OVERVIEW AIM To evaluate the charge state separation capability of a new type of ion mobility spectrometry device which is based on a stacked-ring RF ion guide with a travelling voltage wave. METHODS A modified Q-ToF has been constructed with a travelling wave based collision cell together with enhanced acquisition software to provide arrival time data of the ions at the ToF analyzer. RESULTS The travelling wave based mobility separator exhibits good separation characteristics at an operating pressure of 0.2 mbar with no loss in sensitivity. Charge state separation parameters were derived from a peptide mix and efficient singly charged species removal was obtained. Application of the charge state separation parameters to a mobility separated protein digest yielded efficient removal of the singly charged ions indicating the approach to be generic. INTRODUCTION The improvement in instrumental sensitivity specifically in the field of proteomics is a major driving force for manufacturers. However, quite often the doubly charged peptide ions of interest for MS/MS analysis from a protein digest can be present at analytically useful levels but are swamped by singly charged background, preventing their detection. The utility of mobility separation of ions with similar m/z values but different charge states to enhance detection limits has been shown to be effective when used in a linked scan mode with a quadrupole operating in RF-only mode [1]. In this study it was shown that the observed systematic difference in drift times for ions of different charge state can be used to effect a low-mass cut-off for removal of singly charged species. This is shown schematically in Figure 1. Presented here are the investigations into the charge state separation capabilities of a novel travelling wave based ion mobility separator in conjunction with an oa-ToF. Low m/z Cut-Off Multiply Charged Singly Charged Drift Time Mass-to-Charge Ratio Figure 1. Schematic showing low mass cut-off capability with ion mobility separation. EVALUATION OF CHARGE STATE SEPARATION USING A NOVEL TRAVELLING WAVE-BASED ION MOBILITY DEVICE COUPLED WITH AN OA-TOF Kevin Giles, Jason L. Wildgoose, Steven D. Pringle, Robert H. Bateman Waters Corporation, Manchester, UK Quadrupole Ions from Electrospray Source Ion Gate Travelling Wave Mobility Separator (TWIMS)/ Collision Cell Source Ion Guide RF-Only Hexapoles Pusher Detector Reflectron 250 l/s Turbo 250 l/s Turbo 250 l/s Turbo 250 l/s Turbo Figure 3. A schematic diagram of the modified hybrid quadrupole oa-ToF instrument. Printed Circuit Boards Gas In Ion Transmission Aperture End Plate Side Plate Ring Electrodes Figure 4. Travelling wave collision cell. OVERVIEW OF TRAVELLING WAVE MOBILITY SEPARATION Previous work has shown that superposition of a travelling voltage wave on the confining RF of a stacked ring ion guide efficiently propels ions through a collision cell [2], reducing their residence time by ions ‘surfing’ on the wave as shown in Figure 2(a). However, by reducing the wave amplitude or increasing the velocity ions of lower mobility can no longer keep up with the wave and roll over the top, Figure 2(b). This process is repeated as successive waves pass through the cell and produces a mobility-based separation of the entrained ions. Through appropriate mass range selection based on ion arrival time at the ToF analyser following mobility separation, discrimination against singly charged species can be achieved. EXPERIMENTAL The experiments were performed on a modified Waters hybrid Quadrupole oa-ToF instrument, shown in Figure 3. The Travelling Wave Ion Mobility Separator (TWIMS) is based on a Waters Quattro Premier T-Wave collision cell, a photograph of which is shown in Figure 4. The TWIMS is housed in a separately pumped