Analytical Methodologies for Per/Poly Fluoroalkyl Substances (PFAS) Measurement in Environmental Waters by LC/MS/MS

Tarun Anumol, Ph.D.Applications Development ScientistAgilent Technologies Inc.Wilmington, DE, USA

Contact: [email protected]

Page 2: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Agilent ASTS-Houston

‘Fit for Purpose’ Analysis

8 August 20172

Page 3: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Considerations when selecting analytical technique to employ

8 August 2017 Agilent ASTS-Houston 3

I. Types of analytesII. Concentrations required (ppm,

ppb, ppt etc.)III. Type of matrix (drinking water,

surface water, wastewater, sludge, food etc.)

IV. Type of sample clean-up required V. Regulatory considerations (EPA

methods, EU & ASTM methods etc.)

Factors in deciding analytical method selection

Page 4: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Terminology

8 August 2017 Agilent ASTS-Houston 4

• Terminology of these compounds can be confusing but it is critical to understand the differences

Page 5: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

TerminologyFluorinated Compounds

8 August 2017 Agilent ASTS-Houston 5

Any organic or inorganic compound that has a fluorine atom incorporated in it

Page 6: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

TerminologyPer/PolyfluoroAlkyl Substances (PFAS)

8 August 2017 Agilent ASTS-Houston 6

R: Continuation of alkyl chain or functional group

When one or more Carbons (C) in an alkyl chain have the Hydrogen (H) replaced by Fluorine (F)

Page 7: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

8 August 2017 Agilent ASTS-Houston 7

Terminology

Polyfluoroalkyl substance

ALL H atoms linked to C in alkyl chain are substituted with F

SOME (but not all) H atoms linked to C in alkyl chain are substituted with F

Perfluoroalkyl substance

Page 8: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

8 August 2017 Agilent ASTS-Houston 8

TerminologyPerfluoroalkyl acids (PFAAs)

• Perfluorocarboxylic acids (PFCAs)

• Perfluorosulfonic acids (PFSAs)

Page 9: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

TerminologyPolyfluorinated compounds - Fluorotelomers

8 August 2017 Agilent ASTS-Houston 9

6:2 FTUCA

8:2 FTUCA

6-C alkyl chain: Hydrophobic 2-C: Hydrophilic

• Formed during the telomerization process in manufacturing

Page 10: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Per/Polyfluoroalkyl substancesUnique Properties

8 August 2017 Agilent ASTS-Houston 10

• Thermal & Chemical stability: grease-proof food packaging, stain repellents

• Zwitterionic properties: surfactants • Surface-tension lowering: fire-fighting foams

Page 11: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Per/Polyfluoroalkyl substancesToxicology & Regulations in US

8 August 201711

Toxicological data:

• PFAS exposure can have adverse effects to the neurological and reproductive systems

• Long chain (>7-C chain) PFASs are bio accumulative, and persistent• Long half-lives for longer chain PFAS• Very difficult to degrade the final products (PFCAs, PFSAs) in the environment

Regulatory Overview:

• USEPA has a public health guidance for PFOA and PFOS at 70 ng/L in drinking water• Many states have their own health guidelines ranging from 20-400 ng/L for PFOS,

PFOA, PFNA and PFHxA

Agilent ASTS-Houston

Page 12: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

PFAS Analysis – LC Instrument SetupEliminate Background Contamination

8 August 2017 Agilent ASTS-Houston 12

Application Note: Recommended Plumbing Configurations for Reduction in Per/Polyfluoroalkyl Substance Background with Agilent 1260/1290 Infinity (II) LC’s (5991-7863EN)

Page 13: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

8 August 2017 Agilent ASTS-Houston 13

PFAS Analysis – LC SetupEliminate Background Contamination

Page 14: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

PFAS Instrument SetupBackground Contamination

8 August 2017 Agilent ASTS-Houston 14

Actual PFOS peak

5 ng/L standard

Mili-Q Water (‘Blank’)

Background PFOS from LC system

• The use of a Delay column is convenient when running multiple methods on the same instrument.

• “Delay” column and not a “Trapping” column. I.e. the background peak will be retention time separated but will be present.

