Upload
michael-armstrong
View
213
Download
1
Embed Size (px)
Citation preview
RAPID COMMUNICATIONS IN MASS SPECTROMETRY
Rapid Commun. Mass Spectrom. 2007; 21: 2717–2726
) DOI: 10.1002/rcm.3124
Published online in Wiley InterScience (www.interscience.wiley.comAnalysis of 25 underivatized amino acids in human
plasma using ion-pairing reversed-phase liquid
chromatography/time-of-flight mass spectrometry
Michael Armstrong1, Karen Jonscher2 and Nichole A. Reisdorph1*1Department of Immunology, National Jewish Medical and Research Center, Denver, CO 80206, USA2University of Colorado at Denver and Health Science Center, Clinical Nutrition Research Unit, Department of Anesthesiology, Denver,
CO 80262, USA
Received 1 December 2006; Revised 25 May 2007; Accepted 26 May 2007
*Corresponology, NJackson SE-mail: RContract/Unit; con
Amino acids in biological fluids have previously been shown to be detectable using liquid
chromatography/electrospray ionization mass spectrometry (LC/ESI-MS) with perfluorinated acids
as ion-pairing agents. To date, these studies have used precursor mass, retention time and tandem
mass spectrometry (MS/MS) to identify and quantify amino acids. While this is a potentially
powerful technique, we sought to adapt the method to time-of-flight (TOF)MS. A new application
of a recently described liquid chromatographic separation method was coupled with TOFMS to
employ accurate mass for qualitative identification; resulting in additional qualitative data not
available with standard single quadrupole data. In the current study, we evaluated 25 physiological
amino acids and one dipeptide that are routinely quantified in human plasma. Accuracy and
precision of the method was evaluated by spiking human plasma with a mix of the 25 amino acids;
in addition, the inclusion of a cation-exchange cleanup step was evaluated. The calibration curves
were linear over a range from 1.56 to 400 mM. The dynamic range was found to be within
physiological levels for all amino acids analyzed. Accuracy and precision for most of the amino
acids was between 80–120% spike recovery and<10% relative standard deviation (RSD). The LC/MS
technique described in this study relies on mass accuracy and is suitable for the quantitation of
free amino acids in plasma. Copyright # 2007 John Wiley & Sons, Ltd.
Free amino acid analysis has applications in a variety
of areas, including the diagnosis of inherited metabolic
disorders,1–4 and nutritional studies of neonates.5–8 Tradi-
tionally, free amino acids in plasma have been analyzed by
ion chromatography (IC) using ninhydrin post-column
derivatization,9 or by cation-exchange solid-phase extraction
followed by derivatization and analysis by gas chromato-
graphy/mass spectrometry (GC/MS).10–15 Both of these
methods have disadvantages, including long run times and
extensive sample preparation, respectively.
Although useful for a broad range of compounds, neither
high-performance liquid chromatography (HPLC) nor MS
techniques were generally employed for amino acid analysis
due to the inability to separate more polar amino acids using
reversed-phase (RP)-HPLC,16 and amino acid signal sup-
pression in electrospray ionization (ESI)-MS caused by
co-elution of components in complex biological matrices.17,18
Some approaches to enable the application of LC/MS to
amino acid analysis include dedicated amino acid analysis
kits such as Waters AccQtag,19 and tandem mass spectrom-
ndence to: N. A. Reisdorph, Department of Immu-ational Jewish Medical and Research Center, 1400
treet K924, Denver, CO 80206, [email protected] sponsor: Colorado Clinical Nutrition Research
tract/grant number: NIH/NIDDK P30 DK048520-09.
etry (MS/MS) utilizing flow injection analysis,4 each of
which has certain disadvantages. For example, buffers
included in kits may be incompatible with ESI.
Recently published methods have described the use of
perfluorinated acids as ion-pairing agents to improve the
separation of amino acids on C18 columns without the
requirement for specialty columns or pre-/post-column
derivatization.16,20–23 Piraud et al.23 utilized HPLC and
tandem mass spectrometry (LC/MS/MS) with tridecafluor-
oheptanoic acid (TDFHA) as an ion-pairing agent with a C18
column. While TDFHA improved separation, all amino acids
of interest were not completely resolved by HPLC. Therefore,
multiple reaction monitoring (MRM) was used to improve
the specificity of the method by monitoring specific transi-
tions of precursor to product ions (e.g. glutamine¼147>84).
The method of Piraud et al. was used to quantitate 76 amino
acids of biological interest and the quantitation of 16 amino
acids was validated.23
In the current study, we have adapted Piraud’s method of
ion-pairing reversed-phase liquid chromatography (IPRP-
LC) for use with an electrospray ionization time-of-flight
(ESI-TOF) mass spectrometer equipped with an analog-
to-digital converter (ADC) for signal processing. Sample
Copyright # 2007 John Wiley & Sons, Ltd.
2718 M. Armstrong, K. Jonscher and N. A. Reisdorph
preparation involves precipitation of proteins from plasma
using methanol fortified with stable isotope labeled internal
standards followed, in some cases, by cation exchange.
Extremely accurate mass measurements (approximately
�1–10 ppm) obtained with a TOF were used to deduce the
identity of an analyte with a much higher degree of certainty
than a standard, high-resolution single quadrupole mass
spectrometer. With the exception of isobaric molecules such
as leucine and isoleucine, the mass-to-charge (m/z) of amino
acids can be verified to within 1–5 ppm, reducing mis-
identification of target amino acids in the presence of
co-eluting matrix components of similar molecular weight.
We found that IPRP-LC/ESI-TOF provides a quick, simple,
reproducible alternative to MS/MS analysis and used this
technique for the analysis of 25 amino acids in human
plasma. The method described in this study uses minimal
amounts of standards, reagents, and sample, can be applied
to any amino acid that ionizes by ESI, and can easily be
adapted to high-throughput sample analysis.
EXPERIMENTAL
ReagentsNanopure water (18.2VOhms) was used for sample pre-
paration. Water (HPLC grade) and acetonitrile (UV) used for
HPLC mobile phases was obtained from Burdick and
Jackson (Morristown, NJ, USA). HPLC-grade methanol
was obtained from Fisher Scientific (Hampton, NH, USA).