chamber to allow use of elevated pressures in the cell. The cell has 122 ring electrodes mounted between two printed circuit boards (PCB). Each ring has an internal diameter of 5 mm, a thickness of 0.5 mm and is spaced 1.0 mm from the next. The cell is terminated by DC only lenses with 2 mm diameter apertures. The electrodes in the cell are divided into repeat sections of seven pairs with interconnections made on the supporting PCBs. Each pair of electrodes are connected to separate DC amplifiers which supply both the travelling wave voltage and any DC offset. The confining phase/anti-phase RF voltages for each pair of electrodes are superimposed onto the DC voltages via secondary windings of an RF transformer. The travelling wave pulse pattern is controlled via a programmable logic device (PLD) connected to a PC. Figure 5 shows a functional diagram of the TWIMS control system with the electrode grouping illustrated using matching colours. Ion mobility spectra are acquired by storing ions in the source ion guide and gating them periodically to the TWIMS using the ion gate. Temporal acquisition of the mobility separated ions is provided by the ToF data acquisition system. This has been modified such that following a trigger from the ion gate, successive ‘pushes’ or mass spectra are recorded separately in sequence until 200 have been acquired; then the next gate pulse starts the process again, with the new set of 200 spectra added to the previous 200 and so on. The timings are shown in Figure 6. Through analysis of the mass spectral data as a function of drift time a low mass cut-off curve can be deter- mined (see Fig. 1) to process the data for removal of singly charged ions. 100 200 300 400 500 600 700 800 900 1000 m/z 0 100 % 556.26 255.09 136.07 185.10 513.29 416.20 343.21 450.23 647.35 943.75 787.39 696.40 745.93 879.98 844.45 Figure 10. Mass spectrum of the 11 peptide mix with no mobility separation. 0 2 4 6 8 10 12 14 Drift Time (msec) Intensity A B C D Figure 11. Mobility separation of the peptide mix using a pulse height of 6 V. 100 200 300 400 500 600 700 800 900 1000 1100 m/z 0 100 % 0 100 % 0 100 % 0 100 % 255.09 571.34 419.24 371.21 639.85 712.44 556.25 513.28 787.40 647.34 943.74 784.41 676.36 647.37 710.33 949.60 Multiply Charged 325.68 (A) (D) (C) (B) Figure 12. Mass spectra obtained from the drift time regions identified in Figure 11 showing distinct temporal separation of singly / multiply charged species. 100 200 300 400 500 600 700 800 900 1000 1100 0 2 4 6 8 10 12 14 Drift Time (msec) Low Mass Cut Off (Da) Figure 13. Low mass cut-off curve as a function of drift time for singly charged ion removal. 100 200 300 400 500 600 700 800 900 1000 m/z 0 100 % 571.32 523.76 450.23 392.71 325.67 625.33 943.76 787.39 696.90 745.92 879.98 844.45 Figure 14. Peptide mix mass spectrum after singly charged ion removal. 500 505 510 515 520 525 530 535 540 545 550 555 560 m/z 0 100 % 0 100 % 556.26 513.29 501.25 506.28 501.76 504.34 507.29 523.76 514.29 515.29 524.28 534.27 530.79 527.32 541.66 535.29 555.30 557.29 558.31 523.76 501.25 501.76 514.77 502.27 524.28 530.79 524.78 541.66 531.29 542.34 557.36 (B) with singly charged ion removal (A) without singly charged ion removal Figure 15. Comparison of data without/with singly charged ion removal illustrating the effectiveness of the technique. 100 200 300 400 500 600 700 800 900 1000 m/z 0 100 % 878.46 814.48 644.84 560.31 478.28 392.20 391.19 322.19 234.14 175.12 157.04 246.17 390.21 331.24 448.28 520.33 632.37 580.31 807.43 672.42 707.40 815.47 821.44 878.97 911.47 912.47 920.97 971.44 972.42 Figure 16. Mass spectrum of Yeast Enolase tryptic digest without mobility separation. 