8
Acura® manual pipettes are precision instruments combining high-tech materials with exceptional, user friendly ergonomics. A choice of models brings much more than superior performance and makes pipetting safe and agreeable. The Acura® line features: Effortless activation Smart and reliable volume adjustment Precision digital display visible at all time Justip TM adjustable tip ejector fitting most tips Swift-set user calibration system Shock, UV-light and autoclaving resistance CE certified IVD 98/79 EEC The Acura® manual models 825 826 XS micro micro 815 835 855 fixed macro multi digital reading micropipettes Acura manual your expectations – our tradition ®

Acura manual your expectations – digital reading micropipettes · and makes pipetting safe and agreeable. ... • Smart and reliable volume adjustment • Precision digital display

Embed Size (px)

Citation preview

Acura® manual pipettes are precision instruments combining high-tech materials with exceptional, user friendly ergonomics. A choice of models brings much more than superior performance and makes pipetting safe and agreeable.

The Acura® line features:

• Effortless activation

• Smart and reliable volume adjustment

• Precision digital display visible at all time

• JustipTM adjustable tip ejector fitting most tips

• Swift-set user calibration system

• Shock, UV-light and autoclaving resistance

• CE certified IVD 98/79 EEC

The Acura® manual models

825 826 XS micro micro

815 835 855 fixed macro multi

digital reading micropipettes

Acura manualyour expectations – our tradition

®

AN

DE

FR

IT

ESPT

Universal micropipettesThe classic Acura® manual 825 is intended for all applications. It brings convenience, comfort, safety and sturdiness. For sure the wise choice for any laboratory.

Digital display always visible at a glance

Located on the front of the instru- ment, the display window allows set volume to be visible at all times during pipetting. A comforting reassurance for the user.

Acura® manual 825/835 Triopack

Nine different packs up to 10 mL, each containing three adjustable micro -pipettes, to cover all indivi-dual requirements in any research or routine application. Available at attractive saver prices, they match with today’s budget constraints.

Smart and reliable volume adjustment

State-of-the-art micrometric setting is performed smoothly and precisely by turning push button. Finely machined click-stops and free rotating cap prevent any unwanted volume alteration.

Most comfortable pipettingThe line fits everybody’s hand with or without glove. Ergonomic shape, light weight and soft plunger travel care for optimal comfort.

2

Acura manualyour expectations – our tradition

®

825micro

0.1 – 2 µL0.5 – 10 µL1 – 10 µL2 – 20 µL5 – 50 µL10 – 100 µL20 – 200 µL100 – 1000 µL

AN

DE

FR

IT

ESPT

Work station Twister universal 336Rotary stand holds up to six single chan-nel pipettes of most brands. Seven trans-lucent colours to choose from.

Universal work station 337Space saving stand holds up to seven single channel pipettes of most brands. Anti-slip pads guarantee bench stability. Available in five pastel colours.

Acura® manual XS 826TwiXS pack

Six different packs up to 1000 µL, each containing two adjustable micropipettes and a free pipette holder, to cover all individual requirements in any research application. Available at attractive saver prices, they match with today’s budget constraints.

3

Narrow conical shaft endSmall diameter provides for easy access to microtubes.

Short shaft

A must for research scientistsThe extra sharp pipette is optimal in any fine applications such as pipetting in microtubes or microplates.

Short shaftLength reduction greatly enhances precision of hand movement and guarantees superior drivability.

826 XSmicro extra sharp

0.1 – 2 µL0.5 – 10 µL1 – 10 µL2 – 20 µL5 – 50 µL10 – 100 µL20 – 200 µL100 – 1000 µL

AN

DE

FR

IT

ESPT

Extra smooth activation Unique tightness lip seal and new springs provide for ultra-soft pipetting, reducing hand fatigue while working.

Above forces, measured on a 20 – 200 µL model, are indicative of very limited finger efforts.

Yet, the overshoot stop offers a clear tactile indicator.

* 1 Newton (N) ~ 0.1 kilogramme force (kgf)

Tip ejection – easier than ever Most ergonomically positioned, the large surface, soft padded ejector button provides for low-pressure activation. In addition to ease of tip ejection, the shaft height adjustment system called Justip™ (4 mm span) – controlled by efficient click-

stops – allows a wide selection

of tips to tightly fit the nozzle.

