Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
1
April 2013
STABILITY METER for HYDROGEN PEROXIDE
S.H.P.
S.H.P. 01_1 - OPERATION MANUAL
P/N: STA 800.0
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
2
INDEX
1. SCOPE
2. SPECIFICATION & INTERFACE
2.1 ELECTRIC INTERFACE
2.2 MECHANIC INTERFACE
2.3 HYDRAULIC INTERFACE
2.4 ENVIRONMENT CONDITIONS
2.5 COMPONENTS SPECIFICATION
3. COMPONENTS
4. PRINCIPLE OF OPERATION
5. USE & OPERATION
6. THEORY OF OPERATION
7. FORMULAS
8. ASSEMBLY INSTRUCTIONS
9. MAINTENANCE INSTRUCTIONS
10. EC DECLARATION OF CONFORMITY
REFERENCES
PROVISIONAL SPARES LIST
APPENDIX 1 – HEAING IMMERSION CIRCULATOR
APPENDIX 3 – DIFFERENTIAL PRESSURE INDICATOR
FIGURES:
1.1. SHP FRONT VIEW
3.1.-3.2.-3.3. COMPOSITION
4.1.-4.2.-4.3. ELECTRIC DIAGRAM & CABLING
5.1.-5.2.-5.3. FLOW CHARTS
8.1.-8.2.-8.3. ASSEMBLY
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
3
1. SCOPE
The SHP device measures the short-term stability or rate of decomposition of Hydrogen Peroxide.
To this purpose a sample solution of Hydrogen Peroxide of known title is maintained at constant
temperature by immersion in a liquid bath, heated by an Heating Immersion Circulator. The bath is
filled with deionized water. A Water Level Device keeps automatically constant the water level
inside the bath.
After a fixed time interval (WARM UP interval), during which the inside pressure is maintained
at ambient pressure level, the device counts, during a fixed time interval (TEST interval), the
number of elementary increments of pressure due to the Hydrogen Peroxide decomposition. At the
end of this time interval, the sample automatically emerges from the bath and the computer device
stores the final number of elementary pressure increments, which gives a measure of the Hydrogen
Peroxide rate of decomposition.
The device is represented (front view) in Figure 1.1.
It can operate, separately and independently, on two samples.
Its functionning is controlled by an industrial PLC, which guarantees:
- automatic independent operation on two samples (instruments);
- immersion and emersion from the bath (through electrical linear actuators);
- preset of Warm Up and Test interval;
- count of number of pressure increments;
- storage of measured values;
- calibration tests to check the accuracy of pressure increment measurements.
The Stability Meter for Hydrogen Peroxide Device is built and certified according to CE
standards.
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
4
2. SPECIFICATION & INTERFACE
2.1 ELECTRIC INTERFACE
The SHP device is supplied at 230 Vac +/-10%. Power consumption is less than 2 KWatt.
2.2 MECHANIC INTERFACE
SHP outside dimensions are: 605 x 800 x 430 mm (LxHxD)
SHP weight is less than 40 Kg.
2.3 HYDRAULIC INTERFACE
The SHP device requires industrial refrigerating max 2 liter/sec, pressure < 2 Bar (rel.) and
deionized leveling water, max 5 liters/hour.
2.4 ENVIRONMENT CONDITIONS
The equipment is designed to operate:
- internal use (laboratory conditions)
- ambient temperature between 5 and 40 °C
2.5 COMPONENTS SPECIFICATION
The SHP device includes the following basic components.
- N° 1 PLC Model ELCON EC601Xwith the following characteristics:
- Central Unit H83003, 16 Mhz
- Monitor Back-Lighted 4x20
- Keyboard 32 Keys
- Memories: 2 Mbit RAM; 2 Mbit Flash
- Inputs: 12 Digital
- Outputs: 12 Digital
- Supply: 24 Vdc, 20 Watt
- N° 2 Differential Pressure Switches - Model DPI 2500-2R-AZ-D (switch at 80 Pascal).
Calibration and settings are described in Appendix 2.
- N° 2 Electrically operated (24 VDC) linear actuators for sample immersion/emersion operation -
Model SKF composed of:
- actuator: CALA36X150X4G6/D24T2
- end-course switch: CAXC 33
- N° 1 Heating Immersion Circulator Model JULABO EH JU9.118.0 /1.1 KWatt
- Operating temperature range: Ambient to +200°C.Selected operating temperature = 96°C.
