20
A & E Lab Instruments (Guangzhou) Co.,Ltd E-mail: [email protected] Website: www.aelabgroup.com AE-P series Portable Pressure Steam Sterilizer Product Features Intelligent microcomputer control system: Using LED digital control system; After sterilization finished, sterilizer will automatic alarm and shut down. Interlock device: The control system can real time supervision the sterilizer chamber’s temperature and pressure. Only under the safety temperature and pressure can open the sterilizer’s cover. Make sure the operator’s safety. The structure of cover’s opening: Using shift place type of quick-open cover to improve sealing and safety performance. Preset program: Can preset fixed program that sterilize the liquid and solid body. Memory storage system: Store the last operation program so convenient next operation. Digital display screen: Digital display screen and working indicator light display running state. Safety devices: Interlock device, overtemperature and over pressure protecting system, burned protection, safety relief valve, current leakage protector and automatic fault detection. Technical Parameter Parameter/Model AE-P18A AE-P18B Configuration Basic Configuration Automatic Control Material/Pot wall thickness(mm) Stainless steel/2.0 Stainless steel/2.0 Volume/GW 18L/12Kg 18L/12Kg Voltage/Power 220V/2Kw 220V/2Kw Packing size(L x W x H mm) 405 x 405 x 615 405 x 405 x 615 Sterilization chamber volume(D x H mm) 280 x 395 80 x 395 Time setting range 0-999 min Rated working pressure 0.142MPa 0.142MPa Sterilization temperature setting range 5012650126

AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

  • Upload
    others

  • View
    21

  • Download
    0

Embed Size (px)

Citation preview

Page 1: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co.,Ltd E-mail: [email protected] Website: www.aelabgroup.com

AE-P seriesPortable Pressure Steam Sterilizer

Product Features

Intelligent microcomputer control system: Using LED digital control system; After sterilization finished,sterilizer will automatic alarm and shut down.Interlock device: The control system can real time supervision the sterilizer chamber’s temperature andpressure. Only under the safety temperature and pressure can open the sterilizer’s cover. Make sure theoperator’s safety.The structure of cover’s opening: Using shift place type of quick-open cover to improve sealing and safetyperformance.Preset program: Can preset fixed program that sterilize the liquid and solid body.Memory storage system: Store the last operation program so convenient next operation.Digital display screen: Digital display screen and working indicator light display running state.Safety devices: Interlock device, overtemperature and over pressure protecting system, burned protection,safety relief valve, current leakage protector and automatic fault detection.

Technical Parameter

Parameter/Model AE-P18A AE-P18BConfiguration Basic Configuration Automatic Control

Material/Pot wall thickness(mm) Stainless steel/2.0 Stainless steel/2.0Volume/GW 18L/12Kg 18L/12Kg

Voltage/Power 220V/2Kw 220V/2KwPacking size(L x W x H mm) 405 x 405 x 615 405 x 405 x 615

Sterilization chamber volume(D x H mm) 280 x 395 80 x 395Time setting range - 0-999 min

Rated working pressure 0.142MPa 0.142MPaSterilization temperature setting range 50~126℃ 50~126℃

Page 2: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co.,Ltd E-mail: [email protected] Website: www.aelabgroup.com

AE-P seriesPortable Pressure Steam Sterilizer

Product Features

Intelligent microcomputer control system: Using LED digital control system; After sterilization finished,sterilizer will automatic alarm and shut down.Interlock device: The control system can real time supervision the sterilizer chamber’s temperature andpressure. Only under the safety temperature and pressure can open the sterilizer’s cover. Make sure theoperator’s safety.The structure of cover’s opening: Using shift place type of quick-open cover to improve sealing and safetyperformance.Preset program: Can preset fixed program that sterilize the liquid and solid body.Memory storage system: Store the last operation program so convenient next operation.Digital display screen: Digital display screen and working indicator light display running state.Safety devices: Interlock device, overtemperature and over pressure protecting system, burned protection,safety relief valve, current leakage protector and automatic fault detection.

Technical Parameter

Parameter/Model AE-PB24 AE-PB30Configuration Automatic Control Automatic Control

Material/Pot wall thickness(mm) Stainless steel/2.0 Stainless steel/2.0Volume/GW 24L/25Kg 30L/27.5Kg

Voltage/Power 220V/2Kw 220V/2KwPacking size(L x W x H mm) 500 x 500 x 800 500 x 500 x 900

Sterilization chamber volume(D x H mm) 280 x 495 280 x 585Time setting range 0-999 min 0-999 min

Rated working pressure 0.142MPa 0.142MPaSterilization temperature setting range 50~126℃ 50~126℃

Page 3: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co.,Ltd E-mail: [email protected] Website: www.aelabgroup.com

AE-V seriesVertical Pressure Steam Sterilizer

Product Features

Technical Parameter

Parameter/Model AE-VB40/60/80 AE-VC40/60/80 AE-VD40/60/80

Configuration Basic ConfigurationSteam emission byinternal circulation

Drying

The main material Stainless steel Stainless steel Stainless steel

Volume/GW40L/95Kg;60L/105Kg;80L/120Kg

40L/95Kg;60L/105Kg;80L/120Kg

40L/95Kg;60L/105Kg;80L/120Kg

Voltage/Power 220V/3.5Kw 220V/3.5Kw 220V/3.5Kw

Packing size(L x W x H mm)660 x 800 x 900;

