0535These instructions for use, Accutron MR, version 2 GB,

apply to the high-pressure injector Accutron MR.

05/2008

Page 3: AccutronMR GA V2 GB - Weebly

Contents1 IntroductionRegarding these instructions for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5

Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6Manufacturer's plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8General conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8

2 Safety instructionsProtection of people and property . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11

3 Product descriptionGeneral features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13Items supplied and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14

Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14Consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14

Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16

Connections on the control panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16Controls mounted on the control panel . . . . . . . . . . . . . . . . . . . . . . . . . . .17The touch screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18The functions of the touch screen buttons . . . . . . . . . . . . . . . . . . . . . . . .20On/Off switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22Connection for the charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22Castors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22Remote control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22Handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22

4 Commissioning the injectorCharging the battery cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23Switching ON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24

Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24Stand-by function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25Display state of charge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25

Setting the touch screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26Setting the touch screen brighter or darker . . . . . . . . . . . . . . . . . . . . . . . .26Selecting a language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27Calibrating touch screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27Setting stand-by time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28Setting idling speed of the pistons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28Setting filling speed of the pistons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28

Preparing injector head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30Moving the pistons back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30

Page 4: AccutronMR GA V2 GB - Weebly

Inserting the syringes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32Connecting the tube system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

5 Operating the injectorFilling injector with contrast medium and NaCl . . . . . . . . . . . . . . . . . . . . . .35

Move both pistons forward . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Venting the tube system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

Changing the maximum operating pressure . . . . . . . . . . . . . . . . . . . . . . . .43Selecting an injection profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Changing an injection profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45

Storing changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46Overwriting an existing profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47Entering a profile name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48Saving changes under a new profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49Creating a new profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Adding a new phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52Deleting a phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53

Performing an injection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54Maintaining vein access with the KVO function . . . . . . . . . . . . . . . . . . . . .55Starting the injection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56Interrupting the injection on the remote control . . . . . . . . . . . . . . . . . . . . .57Interrupting the injection on the injector . . . . . . . . . . . . . . . . . . . . . . . . . . 57Injection end . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

Replenishing injector with contrast medium/NaCl . . . . . . . . . . . . . . . . . . . . 59Enter filling volume and fill . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60Venting after replenishing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61

Post-examination work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64

6 AnnexWaste disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65Safety inspections and maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65Cleaning and storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65System messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

Page 5: AccutronMR GA V2 GB - Weebly

1 IntroductionThis chapter contains some preliminary remarks on the use of the injector Acc-utron MR with Easy Loading Syringe (ELS) as well as explanations on the structure of these instructions for use and on the use of symbols and text markings.

Regarding these instructions for useThese instructions for use explain how to correctly commission and operate the Accutron MR for your application.

MEDTRON wants you to be able to use your Accutron MR safely, correctly and efficiently. To this end, it is essential that you read these instructions for use thoroughly before you use the injector. They contain important instructions designed to assist you in avoiding risks and, at the same time, increasing both the reliability and service life of the injector and its accessories.

For your own safety you should read the section Safety instructions. Comply exactly with all the instructions to ensure that you do not put yourself, other employees or patients at risk and prevent damage to the injector.

If you have any questions concerning the handling of the Accutron MR not answered by these operating instructions, please contact:

Hauptstraße 255 D-66128 Saarbrücken

Telephone +49 (0)6 81/970 17-0 Telefax +49 (0)6 81/970 17-20

Accutron MR 05/2008 5

Page 6: AccutronMR GA V2 GB - Weebly

6

Regarding these instructions for use 1 Introduction

Intended useThe injector Accutron MR is intended exclusively for the injection of contrast media and physiological saline solution into patients for examinations involv-ing magnetic resonance imaging (MRI). Any use above and beyond this scope is deemed as inappropriate. Never use the Accutron MR for other purposes!

Manufacturer's plateThe manufacturer's plate of the Accutron MR bearing the model name and serial number is located on the rear of the operating unit. You will require this information if you contact the MEDTRON Customer Service, e.g. if you want to order accessories or spare parts.

Enter the model name and serial number here so that you have them at hand when required:

Model name: ____________________________________________ Serial number: ___________________________________________

2006

05/2008 Accutron MR

Page 7: AccutronMR GA V2 GB - Weebly

1 Introduction Regarding these instructions for use

ContentsThis manual describes the commissioning and operation of the high-pressure injector Accutron MR. It consists of the following chapters:– The chapter Safety instructions contains important safety instructions

which are essential to observe for the safety of people and the system.– The chapter Product description introduces you to the injector with its

functions and controls.– The chapter Commissioning the injector describes how to set the touch

screen to suit your needs, how to prepare the injector heads and connect the tube system.

– The chapter Operating the injector explains how to fill the contrast medium and saline solution, how to deal with the injection profiles and how to perform the injection.

– In the Annex you will find notes on inspection and maintenance work, cleaning and storage, a table with system messages and errors and their remedy as well as the technical specifications of the injector and the declaration of conformity.

– The Index helps you to quickly find the desired text passages you want.

Accutron MR 05/2008 7

Page 8: AccutronMR GA V2 GB - Weebly

8

Regarding these instructions for use 1 Introduction

Symbols usedYou will find the following two symbols identifying important safety instructions at several points in this manual:

The following symbol indicates tips and particularities which facilitate the oper-ation of the device.

General conventions

Bold type print

Designations and lettering of menus, menu items and buttons are printed in bold type.

Italic type print

Cross references to other chapters or text passages are printed in italics.

Inverted commas

Device functions are placed in "inverted commas".

Control steps

Control steps are identified as follows:• Make sure that the piston is moved back.

• Push the syringe into the recipient until it engages audibly.

• Remove the protective cap from the syringe.

Listings

Listings are marked as follows:– Variant A– Variant B– Variant C

CAUTION! This symbol indicates risks which could involve fatal or serious personal injuries or severe damage to property.

ATTENTION! This symbol highlights risks which could involve material damage.

Note! If the unit has been out of service for a prolonged period, charge the battery cells before starting it up again.

05/2008 Accutron MR

Page 9: AccutronMR GA V2 GB - Weebly

2 Safety instructions

– Do not connect the patient to the system until you have vented the syringes and the tube system.

– Vent the syringe and the tube system every time after filling the injector with contrast medium or saline solution.

– Make sure before every injection that there are no air inclusions in the syringes and the connected tubes.

– Only start an injection when the syringes are in the lowered position (injection position).

– Make sure before every injection that the injection parameters shown on the display do not endanger the patient.

– Maintenance and repair work may only be performed by the MEDTRON Customer Service or persons authorised by MEDTRON. Authorised persons will receive the documents necessary for the maintenance and safety inspections from MEDTRON.

– Make sure that the injector is no longer operated when the display of the injector has only failed slightly or fails completely.

Read the following safety instructions thoroughly and adhere to them closely. They are intended to protect your own safety and the safety of colleagues and patients as well as to avoid damage to the Accutron MR and its accessories.

Protection of people and propertyCAUTION! Risk of fatal or serious injuries due to air embolisms!

CAUTION! Risk of uncontrolled supply of contrast medium due to hydrostatic pressure compensation! Make sure that the patient and the syringes are at the same level.

Accutron MR 05/2008 9

Page 10: AccutronMR GA V2 GB - Weebly

10

Protection of people and property 2 Safety instructions

– Replace all disposable articles with new ones after they have been used once. Otherwise, you jeopardise your patients' health.