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
5
- N° 2 Cooling Column pirex sets, water refrigerable, to act as sample expansion chamber.
- N° 2 Pirex Flask (50 ml capacity) to be filled with the Hydrogen Peroxide solution sample (25 cc
nominal value).
- N° 2 Compensation Electrovalves Model CKD -3PA210-06G-B0-24VDC.
- N° 2 Proximity Switches Model SCHNEIDER RS 444-3243 to control the bath door opening
- N° 4 Micro Switches RS 197-7709 for safety during column movements.
- N° 1 PCB card for signal conditionning towards PLC (in/out).
- Outer P.V.C. case, mounted on a metallic frame.
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
6
3. COMPOSITION
The equipment S.H.P, P/N STA 800.0_0 is built according to the following Part Lists. N° P/N Level 0 Level 1 Level 2 Description _______________________________________________________________________ 1 400-02 X LP400.2 Assembly S.H.P. 2 401-02 X LP401.2 Main Frame 3 402-02 X LP402.2 Bath & Heating 4 403-02 X LP403.2 Instruments Plate 5 404-02 X LP404.2 Actuator 6 405-02 X LP405.2 Electronics 7 406-02 X LP406.2 Water Level 8 407-01 X LP407.1 Calibration Box
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
7
PART LIST LP400.2
LEVEL 1 ASSEMBLY: S.H.P.
P/N 400-02 Rif: Fig. 1.1
N° Q.ty DESCRIPTION P/N Rif. -------------------------------------------------------------------------------------------------------------------------- 1 1 MAIN FRAME 401-02 Fig.3.1 2 1 BATH & HEATING 402-02 Fig.3.1 3 2 INSTRUMENTS PLATE 403-02 Fig.3.1 4 2 ACTUATOR 404-02 Fig.3.2 5 1 ELECTRONICS 405.02 Fig.3.2 6 1 WATER LEVEL 406.02 Fig.3.1 7 1 CALIBRATION BOX 407.01 Fig.3.3
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
8
PART LIST LP401.1
LEVEL 2 ASSEMBLY: MAIN FRAME
P/N 401-02 Rif: Fig. 3.1
N° Q.ty DESCRIPTION P/N - Code - Drawing -------------------------------------------------------------------------------------------------------------------------- 1 1 Metal Base Frame 970101_C 2 1 PVC Base Frame 970103_B 3 1 Left Side 970104_B 4 1 Right Side 970105_B 5 1 Instrument Front 970106 6 1 Bath Front 970107_B 7 1 Top Cover 970108 8 1 Staff 970109_B 9 1 Rear Cover 970110 10 13 Self-tapping screw TBIC fi3.5x16 11 5 Self-tapping screw TBIC fi3.5x25 12 64 Self-tapping screw TSIC fi3.5x20 13 20 Self-tapping screw TBIC fi2.9x16 14 4 Levelling foot RS331-4166 (M8x25) 15 2 Water connections Inner Diameter = 8mm 16 1 Label CE 17 1 Label High Temperature 18 2 Label IN – OUT 19 1 Label Gamba e Botteghi snc 20 2 Actuator Safety Switch RS 122-9280
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
9
PART LIST LP402.2 LEVEL 2
ASSEMBLY: BATH & HEATING P/N 402-02 Rif: Fig. 3.1
N° Q.ty DESCRIPTION P/N - Code - Drawing -------------------------------------------------------------------------------------------------------------------------- 1 1 Heating Immersion Circulator JULABO-EH JU9.118.0 2 2 Proximity Switch SCHNEIDER rs 444-3243 3 1 Bath 970130_C (BBC) 4 1 Bath Cover 970112_B (BRUSA) 5 1 Thermostat Support 970115 6 1 Forward Doors Guide 970118_B 7 1 Backward Doors Guide 970119_C 8 1 Doors Separation 970125 9 2 Door 970126_B 10 1 Door 1 Handle 970128_B 11 1 Door 2 Handle 970127_B 12 4 Screw TCEI M4x20 13 2 Screw TCIC M5x15 14 18 Screw TCEI M3x15 15 18 Helicoil fi 3 16 4 Screw TCEI M6x40 17 1 Screw TCEI M4x10 18 4 Screw TSEI M3x10 19 4 Self locking bush M6
