http://anapa.com
New Generation
Gas-Pressured Infusion Device
anapa & anaplus
More Convenient!
MoreAccurate!
MuchSafer!
ananpa & anaplus
2
• Repetitive filling of medication by
syringe
• Difficult due to the internal pressure of
pump
No-resistance Filling
Maximizing
convenience
anapaExisting Pumps
• Long Priming time
• Long preparation time and inefficiency
World’s 1st
about 2minuts Priming
• Flow change due to the position of
Chamber
• Not optimized for treatment due to the
inaccurate flow rate
More accurate flow rate
by increased pressure
ananpa & anaplus
3
1. No-resistance Filling
2. Short Priming time
4. Monitoring the injected volume accurately
5. Prevention of over-dose
6. Minimizing residual volume
3.Minimizing the change of flow rate by chamber
position
7. Minimizing the risk of contamination
ananpa & anaplus
5
World 1st
CO2 PressuredInfusion device
SpecialRelief Valve(R/V)for Maintaining
CO₂ gas pressure
Constant flow rate by
Glass capillary
Gas-Pressured Infusion Device
ananpa & anaplus
ananpa & anaplus
6
Citric acid
3Na2CO3
SodiumCarbonate
2C6H8O7
Generation of CO2
gas
3H2O+ 3CO2↑+ 2Na3C6H5O7
ananpa & anaplus
7
Membrane Filter
• Hydrophobic Filter : Pass gas, Keep liquid
• Generated gas passes membrane filter and moves to Chamber, pushes
Piston.
ananpa & anaplus
8
Relief Valve(R/V)
The pressure inside Chamber is
controlled at set-value,
Extra CO₂ gas is released outside
Chamber
=> Accurate flow rate
0bar
0.1bar
0.2bar
0.3bar
0.4bar
0.5bar
0.6bar
0.7bar
0ml
1ml
2ml
3ml
4ml
0hr 8hr 16hr 24hr 32hr 40hr
Pre
ssure
[ bar
]Flo
wR
ate
[ m
l/hr
]
Time [ hr ]
2.2ml
1.8ml
48hr 56hr
ananpa & anaplus
9
ModelAvg. Flow rate
(Criteria ± 15%)
Instantaneous Flow Rate (Criteria ± 50%)
Max. Flow rate Min. Flow rate Difference
anapa 0.5% (2.01ml/hr) 2.0% (2.04 ml/hr) -3.0% (1.94 ml/hr) 5%
"A" 16.5% (2.33ml/hr)50.0% (3.00 ml/hr
)6.0% (2.12 ml/hr) 44%
“B" -8.5% (1.83ml/hr)10.0% (2.20 ml/hr
)-20.0% (1.60 ml/hr) 30%
“C" -18.0% (1.23ml/hr)16.0% (1.74 ml/hr
)-29.3% (1.06 ml/hr) 45.3%
• - The average flow rate shall have a tolerance of ± 15%
compared to the nominal flow rate.
• - At least 80% of the nominal volume shall be delivered at
an instantaneous flow rate within ± 50% of the
In accordance with ISO 28620:2010_Non-electrically driven portable infusion device
ananpa & anaplus
10
• Instantaneous Flow Rate both at the beginning and end of injection should be in ± 10% variation.
• The accuracy is key criteria of Infusion pump.
0bar
0.1bar
0.2bar
0.3bar
0.4bar
0.5bar
0.6bar
0.9ar∨
0ml
1ml
2ml
3ml
4ml
0hr 8hr 16hr 24hr 32hr 40hr
anapa Flow Rate (ml/hr)
Balloon Flow Rate (ml/hr)
anapa Pressure (bar)
Balloon Pressure (bar)
Pre
ssure
[ bar
]
Flo
wR
ate
[ m
l/hr
]
Time [ hr ]
2.2ml
1.8ml
48hr 56hr
+15%
-15%- 10%
+10%
ananpa & anaplus
11
• The first internal pressure is “0” as the inside and outside of Chamber connected.
• The air inside Chamber goes outside upon filling.
• Totally different from balloon pumps which much power is required during filling procedure.
• Minimizing the preparation road and time of clinicians
ananpa & anaplus
12
Fillin
g P
ressure
Balloonpumps
anapa
Filling Pressure
A : 9.28kgf
B : 9.08kgf
C : 8.04kgf
No resistance
anaplus
1.50kgf
D : 5.40kgf
ananpa & anaplus
13
• Change on the flow rate by the position of Chamber.
ex) hanging on rack, put in pouch and wear on waist etc.
