The system can be started now. The automation control system will establish a con-nection to the configured devices. The device parameters which are relevant for the process can now be set using the Class 2 master, e.g. with Commuwin II (see Page 87).
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 99
6.4.2 Configuration examples with Simatic S7 HW-Konfig
Example 1:Full configuration using the manufacturer-specific GSD file
This form of configuration activates all data blocks which are supported by Promag 53. The significance of SET_TOT and MODE_TOT is described on Page 97.
Configuration data
Byte length(Input)
Byte length
(Output)
Data blocks Status Access type GSD block
designation
GSDExtended
blockidentification
GSDStandard
blockidentification
0...4 Volume flow + Status
active read AI 0x42, 0x84,0x08, 0x05
0x94
5...9 0 + 1 Totalizer 1 +Status +Control variable
active read+
write
SETTOT_MODETOT_
TOTAL
0xC1, 0x81, 0x84, 0x85
0xC1, 0x81,0x84, 0x85
10...14 2 + 3 Totalizer 2 +Status +Control variable
active read+
write
SETTOT_MODETOT_
TOTAL
0xC1, 0x81, 0x84, 0x85
0xC1, 0x81,0x84, 0x85
15...19 4 + 5 Totalizer 3 +Status +Control variable
active read+
write
SETTOT_MODETOT_
TOTAL
0xC1, 0x81, 0x84, 0x85
0xC1, 0x81,0x84, 0x85
20...24 – Mass flow + Status
active read AI 0x42, 0x84,0x08, 0x05
0x94
– 6...10 Display value + Status
active write DISPLAY_VALUE
0x82, 0x84,0x08, 0x05
0xA4
– 11 Control variable
active write CONTROL_BLOCK
0x20 0x20
Slot 0 → AI volume flowSlot 1 → Totalizer 1 + Control variable (SET_TOT, MODE_TOT)Slot 2 → Totalizer 2 + Control variable (SET_TOT, MODE_TOT)Slot 3 → Totalizer 3 + Control variable (SET_TOT, MODE_TOT)Slot 4 → AI mass flowSlot 5: Display value (DISPLAY_VALUE)Slot 6: Control block (CONTROL_BLOCK) manufacturer-specific
F06-
53P
Bxx
xx-2
0-xx
-xx-
xx-0
00
6 Commissioning Promag 53 PROFIBUS-DP/-PA
100 Endress+Hauser
Example 2:Replacing measured variables with placeholders (EMPTY_MODULE) using the manu-facturer-specific GSD file:
This form of configuration activates the volume flow, totalizer 1, display value and the manufacturer-specific control blocks.Totalizer 1 is configured “without control variable”. In this example it provides a meas-ured value and cannot be controlled. It is not possible to reset or stop the totalizer.
Configuration data
Byte length(Input)
Byte length(Out-put)
Datablocks
Status Accesstype
GSD block
designation
GSDExtended
blockidentifica-
tion
GSDStandard
blockidentifica-
tion
0...4 – Volume flow +Status
active read AI 0x42, 0x84,0x08, 0x05
0x94
5...9 – Totalizer 1 + Status
active read TOTAL. 0x41, 0x84,0x85
0x41, 0x84,0x85
– – Place-holder
disa-bled
– EMTPY_MODULE
0x00 0x00
– – Place-holder
disa-bled
– EMTPY_MODULE
0x00 0x00
– – Place-holder
disa-bled
– EMTPY_MODULE
0x00 0x00
– 0...4 Display value
active write DISPLAY_VALUE
0x82, 0x84,0x08, 0x05
0xA4
– 5 Control variable
active write CONTROL_BLOCK
0x20 0x20
Slot 0: AI volume flowSlot 1: Totalizer 1 “without control variable”Slot 2: Placeholder (EMPTY_MODULE)Slot 3: Placeholder (EMPTY_MODULE)Slot 4: Placeholder (EMPTY_MODULE)Slot 5: Display value (DISPLAY_VALUE)Slot 6: Control block (CONTROL_BLOCK) manufacturer-specific
F06-
53P
Bxx
xx-2
0-xx
-xx-
xx-0
02
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 101
Example 3:Configuration of the measured variables without placeholders (EMPTY_MODULE) using the manufacturer-specific GSD file.
This form of configuration transfers the volume flow - totalizer 1 with control variable (SET_TOT).
Note:If no further measured variables are required, the placeholders will not have to be used. This only applies if no manufacturer-specific control blocks are used.
F06-
53P
Bxx
xx-2
0-xx
-xx-
xx-0
01Slot 0: AI volume flowSlot 1: Totalizer 1 + Control variable (SET_TOT)
Configuration data
Byte length(Input)
Byte length(Out-put)
Datablocks
Status Access type
GSD block
designation
GSDExtended
blockidentifica-
tion
GSDStandard
blockidentifica-
tion
0...4 – Volume flow
+ Status
active read AI 0x42, 0x84,0x08, 0x05
0x94
5...9 0 Totalizer 1+ Status+ Control variable
active read + write
SETTOT_TOTAL
0xC1, 0x80, 0x84, 0x85
0xC1, 0x80,0x84, 0x85
6 Commissioning Promag 53 PROFIBUS-DP/-PA
102 Endress+Hauser
Example 4:Full configuration using the Profile GSD files PA039741.gsd (RS 485) and PA139741.gsd (IEC 61158-2):
This form of configuration transfers the volume flow - totalizer 1 and the mass flow with control variable.
Note:This GSD file contains two AI blocks and a Totalizer block.The first AI block is always assigned to the volume flow. This guarantees that the first measured variable agrees with the field equipment of other manufacturers.The second AI block can be freely selected as Promag 53 is capable of producing a calculated mass flow.
Note:There is no difference between the block ID's of the two Profile GSD files PA039741.gsd (RS 485) and PA139741.gsd (IEC 61158-2).There is, however, a difference in the baud rates that are supported and the Min_Slave_Interval.
F06-
53P
Bxx
xx-2
0-xx
-xx-
xx-0
03
Slot 0 → AI volume flowSlot 2 → Totalizer 1 + Control variable (SET_TOT, MODE_TOT)Slot 1 → AI mass flow
Configuration data
Byte length(Input)
Byte length(Out-put)
Datablocks
Status Access type
GSD block
designation
GSDExtended
blockidentifica-
tion
GSDStandard
blockidentifica-
tion
0...4 – Volume flow
+ Status
active read AI – 0x94
5...9 0...1 Totalizer 1 + Status + Control
active read +write
SETTOT_MODETOT_
TOTAL
– 0xC1, 0x81, 0x84, 0x85
10...14 – Mass flow + Status
activeread
AI – 0x94
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 103
Status codeThe status codes which are supported by the AI blocks (Analog Input), TOT (Totalizers 1...3) and display value are listed in the following table.The coding of the status corresponds to “PROFIBUS-PA Profile for Process Control Devices - General Requirements” V 3.0:
Status Code Meaning Device status Limits
0x1C0x1D0x1E0x1F
out of service bad
OKLOW_LIMHIG_LIMCONST
0x11 below sensor limit bad LOW_LIM
0x12 sensor limit exceeded bad HIG_LIM
0x0C0x0D0x0E0x0F
device failure bad
OKLOW_LIMHIG_LIMCONST
0x080x090x0A0x0B
not connected(function block not available)
bad
OKLOW_LIMHIG_LIMCONST
0x400x410x420x43
non-specific uncertain
OKLOW_LIMHIG_LIMCONST
0x440x450x460x47
last useable value uncertain
OKLOW_LIMHIG_LIMCONST
0x480x490x4A0x4B
substitute set uncertain
OKLOW_LIMHIG_LIMCONST
0x4C0x4D0x4E0x4F
initial value(values which are not saved after the device or parameters have
been reset)
uncertain
OKLOW_LIMHIG_LIMCONST
0x500x510x520x53
sensor conversion not accurate(measured value of sensor
inaccurate)uncertain
OKLOW_LIMHIG_LIMCONST
0x600x610x620x63
simulated value uncertain
OKLOW_LIMHIG_LIMCONST
0x800x810x820x83
measuring system OK good
OKLOW_LIMHIG_LIMCONST
0x840x850x860x87
update event(change of parameters)
good
OKLOW_LIMHIG_LIMCONST
0x8C0x8D0x8E0x8F
active critical alarm(critical alarm:
alarm limit exceeded)good
OKLOW_LIMHIG_LIMCONST
0x880x890x8A0x8B
active advisory alarm(warning: advance warning limit
exceeded)good
OKLOW_LIMHIG_LIMCONST
6 Commissioning Promag 53 PROFIBUS-DP/-PA
104 Endress+Hauser
6.4.3 Cycle times
Measurement value processing and data communication is effected by Promag in three steps:
1st step: Processing measurement valuesThis involves the calculation of the primary measured variable “volume flow” on the basis of sensor signals.The duration of the sampling intervals will depend on the type of sensor that is being used, the nominal diameter and the energy supply (50 Hz, 60 Hz, DC). The typical processing times of Promag 53 can be found in the following table:
2nd step: AI block calculationThe default values of the AI block and the totalizers are calculated on the basis of the measured variable ascertained while processing the measurement values (volume flow and calculated mass flow) and are then copied to a cyclic data telegram.The AI block calculation is effected within a period of max. 50 ms per block.
Note:Each measurement only involves the calculation of the AI blocks or totalizer blocks. This AI block or totalizer block will only be calculated if it has been activated using the con-figuration software (see Page 97). The deactivation of parameters which are not required in the cyclic data telegram will improve the real time performance of the flowmeter.
Sensor DN Scan interval in [ms]
[mm] 50 Hz 60 Hz DC
Promag WPromag P
15...5065
80...100125...200250...400450...500
600700...750800...900
1000...10501200
1350...14001500
1600...170018002000
60.060.060.0
100.0120.0140.0160.0200.0240.0300.0300.0340.0380.0420.0480.0500.0
50.050.066.7
100.0116.7150.0166.7200.0250.0300.0300.0350.0400.0450.0500.0500.0
55.072.572.5
109.0126.6145.0164.0200.0217.0256.0294.0323.0370.0400.0435.0435.0
Promag H 2...2540...65
80...100
40.060.060.0
33.350.066.7
-36.036.0
AI-Block-berechnung
.................
Firstblock
Data telegram
Secondblock
Mass flow
Volume flow
Sensorsignals
1. Step 2. Step 3. Step
PROFIBUSProtocol chip
Inquiry fromMaster
AI blockcalculation
Totalizer blockcalculation
Volume flow
Signal processing
Mass flow
F06-
x3P
Bxx
xx-0
2-xx
-xx-
en-0
00
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 105
3rd step: PROFIBUS protocol chipThe cyclic data telegram is transferred to the protocol chip and, following an inquiry from the master, is sent to the master in accordance with the data transfer rate.
ms
515
2535
4555
6575
8595
105
115
125
135
145
155
165
175
185
195
205
215
100
2030
4050
6070
8090
100
110
120
130
140
150
160
170
180
190
200
210
9.8
9.
8
9.8
9.
8
9.8
10.
0 1
0.0
10.
0 1
0.0
10.
010
.2 1
0.2
10.
2 1
0.2
10.
2 1
0.5
10.
5 1
0.5
10.
5 1
0.5
Pro
mag
HD
N 2
5, 5
0 H
zS
cann
ing
inte
rval
l <
40
ms
2 ac
tivat
ed b
lock
sA
I blo
ck c
alcu
latio
n / T
otal
izer
blo
ck c
alcu
latio
n: 5
3 m
s p
er b
lock
Exa
mp
le f
or t
he
tim
ing
of
the
blo
ck c
alcu
lati
on a
nd
th
e si
gn
al p
roce
ssin
g
Tota
l pro
cess
ing
tim
efo
r on
e fu
nctio
n b
lock
Tota
l pro
cess
ing
tim
efo
r tw
o fu
nctio
n b
lock
s
Blo
ck c
alcu
latio
n
Sig
nal p
roce
ssin
g
Volu
me
flow
Tota
lizer
Volu
me
flow
Tota
lizer
Sig
nal p
roce
ssin
g
Func
tion
blo
ck c
alcu
latio
n
Mea
surin
g v
alue
tran
smis
sion
to th
e au
tom
atio
n co
ntro
l sys
tem
New
mea
surin
gva
lue
New
mea
surin
gva
lue
New
mea
surin
gva
lue
sam
e va
lue
sam
e va
lue
sam
e va
lue
New
mea
surin
gva
lue
sam
e va
lue
F06-
5xH
PB
xxx-
02-x
x-xx
-en-
000
6 Commissioning Promag 53 PROFIBUS-DP/-PA
106 Endress+Hauser
6.5 Empty-pipe/full-pipe calibration
Flow cannot be measured correctly unless the measuring tube is full. The Empty Pipe Detection function (EPD = Empty Pipe Detection) allows this status to be monitored con-tinuously.The EPD function cannot be activated until after empty/full pipe calibration has been performed. This adjustment procedure is described below.
Note:• These settings can also be made in accordance with the manufacturer's specific
matrix using Commuwin II.• The EPD function is only available if the sensor is equipped with an EPD
electrode.• If the EPD electrode is installed, the sensor is calibrated ex-works with water, i.e. at
approximately 500 µS/cm. If the fluid conductivity differs from this reference, empty pipe/full pipe calibration has to be performed again on site.
• The default setting for EPD when the devices are delivered is OFF; the function has to be activated if required.
• The EPD process error is issued via the status and diagnosis messages.
Response to partially filled pipeIf EPD is on and trips in response to a partially filled or empty measuring tube, the mes-sage “PARTIALLY FILLED PIPE” appears on the local display. The message is also transferred in the status and diagnosis message to the control system during cyclic data exchange → Page 115 ff.If the pipe is partially filled or empty and EPD is not on, response can vary in identically configured systems:• Flow reading fluctuates• Zero flow• Excessively high flow readings
Perform empty and full pipe calibration (EPD)
1. Select the appropriate function in the function matrix: HOME → →→ BASIC FUNCTION → →→ PROCESS PARAMETER → →→ ADJUSTMENT → → EPD ADJUSTMENT
2. Empty the piping. Make sure that the wall of the measuring tube is wetted with fluid for the calibration procedure.
3. Start empty pipe calibration:Select EMPTY PIPE ADJUSTMENT setting and confirm with .
4. Fill the pipe with fluid.5. Start full pipe calibration with the fluid at a standstill:
Select FULL PIPE ADJUSTMENT setting and confirm with .6. Switch on Empty Pipe Detection after calibration → Select “ON” and confirm
with .
Caution:The calibration coefficients must be valid before you can activate the EPD function. If calibration is incorrect the following messages might appear on the display:– EPD ADJUSTMENT FULL = EMPTY
The EPD calibration values for empty pipe and full pipe are identical.– EPD ADJUSTMENT NOT OK
EPD calibration is not possible because the fluid's conductivity is out of range.
In cases of this nature you must repeat empty-pipe or full-pipe calibration!These messages will also be displayed in the cyclic status and in the diagnosis of the control system (error messages → Page 115 ff.).
Promag 53 PROFIBUS-DP/-PA 6 Commissioning
Endress+Hauser 107
6.6 Data storage device (DAT, F-Chip™)
S-DAT™ (Sensor-DAT)The S-DAT™ is an exchangeable data storage chip in which all sensor relevant param-eters are stored, i.e., diameter, serial number, calibration factor, zero point.
