Brain Products is proud to present actiCAP, another ground-
breaking development aimed at raising EEG & ERP research
to the next level. actiCAP combines active electrodes with a
new type of integrated noise subtraction circuits delivering
even lower noise levels than any other „normal“ active
electrodes achieve.
But that‘s not all: actiCAP can be connected to nearly any
existing research EEG amplifier system. All of our amplifiers
as well as most of our competitors’ amplifiers will work with
actiCAP. If you decide to switch from passive to active electrodes
you will no longer be forced to exchange your entire recording
system. Simply keep your existing hardware, and the
acquisition and analysis software you are used to working with
and add the actiCAP to reduce your preparation time e.g. for 64
channels down to 5 minutes.
The third generation of active electrodes
Enlighten your ResearchRecording EEG the progressive way
Technical Specification:
Classification MDD 93/42/EEC, Class IIa
Patient Safety Protection class II, EMC tested, Type BF (DIN EN IEC 60601)
Safety requirements:
The third generation of active electrodes
actiCAP control software
ww
w.b
rain
pro
du
cts
.co
m
Amplification 1 (Tolerance < 0,001%)
Input impedance > 200 MOhm
Bandwith DC - 5000 Hz
Noise < 2 µV p.p. (0.1 - 35 Hz)
Dynamic range ±1000 mV
Impedance measurement frequency
30 Hz
Impedance working range 1 to 120 kOhm
Impedance display LED at electrode(colors: red, yellow, green)
Default impedance values green < 25 kOhmyellow 20-60 kOhmred > 60 kOhm
Test signal generation Sinewave 200 µV
actiCAP control software Threshold selection, impedance display (values storable), topographic verification setup, USB connection
Configuration 16, 32, 64, 96, 128 channels
Power 4 AA rechargeable NIMH batteries (1.5 V) or USB powered
Power indication Good/Bad (less than 2 hours remaining)
Working temperature 0°C / +40°C
Software requirements Windows XPWindows Vista
Other Specifications:
Technical changes can be made without prior notification.
ww
w.b
rain
pro
du
cts
.co
m
Version 005 01/2010© Brain Products GmbH
Further information can be obtained from your local distri-butor or directly from:
Brain Products GmbHZeppelinstrasse 782205 Gilching (Munich)Germany
Tel. +49 (0) 8105 733 84 0Fax +49 (0) 8105 733 84 [email protected]
www.brainproducts.com
Worldwide Caution:Limited by United States Law to Investigational Use. ID in accordance with 801.1 For research use only.
ISO 13485:2007
MDD 93/42/EEC
Components:actiCAP control box, Electrode sets (16-/32-/64-/96-/ 128-channel sets), Ground/reference electrode, Cap with holders, USB cable, Batteries & charger, Amplifier connection cable and adapter, acti-CAP control software, Manual