AO Instrumentation

Vision Science

Austin Roorda, PhDUniversity of California, Berkeley

CFAO Summer School 2007

Page 2: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

IntroductionAs a vision scientist, you need first to concentrate on your science goals. These goals should drive your decision on the type of adaptive optics system you want to develop.

Basically, AO is an optical technique that offers microscopic optical access to living retina. Whether it is for stimulating a single cone, forming images of small features on the retina, measuring dynamic functional activity, or improving someone’s visual acuity to theretinal sampling limit will dictate exactly how and into what kind of system you implement adaptive optics.

In this talk, I will describe instrumentation that uses adaptiveoptics, from the simple to the most complex and describe the range of basic and clinical science applications associated witheach.

Once you’ve decided on your modality, then you need to know how to actually build the AO system…that’s Don Miller’s talk.

Page 3: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

AO system for flood-illuminated imaging and

vision testing

Page 4: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

imaging

wavefront sensing

wavefront correction

laser beacon

illumination

AO for AO for the eyethe eye

Page 5: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

imagingvision testing

wavefront sensing

wavefront correction

laser beacon

illumination

AO for AO for the eyethe eye

Page 6: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

No AO With AO

JW right eye1 deg eccentricity

image wavelength = 550 nm

Adaptive Optics Makes it Possible to See Microscopic Retinal Features

multiple AO frames

Page 7: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Courtesy of Heidi Hofer, Matt McMahon, David Williams

Page 8: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

MD JP JC

YY

*

*

*

HS AP nasalAN

RS JW temporal BSJW nasal

AP temporal

5 arcmin Roorda and Williams, Nature, 1999Hofer et al, J Neurosci, 2005University of Rochester

Page 9: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Waveguide Properties of Cones

cone efficiency reduces with input angle

11

Page 10: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Light Delivery

EyeTranslating artificial pupil

Adaptive Optics Compensation

CCD

Page 11: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Light Delivery

EyeTranslating artificial pupil

Adaptive Optics Compensation

CCD

Page 12: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Light Delivery

EyeTranslating artificial pupil

Adaptive Optics Compensation

CCD

Page 13: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

JP right eye1º eccentricity2 mm illumination pupil

T N

S

I

0,0

0.866,-1.5-0.866,-1.5

-1.73,0

-0.866,1.5 0.866,1.5

1.73,0

Page 14: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Cone Disarray Plot

T N

S

I

There is a systematic and measurable disarray among the photoreceptors …..

Page 15: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

JP (275 cones) GY (200 cones)

Cone Disarray Projected into the Pupil PlaneCone Disarray Projected into the Pupil Plane

T N

S

I

1

2

0

1

2

0

…but all photoreceptors point in nearly the same direction

Page 16: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

MM – all M conesNC – M cones absent

Has “normal” gene array Missing all L gene(s)

University of Rochester Carroll, Neitz, Hofer, Neitz, Williams, PNAS, 2004

Genotype-Phenotype comparisons

Page 17: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

AO for Vision Testing

Page 18: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

The eye may not prefer to have all of its aberrations corrected

Page 19: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Large stroke adaptive optics for abnormal eyes

52 channelDeformable mirror

WavefrontSensor

Eye

VisualStimulus

Customized Refraction Lab, Geunyoung Yoon, University of Rochester

Page 20: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Large stroke AO can provide high quality optics for both normal and keratoconic eyes.

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

1.8

Totalaberration

HOaberration

Wav

efro

nt rm

s (m

icro

ns)

Normal eyes

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

1.8

Totalaberration

HOaberration

Keratoconic eyes

Without AO

With AO

Page 21: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

With AO, visual acuity for keratoconic eyes was significantly worse than normal eyes.