T-DAT™ (Transmitter-DAT)The T-DAT™ is an exchangeable data storage chip in which all transmitter parameters and settings are stored.The user has to transfer specific parameterisation values from EEPROM to T-DAT™ and vice versa. See the “Description of Device Functions” manual for details (“T-DAT ADMINISTRATION” function, No. 1009).
F-Chip™ (Function-Chip)The F-Chip™ is a microprocessor chip that contains additional software packages that extend the functionality and application possibilities of the transmitter. In the case of a later upgrade, the F-Chip™ can be ordered as an accessory (see Page 111) and can simply be plugged on to the I/O board (see Page 122). After start up, the software is immediately made available to the transmitter.
Caution:To ensure an unambiguous assignment, the F-Chip™ is coded with the transmitter serial number once it is plugged in. Thus, it can not be reused with other transmitters.
6 Commissioning Promag 53 PROFIBUS-DP/-PA
108 Endress+Hauser
Promag 53 PROFIBUS-DP/-PA 7 Maintenance
Endress+Hauser 109
7 Maintenance
The Promag 53 flow measuring system requires no special maintenance.
Exterior cleaningWhen cleaning the exterior of flowmeters, always use cleaning agents that do not attack the surface of the housing and the seals.
SealsThe seals of the Promag H sensor must be replaced periodically, particularly in the case of gasket seals (aseptic version). The period between changes depends on the fre-quency of cleaning cycles, the cleaning temperature and the fluid temperature.
Replacement seals (accessories) → Page 111
7 Maintenance Promag 53 PROFIBUS-DP/-PA
110 Endress+Hauser
Promag 53 PROFIBUS-DP/-PA 8 Accessories
Endress+Hauser 111
8 Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are avail-able for the transmitter and the sensor. The E+H service organisation can provide detailed information on the order codes of your choice.
Accessory Description Ordering code
Transmitter Promag 53
Transmitter for replacement or for stock. Use the order code to define the following specifi-cations:– Approvals– Degree of protection / version– Cable type for the remote version– Cable entries– Display / operation– Software
53XXX − XXXXX * * * * * * * *
Software package forPromag 53
Software add-ons on F-Chip™, can be ordered individually:– Extended diagnosis– Electrode Cleaning Circuitry (ECC)
DK5SO – *
Mounting kit for transmitter
Mounting set for wall mounted transmitter housing. Suitable for:– Wall mounting– Pipe mounting– Installation in control panel
DK5WM − *
Cable for remote version Coil and signal cables, various lengths. Reinforced cable on request.
DK5CA − * *
Grounding cable for Promag W/P
Set consists of two grounding cables. DK5GC − * * *
Grounding disk for Promag W, P
Grounding disk for potential equalisation DK5GD – * * * *
Mounting kit for Promag H
Mounting kit for Promag H, comprising:– 2 process connections (see Page 160 ff.)– Threaded fasteners– Seals
DKH * * − * * * *
Adapter connection for Promag A/H
Adapter connections for installing Promag 53 H instead of a Promag 30/33 A or Promag 30/33 H / DN 25.
DK5HA − * * * * * *
Ground rings for Promag H
If the process connections are made of PVC or PVDF, ground rings are necessary to ensure that potential is matched. Set of ground rings, comprising:– 2 ground rings– 2 seals
DK5HR − * * * *
Set of seals forPromag H
For regular replacement of the seals of the Promag H sensor.
DK5HS − * * *
Wall mounting kit forPromag H
Wall-mounting kit for the Promag H sensor DK5HM − * *
Welding jig forPromag H
Weld nipple as process connection:welding jig for installation in pipe.
DK5HW − * * *
8 Accessories Promag 53 PROFIBUS-DP/-PA
112 Endress+Hauser
Applicator ™ Software for selecting and configuring flow-meters.Applicator™ can be downloaded from the Internet or ordered on CD-ROM for installa-tion on a local PC.
Contact your E+H representative for more information.
DKA80 − *
FieldTool ™ Configuration and service software for flow-meters in the field:– Commissioning, maintenance analysis– Configuring devices– Service functions– Visualisation of process data– Trouble-shooting– Controlling the “FieldCheck™” tester/simu-
lator
Contact your E+H representative for more information.
DXS10 − * * * * *
FieldCheck ™ Tester/simulator for testing flowmeters in the field. When used in conjunction with the “Field-Tool™” software package, test results can be imported into a database, printed and used for official certification.
Contact your E+H representative for more information.
DXC10 − * *
Accessory Description Ordering code
Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 113
9 Trouble-shooting
9.1 Trouble-shooting instructions
Always start trouble-shooting with the checklist below, if faults occur after start-up or during operation. The routine takes you directly to the cause of the problem and the appropriate remedial measures.
Caution:In the event of a serious fault, a flowmeter might have to be returned to the manufacturer for repair. The procedures on Page 6 must be carried out before you return a device to Endress+Hauser. Always enclose a duly completed “Safety regulation” form. You will find a preprinted form at the back of this manual.
Check the display
No display visible and no output signals present.
1. Check the power supply → terminals 1, 22. Check device fuse → Page 127
85...260 V AC: 0,8 A slow-blow / 250 V20...55 V AC and 16...62 V DC: 2 A slow-blow / 250 V
3. Measuring electronics defective → order spare parts → Page 122
No display visible, but out-put signals are present.
1. Check whether the ribbon-cable connector of the display module is correctly plugged into the amplifier board → Page 124, 126
2. Display module defective → order spare parts → Page 1223. Measuring electronics defective → order spare parts → Page 122
Display texts are in a foreign language.
Switch off power supply. Press and hold down both the +/− buttons and switch on the measuring device. The display text will appear in English (default) and is displayed at maximum contrast.
Error messages on display
Errors which occur during commissioning or measuring operation are displayed immediately.Error messages consist of a variety of icons. The meanings of these icons are as follows:– Error type: S = system error, P = process error– Error message type: = fault message, ! = notice message– PARTIALLY FILLED PIPE = error designation (e.g. measuring tube is only partly filled)– 03:00:05 = duration of error occurrence (in hours, minutes and seconds)– #401 = error number
Caution:• See the information on Page 69 ff.!• Simulations and positive zero return are interpreted by the measuring system as being system errors-
but are only displayed as notice messages.
Error number:No. 001 – 111No. 205 – 321No. 601 – 692
System error (device failure) has occurred → Page 115 ff.
Error number:Nr. 401 − 463
Process error (application error) has occurred → Page 115 ff.
9 Trouble-shooting Promag 53 PROFIBUS-DP/-PA
114 Endress+Hauser
Faulty connection to control system
No connection can be made between the control system and the device.Check the following points:
Supply voltageTransmitter
Check the supply voltage → terminals 1/2
Device fuse Check device fuse → Page 12785...260 V AC: 0,8 A slow-blow / 250 V20...55 V AC and 16...62 V DC: 2 A slow-blow / 250 V
Fieldbus connection PROFIBUS-DP:Check the data lines: A = terminal 26, B = terminal 27
Field bus connector (only for PROFIBUS-PA, option)
– Check pin assignment/wiring → Page 62– Check connection between connector/fieldbus port.
Is the coupling ring tightened correctly?
Field bus voltage(only for PROFIBUS-PA)
Check that a min. bus voltage of 9 V DC is present at terminals 26/27. Permissible range: 9 ... 32 V DC
Network structure Check permissible fieldbus length and number of spurs → Page 46
Base current (only for PROFIBUS-PA)
Is there a basic current of min. 11 mA?
Bus address Check bus address: make certain that there are no double assignments!
Terminators Has the PROFIBUS network been terminated correctly?Each bus segment must always be terminated with a bus terminator at both ends (start and finish). Otherwise interference may affect communi-cation.
Current consumptionPermissible feed current(only for PROFIBUS-PA)
Check the current consumption of the bus segment:The current consumption of the bus segment in question (= total of basic currents of all fieldbus stations) must not exceed the max. permissible feed current of the fieldbus power supply.
System or process error messages
System or process errors which occur during commissioning or during a measuring operation can also be displayed in the manufacturer-specific device controls using the Commuwin II Operating Program → Page 115
Other error (without error message)
Some other error has occurred.
Diagnosis and rectification → Page 121
Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 115
9.2 System and process error messages
General notesSystem and process errors are permanently assigned to two different error message types and are thus given different priorities:
Error message type “Fault message”: • This signal will cause the measurement operation to be immediately interrupted or
stopped!• Presentation on the PROFIBUS → Fault signals will be reported to the next function
block or the control system by setting the status of the corresponding process value to “BAD”.
• Local display → A blinking lightning bolt () will appear.
Error message type “Notice message”:• The measurement operation will continue as usual in spite of this message!• Presentation on the PROFIBUS → Notice messages will be reported to the next func-
tion block or the control system by setting the status of the corresponding process value to “UNCERTAIN”.
• Local display → A blinking exclamation point (!) will appear.
Serious system errors, e.g. a fault in the electronic module, are always evaluated by the instrument as a “fault signal” and will be displayed! Simulations and the suppression of the measured value are only reported with a “Notice message”.
Error messages in the configuration program (Class 2 Master) → see TableIn the Promag 53 system/process errors are recognised and reported in the Transducer Block and Analog Input Block. The following table contains a listing of the device status messages for the Analog Input blocks (PROFIBUS-Profile 3.0) and a description of the device status messages which may appear on the display (Measured value Q = Meas-urement Quality).
Error messages on the local display → see Table.
Device status messageDiagnosis message (Control system)
Device status message(Display)
No. Initial status Analog Input Block/Totalizer Block
Measured value Q /Substatus/alarm limit
Cause/remedy
ROM / RAM failure S CRITICAL FAILURE # 001
1 device failure BAD0x0Fconstant
Cause:System error. ROM / RAM error.Error when accessing the program memory (ROM) or random access memory (RAM) of the processor.
Remedy:Replace the amplifier board. Spare parts → Page 122
Amplifier EEPROMfailure
S AMP HW EEPROM # 011
11 device failure BAD0x0Fconstant
Cause:System error. Measuring amplifier has faulty EEPROM
Remedy:Replace the amplifier board. Spare parts → Page 122
9 Trouble-shooting Promag 53 PROFIBUS-DP/-PA
116 Endress+Hauser
Amplifier EEPROM data inconsistent
S MP SW-EEPROM # 012
12 device failure BAD0x0Fconstant
Cause:System error. Error when accessing data of the measuring amplifier EEPROM
Remedy:Execute a warm start (= start up the measuring system without net-work interruption).• PROFIBUS (Commuwin II):
Manufacturer-specific trans-ducer block → Device matrix (V8H5)
• Local display:SUPERVISION → SYSTEM → OPERATION → SYSTEM RESET (→ RESTART SYSTEM)
S-DAT failure / S-DAT not inserted
S SENSOR HW DAT # 031
31 sensor failure BAD0x10no limits
Cause:System error1. S-DAT™ is defective.2. S-DAT™ is not plugged into I/O
board (missing)
Remedy:1. Replace the S-DAT™.
Spare parts → Page 122Check the spare part set num-ber to ensure that the new, replacement DAT is compatible with the measuring electronics.
2. Plug S-DAT™ into I/O-board → Page 124, 126
S-DAT data inconsistent
S SENSOR SW DAT # 032
32 sensor failure BAD0x10no limits
Cause:System error. Error accessing the calibration values stored in the S-DAT™.
Remedy:1. Check whether the
S-DAT™ is correctly plugged into the amplifier board → Page 124, 126
2. Replace the S-DAT™ if it is defective. Spare parts → Page 122Check the spare part set num-ber to ensure that the new, replacement DAT is compatible with the measuring electronics.
3. Replace measuring electronics boards if necessary. Spare parts → Page 122
Device status messageDiagnosis message (Control system)
Device status message(Display)
No. Initial status Analog Input Block/Totalizer Block
Measured value Q /Substatus/alarm limit
Cause/remedy
Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 117
T-DAT failure S TRANSM. HW DAT # 041
41 device failure BAD0x0Fconstant
Cause:System error1. T-DAT™ is faulty2. T-DAT™ is not plugged into I/O
board (missing).
Remedy:1. Replace the T-DAT™.
Spare parts → Page 122Check the spare part set num-ber to ensure that the new, replacement DAT is compatible with the measuring electronics.
2. Plug T-DAT™ into I/O-board → Page 124, 126
T-DAT data inconsistent
S TRANSM. SW DAT # 042
42 device failure BAD0x0Fconstant
Cause:System error. Error accessing the calibration values stored in the T-DAT™.
Remedy:1. Check whether the T-DAT™ is
correctly plugged into the I/O board → Page 124, 126
2. Replace T-DAT™ if faulty.Spare parts → Page 122Check the spare part set num-ber to ensure that the new, replacement DAT is compatible with the measuring electronics.
3. Replace measuring electronics boards if necessary. Spare parts → Page 122
TOT could not be restarted
S CHEKSUM. TOTAL. # 111
111 device failure BAD0x0Fconstant
Cause:System error. Totalizer checksum error
Remedy:1. Restart the device2. Replace the amplifier board if
necessary. Spare parts → Page 122
Save to T-DAT failed S LOAD T-DAT # 205
205 device failure BAD0x0Fconstant
Cause:System error. Data backup to T-DAT failed, or error when acces-sing the calibration values stored in the T-DAT™.
Remedy:1. Check whether the T-DAT™ is
correctly plugged into the I/O board → Page 124, 126
2. Replace T-DAT™ if faulty. Spare parts → Page 122Check the spare part set num-ber to ensure that the new, replacement DAT is compatible with the measuring electronics.
3. Replace measuring electronics boards if necessary. Spare parts → Page 122
Device status messageDiagnosis message (Control system)
Device status message(Display)
No. Initial status Analog Input Block/Totalizer Block
Measured value Q /Substatus/alarm limit
Cause/remedy
9 Trouble-shooting Promag 53 PROFIBUS-DP/-PA
118 Endress+Hauser
Restore from T-DAT failed
S SAVE T-DAT # 206
206 device failure BAD0x0Fconstant
Cause:System error. Data backup to T-DAT failed, or error when acces-sing the calibration values stored in the T-DAT™.
Remedy:1. Check whether the
T-DAT™ is correctly plugged into the I/O-board → Page 124, 126
2. Replace T-DAT™ if faulty.Spare parts → Page 122Check the spare part set num-ber to ensure that the new, replacement DAT is compatible with the measuring electronics.
3. Replace measuring electronics boards if necessary. Spare parts → Page 122
Communication failure S COMMUNICAT. I/O # 261
261 no communication BAD0x18no limits
Cause:System error. Communication error. No data reception between ampli-fier and I/O board or faulty internal data transfer.
Remedy:Check, whether the electronics boards are correctly inserted in their holders → Page 124, 126
Coil Current out of tolerance
S TOL. COIL CURR. # 321
321 device failure BAD0x0Fconstant
Cause:System error. The coil current of the sensor is outside the tolerance.
Remedy:1. Switch off the power supply and
check the connectors of the coil cable → Page 124, 126
2. Replace measuring electronics boards if necessary.Spare parts → Page 122
Empty Pipe detected P EMPTY PIPE! # 401
401 sensor conversion not accurate (measured value from sensor not accurate)
UNCERTAIN0x50no limits
Cause:Process error. Alarm by Empty Pipe Detection (EPD). Measuring tube partially filled or empty.