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
10
PART LIST LP403.2
LEVEL 2 ASSEMBLY: INSTRUMENTS PLATE
P/N 403-02 Rif: Fig. 3.1
N° Q.ty DESCRIPTION P/N - Code - Drawing -------------------------------------------------------------------------------------------------------------------------- 1 1 Base Plate 970113_B 2 2 Column Support 970116_B 3 2 Column Support Closure 970124 4 1 Cooling Column TECNOVETRO 5 1 Flask TECNOVETRO- 50 cc – CONE 14/23 6 1 Pressure Switch DPI 2500-2R-AZ D(Switch at 80Pa = 8 mm H2O = 0.8 mBar) 7 1 Electrovalve CKD - 3PA210-06G-B0-24VDC 8 4 Screw TCEI M5x20 9 2 Screw TCEI M5x12 10 4 Screw TSEI M5x12 11 2 Screw TSEI M4x12 12 2 Screw TSEI M3x12 13 2 Nut M4 14 1 Plexiglass Cover 220213_B 15 2 Pommel RS 702-7532 (fi22 M5x10) 16 2 Emergency Micro Switch RS 197-7709 17 2 Silicon Tubes Ext.Diam. = 6mm 18 1 Label “1” or “2” 19 1 Coupling Ring Metallic 20 1 Rubber Pass-through 21 4 Rubber Strip
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
11
PART LIST LP404.2
LEVEL 2 ASSEMBLY: ACTUATOR
P/N 404-02 Rif: Fig. 3.2
N° Q.ty DESCRIPTION P/N - Code - Drawing -------------------------------------------------------------------------------------------------------------------------- 1 1 Bearing Slot ROLLON-AS35T-400 2 1 Angle Slot 970114 3 2 Instrument Plate Support 970117 4 1 Electric Actuator SKF - CALA36X4G6/D24T2 5 2 End Course Sensor SKF - CAXC 33 6 1 Actuator Base Plate 970120 7 1 Actuator Register 970121 8 1 Actuator Base Pivot 970122 9 1 Actuator High Pivot 970123 10 3 Screw TSEI M6x10 11 4 Screw TCEI M6x10 12 4 Washer fi 6 13 4 Screw TCEI M5x10 14 4 Washer fi 5 15 4 Screw TCEI M6x15 16 1 Nut M8 17 1 Washer fi 8 18 1 Nut M12 19 2 Seeger shaft fi 9.6
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
12
PART LIST LP405.2
LEVEL 2 ASSEMBLY: ELECTRONICS
P/N 405-02 Rif: Fig. 3.2
N° Q.ty DESCRIPTION P/N - Code - Drawing -------------------------------------------------------------------------------------------------------------------------- 1 1 PLC ELCON - EC 601X 2 1 EMI Filter RFI 1 phase – RS 219-2819 3 1 Transformer 160VA-2x18 – RS 223-8774 4 1 Printed Circuit Board CAD LINE 5 1 Mag.therm.Diff.Switch BTICINO C40 V(400V-25A-30mA) 6 4 Nut M4 7 4 Washer fi 4 8 2 Self-tapping screw TCIC fi3.5x8 9 1 Screw TCEI M5x65 10 1 Nut M5 11 1 Washer fi 5 12 2 Screw TSEI M6x70 13 4 Nut M6 14 1 Omega Staff L=125 15 1 Cable-press fi 22 16 1 Diode Bridge 8A-200V (RS227-8564) 17 6 Relais 8A-24V - RS680-3678 18 8 Diode 1N4004-400V – RS 628-9029 19 1 Electrolytic Capacitor 10000uF – RS339-7048 20 1 Capacitor 0.1uF 21 1 Fuse 2A-250V
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
13
PART LIST LP406.2
LEVEL 2 ASSEMBLY: WATER LEVEL
P/N 406-02 Rif: Fig.
