Page 1: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

USING SOLID PHASE EXTRACTION AS A PRE-

CONCENTRATION TECHNIQUE

ANIZAH BINTI MAHMOD

UNIVERSITI TEKNOLOGI MALAYSIA

Page 2: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE USING SOLID PHASE EXTRACTION AS A PRE-

CONCENTRATION TECHNIQUE

ANIZAH BINTI MAHMOD

A dissertation submitted in partial fulfilment of the

requirements for the award of the degree of

Master of Science (Forensic Science)

Faculty of Science

Universiti Teknologi Malaysia

APRIL 2016

Page 3: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

To my family, friends and nation

Page 4: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

x

LIST OF TABLES

TABLE NO. TITLE PAGE

2.1 Statistics of Explosive Cases in Malaysia 6

2.2 Reasons of the criminals to use explosives 7

4.1 Limit of detection and correlation coefficient of analytes 42

under study

4.2 Relative standard deviation for within-day and day-to-day for 43

each analyte

4.3 Pre-concentration factor for each analyte 50

4.4 Concentration of cations in pyrotechnic samples 53

Page 5: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

xi

LIST OF FIGURES

FIGURE NO. TITLE PAGE 2.1 SPE operations 20

2.2 SPE adsorbent packing for ion-exchange 21

2.3 Schematic diagram of CE system 24

2.4 Basic IC components 26

3.1 Physical appearance of all explosive samples 35

3.2 Flow chart for experimental design 38

4.1 IC chromatogram for 1 mg/L standard cations 40

4.2 Calibration graph of the cations for 1 to 5 mg/L of

standard solution 41

4.3 Effect of flow rate on SPE recoveries of target cations using

Supelco MCAX, benzenesulphonic acid (H+) 500 mg cartridge 45

4.4 Effect of eluent type using Supelco MCAX, benzenesulphonic

acid (H+) 500 mg cartridge 46

4.5 Effect of the eluent volume using Supelco MCAX,

benzenesulphonic acid (H+) 500 mg cartridge 48

Page 6: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

xii

4.6 IC chromatogram of blank SPE extract 51

4.7 Comparison of IC profiles of pre-blast samples for (a) MB,

(b) MK, (c) MS and (d) HB 52

4.8 IC chromatogram of target cations in (a) post-blast with

cotton swab method and (b) post-blast with cotton

swab followed by SPE method of MB sample 56

4.9 IC chromatogram of target cations in (a) post-blast with

cotton swab method and (b) post-blast with cotton

swab followed by SPE method of MK sample 57

4.10 IC chromatogram of target cations in (a) post-blast with

cotton swab method and (b) post-blast with cotton

swab followed by SPE method of MS sample 58

4.11 IC chromatogram of target cations in (a) post-blast with

cotton swab method and (b) post-blast with cotton

swab followed by SPE method of sample HB 59

Page 7: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

xiii

LIST OF ABBREVIATIONS

CE – Capillary Electrophoresis

DDDW – Double Distilled Deionized Water

FTIR

– Fourier Transform Infrared

GC

– Gas Chromatography

HPLC

– High Performance Liquid Chromatography

IC – Ion Chromatography

IED

– Improvised Explosive Device

MCAX

– Mixed Cation Acidic Exchange

ND – Not detected

RSD – Relative Standard Deviation

RMP – Royal Malaysia Police

SPE – Solid Phase Extraction

SPME – Solid Phase Microextraction

Page 8: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

xiv

LIST OF APPENDICES

APPENDIX TITLE PAGE

A Abstract for National Convention Forensic Science 2015 68

Page 9: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

iv

ACKNOWLEDGEMENT

In completion of my dissertation, I wish to thank the supervisor, Dr Umi

Kalthom binti Ahmad for her support and guidance throughout the years. I would

like to thank my co-supervisor, DSP Satwant Singh A/L Karam Singh for his help

and support. I am truly indebted to both of you.