Remedy:– Check the process conditions of
the plant– Fill the measuring tube
EPD adjustment not possible
P EPD ADJ.N.OK! # 461
461 non-specific UNCERTAIN0x40no limits
Cause:Process error. EPD calibration not possible because the fluid's con-ductivity is either too low or too high.
Remedy:The EPD function cannot be used with fluids of this nature.
Device status messageDiagnosis message (Control system)
Device status message(Display)
No. Initial status Analog Input Block/Totalizer Block
Measured value Q /Substatus/alarm limit
Cause/remedy
Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 119
EPD adjustment wrong P EPD FULL=EMPTY # 463
463 non specific UNCERTAIN0x40no limits
Cause:Process error. The EPD calibration values for empty pipe and full pipe are identical, therefore incorrect.
Remedy:Repeat EPD calibration, making sure procedure is correct → Page 106
Positive zero return active
S POS.ZERO-RET.! # 601
601 sensor conversion not accurate (measured value from sensor not accurate)
UNCERTAIN0x53constant
Cause:System errorPositive zero return is activated
Remedy:Switch off positive zero return• PROFIBUS (Commuwin II):
Manufacturer-specific trans-ducer block → Device matrix (V8H5)
• Local display:BASIC FUNCTION → SYSTEM PARAMETER → CONFIGURA-TION → POSITIVE ZERO RETURN (→ OFF)
Simulation failsafe active
S SIM. FAILSAFE! # 691
691 substitute set UNCERTAIN0x48...0x4Blow/highconstant
Cause:System errorSimulation of response to error is active
Remedy:Switch off simulation• PROFIBUS (Commuwin II):
Manufacturer-specific trans-ducer block → Service & Analy-sis (V4H2)
• Local display:SUPERVISION → SYSTEM → OPERATION → SIM. FAILSAFE MODE (→ OFF)
Simulation Volume flow active
S SIM. VOL. FLOW! # 692
692 simulated value(manually defined value)
UNCERTAIN0x60...0x63low/highconstant
Cause:System errorSimulation is active
Remedy:Switch off simulation• PROFIBUS (Commuwin II):
Manufacturer-specific trans-ducer block → Service & Analy-sis (V4H0)
• Local display:SUPERVISION → SYSTEM → OPERATION → SIM. MEASU-RED VARIABLE (→ OFF)
Device status messageDiagnosis message (Control system)
Device status message(Display)
No. Initial status Analog Input Block/Totalizer Block
Measured value Q /Substatus/alarm limit
Cause/remedy
9 Trouble-shooting Promag 53 PROFIBUS-DP/-PA
120 Endress+Hauser
Simulation mass flow active
S SIM. MASS FLOW! # 493
693 simulated value(manually defined value)
UNCERTAIN0x60...0x63low/highconstant
Cause:System errorSimulation is active
Remedy:Switch off simulation• PROFIBUS (Commuwin II):
Manufacturer-specific trans-ducer block → Service & Analy-sis (V4H0)
• Local display:SUPERVISION → SYSTEM → OPERATION → SIM. MEASU-RED VARIABLE (→ OFF)
Device status messageDiagnosis message (Control system)
Device status message(Display)
No. Initial status Analog Input Block/Totalizer Block
Measured value Q /Substatus/alarm limit
Cause/remedy
Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 121
9.3 Process errors without messages
Symptoms Rectification
Remark:You may have to change or correct certain settings in parameters in order to rectify the fault. The following parameters are described in detail in the “Description of Device Functions” manual.
Flow values are negative, even though the fluid is flow-ing forwards through the pipe.
1. In case of remote version:Switch off power supply and check wiring → Page 50 ff.
2. Alter setting in INSTALLATION DIR. SENSOR function accordingly
Measured value reading fluctuates even though flow is steady.
1. Check grounding and potential equalisation → Page 58 ff.2. The fluid is not homogenous. Check the following fluid properties:
– Proportion of gas bubbles – too high?– Proportion of solids – too high?– Conductivity fluctuations – too high?
Measured value reading or measuring value output pul-sates or fluctuates, e.g. because of reciprocating pump, peristaltic pump, dia-phragm pump or pump with similar delivery characteris-tic.
Increase the system damping value:• PROFIBUS (Commuwin II):
– Manufacturer-specific transducer block → Device matrix → SYSTEM DAMPING (V8H1)
– Analog Input function block → RISING TIME (V1H8)• Local display:
BASIC FUNCTIONS → SYSTEM PARAMETERS → CONFIGURATION → SYSTEM DAMPING
Will a slight flow be dis-played even though the fluid is at a standstill and the measuring tube is full?
1. Check grounding and potential equalisation → Page 58 ff.2. Check the fluid for presence of gas bubbles.3. Enter a value for the low flow cutoff or increase this value:
– PROFIBUS (Commuwin II): manufacturer-specific transducer block → Device matrix → ON VALUE LOW FLOW CUTOFF (V3H2)
– Local display: BASIC FUNCTIONS → PROCESS PARAMETERS → CONFIGURATION → ON VALUE LOW FLOW CUTOFF
Will a measured value be displayed even though measuring tube is empty?
1. Perform empty-pipe/full-pipe calibration and then switch on Empty Pipe Detection → Page 106
2. Fill the measuring tube.
The fault cannot be rectified or some other fault not described above has occurred. In these instances, please contact your E+H service organisation.
The following options are available for tackling problems of this nature:
Request the services of an E+H service technicianIf you contact our service organisation to have a service technician sent out, please be ready with the following information:– Brief description of the fault– Nameplate specifications (Page 7 ff.): Order code and serial number
Returning devices to E+HThe procedures on Page 6 must be carried out before you return a flow-meter requiring repair or calibration to Endress+Hauser. Always enclose a duly completed “Safety regulation” form with the device. You will find a preprinted “Safety regulation” form at the back of this manual.
Replace transmitter electronicsComponents in the measuring electronics defective → order replacement → Page 122
9 Trouble-shooting Promag 53 PROFIBUS-DP/-PA
122 Endress+Hauser
9.4 Spare parts
Chap. 9.1 contains a detailed trouble-shooting guide. The measuring device, moreover, provides additional support in the form of continuous self-diagnosis and error messages.Fault rectification can entail replacing defective components with tested spare parts. The illustration below shows the available scope of spare parts.
Note:You can order spare parts directly from your E+H service organisation by providing the serial number printed on the nameplates (Page 7 ff.).
Spare parts are shipped as sets comprising the following parts:• Spare part• Additional parts, small items (fasteners, etc.)• Mounting instructions• Packaging
Fig. 53: Spare parts for Promag 53 transmitter (field and wall-mount housings)
1 Power supply board2 Amplifier board3 I/O board (Type PROFIBUS-DP/-PA)4 S-DAT™ (sensor data memory)5 T-DAT™ (transmitter data memory)6 F-Chip™ (function chip for optional software)7 Display module8 Fieldbus connectors comprising protective cap, connector, adapter PG 13.5/M20.5
(only for PROFIBUS-PA, Order no. 50098037)
1
2
3
6
7
5
4
8
F06-
53xP
Bxx
x-03
-06-
06-x
x-00
0
Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 123
9.5 Removing and installing printed circuit boards
Field housing: removing and installing printed circuit boards (Fig. 54)
Warning:• Risk of electric shock! Exposed components carry dangerous voltages. Make sure
that the power supply is switched off before you remove the cover of the electronics compartment.
• Risk of damaging electronic components (ESD protection). Static electricity can damage electronic components or impair their operability. Use a workplace with a grounded working surface purposely built for electrostatically sensitive devices!
1. Remove the cover of the electronics compartment.2. Remove the local display (1) as follows:
– Press in the latches (1.1) at the side and remove the display module.– Disconnect the ribbon cable (1.2) of the display module from the amplifier board.
3. Remove the screws and remove the cover (2) from the electronics compartment.4. Remove power supply board and I/O board (4, 6):
Insert a thin pin into the hole (3) provided for the purpose and pull the board clear of its holder.
5. Remove amplifier board (5):– Disconnect the plug of the sensor signal cable (5.1) including S-DAT™ (5.3) from
the board.– Disconnect the plug of the coil current cable (5.2) from the board.– Insert a thin pin into the hole (3) provided for the purpose and pull the board clear
of its holder.6. Installation is the reverse of the removal procedure.
Caution:Use only original Endress+Hauser parts.
9 Trouble-shooting Promag 53 PROFIBUS-DP/-PA
124 Endress+Hauser
Fig. 54: Field housing: removing and installing printed circuit boards
1 Local display1.1 Latch1.2 Ribbon cable (display module)2 Screws of electronics compartment cover3 Aperture for installing/removing boards4 Power supply board5 Amplifier board5.1 Signal cable (sensor)5.2 Coil current cable (sensor)5.3 S-DAT™ (sensor data memory)5.4 T-DAT™ (transmitter data memory)6 I/O board (Type PROFIBUS-DP/-PA)6.1 F-Chip™ (function chip for optional software)
4
5
6
3
3
3
2
1
1.2
5.1
5.3
5.4
5.2
1.1
6.1
F06-
53xP
Bxx
x-03
-06-
06-x
x-00
1
Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 125
Wall-mounted housing: removing and installing printed circuit boards (Fig. 55)
Warning:• Risk of electric shock. Exposed components carry dangerous voltages. Make sure
that the power supply is switched off before you remove the cover of the electronics compartment.
• Risk of damaging electronic components (ESD protection). Static electricity can damage electronic components or impair their operability. Use a workplace with a grounded working surface, purposely built for electrostatically sensitive devices!
1. Remove the screws and open the hinged cover (1) of the housing.2. Remove the screws securing the electronics module (2). Then push up electronics
module and pull it as far as possible out of the wall-mounted housing.3. Disconnect the following cable plugs from amplifier board (7):
– Unplug sensor signal cable (7.1) including S-DAT™ (7.3)– Unplug coil current cable (7.2)– Unplug ribbon cable (3) of the display module
4. Remove the screws and remove the cover (4) from the electronics compartment.5. Extension of boards (6, 7, 8):
Insert a thin pin into the hole (5) provided for the purpose and pull the board clear of its holder.
6. Installation is the reverse of the removal procedure.
Caution:Use only original Endress+Hauser parts.
9 Trouble-shooting Promag 53 PROFIBUS-DP/-PA
126 Endress+Hauser
Fig. 55: Wall-mounted housing: removing and installing printed circuit boards
1 Housing cover2 Electronics module3 Ribbon cable (display module)4 Screws of electronics compartment cover5 Aperture for installing/removing boards6 Power supply board7 Amplifier board7.1 Signal cable (sensor)7.2 Coil current cable (sensor)7.3 S-DAT™ (sensor data memory)7.4 T-DAT™ (transmitter data memory)8 I/O board (Type PROFIBUS-DP/-PA)8.1 F-Chip™ (function chip for optional software)
34
1
2
3
6
7
8
5
5
7.1
7.3
7.4
7.2
5
8.1
F06-
53xP
Bxx
x-03
-03-
06-x
x-00
0
Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 127
9.6 Replacing the device fuse
Warning:Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power supply is switched off before you remove the cover of the electronics com-partment.
The main fuse is on the power supply board (Fig. 56).The procedure for replacing the fuse is as follows:
1. Switch off power supply.2. Remove the power supply board → Page 123, 1253. Remove cap (1) and replace the device fuse (2).
Use only fuses of the following type:– Power supply 20...55 V AC / 16...62 V DC → 2.0 A slow-blow /
250 V; 5.2 x 20 mm– Power supply 85...260 V AC → 0.8 A slow-blow / 250 V; 5.2 x 20 mm– Ex-rated devices → see the Ex documentation.
4. Assembly is the reverse of the disassembly procedure.
Caution:Use only original Endress+Hauser parts.
Fig. 56: Replacing the device fuse on the power supply board
1 Protective cap2 Device fuse
F-06
xxxx
xx-0
3-xx
-06-
xx-0
00
9 Trouble-shooting Promag 53 PROFIBUS-DP/-PA
128 Endress+Hauser
9.7 Replacing exchangeable measuring electrodes
The Promag W sensor (DN 350...2000) is available with exchangeable measuring elec-trodes as an option. This design permits the measuring electrodes to be replaced or cleaned under process conditions (see Page 129).
Fig. 57: Apparatus for replacing exchangeable measuring electrodes (Replacing → Page 129)
a Allen screwb Handlec Electrode cabled Knurled nut (locknut)e Measuring electrodef Stop cockg Retaining cylinderh Locking pin (for handle)i Ball valve housingj Seal (retaining cylinder)k Coil spring
F06-
5xW
xxxx
x-00
-05-
xx-x
x-00
0
Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 129
Removing the electrode Installing the electrode
1 Loosen Allen screw (a) and remove the cover.
1 Insert new electrode (e) into retaining cylin-der (g) from below. Make sure that the seals at the tip of the electrode are clean.
2 Remove electrode cable (c) secured to handle (b).
2 Mount handle (b) on the electrode and insert locking pin (h) to secure it in position.
Caution:Make sure that coil spring (k) is inserted. This is essential to ensure correct electrical contact and correct measuring signals.
2 Loosen knurled nut (d) by hand. This knurled nut acts as a locknut.
2 Pull the electrode back until the tip of the electrode no longer protrudes from retain-ing cylinder (g).
3 Remove electrode (e) by turning handle (b). The electrode can now be pulled out of retaining cylinder (g) as far as a defined stop.
Warning:Risk of injury! Under process conditions (pressure in the piping system) the elec-trode can recoil suddenly against its stop. Apply counter-pressure while releasing the electrode.
3 Screw the retaining cylinder (g) onto ball-valve housing (i) and tighten it by hand. Seal (j) on the cylinder must be correctly seated and clean.
Note:Make sure that the rubber hoses on retain-ing cylinder (g) and stop cock (f) are of the same color (red or blue).
4 Close stop cock (f) after pulling out the electrode as far as it will go.
Warning:Do not subsequently open the stop cock, in order to prevent fluid escaping.
4 Open stop cock (f) and turn handle (b) to screw the electrode all the way into the retaining cylinder.
5 Remove the electrode complete with retaining cylinder (g).
5 Screw knurled nut (d) onto the retaining cyl-inder. This firmly locates the electrode in position.
6 Remove handle (b) from electrode (e) by pressing out locking pin (h). Take care not to lose coil spring (k).
6 Use the Allen screw to secure electrode cable (c) to handle (b).
Caution:Make sure that the machine screw securing the electrode cable is firmly tightened. This is essential to ensure correct electrical con-tact and correct measuring signals.
7 Remove the old electrode and insert the new electrode. Replacement electrodes can be ordered separately from E+H.
7 Reinstall the cover and tighten (a) Allen screw.
9 Trouble-shooting Promag 53 PROFIBUS-DP/-PA
130 Endress+Hauser
9.8 Software history
Software version / date Altering the software Documentation:Alterations / supplements
Amplifier:V 1.01.01
I/O boardType PROFIBUS-DP/-PA:V 1.00.00
Original software.
Compatible with:– FieldTool ™– Commuwin II (version 2.07.XX and
higher)– PROFIBUS-DP/-PA
PROFILE version 3.0
−
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 131
10 Technical data
10.1 Technical data at a glance
Application
Application Measuring the flow rate of fluids in closed piping systems. A minimum conductivity of ≥ 5 µS/cm is required for measuring; the minimum conductivity required in the case of demineralised water is ≥ 20 µS/cm.Applications in measuring, control and regulation technology for monitoring processes, filling and batching operations.
Function and system design
Measuring principle Electromagnetic flow measurement on the basis of Faraday’s Law.
