Advancements in Handheld LIBS for Metal Alloy Analysis Don Sackett, CEO SciAps, Inc. Woburn, MA Confidential and Proprietary SciAps Inc 1

Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Embed Size (px)

Citation preview

Page 1: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Advancements in Handheld LIBS for Metal Alloy Analysis

Don Sackett, CEOSciAps, Inc. Woburn, MA

Confidential and Proprietary SciAps Inc 1

Page 2: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Topics

• How LIBS works,• Recent Advancements in HH LIBS:

• Argon purge – why it’s essential,• Laser technology improvements,• Laser power improvements,

• Safety and Regulatory advantages of laser‐based measurements,• Application Stories,• Questions

Confidential and Proprietary SciAps Inc 2

Page 3: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Introduction

• LIBS (Laser Induced Breakdown Spectroscopy)• Bench top technology for the past 30 years or so 

• Handheld LIBS is a new technology that is advancing rapidly,• Advantages include:

• Wide element range• Light Elements (Li, Be, B, C, N, O, F), • Improved performance on Mg, Al, Si, K, Ca, • Fast,• No X‐Rays

• Early users of handheld LIBS have complained of • Poor repeatability,• Reliance on sample preparation,• Weaker performance on key alloying refractory elements (e.g. Mo, V, W, Ta, Co)

• New advancements in both hardware and software are overcoming these issues 

Confidential and Proprietary SciAps Inc 3

Page 4: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Laser Induced Breakdown Spectroscopy (LIBS!)

Confidential and Proprietary SciAps Inc 4

• Similar to other Optical Emission techniques like Arc/Spark but utilizes a laser to excite sample

• Can measure any sample medium (powder, solid, liquid, nonconductive),

• Laser is about 50 um diameter

Page 5: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

One Advancement – Argon Purge

Confidential and Proprietary SciAps Inc 5

Page 6: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Argon Purge

• Long established spark optical emission technology, 

• A key to repeatable, quantitative chemistry is operating in an argon purge,

• LIBS is very similar to optical emission, but with a laser to produce the plasma rather than electric spark. 

Confidential and Proprietary SciAps Inc 6

Typical mobile spark OES analyzer, with tank of argon gas

Page 7: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Argon Purge Benefits

• Operating in an oxygen free environment reduces quenching of the plasma,

• Oxygen absorbs the plasma light emissions much more than an inert gas, 

• Depending on element/wavelength there is 3x‐10x more light when using argon,

• Eliminate interfering lines from vaporized elements forming oxides.  

Confidential and Proprietary SciAps Inc 7

Argon PurgeNo Argon Purge

Stainless 316

Page 8: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Argon, cont’d

Confidential and Proprietary SciAps Inc 8

• The Z uses a small spot, pulsed laser, • Argon flow rates are very low, • This supports a small, user‐replaceable argon cartridge in the analyzer’s handle. 

• Each canister supplies 250 tests,• Cost/test is 2 cents,• Analyzer can be connected to larger tanks for stationary operation. 

Page 9: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Another Advancement – Laser Cleaning and Rastering

Confidential and Proprietary SciAps Inc 9

Page 10: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Laser Cleaning

• LIBS is very sensitive to sample surface conditions,• Nobody in the alloy business wants to take the time to grind samples, • Laser cleaning means firing the laser at the same location multiple times to “burn off” surface material. 

• Fire the laser several times to clean, then several more times to collect spectral data for analysis. 

Confidential and Proprietary SciAps Inc 10

Page 11: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Cleaning Pulses:  Change in Ca signal

• By firing the laser in a single location, a change in the Calcium signal can be observed as the laser ablates away the surface contaminants,

• Firing at 10 Hz (10 laser pulses/second) is too slow – 1 sec. of cleaning time.

Confidential and Proprietary SciAps Inc 11

‐0.5

0

0.5

1

1.5

0 5 10 15 20Ca Signal (no

rmalize

d)

Pulse #

Calcium Signal with Consecutive Laser PulsesLocation 1Location 2Location 3

Page 12: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Advancement:  Bursting the laser

Confidential and Proprietary SciAps Inc 12

• Burst the laser at 50 Hz (50 pulses/second) for short periods of time,

• In 0.2 s, fire 10 cleaning shots,• Rapidly burn off material, more thoroughly, 

• Reduce to 10 Hz for data collection (0.3 s), 

• One location analyzed on sample in 0.5 sec. 

Page 13: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

The Challenge of Real‐world Samples

Confidential and Proprietary SciAps Inc 13

• Measuring Mg in an aluminum alloy,

• In OES and LIBS, you calibrate to intensity ratio’s:

Intensity of Mg lightIntensity of base metal light. • Poor cleaning affects both numerator and denominator,

• Mg measurement > 2x high without good surface cleaning.

Page 14: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Laser Rastering

• Why? Benchtop LIBS research shows better analytical results by averaging several locations,

• For alloys, theories are alloys not entirely homogeneous at the 50 um level (laser beam size),

• Grain boundaries, inclusions. • Recommendations from 30 years of LIBS research is to average results from multiple locations to improve repeatability. 