0.1 - 2 µL0.5 - 10 µL1 - 10 µL2 - 20 µL5 - 50 µL10 - 100 µL20 - 200 µL100 - 1000µL

Volume

88 gr88 gr88 gr88 gr90 gr92 gr90 gr95 gr

Weight Inaccuracy (E%) Imprecision (CV%)

825

84 gr84 gr85 gr85 gr86 gr87 gr86 gr88 gr

<+/- 6.0 %1)

<+/- 2.5 %2)

<+/- 2.5 %<+/- 2.5 %<+/- 1.5 %<+/- 1.5 %<+/- 1.5 %<+/- 1.5 %

<+/- 4.0 %<+/- 1.8 %<+/- 1.8 %<+/- 1.8 %<+/- 1.3 %<+/- 1.2 %<+/- 1.1 %<+/- 1.0 %

<+/- 2.0 %<+/- 1.0 %<+/- 1.0 %<+/- 1.0 %<+/- 1.0 %<+/- 0.8 %<+/- 0.6 %<+/- 0.5 %

< 5.0 %1)

< 1.8 %2)

< 2.5 %< 1.7 %< 1.0 %< 1.0 %< 0.6 %< 0.5 %

< 3.3 %< 1.2 %< 1.6 %< 1.0 %< 0.7 %< 0.6 %< 0.4 %< 0.4 %

< 1.5 %< 0.5 %< 0.7 %< 0.5 %< 0.4 %< 0.2 %< 0.2 %< 0.2 %

826 Min. vol. Mid. vol. Max. vol. Min. vol. Mid. vol. Max. vol.

Performance values obtained with bidest. water at constant temperature (± 0.5°C) comprised between 20 and 25°C in accordance with ISO 8655. 1) measured at 0.5 µL 2) measured at 1 µL

Performance and weight Acura® manual 825 and 826 XS

Extra sharp micropipettesDerived from the Acura® manual 825, the 826 XS line was designed with the research scientists in mind. Shorter, smoother, lighter in weight, the eight models offer key features aiming at excellence in pipetting.

ULTRALIGHT

décomposé

Shelf pipette holder 332for two instruments

Fitting micro, macro and/or multi-channel pipettes, its attractive design accom modates all Acura® and Calibra® instruments. The self-adhesive strip supplied ensures lasting attachment on any surface.

4

1 µL5 µL10 µL10Y µL15 µL20 µL25 µL30 µL32 µL40 µL50 µL60 µL70 µL75 µL80 µL90 µL100 µL120 µL

85 gr85 gr85 gr85 gr85 gr85 gr89 gr89 gr89 gr89 gr89 gr89 gr89 gr89 gr89 gr89 gr89 gr89 gr

89 gr89 gr93 gr93 gr93 gr93 gr93 gr93 gr93 gr93 gr93 gr93 gr

105 gr111 gr111 gr120 gr

150 µL200 µL250 µL300 µL400 µL500 µL600 µL700 µL750 µL800 µL900 µL1000

Volume

Acura® manual 815

Weight Inaccuracy(E%)

Imprecision(CV%)

Volume Weight Inaccuracy(E%)

Imprecision(CV%)

<+/- 2.0 %<+/- 1.4 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.75 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %

< 1.0 %< 1.0 %< 0.7 %< 0.8 %< 0.5 %< 0.5 %< 0.6 %< 0.6 %< 0.6 %< 0.5 %< 0.4 %< 0.5 %< 0.4 %< 0.4 %< 0.35 %< 0.3 %< 0.3 %< 0.4 %

<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.7 %<+/- 0.6 %

< 0.3 %< 0.3 %< 0.4 %< 0.4 %< 0.3 %< 0.3 %< 0.3 %< 0.2 %< 0.2 %< 0.2 %< 0.2 %< 0.2 %

2 mL2.5 mL5 mL10 mL

<+/- 0.5 %<+/- 0.8 %<+/- 0.6 %<+/- 0.5 %

< 0.2 %< 0.5 %< 0.3 %< 0.2 %

Performance values obtained with bidest. water at constant temperature (± 0.5°C) comprised between 20 and 25°C in accordance with ISO 8655.

Acura® manual 815

Acura® manual 835F

Performance and weight Acura® manual 815 and 835F

Nozzle protection filterMacro models accommodate inter-changeable nozzle filters for added protection against liquid overflow and barrel contamination.

Pasteur pipette adapterOptional adapters allow fitting of glass Pasteur pipette on 2 and 5 mL models in addition to Polypropylene tips. The adapters include two O-rings guaranteeing proper holding and tightness of any standard 2 mL Pasteur pipettes (Ø 7 mm).

