ASEAN AnalValdn 26Sep07 Effect 7May08

Embed Size (px)

Citation preview

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    1/76

    ASEAN GUIDELINES FOR

    VALIDATION OF

    ANALYTICALPROCEDURES

    SupaneeDuangteraprecha, Ph.D.Bureau of Drug and Narcotic

    Department of Medical Sciences

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    2/76

    Objective

    The objective of validation of an analytical

    procedure is to demonstrate that it is

    suitable for its intended purpose.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    3/76

    Scope

    provide guidance and recommendation ofvalidation of the analytical procedures forsubmission as part of registration applicationswithin ASEAN.

    it mainly adopts two ICH guidelines

    Q2A:Validation of Analytical Methods: Definitions andTerminology, 27 October 1994 and ICH Q2B:Validation of Analytical Procedure: Methodology, 6November 1996.

    the methodology applied for biological andbiotechnological products may be approacheddifferently than chemical entities.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    4/76

    Types of Analytical Procedures

    to be Validated

    Identification tests.

    Quantitative tests for impurities' content.

    Limit tests for the control of impurities. Quantitative tests of the active moiety in

    samples of drug substance or drug product

    or other selected component(s) in the drugproduct.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    5/76

    Criteria For Analytical Test Validation

    PRECISION

    LIMITS

    ACCURACY

    LINEARITY RANGE

    SPECIFICITY

    REPEATABILITY

    QUANTIFICATION

    INTERMEDIATEPRECISION

    DETECTION

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    6/76

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    7/76

    Symbols

    - signifies that this characteristic is not normallyevaluated

    + signifies that this characteristic is normallyevaluated

    (1) in cases where reproducibility (see glossary)has been performed, intermediate precision is notneeded

    (2) lack of specificity of one analytical procedure

    could be compensated by other supportinganalytical procedure(s)

    (3) may be needed in some cases

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    8/76

    Specificity

    Is this analyticalprocedure specific

    for the drug under test?

    Suppose we altertest conditions

    slightly?

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    9/76

    Specificity (1)

    Specificity is the ability to assessunequivocally the analyte in the presence ofcomponents which may be expected to be

    present. Typically these might include impurities,

    degradants, matrix, etc.

    Lack of specificity of an individualanalytical procedure may be compensatedby other supporting analytical procedure(s).

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    10/76

    Specificity (2)

    An investigation of specificity should be conductedduring the validation of identification tests, thedetermination of impurities and the assay. The procedures used to demonstrate specificity will

    depend on the intended objective of the analytical

    procedure.

    It is not always possible to demonstrate that ananalytical procedure is specific for a particularanalyte (complete discrimination).

    In this case a combination of two or more analyticalprocedures is recommended to achieve the necessarylevel of discrimination.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    11/76

    Identification

    Suitable identification tests should be able todiscriminate between compounds of closely related

    structures which are likely to be present.

    The discrimination of a procedure may be confirmed by

    obtaining positive results (perhaps by comparison with

    a known reference material) from samples containing

    the analyte, coupled with negative results from samples

    which do not contain the analyte.

    In addition, the identification test may be applied to

    materials structurally similar to or closely related to the

    analyte to confirm that a positive response is not

    obtained.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    12/76

    Assay and Impurity Test(s)

    For chromatographic procedures, representativechromatograms should be used to demonstrate specificityand individual components should be appropriatelylabelled.

    Critical separations in chromatography should be

    investigated at an appropriate level. For critical separations, specificity can be demonstrated by the

    resolution of the two components which elute closest to each other.

    In cases where a non-specific assay is used, other

    supporting analytical procedures should be used todemonstrate overall specificity.

    For example, where a titration is adopted to assay the drugsubstance for release, the combination of the assay and a suitabletest for impurities can be used.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    13/76

    Impurities are available (1)

    For the assay , this should involve demonstrationof the discrimination of the analyte in the presenceof impurities and/or excipients;

    practically, this can be done by spiking puresubstances (drug substance or drug product) withappropriate levels of impurities and/or excipientsand demonstrating that the assay result is

    unaffected by the presence of these materials (bycomparison with the assay result obtained onunspiked samples).