Special thanks to the lab staffs, Madam Mariam Hasan and Madam Linda for

their endless support and for all the opportunities and resources they has given me in

order for me to complete this dissertation. Thank you for everything, both of you

have taught me about IC and answering all of my questions.

Thanks to my sponsor, MyBrain15 under Ministry of Education for financial

aids. I appreciate my family that always give supports through my ups and downs.

Thanks to all my friends especially my classmates for your help, supports and joy

during my graduate studies.

Page 10: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

v

ABSTRACT

Solid Phase Extraction (SPE) is a very powerful technique for clean-up and

pre-concentration of the sample. Even though there is no universal method for SPE,

there have been an increasing demand for a reliable and accurate forensic trace

analysis. Samples received by a forensic laboratory are often in small amounts and

there is a need to develop methods that are capable to detect lower concentration. The

objectives of this study were to develop, evaluate and compare an SPE procedure as a

pre-concentration step with conventional cotton swabbing method for analysis of ions

in explosive residue. Four samples of pyrotechnics were analysed in this study. The

samples were Mercun Bola (MB), Mercun Ketupat (MK), Happy Bomb (HB) and

Monster Bomb (MS). The results of this study showed that all six standard cations

(lithium, sodium, ammonium, potassium, magnesium and calcium) were successfully

separated within 45 minutes using Metrosep C4 column (250 x 4 mm i.d) with 5 µm

particle size. The eluent used were 0.7 mmol/L dipicolinic acid and 1.7 mmol/L nitric

acid with a flow rate of 1.0 mL/min and pressure of 9.5 MPa. The calibration graph

gave a good linearity with r2 ≥ 0.992. The detection limits were 0.32 µg/L to 1.32 µg/L

and quantitation limits in the range of 1.26 µg/L to 4.4 µg/L. The SPE method was

optimized using several parameters before the method was applied to post-blast

residues. The SPE cartridge used was Supelco MCAX with 500 mg sorbent bed packed

with sulfonilic functional group. The SPE parameters that have been chosen were

sample flow rate, type of eluents and volume of eluent. The results indicated that SPE

procedure gave good pre-concentration of the analytes of interest with the pre-

concentration factor in the range of 7 to 74.

Page 11: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

vi

ABSTRAK

Pengestrakan fasa pepejal (SPE) adalah suatu teknik pembersihan dan

prapemekatan yang baik. Walaupun tiada kaedah yang universal untuk SPE, keperluan

untuk menghasilkan keputusan analisis forensik yang baik dan tepat sentiasa

diperlukan. Sampel yang diterima oleh makmal forensik kebiasaannya adalah pada

kepekatan yang rendah dan adanya keperluan untuk membina kaedah yang dapat

mengenalpasti kepekatan sisa bahan letupan pada kepekatan yang rendah. Objektif

kajian ini adalah untuk membentuk, menilai dan membuat perbandingan antara SPE

sebagai kaedah prapemekatan terhadap kaedah konvensional sapuan kapas bagi

analisis sisa bahan letupan. Empat sampel piroteknik dianalisis di dalam kajian ini iaitu

Mercun Bola (MB), Mercun Ketupat (MK), Happy Bomb (HB) dan Monster Bomb

(MS). Hasil kajian ini mendapati kesemua enam piawai kation (litium, natrium,

amonium, kalium, magnesium dan kalsium) berjaya dipisahkan dalam masa 45 minit

menggunakan turus Metrosep C4 (250 x 4 mm i.d) dengan saiz zarah 5 µm. Eluen yang

digunakan adalah 0.7 mmol/L asid dipikolinik dan 1.7 mmol/L asid nitrik dengan

kadar aliran 1.0 mL/min dan tekanan 9.5 MPa. Graf penentukuran menunjukkan garis

lurus yang bagus dengan r2 ≥ 0.992. Had pengesanan adalah 0.32 µg/L hingga 1.32

µg/L dan had kuantitasi adalah antara 1.26 µg/L hingga 4.4 µg/L. Kaedah SPE

dioptimumkan dengan beberapa parameter terlebih dahulu sebelum diaplikasikan

kepada residu bahan letupan. Kartrij SPE menggunakan Supelco MCAX dengan 500

mg penjerapan dan kumpulan berfungsi asid sulfonilik. Parameter yang dipilih adalah

kadar aliran sampel, jenis eluen dan isipadu eluen. Keputusan menunjukkan prosedur

SPE memberikan analisis yng bagus dengan faktor prapemekatan antara 7 hingga 74.