660 x 800 x 1100; 660x 800 x 1250

660 x 800 x 900;660 x 800 x 1100;660 x 800 x 1250

660 x 800 x 900;660 x 800 x 1100;660 x 800 x 1250

Sterilization chamber volume(D x H mm)

370 x 380;370 x 560;370 x 750

370 x 380;370 x 560;370 x 750

370 x 380;370 x 560;370 x 750

Time setting range 0-99 h 0-99 h 0-99 hRated working pressure 0.217MPa 0.217MPa 0.217MPaSterilization temperature

setting range50~134℃ 50~134℃ 50~134℃

Intelligent microcomputer control system: Using LED digital control system; Aftersterilization finished, sterilizer will automatic alarm and shut down.Electric interlock device: The control system can real time supervision the sterilizerchamber’s temperature and pressure. Only under the safety temperature and pressurecan open the sterilizer’s cover. Make sure the operator’s safety.The structure of cover’s opening: Using 8 rotary extendable stick to improve sealing andsafety performance.Preset program: Can preset fixed program that sterilize and dry the liquid and solid body.Memory storage system: Store the last operation program so convenient next operation.Exhaust steam control: After sterilization, exhaust steam valve automatically work whensterilization time run out(apply to Auto exhaust and dry configuration).Interface check: Provide temperature and pressure check interface; convenient to checkmultiple position temperature measurement.Digital display screen: Digital display screen and working indicator light display runningstate.Safety devices: Electric interlock device, overtemperature protecting system, burnedprotection, safety relief valve, current leakage protector and automatic fault detection.Print function: Print the current temperature and sterilization time.

Page 4: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co.,Ltd E-mail: [email protected] Website: www.aelabgroup.com

AE-V seriesVertical Pressure Steam Sterilizer

Product Features

Technical Parameter

Parameter/Model AE-VB30/50/75 AE-VE30/50/75 AE-VD30/50/75Configuration Basic Configuration Auto exhaust Drying

The main material Stainless steel Stainless steel Stainless steel

Volume/GW30L/80Kg;50L/90Kg;75L/100Kg

30L/80Kg;50L/90Kg;75L/100Kg

30L/80Kg;50L/90Kg;75L/100Kg

Voltage/Power 220V/3.5Kw 220V/3.5Kw 220V/3.5Kw

Packing size(L x W x H mm)660 x 610 x 970;660 x 610 x 1120;660 x 610 x 1270

660 x 610 x 970;660 x 610 x 1120;660 x 610 x 1270

660 x 610 x 970;660 x 610 x 1120;660 x 610 x 1270

Sterilization chamber volume(D x H mm)370 x 320;370 x 500;370 x 690

370 x 320;370 x 500;370 x 690

370 x 320;370 x 500;370 x 690

Time setting range 0-99 h 0-99 h 0-99 hRated working pressure 0.217MPa 0.217MPa 0.217MPa

Sterilization temperature setting range 50~134℃ 50~134℃ 50~134℃

Intelligent microcomputer control system: Using LED digital control system;After sterilization finished, sterilizer will automatic alarm and shut down.Interlock device: The control system can real time supervision the sterilizerchamber’s temperature and pressure. Only under the safety temperatureand pressure can open the sterilizer’s cover. Make sure the operator’ssafety.The structure of cover’s opening: Using rotary type cover to improvesealing and safety performance.Preset program: Can preset fixed program that sterilize the liquid and solidbody.Memory storage system: Store the last operation program so convenientnext operation.Interface check: Provide temperature and pressure check interface;convenient to check multiple position temperature measurement.Digital display screen: Digital display screen and working indicator lightdisplay running state.Safety devices: Interlock device, overtemperature and over pressureprotecting system, burned protection, safety relief valve, current leakageprotector and automatic fault detection.Print function: Print the current temperature and sterilization time.

Page 5: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co.,Ltd E-mail: [email protected] Website: www.aelabgroup.com

AE-V seriesVertical Pressure Steam Sterilizer

Product Features

Intelligent microcomputer control system: Using LED digital control system; After sterilization finished,sterilizer will automatic alarm and shut down.Interlock device: The control system can real time supervision the sterilizer chamber’s temperature andpressure. Only under the safety temperature and pressure can open the sterilizer’s cover. Make sure theoperator’s safety.The structure of cover’s opening: Using hand wheel rotary type cover to improve sealing and safetyperformance.Preset program: Can preset fixed program that sterilize the liquid and solid body.Memory storage system: Store the last operation program so convenient next operation.Interface check: Provide temperature and pressure check interface; convenient to check multiple positiontemperature measurement.Digital display screen: Digital display screen and working indicator light display running state.Safety devices: Interlock device, overtemperature and over pressure protecting system, burned protection,safety relief valve, current leakage protector and automatic fault detection.Print function: Print the current temperature and sterilization time.

