Aplio 400 Operator Manual

Embed Size (px)

Citation preview

  • 7/23/2019 Aplio 400 Operator Manual

    1/441

    No. 2B771-044EN*D

    IMPORTANT!

    Read and understand this manual before operatingthe equipment. After reading, keep this manual in aneasily accessible place.

    TOSHIBA MEDICAL SYSTEMS CORPORATION 2011-2012

    ALL RIGHTS RESERVED

    OPERATION MANUALFOR

    DIAGNOSTIC ULTRASOUND SYSTEM

    MODEL TUS-A400

    MEASUREMENTS VOLUME

    (2B771-044EN*D)

  • 7/23/2019 Aplio 400 Operator Manual

    2/441

    No. 2B771-044EN*D

    Trademarks

    Microsoft Windows XP is a trademark of Microsoft Corporation in the USA and other

    countries.

    APLIO, Dynamic Flow, ApliPure, MicroPure, and TwinView are trademarks of Toshiba MedicalSystems Corporation.This manual may include trademarks or registered trademarks of other companies.

    Note that the trademark symbol "" and the registered trademark symbol "" may or may notbe used in this manual.

    IMPORTANT!

    1. No part of this manual may be copied or reprinted, in whole or in part,without prior written permission.

    2. The contents of this manual are subject to change without prior noticeand without legal obligation.

    3. The contents of this manual are correct to the best of our knowledge.Please inform us of any ambiguous or erroneous descriptions, missinginformation, etc.

  • 7/23/2019 Aplio 400 Operator Manual

    3/441

    No. 2B771-044EN*D

    - a -

    Table of Contents

    1. Outline of Measurement Functions.................... 1-1

    1.1 Switches Used for Measurement................................................. 1-1

    1.2 Setting the Measurement................................................................... 1-2

    1.3 Modifying a Measurement................................................................. 1-3

    1.4 Deleting a Measurement..................................................................... 1-6

    1.4.1 Deleting a measurement using the touch panel............................... 1-6

    1.4.2 Deleting the selected measurement result......................................... 1-7

    1.5 Changing the Result Display Position..................................... 1-8

    1.6 Worksheet Screen................................................................................... 1-9

    2. 2D-Mode Measurements.................................................... 2-1

    2.1 Distance Measurement (Distance).............................................. 2-2

    2.2 Curved Line Distance Measurement (Trace Length)..... 2-3

    2.2.1 Continuous Trace.......................................................................................... 2-3

    2.2.2 Spline Trace..................................................................................................... 2-4

    2.3 Angle Measurement (Angle)............................................................ 2-5

    2.4

    Area/Circumference Measurement (Area)............................. 2-6

    2.4.1 Continuous Trace.......................................................................................... 2-6

    2.4.2 Spline Trace..................................................................................................... 2-6

    2.4.3 Ellipse................................................................................................................. 2-7

    2.4.4 Cross................................................................................................................... 2-8

    2.5 Volume Measurement (Volume).................................................. 2-10

    2.6 Joint Measurement............................................................................... 2-11

  • 7/23/2019 Aplio 400 Operator Manual

    4/441

    No. 2B771-044EN*D

    - b -

    2.7 MeanIMT Measurement..................................................................... 2-12

    2.8 %Stenosis (Distance) Measurement....................................... 2-14

    2.9 %Stenosis (Area) Measurement................................................. 2-15

    3. M-mode Measurements....................................................... 3-1

    3.1 M-Mode Distance Measurement (Distance)......................... 3-2

    3.2 Slope Measurement (Slope)............................................................. 3-3

    3.3 Time Measurement (Time)................................................................. 3-4

    3.4 Heart Rate Measurement (HR)........................................................ 3-5

    4. Doppler-Mode Measurements.................................... 4-1

    4.1 Velocity Measurement (Velocity)................................................. 4-2

    4.2 Acceleration Measurement (Accel)............................................ 4-3

    4.3 Time Measurement (Time)................................................................. 4-3

    4.4 Heart Rate Measurement (HR)........................................................ 4-3

    4.5 Velocity Trace Measurement (Vel Trace)............................... 4-4

    4.5.1 Continuous Trace.......................................................................................... 4-5

    4.5.2 Spline Trace..................................................................................................... 4-7

    4.5.3 Line Trace......................................................................................................... 4-8

    4.5.4 Range.................................................................................................................. 4-9

    4.5.5 Auto Range..................................................................................................... 4-11

    4.6 PI Measurement (PI Trace).............................................................. 4-12

    4.7 RI Measurement...................................................................................... 4-13

    4.8 S/D Measurement................................................................................... 4-14

    4.9 Flow Volume Trace Measurement (Flow Volume)......... 4-15

  • 7/23/2019 Aplio 400 Operator Manual

    5/441

    No. 2B771-044EN*D

    - c -

    5. Application Measurement (Cardiac).................. 5-1

    5.1 Introduction.................................................................................................. 5-1

    5.1.1 Starting the application measurement menu..................................... 5-1

    5.1.2 Heart rate acquisition methods (Cardiac)........................................... 5-1

    5.2 2D-Mode Measurements (Cardiac)............................................. 5-3

    5.2.1 LV measurement (MOD Simpson method)/LA measurement (MOD Simpson method)/LA measurement (3-Axis method).......................................................... 5-3

    5.2.2 LV measurement (Teichholz method)................................................. 5-11

    5.2.3 LV measurement (Cube method).......................................................... 5-16

    5.2.4 LV measurement (Gibson method)...................................................... 5-16

    5.2.5 LV measurement (Single-plane (aka Area-Length) method)..... 5-17

    5.2.6 LV measurement (Biplane method)..................................................... 5-19

    5.2.7 LV measurement (Bullet method)......................................................... 5-21

    5.3 M-Mode Measurement (Cardiac)................................................ 5-23

    5.3.1 Aortic valve measurement (Aortic Valve).......................................... 5-23

    5.3.2 Mitral valve measurement (Mitral Valve)............................................ 5-25

    5.3.3 LV measurement (Teichholz method)................................................. 5-27

    5.3.4 LV measurement (Cube method).......................................................... 5-33

    5.3.5 LV measurement (Gibson method)...................................................... 5-33

    5.4 Doppler-Mode Measurement (Cardiac).................................. 5-34

    5.4.1 Aortic valve measurement (Aortic)...................................................... 5-34

    5.4.2 Mitral valve measurement (Mitral)........................................................ 5-42

    5.4.3 Pulmonary vein blood flow measurement (Pulmo. Vein)........... 5-49

    5.4.4 Tricuspid valve measurement (Tricus.).............................................. 5-51

    5.4.5 Pulmonary valve measurement (Pulmo.)........................................... 5-55

    5.5 Extra Measurements............................................................................ 5-60

  • 7/23/2019 Aplio 400 Operator Manual

    6/441

    No. 2B771-044EN*D

    - d -

    5.5.1 LV Mass AL (Area-Length)....................................................................... 5-60

    5.5.2 LV Mass TE (Truncated Ellipsoid)........................................................ 5-62

    5.5.3 PISA................................................................................................................... 5-64

    5.5.4 Coronary.......................................................................................................... 5-66

    6. Application Measurement (Vascular).............. 6-1

    6.1 Carotid1........................................................................................................... 6-1

    6.1.1 2D-mode measurement............................................................................... 6-1

    6.1.2 Doppler-mode measurement.................................................................... 6-3

    6.1.3 Measurement value display on the Worksheet screen.................. 6-4

    6.2 Carotid1 (Auto-IMT Measurement).............................................. 6-5

    6.2.1 Touch panel menu......................................................................................... 6-5

    6.2.2 Measurement method.................................................................................. 6-7

    6.2.3 Worksheet screen.......................................................................................... 6-9

    7. Obstetric Measurement (OB)....................................... 7-1

    7.1 Outline of OB Measurement............................................................ 7-1

    7.2 Patient Registration for OB Measurement............................ 7-2

    7.3 OB Measurement Menu on the Touch Panel....................... 7-3

    7.4 Measurement Methods Specific toOB Measurements................................................................................... 7-5

    7.4.1 Amniotic fluid index (AFI).......................................................................... 7-5

    7.4.2 NB (Nasal Bone)/NT (Nuchal Translucency)/NF (Nuchal Fold)measurement................................................................................................... 7-6

    7.4.3 CTAR measurement...................................................................................... 7-7

    7.4.4 Maxilla Angle (Facial Maxilla Angle) measurement........................ 7-9

    7.4.5 DV (Ductus Venosus) measurement..................................................... 7-9

    7.5 Calculation Items................................................................................... 7-10

  • 7/23/2019 Aplio 400 Operator Manual

    7/441

    No. 2B771-044EN*D

    - e -

    7.6 List of OB Measurements andOB Calculation Items.......................................................................... 7-14

    7.7 Measurement Items That Can Be Selected in theGyn. Menu................................................................................................... 7-22

    7.8 Updating the OB Date Information During anExamination (GA Input)..................................................................... 7-23

    7.9 Trend Graph (Trend)............................................................................ 7-24

    7.10 OB Worksheet.......................................................................................... 7-26

    7.10.1 OB 2D Worksheet........................................................................................ 7-27

    7.10.2

    OB Doppler Worksheet.............................................................................. 7-28

    7.10.3 Gyn Worksheet............................................................................................. 7-29

    7.10.4 Comment screen.......................................................................................... 7-30

    7.10.5 Anatomy function........................................................................................ 7-32

    7.10.6 Data List........................................................................................................... 7-34

    7.11 Data Storage.............................................................................................. 7-39

    8. Measurements on a Stored Image....................... 8-1

    8.1 Measurements on ER Images......................................................... 8-3

    8.2 Measurements on ER Images for Which ManualCalibration Is Required........................................................................ 8-5

    8.3 Manual Calibration Required Before the Start ofEach Measurement............................................................................... 8-10

    8.4 Worksheet Display................................................................................ 8-11

    8.5 Accuracy of Measurement When Manual CalibrationIs Used........................................................................................................... 8-12

    9. Measurement Data Storage........................................... 9-1

    9.1 Storing Measurement Data in the DICOM SR Format... 9-1

    9.1.1 Data transfer.................................................................................................... 9-1