Measuring system The measuring system consists of a transmitter and a sensor.Two versions are available:– Compact version– Remote version
Transmitter• Promag 53 (Type PROFIBUS-DP/-PA)
Sensor• Promag W (for water / waste water applications)
DN 25...2000, hard rubber or polyurethane lining• Promag P (for chemical and process applications)
DN 15...600, PFA or PTFE lining• Promag H (for hygenic applications)
DN 2...100, PFA lining
Input
Measured variable Flow rate (proportional to induced voltage)
Measuring range Typically v = 0.01...10 m/s with the specified measuring accuracy
Operable flow range Over 1000 : 1
Output
Output signal PROFIBUS-DP/-PA interface:PROFIBUS-DP/-PA in acc. with EN 50170 Volume 2, IEC 61158-2, Profile Version 3.0galvanically insulated
Signal on alarm PROFIBUS-DP/-PA interface:Status and alarm messages in acc. with PROFIBUS-PA Profile Version 3.0
Current consumption 11 mA
Permissible fieldbus voltage
9...32 V, not intrinsically safe
FDE (Fault Disconnec-tion Electronic)
0 mA
Data transfer rate PROFIBUS-PA: supported baud rate = 31.25 kBaudPROFIBUS-DP: supported baud rate = 9.6 kBaud...12 MBaud
Signal coding PROFIBUS-PA: Manchester IIPROFIBUS-DP: NRZ Code
10 Technical data Promag 53 PROFIBUS-DP/-PA
132 Endress+Hauser
Power supply
Electrical connection see Page 45 ff.
Potential equalisation see Page 58 ff.
Cable entries Power supply and signal cables (inputs/outputs):– Cable entry M20 x 1.5 (8...12 mm)– Threads for cable entries PG 13.5 (5...15 mm), 1/2" NPT, G 1/2"
Connecting cable for remote version:– Cable entry M20 x 1.5 (8...12 mm)– Threads for cable entries PG 13.5 (5...15 mm), 1/2" NPT, G 1/2"
Cable specifications see Page 52
Supply voltage 85...260 V AC, 45...65 Hz20...55 V AC, 45...65 Hz16...62 V DC
Power consumption AC: <15 VA (including sensor)DC: <15 W (including sensor)
Switch-on current– max. 13.5 A (< 50 ms) at 24 V DC– max. 3 A (< 5 ms) at 260 V AC
Power supply failure Lasting min. 1 power cycle• EEPROM or T-DAT™ saves measuring system data if power supply fails• S-DAT™: exchangeable data storage chip which stores the data of the
sensor: nominal diameter, serial number, calibration factor, zero point, etc.
Performance characteristics
Reference operating conditions
To DIN 19200 and VDI/VDE 2641:• Fluid temperature: +28 °C ± 2 K• Ambient temperature: +22 °C ± 2 K• Warm-up time: 30 minutes
Installation:• Inlet run >10 x DN• Outlet run > 5 x DN• Sensor and transmitter grounded.• Sensor centered relative to the pipe.
Maximum measured error
± 0.2% o.r. ± 2 mm/s (o.r. = of reading)Supply-voltage fluctuations have no effect within the specified range.
Max. measured error in [%] of reading
Repeatability ± 0.1% o.r. ± 0.5 mm/s (o.r. = of reading)
F06-
53xx
xxxx
xx-0
5-xx
-xx-
xx-0
01
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 133
Operating conditions
Installation
Installation instructions Any orientation (vertical, horizontal)Restrictions and additional installation instructions → Page 13 ff.
Inlet and outlet runs Inlet run: typically ≥ 5 x DNOutlet run: typically ≥ 2 x DN
Length of connecting cable
Permissible cable length Lmax depends on the fluid conductivity. A minimum conductivity of 20 µS/cm is required for measuring demineralised water.
Permissible range = grey shaded area
Environment
Ambient temperature −20...+60 °C (sensor, transmitter)
Note the following points:• Install the device at a shady location. Avoid direct sunlight, particularly in
warm climatic regions.• If both fluid and ambient temperatures are high, install the transmitter at a
remote location from the sensor (→ “Fluid temperature range”).
Storage temperature –10...+50 °C (preferably +20 °C)
Degree of protection • Standard: IP 67 (NEMA 4X) for transmitter and sensor• Optional: IP 68 (NEMA 6P) for remote version of Promag W and P sensor
Resistance to vibrations and shock
Acceleration up to 2 g in accordance with IEC 68-2-6(High-temperature version: no corresponding information available)
CIP cleaning Promag W: not possiblePromag P: possible (note max. temperature)Promag H: possible (note max. temperature)
SIP cleaning Promag W: not possiblePromag P: possible with PFA (note max. temperature)Promag H: possible (note max. temperature)
Electromagnetic compatibility (EMC)
To EN 61326 and NAMUR recommendations NE 21
200
100
5
10 100
L max
200[m]
[ S/cm]µ
L max
F06-
xxxx
xxxx
-05-
xx-x
x-xx
-006
10 Technical data Promag 53 PROFIBUS-DP/-PA
134 Endress+Hauser
Process
Medium temperature range
The permissible fluid temperature depends on the lining of the measuring tube:
Promag W0...+80 °C in the case of hard rubber (DN 65...2000)–20...+70 °C in the case of polyurethane (DN 25...2000)
Promag P–40...+130 °C in the case of PTFE (DN 15...600), restrictions → see diagrams–20...+180 °C in the case of PFA (DN 25...200), restrictions → see diagrams
Compact version Promag P (PFA or PTFE lining)TA = ambient temperature, TF = fluid temperature HT = high-temperature version, with insulation
Remote version Promag P (PFA or PTFE lining)TA = ambient temperature, TF = fluid temperatureHT = high-temperature version, with insulation
0
0
-20 20
20
40
60
TA [°C]
40 60 80
TF [°C]
100 120 140 160 180
HT
PT
FE
PFA
F06-
5xP
xxxx
x-05
-xx-
xx-x
x-00
0
0
0
-20 20
20
40
60
TA [°C]
40 60 80
TF [°C]
100 120 140 160 180
HT
PT
FE
PFA
F06-
xxP
xxxx
x-05
-xx-
xx-x
x-00
0
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 135
Promag HSensor:• DN 2...25: –20...+150 °C (+180 °C in prep.)• DN 40...100: –20...+150 °C
Seal:• EPDM: –20...+130 °C• Silicone: –20...+150 °C• Viton: –20...+150 °C• Kalrez: –20...+150 °C
Conductivity Minimum conductivity:• ≥ 5 µS/cm for fluids generally• ≥ 20 µS/cm for demineralised water
Note that in the case of the remote version, the requisite conductivity is also influenced by the length of the connecting cable → Page 133
Fluid pressure range(Nominal pressure)
Promag W:• DIN 2501: PN 6 (DN 1200...2000), PN 10 (DN 200...2000),
PN 16 (DN 65...2000), PN 25 (DN 200...1000), PN 40 (DN 25...150)• ANSI B16.5: Class 150 (1...24"), Class 300 (1...6")• AWWA: Class D (28...78")• JIS B2238: 10K (DN 50...300), 20K (DN 15...300)
Promag P:• DIN 2501: PN 10 (DN 200...600), PN 16 (DN 65...600),
PN 25 (DN 200...600), PN 40 (DN 15...150)• ANSI B16.5: Class 150 (1/2...24"), Class 300 (1/2...6")• JIS B2238: 10K (DN 50...300), 20K (DN 15...300)
Promag H:The permissible nominal pressure depends on the process connection and seal:• 40 bar: flange, weld nipple (with O-ring-seal)• 16 bar: all other process connections
Resistance to partial vacuum of measuring tube lining
see Page 142
Limiting flow see Page 18
Pressure loss • No pressure loss if the sensor is installed in a pipe of the same nominal diameter (Promag H: only DN 8 and larger).
• Pressure losses for configurations incorporating adapters acc. to (E) DIN EN 545 → Page 18
10 Technical data Promag 53 PROFIBUS-DP/-PA
136 Endress+Hauser
Mechanical construction
Design / dimensions see Page 145 ff.
Weight see Page 143 ff.
Materials Promag W
Transmitter housing:• Compact housing: powder-coated die-cast aluminum or stainless steel
field housing (1.4301 / 316L)• Wall-mounted housing: powder-coated die-cast aluminium
Sensor housing:• DN 25...300: powder-coated die-cast aluminium• DN 350...2000: painted steel
Measuring pipe:• DN < 350: stainless steel1.4301 or 1.4306 (304L); for non-stainless flange
material: with Al/Zn protective coating• DN > 300: stainless steel 1.4301 (304); for non-stainless flange material:
Amerlock 400 lacquer
Flanges:• DIN: stainless steel 1.4571; ST37-2 (DN < 350: with Al/Zn protective coat-
ing; DN > 300 with Amerlock 400 lacquer• ANSI: A105; 316L (DN < 350 with Al/Zn protective coatiing, DN > 300 with
Amerlock 400 lacquer)• AWWA: A105• JIS: S20C, SUS 316L (DN < 350 with Al/Zn protective coatiing, DN > 300
with Amerlock 400 lacquer)
Ground disks: 1.4435 (316L) or Alloy C-22Electrodes: 1.4435, Alloy C-22, tantalumSeals: seals in acc. with DIN 2690
Promag P
Transmitter housing:• Compact housing: powder-coated die-cast aluminum or stainless-steel
field housing• Wall-mounted housing: powder-coated die-cast aluminium
Sensor housing:• DN 25...300: powder-coated die-cast aluminium• DN 350...600: painted steel
Measuring pipe:• DN < 350: stainless steel1.4301 or 1.4306 (304L); for non-stainless flange
material: with Al/Zn protective coating• DN > 300: staininless steel 1.4301 (304); for non-stainless flange material:
Amerlock 400 lacquer
Flanges:• DIN: stainless steel 1.4571; ST37-2 (DN < 350: with Al/Zn protective coat-
ing; DN > 300 with Amerlock 400 lacquer• ANSI: A105; 316L (DN < 350 with Al/Zn protective coatiing, DN > 300 with
Amerlock 400 lacquer)• JIS: S20C, SUS 316L (DN < 350 with Al/Zn protective coatiing, DN > 300
with Amerlock 400 lacquer)
Ground disks: 1.4435 (316L) or Alloy C-22Electrodes: 1.4435; platinum/rhodium 80/20; Alloy C-22; tantalumSeals: seals in acc. with DIN 2690
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 137
Promag H
Transmitter housing:• Compact housing: powder-coated die-cast aluminum or stainless-steel
field housing (1.4301 / 316L)• Wall-mounted housing: powder-coated die-cast aluminium
Sensor housing: 1.4301Wall assembly set: 1.4301Measuring pipe: stainless steel 1.4301 or 1.4306 (304L)
Flanges:• All connections 1.4404 / 316L • Flanges (DIN, ANSI, JIS) also available in PVDF• Adhesive fitting made of PVC
Ground rings: 1.4435 (316L), tantalum, platinum (base material: titanium grade 2, platinum coating min. 12 µm), Alloy C-22
Electrodes:• Standard: 1.4435• Optional: Alloy C-22, tantalum, platinum/rhodium 80/20 (only to DN 25)
Seals:• DN 2...25: O-ring (EPDM, Viton, Kalrez) or gasket seal (EPDM, silicone,
Viton)• DN 40...100: gasket seal (EPDM, silicone)
Material load diagram The material load diagrams (pressure-temperature graphs) for the process connections are to be found in the following documents:– Technical Information “Promag 50/53 W” (TI 046D/06/en)– Technical Information “Promag 50/53 P” (TI 047D/06/en)– Technical Information “Promag 50/53 H” (TI 048D/06/en)
Fitted electrodes Promag W:Measuring, reference and EPD electrodes– Standard available with 1.4435, Alloy C-22, tantalum– Optional: exchangeable measuring electrodes made of
1.4435 (DN 350...2000)
Promag P:Measuring, reference and EPD electrodes– Standard available with 1.4435, Alloy C-22, tantalum– Optional: reference electrode and EPD electrodes made of
platinum/rhodium 80/20
Promag H:Measuring electrodes and EPD electrodes– Standard available with 1.4435, Alloy C-22, tantalum, platinum/rhodium– DN 2...8: without EPD electrode
Process connections Promag W:Flange connection: DIN (fitting length according to DIN 2501), ANSI, AWWA, JIS
Promag P:Flange connection : DIN (fitting length according to DIN 2501), ANSI, JIS
Promag H:– With O-ring: weld nipples (ISO 2463, IPS), flanges (DIN, ANSI, JIS), PVDF
flanges (DIN, ANSI, JIS), external pipe thread, internal pipe thread, hose connection, PVC adhesive fittings
– With gasket seals: weld nipples (DIN 11850, ODT), clamps (ODT, ISO 2852, DIN 32676), threaded fasteners (DIN 11851, DIN 11864-1, ISO 2853, SMS1145), flanges (DIN 11864-2)
10 Technical data Promag 53 PROFIBUS-DP/-PA
138 Endress+Hauser
Surface roughness • Measuring tube lined with PFA: ≤ 0.3 µm• Electrodes:
– 1.4435, Alloy C-22: ≤ 0,4 µm– Tantal, Platin/Rhodium: ≤0,8 µm
• Process connection Promag H: < 0.8 µm
(all details refer to parts which come into contact with the fluid)
Human interface
Display elements • Liquid-crystal display: illuminated, four lines with 16 characters per line• Custom configurations for presenting different measured value and status
variables• 3 totalizers
Operating elements Local operation with three optical keys (–, +, E)
Remote operation Operation via PROFIBUS-DP/-PA
Certificates and approvals
Ex Approvals Information presently available in Ex versions (ATEX, FM, CSA) can be sup-plied by your E+H Sales Centre on request. All information relevant to explo-sion protection is available in separate Ex documents that you can order as necessary.
Sanitary compatibility Promag W:No applicable approvals or certification
Promag P:No applicable approvals or certification
Promag H:• 3A authorization and EHEDG-tested• Seals in conformity with FDA (except Kalrez seals)
CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
Certification PROFIBUS-DP/-PA
The Promag 53 flowmeter has passed all the test procedures implemented and has been certified and registered by the PROFIBUS User Organisation. The flowmeter thus meets all the requirements of the specifications listed below:• Certified in acc. with PROFIBUS
Device Certification Number: upon request• The device meets all PROFIBUS PROFILE Version 3.0 specifications.• The device may also be operated using certified devices from other manu-
facturers (interoperability).
Other standards andguidelines
EN 60529:Degrees of protection by housing (IP code)
EN 61010:Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures.
EN 61326 (IEC 1326):Electromagnetic compatibility (EMC requirements)
NAMUR NE 21:Association for Standards for Control and Regulation in the Chemical Industry
PROFIBUS-PA:PROFIBUS Volume 2, Profile 3.0
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 139
Ordering information
The E+H service organisation can provide detailed ordering information and information on specific order codes on request.
Accessories
Various accessories, which can be ordered separately from E+H, are availa-ble for the transmitter and the sensor (see Page 111). The E+H service organisation can provide detailed information on the order codes of your choice.
Supplementary documentation
• System Information Promag (SI 028D/06/en)• Technical Information Promag 50/53 W (TI 046D/06/en)• Technical Information Promag 50/53 P (TI 047D/06/en)• Technical Information Promag 50/53 H (TI 048D/06/en)• Description of Device Functions Promag 53 (BA054D/06/de)• Supplementary documentation on Ex-ratings: ATEX, FM, CSA, etc.