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    14/76

    Impurities are available (2)

    For the impurity test, the discrimination

    may be established by:

    spiking drug substance or drug product with

    appropriate levels of impurities and

    demonstrating the separation of these

    impurities individually and/or from other

    components in the sample matrix.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    15/76

    Impurities are not available

    If impurity or degradation product standards areunavailable, specificity may be demonstrated bycomparing the test results of samples containingimpurities or degradation products to a secondwell-characterized procedure e.g.:

    pharmacopoeial method or other validatedanalytical procedure (independent procedure).

    As appropriate, this should include samples storedunder relevant stress conditions: light, heat, humidity, acid/base hydrolysis and

    oxidation.

    for the assay, the two results should be compared.

    for the impurity tests, the impurity profiles should becompared.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    16/76

    Linearity and Range

    Know that its a straight line

    vs

    For what concentrations is it astraight line

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    17/76

    Linearity and Range

    Know that its a straight lineversus

    For whatconcentrations is it a straight line Is it a straight line between 0.4 & 0.6 mg/mL?

    Over what range is it a straight line?

    Answer: approx 0.25-0.70 mg/mL

    Concentration

    mg/mL

    Response

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    18/76

    LINEARITY (1)

    A linear relationship should be evaluatedacross the range of the analytical procedure.

    It may be demonstrated directly on the drug

    substance by: dilution of a standard stock solution and/or

    separate weighings of synthetic mixtures of thedrug product components

    using the proposed procedure.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    19/76

    LINEARITY (2)

    Linearity should be evaluated by visual inspection of a plot ofsignals as a function of analyte concentration or content.

    If there is a linear relationship, test results should beevaluated by appropriate statistical methods, for example, bycalculation of a regression line by the method of least squares.

    In some cases, to obtain linearity between assays and sampleconcentrations, the test data may need to be subjected to amathematical transformation prior to the regression analysis.Data from the regression line itself may be helpful to providemathematical estimates of the degree of linearity.

    For the establishment of linearity, a minimum of 5concentrations is recommended.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    20/76

    Example

    Taken from:

    ASEAN Operational Manual for Implementationof GMP ed. 2000 p.403

    Seven solutions containing differentconcentrations (0.2800.520) mg/ml of ketotifenfumarate in tablet batch no. 2506 VAMG were

    assayed using HPLC

    The results were evaluated statistically and theresults shown on the following slide

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    21/76

    Example (continued)

    Concentration of ketotifen fumarate Area detected Acceptance

    criteriamg/ml %

    0.280

    0.320

    0.360

    0.400

    0.440

    0.480

    0.520

    70

    80

    90

    100

    110

    120

    130

    1473566

    1677013

    1904848

    2091215

    2293647

    2518976

    2670144

    Regression: y = ax + b

    a = 5055766.964

    b = 67608.786

    r2 = 0.9984

    0.9981.002

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    22/76

    RANGE

    The specified range is normally derived from

    linearity studies and depends on the intended

    application of the procedure.

    It is established by confirming that the analyticalprocedure provides an acceptable degree of

    linearity, accuracy and precision when applied to

    samples containing amounts of analyte within or

    at the extremes of the specified range of theanalytical procedure.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    23/76

    Minimum Specified Ranges (1)

    for the assay of a drug substance or a finished(drug) product: normally from 80 - 120 % of the test

    concentration

    for content uniformity, covering a minimum of70 - 130 % of the test concentration

    for dissolution testing: +/-20 % over thespecified range; e.g., if the specifications for a

    controlled released product cover a region from20%, after 1 hour, up to 90%, after 24 hours, the

    validated range would be 0-110% of the label claim

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    24/76

    Minimum Specified Ranges (2)

    for the determination of an impurity: from the reportinglevel of an impurity to 120% of the specification; forimpurities known to be unusually potent or to producetoxic or unexpected pharmacological effects, the

    detection/quantitation limit should be commensurate withthe level at which the impurities must be controlled.

    if assay and purity are performed together as one test andonly a 100% standard is used, linearity should cover the

    range from the reporting level of the impurities to 120% ofthe assay specification

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    25/76

    Accuracy vs precision

    What youwould like

    to see!