  • 7/23/2019 Aplio 400 Operator Manual

    8/441

    No. 2B771-044EN*D

    - f -

    9.1.2 Displaying the stored data......................................................................... 9-3

    10. OB Measurement Chart Data..................................... 10-1

    10.1 Gestational Age Chart...................................................................... 10-11

    10.2 Growth Analysis Chart.................................................................... 10-68

    10.3 Identical Chart...................................................................................... 10-129

    10.4 Charts for Creating Trend Graphs........................................ 10-178

    11. Presetting............................................................................................. 11-1

    11.1 Setting Presets........................................................................................ 11-1

    11.2 Details of Each Preset........................................................................ 11-3

    11.2.1 Appl Meas....................................................................................................... 11-3

    11.2.2 Meas. Window............................................................................................... 11-4

    11.2.3 2D-Mode Meas.............................................................................................. 11-6

    11.2.4

    D-Mode Meas................................................................................................. 11-7

    11.2.5 D-Mode Meas. 2.......................................................................................... 11-13

    11.2.6 Worksheet..................................................................................................... 11-14

    11.2.7 Meas. Marker................................................................................................ 11-16

    11.2.8 Meas. Marker2............................................................................................. 11-17

    11.2.9 Meas. Operation......................................................................................... 11-18

    11.2.10 Annotation.................................................................................................... 11-20

    11.2.11 LV (2D)............................................................................................................ 11-21

    11.2.12 LV (M).............................................................................................................. 11-22

    11.2.13 Cardiac Doppler.......................................................................................... 11-23

    11.2.14 LV Mass.......................................................................................................... 11-24

    11.2.15 Carotid1 Doppler........................................................................................ 11-25

    11.2.16 Carotid1 Meas. Type................................................................................. 11-27

  • 7/23/2019 Aplio 400 Operator Manual

    9/441

    No. 2B771-044EN*D

    - g -

    11.2.17 Carotid1 Meas. Tool.................................................................................. 11-28

    11.2.18 Auto IMT......................................................................................................... 11-29

    11.2.19 Body Mark..................................................................................................... 11-30

    11.2.20 Meas. Data Display.................................................................................... 11-31

    11.2.21 OB/Gyn Report............................................................................................ 11-32

    11.2.22 OB/Gyn Report 2........................................................................................ 11-33

    11.2.23 OB Evaluated - OB Calc.......................................................................... 11-34

    11.2.24 OB Evaluated - OB Calc. 2..................................................................... 11-35

    11.2.25 OB Evaluated - OB Calc. 3..................................................................... 11-37

    11.2.26 OB Evaluated - D OB Calc...................................................................... 11-39

    11.2.27 OB Display.................................................................................................... 11-40

    11.2.28 OB Behavior................................................................................................. 11-41

    11.2.29 OB 2D Tool................................................................................................... 11-42

    11.2.30 OB/Gyn D Meas.......................................................................................... 11-43

    11.2.31

    Gyn Meas...................................................................................................... 11-44

    11.3 Confirmation of the Automatic MeasurementStart Item.................................................................................................... 11-45

    12. Registration of User-ProgrammedMeasurements............................................................................... 12-1

    *

  • 7/23/2019 Aplio 400 Operator Manual

    10/441

    No. 2B771-044EN*D

    1-1

    1. Outline of Measurement Functions

    1.1 Switches Used for Measurement

    No. Switch Function

    Moves the marker.

    The trace lines can be modified using the dial located aroundthe trackball.

    Sets the start and end points.

    Starts the next measurement.

    Starts a basic measurement. (The default measurement that isstarted differs according to the mode.)

    2D mode : Distance

    M mode : Distance

    PW/CW mode : Velocity

    Starts a basic measurement. (The default measurement that isstarted differs according to the mode and the customizedsetting.)

    2D mode : Area

    M mode : Slope

    PW/CW mode : Vel Trace

    Starts an application measurement. (The measurement to bestarted must be preset in advance.)

    Modifies trace lines.

    Rotates the line marker to adjust the angle.

    Freezes the image.

    Displays the application measurement results on the Worksheetscreen. (The screen to be displayed differs depending on thesetting.)

  • 7/23/2019 Aplio 400 Operator Manual

    11/441

    No. 2B771-044EN*D

    1-2

    NOTE: 1. The measurements that start when the corresponding switch is pressedand the methods for setting the measurement data differ depending on thepreset.The data used for calculation and the display of calculation results alsodiffer depending on the preset.

    Refer to section 11 "Presetting" and set the desired presets before usingthe above switches.

    2. The default basic measurement can be changed as needed. Contact yourToshiba service representative.

    1.2 Setting the Measurement

    When the measurement is set on an image in the current examination, the color of themeasurement markers is changed to cyan.

    When the measurement is set on an image stored for a previous examination, the colorof the measurement markers is changed to white.

    CAUTION: Be sure to perform measurement within the displayed images.If an area outside an image is included in a measurement, anincorrect result may be output.

    NOTE: 1. Both and can be used to set (register) the measurement.

    When is used, the measurement result is set and the next

    measurement is started.

    However, when "Fixed by [SET] only" in [Vel, Auto Range Tool] of theMeas. Operation preset is set to ON, the preset setting takes precedence

    and must be used to set the measurement result.

    Note that pressing a measurement item switch on the touch panel (otherthan Velocity measurement of the basic measurements) also sets themeasurement result.

    2. The application measurement results are not displayed on the Worksheetscreen if the measurement data has not been set.

    3. In basic measurement mode, the measured data is lost when the systempower is turned OFF or the exam data has been changed.

    4. The results of basic measurements are not displayed on the Worksheetscreen.

    5. Only a part of the calculation results are displayed on the measurementresult screen in application measurement mode. To check all themeasurement results (calculation results), use the Worksheet screen.

    6. It is possible to preset whether or not the line connecting the start point and

    end point is displayed before the measurement is set in distancemeasurement and area measurement.

  • 7/23/2019 Aplio 400 Operator Manual

    12/441

    No. 2B771-044EN*D

    1-3

    1.3 Modifying a Measurement

    A measurement data item can be modified by pressing even after the

    measurement data is set and the measured value is displayed, provided that or

    another measurement switch has not been pressed.

    Perform the procedures below to modify the measurement data.

    Step 1:

    Start modification.

    Step 2:

    Move the startpoint.

    Step 3:

    Finish modificationof the start point.

    Step 4:

    Modify the endpoint.

    Press .

    The start point(first point) changes toa marker.

    Change theposition of the

    start point using

    .

    A

    Press .

    A

    The newly measureddata is set.

    To modify the endpoint, press

    and follow

    the sameprocedure.

  • 7/23/2019 Aplio 400 Operator Manual

    13/441

    No. 2B771-044EN*D

    1-4

    Step 1:

    Start modification.

    Step 2:

    Undo the trace tothe desired point.

    Step 3:

    Draw the traceagain.

    Step 4:

    Set the newlymeasured data.

    Press .

    The start point (firstpoint) changes to amarker.

    Rotate to

    undo the trace tothe desired point.

    From that point,draw the traceagain using

    .

    Press .

    The newly measureddata is set.

    To modify the endpoint, press

    and follow

    the sameprocedure.

    * During velocity trace measurement in PW/CW mode, rotate counterclockwise

    to undo the trace.

  • 7/23/2019 Aplio 400 Operator Manual

    14/441

    No. 2B771-044EN*D

    1-5

    The above procedures cannot be used to modify measurements obtained using theSpline Trace method.

    To modify such measurements, press and then move the marker.

    Step 1:

    Start modification.

    Step 2:

    Select the markerto be modified.

    Step 3:

    Move the marker.

    Step 4:

    Set the newlymeasured data.

    Press .

    An arrow is displayed.

    Move the arrownear the marker tobe modified.

    The color of theselected markerchanges to red.

    Press .

    Move the markerto the desiredposition using

    .

    Press .

    The newly measureddata is set.

  • 7/23/2019 Aplio 400 Operator Manual

    15/441

    No. 2B771-044EN*D

    1-6

    1.4 Deleting a Measurement

    1.4.1 Deleting a measurement using the touch panel

    *1: Refer to subsection 7.3 for the function of [Clear All] in OB measurement mode.

  • 7/23/2019 Aplio 400 Operator Manual

    16/441

    No. 2B771-044EN*D

    1-7

    1.4.2 Deleting the selected measurement result

    Perform the procedures below to delete the selected measurement result.

    Step 1:

    Start

    Step 2:

    Select the measurementresult to be deleted.

    Step 3:

    Delete themeasurement result.

    Press .

    An arrow is displayed.

    Move the arrow nearthe marker for themeasurement result tobe deleted.

    A B

    The color of the selectedmarker changes to red.

    Press .

    The measurement resultassociated with theselected marker is deleted.

  • 7/23/2019 Aplio 400 Operator Manual

    17/441

    No. 2B771-044EN*D

    1-8

    1.5 Changing the Result Display Position

    is displayed during measurement. Press this switch to display the menu for

    changing the measurement result display position.

    *1: Each time is pressed, the result display position is changed as

    shown below.

  • 7/23/2019 Aplio 400 Operator Manual

    18/441

    No. 2B771-044EN*D

    1-9

    1.6 Worksheet Screen

    After performing application measurement, press to display the Worksheet

    screen*1.

    *1: If the system is set not to start up the Worksheet screen by pressing , press

    on page 3/3 of the OTHER menu.

    On this screen, the measurement and calculation results obtained in the applicationmeasurement are displayed.

    CAUTION: 1. If the measurement data displayed on the Worksheet screen is edited,the data that precedes editing is overwritten and lost.

    2. When is pressed, the message "Calculating" is displayed until the

    Worksheet screen is displayed. If many items have been measured,"Calculating" may remain displayed for several tens of seconds toseveral minutes. Do not operate the main panel or touch panel whilethis message is displayed.

    3. When more than one measurement has been performed using differentapplication presets, the application preset for which the results are to bedisplayed can be selected from the pull-down menu on the Worksheetscreen. If the number of application presets used is large, there may bea delay of several tens of seconds to several minutes before the resultsfor the selected application preset are displayed. Do not operate themain panel or touch panel during this period.

    NOTE: 1. If "Mean of Measurements" is selected for the "Measurement Value DisplayMethod" on the Worksheet tab of the Preset menu, the mean of themeasurements is displayed in the worksheet.If only one measurement has been executed, the result of themeasurement is displayed in the Mean field.