10 Technical data Promag 53 PROFIBUS-DP/-PA
140 Endress+Hauser
10.2 Measuring tube specifications
Promag WNominal diameter
Pressure rating Inside diameter of measuring tube
DIN[mm]
ANSI[inch]
DIN[bar]
ANSI[lbs]
AWWA JIS Hard rubber Polyurethane
25 1" PN 40 Cl 150 − 20K − 24
32 − PN 40 − − 20K − 32
40 1 1/2" PN 40 Cl 150 − 20K − 38
50 2" PN 40 Cl 150 − 10K − 50
65 − PN 16 − − 10K 66 66
80 3" PN 16 Cl 150 − 10K 79 79
100 4" PN 16 Cl 150 − 10K 102 102
125 − PN 16 − − 10K 127 127
150 6" PN 16 Cl 150 − 10K 156 156
200 8" PN 10 Cl 150 − 10K 204 204
250 10" PN 10 Cl 150 − 10K 258 258
300 12" PN 10 Cl 150 − 10K 309 309
350 14" PN 10 Cl 150 − − 342 342
400 16" PN 10 Cl 150 − − 392 392
450 18" PN 10 Cl 150 − − 437 437
500 20" PN 10 Cl 150 − − 492 492
600 24" PN 10 Cl 150 − − 594 594
700 28" PN 10 − Class D − 692 692
− 30" − − Class D − 742 742
800 32" PN 10 − Class D − 794 794
900 36" PN 10 − Class D − 891 891
1000 40" PN 10 − Class D − 994 994
− 42" − − Class D − 1043 1043
1200 48" PN 6 − Class D − 1197 1197
− 54" − − Class D − 1339 1339
1400 − PN 6 − − − 1402 1402
− 60" − − Class D − 1492 1492
1600 − PN 6 − − − 1600 1600
− 66" − − Class D − 1638 1638
1800 72" PN 6 − Class D − 1786 1786
2000 78" PN 6 − Class D − 1989 1989
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 141
Promag PNominal diameter
Pressure rating Inside diameter of measuring tube
DIN[mm]
ANSI[inch]
DIN[bar]
ANSI[lbs]
JIS with PFA[mm]
with PTFE[mm]
15 1/2" PN 40 Cl 150 20K − 15
25 1" PN 40 Cl 150 20K 23 26
32 − PN 40 − 20K 32 35
40 1 1/2" PN 40 Cl 150 20K 36 41
50 2" PN 40 Cl 150 10K 48 52
65 − PN 16 − 10K 63 67
80 3" PN 16 Cl 150 10K 75 80
100 4" PN 16 Cl 150 10K 101 104
125 − PN 16 − 10K 126 129
150 6" PN 16 Cl 150 10K 154 156
200 8" PN 10 Cl 150 10K 201 202
250 10" PN 10 Cl 150 10K − 256
300 12" PN 10 Cl 150 10K − 306
350 14" PN 10 Cl 150 − − 337
400 16" PN 10 Cl 150 − − 387
450 18" PN 10 Cl 150 − − 432
500 20" PN 10 Cl 150 − − 487
600 24" PN 10 Cl 150 − − 593
Promag HNominal diameter
Pressure ratings * Inside diameter of measuring tube **
DIN[mm]
ANSI[inch]
DIN[bar]
PFA
2 1/12" PN 16 / PN 40 2.25
4 5/32" PN 16 / PN 40 4.5
8 5/16" PN 16 / PN 40 9.0
15 1/2" PN 16 / PN 40 16.0
25 1" PN 16 / PN 40 22.6
25 1" PN 16 / PN 40 26.0
40 1 1/2" PN 16 / PN 40 35.3
50 2" PN 16 / PN 40 48.1
65 2 1/2" PN 16 / PN 40 59.9
80 3" PN 16 / PN 40 72.6
100 4" PN 16 / PN 40 97.5
* The pressure rating depends on the process connections and the seals (see Page 135).** Inside diameters of process connections → Page 160 ff.
10 Technical data Promag 53 PROFIBUS-DP/-PA
142 Endress+Hauser
10.3 Resistance to partial vacuum of measuring tube lining
Promag WNominal diameter
Measuring tube lining
Resistance to partial vacuum of measuring tube liningLimit values for abs. pressure [mbar] at various fluid temperatures
[mm] [inch] 25 °C 70 °C 80° C 100 °C 130 °C 150 °C 180 °C
25...2000 1...78" Polyurethane 0 0 − − − − −
65...2000 3...78" Hard rubber 0 0 0 − − − −
Promag PNominal diameter
Measuring tube lining
Resistance to partial vacuum of measuring tube liningLimit values for abs. pressure [mbar] at various fluid temperatures
[mm] [inch] 25 °C 80° C 100 °C 130 °C 150 °C 180 °C
15 1/2" PTFE 0 0 0 100 − −
25 1" PTFE / PFA 0 / 0 0 / 0 0 / 0 100 / 0 − / 0 − / 0
32 − PTFE / PFA 0 / 0 0 / 0 0 / 0 100 / 0 − / 0 − / 0
40 1 1/2" PTFE / PFA 0 / 0 0 / 0 0 / 0 100 / 0 − / 0 − / 0
50 2" PTFE / PFA 0 / 0 0 / 0 0 / 0 100 / 0 − / 0 − / 0
65 − PTFE / PFA 0 / 0 * 40 / 0 130 / 0 − / 0 − / 0
80 3" PTFE / PFA 0 / 0 * 40 / 0 130 / 0 − / 0 − / 0
100 4" PTFE / PFA 0 / 0 * 135 / 0 170 / 0 − / 0 − / 0
125 − PTFE / PFA 135 / 0 * 240 / 0 385 / 0 − / 0 − / 0
150 6" PTFE / PFA 135 / 0 * 240 / 0 385 / 0 − / 0 − / 0
200 8" PTFE / PFA 200 / 0 * 290 / 0 410 / 0 − / 0 − / 0
250 10" PTFE 330 * 400 530 − −
300 12" PTFE 400 * 500 630 − −
350 14" PTFE 470 * 600 730 − −
400 16" PTFE 540 * 670 800 − −
450 18" PTFE
Partial vacuum is impermissible500 20" PTFE
600 24" PTFE
* No value can be quoted.
Promag HNominal diameter
Measuring tube lining
Resistance to partial vacuum of measuring tube liningLimit values for abs. pressure [mbar] at various fluid temperatures
[mm] [inch] 25 °C 80° C 100 °C 130 °C 150 °C 180 °C
2...100 1/12...4" PFA 0 0 0 0 0 0
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 143
10.4 Weight details
Weight details Promag W in kg
Nominal diameter
Compact version Remote version (without cable)
[mm] [inch] DIN ANSI / AWWA SensorWall
housing
25 1"
PN
40
7.3
Cla
ss 1
50
7.3
PN
40
5.3
Cla
ss 1
50
5.3 6.0
32 1 1/4" 8.0 – 6.0 – 6.0
40 1 1/2" 9.4 9.4 7.4 7.4 6.0
50 2" 0.6 10.6 8.6 8.6 6.0
65 2 1/2"P
N 1
612.0 –
PN
16
10.0 – 6.0
80 3" 14.0 14.0 12.0 12.0 6.0
100 4" 16.0 16.0 14.0 14.0 6.0
125 5" 21.5 – 19.5 – 6.0
150 6" 25.5 25.5 23.5 23.5 6.0
200 8"
PN
10
45 45
PN
10
43 43 6.0
250 10" 65 75 63 73 6.0
300 12" 70 110 68 108 6.0
350 14" 115 175 113 173 6.0
400 16" 135 205 133 203 6.0
450 18" 175 255 173 253 6.0
500 20" 175 285 173 283 6.0
600 24" 235 405 233 403 6.0
700 28" 355
Cla
ss D
400 353C
lass
D398 6.0
– 30" – 460 – 458 6.0
800 32" 435 550 433 548 6.0
900 36" 575 800 573 798 6.0
1000 40" 700 900 698 898 6.0
– 42"
PN
6
– 1100
PN
6
– 1098 6.0
1200 48" 850 1400 848 1398 6.0
– 54" – 2200 – 2198 6.0
1400 – 1300 – 1298 – 6.0
– 60" – 2700 – 2698 6.0
1600 – 1700 – 1698 – 6.0
– 66" – 3700 – 3698 6.0
1800 72" 2200 4100 2198 4098 6.0
– 78" – 4600 – 4598 6.0
2000 – 2800 – 2798 – 6.0
Transmitter Promag (compact version): 3.4 kgHigh-temperature version: +1.5 kg(Weight details apply for standard pressure ratings without packing material)
10 Technical data Promag 53 PROFIBUS-DP/-PA
144 Endress+Hauser
Weight details Promag P in kg
Nominal diameter
Compact version Remote version (without cable)
[mm] [inch] DIN ANSI / AWWA SensorWall
housing
15 1/2"
PN
40
6.5
Cla
ss 1
50
6.5
PN
40
4.5
Cla
ss 1
50
4.5 6.0
25 1" 7.3 7.3 5.3 5.3 6.0
32 1 1/4" 8.0 – 6.0 – 6.0
40 1 1/2" 9.4 9.4 7.4 7.4 6.0
50 2" 10.6 10.6 8.6 8.6 6.0
65 2 1/2"P
N 1
612.0 –
PN
16
10.0 – 6.0
80 3" 14.0 14.0 12.0 12.0 6.0
100 4" 16.0 16.0 14.0 14.0 6.0
125 5" 21.5 – 19.5 – 6.0
150 6" 25.5 25.5 23.5 23.5 6.0
200 8"
PN
10
45 45
PN
10
43 43 6.0
250 10" 65 75 63 73 6.0
300 12" 70 110 68 108 6.0
350 14" 115 175 113 173 6.0
400 16" 135 205 133 203 6.0
450 18" 175 255 173 253 6.0
500 20" 175 285 173 283 6.0
600 24" 235 405 233 403 6.0
Transmitter Promag (compact version): 3.4 kgHigh-temperature version: +1.5 kg(Weight details apply for standard pressure ratings without packing material)
Weight details Promag H in kg
Nominal diameter
Compact version Remote version (without cable)
[mm] [inch] DIN Sensor Wall housing
2 1/12" 5.2 2.5 6.0
4 5/32" 5.2 2.5 6.0
8 5/16" 5.3 2.5 6.0
15 1/2" 5.4 2.6 6.0
25 1" 5.5 2.8 6.0
40 1 1/2" 6.5 4.5 6.0
50 2" 9.0 7.0 6.0
65 2 1/2" 9.5 7.5 6.0
80 3" 19.0 17.0 6.0
100 4" 18.5 16.5 6.0
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 145
10.5 Dimensions of wall-mount housing
Fig. 58: Dimensions of wall-mounted housing (control panel mounting and pipe mounting → Page 42)
F06-
x3xx
xxxx
-06-
03-x
x-xx
-000
10 Technical data Promag 53 PROFIBUS-DP/-PA
146 Endress+Hauser
10.6 Dimensions Promag 53 W
Promag W / DN ≤ 300 (compact version)
Fig. 59: Dimensions Promag W / DN ≤ 300 (compact version)
DN L A B C K E
DIN[mm]
ANSI[inch] [mm] [mm] [mm] [mm] [mm] [mm]
25 1" 200 341 257 84 120 94
32 − 200 341 257 84 120 94
40 1 1/2" 200 341 257 84 120 94
50 2" 200 341 257 84 120 94
65 − 200 391 282 109 180 94
80 3" 200 391 282 109 180 94
100 4" 250 391 282 109 180 94
125 − 250 472 322 150 260 140
150 6" 300 472 322 150 260 140
200 8" 350 527 347 180 324 156
250 10" 450 577 372 205 400 156
300 12" 500 627 397 230 460 166
The fitting length (L) is always the same, regardless of the pressure rating.
F06-
53Fx
xxxx
-06-
00-x
x-xx
-000
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 147
Promag W / DN ≤ 300 (remote version)
Fig. 60: Dimensions Promag W / DN ≤ 300 (remote version)Dimensions of wall-mount housing → see Page 145
DN L A B C K E
DIN[mm]
ANSI[inch] [mm] [mm] [mm] [mm] [mm] [mm]
25 1" 200 286 202 84 120 94
32 − 200 286 202 84 120 94
40 1 1/2" 200 286 202 84 120 94
50 2" 200 286 202 84 120 94
65 − 200 336 227 109 180 94
80 3" 200 336 227 109 180 94
100 4" 250 336 227 109 180 94
125 − 250 417 267 150 260 140
150 6" 300 417 267 150 260 140
200 8" 350 472 292 180 324 156
250 10" 450 522 317 205 400 156
300 12" 500 572 342 230 460 166
The fitting length (L) is always the same, regardless of the pressure rating.
F06-
xxFx
xxxx
-06-
05-x
x-xx
-000
10 Technical data Promag 53 PROFIBUS-DP/-PA
148 Endress+Hauser
Promag W / DN ≥ 350 (compact version)
Fig. 61: Dimensions Promag W / DN ≥ 350 (compact version)
DN L A B C K E
DIN[mm]
ANSI[inch] [mm] [mm] [mm] [mm] [mm] [mm]
350 14" 550 738.5 456.5 282.0 564 276
400 16" 600 790.5 482.5 308.0 616 276
450 18" 650 840.5 507.5 333.0 666 292
500 20" 650 891.5 533.0 358.5 717 292
600 24" 780 995.5 585.0 410.5 821 402
700 28" 910 1198.5 686.5 512.0 1024 589
750 30" 975 1198.5 686.5 512.0 1024 626
800 32" 1040 1241.5 708.0 533.5 1067 647
900 36" 1170 1394.5 784.5 610.0 1220 785
1000 40" 1300 1546.5 860.5 686.0 1372 862
1050 42" 1365 1598.5 886.5 712.0 1424 912
1200 48" 1560 1796.5 985.5 811.0 1622 992
1350 54" 1755 1998.5 1086.5 912.0 1824 1252
1400 56" 1820 2148.5 1161.5 987.0 1974 1252
1500 60" 1950 2196.5 1185.5 1011.0 2022 1392
1600 64" 2080 2286.5 1230.5 1056.0 2112 1482
1650 66" 2145 2360.5 1267.5 1093.0 2186 1482
1800 72" 2340 2550.5 1362.5 1188.0 2376 1632
2000 78" 2600 2650.5 1412.5 1238.0 2476 1732
The fitting length (L) is always the same, regardless of the pressure rating.
F06-
53Fx
xxxx
-06-
00-x
x-xx
-001
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 149
Promag W / DN ≥ 350 (remote version)
Fig. 62: Dimensions Promag W / DN ≥ 350 (remote version)Dimensions of wall-mount housing → see Page 145
DN L A B C K E
DIN[mm]
ANSI[inch] [mm] [mm] [mm] [mm] [mm] [mm]
350 14" 550 683.5 401.5 282.0 564 276
400 16" 600 735.5 427.5 308.0 616 276
450 18" 650 785.5 452.5 333.0 666 292
500 20" 650 836.5 478.0 358.5 717 292
600 24" 780 940.5 530.0 410.5 821 402
700 28" 910 1143.5 631.5 512.0 1024 589
750 30" 975 1143.5 631.5 512.0 1024 626
800 32" 1040 1186.5 653.0 533.5 1067 647
900 36" 1170 1339.5 729.5 610.0 1220 785
1000 40" 1300 1491.5 805.5 686.0 1372 862
1050 42" 1365 1543.5 831.5 712.0 1424 912
1200 48" 1560 1741.5 930.5 811.0 1622 992
1350 54" 1755 1943.5 1031.5 912.0 1824 1252
1400 56" 1820 2093.5 1106.5 987.0 1974 1252
1500 60" 1950 2141.5 1130.5 1011.0 2022 1392
1600 64" 2080 2231.5 1175.5 1056.0 2112 1482
1650 66" 2145 2305.5 1212.5 1093.0 2186 1482
1800 72" 2340 2495.5 1307.5 1188.0 2376 1632
2000 78" 2600 2595.5 1357.5 1238.0 2476 1732
The fitting length (L) is always the same, regardless of the pressure rating.