N° Q.ty DESCRIPTION P/N - Code - Drawing -------------------------------------------------------------------------------------------------------------------------- 1 1 Body 970131 2 1 Level Column 970132 3 1 Level Control 970133 4 1 Water Supply 970134 5 3 Self-tapping screw TSIC fi4.8x40 6 1 Washer fi 12/20 7 4 OR Parker 5-612 8 1 OR Parker 2-11 9 2 Water connections Inner Diameter = 8mm
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
14
PART LIST LP407.1 LEVEL 2
ASSEMBLY: CALIBRATION BOX P/N 407-01
Rif: Fig. 3.xxx
N° Q.ty DESCRIPTION P/N - Code - Drawing -------------------------------------------------------------------------------------------------------------------------- 1 1 Watertight suitcase RS 6662297 2 1 Dosimetric Pump STEPDOS03/S KNF FEM 03 TT-18 S 3 1 Bottle 100 ml Franceschi BTLSVL100 4 1 Set of tubes ……. 5 1 Label Gamba e Botteghi snc
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
15
4. PRINCIPLE OF OPERATION
The SHP operations are controlled by a PLC, which interfaces with device sensors and actuators
through its digital input and output and through a PCB card. The PLC controls the operations of two
instruments: N°1 (left) and N°2 (right) at the same time.
With reference to Fig. 4.1- ELECTRIC DIAGRAM, the 230 VAC power supply is routed through
an ON/OFF switch (with ELCB internal protection) to supply directly the Heating Immersion
Circulator and through an EMI Filter the PCB card, Fig. 4.2-PCB Schematic. The card sends
through a fuse the input power to a transformer which a 18 VAC output. The output goes to the
diode bridge rectifier D9 and capacitors C1, C2 for a 24 VDC output voltage which supplies the PC
board, the PLC, the two pressure switches, relais and actuators.
Fig. 4.3 details the PCB Layout and Fig. 4.4. the SHP cabling.
Digital Output from PLC are
OUT 1 (O1): 1=> ACTUATOR N°1 UP
OUT 2 (O2): 1=> ACTUATOR N°1 DOWN
OUT 3 (O3): 1=> ACTUATOR N° 2 UP
OUT 4 (O4): 1=> ACTUATOR N° 2 DOWN
OUT 5 (O5): 1=> ELECTROVALVE N° 1 CLOSED
OUT 6 (O6): 1=> ELECTROVALVE N° 2 CLOSED
Drives to the actuators are routed through relais K1, K2, K3, K4, with smoothing diodes D1, D2,
D3, D4 in parallel. O5 and O6 drive directly the valves, with smoothing diodes D5, D6 in parallel.
Digital Inputs to PLC are:
INP 1 (I1): 1=> DOOR N° 1 OPEN
INP 2 (I2): 0=> END COURSE LOW ACTUATOR N°1 ON
INP 3 (I3): 0=> END COURSE HIGH ACTUATOR N°1 ON
INP 4 (I4): 1=> DOOR N°2 OPEN
INP 5 (I5): 0=> END COURSE LOW ACTUATOR N°2 ON
INP 6 (I6): 0=> END COURSE HIGH ACTUATOR N°2 ON
INP 7 (I7): 1=> PRESSURE SWITCH N° 1 ON
INP 8 (I8): 1=> PRESSURE SWITCH N° 2 ON
INP 9 (I9): 0=> EMERGENCY PANEL N° 1
INP 10 (I10): 0=> EMERGENCY PANEL N° 2
Inputs from the proximity switches of the doors are passed through relais K5 and K6, with
smoothing diodes, before entering the PLC.
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
16
5. USE & OPERATION
To operate the S.H.P. device is needed to connect electric (230 V.A.C.) and water supply.
Switching-on power (the switch is on the left side of the device), the instrument, including
Circulator, is automatically activated. Before starting with test, wait some minutes for reaching
thermostat set temperature (programmable, nominal value = 96°C).
The annex flowcharts detail the way to operate the device and are mostly self explaining. Red
rectangles, written in corsive, represent displayed data. Black rectangles represent actions
performed by the device. Green rectangles represent actions to be performed by the operator.
At switch-on, the device raises the two water cooled columns up to high endcourse switches and
opens the two elettrovalves (columns connected to the atmosphere) =>RESET RUNNING –
WAITING EC’S. Then the display shows the ways we can operate:
<1>-->TEST.1……………
<2>-->TEST.2……………
<3>-->CALIBRATION….
<4>-->SERVICE………..