Page 12: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

vii

TABLE OF CONTENTS

CHAPTER TITLE PAGE

DECLARATION ii

DEDICATION iii

ACKNOWLEDGEMENT iv

ABSTRACT v

ABSTRAK vi

TABLE OF CONTENTS vii

LIST OF TABLES x

LIST OF FIGURES xi

LIST OF ABBREVIATIONS xiii

LIST OF APPENDICES xiv

1 INTRODUCTION 1

1.1 Background of Study 1

1.2 Problem Statement 2

1.3 Research Objectives 3

1.4 Scope of Study 3

1.5 Significant of Study 3

2 LITERATURE REVIEW 5

2.1 Preamble 5

2.2 Explosives 6

Page 13: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

viii

2.2.1 Classification of Explosive 7

2.2.2 Types of Explosive 8

2.2.2.1 High Explosive 9

2.2.2.2 Low Explosive 10

2.2.2.3 Pyrotechnics 11

2.3 Forensic Sampling of Explosive Residue 14

2.4 Sample Preparation of Explosive Residue 15

2.4.1 SPME 15

2.4.2 Solid Phase Extraction 16

2.4.2.1 Basic Principle of Solid Phase

Extraction 19

2.4.2.2 Types of Solid Phase Extraction 19

2.4.2.3 Off-line vs On-line SPE 21

2.5 Analytical Techniques for Inorganic Analytes 22

2.5.1 Capillary Electrophoresis 22

2.5.2 Ion Chromatography 24

2.5.2.1 Basic Principle of IC 27

2.5.2.2 Suppressor and Non-Suppressor

Ion Chromatography 28

2.5.2.3 Single Column and Double

Column 30

2.6 Trace Detection Challenges 31

3 EXPERIMENTAL 33

3.1 Introduction 33

3.2 Apparatus 33

3.3 Chemicals and Materials 33

3.4 Sample Preparation 34

3.4.1 Explosive Sampling 34

3.4.2 Unexploded Explosive Sample 35

3.4.3 Cotton Swab Method 35

3.4.4 Solid Phase Extraction 36

3.5 Optimization of Experimental Parameters 36

Page 14: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

ix

3.5.1 Effect of Type of Eluent 36

3.5.2 Effect of Eluent Volume 37

3.5.3 Effect of Flow Rate 37

3.6 Preparation of Eluent 37

3.7 Preparation of Standard 37

3.8 Instrumentation 37

3.9 Flow Chart of Research Design 38

4 RESULTS AND DISCUSSION 39

4.1 Introduction 39

4.2 IC Separation of Cations 39

4.3 Calibration Graph and Detection Limits 41

4.4 Method Validation 43

4.5 Optimization of SPE 44

4.5.1 Effect of Sample Loading Flow Rate 44

4.5.2 Effect of Eluent Type 46

4.5.3 Effect of Eluent Volume 47

4.6 Determination of Cation from Spiked Cotton Ball 49

4.7 Identification of Inorganic Cations in Explosives

Sample 50

4.7.1 Analysis of Pre-blast Pyrotechnic Samples 51

4.7.2 Analysis of Post-blast Residue 54

4.7.3 Comparison of Post-Blast Extracts with and

without SPE Preconcentration 59

5 CONCLUSION 61

5.1 Conclusion 61

5.2 Suggestion for Future Work 63

REFERENCES 64

Appendix A 68

Page 15: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

CHAPTER 1

INTRODUCTION

1.1 Background of Study

Identification of ions in post-blast residues at crime scenes generally helps the

investigator to know the type and source of explosives used by the criminals. There

are two types of explosives, which are high and low explosives. Low explosives are

usually modified by using improvised explosive device (IED) to give explosion of high

impact.