0 2 4 6 8 10 12 14 Drift Time (msec) Intensity Figure 17. Mobility separation of the Yeast Enolase tryptic digest using a pulse height of 6 V. 100 200 300 400 500 600 700 800 900 1000 m/z 0 100 % 878.46 644.84 643.84 392.20 362.23 333.18 580.31 514.57 392.72 486.23 393.23 457.89 514.92 550.77 789.90 706.90 709.36 709.88 790.40 814.75 815.08 815.41 877.43 824.92 878.97 911.46 912.47 920.97 921.96 938.00 944.47 Figure 18: Yeast Enolase digest mass spectrum after singly charged ion removal. 370 375 380 385 390 395 400 405 410 415 420 m/z 0 100 % 0 100 % 392.20 374.23 377.20 387.20 383.23 404.24 401.30 397.23 409.23 407.24 414.25 412.19 418.24 392.20 383.23 378.74 387.22 404.23 407.76 (B) with singly charged ion removal (A) without singly charged ion removal Figure 20. Comparison of digest data without/with singly charged ion removal. m/z Drift Time Pusher Pulse (0..200) Mass Spectra TIC Gate Pulse Figure 6. Ion mobility acquisition timing diagram. Figure 7. TWIMS separation capability for Gramicidin-S (m/z 571) and Leucine Enkephalin (m/z 556) at 0.2 mbar. 0 2 4 6 8 10 0 1 2 3 4 5 6 7 8 9 10 11 Pulse Height (V) Drift Time (msec) m/z 571 (2+) FWHM m/z 556 (1+) FWHM Pulse Transit Time Mass Chromatograms at 7V Figure 9. Transmission of the m/z 785 peak of Glu- Fibrinopeptide in mobility mode at 0.2 mbar compared with transmission in non-mobility mode. The corrected data account for dead-time effects in the acquisition system. 0 20 40 60 80 100 120 3 4 5 6 7 8 9 10 11 Pulse Height (V) Signal Relative to Non-Mobility (%) m/z 785 (corrected) m/z 785 (uncorrected) 550 555 560 565 570 575 580 585 590 Intensity m/z m/z 556 (1+) m/z 571 (2+) (a) (c) (b) Figure 8. Mass spectra obtained from the mobility separation of Gramicidin-s and Leucine Enkephalin at 0.2 mbar (a) no separation (b) m/z 556 mobility peak and (c) m/z 571 mobility peak from Figure 7. distinct separation between singly and multiply charged ions as a function of drift time. Through combining groups of five spectra (equivalent to 0.375 msec slices) across the complete mobility spectrum and determining the low mass cut-off point in each, a cut-off curve has been generated and is shown in Figure 13. Using the data in Figure 13, a visual basic program was written to process the mobility data of Figure 11 and generate a full mass spectrum with singly charged ion removal for comparison with Figure 10. The result is shown in Figure 14. The spectrum is considerably simpler than that of Figure 10. Closer inspection over a reduced mass range, see Figure 15, illustrates that efficient removal of the singly charged species has been achieved. CHARGE STATE SEPARATION OF A PROTEIN DIGEST To evaluate whether the parameters used above for the travelling wave and low mass cut-off can provide charge state separation for other mixtures, a tryptic digest of Yeast Enolase (500 fmol/μL in 1:1 Acetonitrile:Water 1% Acetic Acid) was infused into the TWIMS system. The mass spectrum obtained without mobility separation is shown in Figure 16, and the mobility separation shown in Figure 17. Processing the mobility data as above generated the mass spectrum shown in Figure 18, which, again, is significantly simpler than that in Figure 16. Comparisons of the spectra in Figures 16 and 18 over reduced mass ranges are shown in Figures 19, 20 and 21. From these it can be seen that efficient singly charged state species removal has been effected over the complete mass range, making identification of multiply charged species easier. CONCLUSIONS A new mobility separation device has been developed which utilises a travelling voltage wave superimposed on the confining RF of a stacked-ring ion guide to provide ion separation. Through appropriate choice of operating conditions the mobility separation can be used to provide charge state separation for efficient removal of singly charged background ions from peptide mixtures, making the identification of low level doubly charged species easier. The charge state removal parameters derived from a peptide mix have been shown to be applicable to a tryptic digest of Yeast Enolase indicating that the parameters may be generic for all peptide mixtures, further work will be needed to establish this. REFERENCES 1. “Charge State Separation by Linked Scanning a Mass Filter with Mobility Separator” J. Hoyes, J. Wildgoose, K. Giles and S. Pringle— Presented at the 51 st ASMS Conference, Montreal, Canada 2003. 