Tridecafluoroheptanoic acid (TDFHA) was obtained from
Aldrich Chemicals (St. Louis, MO, USA). Hydrochloric acid
was obtained from Sigma (St. Louis, MO, USA). The primary
amino acid calibration standard at 2.5 mM (standard ’H’) was
obtained from Pierce (Rockford, IL, USA). Hyp, Gly, Glu,
Ala, Trp, Tau, Asn, Gln, Cit, Ala-Glu, Nor and Orn were
obtained from Sigma. Stable isotope labeled analogs of amino
acids used as internal standards (glutamine-d5, glutamic
acid-d3, methionine-d3, leucine-d10 and tryptophan-d5)
were obtained from Cambridge Isotope Laboratories (And-
over, MA, USA). Outdated human blood plasma was
provided by Bonfil’s Blood Center (Denver, CO, USA).
The use of outdated plasma samples for method validation
and quality control purposes was considered exempt by the
Colorado Multiple Institutional Review Board (COMIRB).
Standards preparation procedureAmino acid calibration and spike standards were prepared at
physiological concentration ranges from pure powder or
commercially available standards. Amino acid mix #1
contained amino acids which are stable in 0.1% hydrochloric
acid solution such as the branched chain amino acids and the
hydroxyl group containing amino acids, including Asp, Hyp,
Ser, Gly, Thr, Glu, Ala, (Cys)2, Pro, Cys, Val, Met, Tyr, Ile,
Leu, Phe, His, Trp, Arg, and Lys. Amino acid mix #2
contained the amino acids which are not stable in an acid
solution, such as Tau, Gln, Asn, Cit, Ala-Gln and Orn. All
calibration stocks and working standards were stored at
�208C until use.
In addition to calibration standards, two separate internal
standard (IS) mixes were used to quantitate. IS mix #1
contained glutamine-d5 and methionine-d3 in water. IS mix
Copyright # 2007 John Wiley & Sons, Ltd.
#2 contained leucine-d10, glutamic acid-d3 and tryptop-
han-d5 in 0.1% hydrochloric acid. The internal standard
working solution was prepared immediately prior to sample
preparation, by adding equal parts of IS #1 and IS #2 to eight
parts methanol (1:1:8). All internal standard stocks and IS
mix #1 and #2 were stored at �208C until use.
Pooled human plasma samples, used for accuracy and
precision measurements, were spiked with amino acid
calibration mixes 1 and 2 (100 mM) and frozen at �808Cfor 1–2 days prior to thawing for extraction and analysis.
Sample preparation procedureCalibration standards were prepared by combining 10mL
each of amino acid mixes 1 and 2 and 100mL of IS working
solution. Standards were vortexed briefly and then centri-
fuged at 10 000 g for 5 min at 48C. An aliquot (70mL) of
supernatant was transferred to a 96-well plate or HPLC vial
containing 30mL of 1.7 mM TDFHA in water, providing a
final concentration of 0.5 mM TDFHA.
Samples were prepared by adding 20mL of plasma to
100mL of IS working solution and briefly vortexing. Samples
were then centrifuged at 10 000 g for 5 min at 48C, resulting in
a protein precipitate that was subsequently discarded. An
aliquot (70mL) of supernatant was transferred to a 96-well
plate or HPLC vial containing 30mL of 1.7 mM TDFHA in
water.
Solid-phase extractionFor some samples, cleanup was performed via solid-phase
extraction (SPE) using a cation-exchange cartridge. Strata
X-C cartridges with a capacity of 30 mg (Phenomenex,
Torrance, CA, USA) were placed on a vacuum SPE manifold,
conditioned with 1 mL of methanol, then equilibrated with
1 mL of 0.1 N HCl in water, as per the manufacturer’s
protocol. Subsequently, 100mL of plasma was mixed by
vortexing with 100 mL of the IS working solution prepared in
0.2 M HCl. The entire sample was then loaded onto the SPE
cartridge and drawn through by vacuum. Afterwards, the
cartridge was washed with 1 mL of methanol, and sample
was eluted into a new test tube using 5% ammonium
hydroxide in methanol. The eluate pH was neutralized by
vacuum evaporation of the ammonium hydroxide. Samples
were then lyophilized to dryness and reconstituted with
100 mL of 50 mM TDFHA in 1:1 methanol/water prior to
analysis. The final volume results in a 5-fold increase in
sample over the samples not extracted by SPE.
High-performance liquid chromatographyLiquid chromatography was carried out using an Agilent
1100 series HPLC system equipped with a binary pump and
a micro wellplate autosampler (Agilent Technologies, Palo
Alto, CA, USA). Amino acids were separated using an
XDB-C18 column (2.1� 50 mm) with a 1.8mM particle size
(Agilent Technologies) operated at ambient temperature.
Buffer A was 0.5 mM TDFHA in HPLC-grade water, and
buffer B was 100% acetonitrile. The initial flow rate was
0.2 mL/min. Separation was accomplished using a gradient
as follows: 0% B for 2 min, then 0% to 15% B from 2 to 3 min,
hold at 15% B from 3 to 8 min, then 15% to 25% B from 8 to
11 min. The column was held at 25% B from 11 to 18 min, and
Rapid Commun. Mass Spectrom. 2007; 21: 2717–2726
DOI: 10.1002/rcm
Analysis of 25 underivatized amino acids in human plasma 2719
then returned to 0% B from 18 to 19 minutes. The flow rate
was then increased to 0.4 mL/min from 19.01 to 29 min to
recondition the column. The flow rate was then returned to
0.2 mL/min at 29.01 min, and allowed to equilibrate for
3 min. A long reconditioning time and re-equilibration time
is required to obtain consistent retention times with this
column. The column was flushed with 100% acetonitrile for
1 h after every 30 injections to wash off accumulated
non-target analytes. Failing to flush the column with
acetonitrile after 30 injections will result in a degradation
in chromatography and retention time drift.