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    26/76

    Accuracy vs precision

    Poor accuracy Good precision

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    27/76

    Accuracy vs precision

    Poor precision Good accuracy

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    28/76

    Accuracy vs precision

    Totally hopeless!

    Poor precision Poor accuracy

    What would youcall this?

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    29/76

    So what definitions do these

    concepts lead us to in thecontext of assay validation?

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    30/76

    ACCURACY (1)

    The accuracy of an analytical procedure

    expresses the closeness of agreement

    between the value which is accepted either

    as a conventional true value or an acceptedreference value and the value found. This is

    sometimes termed trueness.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    31/76

    ACCURACY (2)

    Assay of Drug Substance:a) application of an analytical procedure to an

    analyte of known purity (e.g. reference material);

    b) comparison of the results of the proposedanalytical procedure with those of a second well-characterized procedure, the accuracy of which isstated and/or defined (independent procedure)

    c) accuracy may be inferred once precision, linearityand specificity have been established

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    32/76

    ACCURACY (3)

    Assay of Drug Product:a) application of the analytical procedure to syntheticmixtures of the drug product components to which knownquantities of the drug substance to be analysed have beenadded;

    b) in cases where it is impossible to obtain samples of all drugproduct components, it may be acceptable either to:

    add known quantities of the analyte to the drug product or

    to compare the results obtained from a second, well characterizedprocedure, the accuracy of which is stated and/or defined

    (independent procedure)

    c) accuracy may be inferred once precision, linearity andspecificity have been established.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    33/76

    ACCURACY (4)

    Impurities (Quantitation): Accuracy should be assessed on samples (drug

    substance/drug product) spiked with known amounts ofimpurities.

    In cases where it is impossible to obtain samples of certain

    impurities and/or degradation products, it is consideredacceptable to compare results obtained by an independentprocedure. The response factor of the drug substance canbe used.

    It should be clear how the individual or total impurities areto be determined e.g., weight/weight or area percent, in allcases with respect to the major analyte.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    34/76

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    35/76

    Example:

    Taken from:

    ASEAN Operational Manual for

    Implementation of GMP ed. 2000 p.405

    Nine solutions containing different

    concentrations of ketotifen fumarate

    reference standard added to ketotifen tablet

    batch no. 2506VAMG were assayed

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    36/76

    Example (continued):

    Conc. of ketotifen fumarate Area

    detected

    Recovery

    (%)

    Acceptance

    Criteriamg/ml %

    0.280

    0.320

    0.360

    0.3800.400

    0.420

    0.440

    0.480

    0.520

    70

    80

    90

    95100

    105

    110

    120

    130

    1473566

    1677013

    1904848

    19058622091215

    2180374

    2293647

    2518976

    2670144

    99.32

    99.48

    100.94

    100.51100.06

    100.03

    100.07

    101.01

    98.99

    Mean (recovery) : 100.04

    Standard deviation : 0.699

    Relative standard deviation (RSD) : 0.699 %

    98.0102.0 %

    < 2 %

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    37/76

    PRECISION

    The precision of an analytical procedure expresses the

    closeness of agreement (degree of scatter) between a seriesof measurements obtained from multiple sampling of thesame homogeneous sample under the prescribedconditions.

    Precision may be considered at three levels: repeatability, intermediate precision and reproducibility.

    Precision should be investigated using homogeneous,authentic samples. However, if it is not possible to obtain ahomogeneous sample it may be investigated usingartificially prepared samples or a sample solution.

    The precision of an analytical procedure is usuallyexpressed as the variance, standard deviation orcoefficient of variation of a series of measurements.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    38/76

    Repeatability (1)

    Repeatability expresses the precision under

    the same operating conditions over a short

    interval of time.