2. “Travelling Wave Ion Propulsion in Collision Cells” K. Giles, S. Pringle, K. Worthington and R. Bateman—Presented at the 51 st ASMS Conference, Montreal, Canada 2003. The travelling wave device described here is similar to that described by Kirchner in US Patent 5,206,506 (1993). In these studies, the TWIMS cell was operated with 0.2 mbar of Argon and with an applied RF of 350 V pk-pk at 2.7 MHz. An ion gate pulse width of 200 μs was used and a wave velocity of 300 m/sec. All ions were generated using electrospray ionisation with a sample infusion rate of 5 μL/min. RESULTS SYSTEM EVALUATION Studies were carried out to evaluate the separation capability and transmission efficiency of the TWIMS device. Figure 7 shows the mobility separation of the doubly charged Gramicidin-S ion (m/z 571) and the singly charged Leucine Enkephalin ion (m/z 556) as a function of travelling wave pulse height. Also shown are the mass chromatograms obtained at a pulse height of 7 V. By combining mass spectra separately across each peak, the spectra in Figure 8 are obtained, showing good differentiation between the ion species. Pulse Drivers Secondary Windings RF Generator and Primary Winding (2.7 MHz) Repeat Time Applied DC Voltage PLD Controller PC Ions in Ions out Ring Electrode Ions Figure 5. Functional diagram of the travelling wave control system. Figure 19. Comparison of digest data without/with singly charged ion removal. The data presented in Figure 9 shows the transmission of the doubly charged ion of Glu-Fibrinopeptide (m/z 785) (100 fmol/μL in 1:1 Methanol:Water 1% acetic acid) in mobility separation mode compared to that with no travelling wave/ ion gating. The ‘corrected’ data account for dead-time losses in the detection system due to multiple ion arrival events per mass spectrum. The data indicate that there is essentially no loss in sensitivity when using the mobility mode compared to standard operation with pulse voltages up to 8 V. CHARGE STATE SEPARATION OF A PEPTIDE MIX A mixture of 11 peptides (MW ranges from 555 – 2845 Da) was infused into the TWIMS system and the ToF mass obtained spectrum is shown in Figure 10. Figure 11 shows the mobility separation of the peptide mix using a travelling wave pulse height of 6 V. The mass spectra obtained by combining over the labelled regions A, B, C and D are shown in Figure 12 where it can be seen that there is a 625 630 635 640 645 650 655 660 665 m/z 0 100 % 0 100 % 644.84 632.37 625.33 629.35 637.32 659.37 655.34 651.37 644.84 625.33 623.34 636.37 630.36 640.36 658.83 655.88 (B) with singly charged ion removal (A) without singly charged ion removal Figure 21. Comparison of digest data without/with singly charged ion removal. 770 775 780 785 790 795 800 805 810 815 820 825 830 m/z 0 100 % 0 100 % 814.48 807.43 789.90 783.44 779.42 787.43 785.42 800.40 796.43 805.41 812.42 821.44 824.41 829.46 789.90 787.43 779.42 785.39 814.75 807.44 804.42 800.93 820.92 828.93 (B) with singly charged ion removal (A) without singly charged ion removal Stacked-Ring Ion Guide Ring Electrode Low Mobility Ion High Mobility Ion Time Ions Travelling Wave Pulse (a) (b) Figure 2. Propelling ions using a travelling voltage wave: (a) high pulse height (b) low pulse height. ASMS 2004 May 2004 720000884EN LL-PDF