Technical Parameter

Parameter/Model AE-V50KBS AE-V75KBSConfiguration Automatic Control Automatic Control

The main material Stainless steel Stainless steelVolume/GW 50L/75Kg 75L/98Kg

Voltage/Power 220V/3.5Kw 220V/3.5kwPacking size(L x W x H mm) 570 x 570 x 1220 610 x 610 x 1410

Sterilization chamber volume(D x H mm) 350 x 520 388 x 650Time setting range 0-99 h 0-99 h

Rated working pressure 0.142MPa 0.142MPaSterilization temperature setting range 50~126℃ 50~126℃

Page 6: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co.,Ltd E-mail: [email protected] Website: www.aelabgroup.com

AE-V seriesVertical Pressure Steam Sterilizer

Product Features

Intelligent microcomputer control system: Using LED digital control system; After sterilization finished,sterilizer will automatic alarm and shut down.Electric interlock device: The control system can real time supervision the sterilizer chamber’s temperatureand pressure. Only under the safety temperature and pressure can open the sterilizer’s cover. Make surethe operator’s safety.The structure of cover’s opening: Using 8 rotary extendable stick to improve sealing and safetyperformance.Preset program: Can preset fixed program that sterilize and dry the liquid and solid body.Memory storage system: Store the last operation program so convenient next operation.Exhaust steam control: After sterilization, exhaust steam valve automatically work when sterilization timerun out(apply to Auto exhaust and dry configuration).Interface check: Provide temperature and pressure check interface; convenient to check multiple positiontemperature measurement.Digital display screen: Digital display screen and working indicator light display running state.Safety devices: Interlock device, overtemperature and over pressure protecting system, burned protection,safety relief valve, current leakage protector and automatic fault detection.Print function: Print the current temperature and sterilization time.

Technical Parameter

Parameter/Model AE-V100KBS AE-V150KBS AE-V200KBSConfiguration Automatic Control Automatic Control Automatic Control

The main material Stainless steel Stainless steel Stainless steelVolume/GW 100L/160Kg 150L/180Kg 200L/200Kg

Voltage/Power 380V/6Kw 380V/6kw 380V/6KwPacking size(L x W x H mm) 760 x 780 x 1310 760 x 710 x 1520 750 x 720 x 1690

Sterilization chamber volume(D x H mm) 500 x 550 500 x 760 500 x 1000Time setting range 0-99 h 0-99 h 0-99 h

Rated working pressure 0.217MPa 0.217MPa 0.217MPaSterilization temperature setting range 50~134℃ 50~134℃ 50~134℃

Page 7: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co.,Ltd E-mail: [email protected] Website: www.aelabgroup.com

AE-H seriesHorizontal Steam Sterilizer

Product Features

Intelligent microcomputer control system: Using LED digital control system; After sterilization finished,sterilizer will automatic alarm and shut down.Interlock device: The control system can real time supervision the sterilizer chamber’s temperature andpressure. Only under the safety temperature and pressure can open the sterilizer’s cover. Make sure theoperator’s safety.The structure of cover’s opening: Using 8 rotary extendable stick to improve sealing and safetyperformance.Preset program: Can preset fixed program that sterilize the liquid and solid body.Memory storage system: Store the last operation program so convenient next operation.Exhaust steam control: After sterilization, exhaust steam valve should be opened(Digital display screenreminds exhaust steam) when sterilization time run out.Interface check: Provide temperature and pressure check interface; convenient to check multiple positiontemperature measurement.Digital display screen: Digital display screen and working indicator light display running state.Safety devices: Interlock device, overtemperature and over pressure protecting system, burned protection,safety relief valve, current leakage protector and automatic fault detection.Print function: Print the current temperature and sterilization time.

Technical Parameter

Parameter/Model AE-H120KC AE-H200KC AE-H300KCConfiguration Automatic Control Automatic Control Automatic Control

The main material Stainless steel Stainless steel Stainless steelVolume/GW 120L/230Kg 200L/250Kg 300L/270Kg

Voltage/Power 380V/7.5Kw 380V/7.5kw 380V/7.5KwPacking size(L x W x H mm) 1160 x 622 x 1550 1460 x 722 x 1650 1460 x 822 x 1750

Sterilization chamber volume(D x H mm) 400 x 960 500 x 1020 600 x 1050Time setting range 0-99 h 0-99 h 0-99 h

Rated working pressure 0.217MPa 0.217MPa 0.217MPaSterilization temperature setting range 50~134℃ 50~134℃ 50~134℃

Page 8: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co.,Ltd E-mail: [email protected] Website: www.aelabgroup.com

AE-SD seriesStainless Steel Water Distiller

Product Features

Using double directions regulate pressure valve. Make sure that steam cooling water and filling waterbeing divided.

Water standard meets 2000 edition State Pharmacopoeia Standard. Water supply cut protecting equipment (apply to controlled type). Indicator light display running state (apply to controlled type). With branching control install; solenoid valve control the flow of filling water (apply to controlled type).