There are several techniques of ion determination in ion chromatography (IC);

ion-exchange chromatography, ion-exclusion chromatography, and ion-pair

chromatography. A number of selected cations and anions can be determined by using

ion-exchange chromatography with high sensitivity. For explosives, determining the

type of ions is very valuable in forensic science since it can help the crime scene

investigator to detect the source of the explosives

Cotton swabbing is the most universal method commonly used for sample

extraction in explosives analysis. There are two types of cotton swabbing, which are

dry and wet swab. Water is a good extraction solvent for inorganic substances. There

has been much research that focuses on different kind of solvents for extraction of

organic and inorganic substances rather than finding another method for pre-

concentration of samples. Normally after cotton swabbing, the sample is analysed

directly by using ion chromatography. Detection of explosives types becomes more

Page 16: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

2

challenging with the advent of more IEDs by the criminals. This makes the quest for

the details of the crime more difficult. Hence, more research is needed to solve this

problem.

In most cases, sample preparation steps take up most of the required analysis

time, contributing substantially to the analysis costs (Weiss, 1995). Manipulation of

the sample can falsify the analytical result. Therefore, sample preparation can directly

affect the quality of the results. Pre-concentration of samples by using solid phase

extraction (SPE) has been proven to help in the development of better methods for

analysis of explosives due to its selectivity and sensitivity.

1.2 Problem Statement

The conventional cotton swabbing has several weaknesses in samples

preparation. Sometimes, the amount of samples, such as the real field samples, is not

enough for analysis purpose. Therefore, forensic laboratory needs to develop methods

with low limit of detection (LOD). They are needed to improve detection method for

explosives to get the most accurate results. Currently, there is a lack of analytical

method for inorganic analytes since most research focuses more towards the organic

substances.

Cotton swabbing is still a commonly used method for extraction of inorganic

explosives. However, this method does not clean up the sample and causes the sample

to be not pure. The interferences and unknown peaks may appear in the chromatogram

during analysis, making the data interpretation and peaks identification to become

harder. Hence, a new method that can clean up and remove interferences may be useful

to solve this problem.

Page 17: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

3

1.3 Research Objectives

The objectives of this study are:

i. To develop an SPE procedure as a pre-concentration step for analysis of

cations in explosive residues.

ii. To evaluate the SPE procedure by comparison with the current cotton swab

method.

iii. To apply the developed SPE procedure for post-blast pyrotechnic residues.

1.4 Scope of Study

This study focused on the inorganic components of hand-made low explosives.

Li+, Na+, NH4+, K+, Ca2+, and Mg2+ were detected by using IC. The accuracy of SPE

was tested and compared with the conventional method of cotton swab extraction

protocols.

1.5 Significance of Study

Identification of explosives becomes crucial due to the need of the investigators

to determine its source every time the crimes occur. There are several high and low

explosives that are usually used by the criminals. The SPE method provides fast and

reliable result in determination of various samples including forensic sample. It is

necessary to develop a sample preparation method that enables determination of the

targeted ions at lower concentration level.

This study can be used by Royal Malaysia Police (RMP) to improve the accuracy

of explosive types detection. This is useful for the RMP in tracking the criminal by

getting the accurate data regarding the explosives. Furthermore, profiling and database

are really crucial for RMP due to the increasing number of criminal cases using low

Page 18: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

4

explosives in this country. Further method development is always needed in order to

keep up with the advanced technology and well-educated criminals.

Page 19: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

64

REFERENCES

Akhavan, J. (2007). The Chemistry of Explosives. (2nd Ed). Royal Society of

Chemistry. United Kingdom.

Amin, M., Lim, L. W., and Takeuchi, T. (2007). Tunable Separation of Anions and

Cations by Column Switching in Ion Chromatography. Talanta, 71 (4): 1470–

1475.