ASMS EVALUATION OF CHARGE STATE SEPARATION USING A …€¦ · Gas In Ion Transmission Aperture End Side Plate Plate Ring Electrodes Figure 4. Travelling wave collision cell. OVERVIEW

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: ASMS EVALUATION OF CHARGE STATE SEPARATION USING A …€¦ · Gas In Ion Transmission Aperture End Side Plate Plate Ring Electrodes Figure 4. Travelling wave collision cell. OVERVIEW

A SYSTEM SOLUTION TO AUTOMATED METHOD DEVELOPMENT

OVERVIEW

AIM

• To evaluate the charge state separation capability of a new type of ion mobility spectrometry device which is based on a stacked-ring RF ion guide with a travelling voltage wave.

METHODS

• A modified Q-ToF has been constructed with a travelling wave based collision cell together with enhanced acquisition software to provide arrival time data of the ions at the ToF analyzer.

RESULTS

• The travelling wave based mobility separator exhibits good separation characteristics at an operating pressure of 0.2 mbar with no loss in sensitivity.

• Charge state separation parameters were derived from a peptide mix and efficient singly charged species removal was obtained.

• Application of the charge state separation parameters to a mobility separated protein digest yielded efficient removal of the singly charged ions indicating the approach to be generic.

INTRODUCTION

The improvement in instrumental sensitivity specifically in the field of proteomics is a major driving force for manufacturers. However, quite often the doubly charged peptide ions of interest for MS/MS analysis from a protein digest can be present at analytically useful levels but are swamped by singly charged background, preventing their detection. The utility of mobility separation of ions with similar m/z values but different charge states to enhance detection limits has been shown to be effective when used in a linked scan mode with a quadrupole operating in RF-only mode [1]. In this study it was shown that the observed systematic difference in drift times for ions of different charge state can be used to effect a low-mass cut-off for removal of singly charged species. This is shown schematically in Figure 1. Presented here are the investigations into the charge state separation capabilities of a novel travelling wave based ion mobility separator in conjunction with an oa-ToF.

Low m/z Cut-Off

Multiply Charged

Singly Charged

Drift Time

Mas

s-to-

Cha

rge

Ratio

Figure 1. Schematic showing low mass cut-off capability with ion mobility separation.

EVALUATION OF CHARGE STATE SEPARATION USING A NOVEL TRAVELLING WAVE-BASED ION MOBILITY DEVICE COUPLED WITH AN OA-TOF

Kevin Giles, Jason L. Wildgoose, Steven D. Pringle, Robert H. BatemanWaters Corporation, Manchester, UK

Quadrupole

Ions fromElectrospray

Source

Ion GateTravelling Wave Mobility

Separator (TWIMS)/Collision Cell

Source Ion Guide RF-Only

Hexapoles

Pusher

Detector

Reflectron

250 l/sTurbo

250 l/sTurbo

250 l/sTurbo

250 l/sTurbo

Figure 3. A schematic diagram of the modified hybrid quadrupole oa-ToF instrument.

PrintedCircuitBoards

Gas In

Ion TransmissionAperture

EndPlateSide

Plate

Ring Electrodes

Figure 4. Travelling wave collision cell.

OVERVIEW OF TRAVELLING WAVE MOBILITY SEPARATION

Previous work has shown that superposition of a travelling voltage wave on the confining RF of a stacked ring ion guide efficiently propels ions through a collision cell [2], reducing their residence time by ions ‘surfing’ on the wave as shown in Figure 2(a). However, by reducing the wave amplitude or increasing the velocity ions of lower mobility can no longer keep up with the wave and roll over the top, Figure 2(b).This process is repeated as successive waves pass through the cell and produces a mobility-based separation of the entrained ions. Through appropriate mass range selection based on ion arrival time at the ToF analyser following mobility separation, discrimination against singly charged species can be achieved.

EXPERIMENTAL

The experiments were performed on a modified Waters hybrid Quadrupole oa-ToF instrument, shown in Figure 3. The Travelling Wave Ion Mobility Separator (TWIMS) is based on a Waters Quattro Premier T-Wave collision cell, a photograph of which is shown in Figure 4. The TWIMS is housed in a separately pumped chamber to allow use of elevated pressures in the cell. The cell has 122 ring electrodes mounted between two printed circuit boards (PCB). Each ring has an internal diameter of 5 mm, a thickness of 0.5 mm and is spaced 1.0 mm from the next. The cell is terminated by DC only lenses with 2 mm diameter apertures.