    Repeatability is also termed intra-assay

    precision.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    39/76

    Repeatability (2)

    Repeatability should be assessed using:

    a) a minimum of 9 determinations

    covering the specified range for the

    procedure (e.g. 3 concentrations/3

    replicates each) or

    b) a minimum of6 determinations at

    100% of the test concentration.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    40/76

    Intermediate precision

    Intermediate precision expresses within-laboratoriesvariations: different days, different analysts,different equipment, etc.

    The extent to which intermediate precision should beestablished depends on the circumstances under which the

    procedure is intended to be used.

    The applicant should establish the effects of random eventson the precision of the analytical procedure.

    Typical variations to be studied include days, analysts,

    equipment, etc. It is not considered necessary to studythese effects individually. The use of an experimentaldesign (matrix) is encouraged.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    41/76

    Reproducibility

    Reproducibility is assessed by means of an

    inter-laboratory trial.

    Reproducibility should be considered in

    case of the standardization of an analytical

    procedure, for instance, for inclusion of

    procedures in pharmacopoeias.

    These data are not part of the marketing

    authorization dossier.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    42/76

    Recommended Data

    The standard deviation, relativestandard deviation (coefficient of

    variation) and confidence interval should bereported for each type of precision

    investigated.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    43/76

    Example

    Taken from:

    ASEAN Operational Manual for

    Implementation of GMP ed. 2000 p.403

    The active ingredient, ketotifen fumarate,

    in tablets (batch no. 2506VAMG) was

    assayed seven times using HPLC and the

    reference standard

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    44/76

    Example (continued)

    Sample no. Concentration (mg/ml) Area detected1

    2

    3

    4

    5

    6

    7

    0.4

    0.4

    0.4

    0.4

    0.4

    0.4

    0.4

    1902803

    1928083

    1911457

    1915897

    1913312

    1897702

    1907019

    Mean : 1910896

    Standard deviation : 9841.78

    Relative standard deviation (RSD) : 0.515 %

    Acceptance criteria:

    Relative standard deviation (RSD): not more than 2 %

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    45/76

    Detection limit vs

    Quantitation limit

    Know that its therevsKnow how much is there

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    46/76

    Detection limit(means)

    Is any of it present?

    Is it there?

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    47/76

    Quantitation limit

    How much of it is present???

    How much of it is there?

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    48/76

    DETECTION LIMIT

    The detection limit of an individualanalytical procedure is the lowest amount ofanalyte in a sample which can be detected

    but not necessarily quantitated as an exactvalue

    Several approaches for determining thedetection limit are possible, depending on

    whether the procedure is a non-instrumental or instrumental.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    49/76

    Based on Visual Evaluation

    Visual evaluation may be used for non-

    instrumental methods but may also be used

    with instrumental methods.

    The detection limit is determined by the

    analysis of samples with known

    concentrations of analyte and by

    establishing the minimum level at which theanalyte can be reliably detected .

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    50/76

    Based on Signal-to-Noise

    This approach can only be applied to analyticalprocedures which exhibit baseline noise.

    Determination of the signal-to-noise ratio is

    performed by comparing measured signals fromsamples with known low concentrations of analytewith those of blank samples and establishing theminimum concentration at which the analyte canbe reliably detected.

    A signal-to-noise ratio between 3:1 or 2:1 isgenerally considered acceptable for estimating thedetection limit.

    B d th St d d D i ti f

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    51/76

    Based on the Standard Deviation of

    the Response and the Slope

    The detection limit (DL) may be expressed

    as:

    DL = 3.3 s/Swhere s = the standard deviation of theresponse

    S = the slope of the calibration curve

    The slope S may be estimated from the

    calibration curve of the analyte.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    52/76

    Estimate ofs

    Based on the Standard Deviation of the Blank

    Measurement of the magnitude of analyticalbackground response is performed by analyzing anappropriate number of blank samples and calculating

    the standard deviation of these responses Based on the Calibration Curve

    A specific calibration curve should be studied usingsamples containing an analyte in the range of DL.

    The residual standard deviation of a regression line orthe standard deviation of y-intercepts of regression linesmay be used as the standard deviation.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    53/76

    Recommended Data

    The detection limit and the method used fordetermining the detection limit should bepresented.