• The delay column does not account for any contamination after the pump (the autosampler)

Page 15: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

250 mL sample collected in Polypropylene Bottles with 5 g/L Trizma (to quench residual chlorine)

Solid Phase Extraction and evaporation to final 1 mL extract in ~96/4 MeOH/Water

Analysis by LC-MS/MS in Negative Electrospray mode for 14 analytes (9 PFCAs, 3 PFSAs, 2 FOSAAs) – 37 min analysis (10 µL injection)

EPA 537: Perfluorinated Alkyl Acids in Drinking Water by SPE-LC/MS/MS

8 August 2017 Agilent ASTS-Houston 15

Confidentiality Label

Page 16: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Solid Phase ExtractionUSEPA 537 vs Modified

8 August 2017 Agilent ASTS-Houston 16

Recoveries calculated in Reagent Water

Modified SPE cartridge: Agilent SampliQ WAX cartridge (6 cc, 150 mg)

All recoveries between 70-130 %

Spiking concentrations:

Modified SPE: 4 ng/L in 250 mLEPA 537: 5-21 ng/L in 250 mL

Final extract: 1 mL ~95% MeOH

PFHxA

PFHpA

PFOA

PFNA

PFDA

PFUdA

PFDoA

PFTrDA

PFTeDA

PFBS

PFHxS

PFOS

N-MeF

OSAA

N-EtFOSAA 0

25

50

75

100

125

150EPA 537 (n=7)Modified SPE (n=6)

Rec

over

y (%

)

Page 17: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Agilent ASTS-Houston

Typical LC/TQ Lab Layout

6 feet1.83 meters

8 August 201717

Page 18: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Quick Change Ion Injector

Fast Ion Injector exchange without system venting

Ion Injector Tool VacShield

Easy to use, increased throughput, less downtime for maintenance

8 August 2017 Agilent ASTS-Houston 18

Page 19: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

8 August 2017 Agilent ASTS-Houston 19

Easy Change Detector Assembly

Page 20: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Analysis of PFAS in Drinking Water with UltivoEPA Method 537

8 August 2017 Agilent ASTS-Houston 20

Analysis of 17 PFASs in drinking water with <0.1 ng/mL DLs in extract (includes all comps in EPA 537)

Page 21: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Analysis of PFAS in Drinking Water with UltivoEPA Method 537

8 August 2017 Agilent ASTS-Houston 21

PFOS - 7 Levels, 7 Levels Used, 14 Points, 14 Points Used, 0 QCs

Concentration (ng/

-1 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21

Res

pons

es

3x10

-0.2

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

1.8

2

2.2

2.4

2.6

2.8

3

y = 147.003489 * x + 20.563484 R^2 = 0.99890136 Type:Linear, Origin:Ignore, Weight:NonePFOA - 7 Levels, 7 Levels Used, 14 Points, 14 Points Used, 0 QCs

Concentration (ng/

-1 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21

Res

pons

es

4x10

-0.5

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

6

y = 3046.084833 * x + 1326.038272 R^2 = 0.99894706 Type:Linear, Origin:Ignore, Weight:None

PFOS PFOA

Calibration Curves from 0.1-20 ng/mL for PFOA and PFOS

Post-spike recoveries of 17 PFASs in a water sample extracted according to EPA 537 with Agilent SampliQ WAX cartridges

Ultivo Triple Quad LC/MS

• All 17 PFASs had linear calibration curves, with R2 > 0.99

• Overall recoveries were between 70–125% for both spiking levels

• %RSD was 0.3–10.8 %for all compounds at both spiking levels (1.0 μg/L and 5.0 μg/L)

Page 22: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Analysis of 30 PFASs in Drinking WaterAgilent 1260 Infinity II + 6470 Triple Quadrupole LC-MS

8 August 2017 Agilent ASTS-Houston 22

Application Note: Analysis of Per/Polyfluoroalkyl Substances in Water Using an Agilent 6470 Triple Quadrupole LC/MS (5991-7951EN)

Chromatogram depicting MRM transitions of 30 PFASs post-spiked in a drinking water extract at 1 µg/L

Page 23: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Analysis of 30 PFASs in Drinking WaterAgilent 1260 Infinity II + 6470 Triple Quadrupole LC-MS

8 August 2017 Agilent ASTS-Houston 23

Sensitivity (0.08 ng/L equil. In water) Precision

Page 24: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Analysis of Environmental Contaminants by LC-MS/MSIon Suppression Comparison by Sample Prep Method