ESI-TOFMSDetection of amino acids was accomplished using an
Agilent 1969 orthogonal TOF mass spectrometer coupled to
a positive ESI source with dual spray needles for continuous
infusion of reference mass solution. Heated (3508C) drying
gas flowing at 9.0 L/min, with a nebulizer pressure of 40
psig, was used for droplet desolvation. Spray was induced
with a capillary voltage of 3000 V and the fragmentor
voltage was 100 V. The TOF was tuned and calibrated using
Agilent ESI-TOF calibration and tuning mix (Agilent Tech-
nologies). The data acquisition mass range was 50–350m/z at
10058 transients/scan and 0.93 scans/s. Reference mass
correction on each sample was performed with a continuous
infusion of Agilent TOF biopolymer analysis mix contain-
Table 1. List of experimental parameters for amino acids and si
injection
CompoundMolecularformula
Exact mass[MþH]
Extractedion window
L
Taurine C2H7NO3S 126.0224 126.00–126.04Aspartic acid C4H7NO4 134.0453 134.01–134.05Hydroxyproline C5H9NO3 132.0660 132.03–132.07Serine C3H7NO3 106.0504 106.02–106.06Glycine C2H5NO2 76.0398 76.01–76.05Glutamine-d5� C5H5D5N2O3 152.1000 152.05–152.15Glutamine C5H10N2O3 146.0769 147.04–147.08Asparagine C4H8N2O3 133.0613 133.03–133.07Threonine C4H9NO3 120.0660 120.03–120.07Glutamic acid-d3� C5H6D3NO4 151.1000 151.05–151.15Glutamic acid C5H9NO4 148.0609 148.03–148.07Alanine C3H7N02 90.0555 90.02–90.06(Cysteine)2 C6H12N2O4S2 241.0316 241.01–241.05Citrulline C6H13N3O3 176.1035 176.01–176.05Proline C5H9NO2 116.0711 116.04–116.08Gly-Gln� C7H13N3O4 204.2000 204.10–204.30Ala-Gln C8H15N3O4 218.1140 218.08–218.12Valine C5H11NO2 118.0868 118.05–118.09Methionine-d3� C5H8D3NO2S 153.1000 153.05–153.15Methionine C5H11NO2S 150.0588 150.03–150.07Tyrosine C9H11NO3 182.0817 182.05–182.09Isoleucine C6H13NO2 132.1024 132.08–132.12Leucine-d10� C6H3D10NO2 142.2000 142.00–142.30Leucine C6H13NO2 132.1024 132.08–132.12Phenylalanine C9H11NO2 166.0868 166.06–166.10Histidine C6H9N3O2 156.0773 156.05–156.08Tryptophan C11H12N2O2 205.0977 205.08–205.12Tryptophan-d5� C11H7D5N2O2 210.1000 210.00–210.30Arginine C6H14N4O2 175.1195 175.09–175.13Ornithine C5H12N2O2 133.0977 133.08–133.12Lysine C6H14N2O2 147.1133 147.09–147.13
� Internal standard.
Copyright # 2007 John Wiley & Sons, Ltd.
ing purine (m/z 121.050873) and hexamethoxyphosphazine
(m/z 322.048121) (Agilent Technologies) at 20mL/min.
Ions monitored for quantitationIons monitored for quantitation (see Table 1) were extracted
using Analyst QS software (Applied Biosystems, Foster City,
CA). Signals from internal standards were extracted with a
window ranging from �0.05 to 0.15 Da, while target amino
acids were provided a �0.02 Da extraction window.
Calibration curvesCalibration curves for each amino acid were constructed
using Analyst QS software and prepared so all amino acids
would be within expected physiological concentrations.
Most amino acids were calibrated from 1.56 to 400mM. The
more abundant amino acids (Gln, Glu, Gly and Ala) were
calibrated from 25 to 3200mM (see Table 2). Analyst QS was
used to choose the best fit for the calibration curve. Either a
quadratic or linear fit was applied to quantify most amino
acids.
Method accuracy and precisionTo test method accuracy and precision, pooled human
plasma was analyzed unspiked and spiked at 100 mM
(nominal) of each amino acid. Intra-day accuracy and
precision (n¼ 5) and inter-day accuracy and precision
(n¼ 3) were calculated.
gnal-to-noise (S/N) ratios obtained from a 125pm (nominal)
ow cal std(nM/mL)
High calstd (nM/mL)
S/N ratio(pM injected)
IS usedfor quantitation
1.56 400 851 (125) Glutamine-d51.56 400 52.5 (125) Glutamic acid-d31.56 400 1750 (125) Glutamine-d51.56 400 267 (125) Glutamic acid-d325 3200 22.3 (3125) Glutamic acid-d3
NA NA 637 (1000) NA25 3200 1450 (3125) Glutamine-d5
1.56 400 94.9 (125) Glutamine-d51.56 400 315 (125) Glutamic acid-d3NA NA 15.2 (1000) NA12.5 1600 714 (1562) Glutamic acid-d312.5 1600 345 (125) Leucine-d101.56 400 1160 (125) Methionine-d31.56 400 383 (125) Glutamine-d51.56 400 355 (125) Glutamine-d5NA NA 2560 (1000) NA1.56 400 508 (125) Gly-Gln1.56 400 163 (125) Leucine-d10NA NA 1810 (1000) NA1.56 400 737 (125) Methionine-d31.56 400 722 (125) Leucine-d101.56 400 340 (125) Leucine-d10NA NA 1450 (1000) NA1.56 400 228 (125) Leucine-d101.56 400 1010 (125) Leucine-d101.56 400 1090 (125) Tryptophan-d51.56 400 383 (125) Tryptophan-d5NA NA 1110 (1000) NA1.56 400 1700 (125) Tryptophan-d51.56 400 544 (125) Tryptophan-d51.56 400 448 (125) Tryptophan-d5
Rapid Commun. Mass Spectrom. 2007; 21: 2717–2726
DOI: 10.1002/rcm
Table 2. Amino acids and internal standards used for quantitation, including the curve fit used and the correlation coefficient
obtained
Amino acid
Without cation exchange With cation exchange
IS Curve fit R2 IS Curve fit R2