Technical Parameter

Parameter/Model AE-SD5/10/20 AE-SD5A/10A/20AConfiguration Basic Configuration Automatic Control

The main material Stainless steel Stainless steelOut of water/GW 5L/h /9Kg; 10L/h /9Kg; 20L/h /10Kg 5L/h /9Kg; 10L/h /9Kg; 20L/h /10Kg

Voltage/Power220V/5KwX1;380V/2.5KwX3;380V/4.5KwX3

220V/5KwX1;380V/2.5KwX3;380V/4.5KwX3

Packing size(L x W x H mm)365 x 368 x 845;365 x 368 x 845;365 x 368 x 945

365 x 368 x 845;365 x 368 x 845;365 x 368 x 945

Water consumption 1:10 1:10Water category Regular water Regular water

Page 9: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co.,Ltd E-mail: [email protected] Website: www.aelabgroup.com

AE-TD seriesStainless Steel Water Distiller

Product Features

Using double directions regulate pressure valve. Make sure that steam cooling water and filling waterbeing divided.

Can extract second distillation water; Cooling mode can distill single distillation water. Water standard meets 2000 edition State Pharmacopoeia Standard. Water supply cut protecting equipment. There will no heating when water level too low. Microcomputer control system (apply to Cooling mode). Equip with water, over temperature alarm and protection (only apply to Cooling mode). Third class water standard for national lab.

Technical Parameter

Parameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A

ConfigurationManual Conversion/Automatic

conversionControlled type

The main material Stainless steel Stainless steel

Out of water/GW 7L/h /36Kg5L/h /30Kg; 10L/h/ 30Kg;

20L/h/ 40Kg

Voltage/Power 220V/4.2Kw380V/10.5Kw;380V/13.5Kw;380V/22.5Kw

Packing size(L x W x H mm) 500 x 500 x 1000770 x 380 x 860;770 x 380 x 860;860 x 380 x 860

Water consumption 7:10 7:10Water category Purified water Regular water

Page 10: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co.,Ltd E-mail: [email protected] Website: www.aelabgroup.com

Autoclave

*What is class B & class N autoclave?The class B autoclave has vacuum cycle.It is the highest classautoclave,which is conforming to the most stringent demandsand therefore can be unrestrictedly used for the sterilization ofall possible loads for all kinds of sterilization packaging(solid ,hollow road type A and hollow road instruments withwrapped and unwrapped). A "class B" autoclave must in anycase pass a Helix test as per EN 13060 by test body per EN867-5:2001. According to expert opinion, that can only beachieved with a fractionated pre-vacuum procedure. Postvacuum drying provides complete drying of porous loads andhollow instruments at the end of thesterilization process.The class N autoclave do not have vacuum cycle. If you onlysterilize unwrapped non hollow instruments, you only needan autoclave capable of running class N cycles.

*What is the standard of EN 13060?This European Standard specifies the performancerequirements and test methods for small steam sterilizersand sterilization cycles which are used for medical purposesor for materials that are likely to come into contact with blood or body fluids. This European Standardapplies to automatically controlled small steam sterilizers that generate steam using electrical heaters oruse steam that is generated by a system external to the sterilizer.

*Model numberThe first letter of the model number means class N type or class B type autoclave. The Arabic numerals inthe model number means the capacity of the autoclave. The last letter of the model number means thewhich series of our autoclave.

Page 11: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co., LtdE-mail: [email protected] Website: www.aelabgroup.com

*Graphic of working cycle

*Product FeatureMain Feature:1 .Full automatic table top autoclave with pre-programmed sterilization cycles.2.According to European standard EN13060 European standard . (Class B type)3. High precision control system for perfect sterilization results.4. Works according to 3 times fractionated pre and post vacuum method with a effective vacuum pump.(Class B type)5. Built-in independent steam generator, provides readily available steam for faster cycles.6. Prefect drying effect by vacuum . (only class B type)7. Stainless steel seamless chamber , door with electro-polish finish.8. Friendly user-Interface which can be operated very easily by the user.9. LCD screen, displays continuous information about cycle progress.10. 5 displaying languages which is English, Spanish, German,Polish and Chinese for user selection.11.Open type water tank, which is very easy for fitting and cleaning.12. With liquid sterilization program .13. Separated drying program (optional).

SafetyDoor Locking1 .Control lock-out switch prevents starting a cycle if door is not locked properly.2.Door protection device prevents door from opening at high pressure and high temperature3.Double locking device prevents door from opening while chamber is pressurized. (optional)4.1ndependent temperature and pressure monitoring and protection.5. Electricity-leaking protector will cut the power automatic if somewhere has short circuit.6.Fail alarm indicates cycle failure and error code, door alarm indicates the door unlocked, water levelalarm indicates the max or min water level in the water tank.

Documentation1 .Printer: an external printer is optional for all "icanclave" autoclaves.2.USB port: it is used for connecting a USB stick, all sterilization data are automatically writtenIn the USE stick and can be read directly in anyPC and stored electronically.3.1nternal memory: the last 20 cycles can be automatically stored In the autoclave system, which can beprint out at any time.