– Do not remove the protective caps of the syringes and the tube ends until you have made the connections.

– Only use consumables and accessories which MEDTRON recommends.

– Make sure that the packings of the disposable articles do not show any signs of damage.

– Make sure that the 'use-by' date of the disposable articles has not been passed.

CAUTION! Risk of infection!

CAUTION! Danger from electric current!

– Make sure that no liquid can enter the injector and the peripheral equipment. Should this happen, switch off the device immediately, and secure it from being switched on again.

– Make sure that the electric connection cables are intact and cannot be bent or squeezed. If you discover any damage, pull out the mains plug and secure the remote control from being switched on again and do not continue to use the charger.

ATTENTION! Risk of malfunctions of the Accutron MR! Only use syringes and tubes recommended by MEDTRON.

05/2008 Accutron MR

Page 11: AccutronMR GA V2 GB - Weebly

2 Safety instructions Electromagnetic compatibility

Electromagnetic compatibilityThe injector is subject to special safety precautions as regards EMC. Installation and commissioning are performed by trained staff authorised by MEDTRON.

The injector satisfies the requirements of the standard DIN EN 60601-1-2 for medical equipment in terms of the emitted interference and interfer-ence immunity.

You can request the document Tables on EMC Rev00 from MEDTRON.

ATTENTION! Portable and mobile HF communications equipment may affect the injector. Make sure that the injector is not positioned near such equipment.

ATTENTION! The injector may only be used with the accessories described in the instructions for use. Operation of the injector with other accessories or equipment may result in increased emitted interference or a reduction in the interference immunity.

Accutron MR 05/2008 11

Page 12: AccutronMR GA V2 GB - Weebly

12

Electromagnetic compatibility 2 Safety instructions

05/2008 Accutron MR

Page 13: AccutronMR GA V2 GB - Weebly

3 Product description

General featuresThe Accutron MR is designed for exactly dosed injections of contrast medium (CM) and physiological saline solution (NaCl). You can store the injection parameters you want– Delay time– Volume,– Concentration– Flow rate and– Injection time

in up to 80 profiles and retrieve them at any time. The reproducibility of the injection parameters is therefore reliable and easy. A profile can consist of a maximum of 6 individually programmable injection phases which are per-formed automatically after the program has been started. With the parameter "Concentration", you can, if required, inject contrast medium and NaCl at the same time.

The Accutron MR has two injection units which can be controlled independ-ently of each other and so the contrast medium can be positioned variably and accurately in the examination area with NaCl, depending on the application. This prevents tailings of the contrast medium which previously made delimita-tion of the examination area more difficult. The function "Keep-Vein-Open" (KVO) injects 0.5 ml of NaCl every two min-utes and therefore ensures that access to the vein is retained.

Thanks to the use of microprocessor-controlled electric motors, you can also control the piston speeds during idling and when filling the syringe manually via the keyboard.

The injector runs independently of the mains network as it is wireless and is operated with battery cells. By means of the remote control you can control the high-pressure injector from the control room.

Accutron MR 05/2008 13

Page 14: AccutronMR GA V2 GB - Weebly

14

Items supplied and accessories 3 Product description

Items supplied and accessoriesThe system of the Accutron MR comprises the following elements:

Equipment– Mobile and wireless injector with swivelling injections unit and recipients for

holding syringes, control panel with touch screen, battery cells in the unit base and separate charger, instructions for use

– Touch screen remote control with power supply unit and transmitter, data transmission via radio, instructions for use

Consumables– Disposable syringe 200 ml– MR tube system– Patient tube

Assemblies

1 Injector base with battery cells and connector for charger

2 Receiving aerial for remote control

3 Column stand

4 Bottle holders for contrast medium and NaCl bottles

5 Control panel with touch screen

6 Swivelling injection unit for contrast medium

7 Recipient for holding the syringe for contrast medium

8 Recipient for holding the syringe for NaCl

9 Swivelling injection unit for NaCl

10 On/Off switch

11 Four smooth-running castors, the front two with locking brakes

05/2008 Accutron MR

Page 15: AccutronMR GA V2 GB - Weebly

3 Product description Assemblies

Accutron MR 05/2008 15

Page 16: AccutronMR GA V2 GB - Weebly

16

Controls 3 Product description

ControlsThe following contains a description of the controls of the Accutron MR.

Connections on the control panelThere are three connections on the rear of the control panel which are irrele-vant for the operation of the injector.– Two connections for the LWL cables to the limit switches of the pistons.

The cables always remain connected.– A service connector for the MEDTRON Customer Service

05/2008 Accutron MR

Page 17: AccutronMR GA V2 GB - Weebly

3 Product description Controls

Controls mounted on the control panelThe control panel of the Accutron MR consists of buttons on a keypad and a central touch screen, a combination of display and entry field.

No. Symbol Designation Function1 Left-hand button group for manual piston movements for CM

Piston forward While pushing the button the piston moves forward

Increase piston speed

Increase piston speed on each push of the button

Piston back While pushing the button the piston moves back

2 Touch screen To display and enter injection parameters

3 Right-hand button group for manual piston movements for NaCl

4 Interrupt injection To interrupt the injection

5

Pilot light

– lights up green when the injector is switched on – flashes green during the injection – lights up red in the event of a fault

6 Start injection To start the injection

7 Wake-up button To activate the display

Accutron MR 05/2008 17

Page 18: AccutronMR GA V2 GB - Weebly

18

Controls 3 Product description

The touch screenThe touch screen displays messages and graphics like a normal monitor. In addition, you can make entries and call up functions by touching the buttons displayed. If a button cannot be used at the moment, it is greyed.

Touch the symbol you want lightly with your finger tip. Depending on the but-ton, you can mark or select a value or call up a function and a new window will open.

In addition to entering commands, the main window of the touch screen also serves to display the process information currently set. The profile name is located in the text field in the top left and the parameters of the injection profile currently selected are in the table underneath.

ATTENTION! Damage to the touch screen possible! Do not strike or hit the touch screen. Do not use any items such as pencils or tools. Only touch the buttons lightly with your fingers.

Prof. Nr. 5

Phase Delays

Volumeml

Conc.%

Flowml/s

Inj. times

1

2

3

20

20

20

10

10

10 100

65

100 6.0

6.0

6.0 0

0

0

Not airfree

180 178

05/2008 Accutron MR

Page 19: AccutronMR GA V2 GB - Weebly

3 Product description Controls

The column headers for the injection parameters at the top of the display describe the following parameters for each phase:

During the injection the touch screen also shows status information on the cur-rent injection in two windows. With the Info button you can switch between the windows.

Phase Numbering of the individual phases

Delay (s) Breaks between the injection phases in seconds

Volume (ml) The amount of fluid, in millilitres, to be injected

Concentration (%)

Concentration of the injection media:– 100% means only contrast medium– 40% means 40% contrast medium, 60% NaCl– 0% means only NaCl

Flow (ml/s) Injection speed in millilitres per second

Inj. time (s) Injection time in seconds

The remaining contrast medium in millilitres is shown below the green syringe symbolThe remaining NaCl in millilitres is shown below the blue syringe symbol

The battery symbol shows the current state of charge of the battery cells.