    If DL is determined based on visual evaluation orbased on signal to noise ratio, the presentation ofthe relevant chromatograms is consideredacceptable for justification.

    In cases where an estimated value for the detectionlimit is obtained by calculation or extrapolation,

    this estimate may subsequently be validated by theindependent analysis of a suitable number ofsamples known to be near or prepared at thedetection limit

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    54/76

    QUANTITATION LIMIT

    The quantitation limit of an individual analyticalprocedure is the lowest amount of analyte in asample which can be quantitatively determinedwith suitable precision and accuracy.

    The quantitation limit is a parameter of

    quantitative assays for low levels of compounds insample matrices, and is used particularly for thedetermination of impurities and/or degradationproducts.

    Several approaches for determining thequantitation limit are possible, depending onwhether the procedure is a non-instrumental orinstrumental.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    55/76

    Based on Visual Evaluation

    Visual evaluation may be used for non-instrumental methods but may also be usedwith instrumental methods.

    The quantitation limit is generallydetermined by the analysis of samples withknown concentrations of analyte and byestablishing the minimum level at which the

    analyte can be quantified with acceptableaccuracy and precision.

    B d

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    56/76

    Based on

    Signal-to-Noise Approach

    This approach can only be applied to analytical

    procedures that exhibit baseline noise.

    Determination of the signal-to-noise ratio is

    performed by comparing measured signals fromsamples with known low concentrations of analyte

    with those of blank samples and by establishing

    the minimum concentration at which the analyte

    can be reliably quantified.

    A typical signal-to-noise ratio is 10:1.

    B d th St d d D i ti f

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    57/76

    Based on the Standard Deviation of

    the Response and the Slope

    The quantitation limit (QL) may be

    expressed as:

    QL = 10 s/Swhere s = the standard deviation of theresponse

    S = the slope of the calibration curve

    The slope S may be estimated from the

    calibration curve of the analyte.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    58/76

    Estimate ofs

    Based on Standard Deviation of the Blank Measurement of the magnitude of analytical

    background response is performed by analyzing anappropriate number of blank samples and calculating

    the standard deviation of these responses. Based on the Calibration Curve

    A specific calibration curve should be studied usingsamples, containing an analyte in the range of QL.

    The residual standard deviation of a regression line orthe standard deviation of y-intercepts of regression linesmay be used as the standard deviation.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    59/76

    Recommended Data

    The quantitation limit and the method used

    for determining the quantitation limit

    should be presented.

    The limit should be subsequently validated

    by the analysis of a suitable number of

    samples known to be near or prepared at

    the quantitation limit.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    60/76

    Robustness

    Small changes do not affectthe parameters of the

    assay

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    61/76

    ROBUSTNESS

    The robustness of an analytical procedure is a measure of

    its capacity to remain unaffected by small, but deliberatevariations in method parameters and provides anindication of its reliability during normal usage.

    The evaluation of robustness should be considered duringthe development phase and depends on the type of

    procedure under study.

    If measurements are susceptible to variations in analyticalconditions, the analytical conditions should be suitablycontrolled or a precautionary statement should be includedin the procedure.

    One consequence of the evaluation of robustness should bethat a series of system suitability parameters (e.g.,resolution test) is established to ensure that the validity ofthe analytical procedure is maintained whenever used.

    T i l V i ti

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    62/76

    Typical Variations

    stability of analytical solutions,

    extraction timeLiquid chromatography:

    influence of variations of pH in a mobile phase,

    influence of variations in mobile phase composition,

    different columns (different lots and/or suppliers), temperature,

    flow rate.

    Gas chromatography:

    different columns (different lots and/or suppliers), temperature,

    flow rate.

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    63/76

    SYSTEM SUITABILITY TESTING

    System suitability testing is an integral part ofmany analytical procedures.

    The tests are based on the concept that theequipment, electronics, analytical operations andsamples to be analyzed constitute an integralsystem that can be evaluated as such.

    System suitability test parameters to beestablished for a particular procedure depend onthe type of procedure being validated. They areespecially important in the case ofchromatographic methods.