The electrodes in the cell are divided into repeat sections of seven pairs with interconnections made on the supporting PCBs. Each pair of electrodes are connected to separate DC amplifiers which supply both the travelling wave voltage and any DC offset. The confining phase/anti-phase RF voltages

for each pair of electrodes are superimposed onto the DC voltages via secondary windings of an RF transformer. The travelling wave pulse pattern is controlled via a programmable logic device (PLD) connected to a PC. Figure 5 shows a functional diagram of the TWIMS control system with the electrode grouping illustrated using matching colours.

Ion mobility spectra are acquired by storing ions in the source ion guide and gating them periodically to the TWIMS using the ion gate. Temporal acquisition of the mobility separated ions is provided by the ToF data acquisition system. This has been modified such that following a trigger from the ion gate, successive ‘pushes’ or mass spectra are recorded separately in sequence until 200 have been acquired; then the next gate pulse starts the process again, with the new set of 200 spectra added to the previous 200 and so on. The timings are shown in Figure 6. Through analysis of the mass spectral data as a function of drift time a low mass cut-off curve can be deter-mined (see Fig. 1) to process the data for removal of singly charged ions.

100 200 300 400 500 600 700 800 900 1000m/z0

100

%

556.26

255.09

136.07

185.10

513.29

416.20343.21

450.23

647.35

943.75787.39

696.40745.93

879.98

844.45

Figure 10. Mass spectrum of the 11 peptide mix with no mobility separation.

0 2 4 6 8 10 12 14Drift Time (msec)

Inte

nsity A

B

C

D

Figure 11. Mobility separation of the peptide mix using a pulse height of 6 V.

100 200 300 400 500 600 700 800 900 1000 1100m/z0

100

%

0

100

%

0

100

%

0

100

%

255.09

571.34419.24

371.21

639.85

712.44

556.25

513.28 787.40

647.34

943.74

784.41676.36

647.37710.33 949.60

Multiply Charged

325.68

(A)

(D)

(C)

(B)

Figure 12. Mass spectra obtained from the drift time regions identified in Figure 11 showing distinct temporal separation of singly / multiply charged species.

100

200

300

400

500

600

700

800

900

1000

1100

0 2 4 6 8 10 12 14Drift Time (msec)

Low

Mas

s C

ut O

ff (D

a)

Figure 13. Low mass cut-off curve as a function of drift time for singly charged ion removal.

100 200 300 400 500 600 700 800 900 1000m/z0

100

%

571.32

523.76

450.23

392.71

325.67

625.33

943.76

787.39

696.90745.92

879.98

844.45

Figure 14. Peptide mix mass spectrum after singly charged ion removal.

500 505 510 515 520 525 530 535 540 545 550 555 560m/z0

100

%

0

100

%

556.26

513.29

501.25 506.28

501.76

504.34 507.29

523.76

514.29

515.29

524.28534.27

530.79527.32 541.66

535.29555.30

557.29

558.31

523.76

501.25

501.76

514.77502.27

524.28

530.79

524.78541.66531.29

542.34557.36

(B) with singly charged ion removal

(A) without singly charged ion removal

Figure 15. Comparison of data without/with singly charged ion removal illustrating the effectiveness of the technique.

100 200 300 400 500 600 700 800 900 1000m/z0

100

%

878.46

814.48

644.84560.31

478.28

392.20

391.19322.19

234.14

175.12

157.04246.17

390.21

331.24

448.28

520.33

632.37

580.31

807.43672.42

707.40

815.47

821.44

878.97

911.47

912.47

920.97

971.44

972.42

Figure 16. Mass spectrum of Yeast Enolase tryptic digest without mobility separation.

0 2 4 6 8 10 12 14Drift Time (msec)

Inte

nsity

Figure 17. Mobility separation of the Yeast Enolase tryptic digest using a pulse height of 6 V.