Injected volume CM Contrast medium injected so far

Injected volume NaCl NaCl injected so far

Phase time (s) Duration of the current injection phase

Delay (s) Breaks between the injection phases

Scan delay (s) Delay between injection start and scanner start (can only be set on the remote control)

Injection time Counts the time from pressing the start button

KVO Indicates whether the KVO function (Keep Vein Open) is activated

Accutron MR 05/2008 19

Page 20: AccutronMR GA V2 GB - Weebly

20

Controls 3 Product description

Example of dialogue messages:

You answer dialogue messages by touching the relevant button. In the example shown above you can answer the question "CM-syringe air-free?" either using the button with the green symbol for Confirm or the button with the red cross for No/Reject.

The functions of the touch screen buttons

Symbol Designation Function

Call up profile To call up an existing profile

Save profile To save a profile

Confirm To confirm inquiries and changes

Reject To reject unstored entries

Back To scroll backwards within a menu

Forward To scroll forwards within a menu

Keep Vein Open To activate and deactivate the KVO function in order to maintain access to the vein

Call up filling menu To call up the filling menu

Fill syringes To enter the filling volume in the filling menu

Piston forward To move the piston forward in the filling menuTo scroll upwards within a list

CM-syringe airfree?

Venting

05/2008 Accutron MR

Page 21: AccutronMR GA V2 GB - Weebly

3 Product description Controls

Piston back To move the piston back in the filling menuTo scroll downwards within a list

Add phase To add a new phase to an injection profile

Delete phase To delete a phase

Interrupt injection To temporarily stop the injection

Set system To open the System settings menu

Calibrate touch screen To calibrate the touch screen

Save system settings To save changed system settings

Info To change between the two windows with the injection data

Activate To prepare the injector for the injection

Continue To continue the injection after an interruption with the HOLD button

NaCl volume To select preset filling volumes of NaCl: 50, 100, 200 ml

CM volume To select preset filing volumes of contrast medium: 20, 65, 100 ml

Leave menu To leave the menu back to the main window

Airfree To confirm the previously performed venting operation

Symbol Designation Function

Accutron MR 05/2008 21

Page 22: AccutronMR GA V2 GB - Weebly

22

Controls 3 Product description

On/Off switchThe Accutron MR has an On/Off switch on the injector base. When you have switched on the injector with this button, the integrated lamp of the switch comes on and the pilot light on the control panel lights up green. By pressing the button again, you switch the injector off again and both lights go off.

Connection for the chargerA socket is located on the side of the injector base. Here you can connect the connecting cable to the charger when you want to charge the battery cells.

CastorsYou can conveniently move the injector using the smooth-running castors to the location you want. When it is in the correct position, secure the injector using the two locking castors. To this end press the locking lever downwards with your foot. When you want to release the castors, raise the locking lever slightly with your foot.

Remote controlThe injector can be remotely controlled using the remote control. The data are transferred to the injector by radio. You will find more details in the instructions for use of the touch screen remote control.

HandleThere is a connecting bar between the two swivelling injection units. With this bar you lower the injector heads backwards into the injection position after establishing all the injection parameters. To prepare a new injection, pull the upper injector section on the connecting bar forwards into the vertical position.

Note! Secure the injector position using its locking castors before you swivel the injector head forwards or backwards.

Note! Secure the injector position using its locking castors before you swivel the injector heads forwards or backwards.

05/2008 Accutron MR

Page 23: AccutronMR GA V2 GB - Weebly

4 Commissioning the injectorIn this chapter you learn how to charge the battery cells, set the touch screen, insert the syringes into the injector head and connect the tube system.

Charging the battery cellsPrior to initial commissioning, you must charge the battery cells accommo-dated in the injector base.

• Make sure that the injector is switched off.

• Insert the connector for the charger into the socket on the injector base and connect the charger to the mains network.

The charging operation starts when the red LED on the charger comes on. After some time the yellow LED comes on. The light comes on green when the battery cells are fully charged.

• As soon as the battery cells are fully charged, disconnect the charger from the mains network and pull the connector of the charger out of the socket on the injector base.

The injector is now ready to be switched on, see next section.

Note! If the unit has been out of service for a prolonged period, charge the battery cells before starting it up again.

Note! The battery cells are maintenance-free. They are replaced every two years or when necessary exclusively by the MEDTRON Customer Service.

ATTENTION! The injector must be switched off before the battery cells are charged.

CAUTION! Do not charge the unit in the MR room. The injector must be at least 2.5 m away from the patients when the battery cells are being charged.

Note! The charging time depends on the state of charge of the battery cells. If they are completely discharged, charging takes about 8 hours.

Accutron MR 05/2008 23

Page 24: AccutronMR GA V2 GB - Weebly

24

Switching ON 4 Commissioning the injector

Switching ONBefore being switched on, the injector must be charged and disconnected from the charger.• Press the On/Off switch on the injector base to switch the injector on.

The integrated lamp of the switch comes on and the pilot light on the control panel lights up green.

Self-testAfter being switched on, the unit automatically performs a self-test. As soon as the self-test is completed, you briefly see the display with the MEDTRON logo:

Then the touch screen immediately shows the main window from where you can call up the submenus.

Accutron MR

Prof. Nr. 5

Phase Delays

Volumeml

Conc.%

Flowml/s

Inj. times

1

2

3

20

20

20

10

10

10 100

65

100 6.0

6.0

6.0 0

0

0

Not airfree

180 178

05/2008 Accutron MR

Page 25: AccutronMR GA V2 GB - Weebly

4 Commissioning the injector Switching ON

Stand-by functionIf you have not pressed a button for a preset period, the display switches off automatically. This saves energy and avoids any disruptive electromagnetic radiation during the scan operation. You can reactivate the deactivated display at any time by pressing the Stop button or the Wake-up button.

Display state of chargeIn addition to the battery symbol in the main window, the injector features also a more exact display of the charge state. When the injector is switched on, you can read off the charge of the battery cells in the System settings menu.

The System settings 1/2 window displays the state of charge in percent.

If the state of charge is less than 10 %, you must charge the battery cells. In this case charge the battery cells fully using the charger, see Charging battery cells on page 23.

If you have not yet set the touch screen to your operating habits, you should make some basic settings now. For this purpose remain in the System set-tings menu.

If the touch screen is already adapted to your needs, you can continue with the section Preparing injector head on page 30.

Note! If you reactivate the display during injection, the injection will be interrupted automatically.

• Open System settings menu.

System settings 1/2

State of charge 90 %

Accutron MR 05/2008 25

Page 26: AccutronMR GA V2 GB - Weebly

26

Setting the touch screen 4 Commissioning the injector

Setting the touch screenIn order to be able to use the touch screen easily and without any problems, you should set it at the outset to suit your operating habits and the environmen-tal conditions. The System settings menu consists of two windows where you can make the following settings:– Set touch screen brighter and darker– Select language for the display texts– Adjust touch screen to the tactile characteristics of the operator– Set stand-by time for the injector– Set idling speed of the pistons– Set filling speed of the pistons

In the dialogue window System settings 1/2 the state of charge is displayed in percent:

Setting the touch screen brighter or darkerAdapt the brightness of the screen to your vision or to different ambient lighting conditions. For this purpose:• Touch the button of the slide controller in the top of the dialogue window.

Pull the slider to the left to make the display brighter. Pull the slider to the right to make the display darker.

• Open System settings menu.

System settings 1/2

State of charge 90 %

05/2008 Accutron MR

Page 27: AccutronMR GA V2 GB - Weebly

4 Commissioning the injector Setting the touch screen

The display changes immediately. As soon as you have set the brightness you want:

Selecting a languageIn this dialogue window you can select the language in which the texts in the touch screen are to be displayed. The following can be selected– German– French– Portuguese– English– Spanish– Italian

• Touch button with the relevant flag.