-0.4

-0.35

-0.3

-0.25

-0.2

-0.15

-0.1

-0.05

06 mm 2.5 mm 6 mm 2.5 mm

Pupil size

Visu

al a

cuity

(Log

MA

R) Visual acuity (Snellen)

8

8.9

10

11.2

12.6

14.2

15.9

17.8

20

20/Normal Keratoconus

Page 22: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

NeurosensoryRetina

Choroid

Sclera

RodsCones

Horizontal Cells

Bipolar CellsAmacrine Cells

Ganglion Cells

Pigment Epithelium

Nerve Fibers

Inner Plexiform Layer

Outer Plexiform Layer

The Retina is a Thick, Multi-layered Structure~2

50 m

icro

ns

Page 23: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

AO Scanning Laser Ophthalmoscopy

(AOSLO)

Page 24: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Adaptive Optics Scanning Laser Ophthalmoscope

In an SLO, AO improves:

• Throughput• Resolution• Contrast

1.2 degrees (~ 360 microns)

Page 25: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Confocal AOSLO

AOSLO confocalthrough-focus series

Page 26: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Dynamic Imaging: Leukocyte Velocity Measurement

• No fluorescent dyes required (safe, long term)• Identify ghost vessels (non-perfused capillaries)• 95% confident of velocity changes as small as 0.076 mm/sec

532 nm laser

Joy Martin

stabilized video difference video

Martin & Roorda, Ophthalmology 112(12): 2219-2224 (2005)

Page 27: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Imaging Function: Dynamic Stimulus Delivery

Allows for simultaneous measurement of optical andretinal limits to vision

Poonja et al, Journal of Refractive Surgery 21(5): 575-580 (2005)

Page 28: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Visual Acuity

No AO

AO 5.81mm

SnellenAcuity

20/1220/1120/1020/920/820/720/6

20/1320/13

20/515

20

25

30

35

40

45

subject 1 subject 2 subject 3

MA

R (a

rc s

econ

ds)

Ethan Rossi

Page 29: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Optical Coherence Tomography(OCT)

Page 30: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

http://www.fitzpatrick.duke.edu/biophotonics/research/OCT/slide_2.htm

Page 31: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

OCT retinal cross section

Bigelow, Iftimia, Ferguson, Ustun, Bloom, Hammer. JOSA A, 2007

Page 32: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

AO OFF->ON

AO Optical Coherence Tomography

courtesy of Jack Werner, UC Davis

Page 33: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

cones

NFL

Focus at photoreceptors Focus at nerve fibers

1.25 degrees(375 microns)

AO-OCTB-scans at 2° retinal eccentricity

Courtesy of Donald Miller, Indiana UniversityZhang et al. Optics Express, 14(10): 4380-4394 (May 2006)

Page 34: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

images courtesy of Robert Zawadzki, UC Davis

Page 35: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

330 µm

ILM

IS/OS

Single frame

Avg of 60 frames

B-scan at 3° eccentricity with AO SD-OCT

38 µm

16 µm

16 µm 5 µm

Courtesy of Donald Miller, Indiana University

Page 36: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Resolution vs Contrast

• Optical resolution with AO is sufficient to resolve all cell bodies in the retina

• But most of the cells are transparent (fortunately for vision), so optical scattering methods are ineffective. In other words, the intrinsic contrast is low.

• There are a host of methods to improve contrast of targeted features in the retina

Page 37: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Improving Contrast:

Fluorescence

Page 38: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

AOSLO Fluorescein Angiography(macaque retina)

Registered frame

University of RochesterGray et al, Optics Express 14, 7144-7158 (2006)

AOSLO movie

Page 39: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

In vitro

Fluorescently labeled ganglion cell bodies and dendrites taken with a confocalmicroscope

Retrograde rhodamine labeled ganglion cells in the living macaque taken with AOSLO

Images courtesy of: Dan Gray, Bill Merigan, David Williams, University of Rochester

In vivo

Page 40: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Improving Contrast:

Autofluorescence

Page 41: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Autofluorescence image of the RPE mosaic in a macaque retina

100µm Courtesy of Morgan, Gray, Merigan, Williams, U Rochester

Page 42: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Courtesy of Heidi Hofer, Matt McMahon, David Williams100µm

Flood-illuminated image of the cone mosaic in a human retina

Page 43: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Improving Contrast:

Polarization

Page 44: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Deselecting polarized light removes specular reflections

patient with epiretinal membrane

raw SLO image depolarized component

Page 45: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

AO permits PS-OCT to probe smaller structures.