Smarties capscolour coding at will

Free rotating colour caps protecting against unwanted volume alteration, can be changed for individual, lab or department identifications. Pick among 14 different colours or choose the smartie mix including all of them for unlimited combinations.

815 / 835Ffixed

1 µL to 10 mL

AN

DE

FR

IT

ESPT

Fixed volume micro and macropipettesThey show most stable performance and provide consistent results for anyanalytical or routine tests. No risk of erroneous volume setting.Available in a selection of micro sizes from 1 to 1000 µL and macro volumes models of 2, 2.5, 5 and 10 mL.

Multi and single channel pipette work station 340Innovative design fits three Calibra® and Acura® models. Easy reach of each instrument. Heavy bottom plate and anti-slip pads guarantee high stability.

Acura manualyour expectations – our tradition

®

855multi

0.5 – 10 µL5 – 50 µL10 – 100 µL20 – 200 µL40 – 350 µL

MacropipettesOptimal for in- and outdoor environmental analysis, clinical chemistry and cell culture. They offer superior performance and make pipetting safer than ever.

Multichannel pipettes Instruments extend pipetting possibilities in 96-well microplates and guarantee outstanding performance and results.

5

0.2 - 2 mL0.5 - 5 mL1 - 10 mL

Volume Weight Inaccuracy (E%) Imprecision (CV%)

<+/- 1.5 %<+/- 1.5 %<+/- 1.5 %

<+/- 1.0 %<+/- 1.1 %<+/- 0.7 %

<+/- 0.5 %<+/- 0.6 %<+/- 0.5 %

< 0.5 %< 0.6 %< 0.5 %

< 0.3 %< 0.5 %< 0.3 %

< 0.2 %< 0.3 %< 0.2 %

Min. vol. Mid. vol. Max. vol. Min. vol. Mid. vol. Max. vol.

Volume

108 gr116 gr130 gr

Weight Inaccuracy (E%) Imprecision (CV%)

8x 12x Min. vol. Mid. vol. Max. vol. Min. vol. Mid. vol. Max. vol.

Performance values obtained with bidest. water at constant temperature (± 0.5°C) comprised between 20 and 25°C in accordance with ISO 8655. 1) measured at 1 µL

0.5 - 10 µL5 - 50 µL10 - 100 µL20 - 200 µL40 - 350 µL

155 gr165 gr165 gr165 gr168 gr

184 gr199 gr

201 gr203 gr

<+/- 3.5 %1)

<+/- 1.0 %<+/- 1.0 %<+/- 0.9 %<+/- 1.0 %

<+/- 2.5 %<+/- 0.9 %<+/- 0.9 %<+/- 0.8 %<+/- 0.9 %

<+/- 1.5 %<+/- 0.8 %<+/- 0.8 %<+/- 0.7 %<+/- 0.8 %

< 3.0 %1)

< 1.0 %< 0.9 %< 0.6 %< 0.6 %

< 2.0 %< 0.7 %< 0.6 %< 0.5 %< 0.5 %

< 1.0 %< 0.4 %< 0.4 %< 0.3 %< 0.3 %

Performance and weight Acura® manual 855

Performance and weight Acura® manual 835

AN

DE

FR

IT

ESPT

Optimal working positionIn addition to the ergonomic design and perfect hand fitting of the pipette, the multichannel volumetric module revolves to allow the selection of the best working position.

Adjustable tip ejectorHeight of tip ejector can be adjusted in no time within a 4 mm span to perfectly match to tip used. Bowed shape of ejector head guarantees effort- less, easy sequential tip ejection.

All purpose reservoirsWell adapted for multichannel pipettes, the reagent reservoirs offer various shapes in different configurations at economical cost. See page 6.