A) TEST
With reference to S.H.P. - Flowchart Test Operation, selecting <1> or <2>, one test is activated:
SELECT.DEVICE <1>.<2> The operator is requested: CHECK TEMPERATURE…<ANY> to
check on the thermostat the temperature (nominal value = 96 +/- 1 °C) and to confirm with a
pushbutton. The operator is then requested to open the bath door corresponding to the test in
operation: OPEN DOOR.1…. The operation is checked by a proximity sensor. Then displays:
<ANY>DIVE….
If confirmed by the operator, the column is dived to the endcourse switch low (sample in the
bath) DIVING and warm up timer starts counting down (starting from the warm up set value)
WARMUP…. ALT<4> (i.e. pressing <4> the operation stops). During this phase is also possible to
start another test in section 2: START.<2>.
At the end of warm up time, the electrovalve is closed (column isolated from the atmosphere) and
test timer starts (counting from the test set value). The display shows: 1.TEST….ALT<4>;
1.T:xxxx…HIT:yyy... In fact, during this time, when the pressure difference to the atmosphere
overcomes the value of 80 Pascal = 8 mm column of water (as measured by the differential pressure
device) the electrovalve opens to the atmosphere for the set EV time and then closes again. The
device displays the increased cumulated number of valve openings: HIT:yyy. Also during this phase
is possible to start another test or stop the running test.
At the end of test time, the electrovalve is opened and the column raised to endcourse switch high.
The final number of electrovalve openings is displayed and stored in computer archive.
The display shows also the computed short term stability: STAB = X.YY ml/min.
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
17
B) CALIBRATION
With reference to S.H.P. - Flowchart Calibration Operation, selecting <3> Calibration is
activated. The operator is then requested to select between instrument <1> or <2> . Then according
to paragraph Theory of Operation is requested to connect to the column of the selected device
and drip-feed through a pump in a separated box distilled water at a rate of 0.25 ml per minute.
During this phase, the display shows: CALIB.DEV.1.<ANY>; TIMER DROP:<xxxxx>. Through the
running of the timer, the operator can check that the drops of water are steady state at a rate of
approx. 1 bubble every 10 seconds or 6 bubbles per minute When the rate is estimated regular,
pressing a pushbutton, the operator closes the electrovalve and starts the calibration test. During this
phase the display shows: CALIB.DEV.1 RUNNING; 1.T:xxxx..HIT:yyy. No operator intervention is
requested.After 20 minutes (= Calibration Time = Test Time) are past, from the beginning of the
test, the PLC stops the calibration test. The electrovalve is opened and calibration results are
displayed and archived. The display shows: RESULT.CALIB.DEV.1; HIT: yy;<ANY>RETURN.
Pressing a pushbutton, we can return to the main menu’ for another operation.
C) SERVICE
With reference to S.H.P. - Flowchart Service Operation, selecting <4> Service Operations are
activated. The operator is requested to input a 4 digit key to enter in Service menu’.
If OK, the operator can:
1 - Change (or verify) device parameters, whose nominal values are:
- Test Time = Calibration Time (in seconds): nominal value 1200
- Warm Up Time (in seconds): nominal value 1200
- Electrovalve opening time (in tenths of millisecond): nominal value 50 (= 0.5 sec.)
2 - See and scroll archive data. The last 100 test (or calibration) data are stored. The archive reports:
- Archive position
- Date and time of test
- Device (1 or 2) used, related final number of Hits, computed short term Stability
- Calibration if the test was a calibration test. In this case device reference numbers are 10
or 20 instead of 1 or 2.
3 - Cancel all archive data.
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
18
D) OTHER OPERATIONS
According again to S.H.P. - Flowchart Service Operation, from the Main Menu it is possible to:
1- Stop the Program pressing the key STOP..
2- Update the P.L.C. stored programs switching off, and then simultaneously pressing SHIFT and
SWITCHING ON. This operation requires to connect to a Main Computer through a serial line.
New programs run automatically at the end of loading.
E) SAFETY
Two safety switches (EMERGENCY SWITCH) are located on top and bottom of each Instrument
Panel. They interrupt machine operations if the operator touchs these points or tries to put a finger
inside the machine slots. They have been inserted to avoid possible injury to careless personnel.
Two other magneto-thermic switches located in the left part of the device and in series with
emergency switches, interrupt heavy consumptions (in case of block) of the actuators (ACTUATOR
SAFETY SWITCH). In the two cases on the display appears SWITCH OFF POWER. To restart it is
requested to rearm magnetic switches, switch off and then switch on the machine.