Taurine (TAU) Glutamine-d5 Quadratic 1 Glutamine-d5 Linear 0.9965Aspartic acid (ASP) Glutamic acid-d3 Quadratic 0.9997 Glutamine-d5 Quadratic 0.9996Hydroxyproline (HYP) Glutamine-d5 Quadratic 1 Glutamine-d5 Linear 0.9994Serine (SER) Glutamic acid-d3 Linear 0.9986 Glutamine-d5 Quadratic 0.9999Glycine (GLY) Glutamic acid-d3 Quadratic 0.9996 Glutamine-d5 Linear 0.9978Glutamine (GLN) Glutamine-d5 Quadratic 1 Glutamine-d5 Linear 0.9996Asparagine (ASN) Glutamine-d5 Quadratic 1 Glutamine-d5 Linear 0.9992Threonine (THR) Glutamic acid-d3 Quadratic 0.9997 Glutamine-d5 Quadratic 0.9999Glutamic acid (GLU) Glutamic acid-d3 Quadratic 0.9998 Glutamate-d3 Linear 0.9992Alanine (ALA) Leucine-d10 Quadratic 0.9997 Leucine-d10 Linear 0.9993(Cysteine)2 Methionine-d3 Quadratic 0.9999 Glutamine-d5 Linear 0.9999Citrulline (CIT) Glutamine-d5 Quadratic 1 Glutamine-d5 Linear 0.9988Proline (PRO) Glutamine-d5 Quadratic 0.9999 Glutamine-d5 Linear 0.9984Valine (VAL) Leucine-d10 Quadratic 0.9999 Leucine-d10 Linear 0.9968Methionine (MET) Methionine-d3 Linear 0.9999 Methionine-d3 Linear 0.9997Tyrosine (TYR) Leucine-d10 Linear 0.9999 Leucine-d10 Linear 0.999Isoleucine (ISO) Leucine-d10 Linear 0.9999 Leucine-d10 Linear 0.9989Leucine (LEU) Leucine-d10 Quadratic 1 Leucine-d10 Linear 0.9994Phenylalanine (PHE) Leucine-d10 Quadratic 0.9999 Leucine-d10 Linear 0.997Histidine (HIS) Leucine-d10 Linear 0.9999 Tryptophan-d5 Linear 0.9999Tryptophan (TRP) Tryptophan-d5 Quadratic 1 Tryptophan-d5 Linear 0.9986Arginine (ARG) Leucine-d10 Quadratic 0.9999 Tryptophan-d5 Linear 0.995Ornithine (ORN) Leucine-d10 Quadratic 0.9997 Tryptophan-d5 Linear 0.9997Lysine (LYS) Leucine-d10 Quadratic 0.9998 Tryptophan-d5 Linear 0.9998
2720 M. Armstrong, K. Jonscher and N. A. Reisdorph
Intra-day accuracy and precision (n¼ 5) was also
measured on samples that were prepared using SPE prior
to analysis.
RESULTS
Chromatographic separation of amino acidsA brief comparison was conducted using Agilent XDB-C18
2.1� 50 mm columns with either 1.8 or 3.5mm solid-phase
particle sizes. Under identical gradient conditions the 1.8mm
column showed superior resolution of the early eluting polar
amino acids (data not shown). An attempt was made to
improve the separation efficiency of the 3.5mm column using
different gradient profiles and flow rates; however, the
1.8mm column still appeared to provide the best separation
(data not shown).
Although the ESI-TOF provides excellent specificity via its
high mass accuracy, there are still instances where complete
or partial chromatographic resolution must be obtained for
accurate quantitation. One example is when an isotopomer,
or m þ n (where n¼ the number of Daltons the ion is shifted
from the m þ 0 ion), in the mass spectrum of a compound
adds to the m þ 0 area of another compound (e.g. Gln and
Glu) as a result of co-elution. Another is the differentiation of
isobaric compounds such as Leu and Ile. The chromato-
graphic resolution obtained using the 1.8mm column was
>90%, separating Gln from Glu, and allowing for complete
resolution of Ile and Leu isobars.
When the final gradient was optimized and established, all
of the amino acids eluted within 16.5 min (Fig. 1). A relatively
long column re-equilibration time resulted in a total cycle
Copyright # 2007 John Wiley & Sons, Ltd.
time of 32 min. While this method is an improvement in
throughput and specificity over traditional amino acid
analysis methods, the throughput could be almost doubled
by using a quaternary or additional HPLC pump, a
column-switching module, and an additional column to
alternate column regeneration and sample analysis.
Integration reproducibility and signal-to-noiseratioThe ability of the ESI-TOF to maintain consistent mass
accuracy and peak integration over time was assessed. The
extracted ion chromatograms for glutamine in five replicate
spiked plasma samples were integrated and compared
(Fig. 2). The mass window for each replicate sample was
m/z 147.04–147.08. The relative standard deviation (RSD)
over the five replicates was 7.29, showing good sample-
to-sample integration reproducibility.
Signal-to-noise (S/N) ratios were also calculated for all
amino acids (See Table 1). Most S/N ratios were greater than
200:1, with Gly being the lowest at 22:1 and Hyp being the
highest at 1750:1.
Matrix interference in plasma samplesThere were significant differences in retention times for
amino acids from extracted standards versus plasma
samples, particularly for the later eluting compounds.
Retention times for amino acids eluting after 4 min were
shifted to as much as 1.5 min earlier in plasma (e.g. Ile and
Leu). The retention time shift in plasma samples did not
typically result in a decreased chromatographic resolution
except for the peak shape of Orn, which was significantly
Rapid Commun. Mass Spectrom. 2007; 21: 2717–2726
DOI: 10.1002/rcm
Figure 1. Extracted ion chromatograms for 25 amino acids. Amino acid calibration and spike standards were prepared as
described in the Experimental section and analyzed according to the parameters listed in Table 1. Overlaid extracted ion
chromatograms of all 25 amino acids in a 400 nM/mL (nominal concentration) standard are shown. Note the separation
between the isobaric amino acids leucine and isoleucine. All peaks are displayed to scale.
Analysis of 25 underivatized amino acids in human plasma 2721
broadened. The peak shape for Orn was significantly worse
in plasma samples (data not shown).