Page 12: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co., LtdE-mail: [email protected] Website: www.aelabgroup.com

*Technical SpecificationsClass Type Class B Series Class N SeriesModel AE-C18B1 AE-C23B1 AE-C18N1 AE-C23N1Volume 18 liter 23 liter 18 liter 23 liter

Chamber size Ø247×350mm Ø247×450mm Ø247×350mm Ø247×450mmPower 2200W 2200W 2100W 2100W

Net weight 47kgs 51kgs 45kgs 48kgsOptional: 1.Printer 2. USB port and USD stick 3. Double door locking device

4. Independent Drying program

*Pre-programmed cycles

Class N AUTOCLAVE

PROGRAM TEMP (℃) PRESSURE (Mpa) STERILIZATION TIME (Min) TOTAL TIME (min)SOLID 134 210 4 20-45

121 110 20 35-60LIQUID 134 210 10 35-60

121 110 30 35-60WRAPPED 134 210 10 45-65

121 110 30 50-75TEXTILE 134 210 10 45-65

121 110 30 50-75PRION 134 210 18 45-75DRYING ----- ----- ----- 1-20

Class B AUTOCLAVE

PROGRAM TEMP (℃) PRESSURE (Mpa) STERILIZATION TIME (Min) TOTAL TIME (min)SOLID 134 210 4 20-45

121 110 20 35-60LIQUID 134 210 10 35-60

121 110 30 35-60WRAPPED 134 210 4 45-65

121 110 20 50-75TEXTILE 134 210 4 45-65

121 110 20 50-75PRION 134 210 18 45-75DRYING ----- ----- ----- 1-20B&D TEST 134 210 3.5 22-35HELIX TEST 134 210 3.5 22-35

VACUUM TEST ----- ----- ----- 15-20

Page 13: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co., LtdE-mail: [email protected] Website: www.aelabgroup.com

AutoclaveProducts follow the directives and standards◎ Complies with Medical Devices Directives 93/42/EEC,◎ Complies with the requirements of standard EN 13060 "small steam sterilizers" ,◎ Complies with safety requirements according to standard

EN61010-1,EN61010-2-040,◎ Complies with safety requirements according to standard.◎ UL61010-1. IEC 61010-2-040,◎ Complies with EMC directive 2004/108/EC, according to standard EN 61326,

EN61000◎ Complies with Pressure Equipment Directive (PED) 97/23/EC, according to standard

EN 764-1, EN 764-5, EN764-7◎ Applies with quality management system EN13485:2012/AC:2012,◎ Applies with quality management system IS013485:2003,

Main Feature1. It is fully automatic table top autoclave with pre set programs,

with high-precision temperature and pressure controlling;2. The class B autoclaves complies with European standard En13060;3. In built independent rapid steam generator ensures quick sterilization cycles;4. User friendly interface;5. LCD screen displaying time, temperature and pressure, process alert and instrument conditions;6. Choice of language settings including: English, Spanish,German, Polish, French, Magyar, Romanian,

Dutch, Lithuanian, Latvian, Czech,Italian and Chinese;7. Easily accessible water tank.;8. Liquid sterilization program optional;9. Extra drying program optional;

SAFETY1. Double door protection system prevents the cycle from starting if the door is not correctly locked. Thissystem also prevents the door from opening if pressure inside the chamber is not equal to atmosphericpressure outside the chamber;2. Pressure safety valve prevent over pressure in the chamber and steam generator;3. The power is a automatically cut if a short circuit or electrical fault occurs;4. Able to detect and identify the exact cause of any problem, and to issue a specific error code to theoperator;5. Water level switch of water tank controlling the min and max water level;6. Automatic warning for maintenance.

Page 14: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co., LtdE-mail: [email protected] Website: www.aelabgroup.com

Documentation1. Internal printer printsdetailed sterilization data automatically in multiple languages;2. A USB port is used for connecting a usb stick, all sterilization data are automatically written in the usb

stick and can be read directly in any pc and stored electronically;3. The last 20 cycles can be automatically stored in the Autoclave system.

Which can be printed out at any time.Model Type Power Chamber Volume

AE-C18N2 Class N 2100W 18 LitersAE-C18B2 Class B 2200W 18 LitersAE-C23N2 Class N 2100W 23 LitersAE-C23B2 Class B 2200W 23 Liters

Sterilization Record=========================================Program: WRAPPEDTemperature: 134℃Pressure: 206.0kpaDrying Time: 08MinHolding Time: 4.0Min------------------------------------------------------------------

Timp. Temp. PressureStart 12:28:17 089.0℃T1: 12:31:32 087.1℃ -075.0kpaT2: 12:33:43 110.2℃ 052.0kpaT3: 12:36:37 088.9℃ -080.0kpaT4: 12:39:20 114.7℃ 053.7kpaT5: 12:43:37 087.9℃ -080.0kpaT6: 12:50:40 134.8℃ 206.0kpaTS: 134.7℃ 209.5kpaMax.Temperature: 135.2℃Min.Temperature: 134.3℃Max.Pressure: 214.0kpaMin.Pressure: 204.9kpaT7: 12:54:39 134.4℃ 211.4kpa

Chamber Out SizeΦ247×350 W 480mm/H 455mm/D 600mmΦ247×350 W 480mm/H 455mm/D 600mmΦ247×450 W 480mm/H 455mm/D 700mmΦ247×450 W 480mm/H 455mm/D 700mm