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
19
6. THEORY OF OPERATION
The S.H.P. device is utilized to derive the short term stability of Hydrogen Peroxide solutions
through the measurement of the oxygen volume which is developed by the Hydrogen Peroxide
sample, kept at fixed temperature for a fixed time.
The procedure foresees the following steps:
a) a thermal fluid bath, filled with deionized water, kept at fixed Tb = 96 °C temperature;
b) a flask containing a volume (Vol) = 25 cc (+/- 0.1 cc) of Hydrogen Peroxide solution of known
title A (g/Kg). The corresponding weight (W) is :
W (g) = Vol (cc) * ( 1 + 0.4 * A / 1000 )
c) connect the 50cc flask to a cooling column, externally refrigerated by streaming water at
temperature (Tw);
d) deep the flask in the bath, keeping the cooling column connected to the atmosphere (electrovalve
open) for a fixed warm-up time of 20 minutes;
e) isolate the column, by closing the electrovalve for a fixed test time (tt) of 20 minutes. During this
time, according to the oxygen development, the inner column pressure increases over the
atmosphere. When the pressure difference overcomes the value of 80 Pascal = 8 mm column of
water (as measured by a differential pressure device) the valve opens to the atmosphere to
compensate for a fixed EV-open time of 500 milliseconds. The device keeps note of the number of
valve openings to derive the short term stability value.
f) After test time, let emerge the cooling colums from the oil bath and disconnect the flask from the
column.
g) Once per month perform a device calibration according to the following procedure.
Connect an empty flask to the cooling column and let the flash outside from the bath. Open the top
cover of the column and connect to a CALIBRATION BOX, which includes a small bottlet filled
with distilled water and a dosimetric pump. Connect and drip-feed the column from the top at a
nominal rate of 0.25 ml/min. When the rate is constant, start calibration from the S.H.P. device.
After 20 minutes (=CALIBRATION TIME = TEST TIME) are past calibration measurements are
stopped and stored. The stored value is the HITCAL reference value. The dosimetric pump can be
stopped and connection removed.
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
20
7. FORMULAS
Given:
A (g/Kg) : Title of the Hydrogen Peroxide solution;
V (cc) : Volume of gas developed during test time;
tt (sec) : Test time;
tref = 60 sec : Reference time;
Tw (°C) : Temperature of cooling water;
W (g) : Weight of sample under test = Vol (cc) * (1 + 0.4 * A / 1000 );
Wref = 1000 gr : Reference weight;
M = 34 g = 1 mole of Hydrogen Peroxide;
Vref = 22400 cc = volume occupied by one mole at standard conditions;
HIT : number of pressure switches during the test;
HITREF : number of pressure switches during calibration ( VOLREF=5 cc);
P0, Pt (Pa) : Pressure at standard and test conditions;
T0 (=273°K), Tw (°C) : Temperature at standard and test conditions;
we have:
VOLUME OF GAS DEVELOPED DURING TEST: Vt (cc):
(1) Vt (cc) = 5 * HIT / HITREF
VOLUME OF GAS REFERRED TO STD. CONDITIONS: Vtref (cc):
(2) Vtref (cc) = Vt * (Pt/P0) * (T0/Tw)
MASS OF DECOMPOSED HYDROGEN PEROXIDE DURING TEST (STD): Mdg (g):
(3) Mdg (g) = Vtref * 2 * M / 22400
( 1 mole of oxygen derives from two hydrogen peroxide moles)
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
21
DECOMPOSED MASS REFERRED TO 1000 GRAMS OF HYGROGEN PEROXIDE
SOLUTION: Mdg1000 (g):
(4) Mdg1000 (g) = Mdg * 1000 / W
DECOMPOSED MASS (4) REFERRED TO 1000 GRAMS DIVIDED SOLUTION TITLE (A) in
terms of gr/Kg:Mt
(5) Mt (g) = Mdg1000 / A
The H2O2 Stability meter computes and stores volume decomposition rate (dr) per minute during
the test:
STAB = VOLUME DECOMPOSED (ml/min) is given by:
STAB = ( VOLREF*HIT / HITREF ) * (Tref/ tt) = (300 * HIT) / (HITREF * tt)
Another interesting formula, assuming Pt=P0; T0=273; Tw=288;T0/Tw=0.948, is (6), referred to:
- 1 hour (decomposed mass per hour ref. to a 1000 grams solution divided solution title (A),
or:
Mt = 0.948*(68/22400)*(1000/W)*60*STAB/A
For low title solutions:
Mt = 6.9* STAB/A
(developed H2O2 mass(g) per hour, per title, per 1000 gr of low title solution)
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
22
8. ASSEMBLY INSTRUCTIONS
UNPACK
Unpack the equipment and accessories and check for damages incurred during transit. These
should be reported to the responsible carrier and a request for a damage report should be made
.These instructions must be followed fully for us to guarantee our full support of your claim for
protecting against loss from concealed damage.