In order to determine if this retention time shift was due to
column overloading, 10, 5 and 2mL of a spiked plasma
sample were loaded onto the column, and the retention time
for leucine-d10 was compared to that of the calibration
standard. The data show that a reduction in the amount of
sample loaded decreased the shift in retention times,
suggesting that the column is indeed overloaded. However,
decreasing the sample load significantly raised the lower
limit of quantitation for several compounds.
Ion suppression from TDFHA adductsWhile examining data acquired in a wider mass range to
determine the major source of column overload, a very large
peak at a retention time of 8–9 min was observed that could
Figure 2. Glutamine integration reproduciblity. Overlay of glutam
five replicates of spiked plasma, showing sample-to-sample in
Copyright # 2007 John Wiley & Sons, Ltd.
not be detected when using the normal acquisition
parameters. This peak was almost non-existent in calibration
standards, but appeared at an extremely high abundance in
plasma samples. Using the accurate mass obtained from the
ESI-TOF data, the empirical formula for the most abundant
ion in the spectrum was calculated to be C7O2F13Na2. This
empirical formula corresponds to a sodium adduct of
tridecafluoroheptanoate, a product of the binding of sodium
salt in plasma with the ion-pairing agent in the aqueous
buffer. The exact reference corrected mass of this ion was
m/z¼ 408.9483, with a theoretical mass of m/z¼ 408.9486 (see
Fig. 3(A)). The less abundant ions in the spectrum
corresponded to clusters of this compound with
additional C7O2F13Na (m/z¼ 385.9647) subunits.
Alanyl-glutamine (Ala-Gln) dipeptide and Val co-eluted
with this peak, which significantly suppressed the signal of
ine extracted ion chromatograms (m/z 147.04–147.08) of
tegration reproducibility.
Rapid Commun. Mass Spectrom. 2007; 21: 2717–2726
DOI: 10.1002/rcm
Figure 3. A tridecafluoroheptanoate (TDFHA) adduct elutes between 8 and 9min and interferes with the analysis of
Ala-Gln and Val. Plasma samples were spiked with internal standards and analyzed as described in the Experimental
section. The mass spectrum of a large peak eluting between 8 and 9min (A) and extracted ion chromatograms of TDFHA
adduct, Ala-Gln and Val (B) are shown. The exact mass of m/z 408.948579 corresponds to the molecular formula of
disodium tridecafluoroheptanoate, an adduct formed between the TDHFA ion-pairing agent and sodium in plasma. The
Ala-Glu dipeptide and Val are almost completely obscured by the adduct (B).
2722 M. Armstrong, K. Jonscher and N. A. Reisdorph
both analytes (Fig. 3(B)). In an effort to separate Ala-Gln and
Val from this peak, we experimented with (a) increasing the
hold at 15% B from 5 to 8 min, (b) changing the hold from 15%
B to 12 % B, and (c) changing the hold from 12% B to 10% B.
None of these changes improved the separation of Ala-Gln
and Val from the TDHFA adduct peak enough to reduce ion
suppression.
SPE cleanupTo eliminate the TDFHA adduct and additional non-target
compounds from the sample extract, a cation-exchange
cleanup step was performed on a series of calibration
standards, unspiked plasma, and spiked plasma. Although
slightly more time-consuming, the improvement in chro-
matographic performance provided by the cation-exchange
cleanup was significant. Retention time shift between the
calibration standards and plasma samples was virtually
eliminated. The abundance of the TDFHA adduct was also
decreased significantly enough to dramatically improve both
the accuracy and the precision of Val in plasma and spiked
plasma. The accuracy and precision of Ala-Gln was not
improved significantly after cation-exchange cleanup (data
not shown).
Copyright # 2007 John Wiley & Sons, Ltd.
Calibration linearityCalibration linearity was compared between the samples
analyzed with and without cation-exchange cleanup. Over-
all, much better results were obtained with cation-exchange
cleanup. In the calibration without cation-exchange cleanup,
a quadratic regression was selected as the preferred fit for
most of the amino acids whilst, with cation exchange, a linear
regression fit was determined to be optimal. This improve-
ment in performance could also be due to a 5-fold increase
in sample amount used for the preparation with cation-
exchange cleanup.
In the cation-exchange cleanup, glutamine-d5 was used
with much better quantitative results as an internal standard
for several of the early eluting amino acids when compared
to the samples that had not been cleaned up. Also for the later
eluting amino acids, tryptophan-d5 produced much better
quantitative results in the cation-exchange cleanup.
Amino acids which used an isotopically labeled analog for
quantitation (e.g. glutamine/glutamine-d5, methionine/
methionine-d3) produced excellent calibration curves as
expected. Much more accurate results could be obtained with
this method if stable isotope labeled analogs were utilized for
all target amino acids. This would be prohibitively expensive
Rapid Commun. Mass Spectrom. 2007; 21: 2717–2726
DOI: 10.1002/rcm
Table
3.Inter-dayaccuracyandprecisionofaminoacidsin
humanplasmawithoutcation-exchangecleanup
Sam
ple
nam
eA
nal
yte
pea
kn
ame
Av
g.
con
c.(m
M)
Av
g.
con
c.(m
M)
Av
g.
con
c.(m
M)
Inte
r-d
ayIn
ter-
day
Pre
cisi
on
Sp
ike
con
c.In
ter-
day
(n¼
5)(d
ay1)
(n¼
5)(d
ay2)
(n¼
5)(d
ay3)
Av
erag
eS
tdD
ev.
(%R
SD
)(m
M)
Sp
ike
reco
ver
y(%
)
PL
AS
MA
0-5
(Cy
stei
ne)
239
.12
6.37
14.5
020
.00
17.0
585
.28
PL
AS
MA
100-
5(C
yst
ein
e)2
220.
4027
9.00
482.
0032
7.13
137.
2841
.97
200.
0015
3.57
PL
AS
MA
0-5
Ala
nin
e(A
LA
)38
2.40
368.
8037
7.00
376.
076.
851.
82P
LA
SM
A10
0-5
Ala
nin
e(A
LA
)77
2.80
716.
0072
9.00
739.
2729
.76
4.03
400.