F06-
xxFx
xxxx
-06-
05-x
x-xx
-001
10 Technical data Promag 53 PROFIBUS-DP/-PA
150 Endress+Hauser
10.7 Dimensions Promag 53 P
Promag P / DN ≤ 300 (compact version)
Fig. 63: Dimensions Promag P / DN ≤ 300 (compact version)
DN L A B C K E
DIN[mm]
ANSI[inch] [mm] [mm] [mm] [mm] [mm] [mm]
15 1/2" 200 341 257 84 120 94
25 1" 200 341 257 84 120 94
32 − 200 341 257 84 120 94
40 1 1/2" 200 341 257 84 120 94
50 2" 200 341 257 84 120 94
65 − 200 391 282 109 180 94
80 3" 200 391 282 109 180 94
100 4" 250 391 282 109 180 94
125 − 250 472 322 150 260 140
150 6" 300 472 322 150 260 140
200 8" 350 527 347 180 324 156
250 10" 450 577 372 205 400 156
300 12" 500 627 397 230 460 166
The fitting length (L) is always the same, regardless of the pressure rating.
F06-
53Fx
xxxx
-06-
00-x
x-xx
-000
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 151
Promag P / DN ≤ 300 (remote version)
Fig. 64: Dimensions Promag P / DN ≤ 300 (remote version)Dimensions of wall-mount housing → see Page 145
DN L A B C K E
DIN[mm]
ANSI[inch] [mm] [mm] [mm] [mm] [mm] [mm]
15 1/2" 200 286 202 84 120 94
25 1" 200 286 202 84 120 94
32 − 200 286 202 84 120 94
40 1 1/2" 200 286 202 84 120 94
50 2" 200 286 202 84 120 94
65 − 200 336 227 109 180 94
80 3" 200 336 227 109 180 94
100 4" 250 336 227 109 180 94
125 − 250 417 267 150 260 140
150 6" 300 417 267 150 260 140
200 8" 350 472 292 180 324 156
250 10" 450 522 317 205 400 156
300 12" 500 572 342 230 460 166
The fitting length (L) is always the same, regardless of the pressure rating.
F06-
xxFx
xxxx
-06-
05-x
x-xx
-000
10 Technical data Promag 53 PROFIBUS-DP/-PA
152 Endress+Hauser
Promag P / DN ≤ 300 / high-temperature version (compact version)
Abb. 65: Dimensions of high-temperature version (Promag P, DN ≤ 300, compact)Dimensions A1, B1 = A, B of standard version plus 110 mm
Promag P / DN ≤ 300 / high-temperature version (remote version)
Abb. 66: Dimensions of high-temperature version (Promag P, DN ≤ 300, remote)Dimensions A1, B1 = A, B of standard version plus 110 mm
A1
B1
11
0
F06-
5xP
xxxx
x-06
-00-
00-x
x-00
0
A1
B1
11
0
F06-
xxP
xxxx
x-06
-05-
00-x
x-00
0
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 153
Promag P / DN ≥ 350 (compact version)
Fig. 67: Dimensions Promag P / DN ≥ 350 (compact version)
DN L A B C K E
DIN[mm]
ANSI[inch] [mm] [mm] [mm] [mm] [mm] [mm]
350 14" 550 738.5 456.5 282.0 564 276
400 16" 600 790.5 482.5 308.0 616 276
450 18" 650 840.5 507.5 333.0 666 292
500 20" 650 891.5 533.0 358.5 717 292
600 24" 780 995.5 585.0 410.5 821 402
The fitting length (L) is always the same, regardless of the pressure rating.F0
6-53
Fxxx
xx-0
6-00
-xx-
xx-0
01
10 Technical data Promag 53 PROFIBUS-DP/-PA
154 Endress+Hauser
Promag P / DN ≥ 350 (remote version)
Fig. 68: Dimensions Promag P / DN ≥ 350 (remote version)Dimensions of wall-mount housing → see Page 145
DN L A B C K E
DIN[mm]
ANSI[inch] [mm] [mm] [mm] [mm] [mm] [mm]
350 14" 550 683.5 401.5 282.0 564 276
400 16" 600 735.5 427.5 308.0 616 276
450 18" 650 785.5 452.5 333.0 666 292
500 20" 650 836.5 478.0 358.5 717 292
600 24" 780 940.5 530.0 410.5 821 402
The fitting length (L) is always the same, regardless of the pressure rating.
F06-
xxFx
xxxx
-06-
05-x
x-xx
-001
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 155
10.8 Dimensions of ground disks (Promag W, P)
Fig. 69: Dimensions of ground disks DN 25...300 (Promag W, P)
DN 1) A B D H
DIN[mm]
ANSI[inch] [mm] [mm] [mm] [mm]
25 1" 30 62 77.5 87.5
32 − 38.5 80 87.5 94.5
40 1 1/2" 44.5 82 101 103
50 2" 56.5 101 115.5 108
65 − 72.5 121 131.5 118
80 3" 85 131 154.5 135
100 4" 110 156 186.5 153
125 − 135 187 206.5 160
150 6" 163 217 256 184
200 8" 210.5 267 288 205
250 10" 265 328 359 240
300 2) 12" 2) 317 375 413 273
300 3) 12" 3) 317 375 404 268
1) Ground disks can, with the exception of DN 300, be used for all flange norms / pressure ratings.2) PN 10/16, Class 1503) PN 25, JIS 10K/20K
15
10
H
Ø B
Ø A
Ø D
Ø 6.5
2
F06-
xxxx
xxxx
-06-
09-0
0-xx
-001
10 Technical data Promag 53 PROFIBUS-DP/-PA
156 Endress+Hauser
10.9 Dimensions Promag 53 H
Promag H / DN 2...25 (compact version)
Fig. 70: Dimensions Promag H / DN 2...25 (compact version, aluminum field housing)
Fig. 71: Dimensions Promag H / DN 2...25 (compact version, stainless-steel field housing)
DN PN ** DI L K M
DIN[mm]
ANSI[inch]
DIN[bar] [mm] [mm] [mm] [mm]
2 − 16/40 2.25 86 43 M 6x4
4 − 16/40 4.5 86 43 M 6x4
8 − 16/40 9.0 86 43 M 6x4
15 − 16/40 16.0 86 43 M 6x4
– 1" 16/40 22.6 86 53 M 6x4
25 − 16/40 26.0 86 53 M 6x4
Overall fitting length will depend on the process connections → Page 160 ff.
** The permissible nominal pressure depends on the process connection and seal:– 40 bar: flange, welded nipples (with O-ring seal)– 16 bar: all other process connections
F06-
53H
xxxx
x-06
-00-
xx-x
x-00
1
220
K
DI
L
153
171
328
6526
3
M
Esc
E- +
F06-
53H
xxxx
x-06
-00-
xx-x
x-00
2
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 157
Promag H / DN 2...25 (remote version)
Fig. 72: Dimensions Promag H / DN 2...25 (remote version)
Wall mounting set for Promag H / DN 2...25
Abb. 73: Wall mounting set for Promag H / DN 2...25 (A = 125 mm, B = 88 mm, C = 120 mm, D = Ø 7 mm)
DN PN * DI L A B C K M
DIN[mm]
ANSI[inch]
DIN[bar] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
2 − 16/40 2.25 86 213 148 65 43 M 6x4
4 − 16/40 4.5 86 213 148 65 43 M 6x4
8 − 16/40 9.0 86 213 148 65 43 M 6x4
15 − 16/40 16.0 86 213 148 65 43 M 6x4
– 1" 16/40 22.6 86 213 148 65 53 M 6x4
26 − 16/40 26.0 86 213 148 65 53 M 6x4
Overall fitting length will depend on the process connections → Page 160 ff.
* The permissible nominal pressure depends on the process connection and seal:– 40 bar: flange, welded nipples (with O-ring seal)– 16 bar: all other process connections
F06-
xxH
xxxx
x-06
-05-
xx-x
x-00
1
A
B
C
D
F06-
xxH
xxxx
x-06
-07-
06-x
x-00
0
10 Technical data Promag 53 PROFIBUS-DP/-PA
158 Endress+Hauser
Promag H / DN 40...100 (compact version)
Fig. 74: Dimensions Promag H / DN 40...100 (compact version, aluminum field housing)
Fig. 75: Dimensions Promag H / DN 40...100 (compact version, stainless-steel field housing)
DN PN DI L A * B * C K M
DIN[mm]
ANSI[inch]
DIN[bar] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
40 1 1/2" 16 35.3 140 319 (330) 255 (266) 64 128 M 6x4
50 2" 16 48.1 140 344 (355) 267 (278) 77 153 M 8x4
65 2 1/2" 16 59.9 140 344 (355) 267 (278) 77 153 M 8x4
80 3" 16 72.6 200 394 (405) 292 (303) 102 203 M 12x4
100 4" 16 97.5 200 394 (405) 292 (303) 102 203 M 12x4
Overall fitting length will depend on the process connections → Page 168 ff.* () = Dimensions stainless steel field housing
F06-
53H
xxxx
x-06
-00-
xx-x
x-00
0
220
K
DI
L
153
171
A
CB
M
Esc
E- +
F06-
53H
xxxx
x-06
-00-
xx-x
x-00
3
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 159
Promag H / DN 40...100 (remote version)
Fig. 76: Dimensions Promag H / DN 40...100 (remote version)Dimensions of wall-mount housing → see Page 145
DN PN DI L A B C K M
DIN[mm]
ANSI[inch]
DIN[bar] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
40 1 1/2" 16 35.3 140 216 151.5 64.5 129 M 6x4
50 2" 16 48.1 140 241 164.0 77.0 154 M 8x4
65 2 1/2" 16 59.9 140 241 164.0 77.0 154 M 8x4
80 3" 16 72.6 200 290 188.5 101.5 203 M 12x4
100 4" 16 97.5 200 290 188.5 101.5 203 M 12x4
Fitting length depends on process connections → Page 168 ff.
F06-
xxH
xxxx
x-06
-05-
xx-x
x-00
0
10 Technical data Promag 53 PROFIBUS-DP/-PA
160 Endress+Hauser
10.10 Process connections Promag H (DN 2...25)
Front view of sensor Promag H / DN 2...25 (without process connections)
Fig. 77: Dimensions front view of sensor DN 2...25
DN C[mm]
D (DIN)[mm]
D (ANSI)[mm]
2...8 9 – –
15 16 – –
25 – 26 22.6
43 53
66
8.58.524
= =
==
= =
==
A
A A-A B B-B
44
B
28
60 41.6
Ø 3
4
C
Ø 4
4
D70
M6
M6
48.5-0
.1±0.
1
±0.
1±
0.1
±0.1 ±0.1 ±0.1 ±0.1
±0.