835macro

0.2 – 2 mL0.5 – 5 mL1 – 10 mL

6

Capacity Cat. No. Cat. No.QuantitySize QuantityStyle

10 µL

10 µL

10 µL

200 µL

200 µL

200 µL

200 µL

200 µL

200 µL

350 µL

1000 µL

1000 µL

1000 µL

1000 µl

2 mL

5 mL

10 mL

10 µL

10 µL

10 µL

20 µL

20 µL

100 µL

100 µL

200 µL

300 µL

300 µL

1000 µL

1000 µL

309.0010R

302.0010GR

302.0020R

327.0200R

308.0200R

308.0200GR

309.0200R

329.0200R

308.0350R

307.1000R

318.1000R

309.1000R

319.1000R

312.05R

312.10ER

309.0010FR

302.0010FR

302.0010AFR

308.0020FR

308.0020GFR

308.0100FR

308.0100GFR

308.0200LFR

308.0300FR

309.1000FR

309.0010B

302.0020B

307.0200B

327.0200B

308.0200B

309.0200B

329.0200B

308.0350B

307.1000B

318.1000B

309.1000B

319.1000B

312.02

312.05B

312.10

302.0010AFB

308.0020FS

308.0200LFS

308.0300AFB

309.1000FS

309.1000AFB

10 x 96/rack

4 x 204/rack

10 x 96/rack

10 x 96/rack

10 x 96/rack

4 x 204/rack

10 x 96/rack

10 x 96/rack

10 x 96/rack

10 x 60/rack

8 x 60/rack

8 x 60/rack

8 x 60/rack

2 x 50/rack

4 x rack alone

10 x 96/rack

10 x 100/rack

10 x 100/rack

10 x 96/rack

4 x 204/rack

10 x 96/rack

4 x 204/rack

8 x 204/rack

10 x 96/rack

10 x 100/rack

31.0 mm

70.9 mm

46.0 mm

49.4 mm

49.4 mm

51.1 mm

69.0 mm

49.7 mm

49.7 mm

58.8 mm

79.0 mm

69.5 mm

72.0 mm

72.0 mm

115.5 mm

123.2 mm

150.0 mm

31.0 mm

45.6 mm

45.6 mm

51.0 mm

69.0 mm

54.1 mm

69.0 mm

90.9 mm

58.8 mm

58.8 mm

88.6 mm

88.6 mm

1 x 1000/bag

1 x 1000/bag

2 x 500/bag

2 x 500/bag

1 x 1000/bag

2 x 500/bag

2 x 500/bag

1 x 1000/bag

4 x 250/bag

4 x 250/bag

4 x 250/bag

4 x 250/bag

2 x 250/bag

2 x 250/bag

3 x 100/bag

1 x 1000/bag

5 x 100/bag

5 x 100/bag

1 x 1000/bag

5 x 100/bag

1 x 1000/bag

Natural ultra-microtip

Natural gel loading tip

Natural microtip

Natural universal tip

Yellow universal tip

Natural tip

Natural gel loading tip

Natural tip

Yellow tip

Natural tip

Natural universal tip

Blue tip

Natural tip

Blue tip

Natural macrotip

Natural macrotip

Natural macrotip

Size Cat. No. Cat. No.Quantity

Also fit

BT, E, G

B, E, F

B, BT, E

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, G

B, BT, E, F, G

B, BT, E, F

B, BT, E, F, G

B, BT, E, F, G

E

G

F

B, BT, E, F, G

B, E, F

B, E, F

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, G

B, BT, E, G

B, BT, E, F

B, BT, E, F, G

Also fitCapacity QuantityColour, style

Rack packaging Bag packaging

Rack packaging Bag packagingDescription

Description

Sterile ultra-microtip

Sterile microtip

Autoclavable microtip

Sterile tip

Sterile gel loading tip

Sterile tip

Sterile gel loading tip

Sterile extended tip

Sterile tip

Autoclavable tip

Sterile tip

Autoclavable tip

Tip fitting: B=Brand®, BT=Sartorius-Biohit®, E=Eppendorf®, F=Thermo-Finnpipette®, G=Gilson®. Brand names belong to their respective owners.

Qualitips® with filterEfficient protection against aerosols. Include high-density PE filter for complete, uncontaminated sample retrieval. Certified DNAse and RNAse free. Sterile tips are guaranteed pyrogenfree.

All-purpose reagent reservoirs

Qualitips® pipette tipsThese superior pipette tips cover a range between 0.1 µL and 10 mL. All models autoclavable. Optimal tightness fitting on Socorex pipettes and broad compatibility with other brands.

