4
Analysis of synthetic polymers by MALDI-TOF MS Polymer Analysis software for the characterisation of polymers and co-polymers using MALDI data Automatically generate useful information (e.g. monomer mass, residual mass (end groups), polymer statistics, etc.) regarding the sample Application News No. MO329 AXIMA Performance MALDI-TOF Analysis of Polymers by MALDI-TOF Mass Spectrometry

Analysis of Polymers by MALDI-TOF Mass Spectrometry MO329

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Analysis of Polymers by MALDI-TOF Mass Spectrometry MO329

Analysis of synthetic polymers by MALDI-TOF MS

Polymer Analysis software for the characterisation of polymers and co-polymers using MALDI data

Automatically generate useful information (e.g. monomer mass, residual mass(end groups), polymer statistics, etc.) regarding the sample

ApplicationNews

No. MO329

AXIMA Performance MALDI-TOF

Analysis of Polymers by MALDI-TOF Mass Spectrometry

Page 2: Analysis of Polymers by MALDI-TOF Mass Spectrometry MO329

Analysis of Polymers by MALDI-TOF Mass Spectrometry

Introduction

Synthetic polymers are widespread in everyday life.Whether they are used as containers for householdproducts and foodstu�s to the construction of carcomponents and electronic devices, plastics now playimportant roles in our day-to-day activities.

Matrix assisted laser desorption/ionization massspectrometry (MALDI-MS) is becoming increasinglypopular for the analysis of synthetic polymers. ThePolymer Division of the National Institute ofStandards and Technology (NIST) is dedicated togathering information regarding methods used in theanalysis of polymers. One of the resources of thePolymer Division includes a collection of samplepreparation methods speci�cally for samples whichare to be analyzed by MALDI-TOF massspectrometry,(http://polymers.msel.nist.gov/maldirecipes/index.cfm ),highlighting the importance of MALDI in polymeranalysis.

The word polymer is derived from the Greek for‘many parts’ and is used to describe any materialwhich is composed of repeating subunits(‘monomers’). In synthetic polymers, if the repeatingunits are identical (e.g. (CH2CH 2O) n), the polymer isreferred to as a homopolymer . Alternatively, thepolymer can consist of di�erent monomer units(e.g. (C2H4O) m(C 3H6O) n), in which case thepolymer is referred to as a copolymer .

The synthesis of polymers can be di�cult to preciselycontrol. Typically, a polymer sample will not containmolecules with the exact same number of monomerunits, but rather will contain polymer chains ofdi�ering lengths which were formed during thesynthesis. Analysis of such a sample results in thedetection of a polymer distribution (see Figure 1),which is typical in the analysis of polymers. For ahomopolymer, adjacent peaks in the polymerdistribution di�er in length by one monomer unit. Ifthe peaks are su�ciently resolved, it may be possibleto calculate the mass of the monomer units. If thesample contains a copolymer, the spectrum canquickly become very complex due to the overlappingnature of the various chain compositions and chainlengths that may be present in the sample. In suchcases, polymer analysis tools which can process thelarge number of peaks and identify the species whichare present are essential.

In this application note, we present results obtainedfollowing the analysis of two polymer samples. Theexamples shown include MALDI spectra obtained fora homopolymer (PEG) and a more complexcopolymer sample. Results are presenteddemonstrating the Polymer Analysistools forprocessing MALDI polymer data.

Methods

PEG 2000 sample:

All solutions were prepared in 50/50 acetonitrile/0.1%TFA. Polyethylene glycol (average molecular weight =2000) was prepared at a concentration of 10 mg/mL.Sodium tri�uoroacetate (NaTFA) was prepared at10 mg/mL. 20 µL of PEG 2000 solution and 20 µL ofNaTFA solution were mixed together with 160 µL of50/50 acetonitrile/0.1% TFA.

Equal volumes of polymer sample and matrix solution(α -cyano 4-hydroxycinnamic acid; 5 mg/mL in 50/50acetonitrile/0.1% TFA) were mixed and 1 µL wasdeposited onto a stainless steel MALDI target andallowed to air-dry.