Calibrating touch screenEveryone operates a touch screen in their own particular way. A right-handed person, for example, does not position his finger at the same place as a left-handed person. The size and angle of the fingers also play a role. To ensure that the touch screen correctly adopts your entries and commands, you must first calibrate it to your touch habits.

To be able to calibrate the touch screen, you must first log into the system.

The dialogue window Login is displayed. You are requested to select your name.• Touch list entry User.• Enter the identification 1001 on the number buttons.

• Touch OK button.

The calibration window is displayed. Cross-hairs are shown in the centre of the window. Here you are requested in a german text:

"Bitte drücken Sie fest auf die Fadenkreuze. Bleiben Sie solange auf den Fadenkreuzen, bis diese verschwinden."

Which means: "Please press the cross-hairs firmly. Remain on the cross-hairs until they disappear."

• Save the brightness settings.

• Touch cross-hairs in the dialogue window System settings.

Accutron MR 05/2008 27

Page 28: AccutronMR GA V2 GB - Weebly

28

Setting the touch screen 4 Commissioning the injector

• Keep your finger pressed on the cross-hairs in the centre of the screen until they disappear.

The cross-hairs are gradually displayed in each corner of the screen.• Press each cross-hair until it disappears.

To be able to make the further settings, you must open the dialogue window System settings 2/2:

The dialogue window System settings 2/2 is displayed:

Setting stand-by timeHere you can determine after how many seconds the injector is to switch into the stand-by mode when it is not being used or during the injection.

Setting idling speed of the pistonsHere you can set at what speed the empty pistons are to be moved. 1 ... 10 ml/s can be set.

Setting filling speed of the pistonsHere you can set at what speed the liquids are to be filled. 1 ... 5 ml/s can be set.

• Save calibration.

• Open System settings 2/2.

System settings 2/2

Stand-by time

Idling speed

Filling speed

60 s

7 ml/s

4 ml/s

05/2008 Accutron MR

Page 29: AccutronMR GA V2 GB - Weebly

4 Commissioning the injector Setting the touch screen

Make all settings as follows:• Touch the relevant button with your finger.

A pop-up window opens. The value currently set and the entry range are displayed. • Overwrite the marked value using the numeric keypad with a new value.

If you have accidentally entered the wrong number, you can delete it with the DEL key.

The pop-up window closes.

When you have made all the settings, you can now either:

or

When all settings have been made, you can now insert the syringe. For this purpose read the next section Preparing injector head on page 30.

• Confirm the new value.

• change to the window System settings 1/2

• quit the dialogue window System settings to display the main window.

Accutron MR 05/2008 29

Page 30: AccutronMR GA V2 GB - Weebly

30

Preparing injector head 4 Commissioning the injector

Preparing injector head

Moving the pistons backOn delivery the pistons are retracted - however, it may happen that, prior to the routine change of the syringes, the pistons are advanced. In order to be able to insert a new syringe, you must first move the pistons back. For this purpose:

The dialogue message with the confirmation prompt "Move piston CM back?" is displayed.

Note! You can only insert the syringes into the injector head when the pistons of the injector are fully moved back.

CAUTION! Risk of infection! Do not remove the protective caps of the syringes until you have made the connections. Make sure that the packings of the disposable articles do not show any signs of damage.

• Call up filling menu in the main window.

• Touch left button for the CM piston.

• Confirm the message.

• Touch right button for the NaCl piston.

Move piston CM back?

05/2008 Accutron MR

Page 31: AccutronMR GA V2 GB - Weebly

4 Commissioning the injector Preparing injector head

The dialogue message with the confirmation prompt "Move piston NaCl back?" is displayed.

Both pistons move back into their lowest position.

The piston is retracted when it is right at the bottom of the syringe, the syringe is completely green/blue and the remaining volume displayed is 200 ml (see diagram).

• Confirm the message.

• Return to the main window.

Move piston NaCl back?

Prof. No. 5

Rem.vol. CM 200 ml

Rem.vol. NaCl 200 mlNot airfree

Accutron MR 05/2008 31

Page 32: AccutronMR GA V2 GB - Weebly

32

Preparing injector head 4 Commissioning the injector

Inserting the syringes• Make sure that both pistons are moved back.

• Push the syringe from the front into the respective recipient until it engages audibly.

Then you can connect the tube system, see next section.

Note! The left-hand piston is intended for the injection of contrast medium, the right-hand piston for NaCl solution.

05/2008 Accutron MR

Page 33: AccutronMR GA V2 GB - Weebly

4 Commissioning the injector Connecting the tube system

Connecting the tube system

The MR tube system has four valves which function as follows:– When you fill the syringe CM with contrast medium, valve A1 is opened and

valve A2 closed.– When you inject contrast medium out of the syringe CM, valve A1 is closed

and valve A2 opened.

The valves B1 and B2 function in the same way.

The MR tube is connected to the syringes as follows:• Remove the protective caps from the syringe and tube end.

• Screw the MR tube with 1/4 to 1/2 turn onto the syringe.

CAUTION! Risk of infection! Do not remove the protective caps of the tube ends until immediately before making the connections. Make sure that the packings of the disposable articles do not show any signs of damage.

Note! The tube marked with a green stripe is intended for the contrast medium. The tube section between valve A2 and the Y-piece is very short in order to save contrast medium.

ATTENTION! When screwing on the MR tube, you may damage the Luer-Lock thread and therefore make the syringe and tube system unusable. Therefore do not exceed one 1/2 turn (180°)

A1A2

B2

B1

syringe NaCl

to the contrast medium bottle

connection patient tube

to the NaCl bottle

syringe CM

Accutron MR 05/2008 33

Page 34: AccutronMR GA V2 GB - Weebly

34

Connecting the tube system 4 Commissioning the injector

• Repeat steps for the NaCl tube.

In the next step connect the MR tube to the bottle.• Hang the contrast medium bottle and NaCl bottle on the two bottle holders.

• Pull the protective cap off the pin of the drip chamber and press it into the contrast medium bottle (infusion level).

• Open valve cap on the drip chamber.

• Briefly press drip chamber together twice to fill it roughly half-full with contrast medium.

• Repeat steps for the NaCl drip chamber.

• Connect the patient tube.

The injector is now ready to be filled with contrast medium and NaCl.

05/2008 Accutron MR

Page 35: AccutronMR GA V2 GB - Weebly

5 Operating the injectorThis chapter describes how to operate the injector manually, how to work with injection profiles and how to carry out an injection.

Filling injector with contrast medium and NaCl

Both syringes are inserted, left for contrast medium, right for NaCl, the MR-tube system is connected, the injector is switched on. The main display is shown.

In order to fill the syringes, the pistons must first be moved forward.

Note! Also fill the injector with contrast medium first and then with the saline solution. In this way you save contrast medium when venting the tube system.

CAUTION! Risk of fatal or serious injuries due to air embolisms! Make sure that the patient is not yet connected to the system!

• Call up the filling menu in the main window.

Prof. No. 5

Phase Delays

Volumeml

Conc.%

Flowml/s

Inj. times

1

2

3

20

20

20

10

10

10 100

65

100 6.0

6.0

6.0 0

0

0

Not airfree

200 200

Accutron MR 05/2008 35

Page 36: AccutronMR GA V2 GB - Weebly

36

Filling injector with contrast medium and NaCl 5 Operating the injector

The filling menu is displayed.

Move both pistons forward

The dialogue message with the confirmation prompt "Move piston CM for-ward?" is displayed.