T8: 12:57:36 102.1℃ -060.0kpaT9: 12:59:54 098.2℃ -060.0kpaEnd 13:04:07 102.4℃------------------------------------------------------------------Cycle No: 00017Ster.Value: SuccessDate: 2014-06-07SN: E54723Operator:=========================================

Page 15: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co., LtdE-mail: [email protected] Website: www.aelabgroup.com

Pre-programmed Cycles

Class N AutoclavePograms Temperature℃ Pressure Kpa Holding Time(min) Total Time(min)Solid 134 210 4 25-45

121 110 20 30-60Wrapped 134 210 4 35-60

121 110 20 35-65Liquid 134 210 10 35-60

121 110 30 35-65Textile 134 210 8 45-65

121 110 30 50-75Prion 134 210 18 45-70Drying / / / 1-20

Class B AutoclavePograms Temperature℃ Pressure Kpa Holding Time(min) Total Time(min)

Solid 134 210 4 25-45121 110 20 30-60

Wrapped 134 210 4 35-60121 110 20 35-65

Textile 134 210 8 45-65121 110 30 50-75

Prion 134 210 18 50-75Liquid 134 210 10 35-60

121 210 30 35-65B & D Test 134 110 3.5 28-35Heltx Test 134 210 3.5 28-35

Vacuum Test / / / 20-25

Page 16: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co., LtdE-mail: [email protected] Website: www.aelabgroup.com

NKB3000 Series Automatic Kjeldahl Analyzer

NKB3000 series Kjeldahl Analyzer is an intelligent instrument applied to analyzenitrogen and protein content with a built-in computer-control system. It effectsdistillation automatically with parameters set on a large touch screen. Only bypushing or pulling the digestive duct handle can the digestive duct be easily pickedand placed at the working portion, after which the acrylic safety door is closed andthen water and alkali are added automatically. This instrument also featuresautomatic shutdown and warning signal after completing the work.

Application:It has been widely applied to such fields as agriculture, biology, food, chemicalindustry, pharmacy, tobacco, soil fertilizer, environment, material, scientific researchand quality supervision, which can be used to conduct macroanalysis andsemimicro-analysis of nitrogen and protein content in the samples like soil, fertilizer,food, crop, medicine, forage, water, chemical product, rubber, material and solidand liquid wastes. It is also adopted to test ammonium salt, volatile phenol, sulfide,alcohol and soil cation exchange capacity.

Product Features:* A large touch screen for setting distillation time and alkali content, real-time display of condensate watercurrent curve and automatic storage of data.* Easy opeartion and automatic warning system after distillation.* Independent water-added pump and alkali-added pump with high reliability.* Mature vapor generator technology, automatic preheating and water replenishing, protection againstheating without water and automatic cleaning.* Reagent addition through the negative suction system instead of the low-cost high pressure pumping,which could avoid injuring people.* The sensor technology enables the temperature of effusion bottle to immediately display the actual testtemperature without manual test, thus avoiding the release of nitrogen due to excess temperature* Flow sensor enables real-time detection of condensate water’s flow velocity to guarantee stable flowvelocity, taking the place of traditional flow switch and avoiding low sample recovery rate due to lowvelocity and temperature of the liquid being on the high side.

Technical Specifications:Parameter Name Parameter

Model NKB3100 NKB3200Measurement Range 0.1-200mgN 0.1-240mgN

Function Automatic test module of conventionalNitrogen

1.Automatic test module of conventional Nitrogen.2.Automatic test module of Nitrate Nitrogen3.Automatic test module of high Nitrogen

Operation Mode Auto mode Auto or Manual modeDistillation Time 0-100min/sample(adjustable) 0-100min/sample(adjustable)Recovery Rate(%) ≥99.5% ≥99.5%

Sample Solid≤5g liquid≤25ml Solid≤5g liquid≤25mlDistillation Rate 10-30ml/min 10-30ml/minDistillation Power 375W-1500W 375W-1500WCondensate Water

Consumption ≥1.5L/min ≥1.5L/min

Operation Interface 4.3-inch touch screen with highresolution 4.3-inch touch screen with high resolution

Safety Protection Device Level alarm Flow velocity monitoring, level alarm, distillatetemperature monitoring

Page 17: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co., LtdE-mail: [email protected] Website: www.aelabgroup.com

Degree of AutomationWith parameter set, the instrument automatically effects sample dilution, addition of boric acid

and alkali, distillation, absorption and piping cleaningRepeatability RSD≤0.5% RSD≤0.5%

Power AC 220±22V 50-60Hz AC 220±22V 50-60HzExternal Dimension 360 x 400 x 600 360 x 400 x 600

Weight About 15kg About 15kg

NKY6100 Series Fully-automatic Kjeldahl AnalyzerProduct Feature* Titration and distillation are conducted at the same time and the ammoniagas released is neutralized, avoid the overflow of ammonia gas and enhancethe test efficiency.* A high-precision micro-metering pump replaces the traditional syringeplunger pump, remarkably improving precision and reducing cleaning work.* the humanized pick & place setting of digestion duct makes it convenientfor users to operate.* Diaphragm pump is used to add reagent to avoid injure which can becaused by high-pressure pump.* Imported RGB color sensor technology is adopted to eliminate thedisturbance of ambient light and achieve an accurate judgment.* A 8.0-inch LCD touch screen is intended to set distillation time, the amountof diluent, acid liquor and alkali. Such a human-computer interactioninterface is easy to operate* Real-time display of titration curve, alerting users to the working conditionat any time..* It possesses an all-stage warning system which alerts users to any abnormal working condition at anytime.* Wireless network modules is installed, realizing remote data transmission, test data processing and testreport printing.* ARM processor is adopted with built-in Flash chip, which is equipped with a file system, enabling usersto create or delete folders or files and store 1000 routine data.* It is equipped with a miniprinter so that users can print experimental data according to their need.