Amoli, H. S. M. R. and P. S. (2000). Rapid Analysis of Ultra Trace Amounts of Anions

in Water by Solid Phase Extraction and Ion Chromatography. Indian Journal of

Chemistry, 39(October): 1101–1104.

Barron, L., and Gilchrist, E. (2014). Ion Chromatography-Mass Spectrometry: A

Review of Recent Technologies and Applications in Forensic and Environmental

Explosives Analysis. Analytica Chimica Acta, 806: 27–54.

Beveridge, A. (1998). Forensic Investigation of Explosions. Taylor and Francis.

London.

Brasil, B., Bettencourt da Silva, R. J. N., Camões, M. F. G. F. C., and Salgueiro, P. A.

S. (2013). Weighted Calibration with Reduced Number of Signals by Weighing

Factor Modelling: Application to the Identification of Explosives by Ion

Chromatography. Analytica Chimica Acta, 804: 287–95.

Buchberger, W. W. (2001). Detection Techniques in Ion Chromatography of Inorganic

Ions. Trends in Analytical Chemistry 20: 296–303.

Buszewski, B., and Szultka, M. (2012). Past, Present, and Future of Solid Phase

Extraction: A Review. Critical Reviews in Analytical Chemistry 42(3): 198–213.

Buxton, T. L., and Harrington, B. (2003). Trace Explosive Detection in Aqueous

Samples by Solid- Phase Extraction Ion Mobility Spectrometry (SPE-IMS).

57(2): 223–232.

Camel, V. (2003). Solid Phase Extraction of Trace Elements. Spectrochimica Acta,

58:1177–1233.

Deguchi, K., Kohda, K., and Ito, M. (1999). Simultaneous separation of anions and

cations by flow-gradient ion chromatography. Journal of Chromatography A,

845(1-2): 165–170.

Page 20: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

65

Dicinoski, G. W., Shellie, R. A., and Haddad, P. R. (2006). Forensic Identification of

Inorganic Explosives by Ion Chromatography. Analytical Letters, 39(4): 639–

657.

Duran, A., Tuzen, M., and Soylak, M. (2009). Preconcentration of Some Trace

Elements via using Multiwalled Carbon Nanotubes as Solid Phase Extraction

Adsorbent. Journal of Hazardous Materials, 169: 466–471.

Fritz, J. S and Gjerde, D. T. (2009). Ion Chromatography. (4th Ed). Wiley-VCH.

Germany.

Haddad, P. R., Nesterenko, P. N., and Buchberger, W. (2008). Recent Developments

and Emerging Directions in Ion Chromatography. Journal of Chromatography

A, 1184: 456–473.

Henderson, K., and Anderson, M. (1991). Sample Preparation for Ion

Chromatography by Solid-Phase Extraction. Elsevier Science Publishers B, 546:

61–71.

Hutchinson, J. P., Johns, C., Breadmore, M. C., Hilder, E. F., Guijt, R. M., Lennard,

C., Dicinoski, G., and Haddad, P. R. (2008). Identification Of Inorganic Ions in

Post-Blast Explosive Residues using Portable CE Instrumentation and

Capacitively Coupled Contactless Conductivity Detection. Electrophoresis,

29(22): 4593–602.

Jackson, P. E. (2001). Determination Of Inorganic Ions In Drinking Water By Ion

Chromatography. Trends in Analytical Chemistry, 20: 320–329.

Johns, C., Shellie, R. A, Potter, O. G., O’Reilly, J. W., Hutchinson, J. P., Guijt, R. M.,

Breadmore, M. C., Hilder, E. F., Dicinoski, G.W., and Haddad, P. R. (2008).

Identification of Homemade Inorganic Explosives by Ion Chromatographic

Analysis of Post-Blast Residues. Journal of Chromatography. A, 1182(2): 205–

14.

Jordan, L. (2000). Solid-Phase Extraction: Principles, Techniques, And Applications.

Taylor and Francis. USA.

Kobilinsky, L. (2012). Forensic Chemistry Handbook. John Wiley and Sons. New

Jersey.

Kuhn, R., and Hoffstetter-Kuhn, S. (1993). Capillary Electrophoresis : Principles and

Practice. Springer-Verlag.