1
0 -0.10
DN 2...15 DN 25
F06-
5xH
xxxx
x-06
-05-
08-x
x-00
0
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 161
Process connections with O-ring seals (DN 2...25)
Weld nipples ISO 24631.4404 / 316L5*H**-B***********
Sensor Pipe di G L H x B
DN [mm] Tube [mm] [mm] [mm] [mm]
2...8 13.5 x 1.6 10.3 13.5 20.3 60 x 42
15 21.3 x 1.6 18.1 21.3 20.3 60 x 42
25 (DIN) 33.7 x 2 29.7 33.7 20.3 70 x 52
Fitting length = (2 x L) + 86 mm
F06
-xxH
xxxx
x-06
-09-
07-x
x-01
0
Weld nipples IPS1.4404 / 316L5*H**-C***********
Sensor Pipe di G L H x B
DN [mm] Tube (ODT / SMS) [mm] [mm] [mm] [mm]
2...8 13.5 x 2.3 9.0 13.5 20.3 60 x 42
15 21.3 x 2.65 16.0 21.3 20.3 60 x 42
1" (25 ANSI) 33.7 x 3.25 27.2 33.7 22.3 70 x 52
Fitting length = (2 x L) + 86 mm
F06
-xxH
xxxx
x-0
6-0
9-07
-xx-
012
Flange DIN 2635 / PN 401.4404 / 316L5*H**-D***********
Sensor Pipe di G L LK M H x B
DN [mm] Flange [mm] [mm] [mm] [mm] [mm] [mm]
2...8 DN 10 13.6 90 56.2 60 14 60 x 42
15 DN 15 17.3 95 56.2 65 14 60 x 42
25 (DIN) DN 25 28.5 115 56.2 85 14 70 x 52
Fitting length = (2 x L) + 86 mmFitting length to DVGW (200 mm)
F06
-xxH
xxxx
x-0
6-09
-07-
xx-0
14
Flange ANSI 16.5 / Cl 1501.4404 / 316L 5*H**-E***********
Sensor Pipe di G L LK M H x B
DN [mm] Flange [inch] [mm] [mm] [mm] [mm] [mm] [mm]
2...8 1/2" 15.7 89 66.0 60.5 15.7 60 x 42
15 1/2" 16.0 89 66.0 60.5 15.7 60 x 42
1" (25 ANSI) 1" 26.7 108 71.8 79.2 15.7 70 x 52
Fitting length = (2 x L) + 86 mm
F06
-xxH
xxxx
x-06
-09-
07-x
x-01
5
10 Technical data Promag 53 PROFIBUS-DP/-PA
162 Endress+Hauser
Flange JIS B2238 / 20K1.4404 / 316L5*H**-F***********
Sensor Pipe di G L LK M H x B
DN [mm] Flange [mm] [mm] [mm] [mm] [mm] [mm]
2...8 ND 10 10 90 67 65 15 60 x 42
15 ND 15 16 95 67 70 15 60 x 42
25 (DIN) ND 25 26 125 67 95 19 70 x 52
Fitting length = (2 x L) + 86 mm
F06
-xxH
xxxx
x-06
-09-
07-x
x-01
6
Flange DIN 2501 / PN 16 PVDF5*H**-G***********
Sensor Pipe di G L M LK H x B
DN [mm] Flange [mm] [mm] [mm] [mm] [mm] [mm]
2...8 DN 15 15.7 95 57 14 65 60 x 42
15 DN 15 15.7 95 57 14 65 60 x 42
25 (DIN) DN 25 27.3 115 57 14 85 70 x 52
– Fitting length = (2 x L) + 86 mm– Fitting length to DVGW (200 mm)– The requisite ground rings can be ordered as accessories (Order No. DK5HR-****)
F06
-xxH
xxxx
x-06
-09
-07
-xx-
029
Flange ANSI 16.5 / Cl 150 PVDF5*H**-H***********
Sensor Pipe di G L M LK H x B
DN [mm] Flange [inch] [mm] [mm] [mm] [mm] [mm] [mm]
2...8 1/2" 15.7 95 57 16 60 60 x 42
15 1/2" 15.7 95 57 16 60 60 x 42
1" (25 ANSI) 1" 27.3 115 57 16 79 70 x 52
– Fitting length = (2 x L) + 86 mm– The requisite ground rings can be ordered as accessories (Order No. DK5HR-****)
F06
-xxH
xxxx
x-0
6-09
-07-
xx-0
29
Flange JIS B2238 / 10KPVDF5*H**-J***********
Sensor Pipe di G L M LK H x B
DN [mm] Flange [mm] [mm] [mm] [mm] [mm] [mm]
2...8 ND 15 15.7 95 57 15 70 60 x 42
15 ND 15 15.7 95 57 15 70 60 x 42
25 (DIN) ND 25 27.3 125 57 19 90 70 x 52
– Fitting length = (2 x L) + 86 mm– The requisite ground rings can be ordered as accessories (Order No. DK5HR-****)
F0
6-xx
Hxx
xxx-
06-0
9-07
-xx-
029
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 163
External pipe thread ISO 228 / DIN 2999, 1.4404 / 316L5*H**-K***********
Sensor Pipe di G L S H x B
DN [mm]Internal thread
[inch] [mm] [inch] [mm] [mm] [mm]
2...8 R 3/8" 10 3/8" 40 10.1 60 x 42
15 R 1/2" 16 1/2" 40 13.2 60 x 42
1" (25 ANSI) R 1" 25 1" 40 16.5 70 x 52
Fitting length = (2 x L) + 86 mm
G di
L
H x
B
S
F06
-xxH
xxxx
x-06
-09-
07-x
x-02
5
Internal pipe thread ISO 228 / DIN 2999, 1.4404 / 316L5*H**-L***********
Sensor Pipe di G D L S H x B
DN [mm]External thread
[inch] [mm] [inch] [mm] [mm] [mm] [mm]
2...8 Rp 3/8" 8,9 3/8" 22 45 13 60 x 42
15 Rp 1/2" 16,0 1/2" 27 45 14 60 x 42
1" (25 ANSI) Rp 1" 27,2 1" 40 49 17 70 x 52
Fitting length = (2 x L) + 86 mm
D Gd
i
L
H x
B
S
F06
-xxH
xxxx
x-06
-09-
07-x
x-02
7
Hose connection1.4404 / 316L5*H**-M/N/P***********
Sensor Hose di LW L H x B
DN [mm]Inside diameter
[mm] [mm] [mm] [mm] [mm]
2...8 13 10.0 13 49 60 x 42
15 16 12.6 16 49 60 x 42
15 19 16.0 19 49 70 x 52
Fitting length = (2 x L) + 86 mm
F06
-xxH
xxxx
x-06
-09-
07-x
x-0
24
Adhesive fitting PVC5*H**-R/S***********
Sensor Pipe di G L H x B
DN [mm] Glued joint [inch] [mm] [mm] [mm] [mm]
2...8 1/2" 21.5 27.3 28.0 60 x 42
15 20 x 2 20.2 27.0 38.5 60 x 42
– Fitting length = (2 x L) + 86 mm– The requisite ground rings can be ordered as accessories (Order No. DK5HR-****)
F06
-xxH
xxxx
x-06
-09-
07-x
x-02
8
10 Technical data Promag 53 PROFIBUS-DP/-PA
164 Endress+Hauser
Process connections with aseptic gasket seal (DN 2...25 )
Weld nipple DIN 118501.4404 / 316L5*H**-U***********
Sensor Pipe di G L H x B
DN [mm] Tube [mm] [mm] [mm] [mm]
2...8 14 x 2 10 14 23.3 60 x 42
15 20 x 2 16 20 23.3 60 x 42
25 (DIN) 30 x 2 26 30 23.3 70 x 52
– Fitting length = (2 x L) + 86 mm– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06
-xxH
xxxx
x-0
6-09
-07-
xx-0
11
Weld nipples ODT/SMS1.4404 / 316L5*H**-V***********
Sensor Pipe di G L H x B
DN [mm] Tube [mm] [mm] [mm] [mm]
2...8 12.7 x 1.65 9.4 12.7 16.1 60 x 42
15 19.1 x 1.65 15.8 19.1 16.1 60 x 42
1" (25 ANSI) 24.5 x 1.65 22.1 25.4 16.1 70 x 52
– Fitting length = (2 x L) + 86 mm– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F0
6-xx
Hxx
xxx-
06-0
9-07
-xx-
013
Clamp ISO 28521.4404 / 316L5*H**-W***********
Sensor Pipe di G L H x B
DN [inch] Clamp [mm] [mm] [mm] [mm]
1" (25 ANSI) Tube 25.4 x 1.65 (ISO; 1")
22.6 50.5 44.5 70 x 52
– Fitting length = (2 x L) + 86 mm– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06
-xxH
xxxx
x-06
-09-
07-
xx-0
23
Clamp DIN 32676 1.4404 / 316L5*H**-0***********
Sensor Pipe di G L H x B
DN [mm] Clamp [mm] [mm] [mm] [mm]
2...8 Tube 14 x 2(DIN 11850; DN 10)
10 34.0 41.0 60 x 42
15 Tube 20 x 2(DIN 11850; DN 15)
16 34.0 41.0 60 x 42
25 (DIN) Tube 30 x 2(DIN 11850; DN 25)
26 50.5 44.5 70 x 52
– Fitting length = (2 x L) + 86 mm– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06
-xxH
xxxx
x-06
-09
-07
-xx-
019
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 165
Tri-clamp for ODT1.4404 / 316L5*H**-1***********
Sensor Pipe di G L H x B
DN [mm] Tri-Clamp [mm] [mm] [mm] [mm]
2...8 Tube 12.7 x 1.65(ODT 1/2")
9.4 25.0 28.5 60 x 42
15 Tube 19.1 x 1.65(ODT 3/4")
15.8 25.0 28.5 60 x 42
1" (25 ANSI) Tube 24.5 x 1.65(ODT 1")
22.1 50.4 28.5 70 x 52
– Fitting length = (2 x L) + 86 mm– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06
-xxH
xxxx
x-0
6-09
-07-
xx-0
20
Coupling DIN 118511.4404 / 316L5*H**-2***********
Sensor Pipe di G L H x B
DN [mm] Coupling [mm] [mm] [mm] [mm]
2...8 Tube 12 x 1(DN 10)
10 Rd 28 x 1/8" 44 60 x 42
15 Tube 18 x 1 or 1.5(DN 15)
16 Rd 34 x 1/8" 44 60 x 42
25 (DIN) Tube 28 x 1 or 1.5(DN 25)
26 Rd 52 x 1/6" 52 70 x 52
– Fitting length = (2 x L) + 86 mm– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06
-xxH
xxxx
x-06
-09
-07-
xx-0
17
Coupling DIN 11864-11.4404 / 316L5*H**-3***********
Sensor Pipe di G L H x B
DN [mm] Coupling [mm] [mm] [mm] [mm]
2...8 Tube 13 x 1.5(DIN 11850; DN 10)
10 Rd 28 x 1/8" 42 60 x 42
15 Tune 19 x 1.5(DIN 11850; DN 15)
16 Rd 34 x 1/8" 42 60 x 42
25 (DIN) Tube 29 x 1.5(DIN 11850; DN 25)
26 Rd 52 x 1/6" 49 70 x 52
– Fitting length = (2 x L) + 86 mm– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06
-xxH
xxxx
x-0
6-0
9-07
-xx-
021
Flange DIN 11864-21.4404 / 316L5*H**-4***********
Sensor Pipe di G L LK M H x B
DN [mm] Flange [mm] [mm] [mm] [mm] [mm] [mm]
2...8 Tube 13 x 1.5(DIN 11850; DN 10)
10 54 48.5 37 9 60 x 42
15 Tube 19 x 1.5(DIN 11850; DN 15)
16 59 48.5 42 9 60 x 42
25 (DIN) Tube 29 x 1.5(DIN 11850; DN 25)
26 70 48.5 53 9 70 x 52
– Fitting length = (2 x L) + 86 mm– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F0
6-xx
Hxx
xxx-
06-0
9-07
-xx-
022
10 Technical data Promag 53 PROFIBUS-DP/-PA
166 Endress+Hauser
Process connections orderable only as accessories (with O-ring seal, DN 2...25)
Coupling SMS 11451.4404 / 316L5*H**-5***********
Sensor Pipe di G L H x B
DN [mm] Coupling [inch] [mm] [mm] [mm] [mm]
1" (25 ANSI) 1" 22.1 Rd 40 x 1/6" 30.8 70 x 52
– Fitting length = (2 x L) + 86 mm– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06
-xxH
xxxx
x-0
6-09
-07-
xx-0
26
External pipe thread NPT1.4404 / 316LDKH**-GD**
Sensor Pipe di G L S H x B
DN [mm]Internal thread
[inch] [mm] [inch] [mm] [mm] [mm]
2...8 NPT 3/8" 10 3/8" 50 15.5 60 x 42
15 NPT 1/2" 16 1/2" 50 20.0 60 x 42
1" (25 ANSI) NPT 1" 25 1" 55 25.0 70 x 52
Fitting length = (2 x L) + 86 mm
G di
L
H x
B
S
F06
-xxH
xxxx
x-06
-09-
07-x
x-02
5
Internal pipe thread NPT1.4404 / 316LDKH**-GC**
Sensor Pipe di G D L S H x B
DN [mm]External thread
[inch] [mm] [inch] [mm] [mm] [mm] [mm]
2...8 NPT 3/8" 8,9 3/8" 22 45 13 60 x 42
15 NPT 1/2" 16,0 1/2" 27 45 14 60 x 42
1" (25 ANSI) NPT 1" 27,2 1" 40 49 17 70 x 52
Fitting length = (2 x L) + 86 mm
D Gd
i
L
H x
B
S
F06
-xxH
xxxx
x-06
-09-
07-x
x-02
7
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 167
Process connections orderable only as accessories (with gasket seal)
Ground rings (accessories)
Tri-clamp for ODT1.4404 / 316LDKH**-HF***
Sensor Pipe di G L H x B
DN [mm] Tri-Clamp [mm] [mm] [mm] [mm]
15 Tube 25.4 x 1.5(ODT; 1")
22.1 50.4 28.5 60 x 42
– Fitting length = (2 x L) + 86 mm– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06
-xxH
xxxx
x-0
6-09
-07-
xx-0
18
Grounding ring1.4435 (316L) tantalum, platinum or Alloy C-22DK5HR – ****
Sensor di D B C
DN [mm] [mm] [mm] [mm] [mm]
2...8 9.0 33.9 22.0 17.6
15 16.0 33.9 29.0 24.6
25 (DIN) 22.6 43.9 36.5 31.2
1" (25 ANSI) 26.0 43.9 39.0 34.6
Ø C
Ø d
i
Ø B
Ø D
1.9
4.53.4
0.5
3.5
Ø C
Ø d
i
Ø B
Ø D
1.9
4.53.4
0.5
3.5
F06
-xxH
xxxx
x-06
-09-
07-x
x-0
30
10 Technical data Promag 53 PROFIBUS-DP/-PA
168 Endress+Hauser
10.11 Process connections of Promag H (DN 40...100)
Front view of sensor Promag H / DN 40...100 (without process connection)
Fig. 78: Dimensions front view of sensor DN 40...100
DN A B C D E F G H L K
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] Threaded holes
40 122.0 86 71.0 51.0 35.3 M 8 15 18 – 4
50 147.0 99 83.5 63.5 48.1 M 8 15 18 – 4
65 147.0 115 100.0 76.1 59.9 M 8 15 18 6 –
80 197.0 141 121.0 88.9 72.6 M 12 15 20 – 4
100 197.0 162 141.5 114.3 97.5 M 12 15 20 6 –
F06-
5xH
xxxx
x-06
-05-
08-x
x-00
1
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 169
Process connections with gasket seal (DN 40...100)
Weld nipples DIN 118501.4404 / 316L5*H**-U***********
DN[mm]
di[mm]
G[mm]
D[mm]
L[mm]
L1[mm]
LK[mm]
40 38.0 43 92 42 19 71.0
50 50.0 55 105 42 19 83.5
65 66.0 72 121 42 21 100.0
80 81.0 87 147 42 24 121.0
100 100.0 106 168 42 24 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.F06
-xxH
xxxx
x-06
-09-
07-x
x-00
2
Weld nipples ODT1.4404 / 316L5*H**-V***********
DN[mm]
di[mm]
G[mm]
D[mm]
L[mm]
L1[mm]
LK[mm]
40 35.3 40 92 42 19 71.0
50 48.1 55 105 42 19 83.5
65 59.9 66 121 42 21 100.0
80 72.6 79 147 42 24 121.0
100 97.5 104 168 42 24 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.F06
-xxH
xxxx
x-0
6-0
9-07
-xx-
002
Clamp ISO 28521.4404 / 316L5*H**-W***********
DN[mm]
di[mm]
G[mm]
D[mm]
L[mm]
LK[mm]
40 35.6 50.5 92 68.5 71.0
50 48.6 64.0 105 68.5 83.5
65 60.3 77.5 121 68.5 100.0
80 72.9 91.0 147 68.5 121.0
100 97.6 119.0 168 68.5 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F06
-xxH
xxxx
x-06
-09-
07-x
x-00
5
Clamp DIN 32676 1.4404 / 316L5*H**-0***********
DN[mm]
di[mm]
G[mm]
D[mm]
L[mm]
LK[mm]
40 38 50.5 92 61.5 71.0
50 50 64.0 105 61.5 83.5
65 66 91.0 121 68.0 100.0
80 81 106.0 147 68.0 121.0
100 100 119.0 168 68.0 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F06
-xxH
xxxx
x-06
-09-
07-x
x-00
8
10 Technical data Promag 53 PROFIBUS-DP/-PA
170 Endress+Hauser
Tri-clamp for ODT1.4404 / 316L5*H**-1***********
DN di G D L LK
[mm] [inch] [mm] [mm] [mm] [mm] [mm]
40 1 1/2" 34.8 50.4 92 68.6 71.0
50 2" 47.5 63.9 105 68.6 83.5
65 − 60.2 77.4 121 68.6 100.0
80 3" 72.9 90.9 147 68.6 121.0
100 4" 97.4 118.9 168 68.6 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F06
-xxH
xxxx
x-06
-09-
07-x
x-0
04
Coupling DIN 118511.4404 / 316L5*H**-2***********
DN[mm]
di[mm]
G[mm]
D[mm]
L[mm]
LK[mm]
40 38 Rd 65 x 1/6" 92 72 71.0
50 50 Rd 78 x 1/6" 105 74 83.5
65 66 Rd 95 x 1/6" 121 78 100.0
80 81 Rd 110 x 1/4" 147 83 121.0
100 100 Rd 130 x 1/4" 168 92 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F06
-xxH
xxxx
x-06
-09-
07-x
x-0
01
Coupling DIN 11864-11.4404 / 316L5*H**-3***********
DN[mm]
di[mm]
G[mm]
D[mm]
L[mm]
LK[mm]
40 38 Rd 65 x 1/6" 92 71 71.0
50 50 Rd 78 x 1/6" 105 71 83.5
65 66 Rd 95 x 1/6" 121 76 100.0
80 81 Rd 110 x 1/4" 147 82 121.0
100 100 Rd 130 x 1/4" 168 90 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F06
-xxH
xxxx
x-06
-09-
07-x
x-00
6
Flange DIN 11864-2 Form A: 1.4404/316L5*H**-4***********
DN[mm]
di[mm]
G[mm]
D[mm]
L[mm]
LK1[mm]
LK2[mm]
40 38 82 92 64 71.0 65
50 50 94 105 64 83.5 77
65 66 113 121 64 100.0 95
80 81 133 147 98 121.0 112
100 100 159 168 98 141.5 137
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F06
-xxH
xxxx
x-06
-09
-07-
xx-0
07
Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 171
Coupling SMS 11451.4404 / 316L5*H**-5***********
DN[mm]
di[mm]
G[mm]
D[mm]
L[mm]
LK[mm]
40 35.5 Rd 60 x 1/6" 92 63 71.0
50 48.5 Rd 70 x 1/6" 105 65 83.5
65 60.5 Rd 85 x 1/6" 121 70 100.0
80 72.0 Rd 98 x 1/6" 147 75 121.0
100 97.6 Rd 132 x 1/6" 168 70 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F0
6-xx
Hxx
xxx-
06-0
9-07
-xx-
000
Coupling ISO 28531.4404 / 316L5*H**-6***********
DN[mm]
di[mm]
G[mm]
D[mm]
L[mm]
LK[mm]
40 35.6 50.6 92 61.5 71.0
50 48.6 64.1 105 61.5 83.5
65 60.3 77.6 121 61.5 100.0
80 72.9 91.1 147 61.5 121.0
100 97.6 118.1 168 61.5 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F06
-xxH
xxxx
x-06
-09
-07-
xx-0
03
10 Technical data Promag 53 PROFIBUS-DP/-PA
172 Endress+Hauser
Promag 53 PROFIBUS-DP/-PA Index
Endress+Hauser 173
Index
AAccessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111Adapters (installation of sensor) . . . . . . . . . . . . . . . 18Addressing
configuration using local display . . . . . . . . . . . . 86configuring the miniature switch . . . . . . . . . . . . . 84
Ambient temperature . . . . . . . . . . . . . . . . . . . . . . . 133Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131Applicator ™ (selection / configuration software) . 112Auxiliary power (supply voltage) . . . . . . . . . . . . . . 132
BBlock model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
CCable entries
degree of protection . . . . . . . . . . . . . . . . . . . . . . 61technical data . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Cable length (remote version) . . . . . . . . . . . . . . . . . 24Cable specifications
PROFIBUS-DP/-PA . . . . . . . . . . . . . . . . . . . . . . . 45remote version . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Calibration factor (default) . . . . . . . . . . . . . . . . . . . . . 8Cathode protection . . . . . . . . . . . . . . . . . . . . . . . . . 60CE mark (declaration of conformity) . . . . . . . . . . . . . 8Certification PROFIBUS-DP/-PA . . . . . . . . . . . . . . . . . 9CIP cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133Cleaning
