AZPES/APP Permit Field Method Training

Fall 2016

Page 2: AZPES/APP Permit Field Method Training Fall 2016
Page 3: AZPES/APP Permit Field Method Training Fall 2016
Page 4: AZPES/APP Permit Field Method Training Fall 2016

Not Approved Methods

Color Wheels

Test Strips

Pool Kits

Page 5: AZPES/APP Permit Field Method Training Fall 2016

Approved Methods

Standard Methods – 22nd Edition -4500-Cl B (2011) - Iodometric Method I

-4500-Cl C (2011) - Iodometric Method II

-4500-Cl D (2011) - Amperometric Titration

-4500-Cl E (2011) - LL Amperometric Titr.

-4500-Cl F (2011) - DPD Ferrous Titration

-4500-Cl G (2011) - DPD Colorimetric*

Page 6: AZPES/APP Permit Field Method Training Fall 2016

Most other meters by non-Hach companies are also acceptable

Page 7: AZPES/APP Permit Field Method Training Fall 2016

Calibration

Most meters (including Hach) are

factory calibrated and do not need

routine calibration.

Page 8: AZPES/APP Permit Field Method Training Fall 2016

General Procedure

Page 9: AZPES/APP Permit Field Method Training Fall 2016

General Procedure

Page 10: AZPES/APP Permit Field Method Training Fall 2016

General Procedure

*

Page 11: AZPES/APP Permit Field Method Training Fall 2016

General Procedure

Page 12: AZPES/APP Permit Field Method Training Fall 2016

General Procedure

Page 13: AZPES/APP Permit Field Method Training Fall 2016
Page 14: AZPES/APP Permit Field Method Training Fall 2016

Method Required QC Method Blank (or RB): 1 daily. If at or above Reporting Level, immediate corrective action. [DI water + DPD] LFB: At reporting level (control charts 1020B.13 or ADHS will allow recovery ± 10%). At least monthly (Goal – each run). [ADHS will allow 2° Stds, e.g. Spec√s APP Permits only] Duplicates: 1 for each matrix type daily or with each batch of 20 or fewer samples. Control Charts 1020B. Need Record for Each

Page 15: AZPES/APP Permit Field Method Training Fall 2016
Page 16: AZPES/APP Permit Field Method Training Fall 2016

NSI Solutions, Inc.

Page 17: AZPES/APP Permit Field Method Training Fall 2016
Page 18: AZPES/APP Permit Field Method Training Fall 2016

HACH

Page 19: AZPES/APP Permit Field Method Training Fall 2016

0.5 ml Pipetman

Page 20: AZPES/APP Permit Field Method Training Fall 2016

(0.5 ml of standard added) X (25-30 mg/L Cl2)

10.5 ml (sample cell volume + standard added)

= 1.19-1.43 mg/L

(12 ml sample cell volume + 0.5 ml standard added = 1.00-1.2 mg/L)

Formula for Hach Standard

Page 21: AZPES/APP Permit Field Method Training Fall 2016

Questions?