Capacity Cat. No. Cat. No.QuantitySize QuantityStyle

10 µL

10 µL

10 µL

200 µL

200 µL

200 µL

200 µL

200 µL

200 µL

350 µL

1000 µL

1000 µL

1000 µL

1000 µl

2 mL

5 mL

10 mL

10 µL

10 µL

10 µL

20 µL

20 µL

100 µL

100 µL

200 µL

300 µL

300 µL

1000 µL

1000 µL

309.0010R

302.0010GR

302.0020R

327.0200R

308.0200R

308.0200GR

309.0200R

329.0200R

308.0350R

307.1000R

318.1000R

309.1000R

319.1000R

312.05R

312.10ER

309.0010FR

302.0010FR

302.0010AFR

308.0020FR

308.0020GFR

308.0100FR

308.0100GFR

308.0200LFR

308.0300FR

309.1000FR

309.0010B

302.0020B

307.0200B

327.0200B

308.0200B

309.0200B

329.0200B

308.0350B

307.1000B

318.1000B

309.1000B

319.1000B

312.02

312.05B

312.10

302.0010AFB

308.0020FS

308.0200LFS

308.0300AFB

309.1000FS

309.1000AFB

10 x 96/rack

4 x 204/rack

10 x 96/rack

10 x 96/rack

10 x 96/rack

4 x 204/rack

10 x 96/rack

10 x 96/rack

10 x 96/rack

10 x 60/rack

8 x 60/rack

8 x 60/rack

8 x 60/rack

2 x 50/rack

4 x rack alone

10 x 96/rack

10 x 100/rack

10 x 100/rack

10 x 96/rack

4 x 204/rack

10 x 96/rack

4 x 204/rack

8 x 204/rack

10 x 96/rack

10 x 100/rack

31.0 mm

70.9 mm

46.0 mm

49.4 mm

49.4 mm

51.1 mm

69.0 mm

49.7 mm

49.7 mm

58.8 mm

79.0 mm

69.5 mm

72.0 mm

72.0 mm

115.5 mm

123.2 mm

150.0 mm

31.0 mm

45.6 mm

45.6 mm

51.0 mm

69.0 mm

54.1 mm

69.0 mm

90.9 mm

58.8 mm

58.8 mm

88.6 mm

88.6 mm

1 x 1000/bag

1 x 1000/bag

2 x 500/bag

2 x 500/bag

1 x 1000/bag

2 x 500/bag

2 x 500/bag

1 x 1000/bag

4 x 250/bag

4 x 250/bag

4 x 250/bag

4 x 250/bag

2 x 250/bag

2 x 250/bag

3 x 100/bag

1 x 1000/bag

5 x 100/bag

5 x 100/bag

1 x 1000/bag

5 x 100/bag

1 x 1000/bag

Natural ultra-microtip

Natural gel loading tip

Natural microtip

Natural universal tip

Yellow universal tip

Natural tip

Natural gel loading tip

Natural tip

Yellow tip

Natural tip

Natural universal tip

Blue tip

Natural tip

Blue tip

Natural macrotip

Natural macrotip

Natural macrotip

Size Cat. No. Cat. No.Quantity

Also fit

BT, E, G

B, E, F

B, BT, E

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, G

B, BT, E, F, G

B, BT, E, F

B, BT, E, F, G

B, BT, E, F, G

E

G

F

B, BT, E, F, G

B, E, F

B, E, F

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, G

B, BT, E, G

B, BT, E, F

B, BT, E, F, G

Also fitCapacity QuantityColour, style

Rack packaging Bag packaging

Rack packaging Bag packagingDescription

Description

Sterile ultra-microtip

Sterile microtip

Autoclavable microtip

Sterile tip

Sterile gel loading tip

Sterile tip

Sterile gel loading tip

Sterile extended tip

Sterile tip

Autoclavable tip

Sterile tip

Autoclavable tip

Tip fitting: B=Brand®, BT=Sartorius-Biohit®, E=Eppendorf®, F=Thermo-Finnpipette®, G=Gilson®. Brand names belong to their respective owners.