    2. If "Most Recent Measurement" is selected for "Measurement Value DisplayMethod" on the Worksheet tab of the Preset menu, only the most recentmeasurement value is displayed on the Worksheet screen.

    3. If two or more measurements have been performed for a single item, besure to check each measurement result using the worksheet function.If an unexpected measurement value is obtained, it will affect the meanvalue.

    4. If a result for a measurement item is deleted in the worksheet edit windowwith "Most Recent" selected, the result of the immediately precedingmeasurement is displayed.

  • 7/23/2019 Aplio 400 Operator Manual

    19/441

    No. 2B771-044EN*D

    1-10

    NOTE: 5. If a result for a measurement item is deleted in the worksheet edit windowand no other results remain for the item, the measurement item, the meanvalue of the item, and the calculation results obtained using themeasurement value (or mean value) are all deleted automatically.To redisplay these values, the item must be measured again.

    Before deleting the contents of the worksheet edit window, confirm thatrequired data will not be deleted.When entering or editing data for measurement items, be sure to enter thevalues and units correctly.

    6. Measurement or calculation data cannot be saved unless a patient ID isregistered for the patient.Be sure to register the patient ID before starting measurement.

    7. Confirm that no unexpected values are displayed on the Worksheet screenbefore saving the data.

    8. The measured value for some items displayed on the Worksheet screencan be edited, provided that the examination has not yet been completed.Once the measured data is saved, however, the data cannot be edited.Confirm the results on the worksheet before saving the data.The values displayed in white in the blue boxes can be edited.

    9. Do not turn ON the measurement switch while the Worksheet screen isdisplayed.

    10. Only a part of the calculation results are displayed on the measurementresult screen in application measurement mode. To check all themeasurement results (calculation results), use the Worksheet screen.

    *

  • 7/23/2019 Aplio 400 Operator Manual

    20/441

    No. 2B771-044EN*D

    2-1

    2. 2D-Mode MeasurementsThis section describes basic measurements in 2D mode.

    * The Joint measurement switch is not displayed when Heart1, Heart2, or Heart3 isselected as Application Preset.

  • 7/23/2019 Aplio 400 Operator Manual

    21/441

    No. 2B771-044EN*D

    2-2

    2.1 Distance Measurement (Distance)

    The straight-line distance between two points is measured on the 2D-mode image.

    Distance : Dist A 999.9 (mm)

    Step 1:

    Set the start point.

    Step 2:

    Move the end point.

    Step 3:

    Set the end point.

    Select .

    Move the marker to themeasurement start

    point using .

    Press to set the

    start point.

    Move the marker to themeasurement end point

    using .

    Press to set the

    end point.

    A

    * To perform the same type of measurement again, press and then repeat

    steps to .

  • 7/23/2019 Aplio 400 Operator Manual

    22/441

    No. 2B771-044EN*D

    2-3

    2.2 Curved Line Distance Measurement (Trace Length)

    The length of a curved line is measured by tracing along the curved line.

    The trace method can be changed using the dial provided under .

    Method Description

    Continuous Trace The user traces a curved line freehand from the specifiedstart point to the end point using the trackball.

    Spline Trace A curved line is traced by interpolating three or more user-specified points (including the start and end points) using aspline curve.

    Distance : Dist A 999.9 (mm)

    2.2.1 Continuous Trace

    Step 1:

    Set the start point.

    Step 2:

    Draw the trace.

    Step 3:

    Set the trace.

    Press and

    select [ContinuousTrace].

    Move the marker using

    .

    Press .

    Draw the trace using

    . Press .

    A

    During tracing, the trace can be modified by rotating or .

    The trace is deleted.

    The deleted portion ofthe trace is restored.

  • 7/23/2019 Aplio 400 Operator Manual

    23/441

    No. 2B771-044EN*D

    2-4

    2.2.2 Spline Trace

    Step 1:

    Set the start point.

    Step 2:

    Draw the trace.

    Step 3:

    Set the trace.

    Press and

    select [Spline Trace].

    Move the marker using

    .

    Press .

    Move the marker using

    .

    Press .

    Repeat the above

    procedure to setmultiple points.

    Press twice at

    the same point.

    A

    During tracing, the trace can be modified by rotating or .

    The trace is deleted.

    The deleted portion ofthe trace is restored.

  • 7/23/2019 Aplio 400 Operator Manual

    24/441

    No. 2B771-044EN*D

    2-5

    2.3 Angle Measurement (Angle)

    The angle between two straight lines is measured. Both the acute and obtuse anglesare displayed.

    Acute : Angle2 A 99 (deg)

    Obtuse : Angle1 A 999 (deg)

    Step 1:

    Set the first marker.

    Step 2:

    Move the secondmarker.

    Step 3:

    Set the secondmarker.

    Select .

    Move the marker using

    .

    Rotate the marker

    using to adjust

    the angle.

    Press .

    Move the second

    marker using .

    A

    Rotate the marker

    using .

    Press .

    A

    A

  • 7/23/2019 Aplio 400 Operator Manual

    25/441

    No. 2B771-044EN*D

    2-6

    2.4 Area/Circumference Measurement (Area)

    The area and circumference of a region are measured in 2D mode.

    The trace method can be changed using the dial provided under .

    Measurement method Area measurement method

    Continuous Trace The path traced by operating the trackball along thecircumference of the target object is used to calculate theapproximate area/circumference.

    Spline Trace The interpolated curve for the set points is used to calculatethe approximate area/circumference.

    Ellipse Elliptical approximation is used to calculate the approximatearea/circumference.

    Cross Arcs are used to calculate the approximatearea/circumference.

    2.4.1 Continuous Trace

    For the measurement procedure, refer to subsection 2.2 "Curved Line DistanceMeasurement (Trace Length)".

    Area : Area A 999.99 (cm2) Circumference : Circ A 999.9 (mm)

    2.4.2 Spline Trace

    For the measurement procedure, refer to subsection 2.2 "Curved Line DistanceMeasurement (Trace Length)".

    Area : Area A 999.99 (cm2) Circumference : Circ A 999.9 (mm)

  • 7/23/2019 Aplio 400 Operator Manual

    26/441

    No. 2B771-044EN*D

    2-7

    2.4.3 Ellipse

    Area : Area A 999.99 (cm2) Long-axis length : Dist1 A 999.9 (mm)

    Circumference : Circ A 999.9 (mm) Short-axis length : Dist2 A 999.9 (mm)

    Step 1:

    Set the start point.

    Step 2:

    Set the diameterof the circle.

    Step 3:

    Set the ellipse.

    Step 4:

    Set the trace.

    Press

    and select[Ellipse].

    Move the marker

    using .

    Press .

    Adjust thediameter of thecircle using

    .

    Press to

    set the diameter ofthe circle.

    Adjust the size ofthe ellipse using

    .

    Press .

    A

    Step 1:

    Start modification.

    Step 2:

    Select the markerto be modified.

    Step 3:

    Modify the ellipse.

    Step 4:

    Finishmodification.

    Press .

    A

    Move the arrownear the ellipse,place the arrow onthe point to bemodified, and

    press .

    (When the arrowis moved to theellipse, the majoraxis and minoraxis of the ellipseare displayed.)

    A

    Adjust the size ofthe ellipse using

    .

    Press .

    A

    To change the

    length of the otheraxis, repeat thesame procedurefor the other axis.

  • 7/23/2019 Aplio 400 Operator Manual

    27/441

    No. 2B771-044EN*D

    2-8

    2.4.4 Cross

    Area : Area A 999.99 (cm2) Long-axis length : Dist1 A 999.9 (mm)

    Circumference : Circ A 999.9 (mm) Short-axis length : Dist2 A 999.9 (mm)

    Step 1:

    Set the first point.

    Step 2:

    Set the secondpoint.

    Step 3:

    Set the start pointof the

    perpendicular line.

    Step 4:

    Set the end pointof the

    perpendicular line.

    Press

    and select [Cross].

    Move the marker

    using .

    Press .

    Move the marker

    using .

    Press .

    A dashed lineperpendicular to theline connecting thefirst point and thesecond point isdisplayed.

    Adjust the positionand length of theperpendicular line

    using .

    Press .

    Adjust the endpoint position ofthe perpendicular

    line using .

    Press .

    A

  • 7/23/2019 Aplio 400 Operator Manual

    28/441

    No. 2B771-044EN*D

    2-9

    Step 1:

    Start modification.

    Step 2:

    Select the markerto be modified.

    Step 3:

    Move the markerto the desired

    position.

    Step 4:

    Finishmodification.

    Press . Place the arrowon the marker tobe modified.

    Press .

    Move the markerto the desiredposition using

    .

    Press .

    To move another

    marker, repeat the

    same procedurefor the othermarker.

  • 7/23/2019 Aplio 400 Operator Manual

    29/441

    No. 2B771-044EN*D

    2-10

    2.5 Volume Measurement (Volume)

    The volume is measured using elliptical approximation with three axes set.

    Volume : Volume A 999.9 (cm3) Axis : Dist1 A 999.9 (mm)

    Axis : Dist2 A 999.9 (mm)

    Axis : Dist3 A 999.9 (mm)

    Step 1:

    Set the first and secondlines.

    Step 2:

    A perpendicular planeis displayed.

    Step 3:

    Set the third line on theperpendicular plane.

    Select .

    Follow the sameprocedure as fordistance measurement(Distance) to set thesecond line.

    A

    A

    A cross-sectional plane

    perpendicular to theplane on which step 1has been performed isdisplayed.

    Select again.

    Follow the same

    procedure as fordistance measurement(Distance).

    A

    Press .

    * In volume measurement, the measurement values are not cleared (even if the imageis unfrozen) until measurement of the three axes is completed.

    However, the values are cleared if is pressed or Exam Review mode is entered.

  • 7/23/2019 Aplio 400 Operator Manual

    30/441

    No. 2B771-044EN*D

    2-11

    2.6 Joint Measurement

    The angles between the base line and two lines (angles and ) are measured.

    Angle between the base line and the first line : Alpha A 999 (deg)

    Angle between the base line and the second line : Beta A 999 (deg)

    Step 1:

    Set the first marker.

    Step 2:

    Move the secondmarker.

    Step 3:

    Set the secondmarker.

    Select .

    Move the marker using

    .

    Rotate the marker

    using to adjust

    the angle.

    Press .

    The second line isdisplayed with a centermarker.

    Move the second

    marker using .