• The variation on flow rate of Balloon pumps is widely due to the low pushing pressure.
Flow change by the position of Chamber
( +100cm ) 2.626 (+100cm ) 31.3% 21.0%
( +50cm ) 2.488 ( +50cm ) 24.4% 13.6%
( 0cm ) 2.216 ( 0cm ) 10.8% 기준
( -50cm ) 2.006 ( -50cm ) 0.30% -10.5%
AverageFlow Rate
(ml/hr)
Percentage
Deviation(%)
0
1
2
3
4
0 8 16 24 32 40 48 56
유량(+100cm)
유량 0cm
유량 (+50cm)
유량(-50cm)
압력
[bar]
유량
[ml/
hr]
Balloon pumps anapa
( +100cm ) 2.03 (+100cm ) 1.5% 6.0%
( +50cm ) 1.90 ( +50cm ) -5.0% -0.5%
( 0cm ) 1.91 ( 0cm ) -4.5% 기준
( -50cm ) 1.80 ( -50cm ) -10% -5.5%
AverageFlow Rate
(ml/hr)
Percentage
Deviation(%)
0
1
2
3
4
0 8 16 24 32 40 48 56
유량(+100cm)
유량 0cm
유량 (+50cm)
유량(-50cm)압력
[bar]
유량
[ml/
hr]
주입압력 1bar 주입압력 0.3-0.5bar
ananpa & anaplus
14
0bar
0.1bar
0.2bar
0.3bar
0.4bar
0.5bar
0.6bar
0.7bar
0.8bar
0.9bar
1bar
1.1bar
0ml
1ml
2ml
3ml
4ml
5ml
6ml
0hr 8hr 16hr 24hr 32hr 40hr 48hr 56hr
Flow Rate 0cm anapa +50cm
New anapa +50cm anapa -50cm
New anapa -50cm Pressure
Time [ hr ]
Flo
wRate
[ m
l/hr
]
Pre
ssure
[bar]
Higher pushing pressure, Lower impact from Potential
energy
Pushing Pressure : about 1
bar
Pushing Pressure :
about 0.5 bar
+50cm+50cm
-50cm-50cm
Pressure +50cm
Pressure +50cm
anapa New anapa
ananpa & anaplus
15
Switch Valve(S/V)
If CO₂ gas exceeds
Set-volume
Switch Valve
Moves to left
CO₂ gas is released outside automatically
through SpecialAirway
Double Pressure Control System
ananpa & anaplus
17
Product Balloon Pumps
Priming time About 2’30” About15’
Compare with other products
ananpa & anaplus
18
• Priming is procedure to remove air inside
tube line and prevents air injected to
patient and contamination.
• All the air inside PCM is removed and
Priming time is shortened about 2’ 30”.
ananpa & anaplus
19
• Easy to use, Grip type design
• Arched design prevents pushing the button accidently
=> The risk of over-dose is minimized.
Grip형
ananpa & anaplus
20
• There are some cases that PCM clip is not removed after Priming.
• In case not remove PCM clip, Bolus volume keeps injecting to patient
additionally(over-dose) and it cause negative side effect to patient.
• To prevent this case, End-cap and PCM clip are linked by string and
can be removed together.
ananpa & anaplus
21
• Hydrophobic Filter : Only air is passed
• No need to open End-cap to check the
completion of priming => Minimizing the risk of
contamination
ananpa & anaplus
22
Flow restrictor, glass capillary has 3 layer rings to prevent
overflow.
=> Safe and constant flow is available.
ananpa & anaplus
25
item Anaplus Balloon pumps
AccuracyAvg. Flow rate = Instantaneous Flow Rate= +/- 10%
Avg. Flow rate = +/- 15% Instantaneous Flow Rate = +/- 50%
Flow changeby Chamber position
+/- 0.8% per 10cm +/- 10% per 10cm
Priming Time About 2’30” About 10-15min
Filling Pressure <1.5kgf 5.4kgf – 9.3kgf
Residual Volume 1ml > 2ml <
Prevention of Over-dose
1. Remove PCM Clip & End-cap together2. 3-layer ring of Capillary housing
No
Air Removal
4 step system1. Remove air inside Chamber2. Remove air inside PCM 3. Remove air inside Tube line 4. Before patient connection: End-cap
2 step system1. Remove air inside Tube line2. Before patient connection: End-cap