100 200 300 400 500 600 700 800 900 1000m/z0

100

%

878.46

644.84

643.84

392.20

362.23333.18

580.31

514.57392.72

486.23393.23

457.89

514.92

550.77

789.90

706.90

709.36

709.88

790.40

814.75

815.08

815.41

877.43824.92

878.97

911.46

912.47

920.97

921.96

938.00

944.47

Figure 18: Yeast Enolase digest mass spectrum after singly charged ion removal.

370 375 380 385 390 395 400 405 410 415 420m/z0

100

%

0

100

%

392.20

374.23377.20

387.20383.23

404.24

401.30

397.23

409.23

407.24

414.25

412.19418.24

392.20

383.23378.74 387.22 404.23 407.76

(B) with singly charged ion removal

(A) without singly charged ion removal

Figure 20. Comparison of digest data without/with singly charged ion removal.

m/z

Drift Time

Pusher Pulse (0..200)

Mass Spectra

TIC

Gate Pulse

Figure 6. Ion mobility acquisition timing diagram.

Figure 7. TWIMS separation capability for Gramicidin-S (m/z 571) and Leucine Enkephalin (m/z 556) at 0.2 mbar.

0

2

4

6

8

10

0 1 2 3 4 5 6 7 8 9 10 11

Pulse

Hei

ght (

V)

Drift Time (msec)

m/z 571 (2+)FWHMm/z 556 (1+)FWHMPulse Transit Time

Mass Chromatograms at 7V

Figure 9. Transmission of the m/z 785 peak of Glu-Fibrinopeptide in mobility mode at 0.2 mbar compared with transmission in non-mobility mode. The corrected data account for dead-time effects in the acquisition system.

0

20

40

60

80

100

120

3 4 5 6 7 8 9 10 11

Pulse Height (V)

Sign

al R

elat

ive

to N

on-M

obili

ty (%

)

m/z 785 (corrected)

m/z 785 (uncorrected)

550 555 560 565 570 575 580 585 590

Inte

nsity

m/z

m/z 556 (1+)

m/z 571 (2+)

(a)

(c)

(b)

Figure 8. Mass spectra obtained from the mobility separation of Gramicidin-s and Leucine Enkephalin at 0.2 mbar (a) no separation (b) m/z 556 mobility peak and (c) m/z 571 mobility peak from Figure 7.

distinct separation between singly and multiply charged ions as a function of drift time.

Through combining groups of five spectra (equivalent to 0.375 msec slices) across the complete mobility spectrum and determining the low mass cut-off point in each, a cut-off curve has been generated and is shown in Figure 13.

Using the data in Figure 13, a visual basic program was written to process the mobility data of Figure 11 and generate a full mass spectrum with singly charged ion removal for comparison with Figure 10. The result is shown in Figure 14. The spectrum is considerably simpler than that of Figure 10. Closer inspection over a reduced mass range, see Figure 15, illustrates that efficient removal of the singly charged species has been achieved.

CHARGE STATE SEPARATION OF A PROTEIN DIGEST

To evaluate whether the parameters used above for the travelling wave and low mass cut-off can provide charge state separation for other mixtures, a tryptic digest of Yeast Enolase (500 fmol/µL in 1:1 Acetonitrile:Water 1% Acetic Acid) was

infused into the TWIMS system. The mass spectrum obtained without mobility separation is shown in Figure 16, and the mobility separation shown in Figure 17.

Processing the mobility data as above generated the mass spectrum shown in Figure 18, which, again, is significantly simpler than that in Figure 16.

Comparisons of the spectra in Figures 16 and 18 over reduced mass ranges are shown in Figures 19, 20 and 21. From these it can be seen that efficient singly charged state species removal has been effected over the complete mass range, making identification of multiply charged species easier.

CONCLUSIONS

A new mobility separation device has been developed which utilises a travelling voltage wave superimposed on the confining RF of a stacked-ring ion guide to provide ion separation.

Through appropriate choice of operating conditions the mobility separation can be used to provide charge state separation for efficient removal of singly charged background

ions from peptide mixtures, making the identification of low level doubly charged species easier.