Lopez-Ruiz, B. (2000). Advances in the Determination of Inorganic Anions by Ion

Chromatography. Journal of Chromatography A, 881:607–627.

Page 21: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

66

Mccord, B. R., Hargadon, K. A., Hall, K. E., and Burmeister, S. G. (1994). Forensic

Analysis of Explosives using Ion Chromatographic Methods. Analytica Chimica

Acta, 288: 43–56.

Medina, V. F., Larson, S. L., Extine, B., and Bednar, A. (2005). Perchlorate Analysis

using Solid-Phase Extraction Cartridges. Journal of Chromatographic Science,

43(April): 195–200.

Meng, H.-B., Wang, T.-R., Guo, B.-Y., Hashi, Y., Guo, C.-X., and Lin, J.-M. (2008).

Simultaneous determination of inorganic anions and cations in explosive residues

by ion chromatography. Talanta, 76(2): 241–5.

Michalski, R. (2009). Applications of Ion Chromatography for the Determination of

Inorganic Cations. Critical Reviews in Analytical Chemistry, 39(4): 230–250.

Nesterenko, P. N. (2001). Simultaneous Separation and Detection of Anions and

Cations in Ion Chromatography. TrAC - Trends in Analytical Chemistry, 20(6-

7): 311–319.

Royds, D., Lewis, S. W., and Taylor, A. M. (2005). A Case Study in Forensic

Chemistry : The Bali Bombings. Atlanta, 67: 262–268.

Sarzanini, C., and Bruzzoniti, M. C. (2005). New Materials : Analytical and

Environmental Applications n Ion Chromatography. Analytica Chimica Acta,

540: 45–53.

Slingsby, R., and Kiser, R. (2001). Sample Treatment Techniques and Methodologies

for Ion Chromatography. Trends in Analytical Chemistry, 20: 288–295.

Song-im, N., Benson, S., and Lennard, C. (2012a). Establishing a Universal Swabbing

and Clean-Up Protocol for the Combined Recovery of Organic and Inorganic

Explosive Residues. Forensic Science International, 223: 136–147.

Song-im, N., Benson, S., and Lennard, C. (2012b). Evaluation of Different Sampling

Media for Their Potential use as a Combined Swab for the Collection of Both

Organic and Inorganic Explosive Residues. Forensic Science International,

222(1-3), 102–110.

Tachon, R., Pichon, V., Barbe Le Borgne, M., and Minet, J.-J. (2008). Comparison of

Solid-Phase Extraction Sorbents for Sample Clean-Up in the Analysis of Organic

Explosives. Journal of Chromatography. A, 1185(1): 1–8.

Telepchak, M. J., August, T. F., and Chaney, G (2004). Forensic and Clinical

Applications of Solid Phase Extraction. Humana Press. New Jersey.

Page 22: ANALYSIS OF INORGANIC CATIONS IN LOW EXPLOSIVE RESIDUE

67

Türker, A. R. (2012). Separation Preconcentration and Speciation of Metal Ions by

Solid Phase Extraction Separation. Separation and Purification Reviews,

(October 2014): 37–41.

Tyrrell, E., Shellie, R. A, Hilder, E. F., Pohl, C. A, and Haddad, P. R. (2009). Fast Ion

Chromatography using Short Anion Exchange Columns. Journal of

Chromatography. A, 1216(48): 8512–7.

Warren, D., Hiley, R. W., Phillips, S. A., and Ritchie, K. (1999). Novel Technique for

the Combined Recovery , Extraction and Clean-Up of Forensic Organic and

Inorganic Trace Explosives Samples. Science and Justice, 39(1): 11–18.

Weiss, J. (1995). Ion Chromatography. Practical High-Performance Liquid

Chromatography. (2nd Ed). VCH Verlagsgesellschaft. Germany.

Yinon, J. (2000). Explosives. Handbook of Analytical Separations Vol.2: 615–616. Elsevier. Oxford, United Kingdom.