Capacity Cat. No. Cat. No.QuantitySize QuantityStyle

10 µL

10 µL

10 µL

200 µL

200 µL

200 µL

200 µL

200 µL

200 µL

350 µL

1000 µL

1000 µL

1000 µL

1000 µl

2 mL

5 mL

10 mL

10 µL

10 µL

10 µL

20 µL

20 µL

100 µL

100 µL

200 µL

300 µL

300 µL

1000 µL

1000 µL

309.0010R

302.0010GR

302.0020R

327.0200R

308.0200R

308.0200GR

309.0200R

329.0200R

308.0350R

307.1000R

318.1000R

309.1000R

319.1000R

312.05R

312.10ER

309.0010FR

302.0010FR

302.0010AFR

308.0020FR

308.0020GFR

308.0100FR

308.0100GFR

308.0200LFR

308.0300FR

309.1000FR

309.0010B

302.0020B

307.0200B

327.0200B

308.0200B

309.0200B

329.0200B

308.0350B

307.1000B

318.1000B

309.1000B

319.1000B

312.02

312.05B

312.10

302.0010AFB

308.0020FS

308.0200LFS

308.0300AFB

309.1000FS

309.1000AFB

10 x 96/rack

4 x 204/rack

10 x 96/rack

10 x 96/rack

10 x 96/rack

4 x 204/rack

10 x 96/rack

10 x 96/rack

10 x 96/rack

10 x 60/rack

8 x 60/rack

8 x 60/rack

8 x 60/rack

2 x 50/rack

4 x rack alone

10 x 96/rack

10 x 100/rack

10 x 100/rack

10 x 96/rack

4 x 204/rack

10 x 96/rack

4 x 204/rack

8 x 204/rack

10 x 96/rack

10 x 100/rack

31.0 mm

70.9 mm

46.0 mm

49.4 mm

49.4 mm

51.1 mm

69.0 mm

49.7 mm

49.7 mm

58.8 mm

79.0 mm

69.5 mm

72.0 mm

72.0 mm

115.5 mm

123.2 mm

150.0 mm

31.0 mm

45.6 mm

45.6 mm

51.0 mm

69.0 mm

54.1 mm

69.0 mm

90.9 mm

58.8 mm

58.8 mm

88.6 mm

88.6 mm

1 x 1000/bag

1 x 1000/bag

2 x 500/bag

2 x 500/bag

1 x 1000/bag

2 x 500/bag

2 x 500/bag

1 x 1000/bag

4 x 250/bag

4 x 250/bag

4 x 250/bag

4 x 250/bag

2 x 250/bag

2 x 250/bag

3 x 100/bag

1 x 1000/bag

5 x 100/bag

5 x 100/bag

1 x 1000/bag

5 x 100/bag

1 x 1000/bag

Natural ultra-microtip

Natural gel loading tip

Natural microtip

Natural universal tip

Yellow universal tip

Natural tip

Natural gel loading tip

Natural tip

Yellow tip

Natural tip

Natural universal tip

Blue tip

Natural tip

Blue tip

Natural macrotip

Natural macrotip

Natural macrotip

Size Cat. No. Cat. No.Quantity

Also fit

BT, E, G

B, E, F

B, BT, E

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, G

B, BT, E, F, G

B, BT, E, F

B, BT, E, F, G

B, BT, E, F, G

E

G

F

B, BT, E, F, G

B, E, F

B, E, F

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, F, G

B, BT, E, G

B, BT, E, G

B, BT, E, F

B, BT, E, F, G

Also fitCapacity QuantityColour, style

Rack packaging Bag packaging

Rack packaging Bag packagingDescription

Description

Sterile ultra-microtip

Sterile microtip

Autoclavable microtip

Sterile tip

Sterile gel loading tip

Sterile tip

Sterile gel loading tip

Sterile extended tip

Sterile tip

Autoclavable tip

Sterile tip

Autoclavable tip

Tip fitting: B=Brand®, BT=Sartorius-Biohit®, E=Eppendorf®, F=Thermo-Finnpipette®, G=Gilson®. Brand names belong to their respective owners.

PPPPPSPSPSPSPSPS

70 mL125 mL48 mL48 mL48 mL48 mL48 mL48 mL

V-shapedV-shaped8-channel8-channel8-channel12-channel12-channel12-channel

Material

For 2 and 5 mL modelsFor 10 mL model

Description

250 / pk100 / pk

322.05322.10

Packaging Cat. No.

For 2 mL modelFor 5 mL model

Description

1 / pk1 / pk

1.835.6311.835.633

Packaging Cat. No.

Volume Style

NoHinged lid

NoNoNoNoNoNo

Lid

AutoclavableAutoclavable

NoNo

SterilizedNoNo

Sterilized

Sterilized

20/pack20/pack30/pack100/pack

10x 1/pack30/pack100/pack

10x 1/pack

Packaging

330.01330.01L330.08

330.08XL330.08.9330.12

330.12XL330.12.9

Cat. No.

Qualitips®

your premium pipette tips

Nozzle protection filter

PPPPPSPSPSPSPSPS

70 mL125 mL48 mL48 mL48 mL48 mL48 mL48 mL

V-shapedV-shaped8-channel8-channel8-channel12-channel12-channel12-channel

Material

For 2 and 5 mL modelsFor 10 mL model

Description

250 / pk100 / pk

322.05322.10

Packaging Cat. No.

For 2 mL modelFor 5 mL model

Description

1 / pk1 / pk

1.835.6311.835.633

Packaging Cat. No.