EO-PO sample:

The EO-PO sample was provided by Day International(Irlam, Manchester (UK)) and was used as supplied.Equal volumes of sample solution, matrix solution(dithranol; 35 mg/mL in THF) and sodiumtri�uoroacetate (10 mg/mL in THF) were mixed and1 µL was deposited onto a stainless steel MALDItarget and allowed to air-dry.

Samples were analyzed using an AXIMA Performance ™MALDI-TOF MS (Shimadzu Corp.) in positive-ionre�ectron mode. The pulsed extraction (P.Ext) valuewas set to 2000 i.e. delay time optimized for mass2000 Da.

Results

The results of the Polymer Analysissoftwaredemonstrate how it is possible to calculate andreport/visualize distribution information (in polymerdistribution analysis mode) or identify components ofdi�erent copolymers (in copolymer analysis mode),present in a sample.

The PEG 2000 example (Figure 1) shows a typicalMALDI spectrum obtained during analysis of asynthetic homopolymer sample. Adjacent peaks di�erin mass by one monomer unit. Measuring the massdi�erence between adjacent peaks of the same seriesgives the mass of the monomer unit. In the case ofFigure 1, the observed mass di�erence is 44.03,consistent with the expected monomer mass(calc. = 44.03).

Page 3: Analysis of Polymers by MALDI-TOF Mass Spectrometry MO329

If the chemical formula of monomer is known, it can be entered, along with the cation/salt used to prepare thesample, into the Polymer analysis software and used to generate a theoretical polymer distribution. This theoretical distribution can then be compared with the experimantally obtained MALDI spectrum (see �gure 4)

0

10

20

30

40

50

60

70

80

90

100

%Int.

1400 1500 1600 1700 1800 1900 2000 2100 2200 2300 2400 2500 2600 2700 2800 2900m/z

2066.272022.24

2198.36

1890.162242.381846.13

2286.431802.11

2330.461758.082374.49

1714.052418.50

1670.04 2462.54

1581.97 2550.611449.90 2682.72

0

10

20

30

40

50

60

70

80

90

100

%Int.

2050 2060 2070 2080 2090 2100 2110 2120 2130m/z

2066.27 2110.30

44.03

Figure 1: MALDI-MS spectrum obtained for polyethyleneglycol (PEG), average molecular weight = 2000. Thesample was prepared in CHCA with the addition ofNaTFA. Inset: Zoom view showing expanded mass range(2050 - 2130 m/z) of the spectrum. The mass di�erencebetween adjacent peaks gives the mass of the monomerunit (for PEG, ΔM = 44.03 (CH 2CH 2O)).

Figure 2: Polymer Analysis window (distribution mode).The Polymer Analysis software is part of the MALDI-MSAXIMA control software and requires a licence key toaccess this module.

0

10

20

30

40

50

60

70

80

90

100

%Int.

1000 1200 1400 1600 1800 2000 2200 2400 2600 2800 3000m/z

1492.04 1726.17

1610.10 1784.211434.011844.231667.151375.971858.24

1494.03

2004.371317.94 1538.042078.40

1259.911027.76

1201.872208.521377.98

2252.571319.93

1086.80

1029.73 2400.61

Figure 5: MALDI-MSspectrum obtained for theEO-PO sample. Thesample was prepared indithranol with the additionof NaTFA.

0

20

40

60

80

100

%Int.Resolution: 2000

2040 2050 2060 2070 2080 2090 2100 2110 2120 2130 2140m/z

Figure 4: Overlay oftheoretical data (black)and experimentalspectrum (red). (a) massrange = 1400 - 2900 m/z;(b) mass range = 2040 –2140 m/z.In the polymercalculations, the masseswere considered asmonoisotopic.

Figure 3: Distribution calculation results followingprocessing (analysis) of the spectrum shown in Figure 1using the Polymer Analysis software.