Conventional SPEMass on Column :150 pg

Online SPEMass on Column :150 pg

Direct InjectionMass on Column :150 pg

8 August 2017 Agilent ASTS-Houston 24

Page 25: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Agilent 6495 Triple Quadrupole LC/MSUltimate Sensitivity, Extreme Reproducibility

8 August 2017 Agilent ASTS-Houston 25

8/8/2017Agilent 6495 QQQ LC/MS System

Agilent Jet Stream Hexabore Capillary Dual Ion Funnel

• Thermal gradient focusing• Efficient desolvation • Creates an ion rich zone

• Six capillary inlets• Samples x10 times more

ion rich gas

• Removes the gas but captures the ions

• Removes neutral noise

Agilent 6495 LC/TQ

Page 26: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Perfluorinated Alkyl Acids in Drinking Water by LC/TQCollaborative work with Suffolk County Water Authority

• 1290 Infinity LC coupled to a 6490 LC-MS/MS

• 14 PFAAs in 6 min

System

• PFOA, PFOS: <1 ng/L• Others: 1-10 ng/L• 80 µL aqueous injection

MDLs:

• Drinking Water• Surface Water• Ground Water

Matrices:

8 August 2017 Agilent ASTS-Houston 26

2x10

-0.1

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

Cpd 9: PFHpA: -ESI MRM Frag=380.0V CF=0.000 DF=0.000 [email protected] (362.90000 -> 169.00000) Cal Std ppb 2.d

1.66 2.02

1 1

Counts (%) vs. Acquisition Time (min)0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2 2.2 2.4 2.6 2.8 3 3.2 3.4 3.6 3.8 4 4.2

Page 27: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Analysis of PFOA/PFOSDirect Injection; Rapid Analysis (1290 II UHPLC + 6495 LC-MS/MS)

8 August 2017 Agilent ASTS-Houston 27

PFOS

PFOA

10 ng/L standard

• Use Multi-wash function!!!• PFOS sticks to needle and seat especially

with long worklists

Page 28: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Targeted Analysis LC-MS/MS vs LC-Q/TOF

8 August 2017 Agilent ASTS-Houston 28

PFHxSPFOA

PFOS

PFHxSPFOA

PFOS0

20

40

60LC-MS/MSLC-QTOF

Con

cent

ratio

n (n

g/L)

Influent Effluent

Collaborative project with RMIT University, Melbourne (Timothy Coggan ; Prof. Bradley Clarke)

Agilent 6550 LC-Q/TOF

Page 29: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Suspect ScreeningPersonal Compounds Database and Library with Spectra• Custom PFAS database with 106 compounds• MS/MS spectra and Retention time data available for a subset of compounds

8 August 2017 Agilent ASTS-Houston 29

Page 30: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Untargeted Analysis

8 August 2017 Agilent ASTS-Houston 30

Page 31: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Untargeted AnalysisIdentification of Novel PFAS

8 August 2017 Agilent ASTS-Houston 31

Inclusion list for targeted MS/MS and structural elucidation using fragment information using software tools (Molecular structure correlator)

Page 32: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

Conclusions

Background contamination of PFASs is prevalent in ALL systems but can be fixed through systematic elimination of fluoropolymers in the LC.The Ultivo Triple Quad LC/MS offers significantly more sensitivity than

EPA 537 analysis while using a smaller footprint through several innovative technologies.The sensitive analysis of several newer classes of PFASs is possible

using the Ultivo and 6470.Rapid analysis of PFOA/PFOS and PFAAs can be performed by direct

large volume aqueous injection using the 6495 with <1 ng/L MDLs.Identification, screening and quantification of ‘emerging’ PFASs has

been shown using the Agilent LC-QTOF

8 August 2017 Agilent ASTS-Houston 32

Page 33: Analytical Methodologies for Per/Poly Fluoroalkyl ... · PFAS Instrument Setup Background Contamination. 14 8 August 2017 Agilent ASTS-Houston Actual PFOS peak 5 ng/L standard Mili-Q

References

8 August 2017 Agilent ASTS-Houston 33

5991-8156EN

5991-7951EN 5991-7863EN

5991-5669EN