    Rotate the marker

    using .

    Press .

    The second linechenges to a dottedline and the third line isdisplayed with a centermarker.

    Move the third marker

    using .

    Rotate the marker

    using .

    Press .

    AA

    A

  • 7/23/2019 Aplio 400 Operator Manual

    31/441

    No. 2B771-044EN*D

    2-12

    2.7 MeanIMT Measurement

    Three distance measurements are performed and the mean value is calculated.

    This method is intended to measure the thickness of the intima-media complex of thecarotid artery.

    Mean : MeanIMT A 999.9 (mm) Distance 1 : Dist1 A 999.9 (mm)

    Distance 2 : Dist2 A 999.9 (mm)

    Distance 3 : Dist3 A 999.9 (mm)

    NOTE: Result display for IMT measurement

    1. When "Double Digits" is selected for Thickness (IMT) Precision in thepreset menu, the IMT measurement value is displayed to two decimalplaces.Depending on the characteristics of the image, however, it may not bepossible to obtain an accurate value to two decimal places. Use themeasurement and calculation values displayed to two decimal places onlyfor reference.

    2. If "Double Digits" is selected, a single quotation mark (') is added to themeasurement values displayed on the screen and the calculation resultsobtained based on the measurement values.

    Limitations: Although markers can be moved in steps of 1 pixel on thescreen, this does not correspond to an indicated value changeof 0.01 due to the resolution of ultrasound images. Theindicated value changes in increments of greater than 0.01 asa mark is moved.

  • 7/23/2019 Aplio 400 Operator Manual

    32/441

    No. 2B771-044EN*D

    2-13

    Step 1:

    Set the first distance.

    Step 2:

    Set the second andthird distances.

    Step 3:

    Determine themeasurement results.

    Select .

    Move the marker using

    .

    * One graduation on thehorizontal axis represents5 mm.

    Rotate the marker to

    the desired angle using

    .

    Press to set the

    upper marker position.

    Move the lower marker

    using and then

    press .

    Press to display

    the next marker.

    Set the second andthird distances using

    the same procedure asfor the first distance.

    * The second and thirdmarkers are displayed ata position 1 cm to theright or left of the firstmarker.

    When themeasurement for thethird position iscompleted, the meanvalue of the threemeasurements isdisplayed.

  • 7/23/2019 Aplio 400 Operator Manual

    33/441

    No. 2B771-044EN*D

    2-14

    2.8 %Stenosis (Distance) Measurement

    The stenosis ratio (%Stenosis) is calculated based on two distance measurementsusing the formulas below.

    If Dist1 > Dist2

    1/)21(% DistDistDistDistS

    If Dist1 < Dist2

    2/)12(% DistDistDistDistS

    2)1(4/ DistAreal 2)2(4/2 DistArea

    If Dist1 > Dist2

    1/)21(% AreaAreaAreaAreaS

    If Dist1 < Dist22/)12(% AreaAreaAreaAreaS

    %Stenosis : %S Dist A 99.9 (%) Distance : Dist1 A 999.9 (mm)

    Distance : Dist2 A 999.9 (mm)

    %Stenosis : %S Area A 99.9 (%) Area : Area1 A 9999.9 (mm )

    Area : Area2 A 9999.9 (mm )

    Press and then perform measurement.

    Refer to subsection 2.1 "Distance Measurement (Distance)" for the measurementprocedure.

  • 7/23/2019 Aplio 400 Operator Manual

    34/441

    No. 2B771-044EN*D

    2-15

    2.9 %Stenosis (Area) Measurement

    Measure Lumen (luminal area) and then Residual (residual luminal area). In Lumenmeasurement, the area is measured using elliptical approximation. In Residualmeasurement, the area is measured using the Continuous Trace method of manualtracing.

    The stenosis ratio (%Stenosis) is calculated based on these two area measurementsusing the formula below.

    If the residual area is larger than the luminal area (which is impossible in actual clinicalexaminations), the result is displayed as a negative value.

    ALumen/A)ResidualA(LumenAArea%S

    %Stenosis : %S Area A 99.9 (%) Luminal area : Lumen A 999.9 (mm2)

    Residual luminal area : Residual A 999.9 (mm )

    Step 1:

    Measure the lumen(maximum area).

    Step 2:

    Display the marker forresidual luminal area

    measurement.

    Step 3:

    Measure the residualluminal area

    (minimum area).

    Select .

    Follow the same

    procedure for Ellipse inarea/circumferencemeasurement (Area) tomeasure the lumen(maximum area).

    Press to display

    the marker for residualluminal area (minimumarea) measurement.

    Follow the sameprocedure as forContinuous Trace for

    area/circumferencemeasurement (Area) tomeasure the residualluminal area (minimumarea).

    Press to set the

    residual luminal area(minimum area).The %Stenosis (Area)value is displayed.

    *

  • 7/23/2019 Aplio 400 Operator Manual

    35/441

    No. 2B771-044EN*D

    3-1

    3. M-mode Measurements

  • 7/23/2019 Aplio 400 Operator Manual

    36/441

    No. 2B771-044EN*D

    3-2

    3.1 M-Mode Distance Measurement (Distance)

    The distance between two points is measured on the M-mode image.

    Distance : Dist A 9999.9 (mm)

    Step 1:

    Set the start point.

    Step 2:

    Move the end point.

    Step 3:

    Set the end point.

    Select .

    Move the intersectionpoint of the marker

    using .

    Press .

    Move the intersectionpoint of the marker

    using .

    Press .

    A

  • 7/23/2019 Aplio 400 Operator Manual

    37/441

    No. 2B771-044EN*D

    3-3

    3.2 Slope Measurement (Slope)

    The slope of the line between two points is measured.

    Slope =(Distance)

    (Time)

    Slope : Slope 999.9 (mm/s) Time : Time A 9.999 (sec)

    Distance : Dist A 999.9 (mm)

    Step 1:

    Set the start point.

    Step 2:

    Move the end point.

    Step 3:

    Set the end point.

    Select .

    Move the intersection

    point of the marker

    using .

    Press .

    Move the intersectionpoint of the marker

    using .

    Press .

    A

  • 7/23/2019 Aplio 400 Operator Manual

    38/441

    No. 2B771-044EN*D

    3-4

    3.3 Time Measurement (Time)

    The time between two points is measured.

    Time : Time A 9.999 (sec)

    Step 1:

    Set the start point.

    Step 2:

    Move the end point.

    Step 3:

    Set the end point.

    Select .

    Move the marker using.

    Press .

    Move the marker using

    . Press .

    A A

  • 7/23/2019 Aplio 400 Operator Manual

    39/441

    No. 2B771-044EN*D

    3-5

    3.4 Heart Rate Measurement (HR)

    The heart rate (bpm) is calculated based on the heartbeats displayed on the monitor.

    Heart rate : HR 999 (bpm) Number of beats : Beat A 9

    Time : Time A 9.999 (sec)

    Step 1:

    Set the value for Beat.

    Step 2:

    Set the start point.

    Step 3:

    Set the end point.

    Select .

    Rotate under

    to set the

    value for Beat. Set thevalue according to thenumber of heartbeats inthe range to bemeasured.

    Move the marker using

    .

    Press .

    Move the marker using

    .

    Press .

    A A

    *

  • 7/23/2019 Aplio 400 Operator Manual

    40/441

    No. 2B771-044EN*D

    4-1

    4. Doppler-Mode Measurements

  • 7/23/2019 Aplio 400 Operator Manual

    41/441

    No. 2B771-044EN*D

    4-2

    4.1 Velocity Measurement (Velocity)

    The velocity is measured on the Doppler-mode image.

    Velocity : Vel 999.9 (cm/s)

    Velocity : Vel A 999.9 (cm/s) Pressure gradient : PG A 999.9 (mmHg)

    Step 1:Set the measurement point.

    Step 2:Set the measurement.

    Select .

    Move the intersection point

    of the marker using .

    Press .

    A

  • 7/23/2019 Aplio 400 Operator Manual

    42/441

    No. 2B771-044EN*D

    4-3

    4.2 Acceleration Measurement (Accel)

    The velocity slope (acceleration) between two points is measured on the Doppler-modeimage.

    The result display differs depending on the items selected in Application Type.

    Acceleration : Accel A 999.9 (cm/s2) Time : Time A 9.999 (sec)

    Velocity : Vel1 A 999.9 (cm/s)

    Velocity : Vel2 A 999.9 (cm/s)

    Acceleration : Accel A 999.9 (cm/s2) Time : Time A 9.999 (sec)Velocity : Vel1 A 999.9 (cm/s) Pressure gradient : PG1 A 999.9 (mmHg)

    Velocity : Vel2 A 999.9 (cm/s) Pressure gradient : PG2 A 999.9 (mmHg)

    Step 1:

    Set the start point.

    Step 2:

    Move the end point.

    Step 3:

    Set the end point.

    Select .

    Move the marker using

    .

    Press .

    Move the marker using

    . Press .

    A

    "Accel" is displayed for thevelocity in the directionaway from the baseline."Decel" is displayed for thevelocity in the direction

    toward the baseline.

    4.3 Time Measurement (Time)

    The measurement procedure is the same as for M-mode time measurement.

    Refer to subsection 3.3 "Time Measurement (Time)".

    4.4 Heart Rate Measurement (HR)

    The measurement procedure is the same as for M-mode heart rate measurement.Refer to subsection 3.4 "Heart Rate Measurement (HR)".

  • 7/23/2019 Aplio 400 Operator Manual

    43/441

    No. 2B771-044EN*D

    4-4

    4.5 Velocity Trace Measurement (Vel Trace)

    The velocity and other items are measured by tracing the Doppler waveform.

    The trace method can be changed using the dial provided under .

    Trace method Description

    Continuous Trace The user traces the Doppler waveform freehand from thespecified start point to the end point using the trackball.

    Spline Trace The Doppler waveform within the specified range is traced byinterpolating three or more user-specified point using a splinecurve.

    Line Trace The user traces the Doppler waveform using straight line

    segments.Range The Doppler waveform in the specified range is traced

    automatically.

    Trace method Description

    Continuous Trace The user traces the Doppler waveform freehand from thespecified start point to the end point using the trackball.

    Spline Trace The Doppler waveform within the specified range is traced by

    interpolating three or more user-specified points using aspline curve.

    Line Trace The user traces the Doppler waveform using straight linesegments.