0

20

40

60

80

100

%Int.Resolution: 2000

1400 1500 1600 1700 1800 1900 2000 2100 2200 2300 2400 2500 2600 2700 2800 2900m/z

The spectrum shown in Figure 5 was obtained duringthe analysis of a more complex copolymer sample.The zoom view of this spectrum (Figure 6) highlightsthe complexity of the spectrum compared with thatobtained for the homopolymer sample (Figure 1). Theincreased spectrum complexity arises from theoverlapping series corresponding to the di�erent polymer chains present in the sample. Attempting tomanually interpret this data would be a challengingand time consuming task. However using the PolymerAnalysis software in copolymer analysis mode, themasses of di�erent copolymer combinations can be automatically calculated, facilitating the assignment of peaks ( see Figure 8).

Analysis of the MALDI data using the Polymer Analysissoftware (in distribution analysis mode; Figure 2)generated the results shown in Figure 3. The PolymerAnalysissoftware automatically detected a singledistribution (in the mass range 1431 to 2713 m/z). Theaverage calculated monomer mass (calculated by thesoftware) is 44.03, consistent with the PEG monomer.In addition, the residual mass calculated by the software (see Figure 3) is 18.0, consistent with H- and -OH as endgroups

(a) (b)

Page 4: Analysis of Polymers by MALDI-TOF Mass Spectrometry MO329

www.shimadzu.com/an/

Conclusion

MALDI-TOF Mass Spectrometry is a valuable tool for the analysis of synthetic polymers

Polymer Analysis software can be used to facilitate the interpretation of MALDI data derived frompolymer samples

The software can be used to identify polymer distributions or for the analysis of copolymers, facilitatingthe assignment of peaks or ions belonging to the same series

© August 2013 Shimadzu Corp E & OE Ref: MO329 Issue 2

AXIMA is a trademark of Kratos Analytical Ltd., Manchester, UK

0

20

40

60

80

100

%Int.Resolution: 5488

1735 1740 1745 1750 1755 1760 1765 1770 1775 1780m/z

H.(EO)2.(PO)28.OH.Na

Figure 7: Polymer Analysis software window showing thecalculated distributed IDs, chemical formulae, theoreticaland observed masses and errors, calculated within thespeci�ed range of values (copolymer analysis mode). Thehighlighted line corresponds to the distribution shown inthe upper trace of Figure 6.

0

20

40

60

80

100

%Int.Resolution: 5564

1735 1740 1745 1750 1755 1760 1765 1770 1775 1780m/z

H.(EO)6.(PO)25.OH.Na

Figure 8: Examples showing assignment of overlapping peaks from a complex copolymer sample. Red trace: MALDImass spectrum obtained for the copolymer sample; Black trace: theoretical distribution corresponding to (a) peak at1756.2 m/z (H.(EO) 6(PO) 25.OH.Na) and (b) peak at 1754.2 m/z (H.(EO) 2(PO) 28.OH.Na)

0

20

40

60

80

100

%Int.Resolution: 5000

1000 1200 1400 1600 1800 2000 2200 2400 2600 2800 3000m/z

H.(PO)26.OH.Na

0

20

40

60

80

100

%Int.Resolution: 5000

1500 1510 1520 1530 1540 1550 1560 1570 1580 1590 1600m/z

H.(PO)26.OH.Na

Figure 6: MALDI mass spectrum obtained for thecopolymer sample (red trace) and theoretical peakcorresponding to H.(PO) 26.OH.Na (black trace).Inset: Zoom view showing expanded mass range(1500 - 1605 m/z) of the MALDI-MS spectrum shown inFigure 5 with the calculated distribution for the peak at1550 m/z. In the polymer calculations, the masses wereconsidered as monoisotopic.

)b()a(

For Research Use Only. Not for use in diagnostic proceduresThe contents of this publication are provided to you “as is” without warranty of any kind and are subject to change without notice.Shimadzu does not assume any responsibility or liability for any damage whether direct or indirect, relating to the use of this publicationSHIMADZU Corporation

www.shimadzu.com/an/

Second Edition: August, 2013

For Research Use Only. Not for use in diagnostic proceduresThe contents of this publication are provided to you “as is” without warranty of any kind and are subject to change without notice.Shimadzu does not assume any responsibility or liability for any damage whether direct or indirect, relating to the use of this publication