900 µm

34 dB

0 dB

40°

Henle’s fiberlayer induces

phase retardation.

Intensity

Double pass phase retardation

40°0dB

34dB

~490

µm

Higher lateral resolution and smaller speckle size reduce spatial averaging in the polarization analysis.

Cense et al., Indiana University

Page 46: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Improving Contrast:

Wavelength

Page 47: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

830 nm

drusen

543 nm

Retinal vessels

488 nmMacular pigment, nerve fiber layer

633 nm

Some nerve fiber layer

514 nm

Some macularpigment, melanin

Elsner et al., Vision Research, 1996

Courtesy of Steve Burns and Ann Elsner, Indiana University

Page 48: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Red vs Green AOSLO video

660 nm light 532 nm light

Page 49: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Improving Contrast:

Motion Contrast

Page 50: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Imaging Blood Flow

R. Ferguson, D. Hammer, A. Elsner, R. Webb, S. Burns, and J. Weiter, Opt. Express 12, 5198-5208 (2004)

Page 51: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Functional Imaging

• Imaging the anatomy on a microscopic scale with AO can only tell us so much

• There have been many efforts in the last few years to make structure function relationships

Page 52: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Imaging Retinal Function

Intrinsic Retinal Signals

Page 53: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Cone Scintillation Indicates Visual Activity

control

test

Don Miller, Indiana University

Page 54: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Functional Imaging via Reflectance Changes

infrared imaging beam

red stimulus beam

Kate Grieve, UC Berkeley

Page 55: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Origin of Cone Signals

Kate Grieve, UC Berkeley

Page 56: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Imaging Retinal Function

Eye Tracking

Page 57: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Eye Movement Tracking

-15

-10

-5

0

5

10

15

0 0.25 0.5 0.75 1 1.25 1.5Time (seconds)

Posi

tion

(arc

min

utes

)

horizontalvertical

960 Hz eye trace, accurate to one image pixel (0.14 minutes of arc)

original video stabilized video

Scott Stevenson, Houston; David Arathorn, Curt Vogel, Al Parker, Qiang Yang, MSU

Page 58: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

The fixation point is displaced about 10’ of arcfrom the point of maximum cone density

What is the fovea?

50µm

Putnam et al, Journal of Vision 5(7):632-639 (2005) )University of Rochester

Page 59: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Visual Psychophysics

saccades pursuit

Where does the eye place the image immediately after a saccade?

Where does the eye place an image that it is pursuing?

Scott Stevenson, Girish Kumar, University of Houston

Page 60: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

What is the fovea?

Scott Stevenson, Girish Kumar, University of Houston

Page 61: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Where am I looking? (WAIL)

Scott Stevenson, Girish Kumar, University of Houston

Page 62: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Imaging Retinal Function

Small Spot Stimulation

Page 63: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

1.0

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1 arcminute

1.0

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

Without adaptive optics With adaptive optics

Adaptive optics increases the fraction of light absorbed by a single cone

6 mm pupil 550 nm

Courtesy: Heidi HoferHofer et al. University of Rochester

Page 64: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Delivery of AO-corrected Stimuli

All stimuli seen Some stimuli not seen

Makous and Carroll, University of Rochester

Page 65: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

-1.2 -1.0 -0.8 -0.6 -0.4 -0.2 0.0 0.2

Log Luminance

Prop

orti

on o

f D

etec

ted

Flas

hes

normals: open symbols

AO-Microperimetry confirms cone drop-out

Makous et al, IOVS 47(9): 4160-4167 (2006)University of Rochester

Page 66: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Hofer et al. J Vis. 2005 May 19;5(5):444-54.

Page 67: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

AOSLO - Stabilized Stimulus Delivery

Page 68: AO Instrumentation Vision Science - Lick Observatory...AO Instrumentation Vision Science Austin Roorda, PhD University of California, Berkeley ... range of basic and clinical science

Future Applications

• novel anatomical imaging– 2-photon– phase contrast imaging

• novel functional imaging– activity dependent dyes – combined AO imaging and electrophysiology

• novel applications– targeted photodynamic therapy– microsurgery