0090
.80
PL
AS
MA
0-5
Ala
ny
l-G
luta
min
e(A
LA
-GL
N)
0.00
0.00
0.00
0.00
0.00
0.00
PL
AS
MA
100-
5A
lan
yl-
Glu
tam
ine
(AL
A-G
LN
)21
.42
25.3
033
.30
26.6
76.
0622
.71
100.
0026
.67
PL
AS
MA
0-5
Arg
inin
e(A
RG
)69
.40
61.6
064
.90
65.3
03.
926.
00P
LA
SM
A10
0-5
Arg
inin
e(A
RG
)16
5.60
138.
0017
2.00
158.
5318
.07
11.4
010
0.00
93.2
3P
LA
SM
A0-
5A
spar
agin
e(A
SN
)8.
7310
.20
10.0
09.
640.
808.
29P
LA
SM
A10
0-5
Asp
arag
ine
(AS
N)
76.1
277
.00
65.9
073
.01
6.17
8.45
100.
0063
.36
PL
AS
MA
0-5
Asp
arti
cac
id(A
SP
)0.
000.
000.
000.
000.
000.
00P
LA
SM
A10
0-5
Asp
arti
cac
id(A
SP
)10
7.60
98.4
010
9.00
105.
005.
765.
4810
0.00
105.
00P
LA
SM
A0-
5C
itru
llin
e(C
IT)
13.4
011
.00
13.0
012
.47
1.29
10.3
1P
LA
SM
A10
0-5
Cit
rull
ine
(CIT
)97
.82
93.8
094
.60
95.4
12.
132.
2310
0.00
82.9
4P
LA
SM
A0-
5G
luta
mic
acid
(GL
U)
227.
2025
5.00
254.
0024
5.40
15.7
76.
43P
LA
SM
A10
0-5
Glu
tam
icac
id(G
LU
)57
1.40
663.
0068
9.00
641.
1361
.77
9.64
400.
0098
.93
PL
AS
MA
0-5
Glu
tam
ine
(GL
N)
27.6
830
.10
39.1
032
.29
6.02
18.6
3P
LA
SM
A10
0-5
Glu
tam
ine
(GL
N)
773.
0081
9.00
686.
0075
9.33
67.5
58.
9080
0.00
90.8
8P
LA
SM
A0-
5G
lyci
ne
(GL
Y)
410.
2045
0.00
486.
0044
8.73
37.9
28.
45P
LA
SM
A10
0-5
Gly
cin
e(G
LY
)12
18.0
013
50.0
013
60.0
013
09.3
379
.25
6.05
800.
0010
7.58
PL
AS
MA
0-5
His
tid
ine
(HIS
)10
3.94
88.4
086
.40
92.9
19.
6010
.33
PL
AS
MA
100-
5H
isti
din
e(H
IS)
313.
2020
0.00
247.
0025
3.40
56.8
722
.44
100.
0016
0.49
PL
AS
MA
0-5
Hy
dro
xy
pro
lin
e(H
YP
)5.
475.
175.
245.
290.
162.
94P
LA
SM
A10
0-5
Hy
dro
xy
pro
lin
e(H
YP
)72
.10
78.7
065
.10
71.9
76.
809.
4510
0.00
66.6
7P
LA
SM
A0-
5Is
ole
uci
ne
(IS
O)
31.7
428
.80
22.5
027
.68
4.72
17.0
5P
LA
SM
A10
0-5
Iso
leu
cin
e(I
SO
)13
0.20
119.
0088
.10
112.
4321
.80
19.3
910
0.00
84.7
5P
LA
SM
A0-
5L
euci
ne
(LE
U)
54.9
658
.70
52.2
055
.29
3.26
5.90
PL
AS
MA
100-
5L
euci
ne
(LE
U)
153.
4015
4.00
138.
0014
8.47
9.07
6.11
100.
0093
.18
PL
AS
MA
0-5
Ly
sin
e(L
YS
)13
3.20
116.
0013
1.00
126.
739.
367.
39P
LA
SM
A10
0-5
Ly
sin
e(L
YS
)24
2.20
189.
0024
4.00
225.
0731
.25
13.8
810
0.00
98.3
3P
LA
SM
A0-
5M
eth
ion
ine
(ME
T)
12.6
011
.80
10.2
011
.53
1.22
10.6
0P
LA
SM
A10
0-5
Met
hio
nin
e(M
ET
)11
0.20
122.
0012
0.00
117.
406.
325.
3810
0.00
105.
87P
LA
SM
A0-
5O
rnit
hin
e(O
RN
)31
8.00
277.
0023
8.00
277.
6740
.00
14.4
1P
LA
SM
A10
0-5
Orn
ith
ine
(OR
N)
757.
8037
6.00
522.
0055
1.93
192.
6534
.90
100.
0027
4.27
PL
AS
MA
0-5
Ph
eny
lala
nin
e(P
HE
)47
.46
47.6
040
.70
45.2
53.
948.
72P
LA
SM
A10
0-5
Ph
eny
lala
nin
e(P
HE
)15
0.00
163.
0015
1.00
154.
677.
234.
6810
0.00
109.
41P
LA
SM
A0-
5P
roli
ne
(PR
O)
42.6
041
.80
50.8
045
.07
4.98
11.0
5P
LA
SM
A10
0-5
Pro
lin
e(P
RO
)16
8.80
145.
0015
3.00
155.
6012
.11
7.78
100.
0011
0.53
PL
AS
MA
0-5
Ser
ine
(SE
R)
71.4
876
.10
81.2
076
.26
4.86
6.38
PL
AS
MA
100-
5S
erin
e(S
ER
)16
4.60
170.
0019
6.00
176.
8716
.79
9.49
100.
0010
0.61
PL
AS
MA
0-5
Tau
rin
e(T
AU
)3.
415.
454.
454.
441.