Confidential and Proprietary SciAps Inc 14

Video:  An example of a 12 Location raster.  Each location is shot with 2 cleaning bursts, followed by 3 “data” pulses

Page 15: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Laser Advancement• New Proprietary SciAps Laser 

• Provided best signal intensity compared to other similar off the shelf lasers

• 1064 nm,  Class 3B • Pulse Energy of 5‐6 mJ/pulse,  

• Compare to other handhelds that use “micro‐joule lasers” 0.1 mJ/pulse, 

• Shorter pulse duration:  1 ns instead of 6 ns• Better beam quality, more photons in smaller area• Variable Pulse Rates

• Normal operation at 10 Hz• Burst:  10 pulses in only 0.2 s (50 Hz), followed by gated, 10 Hz operation for 0.3 s, then raster to new location. 

Confidential and Proprietary SciAps Inc 15

Page 16: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Comparison to Handheld X‐ray Fluorescence (XRF)

Confidential and Proprietary SciAps Inc 16

Page 17: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Analytical Comparisons

• HH LIBS measures virtually every element in the Periodic Table,• H, Li, Be, B, C, N, O, F, Na  HHXRF cannot measure at all• Mg, Al, Si HHXRF measures poorly compared to 

LIBS• Most difficult elements for HH LIBS:    P and S   (in alloys)

S   (all materials)

Confidential and Proprietary SciAps Inc 17

Page 18: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Regulatory – X‐ray Requirements

• Annual operator training,• Body and ring badges,• Unannounced inspections from State agencies,• Administrative burden for tracking badges, training requirements and overall radiation safety,

Confidential and Proprietary SciAps Inc 18

Page 19: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Continued… 

11/23/2015 Confidential & Proprietary SciAps Inc 19

• Easier to handle small samples,• Hold them in your hand,• Don’t need to have test stand, laptop, etc.

Laser versus X‐ray

Page 20: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Maintenance Costs

• Far lower maintenance costs:• LIBZ typical repairs are $500, maximum $1,500, 

• X‐ray typical repairs are $6,000 ‐ $10,000 due to tube and detector,

• SciAps makes our own lasers and spectrometers and thus can repair at the component level. 

11/23/2015 Confidential & Proprietary SciAps Inc 20

Page 21: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Success Stories

Confidential and Proprietary SciAps Inc 21

Page 22: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Aluminum Alloys• Precise measurements 0.05 – 1% range of key alloying elements Mg, Si, Cr, Cu,• Fast sorting (1‐2 s) of close alloy grades

• 6061/6063/1100,   3003/3005/3105,    7050/7075,• Cast alloys, etc. 

Confidential and Proprietary SciAps Inc 22

Page 23: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Aluminum alloy producers• Looking at 1000 series grades, • Interest in grading alloys by low concentrations Mg, Mn, Cr, Si, etc.  

Confidential and Proprietary SciAps Inc 23

Page 24: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Low Alloy Steels• Looking at 0‐1% Cr, V, Ni, Cu, Mo, Mn in low‐alloy steels,• Evaluated air‐based HH LIBS couldn’t achieve needed precision compared to HHXRF,• Z‐200 successful

Confidential and Proprietary SciAps Inc 24

Page 25: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Nickel Superalloys:

11/23/2015 Confidential & Proprietary SciAps Inc 25

• Evaluated Z‐200 on their 70 grades of Ni, Co and Ti alloys.• Z‐200 reliably identified 70 out of 70, • Far better than any air‐based LIBS unit,• Superalloys very refractive, hard to get into plasma, thousands of lines,

• High laser energy, argon to maximize signal, high resolution spectrometer to minimize interferences. 

Page 26: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Thank you.  Questions?

Confidential and Proprietary SciAps Inc 26

Check out our demo videos:

Scrap Sorting  https://goo.gl/gZlLvnGeneral Apps Available  https://goo.gl/mnSxUO

Page 27: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

SciAps Models

• Z 200• 190 nm – 615 nm• Includes Argon purge for improved sensitivity

• Carbon at 193 nm*

• Z 500• 190 nm – 850 nm,• Mostly for geochem, other non‐alloy applications,

• Adds in O, N, F, Cl, Br, and improved K. 

Confidential and Proprietary SciAps Inc 27

The Argon Difference

Argon PurgeNo Argon Purge

Page 28: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Detect, Measure, Analyze.

Thank You for Attending Today’s Webinar

Your HostTom DouglasSalesRAECO-LIC [email protected]

Our Featured SpeakerDon SackettCEO & [email protected]

Page 29: Advancements in Handheld LIBS for Metal Alloy … in Handheld LIBS for Metal Alloy Analysis Don Sackett ... • For alloys, theories are alloys ... Comparison to Handheld X‐ray Fluorescence

Detect, Measure, Analyze.

Get Social with RAECO

• Our Website• www.raeco.com

• Our Blog• Blog.raeco.com

• Follow us on LinkedIn• www.linkedin.com/company/raeco-lic-llc

• Follow us on Twitter• @RAECOLIC

• Check Out our YouTube Channel• www.youtube.com/user/RaecoVideos