Technical IndexParameter Name ParameterModel NKY6180 NKY6160 NKY6120

Measurement Range 0.1-240mgN 0.1-240mgN 0.1-200mgN

Function1.Automatic test module of conventional Nitrogen.

2.Automatic test module of Nitrate Nitrogen3.Automatic test module of high Nitrogen

Automatic test module of conventionalNitrogen

Recovery Rate(%) ≥99.5% ≥99.5% ≥99.5%Distillation Time 3-8min/sample 3-8min/sample 3-8min/sample

Sample Solid≤5g liquid≤25ml Solid≤5g liquid≤25ml Solid≤5g liquid≤25mlDistillation Mode Double distillation mode Double distillation mode conventional distillation mode

Condensate Water Consumption ≥1.5L/min ≥1.5L/min ≥1.5L/minDistillation Power 375W-1500W 375W-1500W 375W-1500WTitation precision 0.0125 0.2 1Repeatability RSD≤0.3% RSD≤0.5% RSD≤0.5%Stored record 30000 10000 5000

Automatic Function Automatic sample dilution, addition of acid and alkali, distillation, reaction cup emptying and cleaning; autocorrection, calculation and printing. Several conventional methods are equipped.

Result Determination Determined by the terminal color Determined by the terminal color Determined by the terminal color

Page 18: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co., LtdE-mail: [email protected] Website: www.aelabgroup.com

Safety Protection Function

1.Level alarm system2.Alarm system monitoring the temperature of vapor generator3.Flow sensor monitors the whole process with the flow velocity beadjusted freely.4.Temperature measurement of distillate5.Acid and alkali resistant polymer composite materials used for thecabient

1.Level alarm system2.Flow sensor monitors the wholeprocess with the flow velocity beadjusted freely.3.Temperature measurement of distillate4.Acid and alkali resistant polymercomposite materials used for the cabient

Optical sensor RGB programmable optical sensingtechnology

RGB programmable opticalsensing technology

RGB programmable optical sensingtechnology

Display mode 8.0-inch LCD touch screen 8.0-inch LCD touch screen 8.0-inch LCD touch screenTitration mode Imported high-precision micro-metering pump high-precision micro-metering pump

Way of distillation Distillation and titration are carriedout simultaneously

Distillation and titration arecarried out simultaneously

Distillation and titration are carried outsimultaneously

Steam flow adjustable adjustable adjustableOperating Ambient Temperature 10℃-35℃ 10℃-35℃ 10℃-35℃

Water-cooling Device Circulatory condensate system / /Language Chinese/English Chinese/English Chinese/EnglishPower AC220V AC±10% 50Hz AC220V AC±10% 50Hz AC220V AC±10% 50Hz

Rated Power 1500W 1500W 1500WExternal Dimension 360x460x680 360x460x680 360x460x680

Communication Mode Network InterfaceRS232 interface/usb interface

(Enhancement Mode) NetworkInterface

RS232 interface/usb interface

(Enhancement Mode) Network InterfaceRS232 interface/usb interface

Weight About 20kg About 20kg About 20kg

NKD6200 Series Fully-automatic Kjeldahl AnalyzerProduct Feature* the potentiometric blank test and end-point determination method abandons the traditional method of ratingelectrode slope by making use of PH buffer solution, which simplifies the operation procedure, reduces certainmaintenance work and ensures accurate results.* titration and distillation are conducted at the same time and the ammonia gas released is neutralized, avoid theoverflow of ammonia gas and enhance the test efficiency.* a high-precision micro-metering pump replaces the traditional syringe plunger pump, remarkably improvingprecision and reducing cleaning work.* the humanized pick & place setting of digestion duct makes it convenient for users to operate.* disphragm pump is used to add reagent to avoid injure which can be caused by high-pressure pump.* a 8.0-inch LCD touch screen is intended to set distillation time, the amount of diluent, acid liquor and alkali.Such a human-computer interaction interface is easy to operate* real-time display of titration curve, alerting users to the working condition at any time.* ARM processor is adopted with built-in flash chip, which is equipped with a file system, enabling users to createor delete folders or files and store 1000 routine data..* it is equipped with a mini-printer so that users can print experimental data according to their need.Technical Index

Parameter Name ParameterProduct Model NKD6280 NKD6260

Measurement Range 0.1-240mgN 0.1-200mgNFunction 1.Automatic Test Module of Conventional Nitrogen

2.Automatic Test Module of Conventional Nitrogen andNitrate Nitrogen

3.Automatic Test Module of Conventional Nitrogen,Nitrate Nitrogen and High Nitrogen

