MC-000555v1 Key Concepts ARROW ® EZ-IO ® Intraosseous Vascular Access System

ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

  • Upload
    others

  • View
    25

  • Download
    1

Embed Size (px)

Citation preview

Page 1: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

MC-000555v1

Key Concepts

ARROW® EZ-IO® Intraosseous Vascular Access System

Page 2: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

MC-000555v1

2

The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated for

adult and pediatric patients any time vascular access is difficult to obtain in

emergent, urgent or medically necessary situations for 24 hours

Contraindications

Fracture of target bone

Infection at area of insertion

Inability to identify landmarks

IO or attempted IO access in target bone within previous 48 hours

Prosthesis or orthopedic procedure near insertion site

Adults Pediatrics

Proximal humerus

Proximal tibia

Distal tibia

Distal femur

Proximal humerus

Proximal tibia

Distal tibia

Indications

Indications & Contraindications

Page 3: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

MC-000555v1

3Anatomy and Physiology

Highly vascular, non-collapsible access

Anatomy and Physiology

Rapid flush to

displace marrow

Page 4: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

MC-000555v1

4Real-time Fluoroscopy – Human Model

Page 5: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

5

MC-000555

Proximal Humerus Site Identification

Page 6: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

6

MC-000555

Tibia Site Identification

Infant/ChildAdult

Page 7: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

MC-000555v1

7Distal Femur Site Identification - Pediatrics

• Secure the leg out-stretched

• Ensure the knee does not bend

Page 8: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

MC-000555v1

8ARROW® EZ-IO® Needle Set Selection

8

15 mm

15 gauge

Indicated for patients

weighing 3-39 kg

25 mm

15 gauge

Indicated for

patients weighing

3 kg or over

45 mm

15 gauge

Indicated for patients

weighing 40 kg or over,

excessive tissue depth

Clinical judgment should be used to determine appropriate

needle set selection based on patient weight, anatomy and

tissue depth overlying the insertion site

Page 9: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

9

MC-000555

Needle Set Selection

• Estimate tissue depth

• Confirm with 5 mm mark visible above the skin

Page 10: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

10

MC-000555

•Prep site and prepare supplies•Open EZ-Stabilizer® Dressing•Prime the EZ-Connect® Extension Set •Attach Needle Set to Driver

Prepare Supplies

Page 11: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

11

MC-000555

Insertion Technique

• Use minimal pressure

• Stabilize hub when removing driver

• Apply EZ-Stabilizer® Dressing

• Firmly attach primed EZ-Connect® Extension Set

• Immobilize extremity for humerus and femur insertions

Page 12: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

12

MC-000555

Flush/Infusions/Medications

12

Consider for blood typing and other commonly ordered labs

FLUSH

Adults: 5-10 mL

Infants/Children: 2-5 mL

INFUSIONS

• For optimal flow infuse with

pressure

• Administer medications in same

dose, rate and concentration as

given via peripheral IV

Page 13: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

13

MC-000555

IO Infusion Pain Management

Observe cautions/contraindications for lidocaine, confirm dose per institution

LidocaineInitial dose

120 seconds

Dwell60 seconds

Rapid FlushLidocaine

½ initial dose60 seconds

2% lidocaine (preservative-free and epinephrine-free) Adult: Typically 40mg Infant/Child: Typically 0.5 mg/kg (NOT to exceed 40 mg)

≥ 4 minutes total time

www.eziocomfort.com

Page 14: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

MC-000555v1

14Assessment/Documentation/Removal

• Assess frequently to ensure:

• Wrist band is on patient

• No evidence of complications

− Catheter is intact/patent

− No evidence of infiltration/

extravasation

• Dressing & connections are

secure

• Repeat flush as needed

• Need for initial or additional

lidocaine

• Removal

Page 15: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

MC-000555v1

15Connect With Us

24 Hour Clinical Support 800-680-4911

Education Resources

www.teleflex.com/ezioeducation

Connect with the App

for iPhone & Android

15

Page 16: ARROW EZ-IO Intraosseous Vascular Access System arrow-ez-io-key-concepts_pp_mc-000555-rev1 (1).pdfMC-000555v1 2 The ARROW® EZ-IO® Intraosseous Vascular Access System is indicated

MC-000555v1

16

THANK YOU

STERRAD is a registered trademark of Johnson and Johnson Company.

Robertsite is a registered trademark of Halkey-Roberts Corporation.

NeedleVISE is a registered trademark of Atrion Medical Products, Inc.

Level 1 is a registered trademark of Smiths Medical ASD, Inc.

Teleflex, EZ-IO, EZ-Connect and EZ-Stabilizer are trademarks or registered trademarks of Teleflex Incorporated or its

affiliates.

Vidacare LLC is a wholly owned subsidiary of Teleflex Incorporated.

© 2014 Teleflex Incorporated. All rights reserved.

This material is not intended to replace standard clinical education and training by Vidacare LLC, a subsidiary of Teleflex

Incorporated, and should be utilized as an adjunct to more detailed information which is available about the proper use of

the product. View educational resources at www.teleflex.com/ezioeducation or contact a Teleflex clinical professional for

any detailed questions related to product insertion, maintenance, removal and other clinical education information.