    System Suitability in

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    64/76

    System Suitability in

    Chromatography

    To verify that the resolution and reproducibility of thechromatographic system are adequate for the analysis tobe done

    The resolution, R, is specified to ensure that closely eluting

    compounds are resolved from each other

    Replicate injections of a standard preparation arecompared to ascertain whether requirements for precisionare met

    The tailing factor, T, has to meet a certain requirement,because as peak asymmetry increases, integration, andhence precision, becomes less reliable

    Evaluating validation data for an

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    65/76

    Evaluating validation data for anHPLC procedure

    Here are some suggestions

    But please note!

    - The slides that follow do not represent requirements; they

    are suggestions.- There is more than one way to do this!

    - Use judgement.

    If you are unsure, consult with experienced analysts!!

    Specificity

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    66/76

    Specificity(selectivity)

    Use some or all of these procedures:

    - Add a synthetic mixture of excipients to the sample &

    check whether the assay result for the drug is the same

    - Add some known impurities to the test sample & check

    whether they are resolved (separated from) the drug

    - Forcably degrade the active & test whether degradants are

    separated from the intact drug

    - Assess peak purity by diode array

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    67/76

    Linearity

    - Minimum of 5 concentrations

    - r2 >0.99 if possible

    - Intercept NMT 2% of response of 100% the

    working concentration

    - Confirm accuracy & precision over the

    required range

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    68/76

    Accuracy

    - Generally within +2%- Recoveries after spiking, or

    - Comparison with well-established methods & byinference

    - Arguably can be up to +10% forrelated substances

    -

    What is known about the referencestandard?

    Precision

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    69/76

    Precision- repeatability

    System repeatability

    Method repeatability

    %CV (of detectorresponse)

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    70/76

    Precision -intermediate[= ruggedness USP]

    - Use same complete analytical procedure for

    comparisons

    -

    Compare results across different analysts, days,equipment

    - Means preferably within 2%

    - Compare %CV with that for method repeatability

    Precision

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    71/76

    Precision- reproducibility

    - This is not normally a component of a dossier for

    an application to register, but if you do have to

    evaluate these data then

    - For interlab comparisons

    - Means should preferably be within 2%

    - Compare the %CV with that for method

    repeatability

    - Can use an F test, normally with 95%

    confidence

    f d

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    72/76

    Limit of detection

    - Use some or all of these procedures:

    - - Visual evaluation: A clear & symmetrical peak isvisible

    - Signal to noise ratio of 3:1 or 2:1

    - Based on statistical information:- Detection limit =

    - 3.3 x (std dev at that concentration)

    - slope

    L f

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    73/76

    Limit of quantitation

    - Use some or all of these procedures:

    - Visual evaluation: A clear & symmetrical peak

    is visible

    - Signal to noise ratio of 10:1

    - Based on statistical information:- Detection limit =

    - 10 x (std dev at that concentration)

    - slope

    R b tn

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    74/76

    Robustness

    - Use some or all of these procedures:- Compare results after altering HPLC parameters,

    eg mobile phase composition, buffer composition,pH, column type, flow rate:

    - NMT 2% difference in assay

    - Compare results after storage of test solution, egfor 24h at say 250C

    - NMT 2% difference in assay

    Evaluation of analytical validation data

    Th bj ti f th l ti l d

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    75/76

    Are the data concerning analytical validation satisfactory? YES/NOIf NO, recommended questions to the applicant appear in

    (eg page number below, or draft letter to the company on page

    )

    The objective of the analytical procedure

    The analytical technique

    Item Data provided by applicant(very briefly)

    Acceptable or not? (add comments ifnecessary, & reasons if unacceptable)

    Is a chromatogram, spectrum orsimilar provided?

    Specificity

    Linearity

    Range

    Accuracy

    Precision:Repeatability

    Precision:Intermediate

    Precision:Reproducibility

    Detection limit

    Quantitation limit

    Robustness

    System suitability (if necessary)

    Data on the reference standard

    Other evaluator comments:

  • 8/3/2019 ASEAN AnalValdn 26Sep07 Effect 7May08

    76/76

    Thank you