    Range The Doppler waveform in the specified range is tracedautomatically.

    Auto Range The Doppler waveform is automatically traced by recognizingthe number of heartbeats without specifying the range.

  • 7/23/2019 Aplio 400 Operator Manual

    44/441

    No. 2B771-044EN*D

    4-5

    4.5.1 Continuous Trace

    The user traces the Doppler waveform freehand from the specified start point to the end

    point using .

    The result display differs depending on the items selected in Application Type.

    Maximum velocity : Vmax A 999.9 (cm/s) PI : PI A 99.99*1

    Minimum velocity : Vmin A 999.9 (cm/s) RI : RI A 99.99*2

    Velocity at end diastole : Ved A 999.9 (cm/s) S/D : S/D A 99.99

    Mean velocity : Vm A 999.9 (cm/s)

    *1: Whether Vmin or Ved is used for calculating PI can be preset.

    eakVmin)/Vm_p(VmaxPI

    peakVmVedVmaxPI _/)(

    *2: Whether Vmin or Ved is used for calculating RI can be preset.

    Vmin)/Vmax(VmaxRI

    Ved)/Vmax(VmaxRI

    Velocity Time Integral : VTI A 999.9 (cm)

    Peak velocity : VP A 999.9 (cm/s)

    Mean velocity : VM A 999.9 (cm/s)

    Peak pressure gradient : PPG A 999.9 (mmHg)

    Mean pressure gradient : MPG A 999.9 (mmHg)

  • 7/23/2019 Aplio 400 Operator Manual

    45/441

    No. 2B771-044EN*D

    4-6

    Step 1:

    Set the start point.

    Step 2:

    Trace the Dopplerwaveform.

    Step 3:

    Set the trace.

    Press and

    select [ContinuousTrace].

    Move the marker using

    .

    Press .

    Trace the Dopplerwaveform using .

    Press .

    During tracing, the trace can be modified by rotating or .

    The trace is deleted.

    The deleted portion ofthe trace is restored.

  • 7/23/2019 Aplio 400 Operator Manual

    46/441

    No. 2B771-044EN*D

    4-7

    4.5.2 Spline Trace

    The Doppler waveform is traced by setting at least three points, including the start andend points, and connecting them with a spline curve.

    Same as described in subsection 4.5.1 "Continuous Trace".

    Step 1:

    Set the start point.

    Step 2:

    Trace the Dopplerwaveform.

    Step 3:

    Set the trace.

    Press and

    select [Spline Trace].

    Move the marker using

    .

    Press .

    Move the marker forthe second point using

    .

    Press .

    Repeat the aboveprocedure to setmultiple points.

    Press twice at

    the same point.

    During tracing, the trace can be modified by rotating or .

    The trace is deleted.

    The deleted portion ofthe trace is restored.

  • 7/23/2019 Aplio 400 Operator Manual

    47/441

    No. 2B771-044EN*D

    4-8

    4.5.3 Line Trace

    The Doppler waveform is traced by setting at least three points, including the start andend points, and connecting them with line segments.

    Same as described in subsection 4.5.1 "Continuous Trace".

    Step 1:

    Set the start point.

    Step 2:

    Trace the Dopplerwaveform.

    Step 3:

    Set the trace.

    Press and

    select [Line Trace].

    Move the marker using

    .

    Press .

    Move the marker forthe second point using

    .

    Press .

    Repeat the aboveprocedure to setmultiple points.

    Press twice at

    the same point.

    During tracing, the trace can be modified by rotating or .

    The trace is deleted.

    The deleted portion ofthe trace is restored.

  • 7/23/2019 Aplio 400 Operator Manual

    48/441

    No. 2B771-044EN*D

    4-9

    4.5.4 Range

    The Doppler waveform in the specified range is traced automatically.

    The result display differs depending on the items selected in Application Type.

    Maximum velocity : Vmax A 999.9 (cm/s)

    Minimum velocity : Vmin A 999.9 (cm/s)

    Velocity at end diastole : Ved A 999.9 (cm/s)

    Mean velocity for peak trace : Vm_peak A 999.9 (cm/s)

    Mean velocity for mean trace : Vm_mean A 999.9 (cm/s)

    PI : PI A 99.99

    RI : RI A 99.99

    S/D : S/D A 99.99

    *1: Whether Vmin or Ved is used for calculating PI can be preset.

    peakVmVminVmaxPI _/)(

    peakVmVedVmaxPI _/)(

    *2: Whether Vmin or Ved is used for calculating RI can be preset.

    VmaxVminVmaxRI /)(

    Vmax/)Vmax( VedRI

    Velocity Time Integral : VTI A 999.9 (cm)

    Peak velocity : VP A 999.9 (cm/s)

    Mean velocity : VM A 999.9 (cm/s)

    Peak pressure gradient : PPG A 999.9 (mmHg)

    Mean pressure gradient : MPG A 999.9 (mmHg)

  • 7/23/2019 Aplio 400 Operator Manual

    49/441

    No. 2B771-044EN*D

    4-10

    When [Range] is selected, the following switches for setting parameters are displayed.

    Use these switches during measurement as required.

    Step 1:

    Set the start point.

    Step 2:

    Perform auto-tracing.

    Step 3:

    Adjust the trace line.

    Press and

    select [Range].

    Move the marker using

    .

    Press .

    Move the marker using

    .

    Press .

    Auto-tracing is performed.

    Adjust the trace line asrequired using theparameter switches.

  • 7/23/2019 Aplio 400 Operator Manual

    50/441

    No. 2B771-044EN*D

    4-11

    4.5.5 Auto Range

    The Doppler waveform for the specified number of heartbeats is traced automaticallywithout specifying the range.

    Maximum velocity : Vmax A 999.9 (cm/s)

    Minimum velocity : Vmin A 999.9 (cm/s)

    Velocity at end diastole : Ved A 999.9 (cm/s)

    Mean velocity for peak trace : Vm_peak A 999.9 (cm/s)

    Mean velocity for mean trace : Vm_mean A 999.9 (cm/s)

    PI : PI A 99.99

    RI : RI A 99.99

    S/D : S/D A 99.99

    *1: Whether Vmin or Ved is used for calculating PI can be preset.

    peakVmVminVmaxPI _/)(

    peakVmVedVmaxPI _/)(

    *2: Whether Vmin or Ved is used for calculating RI can be preset.

    VmaxVminVmaxRI /)(

    VmaxVedVmaxRI /)(

    The number of waveforms to be recognized is 1 to 8.

    If the number of waveforms specified is greater than that displayed on the monitorwhen [Auto Range] is started, the maximum number of available waveforms isrecognized.

    If the number of heartbeats is set to "8" on the touch panel, but only five waveformsare displayed on the screen, the range of auto-tracing is automatically set to the fiveheartbeats.

    When the number of heartbeats is changed by rotating , the value set in [Beat]

    on the touch panel is updated accordingly.

    The number of heartbeats last set is saved for each application preset.The saved number of heartbeats is used the next time Auto Range measurement isstarted.

    If no waveform is recognized, the screen displays a marker for the trace start point asin Range measurement.

  • 7/23/2019 Aplio 400 Operator Manual

    51/441

    No. 2B771-044EN*D

    4-12

    4.6 PI Measurement (PI Trace)

    The PI and other items are measured by tracing the Doppler waveform.

    Whether Vmin or Ved is used for calculating PI can be preset.

    eakVmin)/Vm_p-(Vmax=PI

    akVed)/Vm_pe-(Vmax=PI

    PI : PI A 9.99

    Maximum velocity : Vmax A 999.9 (cm/s)

    Minimum velocity : Vmin A 999.9 (cm/s)

    Velocity at end diastole : Ved A 999.9 (cm/s)

    Mean velocity : Vm A 999.9 (cm/s) *1

    *1: When "Range" (auto trace) is selected, the mean velocity is displayed in thefollowing format.

    Vm_peak 999.9 (cm/s)

    The measurement procedure is the same as for velocity trace measurement (Vel Trace).

    Refer to subsection 4.5 "Velocity Trace Measurement (Vel Trace)".

  • 7/23/2019 Aplio 400 Operator Manual

    52/441

    No. 2B771-044EN*D

    4-13

    4.7 RI Measurement

    RI is calculated by measuring the velocity at two points.

    Whether Vmin or Ved is used for calculating RI can be preset.

    Vmin)/Vmax-(Vmax=RI

    Ved)/Vmax-(Vmax=RI

    The result display differs depending on the items selected in the preset.

    RI : RI A 9.99

    Maximum velocity : Vmax A 999.9 (cm/s)

    Minimum velocity : Vmin A 999.9 (cm/s)

    RI : RI A 9.99

    Maximum velocity : Vmax A 999.9 (cm/s)

    Velocity at end diastole : Ved A 999.9 (cm/s)

    Step 1:Set the maximumvelocity point.

    Step 2:Set the minimumvelocity point.

    Step 3:Set the trace.

    Select .

    Move the marker to the

    maximum velocity point

    using .

    Press .

    Move the marker to theminimum velocity point

    using .

    Press .

  • 7/23/2019 Aplio 400 Operator Manual

    53/441

    No. 2B771-044EN*D

    4-14

    4.8 S/D Measurement

    S/D is calculated by measuring the velocity at two points.

    Vmax/VedS/D

    S/D : S/D A 9.99

    Maximum velocity : Vmax A 999.9 (cm/s)

    Velocity at end diastole : Ved A 999.9 (cm/s)

    Step 1:

    Set the maximumvelocity point.

    Step 2:

    Set the end-diastolicvelocity point.

    Step 3:

    Set the trace.

    Select .

    Move the marker to themaximum velocity point

    using .

    Press .

    Move the marker to theend-diastolic velocity

    point using .

    Press .

  • 7/23/2019 Aplio 400 Operator Manual

    54/441

    No. 2B771-044EN*D

    4-15

    4.9 Flow Volume Trace Measurement (Flow Volume)

    The flow velocity is measured by tracing the Doppler waveform in Doppler mode, andthe elliptical area is obtained by measuring the lengths of the major axis and minor axisof the vessel in 2D mode. Based on these measurement results, the flow volume iscalculated.

    The following trace methods are provided. Each time is pressed, the trace

    method is switched.

    The default trace method can be changed in the preset.

    The calculation formulas can also be selected in the preset.