Note! Wenn the pistons are fully moved back, the correspondent filling button isn't active and thus marked grey.

• Touch left button for the CM piston.

• Confirm the message.

Prof. No. 5

Rem.vol. CM 200 ml

Rem.vol. NaCl 200 mlNot airfree

Move piston CM forward?

05/2008 Accutron MR

Page 37: AccutronMR GA V2 GB - Weebly

5 Operating the injector Filling injector with contrast medium and NaCl

The dialogue message with the confirmation prompt "Move piston NaCl for-ward?" is displayed.

Both pistons of the syringes move forwards into their uppermost position, the remaining volume displayed becomes lower on both sides, the two symbols for the syringes become white.

When the pistons are fully extended, the filling menu is shown like this:

• Touch right button for the NaCl piston.

• Confirm the message.

Note! When advancing the pistons, the valves may cause a low hissing sound – this is normal.

Move piston NaCl forward?

Prof. No. 5

Rem.vol. CM 0 ml

Rem.vol. NaCl 0 mlNot airfree

Accutron MR 05/2008 37

Page 38: AccutronMR GA V2 GB - Weebly

38

Filling injector with contrast medium and NaCl 5 Operating the injector

Enter filling volume for contrast medium and fill

A dialogue window with number buttons is displayed.

The range which you can enter, e.g. 0.0 ... 200.0, is in the top right of the dia-logue window Enter filling volume and fill. The current (maximum) value is already specified. You can change the value by entering the value you want on the numpad. If you have accidentally entered a wrong number, you can delete it using the DEL button.• Enter the filling volume you want.

The dialogue window Enter filling volume and fill is closed. The piston of the syringe for contrast medium moves back.

Enter filling volume for NaCl and fill

A dialogue window with number buttons is displayed.

The range which you can enter, e.g. 0.0 ... 200.0, is in the top right of the dia-logue window Enter filling volume and fill. The current (maximum) value is already specified. You can change the value by entering the value you want on the numpad. If you have accidentally entered a wrong number, you can delete it using the DEL button.• Enter the filling volume you want.

The dialogue window Enter filling volume and fill is closed. The piston of the syringe for NaCl moves back.

• In the filling menu touch the Fill button in order to enter the filling volume for contrast medium.

• Confirm entry.

• Touch the blue Fill button in the filling menu to enter the filling volume for NaCl.

• Confirm entry.

05/2008 Accutron MR

Page 39: AccutronMR GA V2 GB - Weebly

5 Operating the injector Filling injector with contrast medium and NaCl

While the pistons draw in contrast medium and NaCl, you see the following display:

• Then connect the patient tube to the MR tube.

• When the filling procedure is finished, return to the main window.

CAUTION! Risk of fatal or serious injuries due to air embolisms! At the latest now you must connect the patient tube to the MR tube so that the entire tube system right to the patient is completely vented.

Prof. No. 5

Rem.vol. CM 102 ml

Rem.vol. NaCl 135 mlNot airfree

Accutron MR 05/2008 39

Page 40: AccutronMR GA V2 GB - Weebly

40

Filling injector with contrast medium and NaCl 5 Operating the injector

Venting the tube systemThe syringes and the tube system still contain air. You must vent the system using the buttons for the manual piston movements.

• Make sure that the patient tube is connected to the MR tube.

• Swivel the two injection units into the vertical position.

• Keep venting the CM side until there are no more air inclusions just after the Y-piece.

• Vent the entire tube system incl. the patient tube with the NaCl piston.

• Keep venting until there are no more air inclusions in the entire tube system.

During venting the injection volume actually available appears in the window Rem. vol.(ml).

Note! Firstly, vent the contrast medium side up to the Y-piece and only then the entire system including the patient tube with the piston for NaCl. In this way you save contrast medium.

In the left-hand button group for the manual piston movements press the button Piston forward and keep it depressed to advance the CM piston.

• To gradually increase the piston speed, also press this adjacent button.

• In the right-hand button group for the manual piston movements press the button Piston forward and keep it depressed to advance the NaCl piston.

• To gradually increase the piston speed, also press this adjacent button.

CAUTION! Risk of fatal or serious injuries due to air embolisms! There must be no more air inclusions in the entire tube system.

05/2008 Accutron MR

Page 41: AccutronMR GA V2 GB - Weebly

5 Operating the injector Filling injector with contrast medium and NaCl

The dialogue message with the confirmation prompt "CM-syringe airfree?" is displayed.

• On completion of venting touch the now active button Airfree or the text field Not airfree.

• Confirm completed venting.

Prof. No. 5

Phase Delays

Volumeml

Conc.%

Flowml/s

Inj. times

1

2

3

20

20

20

10

10

10 100

65

100 6.0

6.0

6.0 0

0

0

Not airfree

197 195

CM-syringe airfree?

Venting

Accutron MR 05/2008 41

Page 42: AccutronMR GA V2 GB - Weebly

42

Filling injector with contrast medium and NaCl 5 Operating the injector

The dialogue message with the confirmation prompt "NaCl-syringe airfree?" is displayed.

The parameters of the current injection profile are displayed. The text field indicates "Airfree" in green. The Activate touch button is highlighted in colour and therefore active.• Press injection unit downwards into the injection position.

The injector is now ready for the injection.

• Confirm completed venting.

NaCl-syringe airfree?

Venting

05/2008 Accutron MR

Page 43: AccutronMR GA V2 GB - Weebly

5 Operating the injector Changing the maximum operating pressure

Changing the maximum operating pressureYou can set the desired injection pressure. The pressure actually reached dur-ing injection depends on a number of ambient variables such as the viscosity of the contrast medium, needle diameter, set flow rate etc.

If the set maximum working pressure is reached during the injection, the injec-tion is interrupted.

The maximum operating pressure can only be changed on the remote control supplied with the Accutron MR. Please read the separate instructions for use to see how to proceed for this purpose.

CAUTION! Risk to the patient! Make sure that the maximum pressure set does not endanger the patient.

Accutron MR 05/2008 43

Page 44: AccutronMR GA V2 GB - Weebly

44

Selecting an injection profile 5 Operating the injector

Selecting an injection profileYou can create with the injector up to 80 injection profiles and store them under a profile number. Depending on the application, you call up the profile you want with the injection parameters saved and you can immediately start the injection. For this purpose proceed as follows:

The dialogue window with the injection profiles already available is displayed.

Using the arrow buttons, you can scroll to the right and left and in this way view all profile names.• Touch the syringe symbol of the profile you want to use.

In this way you have selected a profile. The parameters of this profile (here Profile No. 7) are in the main window.

• Call up profile.

Prof. No. 7

Prof. No. 1

Prof. No. 2

Prof. No. 3

Prof. No. 4

Prof. No. 5

Prof. No. 6

Prof. No. 7

Prof. No. 8

Prof. No. 9

Prof. No. 10

05/2008 Accutron MR

Page 45: AccutronMR GA V2 GB - Weebly

5 Operating the injector Changing an injection profile

Changing an injection profileYou can change the parameters of an injection profile as required.

The touch screen displays the injection profile to be changed.

Example: The profile to be changed is displayed. You want to change the volume of a phase.• Touch the relevant value with your finger.

A pop-up window opens. The value currently set and the entry range are displayed.• Overwrite the marked value using the numeric keypad with a new value.

If you have accidentally entered the wrong number, you can delete it with the DEL key.

You edit the other values, such as delay, concentration and flow in the same way.

• Confirm the new value.