The charge state removal parameters derived from a peptide mix have been shown to be applicable to a tryptic digest of Yeast Enolase indicating that the parameters may be generic for all peptide mixtures, further work will be needed to establish this.

REFERENCES

1. “Charge State Separation by Linked Scanning a Mass Filter with Mobility Separator” J. Hoyes, J. Wildgoose, K. Giles and S. Pringle—Presented at the 51st ASMS Conference, Montreal, Canada 2003.

2. “Travelling Wave Ion Propulsion in Collision Cells” K. Giles, S. Pringle, K. Worthington and R. Bateman—Presented at the 51st ASMS Conference, Montreal, Canada 2003. The travelling wave device described here is similar to that described by Kirchner in US Patent 5,206,506 (1993).

In these studies, the TWIMS cell was operated with 0.2 mbar of Argon and with an applied RF of 350 V pk-pk at 2.7 MHz. An ion gate pulse width of 200 µs was used and a wave velocity of 300 m/sec. All ions were generated using electrospray ionisation with a sample infusion rate of 5 µL/min.

RESULTS

SYSTEM EVALUATION

Studies were carried out to evaluate the separation capability and transmission efficiency of the TWIMS device. Figure 7 shows the mobility separation of the doubly charged Gramicidin-S ion (m/z 571) and the singly charged Leucine Enkephalin ion (m/z 556) as a function of travelling wave pulse height. Also shown are the mass chromatograms obtained at a pulse height of 7 V. By combining mass spectra separately across each peak, the spectra in Figure 8 are obtained, showing good differentiation between the ion species.

Pulse Drivers

SecondaryWindings

RF Generatorand Primary

Winding(2.7 MHz)

RepeatTime

AppliedDC Voltage

PLD Controller PC

Ions in Ions out

RingElectrode

Ions

Figure 5. Functional diagram of the travelling wave control system.

Figure 19. Comparison of digest data without/with singly charged ion removal.

The data presented in Figure 9 shows the transmission of the doubly charged ion of Glu-Fibrinopeptide (m/z 785) (100 fmol/µL in 1:1 Methanol:Water 1% acetic acid) in mobility separation mode compared to that with no travelling wave/ion gating. The ‘corrected’ data account for dead-time losses in the detection system due to multiple ion arrival events per mass spectrum. The data indicate that there is essentially no loss in sensitivity when using the mobility mode compared to standard operation with pulse voltages up to 8 V.

CHARGE STATE SEPARATION OF A PEPTIDE MIX

A mixture of 11 peptides (MW ranges from 555 – 2845 Da) was infused into the TWIMS system and the ToF mass obtained spectrum is shown in Figure 10.

Figure 11 shows the mobility separation of the peptide mix using a travelling wave pulse height of 6 V. The mass spectra obtained by combining over the labelled regions A, B, C and D are shown in Figure 12 where it can be seen that there is a

625 630 635 640 645 650 655 660 665m/z

0

100

%

0

100

%

644.84

632.37

625.33

629.35

637.32

659.37

655.34

651.37

644.84

625.33623.34

636.37630.36 640.36

658.83

655.88

(B) with singly charged ion removal

(A) without singly charged ion removal

Figure 21. Comparison of digest data without/with singly charged ion removal.

770 775 780 785 790 795 800 805 810 815 820 825 830m/z0

100

%

0

100

%

814.48

807.43789.90

783.44

779.42

787.43

785.42800.40

796.43

805.41

812.42

821.44

824.41829.46

789.90

787.43

779.42 785.39

814.75

807.44

804.42800.93

820.92

828.93

(B) with singly charged ion removal

(A) without singly charged ion removal

Stacked-RingIon Guide

Ring ElectrodeLow Mobility IonHigh Mobility Ion

TimeIons

Travelling Wave Pulse

(a) (b)

Figure 2. Propelling ions using a travelling voltage wave: (a) high pulse height (b) low pulse height.

ASMS 2 0 0 4

May 2004 720000884EN LL-PDF