    Switch on the touch panel Trace method

    Flow Vol. Vel Trace The user traces the Doppler waveform freehand from the

    specified start point to the end point using the trackball.Flow Vol. Auto Trace The Doppler waveform in the specified range is traced

    automatically.

    Flow Vol. Auto Range The Doppler waveform in the recognized range is tracedautomatically.

  • 7/23/2019 Aplio 400 Operator Manual

    55/441

    No. 2B771-044EN*D

    4-16

    The result display differs depending on the items selected in Application Type.

    Flow volume : Flow Vol A 9.99(L/min) *1Maximum velocity : Vmax A 999.9 (cm/s)

    Minimum velocity : Vmin A 999.9 (cm/s)

    Velocity at end diastole : Ved A 999.9 (cm/s)

    Mean velocity for peak trace : Vm_peak A 999.9 (cm/s) *5

    Mean velocity for mean trace : Vm_mean A 999.9 (cm/s) *5

    Time-mean velocity : Vm A 999.9 (cm/s) *6

    PI : PI A 9.99 *2

    RI : RI A 9.99 *3

    S/V : S/D A 9.99 *4

    Elliptical area : Area A 99.9 (mm2)

    Length of the major axis of the vessel : Dist1 A 99.9 (mm)Length of the minor axis of the vessel : Dist2 A 99.9 (mm)

    *1: Either of the following two methods can be preset to calculate the flow volume.The trace for the item (Vm_peak or Vm_mean) to be used for calculating the flowvolume is displayed in light blue.The trace for the item not used for calculation is displayed in dark blue.

    1000/100/)(min)/(60)/(_min)/( 2mmAreasscmpeakVmLVolFlow

    1000/100/)(min)/(60)/(_min)/( 2mmAreasscmmeanVmLVolFlow

    *2: Whether Vmin or Ved is used for calculating PI can be preset.

    eakVmin)/Vm_p(VmaxPI

    akVed)/Vm_pe(VmaxPI

    *3: Whether Vmin or Ved is used for calculating RI can be preset.

    Vmin)/Vmax(VmaxRI

    Ved)/Vmax(VmaxRI

    *4: S/D calculation formula

    Vmax/VedS/D

    *5: This item is measured and displayed when Auto Trace or Auto Range is selectedas the trace method.

    *6: This item is measured and displayed when Vel Trace is selected as the tracemethod.

  • 7/23/2019 Aplio 400 Operator Manual

    56/441

    No. 2B771-044EN*D

    4-17

    Stroke volume : SV A 99.9 (mL)

    Heart rate : HR A 999 (bpm)

    Cardiac output : CO A 9.99 (L/min)

    Velocity Time Integral : VTI 999.9 (cm/s)Peak velocity : VP A 999.9 (cm/s)

    Peak pressure gradient : PPG A 999.9 (mmHg)

    Mean pressure gradient : MPG A 999.9 (mmHg)

    Elliptical area : Area A 999.9 (mm )

    Length of the major axis of the vessel : Dist1 A 99.9 (mm)

    Length of the minor axis of the vessel : Dist2 A 99.9 (mm)

    * When the ECG unit is used, the heart rate is displayed automatically when the flowvolume measurement is determined.

    Step 1:

    Select the velocitytracing method.

    Step 2:

    Perform tracing using theselected method.

    Step 3:

    Perform measurementfor the luminal area(elliptical areameasurement).

    Press to

    select the method.

    Perform tracing usingthe selected method.

    Performmeasurement for theluminal area (ellipticalarea measurement).Refer to subsection2.4.3 "Ellipse".

    (a) Perform freehandtracing. (a) Press and

    perform tracing

    using .

    Refer to subsection4.5.1 "ContinuousTrace".

    (b) Perform automatictracing. (b) Press to

    specify the range.Refer to subsection4.5.4 "Range".

    (c) Perform automatictracing using theAuto Rangefunction.

    * This is availableonly whenApplication Typeis Heart1, Heart2,or Heart3.

    (c) Press to

    perform automatictracing for thewaveform of thespecified cardiaccycle.Refer to subsection4.5.5 "Auto Range"

    *

  • 7/23/2019 Aplio 400 Operator Manual

    57/441

    No. 2B771-044EN*D

    5-1

    5. Application Measurement(Cardiac)

    5.1 Introduction

    5.1.1 Starting the application measurement menu

    Select Adult Heart on the Patient Registration screen or select one of Heart 1 to Heart 3

    in the Application preset menu. Then press . The menu for application

    measurement is displayed on the touch panel. Press the switch for the desiredmeasurement.

    The application measurement menu that is displayed when is pressed differs

    according to the preset. The measurement marker settings can also be changed by

    presetting.

    5.1.2 Heart rate acquisition methods (Cardiac)

    The heart rate is measured during LV measurement, aortic valve measurement, mitralvalve measurement, and pulmonary valve measurement. The heart rate can beobtained by the following three methods.

    (1) Acquiring the heart rate automatically from the ECG

    When the optional ECG unit is used and measurement is performed for an item inwhich the heart rate is required, the heart rate is displayed automatically.

    (2) Calculating the heart rate based on the Range measurement result (in M mode andDoppler mode)

    Refer to the section describing heart rate measurement in the basic measurements(M mode/Doppler mode).

    (3) Entering the heart rate manually

    Enter the heart rate value.

    Select .

  • 7/23/2019 Aplio 400 Operator Manual

    58/441

    No. 2B771-044EN*D

    5-2

    NOTE: 1. In dual-display mode, the heart rate for the most recently frozen image isdisplayed. For example, if the image on the left was frozen most recently,the heart rate for that image is also displayed on the image on the right.

    2. When a patient with an unstable heart rate, such as a patient with atrial

    fibrillation, is to be examined, enter a heart rate using the Manual Inputdialog box.

    3. The heart rate value displayed on the 2D (including Color) + PW/CWscreen is the heart rate obtained at the time the PW/CW image is stopped.

    For example, when the PW/CW image is stopped by pressing and

    the 2D image is then frozen, the displayed heart rate value is the heart rateat the time the PW/CW image was stopped, not at the time the Freezeswitch was pressed.

    4. Confirm that the ECG waveform is displayed correctly. If the ECGwaveform is not displayed correctly, an incorrect heart rate value may bedisplayed due to noise.

    5. When an ECG is not used, turn OFF the ECG monitor. Otherwise, anincorrect heart rate value may be displayed due to noise from the ECGmonitor.

  • 7/23/2019 Aplio 400 Operator Manual

    59/441

    No. 2B771-044EN*D

    5-3

    5.2 2D-Mode Measurements (Cardiac)

    5.2.1 LV measurement (MOD Simpson method)/LA measurement (MODSimpson method)/LA measurement (3-Axis method)

    LV measurement (MOD Simpson method)

    The endocardial contour of the left ventricle (LV) is traced on the apical long-axiscross-sectional images (two-chamber view and four-chamber view) acquired in 2Dmode and the left ventricular long axis is specified. The LV volume is then calculatedas the sum of the volumes of 20 elliptical disks perpendicular to the LV long axis.LV volume calculation is also possible using only a two-chamber or four-chambercross-sectional image. In this case, the LV volume is calculated as a sum of thevolumes of 20 circular disks set on the corresponding image.

    Switch Item to be measured Result display

    2ch Diast Tracing of the end-diastolic LV endocardial contour on

    the two-chamber cross-sectional image and setting of theLV long-axis position

    LVAd2 (cm2)

    LVLd2 (mm)EDV2 (mL)

    2ch Syst Tracing of the end-systolic LV endocardial contour on thetwo-chamber cross-sectional image and setting of the LVlong-axis position

    LVAs2 (cm2)

    LVLs2 (mm)

    ESV2 (mL)

    4ch Diast Tracing of the end-diastolic LV endocardial contour onthe four-chamber cross-sectional image and setting of theLV long-axis position

    LVAd4 (cm2)

    LVLd4 (mm)

    EDV4 (mL)

    4ch Syst Tracing of the end-systolic LV endocardial contour on thefour-chamber cross-sectional image and setting of the LVlong-axis position

    LVAs4 (cm2)

    LVLs4 (mm)

    ESV4 (mL)

    HR Heart rate HR (bpm)

  • 7/23/2019 Aplio 400 Operator Manual

    60/441

    No. 2B771-044EN*D

    5-4

    Item name Calculation formula Calculation output Unit

    EDV

    1000

    1

    204

    20

    )( 1i

    4i2i DistanceDistanceLVLd

    LVLd : The LV long-axis length at enddiastole in the two-chamber or four-chamber view (whichever is longer)

    Distance2i : Disk diameter in the two-chamberview

    Distance4i : Disk diameter in the four-chamberview

    LV volume at enddiastole

    (mL)

    ESV

    1000

    1

    204

    20

    )( 1i

    4i2i DistanceDistance

    LVLs

    LVLs : The LV long-axis length at endsystole in the two-chamber or four-

    chamber view (whichever is longer)Distance2i : Disk diameter in the two-chamber

    view

    Distance4i : Disk diameter in the four-chamberview

    LV volume at endsystole

    (mL)

    SV ESVEDV Stroke volume (mL)

    CO1000

    1 SVHR Cardiac output (L/min)

    EF )/(100 EDVSV Ejection fraction (%)

    SI BSASV/ SV index (mL/m2)

    CI BSACO/ CO index (L/min/m2

    )SV4 44 ESVEDV (mL)

    CO41000

    14 SVHR (L/min)

    EF4 )4/4(100 EDVSV (%)

    SI4 BSASV /4 (mL/m2)

    CI4 BSACO /4 (L/min/m2)

    SV2 22 ESVEDV (mL)

    C02

    1000

    12 SVHR (L/min)

    EF2 )2/2(100 EDVSV (%)

    SI2 BSASV /2 (mL/m2)

    CI2 BSACO /2 (L/min/m2)

    LVLd Diff2

    )24(100

    LVLd

    LVLdLVLd Difference in LVlong-axis lengthat end diastole

    (%)

    LVLs Diff2

    )24(100

    LVLs

    LVLsLVLs Difference in LVlong-axis lengthat end systole

    (%)

    BSA Refer to the Operation Manual volume. Body surfacearea (m2

    )

  • 7/23/2019 Aplio 400 Operator Manual

    61/441

    No. 2B771-044EN*D

    5-5

    The LA volume can be measured using one of the following two types ofmeasurement.