Note! Take note before entering the concentration:

– 100% means contrast medium only – 40% means 40% contrast medium, 60% NaCl – 0% means NaCl only

Prof. No. 7

Phase Delays

Volumeml

Conc.%

Flowml/s

Inj. times

1

2

3

20

20

20

10

10

10 100

65

100 6.0

6.0

6.0 0

0

0

Not airfree

180 178

Accutron MR 05/2008 45

Page 46: AccutronMR GA V2 GB - Weebly

46

Changing an injection profile 5 Operating the injector

The system displays the changed value in the main window. The profile name has changed automatically. Prof. No. 0 is now shown in the main window.

Storing changesIf you have made changes to the parameters of an injection profile, you must store the profile. However, after a change the original name of the profile is no longer on the screen but Prof.Nr. 0.

Note! Some values have a reciprocal effect on one another. If you change the volume value or the flow, the injection time of this phase changes automatically. If you change the injection time, the flow is changed accordingly.

CAUTION! Risk of serious injuries! Before entering the flow (up to 10 ml/s possible!), check the constitution of the patient to be examined.

ATTENTION! Risk of data loss! Save changes to parameters under a profile number. Otherwise the changes are lost when you switch off the injector.

• Save profile.

05/2008 Accutron MR

Page 47: AccutronMR GA V2 GB - Weebly

5 Operating the injector Changing an injection profile

The dialogue window with the injection profiles already available is displayed.

You can scroll to the right and left using the arrow buttons and in this way view all the profile names.

You have two possibilities:– overwriting an existing profile, see next paragraph, or– saving the changes under a new profile, see next paragraph but one.

Overwriting an existing profileIf you want to overwrite an already existing profile,• touch the appropriate syringe symbol.

The writing in the syringe symbol is green and the name of the selected profile number is now in the text field for the profile name.

Prof. No. 7

Prof. No. 1

Prof. No. 2

Prof. No. 3

Prof. No. 4

Prof. No. 5

Prof. No. 6

Prof. No. 7

Accutron MR 05/2008 47

Page 48: AccutronMR GA V2 GB - Weebly

48

Changing an injection profile 5 Operating the injector

Entering a profile nameIf you want to change the name of this profile:• touch the dark-blue text field.

The touch screen displays the input mask for the profile name.

You can enter the new profile name using the keyboard. You can delete indi-vidual letters with the DEL button or delete the entire name with the Delete button. You can write capital letters with the SHIFT button.

• Enter a new profile name.

• Touch the OK button to confirm the entry.

The changed profile is stored and the main window is displayed.

• Save the new name for the profile.

new name

Delete

ESC

05/2008 Accutron MR

Page 49: AccutronMR GA V2 GB - Weebly

5 Operating the injector Changing an injection profile

Saving changes under a new profileIf you want to save the changed profile under a new profile,• select the empty syringe symbol.

The marking in the syringe symbol is green and is also shown at the top in the dark-blue text field.• Touch the dark-blue text field.

The touch screen displays the input mask for the profile name, see preceding section.• Enter the name for the new profile with the keyboard.

• Touch the OK button to confirm the entry.

The new entry is now in the profile list.

The main window displays the newly created profile.

• Save the new profile with the new profile name.

new name

Prof. No. 1

Prof. No. 2

Prof. No. 3

Prof. No. 4

Prof. No. 5

Prof. No. 6

Prof. No. 7

new name

Accutron MR 05/2008 49

Page 50: AccutronMR GA V2 GB - Weebly

50

Changing an injection profile 5 Operating the injector

Creating a new profileIf you want to create a completely new profile, you must first select the relevant syringe symbol. For this purpose:

The list with the profiles is displayed.

You can scroll to the right and left with the arrow buttons and in this way view all the syringe symbols.• Choose the empty syringe symbol of the new profile and touch it.

In the main window you have opened an empty profile with only one phase which you can now fill with the injection parameters you want.• Enter all the profile data and parameters you want.

• Select the syringe symbol you want again.

• Touch dark-blue text field to enter the profile name.

The touch screen displays the input mask for the profile name, see page 48.• Enter the name for the new profile with the keyboard.

• Call up profile in the main window.

• Save profile.

-

Prof. No. 1

Prof. No. 2

Prof. No. 3

Prof. No. 4

Prof. No. 5

Prof. No. 6

-

05/2008 Accutron MR

Page 51: AccutronMR GA V2 GB - Weebly

5 Operating the injector Changing an injection profile

• Touch the OK button to confirm the entry.

The main window displays the newly created profile.

• Save the new profile with the new profile name.

Accutron MR 05/2008 51

Page 52: AccutronMR GA V2 GB - Weebly

52

Changing an injection profile 5 Operating the injector

Adding a new phaseYou can add a new phase (a maximum of 6 are possible) to an existing profile with, for example, 2 phases. The main window of the relevant profile is displayed.

• Touch parameter fields of the new phase.

• Enter the individual values in the input mask using the number buttons.

The main window of the profile is displayed.

• Touch the appropriate syringe symbol.

The lettering in the syringe symbol is green and the profile number is in the text field at the top.• Enter profile name as described on page 48.

• Touch the OK button to confirm the entry.

The changed profile is stored and displayed in the main window.

• Add a phase.

• Confirm the new value.

• Save the changes.

• Save the profile under the profile name selected.

05/2008 Accutron MR

Page 53: AccutronMR GA V2 GB - Weebly

5 Operating the injector Changing an injection profile

Deleting a phaseYou can completely delete the last phase from a profile. Proceed as follows:

• Touch the appropriate syringe symbol.

The lettering in the syringe symbol is green and the profile number is in the text field at the top.• Enter profile name as described on page 48.

• Touch the OK button to confirm the entry.

The changed profile is stored and the main window is displayed.

• Delete phase in the main window of the correspondent profile.

• Save the changed profile.

• Save the profile under the profile name selected.

Accutron MR 05/2008 53

Page 54: AccutronMR GA V2 GB - Weebly

54

Performing an injection 5 Operating the injector

Performing an injection

– Do not connect the patient to the system until you have vented the syringes and the tube system.

– Vent the syringes and the tube system every time after filling the injector with contrast medium or saline solution.

– Make sure before every injection that there are no air inclusions in the syringes and the connected tubes.

– Only start an injection when the syringes are in the lowered position (injection position).

– Make sure before every injection that the injection parameters shown on the display do not endanger the patient.

• Swivel the injection units out of the vertical position backwards into the lowest position (injection position):

This injection position ensures that air bubbles rise to the piston and are not injected.• Make sure again that the MR tube and the patient tube systems do not

contain any air inclusions.

• Remove protective cap from the patient tube and connect the tube to the patient cannula.

The injector is vented, the Activate button on the touch screen is active.

CAUTION! Risk of fatal or serious injuries due to air embolisms!

CAUTION! Risk of uncontrolled supply of contrast medium due to hydrostatic pressure compensation! Make sure that the patient and the syringes are at the same level.

05/2008 Accutron MR

Page 55: AccutronMR GA V2 GB - Weebly

5 Operating the injector Performing an injection

Maintaining vein access with the KVO functionBefore performing the injection, you can now establish whether you want to activate the Keep-Vein-Open function (KVO).The KVO function ensures that access to the vein is retained during prolonged examinations.

If you activate this function, the injector injects 0.5 ml of saline solution every 2 minutes.

– When the KVO function is activated, the writing in the button is green.– If the KVO function is deactivated, the writing in the button is grey.

• If you want, activate KVO.