Volume Style

NoHinged lid

NoNoNoNoNoNo

Lid

AutoclavableAutoclavable

NoNo

SterilizedNoNo

Sterilized

Sterilized

20/pack20/pack30/pack100/pack

10x 1/pack30/pack100/pack

10x 1/pack

Packaging

330.01330.01L330.08

330.08XL330.08.9330.12

330.12XL330.12.9

Cat. No.PPPPPSPSPSPSPSPS

70 mL125 mL48 mL48 mL48 mL48 mL48 mL48 mL

V-shapedV-shaped8-channel8-channel8-channel12-channel12-channel12-channel

Material

For 2 and 5 mL modelsFor 10 mL model

Description

250 / pk100 / pk

322.05322.10

Packaging Cat. No.

For 2 mL modelFor 5 mL model

Description

1 / pk1 / pk

1.835.6311.835.633

Packaging Cat. No.

Volume Style

NoHinged lid

NoNoNoNoNoNo

Lid

AutoclavableAutoclavable

NoNo

SterilizedNoNo

Sterilized

Sterilized

20/pack20/pack30/pack100/pack

10x 1/pack30/pack100/pack

10x 1/pack

Packaging

330.01330.01L330.08

330.08XL330.08.9330.12

330.12XL330.12.9

Cat. No.

Pasteur pipette adapters

7

Ordering information – Instruments

Instant calibrationAll Acura® manual models include the swift-set user calibration system - with integrated key and locking mechanism - making tedious procedures a thing of the past. Easy and precise, it is in full compliance with international standards and recommendations.

Safety seal stickerThough easily removed when needed, the switch protection seal is autoclaving resistant. Replacement service stickers (5 ea.) included with each pipette.

Minimum maintenance / sterilisation

Easy disassembling for thorough cleaning.

Sterilisation achieved by autoclaving the instrument fully assembled (121°C / 250°F , 20 minutes).

Calibration seal stickers

25 / pk100 / pk

1.825.0251.825.100

Replacement sealsReplacement seals

Volume

0.5 - 10 µL5 - 50 µL

10 - 100 µL20 - 200 µL40 - 350 µL

Division

0.01 µL0.1 µL0.1 µL0.2 µL0.4 µL

Tip style

ultra 10 µL200 µL200 µL200 µL350 µL

8-channel/Cat. No.

855.08.010855.08.050855.08.100855.08.200855.08.350

12-channel/Cat. No.

855.12.010855.12.050

855.12.200855.12.350

Acura® 855, multichannel micropipettes

Pack

C

J

M

O

P

S

T

V

Y

Cat. No.

825.TRIO.C

825.TRIO.J

825.TRIO.M

825.TRIO.O

825.TRIO.P

825.TRIO.S

825.TRIO.T

825.TRIO.V

835.TRIO.Y

2 µL(0.1 - 2)

2 µL(0.1 - 2)10Y µL(1 - 10)10 µL

(0.5 - 10)10Y µL(1 - 10)20 µL

(2 - 20)50 µL

(5 - 50)100 µL

(10 - 100)1000 µL

(100 - 1000)

10 µL(0.5 - 10)

20 µL(2 - 20)50 µL

(5 - 50)100 µL

(10 - 100)100 µL

(10 - 100)200 µL

(20 - 200)200 µL

(20 - 200)1000 µL

(100 - 1000)5 mL

(0.5 - 5)

50 µL(5 - 50)200 µL

(20 - 200)200 µL

(20 - 200)1000 µL

(100 - 1000)1000 µL

(100 - 1000)1000 µL

(100 - 1000)1000 µL

(100 - 1000)5 mL

(0.5 - 5)10 mL(1 - 10)

Contents (Micro/macropipette) Pack

E

G

H

K

N

Q

Cat. No.

826.TWX.E

826.TWX.G

826.TWX.H

826.TWX.K

826.TWX.N

826.TWX.Q

2 µL(0.1 - 2)10 µL

(0.5 - 10)20 µL

(2 - 20)50 µL

(5 - 50)100 µL

(10 - 100)200 µL

(20 - 200)

20 µL(2 - 20)100 µL

(10 - 100)200 µL

(20 - 200)200 µL

(20 - 200)1000 µL

(100 - 1000)1000 µL

(100 - 1000)

Contents (micropipettes)

Triopack Acura® 825 and 835 TwiXS pack Acura® 826 XS

Cat. No.