    LA measurement (MOD Simpson method)

    The endocardial contour of the left atrium (LA) is traced on the apical long-axis cross-

    sectional images (two-chamber view and four-chamber view) acquired in 2D modeand the left atrial long axis is specified. The LA volume is then calculated as the sumof the volumes of 20 elliptical disks perpendicular to the LA long axis.

    LA volume calculation is also possible using only a two-chamber or four-chambercross-sectional image. In this case, the LA volume is calculated as a sum of thevolumes of 20 circular disks set on the corresponding image.

    LA measurement (3-Axis method)

    The LA volume is calculated by measuring the width and length of the left atrium onthe apical long-axis cross-sectional image (four-chamber view) that is acquired in 2Dmode and measuring the depth of the left atrium on the left-ventricular long-axisimage.

  • 7/23/2019 Aplio 400 Operator Manual

    62/441

    No. 2B771-044EN*D

    5-6

    Switch Item to be measured Result display

    LA 4ch Tracing of the LA endocardial contour on thefour-chamber cross-sectional image and setting of the LA

    long-axis position

    LAa4 (cm2)

    LAd4 (mm)

    LAV4 (mL)LA 2ch Tracing of the LA endocardial contour on the

    two-chamber cross-sectional image and setting of the LAlong-axis position

    LAa2 (cm2)

    LAd2 (mm)

    LAV2 (mL)

    LA W Measurement of the width of the left atrium LA W (mm)

    LA H Measurement of the length of the left atrium LA H (mm)

    LA D Measurement of the depth of the left atrium LA D (mm)

    Item name Calculation formula Calculation output Unit

    LAV1000

    4)Distance(Distance

    DiskCount

    LAL

    DisCount

    1=i

    4i2i

    LAL : The LA long-axis length in the two-chamber or four-chamber view(whichever is longer)

    DiskCount : Number of segments for long-axislength (20)

    Distance2i : Disk diameter in the two-chamberview

    Distance4i : Disk diameter in the four-chamberview

    LA volume (mL)

    LAVI LAV / BSA LAV index (mL/m2)

    LAVI2 LAV2 / BSA 2ch viewLAV index

    (mL/m2)

    LAVI4 LAV4 / BSA 4ch viewLAV index

    (mL/m2)

    LAV (AL)*1 8/3/* LAa4 * LAa2 / (L/10)

    L: LAd4 or LAd2 (whichever is shorter)

    LV volume

    (Area-Lengthmethod)

    (mL)

    LAVI (AL)*1 LAV (AL) / BSA LAV (AL)

    Index

    (mL/m2)

    LA_Vol / 6 * LA W * LA H * LA D / 1000 LA volume (mL)

    LA_VI LA_Vol / BSA LA volume index (mL/m2)

    *1: When LA Volume calculation is performed using the Simpson method, calculationusing the Area-Length method is also performed. The calculation result obtainedusing the Area-Length method is displayed on the Worksheet screen only.

  • 7/23/2019 Aplio 400 Operator Manual

    63/441

    No. 2B771-044EN*D

    5-7

  • 7/23/2019 Aplio 400 Operator Manual

    64/441

    No. 2B771-044EN*D

    5-8

    * An example of LV measurement (MOD Simpson method) is provided below. For LAmeasurement (MOD Simpson method), trace the endocardium of the left atrium inthe same manner.

    (1) Display the end-diastolic four-chamber cross-sectional image.

    (2) Press .

    (3) Display the page of the application measurement menu.

    (4) Press .

    (5) Follow steps (a) to (f) to measure the LV volume.

    (a) Move the marker to the start point of the LVendocardial contour on the apical long-axis

    cross-sectional image, and press .

    (b) Trace the endocardial contour using .

    (c) Press to complete tracing.

    The LV long axis (which starts from the

    midpoint of the line connecting the start andend points) and the lines dividing the LVlong axis into 20 segments are displayed.

    (d) Adjust the LV long-axis end point (apex)position.

    (e) Press . The LV long axis and the segmentation lines are set. The LV

    volume is calculated as the sum of the volumes of 20 circular disksperpendicular to the LV long axis.

    (f) The segmentation lines are deleted when is operated.

    Example of continuous tracingSpline tracing (by setting pointsalong the endocardial contour)is also possible.

    The end point can be movedusing the trackball.

  • 7/23/2019 Aplio 400 Operator Manual

    65/441

    No. 2B771-044EN*D

    5-9

    (6) Press to enter cine mode.

    * The color of at the bottom right of the monitor changes to blue.

    (7) Operate on the cine screen to display the end-systolic four-chamber cross-

    sectional image.

    (8) Press .

    (9) Repeat steps (a) to (f) in (5) to calculate the LVvolume on the end-systolic four-chamber cross-sectional image.

    (10) Display the end-diastolic two-chamber cross-sectional image.

    (11) Press .

    (12) Repeat steps (a) to (f) in (5) to measure the LVvolume on the end-diastolic two-chamber cross-sectional image.

    (13) Press to enter cine mode.

    (14) Operate on the cine screen to display the end-systolic two-chamber cross-

    sectional image.

    (15) Press .

    (16) Repeat steps (a) to (f) in (5) to measure the LVvolume on the end-systolic two-chamber cross-sectional image.

  • 7/23/2019 Aplio 400 Operator Manual

    66/441

    No. 2B771-044EN*D

    5-10

    (17) Based on the end-diastolic four-chamber and two-chamber cross-sectional imagesand the end-systolic four-chamber and two-chamber cross-sectional images, theLV volumes at end diastole and end systole are displayed as the sum of theelliptical disks.

    (18) Perform measurement for LA in the same manner as for LV.

    NOTE: 1. The LV endocardium tracing method can be preset as Continuous orSpline Trace.

    2. It is possible to preset whether or not the measurement marker remainsdisplayed on the images during frame-advance playback in cine mode.

  • 7/23/2019 Aplio 400 Operator Manual

    67/441

    No. 2B771-044EN*D

    5-11

    5.2.2 LV measurement (Teichholz method)

    Switch Item to be measured Result display

    Diastole RVD Right ventricular diameter RVD (mm)IVSTd Interventricular septal thickness at end

    diastoleIVSTd (mm)

    LVIDd LV short-axis diameter at end diastole LVIDd (mm)

    LVPWTd LV posterior wall thickness at end diastole LVPWTd (mm)

    Systole IVSTs Interventricular septal thickness at endsystole

    IVSTs (mm)

    LVIDs LV short-axis diameter at end systole LVIDs (mm)

    LVPWTs LV posterior wall thickness at end systole LVPWTs (mm)

    HR Heart rate HR (bpm)

    Item name Calculation formula Calculation output Unit

    EDV 1000/7 3LVIDd / 10/4.2 LVIDd LV volume at enddiastole

    mL

    ESV 1000/7 3LVIDs / 10/4.2 LVIDs LV volume at endsystole

    mL

    SV ESVEDV Stroke volume mL

    CO 10001 SVHR Cardiac output L/min

    EF EDVSV/100 Ejection fraction %

    FS LVIDdLVIDsLVIDd /)(100 Fractionalshortening

    %

    SI BSASV/ SV index mL/m2

    CI BSACO/ CO index L/min/m2

    BSA Refer to the Operation Manual volume.

    Body surfacearea

    m2

    LV MASSd

    (ASE-Cube)

    1000/04.1 33 LVIDdLVPWTdIVSTdLVIDd

    LV myocardial

    mass at enddiastole

    g

    LV MASSd(Penn-Cube)

    1000/04.1 33 LVIDdLVPWTdIVSTdLVIDd6.13

    LV myocardialmass at enddiastole

    g

    LV MASSd(Teichholz)

    10/10/10/4.2

    )10/)((0.704.1

    3

    LVPWTdLVIDdIVSTd

    LVPWTdLVIDdIVSTd

    10/4.2

    )10/(0.7 3

    LVIDd

    LVIDd

    LV myocardialmass at enddiastole

    g

  • 7/23/2019 Aplio 400 Operator Manual

    68/441

    No. 2B771-044EN*D

    5-12

    Item name Calculation formula Calculation output Unit

    LV MASSd(AV-Cube)

    LVIDdLVPWTdIVSTdLVIDd1.040.8 33 [ 0.6]1000/

    LV myocardialmass at enddiastole

    g

    LV MASSs(ASE-Cube)

    1000/04.1 33 LVIDsLVPWTsIVSTsLVIDs LV myocardial

    mass at endsystole

    g

    LV MASSs(Penn-Cube)

    1000/04.1 33 LVIDsLVPWTsIVSTsLVIDs 6.13

    LV myocardialmass at endsystole

    g

    LV MASSs(Teichholz)

    10/10/10/4.2

    )10/)((0.704.1

    3

    LVPWTsLVIDsIVSTs

    LVPWTsLVIDsIVSTs

    10/4.2

    )10/(0.7 3

    LVIDs

    LVIDs

    LV myocardialmass at endsystole

    g

    LV MASSs(AV-Cube) 04.1[8.0 33

    LVIDsLVPWTsIVSTsLVIDs 6.0]1000/

    LV myocardialmass at endsystole

    g

    LV MASSdIndex

    BSAMASSdLV / LV mass index atend diastole

    g/m2

    LV MASSsIndex

    BSAMASSsLV / LV mass index atend systole

    g/m2

    * The calculation methods for LV MASSd and LV MASSs can be selected and presetfrom among ASE-Cube, Penn-Cube, Teichholz, and AV-Cube.

  • 7/23/2019 Aplio 400 Operator Manual

    69/441

    No. 2B771-044EN*D

    5-13

    [LV Teich.]

    * The other method (Cube or Gibson) can be selected as follows.

  • 7/23/2019 Aplio 400 Operator Manual

    70/441

    No. 2B771-044EN*D

    5-14

    (1) Display the end-diastolic LV cross-sectional image.

    (2) Press .

    (3) Press in the application measurement menu.

    (4) Press .

    (5) Rotate the measurement marker using .

    (6) Perform measurement in parallel using the 2D-mode distance measurementprocedure.

    * Either 4-segment (RVD-LVPWTd) or 3-segment (IVSTd-LVPWTd) can be set.

    RVD

    IVSTd

    LVIDd

    LVPWTd

    Diastole

    (7) Display the end-systolic LV cross-sectional image.

    (8) Press .

