Activity 1 Lab report Quan.docx

Embed Size (px)

Citation preview

  • 8/14/2019 Activity 1 Lab report Quan.docx

    1/4

    Activity 1

    Calibration of Pipets

    Briana Halbert

    9/17/2013

    Dr. Kimaro

  • 8/14/2019 Activity 1 Lab report Quan.docx

    2/4

    Purpose:

    The purpose of this experiment was to calibrate three different pipets.

    Objectives:

    The objectives of this laboratory experiment was to learn the usage of volumetric glassware, proper procedures in

    transferring solutions, the difference between each pipet, determining the actual volumes of the three different typeof pipets, and how to use an analytical balance.

    Procedure:

    The introductory sections, Mass Measurements and Volumetric Glassware were referred to, prior to using the

    balance and pipets. About 250 mL of deionized or distilled water was placed in a 400 mL beaker. When the

    temperature of the water was not at room temperature, it was left to be equilibrated to room temperature. The

    temperature of the equilibrated water was recorded using a thermometer. The thermometer was placed in the beakecontaining water, and left in place during the experiment. If the temperature of the water changed during the

    experiment, the changed temperature was recorded. The temperature recorded, prior to each mass measurement, wa

    used to do calculations.The Erlenmeyer flasks were labeled, 1, 2, and 3. Flask 1 was used with volumetric pipet; flask 2 was used with the

    serological pipet; flask 3 was used with the Mohr pipet. A rubber or cork stopper was placed on each flask, and the

    flasks were weighed. The masses were recorded.

    Upon proceeding to the next step, we practiced using the pipets. The volumetric pipets were allowed to drain into tvessel. The tip of the pipet touched the side of the vessel to remove the last drop. Mohr pipet delivered the volumes

    between calibrated marks on the sides. Serological pipet was allowed to completely drain into the vessel, while the

    remaining solution was blown into the vessel.The volumetric pipet was filled with room temperature water to the line. And transferred to flask 1. The cleaned

    serological pipet delivered the water to flask 2. The Mohr pipet delivered the water to flask 3. Each flask had a cork

    stopper placed on it follow by weighing them. The masses were recorded. The necks of the flask were not wet with

    water. Each pipet was used with its respectable flask seven more times. After adding the water each time, the flaskwere weighed. The masses were recorded.

    Data/ Calculations:

    Temperature of distilled water: 22C.

    () () Standard Deviation of Flask 1: Volumetric Pipet

    () ()

    % Standard Deviation

  • 8/14/2019 Activity 1 Lab report Quan.docx

    3/4

    % Standard Deviation of Flask 1: Volumetric Pipet

    Results:

    Temperature of distilled water: 22C.

    Flask 1: Volumetric Pipet

    Volume (mL)Mass of Flask,

    stopper & Water(g)Temperature (C)

    Mass of Water

    Transferred (g)

    Calculated

    Volume (mL)

    0 42.3206 22 0 0

    5 47.3474 22 5.0268 5.0429

    10 52.3433 22 4.9959 5.0119

    15 57.3433 22 5.0133 5.0294

    20 62.3497 22 4.9931 5.0091

    25 67.3480 22 4.9983 5.0143

    30 72.3439 22 4.9959 5.0119

    35 77.3262 22 5.0000 5.0161

    40 82.3154 22 4.9892 5.0051

    Flask 2: Serological Pipet

    Volume (mL)Mass of Flask,

    stopper & Water(g)Temperature (C)

    Mass of Water

    Transferred (g)

    Calculated

    Volume (mL)

    0 40.0396 21 0 0

    5 45.0615 21 5.0219 5.0370

    10 50.0795 21 5.0180 5.0331

    15 55.1089 21 5.0294 5.0445

    20 60.0030 21 4.8941 4.9088

    25 65.0564 21 5.0534 5.0686

    30 70.0541 21 4.9977 5.0127

    35 75.0756 21 5.0215 5.0366

    40 80.0823 21 5.0067 5.0218

  • 8/14/2019 Activity 1 Lab report Quan.docx

    4/4

    Flask 3: Mohr Pipet

    Volume (mL)Mass of Flask,

    stopper & Water(g)Temperature (C)

    Mass of Water

    Transferred (g)

    Calculated

    Volume (mL)

    0 38.4509 22 0 0

    5 43.4478 22 4.9969 5.012910 48.4535 22 5.0057 5.0217

    15 53.4661 22 5.0126 5.0287

    20 58.4639 22 4.9978 5.0138

    25 63.4836 22 5.0197 5.0358

    30 68.5703 22 5.0867 5.103

    35 73.5806 22 5.0103 5.0264

    40 78.5928 22 5.0122 5.0283

    During the volumetric experiment, the temperature read 22C.

    The density was 0.9968 g/mL.

    During the Mohr experiment, the temperature was 22C.

    The density was 0.9966 g/mL.

    During the Serological experiment, the temperature was 21C.

    The density was 0.9970 g/mL.

    Volumetric Serological Mohr

    Standard Deviation 0.0124 0.0479 0.02899

    Average (mL) 5.017875 5.0204 5.033825

    RSD (%) 0.31% 0.95% 0.58%

    Discussion:

    The volumetric pipet was the most accurate in measuring the distilled water. Volumetric pipet had the lowe

    standard deviation. The serological and Mohr pipets were off by 0.04 and 0.03. After calibrating a 5mL pipet givin

    the masses 4.975, 4.9845, 4.9800, and 4.9820g at 26C, the individual volumes of the pipets were 5.0000, 5.0050,

    5.0005, and 5.0025. The average volume was 5.0020, and the standard of deviation was 0.0023.

    Conclusion:

    According to the results of the experiment, the volumetric pipet was the most precise with the lowest standard

    of deviation. The serological was the least precise. Therefore, the Mohr pipet was between the volumetric and

    Serological with the standard of deviation. Moreover, the conclusion of this experiment was drawn that this lab

    experiment is the calibration of the pipets: volumetric, serological, and Mohr pipets are precise but not accurate.

    Better measurement and practice of the experiment would have made the experiment more precise.