Automatic Test Module of Conventional Nitrogen

Distillation Time 3-8min 3-8minRecovery Rate(%) ≥99.5 ≥99.5

Sample solid≤5g liquid≤25ml solid≤5g liquid≤25ml

Condensate Water Consumption ≥1,5L/min ≥1,5L/minDistillation Power 375W-1500W 375W-1500WTitration Precision 0.0125 0.2

Repeatability RSD≤0.3% RSD≤0.5%Stored Record 30000 10000

Page 19: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co., LtdE-mail: [email protected] Website: www.aelabgroup.com

Automatic Design Automatic sample dilution, addition of acid and alkali,distillation, titration, reaction cup emptying and

cleaning; auto correction, calculation and printing.Several conventional methods are equipped

Automatic sample dilution, addition of acid and alkali,distillation, titration, reaction cup emptying and cleaning;

auto correction, calculation and printing.Several conventional methods are equipped

Communication Mode Network Interface RS232 interface/usb interface Network Interface RS232 interface/usb interfaceSafety Protection Function 1.Level Alarm Ststem

2.Alarm system monitoring the temperature of vaporgenerator

3.Flow sensor monitors the whole process with the flowvelocity be adjusted freely

4.Temperature measurement of distillate5.Acid and alkali resistant polymer composite materials

used for the cabient

1.Level Alarm Ststem2.Alarm system monitoring the temperature of vapor

generator3.Flow sensor monitors the whole process with the flow

velocity be adjusted freely4.Temperature measurement of distillate

5.Acid and alkali resistant polymer composite materialsused for the cabient

Result Determination Determined by the potential value Determined by the potential valueElectrode Imported PH electrode Imported PH electrode

Display Mode 8.0-inch LCD touch screen for operation 8.0-inch LCD touch screen for operationTitration Driving Mode Imported high-precision micro-metering pump Imported high-precision micro-metering pumpWay of Distillation Distillation and titration are carried out simultaneously Distillation and titration are carried out simultaneously

Steam Flow adjustable adjustableOperating Ambient Temperature 10℃-35℃ 10℃-35℃

Water-cooling Device Circulatory condensate system /Language Chinese ChinesePower AC220V AC±10% 50Hz AC220V AC±10% 50Hz

Rated Power 1500W 1500WExternal Dimension 360x460x680 360x460x680

Weight About 20kg About 20kg

XTG5000 Series Carbon/Electric Heating Digestion Furnace

Application:It has been widely applied to such fields as agriculture, biology,food, chemical industry, pharmacy, tobacco, soil fertilizer,environment, material, scientific research and qualitysupervision, which can be used to conduct macroanalysis andsemimicro-analysis of nitrogen and protein content in thesamples like soil, fertilizer, food, crop, medicine, forage, water,chemical product, rubber, material and solid and liquid wastes.It is also adopted to test ammonium salt, volatile phenol, sulfide,alcohol and soil cation exchange capacity.

Product Features:* Vaccum heating tube is applied which features high heatingrate, good temperature uniformity, acid and alkali resistance and is free from short trouble.* nanometer thermal insulating material possesses good insulation effect. Working continuously at the hightemperature of 450℃, the surface temperature of the instrument will not exceed 50℃.* PID algorithm is adopted to accurately control the temperature, featuring high heating rate and goodthermal insulation.* a layer of Teflon coating on the surface is resistant to such corrosive liquids as acid and alkali which canprevent rust* a 4.3-inch touch screen charactrizing easy operation, convenience and understandabiltity* real-time display of temperature curve and heat preservation time at the current temperature phase,which makes it convenient to observe and understand the real-time condition..

Page 20: AE-PseriesParameter/Model AE-FD7A/FD7B AE-TD5A/10A/20A Configuration ManualConversion/Automatic conversion Controlledtype Themainmaterial Stainlesssteel Stainlesssteel Outofwater/GW

A & E Lab Instruments (Guangzhou) Co., LtdE-mail: [email protected] Website: www.aelabgroup.com

* data storage function enables the temperature parameters set to store in the storage so that theseparameters can be called directly in the future.* multistage temperature setting function, making digesting sample parameter set according to needs* several warning functions which can alert users to the working condition.Technical Index

Parameter Name ParameterProduct Model XTG5201 XTG5081

Digestion Hole Number 20 8

Digestion Amount/SampleSolid≤5g ≤5g

Fluid≤25ml ≤25mlTemperature Range Indoor temp-500℃ Indoor temp-500℃

Heating Material Carbon fiber/vaccum heatingwire

Carbon fiber/vaccumheating wire

Thermal Insulation Material Nano Cloud Nano CloudHeating Method Infrared radiation Infrared radiation

Temperature measurement accuracy 0.1℃/1℃ 0.1℃/1℃

Rated Power 2200W 1700WWeight 20KG 15KG

External Dimension 580 x 450 x 200 480 x 380 x 200Power AC 220±22V 50-60Hz AC 220±22V 50-60Hz

Form of Control Touch Screen Touch ScreenTemperature control accuracy ±1℃/±2℃ ±1℃/±2℃

No-load heating up time(20℃-440℃) 10min 10min