Note! If the NaCl volume falls below the minimum of an injection volume with the KVO function activated, the KVO function is switched off and in, in addition, a warning appears in the display.

Accutron MR 05/2008 55

Page 56: AccutronMR GA V2 GB - Weebly

56

Performing an injection 5 Operating the injector

Starting the injection

Prof. No. 7

Injector readyKVO on

Ph. Delays

Vol.ml

Conc.%

Flowml/s

Times

1

2

3

20

20

20

10

10

10

100

65

100

6.0

6.0

6.0 0

0

0

0.0 s

The following message is displayed:

• Touch the Activate button.

• Using the Info button, you can change to a second window with more injection data.

Prof. No. 7

Injected volume - CM

Injected volume - NaCl

Phase time

Delay

Injection time

0

0

0.0

20.0

86.7

Injector readyKVO on

0.0 s

05/2008 Accutron MR

Page 57: AccutronMR GA V2 GB - Weebly

5 Operating the injector Performing an injection

The LEDs of the start and stop button on the control panel come on.

The injection starts.

The pilot light on the control panel flashes green.

After the start of the injection, you can follow the individual phases and the details of already injected and remaining contrast medium, phase time, delay, injection time and scan delay on the display of the remote control. The stop-watch in the top right begins to run when you press the start button.

Interrupting the injection on the remote controlYou can interrupt the injection at any time by touching the HOLD button on the remote control. In the confirmation prompt then displayed you have two pos-sibilities:– Touch the End button if you want to interrupt the injection completely, – Touch the Continue button if you want to continue the interrupted injection.

The unit then continues the injection at the point where it was interrupted.

Interrupting the injection on the injectorIf the display is deactivated, you can stop the injection at any time by

or by

The injector stops the injection.

The injection data window is displayed, and you can either continue the injec-tion or stop it completely.

• Press the start button on the control panel of the remote control or on the injector.

Note! The touch screen on the injector switches off on expiry of the set stand-by time and you can follow the course of the injection on the remote control.

• pressing the Stop button on the control panel of the injector

• pressing the Wake-up button on the control panel of the injector.

Accutron MR 05/2008 57

Page 58: AccutronMR GA V2 GB - Weebly

58

Performing an injection 5 Operating the injector

Injection endThe program stops after passing through all the phases. All time values of the injection count backwards to 0.The injected volume counts forwards.

The message "Injection finished" is displayed.

• The main window is displayed in the touch screen.

• When the examination is over, disconnect the patient tube from the cannula and dispose of the cannula.

If you want to perform further injections, change the patient tube, prepare a new injection, replenish contrast medium and vent, as described above.

If you want to stop working with the injector, read and follow the instructions in the chapter Post-examination work on page 64.

Note! The program run stops when all the phases have been passed through. If the KVO function has been activated, NaCl is injected from this time onwards until you confirm the program end.

• Confirm message.

• If necessary, wake up touch screen on the injector.

CAUTION! Risk of infection! Replace the patient tube with a new one after it has been used once. Otherwise, you jeopardise your patients' health.

Observe the applicable hygiene regulations!

05/2008 Accutron MR

Page 59: AccutronMR GA V2 GB - Weebly

5 Operating the injector Replenishing injector with contrast medium/NaCl

Replenishing injector with contrast medium/NaClWhen the contrast medium or NaCl filling of the syringes is coming to an end and you want to perform more injections or if the message "CM-volume too low" appears, top up the syringes.

How to proceed is described here taking the example of contrast medium.• Confirm message "CM-volume too low".

The main window of the touch screen is displayed.• Remove the patient tube.

The filling menu is displayed.

CAUTION! Risk of fatal or serious injuries due to air embolisms! Make sure that the patient is not yet connected to the system!

• Touch the Fill button in the main window.

Prof. No. 5

Rem.vol. CM 28 ml

Rem.vol. NaCl 33 mlNot airfree

Accutron MR 05/2008 59

Page 60: AccutronMR GA V2 GB - Weebly

60

Replenishing injector with contrast medium/NaCl 5 Operating the injector

Enter filling volume and fill

The dialogue window Enter filling volume and fill with number buttons is dis-played.

The range with the possible values you can enter, e.g. 0.0 … 172.0, is in the top right of the dialogue window. The current (maximum) value is already specified, in this example that is 172 ml as 28 ml are available as the remaining volume. You can change the value by entering the value you want on the numeric keypad. If you have accidentally entered the wrong number, you can delete it with the DEL button.• Enter the filling volume you want.

The dialogue window Enter filling volume and fill is closed. The piston of the syringe moves back. While the piston draws in the contrast medium, you see the following display:

• Touch the green Fill button in the filling menu to enter the filling volume for contrast medium.

• After entry, confirm the value you want.

• When the filling procedure is finished, return to the main window.

Prof. No. 5

Rem.vol. CM 172 ml

Rem.vol. NaCl 33 mlNot airfree

05/2008 Accutron MR

Page 61: AccutronMR GA V2 GB - Weebly

5 Operating the injector Replenishing injector with contrast medium/NaCl

Venting after replenishing• Then connect the patient tube to the MR tube.

At present, the syringe and the tube system still contain air. You must vent the system using the buttons for the manual piston movements.

• Make sure that the patient tube is connected to the MR tube.

• Keep venting CM side until there are no more air inclusions downstream of the Y-piece.

With the NaCl side vent the entire tube system including the patient tube.

• Keep venting until there are no more air inclusions in the entire tube system.

CAUTION! Risk of fatal or serious injuries due to air embolisms! At the latest now you must connect the patient tube to the MR tube so that the entire tube system right to the patient is completely vented.

Note! First vent the contrast medium side up to the Y-piece and only then the entire system including the patient tube with the piston for NaCl. In this way you save contrast medium.

In the left-hand button group for the manual piston movements press the button Piston forward and keep it depressed to advance the CM piston.

• To gradually increase the piston speed, also press this adjacent button.

In the right-hand button group for the manual piston movements press the button Piston forward and keep it depressed to advance the NaCl piston.

• To gradually increase the piston speed, also press this adjacent button.

CAUTION! Risk of fatal or serious injuries due to air embolisms! There must be no more air inclusions in the entire tube system.

Accutron MR 05/2008 61

Page 62: AccutronMR GA V2 GB - Weebly

62

Replenishing injector with contrast medium/NaCl 5 Operating the injector

During venting the injection volume actually still available appears in the win-dow Rem.vol.(ml).

The dialogue message with the confirmation prompt "CM-syringe airfree?" is displayed.

• On completion of venting touch the now active button Airfree or the text field Not airfree.

• Confirm completed venting.

Prof. No. 7

Phase Delays

Volumeml

Conc.%

Flowml/s

Inj. times

1

2

3

20

20

20

10

10

10 100

65

100 6.0

6.0

6.0 0

0

0

Not airfree

197 29

CM-syringe airfree?

Venting

05/2008 Accutron MR

Page 63: AccutronMR GA V2 GB - Weebly

5 Operating the injector Replenishing injector with contrast medium/NaCl

The dialogue message with the confirmation prompt "NaCl-syringe airfree?" is displayed.

The parameters of the current injection profile are displayed. The text field indicates "Airfree" in green. The Activate button is highlighted in colour and therefore active.

The injector is now ready for the injection.

• Confirm completed venting.

NaCl-syringe airfree?

Venting

Accutron MR 05/2008 63

Page 64: AccutronMR GA V2 GB - Weebly

64

Post-examination work 5 Operating the injector

Post-examination work• When the examination is over, disconnect the patient tube from the cannula

and dispose of the cannula.