    (9) Rotate the measurement marker using .

    (10) Perform measurement in parallel using the 2D-mode distance measurementprocedure.

    LVIDs

    LVPWTs

    IVSTs

    Systole

  • 7/23/2019 Aplio 400 Operator Manual

    71/441

    No. 2B771-044EN*D

    5-15

    NOTE: 1. Functions of the switch

    The LV measurement items set in a measurement sequence can be

    started sequentially each time is pressed, without pressing the

    corresponding measurement switch on the touch panel.

    The measurement items to be started up differ depending on the preset(Type 1/Type 2). In addition, the measurement item to be started up next

    by pressing differs depending the measurement item that is

    selected first.

    Preset type Type 1 Type 2Default

    selectionswitch

    [Diastole] [RVD] [Diastole] [RVD] [LVIDd]

    [Systole]

    [IVSTd]

    [LVIDd]

    [LVPWTd]

    [IVSTs]

    [LVIDs]

    [LVPWTs]

    [LVIDs]

    [IVSTd]

    [LVIDd]

    [LVPWTd]

    [LVIDs]

    [LVIDs]

    2. Default angle of the measurement marker

    The angle of the measurement marker that is displayed when

    is pressed can be preset.

    3. Parallel measurement items for Diastole

    The segments that are measured in parallel after is

    pressed can be preset.

  • 7/23/2019 Aplio 400 Operator Manual

    72/441

    No. 2B771-044EN*D

    5-16

    5.2.3 LV measurement (Cube method)

    By pressing on the touch panel, the LV measurement method can be

    changed from LV measurement (Teichholz) to LV measurement (Cube). It is alsopossible to set LV measurement (Cube) as the default in the presets.

    The measurement items, calculation items, and measurement procedure (measurementpoints) are the same as for the Teichholz method (described in subsection 5.2.2). Onlythe LV volume calculation formula differs from that used in the Teichholz method.

    LV volume at end diastole

    3

    10

    LVIDd

    EDV

    LV volume at end systole3

    10

    LVIDsESV

    5.2.4 LV measurement (Gibson method)

    By pressing on the touch panel, the LV measurement method can be

    changed from LV measurement (Teichholz) to LV measurement (Gibson). It is alsopossible to set LV measurement (Gibson) as the default in the presets.

    The measurement items, calculation items, and measurement procedure (measurementpositions) are the same as for the Teichholz method (described in subsection 5.2.2).Only the LV volume calculation formula differs from that used in the Teichholz method.

    LV volume at end diastole

    2

    1090.5

    1098.052.0

    LVIDdLVIDdEDV

    LV volume at end systole

    2

    1018.4

    1014.152.0

    LVIDsLVIDsESV

  • 7/23/2019 Aplio 400 Operator Manual

    73/441

    No. 2B771-044EN*D

    5-17

    5.2.5 LV measurement (Single-plane (aka Area-Length) method)

    LV measurement (Single-plane (aka Area-Length)) can be preset instead of LVmeasurement (Teichholz).

    Switch Item to be measured Result display

    LVALd LV long-axis cross-sectional area at end diastole LVALd (cm2)

    LVLd LV long-axis length at end diastole LVLd (mm)

    LVALs LV long-axis cross-sectional area at end systole LVALs (cm2)

    LVLs LV long-axis length at end systole LVLs (mm)

    HR Heart rate HR (bpm)

    Item name Calculation formula Calculation output Unit

    EDV

    )10/(

    )(

    3

    8 2

    LVLd

    LVALd

    LV volume at end diastole mL

    ESV

    )10/(

    )(

    3

    8 2

    LVLs

    LVALs

    LV volume at end systole mL

    SV ESVEDV Stroke volume mL

    CO 1000/SVHR Cardiac output L/min

    EF EDVSV/100 Ejection fraction %

    SI BSASV/ SV index mL/m2

    CI BSACO/ CO index L/min/m2

    BSA Refer to the Operation Manual volume.

    Body surface area m2

  • 7/23/2019 Aplio 400 Operator Manual

    74/441

    No. 2B771-044EN*D

    5-18

    [Diastole]

    LVALd (2D-mode areameasurement)

    LVLd (2D-mode distancemeasurement)

    [Systole]

    LVALs (2D-mode areameasurement)

    LVLs (2D-mode distancemeasurement)

  • 7/23/2019 Aplio 400 Operator Manual

    75/441

    No. 2B771-044EN*D

    5-19

    5.2.6 LV measurement (Biplane method)

    LV measurement (Biplane) can be preset instead of LV measurement (Teichholz).

    Switch Item to be measured Result display

    LVALd LV long-axis cross-sectional area at end diastole LVALd (cm2)

    LVAMd LV short-axis cross-sectional area at the mitral valve levelat end diastole

    LVAMd (cm2)

    LVIDd LV short-axis diameter at end diastole LVIDd (mm)

    LVALs LV long-axis cross-sectional area at end systole LVALs (cm2)

    LVAMs LV short-axis cross-sectional area at the mitral valve levelat end systole

    LVAMs (cm2)

    LVIDs LV short-axis diameter at end systole LVIDs (mm)

    HR Heart rate HR (bpm)

    Item name Calculation formula Calculation output Unit

    EDV 8

    3 10

    LVALd LVAMd

    LVIDd ( / ) LV volume at end diastole mL

    ESV 8

    3 10

    LVALs LVAMs

    LVIDs ( / ) LV volume at end systole mL

    SV ESVEDV Stroke volum mL

    CO 1000/SVHR Cardiac output L/min

    EF EDVSV/100 Ejection fraction %

    SI BSASV/ SV index mL/m2

    CI BSACO/ CO index L/min/m2

    BSA Refer to the Operation Manual volume.

    Body surface area m2

  • 7/23/2019 Aplio 400 Operator Manual

    76/441

    No. 2B771-044EN*D

    5-20

    [LV Biplane]

    [Diastole]

    LVALd (2D-mode areameasurement) LVAMd (2D-mode area

    measurement)

    LVIDd (2D-modedistance measurement)

    [Systole]

    LVALs (2D-mode areameasurement)

    LVAMs (2D-mode areameasurement)

    LVIDs (2D-mode

    distance measurement)

  • 7/23/2019 Aplio 400 Operator Manual

    77/441

    No. 2B771-044EN*D

    5-21

    5.2.7 LV measurement (Bullet method)

    LV measurement (Bullet) can be preset instead of LV measurement (Teichholz).

    Switch Item to be measured Result display

    LVAMd LV short-axis cross-sectional area at the mitral valve levelat end diastole

    LVAMd (cm2)

    LVLd LV long-axis length at end diastole LVLd (mm)

    LVAMs LV short-axis cross-sectional area at the mitral valve levelat end systole

    LVAMs (cm2)

    LVLs LV long-axis length at end systole LVLs (mm)

    HR Heart rate HR (bpm)

    Item name Calculation formula Calculation output Unit

    EDV10/

    6

    5LVLdLVAMd

    LV volume at end diastole mL

    ESV10/

    6

    5LVLsLVAMs

    LV volume at end systole mL

    SV ESVEDV Stroke volume mL

    CO 1000/SVHR Cardiac output L/min

    EF EDVSV/100 Ejection fraction %

    SI BSASV/ SV index mL/m2

    CI BSACO/ CO index L/min/m2

    BSA Refer to the Operation Manual volume.

    Body surface area m2

    [LV Bullet]

  • 7/23/2019 Aplio 400 Operator Manual

    78/441

    No. 2B771-044EN*D

    5-22

    LVLd

    [Diastole]

    LVAMd

    LVLs

    [Systole]

    LVAMs

  • 7/23/2019 Aplio 400 Operator Manual

    79/441

    No. 2B771-044EN*D

    5-23

    5.3 M-Mode Measurement (Cardiac)

    5.3.1 Aortic valve measurement (Aortic Valve)

    Aortic valve measurement is performed on M-mode images.

    ET

    Ao Diam

    LA Diam

    AoV Diam

    Switch Item to be measured Result display

    Ao Diam Aortic diameter Ao Diam (mm)

    LA Diam Left atrial diameter LA Diam (mm)

    ET Ejection time ET (sec)

    AoV Diam Aortic valve diameter AoV Diam (mm)

    Item name Calculation formula Result display

    LA/AO DiamAoDiamLA / LA/Ao

  • 7/23/2019 Aplio 400 Operator Manual

    80/441

    No. 2B771-044EN*D

    5-24

    [Aortic Valve]

    (1) Press .

    (2) Display the page of the application measurement menu.

    (3) Press .

    (4) Measure the aortic diameter using the M-mode distance measurement procedure.

    (5) Press .

    (6) Measure the left atrial diameter using the M-mode distance measurementprocedure.

    (7) Press .

    (8) Measure the aortic valve diameter using the M-mode distance measurementprocedure.

    (9) Press .

    (10) Measure the ejection time (time during which the aortic valve is open) using theM-mode time measurement procedure.

  • 7/23/2019 Aplio 400 Operator Manual

    81/441

    No. 2B771-044EN*D

    5-25

    5.3.2 Mitral valve measurement (Mitral Valve)

    Mitral valve measurement is performed on M-mode images.

    >

    DE Amp

    CE Amp

    CA Amp

    EPSS

    EF Slope

    DE Slope

    Switch Item to be measured Result display

    EPSS Distance between the E-point and theinterventricular septum

    EPSS (mm)

    EF Slope Diastolic descent rate of the mitral valve EF Slope (mm/s)

    CE Amp E-wave amplitude CE Amp (mm)

    CA Amp A-wave amplitude CA Amp (mm)

    DE Amp DE-wave amplitude DE Amp (mm)

    DE Slope Mitral valve opening rate DE Slope (mm/s)

    Item name Calculation formula Result display

    CA/CE AMPCEAMPCA / CA/CE

  • 7/23/2019 Aplio 400 Operator Manual

    82/441

    No. 2B771-044EN*D

    5-26

    [Mitral Valve]

    (1) Press .

    (2) Display the page of the application measurement menu.

    (3) Press .

    Measure the distance between the E-point and the interventricular septum usingthe M-mode distance measurement procedure.

    (4) Press .

    Measure the diastolic descent rate of the mitral valve using the M-mode slopemeasurement procedure.

    (5) Press .

    Measure the E-wave amplitude using the M-mode distance measurementprocedure.

    (6) Press .

    Measure the A-wave