0223
.04
PL
AS
MA
100-
5T
auri
ne
(TA
U)
33.5
242
.40
33.6
036
.51
5.10
13.9
810
0.00
32.0
7P
LA
SM
A0-
5T
hre
on
ine
(TH
R)
68.8
075
.30
74.5
072
.87
3.54
4.86
Copyright # 2007 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2007; 21: 2717–2726
DOI: 10.1002/rcm
Analysis of 25 underivatized amino acids in human plasma 2723
(Continues)
Table
3.(C
ontinued)
Sam
ple
nam
eA
nal
yte
pea
kn
ame
Av
g.
con
c.(m
M)
Av
g.
con
c.(m
M)
Av
g.
con
c.(m
M)
Inte
r-d
ayIn
ter-
day
Pre
cisi
on
Sp
ike
con
c.In
ter-
day
(n¼
5)(d
ay1)
(n¼
5)(d
ay2)
(n¼
5)(d
ay3)
Av
erag
eS
tdD
ev.
(%R
SD
)(m
M)
Sp
ike
reco
ver
y(%
)
PL
AS
MA
100-
5T
hre
on
ine
(TH
R)
166.
2017
2.00
173.
0017
0.40
3.67
2.15
100.
0097
.53
PL
AS
MA
0-5
Try
pto
ph
an(T
RP
)28
.86
27.5
029
.80
28.7
21.
164.
03P
LA
SM
A10
0-5
Try
pto
ph
an(T
RP
)14
1.40
137.
0014
2.00
140.
132.
731.
9510
0.00
111.
41P
LA
SM
A0-
5T
yro
sin
e(T
YR
)37
.38
30.4
025
.50
31.0
95.
9719
.19
PL
AS
MA
100-
5T
yro
sin
e(T
YR
)13
2.60
109.
0096
.40
112.
6718
.38
16.3
110
0.00
81.5
8P
LA
SM
A0-
5V
alin
e(V
AL
)88
.34
72.8
028
.00
63.0
531
.33
49.6
9P
LA
SM
A10
0-5
Val
ine
(VA
L)
189.
6015
1.00
51.7
013
0.77
71.1
454
.40
100.
0067
.72
2724 M. Armstrong, K. Jonscher and N. A. Reisdorph
Copyright # 2007 John Wiley & Sons, Ltd.
to do for all amino acids, but if only a few amino acids are
targeted, the benefits of improved accuracy would con-
ceivably outweigh the costs.
Accuracy and precision of methodTo measure accuracy and precision, human plasma was
prepared and analyzed without amino acids spiked, and
with amino acids spiked at levels equivalent to the level 4
calibration standard (100mM nominal concentrations). The
plasma samples were spiked and frozen at �808C for at least
24 h before thawing, preparing and analyzing. Three
separate aliquots of the unspiked/spiked plasma were
prepared and analyzed on three different days. Results
from the first day were used to calculate the intra-day
accuracy and precision and results from all three days were
used to calculate inter-day accuracy and precision (Table 3).
The same unspiked/spiked samples were prepared using
cation-exchange cleanup, and intra-day precision was
calculated.
Intra-day precision of all amino acids without cation-
exchange cleanup was <20 RSD (n¼ 5) for all amino acids
measured. Only Cit, Gly, His, and Val had RSDs >10.
Intra-day spike recoveries for most amino acids was within
80–120% with the exception of Ala-Gln (21.4%), Asn (67.4%),
His (209%), Hyp (66.6%), Orn (440%), Pro (126%) and Tau
(30.1%). Due to extremely poor chromatography, His and
Orn peaks were poorly integrated, resulting in aberrantly
high recoveries. Ion suppression from plasma co-extractives
eluting in the void volume reduced the recovery of Tau,
while Ala-Gln recovery was affected by ion suppression
from TDFHA adducts.
Inter-day precision of most amino acids was <20 RSD (see
Table 3). However, some amino acids had very high
inter-day RSD due to low endogenous concentration
(Cys2), poor chromatography (Orn), or ion suppression
(Ala-Gln, Val). Inter-day spike recoveries for most amino
acids were within 80–120% with the exception of Cys2
(154%), Ala-Gln (26.7%), Asn (63.4%), His (160%), Hyp
(66.7%), Orn (274%), Tau (32%) and Val (67.7%). Reprodu-
cibility of Val quantitiation was good on the first two days of
the study; however, due to progressively degrading
chromatography, Val eventually co-eluted with the TDFHA
adduct peak and its signal was suppressed. This degradation
in chromatography can be improved through more frequent
washes with 100% acetonitrile, as was noted by Piraud et al.23
Intra-day precision of amino acids following cation-
exchange cleanup was <20 RSD (n¼ 5) for all amino acids
measured. Intra-day accuracy following cation-exchange
cleanup was vastly improved over analysis without
cation-exchange cleanup. Recoveries for all amino acids
were between 78–127%, with only Ala (127%), Asn (78.3%),
Cit (78.3%) and Glu (79.2%) outside of 80–120%. The most
dramatic improvements occurred with Tau (118%), His
(93.4%) and Orn (84.8%).
We attribute the improvement in the results to the removal
of co-extracted non-target analytes and reduction of the
TDFHA adduct that resulted in significantly diminished ion
suppression. The degradation in chromatography over time
was also much less pronounced. While similar intra-day
precision was achieved without cation-exchange cleanup,
Rapid Commun. Mass Spectrom. 2007; 21: 2717–2726
DOI: 10.1002/rcm
Figure 4. Mass accuracy of the ESI-TOF is under 1 ppm for His and under 3 ppm for Lys and Arg. Plasma samples
were prepared and analyzed using RP-LC/ESI-TOF as described in the Experimental section. The mass spectra for
Lys, His and Arg are shown and mass accuracy was calculated using experimental mass and actual mass values as
shown. Overall the mass accuracy is well below 5 ppm, thereby reducing the number of possible empirical formulas for
these peaks and improving the qualitative spectral data.
Analysis of 25 underivatized amino acids in human plasma 2725
the few additional steps involved with the cleanup greatly
improve the overall performance of the method.
DISCUSSION
Analysis of amino acids by ion-pairing RP-LC/ESI-TOF is a
viable alternative to traditional amino acid analysis methods,
such as GC/MS and ninhydrin methods, both of which
require derivatization. While the IC/ninhydrin method
requires very little sample preparation, relatively long
(approximately 1–2 h per sample) analysis times are needed
in order to achieve baseline separation suitable for quanti-
tation. The method also generally requires a dedicated
system for online derivatization of samples in order to obtain
consistent results. Conversely, GC/MS methods tradition-
ally have shorter run times, excellent chromatography, and
increased specificity; however, GC/MS methods require
extensive sample preparation and derivatization. Increased
handling of the sample due to numerous steps is not only
time-consuming, but can potentially lead to increased error
and variability in the results.