• Remove the syringe with tube system. This is possible at every piston position.

• Properly dispose of all disposable articles in accordance with the applicable provisions.

• In the filling menu, move both pistons back, see page 30.

• Press the On/Off switch to switch the injector off.

The green LED of the switch and the pilot light on the control panel go off.• Connect the injector to the charger in order to charge the battery cells for

use the next day.

CAUTION! Risk of infection! Replace the patient tube with a new one after it has been used once. Otherwise, you jeopardise your patients' health.

Observe the applicable hygiene regulations!

Note! To attach a new syringe, the piston has to be in the lowest position.

05/2008 Accutron MR

Page 65: AccutronMR GA V2 GB - Weebly

6 Annex

Waste disposal

Safety inspections and maintenance

MEDTRON prescribes annual maintenance and a safety inspection for the injector. The maintenance work is performed by staff trained and authorised by MEDTRON. Regular maintenance ensures the functionability of the injec-tor.

Safety Inspections and maintenance work should be certified and documented in the medical equipment logbook.

Cleaning and storageRemove any disposable articles from the injector immediately after use and dispose of them properly. Disposable articles must not be sterilised and re-used. Remove all dirt caused by contrast medium with warm water before the dirt dries in.

To disinfect the Accutron MR, we recommend Mikrozid Liquid from Schülke & Mayr GmbH, Germany. Do not use any aggressive disinfectant to disinfect the injector.

In the case of prolonged breaks in operation, store the Accutron MR in a safe place where it is protected against dust and moisture.

Do not put old devices to the normal domestic waste! MEDTRON undertakes to take back its old devices as well as electric and electronic accessories, such as chargers, remote controls and similar. The disposable articles are to be disposed of in a proper manner.

CAUTION! Risk of injury to operator and patient! Maintenance and repair work may only be performed by the MEDTRON Customer Service or people authorised by MEDTRON.

ATTENTION! Risk of malfunctions of the Accutron MR! Do not immerse the injector in water! Do not use any aggressive cleaning agents or solvents. Warm water and a mild soap are sufficient.

Accutron MR 05/2008 65

Page 66: AccutronMR GA V2 GB - Weebly

66

System messages 6 Annex

System messagesThe following table contains system messages which indicate either faults in the injector or incorrect operation. Using the following table, try to eliminate the cause of the fault in question. If the cause remains unclear and the suggestions for remedy prove unsuc-cessful, please contact the MEDTRON Customer Service.

Message/Error Possible cause and remedy

CM VOLUME TOO LOW

There is too little contrast medium in the syringe to implement the injection profile.Acknowledge message and replenish contrast medium, see page 59.

NACL VOLUME TOO LOW

There is too little NaCl in the syringe to implement the injection profile.Acknowledge message and replenish NaCl, see page 59.

No operation possible

The value for the volume or flow is missing in at least one phase. Enter missing values.

CONFIGURATION ON THE REMOTE CONTROL

This message appears while settings are being made on the touch screen of the remote control.

At the same time:CONFIGURATION ON THE REMOTE CONTROLCONFIGURATION ON THE INJECTOR

The injector and remote control were operated at the same time.Press the stop button on the injector.Never operate units at the same time.

The unit was not switched off correctly! Please remove the syringe and activate the OK button. The piston moves back!

05/2008 Accutron MR

Page 67: AccutronMR GA V2 GB - Weebly

6 Annex Technical data

Technical data

The injector satisfies the requirements of the standard DIN EN 60601-1-2 for medical equipment in terms of the emitted interference and interference immu-nity.

Flow rate per unit 0.1 ml/s ... 10 ml/s, increment 0.1 ml/s

Injection volume Contrast medium: 200 ml Saline solution: 200 ml

Partial volume which can be pre-selected

1 ml ... 200 ml, increment 1 ml

Max. injection pressure 21 bar

Filling speed 1 ml ... 5 ml

KVO parameters 0.5 ml of saline solution every 2 minutes.

Injection profiles 80

Injection phases 1 ... 6

Accuracy The accuracy of the contrast medium volume injected is 1% of the syringe volume (only applies to the empty syringes recommended by MEDTRON)

Charger Mascot Type 2040 with medical approval

Input voltage charger 100 ... 240 V~, 50 ... 60 Hz

Power consumption charg-er

< 100 VA

Protection class injector device with internal power supply

Device type CF, identified by

Dimensions (H × W × D) 1180 × 470 × 620 mm in the injection position

Weight 37 kg

Ambient conditions during operation

– operating temperature 10 °C ... 40 °C – relative humidity 30 % ... 75 % – air pressure 700 hPa ... 1060 hPa

Ambient conditions for storage and transport

– storage temperature - 20 °C ... 60 °C – relative humidity up to 75 % – air pressure 700 hPa ... 1060 hPa

Rechargeable batteries 24 V / 17 Ah

Operating time when com-pletely charged

Fully charged batteries permit at least 100 injec-tions of 30 ml at 2.0 ml/s

Radio communication Bluetooth version 2.0

Bluetooth class 1

Frequency range 2.402 ... 2.48 GHz

Accutron MR 05/2008 67

Page 68: AccutronMR GA V2 GB - Weebly

68

Technical data 6 Annex

05/2008 Accutron MR

Page 69: AccutronMR GA V2 GB - Weebly

Index

Index

AAir inclusions 40, 61Ambient light 26

BBattery cells 14

charging 23charging time 23

Battery symbol 19Bluetooth class 67Bottle holders 14Breaks in operation 65Brightness

setting 26

CCalibrating touch screen 27Calibration window 27Castors 14, 22Charger 23Cleaning and storage 65Column stand 14Concentration 19

changing 45Connections 16Contrast medium

filling 35replenishing 59

Contrast medium sideventing 61

Control panel 17Cross-hairs 27

DDelay 19, 46Disinfection 65Disposable articles

disposing of 64Drip chamber 34

EElectromagnetic compatibility 11EMC 11

FFadenkreuze 27Fault messages 66Filling speed 67

setting 28Flow 19, 46

changing 45Frequency range 67

HHandle 22

IIdling speed

setting 28Inj. time 19Injected volume 19Injection

end 58performing 54phases 67position 54pressure 67profiles 67starting 56unit 14

Injection profilestoring changes 46

Injectorcommissioning 23switching ON 24

Inspection and maintenance work 65Intended use 6Items supplied 14

KKeep-Vein-Open function 55KVO 19KVO parameters 67

LLanguage

selecting 27Login 27

Accutron MR 05/2008 69

Page 70: AccutronMR GA V2 GB - Weebly

70

Index

MMain window 18Maintenance work 65Manufacturer’s plate 6Medical equipment logbook 65Model name 6

NNaCl

filling 35replenishing 59

NaCl sideventing 40, 61

OOn/Off switch 14, 22Operating

habits 26pressure 43time 67

Optional accessories 14

PParameters

changing 45Phase 19

adding new 52Phase time 19Pistons

move forward 36moving back 30

Pressurechanging 43

Profilechanging 45selecting 44

Protectionclass 67of people 9of property 9

RRadio communication 67Receiving aerial 14Rechargeable batteries 67Recipient 14Remote control 22Repair work 65

SSafety instructions 9Scan delay 19Serial number 6State of charge

displaying 25Storage 65Switching ON 24Syringe

inserting 32venting 40

Syringe symbol 44, 50System settings 26, 28

TTouch screen 18

calibrating 27

VVenting the syringe 40Volume 19

05/2008 Accutron MR