835.F02835.F02.5835.F05835.F10

Volume

2 mL2.5 mL5 mL

10 mL

Tip style

2 mL5 mL5 mL

10 mL

Acura® 835F fixed volume macropipettes

825.0002825.0010825.0010Y825.0020825.0050825.0100825.0200825.1000

826.0002826.0010826.0010Y826.0020826.0050826.0100826.0200826.1000

Acura® 815, fixed volume micropipettes

Cat. No.

835.02835.02PP835.05

835.05PP835.10

Tip style

2 mL2 mL/Pasteur

5 mL5 mL/Pasteur

10 mL

Division

0.002 mL0.002 mL0.01 mL0.01 mL0.01 mL

Volume

0.2 - 2 mL0.2 - 2 mL0.5 - 5 mL0.5 - 5 mL1 - 10 mL

Acura® 835 adjustable volume macropipettes

Volume

0.1 - 2 µL0.5 - 10 µL1 - 10 µL2 - 20 µL5 - 50 µL

10 - 100 µL20 - 200 µL

100 - 1000 µL

Division

0.002 µL0.01 µL0.01 µL0.02 µL0.1 µL0.1 µL0.2 µL1 µL

Tip style

10 µL10 µL

200 µL200 µL200 µL200 µL200 µL

1000 µL

Acura® 825 and 826 XS, adjustable volume micropipettes

Volume

50 µL60 µL70 µL75 µL80 µL90 µL100 µL120 µL150 µL200 µL

Tip style

200 µL200 µL200 µL200 µL200 µL200 µL200 µL200 µL200 µL200 µL

Cat. No.

815.0050815.0060815.0070815.0075815.0080815.0090815.0100815.0120815.0150815.0200

Volume

250 µL300 µL400 µL500 µL600 µL700 µL750 µL800 µL900 µL1000 µL

Tip style

1000 µL1000 µL1000 µL1000 µL1000 µL1000 µL1000 µL1000 µL1000 µL1000 µL

Cat. No.

815.0250815.0300815.0400815.0500815.0600815.0700815.0750815.0800815.0900815.1000

Cat. No.

Volume

1 µL5 µL10 µL10Y µL15 µL20 µL25 µL30 µL32 µL40 µL

Tip style

10 µL10 µL10 µL200 µL200 µL200 µL200 µL200 µL200 µL200 µL

Cat. No.

815.0001815.0005815.0010815.0010Y815.0015815.0020815.0025815.0030815.0032815.0040

Ordering information – Pipette multipacks

digital reading micropipettes

Acura manualyour expectations – our tradition

®

SOCOREX ISBA SAChamp-Colomb 7a, P.O. Box1024 Ecublens/[email protected]. +41 (0)21 651 6000 B

.AC

UX

.E -

B1

11

6

Shelf pipette holder 332Packaging

1 / pk4 / pk

Cat. No.

320.332320.332.4

Universal pipette work station 337

Smartie plunger caps

Packaging

1 / pk1 / pk1 / pk1 / pk1 / pk

Cat. No.

320.337G320.337R320.337B320.337M320.337Y

Colour

WhiteGreyVanillaLemonYellowOrangeRed

Colour

RosePinkPurpleIce blueBlueMintGreen

Cat. No.

1.825.7191.825.7201.825.7211.825.7221.825.7131.825.7231.825.711

Cat. No.

1.825.7101.825.7151.825.7161.825.7121.825.7171.825.7181.825.714

Assorted colours (14 / pk) Smartie mix 1.825.700

Alternative colour selection (6 / pk)

Colour

Light greyPastel roseIce blueMint greenVanilla yellow

Colour

Transparent blue

Twister universal work station 336Packaging

1 / pk1 / pk1 / pk1 / pk1 / pk1 / pk1 / pk

Cat. No.

320.336W320.336Q320.336Y320.336O320.336R320.336G320.336B

Colour

Diamond whiteQuartz greyCitrine yellowTopaz orangeRubin redEmerald greenSapphire blue

Pipette work station 340Packaging

1 / pk

Cat. No.

320.340

Colour

Light grey

Standards and complianceAll Acura® manual models are designed to perform in accordance with national and international standards such as ISO 8655, GMP, GLP, NCCLS.

They conform to the CE directive IVD 98/79 EEC (in vitro diagnostic medical devices).

QC and warrantyPerformance values are obtained with bi-dest. water at constant temperature (± 0.5°C) between 20 and 25°C, in accordance with ISO 8655.Each micropipette bears its own serial number and passes strict performance control attested by an individual QC certificate.

Three-year warranty.

Ordering information – AccessoriesS+

Your local distributor

AN

DE

FR

IT

ESPT

332 336 337 340