For assembly, reference is made to Fig. 8.1. - Assembly, Fig. 8.2 - Cooling column, Fig. 8.3. -
Assembly (Rear).
POSITIONNING
Place the SHP equipment on a stable, flat workbench. If necessary use the four foots (1) in the
bottom for levelling.
ASSEMBLY INSTRUMENT PLATE
This operation is requested in case the Cooling Column is shipped separately from the Instrument
Plate.
Assembly one instrument plate per time. For this operation first hand unscrew the two pommels
(2) and remove the plexiglass cover. Unscrew and remove the two Column Support Closures (3).
Place in position the Cooling Column, handling with care and keeping it in the upright position
(black plug high). Connect (B) and (A) of the Cooling Column to the rubber tubes exiting from the
instrument plate (IN and OUT of water entry). Mount again the two Column Support Closures (3)
keeping the water tubes horizontal and tightening slightly in order to avoid rupture of the Cooling
Column. Connect the rubber tube exiting from the Electrovalve (4) to (C) of the Cooling Column.
Connect the rubber tube exiting from the Pressure Switch-HI (5) to (D) of the Cooling Column.
Install again the plexiglass cover and hand screw the two pommels (2).
Repeat the operation for the next instrument plate.
WATER CONNECTION
Connect the plant water entry (requirements as per para 2.3) to point (6) - IN of SHP instrument
through a flexible rubber tube (inner diameter = 10 mm). Tighten with a fastening clip.
Connect the deionized water entry through a flexible rubber tube to point (11) - IN. Adjust the tap
for minimum flow.
Connect point (7) and (12) - OUT of SHP equipment to plant water disposal through a similar
flexible rubber tube.
Remove the SHP rear cover (unscrewing the 14 self-tapping screw).
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
23
Let flow water and verify for no water leakage throughout all the water circuitry.
Reinstall the rear cover.
Water can flow permanently.
THERMAL FLUID FILLING
Open the two doors (8) of the bath.
Fill the bath through one door with deionized water. Visually check through the other door for
filling up to approximately 1 inch from the upper edge. 11 to12 liters of fluid are required.
Adjust the Level Control Tube (10) to get the desired level value.
ELECTRIC SUPPLY
Connect a suitable plug to the cable (9) exiting from the SHP equipment - left side. Blu and
Brown wires are line wires. Yellow-green wire is the earth wire. Connect to a socket for 230 V A.C.
supply (requirements as per para 2.1.).
OPERATION
Operating instruction are given in Chapter 5.
For performing a test, the operator has to wait until the thermostat has reached the set temperature
(Appendix 1 and the JULABO Heating Immersion Circulator Manual gives all the instructions to
change the set temperature), then need to connect a 50cc flask, containing the 25 cc Hydrogen
Peroxide sample solution of known title to the cooling column of one of the instruments and fasten
with a clip. Then we need to operate according to PLC instructions.
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
24
9. MAINTENANCE INSTRUCTIONS
The S.H.P. equipment is designed for continuous operation under normal conditions. Periodic
maintenance in not required.
For cleaning the outside of the unit use a wet cloth.
RECOMMENDATIONS
- Connect the unit through the plug to a grounded mains power socket.
- Note the flash point of the bath medium used. The excess temperature protection should be set at
least 10°C below the flash point.
- Pay attention to the thermal expansion of bath oil during heating to avoid overflowing of the
liquid. If necessary, lower the Level Control Tube (10).
- Regularly check the tubing for material defects (e.g. for cracks) and the workbench below the
instrument to verify bath tightness.