CIP/SIP cleaning . . . . . . . . . . . . . . . . . . . . . . . . 133exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . 109
Code entry (function matrix) . . . . . . . . . . . . . . . . . . 68Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
empty-pipe and full-pipe calibration (EPD) . . . 106PROFIBUS interface (with Commuwin II) . . . . . . 87PROFIBUS interface (with local display) . . . . . . 86
Commuwin II operating program . . . . . . . . . . . . . . . 75Conditions for operation . . . . . . . . . . . . . . . . . . . . . 133Conductivity of fluid, minimum . . . . . . . . . . . . . . . . 135Configuration examples
cyclic data exchange . . . . . . . . . . . . . . . . . . . . . 98Simatic S7 HW configuration . . . . . . . . . . . . . . . 99
Connectionsee Electrical connection
Connector (fieldbus connector) . . . . . . . . . . . . . . . . 56Cycle times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104Cyclic measured variables, factory settings . . . . . . 97
DData exchange
acyclic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74cyclic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Data storage (S-DAT™, T-DAT™, F-Chip™) . . . . . 107Declaration of conformity (CE mark) . . . . . . . . . . . . . 8Degree of protection . . . . . . . . . . . . . . . . . . . . 61, 133Design
see dimensions
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Device designation . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Device functions
see “Description of device functions” manualDevice matrix (Commuwin II) . . . . . . . . . . . . . . . . . . 76Dimensions
ground disks (Promag P, W) . . . . . . . . . . . . . . . 155ground ring (Promag H) . . . . . . . . . . . . . . . . . . . 167process connections, Promag H (DN 2...25) . . . 160process connections, Promag H (DN 40...100) . 168Promag 53 H (DN 2...25) . . . . . . . . . . . . . . . . . . 156Promag 53 H (DN 40...100) . . . . . . . . . . . . . . . . 158Promag 53 P (DN ≤ 300) . . . . . . . . . . . . . . . . . . 150Promag 53 P (DN ≥ 350) . . . . . . . . . . . . . . . . . . 153Promag 53 W (DN ≤ 300) . . . . . . . . . . . . . . . . . . 146Promag 53 W (DN ≥ 350) . . . . . . . . . . . . . . . . . . 148wall-mount housing . . . . . . . . . . . . . . . . . . . . . . 145
Displaydisplay and operating elements . . . . . . . . . . . . . 64readings displayed . . . . . . . . . . . . . . . . . . . . . . . 65turning the display . . . . . . . . . . . . . . . . . . . . . . . . 43
Display value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96Documentation, supplementary . . . . . . . . . . . . . . . 139Down pipes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
EElectrical connection
cable specifications (remote version) . . . . . . . . . 52degree of protection . . . . . . . . . . . . . . . . . . . . . . 61length of connecting cable . . . . . . . . . . . . . . . . . 24post-connection check (checklist) . . . . . . . . . . . . 62potential equalisation . . . . . . . . . . . . . . . . . . . . . . 58remote version (connecting cable) . . . . . . . . . . . 50transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Electrode cleaning circuitrysee “Description of device functions” manual
ElectrodesEPD electrode . . . . . . . . . . . . . . . . . . . . . . 15, 106fitted electrodes . . . . . . . . . . . . . . . . . . . . . . . . . 137measuring-electrode plane . . . . . . . . . . . . . . . . . 15reference electrode (potential equalisation) . . . . 15replaceable electrodes (exchanging) . . . . . . . . 128
Electromagnetic compatibility (EMC) . . . . . . . 52, 133Empty Pipe Detection (EPD)
empty pipe/full pipe calibration . . . . . . . . . . . . . 106EPD electrode . . . . . . . . . . . . . . . . . . . . . . 15, 106general remarks . . . . . . . . . . . . . . . . . . . . . . . . . 106
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133Error limits
see Measuring accuracyError messages
acknowledging . . . . . . . . . . . . . . . . . . . . . . . . . . 69system / process errors . . . . . . . . . . . . . . . . . . . 115
Error types (system and process errors) . . . . . . . . . 69Ex certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138Exchange
Index Promag 53 PROFIBUS-DP/-PA
174 Endress+Hauser
device fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127measuring electrodes . . . . . . . . . . . . . . . . . . . . 128printed circuit boards . . . . . . . . . . . . . . . . . . . . 123
Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . 109
FF-CHIP™ (Function-Chip) . . . . . . . . . . . . . . . . . . . 107Fieldbus connector . . . . . . . . . . . . . . . . . . . . . . . . . 56FieldCheck ™ (test and simulation software) . . . . . 112FieldTool ™ (configuration and service software) . 112Fitting length
see DimensionsFlow rate / nominal diameter . . . . . . . . . . . . . . . . . . 18Fluid conductivity, minimal . . . . . . . . . . . . . . . . . . . . 24Fluid pressure range . . . . . . . . . . . . . . . . . . . . . . . 135Fluid temperature ranges . . . . . . . . . . . . . . . . . . . . 134Formats, Standard and Extended . . . . . . . . . . . . . . 92Full pipe calibration
see Empty Pipe DetectionFunction check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Function descriptions
see “Description of device functions” manualFunction matrix (operation via local display) . . . . . . 67Functions, function blocks, function groups . . . . . . 67Fuse, replacing . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
GGefahrgutblatt (für Rücksendung von Geräten) . . 177Ground disks
assembly (Promag P) . . . . . . . . . . . . . . . . . . . . . 32assembly (Promag W) . . . . . . . . . . . . . . . . . . . . . 26potential equalisation . . . . . . . . . . . . . . . . . . . . . 60
Ground rings (Promag H)assembly, application area . . . . . . . . . . . . . . . . . 38potential equalisation . . . . . . . . . . . . . . . . . . . . . 58
Grounding cable . . . . . . . . . . . . . . . . . . . . . . . . . . . 59GSD
producer specific GSD . . . . . . . . . . . . . . . . . . . . 90Profile GSD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
HHardware configuration
device addressing . . . . . . . . . . . . . . . . . . . . . . . 84write protection . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Hazardous substances . . . . . . . . . . . . . . . . . . . . . . . 6High-temperature version (Promag P)
installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33temperature ranges . . . . . . . . . . . . . . . . . . . . . 134
HOME position (operating mode) . . . . . . . . . . . . . . 64
IIEEE floating point number . . . . . . . . . . . . . . . . . . . . 93Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . 11Inlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Input data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95Input variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131Installation
see also Installation conditions / Installing sensorInstallation check (checklist) . . . . . . . . . . . . . . . . . . 44Installation conditions
adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18dimensions, fitting lengths . . . . . . . . . . . . . . . . . 13down pipes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14foundations (DN > 300) . . . . . . . . . . . . . . . . . . . 17inlet and outlet runs . . . . . . . . . . . . . . . . . . . . . . 16location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13orientation (vertical, horizontal) . . . . . . . . . . . . . 15partially filled pipes, drains . . . . . . . . . . . . . . . . 14pumps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13vibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Installing sensoradapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18foundations (DN > 300) . . . . . . . . . . . . . . . . . . . 17ground disks . . . . . . . . . . . . . . . . . . . . . . . . 26, 32ground rings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Promag H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Promag H with welding adapters . . . . . . . . . . . . 39Promag P . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Promag P, high-temperature version . . . . . . . . . 33Promag W . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Installing the wall-mount housing, . . . . . . . . . . . . . . 41Insulating pipes (Promag P) . . . . . . . . . . . . . . . . . . 33
LLength of connecting cable (remote version) . . . . . 24Local display
see displayLow flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
MMaintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109Material load curves . . . . . . . . . . . . . . . . . . . . . . . 137Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136Measured variable . . . . . . . . . . . . . . . . . . . . . . . . 131Measuring accuracy
deviation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132reference conditions . . . . . . . . . . . . . . . . . . . . 132repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
Measuring electrodessee Electrodes
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . 131Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . 131Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . 131Measuring tube
inside diameter . . . . . . . . . . . . . . . . . . . . . . . . . 140lining, temperature range . . . . . . . . . . . . . . . . . 134
Measuring tube specifications . . . . . . . . . . . . . . . 140
NNameplate
sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Nominal diameter / flow rate . . . . . . . . . . . . . . . . . . 18Nominal pressure
see Fluid pressure range
OOperable flow range . . . . . . . . . . . . . . . . . . . . . . . 131Operation
at a glance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Promag 53 PROFIBUS-DP/-PA Index
Endress+Hauser 175
Commuwin II (operating software) . . . . . . . . . . . 75display and operating elements . . . . . . . . . . . . . 64FieldtooI™ (configuration and service software) 75function matrix . . . . . . . . . . . . . . . . . . . . . . . . . . 67local display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Ordering codeaccessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Ordering information . . . . . . . . . . . . . . . . . . . . . . . 139Outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Output data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131Output variables . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
PPerformance characteristics
See Measuring accuracyPotential equalisation . . . . . . . . . . . . . . . . . . . . . . . . 58Power consumption . . . . . . . . . . . . . . . . . . . . . . . . 132Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . 132Pressure loss
adapters (reducers, expanders) . . . . . . . . . . . . . 18general data . . . . . . . . . . . . . . . . . . . . . . . . . . . 135linings resistance to partial vacuum . . . . . . . . . 142
Printed circuit boards (removing/installing)field housing . . . . . . . . . . . . . . . . . . . . . . . . . . . 123wall-mount housing . . . . . . . . . . . . . . . . . . . . . . 125
Process connections . . . . . . . . . . . . . . . . . . . . . . . 137Process error (definition) . . . . . . . . . . . . . . . . . . . . . 69Process error messages . . . . . . . . . . . . . . . . . . . . 115Process errors without messages . . . . . . . . . . . . . 121PROFIBUS-DP/-PA
cable specifications . . . . . . . . . . . . . . . . . . . . . . 45certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9configure device address . . . . . . . . . . . . . . . . . . 84general information . . . . . . . . . . . . . . . . . . . . . . . 70shielding, grounding . . . . . . . . . . . . . . . . . . . . . . 46
Programming modedisable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Pumps, location . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
RRegistered trademarks . . . . . . . . . . . . . . . . . . . . . . . 9Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Repeatability (measuring accuracy) . . . . . . . . . . . 132Replaceable electrodes, exchanging . . . . . . . . . . 128Resistance to shock . . . . . . . . . . . . . . . . . . . . . . . . 133Resistance to vibrations . . . . . . . . . . . . . . . . . . . . . 133Returning devices . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
SSafety icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Safety of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . 138S-DAT™ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107Seals
Promag H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Promag P . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Promag W . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25temperature ranges (Promag H) . . . . . . . . . . . . 135
Sensor installationsee Installation
Serial number . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7, 8Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131SIP cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133Software
amplifier display . . . . . . . . . . . . . . . . . . . . . . . . . . 85versions (history) . . . . . . . . . . . . . . . . . . . . . . . . 130
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122Status code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Supplementary Ex documentation . . . . . . . . . . . . . . . 5Supply voltage (auxiliary power) . . . . . . . . . . . . . . . 132System architecture
PROFIBUS-DP . . . . . . . . . . . . . . . . . . . . . . . . . . . 71PROFIBUS-PA . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
System errordefinition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
System integration . . . . . . . . . . . . . . . . . . . . . . . . . . 90
TT-DAT™ (Transmitter-DAT) . . . . . . . . . . . . . . . . . . . 107Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131Temperature ranges
ambient temperature . . . . . . . . . . . . . . . . . . . . . 133fluid temperature . . . . . . . . . . . . . . . . . . . . . . . . 134storage temperature . . . . . . . . . . . . . . . . . . . . . 133
Torques of threaded fastenerssee Installation
Totalizers, control variables . . . . . . . . . . . . . . . . . . . 97Transmitter
electrical connection . . . . . . . . . . . . . . . . . . . . . . 53installing the wall-mount housing . . . . . . . . . . . . . 41length of connecting cable (remote version) . . . . 24turning the field housing (aluminum) . . . . . . . . . . 40turning the field housing (stainless steel) . . . . . . 40
Transporting the sensor . . . . . . . . . . . . . . . . . . . . . . 11Trouble-shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
VVibrations
countermeasures . . . . . . . . . . . . . . . . . . . . . . . . . 16resistance to vibration and shock . . . . . . . . . . . 133
WWall-mount housing, installing . . . . . . . . . . . . . . . . . 41Weight details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143Welding work
grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Promag H with welding adapters . . . . . . . . . . . . . 39
Wiringsee Electrical connection
Write protection (hardware) . . . . . . . . . . . . . . . . . . . 83
Index Promag 53 PROFIBUS-DP/-PA
176 Endress+Hauser