Although the preparation of samples for LC/MS analysis
using an amino acid kit is relatively simple compared to
derivatization, the kits are designed to be used with the
specific derivative chemistry they were developed for and
may not be optimal for every amino acid, nor for every
detection technique. Analysis of the resultant samples
requires very high flow rates (0.5–1.0 mL/min) and the
use of non-volatile salt buffers which are not readily
compatible with ESI-MS.
Tandem mass spectrometry with flow injection analysis
can be used to identify amino acids without chromatographic
separation through the use of monitoring the transition of
precursor ions to product ions or multiple reaction
monitoring (MRM). While this technique is extremely
Copyright # 2007 John Wiley & Sons, Ltd.
specific and sensitive, some analytes may be subject to ion
suppression due to the complex nature of the sample
matrix,23 resulting in anomalous quantitation levels.
While there have been some promising advances in amino
acid analysis utilizing sample introduction and ionization
methods such as matrix-assisted laser desorption/ionization
(MALDI)24 and high-field asymmetric waveform ion mobi-
lity spectroscopy (FAIMS),25 as of the time of writing neither
of these techniques has been investigated for analysis of
amino acids in biological fluids such as plasma or urine.
In spite of advantages such as high mass accuracy, TOFMS
has not been used extensively for quantitative analysis due to
limitations imposed by time-to-digital converters (TDC),
which have poor dynamic range and can have considerable
dead times when measuring high concentrations of analyte.
An analog-to-digital converter (ADC) can more accurately
measure signal intensity than a TDC. ADC technology allows
TOF mass spectrometers to be used for quantitative analysis
while retaining a high degree of mass accuracy. The mass
accuracy for amino acids obtained by this method of
correction is well below 5 ppm (Fig. 4).
CONCLUSIONS
Ion-pairing reversed-phase chromatography coupled with
the current generation of small particle size columns makes
separation of amino acids possible without derivatization,
allowing for quick and reproducible sample preparation.
Mass accuracy obtained through time-of-flight mass spec-
trometry can be used in addition to retention time to provide
qualitative data that is not available with single quadrupole
or triple quadrupole mass spectrometers. The described
method can be utilized to provide quick and accurate results
for amino acids in human plasma.
Rapid Commun. Mass Spectrom. 2007; 21: 2717–2726
DOI: 10.1002/rcm
2726 M. Armstrong, K. Jonscher and N. A. Reisdorph
AcknowledgementsSupport for this work was generously provided through the
Colorado Clinical Nutrition Research Unit (Funding through
NIH/NIDDK P30 DK048520-09, PI Dr. James Hill). The
authors would like to thank Dr. Patti Thureen for her helpful
comments.
REFERENCES
1. Jellum E, Kvittingen EA, Stokke O. Biomed. Environ. MassSpectrom. 1988; 16: 57.
2. Deng C, Deng Y. J. Chromatogr. B Analyt. Technol. Biomed. LifeSci. 2003; 792: 261.
3. Magni F, Arnoldi L, Galati G, Galli Kienle M. Anal. Biochem.1994; 220: 308.
4. Pitt JJ, Eggingtin E, Kahler SG. Clin. Chem. 2002; 48: 1970.5. Shen X, Deng C, Wang B, Dong L.Anal. Bioanal. Chem. 2006;
384: 931.6. Deng C, Li N, Zhang X. Rapid Commun.Mass Spectrom. 2004;
18: 2558.7. des Robert C, Le Bacquer O, Piloquet H, Roze JC, Darmaun
D. Pediatr. Res. 2002; 51: 87.8. Miller RG, Jahoor F, Reeds PJ, Heird WC, Jaksic T. J. Pediatr.
Surg. 1995; 30: 1325.9. Spackman DH, Stein WH, Moore S. Anal. Chem. 1958; 30:
1185.
Copyright # 2007 John Wiley & Sons, Ltd.
10. Mahwhinney TP, Robinett RS, Atalay A, Madson MA.J. Chromatogr. 1986; 358: 231.
11. Calder AG, Garden KE, Andersen SE, Lobley GE. RapidCommun. Mass Spectrom. 1999; 13: 2080.
12. Wood PL, Amin Khan M, Moskal JR. J. Chromatogr. B 2006;831: 313.
13. Husek JP. J. Chromatogr. 1991; 552: 289.14. Husek JP. J. Chromatogr. B 1998; 717: 57.15. Namera A, Yashiki M, Nishida M, Kojima T. J. Chromatogr. B
2002; 776: 49.16. Petritis K, Chaimbault P, Elfakir C, Dreux M. J. Chromatogr. A
1999; 833: 147.17. Muller C, Schafer P, Stortzel M, Vogt S, Weinmann W.
J. Chromatogr. B 2002; 773: 47.18. Matuszewski BK, Constanzer ML, Chavez-Eng CM. Anal.
Chem. 1998; 70: 882.19. Cohen SA, Michaud DP. Anal. Biochem. 1993; 211: 279.20. Petritis K, Chaimbault P, Elfakir C, Dreux M. J. Chromatogr. A
2000; 896: 253.21. Chaimbault P, Petritis K, Elfakir C, Dreux M. J. Chromatogr. A
1999; 855: 191.22. Piraud M, Vianey-Saban C, Petritis K, Elfakir C, Steghens
JP, Morla A, Bouchu D. Rapid commun. Mass Spectrom. 2003;17: 1297.
23. Piraud M, Vianey-Saban C, Petritis K, Elfakir C, SteghensJP, Bouchu D. Rapid Commun. Mass Spectrom. 2005; 17:1587.
24. Alterman M, Gogichayeva N, Kornilayev B. Anal. Biochem.2004; 335: 184.
25. McCooeye M, Mester Z. Rapid Commun. Mass Spectrom.2006; 20: 1801.
Rapid Commun. Mass Spectrom. 2007; 21: 2717–2726
DOI: 10.1002/rcm