- Before cleanig the unit, disconnect the power from the mains socket.
- Empty the bath before moving the unit.
REPAIRS
For technical assistance report to the manufacter:
GAMBA & BOTTEGHI snc
Via SPAGNA, 38
57017 GUASTICCE (LI)
ITALY
www.gambaebotteghi.com
Tf. +39 0586 983060
Fax +39 0586 983163
e-mail: [email protected]
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
25
10. EC DECLARATION OF CONFORMITY
The undersigned, representing the following manufacturer
GAMBA & BOTTEGHI snc
Via GALILEI, 381
57019 VICARELLO (LI)
herewith declares the the product
S.H.P.
STABILITY METER FOR HYDROGEN PEROXIDE
is in conformity with the provisions of the following EC directives (including all applicable
amendments)
89/336 “Electromagnetic Compatibility”
73/23 “Low voltage directive”
and that the standards and/or technical specifications referenced overleaf have been applied.
Last two digits of the year in which the CE marking was affixed: 98
VICARELLO (LI) 20 Feb. 1998
............................................................. ..................................
(place) (date)
..............................................................
(signature)
MARIO GAMBA - Legal Representative
.................................................................................
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
26
REFERENCES
References of standards and/or technical specifications applied for this declaration of conformity,
or parts thereof:
EN 50081-1 Issue 1 EMC - Generic emission standard - Part 1: Residential, commercial and
light industry
EN 50082-1 Issue 1 EMC - Generic immunity standard - Part 1: Residential, commercial and
light industry
EN 61010-1 Safety requirements for electrical equipment for measurement, control and
laboratory use - Part 1: General requirements
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
27
PROVISIONAL SPARES LIST
P/N NAME QUANTITY
402-02-1 Heating Immersion Circulator TBD
402-02-2 Proximity Switch TBD
403-02-4 Cooling Column TBD
402-02-5 Flask TBD
403-02-6 Pressure Switch TBD
403-02-7 Electrovalve TBD
403-02-14 Plexiglass Cover TBD
403-02-16 Emergency Micro Switch TBD
404-02-4 Electric Actuator TBD
404-02-5 End Course Sensor TBD
405-02-1 PLC TBD
405-02-3 Transformer TBD
405-02-4 Printed Circuit Board TBD
402-02-3 Bath TBD
402-02-4 Bath Cover TBD
401-02-20 Actuator Safety Switch TBD
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
28
Appendix 1
HEATING IMMERSION CIRCULATOR
The heating immersion circulator temperature is factory set at 96°C.
The procedures to switch ON-OFF the circulator and to change the set temperature are described in
the following pages.
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
29
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
30
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
31
Appendix 2
DIFFERENTIAL PRESSURE INDICATOR
Model: DPI 2500-2R-AZ –D : F.S..+/-250 Pa; 2 Rel’ output NC; Self zeroing function: 3” every 10 minutes Rele switch output (NCNO) set at 80Pa = 8 mm H2O = 0.8 mBar. Functions can be set and/or checked scrolling the MENU through the pusbuttons: MENU (to scroll), BACK (return back), OK (confirm), DOWN, UP (modify) and the DISPLAY (corsive) in the following way.
- Press MENU more than 3 seconds; - Unit Pa (Pascal)
(to change setting, press OK, arrows up/down, OK to confirm. To change [Unit to Psel] press the Arrow Down)
- Psel (Full Scale) 250 (Pascal) - rESP 0.5 sec (response time) - rel1 (Relè n.1)
o dirrise (switch with increasing pressure) o S.P.--> 80 Pa ..Set (switch) Point o Hys 2 (Hysteresis = 2 Pascal)
- rel2 ….. idem - zero point calibration: not requested - span point calibration: not requested
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
32
FIGURES
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
33
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
34
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
35
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
36
Fig. 3.3 CALIBRATION BOX
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
37
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
38
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
39
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
40
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
41
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
42
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
43
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
44
www.gambaebotteghi.com
57017 GUASTICCE (LI) – Via Spagna,38 – Italy - Tel.: 0586/983060; Fax:0586/983163
e-mail: [email protected] - Cell: 335/7890511-512 – P.IVA: 01006040495
45
Fig. 8.3. ASSEMBLY (Rear)